This Instruction Manual will explain how to use this product safely and effectively.
Please read and carefully follow the instruction manual in its entirety.
The triangle/exclamation mark symbol alerts the user of the product to important operational, maintenance, and/or warranty requirements.
The triangle/lighting bolt symbol alerts the user of the product to potentially hazardous
electrical exposure.
Failure to adhere to the instructions could result in personal and/or laboratory hazards,
as well as invalidate any warranty. Always turn off the DC power source prior to disconnecting power cords from the product. Disconnect power cords from the power source
rst and then from the product. For maximum safety, always operate this system in an
isolated, low trafc area, not accessible to unauthorized personnel. Never operate dam-
aged or leaking equipment.
WARRANTY AND LIABILITY
This product was produced utilizing the highest practical standards of materials, work-
manship, and design. C.B.S. Scientic warrants that the product has been tested and
will meet or exceed published specications. This warranty is valid only if the product
has been operated and maintained according to the instructions provided.
C.B.S. Scientic warrants this product to be free from defects in materials and workmanship under normal service for one year from date of shipment. If the product
proves defective during this period, C.B.S. Scientic will repair or replace it at our
option, free of charge, if returned to us postage prepaid. This warranty does not
cover: damage in transit, damage caused by carelessness, misuse or neglect, normal
wear through frequent use, damage caused by solvent corrosion, damage caused by
improper handling or user alteration, nor unsatisfactory performance as a result of con-
ditions beyond our control. C.B.S. Scientic shall in no event be liable for incidental nor
consequential damages, including without limitation, lost prots, loss of income, loss
of business opportunities, loss of use and other related damages, however caused, nor
any damage arising from the incorrect use of the product.
3
www.cbsscientic.com
FRANÇAIS INFORMATION IMPORTANTE À L’USAGE DES UTILISATEURS
Le présent manuel d’utilisation explique la manière de se servir efcacement du produit en conditions
de sécurité. Il est recommandé de soigneusement lire la totalité du manuel, avec ses consignes
et ses instructions.
Le triangle avec point d’exclamation est un symbole destiné à avertir l’utilisateur du produit
de l’importance de certaines exigences relatives au fonctionnement, à l’entretien et/ou à
la garantie.
Le triangle avec èche en zigzag est un symbole destiné à avertir l’utilisateur du produit de la
possibilité d’exposition à des décharges avec danger de secousses électriques.
Tout manquement à l’observation des consignes et des instructions peut exposer les personnes
et les biens à des dommages corporels et/ou matériels et peut annuler toute garantie. Il faut toujours
interrompre l’alimentation de courant continu avant de déconnecter les cordons d’alimentation du
produit. Déconnecter d’abord les cordons d’alimentation branchés sur la source de tension (alimentation
de secteur) puis ceux branchés sur le produit. Pour une sécurité maximum, il faut toujours faire
fonctionner ce système dans un lieu isolé, peu fréquenté, où le personnel non autorisé n’a pas accès. Ne
jamais faire fonctionner un matériel endommagé ou affecté par des fuites.
GARANTIE ET RESPONSABILITÉ
Le produit a été fabriqué conformément aux normes applicables les plus exigentes en matière de
matériaux, de main d’oeuvre, de conception et d’ingéniérie. C.B.S. Scientic garantit que le produit a subi
des essais et que ses performances rempliront les conditions des spécications publiées ou leur seront
même supérieures. La présente garantie n’est valide que si le produit a fonctionné et a été entretenu
conformément aux consignes et instructions fournies.
C.B.S. Scientic garantit que le produit sera dépourvu de vices de matériaux et de main d’oeuvre, en
conditions de service normales, pendant un an à compter de la date d’expédition. Au cas où le produit
s’avérerait défectueux pendant cette période de garantie, C.B.S. Scientic réparera ou remplacera le
produit, à sa discrétion et gratuitement, si le produit lui est retourné port payé d’avance. La garantie
ne couvre pas les dommages de transport; les dommages causés par l’imprudence, le manque de
soins, l’abus ou la négligence; l’usure normale résultant d’une utilisation fréquente; les dommages
causés par la corrosion des solvants; et les dommages causés par la manipulation inadéquate ou
des changements apportés par l’utilisateur. La garantie ne couvre pas non plus les performances non
satisfaisantes résultant de conditions hors du contrôle de C.B.S. Scientic. C.B.S. Scientic ne pourra
en aucun cas être tenue responsable de dommages indirects, y compris, de manière non limitative, la
perte de bénéces, le manque à gagner, la perte d’occasions d’affaires, l’impossibilité d’usage ou tous
autres dommages associés, quelle qu’en soit la cause, ni de dommages résultant de l’usage incorrect
du produit.
ESPAÑOL INFORMACIÓN IMPORTANTE PARA EL USUARIO
El presente instructivo explica la manera de usar este producto en forma segura y efectiva. Sírvase
leerlo en su totalidad y seguir detenidamente las indicaciones que contiene.
El símbolo del triángulo con exclamación llama la atención del usuario a requisitos
importantes para el uso y mantenimiento del producto, así como para la validez de
la garantía.
El símbolo del triángulo con rayo llama la atención del usuario a la posibilidad de riesgos
eléctricos.
El incumplimiento de las instrucciones aquí señaladas podría dar lugar a riesgos a la persona, al
laboratorio o a ambos y podría anular toda garantía. Siempre apague la fuente de corriente continua
antes de desenchufar los cables eléctricos del producto. Primero desconecte los cables de la fuente de
energía y después del producto. Para mayor seguridad, siempre use este sistema en un área aislada,
de poco movimiento de personas e inaccesible a personal no autorizado. Jamás use equipo que
presenta algún daño o fuga.
GARANTÍA Y RESPONSABILIDAD
Este producto fue fabricado de acuerdo con las normas más estrictas que sean factibles en cuanto a
materiales, mano de obra y diseño. C.B.S Scientic garantiza que se sometió el producto a pruebas
y que cumplirá o excederá las especicaciones publicadas. Esta garantía será válida únicamente si
se usa y se da servicio de mantenimiento al producto de acuerdo con las instrucciones señaladas.
C.B.S. Scientic garantiza que este producto se encontrará libre de defectos de materiales y mano de
obra por un período de servicio normal de un año a partir de la fecha de embarque. Si el producto resulta
defectuoso durante este período, C.B.S. Scientic lo reparará o lo repondrá, a criterio de C.B.S., libre
de cargos, si se devuelve el producto a C.B.S. porte pagado. Esta garantía no cubre daños sufridos en
tránsito, daños provocados por descuido, mal uso o negligencia, desgaste normal como consecuencia
del uso excesivo, daños atribuibles a corrosión provocada por solventes, daños causados por el uso
indebido o alteraciones realizadas por el usuario ni rendimiento insatisfactorio atribuible a circunstancias
fuera del control de C.B.S. Scientic. C.B.S. Scientic en ningún caso asumirá responsabilidad por
daños incidentales o subsecuentes, incluyendo, en forma no limitativa, la pérdida de utilidades, de
ingresos, de oportunidades comerciales o del uso del producto y otros daños anes, fuere cual fuere su
origen, ni por daños derivados del uso incorrecto del producto.
DEUTSCH WICHTIGE INFORMATION FÜR DEN BENUTZER
Diese Bedienungsanleitung beschreibt wie man dieses Produkt sicher und wirksam benutzt. Bitte lesen
und befolgen Sie alle Anweisungen in dieser Anleitung.
Das Dreieck mit Ausrufezeichen weist den Benutzer des Produktes darauf hin, daß
wichtige Bedienungs-, Wartungs- und/oder Garantievorschriften zu beachten sind.
Das Dreieck mit Zickzackblitz warnt den Benutzer des Produktes vor möglichen
Gefahren durch elektrische Spannungen.
Nichtbeachtung dieser Anweisungen kann zu persönlichen und/oder labortechnischen Schäden
führen und gleichzeitig alle Garantien als nichtig erklären. Die DC Stromzufuhr muß immer, vor dem
Entfernen der Stromkabel vom Produkt, abgeschaltet werden. Die Stromzufuhrkabel müssen zuerst von
der Steckdose und erst dann vom Produkt entfernt werden. Um höchste Sicherheit zu gewährleisten
sollte dieses System in einem abgesonderten und besonders ruhigen Bereich eingesetzt werden und
vor Unbefugten sicher sein.
GARANTIE UND HAFTUNG
Dieses Produkt wurde unter Anwendung von Produkten mit höchster Qualität und aus Materialien mit
bester Verarbeitung und modernstem Design hergestellt. C.B.S Scientic garantiert, daß das Produkt
getestet wurde und alle publizierten Spezikationen übertrifft. Diese Garantie ist jedoch nur gültig, wenn
das Produkt nach der beigefügten Bedienungsanleitung bedient und gewartet wurde.
C.B.S. Scientic garantiert, daß dieses Produkt bei normaler Bedienung aus fehlerfreiem Material
besteht und fehlerfrei in der Ausführung ist. Diese Garantie gilt für ein Jahr ab Lieferdatum. Sollte
das Produkt in diesem Zeiraum fehlerhaft werden, bietet C.B.S. Scientic eine kostenlose Reparatur
bzw. kostenlosen Ersatz, einschließlich freiem Rückporto. Diese Garantie schließt folgendes aus:
Transportschaden, Schaden durch Nachlässigkeit, Mißbrauch oder Vernachlässigung, normale
Abnützung durch regelmäßigen Gebrauch, Schaden durch Säureangriff, Schaden durch falsche
Handhabung, Veränderung des Produktes durch den Benutzer, oder unzureichende Leistungen die
sich nicht im Verantwortungsbereich von C.B.S. Scientic benden. C.B.S. Scientic kommt unter
keinen Umständen für folgende Schäden auf: Sachschadensverlust, Einkommensverlust, Verlust
von Geschäftsmöglichkeiten, Verlust der Anwendung und andere damit verbundene Schäden die auf
irgend eine Art und Weise entstanden sind, oder Schäden die aus falscher Anwendung des Produktes
entstanden sind.
ITALIANO INFORMAZIONI IMPORTANTI PER L’UTENTE
Questo manuale spiega come utilizzare questo prodotto in maniera sicura ed efciente. Si pregai di
leggere e seguire con cautela le istruzioni di ogni parte di questo manuale.
Il triangolo contenete il simbolo di un punto esclamativo avverte l’utente di importanti
requisiti relativi al funzionamento, manutenzione e/o garanzia del prodotto.
Il triangolo contenete il simbolo di un lampo avverte l’utente del prodotto della possibilità
di pericoli dovuti a corrente elettrica.
La mancata osservanza delle istruzioni può essere causa di pericolo alla propria persona ed
al laboratorio, oltre a poter annullare la garanzia. Prima di distaccare il cordone d’alimentazione dal
prodotto, spegnere sempre la sorgente di corrente continua. Distaccare i cordoni d’alimentazione prima
dal lato della sorgente di tensione e poi dal lato del prodotto. Per maggior sicurezza, mettere sempre
in funzione il prodotto in un’area isolata con poco trafco che non sia accessibile al personale non
autorizzato. Non mettere mai in funzione un’apparecchiatura che sia danneggiata o abbia perdite.
GARANZIA E RESPONSABILITÀ
Questo prodotto è stato fabbricato seguendo gli standard più elevati per i materiali, la manodopera e la
progettazione. La C.B.S. Scientic garantisce il prodotto è stato sottoposto a prova e raggiunge o supera
i valori pubblicati per i dati tecnici. Questa garanzia è valida solo se il prodotto è messo in esercizio e
soggetto a manutenzione secondo le istruzioni fornite.
La C.B.S. Scientic garantisce che questo prodotto è libero di difetti di materiali e manodopera, in normali
condizioni d’esercizio, per la durata di un anno dalla data di spedizione. Se, in questo periodo, il prodotto
si dimostrerà difettoso, la C.B.S. Scientic, a suo giudizio, lo riparerà o sostituirà. Questa garanzia non
copre danni in transito, danni causati da negligenza, uso improprio, trascuratezza, normale consumo
derivante da uso frequente, o danni causati da solventi corrosivi, danni causati da maltrattamento o da
modiche apportate dall’utente e non copre prestazioni insoddisfacenti che siano il risultato di condizioni
al di fuori del controllo del fabbricante. La C.B.S. Scientic non sarà in ogni caso responsabile per danni
incidentali o consequenziali, incluso, senza limitazioni, perdite di protto, perdita di entrate, perdita di
opportunità d’affari e altri danni relativi, comunque causati, e per danni risultati da uso incorretto del
prodotto.
4
DCX-900C Large Format Instructions 9-30-13
SECTION 1
General Information
1.1 Introduction and Features
C.B.S. Scientic offers the Large Format Cooled Vertical is a temperature controlled
large capacity system for running SDS PAGE and 2-D analysis. The increased gel sizze
of 20cm(w) x 22cm(l) allows a larger sample volumes, increased number of wells and
longer migration distance.
volume is calculated by
taking 75% of the total
tooth volume.
*Comb conguration is
one large lane and one
marker lane
1.4 Safety
Power to the Large Format Cooled Vertical Electrophoresis System is to be supplied
by an external DC voltage power supply that must be ground isolated so that the DC
voltage output oats with respect to ground. For any power supply used, the maximum specied operating parameters for the units are:
Maximum Limits
300 VDC
30 watts power
190mA current
60°C ambient temperature
Current to the unit, provided from the external power supply, must enter the unit
through the lid assembly, providing a safety interlock to the user. Current to the unit
is broken when the lid is removed. Do not attempt to use the unit without the
safety lid, and always turn the power supply off before removing the lid, or when
working with the unit in any way. Follow safety precautions specied by the
power supply manufacuturer.
6
DCX-900C Large Format Instructions 9-30-13
SECTION 2
Description of Parts
2.1 Unpacking and Components
Please verify that your unit comes complete with the following components:
Large Format Cooled Vertical Electrophoresis System:
• Lower Reservoir with double-sided cooling labyrinth, tubing adapters
and stirring corral
• Safety cover
• Power leads (must be attached before using. See section 3.1.1)
• Core
• 2 sets of Gel Wrap Gasket (not shown)
• 2 sets of Gel Wrap Glass Plates (not shown)
• 2 sets of Gel Wrap Spacers (not shown)
• 2 Combs (not shown)
• 12 White clamps (not shown)
• Conversion plate for running single gels (not shown)
Safety Cover
Lower reservoir with side cooling labyrinths
7
Core
Tubing adaptors
Power leads
www.cbsscientic.com
SECTION 3
Instructions for Slab Gel Electrophoresis
See Section 4 for casting gels using the Gel Wrap method. You may also use
the 4-place Multi-Caster DCX9-MC-4.
3.1 Preparing the Electrophoresis Unit
1. Power leads are shipped separately and
must be attached before using your
Large Format Cooled Vertical. Thread
the power lead ends into the lid receptacles and rotate in clockwise direction to
hand tighten.
2. Place unit in authorized work area. Remove safety cover from the assembled
unit by simultaneously pressing down on
white push pins while lifting up on blue
safety cover as shown in gure 1.
Do not remove safety cover by pulling
up on leads!
1
3. Remove white core from lower reservoir
by grasping core with one hand and lifting
directly up as shown in gure 2.
4. Open doors on the core assembly by pulling up on the white latches.
8
2
DCX-900C Large Format Instructions 9-30-13
5. Slide gel plate set(s) into the core
assembly with the notched plate
facing in towards the upper buffer
reservoir as shown in gure 4.
6. If running one gel, slide white plastic
conversion plate into the side without
the gel.
3
7. Close doors and relatch by pressing
down on the white latches so that
the assembly looks like that shown in
gure 6.
4
5
9
www.cbsscientic.com
3.2 Running the Gel
1. Place stirring bar into bottom of reservoir in
stirring corral (as shown in gure 6). Place
core assembly into lower reservoir. The
anode (red) and cathode (black) electrodes
are color-coded on both the core/cassette
assembly and lower reservoir. Ensure the
red dot on the cassette assembly is on the
same side as the red receptacle on the lower
reservoir. Fill core upper reservoir with freshly
prepared buffer (~ 190mls).
If any buffer is spilled into banana jack
receptacles (outlined in yellow boxes in
gure 6) in lower reservoir, dry completely
using compressed air! Failure to do this
will result in accelerated banana jack corrosion.
2. If using cooling capability of DCX-900C at-
tach external cooler to tubing adapters and
begin circulating coolant through base.
stirring bar
6
3. Pour only enough freshly prepared buffer into
lower chamber so that the nal buffer level
is just below bottom of sample wells. Using
a pipette or syringe, thoroughly ush out the
wells in the glass plate sandwich with buffer.
Load samples. If outer lanes do not contain sample, it is recommended that you run
standards and/or ll outer lanes with loading
buffer to reduce smiling and wrap-around effects.
4. Attach safety cover and turn on magnetic
stirrer. The closed unit ready for power is
shown in gure 7.
5. Connect the leads to the power supply,
matching the color-coded red to red and
black to black. See Section 4.1 for recom-mended power conditions. Begin separation by electrophoresis.
10
7
DCX-900C Large Format Instructions 9-30-13
3.3 Removing the Gel
1. Turn the power supply off and disconnect
the leads from the power supply. Remove
the safety cover from the unit, by placing
thumbs on white posts next to red & black
connectors, then pushing down while pulling
up with ngers under lid as shown in gure
8.
Do not remove safety cover by pulling up
on leads!
2. Pull up on gel door latches, and open gel
door. Remove gel sandwich from Cassette
Assembly. Stain and x according to your
preferred method.
8
11
www.cbsscientic.com
3.4 Gel Casting Using Gel Wrap™ Gasket Casting Method
1. Place all components in an authorized work area. You will need: glass plate set, Gel
Wrap™ Gasket, spacer set, comb, 3 GPC-0002 clamps, and polyacrylamide solution. Prepare and clean glass plates by hand washing both plates with a high quality
lab detergent followed by a complete rinsing with dH2O. Air-dry or use a lint-free
tissue. Spray/wipe the chosen inner surfaces of the plate set with 95% ethanol and
dry with lint-free tissue.
2. Hold the rectangular back plate with
the rounded bottom corners and start
applying the gasket around one side of
the glass plate. Note: one side of the “U”
shaped gasket is at, and the other side
has tubing that will act as a seal around
the spacers.
3. When applying the gasket over the
rounded corners of the back glass plate,
make sure the notches on the gasket
align with the rounded corners of the
glass plate. Once the gasket is pushed
over the bottom edge and corners, work
it down the remaining side.
4. Place the gasketed plate on the lab
bench with the tubing side up, and
extend the bottom of the plate over the
edge of the bench, approximately ¾ of
an inch. Place the spacers along side
the inside edges of the gasket. Be sure
the rounded corner end of each spacer
is facing the outside bottom of the plate,
following the radius of the glass.
12
DCX-900C Large Format Instructions 9-30-13
5. Place the notched plate on top of the bottom
assembly, starting from the bottom edge and
gently easing the plate down. Verify the gasket
is smooth around the edges and then clamp
along the bottom.
6. Lift the assembly and stand it on the base
of the clamps. For leveling, push glass plate
assembly down until it stops against clamp
body. Clamp the sides of the assembly with
additional casting clamps on either side. As
each clamp is attached, be sure the gasket is
aligned between the plates forming a seal.
7. Apply PAGE or agarose solution to gel plate
sandwich using a syringe or pipette. If using
a stacking gel, pour desired height of running
gel, then overlay a small amount of dH2O
or 0.1% SDS solution to top of gel. After
polymerization, rinse with buffer, add stacking
gel solution and insert comb. For regular, unit
percentage gels, add polyacrylamide solution
to correct height, and insert comb. Allow gel
to set, usually 20 minutes. Extra gel solution
in pipette or syringe can be monitored to test
polymerization of gel mix.
8. Disassembly. Hold the clamped plate assembly with one hand. Remove the gasket by
starting at one of the top ends and pulling up
and out on the gasket until it releases from the
plate, up to the bottom of each of the white
clamps. When each clamp is reached DO
NOT remove it, instead feed the gasket down
through the clamp body and repeat pulling up
and out. Continue feeding until the gasket is
fully detached. If gel, is not to be used immediately, wrap entire plate sandwich with plastic
wrap tightly to seal and store at 4°C for up to
a month.
13
www.cbsscientic.com
SECTION 4
Running Conditions
4.1 Recommended Power for Slab Gels:
Precise electrophoresis conditions will vary according to the number and type of
gels used, buffer conditions employed, power input, and the general goal of the
experiment. Refer to reference section 4.2 for in depth discussions on practical and
theoretical approaches to protein gel electrophoresis.
4.1.1 SDS PAGE Gels
Run VoltageStarting CurrentEnding VoltageApprox. Run TIme
90VDC30mA/gel CC110 volts/gel4-5 hours
4.1.2 General Recommendations
If running only one gel, keep the volts the same but reduce the mA’s by half. Keep in
mind that as the thickness of gel increases, the mA’s increase proportionally.
At constant voltage, the proteins will migrate at a constant rate during electrophoresis with adequate heating appropriate for denaturing gels. Increasing the voltage/
mA (for a single gel thickness and percentage) will speed mobility but increase the
risk of overheating.
If using cooling base the power input and the migration rate can be increased. The
joule heating generated by the higher power is offset by the cooling effect of the
buffer between the gels.
4.2 References
1. Hames, B.D. (ed.) (1998). Gel Electrophoresis of Proteins. A Practical
Approach. 3
2. Sambrook, J., Fritsch, Russell, D. (2001). Molecular Cloning. A Laboratory
Manual. 3
New York. A8.40-A8.55
3. Ausubel, F.M., Brent, R., Kingston, R.E., Moore, D.D., Seidman, J.G., Smith,
J.A., Struhl, K. (ed) (1993). Current Protocols in Molecular Biology. Vol. 2,
Greene Publishing Associates, Inc. and John Wiley & Sons, Inc., Ch. 10.
rd
edn. Oxford University Press, Oxford. Ch. 1,3.
rd
edn. Cold Spring Harbor Laboratory Press, Cold Spring Harbor,
14
DCX-900C Large Format Instructions 9-30-13
SECTION 5
Maintenance of Equipment
5.1 Care and Handling
The plastic components of the Large Format Cooled Vertical Electrophoresis Systems
are fabricated from acrylic and polycarbonate. Electrodes and connectors are made
from pure platinum, stainless steel, and chrome plated brass. As with any laboratory
instrument, adequate care ensures consistent and reliable performance.
After each use, rinse buffer chamber, gel tray and combs with de-ionized water. Wipe
dry with a soft cloth or paper towel, or allow to air dry. Whenever necessary, all components may be washed gently with water and a non-abrasive detergent, and rinsed and
dried as above. Never use abrasive cleaners, glass cleaning sprays or scouring pads to
clean the components, as these will damage the unit and components.
Additional precautions:
• Do not autoclave or dry-heat sterilize the apparatus or components.
• Do not expose the apparatus or components to phenol, acetone, benzene,
halogenated hydrocarbon solvents or alcohols.
• Avoid prolonged exposure of the apparatus or components to UV light.
• Do NOT treat with diethylpyrocarbonate (DEPC)-treated water for extended
periods at 37°C. A brief rinse with DEPC-water is sufcient after a thorough
wash, followed by a quick rinse in 50% ethanol.
5.2 Maintenance
The following inspection and maintenance procedures will help maintain the safety and
reliable performance of the Large Format Cooled Vertical Systems. Replacement parts
can be ordered by calling 1-858-755-4959 or by contacting your local distributor.
• Banana plugs and power cords should be inspected regularly. If the banana
plugs become loose or do not feel friction tight replace the plugs or power
cords.
• Should power cord assemblies (connectors, wire or shrouds) show any signs
of wear or damage (e.g. cracks, nicks, abrasions, or melted insulation), replace
them immediately.
• The platinum wire is secured to the banana jack by compression between a
stainless washer and the jack nut. The nut/washer interface should be tight
and free of corrosion.
15
www.cbsscientic.com
SECTION 6
Ordering Information
Large Format Cooled Vertical Ordering Information
COMBS: See page 6.
CAT. # DESCRIPTION
COOLED LARGE FORMAT GEL SYSTEM- 20CM(W)*
DCX-900C
Accessories and Related Products for Cooled Large Format Gel System