Bunnell Life Pulse 203 User manual

Page 1
Life Pulse
High Frequency Ventilator
In-Service Manual
436 Lawndale Drive Salt Lake City, UT 84115
01801-07.12
01513-07.12
Phone: 801-467-0800 Hotline: 800-800-4358 Fax: 801-467-0867 Website: www.bunl.com
Page 2
i
In-service Manual

TABLE OF CONTENTS

HOW TO USE THIS MANUAL ................................................................................................................................................. 1
CHAPTER 1: OVERVIEW .......................................................................................................................................................... 2
HFV WITH CV ..................................................................................................................................................... 3
LIFEPORT ADAPTER ....................................................................................................................................... 4
CHAPTER 2: SETUP .................................................................................................................................................................. 5
REAR PANEL CONNECTIONS ....................................................................................................................... 6
FRONT PANEL CONNECTIONS .................................................................................................................... 7
PATIENT BOX CONNECTIONS ............................................................................................................... 8
SETUP CAUTIONS ............................................................................................................................................ 9
CHAPTER 3: VENTILATOR CONTROLS & TEST PROCEDURE ................................................................................... 10
THE CONTROLS SECTION ............................................................................................................................11
PERFORMING A TEST ....................................................................................................................................12
A FAILED TEST ................................................................................................................................................13
CHAPTER 4: PRESSURE MONITORING ............................................................................................................................. 14
MONITOR DISPLAYS ......................................................................................................................................15
PRESSING ENTER BUTTON .........................................................................................................................16
PURGE PAUSES ..............................................................................................................................................17
CHAPTER 5: HUMIDIFIER ...................................................................................................................................................... 18
GAS FLOW THROUGH HUMIDIFIER ...........................................................................................................19
STARTING THE HUMIDIFIER ........................................................................................................................21
CHANGING THE CIRCUIT ..............................................................................................................................22
CIRCUIT CHANGE PREPARATIONS .................................................................................................................
CHECKLIST ...........................................................................................................................................................
FRONT PANEL DUTIES .......................................................................................................................................
PATIENT BOX DUTIES ........................................................................................................................................
POST-CIRCUIT CHANGE.....................................................................................................................................
AFTER ENTER BUTTON IS PRESSED: ..............................................................................................................
IMPORTANT ................................ ................................................................ ..........................................................
TO COMPLETE A CIRCUIT CHANGE: ..............................................................................................................
IDENTIFYING PROPER HUMIDIFICATION .................................................................................................25
CHAPTER 6: START UP ......................................................................................................................................................... 26
MEASURING & DISPLAYING PRESSURES ...............................................................................................27
CHOOSING STARTING VALUES ..................................................................................................................28
6 STEPS TO START HFV ................................................................................................................................29
CHAPTER 7: PATIENT MANAGEMENT .............................................................................................................................. 32
CLINICAL OBJECTIVES ...................................................................................................................................32
OXYGENATION .................................................................................................................................................33
OXYGENATION OVEREXPANDED LUNGS .................................................................................................35
FINDING OPTIMAL PEEP ...............................................................................................................................36
VENTILATION ...................................................................................................................................................37
UNDERSTANDING SERVO PRESSURE .....................................................................................................39
CHAPTER 8: SUCTIONING THE PATIENT ......................................................................................................................... 40
SUCTION PROCEDURE #1 ...........................................................................................................................41
Page 3
ii
In-service Manual
TABLE OF CONTENTS
SUCTION PROCEDURE #2 ...........................................................................................................................42
READY LIGHT MUST BE ON! ................................ ................................................................ ...............
INSTILL INTO JET PORT THEN RECONNECT LIFE PULSE CIRCUIT. ...............................
ONE PERSON DISCONNECT CV CIRCUIT FROM LIFEPORT. ................................................
SECOND PERSON APPLIES SUCTION GOING INTO AND OUT OF THE ET TUBE. ........
SUCTION PROCEDURE #3 ...........................................................................................................................44
SELECT AN IN-LINE SUCTION SYSTEM COMPATIBLE WITH THE LIFEPORT
ADAPTER .............................................................................................................................................................
SELECT AN IN-LINE SUCTION SYSTEM WITH A CATHETER THAT IS STRAIGHT
FOR THE FIRST FEW INCHES .....................................................................................................................
SUCTION ACCORDING TO PROCEDURE 1 OR 2 DESCRIBED EARLIER IN THIS
SECTION ..............................................................................................................................................................
CHAPTER 9: WEANING .......................................................................................................................................................... 45
GENERAL GUIDELLINES FOR WEANING ....................................................................................................45
GENERAL GUIDELINES FOR WEANING ............................................................................................................................
CHAPTER 10: VENTILATOR ALARMS ............................................................................................................................... 48
READY LIGHT ON .............................................................................................................................................48
ALARM LIMITS ..................................................................................................................................................49
JET VALVE FAULT .........................................................................................................................................50
VENTILATOR FAULT: HFV CONTINUES RUNNING .................................................................................51
VENTILATOR FAULT:HFV STOPS RUNNING ..............................................................................................52
VENTILATOR FAULT: CODE 10 .....................................................................................................................53
LOW GAS PRESSURE ....................................................................................................................................54
CANNOT MEET PIP ..........................................................................................................................................55
LOSS OF PIP .....................................................................................................................................................56
HIGH PIP.............................................................................................................................................................57
CHAPTER 11: INTERPRETING VENTILATOR ALARMS ................................................................................................ 58
ALARM DISPLAY .........................................................................................................................................58
LOOK AT THE PATIENT ..................................................................................................................................59
USE COMMON SENSE ....................................................................................................................................60
KNOW THE LIFE PULSE .................................................................................................................................61
COMMON THINGS HAPPEN COMMONLY .................................................................................................62
USE AVAILABLE RESOURCES .....................................................................................................................63
CHAPTER 12: HUMIDIFIER ALARMS ................................................................................................................................. 64
HUMIDIFIER OPERATION ..............................................................................................................................65
CIRCUIT FAULT: LEVEL ALARM ...................................................................................................................66
WATER LEVEL SENSING ...............................................................................................................................67
CIRCUIT TEMPERATURE ...............................................................................................................................68
TEMPERATURE CONTROLS .........................................................................................................................69
SETTING CARTRIDGE TEMPERATURE .....................................................................................................70
CONDENSATION PROBLEMS .......................................................................................................................71
MANUALLY PURGING .....................................................................................................................................72
HIGH / LOW WATER LEVEL ...........................................................................................................................73
HIGH / LOW TEMPERATURE .........................................................................................................................74
APPENDIX A: POST-TEST QUESTIONS ............................................................................................................................. 75
APPENDIX B: POST-TEST ANSWERS ................................................................................................................................ 89
Page 4
1
Introduction

HOW TO USE THIS MANUAL

Objectives
1. Understand how to use this manual.
2. Know how to navigate to the information you need.
INSTRUCTION BOXES
Boxes containing instructions and summaries of procedures will appear in lightly shaded boxes and will look like this.
WARNING: Warnings will also appear in
lightly shaded boxes and will look like this.
CONTROLS
ON
STANDBY
STANDBYENTER ENTER
SILENCEENTER
Each chapter includes objectives for the reader
Graphics and instruction boxes will appear in the
Text that lists sequential procedures:
1. Is numbered
2. Like this
Otherwise, lists will be:
Bulleted Like this
Text that instructs you to press a button will
highlight the button‟s font like this: press MENU
Each chapter begins with a chapter designation
Each page will list its contents in
large, bold, capitol letters
Text will appear in the right column
Page 5
2
Objectives
1. Know the 5 subsystems of the Life Pulse and their purposes.
2. Understand the relationship between the Life Pulse and the conventional ventilator.
3. Know the structure and function of the LifePort adapter.
NOTE: Graphics that display illuminated
indicator lamps (LEDs) in this manual reflect Life Pulse model 203. For model 203A, please refer to the table at right:
INDICATOR LED MODEL 203 MODEL 203A
STANDBY
Red Yellow
TEST
Red Yellow
JET VALVE OFF
Red Yellow
READY
Red Green
WAIT
Red Yellow
CIRCUIT
Red Yellow
CARTRIDGE
Red Yellow
MONITOR
ALARMS
CONTROLS
PATIENT BOX
HUMIDIFIER
Chapter 1
OVERVIEW
The Life Pulse High-Frequency Ventilator is a microprocessor-controlled infant ventilator capable of delivering and monitoring between 240 and 660 heated, humidified breaths per minute.
The Life Pulse is composed if 5 subsystems:
MONITOR: Displays patient and machine
pressures.
ALARMS: Indicates various conditions that may
require attention.
CONTROLS: Regulates the On-Time, Peak
Inspiratory Pressure, and Rate of the HFV
breaths.
HUMIDIFIER: Monitors and controls the
temperature and humidification of gas flowing through the disposable humidifier circuit to the
patient.
PATIENT BOX: Contains the pinch valve that
breaks the flow of pressurized gas into tiny jet pulses and sends pressure information back to the ventilator‟s microprocessor.
Together, these elements form a system that offers a variety of options for managing patients and the potential for improving blood gases using less pressure.
Page 6
3
VENTILATOR FUNCTIONS
LIFE PULSE
High-frequency breaths Airway pressure monitoring Humidification Alarms
CONVENTIONAL VENTILATOR
Regulate PEEP Fresh gas for spontaneous breathing Supplementary IMV Airway dilation
HFV with CV
The Life Pulse is used in conjunction with a conventional infant ventilator. The conventional ventilator has 4 functions:
provides fresh gas for the patient‟s
spontaneous breathing,
regulates PEEP, provides supplementary IMV when needed,
and
provides periodic dilation of airways when
needed.
The purpose of the supplementary IMV is to provide background breaths sufficient to recruit atelectatic alveoli.
The purpose of the PEEP provided by the conventional ventilator is to maintain the inflation of alveoli with adequate FRC.
Using IMV periodically to dilate airways without interrupting the Life Pulse affords opportunities to ventilate areas downstream from airway restrictions.
Page 7
4
Jet inlet port
Jet port cap
Main port
Pressure monitoring tube
Conventional endotracheal tube
Pressure monitoring port
Jet port 15 mm port
LifePort Adapter
The LifePort adapter allows both the conventional ventilator and the Life Pulse to be connected to a patient. The LifePort has three main features:
15 mm Port: provides the standard
connection to the conventional ventilator.
Jet Port: the entrance for the high­frequency pulses provided by the Life
Pulse.
Pressure Monitoring Tube: allows the Life Pulse to display approximations of distal tip airway pressures.
Accurate pressure measurements through the LifePort are fed back to the ventilator and provide the Life Pulse with information necessary to control peak inspiratory pressure
(PIP) . For example, if the measured PIP is higher
than the set PIP, the Life Pulse stops pulsing. If the PIP suddenly reaches automatically set
criteria, the Life Pulse dumps the stored Servo Pressure to insure that the Life Pulse will not deliver excess gas to the patient.
Page 8
5
Objectives
1. Understand the connections to the electrical power, air, oxygen, and water necessary for the Life Pulse to function.
2. Know how to install the humidifier cartridge and circuit.
Chapter 2
SETUP
The Life Pulse should be positioned so that the displays are easy to read and its controls and the conventional ventilator controls are within easy reach. The cart is on 5 inch, lockable casters for easy portability and stability.
The Bunnell ventilator cart is designed to
carry most of the patient‟s
cardiopulmonary equipment. For example, oxygen analyzers and other monitors can be placed on the top shelf.
The Life Pulse is typically placed on the second shelf, the conventional ventilator on the third, and an uninterruptable power supply on the bottom.
Setup procedures for the Life Pulse include making connections in three places:
the Rear Panel,
the Front Panel, and
the Patient Box.
Page 9
6
CHECKLIST
Rear Panel Connections
Power cord Gas in from blender O
2
Sensor (Only SN‟s 2413 and lower)
Patient Box Alarm volume
DO NOT BLOCK VENT HOLES
ALARM VOLUME
PATIENT BOX
+
+
+
+
+
+
MIXED GAS
INPUT
+
DO NOT BLOCK
HOURS
LIFE PULSE
30 - 60 PSI
DANGER-- EXPLOSION HAZARD:
RISQUE D’EXPLOSION:
CAUTION:
DO NOT USE
IN THE PRESENCE OF FLAMMABLE ANESTHETICS.
NE PAS EMPLOYER
EN PRESENCE D’ANESTHESIQUES INFLAMMABLES
OPERATE ONLY IN ACCORDANCE
WITH MANUFACTURER’S OPERATING MANUAL.
EQUIPMENT SHOULD BE CONNECTED TO AN
EQUIVALENT RECEPTACLE MARKED “HOSPITAL GRADE” OR “HOSPITAL ONLY”. RISK OF
ELECTRICAL SHOCK: DO NOT REMOVE COVER. REFER SERVICING TO AUTHORIZED PERSONNEL. .
THIS EQUIPMENT COMPLIES WITH THE REQUIREMENTS IN PART 15 OF FCC RULES FOR A CLASS A COMPUTING DEVICE. OPERATION OF THIS EQUIPMENT IN A RESIDENTIAL AREA MAY CAUSE UNACCEPTABLE INTERFERENCE TO RADIO AND TV RECEPTION REQUIRING THE OPERATOR TO TAKE WHATEVER STEPS ARE NECESSARY TO CORRECT THE INTERFERENCE.
THIS DEVICE AND ITS USE IS COVERED BY ONE OR MORE OF THE FOLLOWING US PATENTS.
HIGH FREQUENCY VENTILATOR MODEL 203A
SERIAL NO. XXXX
MADE IN THE U.S.A. MANUFACTURED BY
120V / 3A / 60Hz
+
3
C US
®
02641– 00.3
Alarm volume
Patient Box connector
Mixed gas input
Power cord
Circuit breaker
Gas dump port
Life Pulse Model 203
Life Pulse Model 203 A
REAR PANEL CONNECTIONS
1. Plug in the power cord to a standard 110 volt outlet or, preferably, an uninterruptible power supply.
2. Connect a high-pressure oxygen hose from a low flow air/oxygen blender (0-30 L/min) or the output from the low flow port (2-100 L/min) of a standard blender to the Mixed Gas Input fitting. A minimum pressure of 30 psi is required to operate the Life Pulse.
3. Attach an oxygen analyzer to the Oxygen Sensor in order to monitor FiO2 (not applicable to serial number 2414 or higher), or monitor FiO2 from the air/oxygen blender output.
4. With Model 203, plug in one of a variety of recording devices to the Analog Output to monitor airway pressure graphically (Optional).
5. Connect the Patient Box to the ventilator by its electrical cable attached to the multi­pinned connector.
6. Adjust the volume of the audible alarms using the Alarm Volume dial.
Page 10
7
2
1
6
5
4
3
GAS OUT
PURGE
Transfer tube to
water supply
Cartridge
in door
Green gas inlet tube to Gas Out
Purge tube
Water inlet
tube
Water Pump
CHECKLIST
Front Panel Connections
1 Cartridge in and door latched
2 Green gas inlet tube
3 Purge tube
4 Water inlet tube
5 Water pump
6 Water supply bag
WARNING: The water inlet tube of the humidifier cartridge/circuit must be latched into the pump housing to prevent cartridge overfill and delivery of water to the patient by gravity feed.
WARNING: The water supply should be positioned at or below the level of the humidifier cartridge to decrease the potential of overfilling the cartridge by gravity feed.
FRONT PANEL CONNECTIONS
1. Insert the Humidifier Cartridge into the cartridge door. Latch the door shut to make all electrical connections between the cartridge and the Life Pulse.
2. Connect the green gas inlet tube to the
large barbed connector labeled GAS OUT.
3. Connect the small diameter purge tube to
the barbed connector labeled PURGE.
4. Locate the water inlet tube, a clear tube running from the lower right corner of the cartridge with a check valve and a Leur fitting on the end.
5. Install the water inlet tube by placing the clear tube into the pump housing, closing the pump door securely to pinch the water inlet tube inside, and connect the water transfer tube to the Luer connector on the end of the water inlet tube.
6. Tap the sterile water bag with the other end of the water transfer tube and open the tubing clamp.
Use only sterile water for the cartridge water supply.
Page 11
8
CHECKLIST
Patient Box Connections
Pinch tubing in pinch valve Purge tube connect
Purge Tube
Push To Load button
To Patient
Pinch Tube
Pinch Valve
End view
PATIENT BOX CONNECTIONS
1. Locate the soft pinch tubing portion of the circuit just beyond where the red and white wires inside the circuit terminate.
2. Place the pinch tubing in line with the pinch valve assembly.
3. Hold the pinch tubing on each side of the pinch valve assembly.
4. Press the PUSH TO LOAD button with one
finger.
5. Slide the pinch tubing into the pinch valve assembly until you feel it snap into place.
The illustrations on the left show a proper pinch tubing placement. The entire width of the pinch tube should be within the pinch valve jaws.
6. Connect the small purge tube to the
barbed connector labeled FROM PURGE.
The disposable humidifier cartridge and circuit are now ready for operation, and the Life Pulse is ready to be turned on.
Page 12
9
SETUP CAUTIONS
Two cautions should be noted about the Set Up procedure:
First, patients are often placed on the Life Pulse on an emergency basis. You can save time if the ventilator is stored clean and partially set up, ready for use.
Do NOT install the pinch tube in the pinch valve prior to actual patient set-up.
The disposable cartridge/circuit, water transfer tube, and a test lung can be placed with the Life Pulse. If using the Bunnell Cart, everything necessary to begin high-frequency
jet ventilation can be wheeled to the patient‟s
bedside.
Second, notice that the installation description includes only cartridge/circuit tubing supplied by Bunnell Incorporated.
Do NOT make modifications to the supplies or the set up procedure!
Pressures cannot be monitored accurately and the Life Pulse will not work properly if other equipment is teed into the pressure monitoring tube or if leaks are present in the circuit.
Page 13
10
Objectives
1. Be able to attach the Life Pulse circuit to the LifePort adapter.
2. Understand the ON, STANDBY, ALARM SILENCE, TEST, and ENTER buttons.
3. Know how to perform a Life Pulse systems test and know the meaning of VENTILATOR FAULT alarms 02, 03, and 04.
4. Be able to enter NEW settings and convert them to NOW settings.
CONTROLS
ON
SILENCE
TEST
POWER ON
STANDBY
STANDBYENT ER ENTER
SILENCE
ENTER
VENTILATOR CONTROLS &
TEST PROCEDURE
This section discusses how to manipulate the CONTROLS and perform a systems test.
When the ON button is pressed, the Life Pulse
activates into the Standby mode with an audible alarm sounding.
The Standby mode is indicated by a small
light in the corner of the STANDBY button and
an audible alarm every 30 seconds. The alarm
may be silenced by pushing the SILENCE
button.
Once the 60-second alarm silence time has expired, the audible alarm will sound 6 beeps every 30 seconds to remind you that the Life Pulse is in the Standby mode and not operating.
The Life Pulse can monitor conventional ventilator and other high-frequency ventilator pressures in the Standby mode.
This information will be displayed in the
MONITOR section.
Chapter 3
Page 14
11
+
+
+
+
RATE JET VALVE TIMEonon/offPIP
NOW
NEW
CONTROLS
TESTSTANDBY
ENTER
THE CONTROLS SECTION
There are only three setting parameters in the CONTROLS section: PIP (Peak Inspiratory Pressure), RATE, and JET VALVE ON Time,
which is nearly synonymous with Inspiratory Time.
These settings may be changed by pressing the up and down arrow buttons next to their respective displays.
1. Press the ENTER button to transfer the NEW settings into the NOW settings (the
settings the patient will receive).
2. Press the STANDBY button to place the
Life Pulse into the Standby mode. The Life Pulse stops producing high-frequency jet
ventilation and only the NEW settings are
displayed.
A systems test should be performed to insure that the Life Pulse is in good operating condition before connecting the system to a patient.
A test lung must be attached to the patient end of a standard endotracheal tube and LifePort adapter prior to performing the Test.
The test lung can be as simple as the one pictured: a finger cot, or the finger of a rubber glove, taped lightly to the tip of the ET tube.
Page 15
12
TESTSTANDBY
JET VALVE TIMEonon/off
ENTER
seconds
ALARMS
JET VALVE FAULT
LOW GAS PRESS
CANNOT MEET PIP
LOSS OF PIP
HIGH PIP
VENTILATOR FAULT
SILENCERESET
All possible alarms will be illuminated briefly during the systems test.
Press TEST button to begin systems test.
PERFORMING A TEST
The Life Pulse‟s systems test will ensure that the
ventilator is operating according to specifications. To perform the test, follow these steps:
1. Attach a LifePort adapter to an ET tube and test lung.
2. In the Standby mode, connect the patient end of the Life Pulse circuit, coming from the pinch valve, to the Jet port on the side of the LifePort adapter.
3. Connect the clear pressure monitoring tube of the LifePort adapter to the Patient Box
barbed connector labeled MONITORING LUMEN.
4. Press the TEST button. An automatic test
begins which determines the integrity of all the ventilator‟s electronics and valves.
5. Observe the front panel to assure all LEDs and displays are functional, and listen to make sure the audible alarm is functioning properly. If no problems are detected, all the ventilator displays will illuminate 1 through 9 in sequence and all the alarm messages will be displayed.
The Test procedure will end with the Life Pulse in the Standby mode and an audible alarm sounding.
6. Silence the audible alarm by pressing the
SILENCE button. The audible alarm stops for
60 seconds.
7. Once the internal Test passes, perform an operational test using the test lung. A conventional ventilator is not needed and the LifePort adapter 15mm connector is left open to the room.
8. Press the ENTER button to activate the
default control settings (20, 420, 0.02).
9. Once the PIP is stable, the Ready light will activate. Make sure the pressures are stable and the PEEP is reading 0.0 ± 1 cm H2O.
10. If the PEEP is > ± 1 cm H2O, switch out the Patient Box and repeat the operational test.
Page 16
13
SILENCE
ENTER
VENTILATOR FAULT
TESTSTANDBY
JET VALVE TIMEonon/off
ENTER
seconds
CHECKLIST
Fault Code Numbers
02 Purge tube disconnect or pressure
transducer failure
03 Green gas inlet tube disconnect, pinch
tube incorrectly installed, or stuck Servo Pressure control valve
04 Electrical component failure
WARNING: All patient connections to the Life Pulse circuit must only be made while the Life Pulse is in the STANDBY mode. Failure to comply may result in a high volume of gas being delivered to the patient.
A FAILED TEST
If an internal fault is detected, the test
sequence will stop, VENTILATOR FAULT will be displayed in the ALARMS area, and a code number 02, 03, or 04 will appear in the ON/OFF window of the CONTROLS section.
A VENTILATOR FAULT code may or may not
mean the Life Pulse has a serious problem.
For example, if the purge tube is disconnected during the Test, either at the Patient Box or
the front panel, a VENTILATOR FAULT 02 will
be displayed.
VENTILATOR FAULT 02 - check the purge
tube for a disconnect at the front panel or the Patient Box. This fault will also occur if the pressure transducer in the Patient Box has failed.
VENTILATOR FAULT 03 - check the green
gas inlet tube for a disconnect at the GAS OUT connector. This fault may also occur if one of the Servo Pressure control valves is not working properly.
VENTILATOR FAULT 04 - You may observe
the Life Pulse performing this check by watching the code display area closely. As the test is performed, the display
momentarily flashes 04. If the test is passed, the 04 disappears. If it fails, the 04 stays lit, a VENTILATOR FAULT code
appears, and an audible alarm sounds.
Once it has successfully passed the test, and after operating properly on a test lung, the Life Pulse is ready for clinical use.
If the Life Pulse is unable to achieve the desired settings on a test lung, or the cause of
a VENTILATOR FAULT cannot be determined, or an 04 stays displayed at the end of the test,
call the Bunnell Hotline (1-800-800-4358).
Page 17
14
Objectives
1. Appreciate the advantages of the LifePort Adapter and the purpose of each of its 3 ports.
2. Understand the parameters displayed in the MONITOR section and where they are measured.
3. Understand Servo Pressure and its clinical relevance.
4. Understand the function of the PURGE.
PIP
P
PEEP
SERVO PRESS
MAP
cm H O
2
cm H O
cm H O
2
cm H O
2
ON OFF
JET VALVE
MONITOR
PSI (0.145kPa)
Chapter 4
PRESSURE MONITORING
The MONITOR displays reflect the pressures
at the tip of the endotracheal tube and the internal Servo or drive pressure of the Life Pulse. If the Life Pulse is in Standby, and there is no patient connected to the Patient Box, the displays will all read zero.
Once the pressure monitoring tube of the LifePort adapter has been connected to the Patient Box, the Life Pulse, even in its Standby mode, will begin to monitor the pressures being delivered by the conventional ventilator. These pressures are updated in
the MONITOR displays every 10 seconds.
The Life Pulse monitors:
PIP: Peak Inspiratory Pressure,
PEEP: Positive End-Expiratory Pressure,
MAP: Mean Airway Pressure,
P: PIP minus PEEP, roughly equivalent
to tidal volume, and
SERVO PRESSURE: internal driving
pressure; always 0.0 in Standby mode.
Page 18
15
SERVO PRESSURE
Improved compliance and/or resistance Air leak
Disconnected tube
SERVO PRESSURE
Worsening compliance and/or resistance ET tube obstruction Tension pneumothorax Patient needs suctioning
Right mainstem intubation
PIP
SERVO PRESSURE
P (i.e., VT)
MAP
PEEP
MONITOR DISPLAYS
PIP: the average of maximum airway
pressures measured during a set time period
PEEP: the average minimum airway
pressure
MAP: is an average of pressures measured
from the total pressure waveform. It includes pressures produced by high­frequency ventilation, conventional ventilation, or spontaneous breathing.
P: (Delta P) simply the arithmetic
difference between PIP and PEEP. P is roughly proportional to Tidal Volume and its importance in clinical decision making will become more apparent in the discussion of patient management.
SERVO PRESSURE: the internal drive
pressure of the Life Pulse; indicates how much gas flow must be produced to meet the NOW PIP, Rate, and On-Time requested by the operator.
Servo Pressure is regulated by the ventilator‟s microprocessor and is outside the control of the operator. Bigger patients, or those with more compliant lungs, will require higher Servo Pressures whereas those infants with smaller and/or less compliant lungs will require lower Servo Pressures.
Servo Pressure is an indication of how much gas flow the Life Pulse must generate to meet the settings requested. Servo Pressure changes above or below the established operating level for a particular patient may be a result of changes in lung compliance, airway resistance, or lung volume.
Servo Pressure changes give an early
indication that the patient‟s condition may be
improving or worsening.
Page 19
16
ALARMS
SILENCERESET
READY
cm H0
2
PIP
cm H0
2
P
cm H0
2
cm H0
2
PEEP
SERVO PRESS
PSI (0.145kPa)
MAP
MONITOR
JET VALVE
ON OFF
JET VALVE TIME on
on
off
Press ENTER button to begin ventilation
READY light ON
Jet Valve ON/OFF lights flash
PRESSING ENTER BUTTON
When the ENTER button is pressed, the
microprocessor will begin increasing the Servo Pressure from zero to whatever value
will produce the NOW PIP at the end of the NOW ON-TIME at the NOW RATE.
Although it will typically take a short time
for the actual PIP to reach the NOW PIP, the displayed PIP will equilibrate slower because
of the averaging characteristics of the display. Thus, it may take longer (typically
within a minute) for the monitored PIP to reach the NOW PIP.
When the monitored NOW PIP is stable for 20 seconds, the READY light will illuminate. The READY light indicates the Life Pulse is
providing ventilation at the settings you have requested, and alarm conditions that are set automatically have been established.
(The specific criteria which produce the
READY condition are discussed in the ALARMS section, but the READY condition is
mentioned here because the monitoring
specifications change after the READY light is
on: the display is updated every 2 seconds.)
The JET VALVE ON/OFF lights turn on and
off in conjunction with the pinch valve that is
located in the Patient Box. ON means the valve is open. OFF means the valve is closed.
These lights serve as a visual check on the status of high-frequency ventilation in general and on the extent to which the patient or the conventional ventilator may be causing interruptions of the Jet pulses.
The Life Pulse will sense a high pressure and will pause whenever the monitored PIP
exceeds the NOW PIP. Ventilation resumes
when the PIP drops below the set level. Monitor such interruptions visually by
watching the ON/OFF lights.
Page 20
17
cm H0
2
PIP
cm H0
2
P
cm H0
2
cm H0
2
PEEP
SERVO PRESS
PSI (0.145kPa)
MAP
MONITOR
JET VALVE
ON
Extended pauses of the Green Jet Valve ON light indicate an interruption of jet pulses
PURGE PAUSES
The Life Pulse may pause periodically and briefly during operation for no apparent reason. Such pauses can occur during clinical use or when you are operating on a test lung.
These brief pauses are usually a result of the purge of the pressure monitoring tube that takes place once every 15 seconds.
The Purge valve is located next to the pressure transducer inside the Patient Box. When the valve opens it allows a pulse of dry air to flush the pressure monitoring tube.
On rare occasions, the purge may cause high pressure to be measured by the transducer that, in turn, makes the Life Pulse skip a beat in conjunction with its built-in response to high pressures.
Do not be concerned if the Life Pulse appears to hiccup every once in a while. If you time the pauses, you will find that they occur at some multiple of 15 seconds. (This pause usually will only occur on rates of 550 bpm or greater.)
Page 21
18
Objectives
1. Describe the flow of gas from the ventilator through the entire length of the Humidifier Cartridge/Circuit.
2. Understand the purpose of humidification and how it is produced and regulated.
3. Learn to adjust the temperature manually in the cartridge and the circuit.
4. Understand the function of the pinch tube portion of the circuit.
Purge tube
Circuit
Heating Wire
Circuit thermistor
Purge tube
Pinch Tube
Gas inlet tube
Cartridge thermistor
Gas to patient
Cartridge
Water supply tube
Chapter 5
HUMIDIFIER
The Life Pulse humidifier uses a one-piece
disposable cartridge and tubing set called the humidifier cartridge/circuit. This item
is often referred to as “the cartridge”, “the circuit”, or “the patient breathing circuit”.
All are terms that refer to all or part of the humidifier cartridge/circuit. It is pre­assembled and contains the heating wire, thermistors, and all connections needed for operation.
Gas from the GAS OUT connector on the
front panel of the ventilator flows into the humidifier cartridge via the green gas inlet tube.
The shorter clear tube is the water inlet tube and contains a check valve that prevents gas from leaking out when the water supply bag is disconnected.
Water is pumped into the cartridge against the cartridge pressure. Once the water inlet tubing is securely closed in the pump housing, water and air cannot be forced back into the water supply bag.
Page 22
19
Gas inlet
tubing (Green)
Baffles
Temperature
thermistor
Water level
sensing pins
Heating
plate
To Patient
GAS FLOW THROUGH HUMIDIFIER
The gas flows from the Life Pulse to the patient as follows:
The gas enters the cartridge through the green gas inlet tube.
The gas flows down below the water level then back up through a venturi mechanism, which atomizes some of the heated water.
The gas then flows over the heated reservoir of water and past a series of baffles which knock out water droplets.
The humidified gas passes over the cartridge thermistor which measures the temperature.
The microprocessor uses this temperature to regulate the amount of heat delivered to the cartridge through the metal heating plate.
This feedback control system uses the
CARTRIDGE temperature setting on the
humidifier front panel as its set point and controls the humidification of gas being delivered to the patient, not the temperature.
Heated and humidified gas leaves the cartridge and enters the circuit tubing with the red and white wires inside.
The temperature thermistor at the tip of the white circuit wire measures the actual temperature of the humidified gas in the circuit just before the Patient Box. The
value is displayed as CIRCUIT TEMP in the
humidifier display section.
The red wire is a heating element that is turned on and off according to feedback from the circuit temperature.
Page 23
20
Pinch Tube
Pinch Valve
GAS FLOW THROUGH HUMIDIFIER
(cont.)
Using the CIRCUIT temperature setting
on the humidifier front panel as its set point, the microprocessor controls the
temperature of the gas being delivered to the patient.
The heating wire also minimizes the amount of condensation in the tubing to control “rainout.”
The heated and humidified gas flows into the pinch tube section of the circuit where the pinch valve breaks the flow into breaths.
The gas begins to cool as it leaves the pinch valve and condensation occurs as a result of the cooling.
The temperature of the gas that enters the LifePort adapter will be approximately 3o C less than when it was last measured by the circuit thermistor.
Therefore, the CIRCUIT temperature is
automatically set at 40o C on the front panel by the microprocessor.
The intention is to deliver the gas to the patient at close to 37o C, normal body temperature.
Page 24
21
PURGE
ENTER
GAS OUT
HUMIDIFIER
WAIT
Press WAIT button to stop and start humidification
Press ENTER button to initially start humidifier
STARTING THE HUMIDIFIER
The Life Pulse Humidifier requires little operator intervention. The temperature of the gas is regulated by feedback control from the point where it enters the cartridge to the point where it enters the Patient Box.
When the Life Pulse is first turned on, the ventilator comes up in its Standby mode and
the humidifier is in its WAIT mode. The two
modes are equivalent; there is no ventilation and no humidification or heating being done in these modes.
When the ENTER button is pressed, the light in the corner of the WAIT button goes off and
the humidifier automatically begins functioning. The pump has 86 seconds to fill the cartridge to the proper level.
When the STANDBY button is pressed, the light in the corner of the WAIT button is lit
and the humidifier assumes its WAIT mode.
The humidifier WAIT mode may also be entered independently by pressing the WAIT
button. In this case, the light in the corner of
the WAIT button begins blinking on and off. Press the WAIT button again to bring the
humidifier back into operation and reset the 86-second timer for the water pump,
Page 25
22
CIRCUIT CHANGE
PREPARATIONS
Checklist
With Life Pulse Operating:
Position new circuit, Press WAIT button, Clamp H
2
O transfer tube,
Disconnect H
2
O transfer tube from old
circuit,
Reconnect H
2
O transfer tube to new
circuit,
Disconnect Purge tube from the front
of the Life Pulse,
Unlatch and open cartridge door, and Disconnect purge tube from Patient
Box.
FRONT PANEL DUTIES
Press STANDBY button, Disconnect green gas inlet tube,
Open pump door, Remove used cartridge, Insert new cartridge, Connect green gas inlet tube, Install new water inlet tube into
pump housing,
Open clamp on H
2
O transfer tube
Press ENTER button.
CHANGING THE CIRCUIT
When replacing the circuit, place the humidifier
in its WAIT mode to turn off the circuit and
cartridge heaters until the replacement circuit is installed. This procedure is best performed with 2 people. Both people should perform their tasks simultaneously.
The actual changing of the circuit should be performed with the Life Pulse in the Standby mode, but many steps can be taken to prepare for the actual circuit change as long as the Life Pulse is in the READY condition. These steps are as follows:
1. Lay the new circuit next to the circuit in use.
2. With the ventilator still running, press the
humidifier WAIT button.
3. Clamp off the water transfer tube connecting the water supply with the water inlet tube.
4. Disconnect the water transfer tube from the old cartridge and attach it to the water inlet tube of the new cartridge.
5. Disconnect the Purge tube.
6. Open the cartridge door.
The Life Pulse continues to ventilate even with the cartridge door open, because pressure remains in the cartridge and circuit, and the actual pulsing is done in the Patient Box, which is still connected. Alarms can be silenced as necessary.
FRONT PANEL DUTIES
1. The person attending the Patient Box can disconnect the purge tube from the Purge barbed connector and attach the purge tube from the new circuit.
2. When both operators are ready, press the
STANDBY button to stop the Life Pulse.
Page 26
23
PATIENT BOX DUTIES
Disconnect Life Pulse circuit at ET tube connections,
Remove pinch tube from pinch valve, Insert new pinch tube into pinch
valve, Connect new Life Pulse circuit at ET
tube connections, and
Press ENTER button to resume
ventilation.
POST-CIRCUIT CHANGE
After ENTER button is pressed:
Close and latch cartridge door, Reconnect purge tube at Life Pulse
front panel, Reconnect purge tube at Patient Box,
Make sure Humidifier is not in WAIT
mode, Recheck all connections, and
Press ENTER button again if
necessary.
CHANGING THE CIRCUIT
(cont.)
Once the Life Pulse has been placed in the Standby mode, manually ventilate the patient or adjust the conventional ventilator settings to provide the patient with adequate ventilatory support while the Life Pulse is not running. With two people, one at the ventilator and one at the Patient Box, the circuit change can be performed more quickly.
PATIENT BOX DUTIES
1. The person at the Patient Box disconnects the Jet port of the old Life Pulse circuit from the side of the LifePort adapter.
2. Remove the old pinch tube from the jaws of the pinch valve in the Patient Box and install the pinch tube of the new circuit.
3. Attach the new circuit to the Jet port of the LifePort adapter.
FRONT PANEL DUTIES
1. The person at the ventilator can disconnect the gas inlet tube and remove the cartridge from its holder,
2. Place the new cartridge into the cartridge door and latch it,
3. Connect the green GAS OUT tube and
purge tube to their ports,
4. Install the water inlet tube into the pump housing and latch the door securely; with the pump door latched open the clamp on the water transfer tube.
5. Press the ENTER button to reestablish
high-frequency ventilation and clear a
LOSS OF PIP alarm that may result from
tubing disconnections.
Page 27
24
IMPORTANT
To complete a circuit change:
Observe water filling cartridge,
Observe water pump stopping when cartridge fills to proper level, and
Observe humidity appearing in the green portion of the circuit.
CHANGING THE CIRCUIT
(cont.)
6. After the ventilator is running, make sure the purge tube is attached at both the ventilator and Patient Box barbed connectors.
7. If necessary, adjust the conventional ventilator settings back to their previous settings. Lower the peak pressure first to eliminate any unintentional interruptions of the jet pulses, then lower the CV rate back to where it was before the circuit change.
8. Bring the humidifier out of its Wait mode
by pressing the WAIT button after the Life
Pulse has been restarted with the new cartridge/circuit; otherwise, the patient will receive relatively cold and dry gas.
There are no alarms for low temperature or low water level when the humidifier is in the Wait mode.
The circuit change is not complete until the water fills the cartridge, the water pump shuts off, and humidity appears in green portion of the circuit between the Patient Box and the LifePort.
On rare occasions, a defective cartridge might not fill, might over fill, or might not heat properly. It will then need to be replaced.
Page 28
25
IDENTIFYING PROPER HUMIDIFICATION
It is important to identify proper humidification of the gas through the circuit. You can do this by observing the degree of mist in the green portion of the circuit tubing between the Patient Box and the LifePort adapter.
Proper humidification looks similar to the mist you would see when breathing on a mirror or against a car window on a cold winter day.
Excess humidification exhibits a collection of water pooling continuously in the clear portion of the circuit tubing between the cartridge and the Patient Box. The water may even begin to
march into the patient‟s endotracheal tube.
This condition can be alleviated by lowering
the set CARTRIDGE temperature by 1-2
Insufficient humidification will be indicated by the green circuit tubing being dry.
Examples of proper, over, and under humidification are illustrated on the left.
It is important to see condensation in the green portion of the circuit between the Patient Box and the LifePort adapter. This condensation is an indication that the gas has reached 100% relative humidity.
o
C.
Page 29
26
Objectives
1. Learn how to determine initial settings for high-frequency ventilation.
2. Learn to balance the conventional and high frequency ventilators to achieve better blood gases while using less pressure.
3. Understand the changes in monitored PEEP level that can occur when initiating high-frequency ventilation.
Attach to pressure monitoring port on ‘Patient Box
Attach CV circuit
Attach Life Pulse Circuit
The following procedures describe how to prepare the Life Pulse for start up:
1. Secure the caps on the LifePort adapter.
2. Replace the endotracheal tube adapter with the LifePort adapter once a successful test sequence has been completed, and
while the Life Pulse is still in the Standby mode.
3. Reattach the patient to the conventional ventilator by connecting the conventional circuit to the 15-mm opening of the LifePort.
4. Once the patient has been stabilized on the conventional ventilator, make the LifePort adapter connections to the Patient Box and Life Pulse circuit while the Life Pulse is in the Standby mode.
5. Connect the pressure monitoring tube of the LifePort adapter to the Patient Box and connect the Life Pulse circuit to the Jet port on the side of the LifePort adapter.
Chapter 6
START UP
Page 30
27
10 sec.
2nd 80 sec.
display
1st 80 sec.
display
MEASURING & DISPLAYING PRESSURES
Once the pressure monitoring tube of the LifePort adapter has been connected to the Patient Box, the Life Pulse, in its Standby mode, will begin to monitor the pressures being delivered by the conventional ventilator or other high frequency ventilator and update them in the
MONITOR section every 10 seconds.
It will take about a minute and a half for
the MONITOR to display an accurate PIP, PEEP, and MAP being delivered by the
conventional ventilator or HFOV.
In the Standby mode, the Life Pulse monitors and displays the airway pressures as if measured at the distal tip of the ET tube. These pressures may or may not be different from the pressures displayed on the conventional ventilator.
Remember that the conventional ventilator displays pressures measured proximally while the Life Pulse displays approximations of distal pressures. We recommend that start up decisions be based on the pressures displayed in the Life
Pulse‟s MONITOR section.
Page 31
28
CHOOSING STARTING VALUES
The PIP setting chosen for initiation of high-frequency ventilation will depend on the PIP currently being monitored by the Life Pulse in its Standby mode.
The other initial high-frequency settings are usually left at the default values of a
RATE of 420 breaths per minute and an ON Time of 0.02 seconds.
For larger patients the rate may need to be lowered to avoid gas trapping and inadvertent PEEP (240-360 bpm).
Generally, the more pulmonary airleaks
are a concern, the lower you will set the background CV rate, PIP, and I-Time on the conventional ventilator. PEEP is a better way to control oxygenation in patients with airleaks.
The more atelectasis is a concern, the
higher the background CV Rate, PEEP, and I-Time can be set.
Backgrounds CV rates greater than 10 bpm are almost never indicated.
For more detailed information on choosing starting values, see the next few pages.
Page 32
29
SUMMARY OF SIX STEPS TO START-UP
1. Select the starting HFV PIP value.
2. Press the ENTER button to start the Life Pulse.
3. If necessary, lower PIP on the conventional ventilator to prevent interruption of
high-frequency pulses.
4. Lower the conventional ventilator rate to CPAP to 3 breaths for air leak syndromes, 3 to 10 breaths for RDS.
5. If necessary, adjust PEEP setting: e.g., higher to improve atelectasis and
oxygenation, lower if FiO2 is lower than 30% and oxygenation is appropriate.
6. After settings stabilize, if a MAP alarm occurs, press the RESET button to enable the Life Pulse to recalculate correct alarm limits around the NEW settings.
1. Select Starting PIP Value
2. Press ENTER
6 STEPS TO START HFV
Initiation of high-frequency ventilation involves six steps:
1. Once the monitored pressures are stable,
select a starting PIP specific to the patient‟s
pathophysiology. Input the chosen value by using the increase or decrease arrow next to
the NEW PIP display in the CONTROLS
section.
2. After selecting the NEW settings press the ENTER button to begin high-frequency
ventilation.
As the Life Pulse begins to operate, note that the monitored values return to zero and new average values accumulate based on the new conditions.
Page 33
30
3. Lower Conventional PIP
4. Lower Conventional Rate
6 STEPS TO START HFV
(cont.)
The Life Pulse may pause every time the conventional ventilator delivers a breath. These interruptions may be heard, or seen by
observing the JET VALVE ON/OFF lights.
The interruptions in the Jet pulses are caused by the delivery of conventional breaths at
pressures higher than the PIP that has been
requested by the operator and entered as the
NOW PIP.
Except in cases of extremely poor lung compliance, it is usually best to allow the high-frequency pulses to continue uninter-
rupted by lowering the conventional PIP.
3. If desired, lower the conventional PIP by
slowly turning down the PIP knob to just
below the threshold of interruptions; it may need to be lowered even more when treating infants with pulmonary airleaks.
4. Lower the rate of the conventional
ventilator to 0 -5 breaths per minute once
the conventional PIP is set properly.
A conventional ventilator rate of zero (CPAP)
to 4 breaths per minute may be used in cases of pulmonary air leak.
If oxygenation is still a concern after
optimizing PEEP, provide ample opportunity
for the recruitment of collapsed alveoli by
adjusting CV PIP, I-TIME, or by providing up
to 10 conventional breaths per minute. Lower the CV support (e.g., lower the rate to 1-3 bpm) once collapsed alveoli are recruited and
stabilize them with adequate PEEP.
If both problems are of equal concern after
PEEP has been optimized, start with the
lowest number of CV breaths possible and
modest CV PIP and I-TIME settings.
Page 34
31
5. Readjust PEEP
6. Press RESET
Once high-frequency ventilation is initiated, the displayed PEEP may be slightly different than
what is desired. Although PEEP is controlled by
the conventional ventilator, high-frequency ventilation may cause it to rise or fall.
5. Adjust the conventional ventilator PEEP
knob and/or flow rate to bring it to the level desired. Remember that average values are displayed, so give the Life Pulse 20 seconds between adjustments to indicate the true
PEEP value.
Be aware of significant changes in Servo and
mean airway pressure (MAP) that may occur
when manipulating the conventional ventilator settings. For example, a pressure change of 2 cm H
2
O in the PEEP setting will cause a change
of about 2 cm H
2
O in the MAP.
6. If changing PEEP produces a MAP or SERVO PRESS alarm, press the RESET button to
accommodate the change. Or, the limits may be changed manually.
Pressing the RESET button allows the limits around SERVO PRESS and MAP to be recalculated
and new alarm limits to be set.
The READY light will turn off when the RESET
button is pressed. As soon as monitored values have been stable for 20 seconds, new limits will be
set and the READY light will illuminate.
Always wait until the READY light is ON before leaving
the patient’s bedside.
6 STEPS TO START HFV
(cont.)
Page 35
32
Objectives
1. Understand the advantages for patient management of using a conventional
ventilator in tandem with the high-frequency ventilator.
2. Understand the techniques for improving oxygenation and ventilation.
3. Comprehend the relationship between P and tidal volume.
Chapter 7
PATIENT MANAGEMENT
Managing patients on high-frequency ventilation is similar to managing patients on a conventional ventilator.
The main distinction with the Life Pulse is that typically less pressure and much smaller tidal volumes are used to manage the patient.
The conventional ventilator settings will be manipulated most often when oxygenation of the patient is of primary concern.
The Life Pulse settings will be manipulated most often when ventilation (CO2 removal) and/or the consequences of using high airway pressures (e.g., pulmonary air leaks)
are of greatest concern.
Page 36
33
1. Increase PEEP
2. Increase CV Rate
2. Increase CV Rate
OXYGENATION
The main choices for improving oxygenation require increasing mean airway pressure by elevating the:
CV PEEP CV RATE CV PIP CV I-Time
High-frequency PIP and rate would be secondary considerations. Raising high­frequency On-Time has rarely been shown to be effective in clinical trials.
The choices for improving oxygenation due to atelectasis are:
Increases in PEEP are meant to stabilize alveoli.
PEEP changes are made with the conventional ventilator since the Life Pulse has no PEEP control. However, the PEEP adjustment will be displayed on the Life
Pulse in the MONITOR, PEEP display.
Increase background rate from the conventional ventilator. Do not exceed 10 bpm on the background rate. If more CV breaths are needed to oxygenate, it may be an indication that the PEEP is too low.
Always optimize PEEP before increasing CV Rate, PIP, or I-time.
Increase the PIP delivered with the background conventional breaths. Increases in CV PIP are meant to reach the critical opening pressure required to inflate collapsed alveoli.
Adequate PEEP levels are essential for avoiding derecruitment between
conventional breaths.
Page 37
34
4. Increase CV I-Time
OXYGENATION
(cont.)
Increase I-Time, in combination with adequate levels of PEEP and PIP, to
improve atelectasis.
Consider carefully the combined effect of PIP and I-Time increases. Increasing I-Time when CV PIP is set at high levels increases the risks of causing lung injury.
If other approaches to oxygenation have
failed or are contrary to the patient‟s
pathophysiology, increase the Life Pulse PIP by 2 cm H2O at a time until the desired response has been achieved.
Increasing HFV PIP too high may result in hyperventilation and hypocarbia which, in preterm infants may increase the risk of cerebral injuries. Concomitant increases in PEEP may help maintain an appropriate tidal volume and reduce this risk.
If necessary, the conventional ventilator PIP can be increased along with the Life Pulse peak pressure except in severe cases of air leak. However, the CV PIP should
not be raised if it is at an adequate level to reach the critical opening
pressure of the alveoli.
Remember to keep the conventional PIP below
the Life Pulse‟s peak pressure to avoid
interrupting the high-frequency pulses.
Additional possibilities for increasing mean airway pressure are available, but they have not been studied in prospective clinical trials. Increasing high-frequency rate by 50 or 60 breaths per minute at a time has been helpful in some cases, especially in smaller infants.
If the high-frequency rate is increased, be sure to watch the PEEP level. Inadvertent PEEP may develop as the I:E ratio is shortened.
Page 38
35
EXCEPTION!
Overexpanded Lungs
OXYGENATION
Overexpanded Lungs
There is one major exception to this strategy. This exception arises when the patient on conventional ventilation has grossly overexpanded lungs.
If overexpansion is observed on X-ray, the lungs will need to deflate considerably before any improvements in oxygenation will result.
To accomplish this deflation, set the conventional ventilator rate near zero when starting the Life Pulse.
In most cases, DO NOT DECREASE PEEP. Overexpanded lungs are usually a result of gas trapping, not excessive PEEP. Decreasing CV support (Rate, PIP, and I­Time) is usually a more effective strategy. PEEP must be maintained, or even increased, when the CV rate is very low to prevent atelectasis and maintain oxygenation.
However, beware that if the patient initially responds well to this strategy, poor oxygenation may result some time later due to atelectasis. You must be ready to treat that condition as outlined above.
A strategy for identifying optimal PEEP is demonstrated in the flow chart on the next page. It is always important to optimize PEEP regardless of what pathophysiology is being treated.
Page 39
36
Switch to HFV from CV at same MAP by
adjusting PEEP.
Reduce IMV Rate to 5 bpm.
Note current SaO2 on pulse oximeter.
Finding Optimal PEEP during HFV *
Switch CV to CPAP mode.
Does SaO2 drop?
(Wait 1 - 5 min.)
PEEP is too low.
Switch back to IMV
Increase PEEP by 1 - 2
Wait for SaO2 to return
to acceptable value.
(It may take 30 min.)
PEEP is high enough, for the moment.
(hours later)
Use IMV = 0 - 3 bpm with
IMV PIP 20 - 50% < HFV PIP
Does FiO
2
needs to be increased?
Keep PEEP at this level
until FiO2 < 0.30
Don't be shocked if optimal
PEEP = 8 - 12 cm H2O!
1-800-800-4358 www.bunl.com
* when switching from CV to HFV.
Warnings: Lowering PEEP may improve SaO2 in some cases. Optimal PEEP may be lower in patients with active air leaks or hemodynamic problems.
Using IMV with high PEEP is hazardous. Do not assume high PEEP causes over-expansion.
Yes
No
No
Yes
FINDING OPTIMAL PEEP
Page 40
37
1. Increase HFV PIP
VENTILATION
Manipulating the patient‟s arterial PCO2 is
one of the easier tasks when using the Life Pulse. Studies have shown that ventilation (CO2 elimination) during high-frequency ventilation is proportional to the tidal volume squared ( V
2
).
T
Tidal volume on the Life Pulse is roughly proportional to delta P P), the arithmetic difference between PIP and PEEP. Thus, small changes in PIP or PEEP can produce significant changes in a patient‟s PCO2.
The main choices for improving ventilation require increasing minute ventilation by changing:
HFV PIP PEEP HFV Rate
If the patient‟s PO
is acceptable, but his
2
PCO2 is too high, increase high-frequency PIP by 1 to 2 cm H2O at a time.
If the patient‟s PO
and PCO2 are both
2
unacceptable, increasing PIP may address both problems at once.
When PEEP is increased, oxygenation may improve. However, increasing PEEP without a corresponding increase in HFV PIP reduces tidal volume and may result in some degree of CO2 retention.
The PIP must be increased by an equal amount in order to keep the delta P ( P) the same and maintain tidal volume and adequate ventilation.
Page 41
38
2. Decrease PEEP
3. Increase HFV Rate
4. Hyperventilation
VENTILATION
(cont.)
If the primary concern is air leaks or
cardiac compromise and oxygenation is acceptable, lower PEEP in order to increase ΔP (i.e., tidal volume) and increase ventilation.
Be careful not to compromise oxygenation when lowering PEEP.
Increase the high-frequency rate by 40 to
80 breaths per minute at a time to improve ventilation without further increasing PIP.
Increasing high-frequency rate is much less effective than increasing delta P (ΔP) for reducing PCO
.
2
Changes in CV Rate, PIP, and, to a lesser degree, I-Time may also effect HFV minute ventilation. Raising CV Rate or PIP seldom improves ventilation, but it may be helpful in extreme cases if HFV is not interrupted.
The main choices for raising PCO
are, of
2
course, doing the opposite of the above suggestions for lowering PCO
.
2
Decreasing HFV PIP is the most effective way to increase PCO
. However, the
2
concomitant drop in mean airway pressure may cause PO
to fall. Thus, it is very
2
important to raise PEEP in such cases to maintain adequate mean airway pressure.
Reducing HFV Rate will also increase PCO unless inadvertent PEEP is present. If monitored PEEP falls when HFV Rate is dropped, PEEP may need to be increased to maintain adequate oxygenation.
Do not hesitate to use the minimum Life Pulse HFV Rate of 240 bpm when indicated for hyperinflation. The Life Pulse can provide adequate ventilation over its entire range of HFV Rates (240 - 660 bpm).
2
Page 42
39
UNDERSTANDING SERVO PRESSURE
Understanding Servo Pressure, what it is, why it changes, how it changes, can indicate when a patients condition is improving or worsening. The diagram and text below may be helpful in understanding Servo Pressure and how it can help manage patients.
Servo Pressure = driving pressure that automatically regulates flow. Servo Pressure changes as lung volume or mechanics change
Page 43
40
SUCTIONING
3 Techniques
1. Suction with Life Pulse in Standby
mode
2. Suction with Life Pulse running
3. Suction with in-line suction system
Objectives
1. Learn three methods of suctioning a patient while on high-frequency
ventilation, which technique is preferred, and when to use which alternative.
2. Be ready for problems that may possibly be encountered when using each
suctioning method.
Chapter 8
SUCTIONING THE PATIENT
High-frequency ventilation may mobilize and help remove secretions.
Be prepared to suction soon after starting the Life Pulse on a patient.
Suctioning may need to be performed more frequently in the first 4 to 6 hours. Suctioning frequency may then subside.
Suctioning the patient‟s airway may be
accomplished by either of three methods:
with the Life Pulse in the Standby mode
with the Life Pulse running
using an in-line suction catheter
system
Page 44
41
Suctioning in the Life Pulse’s
STANDBY Mode
SUCTION PROCEDURE #1
The first technique for suctioning the patient is the easiest to learn because the procedure is similar to suctioning during conventional ventilation.
1. Place the Life Pulse into the Standby mode.
2. Suction as usual.
2. Press the ENTER button after reattaching
the conventional ventilator circuit, when the suction procedure is complete.
Page 45
42
Suctioning with the Life Pulse
Running
SUCTIONING CHECKLIST
READY light must be on!
Instill into Jet port then reconnect Life Pulse circuit.
One person disconnect CV circuit from LifePort.
Second person applies suction
going into and out of the ET Tube.
Reconnect CV circuit to LifePort.
Press SILENCE button, if necessary
SUCTION PROCEDURE #2
The second technique requires that suction be applied in the endotracheal tube throughout the introduction and withdrawal of the suction catheter:
1. First, make sure the Life Pulse’s
READY light is ON before beginning the
suction procedure.
2. Press the alarm SILENCE buttons on
both ventilators.
3. With the Life Pulse running, disconnect the Life Pulse circuit from the Jet port of the LifePort adapter.
4. Instill irrigation fluid into the Jet port.
5. Reattach the Life Pulse circuit to the Jet port to “jet” the fluid into the patient‟s airways.
6. One person may disconnect the conventional circuit from the 15-mm connection of the LifePort while a second person prepares to introduce the suction catheter.
7. Introduce the catheter with suction applied.
This procedure allows the Life Pulse to continue ventilating the patient with fewer interruptions. If suction is not applied while advancing the suction catheter, the catheter creates an obstruction and the Life Pulse will pause to protect the patient; the Life Pulse will not deliver gas if the exhalation path is obstructed by the suction catheter.
Suctioning all the way into the ET tube and all the way back out allows the patient to receive some jet ventilation throughout the suction procedure.
Page 46
43
SILENCE
ENTER
LOSS OF PIP
READY
If suctioning with the Life Pulse Running, MAKE SURE THE READY LIGHT IS ON!
If a LOSS OF PIP alarm occurs with the READY light on, the Servo Pressure locks to protect the patient.
SUCTION PROCEDURE #2
(cont.)
A LOSS OF PIP alarm usually occurs when
suctioning with the Life Pulse running. As
long as the Life Pulse is in the READY
condition, the Servo Pressure will lock at or near its operating level and the patient will continue to receive appropriate ventilation even though the displayed PIP and PEEP may fluctuate.
8. Reconnect the conventional ventilator
when finished.
9. Provide a few manual breaths with the
conventional ventilator to help the patient recover from the procedure.
10. Press the SILENCE button to stop
audible alarm.
11. If necessary, press the ENTER button to
clear all alarms and reestablish appropriate ventilation and new alarm settings.
If pressure fluctuations continue after the suctioning procedure, it may be necessary to
press the ENTER button again. This action
will activate the Jet‟s purge system to clear
the pressure monitoring tubing of fluids.
If problems persist with pressure
fluctuations and the CANNOT MEET PIP
alarm is displayed, it may be necessary to suction again to remove secretions from the distal tip of the ET tube.
Page 47
44
SUCTIONING CHECKLIST
Select an in-line suction system
compatible with the LifePort adapter
Select an in-line suction system
with a catheter that is straight for the first few inches
Suction according to procedure 1
or 2 described earlier in this section
Call the Bunnell Hotline for more
information or assistance.
Suctioning with an In-Line
Suction System
SUCTION PROCEDURE #3
The third technique may be used in conjunction with techniques 1 and 2. The key to success using this procedure is to choose a compatible in-line suction system.
The two most important factors in choosing an in-line system are:
how the system connects to the LifePort
adapter
An in-line suction system that uses a special adapter which replaces the conventional ET tube adapter cannot be used with the Life Pulse. The Life Pulse requires the LifePort adapter in order to operate. For example, The Ballard
“Neonatal Elbow” system can be use with the Life Pulse; the Ballard “Neonatal „Y‟ ”
system cannot.
if the suction catheter is straight for the
first few inches at its tip
A straight tip on the in-line suction catheter
ensures a “straight shot” into the
endotracheal tube. A curved suction catheter may dead-end against the inner wall of the LifePort adapter making it difficult to advance the catheter down the ET tube.
Once a compatible inline suction system has been selected, it can be used according to the procedures outlined in procedures 1 and 2 described earlier in this section.
Page 48
45
Objectives:
1. Recognize the indications for weaning from the Life Pulse.
2. Know the various options for weaning and the advantages of each.
GENERAL GUIDELINES FOR WEANING
Decrease Minute Ventilation and MAP slowly by lowering PIP on the Life Pulse
and the conventional ventilator.
Lower F
I
O2 gradually to around 30%.
Continue decreasing PIP as blood gases allow, weaning slowly (1-2 cm H
2
O of PIP), unless hyperventilation is occurring, in which case PIP should be weaned faster.
As the PIP is lowered into the teens, begin to allow the conventional ventilator
breaths to interrupt the high-frequency pulses.
As the high-frequency PIP continues to be decreased, increase the conventional
rate as needed to achieve good blood gases.
Place the Life Pulse into the Standby mode.
Chapter 9
WEANING
As the patient improves, he will eventually need to be weaned from the Life Pulse. The goal in most cases will be to wean the patient back to conventional ventilation at much less support than the patient was on before beginning high-frequency ventilation.
However, the Life Pulse may be left on the patient while he is weaned directly to CPAP, a nasal cannula, or an oxygen hood.
Below is a brief summary of weaning guidelines:
Page 49
46
Wean SLOWLY!
Find Optimal MAP
Lower HFV PIP
Adjust PEEP if necessary
Lower FiO
2
WEANING
(cont.)
Wean slowly: The fundamental rule in
weaning is to WEAN SLOWLY, to wean as slowly as the patient allows. The Life Pulse is a very gentle form of ventilation, much more so than conventional ventilation. Therefore, weaning a patient back to conventional ventilation too soon is ill-advised.
Find Optimal MAP: During patient
management, the MAP should be kept as
low as the patient‟s pathophysiology
allows without compromising oxygenation.
As the patient‟s condition improves, blood gases will indicate that pressures can be weaned.
Lower Life Pulse PIP: As weaning
begins in earnest, reduce PIP in increments of only 1 to 2 cm H2O unless PCO2 is below 35 torr.
Small changes in tidal volume have very dramatic results during high-frequency ventilation whether you are increasing or decreasing support. Lower the conventional PIP just enough to avoid interruptions of the high-frequency pulses when these brief interruptions begin to occur.
Adjust PEEP: The PEEP may be
adjusted as necessary to maintain adequate PO2, alveolar inflation, and MAP (which may decrease as PIP is weaned). Do not worry about lowering PEEP until FiO2 is .30 or less.
Lower FiO
weaning of FiO2 when you are comfortable with the patient‟s MAP. Remember to adjust both blenders so the settings remain equal.
: Begin more aggressive
2
Page 50
47
Resume Lowering HFV PIP
Allow Jet Pulse Interruptions
Add CV Support as Necessary
Begin CV Trial by Placing
Life Pulse in STANDBY
Weaning Directly to CPAP or
Nasal CPAP
WEANING
(cont.)
Resume weaning PIP: Do so as tolerated.
If the patient begins to deteriorate with further reductions in PIP, the pressure may be near a mean airway pressure threshold. In this case, cease weaning and consider raising PEEP to stabilize the lungs. Resume weaning when the patient is ready as determined by blood gases and clinical observation. If possible, you may continue weaning FiO2 during the interim.
Allow Jet Pulse Interruptions: When the high-
frequency PIP has been lowered into the mid-teens, allow the conventional ventilator breaths to interrupt the high-frequency pulses (i.e. don’t lower CV PIP further).
Add CV Support: provide more CV support by
increasing the conventional ventilator’s IMV rate as
high-frequency PIP is weaned.
Once the high-frequency PIP is down to approximately 15 cm H2O or less, and the conventional rate has been increased to about 15-20 bpm, the patient’s chest rise will be much more a result of the conventional breaths than the high-frequency pulses.
Attempt CV Trial: Place the Life Pulse into the
Standby mode to begin a trial of conventional ventilation only. The CV PIP should not have to be raised above 20 cm H2O. If this is necessary, the trial has failed. Return to HFV.
If the patient is stable on conventional ventilation,
disconnect and cap the Jet port and the pressure monitoring tube of the LifePort adapter.
Weaning to Nasal CPAP: If the patient is stable on
low Life Pulse and background CV settings, and the patient is breathing on his own, try the patient on CPAP. If tolerated, you may want to extubate the patient to nasal CPAP.
Remove the Life Pulse for cleaning and preparation for
the next patient.
Page 51
48
Objectives
1. Understand how the High and Low alarm limits are set for the Servo Pressure and Mean Airway Pressure.
2. Learn how and when to change the High and Low alarm limits.
3. Know the possible causes of the six types of alarm messages and how to troubleshoot and correct them.
READY Light ON
1. Monitored PIP + 1.5 cm H
2
O of
set NOW PIP for 20 Seconds.
2. SERVO PRESSURE alarm limits
vary according to size of patient.
3. MAP alarms set + 1.5 cm H
2
O.
Chapter 10
VENTILATOR ALARMS
The Life Pulse alarm system alerts the operator, both audibly and visually, to changes in the ventilator or the patient. The alarm statements are not visible until they are lit.
The ALARMS area has three key features:
upper and lower alarm limits for Servo
and Mean Airway Pressure (MAP);
alarm messages for various potentially
hazardous conditions, and;
an alarm SILENCE button.
Both the upper and lower alarm limits are set automatically and can be adjusted manually.
After the ENTER button has been pressed, the
Servo Pressure rises to bring the monitored
PIP up to the NOW PIP setting.
The alarm limits are set automatically when
the READY light comes on indicating that the monitored PIP has come to within ± 1.5 cm
H
O of the NOW PIP and has stabilized there
2
for at least 20 seconds.
The instant the READY light comes on, the
limits around the current Servo Pressure are set and vary according to the size of the patient; wider limits are set for larger patients and tighter limits for smaller patients.
Page 52
49
SILENCE
ENTER
JET VALVE FAULT
VENTILATOR FAULT
LOW GAS PRESSURE
CANNOT MEET PIP
LOSS OF PIP
HIGH PIP
ALARM LIMITS
The limits around the current mean airway pressure (MAP) are set at +1.5 cm H2O.
Press the various limit buttons any time after
the READY light is lit to observe where the
limits have been set.
If the READY light is not lit then the limits have not yet been set, and the SERVO PRESS and MAP displays will not change when you
press one of the limit buttons.
To change the limits and make them tighter or wider in any combination, press the button for the particular limit, hold it down, and press either the increase or decrease button next to it.
The other alarms that are available with the Life Pulse are illuminated only to indicate changing or potentially threatening conditions.
While in the Standby mode, press the TEST
button to observe a display of these alarms.
Do NOT perform the Test while a patient is connected to the Life Pulse.
Alarm messages include:
JET VALVE FAULT VENTILATOR FAULT LOW GAS PRESSURE CANNOT MEET PIP LOSS OF PIP HIGH PIP
Page 53
50
SILENCE
ENTER
JET VALVE FAULT
JET VALVE FAULT
JET VALVE FAULT
A JET VALVE FAULT alarm implies that the pinch valve is out of synchrony with the electrical drive signal. A JET VALVE FAULT alarm will appear in the ALARMS section and the Life Pulse and Patient Box will continue running.
This alarm is extremely rare; if it occurs call the Bunnell Hotline at 1-800-800-4358 for troubleshooting support. The JET VALVE FAULT alarm only applies to electrical failures. If the pinch valve fails mechanically, it will stop cycling.
The following precautions should be followed if the “WhisperJet” Patient Box (Cat # 312) pinch valve stops cycling while on a patient.
In the READY condition:
1. A Loss of PIP alarm will be activated;
check for chest vibration.
2. Do not press the Enter or Reset buttons.
3. If chest is vibrating, do normal
troubleshooting for Loss of PIP alarm (see Operator’s Manual).
4. If chest is not vibrating, check the pinch
valve.
5. If pinch valve is NOT CYCLING, press
the STANDBY button to dump the Servo Pressure and change out the “WhisperJet” Patient Box.
6. Call Bunnell Hotline to report stoppage, get
RA#, and send “WhisperJet” Patient Box to Bunnell for service.
In the Non-READY condition:
1. A Loss of PIP alarm will be activated;
verify pinch valve is cycling.
2. Do not press the Enter or Reset buttons.
3. If pinch valve is cycling, do normal
troubleshooting for a Loss of PIP alarm (see Operator’s Manual).
4. If pinch valve is NOT CYCLING, press
the STANDBY button to dump Servo Pressure and change out “WhisperJet” Patient Box.
5. Call Bunnell Hotline to report stoppage, get
RA#, and send “WhisperJet” Patient Box to Bunnell for service.
Page 54
51
SILENCE
ENTER
VENTILATOR FAULT
VENTILATOR FAULT
VENTILATOR FAULT:
HFV CONTINUES RUNNING
A VENTILATOR FAULT message may or
may not be serious. If the Life Pulse has a serious Ventilator Fault, it will
automatically put itself in the STANDBY
mode with the audible alarm sounding.
Ventilator Fault alarms that place the Life Pulse in Standby will display a code number
in the ON/OFF display window of the CONTROLS section to let the operator or
service rep know where the fault occurred.
If the Life Pulse continues to run with the
VENTILATOR FAULT message, possible
causes are:
One of the ends of the purge tube has
become disconnected, either at the
PURGE connector on the front panel, or at the FROM PURGE connector on the
Patient Box.
The purge tube is kinked or obstructed.
In any case, with the Life Pulse in its Ready condition, the Life Pulse will continue operating because the purge function is not critical for supporting the patient.
Page 55
52
SILENCE
ENTER
VENTILATOR FAULT
TESTSTANDBY
JET VALVE TIMEonon/off
ENTER
seconds
05
VENTILATOR FAULT:
HFV STOPS RUNNING
If the VENTILATOR FAULT places the Life
Pulse into Standby mode and a code number
appears in the ON/OFF window, take the
following steps to determine if the alarm is true:
1. Disconnect the patient from the Life Pulse.
2. Provide ventilatory support to the patient using the conventional ventilator.
3. Turn off the power to the Life Pulse for at least three minutes.
4. Turn on the power, attach a test lung to a Life Pulse circuit, and perform an internal
test by pressing the TEST button.
5. Perform an operational check of the Life Pulse on a test lung with the 15 mm LifePort connector open to the room.
6 If the Life Pulse reaches the set pressure,
the READY light illuminates, and a VENT FAULT alarm doesn‟t occur, it is safe to
resume ventilation.
7. If the Test procedure fails, call the Bunnell
Hotline at 1-800-800-4358.
Page 56
53
TESTSTANDBY
JET VALVE TIMEonon/off
ENTER
seconds
10
SILENCE
ENTER
VENTILATOR FAULT
VENTILATOR FAULT
Code 10
VENTILATOR FAULT
CODE 10
There is one Ventilator Fault that may cause
the Life Pulse to revert to the STANDBY mode
but is immediately recoverable.
If the Servo Pressure rises 3.4 psi (23.44 kPa) above the level necessary to have met the
READY condition, the Life Pulse will revert to STANDBY mode with a VENTILATOR FAULT
and a code 10 displayed in the ON/OFF ratio display in the CONTROL section.
After the condition that caused the alarm (e.g., a kinked pressure monitoring tube) is
corrected, press the ENTER button and the Life Pulse will resume normal operation.
If the VENTILATOR FAULT 10 was caused by a
problem not immediately correctable (e.g., a faulty Servo Pressure control valve), the Life
Pulse, after the ENTER button is pressed, will immediately enter the STANDBY mode and display another VENTILATOR FAULT 10.
If necessary, call the Bunnell Hotline at
1-800-800-4358.
Page 57
54
SILENCE
ENTER
LOW GAS PRESSURE
LOW GAS PRESSURE
LOW GAS PRESSURE
A LOW GAS PRESS alarm indicates that
the gas supply to the Life Pulse has fallen
to a pressure less than 30 psi (206.85 kPa). This alarm could indicate that:
You have a leak in one of the gas supply
lines, the blender, or a failure in the
hospital‟s gas supply system, or
The gas pressure switch in the Life Pulse
is faulty. The Life Pulse will otherwise function properly but will display a
continuous LOW GAS PRESS alarm.
If necessary, call the Bunnell Hotline at 1-800-800-4358
Page 58
55
SILENCE
ENTER
CANNOT MEET PIP
CANNOT MEET PIP
CANNOT MEET PIP
Causes
1. Ready condition not met within 3
minutes
2. Servo Pressure reaches 20 psi (137.90 kPa) before the Ready
condition is met
CANNOT MEET PIP
A CANNOT MEET PIP alarm means one of
two things:
the Life Pulse has been unable to meet
the Ready condition within 3 minutes of pressing the RESET or ENTER button;
or,
the Servo Pressure has risen to 20 psi
(137.90 kPa) and the Ready condition
has not been met.
The first criteria would be met if conditions
keep changing after the RESET or ENTER
button is pressed. An unstable monitored PIP makes it difficult for the Life Pulse to meet the criteria necessary to enter the
Ready condition.
For example, a patient who is taking vigorous spontaneous breaths may initiate this alarm. The alarm may be alerting you that the patient needs comforting, attention, or possibly sedation.
This condition can be simulated with a test lung by periodically creating leaks after
pressing the ENTER button. The CANNOT MEET PIP alarm will occur after 3 minutes.
The second criteria may be met with a relatively large patient on the Life Pulse, or if large leaks are present somewhere in the system. This condition can be simulated by using a test lung with a large air leak.
In either case, the Life Pulse does not give
up trying to meet the Ready condition. This
alarm is simply informing you that the Life Pulse is taking longer than usual to meet the
Ready condition.
If necessary, call the Bunnell Hotline at
1-800-800-4358
Page 59
56
SILENCE
ENTER
LOSS OF PIP
LOSS OF PIP
Causes
1. 25% drop in monitored PIP.
2. Monitored PIP < 3 cm H2O.
3. Monitored PIP and PEEP within 2 cm H2O.
LOSS OF PIP
If necessary, call the Bunnell Hotline at
1-800-800-4358
LOSS OF PIP
The LOSS OF PIP alarm implies an extubation
or a disconnected, obstructed, or kinked tube. The alarm is initiated by one of the following criteria:
The monitored PIP drops below 25% of the
NOW PIP. The monitored PIP is less than 3 cm H The monitored PIP and PEEP are within 2
cm H2O of each other.
The Life Pulse responds to a LOSS OF PIP
alarm in the Non-READY condition as follows:
The Servo valves close. The Servo Pressure and PIP will drop to or
near zero.
The patient will not be receiving
adequate ventilation!
This response is designed to stop gas flow into the patient circuit during start-up, or when changing settings, if inadequate PIP is detected. It also prevents pressure spikes when the Life Pulse circuit is disconnected or kinked, then reconnected or unkinked while
the READY light is off. Gas flow is easily restarted by pressing the
ENTER button. If the LOSS OF PIP alarm occurs in the Ready
condition, the Servo valves are designed to lock and allow the Life Pulse to continue to ventilate the patient with nearly constant tidal volumes.
You should NOT press the ENTER button if the
Servo Pressure display is locked at or near the
established operating level and the patient‟s
level of ventilation is adequate. Eliminating
the cause of LOSS OF PIP will eliminate the
alarm condition.
2
O.
Page 60
57
SILENCE
ENTER
HIGH PIP
HIGH PIP
HIGH PIP
Causes
1. Pressure 5 cm > NOW PIP for 1 sec.
2. PIP 10 cm > NOW PIP for each breath over a 30-second period
3. Monitored PIP > NOW PIP by
30 cm for each breath during a
0.75 sec. period.
4. Instantaneous airway pressure
> 65 cm H2O
HIGH PIP
The HIGH PIP alarm indicates one of the
following conditions:
the monitored pressure has exceeded the
set NOW PIP by at least 5 cm H
second;
the PIP has consistently exceeded the
NOW PIP by 10 cm H
frequency breaths for the past 30 seconds; or,
the monitored PIP for each breath during
a 0.75 second period exceeds the set point by 30 cm H2O, or,
Instantaneous airway pressure > 65
O for all high-
2
cm H2O
This alarm may be observed by pinching off the exhalation limb of the conventional circuit. Note that the Life Pulse Servo Pressure is “dumped” in this condition. Although it has no way of alleviating a condition caused by the conventional ventilator, the Life Pulse does ensure that it will not further aggravate the problem.
All of the alarm conditions described in this section are accompanied by an audible alarm. While the cause of the alarm is being evaluated, the beeping may be silenced by
pressing the Alarm SILENCE button.
It may be necessary to review this section several times before becoming familiar with the conditions that trigger each alarm.
Remember, all alarms are important and should be responded to by the operator.
If necessary, call the Bunnell Hotline at
1-800-800-4358.
O for 1
2
Page 61
58
SILENCE
ENTER
JET VALVE FAULT
VENTILATOR FAULT
LOW GAS PRESSURE
CANNOT MEET PIP
LOSS OF PIP
HIGH PIP
Objectives
1. Understand the usefulness of ventilator alarms.
2. Know the “Five Principles” of ventilator troubleshooting.
Chapter 11
INTERPRETING VENTILATOR ALARMS
ALARM DISPLAY
This section discusses interpreting and correcting the ventilator alarms. It is beyond the scope of this manual to list all possible problems and their solutions.
Many ventilator problems and corrective actions have already been covered. A more comprehensive review is offered in the
Operator‟s and Service manuals. There are,
however, a few general principles that can simplify Life Pulse troubleshooting.
Most of the necessary troubleshooting will be in response to alarms. Ventilator alarms alert the operator to changes in the patient or the Life Pulse. It is recommended that when a patient is on the Life Pulse the operator:
pays particular attention to ventilator
alarms; and,
sets the alarm volume loud enough to
present a sense of urgency when it sounds
There are at least five basic troubleshooting principles that, if followed, will help make working with patients on the Life Pulse
easier.
Page 62
59
1. Look At The Patient
LOOK AT THE PATIENT
First, as with any ventilator, observe the patient. Whenever an alarm occurs, LOOK AT THE PATIENT FIRST.
Is the patient‟s chest rise adequate? How is the patient‟s color? How is the patient‟s external monitoring? Is there an obvious disconnected tube
near the patient?
Has the Servo Pressure locked at or near
its operating level? With the most
common Life Pulse alarm, LOSS OF PIP,
the Servo Pressure will lock and the patient will continue to receive a constant level of ventilation.
The cause of the alarm may then be determined without increasing conventional ventilator support or providing manual ventilation.
However, some alarm conditions may require the conventional ventilator settings to be increased to provide ventilation while an alarm condition is corrected, or the patient may need to be hand bagged.
Page 63
60
2. Use Common Sense
USE COMMON SENSE
Second, USE COMMON SENSE. There is no
substitute for a skilled and alert therapist, nurse, or doctor who takes a logical approach to alarm interpretation.
The Life Pulse‟s rhythmic sounds will
become familiar to you. Learn to recognize when the sounds have changed.
The displayed and monitored values
should remain at consistent levels. Watch for changes in these levels and learn to understand what the changes mean.
For example, the Servo Pressure value reflects the gas flow and pressure required by the Life
Pulse to produce the NOW settings on a
patient. If the Servo Pressure changes, either suddenly or gradually, it may be that:
Lung compliance is changing. A pneumothorax has occurred. Tension is developing on a pneumothorax. A right mainstem intubation has occurred. The patient needs suctioning. There is a leak in the tubing.
Not all alarms require emergency corrective actions. Use common sense. If the Life Pulse alarms are saying one thing, and clinical observation says another, react accordingly.
For example, if the humidifier displays a
TEMP LOW alarm and the circuit and
cartridge feels hot to the touch, place the
humidifier into WAIT and replace the
cartridge/circuit.
Page 64
61
3. Know The Life Pulse
KNOW THE LIFE PULSE
Third, KNOW THE LIFE PULSE. Interpreting alarms is difficult, if not impossible, without a good working knowledge of the ventilator, how the feedback control mechanisms function, and what the various alarms mean.
What conditions can cause an alarm?
How does the Life Pulse respond to an
alarm?
What should the clinical response be?
Review this manual and study the Operator‟s
Manual which, among other information, contains a comprehensive troubleshooting section.
Interpreting alarm conditions is made easier
when you know, for example, the HIGH PIP
alarm means that the Life Pulse has sensed PIPs 10 cm H2O greater than the set PIP for every breath during the last 30 seconds, or a sustained PIP 5 cm H2O greater for 1 second.
Page 65
62
4. Common Things Happen Commonly
COMMON THINGS HAPPEN COMMONLY
Fourth, common things are the most likely cause of the majority of alarms. COMMON THINGS HAPPEN COMMONLY.
Perhaps ninety-five percent of alarms occur
due to kinked, disconnected, or obstructed tubes. Knowing this information will help greatly in troubleshooting.
The solutions to these problems are equally as common:
Unkink the tubes, reconnect the tubes, or
clear the obstruction.
Again, if the Servo Pressure has locked at or near its last operating level, the tubing problem can be corrected without manual ventilation or changing conventional ventilator settings.
Page 66
63
5. Use Available Resources
Examples of
Available Resources
In-service Manual Operator‟s Manual Service Manual Clinical Updates Newsletters Website - www.bunl.com 24 Hour Hotline
USE AVAILABLE RESOURCES
Fifth, USE AVAILABLE RESOURCES.
Bunnell Incorporated offers an Operator‟s
Manual, newsletters, research articles, and a 24 hour Hotline:
1-800-800-4358
When using the Hotline, make sure to gather as much information about the alarm condition as possible. The more facts you have about the situation, the better the assistance a clinical specialist can provide on the Hotline.
If an alarm condition cannot be corrected easily, perform a systems Test. The systems Test can help isolate the source of the alarm. If the Life Pulse passes the Test on the test lung, the cause of the alarm has to be in the ET tube or the patient.
Also run an operational test on a test lung to verify the Life Pulse‟s operation.
Remember to call as soon as possible after determining the alarm condition cannot be corrected.
Do not hesitate to call if there are any question or concern. It is much better to call and have a good experience than not call and have a bad experience.
Page 67
64
WAIT
HUMIDIFIER
TEMPERATURE
SET
CIRCUIT
CARTRIDGE
CIRCUIT TEMP
SILENCE
CIRCUIT
LEVEL
TEMP
FAULT
LOW
HIGH
ALARMS
HIGH FREQUENCY VENTILATOR
LIFE PULSE
Objectives
1. Recognize and correct the conditions that will produce humidifier cartridge/circuit alarms.
2. Understand the purpose and methods of controlling gas temperature in the cartridge and circuit.
3. Understand how cartridge temperature is linked with humidification and when and how to change this temperature.
4. Understand the importance of the CIRCUIT TEMP, CARTRIDGE TEMP, WAIT button, and SILENCE button.
Chapter 12
HUMIDIFIER ALARMS
The humidifier alarms are on the right side of the humidifier control panel.
Humidifier alarms will appear as red LEDs accompanied by an audible alarm. The alarm can be silenced for 60 seconds by
pressing the SILENCE button.
The humidifier alarm system detects high or low water levels and high and low temperature for both the circuit and cartridge.
The system also has an alarm for disrupted electrical connections between the humidifier and the cartridge.
Page 68
65
Gas inlet
tubing (Green)
Baffles
Temperature
thermistor
Water level
sensing pins
Heating
plate
To Patient
HUMIDIFIER OPERATION
Installation of a new humidifier cartridge/circuit includes observing the cartridge function for a few minutes.
After you press the ENTER button:
The humidifier begins operating.
The pump fills the cartridge with water.
Water in the cartridge is heated by the
hot plate behind the cartridge.
The red circuit heater wire begins
heating.
The cartridge fills with water up to the
point where the water contacts the first two level sensing pins.
The pump stops pumping.
Page 69
66
SILENCESET
CARTRIDGE
CIRCUIT
CIRCUIT TEMP
ALARMS
CIRCUIT
LEVEL
TEMP
FAULT
LOW
HIGH
CIRCUIT FAULT: LEVEL Alarm
The humidifier microprocessor knows that the pump should be able to fill the cartridge within eighty-six seconds. If the level sensing pins do not detect water within this period, the water pump will shut off and a
CIRCUIT FAULT: LEVEL alarm will appear in the humidifier ALARMS window.
The alarm may be silenced for 60 seconds by
pressing the humidifier alarm SILENCE
button.
The CIRCUIT FAULT: LEVEL alarm indicates
it is necessary to check the progress of the cartridge filling. If the water has reached the appropriate level as described above, the cartridge may need to be replaced.
If a CIRCUIT FAULT: LEVEL alarm sounds
and the water has not reached it proper level, look for one of the following possible causes:
The water supply and/or water transfer
tubing may not be connected properly.
The water transfer tubing may be
clamped shut.
The water supply may be empty.
If the CIRCUIT FAULT alarm sounds immediately after the ENTER button is
pressed:
The humidifier door may not be closed
properly.
The cartridge/circuit may be faulty.
If none of these conditions exists, call the Bunnell Hotline at 1-800-800-4358.
Page 70
67
Proper water level
is at 2nd level sensing pin
Water level
sensing pins
WATER LEVEL SENSING
While observing the delivery of water to the cartridge, ensure that it does not overfill. The pump stops pumping when the water level reaches the second water level sensing pin.
The cartridge has a third level sensing pin to detect high water level and prevent overfill into the circuit. The water pump is stopped automatically if the water level reaches the third pin.
If the all level sensing pins are defective, the water level isn‟t sensed properly and a set fill time limit (86 seconds) is designed to turn off the pump in time to prevent overfill.
Whenever installing a new cartridge/circuit, observe the cartridge fill until the pump stops.
As the cartridge fills, the water and gas inside are heated by the hot plate behind the cartridge. The temperature is regulated by the white cartridge thermistor wire located near where the circuit tubing attaches to the cartridge.
Page 71
68
WAIT
HUMIDIFIER
TEMPERATURE
SET
CIRCUIT CARTRIDGE
CIRCUIT TEMP
The actual Circuit Temperature will be 40 C and will be indicated by a green LED
o
CIRCUIT TEMPERATURE
The actual temperature of the gas is measured in the circuit just before the Patient Box and is displayed in the humidifier monitor section as
CIRCUIT TEMP with a green light.
As this temperature adjusts, it will approach 40o C, may overshoot slightly and then settle in right at or near 40o C.
There is approximately a 3o C drop in gas temperature from where it is last measured to where it enters the patient‟s trachea.
If the set circuit temperature is 40o C, the actual temperature of the gas as it enters the patient will be 37o C (body temperature).
The lungs are very effective heat exchangers. Unless there is a reason for wanting to raise or
lower the patient‟s core temperature, do not adjust the Circuit temperature to any value other than 40o C.
Page 72
69
WAIT
HUMIDIFIER
TEMPERATURE
SET
CIRCUIT CARTRIDGE
CIRCUIT TEMP
Pressing the SET button repeated will scroll through the Circuit, Cartridge, and Circuit Temp LEDs.
TEMPERATURE CONTROLS
We recommend that the set CIRCUIT
temperature remain at its default setting of 40o C.
Adjust the Circuit Temperature only if the patient‟s temperature is not within an expected range, and if there is a reason to believe the temperature of the delivered gas may be contributing to the problem.
Adjusting the Circuit temperature may affect humidification in the circuit tubing. Be ready to change the Cartridge Temperature to provide appropriate levels of humidification.
The temperature in the cartridge and the circuit tubing are set and controlled
independently. Pressing the SET button
repeatedly causes the LED to alternate
between CIRCUIT, CARTRIDGE, and CIRCUIT TEMP readings.
A red or yellow light indicates a set
temperature.
A green light indicates an actual
temperature.
Page 73
70
TEMPERATURE
SET
CIRCUIT
CARTRIDGE
CIRCUIT TEMP
SETTING CARTRIDGE TEMPERATURE
The humidity levels of the gas delivered to the patient may be increased or decreased by
raising or lowering the set CARTRIDGE
temperature as follows:
press the SET button twice to light the red
or yellow LED next to CARTRIDGE, and
adjust the temperature up or down by using
the adjustment buttons.
The humidifier has no ENTER button.
Wherever you leave the setting will be the temperature at which the heater will maintain the humidified gas in the cartridge.
The humidifier will switch the temperature display back automatically to continuous
CIRCUIT TEMP monitoring ten seconds after a
button is pressed.
Remember that changes to the CARTRIDGE
temperature affect humidification, not the temperature of the gas delivered to the patient.
Page 74
71
CARTRIDGE TEMP too high
CARTRIDGE TEMP too low
SILENCE
ENTER
READY
SERVO
PRESS
PSI
LOSS OF PIP
UPPER
LIMIT
CONDENSATION PROBLEMS
A LOSS OF PIP alarm with the READY light on will lock the Servo Pressure.
The patient will continue to be ventilated appropriately.
If the preset Cartridge Temperature is too high for a particular patient:
Excess condensation will develop in the
clear portion of the Life Pulse circuit between the cartridge and the Patient Box.
The MONITOR section of the Life Pulse
will display fluctuating pressures.
A LOSS OF PIP alarm may occur. As long as the Life Pulse is in the READY
mode, the Servo Pressure will lock at or near its operating level, and the Life Pulse will continue to ventilate the
patient appropriately during a LOSS OF PIP alarm
If the Cartridge temperature is set too low:
The green portion of the Life Pulse
circuit between the Patient Box and the LifePort adapter will be dry.
Page 75
72
Purge pressure monitoring tubing with air from a syringe if condensation is partially obstructing the tube.
Lower Cartridge Temperature if water collects in the circuit.
MANUALLY PURGING
PRESSURE TUBE
The pressure monitoring tubing is purged automatically every 15 seconds. If the Life Pulse monitoring senses that the pressure monitoring tube may be obstructed, the purge fires once per second until the obstruction is cleared.
However, if displayed pressures are erratic, the fluctuating pressures may be a result of excess condensation or mucus partially obstructing the pressure monitoring tubing. You may find that manually purging the pressure monitoring tubing, followed by suctioning of the ET tube may alleviate the problem.
Manually purge the pressure monitoring
tube of the LifePort adapter by flushing 2-3 cc of air from a needless syringe before manipulating the temperature settings.
If the clear portion of the Life Pulse
circuit collects water that migrates into
the patient, lower the CARTRIDGE
temperature.
If condensation enters the endotracheal
tube, it may be necessary to suction the patient. However, most of the water will be evacuated out of the ET tube into the conventional ventilator circuit because of the Life Pulse‟s expiratory flow pattern.
Page 76
73
SILENCESET
CARTRIDGE
CIRCUIT
CIRCUIT TEMP
ALARMS
CIRCUIT
LEVEL
TEMP
FAULT
LOW
HIGH
SILENCESET
CARTRIDGE
CIRCUIT
CIRCUIT TEMP
ALARMS
CIRCUIT
LEVEL
TEMP
FAULT
LOW
HIGH
HIGH / LOW WATER LEVEL
The humidifier alarm system detects high or low water levels for the cartridge.
A LEVEL LOW alarm may indicate that the
humidifier water supply has been
exhausted. The humidifier alarm SILENCE
button can be pressed while a new water supply is connected as follows:
1. Close the clamp on the water transfer
tube and disconnect the empty water supply bag.
2. Connect a new water supply.
3. Open the clamp on the transfer tube to
allow water to flow to the cartridge.
4. Press the WAIT button twice (i.e., off,
on) to give the pump time (86 seconds) to fill the humidifier cartridge.
Once the cartridge is filled and the pump has stopped, the pump is limited to 5 seconds of operation every minute. This restriction serves as a precaution against overfilling the cartridge during normal Life Pulse operation.
A LEVEL HIGH alarm rarely occurs but
may be caused by a faulty humidifier cartridge. Replacing the cartridge/circuit usually eliminates this alarm.
A LEVEL HIGH alarm may occur when the
water supply is hung above the Life Pulse, and the water supply tubing is not properly installed in the Life Pulse‟s water pump.
For this reason, the water supply bag should always be located below the level of
the humidifier so that a LEVEL LOW alarm
will occur when the water supply tubing is not properly installed.
Page 77
74
SILENCESET
CARTRIDGE
CIRCUIT
CIRCUIT TEMP
ALARMS
CIRCUIT
LEVEL
TEMP
FAULT
LOW
HIGH
SILENCESET
CARTRIDGE
CIRCUIT
CIRCUIT TEMP
ALARMS
CIRCUIT
LEVEL
TEMP
FAULT
LOW
HIGH
HIGH / LOW TEMPERATURE
The humidifier alarm system detects high and low temperature for both the circuit and cartridge.
A high or low temperature is indicated by
an audible alarm and a TEMP HIGH or TEMP LOW message in the alarm display.
The TEMP HIGH alarm is activated if:
the monitored temperature strays 3
o
above the set temperature for more than 1 minute in the circuit; or
the monitored temperature strays 3
o
above the set temperature for more than 10 minutes in the cartridge.
The TEMP LOW alarm is activated if: the monitored temperature strays 3
o
below the set temperature for more than 3 minutes in the circuit; or
the monitored temperature strays 3
o
below the set temperature for more than 30 minutes in the cartridge.
Such alarms may be caused by a faulty circuit relaying false information to the microprocessor. For example, if the circuit is quite warm to the touch when the humidifier is indicating that it is too cold, it could result in overheating. The humidifier circuit might need to be replaced in such conditions. Call the Bunnell
Hotline at 800-800-4358 before changing the circuit
Page 78
75
Appendix A
POST-TEST QUESTIONS
Section 1
OVERVIEW
1. The five subsystems of the Life Pulse are ____________, ____________, ___________, ____________, and ____________.
2. Pressures measured in the patient are displayed in the ____________ section.
3. To make ventilator changes in high-frequency Rate, PIP, and On-Time, use the buttons and displays in the ____________ section.
4. The Patient Box is where blended gas is broken up into small “pulses” which
then provide the patient with breaths necessary for ventilation.
T F
5. The Life Pulse is routinely used without attaching a conventional ventilator.
T F
6. The Life Pulse is designed to improve the patient‟s blood gases while using less
pressure.
T F
7. Each ventilator performs distinct functions when using a conventional ventilator in tandem with the Life Pulse. Check which ventilator primarily performs the following functions. Check both ventilators if they provide an equally important function.
Life Pulse Conventional
___ a. Provides gas for patient‟s spontaneous breathing. ___ ___ b. Monitors Mean Airway Pressure (MAP). ___ ___ c. Controls Positive End Expiratory Pressure (PEEP). ___ ___ d. Controls high-frequency Rate and PIP. ___
___ e. Provides supplementary IMV. ___ ___ f. Humidifies blended gas delivered to the patient. ___ ___ g. Blends oxygen and compressed air. ___
8. A major advantage of the LifePort adapter is that reintubation is not required in
order to provide high-frequency ventilation.
T F
Page 79
76
Appendix A
POST-TEST QUESTIONS
Section 2
SET UP
REAR PANEL
1. The Life Pulse‟s power cord may be plugged into a standard electrical outlet or an
uninterruptible power supply.
T F
2. Oxygen and compressed air must be blended before entering the Life Pulse.
T F
3. At least 30 psi (206.85 kPa) of gas pressure must be supplied for the Life Pulse to
operate.
T F
4. Analyzing FiO
T F
5. The Life Pulse alarm volume level is not adjustable.
T F
FRONT PANEL
6. The disposable humidifier cartridge/circuit consists of which of the following:
a. Electrical sensors b. Heating elements c. Purge tubing d. Humidification cartridge
e. All of the above
7. The pinch valve is located in the Patient Box.
T F
8. Either sterile water or normal saline may be used in the humidifier cartridge.
T F
9. Once the humidifier cartridge/circuit has been installed, three tubes will need to be connected to the front panel of the Life Pulse. Attach the small clear tube to the barbed connector labeled PURGE; attach the green gas inlet tube to the barbed connector
labeled GAS OUT, and latch the clear water inlet tube into the pump housing.
T F
10. The pinch tube portion of the humidifier cartridge/circuit is the only tube that should be
between the jaws of the pinch valve.
T F
is best performed by measuring directly from the air/oxygen blender.
2
Page 80
77
Appendix A
POST-TEST QUESTIONS
Section 3
VENTILATOR CONTROLS AND TEST PROCEDURE
1. An immediate audible alarm after the Life Pulse is turned on means the ventilator is
defective.
T F
2. The STANDBY button stops jet ventilation from being delivered to the patient.
T F
3. The STANDBY button:
a. should be pressed each time a parameter is changed in the CONTROLS section.
b. puts the Life Pulse through a systems test.
c. creates a brief audible alarm every 30 seconds.
d. none of the above.
4. A systems TEST may be performed only with the Life Pulse in the Standby mode and
should never be performed while the Life Pulse is attached to the patient.
T F
5. It is important to observe the front panel of the Life Pulse during the TEST procedure to
assure all LEDs and displays are functional.
T F
6. Match each item in the first column with the single most accurate statement in the second column.
1. ENTER button ____ a. Never pressed while the Life Pulse is attached to
the patient
2. NOW settings ____ b. Stops the audible alarm for 60 seconds
3. TEST button ____ c. “Awakens” the Life Pulse into the Standby mode
with an audible alarm
4. ON button ____ d. Transfers NEW settings to NOW settings.
5. NEW settings ____ e. Disconnected purge tubing
6. SILENCE button ____ f. The current or active PIP, RATE, and ON-Time
7. STANDBY button ____ g. Problem with connection at GAS OUT hose barb
or with one of the Servo Pressure control valves
8. VENTILATOR ____ h. Similar to Inspiratory Time
FAULT 02
9. VENTILATOR ____ i. Proposed PIP, RATE, and ON Time FAULT 03
10. JET VALVE ____ j. Stops high-frequency ventilation delivery
ON-TIME
Page 81
78
Appendix A
POST-TEST QUESTIONS
Section 4
PRESSURE MONITORING
1. What pressure is displayed in the MONITOR section when the Life Pulse is in the Standby mode and not attached to a patient?
a. Servo Pressure b. MAP c. proposed PIP d. Zeros will be displayed e. a., b., and c.
2. Once the pressure monitoring tube of the LifePort adapter has been connected to the Patient Box, the Life Pulse, in its Standby mode, will begin to monitor the pressures being delivered by the conventional ventilator. These pressures are measured every 2 milliseconds, averaged over ________ seconds, and updated on the display every ________ seconds.
3. Match one MONITOR feature in the first column with its definition in the second column:
1. MAP a. The internal driving or working pressure required to
ventilate the patient at the NOW settings
2. P b. Average of peak pressures measured in the LifePort.
3. PEEP c. PIP minus PEEP
4. PIP d. Average of pressures measured over the total area of
the pressure waveform
5. SERVO e. Average of minimum airway pressure
4. Bigger patients, or patients with more compliant lungs, require ________ (MORE/LESS) Servo Pressure.
5. The Purge system maintains the patency of which tube?
a. Jet port
b. Main port of the endotracheal tube c. Pressure monitoring tube
Page 82
79
Appendix A
POST-TEST QUESTIONS
Section 5
HUMIDIFIER
1. Lowering the cartridge temperature will reduce the amount of humidity in the gas but
will not affect the temperature of the gas that is delivered to the patient.
T F
2. There are two separate temperature controls for the humidifier cartridge/circuit.
T F
3. Improper regulation of the humidity and temperature of the gas delivered to the
patient may: (choose all that apply)
a. contribute to dehydration. b. contribute to fluid overload. c. raise the patient‟s body temperature. d. lower the patient‟s body temperature. e. contribute to mucus plugging.
4. It is important to check for proper humidification (misty or condensation) in the green
portion of the Life Pulse circuit between the Patient Box and the LifePort.
T F
5. The humidity of the gas delivered to the patient is:
a. primarily controlled by the Circuit temperature. b. determined by the Cartridge temperature. c. controlled in response to measurements taken by the Cartridge thermistor.
6. The temperature regulation in the Patient Circuit tubing:
a. primarily controls the temperature of the gas going to the patient. b. is used to control condensation in the Circuit tubing. c. is controlled in response to measurements taken by the thermistor in the circuit
just before the Patient Box.
d. a. and c. only.
7. The Circuit temperature stays at 40o C and the Cartridge temperature is adjusted as
necessary to control condensation and rainout.
T F
8. If a small red light on the Humidifier‟s WAIT button is flashing, the Humidifier has
been placed in its WAIT mode manually, and the button must be pressed by the
operator to return the humidifier to normal operation.
T F
Page 83
80
Appendix A
POST-TEST QUESTIONS
Section 6
START UP
1. To begin high-frequency Jet ventilation, attach the Life Pulse to the patient, monitor
conventional ventilator pressures with the Life Pulse in Standby mode, and determine
the initial Life Pulse settings.
T F
2. In the Standby mode, the pressures displayed in the MONITOR section are 80-second
running averages updated every 10 seconds. During operation, the displayed
pressures are 10-second running averages updated every 2 seconds.
T F
3. If air leak is the primary concern, select a PIP approximately 90-100% of the
monitored conventional ventilator PIP. If the primary concern is RDS, select a PIP
100-110% of the monitored conventional ventilator PIP.
T F
4. Number the following Start-up steps according to the order in which they occur.
____ a. Adjust PEEP, if necessary ____ b. Select starting PIP ____ c. Lower conventional ventilator Rate ____ d. Lower conventional ventilator PIP if interrupting jet pulses ____ e. Press ENTER button ____ f. Press RESET button to establish new alarm limits, if necessary
5. A 10% drop in PIP with the Life Pulse typically results in a 20 to 25% drop in MAP, a
reduction that may be too great in the RDS patient.
T F
6. After making a change in the conventional ventilator settings, it may be necessary to:
a. make the same changes on the Life Pulse. b. press the RESET button to allow the Life Pulse to recalculate alarm limits around
the Servo and Mean Airway Pressures. c. lower the RATE on the Life Pulse. d. Both a. and c. e. None of the above.
Page 84
81
Appendix A
POST-TEST QUESTIONS
Section 7
PATIENT MANAGEMENT
1. The thought processes and rationale for managing patients on the Life Pulse as
compared with a conventional ventilator are very similar.
T F
2. Which of the following statements regarding patient management is (are) true? (Circle all that apply)
a. Most setting changes to manage PCO2 will be performed on the Life Pulse. b. Most setting changes to manage PO2 will be performed on the conventional
ventilator.
c. The Life Pulse is used in tandem with the conventional ventilator to provide
better blood gases while using less airway pressure.
d. All of the above.
e. a. and b. only.
3. The best approach for dealing with a high PCO
is to increase delta P ( P).
2
T F
4. Because CO2 elimination is proportional to Tidal Volume squared ( V frequency ventilation, small changes in which of the following parameters may
produce significant changes in PCO2? (Circle all that apply)
a. PIP b. PEEP c. Delta P d. All of the above
5. PCO2 would most likely ______________ (INCREASE/DECREASE) if PEEP is increased to improve PO2.
6. Increasing high-frequency PIP is a more effective means of eliminating PCO2 than
increasing high-frequency Rate.
2
) during high-
T
T F
7. The best approach for controlling PO
is to adjust Mean Airway Pressure (MAP).
2
T F 
Page 85
82
Appendix A
POST-TEST QUESTIONS
Section 7
PATIENT MANAGEMENT
Cont.
8. Which of the following strategies may be the most effective for increasing MAP and oxygenation?
a. Increasing PEEP b. Increasing conventional IMV c. Increasing conventional inspiratory time d. Increasing high-frequency On-time e. Increasing high-frequency PIP
In the following two scenarios, match the most appropriate ventilator strategy with the patient‟s specific pathophysiology.
9. A near term infant with a. Increase FiO2
a spontaneous pneumothorax and pneumomediastinum; b. Set initial conventional IMV rates 0 to 3 bpm minimal oxygenation concerns but excessively high PCO2. c. Set initial conventional IMV rates 3 to 5 bpm
____ ____ ____ d. Increase high-frequency PIP by 1 to 2 cm H2O
e. Increase PEEP
10. A premature infant with RDS
requiring high pressures; f. Increase conventional IMV rate to 10 bpm worsening oxygenation. g. Decrease PEEP ____ ____ ____ ____
11. Changing the high-frequency On-Time has not been proven an effective means of
improving oxygenation.
T F
Page 86
83
Appendix A
POST-TEST QUESTIONS
Section 8
SUCTIONING THE PATIENT
1. There are three methods of suctioning a patient while on the Life Pulse. T F
2. Secretions and suctioning frequency may ______________ (INCREASE/DECREASE) during the first 4 to 6 hours of high-frequency ventilation.
3. The technique(s) for suctioning during high-frequency ventilation is (are):
a. to place the Life Pulse in Standby mode and suction normally. b. to leave the Life Pulse running, instill normal saline down the Jet port of the LifePort
adapter, and apply suction in both directions, while introducing and withdrawing the
suction catheter. c. to use a closed-suction system. d. All of the above
4. If the Life Pulse is placed in Standby mode for suctioning, the patient will not require manual resuscitation or increased support from the conventional ventilator. T F
5. With the Life Pulse running, if suction is not applied while introducing the suction catheter, the jet pulses will interrupt and a HIGH PIP alarm may occur, dumping the Servo Pressure safely to atmosphere. T F
6. If the READY light is on while suctioning is performed with the Life Pulse running, the Servo Pressure is designed to lock at or near its baseline operating level and continue to ventilate the patient as requested. T F
Page 87
84
POST-TEST QUESTIONS
Section 9
WEANING
1. Which of the following statements regarding weaning is false?
a. If the Life Pulse is not needed for another patient, a baby may be weaned from
high-frequency ventilation down to CPAP.
b. Weaning too quickly is a vital concern and should be avoided. c. If a separate blender is being used to supply mixed gas to the Life Pulse, both
blenders should be adjusted to wean FiO2 .
d. Conventional pressures and/or breaths may need to be increased gradually when
the Life Pulse pressures have been decreased into the teens.
e. The Life Pulse Rate should always be lowered to its minimum setting of 240 bpm
before returning the patient to conventional ventilation.
Appendix A
2. The Servo Pressure may indicate when a patient‟s compliance is improving.
T F
3. Generally, as PCO2 decreases, high-frequency PIP is lowered in increments of 1 to 2 bpm; as PO2 improves, the FiO
or the parameters that produce MAP are lowered.
2
T F
4. When a patient begins to respond negatively to further decreases in pressure, it may
indicate that the patient has reached a “pressure threshold” and further decreases are
ill-advised until the patient has “acclimated” to the changes.
T F
5. FiO
should be reduced to 40% or less before weaning PEEP.
2
T F
6. Interruptions in the jet pulses should never be allowed during any stage of the
weaning procedure.
T F
7. The conventional ventilator rate and PIP usually need to be increased as Life Pulse
PIP is weaned into the mid-teens.
T F
8. When the patient is stable and most of the chest rise is being provided by the conventional ventilator breaths, a trial of conventional ventilation alone may be
attempted by pressing the STANDBY button on the Life Pulse.
T F 
Page 88
85
Appendix A
POST-TEST QUESTIONS
Section 10
VENTILATOR ALARMS
1. The four components of the ALARMS section are: upper and lower alarm limits, alarm
messages, the SILENCE button, and the RESET button.
T F
2. The displayed values which have manually adjustable upper and lower alarm limits are: (check all that apply)
a. PIP
b. P
c. PEEP d. Servo Pressure e. MAP
3. The READY condition is met when the monitored PIP has come to within ______ cm H2O of the set NOW PIP and has stabilized there for ______ seconds.
4. It is important to never walk away from the Life Pulse until the READY light is
illuminated.
T F
5. The Servo and Mean Airway Pressure alarm limits are set automatically and are not
manually adjustable.
T F
6. A JET VALVE FAULT alarm may indicate that the electronics controlling the pinch
valve have failed.
T F
7. The patient should be left attached to the Life Pulse when pressing the TEST button
to perform a systems test in the Standby mode.
T F
8. If the Life Pulse continues to run with the VENTILATOR FAULT message, chances
are one of the ends of the purge tube has become disconnected.
T F
9. If a VENTILATOR FAULT 10 alarm occurs, the Life Pulse will resume operating
when the ENTER button is pressed.
T F 
Page 89
86
Appendix A
POST-TEST QUESTIONS
Section 10
VENTILATOR ALARMS
cont.
10. If the exhalation tubing of the conventional ventilator circuit becomes kinked, the Life Pulse: (choose two)
a. initiates a HIGH PIP alarm.
b. initiates a LOW GAS PRESSURE alarm.
c. “dumps” the Servo Pressure to reduce the PIP.
d. locks the Servo Pressure at the current value.
11. Match the alarms in the first column with the descriptions in the second column:
1. LOW GAS PRESSURE ____ ____ a. Unable to meet READY condition
within 3 minutes of pressing ENTER button
2. CANNOT MEET PIP ____ ____ b. Monitored PIP and PEEP come to
within 2 cm H2O of each other
3. LOSS OF PIP ____ ____ c. Gas supply drops below 30 psi (206.85
kPa)
4. HIGH PIP ____ ____ d. If occurs in Non-READY condition,
Servo value drops to or near zero
____ ____ e. May occur on a patient who is too
large to be ventilated by Life Pulse
____ ____ f. Leak in oxygen or compressed air hose
____ ____ g. Monitored PIP exceeds NEW setting
by at least 5 cm H2O continuously for 1 second.
____ ____ h. Servo Pressure is exhausted safely
through “dump” valve to atmosphere.
Page 90
87
Appendix A
POST-TEST QUESTIONS
Section 11
INTERPRETING VENTILATOR ALARMS
1. Many alarms can be prevented by learning how the Life Pulse is designed, understanding its operation, and performing careful and thorough ventilator checks. However, when an alarm occurs, the first thing to do is look at the _____________.
2. Some alarm conditions may require manual ventilation or increased conventional
ventilator support while adjustments are made to the Life Pulse.
T F
3. “There is no substitute for a skilled and alert ______________, ______________, or ______________.”
4. Careful observation is one aspect of using common sense when troubleshooting. An example of this skill is:
a. watching for appropriate chest rise. b. looking for proper condensation in the circuit. c. listening to the sounds of the Life Pulse for correct operation. d. knowing baseline operating levels and observing trends. e. All of the above.
5. If Servo Pressure exhibits an upward trend, the patient‟s compliance may be
improving or there may be an air leak, either in the patient or the cartridge/circuit.
T F
6. Knowing the ventilator is an important principle for understanding how to
troubleshoot the Life Pulse.
T F
7. A majority of alarms are caused by kinked, disconnected, or obstructed tubes.
T F
8. One of the most important resources available to clinicians using the Life Pulse is the
trained customer service representative answering the Bunnell 24-hour Hotline.
T F
9. The Operator‟s Manual is a valuable resource for understanding the Life Pulse since it
details information not covered in the In-Service Manual.
T F
Page 91
88
Appendix A
POST-TEST QUESTIONS
Section 12
INTERPRETING HUMIDIFIER ALARMS
1. A humidifier CIRCUIT FAULT alarm may be caused by:
a. broken cartridge or circuit wires. b. impaired contact between electrical connections (e.g., cartridge door open). c. water in the cartridge not reaching the appropriate level within the allotted fill time. d. All of the above.
2. The Circuit temperature is best left at 40o C because the gas temperature drops 3o
from where it is last heated to where it enters the patient.
T F
3. Match the following:
Circuit temperature ____ a. Primarily controls humidity of delivered gas. Cartridge temperature ____ b. Primarily controls temperature of delivered gas.
4. Which of the following alarms is not found in the Humidifier Section?
a. LEVEL LOW b. TEMP HIGH c. LEVEL HIGH d. TEMP LOW e. VENTILATOR FAULT
5. Excessive rainout in the clear portion of the circuit may enter the ET tube and impede
pressure monitoring causing a LOSS OF PIP alarm.
T F
6. If the cartridge door is closed and secured, if all tubing and electrical connections have been made, if water is in the water supply bag, and if the water transfer tubing is patent, and the humidifier appears to be working normally, a TEMP HIGH or LOW or a LEVEL HIGH or LOW alarm usually means the operator should:
a. press the WAIT button and continue delivering high-frequency Jet ventilation. b. adjust the Cartridge and Circuit temperature settings. c. probably replace the Humidifier Cartridge/Circuit. d. call the Bunnell Hotline for additional troubleshooting ideas.
e. c and d only
Page 92
89
Appendix B
POST-TEST ANSWERS
Section 1
OVERVIEW
1. CONTROLS, MONITOR , ALARMS, HUMIDIFIER, AND the Patient Box.
2. MONITOR.
3. CONTROL.
4. True.
5. False. The Life Pulse is always used in tandem with a conventional ventilator.
6. True. The conventional ventilator and the Life Pulse work synergistically to
improve the patient‟s blood gases using less pressure, in most cases.
7. a. Conventional only.
b. Life Pulse and some Conventional. c. Conventional only d. Life Pulse only.
e. Conventional only. f. Both. g. Conventional. (Unless the conventional ventilator has a low flow external
output, a separate oxygen blender provides mixed gas to the Life Pulse)
8. True. The LifePort adapter replaces the standard ET tube adapter and obviates
the need to reintubate the patient.
Page 93
90
Appendix B
POST-TEST ANSWERS
Section 2
SET UP
1. True. However, the uninterruptible power source is more desirable.
2. True. The Life Pulse does not have an internal blender.
3. True. Pressures under 30 psi (206.85 kPa) will cause a LOW GAS PRESS alarm.
4. True.
5. False. The alarm volume knob is located on the rear panel of the Life Pulse.
6. e. All of the above.
7. True.
8. False. Only sterile water should be used to properly humidify the gas and assure
normal function of the cartridge.
9. True. Once the cartridge is placed in the cartridge door and the door is closed, the
tubes must be attached to the Life Pulse.
10. True. The pinch tube is the only circuit section flexible enough to function in the
pinch valve.
Page 94
91
Appendix B
POST-TEST ANSWERS
Section 3
VENTILATOR CONTROLS AND TEST PROCEDURE
1. False. When the Life Pulse is powered up it always “awakens” in the
Standby mode with the alarm sounding.
2. True. Pressing STANDBY is the accepted method of stopping high-frequency
ventilation while still allowing the Life Pulse monitoring to be active.
3. c. In the Standby mode, the Life Pulse beeps six times every thirty
seconds to inform you that it is not ventilating.
4. True. The TEST button should only be pushed with the ventilator attached
to a test lung. Pressing the TEST button will not initiate the systems test when the Life Pulse is running.
5. True.
6. 1-d; 2-f; 3-a; 4-c; 5-i; 6-b; 7-j; 8-e; 9-g; 10-h.
Page 95
92
Appendix B
POST-TEST ANSWERS
Section 4
PRESSURE MONITORING
1. d. “zero pressure.” All displays will be zero.
2. “80”, “10”. However, during operation all pressures are averaged over a 10 second
period and the display is updated every 2 seconds. This approach helps to produces more stable readings.
3. 1-d; 2-c; 3-e; 4-b; and 5-a.
4. MORE.
5. C. Pressure monitoring tube.
Page 96
93
Appendix B
POST-TEST ANSWERS
Section 5
HUMIDIFIER
1. True. The cartridge temperature may be reduced to minimize condensation;
however, the heated wire circuit will assure that the temperature of the gas remains stable.
2. True. Temperature in the cartridge and in the circuit tubing are controlled
separately.
3. a, b, c, d, e. All of these are complications of improperly regulated humidification,
with high-frequency ventilation and conventional ventilation.
4. True.
5. b and c only.
6. d. Condensation is controlled by adjusting the Cartridge temperature.
The Circuit temperature should normally remain at 40° C
7. True.
8. True. If the humidifier is placed in the WAIT mode manually (indicated by a
flashing WAIT button), it must be restarted manually.
Page 97
94
Appendix B
POST-TEST ANSWERS
Section 6
START UP
1. True.
2. True.
3. True. However, these starting criteria may vary depending on the patient‟s
pathophysiology.
4. a-5, b-1, c-4, d-3, e-2, f-6.
5. True.
6. b.
Page 98
95
Appendix B
POST-TEST ANSWERS
Section 7
PATIENT MANAGEMENT
1. True.
2. d.
3. True. Delta P can be increased by raising the Life Pulse PIP or by
decreasing the PEEP.
4. d. Delta P is PIP minus PEEP.
5. EITHER. Raising PEEP might decreases Tidal Volume and may result in CO2
retention if, despite raising PEEP, alveoli remain unstable at the end of expiration. However, if by raising PEEP the alveoli are stabilized at the end of expiration, CO2 may decrease.
6. True.
7. True.
8. a. MAP and PEEP have a 1-to-1 ratio, meaning that raising the PEEP by
1cm H2O raises the MAP by 1 cm H2O.
9. b, d, g. Concern for pulmonary airleaks means limiting tidal volume form
IMV breaths, but increasing minute ventilation using HFJV. Increasing HFJV PIP is more effective than increasing HFJV Rate for lowering PCO2. Decreasing PEEP will increase delta pressure, thus tidal volume, and may help lower PCO2.
10. a, c, e, f. Concern for oxygenation would lead to actions that recruit and
maintain lung volume. Except for increasing FiO2, all options involve increasing Mean Airway Pressure. Using more IMV breaths improves lung volume recruitment, and raising PEEP maintains recruitment. Once oxygenation improves (alveoli are open and stable), wean the IMV rate as tolerated.
11. True. Clinical trials have shown that changing high-frequency On-Time has
little clinical value, except possibly at slow ventilator rates while weaning and trying to maintain a desired I:E ratio.
Page 99
96
1. True.
2. INCREASE.
3. d. (all of the above)
4. False.
5. True.
6. True.
Appendix B
POST-TEST ANSWERS
Section 8
SUCTIONING THE PATIENT
Page 100
97
1. e.
2. True.
3. True.
4. True.
5. True.
6. False.
7. True.
8. True.
Appendix B
POST-TEST ANSWERS
Section 9
WEANING
Loading...