This operating manual contains certain types of text design intended to assist you in
comprehending the significance of the text based on its appearance.
Instructions for actions
•This text instructs you in how to operate your device correctly.
•This text subdivides an action into steps or comments on the action step.
This text shows the result of an action.
Lists
–This text is part of a list.
Menus and buttons
Names of
The operating procedure descriptions also contain references to the buttons that you
have to press and what you can expect to see in the text.
Cross references to other chapters
Cross references to other chapters are highlighted in bold and in small caps.
MENUS and BUTTONS are greyed out and highlighted in small caps.
Warning notes
This manual contains warnings, safety instructions and specific operating instructions
in accordance with liability regulations.
DANGER refers to a situation of acute danger which, if not avoided, could lead to serious or fatal injury.
DANGER
The source of the danger is stated here.
These are the possible consequences!
► The instructions for avoiding the danger are given here.
WARNING refers to a situation of potential danger which, if not avoided, could lead
to serious injury.
WARNING
The source of the danger is stated here.
These are possible consequences!
► The instructions for avoiding the danger are given here.
CAUTION indicates that incorrect operation could lead to minor injuries.
4
13 610 0002 11.2022
Introduction
Page 6
CAUTION
The source of the danger is stated here.
These are the possible consequences!
► The instructions for avoiding the danger are given here.
NOTICE indicates that incorrect operation could lead to damage to the device.
NOTICE
The source of the danger is stated here.
These are possible consequences!
► The instructions for avoiding the danger are given here.
Other instructions
This text indicates additional information concerning special features, etc. and/or operating instructions.
Safety signs and other symbols used in this manual
SymbolNotification
General warning sign
5
Type B applied part
Potential equalisation
Disconnect mains plug
Wear hearing protection
USB connection
Introduction
WEEE mark
13 610 0002 11.2022
Page 7
SymbolNotification
Manufacturer
CE mark
Device serial number
CSA certification mark
It is essential to comply with the operating manual
Fuse
Electromagnetic interference may occur
in the vicinity of devices marked with this
symbol.
Hot surface
Explosion
Electrical warning sign: high voltage
6
13 610 0002 11.2022
Introduction
Page 8
1General Safety Information
The following chapter contains all safety information that has to be followed when
working with the Chattanooga Intelect F-SW.
WARNING
Incorrect handling of the device
Possibility of injuries to the patient and the operating personnel!
► Read this chapter carefully before you start using the Chattanooga Intelect F-SW.
► Read the separate operating manuals for all devices associated with the Chatta-
nooga Intelect F-SW.
1.1Instructions for safe use
1.1.1Intended Use and Operational Safety
To use this device in accordance with its intended use, the user must possess the necessary technical proficiency, and knowledge of the operating manual.
The device is intended exclusively for use by healthcare professionals who have been
trained to use the device (see also 2.2 PRECONDITIONSFOROPERATION).
The device is only allowed to be used for the applications described in chapter 2.1.1INDICATIONS. Only perform treatments approved by the manufacturer!
Furthermore, the device is only allowed to be operated by trained personnel who comply with the preconditions for operation in the chapter 2.2 P
TION.
All status and error messages signaled during treatment must always be attended to
without delay.
7
While applying focused shockwaves at maximum adjustment, do not use more than
6.000 subsequent shocks and stick to a consecutive break of 5 minutes.
RECONDITIONSFOROPERA-
Checks and inspections prior to treatment
Before using the device, the user must make sure it is functioning safely and that it is
in proper condition.
•It is essential to perform the functional checks after switching on the Chattanooga
Intelect F-SW, before starting treatment. Read about this in Chapter 5.14 F
NALCHECKS.
•Have the maintenance procedures recommended by the manufacturer carried out
by personnel suitably authorised. Read about this in Chapter 6.7 MAINTENANCE
AND
SAFETYCHECKS.
No treatment is permitted if a display on the control device or a touch screen fails.
Protection against electrical hazards
Sources of voltage can give rise to currents as a result of body resistance, which not
only flow through the patient but can also impair or even endanger the personnel administering the treatment.
•Devices that are not medical products in accordance with EN 60601 must be set
up outside the patient environment.
•Do not touch electrical connectors while you are touching the patient.
UNCTIO-
General Safety Information
13 610 0002 11.2022
Page 9
•Disconnect the Chattanooga Intelect F-SW from the mains plug before starting
any cleaning or maintenance work.
•Disconnect the connected handpieces from the device before carrying out cleaning and maintenance work. Do not reconnect it until everything has been completely reassembled.
•Do not try to open the device! Risk of electric shocks!
•Always connect the potential equalisation connector of the Chattanooga Intelect
F-SW in accordance with national guidelines.
Protection against high voltage
High-voltage components are identified as follows:
DANGER
Contact with high-voltage parts:
Severe or fatal injury!
► Only operate the device if the housing is intact and closed
► Work in the area of high voltage is only allowed to be performed by personnel suit-
ably authorised by the manufacturer.
Protection against noise
The noise level during administration of shockwaves is within the safe range. Nevertheless, we recommend wearing suitable ear protection during treatment in order to
minimise exposure to noise.
Protection against explosion
The device is not allowed to be used in potentially explosive atmospheres (according
to classification AP and APG of IEC 60601-1) i.e. in the reach of anesthetic gas mixtures
with air, oxygen or nitrogen oxide.
DANGER
Risk of explosion due to flammable and explosive materials
Injuries to patients, people administering the treatment and third parties!
► During operation, avoid using the substances specified in the following section.
► Switch off the device before cleaning and disconnect the mains plug.
► Comply with the information on cleaning in Chapter 6 CLEANING, CAREANDMAIN-
TENANCE
8
13 610 0002 11.2022
General Safety Information
Page 10
The following agents are NOT to be used during operation:
–Highly inflammable and potentially explosive inhalation anaesthetics and mixtures
of the same, such as
•Ether pro narcosi (diethyl ether)
•Cyclopropane.
–Inflammable, highly volatile skin cleaning agents and skin disinfectants which can
form a potentially explosive atmosphere, such as
•Washing ether
•Petrol ether.
The optional foot switch must not be used in potentially explosive atmospheres according to classification AP as per IEC 60601.
1.1.2Safety during treatment of the patient
General note:
–Organs with gas inclusions, in particular parts of the lung, are NOT allowed to be
exposed to shock waves.
–As it passes through tissue, the shock wave’s energy is slightly reduced; this reduc-
tion is significantly weakened by the bone structure.
–Shock waves can give rise to undesirable heart reactions. The patient must be con-
tinuously observed during the treatment and attention must be paid to any reac-
tions experienced by the patient.
–The patient must not be under anaesthetic.
–Only perform treatments approved by the manufacturer!
–The user is responsible for correctly positioning the handpieces and correctly se-
lecting the treatment zone.
9
–Air bubbles reduce the effectiveness of shock waves. Therefore, air bubbles must
always be removed from the shock wave path.
–Risk of transmission of microorganisms! Disinfect the handpiece after each use!
For information about this, see 6 C
While applying focused shockwaves at maximum adjustment, do not use more than
6.000 subsequent shocks and stick to a consecutive break of 5 minutes.
LEANING, CAREANDMAINTENANCE.
General Safety Information
1.1.3Warning against damage to equipment and the device
Any damage to the device resulting from incorrect operation is not covered by the
manufacturer's warranty.
Electromagnetic compatibility
This device complies with the requirements of the applicable standard on electromagnetic compatibility.
Nevertheless, portable and mobile HF communications equipment (e.g. mobile
phones), including antennas, can interfere with medical electrical equipment. They
should not be used less than 30 cm from the device - including the cables specified by
the manufacturer.
This device is subject to special precautions regarding EMC and needs to be installed
according the EMC guidelines in chapter 9.3.1 EMC
DECLARATION
The use of accessories or cables that are not authorised by the manufacturer can result
in increased interference emissions or reduced resistance to interference emissions by
the device.
GUIDELINESANDMANUFACTURER'S
13 610 0002 11.2022
Page 11
The Chattanooga Intelect F-SW is not allowed to be positioned immediately next to,
stacked on or jointly with other devices. If operation near or jointly with other devices
is required, the Chattanooga Intelect F-SW must be tested in that particular environment to ensure operation according to technical specifications.
The device is not to be used within an MR environment. Within proximity of the device
to an Magnetic Resonance source device disturbances may occur. In case of disturbance consider repositioning of the device.
DANGER
Improper connection:
There is a risk of electric shock!
► Do not connect the device to a power supply network unless it has a protective
conductor.
► Never use multi-socket power strips.
Please read Chapter 9.3.1 EMC
GUIDELINESANDMANUFACTURER'SDECLARATION.
If the Chattanooga Intelect F-SW is connected to a 240 V mains supply with a mains
frequency of 60 Hz, the mains supply must be balanced.
Setup and operation
There are ventilation slits on the device which must not be covered by other objects.
•Check that the installation surfaces have sufficient carrying capacity to avoid
equipment damage!
•Check that the device is in perfect working order before each use. Read about this
in Chapter 5.14 F
UNCTIONALCHECKS.
•Never cover the device when in use!
•Make absolutely sure that no liquid can seep into the system housing or handpiece.
Storage and transport
Incorrect storage and transport can result in damage to the device and device failure.
•Completely empty the water circuit and the handpiece before you transport the
device or put it into storage.
Otherwise, there is a risk of the water freezing, which will lead to damage. Please
contact your responsible Service centre if this does happen.
•Comply with the ambient conditions specified in Chapter 9 T
ONS.
ECHNICALSPECIFICATI-
10
•Make sure that no cables are crushed or sheared.
•Make sure that the handpiece cable is not kinked.
•Do not pull on the handpiece or its cable in order to move the device.
•Always disconnect the handpiece from the control device properly.
Disposal
•Comply with national disposal regulations when disposing of the Chattanooga In-
•Comply with the relevant information in the operating manuals for the additional
13 610 0002 11.2022
telect F-SW or individual components.
devices.
General Safety Information
Page 12
11
1.2Manufacturer`s Responsibility
WARNING
No modifications are to be made to this device without the permission of the manufacturer.
The manufacturer of the Chattanooga Intelect F-SW is only responsible for the impact
of its product on safety, reliability and performance if:
–maintenance of the device is performed at the intervals specified by the manufac-
turer
–installation, expansions, conversions, new installations, modifications or repairs
are performed by people authorised by the manufacturer
–the electrical installation in the rooms in question corresponds to the requirements
of DIN/IEC
–the device is used in compliance with the operating manual
The periodic maintenance measures specified by the manufacturer must be performed
on schedule by personnel suitably authorised. Original parts from the manufacturer
must be used; otherwise, the manufacturer's liability shall be rendered null and void.
1.3Owner`s Responsibility
The owner is responsible for complying with the relevant national statutory provisions
governing setting up and operating technical medical equipment. (For Germany, the
Medical Products Act.) It is expressly stated that the use of unauthorised accessories
and/or unauthorised equipment combinations shall render the product liability null
and void.
The device is exclusively allowed to be used with accessories, wearing parts and disposable articles that have been checked by the testing body responsible for testing the
device to ensure that they function without risk.
General Safety Information
13 610 0002 11.2022
Page 13
2Principles
2.1Physical principles
The Chattanooga Intelect F-SW is a universal, compact shock wave generator that generates focused shock waves hereinafter referred to as F-SW .
The Chattanooga Intelect F-SW may be used to perform the following treatments:
–Low-energy focused shock wave therapy with electromagnetically generated
shock waves (energy levels below 0.3 mJ/mm2)
–High-energy focused shock wave therapy with electromagnetically generated
shock waves (energy levels above 0.3 mJ/mm2)
The F-SW have a short pulse length and are concentrated on areas a few millimetres
in diameter, allowing shock waves to be applied to a tightly localised area, even in
deeper tissue layers.
2.1.1Indications
Orthopaedics/pain therapy
–MPS (Myofascial pain syndrome)
•Treatment of deep muscle trigger points
•Treatment of superficial muscle trigger points
•Myofascial trigger points1 / Acupuncture points
•Myofascial pain / muscle pain / fascial pain
•Chronic lower back pain, trapezius muscle, pelvic floor muscle
A sound knowledge of trigger point therapy and trigger point shock wave therapy (TrST) is required for
therapeutic application of the Chattanooga Intelect F-SW in the field of trigger point shock wave therapy.
Qualified training in acupuncture and acupuncture shock wave therapy (AkuST) is required for therapeutic application of the Chattanooga Intelect F-SW in the field of acupuncture.
Principles
Page 14
Dermatology
Wound healing
–Ulcers
•Arterial ulcers
•Venous ulcers
•Diabetic foot
•Pressure sores, Decubital ulcers
–Burns
–Acute and chronic lesions
–Post-traumatic necrosis
–Wounds with disturbed healing
–Postsurgical wounds
–Traumatic wounds
Low-energy focused shock wave therapy (according to Chapter 2.1 PHYSICALPRINCIP-
LES) is not permitted for:
–Brain or spine in the treatment area
–Malignant tumour in the treatment area
–Pregnancy
High-energy shock wave therapy
High-energy focused shock wave therapy (according to Chapter 2.1 P
LES) is not permitted for:
–Air-filled tissue (particularly lung tissue) in the treatment area
–Brain or spine in the treatment area
–Untreated coagulopathies (haemophilia)
–Malignant tumour in the treatment area
–Epiphyseal plate areas in children
–Pregnancy
–Use of anticoagulants, especially Marcumar
–Thrombosis in the treatment area
–Cortisone therapy up to 6 weeks before first treatment
HYSICALPRINCIP-
Principles
13 610 0002 11.2022
Page 15
2.1.3Side effects
CAUTION
After treatment with the Chattanooga Intelect F-SW
side effects may occur.
► Familiarise yourself with the list of side effects.
► Inform the patient of possible side effects.
Treatment with the Chattanooga Intelect F-SW may cause the following side effects:
–Swelling, reddening, haematomas
–Petechiae
–Pain
These side effects generally abate after 5 to 10 days.
2.2Preconditions for operation
2.2.1Operating personnel
DANGER
If treatments and medical procedures are performed by inadequately qualified personnel,
this can result in damage to the health of patients and third parties as well as
fire or explosion hazards.
► Make sure that treatments and cleaning work are carried out only by qualified and
instructed healthcare professionals.
► Observe the preconditions for operation that are detailed in this chapter.
14
The Chattanooga Intelect F-SW is intended exclusively for use by healthcare professionals who have been trained to use the device.
It is expected that this professional has practical knowledge of medical procedures and
applications as well as of the terminology and should be experienced in treating the
indications stated in Chapter 2.1.1 I
The professional must have physical and cognitive prerequisites such as vision, hearing
and reading. Furthermore, the basic functions of the upper extremities must be guaranteed.
The device is designed for a demographic target group between 18 and 65 years.
NDICATIONS.
13 610 0002 11.2022
Principles
Page 16
2.2.2Training of the operator
Operators of the Chattanooga Intelect F-SW must have been adequately trained in using this device. An introduction to the principles of operation will be provided by your
dealer with reference to this operating manual and will be documented in the system
logbook.
The operator must be instructed in the following points :
–Operation and intended use of the device with practical exercises
–Mechanism of action and function of the device and the energies delivered by it
–All component settings
–Indications for use of the device
–Contraindications and side effects
–Explanation of the warnings in all operating modes/states
–Training on how to perform the functional checks
1Monitor
2Power indicator
3Connection for foot switch
4F-SW handpiece connection
16
13 610 0002 11.2022
System description
Page 18
17
Fig. 3-2 Rear view of Chattanooga Intelect F-SW
1Potential equalisation connection
2not used
3Mains connection
4Mains fuse holder
5Mains switch
6USB connection for USB stick, USB mouse, USB keyboard
7Water supply connection
8not used
9Type Plate
System description
13 610 0002 11.2022
Page 19
3.2F-SW handpiece and optional C-ACTOR handpiece
Focused shock waves with a short wavelength that are concentrated on a focal zone
outside the handpiece are administered over the F-SW handpiece or the C-ACTOR
handpiece into the body at the treatment zone that has been established by diagnosis.
Optical difference between F-SW and C-ACTOR : The F-SW has a blue ring around
the coupling diaphragm and the C-ACTOR has a red ring around the coupling diaphragm.
Fig. 3-3 F-SW handpiece or C-ACTOR handpiece
1Trigger button
2Clamping ring
3Fixing screws
4Coupling diaphragm
The coupling diaphragm is fixed by a clamping ring and 3 fixing screws. It can only be
opened from authorised personnel with special tools.
The penetration depth of the shock wave can be varied by stand-off devices.
3.3Use of stand-off devices
The penetration depth of the shock wave can be adjusted by using different stand-off
devices.
18
Fig. 3-4 F-SW handpiece or C-ACTOR handpiece
13 610 0002 11.2022
System description
Page 20
19
Fig. 3-5 Depth of therapeutical effect of F-SW handpiece
Fig. 3-6 Depth of therapeutical effect of C-ACTOR handpiece
System description
•Perform changing of the stand-off devices as described in chapter 6.2.1 CHANGING
THE
STAND-OFFDEVICE.
The stand-off has a limited service life. It should be replaced if there are visible changes
in the material (discolouration, tarnishing, streaks, gas bubbles), deformation of the
surface in the coupling area or leaks.
The stand-off should be replaced at least every 12 months.
13 610 0002 11.2022
Page 21
4Installation Instructions
4.1Scope of Supply
The standard scope of supply of the Chattanooga Intelect F-SW includes the following
items:
Equipment damage due to improper storage and transport
may affect the health of patients and users.
► Before commissioning, check that the delivered items are undamaged.
•Remove the device and accessories from the packaging container. Proceed with
extreme caution.
•Check that all items are included in the packaging container and that they are not
damaged.
•Contact your supplier or the manufacturer/dealer immediately if any delivered
items are missing or damaged.
•Retain the original packaging. It may prove useful for any later equipment transport.
4.3Correct Positioning of the Device
Maintain a minimum distance between the device and the wall so that the mainsplug
can be pulled out without restrictions (disconnected from the power supplynetwork)
and the ventilation slits on the rear are not blocked.
20
13 610 0002 11.2022
Installation Instructions
Page 22
4.4Installing the handpiece holder
The handpiece holder can be mounted on the right as well as on the left side.
•Use a 2.5 mm Allen key for installation
•Screw the handpiece holder onto the right side wall of the Chattanooga Intelect
F-SW, as shown in the picture below.
Fig. 4-1 Mounted handpiece holder
1Handpiece holder mounted to right side wall
21
4.5Installing the holding arm
To facilitate handling of the F-SW handpiece, you can hook the handpiece onto the
optionally available holding arm.
•Use a 2.5 mm Allen key for installation.
•Screw the holder for the arm firmly onto the holes provided for it on the left of the
instrument (see picture below).
Fig. 4-2 Attachment holes for the holding arm
•Place the holding arm into the holder.
Fig. 4-3 Holding arm attached
Installation Instructions
13 610 0002 11.2022
Page 23
4.6Connecting the electrical power supply
DANGER
Improper connection:
There is a risk of electric shock!
► Do not connect the device to a power supply network unless it has a protective
conductor.
► Never use multi-socket power strips.
•Connect the mains cable to the mains connector on the rear of the device (see F3-2 ).
•Insert the mains cable into the electrical socket.
NOTICE
► Maintain a minimum distance between the device and the wall so that the mains
plug can be pulled out without restrictions (disconnected from the power supply
network) and the ventilation slits on the rear are not blocked.
IG.
22
13 610 0002 11.2022
Installation Instructions
Page 24
4.7Connecting Handpiece
•Connect the connector of the F-SW handpiece to the handpiece connection provided on the Chattanooga Intelect F-SW and secure it using the black locking
screw. The locking screw must be tightened up to the stop until finger-tight.
Fig. 4-4 Connecting the F-SW handpiece
Fill the water circuit of the Chattanooga Intelect F-SW first when the F-SW handpiece
is first connected after delivery. The instrument will signal “water level too low” when
it is switched on.
4.8Connecting an optional foot switch
23
•Connect the connection cable of the foot switch to the appropriate connection on
the front side of the instrument.
The foot switch is protected against ingress of water according to classification IPX8
as per IEC 60529.
4.9Potential equalisation connector
The Chattanooga Intelect F-SW features a potential equalisation connection.
•Connect one end of the potential equalisation cable to the PE connection on the
Chattanooga Intelect F-SW and the other end to your PE connection.
CAUTION
Risk of
electrical hazard.
► Always connect the potential equalisation connector in accordance with national
guidelines.
4.10USB connection
The USB connection acts as an interface for data input and output.
•Connect if required
–a USB memory stick which supports the USB V1.1 protocol
–a USB mouse
–a USB keyboard.
Installation Instructions
13 610 0002 11.2022
Page 25
The connected instruments must be approved as medical products in accordance with
IEC 60601.
CAUTION
Misuse of the USB connector may cause
malfunction of or damage to the device.
► The use of USB connector not authorized by the manufacturer is not permitted.
► Do not connect any wireless technology such as bluetooth and do not connect any
devices for battery charging.
4.11Transport
NOTICE
The side walls of the device can be bent if it is not transported correctly.
Defect of the touchscreen or other components!
► DO NOT carry the device by means of mounted accessory parts (e.g. F-SW plug).
Make sure that your hands are dry and free of grease.
•Dismount the handpiece holder before transporting the device.
•To transport the instrument, grip the indentations on the side of the housing as
shown in the picture below (1) and lift it carefully.
Fig. 4-5 Transporting the device
•Set the device slantly down in order to avoid squeezing the fingers.
4.12Compatibility
The Chattanooga Intelect F-SW is allowed to be operated with the following accessories:
24
Handpiece F-SW SEPIA LTArt no. 19000
Handpiece C-ACTOR SEPIA LTArt. no. 29204.0001
Foot switch
13 610 0002 11.2022
Installation Instructions
Page 26
5Operation
The Chattanooga Intelect F-SW is operated using a colour TFT LCD monitor with touch
screen function and a graphical user interface.
5.1Graphical user interface
The user interface of the Chattanooga Intelect F-SW is divided into various areas for
displaying different information. The individual controls are arranged in function
groups (see picture below):
25
Fig. 5-1 Controls
1Top navigation bar
2Top navigation bar
3Top navigation bar
4Status bar
5Selection area
6Bottom navigation bar
7Bottom navigation bar
8Bottom navigation bar
9Parameter display (nominal and actual values)
The following functional description refers to control software version 13441.19.x.x or
later (this can be seen in the Info menu).
Navigation bars:
The top and bottom navigation bars (see F
that you can use for navigating through the menus:
IG. 5-1 CONTROLS) contain control buttons
Operation
13 610 0002 11.2022
Page 27
Parameter entry screen
MENUOpen the sub-menu
CONFIGURATIONJump to the LOADCONFIGURATION sub-menu (call up saved
parameter configurations or patient records)
Main and sub-menu
BACKStep back
MENUEXITReturn to parameter entry screen
DELETEDelete configurations
SAVESave configurations
OKConfirm entries, acknowledge messages
The arrow keys can be used for changing (increasing or
decreasing) the parameter values. If you are in a sub-menu
that contains more menu items than can be displayed in
the top part of the display, you can use the arrow keys to
scroll to the bottom of the list (page up/down).
Press the date key on the parameter input page to open
the INFO window.
Status bar:
The flag on the right of the status bar displays the menu
language. Touching the flag icon takes you directly to the
LANGUAGES sub-menu where you can select a different
menu language.
A warning symbol appears at the far left of the status bar
if there is an error. Touching this symbol takes you directly
to the W
sages that are currently active.
ARNINGS sub-menu that displays all warning mes-
26
The name of the loaded configuration/patient record (* indication/patient name) is displayed.
Parameter display:
The treatment parameters are displayed in the parameter display field in the following
sequence:
After the first start-up of the unit as well as after operating mode change, configuration loading and parameter change, the display flashes and must be confirmed by
touching the display field or a parameter.
13 610 0002 11.2022
Operation
Page 28
Selection area
The selection area (see picture below) of the parameter entry screen contains the nominal value selection fields “Energy level”, “Number of shocks” and “Frequency”
Fig. 5-2 Parameter entry screen
–When you open a menu, the name of the opened menu appears in the top line
against a dark blue background. The sub-menu items are indented.
–A sub-menu item is selected by touching the corresponding display area
–The selected sub-menu item appears against a dark blue background.
–Sub-menu items that themselves have an additional sub-menu are identified by a
green arrow to the right (F
–If there are more than 4 menu items, they can be selected using the arrow keys
(FIG. 5-3 /2). If one of the arrow keys disappears, this means no more selections
can be made in this direction.
–Once a sub-menu has been selected, it is opened using the OK button
IG. 5-3 /1).
27
Fig. 5-3 List of the sub-menu items
Operation
13 610 0002 11.2022
Page 29
5.2Function overview
Main menu1st sub menu2nd sub menu
Actual val. resetInfoEnglish
Save configurationLanguageGerman
Load configurationTimeSpanish
SetupTouch-screen calibrationItalian
WarningsDrain waterFrench
Data transferFill waterPortuguese
Exit user interfaceBleed water circuitRussian
1) password protected
Reset water change timePolish
Set mJ/mm² - MPaCzech
Set SW number/energyDutch
Auto-Freq.Norwegian
Warning historySwedish
Software updateDanish
ing the patient record.The keyboard window in the
2nd sub-menu enables you to make your own text entries. However, you can also do this by connecting a
separate USB keyboard.
See 1st sub-menu
–List of current warnings
–Export treatment data (using this sub-menu, it is pos-
sible to transfer the treatment data as files onto a USB
memory stick and open them in Excel)
–Backup settings (backup)
–Restore settings (backup)
•Closes the user interface
28
13 610 0002 11.2022
Operation
Page 30
1st sub-menu
Info–Total shock count and instrument operating hours
(depending on operating mode selected)
–Total number of shocks of the respective handpiece,
data on monitoring software, operating system, hardware serial numbers and modification status
–Information about modules: To view serial numbers
and indexes of the modules, scroll to the second page
of the Info window by using the arrow key.
29
Warning history–List of the last 100 warning and error messages
Language–Setting leads to language submenu
Time–Setting the date and time
Touch screen calibration
Drain water / fill water–The corresponding sequences for emptying or filling
Bleed water circuit–The corresponding sequences for bleeding the water
Reset water change
time
Software update–Transferring new software from the USB memory stick
Set mJ/mm2 - MPa–Changeover between mJ/mm
Set SW number /
energy
Auto-Freq.–Selecting an energy level causes the instrument to
–This function makes it possible to recalibrate the touch
screen, i.e. for correct recognition of the touch coordinates
the water circuit are activated.
circuit are activated.
–Reset the reminder function for the water renewal
2
- MPa for displayed
energy
–Changeover between shock number and total energy
nominal value specification
switch to the maximum frequency automatically.
If this function is not activated, the selected frequency
is not exceeded when the energy level is changed.
However, it is adapted according to the energy level.
Operation
13 610 0002 11.2022
Page 31
5.3Starting the instrument
•Switch the control device on or off at the mains switch on the rear (see FIG. 3-2 ).
The device is operated via the touchscreen display.
WARNING
If a control panel display or a touchscreen / operating monitor should fail, the safety
of the patient can no longer be ensured
Risk of patients being placed under strain due to ineffective treatment or even
impairments to their health!
► Abort the treatment.
► Inform your service centre.
Filling the water circuit:
The first time the instrument is switched on and each time the F-SW handpiece is replaced, the instrument will display the message “Fill water circuit”.
•Touch OK to confirm the message.
The device is operated via the touchscreen display.
The instructions on the display will guide you through the steps required:
– Connect the full water bag
– Filling the water circuit
– Remove the water bag
A detailed description can be found in Chapter 6.3.2 F
ILLINGTHEWATERCIRCUIT
Warm-up phase
Once a day, the Chattanooga Intelect F-SW starts a warm-up phase lasting about 3
minutes, the progress of which is shown in the progress indicator.
The water circuit is bled.
•Check that the F-SW handpiece is correctly positioned in the holder and that no
stand-off is fitted.
Fig. 5-4 Warm-up phase
No F-SW shock triggering is possible during the warm-up phase. All other functions
of the instrument can be used, however.
30
13 610 0002 11.2022
Operation
Page 32
High voltage test
A high voltage test is performed once a day when the Chattanooga Intelect F-SW is
switched on for the first time. This test takes place after the warm-up phase.
•When prompted to do so, briefly touch the trigger button on the F-SW handpiece
or the foot switch.
Fig. 5-5 High voltage test
5.4Setting treatment parameters
31
Once the unit has been started, the display automatically shows the last setting.
•Touch the flashing parameter display or one of the parameter selection fields to
confirm the operating mode.
•Select the line of the parameter that you would like to change.
•Set the value using the arrow keys.
•Release shocks.
The treatment will be carried out with the displayed values.
The maximum possible frequency for generating shock waves depends on the selected
energy level. When increasing the energy level, the shock wave frequency is reduced
if necessary (see Chapter 5.1 GRAPHICALUSERINTERFACE).
The selection of energy levels is based on the medical opinion of the healthcare professional administering treatment. The treatment must never be allowed to cause the
patient to experience any excessive amount of pain.
Operation
13 610 0002 11.2022
Page 33
F-SW
Energy flux density in mJ/mm
2
0.553 Hz
0.503 Hz
0.453 Hz
0.403 Hz
0.354 Hz
0.304 Hz
0.254 Hz
0.205 Hz
0.156 Hz
0.126 Hz
0.106 Hz
0.076 Hz
0.057 Hz
0.038 Hz
0.028 Hz
0.018 Hz
Tab. 5-1 Treatment parameters in F-SW mode
Maximum frequency Handpiece
C-Actor
Energy flux density in mJ/mm
2
1.243 Hz
1.143 Hz
1.023 Hz
0.883 Hz
0.764 Hz
0.694 Hz
0.564 Hz
0.455 Hz
0.336 Hz
0.256 Hz
0.137 Hz
0.088 Hz
0.058 Hz
0.038 Hz
Tab. 5-2 Treatment parameters in C-ACTOR mode
Maximum frequency Handpiece
32
13 610 0002 11.2022
Operation
Page 34
5.5F-SW energy display task
To make sure that the energy level is correctly displayed at all times, the system includes a self monitoring function. Therefore, during shockwave release the system constantly compares the nominal energy value with the actual energy value. If these values
do not match the energy level is displayed in grey and turns white as soon as the required set value is reached.
Fig. 5-6 Energy level not yet reached
If the difference persists shock wave release is disabled and an error message is displayed.
33
Fig. 5-7 Error: Energy level not set
In case the warning appears, you can acknowledge it by touching OK . Inform your
service centre if the fault continues.
Operation
13 610 0002 11.2022
Page 35
5.6Store treatment notes
•Touch the MENU button.
•Select the S
AVECONFIGURATION function as shown in the picture below to save the
current setting of the treatment parameters (FIG. 5-8 STORINGTHETREATMENTPA-
RAMETERS/1).
•Touch the OK button.
Fig. 5-8 Storing the treatment parameters
A list with a total of 100 memory locations appears on the touch screen display in
the SAVECONFIGURATION sub-menu. The system automatically stores the new parameter configurations at the end of the list with the corresponding creation date
and time as shown in the picture below.
•Touch the S
CONFIGURATION
AVE key to save the current setting (see picture below FIG. 5-9 SAVE
SUB-MENU/1).
Fig. 5-9 Save configuration sub-menu
If you select a field that is already occupied, you are asked if you want to overwrite the
content. Confirm by touching OK or revoke your selection by touching the BACK key.
•To rename the configuration, touch the button again that has already been selected. This activates the keyboard window.
34
13 610 0002 11.2022
Operation
Page 36
35
Fig. 5-10 Keyboard window
You can save your parameter setting either as an indication or under a pa-
tient’s name.
•To save the parameters as an indication, place an “*” before the name of the indication or leave it in place (“*Indication name”).
The saved and selected or loaded indication appears in the status bar. This dis-
play disappears if a parameter is subsequently changed
•To save the parameters for a particular patient (patient record), store the setting
directly under the name of the patient (“last name, first name”).
The configuration stored for a patient name is also displayed in the status bar.
The display of patients’ names does not disappear when the parameters are
changed. All parameter changes are logged in a table.
The patient record is closed when:
– a new patient record is called up (loaded),
– an indication is loaded,
– a parameter reset is performed (actual value),
–the unit is switched off.
•Confirm each of your entries by touching the OK button.
•Delete a stored configuration that is no longer required using the D
IG. 5-9 SAVECONFIGURATIONSUB-MENU/3).
(F
Up to 1000 treatments can be stored.
ELETE button
Operation
13 610 0002 11.2022
Page 37
5.7Loading treatment parameters
The alphabetical list of treatment parameters that have already been stored or of the
patient record can be opened either directly from the parameter entry screen or from
the main menu screen
•If you are in the parameter entry screen, touch the button C
(FIG. 5-6 ENERGYLEVELNOTYETREACHED)
•If you are in the main menu, select the LOADCONFIGURATION function from the list
(FIG. 5-8 STORINGTHETREATMENTPARAMETERS/2).
The “Load configuration” menu contains the following indication groups:
In-house applications
Orthopaedic indications
Urological indications
Aesthetics indications
Wound healing
ONFIGURATION .
5.7.1Pre-programmed indications from the manufacturer
•Touch the button on which the required application area is displayed (see picture
below).
•Touch the OK button.
36
Fig. 5-11 Loading a configuration I
•Select the required indication
Fig. 5-12 Loading a configuration II
•Prior to loading an indication, you can view further information on the selected in-
13 610 0002 11.2022
dication.
Operation
Page 38
•To accomplish this, touch NOTE .
The treatment notes will be displayed.
•To load the indication, touch BACK to return to the previous screen.
•Touch L
The indication has been loaded successfully when the loaded indication is dis-
played on the grey status bar (see picture below).
Fig. 5-13 Loaded indication
•To review the treatment notes, touch the name of the indication on the grey status
bar.
The loaded indication is exited by
–Opening a new indication
–Changing a treatment parameter range
–Switching off the instrument.
OAD .
37
5.7.2In-house applications
•Touch the IN-HOUSEAPPLICATION button (see picture below).
•Touch OK .
Fig. 5-14 In-house applications
•Touch the button for the indication required (see picture below).
Operation
13 610 0002 11.2022
Page 39
Fig. 5-15 In-house indications
If additional information for the selected indication has been saved, this can be accessed by touching NOTE .
•To add additional information, touch the text box (see picture below) to display
the on-screen keyboard.
Fig. 5-16 Text box for treatment notes
•Save the text by touching OK .
•Touch the BACK button to view the list of in-house applications.
•Touch the L
The highlighted indication will be loaded. The indication has been loaded suc-
cessfully when the loaded indication is displayed on the grey status bar.
•To review the treatment notes, touch the grey status bar.
The loaded indication is exited by
–Opening a new indication
–Changing a treatment parameter range
–Switching off the instrument
OAD button.
38
13 610 0002 11.2022
Operation
Page 40
5.8Loading patient data
•Touch the IN-HOUSEAPPLICATIONS button (see picture below).
•Touch OK .
•Touch the button on which the required patient name is displayed.
Fig. 5-17 Loading a patient record
•Touch the PROTOCOL button.
The patient record will be displayed.
39
Fig. 5-18 Patient record - treatment details
A patient record consists of treatment details and a table of treatment parameters that is created by the instrument automatically.
Each time a patient is accessed, a new treatment with the current date is saved
to his or her patient record.
Fig. 5-19 Treatment parameters
Operation
13 610 0002 11.2022
Page 41
•To add additional treatment details, touch the text field to display the on-screen
keyboard.
•Save the text by touching OK .
•Touch the B
•Touch the LOAD button.
The treatment parameters for the highlighted patient will be loaded.
The treatment parameters have been loaded successfully when the patient’s
name is displayed on the grey status bar on the protocol screen.
•To review the patient record, touch the grey status bar.
The patient record is closed by
– Opening a new patient record or indication
– Resetting the shock counter
– Switching off the instrument.
ACK button to view the list of in-house applications.
5.9Visual analogue scale (VAS)
The visual analogue scale in the patient record can be used for assessing the progress
of the therapy. The VAS measures the patient’s subjective pain sensation on a scale
from 0 to 10, within which the patient can classify his or her pain intensity. The starting
point (0) stands for “no pain” while the ultimate point (10) stands for the “worst imaginable pain”. In each therapy session, the patient is asked once again to assign a value
to the pain he/she has felt since the last treatment. The reduction in VAS values over
the course of the therapy gives an indication of the success of the treatment.
•Touch and drag the arrow to move it to the point on the scale (see picture below)
where the patient has assigned his or her pain intensity.
•Touch OK to fix the arrow.
40
Fig. 5-20 Setting the VAS value
13 610 0002 11.2022
The arrow can then no longer be moved and the set value appears at the left-
hand edge of the VAS scale.
Fig. 5-21 Set VAS value
Operation
Page 42
5.10Data transfer
Exporting treatment data
Using this function, treatment data can be exported onto a USB memory stick in a format that can be opened in Excel. Also, operating data can be saved (backup) or restored following a repair or if the instrument is replaced.
•Ensure that your USB memory stick supports the USB V1.1 protocol. You can order
a validated USB stick from your dealer
•Load a patient-specific parameter record.
•Select the DATATRANSFER / EXPORTTREATMENTDATA function in the 1st sub-menu
(see picture below).
Fig. 5-22 Data export
41
•Connect the memory stick to the USB port as soon as you are prompted to do so
(see picture below) and confirm by touching OK .
Fig. 5-23 Data export II
Operation
13 610 0002 11.2022
Page 43
Fig. 5-24 Establishing the USB connection
The USB connection is established.
The data is transferred once the USB connection has been established. The export file name of the patient record is protocol_name.csv.
All data is exported if no patient record or no indication has been opened.
The export file name of the record data is protocol_DateTime.csv
•Wait until the “Export completed” message appears on the display (see picture below), then remove the memory stick.
Fig. 5-25 Data export complete
Backing up the settings
42
Using the B
indication data onto a USB memory stick as a backup (in a file format that can only be
read by the instrument).
•Select the DATATRANSFER / BACKUPSETTINGS function in the 1st sub-menu (FIG. 5-
•Connect the memory stick to the USB connector as soon as you are prompted to
•Remove the USB memory stick.
13 610 0002 11.2022
ACKUPSETTINGS function, you can save configuration settings, patient and
22 /2).
do so and confirm by touching OK .
After the USB connection has been established, the data backup is performed
and the text window shows the name of the backup file.
Operation
Page 44
Restoring the settings
The system is restored to the data status of the last backup using the RESTORESETTINGS
function.
•Select the D
22 /3).
•Connect the memory stick with the backup file to the USB port as soon as you are
prompted to do so and confirm by touching OK .
The backup file is loaded onto the system once the USB connection has been
established. You are prompted to restart the system when the loading procedure has finished.
•Remove the USB stick and restart the instrument.
ATATRANSFER / RESTORESETTINGS function in the 1st sub-menu (FIG. 5-
5.11Resetting the treatment shock counter
•To reset the applied shock counter to “0”, select the ACT. VAL. RESET menu option
(see picture below) or touch the counter display.
43
Fig. 5-26 Resetting the treatment shock counter
Operation
13 610 0002 11.2022
Page 45
5.12"Autofrequency" function
If the autofrequency function is activated, the frequency is automatically increased to
the maximum possible setting when the energy level is reduced in F-SW mode (see
chapter 5.4 S
MODE).
•Select the F-SW operating mode if this function should be deactivated.
•Press the AUTOFREQ. [ON] item in the SETUP menu (see Fig. FIG. 5-27 /1) to deactivate the autofrequency function.
Fig. 5-27 AUTOFREQ. [ON] item
The instrument automatically changes to AUTOFREQ. [OFF] status (see FIG. 5-28
Now, the selected frequency remains the same even if the energy level is
changed.
•Touch the EXIT button to return to the main menu.
This frequency can be reduced manually.
5.13Start-up
•Switch the instrument on as described in chapter 5.3 STARTINGTHEINSTRUMENT.
CAUTION
Equipment damage due to improper storage and transport
may affect the health of patients and users.
► Before commissioning, check that the delivered items are undamaged.
44
13 610 0002 11.2022
Operation
Page 46
•Check that there are no bubbles in the handpiece.
If air bubbles are visible under the coupling diaphragm, proceed as follows:
•Position the handpiece in the handpiece holder.
This ensures that air bubbles will always be sucked out of the handpiece auto-
matically.
•Leave the handpiece in this position for approx. 3 minutes until the suction procedure has finished.
45
•To work in F-SW mode, set the energy of the shocks to an initial value of 0.1 mJ/
mm
2
.
The maximum energy level corresponds to an energy flux density of 0.55 mJ/mm
•Optional: To work in C-ACTOR mode, set the energy of the shocks to an initial
value of 0.33 mJ/mm2.
The maximum energy level corresponds to an energy flux density of 1.24 mJ/mm
The highest permitted frequency is always set when an energy level is selected (see
chapter 5.7 LOADINGTREATMENTPARAMETERS). This frequency can be reduced manually.
•Press the trigger button.
The trigger button functions as an on/off switch when it is pressed briefly (< 1.5 s).
Pressing it for longer (> 1.5 s) causes it to function as a tip switch, i.e. the shocks will
continue until the button is released.
If a nominal shock wave value of less than 1000 shock waves is selected (e.g. 400
shock waves), a window with the following text appears after the nominal value has
been reached: “Number/energy set value reached”.
The message can be acknowledged by touching the OK button or the corresponding
trigger button. Further treatment is possible.
This message is activated again as soon as a multiple of the set nominal value is
reached (e.g. 800 shock waves, 1200 shock waves, etc.).
If a nominal value above 1000 shock waves is selected (e.g. 1700 shock waves), the
instrument automatically triggers a safety stop at 1000 shock waves (see picture below). The next stop occurs when the set nominal value is reached. Following this, the
counter continues to stop at intervals of 1000 (e.g. 2700, 3700, etc.).
2
.
2
.
Operation
If the nominal value is 0 (displayed as “ - ”), the stop only occurs at 19,999 shocks.
Fig. 5-29 Safety stop
13 610 0002 11.2022
Page 47
5.14Functional checks
Perform the following functional checks after the system has been installed:
•Check the control unit and handpieces for damage.
•Start the Chattanooga Intelect F-SW (see chapter 5.13 S
•Set the energy level in F-SW mode to 0.2 mJ/mm2.
Optional: Set the energy level in C-ACTOR mode to 0,69 mJ/mm2.
•Reset the actual number of shocks on the parameter display of the control panel.
See chaper 5.11 RESETTINGTHETREATMENTSHOCKCOUNTER
•Release shocks with a shock frequency of 4 Hz.
•Release shocks by means of the foot switch, if used
•Check that the triggered shocks are correctly counted on the treatment shock
counter.
If necessary, the functional capability of the F-SW handpiece can be checked with the
aid of special Colour sensitive pressure sensors (see chapter 8 ACCESSORIES).
5.15Standard settings
•Before each treatment, make sure that the shock counter is set to ‘0’.
Set the nominal value counter to the required value. The "-" symbol appears if zero is
selected. The instrument then operates without a nominal value specification.
•Start the F-SW treatment at a energy level of 0.1 mJ/mm
•Start the C-ACTOR treatment at a energy level of 0.33 mJ/mm2 and a frequency
of 6 Hz.
TART-UP).
2
and a frequency of 6 Hz.
46
5.16Treatment
Safety information
Before using the device, the user must make sure it is functioning safely and in proper
condition.
•Each time after the device has been transported, make sure that all functional
checks have been performed on the device before you start treatment. Read about
this in chapter 5.14 F
•Read chapter 1 GENERAL SAFETY INFORMATION before beginning treatment.
CAUTION
If the handpiece is not positioned correctly, there is an
impairment to health due to ineffective treatment!
► Define the treatment zone and make sure that the handpiece position always cor-
responds to the treatment zone.
► Make sure that the treatment is only administered by users who meet the condi-
tions in Chapter 2.2 PRECONDITIONSFOROPERATION.
UNCTIONALCHECKS.
13 610 0002 11.2022
Operation
Page 48
For safety reasons, using the device for applications other than those specified in chapter 1 GENERAL SAFETY INFORMATION is not permitted!
•Apply a sufficient amount of coupling gel to the patient’s skin in the treatment
area and to the F-SW coupling diaphragm or the stand off device.
CAUTION
Cleaning and maintenance work during treatment may lead
to injuries to patients and therapists.
► No cleaning and maintenance work is to be carried out while the device is being
used on the patient.
•All status and error messages signaled during treatment must always be attended
to without delay!
5.17Switching Off the Device
•Switch off the Chattanooga Intelect F-SW using the main switch.
• If the system is transported or is unused for a longer period of time, always switch
it off completely, using the button on the backside.
47
Operation
13 610 0002 11.2022
Page 49
6Cleaning, care and maintenance
CAUTION
Injuries to patients and people administering the treatment
► No cleaning and maintenance work is to be carried out while the device is being
used on the patient.
6.1Cleaning the device
Regular cleaning ensures perfect hygiene and operation of the Chattanooga Intelect
F-SW.
The frequency of complete exterior cleaning depends on the frequency of use and
what the device is used for.
DANGER
Electrical hazard
Disconnect the device from the mains before starting any cleaning, maintenance or overhaul work.
► Disconnect the mains plug.
DANGER
Cleaning agents and disinfectants
can form an explosive atmosphere.
► Disconnect the handpiece from the control unit before starting any cleaning or
maintenance work.
► Make sure that treatments and cleaning work are carried out only by qualified and
instructed healthcare professionals.
WARNING
On contact with contaminated surfaces
there is a risk of transmission of infection.
► Clean all parts which come into contact with the patient before and after each
treatment.
48
13 610 0002 11.2022
Cleaning, care and maintenance
Page 50
•Wipe down the device parts with a damp cloth.
For cleaning, use a lukewarm, diluted solution of non-vegetable soapy water.
DANGER
► It is essential that no fluid be permitted to penetrate either the device or its hoses.
Ventilation slots
•Keep the ventilation slits clear.
Monitor and Touch Screen
Only a cloth moistened with water, without any cleaning additives, may be used to
clean the LC displays.
•Stand-off devices must be changed for treatment and service reasons. Check before treatment that the stand-off device is tightened firmly.
49
1To attach the stand-off device
2To release the stand-off device
3Stand-off
4Closing ring
Fig. 6-1 Attaching the stand-off device
•To release: Press the closing ring towards the rear and then unscrew it.
Fig. 6-2 Releasing the closing ring
Cleaning, care and maintenance
13 610 0002 11.2022
Page 51
The stand-off has a limited service life. It should be replaced if there are visible changes
in the material (discolouration, tarnishing, streaks, gas bubbles), deformation of the
surface in the coupling area or leaks.
The stand-off device should be replaced at least every 12 months.
1stand-off I
2stand-off II
3anatomic stand-off (optional)
Fig. 6-3 Stand-offs without ring
•Fit a clean stand-off into the closing ring.
•Apply a drop of silicone oil to the coupling diaphragm as a coupling medium.
•To attach: Screw the closing ring towards the front and tighten it firmly with your
hand.
6.2.2Reprocessing the handpiece and the stand-off devices
After each therapy session all parts of the handpiece which have been in contact with
the patient must be thoroughly cleaned and disinfected for further treatments. Therefore the instruction must be strictly followed in order to avoid damage to the parts and
prevent malfunctions.
Make sure that the following means and tools are available for cleaning and disinfection:
The constituents listed here are non-binding examples. No claims are made regarding
the completeness of the list.
6.2.3Cleaning the optional foot switch
•Clean the foot switch with soapy water or a mild cleaning agent.
The foot switch is protected against ingress of water according to classification IPX8
as per IEC 60529.
Cleaning, care and maintenance
13 610 0002 11.2022
Page 53
6.3Water renewal
The water in the cooling circuit of the Chattanooga Intelect F-SW should be renewed
every 6 months or so. The instrument automatically displays a message to this effect
when it is switched on if the water renewal is due (see the picture below).
Fig. 6-4 Prompt for water renewal
•Touch the OK button to acknowledge this message.
The message no longer appears once the water has been renewed.
6.3.1Draining the water circuit
The water circuit must be drained if the instrument will not be used for several weeks.
•Make sure that the instrument is standing on a smooth surface.
•Activate DRAINWATER operating mode in the SETUP menu (see picture below).
Fig. 6-5 Water renewal
•Connect the water bag to the Chattanooga Intelect F-SW and put it on the floor
as soon as the message appears.
52
13 610 0002 11.2022
Cleaning, care and maintenance
Page 54
53
Fig. 6-6 Draining the water circuit I
The “Please wait” message and a progress indicator appear on the display.
•Allow the remainder of the water to drain out of the handpiece by holding the FSW handpiece vertically above the instrument as soon as you are prompted to do
so. Make sure that the coupling diaphragm of the handpiece is pointing upwards.
Fig. 6-7 Draining the water circuit II
The “Please wait” message and a progress indicator appear on the display.
•Wait until the instrument is ready. The display shows when the water circuit is
empty.
Cleaning, care and maintenance
Fig. 6-8 Draining the water circuit III
•Open the lock on the tube connection and pull the tube out of the tube connector.
•Remove the full water bag and dispose of the contents.
13 610 0002 11.2022
Page 55
6.3.2Filling the water circuit
•Make sure that the instrument is standing on a smooth, horizontal surface.
•Use only deionised water (in compliance with VDE 0510, e.g. water for batteries
or clothing irons) to rinse or fill the water bag.
•Fill the water bag to the brim ( minimum 800ml or 27oz.). In order to add the maximum amount of water to the water bag, place the bag on the palm of your hand
and squeeze it gently at the sides whilst filling.
Fig. 6-9 Filling the water bag
Do not use water that has been distilled more than once!
After the water bag has been filled, there should be as few bubbles as possible in
the connection tube.
•Hold the water bag up with the connection tube hanging vertically downwards to
let the air rise automatically into the water bag.
•Place the F-SW handpiece into the F-SW handpiece holder so that any air bubbles
that form will be immediately sucked up by the bubble trap.
Fig. 6-10 Handpiece in holder
•Activate FILLWATER operating mode in the SETUP menu.
54
13 610 0002 11.2022
Cleaning, care and maintenance
Page 56
55
Fig. 6-11 Filling the water circuit I
• Connect the water bag to the water tube connection on the rear of the instrument as soon as the message appears.
•At the same time, hold the water bag at least 70 cm / 28" above the water tube
connection, so the water can flow out optimally. Hook the bag onto an infusion
stand if necessary.
•Touch OK .
Cleaning, care and maintenance
Fig. 6-12 Filling position and progress display
A progress display with the message “Please wait” appears on the display.
•As soon as the water circuit has been filled, the instrument prompts you to remove
the water bag (see picture below). There may be water left in the bag.
13 610 0002 11.2022
Page 57
Fig. 6-13 Filling the water circuit II
•Push the lock on the tube connector and pull the tube out of the connection.
•Confirm by touching OK .
There might be air bubbles in the system after the water has been changed.
The instrument needs about 15 minutes to remove these air bubbles. A progress bar will be displayed (see FIG. 6-14 ).
Fig. 6-14 Bleeding the water circuit
•Wait for the message to disappear, then return to the parameter entry screen by
touching the MENUEXIT button.
•Check that there are no bubbles under the coupling diaphragm of the F-SW handpiece. If bubbles are present, briefly hold the handpiece pointing downwards in a
vertical position. The air bubbles will then be automatically sucked in by the bubble
trap.
56
6.3.3Bleeding the water circuit
•Select BLEEDWATERCIRCUIT from the SETUP menu (see FIG. 6-6 DRAININGTHEWA-
TERCIRCUIT I, /3).
A progress bar will be displayed.
•Wait for the message to disappear, then return to the parameter entry screen by
touching the M
•Check that there are no bubbles under the coupling diaphragm of the F-SW handpiece. If bubbles are present, briefly hold the handpiece pointing downwards in a
vertical position. The air bubbles will then be automatically sucked in by the bubble
trap.
13 610 0002 11.2022
ENUEXIT button.
Cleaning, care and maintenance
Page 58
6.3.4Resetting the water renewal time
2
1
Every six months, the instrument prompts you to renew the water; the prompt does
not disappear permanently until the water has been renewed.The RESETWATERCHANGE
TIME
function can be selected to cancel this reminder function or to adapt it to a new
date setting.
57
•Activate R
ESETWATERCHANGETIME operating mode in the SETUP menu.
The time when the water change time reminder is triggered is automatically
moved forwards by six months. A window showing the new date of water renewal appears briefly on the display.
•Press the EXIT button to open the parameter entry screen.
Failure to renew the water regularly may shorten the service life of the instrument.
6.4Mains fuse replacement
The mains fuse holder is located on the rear of the Chattanooga Intelect F-SW between the mains connection and the ON/OFF switch.
•Push the clip of the mains fuse holder to the right and take the holder off the housing.
Fig. 6-15 Mains fuse holder
Fig. 6-16 Fuse replacement
1Fuse holder
2Fuse
Cleaning, care and maintenance
•Pull the old fuses out of the mains fuse holder.
•Replace the fuses (T5AH / 250 VAC).
•Push the mains fuse holder back into the opening until it engages.
13 610 0002 11.2022
Page 59
6.5Cybersecurity measures
Like all computer-based systems, the system might be exposed to cybersecurity
threats.
Safety critical core functions of the device, such as shock wave release are decoupled
from software and can be activated by hardware switches only.
In order to minimize the possibility of cyber attacks, it is the user’s responsibility to
make sure that the following protection measures are followed.
1) The product may only be installed, commissioned, maintained, updated and
operated by authorised personnel. These persons are to be employees of the
manufacturer or other authorised third parties.
2) Software updates distributed by the manufacturer (functional or safety-relevant)
must be installed via the enclosed Technical Product Information (TPI). Subsequent
feedback to the manufacturer or its service partner is required.
3) A regular data backup of the system must be carried out. Depending on the number of patients, this should be done daily, but at least weekly.
4) Software that is not distributed by the manufacturer or one of its service partners
may not be used with the product.
5) A virus scan of the USB sticks used must be carried out to check whether they are
free of viruses, malware or dangerous software.
6) All service and maintenance areas are password protected. Such access can only
be granted by the manufacturer to authorized service personnel.
Contact the manufacturer or the authorised customer service in the following
cases:
–In case the product shows unknown or not logical behavior, such as a slow reac-
tion of the software or if the password is not accepted, access to the databases is
not possible or it switches to a wrong user interface dialogue.
–In case you face issues with the IT security of this product.
–In case you lost your password, login credentials or user access.
58
6.6Software updates
For software update, please contact your local dealer.
6.7Maintenance and safety checks
Preventive maintenance is not necessarily required. However, regular maintenance
may help to identify possible defects at an early stage and thus increase the safety and
service life of the device.
Maintenance services can be ordered from our regional representatives in your area.
We recommend that functional and safety checks be performed at least once a year
(see Chapter 5.14 FUNCTIONALCHECKS), in addition to the national accident prevention
regulations and test and inspection intervals prescribed for medical devices that are required to be observed.
The following checks should be performed to ensure that the Chattanooga Intelect FSW operates safely:
1. Earth leakage current test from the chassis in accordance with national regulations.
2. Earth impedance test (including housing and with mains cable) in accordance with
national regulations.
13 610 0002 11.2022
Cleaning, care and maintenance
Page 60
3. Test of essential performance.
For further details on content and performance of the safety checks please contact
your local dealer.
6.8Repair
Repair work on defective devices must only be carried out by personnel suitably authorised by the manufacturer. Only original parts from the manufacturer may be used for
this purpose.
The suitably authorised personnel can be representatives or its agencies and dealers.
6.9Disposal
When disposing of this medical product, no special measures have to be observed.
Please proceed in accordance with applicable country-specific regulations.
After expiration of the service life of the device, dispose of the Chattanooga Intelect
F-SW as waste electronic equipment.
•Please contact the manufacturer or distributing company in relation to this.
•When disposing of wear parts, you must comply with the relevant national disposal regulations.
59
•Comply with the relevant information in the operating manuals for the additional
devices.
6.10Service life
The average expected service life (MTTF) in accordance with IEC 60601-1:2005 +
A1:2012 / EN 60601-1:2006 + A1:2013 is
–15,000 operating hours for the Chattanooga Intelect F-SW
–5 million shocks for the F-SW SEPIA LT handpiece
–5 million shocks for the C-ACTOR SEPIA LT handpiece
The coil is a wear part. After 2 million shocks the coil needs to be exchanged to avoid
secondary damages on the handpiece and on the control device.
Exceeding the service life can be expected to result in a failure of the device and accessories. This also applies to handpieces.No warranty claims shall be accepted beyond
the information given in chapter 10 W
ARRANTYANDSERVICE.
Cleaning, care and maintenance
13 610 0002 11.2022
Page 61
7Status messages, error messages / fault displays
7.1Status messages
CAUTION
Malfunction of the device or its components.
Various injuries are possible!
► Immediately comply with all status and error messages which appear during the
treatment.
Fault descriptionPossible cause and remedy
Specified number of shock waves
reached
Shock wave safety stopAcknowledge message, further treat-
F-SW : load test unsuccessfulRestart the instrument and repeat the
F-SW : charging timeoutAcknowledge message.
F-SW : water temperature too highAcknowledge message,
F-SW : water temperature too lowAcknowledge message,further treat-
F-SW : water level too lowFill the water circuit (see chapter 6.3.2
F-SW : water circuit faultPump defect
F-SW : pump temperature too highAcknowledge message,
F-SW : therapy head overtemperatureAcknowledge message,
F-SW : water temperature sensor failureRestart the instrument.
Acknowledge message, further treatment is possible.
ment is possible.
test.
Do not continue to use the instrument if
the fault continues.
Notify your Service centre.
Inform your Service centre if the fault
continues.
further treatment is possible once the
water temperature has returned to permitted values.
ment is possible once the water temperature has returned to permitted values.
ILLINGTHEWATERCIRCUIT)
F
Treatment is not possible.
Bleed the water circuit (see chapter 6.3.3
BLEEDINGTHEWATERCIRCUIT).
Inform your Service centre if the fault
continues.
further treatment is possible once the
pump temperature has returned to permitted values.
further treatment is possible once the
therapy head temperature has returned
to permitted values.
Inform your Service centre if the fault
continues.
60
13 610 0002 11.2022
Status messages, error messages / fault
Page 62
Fault descriptionPossible cause and remedy
Shock wave limit for current handpiece
reached
F-SW : charging unit not readyAcknowledge message.
USB stick was not recognisedRemove the USB stick, then switch off
Water amount insufficient, please check
supply
Water pump malfunctionRestart the instrument. Drain and re-fill
Water pressure lowRestart the instrument. Call your Service
Shock wave limit for current handpiece
reached.
Contact your service center for coil replacement.
Call your Service centre if the fault continues after a reset.
and restart the instrument.
Reinsert the USB stick.
Check that there is software on the USB
stick.
If the fault persists, check that the USB
stick supports the USB V1.1 protocol.
If it does not, replace the USB stick.
Fill the water bag and check whether water is flowing into the water circuit.
If the fault persists, inform your Service
centre.
water.
Call your Service centre if the fault continues. .
centre if the fault continues.
61
Status messages, error messages / fault
13 610 0002 11.2022
Page 63
7.2Trouble shooting
DANGER
Electrical hazard
Disconnect the device from the mains before starting any cleaning, maintenance or overhaul work.
► Disconnect the mains plug.
Fault descriptionPossible causeCorrective actions
Device does not
work.
No F-SW power output
No F-SW power output
Shock triggering
noise changes after
several shocks
Tab. 7-1 Trouble shooting
Power failure• Check the power supply.
Defective mains fuse• Replace the fuses.
Defective mains plug• Replace the mains cable.
F-SW handpiece defective
Malfunction in control device
Handpiece has not
been recognised
Air in handpiece• Hold handpiece vertically with
• Replace the handpiece.
• Call your service centre
• Check that the black screw is
screwed to the correct tightness.
coupling diaphragm downwards
so that air is sucked out.
62
13 610 0002 11.2022
Status messages, error messages / fault
Page 64
8Accessories
Product DescriptionPart Number
Handpiece F-SW SEPIA LT19000
Handpiece C-ACTOR SEPIA LT29204.0001
Water bag4600
Silicone oil4700
Gel bottle22601
Stand-off I 15 mm19100
Stand-off II 30 mm19200
Stand-off anatomical19150
Closing ring19300
Handpiece holder set13-27268
User manual13-00061
Color sensitive pressure sensors13-18342
Mains cable EU0.0032.012
Trolley/Cart4560
63
Accessories
13 610 0002 11.2022
Page 65
9Technical specifications
Chattanooga Intelect F-SW
F-SW operating mode
C-ACTOR operating mode
F-SW energy selectionin steps from 0.01 to 0.55 mJ/mm
C-ACTOR energy selection in steps from 0.03 to 1.24 mJ/mm
Mains input voltage100 – 240 VAC
Mains frequency50 / 60 Hz
Mains fuseT5AH / 250 VAC
Power consumption max. 450 VA
Ambient temperature during
operation
Ambient temperature during
storage and transport
Ambient air pressure during
operation
Ambient air pressure during
storage and transport
Air humidity during operation5 - 55%, non condensing
The number of the software version of the device can be seen on the touch panel.
Pressing INFO / VERSIONS shows the actual software und hardware versions.
0–65 mm
0–45 mm
0–30 mm
In the event of the medical product being transferred to third parties, the following
must be observed:
–The complete device documentation must be delivered together with the medical
product.
–The medical product may only be exported to a foreign country if the medical
product and the corresponding indications are allowed there.
9.1Type plate
Fig. 9-1 Type plate Chattanooga Intelect F-SW
9.2Conformity with directives
This medical product bears the CE mark in accordance with the Medical Device Directive (MDD) 93/42/EEC.
66
13 610 0002 11.2022
Technical specifications
Page 68
67
9.3Conformity with standards
This device complies with the applicable standards EN/IEC 60601-1, CAN /CSA-C22.2
No.601.1, UL Std. No 60601-1.
Acc. to EN/IEC 60601-1
–Type of protection against electric shocks:Protection class 1
–Applied part of type B
The following essential performance characteristics apply to the Chattanooga Intelect
F-SW:
1. The Medical Equipment shall be free from incorrect display of energy levels.
2. The Medical Equipment shall be free from unintended shock wave release.
The essential performance characteristics of the Chattanooga Intelect F-SW cannot be
disturbed by electromagnetic interference (EMI).
9.3.1EMC guidelines and manufacturer's declaration
Guidelines and manufacturer's declaration Emitted electromagnetic interference
The Chattanooga Intelect F-SW model is intended to be used in the electromagnetic environment
specified below. The customer or the user of the Chattanooga Intelect F-SW should ensure that it
is used in such an environment.
The maximum length of the mains cable for the unit is 3 m.
Interference emission measurements
HF emissions acc. to
CISPR 11
HF emissions acc. to
CISPR 11
Harmonic emissions
according to IEC
61000-3-2
Voltage fluctuations/
flicker emissions acc.
to IEC 61000-3-3
Conducted RF emissions
Radiated RF emissions
Compliance
Group 1
Class B
Class A
Complies
Class B
Class B
Electromagnetic environment – guidelines
The Chattanooga Intelect F-SW uses HF energy only for
its internal functioning. Therefore, its HF emissions are
very low and are not likely to cause any interference in
nearby electronic equipment. As per EN IEC 60601-236:2015 Section 202, this does not apply when triggering and generating the pressure pulse.
The Chattanooga Intelect F-SW is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Technical specifications
13 610 0002 11.2022
Page 69
Guidelines and manufacturer's declaration Resistance to emitted electromagnetic in-
terference
The Chattanooga Intelect F-SW model is intended to be used in the electromagnetic environment
specified below. The customer or the user of the Chattanooga Intelect F-SW should ensure that it
is used in such an environment.
Immunity
tests
Electrostatic
discharge
(ESD) acc. to
IEC 61000-4-2
Electrical fast
transient disturbances /
bursts according to IEC
IEC 60601
test level
±8 kV contact
discharge
±15 kV air discharge
±2 kV for
mains cables
±1 kV for input/output
lines
Compliance
level
±8 kV contact
discharge
±15 kV air discharge
±2 kV for
mains cables
±1kV for input/output
lines
Electromagnetic environment – guidelines
Floors should be made of wood or concrete, or
be covered with ceramic tiles. If floors are covered with synthetic material, the relative air humidity must be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
61000-4-4
Surges according to IEC
61000-4-5
Voltage drops,
short interruptions and voltage variations
on power supply input lines
acc. to IEC
61000-4-11
Power
frequency (50/
60 Hz) magnetic field acc.
±1 kV line(s)
to line(s)
±2 kV line(s)
to earth
< 5% U
a
T
(> 95% drop
in UT) for ½
and 1 period
70 % U
T
(30% drop in
UT) for 25/30
periods
< 5% UT
(> 95% drop
) for 250/
in U
T
300 s
±1 kV line(s)
to line(s)
±2 kV line(s)
to earth
< 5% U
(> 95% drop
in UT) for ½
and 1 period
70 % U
(30% drop in
UT) for 25/30
periods
< 5% UT
(> 95% drop
) for 250/
in U
T
300 s
30 A/m30 A/m
Mains power quality should be that of a typical
commercial or hospital environment.
T
Mains power quality should be that of a typical
commercial or hospital environment.
If the user of the Chattanooga Intelect F-SW requires continued operation during mains power
T
interruptions, it is recommended that the Chattanooga Intelect F-SW be powered from an uninterruptible power supply or a battery.
The mains frequency magnetic fields should be
those of a typical business or hospital environment.
to IEC 610004-8
a. NOTE: UT is the mains alternating voltage prior to application of the test level.
68
13 610 0002 11.2022
Technical specifications
Page 70
69
Guidelines and manufacturer's declaration Resistance to emitted electromagnetic in-
terference
The Chattanooga Intelect F-SW model is intended to be used in the electromagnetic environment
specified below. The customer or the user of the Chattanooga Intelect F-SW should ensure that it
is used in such an environment.
Immunity
tests
IEC 60601
test level
Compliance
level
Electromagnetic environment – guidelines
Portable and mobile RF equipment should be
used no closer to any part of the Chattanooga
Intelect F-SW, including cables, than the recommended safety distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommended safety distance:
Conducted RF
according to
IEC 61000-4-6
Radiated HF
interference
according to
IEC 61000-4-3
/6V
3V
RMS
RMS
150 kHz to 80
MHz
10 V/m
80 MHz to 2.7
GHz
3V
/6V
RMS
150 kHz to 80
MHz
10 V/m
80 MHz to 2.7
GHz
d = 1.2√ P
RMS
d = 1.2√ P
for 80 MHz to 800 MHz
d = 2.3√ P
for 800 MHz to 2.7 GHz
Where P is the rated power of the transmitter in
watts (W) according to the transmitter manufacturer and d is the recommended safety distance
in metres (m).
The field intensity of stationary radio transmit-
a
ters, based on an on-site inspection
less than the compliance level.
should be
b
Interference may occur in the vicinity of devices
marked with the following symbol:
NOTE 1:
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment with respect to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength at the location in which the Chattanooga Intelect F-SW is used exceeds the applicable HF compliance
level indicated above, the Chattanooga Intelect F-SW should be observed to verify normal operation. If abnormal performance
is observed, additional measures may be necessary, such as re-orienting or relocating the Chattanooga Intelect F-SW.
b. Over the frequency range of 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Technical specifications
13 610 0002 11.2022
Page 71
Recommended safety distances between portable and mobile HF communications equip-
ment and the Chattanooga Intelect F-SW
The Chattanooga Intelect F-SW is intended for use in an electromagnetic environment in which radiated HF disturbances are controlled. The operator or the user of the Chattanooga Intelect F-SW
can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile HF communications equipment (transmitters) and the Chattanooga Intelect F-SW,
as recommended below,
according to the maximum output power of the communications equipment.
Rated power of
transmitter [W]
0.010.120.120.23
0.10.380.380.73
11.21.22.3
103.83.87.3
100121223
For transmitters rated at a maximum output power not listed above, the recommended safety distance can be estimated using the equation applicable to the frequency of the transmitter, where P
is the rated power of the transmitter in watts [W] according to the transmitter manufacturer.
NOTE 1:
An additional factor of 10/3 was used for calculating the recommended safety distance of transmitters in the frequency range from 80 MHz to 2.7 GHz in order to reduce the probability that a
mobile/portable communications device brought into the patient area might inadvertently lead to
a malfunction.
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Safety distance according to transmission frequency [m]
150 kHz to 80 MHz
d = 1.2 √P
80 MHz to 800 MHz
d = 1.2 √P
800 MHz to 2.5 GHz
d = 2.3 √P
70
13 610 0002 11.2022
Technical specifications
Page 72
9.4Certificates
71
Fig. 9-2 Declaration of Conformity
Technical specifications
13 610 0002 11.2022
Page 73
9.5Symbols and labels
The following symbols and labels are affixed to the Chattanooga Intelect F-SW:
Fig. 9-3 Symbols and labels at the front cover
1Foot switch connection
2F-SW handpiece connection
3Applied part of type B
LabelNotification
1 foot switch connection
2 F-SW handpiece connection
3 Applied part of type B
Tab. 9-1 Symbols attached to the front side
72
13 610 0002 11.2022
Technical specifications
Page 74
73
Fig. 9-4 Symbols and labels at the back cover
1Potential equalisation
2USB-connection
3Comply with the OM
4Type plate
LabelNotification
1 Potential equalisation
2 USB-connection
3
4
Tab. 9-2 Symbols attached to the rear side
It is essential to comply with the
operating manual
Type plate
Technical specifications
13 610 0002 11.2022
Page 75
The following symbols and labels are on the type plate of the Chattanooga
Intelect F-SW
LabelMeaning
fuse
Applied part of type B
CSA certification
manufacturer
CE mark (in compliance with Medical Device
Directive (MDD) 93/42/EEC)
DataMatrix according to GS1 standards
Unique Device Identification (UDI): Human
Readable Interpretation (HRI)
(01)Global Trade Item Number (GTIN)
(21)Application Identifier (AI): Serial Number (SN)
(11)Application Identifier (AI): Production Date
(PRODDATE)
Tab. 9-3 Device labelling
The following symbols and labels are attached to the Chattanooga Intelect F-SW packaging:
LabelMeaning
Ambient temperature during storage and
transport
Air humidity during storage and transport
Ambient air pressure during storage and
transport
74
Tab. 9-4 Labelling on the packaging
13 610 0002 11.2022
Technical specifications
Page 76
75
10Warranty and service
10.1Warranty for the control device
During the two-year warranty period from the date of invoice, defects will be either
replaced or repaired at the discretion of the manufacturer and at no charge to the customer upon the customer furnishing adequate proof that the defect is due to defects
in material or workmanship. The warranty does not extend to wear parts.
Transport costs and the risk of loss during the shipping of returned products shall be
borne by the customer.
NOTICE
Unauthorised opening, repairs and modifications to the device by unauthorised persons
releases the manufacturer from any liability for the safe system operation.
This will automatically void the warranty even before the end of the warranty
period.
► Do not open the device without authorisation and do not manipulate the device.
10.2Warranty for the handpiece
The F-SW handpiece and the C-ACTOR handpiece are wear parts.
New handpieces that have performed up to 1 million pulses will be either replaced or
repaired at the discretion of the manufacturer and at no charge to the customer upon
the customer providing adequate proof that the defect is due to defects in material or
workmanship.
Transport costs and the risk of loss during the shipping of returned products shall be
borne by the customer.
Warranty claims will only be accepted if the handpiece is returned in its complete and
original state, cleaned and in the case, with the repair label filled in completely.
Missing components will be replaced subject to charge.
Accessories also sent will be checked and, if necessary, replaced after we have assessed
them.
The coil is a wear part. It is not covered by the handpiece’s warranty.
Modifications to the device are not permitted.
Any opening, repair or modification of the handpiece and the stand-off devices are not permitted. Any opening, repair or modification of the instruments
by unauthorised personnel will relieve the manufacturer of its liability and responsibility for safe system operation. This will automatically void the warranty even before the end of the warranty period.
10.3Service
If you have any other questions about the Chattanooga Intelect F-SW, please contact
your certified dealer.
Warranty and service
NOTICE
13 610 0002 11.2022
Page 77
Medical devices identified with this CE mark are in accordance with the regulations in EC Directive 93/42/EEC “Council Directive concerning medical devices”
The operating manual, including all of its parts, is protected by copyright. Any use outside the narrow limits of copyright law without the written consent of the manufacturer is inadmissible and liable to prosecution.
This applies in particular to duplication, translation, microfilming as well as feeding and
distribution in electronic systems.