CHINO AL4000 Specifications

Page 1
AL4000 SERIES
100mm chart MULTI-POINT TYPE
HYBRID MEMORY RECORDER
AL4000 series is a hybrid recorder which employs bright and clear, easy to view LCD display. Measuring value display is prepared as 1 point display, multi-points simultaneous display and digital display + bar graph display. Various measuring and recording settings can be easily done by front key switch and confirmed by LCD digital display.
FEATURES
Corresponds to SD card
Equipped with SD card (sold separately) and it can record data, read and write setting value.
Full multi range
Equipped with DC voltage 10 kinds, T/C 36 kinds, RTD 12 kinds, in total 58 kinds. Easily set the range per channels.
Easy data management by communication interface
Provided with USB port and connect with PC directly. RS232C, RS422A, RS485 and Ethernet communica­tion interface is optionally prepared. When Ethernet is selected, settings from the web and E-mail alarm notifi­cation are added.
Package Software attached.
By Data acquisition software, the use of application expands from recording/management to information processing. *Optional communication interface required. Data analysis software can replay display, wave process, editing and trend display. Parameter setting software can manage the setting information on PC.
Standard alarm display/ Printing function
Set 4 types of alarm per each input points. When alarm occurs, status display “ALM” flashes and measuring value flashes at LCD operation screen.
Chart end detection function available
Can set the alarm operation when chart end is detected.
Various programming function
Process the measured data by programming setting and displayed/recorded data of each channels are shown as programmed result data.
MODELS
AL47□□-□□□-NNN
Input point
06 : 6 points
Communication interface (option)
N:None E:Ethernet R:RS232C A:RS422A/RS485 Q:RS232C+RS485 C:RS422A/RS485+RS485 G:Ethernet + RS422A/RS485
+RS485
Alarm output / remote contacts (option)
0:None 2:Mechanical relay 2 points
(‘a’ contact)
4:Mechanical relay 4 points
(c contact) + remote contact 5 points
A:Mechanical relay 6 points
(‘a’ contact) + remote contact 5 points
Power supply
A:100-240V AC
PSE-367
1
Page 2
AL4000 SERIES
NAME
1 Display
2 Operation/ Setting Key
3 SD card slot 4 Engineering port
1. Graphic LCD display
Display measured data by digital display and analog indication by bar graph display.
1 point display 1 point display + bar graph display
6 points simultaneous display
2. Front key switch
Setting contents can be easily registered by front key switch.
Press Menu key and menu screen (list of setting items) will be displayed to graphic LCD.
(USB communication connector)
3. SD card slot
Save measured data to SD card by designated interval (Fastest 6 points: 1sec). Also, register measuring / recording condition such as range, scale, chart speed and when required, setup the unit by registered conditions.
5. White LED chart illumination
Set ON/OFF/AUTO (OFF after no operation for 3 minutes).
4. Prepare engineering port at the front
Connect with PC by mini-USB cable*. By attached setting software, you can set or change the parameter by PC. *Purchase commercialized product separately.
2
Page 3
RECORDING EXAMPLE
Periodic data printing
Record the data over trace printing by arbitary interval.
Time Digital recording Trace printing
List printing
Print setting data such as range, scale, etc. for each channel.
Chart speed
Range
Scale, tag, burnout
Data print
When the latest data is required, trace printing will stop and record.
Tr ace printingDigital recordingTime
Alarm printing
When alarm activates/reset, prints time, channel no., alarm type and alarm no.
Alarm reset time
Alarm
Alarm reset channel number
Level
Alarm activation time
3
Level
Alarm type
Alarm activation CH No.
Page 4
AL4000 SERIES
INPUT SPECIFICATIONS
Measuring points: 6 Input types: DC voltage --- ±13.8mV, ±27.6mV, ±69.0mV,
DC current --- Max 50mA by external shunt
Thermocouple ---
Resistance thermometer ---
Accuracy ratings: Refer to the table of measuring
Measuring interval:1 second / 6 points Input resolution: About 1/40,000 or better (converted to
Reference junction compensation accuracy:
Burnout: Burnout detection function for thermocouple
Maximum common mode voltage:
Common mode rejection ratio:
Normal mode rejection ratio:
Te r minal board: Removable when wiring.
range/accuracy ratings/display resolution
reference range)
At ambient temperature:23±10 K, E, J, T, N Platinel
Other than above ---
input and RTD input. Upper burnout, lower burnout or burnout disabled is selectable for each input.
30V AC/60V DC
130dB or more (50/60Hz)
50dB or more (50/60Hz)
±200mV, ±500mV, ±1V ±5V, ±10V, ±20V, ±50V
resistor (100, 250) (sold separately)
K, E, J, T, R, S, B, N, U, L, W-WRe26, WRe5-WRe26, PtRh40-PtRh20, NiMo-Ni, CR-AuFe, Platinel , Au/Pt
Pt100, old Pt100, JPt100, Pt50, Pt-Co
--­±0.5℃ or EMF 20µV, whichever greater
±1.0℃ or EMF 40µV, whichever greater
DISPLAY SPECIFICATIONS
Analog display: LCD bar graph 100mm Digital display: Monographic type LCD
Dots: 240 x 48 dots Display area: 106 x 16mm Display item: All channels simultaneous display,
Status display: REC, CARD, ALM
(Backlight AUTO / Always ON settable)
year/month/day, hour/minute, alarm activate channel, chart speed display of measuring value.
ALARM DISPLAY
Alarm display: Status display “ALM” flash, measuring value
Alarm types: Absolute alarm, differential alarm, rate-of-
Alarm settings: Individual settings, Max 4 levels/channel Alarm output: Mechanical relay 2 or 6 points (‘a’ contact)
flash at operation screen
change alarm, FAIL, calendar timer, chart end.
Mechanical relay 4 points (‘c’ contact)
STANDARDS (CONFORMITY PENDING)
CE marking: EN61326-1
UL: UL61010-1 2nd edition CSA (C-UL): CAN/CSA C22.2 No.61010-1-04 Protection: IEC 60529 IP54
EN61010-1 *Under EMC test condition, variation in
indication value is ±20% or ±2mV at maximum,
whichever is larger.
RECORDING SPECIFICATIONS
Dotting interval: 5 seconds/point, 2.5 seconds/point
Recording method:Wire-dot type 6-color ribbon Record/Printed color:
Chart paper: Fan-fold type
Chart speed: 1 to 1500mm / h, in 1mm/h increments
Periodic data printing:
Data printing: When required, interrupt trace printing and
Alarm printing: Alarm activated --- Time, channel no., alarm
List printing: When required, interrupt trace printing and print
Message printing: Print when required
ON/OFF of display and recording:
Subtract printing: Record difference between reference channel
Zone printing: 2 divisions Compressed/Expanded printing:
Automatic range shift printing:
Skip function: No display or printing of channels of which
Interlock to chart speed
Tr ace printing (default colors)
Channel no.
Color Red Black Blue
Channel no.
Color Green Brown Purple
Digital recording
Periodic data printing
Alarm printing
List printing
Total width 114mm, total length 10m, effective chart width 100mm
(12.5mm / h can be set exceptionally)
Digital printing is added to trace printing at month / day, time, channel no., data, unit Interval (hour/time) arbitrary setting.
digital print time and measuring value.
type and level Alarm reset --- Time, channel no., alarm level Memory capacity --- Max. 48 data
date, chart speed and setting information of each channel.
Up to 15 characters/message, register up to 20 characters.
Select ON / OFF of display per each channel, trace recording to chart, digital recording to chart, recording to SD card
and measuring value or between reference value (set value) and measuring value.
Range limit is made non-linear and specific chart recording lower/upper limit is shrunk or expanded.
Recording range is shifted automatically to another set range when measured value exceeds the current range. Overlap function available.
ranges are not set.
132
45 6
Repetition of red, black, blue, green, brown and purple
Activate: Red, Reset: Green
Black (channel each items color are same as trace printing color)
CONNECTIVITY
SD card slot
USB port for PC
PLC
Web browser
SD card
Application
RS422A / RS485 *Option
DBKP
FTP server
Ethernet
LT
4
Page 5
GENERAL SPECIFICATIONS
Rated power voltage:
Maximum power consumption:
Normal operation condition:
Case material: Door --- Aluminum die-casting
Case color: Door--- Black (equivalent of Munsell N3.0)
Mounting: Panel mounting Weight: About 3.0kg Te r minal screw: Power terminal,
100 to 240VAC, 50/60Hz
Max 40VA 100V AC balanced: 20VA, 240V AC balanced: 27VA
Ambient temperature range: 0 to 50ºC (20 to 65%) Ambient humidity range: 20 to 80%RH (5 to 40ºC) Power voltage:90 to 264V AC Power frequency:50/60Hz ±2% Attitude: forward tilting 0º, backward tilting 0 to 30º, left/right 0 to 10º
Front panel --- Glass Case --- Cold-rolled steel plate
Glass--- Clear and colorless Case --- Gray (equivalent of Munsell N7.0)
Protective conductor terminal --- M4.0 Measuring input terminal, alarm output terminal Remote contact terminal --- M3.5 Communication terminal --- M3.0
OPTIONS
Remote contact: By external relay contact signal
Alarm output: Mechanical relay (‘a’ contact) 2 points, 6 points
Communication interface: RS232C, RS422A, RS485, Ethernet
(digital contact: short or open), you can select chart speed or data printing Input points: 5 points Input signal: Digital contact signal or open
collector signal Exterior output: 5V DC/2mA Function: 1. Record start/stop
2. Chart speed 3-speed switch
3. Data printing
4. List printing
5. Message printing
6. Operation record (Record ON/OFF condition to the designate location by bar line)
7. Integration/F value reset
8. Memory card (record start/stop)
9. Alarm output rest
10.Time correction
Max. load 100 to 240VAC 0.2A 30V DC 0.2A Min. load 5V DC 10mA
Mechanical relay (‘c’ contact) 4 points
Max. load 100 to 240VAC 0.2A 30V DC 0.2A Min. load 5V DC 10mA
ACCESSORIES
SD Card
512MB
1GB 2GB
Model : RZ-SMC512 Model : RZ-SMC1G Model : RZ-SMC2G
MEASURING RANGES/ACCURACY RATING/DISPLAY RESOLUTION
Input type Measuring range
-
13.8 to 13.8mV
-
mV
DC voltage
V
K
E
J
T
R
S
Thermocouple
B
N
U
L
W-WRe26
WRe5-WRe26
27.6 to 27.6mV
-
69.0 to 69.0mV
-
200 to 200mV
-
500 to 500mV
-
1to 1V
-
5to 5V
-
10 to 10V
-
20 to 20V
-
50 to 50V
-
200 to 300ºC
-
200 to 600ºC
-
200 to 1370ºC
-
200 to 200ºC
-
200 to 350ºC
-
200 to 900ºC
-
200 to 250ºC
-
200 to 500ºC
-
200 to 1200ºC
-
200 to 250ºC
-
200 to 400ºC
0to1200ºC 0to1760ºC 0to1300ºC 0to1760ºC 0to1820ºC
-
200 to 400ºC
-
200 to 750ºC
-
200 to 1300ºC
-
200 to 250ºC
-
200 to 500ºC
-
200 to 600ºC
-
200 to 250ºC
-
200 to 500ºC
-
200 to 900ºC
0to2315ºC 0to2315ºC 0to 290ºC
NiMo-Ni
0to 600ºC 0to1310ºC 0to 350ºC
Platinel
0to 650ºC 0to1390ºC
PtRh40-PtRh20
CR-AuFe
Au/Pt
Pt100
Old Pt100
RTD
JPt100
Pt50
Pt-Co
Note: The accuracy ratings are converted into the measuring range under reference condition. Thermocouple input does not contain reference junction compensation accuracy.
K, E, J, T, R, S, B, N : IEC584(1977, 1982), JIS C 1602-1995, JIS C 1605-1995 W-WRe26, NiMo-Ni, Platinel , PtRh40-PtRh20, CR-AuFe, Au/Pt : ASTM E1751 WRe5-WRe26 : ASTM E988 U, L : DIN43710-1985 Pt100 : IEC751(1995), JIS C 1604-1997 Old Pt100 : IEC751(1983), JIS C 1604-1989, JIS C 1606-1989 JPt100 : JIS C 1604-1981, JIS C 1606-1986, Pt50 : JIS C 1604-1981 Pt-Co : CHINO
0to1880ºC 0to 280 K 0to1000ºC
-
140 to 150ºC
-
200 to 300ºC
-
200 to 649ºC
-
200 to 850ºC
-
140 to 150ºC
-
200 to 300ºC
-
200 to 649ºC
-
140 to 150ºC
-
200 to 300ºC
-
200 to 649ºC
-
200 to 649ºC
4to 374K
Reference range
±13.8mV ±27.6mV ±69.0mV
±200mV ±500mV
± 1V
± 5V ± 10V ± 20V ± 50V
±13.8mV ±27.6mV ±69.0mV ±13.8mV ±27.6mV ±69.0mV ±13.8mV ±27.6mV ±69.0mV ±13.8mV ±27.6mV ±13.8mV ±27.6mV ±13.8mV ±27.6mV ±13.8mV ±13.8mV ±27.6mV ±69.0mV ±13.8mV ±27.6mV ±69.0mV ±13.8mV ±27.6mV ±69.0mV ±69.0mV ±69.0mV ±13.8mV ±27.6mV ±69.0mV ±13.8mV ±27.6mV ±69.0mV ±13.8mV
±6.9mV
±27.6mV
160 220 340 400 160 220 340 160 220 340 220
220
Accuracy ratings
±0.1% ±1digit
±0.1% ±1digit
±0.2% ±1digit
±0.1% ±1digit
±0.15%
±1digit
Display
resolution
10µV 10µV
10µV 100µV 100µV
10mV 10mV 10mV 10mV 10mV
0.1ºC
0.1ºC 1 ºC
0.1ºC
0.1ºC 1 ºC
0.1ºC
0.1ºC 1 ºC
0.1ºC
0.1ºC 1 ºC 1 ºC 1 ºC 1 ºC 1 ºC
0.1ºC
0.1ºC 1 ºC
0.1ºC
0.1ºC
0.1ºC
0.1ºC
0.1ºC 1 ºC 1 ºC 1 ºC
0.1ºC
0.1ºC 1 ºC
0.1ºC
0.1ºC 1 ºC 1 ºC
0.1 K
0.1ºC
0.1ºC
0.1ºC
0.1ºC
0.1ºC
0.1ºC
0.1ºC
0.1ºC
0.1ºC
0.1ºC
0.1ºC
0.1ºC
0.1 K
5
Page 6
APPLICATION SOFTWARE (standard attached)
Data Acquisition Software
You can acquire data easily to your PC.
*Optional communication interface required
List Data Screen
Trend Data Screen
Data Analysis Software
Open the binary file recorded in the SD card, replay display and edit the trend of acquired data file.
TERMINAL ARRANGEMENT
Alarm relay output (6 points ‘a’ contact) + remote contacts and communication interface
Communication terminal
RS232C RS422A
COM1
RS485
*
COM2
RS485
* RS232C and RS422A/485 are specified on purchase.
12345678
SA SB SG
SG SD RD SG
SDA SDB RDA RDB
SG SA SB SA SB
Measurement input terminals
TC.mV(+), RTD(A) terminals
TC.mV(
-
), RTD(B) terminals
RTD(B) terminals
Short between 5-7, 6-8
Parameter Setting Software
Control the setting information at PC by using communication interface or USB port (standard equipped)
Alarm output terminals (option)
N.O terminal
COM terminal
Power/protective conductive terminals
Remote contact terminals (option)
Alarm relay output (4 points ‘c’ contact) + remote contacts and communication interface
Ethernet connector (option)
Communication terminal
COM1
*
COM2
DIMENSIONS
144
144
18
63 63
RS232C RS422A RS485 RS485
* RS232C and RS422A/485 are specified on purchase.
12345678
SA SB SG
SG SD RD
SDA SDB RDA RDB
SG SG SA SB SA SB
Measurement input terminals
TC.mV(+), RTD(A) terminals
TC.mV(
-
), RTD(B) terminals
RTD(B) terminals
Short between 5-7, 6-8
Max 216
Alarm output terminals (option) N.O terminal
N.C terminal
Ethernet connector (option)
Panel cutout
*
162
137
Power/protective conductive terminals
Remote contact terminals (option)
COM terminal
200
200
1
0
138
27.2
*Max216, When alarm output/remote contacts unit and communication unit are added
154.4
195
32-8 KUMANO-CHO,ITABASHI-KU,TOKYO 173-8632
Telephone : +81-3-3956-2171 Facsimile : +81-3-3956-0915 E-mail : inter@chino.co.jp Website : www.chino.co.jp/
1
138
 0
Unit :mm
Specifications subject to change without notice. Printed in Japan (I) 2011. 12
$
Loading...