Chauvin Arnoux C.A 6511, C.A 6513 User guide

■■
■■
■■
INSULATION CONTROLER
■■
■■
ISOLATIONSPRÜFER
■■
■■
CONTROLADORES DE AISLAMIENTO
■■
■■
CONTROLLORI DI ISOLAMENTO
■■
C.A 6511 C.A 6513
MEGOHMMETER
FRANCAIS ENGLISH DEUTSCH Bedienungsanleitung ITALIANO Manuale di Istruzione ESPANOL Manual de Instrucciones
Mode d'Emploi User's Manual
English
SAFETY PRECAUTIONS
- Before any measurement check that there is no voltage present on the circuit to be tested.
- Use the appropriate leads supplied with the instrument.
- At rest, the needle must indicate 0 on the voltmeter scale. If not, adjust it by means of the central (6) zero reset screw.
- When the insulation measurement is finished, leave the instrument connected for a few seconds to allow automatic discharge of the high tension of the device tested.
- During measurement of continuity or resistance, the voltage should be measured first. The presence of a voltage van activate protection and necessitate changing the fuses.
On the -10W +10W and 1000W ranges, the instrument is protected by an HBC fuse. On the other ranges the instrument is statically protected on all the ranges from a voltage of 600V rms applied permanently between the terminals and from a voltage of 1000V rms accidentally applied for less than 15 seconds between the terminals.
- Position the switch in the OFF position when the instrument is not used.
11
SUMMARY
Pages
DESCRIPTION .................................................................................................................................................................... 13
USE ............................................................................................................................................................................................. 13
Battery test .................................................................................................................................................................. 13
Insulation/voltage measurement ................................................................................................................. 13
Continuity measurement ................................................................................................................................... 14
Resistance measurement (C.A 6513 only) ......................................................................................... 14
MAINTENANCE ................................................................................................................................................................. 15
Changing the batteries/fuse ............................................................................................................................ 15
Fitting the shockproof case ............................................................................................................................. 15
Cleaning ......................................................................................................................................................................... 15
Repairs - maintenance ....................................................................................................................................... 15
SPECIFICATIONS ........................................................................................................................................................... 16
Functional specifications ................................................................................................................................... 16
Insulation ....................................................................................................................................................................... 16
Continuity ...................................................................................................................................................................... 16
Resistance (C.A 6513 only) ............................................................................................................................ 16
Voltage ............................................................................................................................................................................ 17
Reference conditions ........................................................................................................................................... 17
Climatic conditions ................................................................................................................................................. 17
Variations in the nominal working range. .............................................................................................. 18
Electrical specifications ...................................................................................................................................... 18
Power supply .............................................................................................................................................................. 18
Electrical safety ........................................................................................................................................................ 19
Mechanical specifications ................................................................................................................................. 19
STANDARDS RESPECTED .................................................................................................................................... 20
TO ORDER ........................................................................................................................................................................... 21
WARRANTY ......................................................................................................................................................................... 21
12
DESCRIPTION
See page 54 and page 55.
- Measurement terminals colour marked- Log scale on yellow background 0.1 to 1000 MW : Insulation measurement.- Linear scale 0 to 10 W : Continuity measurement
Resistance measurement (C.A 6513 only)
- Linear scale 0 to 600 V AC : Voltage measurement- Red / green scale : battery test- Screw to reset needle to mechanical zero- Rotary switch : 4 positions (C.A 6511)
6 positions (C.A 6513)
- Press button
USE
BATTERY TEST
Before starting a series of measurements, check that the batteries powering the instrument are good. Place the switch in the OFF position, press the yellow button, and display the condition of the batteries on the scale of the galvanometer :
- if the needle is in the green zone, the batteries are good
- if the needle is in the red zone, change the four batteries
INSULATION/VOLTAGE MEASUREMENT
Once you are sure the power supply is good (see "Battery test"), you can proceed with insulation measurements. Connect the instrument to the installation to be tested using the accessories supplied (leads and crocodile clip). It is preferable that the "+" is connected to the earth. Select an insulation measurement position with the switch (500 V for the C.A 6511, 500 or 1000 V for the C.A
6513).
At this stage, the instrument is thus automatically set to AC voltmeter. The instrument
makes a voltage measurement between the + and - terminals (scale of voltages up to 600 V AC).
- If the instrument does not indicate any voltage present, the insulation measurement can be
made.
- If the instrument indicates the presence of voltage on the circuit to be measured, insulation
measurement must not be made. Find the source of this voltage and suppress it. Insulation measurements can only be done on this condition. Proceed with the insulation measurement, by pressing the yellow button. The instrument generates a high voltage between the + and - terminals. The reading is done on the yellow log scale 0.1 to 1000 MW . As soon as the button is released, the instrument switches to voltage measurement. The high tension present on the tested device is discharged via the leads of the instrument. You must therefore leave the instrument connected for a few seconds once the measurement is finished (needle returns to 0 V).
13
CONTINUITY MEASUREMENT
After having connected the instrument to the installation to be tested, you must check that no voltage is present (see "Insulation measurement/voltage").
Set the switch to the "10 W" position. The measurement is done automatically without having to press the yellow button. Read the value on the white scale 0 to 10 W. Then set the switch to the "-10 W" position and check that the needle of the instrument indicates the same continuity value. If this second value is different from the preceding one you must make the following calculation :
R
= R+10 W + R-10 W
continuity
2
To have a better measurement accuracy on the +10 W and -10 W ranges, measure the resistance of the leads by short-circuiting them. Then subtract this value from the measured resistances.
Remarks : - In order to save battery power, it is preferable to disconnect the leads once
the continuity measurement has been finished.
- If the instrument, on continuity measurement, is not connected, or is badly connected, the needle then positions itself to the far right.
RESISTANCE MEASUREMENT (C.A 6513 only)
After having connected the instrument to the installation to be tested, you must check that no voltage is present (see "Insulation measurement/voltage").
Set the switch to the "1000 W" position. The measurement is done automatically without the need to press the yellow button. Read the value on the white scale 0 to 10 W, applying a coefficient x100.
Remarks : - If the instrument, on resistance measurement, is not connected, or is badly
connected, the needle then positions itself to the far right.
- In order to save battery power, it is preferable to disconnect the leads once the resistance measurement is finished.
14
MAINTENANCE
CHANGING THE BATTERIES/FUSE
Check that no terminals are connected and that the switch is in the OFF position before opening the instrument. For safety, the instrument can only be opened if the leads are disconnected from the terminals. Access to the fuses and the batteries is by opening the back of the case. Remove the shockproof case if there is one, then undo the tool release screws with a screwdriver and remove the back of the case. Warning : never press the yellow button when the case is open.
Tool release
screws
Fold away stand
+
-
+
-
+
-
+
-
Fuse 1.6 A
4 batteries
1.5 V LR6
FITTING THE SHOCKPROOF CASE
In order not to damage the window protecting the display, it is preferable to first engage the upper part of the instrument in the case, then to push the lower part in.
CLEANING
Clean the case of the instrument with a soft damp cloth. Use soapy water but not alcohol, solvents or hydrocarbons.
REPAIRS-MAINTENANCE
The maintenance operations must be done with original parts, and by an approved service department. To obtain a good quality measurement, it is preferable to recalibrate every two years, or every year in case of intensive use.
15
SPECIFICATIONS
FUNCTIONAL SPECIFICATIONS
Insulation
Measurement range : 0.1 to 1000 MW
Scale 0,1 MW to 1000 M W Accuracy ± 5% of the measurement
Range 500 V 1000V*
Empty voltage 600V 1200V
Test current ³ 1 mA for ³ 1 mA for
Short circuit current £ 6 mA £ 6 mA
* C.A 6513 only
The time required for the voltage present on the terminals to decrease by 90% of its value, once the MW button is released (= discharge time), is 1s/µF.
Continuity
Measurement range : 0 to 10 W with reversal of the measurement current
R £ 500 kW R £ 1 MW
Scale 0 - 10 W
Accuracy ± 3% of end of scale
Short circuit
current
Empty voltage 4.5 V £ V £ 6.5 V
Resistance (C.A 6513 only)
Measurement range : 0 to 1000 W
Scale 0 - 1000 W
Accuracy ± 3% of end of scale
Short circuit
current
Empty voltage 4.5 V £ V £ 6.5 V
³ 200 mA
³ 2mA
16
Voltage
Measurement range : 0 to 600 V AC
Scale 0 - 600 V
Accuracy ± 3% of end of scale
Input impedance 300 kW
Reference conditions
Distortion quantities Reference conditions
Temperature 20 °C ± 3K
Relative humidity 45 to 55 % RH
Supply voltage 5.5 V ± 0.2 V
Voltage frequency 45 Hz to 65 Hz
Electric field < 1 V/m
Magnetic field < 40 A/m
Position Horizontale ± 5°
Climatic conditions
Humidity in % RH without condensation
Nominal reference range Working range Storage range
17
Temperature
in °C
Variations in the nominal working range.
Distortion quantities
Operating 0 + 90 ° - £ 5 % of
position 0 - 90 ° - the measurement
Temperature -10 to + 55 °C 3 % / 10°C (1) 5 % / 10°C (1)
Humidity 20 to 80 % RH (2) 5 % of 10 % of
Supply voltage 4.5 to 6.5 V 0.1 % of 0.2 % of
Frequency (on voltmeter) 45 to 400 Hz - £ 0.1 % of
Electromagnetic field 0 to 400 A/m (3) - 1/2 class
(1) For measurements in MW, it is a % of the measurement.
For the other measurements, it is end of scale. (2) For insulation measurements from 0.1 to 100 MW , voltage measurements from 0 to 600V, to 1000 W. (3) The magnetic fields envisaged are fields with stable amplitude and direction, and of
frequency between 0 and 60Hz.
ELECTRICAL SPECIFICATIONS
Power supply
The instrument is powered by 4 batteries 1.5V type R6 alkaline The voltage range ensuring correct operation is from 4.5V to 6.5V Check correct operation by testing the battery (see "Use").
continuity measurements from 0 to ± 10 W and resistance measurements from 0
Limits of working
range
the measurement the measurement
the measurement the measurement
% variation of
the measurement
Typical Max.
the measurement
Consumption is approximately : 300mA for R = 0.5 MW on the MW 500V range 600mA for R = 1 MW on the MW 1000V range (C.A 6513 only) 200mA on continuity on the +10 W and -10 W ranges 10mA on ohmmeter on the 1kW range (C.A 6513 only)
The average service life is : 1000 insulation measurements of 10s on the MW 500 V range for R = 500 kW, 200 insulation measurements of 10s on the MW 1000 V range for R = 1 MW (C.A 6513 only) 1500 continuity measurements of 10s on the 10 range.
18
Electrical safety
Dual insulation instrument in accordance with standard IEC 1010-1, for a set live-earth voltage of 600 V, installation category III, and degree of pollution 2.
Reminders of definitions :
- Installation category : classification of installations in accordance with standardised limits for transient access voltages depending on the nominal voltage of the network in relation to
ground.
- Degree of pollution : classification of micro-environments. See IEC 1010.
- Set voltage : nominal voltage in relation to the earth.
MECHANICAL SPECIFICATIONS
Dimensions : 167 x 106 x 55mm Weight : 500g approx/650g with sheath (batteries included)
19
STANDARDS RESPECTED
Standards Titles
VDE 0413-1 VDE specifications for testing protective devices in electrical installations. VDE 0413-4 Part 1 : insulation testers
VDE 0100 Specifications for the creation of high current installations of nominal NFC 15100 Design, production, checking and maintenance of electrical installations
IEC 1010-1 Safety rules for electronic instruments NFC 42 100 Electrical measurement instruments analogue direct movement panel meters
NF EN 50 081-1 Electromagnetic compatibility. Generic emission standard. NF EN 50 082-1 Electromagnetic compatibility. Generic immunity standard. NF EN 60 555 Disturbances produced in power supply networks by electrical domestic equipment
NF EN 55 014 Limits and measurement methods for the specifications of electric
NF EN 55 022 Limits and measurement methods of the specifications of data processing IEC 801 Electromagnetic compatibility for measurement and control equipment in industrial processes.
Publication 160 Normal atmospheric conditions for tests and measurements NF EN 60 529 Degrees of protection obtained by the case.
IEC 68.2.6 Vibrations
on the frequency range from 10 to 55 Hz. The sweep speed is one octave per minute and the duration of the test is 30 min (in accordance with IEC 1010-1 and NF C 42-
100).
IEC 68.2.27 Shocks
IEC 68.2.29 Shakes
at a rate of 1 shake per second.
IEC 68.2.31 Bumps IEC 68.2.32 Free fall
case (in accordance with IEC 1010-1)
UL 94 Self-extinguishing capability.
Part 4 : ohmmeters voltage less than 1000V. powered by a voltage at most equal to 1000V (rms value) on AC current
and at 1500V on DC current.
and their accessories. Part 1 : Definitions and general regulations Part 7 : Special regulations for multi-function instruments.
Part 1 : Residential, commercial and light industry. Part 1 : Residential, commercial and light industry. and analogous equipment.
Part 2 : Harmonics Part 3 : Voltage fluctuations
domestic appliances, portable tools and similar electric instruments, relating to radioelectric disturbances.
equipment relating to radioelectric disturbances.
Part 2 : Sections relating to electrostatic discharges Level 3 : 8 kV close to the instrument, aptitude criteria B. The instrument holds 15 kV with contact with an aptitude criteria A. Part 3 : Sections relating to radiated electric fields Level 2 : 3 V/m from 27 to 500 MHz not modulated, aptitude criteria A Part 4 : Sections relating to rapid electric transients in salvos Level 2 : 1kV peak, wave 5/50ns, repetition frequency 5 kHz, aptitude criteria B. The instrument holds 4 kV with an aptitude criteria A. Part 5 : Sections relating to electric shocks Level 3 : 2 kV. (not defined in NF EN 55 082-1 at present)
Fundamental climatic tests and mechanical sturdiness The instrument in its operating position withstands vertical vibrations of 0.15 mm amplitude
The instrument withstands vibrations of ± 1 mm amplitude (i.e 2.5 g) at 25 Hz and for 20 min (in accordance with VDE 0413).
The instrument withstands 3 shocks of 15 g for 11 ms in both directions on the 3 axes (in accordance with NFC 42-100).
The instrument withstands 1000 shakes of 10 g for 16 ms in both directions on the 3 axes
The instrument withstands drops of 5 cm. (in accordance with VDE 0413). The instrument which is hand-hold portable, withstands free fall of 1 m with or without the
20
TO ORDER
C.A 6511 Megohmmeter .......................................................................................................................... P01.140201
supplied with 1 set of leads, 1 red crocodile clip, 1 spare fuse and 1 User manual
C.A 6513 Megohmmeter .......................................................................................................................... P01.140301
supplied with 1 set of leads, 1 red crocodile clip, 1 spare fuse and 1 User manual
Spares :
Shockproof case ............................................................................................................................................... P01.298016
Leads 181A/181B ............................................................................................................................................ P03.100819
Red crocodile clip ............................................................................................................................................. P01.101803
Fuse 6.3 x 32 1.6A HBC ............................................................................................................................. P01.297022
Set of 4 batteries 1.5 V LR6 .....................................................................................................................P01.100759
Accessory :
Black crocodile clip .........................................................................................................................................P01.1018.02
WARRANTY
Unless dispensation to the contrary, our instruments are guaranteed from any manufacturing defect or matérial defect. They do not bear the specification known as the safety specification. Our guarantee, which may not under any circumstances exceed the amount of the invoiced price, goes no further than the repair of our faulty equipment, carriage paid to our workshops. It is applicable for normal use of our instruments, and does not apply to damage or destruction caused, notably by error in mounting, mechanical accident, faulty maintenance, defective use, overload or excess voltage, calibration performed by third parties. Our responsability being strictly limited to the pure and simple replacement of the faulty parts of our equipment, the buyer expressly renounces any attempt to find ue responsible for damages or losses caused directly or indirectly.
Our guarantee is applicable for twelve (12) months after the date on which the equipment is made available. The repair, modification or remplacement of a part during the guarantee period will not result in this guarantee being extended.
21
Loading...