CARESTREAM INDUSTREX Digital TECHNICAL DATA

Page 1
TECHNICAL DATA / Non-Destructive Testing
CARESTREAM INDUSTREX Digit al Imaging Plates
February 2011 TI-2632
CARESTREAM INDUSTREX Flex XL, GP and HR Digital Imaging Plates (IPs) are designed for computed radiography in non-destructive testing ap p licat ion s.
This latent image is subsequently read out by scanning with a red laser, which stimulates the emission of blue light, in a phenomenon known as photostimulated luminescence.
The imaging plate can be loaded into a flexible or rigid cassette and used with metal screens of appropriate thickness.
They offer the flexibility of film without the need for wet processing, so you can capture and read images quickly and easily, both in the field and in the lab.
The images can be optimized, if needed, and can be stored and shared digitally. These imaging plates work with CARESTREAM CR Systems including the INDUSTREX HPX-1, ACR-2000 and ACR-2000i Digital Systems.
CARESTREAM INDUSTREX Flex XL, GP and HR Digital Imaging Plates are:
Flexible—Wrap around most shapes.
Sharp—Better resolution than high-speed films, and a suitable replacement for many applications.
Available in Multiple Sizes—To cover a wide range of applications.
Portable—With a CARESTREAM INDUSTREX Digital System, image capture in remote areas is easier than ever.
Versatile—Use in a rigid or flexible cassette.
Efficient—Phosphor plates can be erased and reused thousands of times, require less exposure than film, and provide faster image access.
Certified—When used with CARESTREAM INDUSTREX HPX-1, ACR-2000 and ACR-2000i Digital Systems, all plates meet requirements for the highest system class according to EN 14784-1 and ASTM E 2446.
Recommended Uses
CARESTREAM INDUSTREX Digital Imaging Plates are available in many sizes, and are designed for a wide variety of NDT applications, including:
• Castings
• Erosion and corrosion
• Pipeline inspection
• Welded fabrication
• Ordnance
Storage and Handling
The lifetime of the plates is hundreds or thousands of exposure/readout/erase cycles, determined primarily by the care in handling and environmental conditions. Careless handling and very dirty environments can shorten the useful lifetime of an imaging plate.
Important
Inspect phosphor plates every week and replace when wear is evident. Clean every 200 exposures, every 30 days, or as necessary as directed in the cleaning sections of this document.
Handling
Handling the IP and operating the HPX-1 or the ACR-2000/ACR-2000i reader must be done in a darkened room that is free of light contamination. It is not necessary to have complete darkness, but the light source must be a low wattage incandescent bulb. Fluorescent, sodium vapor, mercury vapor, safelights and sunlight must not be allowed in the processing room during handling or scanning of the exposed imaging plates.
Under normal use conditions, phosphor plates will eventually show wear. Plate wear can result in artifacts on radiographs. This may occur from abrasion of the protective overcoat or inadvertent physical damage to the surface or edge. Certain chemical agents, such as non-approved plate cleaners, hand lotions, topical medications, food, and so forth, may also damage the plates.
Handle flexible phosphor plates with care. Hold plates by the edges and by the black side. Avoid contact with the plate’s white (phosphor) side. Make sure hands are clean and dry. For best results, wear nitrile rubber or latex gloves.
© Carestream Health, Inc. 2011
Page 2
Cutting
Cutting the plates is not recommended, as this will compromise the edge seal and make the plates prone to moisture damage at the edges.
Caution
Read and follow instructions in Material Safety Data Sheets (MSDS) for KODAK Screen Cleaner.
Temperature Range
The normal temperature oper at ing ran g e is
-5 to 30 °C (23to86°F) , although lower temperatures for short periods (several hours) will have no detrimental effect on the performance or physical integrity of the plates. In very hot environments, care should be exercised in bending or flexing the plates, as the phosphor layer could crack if flexed at temperatures exceeding 65 °C (149 °F). Between 65 and 100 °C (149 and 212 °F), minor distortions of the polyester support may occur, and ov er 100 °C (212 ° F), damage to the phosphor layer (blistering/ buckling) can occur.
When radiography of very hot objects (such as hot welds) is necessary, a thin fire-resistant insulating layer may be placed between the cassette and the object, if experimentation shows it does not negatively affect image quality.
Again, as a practical matter, the upper temperature limit will be determined by the point at which the cassette begins to burn or the lead identification letters begin to melt.
Cleaning Plates
CARESTREAM INDUSTREX Digital Imaging Plates employ a phosphor that is moisture sensitive. When exposed to water (either liquid or vapor form), the phosphor breaks down and liberates iodine. The iodine can become trapped within the phospho r layer at the interface between the overcoat and phosphor layer. The yellow-brown color in the iodine strongly absorbs the blue light emitted by the phosphor. This results in a loss in plate speed, which causes image quality problems.
The plates are over-coated with a highl y moisture-resistant polymer blend. However, the overcoat is very thin (less than 0.0254 mm [0.001 in.]) and has limited resistance to mechanical abrasion that may occur during cleaning. Thus, it is possible to lower the resistance to water during use or cleaning. This can result in the degradation of the plate.
New imaging plates should be cleaned before first use. This will impart the necessary anti-static properties and prevent the attraction of dust and dirt which can result in image artifacts.
Note: Plates should be wiped routinely to remove any dust and loose dirt. It is also a good idea to clean the interior of flexible cassettes to avoid transfer of dirt or dust to the plate. Use KODAK Screen Cleaner to remove other dirt. Other cleaning solutio ns can contain chemicals that cause visible or hidden damage to the plate and could result in immediate or future image artifacts.
Cleaning Methods
For routine cleaning, place the plate with the black (non-phosphor) side down on a clean, dry, non-abrasive surface, such as the cassette’s cleaned Tube Side. Wipe the plate gently to remove dust. For other dirt, use KODAK Screen Cleaner as follows:
1. Fold a non-abrasive, lint-free wipe or cloth and
dampen with a small amount of the solution.
Important
Do not pour the solution directly onto the plate. Excessive amounts of the screen cleaner may damage the plate. Moisture can cause immediate or future screen damage and image artifacts. Minimize contact with moisture and always dry plates immediately.
2. Wipe the plate thoroughly, one section at a time. Fold the cloth to expose a fresh area for each section of the plate.
3. If necessary for the removal of persistent dirt, apply pressure while cleaning.
4. Thoroughly dry the plate with a clean and dry, non-abrasive, lint-free cloth to remove residual clean er. Do not leave the plate to air dry.
Important
Never insert a storage phosphor plate into a cassette unless it is thoroughly dry.
Important
Water, isopropyl alcohol (Isopropanol, rubbing alcohol), and plate cleaners containing isopropyl alcohol are not recommended.
2 CARESTREAM INDUSTREX Digital Imaging Plates TI-2632
Page 3
Characteristics
0
10
20
30
40
50
60
70
80
90
100
0123456
Spacial Freque ncy (cycle s/mm)
Sharpness (MTF %)
XL Blue
HR & XL
GP
Physical Properties
Thickness
Total
Thickness
(microns
[mils])
XL 428 (16.9) 150 (5.9) 6 (0.24) 18 (0.71) HR 444 (17.5) 150 (5.9) 6 (0.24) 25.4 (1) GP 605 (23.8) 300 (11.8) 12 (0.47) 38.1 (1.5)
CARESTREAM INDUSTREX Digital Imaging Plates are comprised of a layer of barium fluorobromoiodide doped with divalent Europium (BaFBr/I:Eu2+), coated on a 10-mil polyester support.
A polymer overcoat provides protect i on ag ainst normal handling such as fingerprints and moisture.
A black polycarbonate backing layer provides anti-halation protection and curl control, and also contains slip agents for optimum transport through the CR reader hardware.
Flexibility
The plates can be bent around pipes and other curved surfaces down to a radius of one inch, although as a practical matter, the bend radius will be much more limited by the configuration of the cassette and the lead screens, when used.
Phosphor
layer
(microns
[mils])
Overcoat (microns
[mils])
Backing
Layer
(microns
[mils])
Signal Retention
The stored energy in the imaging plate ph os ph o rs will decay over time, resulting in some signal loss. This effect is known as image fading. This loss is also a function of temperature, and will be greater at elevated temperatures.
In most applications this signal decay will be of little importance, given the very wide latitude of stora ge phosphor imaging plates.
In critical applications, there are a couple of measures that can be taken; when the interval between exposing and readout is more than a couple of hours, the plates can be exposed with a higher radiation dose (greater milliamp-seconds) to compensate for signal decay.
For shorter intervals, simply keeping the time between exposing and readout constant will eliminate any decay-induced variability.
The figure below shows the percent signal retention in the first two hours at ambient temperatures after an exposure of 80 kV.
Signal Retention
Sharpness
Sharpness is typically characterized by Modulation Transfer Function, a measure of signal or image modulation as a function of spatial frequency. The figure below shows the MTF curves for all imaging plates, and highlights the improved sharpness of the HR and XL plates.
Modulation Transfer Function
CARESTREAM INDUSTREX Digital Imaging Plates TI-2632 3
H192_3008AA
Page 4
Absorption
X-ray Absorption vs. X-ray Energy
0.01
0.1
1
0 50 100 150
X-ray Energy (kV)
Relativ e (Pho ton) A bsorp tion
GP
XL & XL Blue
GP, HR, XL & XL Blue Emission
0.0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
300 350 400 450 500 550 600 650 700 750 800
Wavele ngth (nm)
Relative Emission
Imaging plates are quite sensitive to low energy scatter, as shown by their x-ray energy absorption curves below.
For this reason, the use of lead screens, especially at energies greater than 100 kV, is important in controlling scatter to produce optimum image quality.
The thickness of the screens is determined by experimentation, however it will generally be two to three times the thickness of a comparable film application.
Absorption
Disposal
Due to the presence of barium, this plate may be considered a hazardous or special waste at the end of its useful life.
For disposal or recycling information, please contact local authorities.
To locate the Article Information Sheet (AIS) for each plate, go to www.carestream.com and navigate to Support and Education > Environment Health & Safety.
H192_3010AA
Emission
The wavelength of light needed to stimulate phosphor luminescence is different from the wavelength of light released by the phosphors. This enables the stimulating light to be filtered from the phosphor luminescence.
The emission is typically stimulated with a red laser diode, while the photostimulated emission of the XL, GP and HR imaging plates is a narrow band of blue light near 400 nm.
Emission
4 CARESTREAM INDUSTREX Digital Imaging Plates TI-2632
H192_3011AA
Page 5
CARESTREAM INDUSTREX Digital Imaging Plates
NOTICE:
While the sensitometric data in this publication are typical of production coatings, they do not
represent standards which must be met by Carestream Health, Inc. Varying storage, exposure, and
processing conditions will affect results. The company reserves the right to change and improve
product characteristics at any time.
INDUSTREX is a trademark of Carestream Health, Inc.
The Kodak trademark and trade dress are used under license from Kodak.
Pub No. TI-2632 Revised 2011-02-16 (February 16, 2011) Printed in the U.S.
Carestream Health, Inc. 150 Verona Street Rochester, NY 14608
Loading...