CHATTANOOGA Intelect Legend XT User manual

© 2006 Encore Medical, L.P.
User Manual
Operation & Installation
Instructions for:
Therapy Systems
2791K- Two Channel Combination System
2795K- Four Channel Combination System
2793K- Two Channel Electrotherapy System
Optional Equipment
2780- Therapy System Cart
2767- NiMH Battery Module
27508 and 27079- User Remote Controls
2781- Channel 3/4 Electrotherapy Module
TABLE OF CONTENTS
Intelect® Legend XT Therapy System
FOREWORD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
PRODUCT DESCRIPTION
ABOUT INTELECT LEGEND XT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
STANDARD PRECAUTIONARY INSTRUCTIONS
Caution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Warning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Danger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Dangerous Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Corrosive. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Spontaneous Combustion. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
Biohazardous Materials. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
CAUTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
WARNINGS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
DANGERS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
ELECTROTHERAPY INDICATIONS, CONTRAINDICATIONS, AND ADVERSE EFFECTS
Indications for Russian, TENS, High Voltage Pulsed Current
(HVPC), Interferential and Premodulated waveforms. . . . . . . . . . . . 7
Additional Indications for Microcurrent, Interferential,
Premodulated, and TENS waveforms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Contraindications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Additional Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Adverse Eff ects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
ULTRASOUND INDICATIONS AND CONTRAINDICATIONS. . . . 9
Indications for Ultrasound. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Contraindications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Additional Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
NOMENCLATURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
. . . . . . . . . . . . . . . 2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
INTELECT LEGEND XT ELECTROTHERAPY AND COMBINATION THERAPY SYSTEMS
Two Channel Electrotherapy System . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Two Channel Combination System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Front Access Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Rear Access Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
USER INTERFACE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
SYMBOL DEFINITIONS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
System Hardware Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
System Software Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
Optional Module and Accessory Symbols. . . . . . . . . . . . . . . . . . . . . . .13
Operator Remote . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
Battery Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Channel 3/4 Electrotherapy Module . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
GENERAL TERMINOLOGY. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
ULTRASOUND . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
SPECIFICATIONS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
SYSTEM SPECIFICATIONS
DIMENSIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Width . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Standard Weight. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Power (Combination and Electrotherapy Units). . . . . . . . . . . . . . . . .15
Electrical Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Regulatory Compliance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
WAVEFORM SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
IFC- Interferential (Traditional 4 Pole) . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Russian. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
TENS- Symmetrical Biphasic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Microcurrent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
. . . . . . . . . . . . . . . . . . . . 10
i
TABLE OF CONTENTS
Intelect® Legend XT Therapy System
Premodulated (Traditional 2 Pole IFC). . . . . . . . . . . . . . . . . . . . . . . . . . .18
High Voltage Pulsed Current (HVPC) . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
ULTRASOUND SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Ultrasound. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
SETUP. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
INTELECT LEGEND XT THERAPY SYSTEMS
THERAPY SYSTEM SETUP. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Clinic Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
Restoring Default Protocols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
Restoring Default Unit Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
Erasing Patient Data Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Setting Date and Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Setting System Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Displaying Unit Version Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Pad Contact Quality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
Selecting Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
Connecting Accessories to the Therapy System. . . . . . . . . . . . . . . . .26
PATIENT PREPARATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
ELECTROTHERAPY PATIENT PREPARATION
Electrode Placement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
DURA-STICK™ Electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
Reusable Carbon Electrodes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
DURA-STICK™ Electrode Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . .29
Connecting Lead Wires . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
Securing Electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
Reusable Carbon Electrodes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Connecting Lead Wires . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
Conductive Medium . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
Securing Electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
. . . . . . . . . . . . . . . . . . 20
. . . . . . . . . . . . . . . . . 27
ULTRASOUND PATIENT PREPARATION. . . . . . . . . . . . . . . . . . . . . . 31
Preparing Treatment Area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
Size of Applicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
Applicator Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
Conductive Medium . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
Treatment Area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
Applicator LED . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
OPERATOR INTERFACE
HOME SCREEN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
ELECTROTHERAPY SCREEN. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34
GENERAL ELECTROTHERAPY WAVEFORM SETUP . . . . . . . . . . .35
Prepare Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
Select Modality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
Select Waveform. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
Edit Waveform Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
Installing Patient Interrupt Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
Patient Interrupt Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
Setting Waveform Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
Intensity Knob Rotation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
Start Treatment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
Pause Treatment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
Stop Treatment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
Save to Patient Data Card. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
ADJUSTING ELECTROTHERAPY CHANNEL PARAMETERS DURING TREATMENT
Selecting Channel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38
Editing Channel Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38
ULTRASOUND . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
ii
TABLE OF CONTENTS
Intelect® Legend XT Therapy System
Preparing Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
Selecting Modality. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
Editing Ultrasound Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
Head Warming. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
Setting Ultrasound Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40
Intensity Knob Rotation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40
Starting Treatment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40
Pausing Treatment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40
Stopping Treatment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41
Saving to Patient Data Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41
Editing Ultrasound from Home Screen. . . . . . . . . . . . . . . . . . . . . . . . . .42
Editing Ultrasound from Treatment Review Screen . . . . . . . . . . . . .42
COMBINATION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Preparing Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
Selecting Modality. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
Accessing Combination Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Editing Ultrasound Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .44
Selecting Waveform . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .44
Using the Patient Interrupt Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .44
Editing Waveform Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45
Setting Waveform Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45
Intensity Knob Rotation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45
Setting Ultrasound Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45
Intensity Knob Rotation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45
Starting Treatment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .46
Pausing Treatment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .46
Stopping Treatment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .46
Saving to Patient Data Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .46
ADJUSTING COMBINATION PARAMETERS DURING TREATMENT
Editing Waveform Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47
Editing Ultrasound Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47
PATIENT DATA CARD SETTING UP A NEW CARD . . . . . . . . . . . . 48
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48
Inserting New Patient Data Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48
Setting up Treatment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48
Setting up a New Patient Data Card . . . . . . . . . . . . . . . . . . . . . . . . . . . .48
Entering Patient ID. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49
Accessing Electrode Placement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50
Setting up Electrodes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50
Electrode Placement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50
Accessing Pain Map. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51
Selecting Pain Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51
Adding Pain Locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51
Selecting Location of Pain . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52
Editing Pain Locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52
Deleting Pain Locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53
Pain Scales . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53
Selecting Pain Scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53
Adjusting Pain Scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53
Saving to Patient Data Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54
EXISTING PATIENT DATA CARD USE . . . . . . . . . . . . . . . . . . . . . . . . .55
Inserting Existing Patient Data Card. . . . . . . . . . . . . . . . . . . . . . . . . . . . .55
Accessing Patient Data Card. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55
Viewing Patient Data Card. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55
Starting a New Treatment from Patient Data Card . . . . . . . . . . . . . .56
Using the Patient Interrupt Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
iii
TABLE OF CONTENTS
Intelect® Legend XT Therapy System
Setting Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56
Intensity Knob Rotation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56
Starting Treatment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
Pausing Treatment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
Erasing Patient Data Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
Stopping Treatment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
CREATING USER PROTOCOLS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58
Selecting Modality. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58
Editing Modality Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Selecting Clinical Resources Library. . . . . . . . . . . . . . . . . . . . . . . . . . . . .58
Entering User Protocol Name. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .59
DELETING USER PROTOCOLS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60
Selecting Clinical Resources Library. . . . . . . . . . . . . . . . . . . . . . . . . . . . .60
Selecting User Protocol to Delete . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Deleting User Protocol. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60
USING USER PROTOCOLS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Accessing User Protocols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61
Selecting User Protocol. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61
Preparing Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61
Editing Modality Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Using the Patient Interrupt Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .62
Setting Modality Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .62
Intensity Knob Rotation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .62
Starting Treatment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63
Pausing Treatment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63
Stopping Treatment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63
Saving to Patient Data Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63
INSTALLATION/REMOVAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
INSTALLING THE CHANNEL 3/4 ELECTROTHERAPY AND NiMH BATTERY MODULE
Nomenclature- Channel 3/4 Electrotherapy Module. . . . . . . . . . . .65
Specifi cations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66
Waveform & Current Specifi cations . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66
Disconnecting Mains Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Removing Lead Wires and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . .67
Removing Therapy System from Cart . . . . . . . . . . . . . . . . . . . . . . . . . . .67
Releasing Ribbon Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68
Positioning Therapy System and Module . . . . . . . . . . . . . . . . . . . . . . .68
Connecting Ribbon Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Setting Therapy System onto Module . . . . . . . . . . . . . . . . . . . . . . . . . .69
Securing Therapy System to Module . . . . . . . . . . . . . . . . . . . . . . . . . . .69
Attaching the Lanyard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69
Installing Lead Wires and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . .70
Installing Front Access Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70
Mounting to Therapy System Cart. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70
Connecting Mains Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70
Turning Therapy System On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .71
REMOVING THE CHANNEL 3/4 ELECTROTHERAPY AND NiMH BATTERY MODULE
Disconnecting Mains Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Removing Lead Wires and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . .72
Removing Therapy System from Cart . . . . . . . . . . . . . . . . . . . . . . . . . . .72
Removing Screws Securing Module . . . . . . . . . . . . . . . . . . . . . . . . . . . .73
Disconnecting Ribbon Cable at Module. . . . . . . . . . . . . . . . . . . . . . . . . 73
Storing and Securing Ribbon Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . .73
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
iv
TABLE OF CONTENTS
Intelect® Legend XT Therapy System
Attaching the Lanyard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .74
Installing Lead Wires and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . .74
Connecting Mains Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .74
Turning Therapy System On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .75
GENERAL INFORMATION OPERATOR REMOTE . . . . . . . . . . . . . 76
Installing the Operator Remote . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .76
GENERAL INFORMATION THERAPY SYSTEM CART . . . . . . . . .77
Nomenclature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .77
Specifi cations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .77
MOUNTING THERAPY SYSTEM TO THERAPY SYSTEM CART 78
Assembling the Therapy System Cart . . . . . . . . . . . . . . . . . . . . . . . . . . .78
Preparing the Therapy System Cart . . . . . . . . . . . . . . . . . . . . . . . . . . . . .78
Mounting Therapy System to Cart. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .78
Connecting Mains Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .79
Installing Storage Bins. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .79
Removing System from Therapy System Cart . . . . . . . . . . . . . . . . . . .79
OPTION OPERATION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
USING THE OPERATOR REMOTE
Nomenclature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .80
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Storing the Operator Remote . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .81
USING THE THERAPY SYSTEM CART. . . . . . . . . . . . . . . . . . . . . . . . . 82
Nomenclature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .82
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
USING THE NiMH BATTERY MODULE. . . . . . . . . . . . . . . . . . . . . . . . 83
Nomenclature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .83
CHARGING BATTERY MODULE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
When to Recharge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Charging Temperature. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .84
BATTERY MODULE SERVICE LIFE . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
STORING THE BATTERY MODULE. . . . . . . . . . . . . . . . . . . . . . . . . . . .85
Short Term Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Long Term Storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .85
TROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
ERROR CODES
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .86
REPLACEMENT ACCESSORIES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
GENERAL INFORMATION
MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
CARING FOR THE THERAPY SYSTEM
Cleaning the Therapy System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .91
Cleaning the Lens. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .91
CALIBRATION REQUIREMENTS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Calibrating Ultrasound Applicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . .91
FACTORY SERVICE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
WARRANTY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .86
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
. . . . . . . . . . . . . . . . . . . . . . . .91
v
FOREWORD
This manual has been written for the users of the Intelect Legend XT Therapy Systems. It contains general information on the operation, precautionary practices, and maintenance information. In order to maximize use, efficiency, and the life of the system, please read this manual thoroughly and become familiar with the controls, as well as the accessories before operating the system.
This manual contains general safety, operating, maintenance, and care instructions as well as installation instructions for the optional Therapy System Cart, Channel 3/4 Electrotherapy and NiMH Battery for the users of the Intelect Legend XT Therapy two channel electrotherapy and combination systems.
Specifications put forth in this manual were in effect at the time of publication. However, owing to Chattanooga Group's policy of continual improvement, changes to these specifications may be made at any time without obligation on the part of Chattanooga Group.
Before administering any treatment to a patient, the users of this equipment should read, understand and follow the information contained in this manual for each mode of treatment available, as well as the indications, contraindications, warnings and precautions. Consult other resources for additional information regarding the application of electrotherapy and ultrasound.
PRODUCT DESCRIPTION
The Intelect Legend XT Therapy Systems are two channel electrotherapy and combination systems with the option of adding additional channels of electrotherapy by installation of the optional Channel 3/4 Electrotherapy Module. Other optional modality modules are available for separate purchase and may be installed by the end user.
Stay current with the latest clinical developments in the field of electrotherapy and ultrasound. Observe all applicable precautionary measures for treatment.
Keep informed of appropriate indications and contraindications for the use of electrotherapy and ultrasound. This equipment is to be used only under the prescription and supervision of a licensed practitioner.
Intelect® Legend XT Therapy System
©2006 Encore Medical Corporation or its affiliates, Austin, Texas, USA. Any use of editorial, pictorial, or layout composition of this publication without express written consent from Chattanooga Group of Encore Medical, L.P. is strictly prohibited. This publication was written, illustrated, and prepared for distribution by Chattanooga Group of Encore Medical, L.P.
1
ABOUT INTELECT LEGEND XT
STANDARD PRECAUTIONARY INSTRUCTIONS
The precautionary instructions found in this section and throughout this manual are indicated by specific symbols. Understand these symbols and their definitions before operating this equipment. The definition of these symbols are as follows:
Dangerous Voltage
Text with a “Dangerous Voltage” indicator serves to inform the user of possible hazards resulting in the electrical
Caution
Text with a “CAUTION” indicator will explain possible safety infractions that could have the potential to cause minor to moderate injury or damage to equipment.
Warning
Text with a “WARNING” indicator will explain possible safety infractions that will potentially cause serious injury and equipment damage.
Danger
Text with a “DANGER” indicator will explain possible safety infractions that are imminently hazardous situations that would result in death or serious injury.
charge delivered to the patient in certain treatment configurations of TENS waveforms.
Corrosive
Text with a “CORROSIVE" indicator will explain possible safety infractions if the chemical components of the battery are exposed to air, skin, or other materials.
Spontaneous Combustion
Text with a “SPONTANEOUS COMBUSTION" indicator will explain possible safety infractions that could create conditions for a Spontaneous Combustion if the material is mishandled and not disposed of properly.
Biohazardous Materials
Text with a “BIOHAZARD” indicator serves to inform the user of possible hazards resulting in improper handling of components and accessories that have come in contact with bodily fluids.
NOTE:
Throughout this manual, “NOTE” may be found. These
Notes are helpful information to aid in the particular area or function being described.
Intelect® Legend XT Therapy System
2
ABOUT INTELECT LEGEND XT
CAUTIONS
Read, understand, and practice the precautionary and operating
instructions. Know the limitations and hazards associated with using any electrical stimulation or ultrasound device. Observe the precautionary and operational decals placed on the unit.
DO NOT operate the Intelect Legend XT Therapy System when
connected to any unit other than Chattanooga Group devices. DO NOT operate this unit in an environment where other devices
are being used that intentionally radiate electromagnetic energy in an unshielded manner.
Ultrasound should be routinely checked before each use to
determine that all controls function normally, especially that the intensity control does properly adjust the intensity of the ultrasonic power output in a stable manner. Also, determine that the treatment time control does actually terminate ultrasonic power output when the timer reaches zero.
DO NOT use sharp objects such as a pencil point or ballpoint pen to
operate the buttons on the control panel. This unit should be operated, transported and stored in
temperatures between 59° F and 104° F (15° C and 40° C), with Relative Humidity ranging from 30%-60%.
Handle Ultrasound Applicator with care. Inappropriate handling of
the Ultrasound Applicator may adversely affect its characteristics. Before each use, inspect Ultrasound Applicator for cracks, which
may allow the ingress of conductive fluid.
Intelect® Legend XT Therapy System
Inspect Applicator cables and associated connectors before
each use. The Intelect Legend XT Therapy System is not designed to prevent
the ingress of water or liquids. Ingress of water or liquids could cause malfunction of internal components of the system and therefore create a risk of injury to the patient.
This equipment generates, uses and can radiate radio frequency
energy and, if not installed and used in accordance with the instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee that interference will not occur in a particular installation. Harmful interference to other devices can be determined by turning this equipment on and off. Try to correct the interference using one or more of the following: reorient or relocate the receiving device, increase the separation between the equipment, connect the equipment to an outlet on a different circuit from that to which the other device(s) are connected and consult the factory field service technician for help.
The Nylatex® Wraps shipped with this unit contain dry natural rubber
and may cause allergic reactions in patients with allergies to latex.
3
ABOUT INTELECT LEGEND XT
WARNINGS
U.S.A. Federal Law restricts these devices to sale by, or on the order
of, a physician or licensed practitioner. This device should be used only under the continued supervision of a physician or licensed practitioner.
For continued protection against fire hazard, replace fuses only with ones of the same type and rating.
Make certain the unit is electrically grounded by connecting only to a grounded electrical service receptacle conforming to the applicable national and local electrical codes.
Care must be taken when operating this equipment around other equipment. Potential electromagnetic or other interference could occur to this or to the other equipment. Try to minimize this interference by not using other equipment in conjunction with it.
The safety of TENS waveforms for use during pregnancy or birth has not been established.
TENS is not effective for pain of central origin. (This includes headache.)
TENS should be used only under the continued supervision of a physician or licensed practitioner.
TENS waveforms have no curative value.
TENS is a symptomatic treatment, and as such, suppresses the sensation of pain which would otherwise serve as a protective mechanism.
The user must keep the device out of the reach of children.
Electronic monitoring equipment (such as ECG monitors and ECG alarms) may not operate properly when TENS stimulation is in use.
Intelect® Legend XT Therapy System
Powered muscle stimulators should be used only with the leads and
electrodes recommended for use by the manufacturer. In the event that an Error message or Warning appears beginning
with a 2 or 3, immediately stop all use of the system and contact the dealer or Chattanooga Group for service. Errors and Warnings in these categories indicate an internal problem with the system that must be tested by Chattanooga Group or a Field Service Technician certified by Chattanooga Group before any further operation or use of the system. Use of a system that indicates an Error or Warning in these categories may pose a risk of injury to the patient, user or cause extensive internal damage to the system.
Use of controls or adjustments or performance of procedures other
than those specified herein may result in hazardous exposure to ultrasonic energy.
Before administering any treatment to a patient you should become
acquainted with the operating procedures for each mode of treatment available, as well as the indications, contraindications, warnings, and precautions. Consult other resources for additional information regarding the application of Electrotherapy and Ultrasound.
To prevent electrical shock, disconnect the unit from the power
source before attempting any maintenance procedures. Keep electrodes separated during treatment. Electrodes in contact
with each other could result in improper stimulation or skin burns. Long term effects of chronic electrical stimulation are unknown.
4
ABOUT INTELECT LEGEND XT
WARNINGS (continued)
Stimulation should not be applied over the anterior neck or mouth.
Severe spasm of the laryngeal and pharyngeal muscles may occur and the contractions may be strong enough to close the airway or cause difficulty in breathing.
Stimulation should not be applied transthoracically in that the
introduction of electrical current into the heart may cause cardiac arrhythmia.
Stimulation should not be applied over swollen, infected, and
inflamed areas or skin eruptions, e.g., phlebitis, thrombophlebitis, varicose veins, etc.
Stimulation should not be applied over, or in proximity to, cancerous
lesions. Output current density is related to electrode size. Improper
application may result in patient injury. If any question arises as to the proper electrode size, consult a licensed practitioner prior to therapy session.
The Intelect Legend XT Therapy System optional modules and associated
accessories are designed for use only with the Chattanooga Group Intelect Legend XT Electrotherapy and Combination Therapy Systems.
Intelect® Legend XT Therapy System
5
ABOUT INTELECT LEGEND XT
DANGERS
Stimulus delivered by the TENS waveforms of this
device, in certain configurations, will deliver a charge of 25 microcoulombs (μC) or greater per pulse and may be sufficient to cause electrocution. Electrical current of this magnitude must not flow through the thorax because it may cause a cardiac arrhythmia.
Patients with an implanted neurostimulation device
must not be treated with or be in close proximity to any shortwave diathermy, microwave diathermy, therapeutic ultrasound diathermy, or laser diathermy anywhere on their body. Energy from diathermy (shortwave, microwave, ultrasound, and laser) can be transferred through the implanted neurostimulation system, can cause tissue damage, and can result in severe injury or death. Injury, damage, or death can occur during diathermy therapy even if the implanted neurostimulation system is turned “off.”
Handle, clean, and dispose of components and accessories
that have come in contact with bodily fluids according to National, Local and Facility rules, regulations and procedures.
Intelect® Legend XT Therapy System
NiMH Batteries contain Class E Corrosive materials. In the event of battery cell rupture or leakage, handle Battery Module wearing neoprene or natural rubber gloves. Contents of a ruptured or leaking battery can cause respiratory irritation. Hypersensitivity to nickel can cause allergic pulmonary asthma. Contents of cell coming in contact with skin can cause skin irritation and/or chemical burns.
Never, under any circumstances, open the Battery Module housing or cells. Should an individual cell from a battery become disassembled, spontaneous combustion of the negative electrode is possible. There can be a delay between exposure to air and spontaneous combustion.
Charge the Battery Module according to the instructions found in this manual. Never attempt to charge the Battery Module on any other charging mechanism.
Use the Battery Module only with the Intelect Legend XT Therapy Systems.
Do not reverse the polarity of the Battery Module. Doing so can increase the individual cell temperature and cause cell rupture or leakage.
Never dispose of Battery Module in fire. Never short circuit the battery. The battery may explode, ignite, leak or get hot causing serious personal injury.
Dispose of NiMH batteries according to national, state and
• local codes and regulations.
6
ABOUT INTELECT LEGEND XT
ELECTROTHERAPY INDICATIONS, CONTRAINDICATIONS, AND ADVERSE EFFECTS
Indications for Russian, TENS, High Voltage Pulsed Current (HVPC), Interferential and Premodulated waveforms
Relaxation of muscle spasms
Prevention or retardation of disuse atrophy
Increase local blood circulation
Muscle re-education
Maintaining or increasing range of motion
Additional Indications for Microcurrent, Interferential, Premodulated, and TENS waveforms
Symptomatic relief and management of chronic, intractable pain
Post-traumatic acute pain
Post-surgical acute pain
Contraindications
This device should not be used for symptomatic local pain relief
unless etiology is established or unless a pain syndrome has been diagnosed.
This device should not be used when cancerous lesions are present in the treatment area.
Stimulation should not be applied over swollen, infected, inflamed areas or skin eruptions (e.g. phlebitis, thrombophlebitis, varicose veins, etc.).
Other contraindications are patients suspected of carrying serious infectious disease and or disease where it is advisable, for general medical purposes, to suppress heat or fevers.
Electrode placements must be avoided that apply current to the carotid sinus region (anterior neck) or transcerebrally (through the head).
Safety has not been established for the use of therapeutic electrical stimulation during pregnancy.
Powered muscle stimulators should not be used on patients with cardiac demand pacemakers.
There should not be any use of TENS waveforms on patients with cardiac demand pacemakers.
Intelect® Legend XT Therapy System
7
ABOUT INTELECT LEGEND XT
ELECTROTHERAPY INDICATIONS, CONTRAINDICATIONS, AND ADVERSE EFFECTS (continued)
With TENS waveforms, isolated cases of skin irritation may
Additional Precautions
Caution should be used for patients with suspected or
diagnosed heart problems.
Caution should be used for patients with suspected or diagnosed epilepsy.
Caution should be used in the presence of the following:
When there is a tendency to hemorrhage following acute trauma or fracture.
Following recent surgical procedures when muscle contraction may disrupt the healing process.
Over a menstruating or pregnant uterus.
Over areas of the skin which lack normal sensation.
Some patients may experience skin irritation or hypersensitivity due to the electrical stimulation or electrical conductive medium. The irritation can usually be reduced by using an alternative conductive medium or an alternative electrode placement.
Electrode placement and stimulation settings should be based on the guidance of the prescribing practitioner.
Powered muscle stimulators should be used only with the lead wires and electrodes recommended for use by the manufacturer.
occur at the site of electrode placement following long-term application.
The effectiveness of TENS waveforms is highly dependent upon patient selection by a person qualified in pain management.
Adverse Effects
Skin irritation and burns beneath the electrodes have been reported with the use of powered muscle stimulators.
Potential adverse effects with TENS are skin irritation and electrode burns.
Intelect® Legend XT Therapy System
8
ABOUT INTELECT LEGEND XT
ULTRASOUND INDICATIONS AND CONTRAINDICATIONS
Indications for Ultrasound
Application of therapeutic deep heat for the treatment of selected sub-chronic and chronic medical conditions such as:
Relief of pain, muscle spasms, and joint contractures
Relief of pain, muscle spasms, and joint contractures that may be associated with:
Adhesive capsulitis
Bursitis with slight calcification
Myositis
Soft tissue injuries
Shortened tendons due to past injuries and scar tissues
Relief of sub-chronic, chronic pain, and joint contractures resulting from:
Capsular tightness
Capsular scarring
Contraindications
This device should not be used for symptomatic local pain relief unless etiology is established or unless a pain syndrome has been diagnosed.
This device should not be used when cancerous lesions are present in the treatment area.
Other contraindications are patients suspected of carrying serious infectious disease and disease where it is advisable for general medical purposes to suppress heat or fevers.
Intelect® Legend XT Therapy System
This device should not be used over or near bone growth
centers until bone growth is complete.
This device should not be used over the thoracic area if the patient is using a cardiac pacemaker.
This device should not be used over a healing fracture.
This device should not be used over or applied to the eye.
This device should not be used over a pregnant uterus.
This device should not be used on ischemic tissues in individuals with vascular disease where the blood supply would be unable to follow the increase in metabolic demand and tissue necrosis might result.
Additional Precautions
Additional precautions should be used when ultrasound is used on patients with the following conditions:
Over an area of the spinal cord following a Laminectomy, i.e., when major covering tissues have been removed
Over anesthetic areas
On patients with hemorrhagic diatheses
9
NOMENCLATURE
INTELECT LEGEND XT ELECTROTHERAPY AND COMBINATION THERAPY SYSTEMS
Two Channel Electrotherapy System
Two Channel Combination System
Intelect® Legend XT Therapy System
4
1
6
5
Two Channel Electrotherapy System
1. User Interface (See Page 12)
2. Front Access Panel
3. Rear Access Panel
4. Patient Data Card access port
5. Multimedia Card (MMC) access port (Unused)
6.
4
1
6
2
3
10
5
Two Channel Combination System
1.
User Interface (See Page 12)
2.
Front Access Panel
3.
Rear Access Panel
4.
Patient Data Card access port
5.
Multimedia Card (MMC) access port (Unused)
6.
Ultrasound Applicator (5cm2 shown)
7
2
3
NOMENCLATURE
INTELECT LEGEND XT ELECTROTHERAPY AND COMBINATION THERAPY SYSTEMS (continued)
Front Access Panel Rear Access Panel
7
Intelect® Legend XT Therapy System
1
1
Front Access Panel Lanyard
1.
When reinstalling the Front Access Panel, make certain the Lanyard does not become kinked.
Operator Remote Control Connector
2. Patient Interrupt Switch Connector
3. Channel 1 Lead Wire Connector
4. Channel 2 Lead Wire Connector
5. Microcurrent Probe Connector
6. Ultrasound Applicator Connector
7.
3
2
4
5
6
7
5
4
Screen Contrast Control
1. Power On/Off Switch
2. Technical Maintenance Port
3. Fuses
4. Mains Power Cord
5. Rear Access Panel
6. Serial Decal
7.
11
2
3
6
NOMENCLATURE
USER INTERFACE
15
14
13
12
11
10
9
8
7
Intelect® Legend XT Therapy System
1
1.
2
3
4
5
Rear Access Panel (See Page 11) Ultrasound Applicator LED Indicator
2.
(Combination only)
Ultrasound Applicator- 5 cm2 Standard
3.
(Optional 1 cm2, 2 cm2, and 10 cm2 applicators available, Combination only)
Intensity knob
4. Cable and Lead Wire Hook
5. Front Access Panel (See Page 11)
6. Start button
7. Pause button
8. Stop button
9. Clinical Resources Library button
10. Home Screen button
11. Back button
12. Patient Data Card Port
13. Multimedia Card (MMC) Port (Unused)
14.
User Setup and Parameter Control buttons
15.
6
5
12
NOMENCLATURE
Intelect® Legend XT Therapy System
SYMBOL DEFINITIONS
Below are the definitions for all of the symbols used in the Intelect Legend XT hardware and software. Study and learn these symbols before any operation of the system.
System
Software Symbols
MOVE UP
MOVE DOWN
MOVE RIGHT
MOVE LEFT
ACCEPT AND RETURN
DO NOT ACCEPT AND RETURN
Optional Module and Accessory Symbols
Operator Remote
Channel 3/4
Electrotherapy
INCREASE INTENSITY
DECREASE INTENSITY
PAUSE TREATMENT
MANUAL
M
STIMULATION
Battery Module
Module
PATIENT INTERRUPT SWITCH
CHANNEL 3 LEAD WIRES
CHANNEL 4 LEAD WIRES
MICROCURRENT PROBE
CONTRAST CONTROL
ON/OFF SWITCH
DATA PORT
MULTIMEDIA CARD*
PATIENT DATA CARD
STOP TREATMENT
System Hardware Symbols
THERAPY INTENSITY CONTROL
HOME
CLINICAL RESOURCES LIBRARY
BACK
PATIENT INTERRUPT SWITCH
CHANNEL 1 LEAD WIRES
CHANNEL 2 LEAD WIRES
MICROCURRENT PROBE
PAUSE TREATMENT
START TREATMENT
* Used for System Software Upgrades
CHANNEL 1/2 OPERATOR REMOTE CONTROL OPTIONAL
ULTRASOUND APPLICATOR
PAD CONTACT QUALITY
SINGLE CHANNEL GRAPH
PAD CONTACT QUALITY
DUAL CHANNEL GRAPH
13
CHARGE LEVEL
BATTERY CHARGING
CHANNEL 3/4 OPERATOR REMOTE CONTROL OPTIONAL
NOMENCLATURE
GENERAL TERMINOLOGY
Below are the definitions for all of the unique terminology used throughout this manual. Study these and become familiar with these terms for ease of system operation and familiarization with the components and control functionality of the Intelect Legend XT Therapy System. Some of these terms and definitions refer to a specific button or control on the system. Refer to page 13 for Symbol Definitions.
Back button
The dedicated button on the Main unit, below the display, that each time pressed takes the user back one screen at a time.
Previous Page button
The button used in some modalities and functions that will take the user back one page when reading multiple pages of text.
UP and DOWN Arrows
Controls used in various modality parameter screens to navigate or change a value up or down within the parameter.
Electrotherapy
Refers to the Electrical muscle or nerve Stimulation modalities of the system.
System
The primary system with all controls and functions.
Module
Any optional modular modality component designed for installation onto the System.
ULTRASOUND
1
2
1. Sound Head
That component of the Applicator that makes contact with the patient during Ultrasound or Combination therapy.
2. Applicator
The assembly that connects to the System and incorporates the Sound Head.
3. Applicator LED
The component of the Applicator which, when illuminated, indicates if the Sound Head is emitting Ultrasound.
Intelect® Legend XT Therapy System
3
14
SPECIFICATIONS
Intelect® Legend XT Therapy System
SYSTEM SPECIFICATIONS
HEIGHT
DEPTH
WIDTH
NOTE: All waveforms except High Voltage Pulsed Current (HVPC) have been designed with a 200 mA current limit. TENS waveform output intensities are measured, specified and listed to peak, not peak to peak.
DIMENSIONS
Width
Combination System . . . . . . . . . . . . . . . . . . . . . . . . . 11.375 in (28.9 cm)
Electrotherapy System . . . . . . . . . . . . . . . . . . . . . . . . . 9.750 in (24.8 cm)
Depth (Combination and Electrotherapy System) . 12.750 in (32.4 cm) Height (Combination and Electrotherapy System) . 8.750 in (22.2 cm) Standard Weight
Two Channel Combination System. . . . . . . . . . . . . . . . . . 7 lbs (3.2 kg)
Two Channel Electrotherapy System . . . . . . . . . . . . . . . . 6 lbs (2.7 kg)
Power (Combination and Electrotherapy Units)
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .100 - 240 V - 1.0 A, 50/60 Hz
Output (Internal Power Supply) . . . . . . . . . . . . . . . . . . . . . . . . +24 V, 7.3 A
Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Two 6.3 A Time Lag
Electrical Class . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .CLASS I
Electrical Type
Ultrasound . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . TYPE B
Electrotherapy and sEMG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TYPE BF
Regulatory Compliance
UL/IEC/EN 60601-1 IEC/EN 60601-1-2 IEC 60601-2-5 IEC 60601-2-10
0413
15
SPECIFICATIONS
WAVEFORM SPECIFICATIONS
Intelect® Legend XT Therapy System
IFC- Interferential (Traditional 4 Pole)
Interferential Current is a medium frequency waveform. Current is distributed through two channels (four electrodes). The currents cross each other in the body at the area requiring treatment. The two currents interfere with each other at this crossing point, resulting in a modulation of the intensity (the current intensity increases and decreases at a regular frequency).
Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Electrodes
Carrier Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2,500-5,000 Hz
Beat Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-200 Hz
Sweep Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 sec
Sweep Low Beat Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-200 Hz
Sweep High Beat Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-200 Hz
Scan Percentage . . . . . . . . . . . . . . . . . . . . . . . . . . Static, 10%, 40%, 100%
Amplitude . . . . . . . . . . . . . . . . . . . . . . . . . . 0-100 mA Peak into 500 ohm
Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 minutes
Available on Channel . . . . . . . . . . . . . . . . . . . . . . . . . 1 & 2, 3 & 4 Option
*CC= Constant Current CV= Constant Voltage
Russian
Russian Current is a sinusoidal waveform, delivered in bursts or series of pulses. This method was claimed by its author (Kots) to produce maximal muscle strengthening effects without significant discomfort to the patient.
Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Electrodes
Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-100 mA
Channel Mode . . . . . . . . . . . . . . . . . . . . . . . . . Single, Reciprocal, Co-Contract
Duty Cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10%, 20%, 30%, 40%, 50%
Mode Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CC or CV*
Anti-Fatigue . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Off or On
Cycle Time . . . . . .5/5, 4/12, 10/10, 10/20, 10/30, 10/50, Continuous
Burst Frequency (Anti-Fatigue Off ) . . . . . . . . . . . . . . . . . . . . 20-100 bps
Ramp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.5, 1, 2, and 5 sec
Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 minutes
Available on Channels . . . . . . . . . . . . . . . . . . . . . . . . 1 & 2, 3 & 4 Option
16
SPECIFICATIONS
WAVEFORM SPECIFICATIONS (continued)
TENS- Symmetrical Biphasic
The Symmetrical Biphasic waveform has a short pulse duration and is capable of strong stimulation of nerve fibers in the skin and in muscle. This waveform is often used in portable muscle stimulation units, and some TENS devices. Because of its short pulse duration, the patient typically tolerates the current well, even at relatively high intensities.
Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Electrodes
Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0-80 mA
Phase Duration . . . . . . . . . . . . . . . . . . . . . . . . . Adjustable 20-1,000 μsec
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-250 Hz
Mode Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CC or CV*
Burst Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-25 bps
Frequency Modulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-250 Hz
Amplitude Modulation . . . . . . . . . . . . Off, 40%, 60%, 80%, and 100%
Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 minutes
*CC= Constant Current CV= Constant Voltage
Intelect® Legend XT Therapy System
Microcurrent
Microcurrent is a monophasic waveform of very low intensity that closely simulates the electrical current generated by the human body. Microcurrent can be applied via electrodes or probe.
Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrodes or Probe
Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0-1000 μA
Polarity . . . . . . . . . . . . . . . . . . . . . . . . . Positive, Negative, or Alternating
Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 minutes
Available on channels . . . . . . . . . . . . . . . . . . . . . . . . 1 & 2, 3, & 4 Option
Stimulus delivered by the TENS waveforms of this device, in certain configurations, will deliver a charge of 25 microcoulombs (μC) or greater per pulse and may be sufficient to cause electrocution. Electrical current of this magnitude must not flow through the thorax because it may cause a cardiac arrhythmia.
17
SPECIFICATIONS
WAVEFORM SPECIFICATIONS (continued)
Intelect® Legend XT Therapy System
Premodulated (Traditional 2 Pole IFC)
Premodulated Current is a medium frequency waveform. Current comes out of one channel (two electrodes). The current intensity is modulated: it increases and decreases at a regular frequency (the Amplitude Modulation Frequency).
Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Electrodes
Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-100 mA
Carrier Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2,500 Hz
Beat Fixed (Sweep Off ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-200 Hz
Sweep Low Beat Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-200 Hz
Sweep High Beat Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . 81-200 Hz
Cycle Time . . . . . . . . . . .Continuous, 5/5, 4/12, 10/10, 10/20, 10/30, 10/50
Mode Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CC or CV*
Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 minutes
Available on Channel . . . . . . . . . . . . . . . . . . . . . . . . .1 & 2, 3, & 4 Option
*CC= Constant Current CV= Constant Voltage
High Voltage Pulsed Current (HVPC)
The High Voltage Pulsed Current (HVPC) has a very brief pulse duration characterized by two distinct peaks delivered at high voltage. The waveform is monophasic (current flows in one direction only). The high voltage causes a decreased skin resistance making the current comfortable and easy to tolerate.
Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrodes or Probe
Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-500 V
Polarity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Positive or Negative
Ramp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.5 sec, 1 sec, 2 sec, 5 sec
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Peak Current or Volts
Sweep . . . . . . . . . . . . . .Continuous, 80/120 pps, 1/120 pps, 1/10 pps
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-120 pps
Cycle Time . . . . . .5/5, 4/12, 10/10, 10/20, 10/30, 10/50, Continuous
Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 minutes
Available on Channels . . . . . . . . . . . . . . . . . . . . . . . .1 & 2, 3, & 4 Option
18
SPECIFICATIONS
ULTRASOUND SPECIFICATIONS
Ultrasound
Intelect® Legend XT Therapy System
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 MHz, ± 5%; 3.3 Mhz, ±5%
Duty Cycles . . . . . . . . . . . . . . . . . . . . . . . . . . . 10%, 20%, 50%, Continuous
Pulse Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 Hz
Pulse Duration . . . . . . . 1 mSec, ±20%; 2 mSec, ±20%; 5 mSec, ±20%
Output Power
10 cm2 Crystal . . . . . . . . . . . . . . . . 0-20 W at 1 MHz, 0-10 W at 3.3 MHz
5 cm2 Crystal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0-10 W, 1 and 3.3 MHz
2 cm2 Crystal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-4 W, 1 and 3.3 MHz
1 cm2 Crystal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-2 W 3.3 MHz Only
Amplitude . . . . . . . . . . . . . . . . . . . . 0 to 2.5 w/cm2 in continuous mode,
0-3 w/cm2 in pulsed modes
Output accuracy . . . . . . . . . . . . . . . . . . . ±20% above 10% of maximum
Temporal Peak to Average Ratio: . . . . 2:1, ± 20%, at 50% Duty Cycle
5:1, ± 20%, at 20% Duty Cycle
9:1, ± 20%, at 10% Duty Cycle
Beam Nonuniformity Ratio. . . . . . . . . . . . . . . . . . . . . . . 5.0 : 1 maximum
Beam Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Collimating
2
Effective Radiating Areas. . . . . . . . . . . . 10 cm
Crystal: 6.8 cm 5 cm2 Crystal: 3.5 cm2 – 5 cm 2 cm2 Crystal: 1.4 cm2 – 2 cm
1 cm2 Crystal: 0.7 cm2 – 1 cm
Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-30 minutes
Head Warming Feature
The Head Warming feature of an Intelect Legend XT Combination Therapy System utilizes Ultrasound output resulting in warming of the Sound Head to increase patient comfort.
With Head Warming enabled, ultrasound is emitted without pressing the Start button. The Applicator LED will not illuminate during the Head Warming period. US Channel will indicate "Head Warming".
Output . . . . . . . . . . . . . . . . . . . . . . . 0 - 50% Cycling of maximum power
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.3 Mhz
Sound Head Temperature . . . . . . . . . 85 °F - 110 °F (29.4 °C - 43.3 °C)
Do not apply the Ultrasound Applicator to the patient during the Head Warming period. Applicator must remain in Applicator Hook during the Head Warming period.
19
2
– 10 cm
2
2
2
2
SETUP
INTELECT LEGEND XT THERAPY SYSTEMS
Remove the Intelect Legend XT Two or Four Channel Therapy System and all accessories from the shipping carton. Visually inspect for damage. Report any damage to the carrier immediately.
Contents of Carton:
Intelect Legend XT Two or Four Channel Electrotherapy or Combination System
Patient Data Cards (5)
Patient Interrupt Switch for Channels 1 and 2
Electrotherapy Lead Kit that includes:
Nylatex® Wrap- 2 rolls- 2.5 in X 24 in (6 cm X 61 cm)
Lead Wires (one for Channel 1 and one for Channel 2)
3 in (7.6 cm) Round Carbon Electrodes (one set of 4 - 2 Black and 2 Red)
DURA-STICK™ II 2.75 in (7 cm) Round Disposable Electrodes (1 pack of 4)
Patient Interrupt Switch for Channels 3 and 4 (Four Channel Systems only)
Lead Wires- (one for Channel 3 and one for Channel 4 - Four Channel Systems only)
2
Ultrasound Applicator (Combination Systems Only)
5 cm Cord Set
• Conductor Transmission Gel - 1 bottle (Combination Systems Only)
• User Manual
Intelect® Legend XT Therapy System
20
SETUP
THERAPY SYSTEM SETUP
Accessing Operator Utilities
Plug unit into wall outlet. Turn system On.
Press the Home and Back buttons simultaneously.
To return to the System Home screen, press the Home button.
Clinic Name
Press Clinic Name button.
POWER SWITCH
CLINIC NAME BUTTON
Select the row of alpha or numeric characters desired by pushing the button beside the corresponding row. Select the desired character in the row by pressing the row button until the desired letter is framed.
NOTE: To add a space, select the area to the left of the letter A.
Once selection is framed, press the Accept and Return Arrow button. The character just chosen will display in the top of the screen and the cursor will advance to the next character.
To go back a character press the Move Left Arrow button. To delete the character, press the Delete button.
Once Clinic Name is completed, press the Save button.
To discard entry, press the Back button.
Intelect® Legend XT Therapy System
SELECT ROW AND CHARACTER BUTTONS
MOVE LEFT
DELETE SAVE
BACK
ACCEPT AND RETURN
21
SETUP
THERAPY SYSTEM SETUP (continued)
Restoring Default Protocols
Press Restore Default Protocols button.
RESTORE DEFAULT PROTOCOLS BUTTON
Restoring Default Unit Settings
Press the Restore Default Unit Settings button to restore the system defaults. This control will neither change the Date and Time nor affect any of the Clinical Protocols stored in the system.
Intelect® Legend XT Therapy System
RESTORE DEFAULT UNIT SETTINGS BUTTON
Press Yes button to restore the Protocols to Factory Settings. This will permanently remove all User Protocols and Sequences.
NOTE: This will permanently remove all User Protocols.
If it is not desired to permanently remove all of the User Protocols and User Sequences from the System, press the No button.
PRESS YES BUTTON TO RESTORE PROTOCOLS
PRESS NO BUTTON TO KEEP PROTOCOLS AS THEY ARE
After the settings have been restored, a message will appear stating that the Default Unit Settings are restored. Press any button to return to Utilities screen.
22
SETUP
THERAPY SYSTEM SETUP (continued)
Erasing Patient Data Card
Install Patient Data Card to be erased into Patient Data Card Access Port on the system.
Press Erase Patient Card button.
INSERT PATIENT CARD
ERASE PATIENT CARD BUTTON
Intelect® Legend XT Therapy System
Setting Date and Time
Press Set Date and Time button.
SET DATE AND TIME BUTTON
Press the Yes button to erase all data from Patient Data Card. Press the No button to keep all data on Patient Data Card.
After Patient Data Card is erased, a verification message will appear. Press any button to return to the Utilities screen.
YES
BUTTON
NO
BUTTON
Press the UP or Down Arrow button for the respective area until desired change is displayed.
After all desired changes are made, press the Back button to return to the Utilities screen.
23
PRESS THE RESPECTIVE UP OR DOWN
ARROW BUTTONS TO CHANGE
YES
BUTTON
BACK
BUTTON
SETUP
THERAPY SYSTEM SETUP (continued)
Setting System Volume
Press Volume button until the desired system volume is achieved. There are six settings: Off, X-Low, Low, Med, High and X-High.
Each time the Volume button is pressed the setting displayed will emit three beep tones at that level.
VOLUME BUTTON
Displaying Unit Version Information
Press the Display Unit Version Information button to show the system software versions installed.
Press the Back button to return the Operator Utilities screen.
Intelect® Legend XT Therapy System
PRESS DISPLAY UNIT
VERSION BUTTON TO VIEW
SOFTWARE VERSIONS
BACK
BUTTON
24
Loading...
+ 69 hidden pages