California Instruments 1001P Users and Programmers Guide

Revision H
May 2000 Copyright 1996,1997,1998,1999,2000
By California Instruments. All rights reserved. P/N 5002-966

801P / 1001P / 1251P Series

AC Power Source
User and Programming Manual
User and Programming Manual - Rev H California Instruments
User's Manual AC Power Source California Instruments Models:
801P 1001P 1251P
Copyright 1996,1997,1998,1999,2000 California Instruments, Rev H, May 2000
P Series May 2000 i
User and Programming Manual - Rev H California Instruments

SAFETY SUMMARY

This power source contains high voltage and current circuits which are potentially lethal. Because of its size and weight, mechanical stability must be ensured. The following safety guidelines must be followed when operating or servicing this equipment. These guidelines are not a substitute for vigilance and common sense. California Instruments assumes no liability for the customer's failure to comply with these requirements.
APPLYING POWER AND GROUNDING
Verify the correct voltage is applied to the unit (100 to 240 VAC Nominal). Verify that the input
power cord is plugged into a properly grounded utility outlet.
FUSES
Use only fuses of the specified current, voltage, and protection speed. Do not short out the fuse holder or use a repaired fuse. The 801P/1001P units use a North American ferrule type fuse rated at 15A and 250Volts. (Fast Acting) The 1251P unit uses a North American ferrule type fuse rated at 20A and 250Volts. (Fast Acting)
DO NOT OPERATE IN A VOLATILE ATMOSPHERE
Do not operate the power source in the presence of flammable gases or fumes.
DO NOT TOUCH ENERGIZED CIRCUITS
Disconnect the power cable before servicing this equipment. Even with the power cable disconnected, high voltage can still exist on some circuits. Discharge these voltages before servicing. Only qualified service personnel may remove covers, replace components or make adjustments.
DO NOT SERVICE ALONE
Do not remove covers, replace components, or make adjustments unless another person, who can administer first aid, is present.
DO NOT EXCEED INPUT RATINGS
Do not exceed the rated input voltage or frequency. Additional hazards may be introduced because of component failure or improper operation.
DO NOT MODIFY INSTRUMENT OR SUBSTITUTE PARTS
Do not modify this instrument or substitute parts. Additional hazards may be introduced because of component failure or improper operation.
MOVING THE POWER SOURCE
When moving the power source, observe the following:
1. Remove all A C power to unit.
2. Use two people to prevent injury.
SURFACE STABILITY
ii May 2000 P Series
User and Programming Manual - Rev H California Instruments
1. Operate the power source only on a level surface.
ONE YEAR WARRANTY
CALIFORNIA INSTRUMENTS CORPORATION warrants each instrument manufactured by them to be free from defects in material and workmanship for a period of one year from the date of shipment to the original purchaser. Excepted from this warranty are fuses and batteries which carry the warranty of their original manufacturer where applicable. CALIFORNIA INSTRUMEN TS will service, replace, or adjust any defective part or parts, free of charge, when the instrument is returned freight prepaid, and when examination reveals that the fault has not occurred because of misuse, abnormal conditions of operation, user modification, or attempted user repair. Equipment repaired beyond the effective date of warranty or when abnormal usage has occurred will be charged at applicable rates. CALIFORNIA INSTRUMENTS will submit an estimate for such charges before commencing repair, if so requested.
PROCEDURE FOR SERVICE If a fault develops, notify CALIFORNIA INSTRUMENTS or its local representative, giving full
details of the difficulty, including the model number and serial number. On receipt of this information, service information or a Return Material Authorization (RMA) number will be given. Add RMA number to shipping label. Pack instrument carefully to prevent transportation damage, affix label to shipping container, and ship freight prepaid to the factory. CALIFORNIA INSTRUMEN TS shall not be responsible for repair of damage due to improper handling or packing. Instruments returned without RMA No. or freight collect will be refused. Instruments repaired under Warranty will be returned by prepaid surface freight. Instruments repaired outside the Warranty period will be returned freight collect, F.O.B. CALIFORNIA INSTRUMENTS, 9689 Towne Centre Drive, San Diego, CA 92121-1964. If requested, an estimate of repair charges will be made before work begins on repairs not covered by the Warranty.
DAMAGE IN TRANSIT
The instrument should be tested when it is received. If it fails to operate properly, or is damaged in any way, a claim should be filed immediately with the carrier. A full report of the damage should be obtained by the claim agent, and a copy of this report should be forwarded to us. CALIFORNIA INSTRUMENTS will prepare an estimate of repair cost and repair the instrument when authorized by the claim agent. Please include model number and serial number when referring to the instrument.
P Series May 2000 iii
User and Programming Manual - Rev H California Instruments

Table of Contents

1. INTRODUCTION ............................................................................................................. 1
1.1 General Description .........................................................................................................1
2. SPECIFICATIONS............................................................................................................ 3
2.1 Electrical ........................................................................................................................3
2.2 Mechanical.....................................................................................................................5
2.3 Environmental.................................................................................................................6
2.4 Regulatory......................................................................................................................6
2.5 Front Panel Controls........................................................................................................6
3. UNPACKING AND INSTALLATION.............................................................................. 10
3.1 Unpacking.................................................................................................................... 10
3.2 Power Requirements ..................................................................................................... 10
3.3 Mechanical Installation .................................................................................................. 10
3.4 Input Wiring.................................................................................................................. 10
3.5 Output Connections .......................................................................................................11
3.6 Output Voltage Ranges ..................................................................................................12
3.7 Functional Test............................................................................................................. 12
4. FRONT PANEL OPERATION....................................................................................... 14
4.1 Front Panel Guided Tour................................................................................................ 14
4.2 How to. .. ......................................................................................................................17
5. PRINCIPLE OF OPERATION ....................................................................................... 20
5.1 General........................................................................................................................ 20
5.2 Overall Description ........................................................................................................ 20
5.3 Power Factor Correction Module (PFC) ........................................................................... 21
5.4 DC - DC Converter Module ............................................................................................. 21
5.5 Oscillator Control Board .................................................................................................22
5.6 DC to AC Power Module................................................................................................ 22
5.7 RS232 Board................................................................................................................ 23
6. CALIBRATION................................................................................................................ 26
6.1 Calibration Equipment ....................................................................................................26
6.2 Routine Calibration........................................................................................................ 26
6.3 Non-Routine Calibration ................................................................................................. 29
7. SERVICE......................................................................................................................... 32
7.1 General........................................................................................................................ 32
7.2 Basic Operation............................................................................................................ 32
7.3 Advanced Troubleshooting.............................................................................................. 34
8. INTRODUCTION TO PGUI/PGUI32 ............................................................................ 38
8.1 About This Program....................................................................................................... 38
8.2 About This Section of the Manual ................................................................................... 38
8.3 Program Requirements.................................................................................................. 38
8.4 RS232C Cable Wiring ....................................................................................................39
9. PGUI/PGUI32 SETUP AND INSTALLATION............................................................... 40
9.1 Connecting the AC Source to the PC ..............................................................................40
9.2 Installing the PGUI Software ...........................................................................................40
9.3 Trouble Shooting - RS232C............................................................................................ 41
9.4 Installed Files ............................................................................................................... 46
9.5 Software Registration..................................................................................................... 47
10. TOP ASSEMBLY REPLACEABLE PARTS.............................................................. 51
11. PROGRAMMING INFORMATION............................................................................. 53
12. INTRODUCTION TO SCPI........................................................................................ 55
12.1 Conventions Used in This Manual ................................................................................... 55
iv May 2000 P Series
User and Programming Manual - Rev H California Instruments
12.2 The SCPI Commands and Messages ..............................................................................55
12.3 Using Queries ...............................................................................................................57
12.4 Structure of a SCPI Message .........................................................................................58
12.5 SCPI Data Formats .......................................................................................................60
13. SYSTEM CONSIDERATIONS....................................................................................63
13.1 RS232C Interface..........................................................................................................63
14. SCPI COMMAND REFERENCE ................................................................................66
14.1 Introduction ..................................................................................................................66
14.2 Subsystem Commands .................................................................................................66
14.3 System Commands ......................................................................................................71
14.4 Common Commands .....................................................................................................73
15. PROGRAMMING EXAMPLES....................................................................................78
15.1 Introduction ..................................................................................................................78
15.2 Programming the Output ................................................................................................78
15.3 Making Measurements..................................................................................................79
16. STATUS REGISTERS ................................................................................................80
16.1 Power-On Conditions.....................................................................................................80
16.2 Standard Event Status Group .........................................................................................80
16.3 Status Byte Register .....................................................................................................81
16.4 Examples.....................................................................................................................81
INDEX...........................................................................................................................86
P Series May 2000 v
User and Programming Manual - Rev H California Instruments

List of Figures

FIGURE 3-1: THE 801P/1001P/1251P AC POWER SOURCE.................................................................. 10
FIGURE 3-2: 1251RP REAR PANEL VIEW........................................................................................... 11
FIGURE 3-3: FUNCTIONAL TEST SETUP............................................................................................... 13
FIGURE 4-1: FRONT PANEL VIEW...................................................................................................... 14
FIGURE 4-2: SHUTTLE KNOB............................................................................................................ 16
FIGURE 5-1: AC SOURCE BLOCK DIAGRAM......................................................................................... 20
FIGURE 5-2: OSCILLATOR AND CONTROLS.......................................................................................... 22
FIGURE 6-1: TEST EQUIPMENT HOOKUP FOR ROUTINE OUTPUT CALIBRATION .............................................. 27
FIGURE 6-2: TEST EQUIPMENT HOOK-UP FOR MEASUREMENT CALIBRATION ............................................ 28
FIGURE 6-3: LOCATION OF INTERNAL ADJUSTMENTS.............................................................................. 30
FIGURE 8-1: RS232C CABLE WIRING ............................................................................................... 39
FIGURE 9-1: SYSTEM PROPERTIES DIALOG BOX .................................................................................. 43
FIGURE 9-2: ADVANCED PORT SETTINGS DIALOG BOX.......................................................................... 44
FIGURE 9-3: COM PORT PROPERTIES DIALOG BOX ............................................................................. 44
FIGURE 12-1: PARTIAL COMMAND TREE............................................................................................. 56
FIGURE 12-2: COMMAND MESSAGE STRUCTURE.................................................................................. 58
FIGURE 13-1: RS232C INTERFACE CABLE WIRING DIAGRAM................................................................... 65
FIGURE 16-1: AC SOURCE STATUS SYSTEM MODEL ............................................................................ 80

List of Tables

TABLE 1: LOGIC BOARD LED'S......................................................................................................... 21
TABLE 2: LOAD AND CURRENT......................................................................................................... 26
TABLE 3: BASIC SYMPTOMS............................................................................................................ 32
TABLE 4: POOR OUTPUT VOLTAGE REGULATION................................................................................... 32
TABLE 5: OVERLOAD LIGHT ON........................................................................................................ 32
TABLE 6: DISTORTED OUTPUT.......................................................................................................... 33
TABLE 7: UNIT SHUTS DOWN AFTER 3-5 SECONDS ................................................................................ 33
TABLE 8: NO OUTPUT AND NO LIGHTS ON FRONT PANEL ......................................................................... 33
TABLE 9: NO OUTPUT BUT "POWER ON" LED IS LIT................................................................................. 33
TABLE 10: REPLACEABLE PARTS..................................................................................................... 51
TABLE 11: BIT CONFIGURATION OF STANDARD EVENT STATUS REGISTER............................................... 75
TABLE 12: *RST DEFAULT PARAMETER VALUES .................................................................................. 76
TABLE 13: STATUS REGISTER POWER ON CONDITION ............................................................................. 76
TABLE A-14: ERROR MESSAGES ..................................................................................................... 84
vi May 2000 P Series
User and Programming Manual - Rev H California Instruments

1. Introduction

This instruction manual contains information on the installation, operation, calibration and maintenance of the P Series power sources with programmable controller.

1.1 General Description

The 801P/1001P/1251P AC source is a high efficiency, light weight 800VA, 1000VA or 1250VA programmable AC power source. The output has two voltage ranges of 0-135V or 0­270V with a frequency range of 16 Hz to 500 Hz. The maximum output current is 9.2 amps at 135 volts range and 4.6 amps at 270 volts range for the 1250VA model. The maximum output current is 7.4 amps at 135 volts and 3.7 amps at 270 volts range for the 1000VA model. The maximum output current is 6. 0 amps at 135 volts and 3.0 amps at 270 volts range for the 800VA model.
The universal nominal input can be from 100 volts to 240 volts at 50 Hz or 60 Hz line frequency. Simple front panel controls enable the voltage and frequency to be changed. An optional
RS232C serial interface is available for applications that require remote control and measurements.
P Series May 2000 1
User and Programming Manual - Rev H California Instruments
page intentionally left blank
2
May 2000 P Series
User and Programming Manual - Rev H California Instruments

2. Specifications

All specifications at 25± 1°C unless noted otherwise.

2.1 Electrical

2.1.1 Input

Line Voltage: 85V to 264V maximum 100V to 240V nominal Line Current: 25 A RMS max. (1251RP) Line Frequency: 47-63 Hertz Efficiency: 80% (typical) depending on line and load Power Factor: 0.95 or greater typical Inrush Current : 70 A peak max. at 260V input Hold-Up Time: 20 ms (with no effect on output) Isolation Voltage: Input to output = 2200 VAC, input to chassis = 1350 VAC

2.1.2 Output

Voltage Range: Low Range is 0 to 135 V rms
High Range is 0 to 270 V rms
Voltage Resolution: 0.1 volt Voltage Accuracy: ±1% or range, 50 to 60 Hz ±2% at 400 Hz Line & Load Regulation: 2%of FS on low range, 1% of FS on Hi range Voltage Distortion: 1% max. THD at 50/60 Hz Total Power: 800VA maximum at full scale voltage either range (model 801P)
1000VA maximum at full scale voltage either range (model 1001P)
1250 VA maximum at full scale voltage either range(model 1251P)
P Series May 2000
3
User and Programming Manual - Rev H California Instruments
Current: 6.0 A rms, 17.8 A peak (low range, model 801P)
3.0A rms, 8.9 A peak (high range, model 801P)
7.4 A rms, 22.2 A peak (low range, model 1001P)
3.7 A rms, 11.1 A peak (high range, model 1001P)
9.2 A rms, 27.6 A peak (low range, model 1251P)
4.6 A rms, 13.8 A peak (high range, model 1251P) Current Limit: 801P 1001P 1251P 135V range: 0.0 to 6.0 7.4 9.2 270V range: 0.0 to 3.0 3.7 4.6 Accuracy: Programmed value +5% of maximum current Frequency Range: Range Resolution
16.0 - 99.9 Hz 0.1 Hz 100 - 500 Hz 1 Hz Frequency Accuracy: ±0.02% of program value DC Offset Voltage: Less than ±25 mV with linear load. Output Noise: <0.2 volts RMS on 135 range, <0.5 volts RMS on 270 range.

2.1.3 Measurements

Current (TRMS) Resolution: 0.1 amp Accuracy: ± 0.2 amp Voltage Accessible only through RS232 0 - 250V 251 to 270V Resolution 1 Volts 2 volts Accuracy: ±4 Volts ±6 volts
4
May 2000 P Series
User and Programming Manual - Rev H California Instruments

2.1.4 System Specification

Non volatile Memory Storage: 8 complete instrument setups [ Accessible through RS232C interface
only ]. RS232C Interface: Bi-directional serial interface [ optional ] 9 pin D-shell connector Handshake: CTS, RTS Data bits: 8 Stopbits: 1 Baud rate: 9600 IEEE 488.2 commands and SCPI

2.1.5 Unit Protection

Input Overcurrent : Electronic current limit with fuse. Input Overvoltage
Transients: Surge protection to withstand EN50082-1 (IEC 801-4, 5) levels. Output Overcurrent : Shutdown after 0.1 second after overcurrent. Output Overvoltage: Shutdown, recycle input power to reset. Output Short Circuit: Peak current limit. Shutdown after 0.1 seconds. Overtemperature: Automatic shutdown.

2.2 Mechanical

Dimensions: 8.5” (216mm) width x 8.25” (210mm) height x 17.5” (445mm) depth
Unit Weight : 30 lbs/13.6 kgms.
chassis size stand-alone configuration.
Material: Aluminum chassis, panels and cover. Finish: Powder coat T241-GR11 platinum Cooling: Fan cooled with air intake on the rear and exhaust to the sides. Internal Construction: Modular sub assemblies.
P Series May 2000
5
User and Programming Manual - Rev H California Instruments
Front panel mounted output connections: CEE 7/7 European socket US domestic Dual 120V line socket

2.3 Environmental

Operating Temp: 0 degrees to +40 degrees Celsius. Storage Temp: 0 degrees to +70 degrees Celsius . Humidity: Operating: 90% RH up to 40° C.
Storage: 90% RH up to 40° C, 75% RH up to 70° C. Creepage and
Clearance: Rated for Pollution Degree 2. Insulation: Rated to Installation Category(Overvoltage Category) II Vibration: Designed to meet NSTA 1A transportation levels. Shock: Designed to meet NSTA 1A transportation levels.

2.4 Regulatory

Electromagnetic Designed to meet EN50081-2 and EN50082-2 European Emissions Emissions and and Immunity standards as required for the “CE” mark. Immunity:
Acoustic Noise: 65 dBA maximum at 0% to 50% load, 75 dBA maximum greater than
Safety: Designed to meet UL3111 and EN61010-1 European safety standards

2.5 Front Panel Controls

Controls: Shuttle knobs:
50% load to 100% load. Measured at one meter.
as required for the “CE” mark.
Allows continuous change of Voltage, Frequency and Current limit. Function keys:
Controls Output state, Voltage range and Display mode.
6
May 2000 P Series
User and Programming Manual - Rev H California Instruments
P Series May 2000
7
User and Programming Manual - Rev H California Instruments
Displays: Displays:
Two, 4 digits, 0.5” LCD display. For viewing programmed
voltage, frequency and current limit. For displaying measured
current. Status indicators: 6 LEDs to indicate: REMOTE, FAULT, OUTPUT (ON/OFF), VOLTAGE RANGE,
FREQUENCY or CURRENT DISPLAY MODE (Refer to paragraph
4.1.2).
8
May 2000 P Series
User and Programming Manual - Rev H California Instruments
page intentionally left blank
P Series May 2000
9
User and Programming Manual - Rev H California Instruments

3. Unpacking and Installation

3.1 Unpacking

Inspect the unit for any possible shipping damage immediately upon receipt. If damage is evident, notify the carrier. DO NOT return an instrument to the factory without prior approval. Do not destroy the packing container until the unit has been inspected for damage in shipment.

3.2 Power Requirements

The AC Power System has been designed to operate from a single phase AC line voltage. The nominal operating voltage is from 100V to 240V line input.
Figure 3-1: The 801P/1001P/1251P AC Power Source
WARNING: Do not connect the unit to a 400-480 service as the result will be a
severely damaged unit.

3.3 Mechanical Installation

The AC Source is a completely self contained power unit. It can be used free standing on a bench. The unit is fan cooled, drawing air in from the rear and exhausting at the sides. The sides of the unit must be kept clear of obstruction and a 6 inch clearance must be maintained to the rear.

3.4 Input Wiring

The AC Source is designed to work from a single utility supply. The IEC 320 input connector will accept a standard IEC line cord with the appropriate mating connector for the utility outlet. The utility outlet must be properly grounded and be capable of supplying 1725 VA at nominal voltage for the 1251P and 1425 VA for the 1001P and 1140 VA for the 801P.
10
May 2000 P Series
User and Programming Manual - Rev H California Instruments
Figure 3-2: 1251RP Rear Panel View

3.5 Output Connections

3.5.1 Output Wiring

When the low voltage range is selected , only the duplex US NEMA 5-15P output socket will be active. If the high voltage range is selected, only the European CEE7/7 socket will be active.
Note: Do not connect these outputs together as this will cause the unit to fault.
P Series May 2000
11
User and Programming Manual - Rev H California Instruments

3.6 Output Voltage Ranges

The AC power source has two standard output voltage ranges 0-135V and 0-270V. The operator may switch from one range to the other at will with no special precautions except to remember that the output voltage will go to zero voltage whenever a range change takes place.
Note: The output changes to the other socket on front panel with a range change.

3.7 Functional Test

CAUTION: Work carefully when performing these test, hazardous voltages are
present on the input and output during this test..
Refer to Figure 3-3 for the test set up.
1. Connect an oscilloscope, voltmeter and/or distortion analyzer to the AC source output at
the 135 Volt output terminal.
2. Connect the AC power input voltage connections to the AC source input terminals. Turn
on the power switch located at the rear panel.
3. Verify that the front panel LCD display reads out the initial start up voltage and frequency.
4. Select the low voltage range. Set the frequency to 60 Hz with the right shuttle. Select the
current function with the Frequency/Current selector. Set the current limit to the maximum value using the right shuttle. Set the output voltage to 135V with the left shuttle.
5. Enable the output by pressing the output “on/off” button in the top right of the front panel.
The green LED above the button will illuminate when the output is on. The output should be a clean 135 volt AC sinewave having less than 1% distortion.
6. Apply full load (refer to table on Figure 3-3) to the output of the source and verify the output
remains within 2% of the initial 135 volt value. The output should still be clean and the distortion should still be less than 1% at 60 Hz.
7. Using the right shuttle set the output current limit value to 6 amps. The system should go
into current limit and give an error message on the display (err. -300) that indicates an output fault condition and the output will go off. Return the current value to the maximum current and disconnect the load.
8. Repeat steps 4 through 7 but set the output for the following: Hi voltage range and the
current limit to maximum value. The output load should be connected to the Hi range output connector. The load value is shown in Figure 3-3, for the 270 volt output.
Note: Output connectors must be changed when changing voltage ranges.
12
May 2000 P Series
User and Programming Manual - Rev H California Instruments
In the unlikely event the power source does not pass the functional test, refer to the calibration procedure on page 6-1 or call California Instrument’s customer satisfaction department for further assistance.
High Range Load
Low Range Load
Load ON/OFF Switch Oscilloscope
AC Output
or Distortion Analyzer
135.0 Unit DMV Under Test
AC Input
Model Range Current Load
1251P 135V 9.2A 14.6 1251P 270V 4.6A 58.6 1001P 135V 7.4A 18.2 1001P 270V 3.7A 72.9
801P 135V 6.0A 22.5 801P 270V 3.0A 90.0
Figure 3-3: Functional Test Setup
P Series May 2000
13
User and Programming Manual - Rev H California Instruments

4. Front Panel Operation

4.1 Front Panel Guided Tour

The front panel can be divided in a small number of functional areas:
Output Sockets
Status Indicator lights
Shuttle knobs.
LCD display
Button controls
Figure 4-1: Front panel view
14
May 2000 P Series
User and Programming Manual - Rev H California Instruments

4.1.1 Output Outlets

The Output Sockets are located on the right side of the front panel. It provides connection to the load from the AC source. When the low voltage range is selected, only the duplex US NEMA 5-15P output socket will be active. If the high voltage range is selected, only the European CEE7/7 socket will be active. Refer to Figure 3-1 and Figure 4-1 for socket locations.

4.1.2 Status Indicator Lights

Six LED status indicators are located directly above the LCD display. These LED’s correspond to the following conditions:
REMOTE The REMOTE LED indicates that the unit is in remote control
mode. If the RS232C interface is used, the REMOTE state can be enabled by the controller using the SYST:REM command. Any time the REMOTE LED is lit, the front panel of the 1001P/1251P unit is disabled. There is no LOCAL button that allows the user to regain control of the front panel. The SYST:LOC command will enable the front panel controls
FAULT The FAULT LED indicates an output overvoltage or
overtemperature condition. An overtemperature is mostly likely caused by poor air flow. Check the air flow intake at the rear of the unit to make sure it is not obstructed.
OUTPUT The Output LED indicates the status of the OUTPUT ON/OFF
button. When the Output LED is not lit, the output voltage is not present at the output socket regardless of the voltage setting.
RANGE The Range LED indicates the selected output voltage range.
When it is illuminated it indicates the high voltage range has been programmed.
FREQUENCY Illuminates when the LCD display shows the programmed
frequency.
CURRENT Illuminates when the LCD displays shows the programmed
current limit or measured current values.
P Series May 2000
15
User and Programming Manual - Rev H California Instruments

4.1.3 The Shuttle Knobs

Counter Clockwise clockwise
DEC. INC.
Figure 4-2: Shuttle Knob
There are two shuttle knobs located below the LCD display which are used to change setup parameters for voltage, frequency and current limit. The mode button controls the function of the right shuttle. The right shuttle will control either the frequency or the current limit as indicated by the indicator above the right LCD.

4.1.4 FUNCTION Buttons

There are three function buttons for the Output Voltage Range, Output State and Shuttle Mode. The following is a description of these buttons:
KEY DESCRIPTION
RANGE The RANGE button is used to change the voltage range
OUTPUT The OUTPUT button will toggle the output to enable or
between the low range (0 to 135 volts) and (0 to 270 volts). The LED above the switch will light to indicate the high voltage range selection. The output voltage will be reset to zero voltage after a range change.
disable the output. The LED above the button will light when the output is on. No output voltage will be present when the OUTPUT button is off despite the level of voltage programmed.
16
May 2000 P Series
User and Programming Manual - Rev H California Instruments
MODE The MODE button will control the function of the right Shuttle
knob and the LCD display. The Shuttle will control the output frequency and the display will show the program frequency value when the mode selection is frequency. The Shuttle knob will program the current limit and the display will show its value in the current mode. The display will revert back to showing the measured current after 3 seconds from the last movement of the shuttle. The measurement is updated 4 times per second. The display mode is indicated by the two LED’s above the LCD display.

4.1.5 LCD Display

The LCD display consists of two 4 digit, 7 segment displays. The voltage display shows the programmed voltage. The Frequency/Current display shows either the programmed frequency or current limit. In the current limit mode the display switches to display the output current after 3 seconds. The Frequency/Current select button will define the operating mode of the frequency/current display.

4.2 How to...

This chapter covers some common tasks that are often performed with an AC power source. These examples are written in a How to... format and provide step by step instructions on how to set up the AC Source for a specific task.

4.2.1 Set the Output

Output parameters are Voltage, Frequency and Current Limit.
1. Disable the output by pressing the OUTPUT button. The LED above the button will turn off.
2. Use the left shuttle to set the output voltage. Clockwise will increase the output, counter clockwise will reduce the output. Thedisplay above the shuttle will show the voltage setting.
3. Use the right shuttle to set the frequency and current limit. The Frequency/Current select button will define the function of the shuttle and the display above it. The Frequency or Current LED will turn on to indicate the function controlled by the right shuttle.
4. Enable the output by pressing the OUTPUT button

4.2.2 Slewing Output Values

The output parameters can be slewed using the shuttles.
1. Enable the output by pressing the OUTPUT button. The LED above it will turn on.
P Series May 2000
17
User and Programming Manual - Rev H California Instruments
2. Use the left shuttle to set the output voltage. Clockwise will increase the output, counter clockwise will reduce the output. The display above the shuttle will show the voltage setting.
3. Use the right shuttle to set the frequency and current limit. The Frequency/Current button will define the function of the shuttle as indicated by the display above it. The Frequency or Current LED will turn on to indicate the function in control.

4.2.3 View Current Measurements

Current measurements can be called up as follows:
1. Press the Frequency/Current button to select the Current function.
2. Immediately the Frequency/Current display will show the measured current.
3. Moving the right shuttle will interrupt the current measurement. The display will show the current limit value.
4. After a short delay the display will revert back to show the measured current.

4.2.4 Voltage Range Change

The voltage range can be changed as follow:
1. Press the HI RANGE button located in the upper left corner. The output voltage will reset to 0 volts.
2. Use the left shuttle knob to set the output voltage.

4.2.5 Output Control

The Output can be disabled or enabled as follows :
1. Pressing the OUTPUT button when the output LED is on will disable the AC source output. The programmed voltage setting will remain at the last program value.
2. Pressing the OUTPUT button again will enable the output voltage and the output will revert to the last programmed value.

4.2.6 Setting the Power on Initialization Values

All P and RP series are supplied with default factory settings when the unit is powered up. The factory settings are:
18
May 2000 P Series
User and Programming Manual - Rev H California Instruments
Voltage range Low Voltage 0 V Frequency 60 Hz Current limit Max available current Display mode Frequency Output OFF It is possible to change the power on initialization values in one of two ways:
1. Using the optional RS232 or IEEE-488 (RP only) interface and the supplied PGUI/PGUI32 program.
2. Using the front panel. (requires main firmware release 1.0 or higher).
To change the power on initialization values from the front panel, proceed as follows:
1. Set the unit up in the desired way from the front. (Range, voltage, frequency, current limit, output relay state). Note: The display mode – frequency or current – is not saved as part of the set up and always defaults to frequency.
2. Press and hold the Select key (normally toggles between F and C readouts).
3. While holding the Select key, press the OUTPUT ON/OFF key. This will save the present front panel settings in non volatile memory register (NVM) no 7 and assign this register as the power on register.
4. Release both keys.
5. This procedure can be repeated as often as needed by the user.
P Series May 2000
19
User and Programming Manual - Rev H California Instruments

5. Principle of Operation

5.1 General

An explanation of the circuits in the AC Source is given in this section. Refer to Figure 5-1 for a block diagram of the system.
AC Power Factor DC - DC DC - AC AC Input Corrector (PFC) Converter Converter Output
Oscillator & Controls

5.2 Overall Description

The AC input is fed to the power factor correction, boost type converter. The converter steps the voltage to 385 VDC while drawing only sinusoidal current from the input power line.
The DC to DC converter provides isolation and changes the voltage to 250 VDC or 400 VDC depending on whether the low output range or high output range is selected.
The DC to AC converter develops an AC sine wave voltage at the output range, frequency and amplitude programmed by the oscillator.
The oscillator board provides the reference signal to the DC to AC converter and has the front panel shuttles and switches to control and view the setting of voltage, frequency and current limit.
Figure 5-1: AC Source block diagram
20
May 2000 P Series
User and Programming Manual - Rev H California Instruments

5.3 Power Factor Correction Module (PFC)

The P.F.C consists of an input EMI filter, input bridge and boost convert er. The boost converter is PWM controlled by a single chip that adjusts the pulse width during the cycle so that only sine wave current is drawn from the supply. An auxiliary winding on the boost inductor provides “bootstrap” power to the logic circuits and is self sustaining.

5.4 DC - DC Converter Module

The 385 VDC from the boost converter is fed to the DC to DC module. It first supplies 385 volts DC to the auxiliary power supply. When the auxiliary power supply starts it, it supplies 6 isolated DC supplies to the AC module. It also supplies 15 volts to the DC converter on the same DC module. The DC to DC converter will then start operating. The DC converter supplies either 250 volts DC or 400 volts DC to the DC to AC module.
The DC to DC converter is a full bridge converter and the auxiliary converter is a two FET forward converter.
All the 8 LEDs on the DC-DC board should be lit under normal operation. The LED DS200 is lit when the DC output is up and within regulation. The amber LEDs DS5, DS4 and DS3 indicate that the 15V power to the optocouplers on the AC converters is within regulation. The red and green LED’s, DS7 and DS6 indicate logic power to the AC converter is ok. The red LED DS8 indicates that logic power to the DC converter is within specification at 15 volts. The LED DS2 shows that 24 volt fan power is available.
Table 1: Logic Board Led's
LED# FUNCTION COMMENTS
DS2 24V ok 24 volt fan power is available. DS3 15V ok 15V power to the optocuplers on the AC converters is within
regulation
DS4 15V ok 15V power to the optocuplers on the AC converters is within
regulation
DS5 15V ok 15V power to the optocuplers on the AC converters is within
regulation DS6 power ok logic power is ok to the AC converter DS7 power ok logic power is ok to the AC converter DS8 power ok logic power is ok to the DC converter
DS200 DC bus ok DC output is up and within regulation
P Series May 2000
21
Loading...
+ 66 hidden pages