Checkline AT-VERIFY User Manual

PosiTest
AT
Verifier
Instruction Manual v. 1.0
The PosiTest AT Verifier kit is ideal for checking the accuracy and operation of adhesion testers and is an important component in fulfilling both ISO and in-house quality control requirements.
Kit includes:
- Smart Sensor Indicator (SSI) Device
- Load Cell with steel 20mm Loading Fixture (dolly)
- Built-in rechargeable Lithium Ion Batteries
- Universal AC power adaptor/charger
- Instruction Manuals (2)
- Certificates of Calibration (2)
- Sample Certificate of Accuracy
- Measurement Uncertainty Worksheet
- Hard shell carrying case
Introduction
Introduction
Load Cell
Smart Sensor Indicator (SSI)
Load Cell to SSI Connection Cable
Loading Fixture
1. Connect the PosiTest AT actuator to the load cell. Ensure the quick coupling is locked in place over the dolly fixture. PosiTest AT-M (manual) models only: Close the pressure relief valve (turn clockwise).
2. Press [Power] button to power-up the Smart Sensor Indicator (SSI).
3. Press and release the [TARE] button to ZERO the SSI.
4. Press and release the [PEAK] button to hold the maximum pressure value at the completion of the test. PEAK will display on the SSI.
5. Power ON the PosiTest AT by pressing the green button.
-Ensure units are set to psi and dolly size to 20 mm.
-Important: PosiTest AT-A (automatic) models: Set the pull rate to 30 psi/sec
6. Press the button to ready the PosiTest AT.
7. Press the button again to start the test.
-PosiTest AT-A models will automatically begin building pressure*
-PosiTest AT-M models: slowly prime the pump until the priming pressure is reached (100 psi)*. Return pump handle to its full upright position and maintain a single slow, continuous pump at approximately 30 psi/sec until a verification point has been reached.
*Make no comparisons between PosiTest adhesion tester and SSI measurements
until the priming pressure (100 psi) has been exceeded.
8. Stop the test when the verification point has been reached.
-PosiTest AT-A users press the red button. Reading will flash and actuator will automatically release pressure.
-PosiTest AT-M users simply release the pump handle and open the pressure relief valve (turn counter-clockwise).
9. Record values from PosiTest AT and Smart Sensor Indicator.
10. Prepare for next test verification point.
-Power-down the SSI by pressing the [Power] button.
-Ensure all pressure has been released within the adhesion tester by pushing the PosiTest AT actuator handle completely down. The actuator should be relaxed over the loading fixture.
Repeat steps 2-10 for each additional verification point. See included Sample Certificate of Accuracy for recommended verification points.
2
How to Perform a Test
How to Perform a Test