Check-line DTS-200, DTS-500, DTS-1000, DTS-2000, DTS-2500 Operating Manual

...
Page 1
CHECK•LINE
®
BY ELECTROMATIC
MODEL DTS
DIGITAL TENSION METER
Operating Manual
Page 2
TABLE OF CONTENTS
1.0 Introduction ................................................................................................................. 02
2.0 Overview of Operating And Display Elements .......................................................... 03
3.0 Set-up ...................................................................................................................... 06
3.1 Charging the battery
3.2 Turning the tension meter ON and OFF
4.0 Tension Meter Settings ............................................................................................... 07
4.1 Procedure
4.2 Setting chart
4.3 Material menu
4.4 Password
4.5 Factory Reset
5.0 Operation .................................................................................................................... 10
5.1 Requirements
5.2 Material thickness compensator
5.3 Zero adjustment of the measurement position
5.4 Inserting and removing the maaterial to be measured
5.5 Damping
5.6 ALARM function
5.7 Cal. adjustment
5.8 Peak Value
6.0 Creating a Material Characteristic .............................................................................. 16
6.1 Calibration procedure
6.2 Verifying the calibration
6.3 Memory functions
6.4 Save data
6.5 Displaying the saved measured values
6.6 Deleting the saved measured values
7.0 Specications ............................................................................................................. 21
7.1 General specications
7.2 Available models
7.3 Guide rollers
8.0 Maintenance and Cleaning ............................................................................................ 23
8.1 Maintenance
8.2 Cleaning
9.0 Warranty...................................................................................................................... 24
WARNING: Tensions that exceed the tension range of the instrument by more than 100% may cause permanent damage to the measuring spring and must be avoided under any circumstances.
Page 3
– 2 –
1.0 INTRODUCTION
The DTS is supplied as a complete kit that includes:
(1) Tension meter with accumulator and AC adapter with country adapters (EU, US, UK) (1) Certicate of compliance with the order 2.1 under EN 10204 (1) Operating Instructions (1) Carrying case
NOTE: Unpack the tension meter and inspect it for shipping damage. Notices of defect must be led, in writing, at the latest within 10 days on receipt of the goods.
2.0 OVERVIEW OF OPERATING AND DISPLAY ELEMENTS
12345678910111213
1. Filament guide
2. Guide rollers
3. Measuring roller
4. Material to be measured
5. Display
6. USB interface / power connector
07. Thumbpiece
08. MEMORY key
09. Arrow key
10. ESC / EXIT key
11. SET / DAMP key
12. POWER / ZERO key
13. Adjustment wheel for material thickness compensator
(subject to model or measuring range, otherwise a cover cap is provided)
FIG. 1
Page 4
– 3 –
Numeric display
Display with Bargraph
Adjusted material diameter
Info on material
Damping
Current reading
Memory mode (data stored in the memory)
Battery status indicator
Measurement unit
MIN/MAX Alarm
Adjusted material diameter
Info on material
Damping
Current reading
MIN.-alarm value
Tension range
Memory mode (data stored in the memory)
Battery status indicator
Measurement unit
Peak values MIN/MAX AlarmMAX.-alarm value
FIG. 2
FIG. 3
Page 5
– 4 –
Graphic display
Adjusted material diameter
Damping
Current reading
MIN.-alarm value
Tension range
Memory mode (data stored in the memory)
Battery status indicator
Measurement unit
Alarm values
MIN/MAX Alarm
MAX.-alarm value
Measured values as graph
Info on material
The Y-axis can be scaled with the and buttons.
3 Scalings: total measuring range, set limits range plus approx. 1/3 of MAX.-alarm value and minus approx. 1/3 of MIN.-alarm value, set limits range.
The scaling is only possible, if limit values are entered and the damping is not activated.
With the or button the X-axis can be scaled. The setting range is 0.5 to 60 sec.
The scaling can also be changed during the measurement, whereby values already displayed are deleted when the X-axis is changed.
FIG. 4
Page 6
– 5 –
3.0 SETUP
The tension meter is delivered with a built-in rechargeable LiPo battery, which has been charged at the factory. The tension meter can only be switched on if the battery is still working, i.e. if the battery has enough charge. If the instrument does not power up or if the battery level indicator shows only one bar after power-up switching the tension meter on, the battery needs to be recharged.
NOTE: To ensure maximum battery life, avoid discharging the battery completely or charging it frequently for short periods. The battery should not be stored for a extended period of time when empty. After a maximum storage period of one year, the battery has to be recharged.
3.1 Charging the Battery
To charge the battery, connect the cable of the AC adapter to the low USB output. The battery can also be charged by connecting the USB cable to a PC. When the battery is fully charged, the battery level indicator will show 5 bars . The charging time is approx. 3 ½ hours (using the AC adapter).
NOTES: It is not possible to overcharge the battery. The battery can only be charged at a temperature between +5 °C and +45 °C. Before you connect the AC adapter, verify that the supply voltage is correct (100 V – 240 V).
3.2 Switching the tension meter ON and OFF
To switch ON
Press and hold the button (for approx. 1 sec.) until the DISPLAY shows the measuring range, the software and hardware versions, e.g. E 1.0, and then “0”. During startup, the tension meter performs an automatic zero adjustment (see Section 5.3).
Holding the button makes the display freeze so that you can read the measuring range and the software and hardware versions.
While switching on the tension meter, make sure not to move it. This will cause the automatic zero adjustment to be faulty
To switch OFF
Automatic Switch Off : After 3 minutes of non-use the tension meter will switch off.
Manual Switch Off : Press and hold the button for ve seconds.
Page 7
– 6 –
4.0 TENSION METER SETTINGS
Thumbpiece positions
Pos. 1Pos. 2 Pos. 0
2 = Measurement position 0 = Adjustment position 1 = Threading position (guide rollers in their forward position
Menu settings can be changed with the thumbpiece in Pos. 0 (Adjustment position). To perform measurements the thumbpiece must be in the measurement position (Pos. 2).
4.1 Procedure
1. Press the and buttons simultaneously to access the main menu.
2. The and buttons can be used to select the various menu items of the main menu, the submenus and the settings menus.
3. Press the button to open the selected menu; by pressing the or button you can close the menu without saving changes.
4. In menus with multi-digit elds (e.g. date) use the and buttons to move forward and backward between digits.
5. Press the button to save the settings and exit the settings menu or press the button to exit the current menu without saving.
6. To exit the main menu press the button.
FIG. 5
Page 8
– 7 –
4.2 Tension Meter settings chart
Main Menu Submenu Settings Menu Description
Material [1] to [11] Section 4.3 Material set up
Cal. Adjustment
[–10%] – [+10%] Section 5.7
Display Section 6.5
[numeric]
[Bargraph]
[Graphic}
• Measured value displayed as number and alarm monitoring
• Measured value displayed as number, bar graph trend display and alarm monitoring
• Measured value displayed as graphical trend, measured value/limit values as numbers
Alarms
[ON], [OFF] Activate/deactivate the alarm for all calibrations.
The alarm of a material characteristic is only active if the [Alarm] menu item in Material Setup is activated as well.
Peak Display [ON], [OFF] Activate/deactivate the peak values in the bar graph
display.
Settings
Tension Unit [cN], [N], [lb], [g], [kg] Set the measurement unit.
The available units may vary depending on the measur­ing range of the device.
Diameter Unirt
[mm], [inch] To select betwen mm and inch.
Contrast [0] – [10] To adjust the contrast
Backlight [ON]. [OFF}, [AUTO] Switch the display light on or off. With the AUTO setting,
the light switches on and off automatically depending on the ambient light.
Brightness [1}, [2], [3] Set the display brightness
Screen Rotation
[0°], 90°], [180°], [270°], [AUTO]
To set the display orientation to be xed or to automati­cally adapt to the current orientation of the device.
Auto Power Off
[ON], [OFF]
Toggle the AUTO POWER OFF function on and off
Language [EN], [DE]
Select English or German user language
Date/Time [Time], [Date], [Time
Format]
Set the time, date, and date/time format
Password [0000], [9999]
Section 4.4
Factory Reset
Reset to the factory settings
Factory settings are in bold text.
Page 9
– 8 –
Material Setup Submenu Settings Menu Description
No. of the characteristic­material curve 1 - 5
[character],
[numbers], special character]
To enter a name for the selected material characteristic.
Damping [1] – [9]; [5] Section 5.5
Alarms
[ON], [OFF] To activate or deactivate the alarm function for the
material characteristic.
High Limit
[0000] - [9999] Activate/deactivate the alarm for all calibrations.
The alarm of a material characteristic is only active if the [Alarm] menu item in Material Setup is activated as well.
Low Limit [0000] - [9999] Activate/deactivate the peak values in the bar graph
display.
Calibration 6.1
Start
To perform a calibration, follow the instructions shown on the display.
Cal. Points
between [10 %] and [100 %] of the tension range
Set three calibration points for which a calibration should be performed.
Weights [cN/daN] [g/kg] Set the unit of the calibration weights used.
4.3 Material Menu
In the material menu you can make the settings for the selected material characteristics and perform the calibration. To perform the calibration, the weights for the selected calibration points must be available.
4.4 Password
Setting a password blocks the access to some menu parts for unauthorized users:
0000 All menus are accessible. 0001 – 0999 The Material, Cal. Adjustment, Settings, and Memory Settings menus are locked 1000 – 9999 The entire menu is locked.
NOTE: To apply a changed password, exit the main menu. Entering 0000 as the password disables the password feature. If you forgot the password, please contact the factory to request the master password. As an alternative, you can perform a factory reset. Please note, that in this case, you will lose any customer-dened calibrations.
4.5 Factory Reset
A factory reset resets the tension meter to its original manufacturer settings. This procedure will delete all settings, including any customer-dened material characteristics (calibrations); the factory calibration, however, will be kept.
Customer calibrations will be deleted.
Page 10
– 9 –
5.0 OPERATION
5.1 Requirements
1. Switch the tension meter on (Section 3.2)
2. Dene the required tension meter settings (Section 4.2)
3. Select the desired material characteristic (Section 4.3)
4. Set the material thickness compensator (Section 5.2), if equipped
5. Bring the tension meter into the desired measuring position and carry out a zero adjustment as described in Section 5.3, if required.
5.2 Material thickness compensator (only for devices with a thumbwheel)
The 3-roller-system for tension measurement relies on the displacement of the middle roller (measuring roller) to give an indication of the line tension. As the material thickness changes, there will be a change in the tension reading, even when the line tension has not changed. To compensate for this effect, high-range instruments are usually equipped, as necessary, with a material thickness compensator, which automatically shifts the outer rollers. To adapt the tension meter to the diameter of the material to be measured, push the thumbpiece into the adjustment position “0” (Section 4.0) and pull the thumbwheel of the material thickness compensator backward to the adjustment position (Figure 6). Turn it clockwise or anti-clockwise until the desired diameter, such as , appears in the upper right part of the display. Then push the thumbwheel of the material thickness compensator back to the measurement position (Figure 6). If you set a diameter that has not been calibrated for the selected material characteristic, the set diameter is displayed with a black background color, e.g., the tension meter can be used anyhow, but please note, that depending on the material to be measured, slight deviations in the measured values might occur.
If you rotate the thumbwheel while the thumbpiece is in position 2 (measuring position Figure 1), the reading on the display will be updated, but the value will not be used by the instrument to determine the tension — false measurement! The maximum material thickness you can set is
2.50 mm. Setting a higher diameter can damage the tension meter.
3 = Measurement position 4 = Adjustment position
Pos. 3
Pos. 4
FIG. 6
Page 11
–10 –
5.3 Zero adjustment of the measurement position
Each time the measurement position is changed, the tension meter will automatically perform a zero adjustment.
If the tension meter does not display zero in its measuring position, perform a manual zero adjustment procedure. For this purpose, no material to be measured must have been inserted yet!
Requirements:
1. The tension meter has been switched on as described in Section 3.2.
To carry out zero adjustment:
1. Bring the tension meter into the desired measuring position and make sure not to move it.
2. Press the button.
The DISPLAY momentarily shows and then switches to The tension meter is now adjusted to its new measuring position and ready to take measurements.
5.4 Inserting and removing the material to be measured
Material to be measured
Filament guide
Measuring roller
Thumbpiece
Guide rollers Guide rollers
Thumbpiece positions
Pos. 1Pos. 2 Pos. 0
2 = Measurement position 0 = Adjustment position 1 = Thread in position (the guide rollers extend forward)
1
2
Page 12
–11 –
To insert the material to be measured:
1. Push the thumbpiece in the direction of the arrow into its threading position 1 (1) until the outer guide rollers extend beyond the lament guide.
2. Position the material to be measured into the tension meter in such a way that it contacts the Filament guide and passes between the outer rollers and the middle measuring roller (1) .
3. Slowly move the thumbpiece back into the adjustment position (2) . It is important to ensure that the material to be measured runs smoothly between the measuring roller and the guide rollers.
4. Push the thumbpiece into the measuring position (2) (2)
5. The display shows the measured tension value.
If the thumbpiece is not in position 2, a corresponding message is issued on the display. If the thumbpiece is not locked, faulty measurements might result.
FIG. 7
Page 13
–12 –
6. Press the and button simultaneously to change the different display modes during the working mode.
Do not let the thumbpiece snap back as this could affect the calibration and damage the instrument.
While measuring the thumbwheel of the material thickness compensator must be set to the measurement position, so that the material diameter is not altered inadvertently.
If you rotate the thumbwheel while the thumbpiece is in position 2 (measuring position), the reading on the display will be updated, but the value will not be used by the instrument to determine the tension – false measurement!
To remove the material to be measured:
1. Push the thumbpiece in the direction of the arrow into its threading position (1)(2).
2. Remove the material to be measured.
3. Slowly set back the thumbpiece into the adjustment position (0) (2).
5.5 Damping
Feature to be used for tensions that vary strongly. In the Damping menu (Section 4.3) you can specify separate damping factors for each material characteristic. Back in the display mode, press the button to activate or deactivate the damping function. This is recommended if the displayed values vary strongly. Press the button to increase the damping value or the button to decrease it.
NOTE: Changes made to the damping factor using the arrow keys are not saved permanently in the material characteristic. If you disable the damping feature or select a different material characteristic, the damping factor will be reset to the value originally saved for the material characteristic.
Numerical Display
Display with bargraph
Graphical Display
FIG. 8 FIG. 9 FIG. 10
Page 14
–13 –
The factory setting for the damping factor is 5. The average shown on the display is calculated as follows:
5 old measured values + 4 new measured values
9
Damping can be changed in 9 steps from 01 = low damping:
1 old measured value + 8 new measured values
9
to 9 = high damping:
8 old measured values + 1 new measured value
9
5.6 Using the ALARM function
Requirements
In the Material Setup menu, make sure that the MIN and MAX limit values have been set for each material characteristic.
The limit value alarm can be enabled in the main menu.
5.7 Cal. adjustment
By performing a calibration adjustment, you can adjust a material characteristic calibrated for a particular material to a different material or diameter without creating a new material characteristic.
Requirements
1. Prepare the measuring setup as shown in Fig. 11
2. Make sure that the thumbpiece is in the adjustment position (Section 5.4).
3. Set the material thickness as described in Section 5.2.
4. Select the desired material (Section4.3).
5. Move the tension meter into the desired measuring position and carry out a zero adjustment, if required.
Adjusting
1. Select the Cal. Adjustment menu item from the main menu (Fig. 12)
2. Insert the material to be measured (Section 5.4) and push the thumbpiece into the measuring position.
Page 15
–14 –
3. Press the or key to perform the calibration adjustment until the value on thedisplay corresponds to the weight suspended to the material. The adjustment can be performed in 1 % steps within the range from + 10% to –10%.
4. Press the button to save the determined value.
Guide roller
Guide roller
Measuring Roller
Material to be measured
Weight
fig. 3.5.6b
NOTE: Please note that this is a general value. So determine it separately for each material characteristic and note it down for later reference. It is not possible to save it per material characteristic.
To disable calibration adjustment, reset the value to 0% in the Cal. Adjustment menu.
5.8 Peak value
During a measurement cycle, the tension meter automatically calculates the Peak Minimum and Peak Maximum values. If these values should not be displayed in the bar graph or in the statistics, disable their display in the main menu. Regardless, the Peak values will be used and shown always in the software Tension Inspect 3.
FIG. 11
FIG. 12
Page 16
–15 –
6.0 CREATING A MATERIAL CHARACTERISTIC
The tension meter has been calibrated on material 1 according our factory procedure for a vertical material path and cannot be deleted or overwritten. The the materials and diameters are given in Section 2. Factory calibrations using customer supplied materials follow the same procedure. In this case, however, the calibration on Schmidt material 1 is omitted
NOTE: The material characteristics 1 has been preset in the factory and cannot be overwritten. For these material characteristic, you can only change the damping factor, the alarm function setting, and the limit values.
6.1 Calibration procedure
Calibrations of the tension meter are performed according to the factory procedure using weights that correspond to 10%, 50%, and 90% of the measuring range. In 95% of all industrial applications, the factory calibration has been proven to provide the best results. In particular, it is suitable for comparative purposes. If the material to be measured differs signicantly from the factory calibration material in material type, diameter, rigidity, shape, etc., we recommend to perform a calibration using customer­supplied material. In addition to the two factory-preset materials, you can save up to 9 additional materials.
NOTE: The material thickness unit used for the calibration is the same as currently selected in the Settings menu. It is not possible to overwrite the factory calibrations.
Calibration Units and Calibration Points
• You can select Newton or Kilogram as the unit of the calibration weights.
• For the calibration of the tension meter, three calibration weights are used. For example, if you select 10%, 40%, 70% for the calibration, the weights must correspond to 10%, 40%, and 70% of the measuring range.
In this example, you need the following weights for the DTX-1000: Calibration in Newton: 100 cN, 400 cN, and 700 cN Calibration in Kilogram: 100 g, 400 g, and 700 g
NOTE: The measured values are displayed in the unit set in the Settings menu, independent from the unit that was used to calibrate the material characteristic.
So that they are available when you verify the calibration later or repeat the calibration after a factory reset.
Material to be
measured
Unit Calibration points
ø 1 ø 2
Steel wire cN 10%, 30%, 50% 0.4 mm 0.6 mm
Page 17
–16 –
Requirements:
1. The thumbpiece must be in the adjustment position(4) (Fig. 6).
2. The material thickness compensator (if existing) must be set to the adjustment position (Section 5.2)
3. The unit for the material thickness must have been set (Section 4.3)
To perform the calibration procedure:
1. Select a new material or edit an existing one (by re-entering the values) from the material menu.
2. Enter a designation (you can use the software supplied with the tension meter alternatively).
Guide roller
Guide roller
Measuring Roller
Material to be measured
Weight
NOTE: While performing the calibration, the tension meter must be xed in such a way that the material to be measured runs smoothly between the guide rollers and the measuring roller. The display gives detailed information on the calibration.
In the Material menu, select Calibration
Step 1:
Set the calibration points and weights, e.g. in Newton using the calibration points 10%, 50%, and 90% of full scale
Start: Select the Start menu item.
Alternative calibration points: 10%, 40%, 70% 10%, 30%, 60% 10%, 30%, 50% 10%, 20%, 40% 10%, 40%, 90% 10%, 60%, 90% 10%, 60%, 100%
We recommend to use the 10%, 50%, 90% setting for an initial calibration. If the measured values are not accurate enough when you check the calibration, repeat the calibration using other calibration points. We recommend that you align the middle of the tension range to be measured with the middle calibration point.
FIG. 13
Page 18
–17 –
Step 2: Set the diameter
This step is only required for tension meters with a material thickness compensator
Step 3:
Perform a zero adjustment with the tension meter in its measuring position
Step 4:
Calibrate calibration point 1 Insert material to be measured
Step 5:
Calibrate calibration point 2
Step 6:
Calibrate calibration point 3
Step 7:
Exit the calibration process or calibrate a second diameter. In this case, continue with step 2.
NOTE: By calibrating a second diameter, you can dene a material thickness range to be used for the material, e.g. 0.1 – 0.3 mm. Therefore the tension meter can be used for a diameter range instead of a specic diameter only.
Once you have completed the calibration, make sure to verify it as described in Section 6.2. In case of a large deviation, repeat the calibration or select different calibration points.
Page 19
–18 –
6.2 Verifying the calibration
When verifying the calibration, make sure to select the same material, calibration position and calibration points as used for creating the associated material characteristic. Otherwise, the precision of the measurements will not be sufcient. The tension meter has been calibrated on material 1 according our factory procedure for a vertical material path and cannot be deleted or overwritten. The diameters and materials are given in Section 7.2. Calibrations of the tension meter are performed according to the factory procedure using weights that correspond to 10%, 50% and 90% of the measuring range.
Guide roller
Guide roller
Measuring Roller
Material to be measured
Weight
1. Attach a weight vertically to the material to be measured that corresponds to the tension to be measured (make sure to select the correct unit). The weight must hang freely. (Always use a fresh portion of the material to be measured.)
2. Insert the material as described in Section 5.4
Do not let the thumbpiece snap back as this could affect the calibration and damage the instrument.
3. Before verifying the calibration, move the instrument slowly up and down to compensate for any mechanical friction losses and thus ensure repeatability of the measurements.
4. The tension value shown on the display should be equal to the mass of the suspended weight.
If the verication of the calibration shows a deviation beyond the allowable tolerance so that reliable operation is no longer possible, the instrument needs to be re-calibrated or returned to the factory for repair.
FIG. 14
Page 20
– 19 –
6.3 Memory functions
You can store and display the statistics of one measuring series (last measured value, the average, the minimum and maximum measured values, the peaks, the standard deviation).
Data is recorded during a user-dened period.
6.4 Save data
1. Press the button to start recording the measured values.
2. While recording is in process, “Rec S” ashes on the display.
3. Press the button again to stop recording.
4. To save the recorded data, press the button (the display shows “Mem S”). Data, stored in the memory before, will be overwritten.
5. Press the button to stop recording the measured values without saving
NOTE: To abort data recording, press the button. This stops the procedure. No data will be saved.
6.5 Displaying the saved measured values
1. Press the Recall and buttons simultaneously to display the saved data.
6.6 Deleting the saved measured values
If data is saved in the tension meter, the display shows “Mem” .
Deleting data:
1. Press the Recall and buttons
2. Then press the button and conrm with the button.
This clears the memory.
The display will only show statistical values
Name of the material to be measured MIN. Date and time of the measurement series PEAK max. Last reading PEAK min. Average Standard deviation (Std. Dev.) MAX. Number of measured values
FIG. 15
Page 21
– 20 –
Calibration According factory procedure
Accuracy For PA from 5 % up to 100 % full scale:
± 0.5 % full scale ± 1 digit remaining tension range and other calibration materials: ± 3 % FS* ± 1 digit
Memory for material curves 1 factory calibration plus 4 for customized calibrations
Measuring units Force (switchable): cN, daN, g, kg, N, lb
Thickness (switchable): mm, inch
Overrange Approx. 10 % full scale, full scale, without accuracy guarantee
Overload protection 100% full scale
Measuring principle Strain guage bridge
Measuring roller deection Max. 0.2mm
Signal processing Digital
Measuring frequency Max. 1 kHz (1000 measuements/sec.), internal 8 kHz
Display Graphic LCD display
3 different displays: numeric, numeric with bargraph, X-Y-diagramm (time-tension)
Display update time 2/sec (Display update)
Memory MIN, MAX, PEAK, AVG and LAST
Damping Adjustable electronic damping (averaging)
Cal. adjustment ± 10 steps in 1% increments
Thickness compensation Max. 2.5 mm (not available for all models)
Auto power off Approx. 3 minutes of non-use
Signal processing 16 bit A/D
Temperature coefficient Gain: less than ± 0.01 % full scale* /°C at 25° C
Zero point: better than ± 0.3 % full scale* /°C at 25° C
Temperature range 10 to 45° C
7.0 SPECIFICATIONS
7.1 General Specifications
Continued on next page
Page 22
– 21 –
Model
Tension Ranges
cN
*Measuring Head Width
mm
**Factory
Calibration
Matretial 1
Polyamide [PA]
Monofilament
***Material
Thickness
Compensator
for Included
DTS-200 0.2 - 200.0 66 0.12 mm Ø
DTS-500 0.5 - 500.0 66 0.12 + 0.20 mm Ø yes
DTS-1000 10 - 1000 66 0.20 + 0.40 mm Ø yes
DTS-2000 20 - 2000 66 0.40 + 0.70 mm Ø yes
DTS-2500 25 - 2500 116 0.40 + 0.70 mm Ø yes
DTS-4000 40 - 4000 66 0.50 + 0.80 mm Ø yes
DTS-5000 50 - 5000 116 0.60 + 1.20 mm Ø yes
DTS-10K 0.1 - 10.00 daN 116 0.80 + 1.40 mm Ø yes
DTS-20K 0.2 - 20.00 daN 116 1.20 + 1.80 mm Ø yes
DTS-30K 0.3 - 30.00 daN 216 1.40 + 2.00 mm Ø yes
DTS-50K 0.5 - 50.00 daN 216 steel rope 1.5 mm Ø
(7 x 7 x 0.15)
DTS-60K-V1 0.6 - 60.00 daN 280 steel rope 2.0 mm
Ø(7 x 7 x 0.30)
7.2 Available models
International unit of tensile force: 1 cN = 1.02 g = 0.01 N, 1 daN = 1.02 kg = 10.0 N
* Depending on model, either width of lament guide or outer distance between outside guide rollers
** Suitable for 95 % of all applications. PA = Polyamide Monolament. If the material to be measured differs signicant from the SCHMIDT calibration material in diameter, rigidity, shape, etc., we recommend calibration using customer supplied material. Instruments with calibration on customer sample are not adjusted and calibrated on material 1. *** The DTSB, DTSE, DTSF and DTSL models do not include a material thickness compensator. They are calibrated only on material 1.
The DTSB is calibrated depending to the range on a textile ribbon or thin tape.
Air humidity 85% RH, max
Power supply LiPo Accumulator (approx. 40 hours continuous use,
chaging time aapro. 3.5 hours
Housing material Die-cast aluminum
Housing dimensions 265mm x 78.5 mm 46 mm (L x W x H)
Weight (net/gross) Up to DTS-50K approx. 875 g / 1550 g
DTS-60K-V1 approx 1040 g / 2700 g
Page 23
–22 –
V-grooved Line Speed
max. m/min
Roller Material
Standard 2000 Hard-coated aluminium
Standard 4000 Hardened steel only for DTXL
Code K 3500 Hard-coated aluminium
Code H 5000 Plasma-coated aluminium
Code T 1000 Plastic (POM) black
Code T 4000 Plastic (PVC) red only for DTSF
Code W 1000 Nickel-plated steel
Code ST 1000 Hardened steel
Code B 1000 Tempered steel for measuring
tire cord
Code CE2 1000 Ceramic-coated aluminium
Asymetrical groove Code ASY
1000 Hard-coated aluminium
(not for DTS-200)
Code ASYB 1000 Tempered steel for measuring tire
cord (not for DTS-200)
Code V1 1000 Hard-coated aluminium
(only for DTS-60K-V1)
U-grooved Code U
2000 Hard-coated aluminium(not for
DTS-200)
Code R1 4000 Hard chrome-plated steel
(radius R5)only for DTSL
7.3 Guide Rollers
Page 24
–23 –
8.0 MAINTENANCE AND CLEANING
8.1 Maintenance
The tension meter is easy to maintain. Depending on operating time and load, the instrument should be checked according to the locally valid regulations and conditions. The use of other test methods than the procedure described in Section 5.0 may cause deviating measuring results.
Rollers
You should regularly inspect the rollers to assure that they are running easily and smoothly. You can replace the rollers yourself, as necessary. Please indicate the tension meter model and the serial number (given on the rear side of the tension meter) in your spare-parts order.
For example, To order spare rollers:
Model: DTS-1000 (given on rear side of tension meter) Serial number: 920 - 888888 (given on rear side of tension meter) Standard rollers: Model number R592004 Delivery: 1 set (3 pcs.) of spare rollers 2000 m/min, incl. mounting tool
or
Model: DTS-1000-K (given on rear side of tension meter) Serial number: 920 - 888888 (given on rear side of tension meter) Code K rollers: Model number R592003 Delivery: 1 set (3 pcs.) of spare rollers 3500 m/min, incl. mounting tool
8.2 Cleaning
For cleaning the unit do not use any AGGRESSIVE SOLVENTS such as trichloroethylene or similar chemicals. NO WARRANTY OR LIABILITY shall be accepted for damage resulting from improper cleaning.
Page 25
– 24 –
9.0 WARRANTY
ELECTROMATIC Equipment Co., Inc. (ELECTROMATIC) warrants to the original purchaser that this product is of merchantable quality and conrms in kind and quality with the descriptions and specications thereof. Product failure or malfunction arising out of any defect in workmanship or material in the product existing at the time of delivery thereof which manifests itself within one year from the sale of such product, shall be remedied by repair or replacement of such product, at ELECTROMATIC’s option, except where unauthorized repair, disassembly, tampering, abuse or misapplication has taken place, as determined by ELECTROMATIC. All returnsfor warranty or non-warranty repairs and/or replacement must be authorized by ELECTROMATIC, in advance, with all repacking and shipping expenses to the address below to be borne by the purchaser.
THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO, THE WARRANTY OF MERCHANTABILITY AND FITNESS FOR ANY PARTICULAR PURPOSE OR APPLICATION. ELECTROMATIC SHALL NOT BE RESPONSIBLE NOR LIABLE FOR ANY CONSEQUENTIAL DAMAGE, OF ANY KIND OR NATURE, RESULTING FROM THE USE OF SUPPLIED EQUIPMENT, WHETHER SUCH DAMAGE OCCURS OR IS DISCOVERED BEFORE, UPON OR AFTER REPLACEMENT OR REPAIR, AND WHETHER OR NOT SUCH DAMAGE IS CAUSED BY MANUFACTURER’S OR SUPPLIER’S NEGLIGENCE WITHIN ONE YEAR FROM INVOICE DATE.
Some State jurisdictions or States do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation may not apply to you. The duration of any implied warranty, including, without limitation, tness for any particular purpose and merchantability with respect to this product, is limited to the duration of the foregoing warranty. Some states do not allow limitations on how long an implied warranty lasts but, not withstanding, this warranty, in the absence of such limitations, shall extend for one year from the date of invoice.
ELECTROMATIC Equipment Co., Inc. 600 Oakland Ave. Cedarhurst, NY 11516—USA Tel: 1-800-645-4330/ Tel: 516-295-4300/ Fax: 516-295-4399
Every precaution has been taken in the preparation of this manual. Electromatic Equipment Co., Inc., assumes no responsibility for errors or omissions. Neither is any liability assumed for damages resulting from the use of information contained herein. Any brand or product names mentioned herein are used for identication purposes only, and are trademarks or registered trademarks of their respective holders
Page 26
ELECTROMATIC
E Q U I P M E N T C O., I N C.
600 Oakland Ave., Cedarhurst, NY 11516 – USA TEL: 516-295-4300 • FAX: 516-295-4399
CHECK•LINE
®
INSTRUMENTS
Loading...