Cardinal APS8, APS12, APS160, APS10, APS15 Setup And Operation Manual

...
APS
ENTERPRISE SCALE
Setup and Operation Manual
CARDINAL SCALE MFG. CO. 8527-0544-2M Rev G 03/15 www.detectoscale.com
Technical Support: Ph: 866-254-8261 techsupport@cardet.com
PO BOX 151 WEBB CITY, MO 64870
PH (417) 673-4631  FAX (417) 673-5001
8527-0544-2M Rev G APS Enterprise Scale I
8527-0544-2M Rev G APS Enterprise Scale II
TABLE OF CONTENTS
SPECIFICATIONS ................................................................................................. Page 1
Options .............................................................................................................. Page 1
Certifications ...................................................................................................... Page 1
SITE PREPARATION REQUIREMENTS .............................................................. Page 2
Site Requirements ............................................................................................. Page 2
Environmental Conditions .................................................................................. Page 2
Electrical Power ................................................................................................. Page 3
Electrical Noise Interference .............................................................................. Page 3
Transient Suppression ....................................................................................... Page 3
RFI Immunity ...................................................................................................... Page 3
INSTALLATION ...................................................................................................... Page 4
ALL MODELS EXCEPT APS250 ....................................................................... Page 4
Unpacking ...................................................................................................... Page 4
Level Adjustment ........................................................................................... Page 4
APS250 ONLY ................................................................................................... Page 5
Unpacking and Setup .................................................................................... Page 5
Level Adjustment ........................................................................................... Page 6
Interconnections ................................................................................................ Page 7
Data Cable Installation ....................................................................................... Page 7
Scale Display/Serial Port 1 ............................................................................ Page 7
Remote Display/Serial Port 2 ........................................................................ Page 7
POS/Serial Port 3 .......................................................................................... Page 7
USB Cable Installation ....................................................................................... Page 7
Power Adapter ................................................................................................... Page 7
Serial Data Cable, 8527-0545-0A ...................................................................... Page 8
KEYPAD FUNCTIONS ........................................................................................... Page 9
ANNUNCIATORS .................................................................................................. Page 9
SETUP AND CONFIGURATION ........................................................................... Page 10
To Enter Setup ................................................................................................... Page 10
type (Scale Type) ........................................................................................
Uns (Motion Range) ......................................................................................
Filter (Adjustable Digital Filtering) ............................................................
Calib (Calibration) ......................................................................................
Port1 (Serial Port 1) ....................................................................................
Port2 (Serial Port 2) ....................................................................................
Port3 (Serial Port 3) ....................................................................................
USB (USB Port) .............................................................................................
HirES (High Resolution Weight Mode) ........................................................
btdis (Button Disable) ................................................................................
ERROR AND STATUS DISPLAYS ........................................................................ Page 15
CARE AND CLEANING ......................................................................................... Page 15
SECURITY SEAL INSTALLATION ........................................................................ Page 16
APSRD (REMOTE DISPLAY) ................................................................................ Page 17
PARTS IDENTIFICATION ...................................................................................... Page 18
ALL MODELS EXCEPT APS250 ....................................................................... Page 18
APS250 ONLY ................................................................................................... Page 22
SCALE DISPLAY AND APSRD (REMOTE DISPLAY) ...................................... Page 26
Page 10 Page 11 Page 11 Page 11 Page 12 Page 12 Page 13 Page 13 Page 14 Page 14
SERIAL NUMBER ____________________ DATE OF PURCHASE ________________ PURCHASED FROM _________________ ___________________________________ ___________________________________ ___________________________________
RETAIN THIS INFORMATION FOR FUTURE USE
PRECAUTIONS
Before using this scale, read this manual and pay special attention to all "NOTIFICATION" symbols:
IMPORTANT ELECTRICAL
8527-0544-2M Rev G APS Enterprise Scale I
WARNING
INTRODUCTION
Thank you for selecting and purchasing the Detecto APS Enterprise Scale. It was built with Detecto quality and reliability at our factory in Webb City, Missouri and incorporates the latest in digital technology and innovative features for the weighing industry.
This manual is provided to guide you through installation, setup/calibration and operation of the Detecto APS Enterprise Scale. Please read it thoroughly before attempting to install the scale and keep it handy for future reference.
FCC COMPLIANCE STATEMENT
This equipment generates uses, can radiate radio frequency, and if not installed and used in accordance with the instruction manual, may cause interference to radio communications. It has been tested and found to comply with the limits for a Class A computing device pursuant to Subpart J of Part 15 of FCC rules, which are designed to provide reasonable protection against such interference when operated in a commercial environment. Operation of this equipment in a residential area may cause interference in which case the user will be responsible to take whatever measures necessary to correct the interference.
You may find the booklet “How to Identify and Resolve Radio TV Interference Problems” prepared by the Federal Communications Commission helpful. It is available from the U.S. Government Printing Office, Washington, D.C. 20402, stock No. 001-000-00315-4.
COPYRIGHT
All rights reserved. Reproduction or use, without expressed written permission, of editorial or pictorial content, in any manner, is prohibited. No patent liability is assumed with respect to the use of the information contained herein.
DISCLAIMER
While every precaution has been taken in the preparation of this manual, the Seller assumes no responsibility for errors or omissions. Neither is any liability assumed for damages resulting from use of the information contained herein. All instructions and diagrams have been checked for accuracy and ease of application; however, success and safety in working with tools depend to a great extent upon the individual accuracy, skill and caution. For this reason, the Seller is not able to guarantee the result of any procedure contained herein. Nor can they assume responsibility for any damage to property or injury to persons occasioned from the procedures. Persons engaging the procedures do so entirely at their own risk.
PROPER DISPOSAL
When this device reaches the end of its useful life, it must be properly disposed of. It must not be disposed of as unsorted municipal waste. Within the European Union, this device should be returned to the distributor from where it was purchased for proper disposal. This is in accordance with EU Directive 2002/96/EC. Within North America, the device should be disposed of in accordance with the local laws regarding the disposal of waste electrical and electronic equipment.
It is everyone’s responsibility to help maintain the environment and to reduce the effects of hazardous substances contained in electrical and electronic equipment on human health. Please do your part by making certain that this device is properly disposed of. The symbol shown to the right indicates that this device must not be disposed of in unsorted municipal waste programs.
8527-0544-2M Rev G APS Enterprise Scale II
SPECIFICATIONS
Model Platform Size Capacity Increments Divisions Load Cell
APS8
APS10
APS12
APS160
APS15
APS20
APS30
APS70
APS150
APS250
Power Requirements: 100 to 240 VAC 50/60Hz 12 VDC 1A wall plug-in UL/CSA listed
6 in x 10 in
15 cm x 25 cm
6 in x 10 in
15 cm x 25 cm
6 in x 10 in
15 cm x 25 cm
10 in x 10 in
25 cm x 25 cm
10 in x 10 in
25 cm x 25 cm
10 in x 10 in
25 cm x 25 cm
12 in x 14 in
31 cm x 36 cm
12 in x 14 in
31 cm x 36 cm
12 in x 14 in
31 cm x 36 cm
18 in x 18 in
46 cm x 46 cm
15 lb 0.01 lb 1500
7.5 kg 0.005 kg 1500 30 lb 0.01 lb 3000
15 kg 0.005 kg 3000
160 oz 0.1 oz 1600 TSP-10kg
160 oz 0.1 oz 1600 TSP-10kg
15 lb 0.01 lb 1500
7.5 kg 0.005 kg 1500 30 lb 0.01 lb 3000
15 kg 0.005 kg 3000
30 lb 0.01 lb 3000
15 kg 0.005 kg 3000
70 lb 0.02 lb 3500
31 kg 0.01 kg 3500
150 lb 0.05 lb 3000
60 kg 0.02 kg 3000
250 lb 0.1 lb 2500
113 kg 0.05 kg 2500
AC power adapter (Cardinal part number 6800-1045)
TSP-10kg
TSP-15kg
TSP-10kg
TSP-15kg
TSP-15kg
TSP-30kg
TSP-100kg
TSP-100kg
Operating Environment: Temperature Range: 14º to 104º F (-10º to +40º C) Display: Five digit, seven segment, 0.875 inch (22 mm) high LCD Interfaces: 2 RS-232 serial, 1 USB-B, 1 OPOS compatible serial, most POS
system protocols included (Cardinal Scale/SMA, Avery Berkel, NCI/Worldship, Mettler PS, Mettler SICS)
Options
APSWIFI – APS Wi-Fi APSRD – Remote Display w/6ft. cable  APSFS – Foot Switch Zero APSPOST – Remote Display Tower
APSPB – Push Button Zero BT1214 / BT1818 – Ball Top Transfer
Certifications
This equipment is certified to comply with the requirements for a Class III device by the:
National Conference on Weights and Measurements (Certificate No. 14-057). Measurement Canada (Approval No. AM-5968
8527-0544-2M Rev G APS Enterprise Scale 1
SITE PREPARATION REQUIREMENTS
The Detecto APS Enterprise Scale is a precision weight-measuring instrument. As with any precision instrument, they require an acceptable environment to operate at peak performance and reliability. This section is provided to assist you in obtaining such an environment.
Site Requirements
In order to keep cooling requirements to a minimum, the indicator should be placed out of direct sunlight and to provide adequate air circulation, keep the area around the indicator clear.
Make certain the indicator is not directly in front of a heating or cooling vent. Such a location will subject the indicator to sudden temperature changes, which may result in unstable weight readings.
Insure that the indicator has good, clean AC power and is properly grounded. In areas subject to lightning strikes, additional protection to minimize lightning damage, such as surge suppressors, should be installed.
Environmental Conditions
The APS Enterprise scale has been designed to operate under the following environmental conditions:
Indoor use only. Ambient temperature range of 14 to 104 °F (-10 to +40 °C Main supply voltage fluctuations not to exceed ±10% of the nominal voltage. Transient overvoltage’s according to Installation Categories II (Overvoltage Categories per
IEC 1010). Temporary voltage spikes on the AC input line that may be as high as 1500V for 115V models and 2500V for 230V models are allowed.
Used in an environment of Pollution degrees 2 (i.e., where normally only non-conductive
atmospheres are present). Occasionally, however, a temporary conductivity caused by condensation must be expected, in accordance with IEC 664.
8527-0544-2M Rev G APS Enterprise Scale 2
SITE PREPARATION REQUIREMENTS, CONT.
Electrical Power
The APS Enterprise scale is powered by a 100 to 240 VAC 50/60Hz 12 VDC 1A wall plug-in UL/CSA listed AC power adapter. Note that a special order is not required for operation at 230/240 VAC.
The socket-outlet supplying power to the scale should be on a separate circuit from the
distribution panel and dedicated to the exclusive use of the scale.
The socket-outlet shall be installed near the equipment and shall be easily accessible.
Note that the wall plug-in AC power adapter serves as the power disconnect.
The wiring should conform to national and local electrical codes and ordinances and should
be approved by the local inspector to assure compliance.
On installations requiring 230 VAC power, it is the responsibility of the customer to have
a qualified electrician install the proper power adapter plug that conforms to national electrical codes and local codes and ordinances.
Electrical Noise Interference
To prevent electrical noise interference, make certain all other wall outlets for use with air conditioning and heating equipment, lighting or other equipment with heavily inductive loads, such as welders, motors and solenoids are on circuits separate from the scale. Many of these disturbances originate within the building itself and can seriously affect the operation of the scale. These sources of disturbances must be identified and steps must be taken to prevent possible adverse effects on the scale. Examples of available alternatives include isolation transformers, power regulators, uninterruptible power supplies, or simple line filters.
Transient Suppression
The following recommendations will help to reduce transients:
Always use shielded cables to connect signal wires to the weight scale. Do not run signal cables from the scale alongside or parallel to wiring carrying AC power. If
unavoidable, position the signal cables a minimum of 24" away from all AC wiring.
Always use arc suppressors across all AC power relay contacts (see recommendations at
http://www.paktron.com/pdf/Quencharch_QRL.pdf).
RFI Immunity
The operation of sensitive electronic equipment can be adversely affected by RF (Radio Frequency) radio transmissions. Digital weight scales are one such type of equipment. Radio transmissions come from things like hand-held radio transmitters and cell phones. One symptom of RFI (Radio Frequency Interference) in a digital weight scale is weight indication instability during a radio transmission.
The APS Enterprise scale was designed with special grounding and RFI shielding to achieve a high degree of immunity to common RFI. To maximize the scale’s immunity to radio transmissions, follow these guidelines:
1. ALWAYS use shielded cable for all I/O (Input/Output) connections to the scale.
2. NEVER operate any radio transmitter within 2 meters (~6ft.) of the scale.
3. NEVER connect un-terminated cables to the scale.
4. KEEP the intended external I/O devices connected to I/O cables entering the scale.
5. ALWAYS connect the shield of the shielded I/O cable at the scale end only. Leave the
shield unconnected at the I/O device.
8527-0544-2M Rev G APS Enterprise Scale 3
INSTALLATION
(ALL MODELS EXCEPT APS250)
Unpacking
Remove the scale from the shipping carton. Remove all packing material and then examine the scale to make certain there is no shipping damage. Should damage be discovered contact the shipping company as soon as possible. The packing material and shipping carton should be retained for possible examination by the shipping company.
Determine where the scale is to be located. It should be a smooth surface capable of supporting both the scale and any load to be placed on the scale. It should not be in direct sunlight nor should it be subject to air flow from heating/cooling vents, fans or similar devices. Plug-in the 12VDC adapter to use scale. Refer to Electrical Power section of this manual for more instruction. The scale is now ready for use.
Level Adjustment
Place the scale in the chosen location and remove the stainless steel platform. The platform can be removed by lifting upward on it removing it from the weighbridge. Locate the small level indicator on the scale base to the right of the display and check to make certain that the scale is level.
NOT LEVEL LEVEL
Bubble is not Bubble is
If the level indicator does not show a level indication, the mounting feet should be adjusted to obtain a level condition. To adjust a mounting foot, first loosen the lock nut (located under the scale base) and then turn the mounting foot in the required direction to raise or lower that corner of the scale. Once a level condition has been achieved, secure the mounting foot settings by tightening each of the lock nuts. Replace the platform on the weighbridge.
within circle within circle
LOCK NUT
MOUNTING FOOT
8527-0544-2M Rev G APS Enterprise Scale 4
INSTALLATION, CONT.
(APS250 ONLY)
Unpacking and Setup
1. Remove the scale display, hardware pack, power supply, scale cover and scale base
assembly from the shipping carton. Remove all packing material and then examine the scale to make certain there is no shipping damage. Should damage be discovered contact the shipping company as soon as possible. The packing material and shipping carton should be retained for possible examination by the shipping company.
2. Remove the four leveling feet from the hardware pack and install in each corner of scale
base as shown. NOTE: Leveling feet must be in place to operate the scale.
3. To mount the display on the scale base, use the two included mounting screws in the hardware pack.
4. Insert the screws through the two holes in the scale display bracket and attach the bracket to the scale base.
5. Route the cable into the scale and plug into scale display port (see interconnection section for port location and more information).
8527-0544-2M Rev G APS Enterprise Scale 5
INSTALLATION, CONT.
(APS250 ONLY)
Unpacking and Setup, Cont.
6. Determine where the scale is to be located. It should be a smooth surface capable of
supporting both the scale and any load to be placed on the scale. It should not be in direct sunlight nor should it be subject to air flow from heating/cooling vents, fans or similar devices.
7. Plug-in 12VDC adapter to use scale. Refer to Electrical Power section of this manual for more instruction.
8. The scale is now ready for use.
Level Adjustment
Place the scale in the chosen location and remove the stainless steel platform. The platform can be removed by lifting upward on it removing it from the weighbridge. Locate the small level indicator on the scale base to the right of the display and check to make certain that the scale is level.
NOT LEVEL LEVEL
Bubble is not Bubble is
If the level indicator does not show a level indication, the mounting feet should be adjusted to obtain a level condition. To adjust a mounting foot, first loosen the lock nut (located under the scale base) and then turn the mounting foot in the required direction to raise or lower that corner of the scale. Once a level condition has been achieved, secure the mounting foot settings by tightening each of the lock nuts. Replace the platform on the weighbridge.
within circle within circle
8527-0544-2M Rev G APS Enterprise Scale 6
LOCK NUT
MOUNTING FOOT
INSTALLATION, CONT.
Interconnections
All input, output and power connections to the APS Enterprise scale are made internally to the main board inside the scale. Connections for the scale display, remote display and OPOS compatible serial device are made using RJ11 snap-in modular connectors. The USB port is a device (or upstream) port using an industry standard Type B connector. The 12VDC 1A wall plug-in UL/CSA listed AC power adapter is connected using a power jack.
POS SERIAL PORT 3
REMOTE
DISPLAY
SERIAL PORT 2
SCALE
DISPLAY
SERIAL PORT 1
12VDC USB
PCB COVER
Data Cable Installation
SCALE DISPLAY/SERIAL PORT 1
Plug the RJ11 connector from the Scale Display cable into the scale base as marked.
REMOTE DISPLAY/SERIAL PORT 2
To use a remote display, plug the remote display RJ11 connector into the scale base where marked.
POS/SERIAL PORT 3
NOTE: The included serial data cable uses an RJ11 phone type connector (scale base) to
a DB9 connector (OPOS compatible serial device).
Plug the RJ11 connector of the included 4’ serial data cable into the scale base where marked and then plug the other end into an available serial port on the POS device.
USB Cable Installation
USB
To use the USB interface, obtain an industry standard Type B connector USB cable and plug it into the scale base where marked and then plug the other end into a USB port on your computer.
Power Adapter
12VDC
To apply power to the scale using the supplied 12VDC 1A wall plug-in UL/CSA listed AC power adapter, insert the plug from the power adapter cable into the power jack on the scale base where marked (refer to previous page) and then plug the power adapter into a proper electrical outlet. The scale is now ready for operation.
8527-0544-2M Rev G APS Enterprise Scale 7
INSTALLATION, CONT.
Serial Data Cable, 8527-0545-0A
(Included 4’ Serial Data Cable, 8527-0545-0A)
4 Feet
(Wiring Diagram, Serial Data Cable, 8527-0545-0A)
8527-0544-2M Rev G APS Enterprise Scale 8
KEYPAD FUNCTIONS
DO NOT operate the keypad with pointed objects (pencils, pens, etc.). Damage to keypad resulting from this practice is NOT covered under warranty.
The keypad of the APS Enterprise Scale has two keys, ZERO (ON/OFF) and UNITS which have the following functions:
This is the ZERO (ON/OFF) key. It serves several purposes. First, when the scale is off, pressing it will turn the scale ON. If the scale is already on, pressing and holding it for approximately 3 seconds will turn the scale OFF.
Pressing this key when the scale is on will zero the scale. This is the UNITS key. This key is used to change the weighing units to the
alternate units of measurement. For example, with pounds displayed, pressing this key will change the weighting units to kilograms.
ANNUNCIATORS
The annunciators are turned on to indicate that the display is in the mode corresponding to the annunciator label or that the status indicated by the label is active.
OZ
This annunciator is turned on to indicate the weight displayed is in ounces.
lb
This annunciator is turned on to indicate that the displayed weight is in pounds.
kg
This annunciator is turned on to indicate the weight displayed is in kilograms.  (STABLE)
This annunciator is turned on when the weight display is stable. When off, it means that the change in successive weight samples is greater than the motion limits selected during setup.
0 (ZERO)
The (0) annunciator is turned on to indicate that the weight is within +/- 1/4 division of the center of zero.
8527-0544-2M Rev G APS Enterprise Scale 9
SETUP AND CONFIGURATION
The keypad of the Enterprise has two keys, ZERO and UNITS which have the following functionality when in setup mode:
ZERO – When a prompt is currently being shown (not the value of the prompt), pressing the
ZERO key will skip to the next prompt. If the value of a prompt is being shown,
pressing the ZERO key will toggle between the acceptable values for the current prompt.
UNITS – When a prompt is currently being shown (not the value of the prompt), pressing the
UNITS key will display the value for that prompt. If the value of a prompt is being
shown, then pressing the UNITS key will save the current value and advance to the next setup prompt.
To Enter Setup
1. Remove the Calibration Switch Access
screw located on the bottom of the scale base.
2. With the scale ON, use a 1/8” or 3 mm Hex Key Wrench or other small tool and press the Calibration Switch.
3. The display will change to show type.
Type (Scale Type)
This will determine the capacity, interval, decimal, and converted units (if applicable). Press the UNITS key to show current setting. If value displayed is acceptable, press the
UNITS key again to save it. Otherwise, press the ZERO key to step through selections for a new value and then press the UNITS key to save it. The following types are selectable for the Enterprise scale:
10
15
30
150
250
160
70
20
12
TEST
= APS10 30 lb x 0.01 lb (15 kg x 0.005 kg) = APS15 15 lb x 0.01 lb (7.5 kg x 0.005 kg) = APS30 30 lb x 0.01 lb (15 kg x 0.005 kg) = APS150 150 lb x 0.05 lb (60 kg x 0.02 kg) = APS250 250 lb x 0.1 lb (125 kg x 0.05 kg) = APS160 160 oz x 0.1 oz = APS70 70 lb x 0.02 lb (31 kg x 0.01 kg) = APS20 30 lb x 0.01 lb (15 kg x 0.005 kg) = APS8 15 lb x 0.01 lb (7.5 kg x 0.005 kg)
8
= APS12 160 oz x 0.1 oz FACTORY TEST SETTINGS
Calibration Switch Access
8527-0544-2M Rev G APS Enterprise Scale 10
SETUP AND CONFIGURATION, CONT.
UnS (Motion Range)
This is the motion range of the scale, or the number of divisions that the scale weight can change without setting the motion indicator.
Press the UNITS key to show current setting. If value displayed is acceptable, press the UNITS key again to save it. Otherwise, press the ZERO key to step through selections for a new value and then press the UNITS key to save it.
Allowable values are 0 through 9.
FILtr (Adjustable Digital Filtering)
This is the adjustable digital weight filtering that is applied to the weight in order to help filter out vibrations or other noise that might make the weight unstable.
Press the UNITS key to show current setting. If value displayed is acceptable, press the UNITS key again to save it. Otherwise, press the ZERO key to step through selections for a new value and then press the UNITS key to save it.
Allowable values are Off, 1, 2 or 3.
= No weight filtering
OFF
= Minimal weight filtering
1
= Moderate weight filtering
2
= Maximum weight filtering
3
CALIb (Calibration)
This is the calibration routine for the scale. With display showing CALib, press the UNITS key. The display will change to show
current setting no. If scale has been previously calibrated and you wish to skip calibration and proceed to PORT1, press the UNITS key again and internal calibration factor will be retained.
To begin calibration, press the ZERO key to select yes and then press the UNITS key. After pressing the UNITS key, the display will change to LOAd.
1. With “LOAd displayed, press the UNITS key. The display will change to show the currently selected scale type’s capacity. NOTE: For convenience, you can press the
ZERO key and change this to half of capacity.
2. Once the test load weight is selected, placed the test load weight on the scale platform and then press the UNITS key.
3. Starting at the right and proceeding left, a series of dashes will appear on the display
and then starting at the left and proceeding right disappear. Next, the display will change to the UnLOd prompt.
4. Remove all weight from the scale platform and then press the UNITS key.
5. Starting at the right and proceeding left, a series of dashes will appear on the display
and then starting at the left and proceeding right disappear. Next, the display will change to the PORT1 prompt. Calibration is complete. Proceed to PORT1.
8527-0544-2M Rev G APS Enterprise Scale 11
SETUP AND CONFIGURATION, CONT.
PORt1 (RS-232 Serial Port 1)
This is the output format for the first serial port. It should default to “SnnA” in order to be used with the Enterprise display.
The serial port parameters for this port are fixed at 9600 baud, 8 data bits, 1 stop bit and No parity.
Press the UNITS key to show current setting. If value displayed is acceptable, press the UNITS key again to save it. Otherwise, press the ZERO key to step through selections for a new value and then press the UNITS key to save it.
Options for PORT1 output format are as follows:
SnnA
nci nntps nntsi
LPL
This selection allows the port to respond to the following commands and will output Weight on Demand in the appropriate format.
Cardinal ENQ (hex 05) Detecto AS-410D ~ (tilde, hex 7E), Pounds ONLY Mode
SMA (W) NCI output format Mettler Toledo PS output format Mettler Toledo SICS output format Lone Peak Labeling
IMPORTANT! Selecting LPL (Lone Peak Labeling) for the port output format will change the behavior of the scale. When the scale settles on “any” positive weight above zero, the display will lock on that weight until the scale returns to zero.
PORt2 (RS-232 Serial Port 2)
This is the second serial port output format selection. It is identical to PORT1. The serial port parameters for this port are fixed at 9600 baud, 8 data bits, 1 stop bit and No parity.
Press the UNITS key to show current setting. If value displayed is acceptable, press the UNITS key again to save it. Otherwise, press the ZERO key to step through selections for a new value and then press the UNITS key to save it.
Options for port2 output format are as follows:
SnnA
nci nntps nntsi
LPL
This selection allows the port to respond to the following commands and will output Weight on Demand in the appropriate format.
Cardinal ENQ (hex 05) Detecto AS-410D ~ (tilde, hex 7E), Pounds ONLY Mode
SMA (W) NCI output format Mettler Toledo PS output format Mettler Toledo SICS output format Lone Peak Labeling
IMPORTANT! Selecting LPL (Lone Peak Labeling) for the port output format will change the behavior of the scale. When the scale settles on “any” positive weight above zero, the display will lock on that weight until the scale returns to zero.
8527-0544-2M Rev G APS Enterprise Scale 12
SETUP AND CONFIGURATION, CONT.
PORt3 (POS Serial Port 3 – OPOS Compatible)
This is the third serial port output format selection. The default format for this port is NCI in order to work with most POS systems that require NCI format.
The serial port parameters for this port are fixed at 9600 baud, 7 data bits, 1 stop bit and Even parity for any output format selected EXCEPT the UPS WorldShip output format.
NOTE: When UPS WorldShip output format is selected for POS Serial Port 3, the serial port parameters are set to 4800 baud, 7 data bits, 1 stop bit and Even parity.
Press the UNITS key to show current setting. If value displayed is acceptable, press the UNITS key again to save it. Otherwise, press the ZERO key to step through selections for a new value and then press the UNITS key to save it.
Options for PORT3 output format are as follows:
SnnA
nci nntps nntsi
LPL
UPS
This selection allows the port to respond to the following commands and will output Weight on Demand in the appropriate format.
Cardinal ENQ (hex 05) Detecto AS-410D ~ (tilde, hex 7E), Pounds ONLY Mode
SMA (W) NCI output format Mettler Toledo PS output format Mettler Toledo SICS output format Lone Peak Labeling UPS WorldShip output format
IMPORTANT! Selecting LPL (Lone Peak Labeling) for the port output format will change the behavior of the scale. When the scale settles on “any” positive weight above zero, the display will lock on that weight until the scale returns to zero.
Usb (USB Port)
The USB port output format selections are similar to the serial port settings. When connected to a PC, this should appear as a virtual com port and can transmit weight using any of the formats listed below.
Press the UNITS key to show current setting. If value displayed is acceptable, press the UNITS key again to save it. Otherwise, press the ZERO key to step through selections for a new value and then press the UNITS key to save it.
Options for USB port output format are as follows:
SnnA
nci nntps nntsi
LPL
This selection allows the port to respond to the following commands and will output Weight on Demand in the appropriate format.
Cardinal ENQ (hex 05) Detecto AS-410D ~ (tilde, hex 7E), Pounds ONLY Mode
SMA (W) NCI output format Mettler Toledo PS output format Mettler Toledo SICS output format Lone Peak Labeling
IMPORTANT! Selecting LPL (Lone Peak Labeling) for the port output format will change the behavior of the scale. When the scale settles on “any” positive weight above zero, the display will lock on that weight until the scale returns to zero.
8527-0544-2M Rev G APS Enterprise Scale 13
SETUP AND CONFIGURATION, CONT.
HirES (High Resolution Weight Mode)
The High Resolution Weight Mode feature will display the weight at 10 times the standard resolution. For example, an APS configured for 250 lb x 0.1 lb will display weight in high resolution mode at 250 lb x 0.01 lb.
Press the UNITS key to show current setting. If value displayed is acceptable, press the UNITS key again to save it. Otherwise, press the ZERO key to step through selections for a new value and then press the UNITS key to save it.
Allowable values are YES or no.
YES = If High Resolution Weight
mode is desired
NOTE: If YES is selected, the scale will exit Setup and Calibration, reset and then display weight in the High Resolution Weight mode.
IMPORTANT! When the scale is in the High Resolution Weight mode, no Units will be displayed and the scale is not Legal For Trade.
To exit the High Resolution Mode, press and hold the ZERO (ON/OFF) key to turn the scale off and then press the ZERO (ON/OFF) to the scale back on
btdis (Button Disable)
This setting allows the keypad buttons to be disabled for use by the operator “during normal operation”. During normal operation, when the operator presses a disabled button, the
button will be ignored.
no = High Resolution Weight mode is
not desired
NOTE: The Button Disable setting only disables the buttons for use during normal operation, not setup and calibration.
Press the UNITS key to show current setting. If value displayed is acceptable, press the UNITS key again to save it. Otherwise, press the ZERO key to step through selections for
a new value and then press the UNITS key to save it. Allowable values are 0, 1, 2 or 3.
= No buttons disabled
0
= Zero button disabled
1
= Tare button disabled
2
= Both buttons disabled
3
NOTE: Disabling the buttons also disables the SMA zero and SMA toggle units commands so this setting may affect Serial/USB port operation if using the SMA commands.
8527-0544-2M Rev G APS Enterprise Scale 14
ERROR AND STATUS DISPLAYS
Display Meaning
-Err-
----
Displayed when the ZERO key is pressed and the scale could not zero. Displayed when scale weight exceeds scale capacity, or if there is an
analog high/low error.
CARE AND CLEANING
NOTE: The APS Enterprise scale contains no user-serviceable parts and maintenance should be limited to an occasional cleaning as required.
Do not submerge the scale in water, pour or spray water directly on it to clean. The
scale is not waterproof and covering it with water will damage it and void the warranty.
Always remove power before cleaning. Do not use wire brushes, abrasives, or cleaning tools such as steel pads and
scrapers, which will scratch the surface. Instead, use soft cloths or plastic scouring pads for cleaning.
When possible use treated water. Hard water can leave behind deposits. Soft water
is much gentler on the stainless steel and painted surfaces.
Avoid the use of acetone, thinner or other volatile solvents and abrasive type cleaners
for cleaning. If required, a mild solvent such as mineral spirits can be used to remove oil, grease, tars, wax, and similar substances. Use a cloth dampened with mineral spirits and only apply to areas that are contaminated. Follow up the use of this mild solvent with detergent cleaning and rinsing.
8527-0544-2M Rev G APS Enterprise Scale 15
SECURITY SEAL INSTALLATION
If your metrology laws require the use of physical sealing, a lead and wire security seal can be installed on the APS Enterprise scale to prevent the Calibration Switch Access screw located on the bottom of the scale base from being removed to gain access to the calibration switch thereby preventing unauthorized access to the calibration adjustments.
NOTE: In order to install the lead and wire security seal, two screws must be replaced with two drilled screws. They are available from the Cardinal/Detecto Parts Department. The part numbers are:
6021-0605 (SCW FILLISTER. MACHINE-SCW 06-32X.375) 6024-0126 (WASHER FLAT) 6021-1108 (SCW FILLISTER. MACHINE-SCW 10-32X.375)
To install a lead and wire security seal, follow the steps provided below:
1. Remove the Calibration Switch Access screw and replace it with the drilled 10-32 screw. Make certain it is securely tighten before proceeding.
2. Remove the PCB Cover screw as shown and replace it with the drilled 6-32 screw and flat washer. Make certain it is securely tighten before proceeding.
3. Thread the sealing wire through the hole in the drilled screws.
4. Pull the wire tight and install the lead seal.
5. The screws cannot be removed without
damaging the seal.
PCB Cover Screw Calibration Switch
Access Screw
6-32 Screw and Flat Washer
10-32 Screw
8527-0544-2M Rev G APS Enterprise Scale 16
APSRD (REMOTE DISPLAY)
The APSRD (Remote Display) can be mounted using three (3) methods.
The bracket can be attached to the wall using
the appropriate hardware for the type of wall construction.
The remote display can be removed from the
bracket and mounted on the wall using the included Velcro
Using the remote display post, it can be
mounted to a table, bench or in a retail environment, the check stand.
®
strips.
8527-0544-2M Rev G APS Enterprise Scale 17
PARTS IDENTIFICATION
(ALL MODEL EXCEPT APS250)
8527-0544-2M Rev G APS Enterprise Scale 18
PARTS IDENTIFICATION, CONT.
(ALL MODEL EXCEPT APS250)
ITEM PART NO. QTY DESCRIPTION
1 1960-B465-08 2 SPACER FOR EB-XXX-XXXX 3 6013-0077 4 HEX NUT
10 6540-1165 4 LEVELLING FOOT, 1.63 BASE X 3/8”-16 X 1.3 STUD,
BLK PU 11 6560-1072 A/R ADHESIVE, LOCTITE 49 14 6690-0001 1 BULLSEYE BUBBLE LEVEL 15 8527-0519-08 1 BASE PLATE, 30 16 8527-0520-0A 1 LOAD CELL ASSY, 10KG (APS160, APS15, APS8, APS12) 17 8527-0521-0A 1 LOAD CELL ASSY, 15KG (APS10, APS20) 18 8527-0521-1A 1 LOAD CELL ASSY, 15KG (APS30) 19 8527-0522-0A 1 LOAD CELL ASSY, 30KG (APS70) 20 8527-0523-08 1 BASE PLATE, 150 21 8527-0524-0A 1 LOAD CELL ASSY, 100KG (APS150) 22 8527-0527-08 1 BASE PLATE, 70 23 8527-0528-08 1 COVER, 12X14 24 8527-0529-08 1 TOP PLATE, 8/10/12 25 8527-0530-08 1 TOP PLATE, 160/15/20 26 8527-0531-08 1 TOP PLATE, 30 27 8527-0532-08 1 TOP PLATE, 70 28 8527-0533-08 1 TOP PLATE, 150 30 8527-C503-08 1 BASE PLATE, 8/10/12 31 8527-C504-08 1 COVER, 6X10 32 8527-C505-08 1 PCB COVER 33 8527-C513-08 1 BASE PLATE, 160/15/20 34 8527-C514-08 1 COVER, 10X10
A/R = AS REQUIRED
8527-0544-2M Rev G APS Enterprise Scale 19
PARTS IDENTIFICATION, CONT.
(ALL MODEL EXCEPT APS250)
ITEM PART NO. QTY DESCRIPTION
2 593GR986 1 SERIAL TAG ASSY 5 6021-0625 3 SCW FLAT-HEAD. MACHINE-SCW 06-32X.375 6 6021-0659 4 SCW FLAT-HEAD. MACHINE-SCW 06-32X1.00 7 6021-1069 2 SCW CUP-POINT. SET-SCREW.. 10-32X.25 NYLON PATCH 8 6021-1108 2 SCW FILLISTER. MACHINE-SCW 10-32X.375
9 6021-1706 4 SCW FLAT-HEAD. CAP-SCREW.. .25-20X.625 12 6650-0087 1 LABEL, MADE IN USA (WITH FLAG) 1” X 1” 35 8527-D510-0A 1 DISPLAY SUB-ASSY, 10 36 8527-D510-1A 1 DISPLAY SUB-ASSY, 160 37 8527-D510-2A 1 DISPLAY SUB-ASSY, 15 38 8527-D510-3A 1 DISPLAY SUB-ASSY, 30 40 8527-D510-5A 1 DISPLAY SUB-ASSY, 70 41 8527-D510-6A 1 DISPLAY SUB-ASSY, 150 42 8527-D510-9A 1 DISPLAY SUB-ASSY, 8 43 8527-D510-AA 1 DISPLAY SUB-ASSY, 12 44 8527-D510-BA 1 DISPLAY SUB-ASSY, 12 45 593R1007 2 FACTORY SET LABEL
8527-0544-2M Rev G APS Enterprise Scale 20
PARTS IDENTIFICATION, CONT.
(ALL MODEL EXCEPT APS250)
ITEM PART NO. QTY DESCRIPTION
4 6021-0623 4 SCW PAN-HEAD.. MACHINE-SCW 06-32X.750 7 6021-1069 2 SCW CUP-POINT. SET-SCREW.. 10-32X.25 NYLON PATCH
9 6021-1706 4 SCW FLAT-HEAD. CAP-SCREW.. .25-20X.625 13 6680-0004 4 WASHER LOCK INT TOOTH #6 TYPE A Z-PL 29 8527-C500-0A 1 ANALOG PCB 45 593R1007 2 FACTORY SET LABEL
8527-0544-2M Rev G APS Enterprise Scale 21
PARTS IDENTIFICATION, CONT.
(APS250 ONLY)
8527-0544-2M Rev G APS Enterprise Scale 22
PARTS IDENTIFICATION, CONT.
(APS250 ONLY)
ITEM PART NO. QTY DESCRIPTION
1 1960-B465-08 2 SPACER FOR EB-XXX-XXXX
3 6013-0077 4 HEX NUT
4 6013-0295 4 NUT #10-32 HEX Z/P
8 6021-0950 4 SCW HEX-HEAD.. MACHINE-SCW 10-32X1.50 Z/P 10 6021-1108 2 SCW FILLISTER. MACHINE-SCW 10-32X.375 12 6540-1123 4 LEVELLER SCREW, 3/8-16 X 1 1/4”, BLACK 13 6560-1072 A/R ADHESIVE, LOCTITE 49 16 6690-0001 1 BULLSEYE BUBBLE LEVEL 17 8527-0524-1A 1 LOAD CELL ASSY, 100KG 18 8527-0525-08 1 BASE PLATE, 250 19 8527-0526-08 1 COVER, 18X18 20 8527-0534-08 1 TOP PLATE, 250 22 8527-C505-08 1 PCB COVER
A/R = AS REQUIRED
8527-0544-2M Rev G APS Enterprise Scale 23
PARTS IDENTIFICATION, CONT.
(APS250 ONLY)
ITEM PART NO. QTY DESCRIPTION
2 593GR986 1 SERIAL TAG ASSY 6 6021-0625 3 SCW FLAT-HEAD. MACHINE-SCW 06-32X.375 7 6021-0659 4 SCW FLAT-HEAD. MACHINE-SCW 06-32X1.00
9 6021-1069 1 SCW CUP-POINT. SET-SCREW.. 10-32X.25 NYLON PATCH 10 6021-1108 2 SCW FILLISTER. MACHINE-SCW 10-32X.375 11 6021-1706 4 SCW FLAT-HEAD. CAP-SCREW.. .25-20X.625 14 6650-0087 1 LABEL, MADE IN USA (WITH FLAG) 1” X 1” 23 8527-D510-7A 1 DISPLAY SUB-ASSY, 250
8527-0544-2M Rev G APS Enterprise Scale 24
PARTS IDENTIFICATION, CONT.
(APS250 ONLY)
ITEM PART NO. QTY DESCRIPTION
5 6013-0315 4 NUT #6-32 HEX SMALL PATTERN Z/P 11 6021-1706 4 SCW FLAT-HEAD. CAP-SCREW.. .25-20X.625 15 6680-0004 4 WASHER LOCK INT TOOTH #6 TYPE A Z-PL 21 8527-C500-0A 1 ANALOG PCB
8527-0544-2M Rev G APS Enterprise Scale 25
PARTS IDENTIFICATION, CONT.
SCALE DISPLAY and APSRD (REMOTE DISPLAY)
ITEM PART NO. QTY DESCRIPTION
1 5930-0146-08 1 CAPACITY LABEL 2 6021-2078 4 SCW PAN-HEAD PHILLIPS #1-32X3/8” THD FORM, SS 3 8527-0515-08 1 APS KEYPAD 4 8527-C502-0A 1 PCB ASSY APS10 DISPLAY 5 8527-C502-1A 1 PCB ASSY APS15/160 DISPLAY 6 8527-C502-2A 1 PCB ASSY APS30/70/150 DISPLAY 7 8527-C502-3A 1 PCB ASSY APS250 DISPLAY 8 8527-C502-4A 1 PCB ASSY APSRD DISPLAY
9 8527-D506-08 1 ENCLOSURE FRONT 10 8527-D507-08 1 ENCLOSURE BACK 11 8527-D508-08 1 ENCLOSURE MOUNT
DISPLAY EXTENSION CABLE
ITEM PART NO. QTY DESCRIPTION
6600-2007 1 CABLE 7FT APS DISPLAY EXTENSION
Not Shown
8527-0544-2M Rev G APS Enterprise Scale 26
STATEMENT OF LIMITED WARRANTY
Detecto Scale warrants its equipment to be free from defects in material and workmanship as follows: Detecto warrants to the original purchaser only that it will repair or replace any part of equipment which is defective in material or workmanship for a period of two (2) years from date of shipment. Detecto shall be the sole judge of what constitutes a defect.
During the first ninety (90) days Detecto may choose to replace the product at no charge to the buyer upon inspection of the returned item.
After the first ninety (90) days, upon inspection of the returned item, Detecto will repair or replace it with a remanufactured product. The customer is responsible for paying for the freight both ways.
This warranty does not apply to peripheral equipment not manufactured by Detecto; this equipment will be covered by certain manufacturer’s warranty only.
This warranty does not include replacement of expendable or consumable parts. This does not apply to any item which has deteriorated or damaged due to wear, accident, misuse, abuse, improper line voltage, overloading, theft, lightning, fire, water or acts of God, or due to extended storage or exposure while in purchaser’s possession. This warranty does not apply to maintenance service. Purchased parts will have a ninety (90) day repair or replacement warranty only.
Detecto may require the product to be returned to the factory; item(s) must be properly packed and shipping charges prepaid. A return authorization number must be obtained for all returns and marked on the outside of all returned packages. Detecto accepts no responsibility for items lost or damaged in transit.
Conditions Which Void Limited Warranty
This warranty shall not apply to equipment which: A.) Has been tampered with, defaced, mishandled or has had repairs and modifications not authorized by Detecto. B.) Has had serial number altered, defaced, or removed. C.) Has not been properly grounded according to Detecto’s recommended procedure.
Freight Carrier Damage
Claims for equipment damaged in transit must be referred to the freight carrier in accordance with freight carrier regulations.
This warranty sets forth the extent of our liability for breach of any warranty or deficiency in connection with the sale or use of the product. Detecto will not be liable for consequential damages of any nature, including but not limited to, loss of profit, delays or expenses, whether based on tort or contract. Detecto reserves the right to incorporate improvements in material and design without notice and is not obligated to incorporate improvements in equipment previously manufactured.
The foregoing is in lieu of all other warranties, express or implied including any warranty that extends beyond the description of the product including any warranty of merchantability or fitness for a particular purpose. This warranty covers only those Detecto products installed in the forty-eight (48) contiguous continental United States.
Ph. (800) 641-2008
E-mail: detecto@cardet.com
203 E. Daugherty
Webb City, MO 64870
08/27/2014
Printed in USA
D268-WARRANTY-DET-A
8527-0544-2M Rev G APS Enterprise Scale 28