Important User Information 3-4
Section 1 General Information
1.1 Introduction 5
1.2 Specifications 5
1.3 Safety 6
Section 2 Description of parts
2.1 Unpacking—U.S. Version 7
2.2 Components/Assembly—U.S. Version 7
2.3 Unpacking---European Version 8
2.4 Components/Assembly---European Version 8
Section 3 Instructions for Use
3.1 Slab Gel Unit Preparation 9
3.2 Preparation/Cleaning of glass plates 9
3.3 Gel Casting 9-11
3.4 Running the Gel 11-12
3.5 Removing the Gel 12
3.6 Maximum Well/Comb volumes 12
Section 4 Running Conditions
4.1 Recommended Power 13
4.2 Recommended Buffer & Reagents 13-14
4.3 References 14
Section 5 Maintenance of Equipment
5.1 Care and Handling 15
5.2 Maintenance 15
Section 6 Slab Gel Kits and Accessories 16-19
C. B. S. Scientific 2 C.B.S. LITE Slab Gels
Page 3
ENGLISH
IMPORTANT USER INFORMATION
This Instruction Manual will explain how to use this product safely and effectively. Please read and
carefully follow the instruction manual in its entirety.
The triangle/exclamation mark symbol alerts the user of the product to important
operational, maintenance, and/or warranty requirements.
The triangle/lighting bolt symbol alerts the user of the product to potentially hazardous
Failure to adhere to the instructions could result in personal and/or laboratory hazards, as well as
WARRANTY AND LIABILITY
This product was produced utilizing the highest practical standards of materials, workmanship, and
design. C.B.S. Scientific warrants that the product has been tested and will meet or exceed published
specifications. This warranty is valid only if the product has been operated and maintained according to
the instructions provided.
C.B.S. Scientific warrants this product to be free from defects in materials and workmanship under normal
service for one year from date of shipment. If the product proves defective during this period, C.B.S.
Scientific will repair or replace it at our option, free of charge, if returned to us postage prepaid. This
warranty does not cover: damage in transit, damage caused by carelessness, misuse or neglect, normal
wear through frequent use, damage caused by solvent corrosion, damage caused by improper handling
or user alteration, nor unsatisfactory performance as a result of conditions beyond our control. C.B.S.
Scientific shall in no event be liable for incidental nor consequential damages, including without limitation,
lost profits, loss of income, loss of business opportunities, loss of use and other related damages,
however caused, nor any damage arising from the incorrect use of the product.
invalidate any warranty. Always turn off the DC power source prior to disconnecting power cords
from the product. Disconnect power cords from the power source first, and then from the
product. For maximum safety, always operate this system in an isolated, low traffic area, not
accessible to unauthorized personnel. Never operate damaged or leaking equipment.
electrical exposure.
C. B. S. Scientific 3 C.B.S. LITE Slab Gels
Page 4
FRANÇAIS INFORMATION IMPORTANTE À L'USAGE DES UTILISATEURS
Le présent manuel d'utilisation explique la manière de se ser vir efficacement du produit en c onditions
de sécurité. Il est recommandé de soigneusem ent lire la totalité du manuel, avec ses consignes
et ses instructions.
Le triangle avec point d'exclamation est un symbole destiné à avertir l'utilisateur du produit de
l'importance de certaines exigences relatives au fonctionnement, à l'entretien et/ou à la
garantie.
Le triangle avec flèche en zigzag est un symbole destiné à avertir l'utilisateur du produit de la
possibilité d'exposition à des décharges avec danger de secousses électriques.
Tout manquement à l'observation des consignes et des instruc tions peut exposer les personnes
et les biens à des dommages corporels et/ou m atériels et peut annuler toute garantie. Il faut toujours
interrompre l'alimentation de courant continu avant de déconnecter les cordons d'alimentation du
produit. Déconnecter d'abord les cordons d'alimentation branchés sur la source de tension
(alimentation de secteur) puis ceux branchés sur le produit. Pour une sécurité m aximum, il faut toujours
faire fonctionner ce système dans un lieu isolé, peu fréquenté, où le personnel non autorisé n'a pas
accès. Ne jamais faire fonctionner un matériel endommagé ou affecté par des fuites.
GARANTIE ET RESPONSABILITÉ
Le produit a été fabriqué conformément aux norm es applicables les plus exigentes en matière de
matériaux, de main d'oeuvre, de conception et d'ingéniérie. C.B.S. Scientific garantit que le produit a
subi des essais et que ses performances r empliront les conditions des spéc ifications publiées ou leur
seront même supérieures. La présente garantie n'est valide que si le produit a fonctionné et a été
entretenu conformément aux consignes et instructions fournies.
C.B.S. Scientific garantit que le produit sera dépourvu de vices de matériaux et de main d'oeuvre, en
conditions de service normales, pendant un an à com pter de la date d'expédition. Au cas où le pr oduit
s'avérerait défectueux pendant cette période de garantie, C.B.S. Scientific réparera ou r emplacera le
produit, à sa discrétion et gratuitement, si le produit lui est retourné port payé d'avance. La garantie ne
couvre pas les dommages de transport; les dommages causés par l'im prudence, le manque de soins,
l'abus ou la négligence; l'usure normale résultant d'une utilisation fréquente; les domm ages causés par
la corrosion des solvants; et les dommages causés par la manipulation inadéquate ou des
changements apportés par l'utilisateur. La garantie ne couvre pas non plus les perf ormances non
satisfaisantes résultant de conditions hors du contrôle de C.B.S. Scientific. C.B.S. Scientific ne pourr a
en aucun cas être tenue responsable de dommages indirects, y compris, de m anière non limitative, la
perte de bénéfices, le manque à gagner, la perte d'occasions d'aff aires, l'impos sibilité d'usage ou tous
autres dommages associés, quelle qu'en soit la cause, ni de dom mages résultant de l'usage incorrect
du produit.
ESPAÑOL INFORMACIÓN IMPORTANTE PARA EL USUARIO
El presente instructivo explica la manera de usar este producto en f orma segura y efectiva. Sírvase
leerlo en su totalidad y seguir detenidamente las indicaciones que contiene.
El símbolo del triángulo con exclamación llama la atenc ión del usuario a requisitos
importantes para el uso y mantenimiento del producto, así como para la validez de la
garantía.
El símbolo del triángulo con rayo llama la atención del usuario a la posibilidad de
riesgos eléctricos.
El incumplimiento de las instrucciones aquí señaladas podría dar lugar a riesgos a la persona, al
laboratorio o a ambos y podría anular toda garantía. Siempre apague la fuente de corriente continua
antes de desenchufar los cables eléctric os del producto. Primero desc onecte los cables de la fuente
de energía y después del producto. Para mayor seguridad, siempre use este sistem a en un área
aislada, de poco movimiento de personas e inaccesible a personal no autorizado. Jamás use equipo
que presenta algún daño o fuga.
GARANTÍA Y RESPONSABILIDAD
Este producto fue fabricado de acuerdo con las normas más estrictas que sean factibles en cuanto a
materiales, mano de obra y diseño. C.B.S Scientific garantiza que se sometió el producto a pruebas y
que cumplirá o excederá las especif icaciones publicadas. Esta garantía s erá válida únicamente si se
usa y se da servicio de mantenimiento al producto de acuerdo con las instrucciones señaladas.
C.B.S. Scientific garantiza que este producto se encontrará libre de defectos de m ateriales y mano de
obra por un período de servicio normal de un año a partir de la fecha de embar que. Si el producto
resulta defectuoso durante este período, C.B.S. Scientific lo reparará o lo repondrá, a criterio de C.B.S.,
libre de cargos, si se devuelve el producto a C.B.S. porte pagado. Esta garantía no cubre daños
sufridos en tránsito, daños provocados por descuido, mal uso o negligencia, desgaste norm al como
consecuencia del uso excesivo, daños atribuibles a corrosión provocada por solventes, daños
causados por el uso indebido o alteraciones realizadas por el usuario ni rendimiento insatisfac torio
atribuible a circunstancias fuera del control de C.B.S. Scientific. C.B.S. Scientific en ningún caso
asumirá responsabilidad por daños incidentales o subsecuentes, inc luyendo, en forma no limitativa, la
pérdida de utilidades, de ingresos, de oportunidades comerciales o del uso del pr oducto y otros daños
afines, fuere cual fuere su origen, ni por daños derivados del uso incorrecto del producto.
DEUTSCH WICHTIGE INFORMATION FÜR DEN BENUTZER
Diese Bedienungsanleitung beschreibt wie man dieses Produkt sicher und wirksam benutzt. Bitte lesen
und befolgen Sie alle Anweisungen in dieser Anleitung.
Das Dreieck mit Ausrufezeichen weist den Benutzer des Produktes darauf hin, daß
wichtige Bedienungs-, Wartungs- und/oder Garantievorschriften zu beachten sind.
Das Dreieck mit Zickzack blitz warnt den Benutzer des Produktes vor möglichen
Gefahren durch elektrische Spannungen.
Nichtbeachtung dieser Anweisungen kann zu persönlichen und/oder labortechnisc hen Schäden
führen und gleichzeitig alle Garantien als nichtig erklären. Die DC Stromzufuhr muß imm er, vor dem
Entfernen der Stromkabel vom Produkt, abges chaltet werden. Die Stromzufuhrkabel müssen zuerst
von der Steckdose und erst dann vom Produkt entfernt werden. Um höchste Sicherheit zu
gewährleisten sollte dieses System in einem abgesonderten und besonders ruhigen Ber eich eingesetzt
werden und vor Unbefugten sicher sein.
GARANTIE UND HAFTUNG
Dieses Produkt wurde unter Anwendung von Produkten mit höchster Qualität und aus Materialien m it
bester Verarbeitung und modernstem Des ign hergestellt. C.B.S Scientific gar antiert, daß das Produk t
getestet wurde und alle publizierten Spezifikationen übertrifft. Diese Garantie ist jedoch nur gültig,
wenn das Produkt nach der beigefügten Bedienungsanleitung bedient und gewartet wurde.
C.B.S. Scientific garantiert, daß dieses Produkt bei normaler Bedienung aus fehlerfreiem Material
besteht und fehlerfrei in der Ausführung ist. Diese Garantie gilt für ein Jahr ab Lieferdatum. Sollte das
Produkt in diesem Zeiraum fehlerhaft werden, bietet C.B.S. Scientif ic eine kostenlose Reparatur bzw.
kostenlosen Ersatz, einschließlich freiem Rückporto. Diese Garantie schließt folgendes aus:
Transportschaden, Schaden durch Nachlässigkeit, Mißbrauch oder Vernachlässigung, normale
Abnützung durch regelmäßigen Gebrauch, Schaden durch Säureangriff, Schaden durch falsche
Handhabung, Veränderung des Produktes durch den Benutzer, oder unzureichende Leistungen die sich
nicht im Verantwortungsbereich von C.B.S. Scientific befinden. C.B.S. Sc ientific kommt unter keinen
Umständen für folgende Schäden auf: Sachschadensverlust, Einkommensverlust, Verlust von
Geschäftsmöglichkeiten, Verlust der Anwendung und andere damit verbundene Schäden die auf irgend
eine Art und Weise entstanden sind, oder Schäden die aus falscher Anwendung des Produktes
entstanden sind.
ITALIANO INFORMAZIONI IMPORTANTI PER L’UTENTE
Questo manuale spiega come utilizzare questo prodotto in maniera sicura ed eff iciente. Si pregai di
leggere e seguire con cautela le istruzioni di ogni parte di questo manuale.
Il triangolo contenete il simbolo di un punto esclamativo avverte l’utente di impor tanti
requisiti relativi al funzionamento, manutenzione e/o garanzia del prodotto.
Il triangolo contenete il simbolo di un lampo avverte l’utente del prodotto della
laboratorio, oltre a poter annullare la garanzia. Prima di distaccare il cordone d’alimentazione dal
prodotto, spegnere sempre la sorgente di corrente continua. Distacc are i cordoni d’alimentazione
prima dal lato della sorgente di tensione e poi dal lato del prodotto. Per maggior sicurezza, mettere
sempre in funzione il prodotto in un’area isolata con poco traffico che non s ia acces sibile al personale
non autorizzato. Non mettere mai in funzione un’apparecchiatura che sia danneggiata o abbia perdite.
GARANZIA E RESPONSABILITÀ
Questo prodotto è stato fabbricato seguendo gli s tandar d più elevati per i materiali, la manodopera e la
progettazione. La C.B.S. Scientific garantisce il prodotto è stato sottopos to a prova e raggiunge o
supera i valori pubblicati per i dati tecnici. Questa garanzia è valida solo se il prodotto è messo in
esercizio e soggetto a manutenzione secondo le istruzioni fornite.
La C.B.S. Scientific garantisce che questo prodotto è liber o di difetti di materiali e manodopera, in
normali condizioni d’esercizio, per la durata di un anno dalla data di spedizione. Se, in questo periodo, il
prodotto si dimostrerà difettoso, la C.B.S. Scientific, a suo giudizio, lo riparerà o sostituirà. Questa
garanzia non copre danni in transito, danni causati da negligenza, uso improprio, trascuratezza,
normale consumo derivante da uso frequente, o danni causati da s olventi corrosivi, danni causati da
maltrattamento o da modifiche apportate dall’utente e non copre pres tazioni insoddisfacenti c he siano il
risultato di condizioni al di fuori del controllo del fabbricante. La C.B.S. Scientific non sarà in ogni caso
responsabile per danni incidentali o consequenziali, incluso, senza limitazioni, perdite di profitto, perdita
di entrate, perdita di opportunità d’affari e altri danni relativi, comunque causati, e per danni risultati da
uso incorretto del prodotto.
possibilità di pericoli dovuti a corrente elettrica.
La mancata osservanza delle istruzioni può essere causa di pericolo alla propria persona ed al
C. B. S. Scientific 4 C.B.S. LITE Slab Gels
Page 5
SECTION 1
General Information
1.1 Introduction
C.B.S. Scientific offers the Slab Gel Systems designed to meet a wide variety of applications.
These systems are designed for separating and characterizing protein and can be adapted for
short sequencing runs with the appropriate accessories. Gel casting can be accomplished using
our convenient and unique Gel Wrap™, or by using our multiple gel casting chambers (see
Section 6, Slab Gel Kits and Accessories).
Other applications include; mobility shift assays, antibody super shift assays, Rnase protection
assays, DNA footprinting, differential display, DNA paternity testing and forensic analysis.
Power to the Slab Gel units is to be supplied by an external DC voltage power supply that must
be ground isolated so that the DC voltage output floats with respect to ground. For any power
60 watts power
80mA current
55°C ambient temperature
supply used, the maximum specified operating parameters for the units are:
Maximum Limits
50VDC/per cm gel
Current to the unit, provided from the external power supply, must enter the unit through the lid
assembly, providing a safety interlock to the user. Current to the unit is broken when the lid is
removed. Do not attempt to use the unit without the safety lid, and alw ays turn the power
supply off before removing the lid, or when working w ith the unit in any w ay. Follow
safety precautions specified by the power supply manufacturer.
SECTION 2
C. B. S. Scientific 6 C.B.S. LITE Slab Gels
Page 7
Description of Parts
2.1 Unpacking - U.S. version
Please verify that your unit comes complete with the following components:
Slab Gel Kit:
• Slab Gel electrophoresis unit • 1 set glass plates
• Safety cover with attached DC power leads • 1 set spacers
• 2ea GPC-0001, white spring clamps • 1ea Gel Wrap
• 6ea GPC-0002, white spring clamps • Bubble level, attached to base
• 1ea comb • 4ea leveling screws, nylon
2.2 Components/Assembly-U.S. version
SECTION 2---continued
C. B. S. Scientific 7 C.B.S. LITE Slab Gels
Page 8
Description of Parts
2.3 Unpacking --- European version
Please verify that your unit comes complete with the following components:
Slab Gel Kit:
• Slab Gel electrophoresis unit • 1 set glass plates
• Safety cover with attached DC power leads • 1 set spacers
• 2ea GPC-0001, white spring clamps • 1ea Gel Wrap
• 6ea GPC-0002, white spring clamps • Bubble level, attached to base
• 1ea comb • 4ea leveling screws, nylon
2.4 Components/Assembly ---European version
SECTION 3
C. B. S. Scientific 8 C.B.S. LITE Slab Gels
Page 9
Instructions for Use
3.1 Slab Gel Unit Preparation
3.2 Preparation/Cleaning of Glass Plates
3.3 Gel Casting Using Gel Wrap™ Gasket Casting method
1. Place Slab Gel unit on a level work surface in an authorized work area. Level the
unit by adjusting the four white nylon thumbscrew until the bubble is centered in the
level. Be sure that all four of the screws are touching the bench.
2. Verify that the comb, spacer set and Gel Wrap™ gasket are the same thickness, by
assembling without gel casting.
Hand wash both plates with a high quality lab detergent followed by a complete rinsing
with dH
O. Air dry or use a lint-free tissue. Spray/wipe the chosen inner surfaces of the
2
plate set with 95% ethanol and dry with lint-free tissue.
1. Start by holding the rectangular back plate with
the rounded bottom corners and start applying
the gasket around one side of the glass plate.
Note: one side of the “U” shaped gasket is flat,
and the other side has tubing that will act as a
seal around the spacers.
2. When applying the gasket over the rounded
corners of the back glass plate, make sure the
notches on the gasket align with the rounded
corners of the glass plate. Once the gasket is
pushed over the bottom edge and corners, work
it down the remaining side.
3. Place the gasketed plate on the lab bench with
the tubing side up, and extend the bottom of the
plate over the edge of the bench, approximately
¾ of an inch. Place the spacers along side the
inside edges of the gasket. Be sure the rounded
corner end of each spacer is facing the outside
bottom of the plate, following the radius of the
glass.
C. B. S. Scientific 9 C.B.S. LITE Slab Gels
Page 10
4. Place the notched plate on top of the bottom
assembly, starting from the bottom edge and
gently easing the plate down. Verify the gasket
is smooth around the edges and then clamp
along the bottom.
5. Lift the assembly and stand it on the base of the
clamps. For leveling, push glass plate assembly
down until it stops against clamp body. Clamp
the sides of the assembly with additional casting
clamps on either side. As each clamp is
attached, be sure the gasket is aligned between
the plates forming a seal.
6. Apply PAGE or agarose solution to gel plate
sandwich using a syringe or pipette. If using a
stacking gel, pour desired height of running gel,
then overlay a small amount of dH
O or 0.1%
2
SDS solution to top of gel. After
polymerization, rinse with buffer, add stacking
gel solution and insert comb. For regular, unit
percentage gels, add polyacrylamide solution
to correct height, and insert comb. Allow gel to
set, usually 20 minutes. Extra gel solution in
pipette or syringe can be monitored to test
polymerization of gel mix.
7. Disassembly. Hold the clamped plate assembly with one hand. Remove the gasket
by starting at one of the top ends and pulling up and out on the gasket until it
releases from the plate, up to the bottom of each of the white clamps. When each
clamp is reached DO NOT remove it, instead feed the gasket down through the
clamp body and repeat pulling up and out. Continue feeding until the gasket is fully
detached. If gel, is not to be used immediately, wrap entire plate sandwich with
plastic wrap tightly to seal and store at 4°C for up to a month.
C. B. S. Scientific 10 C.B.S. LITE Slab Gels
Page 11
3.4 Running the Gel
1. Attach plate set(s) to unit with
smaller or notched plate facing
towards the upper buffer reservoir. If
using a pre-cast gel stored at 4°C,
allow to warm to room temperature.
Use white clamps with mouth open
(cat. # GPC-0001) to attach plate
set to unit, clamping each side to
the sandwich.
2. Pour freshly prepared buffer in
upper and lower chamber. Using a
pipette or syringe, thoroughly flush
out the wells in the glass plate
sandwich with buffer from the upper
chamber. Boil or heat samples and
then place immediately on ice. Load
samples. If outer lanes do not
contain sample, it is recommended
that you run standards (Rainbow™
standards can be monitored easily
during run) and/or fill outer lanes
with loading buffer to reduce smiling
and wrap-around effects.
3. Attach safety cover.
4. Connect the leads to the power
supply, matching the color-coded red
to red and black to black. See
Section 4.1 for recommended
power conditions.
Begin separation by electrophoresis.
C. B. S. Scientific 11 C.B.S. LITE Slab Gels
Page 12
3.5 Removing the Gel
1. Turn the power supply off and disconnect the leads from the power supply. Remove the
safety cover from the unit, by placing thumbs on white posts next to red & black
connectors, then pushing down while pulling up with fingers under lid. DO NOT pull on power cords.
2. Drain upper reservoir using a pipette or syringe. Detach side clamps and remove gel
sandwich. Carefully separate, making sure that the gel remains attached to one plate.
Stain and fix according to your preferred method. Dispose of buffer according to your
institutions safety requirements.
3.6 Maximum Well/Comb Volumes
NOTE: To calculate sample well volume expressed in millimeters (mm) of height
Precise electrophoresis conditions will vary according to the number and type of gels
used, buffer conditions employed, power input, and the general goal of the experiment.
Refer to the reference section for in depth discussions on practical and theoretical
approaches to protein gel electrophoresis.
Using standard SDS-PAGE buffer systems apply 1-10 VDC/cm of gel. For sequencing
applications use 50 VDC/cm. If running two gels in the Dual Units, keep the volts the
same but double the mA. It is also true that if the thickness of gel increases, increase the
mA proportionally.
At constant voltage, the proteins will migrate at a constant rate during electrophoresis
with adequate heating appropriate for denaturing gels. Increasing the voltage/mA (for a
single gel thickness and percentage) will speed mobility but increase the risk of
overheating.
The sample migration rate can be increased by raising the input power. This can be done
son systems which employ “active” temperature control such as Dual Slab Gel Units and
Dual Mini-Vertical Gel Units. The joule heating generated by the higher input power is
offset by the cooling effect of the water jacket between the gels. Exact conditions should
be determined empirically but could be increased at least in the 20% range.
4.2 Recommended Buffers and Reagents
Pre-mixed acrylamide stock solutions are the method of choice. Use according to
manufacturer’s instructions.
Typical ‘scratch’ recipe for a 4% acrylamide gel:
10mls 40% acrylamide
6.6mls 2% Bis-acrylamide stock
5mls 10X TBE
78.4mls dH
750ul 10% APS
50ul TEMED
1. Make up 0.5X TBE buffer
2. After gel apparatus is set-up and ready for the gel to be poured, add 750ul
fresh 10% APS solution to the acrylamide solution.
3. Add 5 to 10 ul TEMED and using a 10 to 25ml pipette, quickly “pour” the gel.
4. Allow the gel to polymerize at least 60 minutes.
5. Remove the comb after polymerization and wash out wells with 0.5X TBE
(acrylamide will seep into the wells).
6. Fill upper and lower chambers with 0.5X TBE.
7. Pre-electrophorese gel, if needed, 20-30 minutes.
8. Load wells with samples.
9. Monitor migration with dye markers.
0
2
C. B. S. Scientific 13 C.B.S. LITE Slab Gels
Page 14
4.2 Recommended Buffers and Reagents-continued
BUFFERS:Agarose Page/Agarose Slab
TBE (1X solution):
0.089M Tris base 0.04M Tris-acetate
0.89M Boric acid 0.001M EDTA
0.002M EDTA pH 8.0
pH 8.3
Protein Denaturing Denaturing/Non-Denaturing/
TG-SDS(1X):
0.025M Tris base TT (Tris-Taurine( (1X): TT-SDS (1X):
1. Hames, B.D., Rickwoood, D. (ed.) (1990). Gel Electrophoresis of Proteins. A
Practical Approach. 2
2. Sambrook, J., Fritsch, E.F., Maniatis, T. (1989). Molecular Cloning. A Laboratory
Manual. 2nd edn. Cold Spring Harbor Laboratory Press, Cold Spring Harbor, New
York. 18.47-18.61.
nd
edn. IRL Press, Oxford. Ch. 1 & 3.
3. Ausubel, F.M., Brent, R., Kingston, R.E., Moore, D.D., Seidman, J.G., Smith, J.A.,
Struhl, K. (ed) (1993). Current Protocols in Molecular Biology. Vol. 2, Greene
Publishing Associates, Inc. and John Wiley & Sons, Inc., Ch. 10.
C. B. S. Scientific 14 C.B.S. LITE Slab Gels
Page 15
SECTION 5
Maintenance of Equipment
5.1 Care and Handling
Additional precautions:
The plastic components of the Slab Gel units are fabricated from acrylic and
polycarbonate. Electrodes and connectors are made from pure platinum, stainless steel,
and chrome plated brass. As with any laboratory instrument, adequate care ensures
consistent and reliable performance.
After each use, rinse buffer chamber, glass plates, spacers and combs with de-ionized
water. Wipe dry with a soft cloth or paper towel, or allow to air dry. Whenever necessary,
all components may be washed gently with water and a non-abrasive detergent, and
rinsed and dried as above. Never use abrasive cleaners, glass cleaning sprays or
scouring pads to clean the components, as these will damage the unit and components.
• Do not autoclave or dry-heat sterilize the apparatus or components.
• Do not expose the apparatus or components to phenol, acetone, benzene,
halogenated hydrocarbon solvents or alcohols.
• Avoid prolonged exposure of the apparatus or components to UV light.
• Do NOT treat with diethylpyrocarbonate (DEPC)-treated water for extended
periods at 37°C. A brief rinse with DEPC-water is sufficient after a thorough
wash.
5.2 Maintenance
The following inspection and maintenance procedures will help maintain the safety and
reliable performance of the Slab Gel systems. Replacement parts can be ordered by
calling 1-858-755-4959 or by contacting your local distributor.
• Banana plugs and power cords should be inspected regularly. If the banana
plugs become loose or do not feel friction tight replace the plugs or power cords.
• Should power cord assemblies (connectors, wire or shrouds) show any signs of
wear or damage (e.g. cracks, nicks, abrasions, or melted insulation), replace
them immediately.
• The platinum wire is secured to the banana jack by compression between a
stainless washer and the jack nut. The nut/washer interface should be tight and
free of corrosion.
C. B. S. Scientific 15 C.B.S. LITE Slab Gels
Page 16
SECTION 6
Slab Gel Kits and Accessories
Cat. # Item
SG-125-02 Slab Gel Kit, 145mm height. Includes 1 comb, 1 Gel Wrap spacer set, 1 Gel Wrap™ Glass plate
set, 1 each Gel Wrap™ and 6 clamps.
SG-160-02 Slab Gel Kit, 160mm height. Includes 1 comb, 1 Gel Wrap spacer set, 1 Gel Wrap™ Glass plate
set, 1 each Gel Wrap™ and 6 clamps.
SG-170-02 Slab Gel Kit, 170mm height. Includes 1 comb, 1 Gel Wrap spacer set, 1 Gel Wrap™ Glass plate
set, 1 each Gel Wrap™ and 6 clamps.
SG-190-02 Slab Gel Kit, 190mm height. Includes 1 comb, 1 Gel Wrap spacer set, 1 Gel Wrap™ Glass plate
set, 1 each Gel Wrap™ and 6 clamps.
SG-200-02 Slab Gel Kit, 220mm height. Includes 1 comb, 1 Gel Wrap spacer set, 1 Gel Wrap™ Glass plate
set, 1 each Gel Wrap™ and 6 clamps.
SG-250-02 Slab Gel Kit, 280mm height. Includes 1 comb, 1 Gel Wrap spacer set, 1 Gel Wrap™ Glass plate
set, 1 each Gel Wrap™ and 6 clamps.
Accessories:
COMBS (Teflon) COMBS (Teflon)
Cat. # Comb dimensions Cat. # Comb dimensions
VGC-7501 0.75mm x 1 well VGC-2001 2.0mm x 1 well
VGC-7502 0.75mm x 2 well VGC-2002 2.0mm x 2 well
VGC-7503E 0.75mm x 3 equal wells VGC-2003E 2.0mm x 3 equal wells
VGC-7503M 0.75mm x 2 well + middle marker VGC-2003M 2.0mm x 2 well + middle marker
VGC-7504 0.75mm x 4 well VGC-2004 2.0mm x 4 well
VGC-7505 0.75mm x 5 well VGC-2005 2.0mm x 5 well
VGC-7506 0.75mm x 6 well VGC-2006 2.0mm x 6 well
VGC-7508 0.75mm x 8 well VGC-2008 2.0mm x 8 well
VGC-7510 0.75mm x 10 well VGC-2010 2.0mm x 10 well
VGC-7512 0.75mm x 12 well VGC-2012 2.0mm x 12 well
VGC-7514* 0.75mm x 14 well VGC-2014* 2.0mm x 14 well
VGC-7515 0.75mm x 15 well VGC-2015 2.0mm x 15 well
VGC-7516 0.75mm x 16 well VGC-2016 2.0mm x 16 well
VGC-7520 0.75mm x 20 well VGC-2020 2.0mm x 20 well
VGC-7529* 0.75mm x 29 well VGC-2029* 2.0mm x 29 well
VGC-7530 0.75mm x 30 well VGC-2030 2.0mm x 30 well
VGC-1001 1.0mm x 1 well VGC-3001 3.0mm x 1 well
VGC-1002 1.0mm x 2 well VGC-3002 3.0mm x 2 well
VGC-1003E 1.0mm x 3 equal wells VGC-3003E 3.0mm x 3 equal wells
VGC-1003M 1.0mm x 2 well + middle marker VGC-3003M 3.0mm x 2 well + middle marker
VGC-1004 1.0mm x 4 well VGC-3004 3.0mm x 4 well
VGC-1005 1.0mm x 5 well VGC-3005 3.0mm x 5 well
VGC-1006 1.0mm x 6 well VGC-3006 3.0mm x 6 well
VGC-1008 1.0mm x 8 well VGC-3008 3.0mm x 8 well
VGC-1010 1.0mm x 10 well VGC-3010 3.0mm x 10 well
VGC-1012 1.0mm x 12 well VGC-3012 3.0mm x 12 well
VGC-1014* 1.0mm x 14 well VGC-3014* 3.0mm x 14 well
VGC-1015 1.0mm x 15 well VGC-3015 3.0mm x 15 well
VGC-1016 1.0mm x 16 well VGC-3016 3.0mm x 16 well
VGC-1020 1.0mm x 20 well VGC-3020 3.0mm x 20 well
VGC-1029* 1.0mm x 29 well VGC-3029* 3.0mm x 29 well
VGC-1030 1.0mm x 30 well VGC-3030 3.0mm x 30 well
VGC-1501 1.5mm x 1 well
VGC-1502 1.5mm x 2 well
VGC-1503E 1.5mm x 3 equal wells
VGC-1503M 1.5mm x 2 well + middle marker
VGC-1504 1.5mm x 4 well
VGC-1505 1.5mm x 5 well
VGC-1506 1.5mm x 6 well
VGC-1508 1.5mm x 8 well
VGC-1510 1.5mm x 10 well
VGC-1512 1.5mm x 12 well
VGC-1514* 1.5mm x 14 well
VGC-1515 1.5mm x 15 well
VGC-1516 1.5mm x 16 well
VGC-1520 1.5mm x 20 well
VGC-1529* 1.5mm x 29 well
VGC-1530 1.5mm x 30 well
* Designed for multi-channel pipettor loading.
Gel Wrap Gasket Accessories:
C. B. S. Scientific 16 C.B.S. LITE Slab Gels
Page 17
SPACERS for GEL WRAP (PVC) GEL WRAP™ GASKET
Cat. # Spacer dimensions Cat. # Gasket thickness x plate size
VGS-7512R 0.75mm x 14.5cm VGE-7512 0.75mm thick x 14.5cm
VGS-7516R 0.75mm x 16cm VGE-7516 0.75mm thick x 16cm
VGS-7517R 0.75mm x 17cm VGE-7517 0.75mm thick x 17cm
VGS-7519R 0.75mm x 19cm VGE-7519 0.75mm thick x 19cm
VGS-7520R 0.75mm x 22cm VGE-7520 0.75mm thick x 22cm
VGS-7525R 0.75mm x 28cm VGE-7525 0.75mm thick x 28cm
VGS-7540R 0.75mm x 40cm VGE-7540 0.75mm thick x 40cm
VGS-1012R 1.0mm x 14.5cm VGE-1012 1.0mm thick x 14.5cm
VGS-1016R 1.0mm x 16cm VGE-1016 1.0mm thick x 16cm
VGS-1017R 1.0mm x 17cm VGE-1017 1.0mm thick x 17cm
VGS-1019R 1.0mm x 19cm VGE-1019 1.0mm thick x 19cm
VGS-1020R 1.0mm x 22cm VGE-1020 1.0mm thick x 22cm
VGS-1040R 1.0mm x 40cm VGE-1040 1.0mm thick x 40cm
VGS-1512R 1.5mm x 14.5cm VGE-1512 1.5mm thick x 14.5cm
VGS-1516R 1.5mm x 16cm VGE-1516 1.5mm thick x 16cm
VGS-1517R 1.5mm x 17cm VGE-1517 1.5mm thick x 17cm
VGS-1519R 1.5mm x 19cm VGE-1519 1.5mm thick x 19cm
VGS-1520R 1.5mm x 22cm VGE-1520 1.5mm thick x 22cm
VGS-1525R 1.5mm x 28cm VGE-1525 1.5mm thick x 28cm
VGS-1540R 1.5mm x 40cm VGE-1540 1.5mm thick x 40cm
VGS-2012R 2.0mm x 14.5cm VGE-2012 2.0mm thick x 14.5cm
VGS-2016R 2.0mm x 16cm VGE-2016 2.0mm thick x 16cm
VGS-2017R 2.0mm x 17cm VGE-2017 2.0mm thick x 17cm
VGS-2019R 2.0mm x 19cm VGE-2019 2.0mm thick x 19cm
VGS-2020R 2.0mm x 22cm VGE-2020 2.0mm thick x 22cm
VGS-2025R 2.0mm x 28cm VGE-2025 2.0mm thick x 28cm
VGS-2040R 2.0mm x 40cm VGE-2040 2.0mm thick x 40cm
VGS-3012R 3.0mm x 14.5cm VGE-3012 3.0mm thick x 14.5cm
VGS-3016R 3.0mm x 16cm VGE-3016 3.0mm thick x 16cm
VGS-3017R 3.0mm x 17cm VGE-3017 3.0mm thick x 17cm
VGS-3019R 3.0mm x 19cm VGE-3019 3.0mm thick x 19cm
VGS-3020R 3.0mm x 22cm VGE-3020 3.0mm thick x 22cm
VGS-3025R 3.0mm x 28cm VGE-3025 3.0mm thick x 28cm
VGS-3040R 3.0mm x 40cm VGE-3040 3.0mm thick x 40cm
VGE-75XX 0.75mm x foot
VGE-10XX 1.0mm x foot
VGE-15XX 1.5mm x foot
VGE-20XX 2.0mm x foot
VGE-30XX 3.0mm x foot
(1/8” back plate with rounded corners) (1/8” back plate with rounded corners)
Cat. # Glass Plate dimensions (w x h)
NGP-125NR 16.5cm x 14.5cm NGP-125BR 16.5cm x 14.5cm
NGP-160NR 16.5cm x 16cm NGP-160BR 16.5cm x 16cm
NGP-170NR 16.5cm x 17cm NGP-170BR 16.5cm x 17cm
NGP-190NR 16.5cm x 19cm NGP-190BR 16.5cm x 19cm
NGP-200NR 16.5cm x 22cm NGP-200BR 16.5cm x 22cm
NGP-250NR 16.5cm x 28cm NGP-250BR 16.5cm x 28cm
NGP-400NR 16.5cm x 40cm NGP-400BR 16.5cm x 40cm
* Glass plate ordering options
Note: Although glass plates are listed as sets, individual pieces can be ordered. To order a notched only plate, add a “-NO” to the end of
the cat. #. To order a back only (rectangular) plate, add a “-BO” to the end of the cat. #.
: To order an unnotched plate set and add a “UN” to the end of the catalog # (rubber ears included).
Cat. # Glass Plate dimensions (w x h)
Standard Accessories for Multiple Vertical Gel Casters:
Cat. # Glass Plate dimensions (w x h) Cat. # Glass Plate dimensions (w x h)
NGP-125N 16.5cm x 14.5cm NGP-125B 16.5cm x 14.5cm
NGP-160N 16.5cm x 16cm NGP-160B 16.5cm x 16cm
NGP-170N 16.5cm x 17cm NGP-170B 16.5cm x 17cm
NGP-190N 16.5cm x 19cm NGP-190B 16.5cm x 19cm
NGP-200N 16.5cm x 22cm NGP-200B 16.5cm x 22cm
NGP-250N 16.5cm x 28cm NGP-250B 16.5cm x 28cm
NGP-400N 16.5cm x 40cm NGP-400B 16.5cm x 40cm
* Glass plate ordering options: To order an unnotched plate set and add a “UN” to the end of the catalog # (rubber ears included).
Note: Although glass plates are listed as sets, individual pieces can be ordered. To order a notched only plate, add a “-NO” to the end of
the cat. #. To order a back only (rectangular) plate, add a “-BO” to the end of the cat. #.
STANDARD SPACERS (PVC)
(2 side spacers, 1 bottom, + silicone tubing)
C. B. S. Scientific 17 C.B.S. LITE Slab Gels
Page 18
Cat. # Spacer dimensions Cat. # Spacer dimensions
VGS-7512 0.75mm x 14.5cm VGS-1512 1.5mm x 14.5cm
VGS-7516 0.75mm x 16cm VGS-1516 1.5mm x 16cm
VGS-7517 0.75mm x 17cm VGS-1517 1.5mm x 17cm
VGS-7519 0.75mm x 19cm VGS-1519 1.5mm x 19cm
VGS-7520 0.75mm x 22cm VGS-1520 1.5mm x 22cm
VGS-7525 0.75mm x 28cm VGS-1525 1.5mm x 28cm
VGS-7540 0.75mm x 40cm VGS-1540 1.5mm x 40cm
VGS-1012 1.0mm x 14.5cm VGS-2012 2.0mm x 14.5cm
VGS-1016 1.0mm x 16cm VGS-2016 2.0mm x 16cm
VGS-1017 1.0mm x 17cm VGS-2017 2.0mm x 17cm
VGS-1019 1.0mm x 19cm VGS-2019 2.0mm x 19cm
VGS-1020 1.0mm x 22cm VGS-2020 2.0mm x 22cm
VGS-1025 1.0mm x 28cm VGS-2025 2.0mm x 28cm
VGS-1040 1.0mm x 40cm VGS-2040 2.0mm x 40cm
VGS-3012 3.0mm x 14.5cm VGS-6012 6.0mm x 14.5cm
VGS-3016 3.0mm x 16cm VGS-6016 6.0mm x 16cm
VGS-3017 3.0mm x 17cm VGS-6017 6.0mm x 17cm
VGS-3019 3.0mm x 19cm VGS-6019 6.0mm x 19cm
VGS-3020 3.0mm x 22cm VGS-6020 6.0mm x 22cm
VGS-3025 3.0mm x 28cm VGS-6025 6.0mm x 28cm
VGS-3040 3.0mm x 40cm VGS-6040 6.0mm x 40cm
VGC-0201 0.2mm x 1 well VGC-0401 0.4mm x 1 well
VGC-0202 0.2mm x 2 well VGC-0402 0.4mm x 2 well
VGC-0203E 0.2mm x 3 equal wells VGC-0403E 0.4mm x 3 equal wells
VGC-0203M 0.2mm x 2 well + middle marker VGC-0403M 0.4mm x 2 well + middle marker
VGC-0204 0.2mm x 4 well VGC-0404 0.4mm x 4 well
VGC-0205 0.2m x 5 well VGC-0405 0.4mm x 5 well
VGC-0206 0.2mm x 6 well VGC-0406 0.4mm x 6 well
VGC-0208 0.2mm x 8 well VGC-0408 0.4mm x 8 well
VGC-0210 0.2mm x 10 well VGC-0410 0.4mm x 10 well
VGC-0212 0.2mm x 12 well VGC-0412 0.4mm x 12 well
VGC-0214* 0.2mm x 14 well VGC-0414* 0.4mm x 14 well
VGC-0215 0.2mm x 15 well VGC-0415 0.4mm x 15 well
VGC-0216 0.2mm x 16 well VGC-0416 0.4mm x 16 well
VGC-0220 0.2mm x 20 well VGC-0420 0.4mm x 20 well
VGC-0229* 0.2mm x 29 well VGC-0429* 0.4mm x 29 well
VGC-0230 0.2mm x 30 well VGC-0430 0.4mm x 30 well
VGC-0501 0.5mm x 1 well
VGC-0502 0.5mm x 2 well
VGC-0503E 0.5mm x 3 equal wells
VGC-0503M 0.5mm x 2 well + middle marker
VGC-0504 0.5mm x 4 well
VGC-0505 0.5mm x 5 well
VGC-0506 0.5mm x 6 well
VGC-0508 0.5mm x 8 well
VGC-0510 0.5mm x 10 well
VGC-0512 0.5mm x 12 well
VGC-0514* 0.5mm x 14 well
VGC-0515 0.5mm x 15 well
VGC-0516 0.5mm x 16 well
VGC-0520 0.5mm x 20 well
VGC-0529* 0.5mm x 29 well
VGC-0530 0.5mm x 30 well
SK16.5-0201 0.2mm x 1 well VGS-0216 0.2mm x 16cm
SK16.5-0220 0.2mm x 20 well VGS-0217 0.2mm x 17cm
Cat. # Comb dimensions
C. B. S. Scientific 18 C.B.S. LITE Slab Gels
Page 19
SK16.5-0226* 0.2mm x 26 well VGS-0219 0.2mm x 19cm
SK16.5-0230 0.2mm x 30 well VGS-0220 0.2mm x 22cm
SK16.5-0242* 0.2mm x 42 well VGS-0225 0.2mm x 28cm
SK16.5-0258* 0.2mm x 58 well VGS-0240 0.2mm x 40cm
SK16.5-0401 0.4mm x 1 well VGS-0416 0.4mm x 16cm
SK16.5-0420 0.4mm x 20 well VGS-0417 0.4mm x 17cm
SK16.5-0426* 0.4mm x 26 well VGS-0419 0.4mm x 19cm
SK16.5-0430 0.4mm x 30 well VGS-0420 0.4mm x 22cm
SK16.5-0442* 0.4mm x 42 well VGS-0425 0.4mm x 28cm
SK16.5-0458* 0.4mm x 58 well VGS-0440 0.4mm x 40cm
SK16.5-0501 0.5mm x 1 well VGS-0516 0.5mm x 16cm
SK16.5-0520 0.5mm x 20 well VGS-0517 0.5mm x 17cm
SK16.5-0526* 0.5mm x 26 well VGS-0519 0.5mm x 19cm
SK16.5-0530 0.5mm x 30 well VGS-0520 0.5mm x 22cm
SK16.5-0542* 0.5mm x 42 well VGS-0525 0.5mm x 28cm
SK16.5-0558* 0.5mm x 58 well VGS-0540 0.5mm x 40cm
ADDITIONAL VERTICAL GEL ACCESSORIES:
BAR CLAMP
Cat. # Item
GPC-1650 Bar clamp for 16.5cm wide units
MULTIPLE GEL CASTING CHAMBERS
Cat. # Item
GCC-404 Multiple Vertical Gel Casting chamber, casts up to 4 gels simultaneously
GCC-410 Multiple Vertical Gel Casting chamber, casts up to 10 gels simultaneously
GEL CASTING STAND
Cat. # Item
GCS-250 Gel Casting Stand, includes stand with leveling base and 4 clamps
Cat. # Item
GPC-0002 Casting clamp, jaws closed
GPC-0001 Larger white clamp, jaws slightly open
POWER SUPPLY
Cat. # Item
EPS-300-II Mini Power Supply, with timer, CV or CC, 10-300V, 110V/60Hz, current range:4-500mA, 90 watts
EPS-300-IIV Mini Power Supply, with timer, CV or CC, 10-300V, 220V/50Hz, current range:4-500mA, 90 watts
EPS-3000-III Constant Power Supply, 0-3000 Volt, 110V/60Hz, Series II
EPS-3000-IIV Constant Power Supply, 0-3000 Volt, 220V/50Hz, Series II
C. B. S. Scientific 19 C.B.S. LITE Slab Gels
Page 20
CONTACT US
Telephone:
Local or International
858-755-4959
Toll Free: 858-755-0733
Fax: 858-755-0733
Check out our new and
improved online ordering!
www.cbsscientific.com
E-mail address:
sales@cbssci.com
Mailing address:
C.B.S. Scientific Company
P.O. Box 856
Del Mar, CA 92014
Shipping address:
C.B.S. Scientific Company
10805 Vista Sorrento Pkwy
Suite 100
Solana Beach, CA 92121