Chino IR-M22 SERIES, IR-M21 SERIES Instructions Manual

Page 1
INST No INE -403-0P0
IR-M2
□□□
Series
INFRARED MOISTURE METER DESIGNED AS A FIBER OPTIC
(Model : IR-M21
□□
,IR-M22
□□
)
Please be sure to deliver these instructions with the unit to the end user.
Always keep these instruction with the unit.
Page 2
INST No INE -403-0P0
C1
Request to designers, instrument controllers, and sale agents
Delivery this instruction manual to the operator of this instrument without fail.
Request to the operator of this instrument
This instruction manual is necessary for maintenance, too. Keep this manual with due care until the instrument is discarded.
FOR SAFE USE
For the purpose of using this unit safely, observe the following cautions for safety during the operation, maintenance, and repair of this unit without fail. Take safety measures separately against wrong operation, troubles, and other circumstances of this unit.
The following warning mark is struck to this unit to enable users to use
this unit safely.
 
This instruction manual uses the following symbol marks for safe use.
           
PREFACERequests and notices
Thank you for your purchase of IR-M21□□□□, IR-M22□□□□series Infrared m oisture meter designed as a fiber optic. Please read this instruction manual without fail for using this instrument correctly and safely and also preventing troubles in advance.
1.For the contents of this manual, alteration is reserved without notice in the future.
2.This manual has been prepared by making assurance doubly sure about its contents.
However, if any question arises or if any error, an omission, or other deficiencies were found, please inform your nearest CHINO’s sales agent of them.
3.Y ou are requested to understand that CHINO is not responsible for any operation
results.
Note
This warning mark is struck to the portions being subject to an electric shock
acc
ident. Turn off the power supply of the feed source without fail before starting
wiring, maintenance, and repair.
Observe this caution item strictly for avoiding a danger, otherwise negligence of this caution may possibly cause troubles of this unit and accidents
Observe this caution item strictly, otherwise negligence of this caution may cause the damage of the unit or measuring errors.
This mark is described in each title covering      or   instructions.
War ning
Caution
Warning
DANGER
Refere nce
Information that you can use as a reference.
Page 3
INST No INE -403-0P0
−C2
Power supply
Make sure that the power voltage of this unit meets the feed power voltage without fail before feeding power to this unit.
Protective grounding
Perform the protective grounding without fail before turning on the power supply of this unit for the purpose of preventing an electric shock accident.
Necessity of protective grounding
Don’t cut the internal or external protective grounding wire of this unit nor disconnect the wiring at the protective grounding terminal.
A defect of protective functions
Don’t operate this unit, if the protective grounding, fuse, or other protective functions seem to be defective. Check the protective functions for defects before operating this unit.
Use in a gaseous atmosphere
Never operate this unit at a place where a combustible gas, an explosive gas or their vapors exist., otherwise a very dangerous accident will occur.
Never touch the interior by hand.
This unit is internally provided with high-voltage circuits. Never touch the interior by hand when the voltage is applied to this unit.
External connections
Perform the protective grounding securely without fail before connecting this unit to a measuring object and/or an external control circuit.
      
CAUTIONS FOR SAFETY
Page 4
INST No INE -403-0P0          
−C3
CONTENTS
1.General
・・・・・・・・・・・・・・・・・・・・・・・・・・・・
1
1.1Summary
・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・
1
1.2Configuration
・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・
1
2 .Model s and accessories
・・・・・・・・・・・・・・・・・
2
2.1Models・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 2
2.1.1 Detector unit
・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・
2
2.1.2 Reflective type fiver
・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・
2
2.1.3 Output checker plate for reflective type fiver optics
・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・
2
2.1.4 Reflective type fiver opt ics with lens attachment 2
2.1.5 Output checker plate for reflective type fiver optics with lens attachment
・ ・・・ ・・・ ・
2
2.1.6 Transparent type fiver optics
・ ・・・ ・・・ ・・・ ・・・
2
2.1.7 Measuring cell for liquids
・ ・・・ ・・・ ・・・ ・・・ ・・
2
2.1.8 Connection cable
・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・
2
2.2Accessories
・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・
2
3.Setting
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・
3
3.1Settin g of dispswitches ・・・・・・・・・・・・・・・・・・ 3
3.2Installation of reflective type fiver optics ・・・・・・ 5
3.2.1 Installation of tip part・・・・・・・・・・・・・・・・・・・ 5
3.2.2 Air purge(Only IR-WCRN)・・・・・・・・・・・・・・ 6
3.2.3 Connections with detector unit ・・・・・・・・・・・・ 6
3.2.4 Optical path ・・・・・・・・・・・・・・・・・・・・・・・・・ 7
3.2.5 Measuring place
(for objects of pulverized or granul ar substance) 7
3.2.6 For the objects in sheet state such as paper, etc.7
3.3Installation of transparent type fiver optics ・・・・ 8
3.3.1 Installation of cell for liquids ・・・・・・・・・・・・・・ 8
3.3.2 Connections with detector unit ・・・・・・・・・・・・ 9
3.3.3 Disassemble cleaning ・・・・・・・・・・・・・・・・・・・ 9
3.4Installation of detector unit ・・・・・・・・・・・・・・・10
3.4.1 Outside and mounting dimensions of
detector unit ・・・・・・・・・・・・・・・・・・・・・・・・・10
4. Connections
・・・・・・・・・・・・・・・・・・・・・・・・・
11
4.1Connections
・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・
11
4.2Layout of connection cable
・ ・・・ ・・・ ・・・ ・・・ ・・
11
4.3Multi -detector connections
・ ・・・ ・・・ ・・・ ・・・ ・・
12
4.3.1 Connections through relay box
・ ・・・ ・・・ ・・・ ・・・
12
4.3.2 Connections without rela y box
・ ・・・ ・・・ ・・・ ・・・
13
5.Names and functions of component parts
・・・
14
5.1Names and functions of detector unit
・ ・・・ ・・・ ・
14
5.2Names and functions of setting display part
・ ・・・
15
6. Operation
・・・・・・・・・・・・・・・・・・・・・・・・・・・・
16
6.1Confirmation before operation
・ ・・・ ・・・ ・・・ ・・
16
6.2Setting of detector unit number
・ ・・・ ・・・ ・・・ ・・
16
6.3Setting of calibration curve data
・ ・・・ ・・・ ・・・ ・・
16
6.3.1 Setting data
・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・
17
6.3.2 Setting of calibration curve data
・ ・・・ ・・・ ・・・ ・
18
6.4Preparation of calibration curbe by knee point line approximation
・ ・・・ ・・・ ・・・ ・
18
6.4.1 Setting of knee point data
・ ・・・ ・・・ ・・・ ・・・ ・・・
18
6.4.2 Knee point line setting data
・ ・・・ ・・・ ・・・ ・・・ ・・
19
6.4.3 Setting of calibration curbe data
for knee point line approximation
・ ・・・ ・・・ ・・・
20
6.5 Setting of operating conditions
・ ・・・ ・・・ ・・・ ・・・
21
6.5.1 Setting data
・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・
21
6.5.2 Setting
・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・
23
6.6Calibration
・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・
23
6.7Setting of calibration curve number・・・・・・・・・24
6.8Correction of calibration curve
・ ・・・ ・・・ ・・・ ・・
24
7.Preparation of calibration curve
・・・・・・・・・・
25
7.1Sample preparation ・・・・・・・・・・・・・・・・・・・・ 25
7.1.1 Powder or guranular
・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・
25
7.1.2 Paper(sheet substance)
・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・
26
! !
Page 5
INST No INE -403-     
−C4−
7.1.3 Liquid
・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・
26
7.2Sample measuremernt ・・・・・・・・・・・・・・・・・・ 27
7.2.1 Powder or granular
・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・
27
7.2.2 Paper(sheet substance)
・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・
28
7.2.3 Liquid
・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・
28
7.3Preparation of calibration curves・・・・・・・・・・・ 29
7.3.1 Computation of regression expression
・ ・・・ ・・・ ・
29
7.3.2 Confirmation of regression expression
・ ・・・ ・・・
30
8.Correction function by sample temprature from external input (option)
・・・・・・・・・・・・
31
8.1Summary ・・・・・・・・・・・・・・・・・・・・・・・・・・・ 31
8.2Model・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 31
8.3Wiring of detector unit ・・・・・・・・・・・・・・・・・・ 31
8.4S etting of dipswitch ・・・・・・・・・・・・・・・・・・・・ 31
8.5Setting of correction enabled/disabled
by sample temperature ・・・・・・・・・・・・・・・・・ 32
8.5.1 Setting at the detector unit
・ ・・・ ・・・ ・・・ ・・・ ・・・
32
8.5.1 Setting at the display unit
・ ・・・ ・・・ ・・・ ・・・ ・・・
32
8.6Setting of sample temperature scaling (Only 4 to 20mA DC) ・・・・・・・・・・・・・・・・・・・ 33
8.6.1 Setting at the det ector unit
・ ・・・ ・・・ ・・・ ・・・ ・・・
33
8.6.2 Setting at the display unit
・ ・・・ ・・・ ・・・ ・・・ ・・・
33
8.7Setting of sample temperature ・・・・・・・・・・・・ 33
8.8Setting of calibration curve data ・・・・・・・・・・・ 34
8.9Measurement ・・・・・・・・・・・・・・・・・・・・・・・・ 35
8.10Display of sample temperature Ts ・・・・・・・・・ 35
8.10.1 Display at the detector unit
・ ・・・ ・・・ ・・・ ・・・ ・
35
8.10.2 Display at the set ting display unit
・ ・・・ ・・・ ・・・
35
9.Inspection and maintenance
・・・・・・・・・・・・・
35
9.1Periodical inspection ・・・・・・・・・・・・・・・・・・・ 35
9.2Self-diagnosis function ・・・・・・・・・・・・・・・・・・ 36
9.3 Measures against troubles not included in
self-diagnosis ・・・・・・・・・・・・・・・・・・・・・・・・・ 37
9.4Maintenance parts・・・・・・・・・・・・・・・・・・・・・ 38
9.4.1 List of maintenance parts
・ ・・・ ・・・ ・・・ ・・・ ・・・
38
9.4.2 Repla cement of light source lamp
・ ・・・・・・・・・ ・
38
9.4.3 Replacement of motor
・ ・・・・・・・・・ ・・・・・・・・・
39
9.4.4 Replacement of lithium batterty
・ ・・・ ・・・ ・・・ ・・
40
9.4.5 Replacement of power supply unit
・ ・・・ ・・・ ・・・
41
9.4.5 Ram clearing
・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・ ・・・
41
10.Specifications
・・・・・・・・・・・・・・・・・・・・・・・
42
10.1Detector unit ・・・・・・・・・・・・・・・・・・・・・・・・42
10.2Fiver optics ・・・・・・・・・・・・・・・・・・・・・・・・・43
10.2.1 Reflective type fiver optics IR-WCRN
・ ・・・ ・・・
43
10.2.2 Output checker plate for reflective type fiver optics IR-WCNB
・ ・・・ ・・・ ・・・ ・・・ ・・・ ・
43
10.2.3 Raflective type fiver optics with lens IR-WCRE 43
10.2.4 Output checker plate for reflective type
fiver optics with lens IR-WCREB
・ ・・・ ・・・ ・・・
43
10.2.5 Measuring cell for liquid IR-WCC1
・ ・・・ ・・・ ・
44
10.2.6 Transparent type fiver optics IR-WCT
・ ・・・ ・・
44
10.3Accessories ・・・・・・・・・・・・・・・・・・・・・・・・・44
10.3.1 Power supply unit (conforming to CE) IR-WEP 44
10.3.2 Connection cable IR-WERT
・ ・・・ ・・・ ・・・ ・・・
44
CONTENTS
Page 6
INST No. INE -403-0P0
1
1.1 Summary
The infrared moisture meter IR-M2 series is designed as a fiber optic moisture meter for on-line moisture meter utilizing infrared absorption of moisture, and it consists of a detector unit, a fiber optics, a setting display unit and accessories. The detector unit is provided with 99 built-in calibration curves to output linear analog signals according to the moisture. Since the detector unit provides keys, digital displays and communications function, it can be used as a single unit or combined with a personal computer. Two kinds of fiber optics, reflective type and transparent type, are available. The reflective type fiber optics is for moisture measurement in powder state or in sheet state objects and the transparent type fiber optics is for moisture measurement in liquids by combining with a measuring cell for liquids. Up to 9 detector unit can be connected with the setting display unit which can set various parameters of the detector unitand displays the moisture value digitally. Accessories including cables and output checker plate are available for various applications. This manual describes about the detector unit, the fiber optics (reflective type and transparent type), the measuring cell for liquids and accessories. For the setting display unit, refer to a separate instruction manual for it.
1.2 Configuration
1. General
C H DATA
Transpa rent type fib er op tics
IR-WCT
Reflective type fiber opticswith lens attachment IR-WCRE
Reflective type fiber opticsIR- WCRN
Measuring cell fo rliquids
IR-WCC1
Detector:Model for IR-M2
Analog output
IR-WERT (connection cable)
24V DC
IR-WEP (power supply unit)
IR- GMEG1 (settin g displayunit)
Optical fiber
RS- 485
Communication and Analogoutput
Page 7
INST No. INE -403-0P0
2
2.1 Models
2.1.1 Detector unit IR-M2    
Measurement object
1: Universal type 2: High moisture type
Fiber optics combined
0: Reflective type 5: Transparent type
External input/output
0: Moisture output 4 to 20mA (Standard) 5: Correction input 4 to 20mA (Option) *1
8 :Correction input Pt100 (Option) *1
Communications interface
S : RS-485 (Standard) *2 R : RS-232C (Option) A : RS-422A (Option) *1
Special specification
Blank: None
x: Other special specification
2.1.2 Reflective type fiber IR-WCRN
   
2.1.3 Output checker plate for IR-WCRNB
reflective type fiber optics
2.1.4 Reflective type fiber optics with lens attachment IR-WCRE
2.1.5 Output checker plate for reflective
type fiber optics with lens attachment      IR-WCREB
2.1.6 Transparent type fiber optics          IR-WCT
2.1.7 Measuring cell for liquids     IR-WCC1
2.1.8 Connection cable          IR-WERT
2.2 Accessories
Component name Quantity
Remarks
Power supply unit (IR-WEP) 1no. S82K-05024 (Omron) Fuse 2 nos. SMP20 (Daito) Instruction manual 1 copy This document Communications instruction manual 1 copy Attached to IR-M2R/A only Inspection certificate 1 copy
2. Models and accessories
*1:
Combination of “5”
or “8” of
optional
external I/O and communications interface “A” is not
possible.
*2:Specify “S” for
combination with the setting display unit.
Page 8
INST No. INE -403-0P0
3
3.1 Setting of dipswitch es
Set the internal dipswitches before installation of the detector unit.
1) Open a door, remove 4 screws for fixing a setting display part and then remove the setting display part.
2) The setting display part is connected to the interior through cables. Lift the setting display part with due care so as not to damage the cable and disconnect the connector for the setting display part at the position shown in the figure.
3) Set the dipswitches according to the working conditions.
3. Setting
Turn off the power supply before removing the setting display part. War ning
Loosen the se fou rscrews
CH
DATA
SEL
ENT
CH/
DIP switch 1
Connector fordisplay
Page 9
INST No. INE -403-0P0
4
3.1 Setting of dipswitch es
SW Dipswitch
No
Function
OFF ON
Defaults
1 Measurement/
Calibration mode
Measurement mode
Calibration mode
OFF (Don’t change.)
2 Mathematical
operation format
Logarithm operation
Ratio operation ON (Don’t change.)
3 No.2 calibration
enable/disable
Disable Enable OFF
4 Connection
destination
Setting display
unit
PC
OFF : IR-M2 S ON : IR-M2R   IR-M2A
5 Key/display
enable/disable
Enable Disable OFF
6 Not used OFF 7
Correction input
enable/disable
Disable Enable OFF
8 Correction input 4 to 20mADC Pt100 OFF
3. Setting
1 2 3 4 5 6 7 8
O N
DIP switch 1
Page 10
INST No. INE -403-0P0
5
3.2 Installation of reflective type fiber optics
There are two types of reflective type fiber optics. One is “Without lens (IR-WCRN)” and the other is “With lens (IR-WCRE)”.
3.2.1 Installation of tip part
Set an attached holder to the installation place first. Insert the tip of fiber optics into the hole of the holder and fix it by tightening the tip fixing screw. Two types of holder, horizontal mounting type and flange type are attached.
The tip of reflection type fiberoptics with lens attachment
The tip of reflection type
fiber opticswith lens attachment
23 27
84
64
(
Ø
3
0
)
(
Ø
3
0
)
64 80
12
30
42
12
70
28
2-Ø9Mountinghole
2-Ø9Mounting hole
Tipfixing screw
Tipfixing screw
Place the tip of fiber optics with t
he following measurement distance from
an object. Reflective type fiber optics 15 to 50mm
Reflective type fiber optics with lens 25 to 100mm
War ning
Fix the fiber optics with appropriate metal tools at several places
for
avoiding affect with vibrations, etc. Make sure not to bend the fiber
optics
below its minimum bending radius (100R).
Warning
[ Model for IR-WCRN ]
15mm
Ø
50
Ø
20
Ø
6
[ Model for IR-WCRE ]
100m
25mm
Ø
40
Ø
25
Ø
20
Tipfixing screw
17
14
40 50 16
7
The tip of reflectiontype fiberoptic s
(
Ø
1
6
)
2-Ø6.6Mountingho le
(
Ø
1
6
)
2-
Ø5.5
Mountinghole
The tipof reflection
typefiber optics
44
8
27
16
48
38
25
Tipfixingsc rew
1)Reflection-typefiber
2)Reflectiontypefiberoptics withlensattachment
3. Setting
Page 11
INST No. INE -403-0P0
6
Purgeair inlet
Purge air flow:50 to 200Nl/min Purge air pressure:Lower than200kPa
3.2.2 Air purge (Only IR -WCRN)
When air purging is required, supply dry air not containing oil, dust, etc. from a purge air inlet. Connect a nylon tube of φ6mm in outer diameter and φ4mm in inner diameter to the purge air inlet. (When the reflection type fiber optics is used in ambient temperature more than 60°C, remove the purge air inlet and pipe to the screw hole of RC1/8 directly .
Purge air flow: 5 to 20NI/min Purge air pressure: Less than 200kPa (2kgf/cm2)
3.2.3 Connection s with detector unit
Connect the side being separated into two parts of light projecting side and light receiving side with the fiber optics connectors of the detector unit.
The fiber optics has “Light project ing
side (D)”and “Light receiving side
(E)”and each one has a carved mark. Connect the fiber optics through the rubber bush as shown in the following figure.
War ning
3. Setting
Connector forfiber [Light projecting side]
Connector forfiber [Light receivingside]
Dent department
Protrusiondepartment
D E
Rubberbush
Bushfittingring
Connector ring
Optical fiber
D:Connect tolight projectingside E:Connectto light receivingside
Page 12
INST No. INE -403-0P0
7
3.2.4 Optical path
Select a place where water drops, dust particles, smoke, water steam, etc. will not enter between the tip of fiber optics and an object. If these substances cannot be prevented and their influences are negligible, purge them with air, or other remedial measure is necessary. If direct current (DC) light like the natural light, which does not change its intensity momentarily , comes to an object, it does not cause any disturbance in measurement, but alternate current (AC) light like as electric bulb, flame, etc., causes disturbance. Further, strong light such as direct sunlight also causes disturbance. In such cases, it is necessary to shade the disturbing light to prevent it from direct shining on an object.
3.2.5 Measuring place (for objects of pulverized or granular substance)
For installation of the tip of fiber optics on on-line, select a place where the moisture difference is small between the surface and the inside of an object, like as at a hopper outlet or just behind the conveyor belt transfer place. If such a place cannot be selected, face the inside to the surface by using a baffle plate during measurement as shown in the figure.
3.2.6 For the objects in sheet state such as paper, etc.
Tilt the tip of fiber optics by 15° from the vertical face as shown in the figure.
3. Setting
Baffle plate
Measuringobject
Belt conveyor
Paperor othersheets
15°
Page 13
INST No. INE -403-0P0
8
3. Installation of transparent type fiber optics
3.3.1 Installation of cell for liquids
Install a measuring cell to flow a sample liquid from down to up. Connect with the fiber optics between the detector unit and the measuring cell. Make sure to connect the light projecting side and the light receiving side of the fiber optics to the correct sides of the detector unit and the measuring cell.
Fix the fiber optics with appropriate metal tools at several places for
avoiding affect with vibrations, etc. Make sure not to bend the fiber optics below its minimum bending
radius
(250R).
War ning
C H D ATA
Detector:Mode l for IR-M2
Power supply unit:
Model for IR-WEP
Setting display unit:
Model for IR- GMEG1
Measuring cell
Output
Input of sample liquid
Light receiving side fiber optics
Light projecting side fiber optics
Fixthe fiberoptics with appropriatemetaltools atseveral places.
3. Setting
Page 14
INST No. INE -403-0P0
9
3.3.2 Connection with detector unit
Connect the fiber optics, as shown in the following figure, through the rubber bush attached to the detector unit .
3.3.3 Disassemble cleaning
For accurate measurements, recommend periodical cleaning of the cover glass in a liquid contacting part of the measur ing cell.
1) Remove the fiber optics (both of light projecting side and light receiving side), from the measuring cell.
2) Next, remove 4 screws fixing a connecting part and remove the connecting part from the liquid-contacting part.
3) Remove a fitting ring by using the tool A. Then after screwing the tool B into the cover glass, pull out the cover glass from the liquid-contacting part.
4) Wipe off dirt on the glass surface with a gauze, etc.
5) Confirm that an O ring is inserted in liquid-contacting part and push in the cover glass to the liquid-contacting part.
6) Match the positions of E and D characters of the connecting part and the liquid contacting part, and fix the connecting part.
7) After the cleaning of the cover glass, make sure to re-calibrate your moisture meter.
3. Setting
OUT
I N
Connectingpart
4- Fixingscew
Liquidcontacting part
Tool A
ToolB
Fittingring
Curler
Cover glass
Oring
Connector for fiber [Lightprojectingside ]
Connector for fiber [Lightreceiving side]
Dent department
Protrusion department
D
E
Rubberbush
Bushfittingring
Connector ring
Optical fiber
D:Connect to light projectingside E:Connectto light receivingside
Page 15
INST No. INE -403-0P0
10
3.4 Installation of detector unit
The detector unit is a wall-mounting type with 4 mounting holes. Fix the detector unit to wall surface, etc. with 4 pieces of M8 screw. The outside and mounting dimensions are shown in the figure.
3.4.1 Outside and mounting dimensions of detector unit
115
2
270
180
3 03 0
98
10
12
2323
74.5
47
47
165
145
125
4-M8Mountin g hole
Vibrations and impacts
Use the detector unit in a place where is free of any noticeable
vibrations
or impacts.
War ning
Induction
The detector unit is designed for anti-induction. However separate
the
detector unit from induction heating oscillators or power
lines as far as
possible.
War ning
Ambient temperature and humidity The ideal ambient temperature of the detector unit is 10 to 30°C
and the
ambient humidity as low as possible is recommended. In high
humidity
environment,
you may have error in measurement due to condensation of
steam by change in temperature.
War ning
3. Setting
Page 16
INST No. INE -403-0P0
11
4.1 Connections
The terminal board is accessible by opening the door of the detector unit. Lead in cables through a small sized cable inlet at the lower part and connect them to terminals. Lead in the exclusive cable IR-WERT through the large cable inlet.
4.2 Layout of connection cable
For connection of cables, be careful of the following points.
4. Connections
For preventing electric shock,
make sure to turn the power supply off
before wiring.
Warning
Caution
Use a cable conforming to the applicable outer diameter (φ4.5 to 6.5
mm)
of the small cable inlet; otherwise waterproofing of the
detector unit may
deteriorate. Tighten the nut at the cable inlet securely after connections.
Separate the cables from induction heating oscillator and power lines. Keep the cables free of deposit of water, oil, etc.Don’t bend the cables extremely or apply any excessive force to them .
For permanent layout, protect the cables with conduits, etc.
SA (Black)
SB (White)
SG (Brown)
G*1 (Green)
P+ (Red)
P­(Blue)
P­(Blue)
P+ (Red)G(Green)
V+ V-
LN
24V DC
Ground
Power supply
(IR-WEP)
100-120V AC 200-240V AC
Terminal
board
Cable inlet (small): Outer diameter of
applicable cableØ4.5toØ6.5
Cable inlet (large):
[For IR-WERT]
Outer diameter of applicable cableØ8.5 to
Ø10
.5
Connection cable
(IR-WERT)
Communication output forIR-GMEG1, pe rsonal
computerand PLC.
Communication output RS-485 (Optional:RS-232C,RS-42 2A)
*1:
Don't connect this cable usually,
but connect it, if the unitis interfared
withnoises.
SA
SB
SG
OUT
Measuredvalue output :
4 to 20mADC
Loadresistance:Lower than 500
Ω
(Optional:RS- 232C, RS-422A)
P+
P-
24VDC
1.5A SUPPLY
Page 17
INST No. INE -403-0P0
12
4.3 Multi-detector connections
By using the setting display unit, you can connect the detector unit up to 9 sets. Two wiring methods are available. One is wiring through a relay box (IR-WEE) and the other is without the relay box (IR-WEE).
4.3.1 Connections through relay box
4. Connection
Caution
For
both wiring methods, before connecting with the setting display unit or
before turning on the power supply of the setting display unit,
number
setting to the detector units is required. For the number setting, be careful not to overlap any detector unit number of each detector unit. Refer to [6.2 Setting of detector unit number ] for the details.
Caution
1. Shorten the length of exclusive cable (IR-WERT * mark) as much as
possible.
2. For wiring to the relay box, refer to “Instruction manual of relay box”.
Detecto r
IR-WERT*
IR-WERT*
IR-WERT*
Relayin g box
(Power supply unit is built in)
Relayin g box(Power supplyunit isbuilt in)
Relayin g box(Power supplyunit isbuilt in)
2-conductor twisted cable(withshield)
2-conductor twisted cable(withshield)
2-conductor twisted cable(withshield)
100-240 V AC
Detecto r
Detecto r
Setting diplayunit
Page 18
INST No. INE -403-0P0
13
4.3.2 Connections without relay box
4. Connection
Detector
Detecto r
Detecto r
IR-WERT*
IR-WERT*
IR-WERT*
Setting diplay unit
100-240 V AC
2-conductor twisted cable(withshield)
2-conductor twisted cable(withshield)
2-conductor twisted cable(withshield)
Taping
Taping
Taping
Connect theP+(Red), P-(Blue), and G(external shield wire) of IR-WERT to power supply. And connect the SA (Black), SB(White)and SG(internal shield brown) of signal wire to the 2-conductor twistcable(with shield)
Fasten the signal wires in IR-WERT (SA, SB andSG) and thesignal wirein 2-conductor twist cable(SA,SB and SG)together at the terminal board of the detector.
Page 19
INST No. INE -403-0P0
14
5.1 Names and functions of detector unit
(1) Hole for installation of detector unit: Four M8 holes for installation of the detector unit
to wall surface, etc.
(2) Data display window: Displays calibration curve number and measured value. (3) Door: By opening the door, key switches and the terminal board are accessible. (4) Cable inlet (small): Cable inlet for connection cable. Outside diameter of connection
cable applicable to this inlet: φ4.5 to 6.5mm.
(5) Cable inlet (large) Cable inlet for exclusive cable IR-WERT. Outside diameter: φ10mm. (6) Serial No plate: Serial No plate of this detector. (7) Connector on light projecting side : Connector for the light projecting side of the fiber
optics. Rubber bush is prepared for the connection.
(8) Connector on light receiving side : Connector for the light receiving side of the fiber
optics. Rubber bush is prepared for the connection.
(9) Setting display: Display part of this detector. (10) Measurement part : Optical assembly, detector element and electric circuits are built-in. (11) Lamp housing : Lamp is located inside. (12) Power supply part: Terminal board for wiring, power fuse, power supply unit and
output board are built in.
(13) Power fuse: A power fuse is mounted to the fuse holder.
Power fuse: Insertion type, 2A, time lag fusion type
(14) Terminal board: For wiring of connection cable. Terminus screw size: M4.
5. Names and functions of component parts
DATACH
SEL CH/・ ENT
DATACH
(1) Detector mounting
screw hole- 2
(2) Data display
window
(3) Door
(5) Connection cableinlet (small)
(6) Connection cableinlet (large)
(6) Manufacturing
nameplate
(8) Connector on lightreceiving side
(7) Connector on lightprojecting side
(11) Lamp (9) Setting display
part
(14) Terminal board
(10) Measuring part
(12) Power supply part
(13) Power
supply
fuse
For preventing electric shock,
make sure to turn the power supply off
before wiring.
War ning
Page 20
INST No. INE -403-0P0
15
5.2 Names and functions of setting display part
(1) CH display: Displays a calibration curve number in measurement mode and a parameter
item in setting mode.
(2) Data display Displays a measured value in the measurement mode and a parameter in
the setting mode.
(3) Setting keys
SEL : For entering into the setting mode and switching a setting screen. CH/ : For setting a calibration curve number in the measurement mode and for
entering a decimal point in the setting mode. When the key is pressed again, the decimal point disappears.
: For entering into a parameter setting in the setting mode and changing the digit
of the parameter. : For changing the parameter in the setting mode. : For changing the parameter in the setting mode.
ENT : For storing the parameter in the setting mode.
5. Names and functions of component parts
CH/・
SEL
ENT
CH
DATA
Setti ng display part
(1) CH Display
(2) DataDisplay
(3) Setting
keys
Page 21
INST No INE -403-0P0
16
6.1 Confirmation before operation
Check the following two points before turning on the power supply.
1) Have the dipswitches been set? (Refer to [3.1 Setting of dipswitches])
2) Is wiring correct? (Refer to [4.1 Wiring]).
6.2 Setting of detector unit number
Detector unit number s are for identifying each detector unit when multiple detector units are connected to the setting display unit. (Refer to [4.3 Multi-detector connection]). Set a detector unit number before connecting the setting display unit.
1) Enter into the operation condition setting mode by pressing SEL key and ENT key simultaneously to
set CH display “t” for 2 seconds.
2) Press SEL key 4 times to set CH display to “Hn”.
3) By pressing key, the numeric blinks for entering a detector unit number.
4) Change the numeric by key or key.
When key is pressed, the num eric is increased. When key is pressed, the num eric is decreased. When key is pressed, the blinking digit shifts.                  
5) When ENT key is pressed, the numeric stops
blinking and the detector number is stored.
6) Return to the measurement mode by pressing SEL
key for 2 seconds.
6.3 Setting of calibration curve data
The output characteristic of infrared moisture meter depends upon measuring objects. It may also change according to the process conditions and sample moisture measuring conditions. Therefore, it is necessary for accurate moisture measurements to perform a sample test of the object and obtain the correlation (This is called as calibration curve.) between the moisture value (%H2O) obtained by a drying method or other measuring methods and the absorbance “x” measured by the moisture meter. This paragraph describes the setting method for calibration curve data, assuming that calibr ation curves have been already created. When the calibration curves have not been created yet, refer to [7. Creation of calibration curves].
Refere nce
The default number of the detector unit is “1”.
For using it with a number other than 1, set the detector unit number
by
following order.
CH
DATA
t 0.0
CH
DATA
Hn
Detector No
6. Operation
Page 22
INST No INE -403-0P0
17
6.3.1 Setting data
The followings are the setting data of calibration curve.
Data name Disp
lay
Setting range
Defaul t value
Remarks
(1) Calibration curve No. CH 1 to 99 1 (2) Polynomial coefficient a0 A0 0 to ±99999 0.0000 Effective numeric: 5 digits (2) Polynomial coefficient a1 A1 0 to ±99999 0.0000 Effective numeric: 5 digits (2) Polynomial coefficient a2 A2 0 to ±99999 0.0000 Effective numeric: 5 digits (2) Polynomial coefficient a3 A3 0 to ±99999 0.0000 Effective numeric: 5 digits (3) Correction expression coefficient b0 B0 0 to ±9999.9 0.0000 Effective numeric: 5 digits (3) Correction expr ession coefficient b1 B1 0 to ±9999.9 1.0000 Effective numeric: 5 digits (4) Output low limit moisture Hi 0 to 9999.9 0.00
Decimal digit depends on
specification.
(4) Output high limit moisture Hi 0 to 9999.9 100.00
Decimal digit depends on
specification.
(5) Low limit alarm moisture AL 0 to 9999.9 0.00
Decimal digit depends on
specification.
(5) High limit alarm moisture AH 0 to 9999.9 100.00
Decimal digit depends on
specification.
(6) Preset moisture PS 0 to 9999.9 0.00 Decimal digit depends on
specification.
(1) Calibration curve number
An optional number from 1 to 99 can be designated as the calibration curve. It is
designated sequentially from “1” usually.
(2) Polynomial coefficient “a0 to a3”
Assume that the calibration curve is represented by third-degree or lower-degree polynomial as shown below.
y = a3x3 + a2x2 + a1x + a0
where, “y” is the moisture value (%H2O) and “x” is the absorbance (“CH0” data of this moisture meter).
(3) Correction expression coefficient “b0 to b1”
These coefficients are provided for primary expression correction (zero and span correction) with reference to moisture value “y”. Assuming that the moisture value after correction be “Y”, we obtain, y = b1y + bo
(4) Output low limit moisture, output high limit moisture
These moisture values are specified as the output low limit and output high limit on the analog output scaling. These values are scaled to be 4mA at the output low limit moisture and 20mA at the output high limit moisture.
(5) Low limit alarm moisture, high limit alarm moisture
The low limit alarm activates when a measured value is lower than the low limit alarm moisture, while the high limit alarm activates when a measured value is higher than the high limit alarm moisture. Since the detector unit does not provide any alarm output function, the setting display unit is required when an alarm output is necessary.
(6) Preset moisture
By turning on the preset output by key operations or through communications, the moisture value being set as preset moisture is displayed and output, regardless of the measured value.
6. Operation
Page 23
INST No INE -403-0P0
18
y
1
y
2
y
3
y
4
y
5
y
7
99.999
y
6
-10
0
10
x
1
x2x3x
4
x5x
6
x
7
6.3.2 Setting of calibration curve data
1) Press SEL key for 2 seconds to enter into the calibration curve data setting mode.
2) By pressing key, a numeric blinks for entering a
calibration curve number.
3) Change the numeric by key, key or key.
When key is pressed, the numeric is increased. When key is pressed, the numeric is decreased. When key is pressed, the blinking digit shifts.
4) When ENT key is pressed, the numeric stops blinking and the calibration curve number is stored.
5) Press SEL key to enter into the setting mode for
polynomial coefficient a0.
6) Enter into a numeric setting mode by pressing key. Change the numeric by key or key and store it by
pressing ENT key. For entering a decimal point, press CH/・ key at the
blinking digit.
7) Press SEL key to enter into the next data setting mode.
8) Repeat the above procedure to set all data.
9) After the above setting procedure is completed, return to the measurement mode by pressing SEL key for 2 seconds. The automatic return is made if no key is pressed for one minute.
 
6.4 Preparation of calibration curve by knee point line approximation
When a calibration curve is obtained in a graph but it cannot be expressed in third-degree or lower-degree polynomial, the calibration curve can be created by broken line approximation.
6.4.1 Setting of knee point data
1) Supposing that the calibration curve graph is as shown in the following figure.
2) Divide the calibration curve graph
appropriately and express it with broken line approximation.
3) In this case, the first junction point and the last junction point should sufficiently include the measuring range. You can plot up to 20 junction points including the first junction point and the last junction point. Determine intermediate junction points appropriately according to the approximation accuracy.
4) Supposing that the junction point data obtained by the above are as shown below.
First junction point D1: (x1, y1) D2: (x2, y2) Last junction point Dn: (xn, yn)
Setting data number
displayed
CH
DATA
CH 0
Blinking Detector No
CH
DATA
CH 0.0000
Setting data
displayed
CH
DATA
CH 0
6. Operation
Data number
Page 24
INST No INE -403-0P0
19
6.4.2 Knee point line setting data
The following data is necessary for the creation of the calibration curve by knee point line approximation.
Data name Display Setting
range
Default value
Remarks
(1) Calibration curve number
CH 1 to 99 1
(2) Data count N 3 to 20 3 (3) Data number Dn 1 to 20 1
(3) Sample data
xn
Hn 0 to ±9999.9 0.0000 Valid number 5 digits
(3) Sample data
yn
yn 0 to ±9999.9 0.0000 Valid number 5 digits
(4) Regression/Broken line
Pb 1 to 4 1
1 to 3:
Polynomial expression degree
(Regression operation)
4: Broken line approximation
(5) Standard deviation Sd 0 to 999.99 0.0000
Setting not required in broken line
approximation
(1) Calibration curve number: Calibration curve number can be specified in optional
numbers from 1 to 99 but usually it is specified in order from 1.
(2) Data count: It is the number of knee point data and the maximum is 20 points. (3) Data number, sample data x
n,
sample data yn:
Enter the knee point data obtained in [6.4.1 Setting of knee point data].
(4) Regression/knee point line: With entering 1, 2 or 3, the polynomial expression with that
degree can be obtained according to the above sample data by regression operation. When 4 is entered, knee point line approximation is obtained.
(5) Standard deviation: When the regression operation is specified, standard deviation is
displayed. This setting is not required in knee point line approximation.
  
6. Operation
Page 25
INST No INE -403-0P0
20
6.4.3 Setting of calibration curve data for knee point line approximation (1) Setting of calibration curve number
1) Press SEL key and CH/. key simultaneously for 2 seconds to enter into the sample data setting mode.
2) By pressing key, a numeric blinks for entering a
calibration curve number.
3) Change the numeric by key, key or key.
When key is pressed, the numeric is increased. When key is pressed, the numeric is decreased.
When key is pressed, the blinking digit shifts.
4) When ENT key is pressed, the numeric stops blinking and the calibration curve number is stored.
(2) Setting of data count
1) Press SEL key to enter into the data count setting
mode.
2) Change a numeric by the procedure mentioned in the above “(1) Setting of calibration curve number 2) and
3)” and store it with ENT key.
(3) Setting of data number, sample data xn and
sample data yn
1) Press SEL key to enter into the data number setting mode.
2) Change a numeric by the procedure mentioned in the
above “(1) Setting of calibration curve number 2) and
3)” and store it with ENT key.
3) After setting the above, press SEL key for entering into the sample data xn setting mode.
4) After setting the above, press SEL key for entering
into the sample data yn setting mode.
5) After setting the above, press CH/. key to return to
“1) Data number setting mode” and then set the next data number.
6) Repeat the above procedures for setting all junction point
data.
(4) Regression/Knee point line
1) After completing the setting of knee points, press SEL key to enter into the regression/knee point line setting mode.
2) Enter “4” in the selection of regression/knee point line.
3) After the settings are completed, press SEL key for 2 seconds to return to the measurement mode.
Automatic return to the measurement mode is made if
any key is operated for 1 minute.
CH
DATA
Dn 0.0000
CH
DATA
yn 0.0000
CH/・ SEL
CH
DATA
pb 4
CH
DATA
CH 1
CH
DATA
CH 01
Detector No
Blinking
CH
DATA
n 4
Setting data
displayed
Setting data number
displayed
CH
DATA
Dn 1 
6. Operation
Page 26
INST No INE -403-0P0
21
6.5 Setting of operating conditions
Set the operating conditions including time constant and decimal place in the display.
6.5.1 Setting data
Set the following data in operating conditions. However, the number of displayed setting data varies depending on the setting of dipswitches.
Data name Display Setting range Default value Remarks (1) Time constant t 0 to 9, 9 to 99 0.0 Unit: Seconds (2) Hold ON/OFF Hd ON, OFF OFF (3) Preset output ON/OFF PS ON, OFF OFF (4) Decimal place in display Dt 0 to 4 1 (5) Detector unit number Hn 1 to 9 1 (6) Communications
speed *1
SP
1200, 2400, 4800,
9600, 19200
9600
(6) Parity *1 P None
odd EVEn
EVEn nonE :Non parity
odd : Odd parity
EVEn :Even parity (6) Data length *1 D 7, 8 7 (6) Stop bit length *1 B 1 to 2 1 (6) BCC enable/disable *1 Bc ON, OFF OFF (disable) (7) Sample temperature
correction enable/disable
St ON, OFF OFF (disable)
(8) Correction input
scaling L*2
TL -999 to 999 0
(8) Correction input
scaling H*2
TH -999 to 999 100
〔Engineering mode
(9) Weightα
Af 0 to 1.000 0.500
(10) Calibration constant k1 k1 0 to ±9.9999 1.0000 (11) Calibration constant
k2 *3
k2 0 to ±9.9999 0.0000
Note) The setting data of *1 to *3 is displayed or not depending on the setting of
dipswitches.
Displayed data Setting of dipswitch 1 Contents *1 Setting data No. 4 : ON Connection destination: Personal computer
*2 Setting data No. 7 : ON No. 8 : OFF Corrected input: Enable
Corrected input specification: 4 to 20mADC
*3 Setting data No. 3 : ON No. 2 Calibration: Enable
6. Operation
Page 27
INST No INE -403-0P0
22
(1) Time constant:
If a measured value fluctuates noticeably, the signal can be set to be dull by delaying the response of this moisture meter. Set the smoothing time (equivalent to the time constant of analog instrument) in 0.1-second increment for less than 10 seconds and in 1-second increment for more than 10 seconds.
(2) Hold ON/OFF:
When holding of measured value is desired, turn it ON. At this time, analog output will also be held.
(3) Preset output ON/OFF:
By turning on the preset output, this moisture meter displays the preset moisture value and the analog output becomes the value corresponding to the preset moisture value.
(4) Decimal place in moisture display:
The decimal place can be set for easily monitoring the moisture value measured.
(5) Detector unit number:
This is the number for identifying the detector unit for connection of multiple detector units. This setting has been completed in [6.2 Setting of detector unit number].
(6) Communications speed, Parity, Data length, Stop bit length, BCC enable/disable:
These are for communications with a personal computer, a sequencer or other similar devices. These data are not necessary to set for combining this moisture meter with the setting display unit.
(7) Sample temperature correction enable/disable:
These are used for sample temperature correction enabled.
(8) Correction input scaling L, Correction input scaling H
These are used for scaling of 4 to 20mA correct ion input.
(9) Weight á
In 3-wavelength operation, the ratio of 2 comparative wavelengths is determined by this
value.
(10) Calibration constant K1:
This is the calibration constant at the time of zero side calibration.
(11) Calibration constant k2:
This is the calibration constant at the time of span side calibration.
6. Operation
Page 28
INST No INE -403-0P0
23
6.5.2 Setting
1) Press SEL key and ENT key simultaneously for 2 seconds to enter into the operation condition setting
mode.
2) By pressing key, a numeric blinks for entering a data.
3) Change the numeric by key, key or key.
When key is pressed, the blinking digit shifts.
For entering the decimal point, press CH/. key.
4) When ENT key is pressed, the numeric stops blinking
and the data is stored.
5) Press SEL key to enter into the setting mode of the
next data. Repeat the above procedures to set all data except the engineering mode.
6) For setting of data in the engineering mode, press ENT
key in the screen shown in the right side to enter into the setting mode of the engineering mode.
7) After the settings are completed, press SEL key for 2
seconds to return to the measurement mode. Automatic return to the measurement mode is made if any key is operated for 1 minute.
6.6Calibration
For using this moisture meter accurately, regular calibration once in three months is requested.
1) Supply power more than one hour to this moisture meter before calibration.
2) Turn the calibration curve number to 0. (For the change in calibration curve number, refer to [6.7 Setting of calibration curve number]. By this setting, absorbance “x” is displayed as a data.
3) Execute the calibration by pressing the three keys SEL , , and ENT simultaneously .
Make sure that light absorbance “x” is within 0.00000±0.0010. The calibration is
completed.
CH
DATA
Eng
CH DATA
t 0.0
Setting data
displayed
Setting data number
displayed
CH
DATA
  
0 0.0021
Refere nce
For the reflective type fiber, place the output checker plate
at the tip of
fiber to see the plate.
Refere nce
For the transparent type fiber, disassemble the measuring cell
to be
empty.
6. Operation
Page 29
INST No INE -403-0P0
24
6.7 Setting of calibration curve number
Set the calibration curve number according to the measuring object.
1) By pressing CH/. key, a numeric blinks for entering a calibration curve number.
2) Change the numeric by key, key or key.
3) When ENT key is pressed, the numeric stops blinking and this moisture meter becomes the measurement mode
at the calibration curve number changed.
6.8 Correction of calibration curve
If the actual moisture value is different from the moisture value being preset due to the difference of one-line and off-line, or other causes, correct it by the correction coefficients b0 and b1 in [6.3 Setting of calibration curve data]. Usually, the shift correction is done with b0 only.
CH
DATA
01 10.0
Blinking
Refere nce
[Example]
When the measured moisture before correction is 15(%H2O)
and actual
moisture value is 13 (%H2O),
Set b0 = -2(%H2O) .
6. Operation
Page 30
INST No INE -403-0P0
25
7. Preparation of calibration curves
The output characteristics of the moisture meter depend upon the measuring objects. Also, the output characteristics may also change according to the process conditions and moisture measurement conditions of samples for certain measuring objects. Therefore, for accurate measurement of moisture, it is necessary to carry out the sample tests of each measuring object in advance and obtain the relative relation (this is called as calibration curve) between the moisture value (%H2O) obtained by the drying method or other measuring methods and the absorbance "x" measured by the moisture meter This chapter describes the measurement of the moisture value by the drying method. However, the method of creation of the calibration curves is same even if the moisture value is measured by the Karl Fischer method or other methods.
7.1 Sample preparation
7.1.1 Powder or granular
(1) Take a sample of about 2 to 4 liters from the measuring object. (2) Dry up the sample up to the absolute dry condition by a dryer. Particularly be careful with
the heating temperature so as not to denature the sample.
(3) Divide the dry sample every 100 to 200cc into 5 to 6 samples separately, although it is
recommendable to divide the sample into many samples.
(4) Add water to these 5 to 6 samples bit by bit so that the moisture values of each sample
divides the measuring range almost evenly.
(5) Stir each sample with water addition sufficiently, and put it into a polyethylene bag and
seal it tightly.
(6) Leave each sample for about 2 days until its moisture is stabilized.
If a sample cannot be heated or if it is hardened by adding water and thus
can ’
t be measured, adjust the moisture correspondingly according to the
procedure in case of a paper s hown below.
Caution
7. Preparation of calibration curves
1
2 3 N
Page 31
INST No INE -403-0P0
26
7.1.2 Paper (sheet substance)
(1) Cut a sample to have a size of about 100 x 100mm by using a cutter or the like. (Number
of samples: 9 sheets)
(2) Divide nine sheets of the sample into 3 groups with 3 sheets in each group. Adjust the
moisture of each group under the following conditions.
Drying : Leave the samples in a desiccator containing silica gel for 2 days. In case of
papers, the moisture becomes about 4%H2O.
In the air: Leave the samples in an indoor atmosphere.
In case of papers, the moisture becomes about 7%H2O in summer or about 5%H2O in winter.
Moistening: Leave the samples in a desiccator containing salt water for 2 days.
In case of papers, the moisture becomes 8 to 9%H2O usually or maximum about 11% H2O.
7.1.3 Liquid
(1) Prepare 500 to 1000cc of sample dried nearly up to the lower limit moisture of the
measuring range.
(2) Prepare 5 to 10 containers (with lid) of 100 to 200cc capacity and put the above sample
till the containers are full.
(3) Add water to these 5 to 10 samples bit by bit so that the moisture values of each sample
divides the measuring range almost evenly.
(4) Stir each sample with water addition sufficiently, and put a lid on the container tightly.
Wear clean gloves when handing samples so as not to touch them by
naked hands directly. It is convenient for you to mark the samples for identifying
its front and
back surfaces and its flow direction.
Caution
If a sample is with strong moisture absorption and low moisture less than
few % is required, even if water is not added,
the moisture of the sample
may reach to its level by absorption of moisture only. For such samples, adjust moisture by changing a moisture absorption time.
Caution
7. Preparation of calibration curves
1
2
3
4
5
6
Withsilicagel
Withsalt water
Page 32
INST No INE -403-0P0
27
w1-w2
w1-w0
×100
(%H2O
)
7.2 Sample measurement
Be careful with the following cautions during measurement.
(1) Before measurement, warming-up of this moisture meter more than 1 hour is required. (2) Place a sample at the same position as in actual measurement, set the calibration curve
number to 0, and read the display (absorbance "x").
(3) Perform the measurement rapidly.
7.2.1 Powder or granular
(1) Prepare the trays (Diameter 100 to 150mm, Depth about 20mm) by the same quantity as
the measuring sample quantity. (2) Weigh the weight w0 of each tray. (3) Spread the sample being sealed in the polyethylene bag on the tray it till the surface
becomes flat and the bottom cannot be seen. (4) Put the tray at the measuring position, set the calibration curve number to 0, and read the
display (absorbance "x"). If the measured value cannot be read easily, delay the response
by the smoothing function. (5) Stir the sample rapidly with a spoon and measure it again. Repeat this measurement 2 or
3 times per sample. (6) Weigh the weight w1 of the tray together with the sample. (7) Perform the same measurement sequentially about all samples. (8) Put each sample together with its tray into a dryer, and dry it out to be the absolute dry
condition (for longer than 2 hours at 105 to 110ºC usually). (9) Cool down the sample after drying, and weigh the weight w2 of each tray together with
the sample. (10) Record the measured value in the following table every measurement, and obtain the
moisture value by the following formula.
No. Absorbance "x" w0 W1 w2 Moisture value (%H2O)
1 x1 y1 2 x2 y2 3
:
x3 :
y3
:
N xn y4
7. Preparation of calibration curves
Page 33
INST No INE -403-0P0
28
7.2.2 Paper (Sheet substance)
(I) Weigh the weight w1 of a sample before measurement. (2) Hold the sample by a paper holder, put it by tilting at 15º to the measuring position, set
the calibration curve number to 0, and read its display (absorbance "x"). In this case, match its front and back surfaces and flo w direction of the samples.
(3) Weigh the weight w
2
of the sample after measurement. Perform the measurement from
(1) to (3) as quickly as possible. (4) Perform the same measurement sequentially for all samples. (S) Put each sample into a dryer, and dry it out to be the absolute dry condition (for longer
than 2 hours at 105ºC usually by using paper stand). (6) Cool down all samples in a desiccator containing silica gel after drying. (7) Weigh the weight w3 of each sample. (8) Record the measured value in the following table every measurement, and obtain the
moisture value by the following formula.
No. Absorbance "x" w0 W1 w2 Moisture value (%H2O)
1 x1 y1 2 x2 y2 3
:
x3 :
y3
:
N xn y4
7.2.3 Liquid
Execute the measurement in order from the object of low moisture to the object of high moisture.
1) Put a sample into the measuring cell, set the calibration curve number to 0, and read the display (absorbance "x").
If the measured value cannot be read easily, delay the response by the smoothing function.
2) With the Karl Fischer method or other methods, measure moisture value of the sample put in the measuring cell.
3) Repeat the above measurements of 1) and 2) for 2 or 3 times per sample.
4) Record the measured value of all samples by the same measuring procedure in the following table, and obtain the average value.
Absorbance "x" No. 1 2 3 Average
Moisture (%H2O)
1 x1 y1 2 x2 y2 3
:
x3
:
y3
:
N xn y4
(w1+w2)/2-w3
(w1+w2)
×100
(% H2O
)
For taking out a sample from a
container, shake the container well to
uniform the moisture in the container and then take the sample out.
When the next sample is measured, throw away the sample
already
contained completely and wash the measurement cell
for two or
three times for the next measurement.
Caution
7. Preparation of calibration curves
Page 34
INST No INE -403-0P0
29
Absorbance (x)
M oisture value (y)
(%H2O
)
7.3 Preparation of calibration curves
Create a calibration curve by plotting data obtained by [7.2 Sample measurement] on a graph. Plot the absorbance "x" of this moisture meter on the X-axis and also plot the moisture value (%H2O) on the Y-axis.
Obtain a curve with minimize errors from this graph as a calibration curve. Usually the curve with minimize errors is obtained by the regression operation. For the creation of a calibration curve by the regression operation, refer to [7.3.1 Computation of regression expression] and [7.3.2 Confirmation of regression expression].
If a graph contains inflection points, use the creation of a calibration curve by broken line approximation. For the creation of a calibration curve by broken line approximation, refer to [6.4 Creation of calibration curve by broken line approximation].
7.3.1 Computation of regression expression
This moisture meter has regression operation function by the minimum mean square method.
By this function, first-degree to third-degree polynomials can be obtained by entering the data
obtained in [7.2 Sample measurement].
Assume that there are “n” sets of sample data ("xn, yn). (in condition that “n” is 20 or less)
(1) Setting of calibration curve number
1) Press SEL key and CH/. key simultaneously for 2 seconds to enter into the sample data setting mode.
2) By pressing key, a numeric blinks for entering a calibration curve number.
3) Change the numeric by key, key or key.
When key is pressed, the numeric is increased. When key is pressed, the numeric is decreased. When key is pressed, the blinking digit shifts.
4) When ENT key is pressed, the numeric stops blinking
and the calibration curve number is stored.
(2) Setting of data count
1) Press SEL key to enter into the data count setting
mode.
2) Change a numeric by the procedure mentioned in the
above “(1) Setting of calibration curve number 2) and
3)” and store it with ENT key.
CH
DATA
CH 0
CH
DATA
CH 00
CH
DATA
n 3
Setting data
displayed
Setting data number
displayed
Blinking
Calibration
Curve number
7. Preparation of calibration curves
Page 35
INST No INE -403-0P0
30
(3) Setting of data number, sample data xn and sample
data yn
1) Press SEL key to enter into the data number setting mode.
2) Change a numeric by the procedure mentioned in the above “(1) Setting of calibration curve number 2) and
3)” and store it with ENT key.
3) After setting the above, press SEL key for entering
into the sample data xn setting mode.
4) After setting the above, press SEL key for entering
into the sample data yn setting mode.
5) After setting the above, press CH/. key to return to “1) Data number setting mode” and then set the next data number.
6) Repeat the above procedures for setting all sample data.
(4) Computation of regression expression
1) After completing the setting of sample data, press SEL
key to enter into the regression/ broken line setting mode.
2) Enter “1, 2 or 3” in the selection of regression/broken line.
A polynomial with corresponding degree can be obtained.
3) By pressing SEL key, the standard deviation
(approximate accuracy of the regression expression) of the above regression expression is displayed.
3) After the settings are completed, press SEL key for 2
seconds to return to the measurement mode. Automatic return to the measurement mode is made if any key is operated for 1 minute.
7.3.2 Confirmation of regression expression
The coefficient of the regression expression obtained by
the above can be confirmed by the following procedure.
1) Press SEL key for 2 seconds to enter into the calibration curve setting mode.
2) Set calibration curve number, for which the regression
expression is obtained.
3) Press SEL key to display the coefficient a0 of the
regression expression.
4) For displaying a1 to a3, press SEL key in the same way.
5) After the confirmation is completed, press SEL key for 2
seconds to return to the measurement mode. Automatic return to the measurement mode is made if any key is operated for 1 minute.
CH
DATA
dn 1
CH DATA
Hn 1
CH
DATA
yn 1
CH/・ SEL
CH
DATA
pb 1
CH
DATA
Sd 0.12
CH
DATA
CH 1
CH DATA
A0 1.2544
7. Preparation of calibration curves
Page 36
INST No INE -403-0P0
31
Absorb ance (x)
M oisture value (y)
Calibrationcurveof sample temperature"T3"
Calibrationcurveof
sampletemperature"T2"
Calibrationcurveof sample temperature"T1"
8.1 Summary
When a calibration curve changes at the sample temperature Ts, this function enables to change the calibration curve to other one at this sample temperature. For the input of the sample temperature, a platinum resistance thermometer (Pt100), a radiation thermometer or a temperature/voltage converter with the output of 4 to 20mADC can be used. By utilizing this function, 10 calibration curves in one group (9 curves only in the group 10) are changed with the sample temperature Ts. Since 99 calibration curves can be stored, up to 10 groups can be set.
8.2 Model
This is an optional function. The model number of the detector unit with this optional function is as follows: IR-M2     External input 5: Correction input 4 to 20Ma 8: Correction input Pt100 Communications S: RS-485 R: RS-232C
8.3 Wiring of detector unit
8.4 Setting of dipswitch (Ref to [3.1 Setting of dipswitches])
(1) Turn off the power supply of the detector unit and the setting display unit. (2) Switch the dipswitch 7 (Correction input enable/disable) of the detector unit to ON side. (3) For 4 to 20mA input: Switch the dipswitch 8 (Correction input)
to OFF side
(4 to 20mADC).
For Pt100 input: Switch the dipswitch 8 to ON side (Pt100).
(4) Turn on the power supply of the detector unit and then turn on the power supply of the
setting display unit.
[For Pt100
input]
SA
SB
SG
A
B
B
Correction input Pt100
[For 4 to 20mAinput ]
SA
SB
SG
-
Correction input 4 to 20mA
+
I N
(mA)
8. Correction function by sample temperature from external input
Page 37
INST No INE -403-0P0
32
8.5 Setting of correction enabled/disabled by sample temperature
Execute this setting to either the detector unit or the setting display unit. When the correction is disabled, the operation is normal which means the calibration curve is not changed according to the sample temperature. When it is enabled, the calibration curve is changed with the sample temperature.
8.5.1 Setting at the detector unit
(1) Enter into the operating condition setting mode.
SEL + ENT for 2 seconds (2) Enter into the correction enable/disable setting mode (St). SEL Several times (3) Select the correction enabled or disabled.
ENT (OFF) Correction OFF (disable)
ENT (ON)Correction ON (enable)
* For the setting of sample temperature scaling as the next step, skip (4) and step to (2) of [8.4
Setting of sample temperature scaling]. (4) Return to the measurement mode. SEL for 2 seconds
8.5.2 Setting at the setting display unit
(1) Set the calibration curve number to 0.
CH 0 ENT
(2) Enter into the correction enable/disable setting mode (St).
MODE 1 9 ENT
(3) Select the correction enabled or disabled.
0 ENT (OFF) Correction OFF (disable) 1 ENT (ON) Correction ON (enable)
* For the setting of sample temperature scaling as the next step, skip (4) and step to (2) of [8.4
Setting of sample temperature scaling].
(4) Return the calibration curve number to the original one.
8. Correction function by sample temperature from external input
Page 38
INST No INE -403-0P0
33
8.6 Setting of sam ple temperature scaling (Only 4 to 20mA DC)
In case of using radio thermometer or thermoelectric converter, it is necessary to match the temperature range of their 4 to 20mA DC output and the temperature range of 4~20mA DC input of this instrument. Carry out scaling in either detector unit side or setting display unit side, following the main points given below.
8.6.1 Setting at the detector unit
(1) Enter into the operating condition setting mode.
SEL + ENT for 2 seconds
(2) Enter into the setting mode of sample temperature scaling low limit (tL).
SEL Several times (3) Set the low limit (tL) of sample temperature scaling. (Setting range: -999.9 to 999.9) (4) Enter into the setting mode of sample temperature scaling high limit (tH). SEL Once (5) Set the high limit (tH) of sample temperature scaling. (Setting range: -999.9 to 999.9)
8.6.2 Setting at the setting display unit
(1) Set the calibration curve number to 0.
CH 0 ENT
(2) Enter into the sample temperature scaling setting mode
MODE 2 1 ENT
(3) Set the low limit (tL) of sample temperature scaling. (Setting range: -999.9 to 999.9) (4) Enter into the setting mode of sample temperature scaling high limit (tH). SEL Once (5) Set the high limit (tH) of sample temperature scaling. (Setting range: -999.9 to 999.9)
8.7 Setting of sample temperature
The key operation for this setting is common on the detector unit and the setting display unit.
(1) Enter into the calibration curve setting mode.
SEL for 2seconds
(2) Enter into the sample temperature set ting mode (tS).
SEL several times
(3)Set the sample temperature.Setting range: -999.9 to 999.9
8. Correction function by sample temperature from external input
Page 39
INST No INE -403-0P0
34
8.8 Setting of calibration curve data
(1) Specify a group for calibration curve number to be used.
The groups are from 1 to 10 and each one corresponds to the following calibration curve
numbers respectively. Group 1: Calibration curve number 1 to 10 Group 6: Calibration curve number 51 to 60 Group 2: Calibration curve number 11 to 20 Group 7: Calibration curve number 61 to 70 Group 3: Calibration curve number 21 to 30 Group 8: Calibration curve number 71 to 80 Group 4: Calibration curve number 31 to 40 Group 9: Calibration curve number 81 to 90 Group 5: Calibration curve number 41 to 50 Group 10: Calibration curve number 91 to 99
(2) Example: Select the group 1.
Set the following calibration curve at the sample low limit temperature T1 to the calibration curve No. 1.
Y = a3X3 + a2X2 + a1x + a0
Only the calibra tion curve by polynomial expression can be used. (The calibration curve
by broken line approximation cannot be used.)
Further, set the sample temperature T1. For its setting, refer to [8.7 Setting of sample
temperature]. (3) Set the calibration curves at the sample temperature Ti ( i = 2 to 10) to the calibration
curve numbers 2 to 10.
However, Ti should fulfill the following expression. T1 (Sample low limit temperature) T2  ……….  T9 T10 (Sample high limit
temperature)
(4) Enter 999.9 at the Ti, which is not used. Be careful that, if this value is not entered, a
calibration curve not used is computed as a data.
* The default of the sample temperature Ti is 999.9. (5) For the correction expression coefficient s b1 and b0 and the constants (low limit range,
high limit range, low limit alarm, high limit alarm and preset value), use the values for the first calibration curve number in a group.
Group 1: Calibration curve number 1 Group 6: Calibration curve number 51 Group 2: Calibration curve number 11 Group 7: Calibration curve number 61 Group 3: Calibration curve number 21 Group 8: Calibration curve number 71 Group 4: Calibration curve number 31 Group 9: Calibration curve number 81 Group 5: Calibration curve number 41 Group 10: Calibration curve number 91
8. Correction function by sample temperature from external input
Page 40
INST No INE -403-0P0
35
8.9 Measurement
(1) When the correction by sample temperature ON (enable) is selected and one calibration
curve number in a group is set, the calibration curve number changes automatically corresponding to sample temperature. For example, when the sample temperature Ts fulfills the following expression, the calibration curve number becomes “i”.
T1  …..  Ti Ts < Ti + 1  …..  T10
(2) However, the measured value Y is not the measured value T1 on the calibration curve No.
“i”. The measured value Y becomes the following value interpolated from the measured value Yi and the measur ed value Y
i+1
on the calibration curve No. “i+1”.
(3) When the sample temperature Ts is lower than the sample temperature of the first
calibration curve number in the group or when it is higher than the sample temperature of the last calibration curve number, the following error display appears.
However, the calibration curve, of which sample temperature is set as 999.9, is ignored. * For the sample temperature lower than the sample temperature of the first calibration
curve number in the group Er15 (Sample temperature too low)
* For the sample temperature higher than the sample temperature of the first calibration
curve number in the group Er14 (Sample temperature too high)
8.10 Display of sample temperature Ts
8.10.1 Display at the detector unit
(1) Set the calibration curve number to 0. CH 0 ENT (2) For canceling of the sample temperature display SEL for 2 seconds or CH 2 for seconds
8.10.2 Display at the setting display unit
(1) Set the calibration curve number to 0. CH 0 ENT (2) For displaying the sample temperature display MODE 2 0 ENT (3) For canceling of the sample temperature display SEL for 2 seconds or CH for 2 seconds
9.1 Periodical inspection
The following inspection is required periodically.
(1) Cleaning of tip of fiber optics (for reflective type fiber optics)
Keep the tip of fiber optics clean. If any dirt or dust is noticed, remove it with gauze etc. If the gauze is dipped in alcohol, it
is more effective.
(2) Disassemble cleaning of measuring cell (for transparent type fiber optics)
Regular cleaning of the cover glass in a liquid-contacting part of the measuring cell is required by disassembling the measuring cell.
(3) Installation
Check the detector unit is installed firmly.
(4) Wiring
Check the wirings to the terminals of the detector unit, the setting display unit and other
instruments are connected completely.
(5) Calibration
Regular calibration one in 3 months is required.
8. Correction function by sample temperature from external input
9. Inspection and maintenance
Page 41
INST No INE -403-0P0
36
9.2 Self-diagnosis function
Error message for abnormal operation is displayed by the self-diagnosis function. The same error message is displayed in the detector unit and the setting display unit.
Erro
r
No.
Item Contents Measures
Self -
diagnosed
abnormality
----- High limit over range Measured value is higher than display range.
Check calibration curve
data. Check output b
y checker
plate.
----- Low limit over
range
Measured value is lower than display range.
Check calibration curve
data.
Check output by checker
plate.
Er01 Calibration data error Calibration data is
damaged.
Clear RAM.
Er02 RAM error Detector unit ram data is
damaged.
Clear RAM.
Er03 Motor rotation
abnormal
Motor stops or rotates abnormally.
Exchange motor.
Er04 Lamp power supply
abnormal
Lamp power supply
is
damaged.
Exchange lamp power supply unit.
Er05 A/D abnormal A/D is over range. Check output by checker
plate.
Er06 Computation error Computation is
overflow.
Check output by checker plate.
Er07 Element
temperature
abnormal
Element temperature
is
abnormal.
Need to return detector unit to CHINO.
Er10 Calibration curve
data error
Calibration curve data is not appropriate.
Reset calibration curve data.
Er11 Knee point line
approximation error
Absorbance x is out of range of broken line approximation.
Reset broken line approximation data.
Er12 Ambient
temperature
too high
Ambient tempera
ture of
detector unit is too high
Lower ambient temperature.
Er13 Ambient
temperature
too low
Ambient
temperature of
detector unit is too low
Raise temperature.
Er14 Sample temperature
too high
Sample temperature is higher than correction range
Widen correction range.
Er15 Sample temperature
too low
Sample temperature is lower than correction range
Widen correction range.
Er16 Sample low
reflection
Reflection rate of sample is low or the lamp is damaged.
Close distance to sample. If the lamp is damaged, exchange it.
Er17
Regression operation
error
Regression operation
is
abnormal.
Check calibration curve sample data.
(Note) For the items filled with ¡ in the self-diagnosed abnormality, an alarm output for
self-diagnosed abnormality is activated from the setting display unit.
9. Inspection and maintenance
Page 42
INST No INE -403-0P0
37
9.3 Measures against troubles not included in self-diagnosis
If a trouble occurred, take remedial measures referring to the corresponding items of this instruction manual after checking the following points. For the troub les of the setting display unit, refer to the separate instruction manual the setting display unit.
[1.] No display appears in detector unit.
(1) Is the power supply normal? (2) Is the fuse normal? (3) Are cables connected completely? (4) Are connectors connecting the setting display part and the internal CPU board connected
securely? (S) Turn off the power supply once and turn it on again. (6) If the trouble cannot be repaired by the above check, the power supply unit in the
detector may be defective.
[2.] Measured value remains unchanged.
(I) Is HOLD set to OFF? (2) Is PRESET set to OFF? (3) Are the calibration curve data correct? (4) Is the sample liquid flowing in the measuring cell? (5) Turn off the power supply once and turn it on again. (6) By using the output checker plate, check if the display at the calibration curve number 0
changes.
[3.] Measured value fluctuates.
(1) Is the measuring surface flat? (for reflective type fiber optics) (2) Is the measuring position correct? Make sure that the detector unit does not measure a
bottom face like as a belt conveyor.
(3) Make sure that the disturbance light does not radiate onto the measuring surface. (for
reflective type fiber optics)
(4) Make sure that bubbles do not generate in the sample liquid. (for transparent type fiber
optics)
(5) Is the fiber optics fixed firmly? Make sure to fix it at a place without any vibration.
(6) Is the smoothing time proper? Certain smoothing time is necessary for stable
measurement.
(7) Make sure that any noise source does not exit around the detector unit or connection
cables.
(8) By using the output checker plate, check if the display at the calibration curve number 0
fluctuate. (for reflective type fiber optics)
(9) With the measuring cell emptied, check if the display at the calibration curve number 0
fluctuate. (for transparent type fiber optics)
(10) If the display fluctuates in (8) or (9), clear RAM. If the trouble is not recovered yet,
send back it to CHINO for repair.
(11) Make sure that the fiber optics is not disconnected.
[4.] Measured value is slightly higher than or lower than the actual value.
(1) Is the calibration curve number correct? (2) Are the calibration curve data correct? (3) Are HOLD and PRESET set to OFF? (4) Is the measuring position correct? (for reflective type fiber optics) (5) Is the top of the fiber optics free of dirt and foggy? (for reflective type fiber optics) (6) Is the sample liquid flowing in the measuring cell? (for transparent type fiber optics) (7) Correct the calibration curve. (Refer to [6.8 Correction of calibration curve])
9. Inspection and maintenance
Page 43
INST No INE -403-0P0
38
Protrusion department ofcase
Hexagonal screw plugs
Connector
Hole
Detector(Aview fromside)
Lamp suppoting part
Lampho lder
M4
Hexagonal screw plugs
Dent department
Ring
Source lamp
M3
Hexagonal screw plugs
Lamp plate
9.4 Maintenance parts
9.4.1 List of maintenance parts
The replacing intervals are reference values. They var y according to the working condition.
Article name Quantity Replacing
interval
Remarks
Light source
lamp (With
connector)
1 5 years
Motor (with connector) 1 3 years Fuse 2 Attached as an accessory
Lithium battery (mounted
on a printed circuit board)
1 3 years 3 years when no power is fed
Power supply unit (Note) 1 5 years 3 years when the ambient
temperature is 40°C or higher
(Note): This is built in the detector unit. (not IR-WEP)
9.4.2Replacement of light source lamp
(1) Turn off the power source to the detector
unit.
(2) Open the door of the detector unit and
remove the connector for the lamp.
(3) While holding the lamp supporting part,
loosen the hexagonal screw plugs on the upside of the detector unit with a hexagonal wrench and take out the lamp assembly.
(4) Loosen two M4 hexagonal fixing screws
of the lamp holder and take out the lamp holder.
(5) Loosen two M3 hexagonal fixing screws
of the lamp tube and take out the light source lamp as shown in the figure.
(6) Put a new light source lamp in the lamp
tube and insert a ring. Then, fix the ring with two M3 hexagonal screws.
(7) Mount the lamp holder by placing its slit
to the side of the lamp supporting part and fix it with two M4 hexagonal screws.
(8) By holding the lamp supporting part of the
lamp assembled, place the holes of the lamp holder to the projecting parts of the
detector unit case and then fix it to the detector unit with the hexagonal screw plugs.
(9) Connect the connector for the lamp.
War ning
For replacing maintenance parts, make sure to turn off the power source.
Fix the ring to close to the lamp side.
W a r n ing
9. Inspection and maintenance
Page 44
INST No INE -403-0P0
39
9.4.3 Replacement of motor
(1) Turn off the power source to the
detector unit.
(2) Open the door of detector unit and take
out the setting display part by removing four screws.
(3) The setting display part is connected
with the interior by a cable. Lift the cable with care not to damage it and disconnect the connector for the setting
display part as shown in the figure. (4) Disconnect the connector for the motor. (5) Loosen fixing screw s for the motor and
take out the motor and the motor band
by pulling them forward. (6) Make the motor unit stand as shown in
the figure and hold the sector lightly.
Then separate the sector from the
motor by removing four M2 screws. (7) Make a new motor stand, too, and
attach it to the sector. Be careful that
the sector has front and rear sides. (8) Place the motor unit to make the top of
the motor touch the stopper of the
motor mounting base and wear the
motor band on the motor. (In this case,
place the motor lead wire to the
position shown in the figure.) (9) Fix the motor band with motor fixing
screw. (10) Insert the motor lead wire into the
connector. (11) Return the setting display part to its
origin.
Connector fordisplay
Motor
Motormountingbase
Motor
Connector for
themo tor
Stopper
Motorfixing
screws
Filter
4-M2screws
Motor band
[ Motor unit ]
Caution
Never scratch the sector filter during replacement.
9. Inspection and maintenance
Page 45
INST No INE -403-0P0
40
Connectorfor display
Guide forthe printed
circuit board
Fixing screws ofholderfor
the printed circuit board
Holder for theprinted
circuit board
The printedcircuit
board for battery
A/D Board
9.4.4 Replacement of lithium battery
The detector unit has a built-in lithium battery for memory backup . The setting data are kept, even if electricity is not supplied for 3 years or longer. When the battery has been co nsumed after 3 years or longer without supplying electricity, replace it with a new battery by the following procedure.
(1) Turn off the power source to the
detector unit.
(2) Open the door of detector unit and
take out the setting display part by removing four screws fixing it.
(3) The setting display part is
connected with the interior by a cable. Lift the cable with care not to damage it and disconnect the connector for the setting display part as shown in the figure.
(4) Take out the holder for the printed
circuit board by loosing 2 M2.6
pan head screw. (5) Pull the A/D board forward. (6) Replace the lithium batter with
the mounting board by removing
fixing screws for the mounting
board placed on the A/D board.
(7) Insert the A/D board to the lower
connector along the guide completely.
Return the holder for the printed
circuit board and the door to the original positions.
Make sure polarities of the
battery are correct.
W arn ing
9. Inspection and maintenance
Page 46
INST No INE -403-0P0
41
9.4.5 Replacement of power supply unit
(1) Turn off the power source to the
detector unit. (2) Open the door of detector unit. (3) Disconnect the connection cable
and take out the connection cable from the cable inlet.
(4) Pull out the output board from the
case inside by loosing two M3 pan head screws and by disconnecting the connector (14-pin).
(5) Pull out the power supply unit by
l oosing M3 nut and two M3 pan head screws and by disconnecting two connectors (10-pin, 2-pin).
(6) Connect the connector to a new
power supply unit and assemble it by reversing the removing procedure. Be careful not to step the connection cable.
(7) Connect the connector to the output
board taken out and fix it to the case inside.
(8) Connect the connection cable and return the door to the original position.
9.5 Ram clearing
If the abnormal condition of the detector unit is not recovered, initialization of the RAM is required.
(1) Turn on the power supply after pressing SEL key and key simultaneously.
(2) The contents of ram will be initialized. Set the calibration curve data, etc. again.
Output board fixing screws
(2-M3 pan head screws)
Output board
Output board connector(14-pin)
Power supply fixing screws
(2-M3 pan head screws)
Power supply
connector(10-pin)
M3 nuts and spring washers
Power supply connector(2-pin)
Power supply
Fix it so that this part enters into the dent department of out put board when attaching the out put board.
9. Inspection and maintenance
Page 47
INST No INE -403-0P0
42
10.1 Detector unit
  
IR-M

Measuring system : Infrared reflection type 3-wavelength system Reproducibility : Within detector unit output “x” ±0.003 with output checker plate  (Under same ambient temperature and humidity
The above reproducibility is specified with the output checker plate and it is different from the
reproducibility in actual measurement. Moisture output : 4 to 20mADC Load resistance 500Ù or lower Output accuracy : ±0.5FS Correction input (option) : 4 to 20mADCPt100 3-wire type (For corrections of sample temperature, measuring distance, etc.)
The moisture output is notavailable in the detector unit with correction input.
Communications output : RS-485RS-232C or RS-422A (to be specified) For combination with the setting display unit : RS-485 Output update cycle : 28ms Display : Data LED 5 digits Calibration curve number LED 2 digits Setting : By keys or through communications Calibration curve : First to third degree expression or broken line Calibration curve count : 99 Calibration curve correction: Zero and span correction Calibration curve creation : Regression operation Smoothing : 0 to 99 seconds Shorter that 10 seconds, 0.1 second increment      10 seconds or longer, 1 second increment Calibration : Executed with the output checker plate (with the cell emptied for
transparent type)
Self -diagnosis : Output of calibration data error, RAM error, abnormal motor rotation,
abnormal lamp light source, abnormal communications, ambient temperature high/low, sample low radiation etc. on display or
through communications Working temperature range : 0 to 50ºC Power voltage : 24VDC Allowable voltage fluctuation: 18 to 30VDC Power consumption : Maximum 36VA Allowable fluctuation : Less than 3G Connection : Fiber optics … Connector connection
Power supply, signal line …Terminal connection
Case : Steel plate Dust and water protection structure Conforming to
IEC529 IP65
Mounting : Wall-mounting M8 bolts 4 Weight : About 4kg
Accessories : Power supply unit IR-WEP 1piece Fuse 2 pieces
10. Specifications
Page 48
INST No INE -403-0P0
43
10.2 Fiber optics
10.2.1 Reflective type fiber optics
  
IR-WCRN
Measuring diameter and distance :
ö20/15mm to ö50/50mm
Fiber length : Standard 1.5m
(Maximum 10m) Minimum bending radius: R100mm Working temperature range :
0 to 150ºC Fiber protection : Stainless hose Purged airflow : 5 to 20Nl/min Accessory : Horizontal mounting
holder, flange holder
10.2.2 Output checker plate for reflective type fiber optics IR-WCRNB
It is used for checking output of the detector unit on site. It is used by mounting it on the top of the fiber optics.
10.2.3 Reflective type fiber
optics with lens IR -WCRE
Measuring diameter and distance :
ö20/25mm to ö40/100mm
Fiber length : Standard 1.5m
Maximum 10m
Minimum bending radius: R100mm Working temperature range:
0 to 150ºC Fiber protection: Stainless hose Air purge function: Not provided Accessory : Horizontal mounting
holder, flange holder
10.2.4 Output checker plate for reflective type fiber optics with lens IR-WCREB
It is used for checking output of the detector unit on site. It is used by mounting it on the top of fiber optics.
10. Specifications
ºC
The tip of reflection
type fiber optics
Ø1 6
32
Pu rge airinlet 1:Max60
Esideco nnector
30 040
Len gth isas de signa ted. Ma x1500 mm)
Ø10
Ø8
Dsideco nnecto r
Purgeair inlet2 (R c1/8).
Unit: mm
(Outerdiameterof applicableØ6nylontubeetc.)
Ø
40
Fitting sc rew for output checker pl ate
The tip of reflection type fiber optics
Output checker plate
Guide pin
47 .5
(34 )
Unit:
mm
82.5
40 300
The tip ofreflection type fiber
optics w ith lens attachment
Ø8
Ø10
Ø30
Dsideco nnector
Esideconn ector
Un it:
mm
Len gth isas de signa ted. Ma x1500 mm)
Output checker plate
Fitting screw for output checker plate
The tip of reflection type fiber
optics with lens attachment
Un it:
mm
Guide pin
Ø 40
41
(38)
Page 49
INST No INE -403-0P0
44
10.2.5 Measuring cell for liquid
IR-WCC1
Pipe diameter : PT1/4 Material : SUS316 Temperature of measuring liquid:
0 to100ºC
Pressure of measuring liquid : Maximum 3.9Mpa40kg/cm2)
10.2.6 Transparent type fiber
optics IR -WCT
Fiber length : Standard 2m
Maximum 20m
Ambient temperature: 0 to 150ºC Fiber protection Stainless hose Minimum bending radius:R250mm
10.3 Accessories
10.3.1 Power supply unit
(conforming to CE) IR-WEP
Output voltage : 24VDC Output current : 2.1A Working temperature range:
-10 to 50ºC
Power supply: 100 to 240V/AC 47 to 450Hz Allowable voltage fluctuation : +10% to (-) 15% of rated value Power consumption :
Maximum about 160VA
Case: Horizontal mounting holder,
flange holder
Mounting : Wall mounting type
(DIN rail
WeightAbout 380g
10.3.2 Connection cable
IR-WERT
It is the exclusive cable used for the connection between the detector unit and the setting display unit/power supply unit.
Structure 4-core cabtyre cable (With double shield) External diameter: ö10mm
Length :Maximum 200m Connection:Both end tips
10. Specifications
OUT
I N
Liquid outlet(R 1/4)
Liquid inlet
(R1/4)
2-M5 Mounting s crew hole
dipth 5 (both sides)
30
142
6 0
Un it:
mm
Flexible tube(SU S)
Connector
Nu ts
Len gth isas designa ted. Ma x2000 mm)
Ø9
Ø10
Unit:
mm
4- Ø
4.5
90
75
80
±0.2
60
7.5
45
3 5.3
591
±0.2
M3.5 Terminal screws
Unit:
mm
To detector
30
11 0
20
200
Len gth isasdesignated. Max200m)
Ø1 0
To power supply and
setting display unIt
UNit:mm
Page 50
32-8, KUMANO-CHO, ITABASHI-KU, TOKYO 173-8632
Telephone: +81-3-3956-2171 Facsimile: +81-3-3956 -0915 E-mail: inter@chino.co.jp
Web site http://www.chino.co.jp/
INE- 403--0P0 Sep -'01 IR-M2□□□Series INFRARED MOISTURE METER DESIGNED AS A FIBER OPTIC(IR-M21□□, IR-M22□□)Printed in Japan ( Kuki )
Loading...