Cardinal Detecto PS-5 Owner's Manual

Page 1
DIGITAL PORTION SCALE
(
PS-5
OWNER’S Manual
8528-M013-O1 Rev G 09/00
CARDINAL SCALE MFG. CO.
PO BOX 151 WEBB CITY, MO 64870
417) 673-4631 FAX (417) 673-5001
PH
Web Site - http://www.detectoscale.com
0
Printed in USA
Page 2
2
Page 3
INTRODUCTION
A
f
r r r
f
r
TABLE OF CONTENTS
The PS-5 Digital Portion Scale is a compact solid scale, designed to fill a variety of weighing needs. Please read this manual before attempting to put scale into use and keep it handy for future reference.
FCC COMPLIANCE STATEMENT
WARNING! This equipment generates, uses
and can radiate radio frequency and if not installed and used in accordance with the instruction manual, may cause interference to radio communications. It has been tested and found to comply with the limits for a Class A computing device pursuant to Subpart J of Part 15 of FCC rules, which are designed to provide reasonable protection against such interference when operated in a commercial environment. Operation of this equipment in a residential area may cause interference in which case the user will be responsible to take whatever measures necessary to correct the interference.
ll rights reserved. Reproduction or use, without expressed written permission, o editorial or pictorial content, in any manner, is prohibited. No patent liability is assumed with respect to the use of the information contained herein. While every precaution has been taken in the preparation of this manual, the Selle assumes no responsibility for errors o omissions. Neither is any liability assumed fo damages resulting from use of the information contained herein. All instructions and diagrams have been checked for accuracy and ease o application; however, success and safety in working with tools depend to a great extent upon the individual accuracy, skill and caution. For this reason the Seller is not able to guarantee the result of any procedure contained herein. Nor can they assume responsibility for any damage to property o injury to persons occasioned from the procedures. Persons engaging the procedures do so entirely at their own risk.
Specifications -----------------------1
Installation -------------------------2
Battery Replacement ------------------2
Operation --------------------------2
Automatic Shut-Off Feature -------------2
Display Codes ----------------------3
Setup and Calibration -----------------3
Setup Review -----------------------4
SPECIFICATIONS
Dimensions - - - 7 1/2" W x 7" D x 2 3/4" H
Weight -------4 lb (without batteries) Construction - - - Steel enclosure with 304
stainless steel tray
Power -------6 VDC four AA alkaline
batteries 115 VAC 50/60 Hz with optional AC adapter
Display -------3 1/2 digits in an LCD display
with low battery indication and LED capacity indicators
Capacity ------Dual key-selectable
capacities: 70 x 0.1 oz/1998 x 2g
Zero ---------Established during power-on
lamp test and maintained automatically by zero-tracking circuitry
Tare -------- Key actuated full capacity tare
feature
CAUTION!!
This sensitive weighing instrument is equipped with a precision transducer. Properly handled, it will provide years of accurate weighing operation. Although this instrument is protected against moderate overloads, extreme overloading, dropping the instrument, or subjecting it to physical shock or abuse may damage or destroy the instrument.
Serial Number _______________________
Date of Purchase_____________________
Purchased From _____________________
___________________________________
___________________________________
RETAIN THIS INFORMATION FOR FUTURE USE
PRECAUTIONS
Before using this instrument, read this manual and pay special attention to all "WARNING" symbols:
IMPORTANT ELECTRICAL
WARNING
1
Page 4
INSTALLATION
1. Remove the scale from its shipping container and inspect it for shipping damage. Should
evidence of damage be found, the carrier should be contacted at once. Retain the shipping container for future storage or shipment of the scale.
2. Install the batteries. Refer to BATTERY REPLACEMENT section. If batteries are not to be
used, plug the optional AC power adapter into the power jack on the back of the scale.
3. Place the scale on a stable, vibration-free location away from direct sunlight and any rapidly
moving air source. If the AC power adapter is being used, make certain the cord is stored out of the way of normal traffic.
4. Plug the optional AC power adapter (if used) into a 115 VAC 50/60 HZ grounded wall outlet.
5. The scale is now ready for use.
BATTERY REPLACEMENT
This instrument is designed for operation from either four (4) AA Alkaline batteries or an optional AC power adapter. A low battery signal is provided in the weight display to indicate when the batteries should be replaced. Remember to use only size AA Alkaline batteries in the instrument.
1. Turn the scale bottom side up and remove the single screw from the battery compartment cover. Remove the cover from the base by lifting up and pulling outward as shown.
2. Remove the old batteries and replace with fresh ones.
Make certain the + and - ends of the batteries match the positions shown on the scale circuit boards as shown. Replace the battery compartment cover and retaining screw.
Battery Compartment Cover
Pull cover out and up
OPERATION
1. TURN THE SCALE ON - Press the ON-TARE key.
2. SELECT THE WEIGHING UNIT - Press the WT-UNIT
key to alternate between ounces and gram weight displays. A red indicator light will illuminate to show the selected units. NOTE! In Grams only mode, the
WT-UNIT key is disabled.
3. WEIGH - Place the item to be weighed on the tray and
read the weight in the digital display.
4. TO RE-ZERO THE WEIGHT DISPLAY - Press the
ON-TARE key. The instrument will zero up to the full capacity of the scale.
The membrane keyboard is not to be operated with pointed objects (pencils, pens, fingernails, etc). Damage to keyboard resulting from this practice will NOT be covered under warranty.
Batteries
AUTOMATIC SHUT-OFF FEATURE
This scale is equipped with an automatic shut-off feature that helps prolong battery life by turning the scale off after a predetermined period of inactivity. To turn the scale back on you must press the ON-TARE key. The automatic shut-off feature is activated during the setup and calibration procedure and can be set for 1, 2, 5, 10, or 15 minutes of inactivity before the scale will automatically shut off. The automatic shut-off feature may be disabled by selecting OFF during setup and calibration procedure.
2
Page 5
DISPLAY CODES
CALIBRATION REQUIRED - Indicates improper stored calibration data, calibration
CALCAL
OCOC
is necessary.
OVER CAPACITY - Indicates a weight exceeding the capacity of scale has been placed on the scale tray.
ERROR - NOV-RAM test failure.
ErrErr
ANALOG FAILURE - This could mean a damaged or defective load cell, a failed
ErAErA
Your scale has been thoroughly tested and calibrated before being shipped to you and should not require adjustment. In the event the scale should need re-calibration, the following calibration procedure should be performed. To maintain the scale's high degree of accuracy, it is recommended that a qualified technician with certified test weights re-calibrate the scale.
In order to perform Setup and Calibration, it is necessary to remove the "Do Not Remove" sticker from the lower right side of the scale. With the sticker removed, locate the Setup and Calibration switch in the opening closest to the front of the scale. Refer to the figure on the next page.
component in the analog circuity of the scale or the platform isn't on or is on incorrectly.
SETUP AND CALIBRATION
NOTE: If operating from batteries, install the batteries making certain that they are positioned in accordance with the drawing on the circuit board. If using the optional AC power adapter, insert the plug into the jack on the back of the scale.
BEGIN SETUP and CALIBRATION
1. Making certain the scale tray is empty, press and hold the setup and calibration switch located on the lower right side of the scale. Refer to the figure on the next page.
2. Press the ON/TARE key to turn the scale on. Release the setup and calibration switch when
the display is showing SCASCA.
SCALE MODEL
The display will show SCASCA. Press the WT/UNIT key to display the current setting. For a PS-5 the display should show PP--55. Press the WT/UNIT key until PP--55 is displayed. Press the
ON/TARE key to save the setting.
CALIBRATION
The display will show CALCAL. Press the WT/UNIT key to display the current setting. To calibrate the scale, press the WT/UNIT key until YESYES is displayed, then press the ON/TARE key. The display will change to LOdLOd.
To skip calibration, press the WT/UNIT key until nono is displayed, then press the ON/TARE key to continue the setup.
LOAD WEIGHT
With the display will showing LOdLOd, place a (2) lb certified test weight on the scale tray. Allow approximately five seconds for the scale to stabilize. Press the ON/TARE key to start the calibration. Dashes will be displayed from the left to the right of the display momentarily, then the display will change to uLduLd.
3
Page 6
SETUP AND CALIBRATION, CONT.
UNLOAD WEIGHT
With the display showing uLduLd, remove the 2 lb test weight from the scale tray. Allow approximately five seconds for the scale to stabilize, then press the ON/TARE key to continue the calibration. Dashes will be displayed from the left to the right of the display momentarily, then the display will change to untunt.
If ErAErA is displayed during the LOdLOd or ulduld calibration sequence, remove the test weight from the scale and verify that it is the correct weight (2 lbs) or insure that the scale is empty and nothing is touching the scale tray. The display will return to LOdLOd or ulduld allowing you to continue with calibration and setup.
WEIGHING UNITS
With the display showing untunt, press the WT/UNIT key to display the current setting for the default weighing unit. The display will show ------ and turn on the red indicator for the weight unit (oz or gm) or 11 and the red indicator for (gm) for the "Grams only" mode. Press the WT/UNIT key until the desired weight unit is indicated. Press the ON/TARE key to save the setting.
AUTOMATIC SHUTOFF
The display will show ASHASH. Press the WT/UNIT key to display the current setting. Press the WT/UNIT key until the desired number of minutes of inactivity before the scale will
automatically shutoff is displayed. Values of 1, 2, 5, 10, or 15 minutes or OFF are available. Selecting OFF will disable the automatic shutoff feature. Press the ON / TARE key to save the setting.
POWER UP ZERO
The display will show PuOPuO. Press the WT/UNIT key to display the current setting. Press the WT/UNIT key to select YESYES to enable or no no to disable. Press the ON / TARE key to save the
setting.
DONE
The scale will briefly display dnEdnE, then reset to the Operation weight mode.
SETUP REVIEW
The Setup Review mode allows you to review and change several operational parameters as desired without having to enter the setup and calibration mode. To enter the setup review mode, you need to turn the scale off by pressing the WT/UNIT and ON-TARE keys simultaneously and then pressing and holding the ON-TARE key until the red indicator of the selected weight units begins to flash. After the red indicator begins to flash, release the ON-TARE key. The display will show untunt, the prompt for the weighing units. Refer to the instructions listed in the Setup and Calibration section for information on how to change these parameters. The parameters in the setup review that can be change are as follows:
WEIGHING UNITS AUTOMATIC SHUTOFF
Location of Calibration Switch
4