Chattanooga Revolution Wireless User Manual

Revolution Wireless
User Manual-EN
REVOLUTION WIRELESS™
QUICK START GUIDE
Note
- For detailed information on usage read the entire user manual.
1. Turn the Remote Control on, by pressing the On/ Off button.
EN
2. Upon activation the screen displays a list that gives you access to the categories of programs.
3. Select a program category and a program within the category by using the navigation pad (up/down).
4. Confirm your choice with the center button.
5. Stick the electrodes on the patient and connect the modules.
6. Turn on the modules, being careful to respect the order of activation of the modules, the turning-on order corresponds to the channel numbering.
REVOLUTION WIRELESS™
QUICK START GUIDE
7. Validate everything by pressing the button below the START symbol.
8. Start the stimulation by increasing the intensities of the channels.
To pause the device, press the center button.
EN
9. At the end of the program press the center button to return to the main menu or press the On/Off button to turn off the device.
REVOLUTION WIRELESS™
EN
CONTENTS
1. How to use the medical equipment (Intended Use) .....................................................................................5
1.1 Treatment environment ............................................................................................................................................................... 5
1.2 Therapy objectives ......................................................................................................................................................................... 5
1.3 Indications .........................................................................................................................................................................................6
1.4 Contraindications ...........................................................................................................................................................................6
1.5 Adverse Eects ................................................................................................................................................................................ 7
2. Safety Information ............................................................................................................................................... 8
3. Description of the Revolution Wireless ......................................................................................................... 18
3.1 Device components and accessories ...................................................................................................................................18
3.2 Explanation of symbols (connections and nameplates) ...........................................................................................20
3.3 Description of the device components ..............................................................................................................................22
3.4 Description of key accessories .............................................................................................................................................24
3.5 Description of key Displays shown ......................................................................................................................................25
4. Device Setup ......................................................................................................................................................... 31
4.1 Smart 4CH Docking station - Connecting the unit, performance check .......................................................... 31
4.2 Charging the Remote Control and the Modules ..........................................................................................................32
4.3 Synchronization between remote and modules ......................................................................................................... 34
5. How to Perform a Treatment, Performance Check ...................................................................................35
5.1 Select a Program ..........................................................................................................................................................................35
5.2 Adjust treatment options ....................................................................................................................................................... 36
5.3 Electrode placement ................................................................................................................................................................. 36
5.4 Patient positioning .................................................................................................................................................................... 38
5.5 Connecting the modules to the electrodes .................................................................................................................... 39
5.6 Starting the treatment ............................................................................................................................................................ 40
5.7 Ending the treatment ............................................................................................................................................................... 42
5.8 Performance Check ................................................................................................................................................................... 42
6. Treatment Options, Functions and Device Settings ..................................................................................43
6.1 Triggering of contraction ......................................................................................................................................................... 43
6.2 Available Programs and Functions .....................................................................................................................................44
6.3 Available Device Settings ........................................................................................................................................................ 48
7. Troubleshooting ..................................................................................................................................................61
7.1 Errors shown on display ............................................................................................................................................................ 61
7.2 Behavior of the Module LEDs ...............................................................................................................................................64
7.3 Behavior of the Docking Station LEDs .............................................................................................................................. 65
7.4 Others ..............................................................................................................................................................................................66
REVOLUTION WIRELESS™
EN
CONTENTS
8. Care, Maintenance, Transport, Environmental Statement ..................................................................... 69
8.1 Care ................................................................................................................................................................................................... 69
8.2 Maintenance .................................................................................................................................................................................70
8.3 Transport ......................................................................................................................................................................................... 71
8.4 Environmental Statement, Expected Life .........................................................................................................................72
9. Technical Data, Standards, Guarantee, Patents ..........................................................................................73
9.1 Technical Data ...............................................................................................................................................................................73
9.2 Standards ....................................................................................................................................................................................... 78
9.3 Guarantee ...................................................................................................................................................................................... 78
9.4 Patents ............................................................................................................................................................................................ 78
10. EMC Tables ......................................................................................................................................................... 79
10.1 Electromagnetic emissions ................................................................................................................................................... 79
10.2 Electromagnetic immunity ..................................................................................................................................................80
10.3 Recommended separation distances .............................................................................................................................. 84
11. Contact ................................................................................................................................................................. 84
REVOLUTION WIRELESS™
. HOW TO USE THE MEDICAL EQUIPMENT INTENDED USE
Note
- This manual is considered as accessory of the therapy unit and therefore it should accompany it at all times.
- The specific instructions provided given here are conditions for the intended use and correct operation of the equipment as well as the safety of the patient and the operator using it.
- Please read the entire manual carefully and section 2 in particular, since information concerning several chapters is only given once, before using your Revolution Wireless device.
1.1 Treatment environment
EN
The Revolution Wireless is a stimulator designed for use by healthcare professionals to provide electrical stimulation treatments in pain management (TENS) as well as for neuromuscular stimulation (NMES).
The Revolution Wireless unit is used in medical and therapeutic treatment environments such as hospitals, clinics, general practices and other health care locations by a skilled clinician. This device is not intended to be used in a home environment.
1.2 Therapy objectives
The Revolution Wireless is a multifunctional electrotherapy unit for the post-surgical and conservative treatment of muscle atrophy as well as pain management.
The following therapy forms are provided by the unit:
• TENS (transcutaneous electrical nerve stimulation) for pain management
• NMES (neuromuscular electrical stimulation)
REVOLUTION WIRELESS™
. HOW TO USE THE MEDICAL EQUIPMENT INTENDED USE
1.3 Indications
The Chattanooga Revolution Wireless is a clinical electrotherapy device intended for use under the supervision of a Healthcare Professional.
Indications for Use:
As an NMES device, indications are for the following conditions:
- Retarding or preventing disuse atrophy
- Maintaining or increasing range of motion
- Re-educating muscles
- Relaxation of muscle spasms
- Increasing local blood circulation
- Prevention of venous thrombosis of the calf muscles immediately after surgery
EN
As a TENS device, indications are for the following conditions:
- Symptomatic relief and management of chronic, intractable pain
- Post-surgical and post-trauma acute pain
As a pulsed-current device, indications are for the following conditions:
- Relaxation of muscle spasm
-Increasing local blood circulation
-Retardation or prevention of disuse atrophy
-Maintenance or increase of range of motion
1.4 Contraindications
Do NOT use the Revolution Wireless under the following conditions:
• Patients with implanted electronic devices. Do not use the device on patients with a cardiac demand pacemaker, implanted defibrillator or other implanted electronic/electrical device.
• Patients with body worn electro mechanical medical devices, i.e., insulin pump.
• Serious arterial circulation problems in lower limbs.
• Abdominal or inguinal hernia.
• For symptomatic local pain relief unless etiology is established or unless a pain syndrome has been diagnosed.
• Do not use chest stimulation on patients with cardiac arrhythmia.
REVOLUTION WIRELESS™
. HOW TO USE THE MEDICAL EQUIPMENT INTENDED USE
Additional Precautions
- Use caution for patients with suspected or diagnosed heart problems.
- Use caution for patients with suspected or diagnosed epilepsy.
- Use caution in the presence of the following:
• When there is a tendency to hemorrhage following acute trauma or fracture.
• Following recent surgical procedures when muscle contraction may disrupt the healing process.
• Over a menstruating or pregnant uterus.
• Over areas of the skin that lack normal sensation.
EN
1.5 Adverse Effects
- Skin irritation and burns beneath the electrodes have been reported with the use of powered muscle
stimulators.
- Potential adverse eects with TENS are skin irritation and electrode burns.
REVOLUTION WIRELESS™
EN
. SAFETY INFORMATION
Definitions
The precautionary instructions found in this section and throughout this manual are indicated by specific symbols. Understand these symbols and their definitions before operating this equipment. The definitions of these symbols are as follows
Danger!
Text with a “DANGER” indicator will explain possible safety infractions that are imminently hazardous situations that would result in death or serious injury.
Warning!
Text with a “WARNING’ indicator explains possible safety infractions that will potentially cause serious injury and equipment damage.
Caution!
Text with a “CAUTION” indicator explains possible safety infractions that have potential to cause minor or moderate injury or damage to the equipment.
Dangerous Voltage
Text with a “Dangerous Voltage” indicator serves to inform the user of possible hazards resulting in the electrical charge delivered in certain program configurations of waveforms.
Explosion Hazard
Text with an “Explosion Hazard” indicator will explain possible safety infractions if this equipment is used in the presence of flammable anesthetics.
BioHazardous materials
Text with a “Biohazard” indicator serves to inform the user of possible hazards resulting in improper handling of components and accessories that have come in contact with bodily fluids
REVOLUTION WIRELESS™
. SAFETY INFORMATION
DANGER!
Explosion Hazard — Revolution Wireless is not designed tor use in areas where an explosion hazard may occur. An explosion hazard may result from the use of flammable anesthetics, oxygen-rich environments, skin cleansing agents and disinfectants.
Handle, clean, and dispose of components and accessories that have come in contact with bodily fluids according to National, Local, and Facility rules, regulations and procedures.
EN
Stimulus delivered by the waveforms of this device, in certain configurations, will deliver a charge of 25 microcoulombs (µC) or greater per pulse and may be sucient to cause electrocution. Electrical current of this magnitude must not flow through the thorax because it may cause a cardiac arrhythmia.
Users with an implanted neurostimulation device must not be treated with or be in close proximity to any shortwave diathermy, therapeutic ultrasound diathermy, or laser diathermy anywhere on their body. Energy from diathermy (shortwave, microwave, ultrasound, and laser) can be transferred through the implanted neurostimulation system, can cause tissue damage, or can result in severe injury or death. Injury, damage, or death can occur during diathermy therapy even if the implanted neurostimulation system is turned o.
The solvents of adhesives and flammable solutions used for cleaning and disinfecting should be allowed to evaporate before the unit is used.
DO NOT connect the unit to an electrical supply without first verifying that the power supply is the correct voltage. Incorrect voltage may cause unit damage, malfunction, electrical shock, fire, or personal injury. Your unit was constructed to operate only at the elctrial voltage specified on the Voltage Rating and Serial Number plate. Contact your DJO dealer if the unit is not properly rated.

REVOLUTION WIRELESS™
. SAFETY INFORMATION
Warning!
Patient hazard –
− U.S.A Federal Law restricts these devices to sale by, or on the order of, a physician or
licensed practitioner. This device should be used only under the continued supervision of a physician or licensed practitioner.
− Be sure to read all instructions for operation before treating a patient.
− Only authorized individuals are allowed to operate the Revolution Wireless. Individuals
are authorized after receiving training in the operation of the unit and reading this operating on manual.
− The long-term eects of chronic electrical stimulation are unknown.
− TENS is not eective for pain of central origin (including headache).
− TENS waveforms have no curative value.
− TENS is a symptomatic treatment, and as such, suppresses the sensation of pain which
would otherwise serve as a protective mechanism.
− Before using the therapy unit, the operator must ascertain that it is in correct working
order and operating condition. The cables and connectors, in particular, must be checked for signs of damage. Damaged parts must be replaced immediately, before use.
− Stop therapy immediately if you have doubts about the device settings and/or the
therapy protocol.
− Patients must be fully conscious while being instructed in the use of the therapy unit and
during therapy.
− Do not treat through clothing.
− Stop treatment immediately if patient experiences discomfort or pain.
− The choice of the therapy parameters to program, of the therapy protocols and electrode
placement is restricted to the responsible physician or therapist. It is the physician’s or therapist’s decision whether or not to use the unit on a specific patient.
− The patient must be familiar with the functions of the Revolution Wireless remote
control with the modules and the remote control must be within easy reach of the patient, allowing them to stop therapy, if needed. Patients unable to operate the emergency stop function (either by stopping on the remote control or by turning o the modules), e.g. paralytic patients, must never be left unattended during therapy.
− Use of accessories other than those specified in this User Manual may increase electrical
emissions, decrease electrical immunity or lead to unsafe conditions.
− The utmost caution is advised under the following conditions. Depending on the
judgement of the responsible physician, the unit may only be applied under supervision and with the parameters defined by the responsible physician. Otherwise the exercise may be too strenuous for the patients with:
EN
1. hypertension (> stage 2), ischemic heart disease and cerebrovascular diseases
2. cardiovascular diseases

REVOLUTION WIRELESS™
. SAFETY INFORMATION
3. Pregnancy. Safety of powered muscle stimulators for use during pregnancy has not been
established.
4. under 16 years of age
− Never apply the electrodes as follows:
• Stimulation should not be applied transcerebrally, or on the eyes.
• Stimulation should not be applied over the anterior neck or mouth. Severe spasm of the
laryngeal and pharyngeal muscles may occur and the contractions may be strong enough to close the airway or cause difficult in breathing.
• Counter-laterally, i.e. do not use two poles connected to the same channel on
opposite sides of the body.
• Stimulation should not be applied over swollen, infected, or inflamed areas or skin
eruptions, e.g., phlebitis, thrombophlebitis, varicose veins, etc.
• Stimulation should not be applied transthoracically in that the introduction of electrical
current into the heart may cause cardiac arrhythmia.
• Stimulation should not be applied over the carotid sinus nerve particularly in patients
with a known sensitivity to the carotid sinus reflex.
• Stimulation should not be applied over, or in proximity to, cancerous lesions.
• If the person is pregnant or menstruating do not place electrodes directly on the uterus
area or connect pairs of electrodes on either side of the abdomen to avoid any risk for the mother and/or the baby.
• Keep electrodes separated during treatment. Electrodes in contact with each other could
result in improper stimulation or skin burns
EN
Warning!
- Extreme caution should be taken when in use around small children and babies!
Sucient distance to the device and its accessories is mandatory for their safety!
- Never leave the device unattended when it is switched on! Switch the device o and
disconnect the electrodes from the modules!
- After use, store the device in a safe place to avoid other people not informed to use the
device!
- This device is not a toy but a medical device that misunderstanding use can cause
damages!

REVOLUTION WIRELESS™
. SAFETY INFORMATION
Warning!
Shock hazard — Strictly observe the following warnings. Failure to do so could endanger
the lives of the patient, the user and other persons involved.
Before use allow the Revolution Wireless to reach room temperature. If the unit
has been transported at temperatures below 0 °C (32°F), leave it to reach at room temperature for about 2 hours, until any condensation has disappeared.
Electrosurgical equipment or defibrillators. Disconnect the electrodes from the device
before using electrosurgical equipment, or a defibrillator, to avoid cutaneous burns from the electrodes and destroying the device.
Electronic surveillance equipment. Do not apply stimulation near electronic surveillance
equipment (e.g. cardiac monitors, ECG alarms), as there is a risk they may not work properly whilst the electrical stimulation device is being used.
Electromagnetic radiation. Do not use the stimulator in areas in which unprotected
devices are used to emit electromagnetic radiation. Portablecommunications equipment can interfere with the device.
Cancer. Do not apply stimulation if you have progressive cancer or near any cancerous
tumour. The increased metabolism, caused by certain modes of stimulation, is likely to encourage cancer cells to spread
Muscle shortening. During the muscular contraction phase it is recommended to hold
the extremities of the stimulated limbs to avoid any shortening of the muscle during contraction, which could cause cramps.
Contralateral stimulation. Do not use two terminals connected to the same channel
on opposite segments of the body (for example, a positive terminal on the left arm and a negative terminal on the right arm).
Loss of sensation. Proceed with caution if stimulation is applied to areas of the skin
whose level of sensation is lower than normal. Do not apply stimulation to a person who cannot express themselves.
Battery leakage. If there is leak from a component, take steps to ensure the liquid does
not come into contact with skin or eyes. Should this occur, wash the aected area with water and consult a doctor.
Strangulation. Do not wind cables around the neck. Tangled cables can cause
strangulation.
Post-surgery. Proceed with caution after recent surgery.
Accessibility of the power adaptor. The plug socket must be close to the power adaptor
and be easily accessible.
Internal bleeding. Proceed with caution if you are prone to internal bleeding; for
example, after an injury or a fracture.
EN

REVOLUTION WIRELESS™
. SAFETY INFORMATION
The Revolution Wireless must only be operated in dry rooms.
Do not use the Revolution Wireless in water or in a humid atmosphere (sauna, Bath,
Shower etc.) that would cause electronic failure.
Water protection. The unit is not protected from the ingress of water
− When connecting the unit to other equipment or when creating a medical system, check
that the sum of leakage currents will not cause any hazard. Please contact DJO GLOBAL if you have questions regarding this matter.
− No modification of this equipment is allowed.
− Do not open the product and its accessories as there is risk of electrocution
− Before cleaning and service interventions, turn the remote control and the modules o
and disconnect the tablet from the power line by removing the power cord from the wall outlet.
Liquids and foreign material (such as dust, metal etc.) must not be allowed to
enter the remote control, the modules, the charging tablet or the power supply. If such material has entered into the units, it must be immediately checked by a service technician, before it can be reused.
Electricity supply. Never connect the stimulation cables to an external power supply, as
there is a risk of electrocution
− Do not apply stimulation near the area of an implant, such as cochlear implants,
pacemakers, skeletal anchorage or electric implants. This could cause an electrical shock, burns, electrical interference or death
− Never use the Revolution Wireless or the AC adaptor if it is damaged or open. There is a
risk of electric shock.
− Disconnect the AC adaptor immediately if there is abnormal heating or smell, or if smoke
comes from the AC adaptor or the device.
− Do not place the docking station in a confined space (carrying case, drawer etc.) when
charging the device. There is a risk of electrocution
EN

REVOLUTION WIRELESS™
. SAFETY INFORMATION
Warning!
Equipment malfunction — these warnings refer to potential equipment mailfunctions that
could result in patient hazards
− Magnetic and electrical fields are capable of interfering with the proper performance
of the unit. For this reason make sure that all external devices operated in the vicinity of the unit comply with the relevant EMC requirements. ¬ Devices that intentionally radiate electromagnetic energy, such as X-ray equipment, MRI devices, ¬radio systems and cell phones are possible sources of interference, as well as RFID equipment (e.g. electromagnetic security systems).
Keep the unit away from such equipment and verify its performance before use. For
additional information, refer to the EMC section of this Manual.
− Do not use the Revolution Wireless within 3 feet of short wave or microwave devices as
this could alter the currents generated by the stimulator. If you are in any doubt as to the use of the stimulator in close proximity to another medical device, seek advice from the manufacturer of the latter or from your doctor.
− Exercise caution when using electrotherapy while the patient is connected to monitoring
equipment with electrodes attached to the body. Stimulation could disrupt the signals sent to the monitoring equipment.
− Refer repair and maintenance to authorized persons. Persons are authorized after
training by a specialist trained and commissioned by the manufacturer.
− Do not disassemble or modify the Revolution Wireless or its accessories. This may cause
damage, malfunction, electrical shock, fire, or personal injury.
− Inspect the Revolution Wireless and it’s accessories for damage and loose connections
at least once a year. Damaged and worn parts must be immediately replaced with original spare parts by authorized sta.
- Do not use the device if you are connected to a high-frequency surgical instrument as
this could cause skin irritation or burns under the electrodes
EN

REVOLUTION WIRELESS™
. SAFETY INFORMATION
Caution!
Read, understand, and practice the precautionary and operating instructions found in this manual. Know the limitations and hazards associated with using electrical stimulation. Observe any precautionary and operational decals placed on the unit.
Patient hazard — these cautions need to be observed to avoid the risk of electrical shock or other negative eects to the patient.
− The device is designed to comply with electromagnetic safety standards. This
equipment generates, uses, and can radiate radiofrequency energy and, if not installed and used in accordance with instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee that interference will not occur in a particular installation. Harmful interference to other devices can be determined by turning this equipment on and o. Try to correct the interference by using one or more of the following:
− Reorient or relocate the receiving device
− Increase the separation between the equipment (refer to the EMC section in this
manual for recommended separation distances)
− Connect the equipment to an outlet on a dierent circuit.
− Consult your DJO dealer for assistance.
− Do not apply stimulation close to metal. Remove jewelry, piercings, belt buckles or any
other metallic product or device in the area of stimulation.
− Be careful if the patient has sensitivity problems or is not able to communicate that he
or she feels discomfort, however light.
− Never begin an initial stimulation session on a person who is standing. The first five
minutes of stimulation must always be performed on a person who is sitting or lying down. In rare instances, people of a nervous disposition may experience a vasovagal reaction. This is of psychological origin and is connected with a fear of the muscle stimulation as well as surprise at seeing one of their muscles contract without having intentionally contracted it themselves. A vasovagal reaction causes heart to slow down and blood pressure to drop, which produces a feeling weakness and a tendency towards fainting. If this does occur, all that is required is to stop the stimulation and for the person to lie down with the legs raised until the feeling of weakness disappears (5 to 10 minutes)
− Never allow muscular contraction during a stimulation session to result in movement.
You should always stimulate isometrically; this means that the extremities of the limb in which a muscle is being stimulated must be firmly fixed, so as to prevent any movement that results from contraction.
EN

REVOLUTION WIRELESS™
. SAFETY INFORMATION
− Do not disconnect any module that is switched on during the stimulation session. They
must be switched o first.
− Do not use the stimulator while driving, operating machinery, or during any activity in
which involuntary muscle contractions may put the user at undue risk of injury. Do not apply stimulation during sleep.
− Do not use the stimulator at altitudes of over 9,800 feet.
− Always turn o the stimulator before moving or removing any electrodes during a
session, to avoid electrical shock to the patient.
− Do not try to place electrodes on a body part not directly visible without assistance.
− Attach the electrodes in such a way that their entire surface is in contact with the skin.
− For reasons of hygiene, each patient must have their own set of electrodes. Do not use
the same electrodes on dierent patients.
− Contaminated electrodes or gel can lead to infection.
− Use of electrodes with degraded hydrogel can result in burns to the skin.
Size of electrodes. Do not use electrodes with an active area of less than 16 cm² due to
the risk of associated burning. Proceed systematically with caution when the density of the current is over 2 mA/cm².
− Do not place the electrodes in water.
− Do not apply solvents of any kind to the electrodes.
- Skin irritation. Some patients may experience skin irritation or hypersensitivity due to
the electrical stimulation or electrical conductive medium. The irritation can usually be reduced by using an alternate conductive medium, or alternate electrode placement. If redness of the skin is observed under the electrode, do not start another stimulation session on the same area if the redness is still visible.
- Instructions for electrodes. See the usage and storage instructions displayed on the bag
of electrodes.
EN

REVOLUTION WIRELESS™
. SAFETY INFORMATION
Caution!
Equipment damage –
- Do not let the modules get in contact with massage oils or any other products of the
same type that could damaged the device or its accessories.
- Check that the voltage and frequency ratings of your local power line are those indicated
on the type plate of the power supply.
− Do not expose the Revolution Wireless to direct sunlight, because some of the
components may reach unacceptably high temperatures.
− The unit is not protected from the ingress of water droplets from rain if used outside the
carrying case
− The presence of children, pets and vermin does not normally aect the proper
functioning. However, make sure that these sources do not contaminate the physiotherapy unit and keep them away from it. Also, keep the unit clean and protect it from dust and lint. The safety rules and regulations set forth apply in any case.
− It is recommended to use the transport bag that comes with the unit, for transport of the
device, and to use a proper transport box to ship it.
− Always use the AC adaptor (power supply) provided by the manufacturer to recharge the
unit.
− Do not store the modules and remote control for a long time with uncharged batteries.
− Only use electrodes supplied by the manufacturer. Other electrodes may have electrical
properties that are unsuitable for or may damage the Revolution Wireless.
EN
Note
- For best results and device safety, wash and clean the skin of any oil and dry it before attaching the electrodes.

REVOLUTION WIRELESS™
.DESCRIPTION OF THE REVOLUTION WIRELESS
3.1 Device components and accessories
Model: Revolution Wireless 4CH Part number: 2544660
Your kit contains (included in delivery):
Quantity Description
1 remote control
4 stimulation modules
EN
1 smart 4 CH docking station
1 AC adaptor
1 USB cable
2 bags of small electrodes (5x5 cm 1 snap connection)
2 bags of large electrodes (5x10 cm 1 snap connection)
2 bags of large electrodes (5x10 cm 2 snap connections)
1
1 quick start guide
1 carrying case
1 lanyard
1 white protection sleeve
CHATT REVOLUTION WIRELESS USB STICK ASSEMBLY with User manual and Quick Start Guide
1 Set of colored module clips

REVOLUTION WIRELESS™
. DESCRIPTION OF THE WIRELESS PROFESSIONAL
The following parts can be re-ordered if needed
EN
Re-Order Part Number
101140 remote control
101141 stimulation modules 4
101142
101143
649028 AC adaptor
42204 4 small electrodes (5x5 cm 1 snap connection)
42203 2 large electrodes (5x10 cm 2 snap connections)
42223 2 large electrodes (5x10 cm 1 snap connection)
680050 carrying case
5529040 white protection sleeve
Description
smart 4 CH docking station including tablet
5529220 Set of colored module clips

20xx
20xx
LATEX
FREE
20xx
20xx
20xx
20xx
REVOLUTION WIRELESS™
.DESCRIPTION OF THE REVOLUTION WIRELESS
3.2 Explanation of symbols (connections and nameplates)
3.2.1 Symbols on Remote Control and Modules
Read the user manual or operating instructions
Caution! Observe warnings set forth in operation manual!
EN
REF
20xx
The Revolution Wireless is a class II device with internal electric power and type BF applied parts.
The power switch On/OFF button is a multi-function button.
The name and address next to this factory symbol is the manufacturer. The date is the manufacturing date.
The number next to this symbol is the article reference number
The number next to this symbol is the serial number
Indicates separate treatment from general waste at end of life.
Keep dry
Non-ionizing radiation
Keep away from direct sunlight

20xx
LATEX
FREE
20xx
20xx
20xx
REVOLUTION WIRELESS™
.DESCRIPTION OF THE REVOLUTION WIRELESS
3.2.2 Symbols on Charging Tablet/Docking Station and AC Power Supply
Alternating current input on AC power supply
Direct current output from power supply
EN
REF
20xx
Protection class II equipment. The AC Power Supply device has double insulation.
Type BF applied part
The name and address next to this factory symbol is the manufacturer. The date is the manufacturing date.
The number next to this symbol is the article reference number (Artn.)
Read the user manual or operating instructions
Caution! Observe warnings set forth in operation manual!
The number next to this symbol is the serial number
Do not dispose with unsorted municipal waste
Keep dry (not on type plates)
Keep away from direct sunlight

REVOLUTION WIRELESS™
.DESCRIPTION OF THE REVOLUTION WIRELESS
3.3 Description of the device components
3.3.1 Remote Control
A
EN
B
C
D
E
A. On/O button (press briefly to switch on, press and hold for more than 2 seconds to switch o, while
browsing the lists press briefly to return to the main menu)
B. 2 multifunction buttons:
• Functions related to icons are located on the screen (e.g. info, main menu, placement of electrodes, etc.)
• Selection of stimulation channel to increase or decrease the intensity level of stimulation
C. Navigation pad to scroll up/down or next/back during programming, or to increase/decrease the
intensity level of stimulation
D. Validation of pause button during stimulation, confirm a selection during program selection or change
a setting during programming
E. Port for the USB cable or the docking station connector
Note
Emergency stop function: By pressing the central button or the On/O button on one of the
modules during stimulation, the device pauses.

REVOLUTION WIRELESS™
.DESCRIPTION OF THE REVOLUTION WIRELESS
3.3.2 Modules
EN
A
C
B
1 module is composed of two pods. A On/O button (press briefly to switch on, press for 1 second to switch o, during stimulation press to pause)
• Flashing green LED: ready
• Flashing yellow LED: stimulation on
• Flashing red LED: battery empty
• Flashing red/green LED: no connection to remote control
B Groove to wind up the cable C Pod containing battery
Note
- When the distance among the remote control and the modules is too big, they will lose the connection, stop with the stimulation immediately and the LEDs will flash red and green.
- Emergency stop function: By pressing the central button or the On/O button on one of the modules during stimulation, the device pauses.

REVOLUTION WIRELESS™
.DESCRIPTION OF THE REVOLUTION WIRELESS
3.4 Description of key accessories
3.4.1 Smart 4CH docking station and Removeable tablet
EN
A
B
E
DOCKING STATION A Removable tablet B Connector to charge the remote control C Docking bay to position the modules to be recharged D Port for the AC adapter and for the USB cable connected to the front of the docking station E Storage bin
C
D

REVOLUTION WIRELESS™
.DESCRIPTION OF THE REVOLUTION WIRELESS
3.5 Description of key Displays shown
3.5.1 Display in Program Category selection mode (Home Screen)
A
B
EN
A Header showing date, time and battery status B Current chosen program category (marked with a blue focus, displayed bigger)
Loading...
+ 61 hidden pages