Checkline CTM User Manual

5.5 Non-linearity
Most three-point tension meters employ only linear characterization and have large errors at the midpoints (up to 15%). The CTM uses multi-point segmenting to correct for non-linearity, reducing it to less than 0.2%.
Table of Contents
1.0 Introduction .........................................................................................................................3
5.6 Wire Characteristics
6.0 T roubleshooting
Creep Every material including steel exhibits creep under load. It will
neck down over time, quite quickly over the fi rst few seconds and much slower as time progresses. A wire cable also sees creep from the wire spacing and wind. This effect is seen as a display that drifts lower after it has been clamped in line.
Variations Material that varies in diameter or shape will have different
output at the same tension
Strands The best cable assembly is one that is perfectly round, as
it will not change contact geometry with the wire twist. The closer the wire cable cross section appears to be round, the better the measurement performance will be.
Problem Possible Cause Solution
Powers on momen­tarily and turns off
Does not power on Low battery Replace with high quality alka-
Display is completely dark
Display drifts down­ward once installed
Temperature not accurate
Low battery Replace with high quality alka-
line batteries
line batteries
Batteries installed backwards or no spring contact
Software reset Remove battery cap & reinstall
Display contrast too light
Display contrast too dark
Wire material is creeping and inter­nal friction between wires is relieved.
Instrument changed temperature envi­ronments
Instrument exposed to sun
Insure that positive terminals of both batteries (nub) face in­ward – towards the black cap. Check that spring is attached to the battery cap.
after one minute. Attempt to turn power on again.
Hold the Right Arrow key down while pressing the F2 key several times to increase the display contrast. If nothing oc­curs, release both keys. Press the power button and try again.
Hold the Arrow key down while pressing the F1 key several times to decrease the display contrast.
This is normal behavior of wire. Lower display resolution to mask this effect.
Allow instrument to remain in environment until temp stabi­lizes or enter temp manually
Enter temp manually
1.1 Unpacking ...................................................................................................................3
1.2 Description ..................................................................................................................3
1.3 Important features .......................................................................................................4
2.0 Operation 5
2.1 Typical Operation ........................................................................................................5
2.2 Measurement Practices ..............................................................................................6
2.3 Softkey Functions .......................................................................................................7
3.0 Confi guration Mode ............................................................................................................11
3.1 Accessing the Confi guration Mode ...........................................................................11
4.0 Changing Sheaves .............................................................................................................14
5.1 Accuracy ...................................................................................................................15
5.2 Calibration to Specifi c Wire Type ...............................................................................15
5.3 Loading Error ............................................................................................................15
5.4 Non-repeatability .......................................................................................................15
5.5 Non-linearity ..............................................................................................................16
5.6 Wire Characteristics ..................................................................................................16
6.0 Troubleshooting ..................................................................................................................16
16
1
Specifi cations
Power
2 AA, common alkaline batteries. Operational battery life over 40 hours at room temperature. Battery life is reduced at lower temperatures.
5.0 Achieving Best Accuracy
The CTM is an instrument designed to give accuracy that typically exceeds
5.1 Accuracy
normal requirements for wire tensioning. You should have an understanding of what factors affect tension measurement accuracy.
Display
Operational keys
Operational annunciators
Temperature detection
Display resolution
Dot graphic LCD display
Power, Wire, Escape/Clear (Esc), Next ( ) and four softkeys with chang­ing fuction and label, depending on the specifi c menu in use
Unit of measure, battery level
Temperature accuracy approximately ±2° F (±1°C)
2,000 lbf/ 1,000 kgf/ 10,000 N CTM instrument:
Displayed resolution setting
Low Med High
lbf (pound-force) 10 lbf 5 lbf 2 lbf kgf (kilogram-force) 5 kgf 2 kgf 1 kgf N (Newton) 50 N 20 N 10 N Custom units between 101 &
200 divisions
8,000 lbf/ 3,500 kgf/ 35,000 N CTM instrument:
Displayed resolution setting
Low Med High
lbf (pound-force) 50 lbf 20 lbf 10 lbf kgf (kilogram-force) 20 kgf 10 kgf 5 kgf N (Newton) 200 N 100 N 50 N Custom units between 101 &
200 divisions
between 201 &
500 divisions
between 201 &
500 divisions
between 501 &
1000 divisions
between 501 &
1000 divisions
5.2 Calibration to Specifi c Wire Type
Contact your distributor for any additional calibrations you may need.
While it is best to have the instrument calibrated to the specifi c wire size(s) and type(s) used, the CTM can often work adequately in other situations. If the best tension accuracy is required, it is recommended that a calibration be performed for that specifi c wire size and type.
What calibration choice and sheaves should I use?
Situation
Exact wire size
and type is
shown in wire list
Wire size is
same, but type is
not identical
Wire size is not
the same
Do not use the CTM to measure tension for wires if both of the following are true:
1. No wire calibrations are stored of the same diameter as the wire you are looking to measure, and
2. You do not have sheaves of the same diameter.
If both of these conditions exist, contact your distributor.
Wire calibration
selection
Description of
exact match
Description of
same wire
diameter
Closest diameter
Sheave
selection
Sheaves noted
in list
Sheaves noted
for that wire size Sheaves match-
ing the wire size being measured
Accuracy
Best
Good
Fair
Available options
Operating environment
Dimensions
Weight
For ease of use, the display always counts by a multiple of 1, 2 or 5.
Varied wire sizes
Suitable for outdoor use
10˝ x 23˝ x 3˝ (25 cm x 59 cm x 8 cm) approximately
11 lb (5 kg) approximately
2
5.3 Loading Error
5.4 Non-repeatability
Contact your distributor to improve accuracy for a specifi c wire type by calibrating to it.
A tensiometer works by defl ecting the cable, which makes the cable path longer than when a tensiometer is not installed. When the tensiometer is removed, the wire tension decreases as the cable length is restored. This effect is known as loading error. The CTM design elongates the cable by a mere 0.08 inch (2 mm), making loading errors extremely small.
The CTMs sheave with bearing design provides the best mechanical performance. It is also superior at detecting tension that is being added or removed.
15
4.0 Changing Sheaves
1.0 Introduction
Do not use the CTM with cable larger than indicated on the sheaves. Overload and dam­age to the instrument may result.
Do not mix sheave sizes. This will result in inaccurate mea­surement and possible over­load.
As you use the CTM on different diameter cables you must change to the correct sheave size. To change sheaves, remove the hex head screws pointed out in Figure 6 below. Replace the sheaves with the correct letter sheave and reinsert the screws and tighten.
Insure sheaves installed agree with sheaves noted in the Wire calibration. Exception: Sheaves match the wire diameter of the cable to be measured and alternate calibration is selected as per section 5.2.
Insure that the wire rope is riding in the groove of all three sheaves.
Screws
1.1 Unpacking
This manual covers the setup and operation of the CTM Clamp Line Ten­siometer. The CTM is a simple, accurate strand dynamometer. It is can be clamped onto a cable, accurately determine the wire tension and be re­moved in seconds.
The CTM can handle multiple wire diameters, it can display live tension, dual live/peak tension, average tension captured from several tests, dual tension/ temperature display and a check-tensioning graphical display.
With its battery-powered electronic interface, setup and operation is made simple with on-screen prompts.
This manual covers the following:
Unpacking
Setup
Operation
Maintenance
Troubleshooting
When you receive your CTM, unpack it and inspect the container and the instrument for any damage. Report any problems to the shipping company immediately and save the packing materials.
Insert 2 AA batteries into the battery compartment, shown in Figure 1. Your CTM probably comes from the factory with the proper sheave size installed and calibrated for your application. If not, follow the setup directions later in section 3.0 Confi guration Mode and 4.0 Changing Sheaves.
Figure 7
Changing sheaves
1.2 Description
Figure 1
The CTM is shown in Figure 1 with the parts labeled.
Sheaves
Support arm
Lever arm
Front panel
Battery
compartment
14
3
Loading...
+ 6 hidden pages