2. GAS RECOVERY SYSTEMS - INSTALLATION
2.1.1 The exhaust gas can be exhausted directly to atmosphere, or can be recovered to a lower pressure gas system.
If vented directly to atmosphere, CheckPoint supplies the pump with a muffler to reduce exhaust noise.
2.1.2 If recovering the exhaust gas, connect the gas recovery line from the process to the open 3/8” tubing
connector on the pump. The pressure in the recovery line (“recovery pressure”) must be lower than the supply
pressure. To calculate the recovery pressure given your supply pressure in PSIG, use the following formula:
RECOVERY PRESSURE = SUPPLY PRESSURE -
[DISCHARGE PRESSURE]
- 30 PSIG
[AMPLIFICATION RATIO]
2.1.3
For gas recovery systems, the maximum recovery pressure, based on a supply pressure of 150 PSIG (the
maximum allowable supply pressure) is given by:
MAXIMUM RECOVERY PRESSURE = 120 PSIG -
[DISCHARGE PRESSURE]
[AMPLIFICATION RATIO]
2.1.4 For more information on how to set up a recovery system for your 6200 pump, please contact CheckPoint or
an authorized distributor.
3. PUMP OPERATION
3.1 Setting The Pump Stroke Rate
3.1.1 Setting Pump Speed Using a Calibration Gauge
The following directions are for setting the pump speed using a calibration gauge. There are a variety of calibration
gauges available, including a complete line of appropriately-sized CheckPoint gauges for every CheckPoint pump.
To ensure that your pump is working as it should and that chemical is being delivered at the rate you need, it is
important to use a calibration gauge or a discharge-side flow meter.
3.1.1.1 Most calibration gauges are designed to read properly when one full minute of pumping has taken place.
However, if the liquid level drops too fast to allow for a full minute, shorter periods are acceptable. Try to size the
gauge so that at least a 30 second test can be made, however, or a loss of accuracy will result.
3.1.1.2 Proper gauge placement and plumbing is important. Please refer to Figure 1 for appropriate valving and
placement, and for reference numbers as used in this section. The calibration gauge is labeled as number 5.
3.1.1.3 With the pump either running or stopped, open the Gauge Fill Valve (shown as #4 in Figure 1). The gauge
should begin to fill. Continue filling until the chemical level is at or near the top markings on the gauge, then close
the Gauge Fill Valve.
3.1.1.4 Now ensure that the CheckPoint pump is running. Take note of the level of chemical in the gauge using
the appropriate scale for the volume units you want to measure the pump’s output in. Usually the gauge will show
liters on one scale and quarts or gallons on the other. It is best to write down the number so that you can
calculate flow accurately.
3.1.1.5 Open the Gauge Fill Valve (#4), and simultaneously close the Chemical Supply Valve (shown as #1 in
Figure 1). This isolates the pump and gauge so that the pump is being supplied chemical directly from the gauge.