Chino IM Series, IRMA51 Series, IRMA52 Series Instructions Manual

Page 1
INST.No.INE-475-P1CE Software Version 1.00
                                
Mirror type multiple-constituents meter/ Detector Unit
Model: IRMA51
IRMA52
Store this manual in a readily accessible location for future reference.
This manual should be provided to the end user.
Page 2
INST.No.INE-475-P1CE Software Version 1.00
-Front1-
¡ Regarding this manual
(1) This manual should be provided to the end user. (2) Read this manual carefully to gain a thorough understanding of how to operate this product
before starting operation.
(3) This manual describes the functions and maintenance of this product. Store it in a readily
accessible location for future reference.
(4) If you have unclear points or need technical assistance, please contact your sales agent of
CHINO Corporation.
Notices
1. The contents of this manual are subject to change without notice.
2. Every effort has been made to ensure that the details of this manual are accurate. However, should any errors be found or importance information be omitted, please contact your nearest agent of CHINO Corporation.
3. Under absolutely no circumstances may the contents of this manual, in part or in whole, be transcribed or copied without permission.
Preface
Thank you for purchasing the IM series infrared multiple-constituents analyzer [Mirror type multiple-constituents meter/detector unit IRMA51, IRMA52]. Read this manual carefully to ensure that you use this product correctly and safely.
Page 3
INST.No.INE-475-P1CE Software Version 1.00
-Front2-
Be sure to read the following safety instructions before attempting to install, operate or store this product.
1. Working conditions and environment
1) This product is installed using the mounting screw holes (four M8 screw holes) on its top surface. Refer to [3. Installation] and fix the detector unit firmly.
2) Do not place this product on a desk and the like. This product may fall, causing serious personal injury and serious damage to this product.
3) Do not use this product in a location where volatile, corrosive or flammable gas is
present, or in a location where moisture, chemical or seawater is splashed.
2. Symbols used in this manual
The following symbols are used depending on important degrees of warnings/cautions to operate this product safely and to avoid malfunctions or unexpected situation.
Important
degree
Symbols Contents
1
This symbol is indicated with a title for an explanation with .
2
Indicates important information that must be observed to avoid the risk of fire or electric shock or other dangers that may result in serious personal injury or death, or damage to this product.
3
Indicates important information that must be observed to avoid the risk of personal injury or malfunctions of this product.
4
Indicates supplementary information that the operator is recommended to understand.
5
Indicates supplementary information or a reference to an operation.
Remarks
Reference
Caution
Warning
Warning
Important Operational Instructions
Page 4
INST.No.INE-475-P1CE Software Version 1.00
-Front3-
Please observe the following safety precautions fully to use this product correctly. In addition, please read this manual carefully and store it in a readily accessible location for future reference.
The mark indicates actions that are prohibited during an operation.
Warning (May cause death or serious personal injury)
Do not operate this product in a location where flammable or volatile gas is present.
The use of this product in such environment may result in the risk of explosion.
If this product emits any smoke, abnormal noise or unusual odor, immediately turn off power.
Do not continue to use this product as it may result in the risk of fire or electric shock. Turn off the power source to this product and contact your nearest sales agent of CHINO Corporation. Do not repair this product yourself as it may be dangerous.
Never touch the internal parts of this product.
If you touch the inside of this product, it may result in personal injury by electric circuits or a moving part Even when you open the case of this product for maintenance works, never touch the internal parts.
Do not attempt to repair or modify this product.
Repairing or modification must be carried out by service personnel authorized by CHINO Corporation. If you repair or modify this product yourself, it may not fulfill this product’s functions, or it may result in the risk of electric shock or damage to this product.
Do not touch, with wet hands, the power terminals or plugs of the power cord.
If you touch, with wet hands, the power terminals or plugs of the power cord, it may result in the risk of electric shock.
If you notice something suspicious, immediately turn off power.
If the temperature of the power cord or other cable is high abnormally, turn off the power source to this product immediately and contact your nearest sales agent of CHINO Corporation.
Confirm the grounding.
To avoid the risk of electric shock, before supplying the power to this product, make sure that the power terminals and the protective ground terminal have been wired correctly and reliably.
Caution (May cause personal injury or property damage
Confirm the specifications (Model code) of this product.
Before using this product, check that the model code shown in the serial number plate matches your order.
Separate precautionary safety measures should be provided before using this product.
When this product is used in equipment that needs security measures, provide separate safety measures with the equipment before using this product, responding to phenomena resulting from erroneous operations or malfunctions of this product or malfunctions of related instruments.
Use crimp type lugs for termination of wires.
To avoid dropping-out or short-circuit of wires, use crimp type lugs for their termination.
Confirm the power to this product.
Before supplying the power to this product, make sure that the power voltage matches the rated voltage to this product and that the power terminals and the protective ground terminal have been wired correctly and reliably.
Do not continue to use this product remaining malfunctioned.
Do not continue to use this product remaining malfunctioned as it may result in the risk of fire or electric shock. Contact your nearest sales agent of CHINO Corporation for repairing.
Comply with the descriptions in this manual.
Comply with the descriptions and handling instructions in this manual to ensure that you use this product correctly and safely. Failure to comply with the descriptions and instructions may result in damage to or functional decline of this product or damage to the equipment.
!
Warnings and Cautions
Page 5
INST.No.INE-475-P1CE Software Version 1.00
-C1-
Contents
Preface Important operational instructions W arnings and cauti ons
1. Introduction ·························································· 1
1.1 General··························································1
1.2 Configuration················································1
2. Models and accessories ········································2
2.1 Models··························································2
2.1.1 Detector unit ·················································2
2.1.2 Accessories··················································· 2
2.2 Attachments··················································2
3. Installation····························································· 3
3.1 Setting dipswitches·······································3
3.1.1 Connecting this detector unit to
the operator interface/display unit ········ 3
3.1.2 Connecting this detector unit to
a personal computer······························3
3.2 Mounting the air-purge hood IR-WEA·········4
3.2.1 Mounting the hood on this detector unit·······4
3.2.2 Adjusting the hood length····························· 4
3.2.3 Purge air························································4
3.3 Cooling this detector unit
(water-cooling or air-cooling)··············· 4
3.4 Installing this detector unit ···························5
3.5 Cautions before installation ························6
3.5.1 Vibration or shock·········································6
3.5.2 Magnetic disturbances··································6
3.5.3 Working temperature and humidity··············6
3.5.4 Sight path······················································6
3.5.5 Installation (for objects of
powders or granular substances)··············6
3.5.6 Installation (for objects in sheet
suc h a s pa p e r s , etc . )································6
4. Connections··························································· 7
4.1 Laying cables················································7
4.2 Single-detector-unit connection···················· 7
4.3 Multiple-detector-units connection···············8
4.3.1 Relaying boxes used ·····································8
4.3.2 Relaying boxes not used·······························9
5. Names and functions·········································· 10
5.1 Names and functions of
detector unit components···················· 10
5.2 Names and functions of data display parts· 11
5.3 Terminal portion ········································· 11
6. Operation ·····························································12
6.1 Confirmation before operation ··················· 12
6.2 Setting detector unit numbers·····················12
6.3 Setting calibration curve data····················· 13
6.3.1 Data for setting ··········································· 13
6.3.1-1) List of setting data··································· 13
6.3.1-2) Function of setting data··························· 14
6.3.2 Procedure for setting··································· 15
6.3.3 Correcting calibration curves ····················· 15
6.4 Setting operating conditions······················· 16
6.4.1 Data for setting ··········································· 16
6.4.1-1 Description of setting data························18
6.4.2 Procedure for setting operating conditions· 20
6.4.3 Setting data in the engineering mode ········· 20
6.5 Setting calibration curve number of each constituent number····················· 21
6.5.1 Setting constituent number························· 21
6.5.2 Setting calibration curve number················ 21
6.6 Calibration·················································· 22
6.6.1 Absorbance display mode settings ············· 22
6.6.2 Calibration·················································· 22
6.7 Setting output limit values of
calibration curves································ 23
6.7.1 Data for setting limit values of
calibration curves································ 23
6.7.2 Procedure for setting data for output limit
values of calibration curves················24
6.7.2-1 Setting the calibration curve numbers (CH)··24
6.7.2-2 Enabling/disabling the output
limit processing (LM)························· 24
6.7.2-3 Setting a low limit absorbance value (XL)
and a low limit measured value (YL)·············24
6.7.2-4 Setting a high limit absorbance value (XH)
and a high limit measured value (YH)···········24
6.7.2-5 Setting output limit values to
oth e r cal i b r ati o n c u r v e nu m b ers·············24
Page 6
INST.No.INE-475-P1CE Software Version 1.00
-C2-
Contents
7. Procedure for creating calibration curves·······25
7.1 Sample preparation: For constituent of moisture etc. ··········25
7.1.1 Powders or granular substances··················25
7.1.2 Papers (sheet materials)······························25
7.2 Sample measurement: For constituent of moisture etc. ··········26
7.2.1 Sample measurement for powders or granular substances ··········26
7.2.2 Sample measurement for papers
(sheet materials)································27
7.3 Sample preparation: For constituent of
thickness or coating-thickness ··········28
7.4 Sample measurement: For constituent of thickness or coating-thickness··········· 29
8. Sample temperature correction function ······· 30
8.1 General························································30
8.2 Wiring this detector unit·····························30
8.3 Enabling/disabling the sample temperature correction function······························31
8.3.1 Setting with this detector unit·····················31
8.3.2 Setting with the operator interface/ display unit ·········································· 31
8.4 Setting a sample temperature scaling ·········31
8.4.1 Setting with the detector unit······················31
8.4.2 Setting with the operator interface/ display unit ·········································31
8.5 Setting sample temperature·························32
8.6 Setting calibration curve data ·····················32
8.7 Measurement ··············································33
8.8 Displaying sample temperature Ts··············33
8.8.1 Displaying with this detector unit···············33
8.8.2 Displaying with the operator interface/
display unit········································· 33
9. Inspection and maintenance······························34
9.1 Periodical inspection···································34
9.2 Self-diagnosis function·······························34
9.3 Measures against troubles not included with self-diagnosis······························35
9.3.1 No display on this detector unit appears·····35
9.3.2 Measured value remains unchanged···········35
9.3.3 Measured value fluctuates ··························35
9.3.4 Measured value is slightly higher or lower than the actual value·················35
9.4 Maintenance parts······································· 36
9.4.1 List of maintenance parts···························· 36
9.4.2 Replacing a lamp········································ 36
9.4.3 Replacing a motor······································· 37
9.4.4 Replacing a power board···························· 38
9.4.5 Replacing a fuse ········································· 39
9.5 RAM clearing············································· 39
9.5.1 RAM clearing············································· 39
10. Specifications ·····················································40
10.1 Detector unit: IRMA51, IRMA52 ······ 40
10.2 Attachments················································41
10.2.1 Power unit IR-WEP··································· 41
10.2.2 Connection cable IR-WERP······················41
10.2.3 Air purge hood IR-WEA···························· 41
10.2.4 Outpu t c he cke r pl ate IR-WEB (For IRMA51) IR-WEB3 (For IRMA52)········· 41
10.2.5 Installation adaptor IR-WED1··················· 41
10.2.6 Air-cooling box IR-WEX ·························· 42
10.2.7 Relaying box IR-WEE···························· 42
11.Outside dimensions ············································43
11.1 Detector unit
IRMA51, IRMA52········ 43
11.2 Attachments················································ 44
11.2.1 Power unit IR-WEP··································· 44
11.2.2 Connection cable IR-WERT······················ 44
11.2.3 Air purge hood IR-WEA···························· 45
11.2.4 Output checker plate
IR-WEB (For IRMA51) IR-WEB3 (For IRMA52)······· 45
11.2.5 Installation adaptor IR-WED1···················46
11.2.6 Air-cooling box IR-WEX··························· 47
11.2.7 Relaying box IR-WEE·····························48
11.2.7-1) IR-WEE1:Steel······································48
11.2.7-2) IR-WEE2:SUS304································ 48
Page 7
INST.No.INE-475-P1CE Software Version 1.00
-1-
1.1 General
The IM series infrared multiple-constituents analyzer [Mirror type multiple-constituents meter/detector unit IRMA51/IRMA52] is a reflective type non-contact on-line multiple-constituents meter using the principles of infrared absorption by constituent value of moisture, thickness or coating-thickness. This multiple-constituents meter is configured with a detector unit, an operator interface/display unit and accessories. The detector unit is provided with 99 built-in calibration curves and outputs a linear analog signal proportional to constituent value of moisture, thickness or coating-thickness of a sampled material. As the stand alone detector unit provides function keys, a data display and a digital signal for communication, it can be operated on its own or with a personal computer. The operator interface/display unit can communicate with up to nine (9) detector units, and displays the constituent value of moisture, thickness or coating-thickness digitally as well as accessing to various parameters of the detector unit. Accessories including cables, connecting the detector unit to the operator interface/display unit or to the power unit, the air-purge hood for shielding all extraneous light and for air-purging to keep the measuring window and the sight path clean, and the output checker plate for your own calibration are available. This manual describes about the detector unit and accessories. The instruction manual is separately available for the operator interface/display unit.
1.2 Configuration
IM
SERIES
1. Introduction
Detector unit IRMA51, 52
Communication output Analog output
Analog output
24V DC
RS-485
Connection cable IR-WERP
Power unit IR-WEP (Supplied as an attachment)
Operator interface /display unit IRGMEG2
Air-purge hood IR-WEA
Output checker plate IR-WEB
Page 8
INST.No.INE-475-P1CE Software Version 1.00
-2-
2.1 Models
2.1.1 Detector unit
IRMA51
Model
51: Multi-constituents (NIR) 52: Multi-constituents (Thin-film, Infrared)
Number of wavelength
: 2 to 0 (Designate 2 to 10)
Number of constituent
: 1 to 4 (Designate 1 to 4)
Communication interface *1
S: RS-485 (standard) L: LAN (option)
(*1) To be specified
In case of connecting to IRGMEG2 (IM series Infrared multiple-constituent analyzer operator interface/Display unit) it is RS-485 communications.
Special specifications
Blank: Standard 1: For small diameter 2: Dust laying 3: Special gain
4: P polarized light
2.1.2 Accessories
Name Model
Cable IR-WERP
Air-purge hood IR-WEA
Output checker plate IR-WEB (For IRMA51)
IR-WEB3 (For IRMA52) Installation adapter IR-WED1 Air-cooling box IR-WEX Relaying box IR-WEE
2.2 Attachments
Name Quantity Remarks
Power unit 1 set IR-WEP (Omron S82K-05024)
Fuse 2 pieces
10A Ø5X20mm
(Littelfuse 217010) Flathead screwdriver 1 piece For connection to the terminal board Hexagon wrench (3mm width across flats)
1 piece
For removing/attaching the terminal cover or the
data display cover Hexagon wrench (4mm width across flats)
1 piece For removing/attaching the air-purge hood
Metal cable gland 2 pieces MS-SC13.5 (LAPP) Ferrite core 2 pieces MSF10KEX (Morimiya) Instruction manual 1 copy This document Instruction manual for communications
1 copy
Separate manual “IM series IRMA/IR-GMEG2
Communications”
2. Models and accessories
EN61326+A1+A2 Emission: class A Immunity: Annex A Table A.1
[Conformance to CE-marking]
1. Use the exclusive power supply IR-WEP
2. The connection cable length is up
to 30m in indoors.
Page 9
INST.No.INE-475-P1CE Software Version 1.00
-3-
3.1 Setting dipswitches
3.1.1 Connecting this detector unit to the operator interface/display unit
It is not required to change the settings of the internal dipswitches of this detector unit.
3.1.2 Connecting this detector unit to a personal computer
Setting the dipswitch 2 in this detector unit is required. (1) Loosen four pieces of the M4 hexagon socket cap screws fixing the data display cover. (2) The data display cover is connected with internal components by cables. Remove the data
display cover carefully not to damage to the cables.
(3) Set No.3 of the dipswitch 2 to ON (upper side).
Warning
(*) Do not change the setting of the dipswitch 1.
Set the internal dipswitches before settingthe detector unit.
The above has been set as the default settings at shipment.
Make sure to turn off the power source to this unit before removing the data display cover.
Remarks
Reference
Caution
3. Installation (Refer to [5. Names and functions].)
IM
SERIES
ON
OFF
ON
OFF
1234
1234 1234 1234
For connecting the detector to the operator
interface/display unit (IRGMEG2)
(default settings at shipment)
Dipswitch 2
Dipswitch 1
For connecting the detector to a
personal computer
Dipswitch 2 Dipswitch 1(*)
Dipswitch 2 No.3
Remove these four bolts.
Data display cover
Page 10
INST.No.INE-475-P1CE Software Version 1.00
-4-
IM
SERIES
3.2 Mounting the air-purge hood IR-WEA
The air-purge hood is used to shield against the extraneous light, including electric lamps, flames and direct sunlight, onto the measuring area. In addition, it is used to clear water droplets, dust, smoke and steam from the sight path between this detector unit and the measuring area. The hood length is adjustable.
3.2.1 Mounting the hood on this detector unit
Position 4 pieces of hexagon cap screws of the air-purge hood with the screw holes of the air-purge fixing flange of this detector unit and then tighten the screws with the hexagon wrench (4mm width across flats) to fix the hood.
3.2.2 Adjusting the hood length
Loosen the hood length adjustment ring by turning counterclockwise and adjust the hood length. After the length is adjusted, turn the ring clockwise until tight to fix the hood.
3.2.3 Purge air
When the air-purging is required, supply dry air, not containing oil, dust, etc., from a purge air inlet. Use a nylon tube of φ6mm in outer diameter and φ4mm in inner diameter for connecting to the purge air inlet. The air-purge inlet is PL6-01M (Nihon Pisco). Purge air flow: 5 to 200Nℓ/min Purge air pressure: 200kPa (2kgf/cm2) or lower
3.3 Cooling this detector unit (water-cooling or air-cooling)
The inlet and outlet for cooling are provided on the side of this detector unit. Let the purified cooling water flow through the inlet and outlet. The recommended water volume is about 0.5 to 1Nl/min and the maximum pressure is 200kPa (2kgf/cm2) or lower. In the environment where the cooling water can not be used, this detector unit can be cooled by feeding cooling air through the inlet and outlet. When the cooling air is shared with the purge air, connect the air outlet of this detector unit and the purge air inlet of the air-purge hood with a tube.
The recommended air flow in the case is about 50 to 200Nl/min.
3. Installation (Refer to [5. Names and functions].)
IM
SERIES
Detector unit
Ø86
Ø100
31
Max. 300 ± 3 to Min. 222 ± 3
Purge air inlet (Nihon Pisco PC8-02) Connection: Outer diameter Ø8 Inner diameter Ø6
By a nylon tube, etc.
Hood length adjustment ring
Four M5 hexagon cap screws (For fixing the air-purge hood)
Detector unit
Air inlet
Air piping (Nylon tube)
Inlet/outlet for cooling (For cooling the detector unit: R1/8)
Air-purge hood
Caution
It is recommended that this detector unit is usually cooled with water or air to ensure its reliability. When this detector unit is used in ambient environments higher that 45ºC, be sure to cool it down with water or air. In addition, monitor constantly if the supply of the cooling water or cooling air is adequate.
Page 11
INST.No.INE-475-P1CE Software Version 1.00
-5-
IM
SERIES
(Ø25)
IM
SERIES
測定距離(L)
< Standard type >< Small diameter type>
3.4 Installing this detector unit
Four pieces of the mounting screw holes (M8, 11mm in depth) are provided on the top surface of this detector unit. Fix this detector unit to a mounting plate or alternatives with four M8 bolts. The
external views and mounting dimensions are shown in the following figure.
(1) For the IRMA51 and IRMA52, place this detector unit with the distance of 300mm
from the measuring surface. (This detector unit can be placed from 200 to 400mm.)
(2) The target spot can be confirmed by the visible light irradiated from this detector unit to the
measuring surface.
(3) Prepare a mounting plate in accordance with the conditions at an installation place. Decide the
dimensions of the mounting plate in consideration of the space for wiring, etc.
Relation of distance and target area
3. Installation (Refer to [5. Names and functions].)
Cable gland
(25)
133
52
260
130
120
10
120
65
Inlet/outlet for cooling
2-R1/8
Mounting screw hole 4-M8, depth 11
8
70
120
17
Distance L (mm)
Target area (mm)
Distance
L (mm)
Target area (mm)
200
30
200
30
300
50
300
30
400
70
400
40

Distance
(
L
)
Unit: mm
Page 12
INST.No.INE-475-P1CE Software Version 1.00
-6-
3.5 Cautions before installation
3.5.1 Vibration or shock
Use this detector unit in a location with minimal vibration or shock. If it is forced to use this detector unit in a location where the mounting plate is subject to vibration or is not stable, separate the mounting plate from the vibration source or take measures against vibration by inserting a vibration proof rubber sheet between this detector unit and the mounting plate or other measures.
3.5.2 Magnetic disturbances
This detector unit is designed to withstand magnetic disturbances. However, install it as far away from induction heating oscillators or power lines as possible.
3.5.3 Working temperature and humidity
The working temperature of 10 to 30°C is optimal for this detector unit. When this detector unit is used in ambient environments higher than 45°C, be sure to cool it down by water or air. When the ambient temperature of the detector unit is lower than 10
o
C, prevent condensation from forming on the cover glass.
The low working humidity is preferable. In the high humid environment, condensation may form on the cover glass by water steam and result in measuring errors.
3.5.4 Sight path
Use this detector unit in a location where water droplets, dust, smoke and steam are not contained in the sight path between this detector unit and the measuring area. If these cannot be prevented from being contained and their effects are not eliminated, mount the air-purge hood on this detector unit or use effective alternatives to provide a clear sight path to the measuring area. The dc light like the natural light onto the measuring area will not cause any disturbances in measurements, but incorrect measurements are produced by the ac light including electric lamps and flames or direct sunlight. In
such cases, shield against the disturbance light to prevent it from irradiating directly onto the measuring area.
3.5.5 Installation (for objects of powders or granular substances)
When this detector unit is installed for on-line measurements, select such place as a hopper outlet or a place just after the change of conveyor belts where there is little difference between the surface and the interior of measuring objects. If such a place cannot be selected, measure the objects by facing the interior to the surface by using a baffle plate as shown below.
3.5.6 Installation (for objects in sheet such as papers, etc. )
Tilt this detector unit by 15° from the vertical surface as shown in the figure.
 
3. Installation (Refer to [5. Names and functions].)
I
M
S
E
R
I
E
S
1
5
°
[For objects of powder or granular substances]
[For objects in sheet such as paper, etc.]
300mm
Paper surface
IM
SERIES
Measuring object
Conveyor belt
Baffle plate
300mm
Page 13
INST.No.INE-471-P1CE Software Version 1.00
-7-
4.1 Laying cables
For laying cables, be careful of the following points. (1) Separate cables from induction heating oscillators or power lines. (2) Keep cables not to get water- or oil-stained. (3) Don’t bend cables extremely or apply any
excessive force to them.
(4) For a permanent layout, protect cables with
conduits, etc.
4.2 Single-detector-unit connection
The internal terminal board is accessible by removing the terminal board cover of the detector unit. Lead in cables through the upper cable glands, and connect them to respective terminals. Lead in the exclusive cable IR-WERP (φ10mm outer diameter) through the cable gland 1.
Lead in a signal cable through the cable grand 2.
Warning
Make sure to use the signal cable with the outer diameter of φ7.0 to 12.5mm.
For avoiding the risk of electric shock, make sure to turn off the power source to this detector unit before wiring to the power terminals.
Use cables suiting to the applicable outer diameter of the cable gland; otherwise waterproofing of the detector unit may deteriorate. Tighten the cable glands with the nuts and fix the terminal board cover securely after wiring.
Reference
Caution
P+ P-
OUT+ OUT-
SA SB SG DI
COM
DO
COM
IN+ IN-
24VDC
4-20mADC RS -485
DIGITALIN
DIGITALOUT
4- 20mADC
P+ P-
OUT+ OUT -
SA SB SGDI
COM
DO
COM
IN+IN-
24VDC
4-2 0mAD C
RS-485
DIGITAL IN DIGI TALOUT
4-20mADC
-V
+V
OUTPUT
DC 24V2. 1A
DC ON DC LOW
V.ADJ
INPUT50/60Hz AC100-240V 1.3A
Operator interface/display unit, personal computer or sequencer communication out
put
Recorde
r
, etc.
Analog output: 4-20mA DC
Cable glands
• Outer diameter of applicable cables
(ø7.0 to ø12.5mm)
Cable (IR-WERP)
Communication output :RS-485
Power unit (IR-WEP)
P-
(Blue)
Power supply 100-120V AC 200-240V AC
Protective grounding
Contact input Contact output Correction input
P+
(Red)
G
(Green)
G (Green)
P+
(
Red
)
P- (Blue)
S
A
(Black)
SB (White)
SG (Brown)
4. Connections
For conformance to CE (1) Change 2 pieces of the resin
cable glands (standard) to the supplied metal cable glands.
(2) Attach the supplied ferrite
cores to cables. You are required to attach one ferrite core to one cable at a place as close to the cable gland.
Caution
Page 14
INST.No.INE-471-P1CE Software Version 1.00
-8-
4.3 Multiple-detector-units connection
By using the operator interface/display unit, you can connect the detector unit up to 9 sets.
Two connection systems are available. One is for utilizing the relaying box (IR-WEE) and the other is without using it
4.3.1 Relaying boxes used
For either connecting method, before connecting to the operator interface/display unit or before turning on the power to the operator interface/display unit, set the detector unit numbers not to overlap them each other. Refer to [6.2 Setting detector unit numbers] for the details.
4. Connections
Caution
For wiring to the relaying box, refer to the separate manual for the relaying box.
Caution
SERIES
IM
SERIES
IM
SERIES
IM
Cable IR-WERP *
Detector unit IRMA
Relaying box IR-WEE1 or IR-WEE2 (The power unit IR-WEP is built in.
)
2-conductor twisted cable (with shield)
Operator interface/display unit
IRGMEG2
Line converter
Personal compute
r
100-240V AC
2-conductor twisted cable (with shield)
2-conductor twisted cable (with shield)
Cable IR-WERP *
Cable IR-WERP *
24V DC (Note)
Remarks
(Note): 24V DC at CE marking (In case of using IRGMEG2V)
(*): The connection cable should be indoors and its length should be up to 30m.
Page 15
INST.No.INE-471-P1CE Software Version 1.00
-9-
4.3.2 Relaying boxes not used
4. Connections
SERIES
IM
SERIES
IM
-V +V
OUTPUT
DC24V 2.1A
DCON
DCLOW
V.ADJ
INPUT50/60Hz AC100-240V1.3A
-V +V
OUTPUT
DC24V 2.1A
DCON DCLOW
V.ADJ
INPUT50/60Hz AC100-240V1.3A
SERIES
IM
-V +V
OUTPUT
DC24V 2.1A
DCON
DCLOW
V.ADJ
INPUT50/60Hz AC100-240V1.3A
100-240V AC
Connect the power wires of P+ (red), P- (blue), and G (green external shield) of the exclusive cable IR-WERP to the power unit. Connect the signal wires of SA (black), SB (white) and SG (brown internal shield) to the 2-conductor twisted cable (with shield
)
.
Connect the signal wires of SA (black), SB (white) and SG (brown internal shield) of the exclusive cable I R-WERP and the wires of SA, SB and SG of the 2-conductor twisted cable (with shield) to respective terminals of this detector unit. (For the signal wire of the exclusive cable, take off the rod shaped tip on its top, strip the cover and then insert them to the terminal together with the twisted cable.
Power unit IR-WEP
Power unit IR-WEP
Power unit IR-WEP
Line converte
r
Detector unit IRMA
Cable IR-WERP *
2-conductor twisted cable (with shield)
Cable IR-WERP *
Cable IR-WERP *
Operator interface/display unit IRGMEG2
Personal compute
r
2-conductor twisted cable (with shield)
2-conductor twisted cable (with shield)
24V DC (Note)
Remarks
(Note): 24V DC at CE marking (In case of using IRGMEG2V)
(*): The connection cable should be indoors and its length should be up to 30m.
Page 16
INST.No.INE-471-P1CE Software Version 1.00
-10-
5.1 Names and functions of detector unit components
Names Functions
(1) Detector unit mounting
screw holes
Four M8 screw holes (11mm in depth) for fixing this detector unit to a mounting plate, etc.
(2) Data display
Displays a calibration curve number, a measured value, etc.
(3) Measuring part
The optical assembly, the detector element and the electric circuits are built in.
(4) Terminal board cover
It is fixed with four M4 hexagon socket cap scre ws and can be removed for writing. The terminal board and the fuse are built in. The serial number plate showing the model number is pas ted on the lower side.
(5) Date display cover
It is fixed with four M4 hexagon socket cap screws. The lamp, the motor and the power board, which are consumables, are built in. For the replacement of these parts, refer to [9. Inspection and maintenance 9.4.2-9.4.4].
(6) Air-purge hood
mounting flange
This flange with four M5 screw holes is for mounting the air-purge hood.
(7) Measurement window
It is for irradiating and receiving light.
(8) Cable gland 1
It is for leading in a cable. The outer diameter of its applicable cable is φ8.5 to
10.5mm. Ordinary this cable gland is sealed by a rubber. Take off the rubber for leading in a cable. The exclusive cable IR-WERP is lead in through this cable gland.
(9) Cable gland 2
It is for leading in a signal cable. The outer diameter of its applicable cable is φ8.5 to 10.5mm. Ordinary this cable gland is sealed by a rubber. Take off the rubber for leading in a cable.
(10) Fuse
The fuse (10A in capacity) is fitted on the fuse holder.
(11) Terminal board
This internal terminal board for wiring of connection cables is accessible by opening this terminal board cover.
(12) Serial number plate
Two serial number plates indicating the model number and the serial number of this detector unit are pasted on the inside and outside of this detector unit.
(13) Inlet/outlet for cooling
These inlet and outlet are for water or air used to cool this detector unit. The nominal diameter of the inlet and outlet is R1/8.
5
. Names and Functions
IM
SERIES
IRMA51615 MA03YA001
IRMA5161S MA03YA001
P+ P -
OUT+ OUT-
SA SBSG DI
COMDOCOM IN+ IN-
24VDC
4-20 mADC
RS-485
DIGITAL IN
DIG IT ALOUT 4 -2 0mA DC
CH
DATA
SEL
CH/
ENT
MADEIN JAPAN
(4) Terminal board cover
(1) Detector unit mounting screw hole
(2) Data display
(3) Measuring part
(5) Data display cover
(6) Air-purge hood mounting flange
(7) Measuring window
(12) Serial No. plate
(outside)
(8) Cable gland 1
(9) Cable gland 2
(10) Fuse
(11) Terminal board
(12) Serial No. plate
(
inside
)
(13) Inlet/outlet for cooling
Page 17
INST.No.INE-471-P1CE Software Version 1.00
-11-
5.2 Names and functions of data display parts
Names Functions
(1) CH display
Displays a calibration curve number in the measurement mode or a setting data name in the setting mode.
(2 ) Data display
Displays a measured value in the measurement mode or a setting data in the setting mode.
(3 ) Function keys
SEL For entering into the setting mode or switching setting screens CH/ F or setting a calibration curve number in the measurement mode or
entering a decimal point in the setting mode; When the key is pressed
again, the decimal point disappears. For entering a setting data in the setting mode or changing the digit of a setting data For changing a setting data in the setting mode
For changing a setting data in the setting mode
ENT For storing a setting data in the setting mode
5.3 Terminal portion
The mounting direction of the cable gland is selectable in the following two kinds. Make sure to use the blind cap to cover the place not mounting the cable gland. (1) Vertical direction of a cable: Default direction at shipment (2) Horizontal direction of a cable: It is effective when there is no space above this detector unit.
5
. Names and Functions
IM
SERIES
CH
DATA
SEL
CH/
ENT
MADE IN JAPAN
(1) CH display (2) Data display
(3) Function keys
Cable gland
Blind cap
Cable gland
Blind cap
(1) Cable – Vertical direction
(2) Cable – Horizontal direction
Page 18
INST.No.INE-475-P1CE Software Version 1.00
-12-
6.1 Confirmation before operation
Check the following two points before turning on the power.
1) Have the dipswitches been set? (Refer to [3.1 Setting dipswitches].)
2) Are all of connections correct? (Refer to [4. Connections])
6.2 Setting detector unit number
Detector unit numbers are for identifying each detector unit in the multiple-detector connections. Up to nine (9) detector units can be connected. Set the detector unit numbers before connecting to the operator interface/display unit.
1) Enter into the operation condition setting mode “ t ”by pressing SEL key and ENT key simultaneously for 2 seconds.
2) Press SEL key 4 times to display “Hn” in the CH display.
3) Press key. A numeric blinks for entering a detector unit
number.
4) Change the numeric by key, key or key.
Press key to increase the numeric. Press key to decrease the numeric.
Press key to shift the blinking digit.
5) Press ENT key. The numeric stops blinking and the detector unit number is stored.
6) Press SEL key for 2 seconds. The mode returns to the measurement mode.
CH
DATA
t0. 0
CH
DATA
Hn 1
Blinkin
g
Detector unit number setting mode
CH
DATA
Hn 4
Detector unit number
The settings shown in the next paragraph can also be executed via the operator interface/display unit (sold separately). Refer to the se
p
arate manual for the operator interface/display unit.
6. Operation
The default number of this detector unit at shipment is “1”. When this detector unit is used with a number other than 1, set the detector unit number b
y
following procedure.
Reference
Reference
Caution
[Analog output in case of using Model“IRGMEG2”]
1. The analogue output of IRGMEG2 is 2 outputs of OUT1, OUT2. (However the analogue output of IRGMEG2V is 1 output of OUT1 only.)
2. By reference to [3.5 Analog output and alarm output] (in separate manual of "IM series Infrared multiple-constituent analyzer operator interface/Display unit”), set the detector unit number that you want to output the analog output.
Page 19
INST.No.INE-475-P1CE Software Version 1.00
-13-
6.3 Setting calibration curve data
The output characteristics of this detector unit depend upon measuring objects. It may also change according to the process conditions or the measuring conditions of samples. Accordingly, for accurate moisture measurements, it is required to perform beforehand sample tests of each object and obtain a relationship (This is called as the calibration curve.) between the moisture constituent (%H2O) obtained by a drying method or other measuring methods and the absorbance “x” measured by this detector unit. Also, for accurate thickness and coating- thickness measurements, it is required to perform beforehand sample tests of each object and obtain a relationship (This is called as the calibration curve.) between the thickness and coating-thickness constituent (µm) obtained by a micro meter or other measuring methods and the absorbance “x” measured by this detector unit.
6.3.1 Data for setting
6.3.1-1) List of setting data
The followings are the setting data for calibration curve.
Setting data name Display Setting range of data
Default value
Remarks
(1) Calibration curve No. CH 1 to 99 1 (2) Calculation mode Md 1 to 4 1
a0
A0 0 to ±9999.9 0.0000 a1 A1 0 to ±9999.9 1.0000 a2
A2 0 to ±9999.9 0.0000 a3 A3 0 to ±9999.9 0.0000 a4 A4 0 to ±9999.9 0.0000 a5 A5 0 to ±9999.9 0.0000 a6 A6 0 to ±9999.9 0.0000 a7 A7 0 to ±9999.9 0.0000 a8 A8 0 to ±9999.9 0.0000 a9 A9 0 to ±9999.9 0.0000
(3) Polynomial coefficient
a10 AA 0 to ±9999.9 0.0000
Effective numeric
: 5 digits
b0 b0 0 to ±9999.9 0.0000
b1
b1 0 to ±9999.9 1.0000
(4)
Correction expression coefficient
b2
b2 0 to ±9999.9 0.0000
Effective numeric
: 5 digits
Lo 0 to 9999.9 0.00 (5) Low/ High limit constituent
output
Hi 0 to 9999.9 100.00
Effective numeric
: 5 digits
AL 0 to 9999.9 0.00 (6) Low/ High limit constituent
alarm
AH 0 to 9999.9 100.00
Effective numeric
: 5 digits
(7) Preset constituents PS 0 to 9999.9 0.00 Effective numeric
: 5 digits
(8) Water absorbance Ab 0 to ±9999.9 0.00 Effective numeric
: 5 digits
The settings shown below can also be executedvia this detector unit or the operator interface/display unit (sold separately).
Remarks
This paragraph describes the setting method of calibration curve data, assuming that calibration curves have been already created. When the calibration curves have not been created yet, refer to [7. Procedure for creating calibration curves].
Reference
Reference
The absorbance “x” is the ratio of the absorbed amount of the light having the wavelength S
of 1.43, 1.94 or 3µm, which is absorbedin a measuring object, to that of the light having
the reference wavelength R not absorbed.
6. Operation
Page 20
INST.No.INE-475-P1CE Software Version 1.00
-14-
6.3.1-2) Functions of setting data
Data name Function
(1) Calibration curve
No.
An optional number from 1 to 99 can be designated as the calibration curve. Ordinary It is designated sequentially from “1”.
(2) Calculation mode Select the calculation mode the calibration curve The calculation mode and
wavelength number differs by the models of IRMA.
Calculation mode 1 to 3: The processing becomes the 2-wavelength processing/3-wavelength processing (R1, S, R2)
on the each calculation mode.
Calculation mode 4: Multiple regression calculation that used each wavelength. The wavelength number differs by the models of IRMA.
Wavelength λ1 λ2 λ3 λ4 λ5 λ6 λ7 λ8 λ9 λ10 Calculation group Group 1 Group 2 Group 3
Calculation mode 1 R1 S R2 ― ― ― ― Calculation mode 2 R1 S R2 ― ― ―
Calculation mode 3 R1 S R2
Calculation mode 4 λ1 λ2 λ3 λ4 λ5 λ6 λ7 λ8 λ9 λ10
(3) Polynomial
coefficient “a0 to a10”
Polynomial coefficient is calculated by selected calculation mode.
Calculation mode 1 to 3: Assume that the calibration curve is represented by the third-degree or lower-degree polynomial of
y = a3x3 + a2x2 + a1x + a0, where, “y” is the constituents and “x” is the absorbance (“CH0” data of this detector unit).
Calculation mode 4: Calibration curve is calculated by multiple regression calculation using each wavelength.
y = a0 + a1LOG(λ1) + a
2
LOG(λ2) + ···· + a
10
LOG(λ10)
The wavelength number differs by the models of IRMA.
(4) Correction
expression
coefficient “b
0
to b2”
These coefficients are provided for the quadratic expression correction (zero and span correction) with reference to the constituents “y”. Assuming that the constituents after the correction be “Y”, we obtain “Y = b2y2+b1y + bo”.
(5)
Low/ High limit
constituents output
These specify constituents at a low limit output and a high limit output on the analog output scaling. The output is scaled to 4mA at the low limit constituents and 20mA at the high limit constituents.
(6)
Low/ High limit
constituents alarm
The low limit alarm is activated when the measured value is lower than the low limit alarm set-point of constituents and the high limit alarm is activated when the measured value is higher than the high limit alarm set-point of constituents. Use the operator interface/display unit when the alarm outputs are required.
(7) Preset constituents By turning on the preset output by key operations or through communications, the
constituents being set as a preset value is displayed and outputted, regardless of the measured value.
(8) Water absorbance Decide the water absorbance to compute the surface water ratio for measuring the
moisture content of sands, etc. [This setting is only enable when the surface water ratio computation shown in (11) of “6.4.1 Setting data” is set to ON.] The relation between the measured moisture content (%: mass basis moisture content), the surface water ratio (%) and the water absorbance (%) is;
y Surface water ratio = (Moisture content – Water absorbance) / (1 + (Water
absorbance / 100)
6. Operation
Page 21
INST.No.INE-475-P1CE Software Version 1.00
-15-
6.3.2 Procedure for setting
1) Press SEL key for 2 seconds to enter into the calibration curve setting mode “CH”.
2) Press key. A numeric blinks for entering a calibration
curve number.
3) Change the numeric by key, key or key.
Press key to increase the numeric.
Press key to decrease the numeric.
Press key to shift the blinking digit.
4) Press ENT key. The numeric stops blinking and the
calibration curve number is stored.
5) Press SEL key to enter into the setting mode “A0” for
polynomial coefficient a
0
.
6) Enter into a numeric setting mode by pressing key.
Change the numeric by key or key and store it by pressing ENT key. For entering a decimal point, press
CH/ key at the blinking digit.
7) Press SEL key to enter into the next data setting mode.
8) Repeat the above procedure from 2) to 7) to set all data.
9) After the above setting procedure is completed, return to the measurement mode by pressing
SEL key for two seconds. The automatic return is made if no key is pressed for one minute.
6.3.3 Correcting calibration curves
If the actual constituents does not correspond with the constituents on the calibration curve being preset due to the difference of one-line and off-line measurements, or other causes, correct it with the correction coefficients b0, b1 and b2 in [6.3 Setting calibration curve data]. These coefficients are provided for the quadratic expression correction against the measured value “y”. Assuming that the measured value after correction be “Y”, we obtain,
Y = b2y2 + b1y + bo
Usually, the shift correction is done with b0 only. (b1 = 1, b2 = 0)
CH
DATA
C H 0
Setting data name
CH
DATA
A
0 0. 0 0 0 0
Setting data
Blinking
Calibration curve number entering mode
CH
DATA
C H 0
CH
DATA
CH
Calibration curve number setting
The settings shown in the next paragraph can also be executed via the operator interface/display unit (sold separately).
[Example 1: Moisture measuring] When the measured moisture content before correction is 15 (%H2O) and the actual moisture content is 13 (%H2O), “-2%H2O should be set.
(b2 = 0, b1 = 1, b0 = -2)
6. Operation
Remarks
Remarks
[Example 2: Thickness and coating-thickness measuring] When the measured thickness or coating-thickness content before correction is
25.0 (
µm) and the actual thickness and coating-thickness is 23.5(µm ), “-1.5 (µm)
should be set.
(b
2
= 0, b1 = 1, b0 = -1.5)
Remarks
Page 22
INST.No.INE-475-P1CE Software Version 1.00
-16-
6.4 Setting operating conditions
Set the operating conditions including time constant and hold function.
6.4.1 Data for setting
Set the following data in the operating condition setting mode.
Setting data name Display Setting range
of data
Default value
Remarks
Constituent 1 t1 Constituent 2 t2 Constituent 3 t3
(1) Time constant
Constituent 4 t4
0.0 to 99.9 0.2 Unit: Second
(2) Hold ON/OFF Hd ON, OFF OFF (3)
Preset output ON/OFF PS ON, OFF OFF
(4) Constituent number dC 0 to 4 1
0: Circulation display
1: Number of constituents Constituent 1 d1 Constituent 2 d2 Constituent 3 d3
(5)
Number of decimal places
Constituent 4 d4
0 to 4 1
(6)
Constituents number/CH number display selection
dK ON, OFF OFF
OFF: Constituents number
display
ON :CH number display
(7) Constituents number of
analog output/Contact output
oC
1 to 4 1
(8) Contact output: Selection of
contact output enabled/disabled and output item
do
nonE ALArM Error
nonE
NonE: No output
ALArM: High/low alarm
Error: Self-diagnosed error
(9) Contact input: Selection of
contact input enabled/disabled and input item
di none
HoLd PrSEt SMt.rL
nonE NonE: No output
HoLd: Hold
PrSEt: Preset
SMt.rL: Smoothing/real
(10) Communication protocol Pt rtU
Priv*
rtU rtU: MODBUS RTU mode
Priv:*
(11) Detector number Hn 1 to 9
1 to 247 1 to 31*
1 Operator interface/display unit
MODBUS RTU mode
Communication speed SP 9600, 19200 9600 Parity P nonE
odd EVEn
EVEn nonE: No parity
odd: Odd parity
EVEn: Even parity
Data length d 7, 8* 8 Stop bit length Sb 1, 2* 1
(12)
BCC enabled/disabled bC ON, OFF* OFF
These parameters are for the
private protocol “Priv” in the
above [(7) Communication
protocol] and are not used with
this detector unit.
* : Not using in this detector
6. Operation
The following paragraph can also be executed from “IRGMEG2” (sold separately).
Remarks
Page 23
INST.No.INE-475-P1CE Software Version 1.00
-17-
Setting data name Display Setting range
of data
Default value
Remarks
Sample temperature correction ON/OFF
St ON, OFF OFF
The following settings can be executed at "ON".
Correction input scaling L tL -999 to 999 0
(13)
Correction input scaling H tH -999 to 999 100
(14) Surface water ratio
computation ON/OFF
Sd ON, OFF OFF
Engineering mode ENG
The following settings can be executed by pressing [ENT] key.
Processing mode 1:
Weight α1
1A 0 to 1.000 0.500 0.5: 3-wavelength processing
0 or 1: 2-wavelength processing
Processing mode 1:
Calibration constant k1
1K 0 to ±9.9999 1.0000
Processing mode 2:
Weight α2
2A 0 to 1.000 0.500 0.5: 3-wavelength processing
0 or 1: 2-wavelength processing
Processing mode 2:
Calibration constant k2
2K 0 to ±9.9999 1.0000
Processing mode3:
Weight α3
3A 0 to 1.000 0.500 0.5: 3-wavelength processing
0 or 1: 2-wavelength processing
(15)
Processing mode 3:
Calibration constant k3
3K 0 to ±9.9999 1.0000
6. Operation
Page 24
INST.No.INE-475-P1CE Software Version 1.00
-18-
6.4.1-1) Descriptions of setting data
Setting data name Description
(1) Time constant If the measured value fluctuates noticeably, the signal can be set to
be dull by delaying the response of this detector unit. The smoothing time (equivalent to the time constant of analog instruments) can be set in 0.1-second increment.
(2) Hold ON/OFF By setting it to ON, the measured value is held.
At the same time, the analog output is also held.
(3) Preset output
ON/OFF
By turning on the preset output, this detector unit displays the preset value and the analog output becomes the value corresponding to the preset value.
(4) Constituents number
For the setting the number of constituents displayed. It skips in the case of constituents 1. The setting of constituent number 1 to 4 is possible and
displays it in
circulation with 0. Defalut value is 1.
(5)
Number of decimal places
For the setting the number of decimal places of displayed data
(6)
Constituents number/CH number display selection
Select constituent number/CH number display.
Set it to ON for
changing the CH display from
constituent number to calibration curve number in measuring mode. Defalut value is OFF.(constituent number
display)
(7) Constituents number of
analog output/Contact output
The setting of constituent number of analog output and contact output is possible. It skips in the case of constituents 1. The setting of constituent number 1 to 4 is possible. Defalut value is 1.
(8) Contact output : Selection
of contact output enabled/disabled and output item
Select whether the contact output is enabled or disabled, and select the output item.
(1) NonE :Disabled (2) ALArM :Outputs the high/low alarm, by the contact, of the
calibration curve number being set
(3) Error :It is outputted when an abnormality is detected by the
self-diagnosis function shown in [9.2 Self-diagnosis function].
(9) Contact input : Selection
of contact input enabled/disabled and output item
Select whether the contact input is enabled or disabled, and select the output item. (1) NonE: Disabled (2) HoLd: The analog signal is held by the latest value.
(Holding by the signal H)
(3) PrSEt: The analog signal is set to a fixed value of 4-20mA.
(Fixing by the signal H)
(4) SMt.rL: Switching of Smoothing or Real (Real by signal H)
(10) Communication protocol rtU: MODBUS RTU mode
Priv: Exclusive mode - *This is not used with this detector unit.
(11) Detector number This is a number for identifying each detector in the
multiple-detector-units connection. This setting has been completed in [6.2 Setting detector unit numbers].
6. Operation
Page 25
INST.No.INE-475-P1CE Software Version 1.00
-19-
Setting data name Description
(12) Communications speed,
Parity, Data length, Stop bit length, BCC enabled /disabled
When this detector unit is connected with the operator interface/display unit, skip these settings. These are for communications with a personal computer, a sequencer or other similar devices. Set them corresponding to a master unit.
• Communications speed: Select 9600 or 19200bps.
• Parity: Select No, Even or Odd.
• Data length: The data length of the MODBUS RTU mode is fixed to 8 bits. No need to select.
• Stop bit length: The stop bit length of the MODBUS RTU mode
is automatically set with the setting of the parity.
(EVEN or odd parity: 1 bit, No parity: 2 bits)
• BCC enabled/disabled: *This is not used with this detector unit.
(13) Sample temperature
correction ON/OFF
When calibration curves change by sample temperatures, this function enables to change the calibration curve to the other one by measuring the sample temperature. As the input of the sample temperature, a radiation thermometer or a temperature/voltage converter with the output of 4 to 20mADC can be used. When the sample temperature correction is set to ON, the following
settings are required. Correction input scaling L, Correction input
scaling H
These are used for scaling of 4 to 20mA correction input.
(14) Surface water ratio
computationON/OFF
For the computation of the moisture content of sands, etc., select the surface water coefficient computation is enabled or disabled. Set it to ON for enabling.
(15) Engineering mode Set the “weight α1” and “calibration constant” to processing mode 1
to processing mode 3.
(1) Weight α1: The 3-wavelength processing or 2-wavelength
processing can be switched by changing of the
setting of the weight α1. When α1 is set to 0.5, the processing becomes the 3-wavelength processing that is the ratio processing of three (3) wavelengths. When α1 is set to 0 or 1, the processing becomes the 2-wavelength processing. (α1: 1 - the ratio processing of λ1 and λ2, α1: 0 - the ratio processing of λ2 and λ3)
(2) Calibration constant K1: This is the value obtained in [6.5
Calibration] and has been stored automatically.
6. Operation
Reference
·Please replace α1 to α2 and k1 to k2 in processing mode2.
·Please replace α1 to α3 and k1 to k3 in
p
rocessing
mode3.
Refer to [6.3.1-2) Functions of setting data
Page 26
INST.No.INE-475-P1CE Software Version 1.00
-20-
6.4.2 Procedure for setting operation conditions
1) Press SEL key and ENT key simultaneously for 2 seconds to enter into the operation condition setting mode.
2) Press key. A numeric blinks for entering a setting data.
3) Change the numeric by key, key or key. Press key to increase the numeric. Press key to decrease the numeric. Press key to shift the blinking digit. For entering the decimal point, press CH/. key at the
blinking digit.
4) Press ENT key. The numeric blinking stops and the data is stored.
5) Press SEL key to enter into the setting mode of the next setting data. Repeat the above procedures from 2) to 4) to set all setting data except the engineering mode.
6.4.3 Setting data in the engineering mode
1) For the setting of the [Weight α1] and the [Calibration constant k1] in the engineering mode, press SEL key several times to enter into the engineering mode “Eng” shown right.
2) Press ENT key in this engineering mode to enter into the
setting mode “1A” of the weight α1.
3) Change the numeric by key, key or key. Press key to increase the numeric. Press key to decrease the numeric. Press key to shift the blinking digit.
For entering the decimal point, press CH/. key at the
blinking digit.
4) Press ENT key to store the weight α1.
5) Then press SEL key to enter into the setting mode “1k” of the calibration constant K1.
6) Change the numeric by key, key or key. Press key to increase the numeric. Press key to decrease the numeric. Press key to shift the blinking digit. For entering the decimal point, press CH/. key at the
blinking digit.
7) For the setting of the [Weight α2] and the [Calibration constant k2] press SEL key several times to
enter into the engineering mode “Eng” shown right. Press ENT key in this engineering mode to enter into the setting mode “2A” of the weight α2. Repeat the above procedure from 3) to 7) to set [Weight α2] and the [Calibration constant k2].
8) For the setting of the [Weight α3] and the [Calibration constant k3] press SEL key several times to
enter into the engineering mode “Eng” shown right. Press ENT key in this engineering mode to enter into the setting mode “3A” of the weight α3. Repeat the above procedure from 3) to 7) to set [Weight α3] and the [Calibration constant k3].
Setting data name
Blinking
t 0.0
CH
DATA
t 0. 0
CH
DATA
t 0. 0
CH
DATA
H d O N
Blinking
Next setting data name
CH
DATA
E n g
Engineering mode
Press ENT key.
6. Operation
CH
DATA
1 A 0. 5 0 0
Weight α1 setting mode
Press ENT key.
Press SEL key.
CH
DATA
1 K 1. 0 0 0
Calibration constant K1 setting mode
Page 27
INST.No.INE-475-P1CE Software Version 1.00
-21-
6.5 Setting calibration curve number of each constituent number
Setting data name
Display
Setting range of data
Default value
Remarks
Constituent number
C0 0 to 4 1
0: Absorbance display mode 1 to 4: Constituent number setting mode
Calibration curve number
CH 1 to 99 1
Correspondence to 1 to 4 each constituent number
6.5.1 Setting constituent number
Set the calibration curve number of constituent number 1 to 4 in accordance with the measurement object.
1) Press CH/. key to enter into the constituent number "C0"
setting mode.
2) Press key to enter a setting data.
3) Change the numeric by key, key or key.
Press key to increase the numeric.
Press key to decrease the numeric.
Press key to shift the blinking digit.
4) Press ENT key. The numeric blinking stops and the data is displayed.
6.5.2 Setting calibration curve number
1) Press SEL key simultaneously for 2 seconds to enter into the calibration curve number "CH" setting mode.
2) Press key to enter a setting data.
3) Change the numeric by key, key or key.
Press key to increase the numeric.
Press key to decrease the numeric.
Press key to shift the blinking digit.
4) Press ENT key. The numeric blinking stops and the calibration curve number is displayed.
5) To set the calibration curve number of constituent number 1 to 4, repeat the above procedure from 1) to 4) to set all data.
6) Press SEL key for 2 seconds. The mode returns to the measurement mode.
6. Operation
CH
DATA
CO 1
CH
DATA
CO 1
Blinking
Constituent number setting mode
CH
DATA
CO 2
The constituent number that is able to set it differs by the models of IRMA. Confirm the specification of your IRMA series.
Remarks
CH
DATA
Calibration curve number setting
Calibration curve number of constituent number
CH 3
CH
DATA
Blinking
Calibration curve number setting mode
CH 1
Caution
[Analog output in case of using Model“IRGMEG2
”]
1. The analogue output of IRGMEG2 is 2 outputs of OUT1, OUT2. (However the analogue output of IRGMEG2V is 1 output of OUT1 only.)
2. By reference to [3.5 Analog output and alarm output] (in separate manual of "IM series Infrared multiple-constituent analyzer operator interface/Display unit”), set the constituent number that you want to output the analog output.
Page 28
INST.No.INE-475-P1CE Software Version 1.00
-22-
6.6Calibration : Continuously set up [6.6.1 Absorbance display mode settings] and [6.6.2
Calibration] For using this detector unit accurately, the periodical calibration (once in three months) is recommended. The calibration is executed by using the output checker plate.
Mount the output checker plate to the top of the air-purge hood and measure the No. 1 side of the output checker plate.
6.6.1 Absorbance display mode settings
With the following procedure, enter into the absorbance display mode by setting the calibration curve number to “0”.
1) Press CH/. key to enter into the constituent number "C0"
setting mode.
2) Press key to enter a numeric.
3) Change the numeric to “0” by using key.
4) Press ENT key. The numeric stops blinking and
constituent number setting mode "C0" is displayed.
5) Press SEL key for 2 seconds and display the absorbance
display "M1" setting mode.
6.6.2 Calibration
1) Execute the calibration by pressing three keys of SEL , , and ENT simultaneously.
When the displayed absorbance “x” is within 0.0000 ±
0.0050 for IRMA5100 and within 0.0000 ± 0.0100 for IRMA5200, the calibration is completed.
2) By pressing CH/. key, display the absorbance display constituent number setting mode "C0". Set the constituent number in original number (1 to 4).
3) Press SEL key for 2 seconds and cancel the absorbance display mode. The measured value of
calibration curve set to the constituent value is displayed.
Supply the power to this detector unit for one hour or longer before calibration.
6. Operation
Remarks
IM
SERIES
Detector unit
Air-purge hood
Output checker plate
Glass side (No. 1 side) of output checker plate
Output checker plate (Insert it until its top is touched to the protrusion of the holder part.)
Set the glass side (No. 1 side) to moisture measurements and insert it into the holder part of the air-purge hood.
CH
DATA
M1 0.0000
Display of absorbance “x”
Absorbance display setting mode
CH
DATA
M1 0.0021
CH
DATA
CO 1
Constituent number setting mode
Constituent number blinkin
g
CH
DATA
CO 0
Absorbance display mode
Page 29
INST.No.INE-475-P1CE Software Version 1.00
-23-
6.7 Setting output limit values of calibration curves
The measured value can be fixed by low limit and high limit absorbance values.
When the absorbance becomes less than the low limit absorbance value, the measured value is fixed to YL as shown right. Also, when the absorbance becomes more than the high limit absorbance value, the measured value is fixed to YH.
6.7.1 Data for setting limit values of calibration curves
Set the following data for the calibration curve output limit values.
Setting data name Display Setting range
of data
Default value
Remarks
(1) Calibration curve number CH 1 to 99 1 (2) Output limit processing
enabled/disabled
LM OFF, ON OFF 0: OFF, 1: ON
Low limit absorbance value XL 0 to ±9999.9 0.0000 Effective numeric: 5 digits
(3)
Low limit measured value YL 0 to ±9999.9 0.0000 Effective numeric: 5 digits High limit absorbance value XH 0 to ±9999.9 9.9999 Effective numeric: 5 digits
(4)
High limit measured value YH 0 to ±9999.9 9999.9 Effective numeric: 5 digits
Setting data name Description
(1) Calibration curve number Any calibration curve number in 1 to 99 can be
specified, but it is normally specified from 1 sequentially.
(2) Output limit processing enabled/
disabled
Set the output limit processing enabled or disabled. When ON (enabled) is set, the measured value is outputted corresponding to the limit values being set. When OFF (disabled) is set, the measured value is outputted corresponding to the calibration curve data.
(3) Low limit absorbance value XL
Low limit measured value YL
Enter data to fix the measured value at the low limit side.
(4) High limit absorbance value XH
High limit measured value YH
Enter data to fix the measured value at the high limit side.
6. Operation
YH
YL
XL
XH
Measured
value
Absorbance
0
Page 30
INST.No.INE-475-P1CE Software Version 1.00
-24-
6.7.2 Procedure for setti ng data for output limit valu es of calibration curves
6.7.2-1 Setting the calibration curve numbers (CH)
1) Press SEL key and CH/. key simultaneously for 2 seconds to display "CH" in the CH display for entering the calibration curve number setting mode for the calibration curve output limit values.
2) Set a calibration curve number.
 ENT
6.7.2-2 Enabling/disabling the output limit processing (LM)
1) Press SEL key to enter the output limit processing enabled/disabled "LM" setting mode.
Press key to enter a setting data. Change the setting data by key, key, or key.
Press ENT key to store the setting data.
6.7.2-3 Set t i n g a l o w l i m i t a b s o r ban c e v a l u e (XL) and a low limit measured value (YL)
1) After the above setting, press SEL key to enter the low limit absorbance value "XL" setting mode. Press key to enter a setting data.
Change the setting data by key, key, or key.
Press ENT key to store the setting data.
2) After the above setting, press SEL key to enter the low limit measured value "YL" setting mode. Press key to enter a setting data. Change the setting data by key, key, or key.
Press ENT key to store the setting data.
6.7.2-4 Set t i n g a hi gh l i m i t a b s o r b a n c e v a l u e (XH) and a high limit measured value (YH)
1) After the above setting, press SEL key to enter the high limit absorbance value "XH" setting mode.
Press key to enter a setting data.
Change the setting data by key, key, or key.
Press ENT key to store the setting data.
2) After the above setting, press SEL key to enter the high limit measured value "YH" setting mode.
Press key to enter a setting data. Change the setting data by key, key, or key.
Press ENT key to store the setting data.
6.7.2-5 Setting output limit values to other calibration curve numbers
1) After the above setting (6.6.2-1 to 6.6.2-4), press SEL key to return to the calibration curve number setting mode.
Enter other calibration curve number for setting output limit values. Repeat the above procedure (6.6.2-1 to 6.6.2-4) to set output limit values to other calibration
curve numbers.
CH
DATA
X
H 1. 000
CH
DATA
YL 0.0 
CH
DATA
Y H 5 0. 0
CH
DATA
LM O F F
Setting data
Setting data name
CH
DATA
C H 1
Calibration curve number setting for calibration curve output limit value
Calibration curve number
CH
DATA
X
L 0.0 
6. Operation
After the above setting procedure is completed, return to the measurement mode by pressing SEL key for two seconds. The automatic return is made if no key is
p
ressed for one minute.
Remarks
Page 31
INST.No.INE-475-P1CE Software Version 1.00
-25-
7.1 Sample preparation: For constituent of moisture etc.
The output characteristics of the infrared multi-constituents meter depend upon measuring objects. It may also change according to the process conditions or the measuring conditions of samples. Accordingly, for accurate measurements, it is required to perform beforehand sample tests of each object and obtain a relationship (This is called as the calibration curve.) between the moisture content (%H2O) obtained by a drying method or other measuring methods and the absorbance “x” measured by this detector unit. This chapter describes the measurements of the moisture content by the drying method. However,
the same creation method of the calibration curves is applied to the measurements of the moisture content by the Karl Fischer method or other methods.
7.1.1 Powders or granular substances
1) Take a sample of about 2 to 4 liters from the measuring object.
2) Dry up the sample up to the absolute dry condition by a dryer. Particularly be careful with the heating temperature so as not to denature the sample.
3) Divide the dried sample into every 100 to 200cc. It is recommended to divide it into many samples, but 5 to 6 samples are ordinary used.
4) Add water to these 5 to 6 samples bit by bit so that the moisture content of each sample divides the measuring range almost evenly.
5) Stir each sample with water added sufficiently, put it into a polyethylene bag and then seal it
tightly. Leave each sample for about 2 days until its moisture content is stabilized.
7.1.2 Papers (sheet materials)
1) Cut a sample to have a size of about 100 x 100mm. (Number of samples: 9 sheets)
2) Divide 9 sample sheets into 3 groups with 3 sheets in each group. Adjust the moisture content of each group under the following conditions.
Drying: Leave the samples in a desiccator containing silica gel for 2 days. In case of
papers, the moisture content becomes about 4%H2O.
In the air: Leave the samples in an indoor atmosphere.
In case of papers, the moisture content becomes about 7%H2O in summer or about 5%H2O in winter.
Moistening: Leave the samples in a desiccator containing salt water for 2 days.
In case of papers, the moisture content becomes 8 to 9%H2O normally or maximum about 11% H2O.
1
2
3
N
1
2
3
4
5
6
With silicagel
With salt water
For a sample cannot be measured because it cannot be heated or it is hardened by adding water, adjust the moisture contents according to the procedure for papers shown in Para. [7.1.2]
Wear clean gloves not to touch the samples by hands directly. It is convenient for you to mark the samples for identifying their front/back surfaces and their measurement sequence.
7. Procedure for creating calibration curves
Caution
Caution
Page 32
INST.No.INE-475-P1CE Software Version 1.00
-26-
7.2 Sample measurement: For constituent of moisture etc.
Be careful with the following cautions in measurements.
1) Before measurements, warm up this detector unit more than 1 hour.
2) Place samples at the same position as in actual measurements, set the calibration curve number to 0, and then read the display (absorbance "x").
3) Perform the measurements rapidly.
4) The visible light is irradiated from this detector unit. Confirm that the samples cover the measurement area completely.
7.2.1 Sample measurement for powders or granular substances
1) Prepare the same quantity of trays (Ø100 to 150mm in diameter, about 20mm in depth) as
samples.
2) Weigh each tray (weight W
0
).
3) Empty the sample sealed from the polyethylene bag and then spread it on the tray until its surface becomes flat and its bottom cannot be seen.
4) Put the tray at the measuring position, set the calibration curve number of this detector unit to 0, and then read its display (absorbance "x"). If the measured value cannot be read easily, delay the response with the smoothing function.
5) Stir the sample rapidly with a spoon and measure it by this detector unit again. Repeat this measurement 2 or 3 times per sample.
6) Weigh the tray with the sample put on (weight W
1
).
7) Perform the same measurements on all samples.
8) Put all samples with trays into a dryer, and dry them until they become in the absolute dry condition (Drying for more than 2 hours at 105 to 110ºC is usually required.).
9) Cool them down after drying and weigh each tray with the sample put on (weight W2).
10) Record the measured values in the following table every measurement and obtain the moisture contents by the following formula.
X 100 (%H2O)
No. Absorbance "x"
W
0
W1 W2
Moisture content (%H
2
O)
1
x1
y1
2
x2
y2
3
:
x3
:
y3
:
n
xn
yn
7. Procedure for creating calibration curves
W1 ─ W2 W
1
W0
Page 33
INST.No.INE-475-P1CE Software Version 1.00
-27-
7.2.2 Sample measurement for papers (sheet materials)
1) Weigh a sample before measurements (weight W1).
2) Hold the sample by a paper holder, put it by tilting at 15º to the measuring position, set the calibration curve number of this detector unit to 0, and then read its display (absorbance "x").
3) Weigh the sample after the above measurement (weight W2). Perform the measurements from (1)
to (3) as quickly as possible.
4) Perform the same measurements on all samples. For these measurements, the same surface (front
or back) of the samples should be measured and the same measurement sequence should be taken.
5) Put all samples into a dryer, and dry them until they become in the absolute dry condition (Drying
for more than 2 hours at 105ºC is usually required.).
6) After drying, put all samples into a desiccator containing silica gel, and allow them natural-
cooling.
7) Weigh each sample (weight W
3
).
8) Record the measured value in the following table every measurement, and obtain the moisture
contents by the following formula.
X 100 (%H2O)
No. Absorbance "x"
W1 W2 W3
Moisture content (%H2O)
1
x1
y1
2
x2
y2
3
:
x3
:
y3
:
n
xn
yn
7. Procedure for creating calibration curves
(W1 + W2)/2 ─W
3
(W1 + W2)/2
Page 34
INST.No.INE-475-P1CE Software Version 1.00
-28-
7.3 Sample preparation: For constituent of thickness or coating-thickness
The output characteristics of the infrared thickness (coating-thickness) meter depend upon measuring objects. It may also change according to the process conditions or the measuring conditions of samples. Also, for accurate thickness and coating- thickness measurements, it is required to perform beforehand sample tests of each object and obtain a relationship (This is called as the calibration curve.) between the thickness and coating-thickness constituent (µm) obtained by a micro meter or other measuring methods and the absorbance “x” measured by this detector unit.
1) 3 to 5 pieces standard is required for measuring the thickness and coating- thickness.
3 pieces standard is the 3 samples that thickness and coating- thickness differs.
(Example) The thinnest thing, thickest thing or the intermediate thing.
2) 3 to 5 pieces of100 mm x 100 mm size is required for 1 standard.
3) The struggling of thickness and coating- thickness sample is required demand accuracy of 1/3 or less
In case that thickness and coating- thickness is 100µm and demand accuracy is ±3µm, the thickness and coating- thickness of sample is required within 99 to 101µm.
7. Procedure for creating calibration curves
Caution
Wear clean gloves on the occasion of handling, so as not to touch the sample directly. If checking the front or rear side and flow directions of the sample, it is convenient in measurement.
Page 35
INST.No.INE-475-P1CE Software Version 1.00
-29-
7.4 Sample measurement: For constituent of thickness or coating-thickness
Please measure and pay attention below.
1) Place samples at the same position as in actual measurements, set the calibration curve number to
0, and then read the display (absorbance "x").
2) The visible light is irradiated from this detector unit. Confirm that the samples cover the
measurement area completely.
        
3) Put the sample tilt by 15° from the vertical surface and read the absorbance "x")
4) Measure the thickness and coating- thickness “y” by a micro meter or other measuring methods. The thickness and coating- thickness “y” in this case is average value measured of 9 points in measurement area.
Measured value by micro meter
No.
Absorbance
"x"
Thickness
“y”
1 2 3 4 5 9
1 2 3
n
I
M
S
E
R
I
E
S
1
5
°
300mm
Measuring objects
Measurement area
Measured points by micro meter
Caution
Conform the front or rear side to flow directions of the sample.
Page 36
INST.No.INE-475-P1CE Software Version 1.00
-30-
8.1 General
When a calibration curve changes depending on the sample temperature Ts, this function enables to switch the calibration curve to other one corresponding to the sample temperature measured. For the input of the sample temperature, a radiation thermometer or a temperature/voltage converter with the output of 4 to 20mADC can be used. By utilizing this function, 10 calibration curves in one group (9 calibration curves only in the group 10) can be switched corresponding to the sample temperature Ts. 99 calibration curves are available and up to 10 groups can be set accordingly.
8.2 Wiring this detector unit
               
8. Sample temperature correction function
Moisture content “Y”
Absorbance "X"
Calibration curve of sample temperature "T3"
Calibration curve of sample temperature "T2"
Calibration curve of sample temperature "T1"
Correction input 4 to 20mA DC
P+ P-
OUT+ OUT-
SA SB SG DI COM DO
COM IN+ IN-
24VDC
4- 20mADC
RS-485 DI G IT A LIN
DIGITAL OUT
4-20mADC
Page 37
INST.No.INE-475-P1CE Software Version 1.00
-31-
8.3.1 Setting with this detector unit
(1) Press SEL key and ENT key simultaneously
for 2 seconds to enter into the operating condition setting mode.
(2) Press SEL key several times to enter into the
sample temperature correction enabled/ disabled setting mode (St).
(3) Select the sample temperature correction
enabled or disabled.
ENT (OFF) Correction OFF (disabled)
ENT (ON) Correction ON (enabled)
* When you go on setting the sample
temperature scaling, skip the next (4) and proceed to (2) of [8.4 Setting a sample temperature scaling (8.4.1 Setting with this detector unit)].
(4)Press SEL key for 2 seconds. The mode
returns the measurement mode.
8.3 Enabling/disabling the sample temperature correction function
Execute this setting with either this detector unit or the operator interface/display unit. When the correction is disabled, the operation is normal which means the calibration curve is not switched corresponding to the sample temperature measured. When it is enabled, the calibration curve is
switched with the sample temperature.
8.4 Setting a sample temperature scaling
When a radiation thermometer or a thermoelectric converter is used, matching the temperature range for its 4 to 20mADC output to the temperature range for 4 to 20mADC input of this detector unit is required. Carry out the scaling with either this detector unit or the operator interface/display unit, following the procedure shown below.
     
8. Sample temperature correction function
8.3.2 Setting with the operator interface/display unit
(1) Set the calibration curve number to 0.
CH 0 ENT
(2) Enter into the sample temperature correction
enabled/ disabled setting mode (St).
MODE 1 9 ENT
(3) Select the sample temperature correction
enabled or disabled.
0 ENT (OFF) Correction OFF (disabled) 1 ENT (ON) Correction ON (enabled)
* When you go on setting the sample
temperature scaling, skip the next (4) and proceed to (2) of [8.4 Setting a sample temperature scaling (8.4.2 Setting with the operator interface/display unit)].
(4) Return the calibration curve number to the
original one.
CH
DATA
S t O F F
8.4.1 Setting with the detector unit
(1) Press SEL key and ENT key simultaneously
for 2 seconds to enter into the operating condition setting mode.
(2) Press SEL key several times to enter into the
sample temperature scaling low limit (tL) setting mode.
(3) Set the sample temperature scaling low
limit (tL). (Setting range: -999.9 to 999.9)
(4) Press SEL key once to enter into the sample
temperature scaling high limit (tH) setting mode.
(5) Set the sample temperature scaling high
limit (tH). (Setting range: -999.9 to 999.9)
CH
DATA
t L 0. 0
8.4.2 Setting with the operator interface/display unit
(1) Set the calibration curve number to 0.
CH 0 ENT
(2) Enter into the sample temperature scaling
setting mode
MODE 2 1 ENT
(3) Set the sample temperature scaling low limit
(tL). (Setting range: -999.9 to 999.9)
(4) Press SEL key once to enter into the sample
temperature scaling high limit (tH) setting mode.
(5) Set the sample temperature scaling high limit
(tH). (Setting range: -999.9 to 999.9)
Page 38
INST.No.INE-475-P1CE Software Version 1.00
-32-
8.5 Setting sample temperature
The following key operation is common to this detector unit or the operator interface/display unit. (1) Press SEL key for 2 seconds to enter into the calibration curve
setting mode.
(2) Press SEL key several times to enter into the sample
temperature setting mode (tS).
(3) Set the sample temperature.Setting range: -999.9 to 999.9
8.6 Setting calibration curve data
(1) Specify a group for calibration curve number to be used.
The groups are from 1 to 10 and each one corresponds to the following calibration curve
numbers respectively. Group 1: Calibration curve number 1 to 10 Group 6: Calibration curve number 51 to 60 Group 2: Calibration curve number 11 to 20 Group 7: Calibration curve number 61 to 70 Group 3: Calibration curve number 21 to 30 Group 8: Calibration curve number 71 to 80 Group 4: Calibration curve number 31 to 40 Group 9: Calibration curve number 81 to 90 Group 5: Calibration curve number 41 to 50 Group 10: Calibration curve number 91 to 99
(2) Example: Assuming that the group 1 has been selected,
Set the following calibration curve at the sample low limit temperature T1 to the calibration
curve No. 1.
y = a3x3 + a2x2 + a1x + a0
Only the calibration curve by polynomial expression can be used.
Further, set the sample temperature T1. For its setting, refer to [8.5 Setting sample temperature].
(3) Set the calibration curves at the sample temperature Ti (i = 2 to 10) to the calibration curve
numbers 2 to 10.
However, Ti should fulfill the following expression.
T1 (Sample low limit temperature) T2 ≦ ………. ≦ T9 ≦ T10 (Sample high limit temperature)
(4) Enter 999.9 into the Ti not used. Be careful that, if this value is not entered, a calibration curve
not used is computed as a data.
* The default of the sample temperature Ti is 999.9.
(5) For the correction expression coefficients b1 and b0 and the constants (low limit range, high limit
range, low limit alarm, high limit alarm and preset value), the values for the first calibration
curve number in a group are used.
Group 1: Calibration curve number 1 Group 6: Calibration curve number 51
Group 2: Calibration curve number 11 Group 7: Calibration curve number 61
Group 3: Calibration curve number 21 Group 8: Calibration curve number 71
Group 4: Calibration curve number 31 Group 9: Calibration curve number 81
Group 5: Calibration curve number 41 Group 10: Calibration curve number 91
CH
DATA
t S 9 9 9. 9
8. Sample temperature correction function
Page 39
INST.No.INE-475-P1CE Software Version 1.00
-33-
8.7 Measurement
(1) When ON (enabled) is selected in the sample temperature correction and one calibration curve
number in a group is set, the calibration curve number is automatically switched corresponding to the sample temperature. For example, when the sample temperature Ts fulfills the following expression, the calibration curve number becomes “i”.
T1 ≦ ….. ≦ Ti Ts < Ti + 1 ≦ ….. ≦ T10
(2) However, the measured value Y is not the measured value Y1 on the calibration curve No. “i”
but becomes the following value interpolated from the measured value Yi and the measured value Y
i+1
on the calibration curve No. “i+1”.
Y = (Y
i+1
- Y
i
) ×

+ Y
i
(3) When the sample temperature Ts is lower than the sample temperature of the first calibration
curve number in the group or when it is higher than the sample temperature of the last calibration curve number, the following error display appears.
However, the calibration curve, of which sample temperature is set as 999.9, is ignored.
* For the sample temperature lower than the sample temperature of the first calibration curve
number in the group ….. Er15 (Sample temperature too low)
* For the sample temperature higher than the sample temperature of the last calibration curve number
in the group ….. Er14 (Sample temperature too high)
Ts - Ti
T
i+1
- Ti
8.8.1 Displaying with this detector unit
(1) Set the calibration curve number to 0.
CH 0 ENT (2) Press key for 2 seconds to display the
sample temperature.
(3) For canceling of the sample temperature
display, press SEL key for 2 seconds or
press CH key for 2 seconds
8.8.2 Displaying with the operator interface/dis p lay u n i t
(1) Set the calibration curve number to 0.
CH 0 ENT
(2) Display the sample temperature.
MODE 2 0 ENT (3) For canceling of the sample temperature
display, press SEL key for 2 seconds or press CH key for 2 seconds
8.8 Displaying sample temperature Ts
8. Sample temperature correction function
Page 40
INST.No.INE-475-P1CE Software Version 1.00
-34-
9.1 Periodical inspection
The following inspections are required periodically.
(1) Cleaning the cover glass surface
Keep the cover glass surface of this detector unit clean all times. Wipe off dirt and flogging with gauze or cloths like that.
It is more effective to impregnate the gauze with alcohol.
(2) Installation
Check this detector unit is correctly installed and fixed securely.
(3) Connections
Check the connections to all terminals of this detector unit, the operator interface/display unit, a receiving instrument or other instruments are fixed securely.
(4) Purge air
Check flow, pressure and cleanness of air supplied for air-purging are adequate.
(5) Calibration
For maintaining the accuracy of this detector unit, the calibration of the unit once in 3 months is recommended by using the output checker plate (Model IR-WEB or IR-WEB3: sold separately).
9.2 Self-diagnosis function
The self-diagnosis function is built in this detector unit. When an abnormal condition occurs, the corresponding error number is displayed. All error numbers are common to this detector unit and the operator interface/d isplay unit.
Error
No.
Item Contents Measures
Diagnosed abnormal condition*
-----
High limit over range Measured value is higher than
the displaying range.
Check the calibration curve data. Check the output by the output checker plate.
-----
Low limit over range Measured value is lower than the
displaying range.
Check the calibration curve data. Check the output by the output checker plate.
Er01 Nonvolatile memory
abnormal
Writing to or reading from the nonvolatile memory is disabled.
Return this detector unit to CHINO.
{
Er03 Motor rotation
abnormal
The motor stops or rotates abnormally.
Replace the motor.
{
Er06 Computation error Overflow happed in computation Check the output by the output
checker plate.
{
Er07 Element temperature
abnormal
The element temperature is abnormal.
Return this detector unit to CHINO.
{
Er12 Ambient temperature
too high
The ambient temperature of the detector unit is high
Lower the ambient temperature.
Er13 Ambient temperature
too low
The ambient temperature of the detector unit is low
Raise the ambient temperature.
Er14 Sample temperature
too high
The sample temperature is higher than the correction range
Widen the correction range.
Er15 Sample temperature
too low
The sample temperature is lower than the correction range
Widen the correction range.
Er16 Sample low
reflection
The reflection rate of the sample is low or the lamp is burned out.
Close the distance to the sample. If the lamp is burned out, replace it.
Er17 Sample high
reflection
The reflection rate of the sample is high.
Incline or extend the distance to the sample.
Er20 Number of calibration
curves insufficient for sample temperature correction
The number of calibration curves is insufficient for the sample temperature correction..
Set two calibration curves setting sample temperature or more.
9. Inspection and maintenance
(Note) For the items filled with { in the self-diagnosed abnormal conditio n, a self-diagnosed abnormal
condition alarm is out
p
utted from this detector unit and the operator interface/display unit.
Page 41
INST.No.INE-475-P1CE Software Version 1.00
-35-
9.3 Measures against troubles not included with self-diagnosis
If a trouble occurred, take remedial measures referring to the corresponding items of this manual after checking the following points. For the troubles of the operator interface/display unit, refer to the separate manual of the operator interface/display unit.
9.3.1 No display on this detector unit appears.
(1) Is the power normal? (2) Is the fuse normal? (3) Are cables connected securely? (4) Is the connector connecting the data display cover and the internal CPU board connected
securely? (5) Turn off the power once and then turn on it. (6) If the trouble cannot be repaired by the above checks, the power unit in this detector unit may
be defective.
9.3.2 Measured value remains unchanged.
(1) Is HOLD set to OFF? (2) Is PRESET set to OFF? (3) Are the calibration curve data correct? (4) Turn off the power once and then turn on it.
(5) Does the data display with the calibration curve number 0 by using the output checker plate
change?
(6) When the data remains unchanged, initialize the RAM. If this phenomenon is not still recovered,
return this detector unit to CHINO.
9.3.3 Measured value fluctuates.
(1) Is the measuring surface flat? (2) Is the measuring position correct? Make sure that the detector unit does not measure such
bottom face as a conveyor belt, etc. (3) Make sure that any disturbance light is not irradiated onto the measuring surface. (4) Is the smoothing time proper? Certain smoothing time is necessary in stable measurements. (5) Make sure that any noise source does not exist around this detector unit or connection cables. (6) Does the data display with the calibration curve number 0 by using the output checker plate
fluctuate? (7) When the data fluctuates, initialize the RAM. If this phenomenon is not still recovered, return
this detector unit to CHINO.
9.3.4 Measured value is slightly higher or lower than the actual value.
(1) Is the calibration curve number correct? (2) Are the calibration curve data correct? (3) Are HOLD and PRESET set to ON? (4) Is the measuring position correct?
(5) Is the cover glass surface of this detector unit clean? (6) Correct the calibration curve. (Refer to [6.3.3 Correcting calibration curves].)
9. Inspection and maintenance
Page 42
INST.No.INE-475-P1CE Software Version 1.00
-36-
9.4 Maintenance parts
9.4.1 List of maintenance parts
The replacing intervals are reference values. They are subjected to the working conditions.
Article name Quantity Replacing interval Remarks Lamp (with rubber sheet) 1 5 years Motor (with connector) 1 3 years Power board 1 5 years 3 years when the ambient temperature
is 40°C or higher
Fuse 2 Supplied as an attachment
10A Ø5X20mm (Littelfuse 217010)
9.4.2 Replacing a lamp
(1) Turn off the power source to this detector
unit.
(2) Loosen four M4 hexagon socket cap screws
fixing the data display cover and remove the data display cover slightly.
(3) Disconnect the cable connector (26 pins) at the
display board and the connector (7 pins) at the
power board, and then remove the data
display cover from this detector unit
completely. (4) Loosen M3 pan screws fixing the lamp cover,
and remove the lamp cover upward.
(5) Turn the lamp bulb counterclockwise viewed
from the filament side about 30 degree, and pull the lamp from the socket. (If the lamp is not turned easily, turn the lamp bulb by wrapping the rubber sheet around it, which is attached to a new replacement lamp.)
(6) Position the new lamp with the crena portion
of the flange, insert it into the back of the
socket, and then turn it clockwise viewed
from the filament side about 30 degree. (7) Mount the lamp cover, and then mount the
data display cover (including the connections of 2 connectors). The replacement of the lamp is completed.
9. Inspection and maintenance
For replacing maintenance parts, ma ke sure to t urn off the power so urce t o this detect or uni t.
For replacing the lamp, make sure to turn off the power source to this detector unit.
Warning
As the data display cover is
connected with internal components
b
y cables, carefully remove the cover
not to scratch the cables.
As the lamp bulb becomes hot after energization, replace it after cooling it down enough.
Warning
Caution
Caution
Lamp cover
Lamp
Display board connector (26 pins)
Data display cover
Powerboard connector (7 pins)
Page 43
INST.No.INE-475-P1CE Software Version 1.00
-37-
9.4.3 Replacing a motor
(1) Turn off the power source to this detector
unit.
(2) Loosen four M4 hexagon socket cap screws
fixing the data display cover and remove the data display cover slightly.
(3) Disconnect the cable connector (26 pins) at
the display board and the connector (7 pins) at the power board, and then remove the data display cover from this detector unit
completely. (4) Remove the motor connector (2 pins). (5) Loosen the fixing screws (M3 pan screws) of
the motor band fixing the motor and remove
the motor by pulling the motor band upward
while pressing the motor. (6) Make the motor unit stand as shown in the
figure and hold the sector lightly. Remove 4
pieces of M2 screws and separate the sector
from the motor. (7) Make the new motor stand, too, and mount it
to the sector. Note that the sector has front
and back sides. (8) Place the motor unit to make the top of the
motor touch the stopper of the motor
mounting base, cover the motor with the
motor band, and then fix the motor by
screwing the fixing screws of the motor
band. (9) Connect the motor connector, and then
mount the data display cover (including the
connections of 2 connectors). The
replacement of the motor is completed.
For replacing the motor, make sure to turn off the power source to this detector unit.
As the data display cover is co nn ected w ith interna l components
b
y cables, carefully remove the
cover not to scratch the cables.
Never scratch the optical filter attached to the sector during the replacement of the motor.
9. Inspection and maintenance
Warning
Sector
M
otor
Caution
Caution
Motor connector (2 pins)
Motor band
Motor unit
Sto
pp
er of the motor
Display board connector (26 pins)
Data display cover
Powerboard connector (7 pins)
Page 44
INST.No.INE-475-P1CE Software Version 1.00
-38-
9.4.4 Replacing a power board
(1) Turn off the power source to this detector
unit.
(2) Loosen four M4 hexagon socket cap screws
fixing the data display cover and remove the data display cover slightly
(3) Disconnect the cable connector (26 pins) at
the display board and the connector (7 pins) at the power board, and then remove the data display cover from this detector unit completely.
(4) Remove the power board by loosing 4 pieces
of M3 pan screws fixing it. At the same time, replace the heat radiation sheet (about 60mm2, 0.5mm thickness), which is slipped with the power board, with a new heat radiation sheet.
(5) Fix the new power board, while slipping the
new heat radiation sheet, with 4 pieces of M3 pan screws.
(6) Connect the cable connector (26 pins) at the
display board and the connector (7 pins) at
the power board, and then fix the date
display cover. The replacement of the power
board is completed.
9. Inspection and maintenance
For replacing the power board, make sure to turn off the power source to this detector unit.
As the data display cover is connected with internal components
b
y cables, carefully remove the
cover not to scratch the cables.
Warning
Caution
Display board connector (26 pins)
Data display cover
Power board connector (7 pins)
Data display cover
Power board
Display board
Heat radiation sheet
4
p
ieces of M3 pan screws
Power board
Page 45
INST.No.INE-475-P1CE Software Version 1.00
-39-
9.4.5 Replacing a fuse
(1) Turn off the power source to this detector unit. (2) Loosen four M4 hexagon socket cap screws
fixing the terminal board cover and remove the
terminal board cover. (3) Push the socket of the fuse holder a little with a
flat head scre wdriver and turn it counterclockwise
about 90 degrees. (4) Pull out the fuse forward. (5) Insert the new fuse into the back of the socket,
push the socket a little with a flathead screwdriver,
and then turn it clockwise about 90 degrees. (6) Mount the terminal board cover. The r e p l acement
of the fuse is completed.
9.5 Ram clearing
If an abnormal condition of this detector unit is not recovered by all means, initialization of the RAM is required.
9.5.1 Ram clearing
(1) Turn on the power by pressing SEL key and key simultaneously. (2) The contents of the RAM are initialized. Set the calibration curve data, etc. again.
 
9. Inspection and maintenance
For replacing the fuse, make sure to turn off the power source to this detector unit.
Warning
For initializing the RAM, make sure to turn off the power source to this detector unit. As the stored contents of the RAM are initialized, record the calibration curve data, etc. before initialization.
FUSE
P+ P- OUT+ OUT-SA SB SG DI
COM D O
COM I N+ I N-
24VDC
4- 20mA DC
RS-485
DIGITAL IN
DIGITAL OUT
4-20mADC
F
use holder
Terminal board
Caution
Fuse
Socket
Page 46
INST.No.INE-475-P1CE Software Version 1.00
-40-
10.1 Detector unit: IRMA51, IRMA52
Model IRMA51 IRMA52
Measuring system Infrared reflection type 3-wavelength
system (NIR)
Infrared reflection type multiple-wavelength system (Thin-film, Infrared)
Lamp Tungsten-filament lamp
Measurement distance 300mm (200 to 300mm adjustable)
Measurement area 50 x 50mm (at the measuring distance 300mm)
With the output checker plate Detector unit output (absorbance “x”):
Within ±0.015 Within ±0.030
Reproducibility
(Under same ambient temperature and the humidity) The above reproducibility is specified with the output checker plate and it is different from the reproducibility in the actual measurement.
Stability under the EMC test environment
Detector unit output (absorbance “x”) : Within ±0.05
Detector unit output (absorbance “x”) : Within ±0.1
Output signal Analog signal: 4-20mADC and ±0.2% of full scale (Load resistance: 500 or less)
Communication output RS-485 (MODBUS) … Standard Ethernet (LAN) … Option
One of the above outputs is to be specified.
Output update cycle 28ms. Display Data: LED 5 digits, Calibration curve number, others: LED 2 digits
Setting By keys or through communication. Processing function 2-color, 3-color ratio processing and multiple regression processing
Number of calibration curve 99 (maximum)
Calibration curve First to third degree to cubic polynomial and multiple regression
With calibration curve correction (First to second-degree correction)
Smoothing processing 0 to 99.9 seconds Calibration The calibration is enabled by the output checker plate. Detector unit number setting
For the multiple-detector-units connection, detector unit numbers can be specified by key operation.
CH number for the calibration curve setting
CH number for the calibration curve can be specified by key operation.
Self-diagnosis function
Contact and communication output at self-diagnosed abnormal conditions
Correction input function
Correction of measured data with an external 4-20mADC input (1 input) (Sample temperature correction, others)
External Di/o Di (contact input): One function selected from the preset, the data hold or the
real/smooth switching can be inputted via the contact input.
Do (contact output): One function selected from the self-diagnosis function (1 Form
B) or the high/low alarms (1 Form A) can be outputted as the contact output.
Working temperature range
0 to 50°C [The cooling by water or air is necessary if the working temperature is higher than 45°C.] Air is dry-air and air temperature is less than 30°C
Power voltage 24VDC [The power voltage is supplied from the power unit IR-WEP (supplied as an
attachment). Power to the power unit: 100 to 240VAC, 50/60Hz]
Permissible voltage fluctuation
18 to 30V DC (For products conforming to CE: Within 24VDC±10%)
Power consumption Approx. 36VA (maximum) Connection Terminal connections
Case Aluminum dust- and splash-proof structure (IEC529, IP65) Weight Approx. 4.3kg Installation method Bolt suspension method with 4 pieces of M8 bolts
CE EN61326A1A2, Emission : classA Immunity : Annex A Table A. 1
Conditions for CE conformance
1. The connection cable should be used indoors and its length should be up to 30m.
2. One(1) set for DC power source unit should be to one(1) set of this unit.
10. Specifications
Page 47
INST.No.INE-475-P1CE Software Version 1.00
-41-
10.2 Attachments
10.2.1 Power unit IR-WEP
The power of 24VDC is supplied to the detector unit. This unit is supplied with the detector unit as an attachment.
Output voltage 24VDC
Output current 2.1A Working temperature range -10 to 50°C
Power supply 100 to 240VAC (universal power supply), 47 to 450Hz
Allowable voltage fluctuation 85 to 264V AC
Power consumption Approx. 160VA (maximum)
Case Resin Mounting Wall mounting type (DIN rail mounting)
Weight Approx. 380g
10.2.2 Connection cable IR-WERP
It is the exclusive cable for connecting the detector unit to the operator interface/display unit or the power unit.
Structure 4-core cab tire cable (with double shield)
Outside diameter Ø10.5mm
Length Maximum length 200m
Connection Both ends - tips
10.2.3 Air-purge hood IR-WEA
It is used to shield against extraneous lights and clear the measurement window and the sight path.
The output checker plate can be attached at the top of the hood.
Purge air Flow: 5 to 200Nℓ/min
Pressure: Maximum 200kPa (Use instrumentation air not containing oil or dust.)
10.2.4 Output checker plate IR-WEB (For IRMA51) IR-WEB3 (For IRMA52)
Use it to check the detector output at site. Attach it to the top of the air-purge hood.
10.2.5 Installation adaptor IR-WED1
It is used to replace an IR-M series moisture meter, which has already been installed, with the IM series infrared multiple constituents analyzer without any change of the installation place.
Material Aluminum Weight About 0.8kg
10. Specifications
Page 48
INST.No.INE-475-P1CE Software Version 1.00
-42-
10.2.6 Air-cooling box IR-WEX
It is used to protect the detector unit from dusts and cool the detector unit by air. (The detector unit is housed in this air-cooling box.)
Working temperature 0 to 55°C
Material Air-cooling portion: SUS304
Mounting metal portion: Iron
Air Flow: 100 to 500N/min Weight About 14kg
10.2.7 Relaying box IR-WEE
It is used as a relaying terminal for the multiple-detector-units connection and a power unit is built in. The steel case and the SUS case are available. The SUS case is the splash-proof construction.
IR-WEE1 IR-WEE2
Working temperature range 0 to 50°C 0 to 50°C
Material Steel SUS304 Color 5Y7/1 (light beige) -
Weight Approx. 4kg
(including the power unit)
Approx. 4kg (including the power unit)
10. Specifications
Page 49
INST.No.INE-475-P1CE Software Version 1.00
-43-
11.1 Detector unit IRMA51, IRMA52
11. Outside dimensions
IM
SERIES
(Ø25)
(25)
133
52
260
130
120
10
120
65
For mounting the air-purge hood 4-M5, depth 6.5
Screw hole for mounting the detector unit 4-M8, depth11
Cable gland
8
Unit: mm
Page 50
INST.No.INE-475-P1CE Software Version 1.00
-44-
11.2 Attachments
11.2.1 Power unit (conforming to CE) IR-WEP
11.2.2 Connection cable IR-WERP
11. Outside dimensions
‑V
+V
OUTPUT
DC24V2.1A
L
N
AC100‑240V1.3A
90
80
±0.
2
4-Ø4.5
75
60
±
0
.
2
M3.5 terminal screw
7.5
5
91
45
35.3
Detector unit side
Power unit or operator interface/display unit side
30
110
20
200
Ø10
Length is to be specified. (up to 200m)
Insulation coating
Maker tube
Unit: mm
Unit: mm
Page 51
INST.No.INE-475-P1CE Software Version 1.00
-45-
11.2.3 Air-purge hood IR-WEA
11.2.4 Output checker plate IR-WEB (For IRMA51) IR-WEB3 (For IRMA52)
11. Outside dimensions
31
Ø86
Ø100
Purge air inlet
4-M5 hexagon socket cap screw
*1 Max 300 ± 3 to Min 222 ± 3 *2 Max 156 ± 5
*1: IR-WEA *2: IR-WEA3
Unit: mm
10
120
65
Unit: mm
Page 52
INST.No.INE-475-P1CE Software Version 1.00
-46-
11.2.5 Installation adaptor IR-WED1
4-Ø9 (Screw hole for installation)
140
120
35
45
159
100
80
70
(179)
36
6
8
Detector unit
4-M8 (Mounting hole for the detector unit)
11. Outside dimensions
Unit: mm
Page 53
INST.No.INE-475-P1CE Software Version 1.00
-47-
11.2.6 Air-cooling box IR-WEX
11. Outside dimensions
6-Ø10 (for installation)
230 200
363 340
(393)
110
(86)
160
330
260
4.5
44
Measurement window portion of the detector unit
(Measurement distance 300
Hood length adjustable
(222 to 300)
Unit: mm
Page 54
INST.No.INE-475-P1CE Software Version 1.00
-48-
11.2.7 Relaying box IR-WEE
11.2.7-1) IR-WEE1:Steel
11. Outside dimensions
4-Ø8 (installation hole)
(Op
en the door and install.)
250
200
40
40
40
40
80
160
192
(20)
5-Ø22 cable gland (Grommet with membrane)
130
250
270
290
200
40
40
40 40
80
160
Mounting hole for 4-M8
Unit: mm
11.2.7-2) IR-WEE2:SUS304
Unit: mm
Page 55
32-8, KUMANO-CHO, IT ABASHI-KU, TOKYO 173-8632
Telephone: +81-3-3956-2171 Facsimile: +81-3-3956-0915 Web site http://www . chino.co.jp/
INE-475-P1CE Nov-'04 IM Series Infrared Multiple-constituents Analyzer Mirror T ype Multiple-constituents meter/Detector Unit
Models IRMA51, IRMA52 Printed in Japan
Loading...