This Instruction Manual will explain how to use this product safely and effectively.
Please read and carefully follow the instruction manual in its entirety.
The triangle/exclamation mark symbol alerts the user of the product to important operational, maintenance, and/or warranty requirements.
The triangle/lighting bolt symbol alerts the user of the product to potentially hazardous
electrical exposure.
Failure to adhere to the instructions could result in personal and/or laboratory hazards,
as well as invalidate any warranty. Always turn off the DC power source prior to disconnecting power cords from the product. Disconnect power cords from the power source
rst and then from the product. For maximum safety, always operate this system in an
isolated, low trafc area, not accessible to unauthorized personnel. Never operate dam-
aged or leaking equipment.
WARRANTY AND LIABILITY
This product was produced utilizing the highest practical standards of materials, work-
manship, and design. C.B.S. Scientic warrants that the product has been tested and
will meet or exceed published specications. This warranty is valid only if the product
has been operated and maintained according to the instructions provided.
C.B.S. Scientic warrants this product to be free from defects in materials and workmanship under normal service for one year from date of shipment. If the product
proves defective during this period, C.B.S. Scientic will repair or replace it at our
option, free of charge, if returned to us postage prepaid. This warranty does not
cover: damage in transit, damage caused by carelessness, misuse or neglect, normal
wear through frequent use, damage caused by solvent corrosion, damage caused by
improper handling or user alteration, nor unsatisfactory performance as a result of con-
ditions beyond our control. C.B.S. Scientic shall in no event be liable for incidental nor
consequential damages, including without limitation, lost prots, loss of income, loss
of business opportunities, loss of use and other related damages, however caused, nor
any damage arising from the incorrect use of the product.
3
www.cbsscientic.com
FRANÇAIS INFORMATION IMPORTANTE À L’USAGE DES UTILISATEURS
Le présent manuel d’utilisation explique la manière de se servir efcacement du produit en conditions
de sécurité. Il est recommandé de soigneusement lire la totalité du manuel, avec ses consignes
et ses instructions.
Le triangle avec point d’exclamation est un symbole destiné à avertir l’utilisateur du produit
de l’importance de certaines exigences relatives au fonctionnement, à l’entretien et/ou à
la garantie.
Le triangle avec èche en zigzag est un symbole destiné à avertir l’utilisateur du produit de la
possibilité d’exposition à des décharges avec danger de secousses électriques.
Tout manquement à l’observation des consignes et des instructions peut exposer les personnes
et les biens à des dommages corporels et/ou matériels et peut annuler toute garantie. Il faut toujours
interrompre l’alimentation de courant continu avant de déconnecter les cordons d’alimentation du
produit. Déconnecter d’abord les cordons d’alimentation branchés sur la source de tension (alimentation
de secteur) puis ceux branchés sur le produit. Pour une sécurité maximum, il faut toujours faire
fonctionner ce système dans un lieu isolé, peu fréquenté, où le personnel non autorisé n’a pas accès. Ne
jamais faire fonctionner un matériel endommagé ou affecté par des fuites.
GARANTIE ET RESPONSABILITÉ
Le produit a été fabriqué conformément aux normes applicables les plus exigentes en matière de
matériaux, de main d’oeuvre, de conception et d’ingéniérie. C.B.S. Scientic garantit que le produit a subi
des essais et que ses performances rempliront les conditions des spécications publiées ou leur seront
même supérieures. La présente garantie n’est valide que si le produit a fonctionné et a été entretenu
conformément aux consignes et instructions fournies.
C.B.S. Scientic garantit que le produit sera dépourvu de vices de matériaux et de main d’oeuvre, en
conditions de service normales, pendant un an à compter de la date d’expédition. Au cas où le produit
s’avérerait défectueux pendant cette période de garantie, C.B.S. Scientic réparera ou remplacera le
produit, à sa discrétion et gratuitement, si le produit lui est retourné port payé d’avance. La garantie
ne couvre pas les dommages de transport; les dommages causés par l’imprudence, le manque de
soins, l’abus ou la négligence; l’usure normale résultant d’une utilisation fréquente; les dommages
causés par la corrosion des solvants; et les dommages causés par la manipulation inadéquate ou
des changements apportés par l’utilisateur. La garantie ne couvre pas non plus les performances non
satisfaisantes résultant de conditions hors du contrôle de C.B.S. Scientic. C.B.S. Scientic ne pourra
en aucun cas être tenue responsable de dommages indirects, y compris, de manière non limitative, la
perte de bénéces, le manque à gagner, la perte d’occasions d’affaires, l’impossibilité d’usage ou tous
autres dommages associés, quelle qu’en soit la cause, ni de dommages résultant de l’usage incorrect
du produit.
ESPAÑOL INFORMACIÓN IMPORTANTE PARA EL USUARIO
El presente instructivo explica la manera de usar este producto en forma segura y efectiva. Sírvase
leerlo en su totalidad y seguir detenidamente las indicaciones que contiene.
El símbolo del triángulo con exclamación llama la atención del usuario a requisitos
importantes para el uso y mantenimiento del producto, así como para la validez de
la garantía.
El símbolo del triángulo con rayo llama la atención del usuario a la posibilidad de riesgos
eléctricos.
El incumplimiento de las instrucciones aquí señaladas podría dar lugar a riesgos a la persona, al
laboratorio o a ambos y podría anular toda garantía. Siempre apague la fuente de corriente continua
antes de desenchufar los cables eléctricos del producto. Primero desconecte los cables de la fuente de
energía y después del producto. Para mayor seguridad, siempre use este sistema en un área aislada,
de poco movimiento de personas e inaccesible a personal no autorizado. Jamás use equipo que
presenta algún daño o fuga.
GARANTÍA Y RESPONSABILIDAD
Este producto fue fabricado de acuerdo con las normas más estrictas que sean factibles en cuanto a
materiales, mano de obra y diseño. C.B.S Scientic garantiza que se sometió el producto a pruebas
y que cumplirá o excederá las especicaciones publicadas. Esta garantía será válida únicamente si
se usa y se da servicio de mantenimiento al producto de acuerdo con las instrucciones señaladas.
C.B.S. Scientic garantiza que este producto se encontrará libre de defectos de materiales y mano de
obra por un período de servicio normal de un año a partir de la fecha de embarque. Si el producto resulta
defectuoso durante este período, C.B.S. Scientic lo reparará o lo repondrá, a criterio de C.B.S., libre
de cargos, si se devuelve el producto a C.B.S. porte pagado. Esta garantía no cubre daños sufridos en
tránsito, daños provocados por descuido, mal uso o negligencia, desgaste normal como consecuencia
del uso excesivo, daños atribuibles a corrosión provocada por solventes, daños causados por el uso
indebido o alteraciones realizadas por el usuario ni rendimiento insatisfactorio atribuible a circunstancias
fuera del control de C.B.S. Scientic. C.B.S. Scientic en ningún caso asumirá responsabilidad por
daños incidentales o subsecuentes, incluyendo, en forma no limitativa, la pérdida de utilidades, de
ingresos, de oportunidades comerciales o del uso del producto y otros daños anes, fuere cual fuere su
origen, ni por daños derivados del uso incorrecto del producto.
DEUTSCH WICHTIGE INFORMATION FÜR DEN BENUTZER
Diese Bedienungsanleitung beschreibt wie man dieses Produkt sicher und wirksam benutzt. Bitte lesen
und befolgen Sie alle Anweisungen in dieser Anleitung.
Das Dreieck mit Ausrufezeichen weist den Benutzer des Produktes darauf hin, daß
wichtige Bedienungs-, Wartungs- und/oder Garantievorschriften zu beachten sind.
Das Dreieck mit Zickzackblitz warnt den Benutzer des Produktes vor möglichen
Gefahren durch elektrische Spannungen.
Nichtbeachtung dieser Anweisungen kann zu persönlichen und/oder labortechnischen Schäden
führen und gleichzeitig alle Garantien als nichtig erklären. Die DC Stromzufuhr muß immer, vor dem
Entfernen der Stromkabel vom Produkt, abgeschaltet werden. Die Stromzufuhrkabel müssen zuerst von
der Steckdose und erst dann vom Produkt entfernt werden. Um höchste Sicherheit zu gewährleisten
sollte dieses System in einem abgesonderten und besonders ruhigen Bereich eingesetzt werden und
vor Unbefugten sicher sein.
GARANTIE UND HAFTUNG
Dieses Produkt wurde unter Anwendung von Produkten mit höchster Qualität und aus Materialien mit
bester Verarbeitung und modernstem Design hergestellt. C.B.S Scientic garantiert, daß das Produkt
getestet wurde und alle publizierten Spezikationen übertrifft. Diese Garantie ist jedoch nur gültig, wenn
das Produkt nach der beigefügten Bedienungsanleitung bedient und gewartet wurde.
C.B.S. Scientic garantiert, daß dieses Produkt bei normaler Bedienung aus fehlerfreiem Material
besteht und fehlerfrei in der Ausführung ist. Diese Garantie gilt für ein Jahr ab Lieferdatum. Sollte
das Produkt in diesem Zeiraum fehlerhaft werden, bietet C.B.S. Scientic eine kostenlose Reparatur
bzw. kostenlosen Ersatz, einschließlich freiem Rückporto. Diese Garantie schließt folgendes aus:
Transportschaden, Schaden durch Nachlässigkeit, Mißbrauch oder Vernachlässigung, normale
Abnützung durch regelmäßigen Gebrauch, Schaden durch Säureangriff, Schaden durch falsche
Handhabung, Veränderung des Produktes durch den Benutzer, oder unzureichende Leistungen die
sich nicht im Verantwortungsbereich von C.B.S. Scientic benden. C.B.S. Scientic kommt unter
keinen Umständen für folgende Schäden auf: Sachschadensverlust, Einkommensverlust, Verlust
von Geschäftsmöglichkeiten, Verlust der Anwendung und andere damit verbundene Schäden die auf
irgend eine Art und Weise entstanden sind, oder Schäden die aus falscher Anwendung des Produktes
entstanden sind.
ITALIANO INFORMAZIONI IMPORTANTI PER L’UTENTE
Questo manuale spiega come utilizzare questo prodotto in maniera sicura ed efciente. Si pregai di
leggere e seguire con cautela le istruzioni di ogni parte di questo manuale.
Il triangolo contenete il simbolo di un punto esclamativo avverte l’utente di importanti
requisiti relativi al funzionamento, manutenzione e/o garanzia del prodotto.
Il triangolo contenete il simbolo di un lampo avverte l’utente del prodotto della possibilità
di pericoli dovuti a corrente elettrica.
La mancata osservanza delle istruzioni può essere causa di pericolo alla propria persona ed
al laboratorio, oltre a poter annullare la garanzia. Prima di distaccare il cordone d’alimentazione dal
prodotto, spegnere sempre la sorgente di corrente continua. Distaccare i cordoni d’alimentazione prima
dal lato della sorgente di tensione e poi dal lato del prodotto. Per maggior sicurezza, mettere sempre
in funzione il prodotto in un’area isolata con poco trafco che non sia accessibile al personale non
autorizzato. Non mettere mai in funzione un’apparecchiatura che sia danneggiata o abbia perdite.
GARANZIA E RESPONSABILITÀ
Questo prodotto è stato fabbricato seguendo gli standard più elevati per i materiali, la manodopera e la
progettazione. La C.B.S. Scientic garantisce il prodotto è stato sottoposto a prova e raggiunge o supera
i valori pubblicati per i dati tecnici. Questa garanzia è valida solo se il prodotto è messo in esercizio e
soggetto a manutenzione secondo le istruzioni fornite.
La C.B.S. Scientic garantisce che questo prodotto è libero di difetti di materiali e manodopera, in normali
condizioni d’esercizio, per la durata di un anno dalla data di spedizione. Se, in questo periodo, il prodotto
si dimostrerà difettoso, la C.B.S. Scientic, a suo giudizio, lo riparerà o sostituirà. Questa garanzia non
copre danni in transito, danni causati da negligenza, uso improprio, trascuratezza, normale consumo
derivante da uso frequente, o danni causati da solventi corrosivi, danni causati da maltrattamento o da
modiche apportate dall’utente e non copre prestazioni insoddisfacenti che siano il risultato di condizioni
al di fuori del controllo del fabbricante. La C.B.S. Scientic non sarà in ogni caso responsabile per danni
incidentali o consequenziali, incluso, senza limitazioni, perdite di protto, perdita di entrate, perdita di
opportunità d’affari e altri danni relativi, comunque causati, e per danni risultati da uso incorretto del
prodotto.
4
Cipher DGGE Instructions 2/28/14
SECTION 1
General Information
1.1 Introduction
Denaturing Gradient Gel Electrophoresis (DGGE) is a powerful genetic analysis technique
that can be used for detecting single base changes and polymorphisms in genomic (1, 2),
cloned, and PCR amplied DNA (3,4). Two of the most valuable uses for DGGE in human
genetics are in directly detecting single base changes that cause disease and in detecting
polymorphisms with DNA probes for genetic-linkage analysis. Also, DNA fragment melting
points can be determined using Perpendicular DGGE (1).
C.B.S. Scientic has designed four different Cipher DGGEK Systems, which are reliable and
easy-to-use. The DGGEK-1001 is a 2 gel system and features one dual gel cassette. The
DGGEK-2001 features two single gel cassettes. The DGGE-2401 has the same buffer tank
as the DGGE-1001 and DGGE-2001 but features two dual gel cassettes, for a capacity of 4
gels per run. Finally there is the DGGE-4801, which features the larger buffer tank and four
dual cassettes enabling the researcher to run up to 8 gels simultaneously. Improvements to
the technique include: a simplied method for casting perpendicular and vertical gels using
Gel Wrap™. Single or dual gel cassettes have been redesigned eliminating the use of agarose
plugs, the need for an external peristaltic pump for buffer cycling has been eliminated, bulky
screw clamps have been replaced with polypropylene spring clamps, buffer tank dimensions
have been changed to use bench space more efciently, and a safety cover with an electrical
interlock has been added for protection to help maintain temperature and reduce evaporation.
1.2 Specications
Constructions
DGGE Tank Acrylic
DGGE Lid Sodalime glass, acrylic
Power to the DGGE systems is to be supplied by an external DC voltage power supply
that must be ground isolated so that the DC voltage output oats with respect to
ground. For any power supply used, the maximum specied operating parameters for
the units are: Maximum Limits
250 VDC voltage
30 watts power
80 mA current
70°C ambient temperature
Current to the unit, provided from the external power supply, must enter the unit
through the lid assembly, providing a safety interlock to the user. DC current to the unit
is broken when the lid is opened. Do not attempt to use the unit without the safety
lid. Always turn the power supply off before removing the lid, or when working
with the unit in any way. Follow safety precautions specied by the power supply
manufacturer.
Input Power: Mains to safety interlock
Depending on country of destination, input voltage from mains electrical supply are as
follows: 110-120VAC @ 50/60Hz/5 Amps or 200-240VAC @ 50/60Hz/5 Amps. Country
specic power cords or CE approved adapter kits are supplied with each system.
Warning: Do NOT turn on Heater/Stirrer until tank has been lled with buffer!
SECTION 2
DGGEK Unit Set-Up
2.1 Unpacking Instructions
1) Unpack the DGGE Tank Assembly and place on level surface in an approved location.
Using gure 2-1 verify that your DGGEK-1001 or DGGEK-2001 System comes
complete with the following components:
• EPS-300 X Power Supply (not pictured)
• Peristaltic Mini-Pump for gel casting
• Lower Reservoir/Acrylic tank/safety interlock/ with 2 black leads inside lid
• Heater/Stirrer/By-pass pump (Controller) - factory-mounted onto DGGE tank
• Gel cassette(s), combs, spacer sets, Gel Wrap™ gaskets, and glass plates sets as
specied in chart below:
Type of Gel
Cassette(s)
DGGEK-10011 - dual222242
DGGEK-20012- single222242
DGGEK-24012 - dual444484
• 1 Gradient maker, 20mls per side, Cat. # GM-40
• Buffer displacement tank (DGGEK-1001 only, not pictured)
• White spring clamps
received with system will vary depending on format of DGGE)
• Buffer siphon pump with tubing, Cat# BSP-1000
• 1 Conversion Plate (not shown) for running 1 gel at a time on dual cassette (for
DGGEK-1001, DGGEK-2401 only)
# of
Sample
Well
Combs
# of
1-well
Combs
6
# of Gel
Wrap ™
Spacer
Sets
# of Perpendicular
Spacer
Sets
Cipher DGGE Instructions 2/28/14
# Gel
Wrap™
Gaskets
# glass
Plate Sets
Figure 2-1
Heater
Stirrer
Buffer Siphon
Pump Assembly
VaporShield
Thermometer
DGGE
Tank
Single
Cassetes
Comb (2)
Spacers
(2 pairs)
Glass Plate
Assembly(2)
Clamps
Mini-Pump
(Optional)
GM-40
Gelwrap
®
(2)
Buffer recycling
manifold
2. Open the lid and conrm the following:
a) Two lengths of small bore silicone tubing for buffer cycling are already attached
to the buffer recycling manifold outlets
b) A single length of large bore tubing from the By-Pass Pump is connected to the
2.2 Filling DGGE tank with Buffer
barb-tting on the plastic manifold.
DGGEK-2001
1) Fill the tank with running buffer (see Section 6 for buffer formulations) to optimal
level for number and type of cassettes used. Refer to chart below for exact buffer
volumes required: This requires 20 to 21.2 liters of buffer. See chart below:
1) Attach power cord to heater/stirrer and plug into wall outlet.
(comes factory-mounted onto DGGE tank)
2.3.1 Front and Rear Panels
1
2
3
4
10
11
9
8
7
12
6
5
13
Front View Rear View
1. Display 11. Pump Speed Switch
2. Preset Temperature Buttons 12. Identification Label
3. Safety Set Reset Button 13. Pump Inlet
4. Safety Set Indicator Knob 14. Reservoir Purge
5. Cooling Light
1
15. Pump Ou tlet 2
6. Power ON / OFF Button 16. AC Input
7. Heating Light 17. Circuit Breaker / AC Power Switch
8. °F Light
9. Select/Set Knob (Push to set, turn to adjust)
10. °C Light
1
Present on R efrigera ting/Hea ting mode ls only
2
Built in on reservoir models only
2
2
8
Cipher DGGE Instructions 2/28/14
1.Temperature Sensor
2.Pump Shaftand Impeller
3.PumpOutletNozzle
1
4.Over-Temperature Sensor
5.HeaterCoils
1
Rotateson Immersion Circulators
5
3
1
2
4
2.3.2 Controller Assembly
1. Temperature Sensor
2. Pump Shaft and Impeller
3. Buffer By-Pass Tubing Adapter
4. Over-Temperature Sensor
5. Heater Coils
2.3.3 Controller Specications
Temperature Stabil ity ±0.05°C
Readout Accuracy ±0.5°C
Over-Tem peratur e Protection Yes, user- adjustabl e
Low-Liqu id Protection Yes
Heater 1100W – 115V,1600 W – 240V
Pump Speeds 2-speed, 9 liters per minute or 15 liters p er minut e
Environmental Conditions:
Indoor Use Only Over Voltage: Category ll
Maximum Altitude: 2000 meters Operating Ambient: 5° to 30°C
Relative Humidity: 80% for temperatures to 30°C Pollution Degree: 2
Class 1: Residential, Commercial , Light Industrial Class 2: Heavy Industrial
3.
1.
4.
2.
5.
9
www.cbsscientic.com
2.3.4 Controller Operation
2.3.4.a Reservoir Liquid Level
Fill the reservoir with the appropriate buffer. The DGGEK Systems are designed to be used
with reagent buffers such as 1xTAE or 1xTBE. On Circulating Baths, the liquid level should
be sufcient to cover the heating coils, pump, over-temperature sensor, and at least one inch
(25mm) of the temperature sensor. On Immersion Circulators, the liquid level should be suf-
cient to fully immerse the heater coils, over-temperature sensor, and pump outlet nozzle.
2.3.4.b Power
An IEC power cord is provided with the Circulator. This power cord should be plugged into
the IEC receptacle on the rear of the Controller and then plugged into a properly grounded
outlet. Make sure that the power outlet is the same voltage and frequency indicated on the
identication label on the back of the Controller. The use of an extension cord is not recommended. However, if one is necessary, it must be properly grounded and capable of handling
the total wattage of the unit. The extension cord must not cause more than a 10% drop
in voltage to the Circulator. Once the unit has been connected to an appropriate electrical
outlet, place the Circuit Breaker/Power Switch on the rear of the Controller in the ON position.
Four decimal points (….) will appear on the digital display. DO NOT place the Power Switch
on the front of the Controller ON until the Safety Set has been adjusted to the desired tem-
perature (see 3.6 below).
Indicated Voltage:
Volts/Phase/FrequencyOperational Voltage RangePhaseFrequency
100/ 1 / 6090 to 110 Voltssingle60 Hz
100/ 1 / 5090 to 110 Voltssingle50 Hz
120/ 1 / 60110 to 130 Voltssingle60 Hz
230/ 1 / 60208 to 230 Voltssingle60 Hz
240/ 1 / 50220 to 240 Voltssingle50 Hz
2.3.4.c Setting the Safety Set Set Point - NOTE: DGGEK is factory set to 80ºC
The Safety Set feature automatically disconnects Controller power to the heater and pump in
the event that the reservoir liquid level drops too low or the Controller fails. The Safety Set is
user-adjustable between approximately 40° and 210°C. It should be set at least 5°C higher
than the Software High Limit temperature. Use a at blade screwdriver to rotate the Safety
Set Indicator Knob to the desired temperature. Do not force the knob beyond the stops at either end of the temperature value range. Once the Safety Set temperature has been set, turn
power to the Controller ON by pressing the Power Switch on the front of the Controller. The
pump will begin operating, the display will ash the current temperature set point (tx.xx), the
°C LED will light, and the current bath temperature will appear on this display. Pump speed
selection is made using the Pump Speed Selection Switch on the rear of the Controller (see
Section 2.3 or 3.3). If power is disrupted because the Safety Set temperature was exceeded,
place the Circuit Breaker/Power Switch in the OFF position, press the Safety Set Reset Button, correct the problem (low liquid level, incorrect Safety Set temperature, etc.), and then
restore power. Activation of the Safety Set during normal operation will display a fault (FLt 3)
on the readout.
2.3.4.d Selecting Temperature Units
The control set point and actual bath temperatures may be displayed in either °C or °F. The
factory-default is °C. To change from °C to °F, place the Circuit Breaker/Power Switch on the
rear of the Controller in the OFF position and then press and hold the P2 Button while turning
the power back ON.
10
Cipher DGGE Instructions 2/28/14
To change from °F to °C, place the Circuit Breaker/Power Switch in the OFF position and then
press and hold the P3 Button while turning the power back ON.
NOTE: When the temperature display units are changed, the Software High Limit value and all
temperature presets revert to the factory-default values. If a calibration value has been entered,
the value will be retained.
2.3.4.e Setting the Software High Limit - NOTE: DGGEK is factory set to 75ºC
This feature provides additional safety and protection by allowing a selectable upper temperature limit set point. To avoid an unwanted shutdown during regular operation, the high
limit value should be set at least 5°C higher than the selected control temperature To set the
Software High Limit temperature set point, press the P2 and P3 keys simultaneously and repeat
until (Hxxx) appears on the display. This is the current Software High Limit value. It is factory set
at 75°C. To change the displayed value, press and turn the Select/Set Knob until the desired
Software High Limit set point value is displayed. A clockwise rotation increases the value; a
counterclockwise rotation decreases the value. Press the Select/Set Knob a second time to
accept the new value and return to normal operation. If the Software High Limit value meets or
exceeds the control temperature set point, (E-H1) will ash on the display. If this occurs, enter
a higher value for the Software High Limit or reduce the control temperature set point. If the
actual bath temperature reaches the Software High Limit setpoint, (FLt1) will ash on the display. Should this occur, the Controller will automatically remove power from the heater and, in
Refrigerating/Heating units, the compressor as well. The pump will continue to operate. Once
the problem is corrected (bath temperature reduced or Software High Limit value increased),
press the Power button to clear the message.
2.3.4.f Setting the Set Point Temperature
Press and release the Select/Set Knob. The decimal point ashes to indicate that the set point
temperature can be changed. Turn the Select/Set Knob until the desired temperature set point
is displayed. A clockwise rotation increases the setting; a counterclockwise rotation decreases
the setting. Press the Select/Set Knob a second time to accept the displayed value. The deci-
mal point stops ashing and the display will indicate the actual bath temperature. Allow sufcient time for the bath to stabilize at the desired temperature before making any adjustments
to set point temperature. NOTE: The unit will automatically accept the displayed set point after
approximately 10 seconds of inactivity, even if the Select/Set Knob was not pressed. The set
point temperature may be checked at any time by pressing the Select/Set Knob. If the set
point temperature cannot be raised, it is possible that the Software High Limit value is set lower
than the desired control temperature set point. Reset the Software High Limit value to 5°C or
more above the desired set point temperature.
2.3.4.g User-Denied Preset Temperatures
With the unit on, press the desired Preset Button —
P1, P2, or P3. The LED associated with the selected
Preset Button will begin to ash. Rotate the Select/
Set Knob to the desired temperature set point. Press
the selected Preset Button a second time to enter the
new set point. The new set point temperature will not
be saved unless the Preset Button is pressed. The
LED associated with Preset Button lights continuously whenever that preset value is controlling bath
temperature. If more than one Preset Button is set
at a given temperature set point, the LED associated
with all Preset Buttons with that set point will light.
11
www.cbsscientic.com
2.3.4.h Local Lockout Feature
This feature enables the user to lock all controls on the controller. While the feature is activated, the unit will remain running at the current settings. To activate the local lockout feature,
press and hold the Select/Set Knob for 10 seconds. Once locked, the screen will read LLo.
When locked, the setpoint decimal point will not ash as usual. Press and hold the Select/Set
Knob again for 10 seconds to unlock the controls. Once unlocked, the screen will read CAn.
2.3.4.i Stick-on Strips for Preset Buttons
The Standard Controller is supplied with two
stick-on strips that can be applied beneath the
Preset Buttons. These strips are removable and
can be reapplied. The user can write the set
point temperature associated with each Preset
Button on this strip. The use of an erasable
medium, such as dry erase marker or air tip
pen is recommended. The former can be wiped
off with a dry cloth or tissue; the later can usually be removed with simple glass cleaner and a
cloth or tissue. Do not use a permanent marker
or ballpoint pen. Always do an ink test before
writing on these strips.
The stick-on strip ts beneath the controller’s
Preset Buttons. It is removable and reapplicable.
2.3.5 Controller Display Messages
Display Description Action Required
….
tx.xx Power up s elf-test Normal — Appears momentarily at startup
oCx.x Calibration offset v alue Normal — C urrent c alibration offset value; refer to section 6.1
Hxxx Software High Limit value Normal — C urrent S oftware High Limit value; refer to section 5.6
Axxx Auto-Refrigeration set point value
E-H1 Softwar e High Limit set point too low
FLt 1 Software High Limit valu e exceeded
FLt 2 EEPROM reset
FLt 3 Safety Set temperatur e exceeded
FLt 4 H eating Triac f ailure Error — Servic e required
FLt 5 Pr obe failure Error — Service required
FLt 6
Standby mode
i2c error — Commun ication f ailure to
modulati on board
Normal — Indic ates that the Circuit Breaker/P ower Switc h is ON
and the Controller Power Switch is OFF
Normal — Curr ent Auto-Refrigeration set point; refer to section
5.10 (Appears only on Refrigerating/H eating Circulators)
Error — The value entered is below the c ontrol temperature set
point. Refer to Setting Software High Limit, s ection 5.6
Error — Set a Software High Limit value higher, then turn main
power to the unit OFF and back ON; refer to section 5.6
Error — Turn Circuit Breaker/Power Switch O FF, hold P3 B utton,
and then turn th e Circuit Break er/Power Switch back ON
Error — Check fluid level
Check flui d temper ature and s et point
Ensure th at OTP s et point is higher than fluid set point or
increase to maximu m
Turn Circuit Breaker/Power Switch OFF, press S afety Set R eset
Button, and then turn the Circuit Breaker/P ower Switch back ON
Error — Service required (App ears onl y on Refrig erating/H eating
Circulators)
12
Cipher DGGE Instructions 2/28/14
2.3.6 Controller Calibrations and Maintenance
2.3.6.a Calibration
Calibration allows the user to match the Controller’s bath temperature display to an external
reference thermometer. Calibration is performed as follows: Set the desired operating uid temperature set point and allow temperature to stabilize. Press the P2 and P3 simultaneously and
release and repeat until the display reads (oCx.x). Press P1 and hold until (Cal) is displayed. This
will take about 2 seconds. At one second intervals, the displayed value will alternate between
the actual bath uid temperature and the current offset value, which is the difference between
the factory calibration setting and the user’s reference temperature sensor. The maximum offset
is ±0.9°C from factory calibration. To change the calibration offset value, rotate the Select/Set
Knob until the display matches the reading on the reference temperature sensor. The display
will continue to alternate between the offset value and the calibrated display temperature. Press
the Select/Set Knob or the P1 Button to accept the entered value. When the new calibration is
stored and the mode is exited, (dONE) will appear on the display. NOTE: The displayed offset
value will also be accepted if there no activity for 20 seconds. 6.2 Heater The heater should be
kept clean. If deposits build up on the heater, they may be removed by scrubbing with a nonmetallic (plastic) abrasive pad. Do not use steel wool.
2.3.6.b Pump Motor
The pump bearings are permanently lubricated with high-temperature silicone grease and do not
require additional lubrication. Should the bearings become noisy, replacement of the entire pump
motor is recommended. This will reduce repair labor costs and retain uid pumping reliability. A
replacement pump and motor mounting kit is available (see Section 10 - Replacement Parts). 6.4
Cleaning Only mild detergents and water or an approved cleaner should be used on the painted
and stainless steel surfaces of the Circulator. Do not allow cleaning liquids or sprays to enter the
Controller vents. A concentrated bath cleaner is available that can be used to remove mineral
deposits from the reservoir. See Section 10 - Replacement Parts.
2.3.7 Controller Reservoir Fluids
For your buffers use distilled or RO water only!
DO NOT USE the following uids:
1. Automotive antifreeze with additives**
2. Hard tap water**
3. Deionized water with a specic resistance > 1 meg ohm
4. Any ammable uids
5. Concentrations of acids or bases
6. Solutions with halides: chlorides, uorides, bromides, iodides or sulfur
7. Bleach (Sodium Hypochlorite)
8. Solutions with chromates or chromium salts
9. Glycerine
10. Syltherm uids
** At temperatures above 40°C, additives or mineral deposits can adhere to the heater. If
deposits are allowed to build up, the heater may overheat and fail. Higher temperatures
and higher concentrations of additives will hasten deposit build up
Warning: Do not use a ammable liquid as a bath medium as a re hazard
may result.
13
www.cbsscientic.com
DGGEK Unit Set-Up, con’t.
2.4 Setting Buffer Temperature
Warning: Do NOT turn on Heater/Stirrer until tank has been lled with buffer!
1. Turn Heater/Stirrer “On” and set temperature to 60°C according to
Section 2.3.4.f Setting the Set Point Temperature (page 11).
2.5 Connecting the Power Supply to Safety Interlock
2. Position EPS-300 X Power Supply on a shelf above the tank. Refer
to Figure 2-2 to complete the steps below:
a. Attach the power supply to the safety interlock.
b. Attach the safety interlock to a wall (mains) outlet.
c. Do NOT attach power leads (from power supply to safety
interlock) until gels are loaded and installed in the tank.
14
Cipher DGGE Instructions 2/28/14
Power Supply Connection
To mains
(wall outlet)
c
To mains
(wall outlet)
d
Figure 2-2
15
b
a
www.cbsscientic.com
SECTION 3
Gel Casting Options
3.1 Preparation/Cleaning of Glass Plate for Gel Casting
Hand wash both plates with a high quality lab detergent followed by a complete
rinsing with dH2O. Air dry or use a lint-free tissue. Spray/wipe the chosen
inner surfaces of the plate set with 95% ethanol and dry with lint-free tissue.
3.2 Gel Casting Techniques.
A. Gel Wrap™ Gasket Casting Method.
B. Vertical gradient gel casting using GM-40 gradient maker and gravity ow.
C. Vertical gradient gel casting using GM-40 gradient maker and a Mini-pump.
D. Vertical gradient gel casting using GM-40 gradient maker, Mini-pump and
Multi-gel Caster.
NOTE: Prior to casting, mark the notched glass plate to designate which side is the
‘inside’ and ‘outside’. The same side of the notched plate should always sit against the
reservoir gasket.
A. Gel Casting using Gel Wrap™ Gasket Casting method
For Vertical DGGE, use the set of spacers which do not have the small hole or channel
milled into the lower end, these are for casting
perpendicular gels.
1. Start by holding the rectangular back
plate with the rounded bottom corners
and start applying the gasket around one
side of the glass plate. Note: one side
of the “U” shaped gasket is at, and the
other side has tubing that will act as a
seal around the spacers.
2. When applying the gasket over the
rounded corners of the back glass plate,
make sure the notches on the gasket
align with the rounded corners of the
glass plate. Once the gasket is pushed
over the bottom edge and corners, work
it down the remaining side.
16
Cipher DGGE Instructions 2/28/14
3. Place the gasketed plate on the lab
bench with the tubing side up, and
extend the bottom of the plate over the
edge of the bench, approximately ¾ of
an inch. Place the spacers along side
the inside edges of the gasket. Be sure
the rounded corner end of each spacer
is facing the outside bottom of the
plate, following the radius of the glass.
4. Place the notched plate on top of the
bottom assembly, starting from the
bottom edge and gently easing the
plate down. Verify the gasket is smooth
around theedges and then clamp along
the bottom.
5. Lift the assembly and stand it on the
base of the clamps. For leveling, push
glass plate assembly down until it
stops against clamp body. Clamp the
sides of the assembly with additional
casting clamps on either side. As each
clamp is attached, be sure the gasket
is aligned between the plates forming a
seal.
17
www.cbsscientic.com
3.2 Gel Casting Techniques- continued
C-1
C-2
V-1
V-2
B. Vertical gradient gel casting using GM-40 gradient maker and gravity ow.
1. Place the GM-40, gradient maker, on an elevated magnetic stirrer with a small
“ea” stir bar in the cylinder (C-2) closest to the outlet. The gradient maker
should be tted with a leur valve, (V-2) and a 20ga needle with attached tubing
to deliver acrylamide to the gel plate sandwich as shown in gure (3.5). With
the height differential between the leur valve (V-2) and the top of the gel plate
sandwich between 2-4 inches, it takes approximately 5-9 minutes to cast the
gel. Raising the height of the gradient maker can increase ow rate. Flow can
be regulated by turning leur valve (V-2). Flow rate appropriate for each gradient
should be determined empirically.
Fig. 3-5
18
Cipher DGGE Instructions 2/28/14
3.2 Gel Casting Techniques- continued
C. Vertical gradient gel casting using gradient maker (GM-40) and a mini-
pump.
1. Alternatively, you may choose to use a “mini-pump” or other peristaltic
pump to cast gels as shown in g (3.6). If so, secure the gradient maker to
a ring stand and connect the outlet tubing to the mini-pump tubing adapter.
Connect tubing from mini-pump to a 20ga. needle for afxing between glass
plates.
Fig. 3-6
19
www.cbsscientic.com
3.2 Gel Casting Techniques- continued
D. Vertical gradient gel casting using GM-40 gradient maker, Mini-Pump and
Multi-Gel Caster.
1. Clean the gel caster with soap and water and the glass plates with alcohol.
2. Place a separating sheet in the gel caster.
3. Place the back plate (non-eared) on the sheet.
4. Place the spacers on each side of the back plate.
5. Place the top plate (eared) on the spacers to form 1st sandwich. Repeat as
required. Continue to Section 3.3 Vertical Gradient Formation.
Fig. 3-7
20
Cipher DGGE Instructions 2/28/14
3.3 Vertical Gradient Formation
To determine the range of gradient appropriate for your fragment analysis,
please read the enclosed paper by Myers, Shefeld and Cox, especially
section 6.1.3 through 6.2.1 (pages 124 to 126). This gives you an excellent
overview of the determination of melting behavior of your fragments.
1. The following is a typical protocol for casting a 40%-60% gradient gel. Refer to
Section 3.2 Vertical Gradient Gel Casting, gure 3.5, for apparatus assembly.
In an ice bucket, place two 50ml conical tubes labeled “A” and “B”. Add to tube
“A”
• 11.5ml of 40%/7.5%
• 80 µI (10%) APS
• 5 µl TEMED
2. Add to tube “B”:
• 11.5ml of 60%/7.5%
• 80 µI (10%) APS
• 5 µI TEMED
3. Pour solution "B" into right side of gradient maker (C-2),(GM-40) as
shown in g. 3.5, and open interior valve to allow air bubble to escape. Let
as much as 1 ml "B" solution BACKFLOW into left side of gradient maker.
Decant the 1 ml back into right side with pasteur pipette. Remove any
residual solution from (C-1) with absorbent paper.
4. Add solution "A" to left side of gradient maker.
5. Turn on magnetic stirrer.
6. Exit tube with needle should be secured to gel plate notch with tape.
7. Open inside valve rst (V-1), then outside (V-2) valve to start gravity
ow. Optionally, Turn on mini-pump and pump acrylamide at medium speed.
Determine optimal speed empirically.
8. Gel volume is 23ml, using the 0.75mm spacers. If gel volume is not
enough to ll sandwich, use 0% to "top-off". If using water saturated butanol
for overlay, leave 0.5cm void to create at interface.
9. Insert the sample well comb. Allow the gel to polymerize for 20-30
minutes.
Go to Section 4 for Gel Running Preparation.
21
www.cbsscientic.com
FIGURE 1
FIGURE 2
1.0mm AND THICKER
.75mm AND THINNER
PERPENDICULAR GEL CASTING
TO GRADIENT MAKER
20 GAUGE NEEDLE
SEAL WITH AGAROSE
HOLE THROUGH SPACER
GEL PLATE
CLAMPS
TO GRADIENT MAKER
20 GAUGE NEEDLE
SEAL WITH AGAROSE
GEL PLATE
CLAMPS
3.4 Perpendicular Gel Casting
NOTE: Prior to casting, mark the notched glass plate to designate which side is the
‘inside’ and ‘outside’. The same side of the notched plate should always sit against the
reservoir gasket.
1. For perpendicular gel casting, locate the spacer which has a channel machined
into it on one end which will be referred to as the “channel spacer”. The gel
sandwich will be cast on its side, with the channel spacer on top and the channel
further away from the single well comb (see gure 1 for casting gels 1.0mm and
thicker, see gure 2 for casting gels .75mm or thinner).
2. Follow the Gel Wrap instructions 1-5 (pages 16-17), making sure to use the
channel spacer.
3. Insert single well comb into glass notch. A small amount of agarose is needed in
two places once the gel sandwich is placed on its side, both places are located
on the lower corner where the single well comb interfaces the spacer, and on the
other side of the spacer where it interfaces the gel wrap.
4. Once the agarose has
set, insert the gradient
maker outow tube tipped
with a 20ga. through the
Gel Wrap and into the
“channel” (gure 1), or
space created between
the horizontal spacer and
the comb (gure 2). You
may use the same piece of
Gel Wrap many times (20-
30) for the perpendicular
gels: do not use a piece
of Gel Wrap Gasket for
Vertical casting after you
have poked a hole in it for
perpendicular casting—it
will leak.
22
Cipher DGGE Instructions 2/28/14
3.5 Perpendicular Gradient Formation
To determine the range of perpendicular gradient appropriate for your fragment
analysis, please read the enclosed paper by Myers, Shefeld and Cox, as well
as the Methods and Enzymology V. 212 paper by Abrams and Stanton. This
gives you an excellent overview of the determination of melting behavior of your
fragments.
2. The following is a typical protocol for casting a 40%-60% gradient gel. Refer to
Section 3.2 Vertical Gradient Gel Casting, gure 3.5 for apparatus assembly. In
an ice bucket, place two 50ml conical tubes labeled “A” and “B”. Add to tube “A”
• 11.5ml of 40%/7.5%
• 80 µI (10%) APS
• 5 µl TEMED
2. Add to tube “B”:
• 11.5ml of 60%/7.5%
• 80 µI (10%) APS
• 5 µI TEMED
3. Pour solution "B" into right side of gradient maker,(GM-40), and open interior
valve to allow air bubble to escape. Let as much as 1 ml "B" solution BACKFLOW into left side of gradient maker. Decant the 1ml back into right side with
pasteur pipette. Remove any residual solution with absorbent paper.
4. Add solution “A” to left side of gradient maker.
5. Turn on magnetic stirrer.
6. Exit tube should be attached to near side of gel plate with tape.
7 Open inside (V-1) valve rst, then outside (V-2) valve to start ow.
8. Gel volume is 23ml, using the 0.75mm spacers. Allow 20-30 minutes for gel
polymeraztion. If gel volume is not enough to ll gel sandwich, use 0% to “topoff”.
Go to Section 4 for Gel Running Preparation.
23
www.cbsscientic.com
SECTION 4
Gel Running Preparations
4.1 Preparation of the Cassettes
1. After polymerization rinse gel plate assembly with D.I. water to remove
excess acrylamide or denaturants from plate exterior.
2. Remove comb and quickly transfer gel
sandwich to cassette by removing all #2
clamps. LEAVE GEL WRAP IN PLACE.
The Gel Wrap acts as a barrier and
prevents perpendicular electrical elds
from interfering with outside lanes. Reclamp to cassette using #1 clamps/4
per side. Dislodge bottom portion of Gel Wrap from the plates to allow buffer
contact with the gel. See gure 4-1a.
NOTE: If running only 1 gel on a dual
cassette, attach conversion plate with
clamps to second side of cassette.
Fig. 4-1a
3. Immerse single (g. 4-1b) or dual cassette(s) (g. 4-1c) into pre-heated buffer
tank.
Fig. 4-1b
Fig. 4-1c
24
Cipher DGGE Instructions 2/28/14
4.2 Buffer Cycling Connections
1. Fill upper reservoir(s) by attaching recirculating tubing to barb tting on each
upper reservoir (g. 4-2c,d,e). Rinse loading interface of gel with tank buffer
to remove non-polymerized acrylamide or excess urea. Adjust nal level of
lower reservoir buffer by lling the tank until the level of buffer reaches the
underside of the upper reservoir.
Buffer Cycling Connections – Two Single Cassettes
Fig. 4-2c
25
Fig. 4-2d
www.cbsscientic.com
4.2 Buffer Cycling Connections-continued.
Buffer Cycling Connections – Dual Cassettes
Fig. 4-2e
SECTION 5
Running Conditions
5.1 Running the Gels
1. Load samples at 1:1 with neutral dye. Load 5-10 ug Genomic DNA/well or 1-2 ug
cloned (B-globin)/well. Determine concentration by O.D. 260.
2. Attach black power leads to cassettes as shown in Fig. 5-1.
3. Close lid to engage safety interlock. Turn on power supply to 150V (40mA)
constant Volts for 5-7 hours.
Fig. 5-1
26
Cipher DGGE Instructions 2/28/14
5.2 Removing the Gels
3. After the run is completed, turn off power supply, disconnect recirculating tubes
and power leads, remove cassettes, remove glass sandwiches from cassettes.
4. Using a wedge plate separator,cat. # WPS-100, pry plates apart and immerse gel/
plate in buffer tray. Stain with Safe Stain or EtBr (.5ug/ml) for 5-10 minutes. Lift oating gel on plate out of tray and rinse with ddH20. Flip over onto saran wrap and peal
off glass plate. View bands on transilluminator. Electro-Blot or photograph.
5.3 Recommended Power
1. 150V (40mA) CV for 5-7 hours
5.4 Disposal of Buffer Using Buffer Siphon Pump
NOTE: Buffer can be used several times before discarding. Determine empirically.
Dispose of solutions in accordance with the safety regulations of your institution.
1. Attach pump assembly to corner of
tank as shown in gure 5.4a.
2. Place the discharge tubing into a
waste carboy located LOWER than
the pump inlet.
Fig. 5-4a
Fig. 5-4b
3. Begin siphon with half strokes until
liquid starts to ow. Leave shaft fully
extended as shown in gure 5.4b to
allow siphon to operate.
4. To stop siphon, depress handle and
remove pump from tank.
57.1ml Glacial acetic acid
100ml .5 M EDTA, pH 8.0
Add dH2O (or RO-H2O) to 1 liter
Denaturant Stock Solution (7.5% gel)
For 100ml use:
Percentage102030405060708090100
40% Acrylamide/
Bis (ml)
50% TAE Buffer
(ml)
Formamide (ml)481216202428323640
Urea (g)
18.818.818.818.818.818.818.818.818.818.8
2222222222
4.28.412.616.82125.229.433.637.842
Add dH20 (or RO-H2O) to 100ml
Store denaturant stock solutions at 4º C in amber bottles
Polymerization Catalysts:
1. TEMED (10ul/ml gel solution)
2. APS Stock Solution (.6ul/ml gel solution)
For 50 ml of APS stock (10%) use 5 g of ammonium persulphate and water to
50ml. Store at 4ºC.
Neutral Loading Solution
For preparation of 10ml:
20% Ficoll or sucrose, 2 g of Ficoll 400 or sucrose
10mM Tris-HCl, pH 7.8, 100ul of 1.0 M Tris-HCl, pH 7.8
1mM EDTA, 20 ul of 0.5 M EDTA, pH 8.0
0.1% dye (BPB), 10 mg of dye
Water to 10ml
28
Cipher DGGE Instructions 2/28/14
SECTION 7
Maximum Well/Comb Volumes
Vertical Comb Options, short wells, overall length = 13.1cm
5. Guldberg, P., Henriksen, K.F., and Guttler, F. (1993). “Molecular analysis of phenylke-“Molecular analysis of phenylke-
tonia in Denmark: 99% of the mutations detected by denaturing gradient gel electrophoresis.” Genomics 17:141-146
6. Moyret, C., Thellet, C., Puig, P.L., Moles, J-P., Thomas, G. and Hamelin, R. (1994).
“Relative efciency of denaturing gradient gel electrophoresis and single-strand
conformation polymorphisms in the detection of mutations in exons 5 to 8 of the p53
gene.” Oncogene 9:1739-1743
7. Shefeld, V.C., Beck, J.S., Kwitek, A.E., Sandstrom, D.W., and Stone, E.M. (1993).
“The sensitivity of single-strand conformation polymorphism analysis for the detection
of single bas substitutions.” Genomics 16: 325-332.
8. Abrams, E.S. Stanton, V.P. Jr., (1992) “Use of Denaturing Gradient Gel Electrophoresis to Study Conformational Transitions in Nucleic Acids” Methods in Enzymology
9.1 Unit Will Not Operate (no heat, cooling, or pumping)
• Check that the power cord is plugged in to an operating electrical outlet.
• Check that the Circuit Breaker/Power Switch is ON.
• Check that the Controller Power Switch is ON.
• Check that the Safety Set temperature is higher than the control temperature set
point.
• Turn power off, press the Safety Set Reset button, and restore power.
9.2 No Pumping
• Check the uid level of the bath to be sure the pump head is covered with uid.
• If the pump motor does not spin, verify that the Pump Speed Switch is either in the
HIGH or LOW position. If it is in the middle, the pump will not receive power.
• Check the pump impeller for obstructions (it should turn freely).
9.3 Slow or Insufcient Pumping
• Check for low line voltage, particularly when the heater is on.
9.4 No Heating
• Verify that the unit is pumping properly.
• If the heat light is not lit, check the control temperature set point and bath temperature to verify that heating is required.
• Verify that the liquid in the bath covers the heater coils.
• Check that the Software High Limit value has not been exceeded (FLt 1 appears
on the display).
9.5 Insufcient Heating
• Verify that the unit is pumping properly.
• Check for proper line voltage.
• Check for excessive external cooling load on unit.
• At higher temperatures, insufcient heating could be due to excessive heat loss
from tanks, hoses, or vapor from the tank.
• Check that the heat demand from connected devices or experiments does not
exceed the heating capabilities of the unit.
• Allow sufcient time for bath temperature to stabilize when changes in heat load or
set point are made.
9.6 Triac Failure
• (FLt4) appears on the display, indicating that the heater triac has failed or line supply voltage has a source of extreme interference from other equipment. Plug the
unit into another power source. If it still displays triac failure, a triac or triac driver
needs replacement.
9.7 Samples Not Entering Gel After Loading
• When running one gel on a dual cassette, the conversion plate must be clamped
to the unused side to complete the cathode buffer chamber.
31
www.cbsscientic.com
SECTION 10
Maintenance of Equipment
10.1 Care and Handling
The plastic components of the DGGE Systems are fabricated from acrylic
and polycarbonate. Electrodes and connectors are made from pure platinum,
stainless steel, and chrome plated brass. As with any laboratory instrument,
adequate care ensures consistent and reliable performance.
After each use, rinse gel cassettes, spacers and combs with de-ionized
water. Wipe dry with a soft cloth or paper towel, or allow to air dry. Whenever
necessary, all components may be washed gently with water and a nonabrasive detergent, and rinsed and dried as above. Never use abrasive
cleaners, glass cleaning sprays or scouring pads to clean the components, as
these will damage the unit and components.
Additional precautions:
• Do not autoclave or dry-heat sterilize the apparatus or
components.
• Do not expose the apparatus or components to phenol,
acetone, benzene, halogenated hydrocarbon solvents or
undiluted alcohol's.
• Avoid prolonged exposure of the apparatus or components to
UV light.
• Do NOT treat with diethylpyrocarbonate (DEPC)-treated water
for extended periods at 37°C. A brief rinse with DEPC-water
is sufcient after a thorough wash.
10.2 Maintenance
The following inspection and maintenance procedures will help maintain the
safety and reliable performance of the Cipher DGGEK systems. Replacement
parts can be ordered by calling 858-755-4959 or by contacting your local
distributor.
• Banana plugs and power cords should be inspected regularly.
If the banana plugs become loose or do not feel friction tight
replace the plugs or power cords.
• Should power cord assemblies (connectors, wire or shrouds)
show any signs of wear or damage (e.g. cracks, nicks,
abrasions, or melted insulation), replace them immediately.
• The platinum wire is secured to the banana jack by compression
between a stainless washer and the jack nut. The nut/washer
interface should be tight and free of corrosion.
32
Cipher DGGE Instructions 2/28/14
SECTION 11
Ordering Information
CAT. # DESCRIPTION
CIPHER DGGE ELECTROPHORESIS SYSTEMS
DGGEK-1001
DGGEK-2001
DGGEK-2401
DGGEK-4801
Related Products
VGS0720RC-177Spacer Set for Perpendicular DGGE, standard spacer + injection port spacer, 0.75mm thickness x 22cm(l)
VGS1020RC-177Spacer Set for Perpendicular DGGE, standard spacer + injection port spacer, 1.0mm thickness x 22cm(l)
VGS1520RC-177Spacer Set for Perpendicular DGGE, standard spacer + injection port spacer, 1.5mm thickness x 22cm(l)
VGC-0701-177Comb for Perpendicular DGGE, 0.75mm thick x 1 well
VGC-1001-177Comb for Perpendicular DGGE, 1.0mm thick x 1 well
VGC-1501-177Comb for Perpendicular DGGE, 1.5mm thick x 1 well
VGS-0720R-177Gel Wrap® Spacer Set, 0.75mm thick x 22cm with rounded bottom corners
VGS-1020R-177Gel Wrap® Spacer Set, 1.0mm thick x 22cm with rounded bottom corners
VGS-1520R-177Gel Wrap® Spacer Set, 1.5mm thick x 22cm with rounded bottom corners
VGE-0720-177Gel Wrap® Gasket to cast one DGGE gel, 0.75mm thick
VGE-1020-177Gel Wrap® Gasket to cast one DGGE gel, 1.0mm thick
VGE-1520-177Gel Wrap® Gasket to cast one DGGE gel, 1.5mm thick
NGP-17720NRGel Wrap® Glass plate set, back plate has rounded bottom corners, 17.7cm (w) x 22 cm (h)
NGP-17720DNRGel Wrap® Glass plate set, Borosilicate back plate with rounded bottom corners, 17.7cm (w) x 22 cm (h)
NGP-17720-N-LPFGel Wrap® Glass Plate Set, both plates are Borosilicate. Back plate has rounded bottom corners, 17.7cm
GCC-404-177KMultiple Gel Casting Chamber Kit, 4-place, casts up to 4 each 15cm x 20cm gels
VGS-0720C-177Multiple Gel Casting Spacer Set, 0.75mm thick x 22m(l)
VGS-1020C-177Multiple Gel Casting Spacer Set, 1.0mm thick x 22m(l)
VGS-1520C-177Multiple Gel Casting Spacer Set, 1.5mm thick x 22m(l)
NGP-17720-NMultiple Gel Casting Chamber Glass Plate Set, 7.7cm(w) x 22cm(h)
NGP-17720-DNMultiple Gel Casting Chamber Glass Plate Set, with Borosilicate back plate, 7.7cm(w) x 22cm(h)
NGP-17720-N-LPFMultiple Gel Casting Chamber Glass Plate Set, both plates are Borosilicate, 7.7cm(w) x 22cm(h)
Cipher DGGE Electrophoresis System, 2-place, 1 dual cassette, Specify 110V or 220V. Gel size 17.7cm
(w) x 22cm (l). Includes Electrophoresis tank, 1 dual cassette, heater/stirrer/buffer cycler, gradient maker,
4 combs (2 each 1 well and 2 ea 16 well combs), 4 sets of spacers, (2 sets for vertical DGGE and 2 sets
for perpendicular DGGE), 2 glass plate sets, 4 pieces gel wrap gasket, 12ea #2 clamps, and 18ea #1
clamps, glass safety cover, buffer tank drain valve, power leads, gel tape, buffer displacement tank, EPS300 X Power Supply and MPP-100 Mini-Perstaltic Pump. Please specify thickness of combs and spacers.
Cipher DGGE Electrophoresis System, 2-place, 2 single cassettes, Specify 110V or 220V. Includes 2
single cassettes (instead of 1 dual cassette) and no buffer displacement tank, otherwise includes same as
DGGEK-1001.
Cipher DGGE Electrophoresis System, 4-place, 2 dual cassettes, Specify 110V or 220V. Gel size 17.7cm
(w) x 22cm (l). Includes Electrophoresis tank, 2 dual cassettes, heater/stirrer/buffer cycler, gradient
maker, 8 combs (4 each 1 well and 4 ea 16 well combs), 8 sets of spacers, (4 sets for vertical DGGE and
4 sets for perpendicular DGGE), 4 glass plate sets, 8 pieces gel wrap gasket, 24ea #2 clamps, and 36ea
#1 clamps, glass safety cover, buffer tank drain valve, power leads, gel tape, EPS-300 X Power Supply
and MPP-100 Mini-Perstaltic Pump. Please specify thickness of combs and spacers.
Cipher DGGE Electrophoresis System, 8-place, 4 dual cassettes, Specify 110V or 220V. Gel size 17.7cm
(w) x 22cm (l). Includes Electrophoresis tank, 4 dual cassettes, heater/stirrer/buffer cycler, gradient
maker, 16 combs (8 each 1 well and 8 ea 16 well combs), 16 sets of spacers, (8 sets for vertical DGGE
and 8 sets for perpendicular DGGE), 8 glass plate sets, 16 pieces gel wrap gasket, 48ea #2 clamps, and
72ea #1 clamps, glass safety cover, buffer tank drain valve, power leads, gel tape,EPS-300 X Power
Supply and MPP-100 Mini-Perstaltic Pump. Please specify thickness of combs and spacers.