9.6Fault tracing using the error codes .............................................................. 5
10ENGINEERING CHANGE HISTORY PV501 AND PV501-2 .................................... 1
10.1Service record for BREAS PV501/PV501-2
Doc. No.: 1523EN Issue: N-1BREAS MEDICAL1 – 2
Page 4
Service Manual PV 501/501-2INTRODUCTION
1INTRODUCTION
1.1The scope of this manual
This manual provides information for the maintenance and service of both the PV 501
and PV 501-2 ventilators. The PV 501-2 is a further development of the PV 501. The
main differences are that the PV 501-2 delivers a maximum patient pressure of 60 mbar
(compared to the PV 501’s 50 mbar) and that theTrigger setting is adjustable between -
-2 to 6 mbar (compared to the PV 501’s -4 to 6 mbar).
The PV 501and PV 501-2 are designed to give many years of trouble-free breathing
assistance to the user provided that preventative maintenance is done at the specified
intervals described in this manual. Correctly performed maintenance will increase the
ventilator's service life considerably.
All points to be checked and service instructions for the PV 501 and the PV 501-2 are
described in this manual. Where information is only relevant to one of the models it is
clearly marked which model the text/figure refers to, otherwise all information is valid for
both models.
It is also important that any peripheral equipment is checked at the same time as the
maintenance service is done.
Also included is a reference copy of the Operation Manual.
1.2Intended use of the ventilator
PV 501 and the PV 501-2 are volume-controlled, pressure-limited ventilators, specially
designed for long term ventilator assistance at home.
The internal batteries are connected automatically should the mains supply fail or is disconnected. Fully-charged batteries will give a running time of 4 hours.
Thanks to its design, the PV 501 and the PV 501-2 are easy to operate. It does not
require any gas, and together with its low weight, it is ideal for breathing therapy both in
hospital and while travelling.
The PV 501and the PV 501-2 are not intended for intensive care and are aimed for
adults and children weighing about 30 kg or more.
1.3Design and function of the ventilator
The PV 501and the PV 501-2 are built around a bellows that is compressed and drawn
out by a ball screw driven by an electronically controlled servo motor. The micro-processor controlled electronics calculate the correct speed and running time for the motor by
reading the settings for tidal volume, frequency and the I/E-relationship. The setting limits for pressures and the trigger level are monitored. If the mains supply or the external
battery supply should fail during operation, the internal batteries are automatically connected and an indication of this is given on the front panel. If the battery supply voltage
drops too low, both audible and visual alarms are given. The ventilator's modular design
makes maintenance easy.
1.4Intended audience
This Service Manual is intended for technicians who have medical/technical training and
knowledge about the construction and function of the ventilator.
It is not intended for clinic personnel or patients.
Doc. No.: 1523EN Issue: N-1BREAS MEDICAL1 – 1
Page 5
INTRODUCTIONService Manual PV 501/501-2
Breas Medical reserves the right to make changes to the product and the contents of
this manual without prior notice.
1.5Service personnel's training requirements
Thanks to their simple construction using a modular system, no special competence is
required other than general medical technical training on ventilators.
Always contact BREAS MEDICAL if there are any questions or if training is required.
All service must, however, be performed according to the instructions in this manual.
1 – 2BREAS MEDICALDoc. No.: 1523EN Issue:N-1
Page 6
Service Manual PV 501/501-2MAINTENANCE SERVICE INSTRUCTIONS
3MAINTENANCE SERVICE INSTRUCTIONS
All routine maintenance checks and additional se rvice instructions for the PV
501and PV 501-2 are described in this chapter. For information about fault-tracing, detailed drawings, board schematics, spare parts etc, please refer to the
respective chapters in this Service Manual.
The checks described in the Patient Instruction delivered with the ventilator
should be followed by the patient and/or care provi ders.
3.1Verifying the components and software installed
Check the Engineering Change History document in the Appendices section
(Chapter 10) for a history of all the changes made and at which serial No they
were introduced.
If in any doubt, read the component designation on the motor unit, circuit boards
and PROMs as upgrades can have been made but not recorded.
3.2Maintenance Service Schedule
IMPORTANT!
A complete maintenance service (as described in this chapter) must be done
every 12th month. If the ventilator is used for continuous operation (24 hours
per day) a complete maintenance service must be done every 6th month.
IntervalAction
At every 500th operation hour or more frequent if
necessary (especially in town environments).
Every 6 months if the ventilator is used for continuous operation i.e. 24 hours per day, otherwise
every 12th month.
Additional service action (when required)
At 20.000 running hoursReplace motor unit and hoses.
The internal batteries should be replaced every 24
months.
•Replace the alarm batteries every 5th year
counted from date of delivery or when
required.
Replace the patient air inlet filter.
Done by the patient/care provider.
Replace during maintenance service.
Complete maintenance service according to this instruction.
Replace the internal batteries
Replace the alarm batteries.
Doc. No.: 1523EN Issue: N-1BREAS MEDICAL3 – 1
Page 7
MAINTENANCE SERVICE INSTRUCTIONSService Manual PV 501/501-2
Every 12th month or every 6th month if the ventilator is used
for continuous operation (24 hours per day)
Replace the patient air inlet filter.2
Clean the ventilator air intake filter2
Motor Unit
Lubricate the ball screw7.2
Replace the check valve membranes in bellows end cover7.3
Leakage test of motor unit and tubes7.4
Electronics
Checking the pressure sensor offset and gain. 8.5.1
Check operation using internal battery8.9
Check operation using external battery2
Check electrical safety levels8.13
Internal Batteries (lead-ac id)
The internal batteries should be replaced every 24 months or when necessary.
Section No.
8.10
See Ch./
Alarm Batteries
Replace every 5th year counted from date of delivery or when last
replaced.
Accessories (where applicable)
Inspect patient circuit2
Replace membrane in exhalation valve2
Clean PEEP adapter, replace O-ring2
Every 20000 operation hours
Replace the complete Motor unit6.6
3.3Special safety precautions
•Avoid working with the mains c onnect ed when the casi ng is r emoved. Always tes t
run using internal battery power.
•Insulate the wires for the inter nal batte ri es when they are di sconne cted to prev ent
short circuits.
•Explosive gases or liquids must not be kept or used close to the ventilator.
8.11
•Always follow good ESD practices when working with the ventilator.
3 – 2BREAS MEDICALDoc. No.: 1523EN Issue: N-1
Page 8
Service Manual PV 501/501-2MAINTENANCE SERVICE INSTRUCTIONS
3.4Equipment required
•Test lung (1L) with exhalation valve.
•Measuring equipment for tidal vol ume and minut e volume/frequency. (Timeter,
Spirometeor or equivalent).
•Voltmeter.
3.5Replacement parts required
Have the following spare parts at hand:
Part No.
000 139Internal battery (lead)
000 036Alarm battery (Ni-Cd)
001 141Motor unit, BEI
000 557Grease (BREAS 283 AZ)
000 004Motor unit, Maxon
000 248Filter, patient air
002 123Service kit, check valves
002 178Membrane kit, exhalation valve
Description
3.6Maintenance instructions
3.6.1Registration
•Fill out the registration part of the Service Record. Check that the following markings can be read:
•Make, model designation and serial number.
•All warning texts on labels.
•Any inventory control markings.
•Document the current patient settings.
•Note the number of running hours. Does the motor unit need to be replaced?
•Check any comments or events made on the previous service record.
3.6.2Information from the user
•Before starting the service request the following information from the patient:
•Has the ventilator functioned problem-free, such as running failure?
•How does the patient check the function of the ventilator? How often?
•How often is the filter changed?
•What is the patient's need of filters unt il the next service point?
3.6.3Validity of documentation
•Check that the patient instructions are up-to-date.
•Check if any modification or update of the ventilator is to be performed during the
service.
Doc. No.: 1523EN Issue: N-1BREAS MEDICAL3 – 3
Page 9
MAINTENANCE SERVICE INSTRUCTIONSService Manual PV 501/501-2
3.7External checks
3.7.1External damage and wear
•Clean the surface with a mild detergent.
•Clean the filter for the machine ventila tor air inlet. Use a vacuum cleaner or wash
and dry the filter before fitting.
•Check that there is no visible damage on the surfaces and other components.
•Check that all texts for controls etc. are readable.
•Turn all knobs and check that their operation feels OK.
•Check that nothing is loose, (including the handles).
•Check that the cover for the adjustment panel is secu red properly.
3.7.2Power cables and plugs
•Check the power cable and its plugs, and the power socket in the rear panel.
•Check that the cable securing clamp is undamaged.
•Check other external battery cables, where applicable.
3.7.3Minimum function check
•Connect the ventilator to the mains and check that the LED for Mains/Charging is
lit.
•Start the ventilator and check that it oper ates normally and no abnormal noise is
heard. Check that the lighting for the press u re gauge is on.
•Check the condition of the internal batteries. They should, when fully charged, be
able to run the PV 501 for at least 2 hours. If not they must be replaced.
3.8Internal Checks
3.8.1Open hood and rear panel
•Remove the mains power cable.
•Open the hood and rear panel. Remove stabiliser bar, see chapter 6.
3.8.2Cleaning
•Remove any dirt or dust that has collected in the ventil ator, also from the folds in
the bellows.
3.8.3Cables and connectors
•Check all cables and thei r connectors for an y damage. Check at the fr ont and rear
panels where cables can be pinched.
3.8.4Component fastening
•Check that the batteries are properly fastened.
•Check that the motor unit is properly fastened.
•Check that all parts on the front and rear panels are properly fastened.
3 – 4BREAS MEDICALDoc. No.: 1523EN Issue: N-1
Page 10
Service Manual PV 501/501-2MAINTENANCE SERVICE INSTRUCTIONS
3.8.5Mains supply
•Check that the power inlet socket is undamaged and that it is properly fastened.
•Check that the touch protection cover is undamaged and that it is properly fastened over the power inlet socket.
•Check that the transformer is properly fastened.
•Check the wiring to the transformer.
3.8.6Remove and check / replace the motor unit.
The motor unit must be checked at each service. At 20.000 running hours the
entire motor unit assembly must be replaced. See Chapter 6 for removing the
motor unit.
Checking the motor unit
•Check that the motor is properly fastened.
•Check that the shaft coupling screws are properly tightened.
•Check that the encoders are undamaged and properly fastened.
3.8.7Grease the ball screw
Use only BREAS grease type 283 AZ. See chapter 7.
3.8.8Replace the membranes in the check valves
•The membranes must be changed at each maintenance service. See chapter 7.
3.8.9Leakage check of hoses and bellows
•Check the function of the outlet check valves. See chapter 7.
3.8.10 Reassemble the motor unit
Before reassembling the motor unit, check if the internal batteries need to be
replaced as this is best done with the motor unit removed, see 3.8.6.
•Reassemble the motor unit. See Chapter 6.
3.8.11Checking the offset and gain of the pressure transducer
•See Chapter 8, Section 8.5.1 for detail ed inf o rmation.
3.8.12 Checking the instrument accuracy
•Check using a manometer conne cted to the exh. valve connec tion t hat the i ndic ating instrument gives the correct indicati on and that the needle moves "softly".
See Chapter 8, Section 8.5.1.1.
NOTE! Make small pressure changes, the indicating instrument should be
"slower" compared with the manometer.
3.8.13 Internal battery test
The internal batteries should be fully charged or charged for at l east 12 hours with
the ventilator turned off before this test is carried out. The batteries, when fully
charged, should operate the PV 501 for at least 2 hours, if not t he batteries must
be replaced. See Chapter 8, Section 10.
Doc. No.: 1523EN Issue: N-1BREAS MEDICAL3 – 5
Page 11
MAINTENANCE SERVICE INSTRUCTIONSService Manual PV 501/501-2
3.8.14 Reassemble the casing
See Chapter 6 for information.
3.9Electrical safety
Follow the instructions for the following checks in Chapter 8.
•Check mains supply value
•Check isolation against the value obtained at the delivery inspection
•Check leakage current. Do not forget to reverse the polarity.
•Check patient leakage current.
•Leakage current at mains voltage at patient connected part
3.10Final checks before delivery
3.10.1 Function check/leakage
•Start the ventilator with the pat ient circuit and reservoir bag connected as shown
in the figure below. Check that everything seems normal. Create a pressure of
approximately 30 cm H
0 and listen for any leakage.
2
3.10.2 Check the tidal volume/frequency
Set: Tidal vol ume:1L
Frequency:12
I/E:1:2
•Measure the tidal volume and that the volume per minute is cor rect.
(Accuracy ± 10%).
When checking using a volume monitor, an exhalation valve with a peep valve
connector can be required. The volume monitor is then connected to the peep
valve outlet on the exhalation valve.
LOW PRESSURE
HIGH PRESSURE
LOW BATTERY
FUNCTION FAIL URE
TIDAL VOLUME
Litre
Setting error
Setting error
Setting errorSetting error
PATIENT PRESSUREALARMS
Peep valve connector
EXHALATION
VALVE
BPM
BPMcm HO
BPMBPM
POWER
PATIEN T
AIR
ON
OFF
LOW PRESSURE
cm HO
cm HOcm HO
POWER
AC/CHARGING
EXTERNAL BATTERY
INTERNAL BATTERY
TRIGGER
Off
Off
OffOff
TRIGGER
cm HO
2
HIGH PRESSURE
2222
Signal
Signal
SignalSignal
I/EBREATH RATE
cm HO
cm HO
Ratio
Ratio
cm HOcm HO
RatioRatio
2222
mbar
Volume monitor
Test lung / Reservoir bag
Fig. 3-1Tidal volume check
3 – 6BREAS MEDICALDoc. No.: 1523EN Issue: N-1
Page 12
Service Manual PV 501/501-2MAINTENANCE SERVICE INSTRUCTIONS
3.10.3 Calculating the tidal volume/patient pressure
The tidal volume delivered is always slightly less than the tidal volume setting.
This difference is due to the effect of the patient pressure on the compliance of
the bellows.
The diagram below shows these differences at different patient pressures.
Settings: I/E = 1:2, Breath rate = 15 breaths/min. with the patient circuit con-
nected.
Set volume/l itr e s
Fig. 3-2Volume/Patient pressure diagramme
Example:
A patient needs a tidal volume of 0,7 litres and has an airway pressure of 20 cm
O. Start from 0.7 on the axis for “ Delivered volume”. Draw a horizont al line unti l
H
2
it crosses the line for 20 cm H
O patient pressure. From this point, go straight
2
down and read off the value the tidal vol ume setti ng should be. In this case the
ventilator setting is 0.8 litres.
Doc. No.: 1523EN Issue: N-1BREAS MEDICAL3 – 7
Page 13
MAINTENANCE SERVICE INSTRUCTIONSService Manual PV 501/501-2
1
3.10.4 Checking the pressure limit
Set:Low pressure10 cmH2O
High pressure40 cmH
•Check the accuracy (±10%). Check that LED indicati on is given as well as an
audible alarm.
•Also check that the solenoid valve for high pressure "clicks". You can clearly hear
the solenoid valve activate at the following inhalation.
O
2
3.10.5 Checking low pressure alarm/alarm mute
•Turn the vent ilator on without anything connected to the patient air outlet. Wait 15
sec until the alarm for "Low pressure" is activated. Press the alarm mute button
and check that the alarm is heard again after approx. 2 min.
3.10.6 Trigger
•Set the trigger knob to -1 c m H20. Create a negative pressure an d check that ther e
is a triggered inhalation. The green LED should light.
3.10.7 Battery operation
•Disconnect the mains cable while the ventilat or is running. Check that the ventilator automatically changes over to internal battery and an alarm is given.
•Reconnect the mains power and check that indication "MAINS" comes on.
3.10.8 Checking accessories
Patient circuit
•Inspect the patient circuit and replace it if necessary.
Replacing the membrane assembly in the exhalation valve
This instruction applies to BREAS exhalation valves, see figure below.
Fig. 3-3Mounting the exhalation valve
•Remove the PEEP adapter , if installed. See "Cleaning the PEEP adapter".
3 – 8BREAS MEDICALDoc. No.: 1523EN Issue: N-1
Page 14
Service Manual PV 501/501-2MAINTENANCE SERVICE INSTRUCTIONS
•Unscrew and remove the complete membrane assembly (1).
•Clean the inside of the exhalation valve using a moist rag or wash in hot water
using a washing-up liquid. Let it air dry.
•Screw on the new membrane assembly
•If a PEEP valve is to be used, fit the O-ring seal for the PEEP valve as shown in
fig. 3-5.
•Connect the exhalation valve to a test lung. Check that no leakage occurs during
the exhalation phase.
3.10.9 Cleaning the PEEP adapter
1
2
3
Fig. 3-4Mounting the exhalation valve with PEEP adapter
•Remove the plastic nut (1) holding the PEEP adapter.
•Pull the adapter (2) up from the exhalation valve.
•Clean using a moist rag or wash in hot water using a washing-up l iqui d. Let the
adapter air dry.
•Check that the O-ring seal (3) is properly in place on the exhalation valve (under
the edge of the membrane assembly cover as shown in the figure below).
Note! Do not fit the O-r ing to the membrane assembly before screwing it on.
Doc. No.: 1523EN Issue: N-1BREAS MEDICAL3 – 9
Page 15
MAINTENANCE SERVICE INSTRUCTIONSService Manual PV 501/501-2
Fig. 3-5Mounting the O-ring for the PEEP adapter
•Fit the PEEP adapter to the exhalation valve and screw on the plastic nut
•Check any other accessories.
3.10.10Set the correct values for the patient
•Adjust the settings to the prescribed values for the patient.
•Change the air filter. Check that the patient has enough filters to last until the next
service.
The patient settings should be checked by a physician at least once a year.
3 – 10BREAS MEDICALDoc. No.: 1523EN Issue: N-1
Page 16
Service Manual PV 501/501-2REPLACEMENT PARTS
4REPLACEMENT PARTS
4.1OV ERVIEW DIAGRAM
The illustration belo w shows the PV501/PV501-2, seen f rom above, with the ho od
and rear panel unscrewed and laid flat against the work surface.
The figure below shows the main units making up the PV501/PV501-2. The following pages contain detailed exploded drawings for each of the main units and
model. Last in this chapter is a complete list of part numbers.
1
Pos. No.
1Hood assembly
2Motor unit
3Base plate assembly
4Rear panel assembly
Component
2
3
4
Doc. No.: 1523EN Issue: N-1BREAS MEDICAL4 – 1
Page 17
REPLACEMENT PARTSService Manual PV 501/501-2
4.2Exploded drawing, Base Plate Assembly PV 501
4 – 2BREAS MEDICALDoc. No.: 1523EN Issue: N-1
Page 18
Service Manual PV 501/501-2REPLACEMENT PARTS
4.3Exploded drawing, Base Plate Assembly PV 501-2
Doc. No.: 1523EN Issue: N-1BREAS MEDICAL4 – 3
Page 19
REPLACEMENT PARTSService Manual PV 501/501-2
4.4Exploded drawing, Motor Unit Assembly PV 501
4 – 4BREAS MEDICALDoc. No.: 1523EN Issue: N-1
Page 20
Service Manual PV 501/501-2REPLACEMENT PARTS
4.5Exploded drawing, Motor Unit Assembly PV 501-2
Doc. No.: 1523EN Issue: N-1BREAS MEDICAL4 – 5
Page 21
REPLACEMENT PARTSService Manual PV 501/501-2
4.6Exploded drawing, Rear Panel Assembly PV 501
4 – 6BREAS MEDICALDoc. No.: 1523EN Issue: N-1
Page 22
Service Manual PV 501/501-2REPLACEMENT PARTS
4.7Exploded drawing, Rear Panel Assembly PV 501-2
Doc. No.: 1523EN Issue: N-1BREAS MEDICAL4 – 7
Page 23
REPLACEMENT PARTSService Manual PV 501/501-2
4.8Exploded drawing, Hood Assembly 501
∂ρ#ελλερ
4 – 8BREAS MEDICALDoc. No.: 1523EN Issue: N-1
Page 24
Service Manual PV 501/501-2REPLACEMENT PARTS
4.9Exploded drawing, Hood Assembly 501-2
Doc. No.: 1523EN Issue: N-1BREAS MEDICAL4 – 9
Page 25
REPLACEMENT PARTSService Manual PV 501/501-2
4.10Part Number List
Part No.DescriptionNotes
2Base plate, complete
4 Motor unit, Maxon - complete
5 Hood, complete, SE
6 Complete, Eng
7 Complete, Ger
8 Rear panel, complete, Eng
9 Rear panel, complete, Ger
10 Hood, complete, Eng
11 Hood, complete, Ger
12 Rear panel, blank
13 Potentiometer box
14 Magnet - potentiometer box
15 Air-hose connection - patient exhale
17Side unit - motor bracket
18Motor plate
19Rear panel - Swedish text
20Hood
21Bottom plate
22Angle bracket - front / hood
24Cooler angle element
25Angle bracket - reading prong (l imit position)
26Ball screw - complete
27Motor bracket
29Panel - potentiometer panel Swedish text
30Panel - patient air Swedish text
31Rubber mount, shock absorbing (opto-switch)
32Pressure sensor
33Circuit board - CPU, PV501
34Circuit board- MDA, PV501
35Circuit board- potentiometer board
36Battery Ni/Cd or Ni/Mh alarm back-up
44Decal - "PATIENT PRESSURE"
45Decal - "ON / OFF"
50Cylomer - G12
51Cable marking - plus
52Cable marking - minus
53Transf ormer - 115V
54Opening - front
55Cross bar
56Spacing- motor bracket
57Software
4 – 10BREAS MEDICALDoc. No.: 1523EN Issue: N-1
Page 26
Service Manual PV 501/501-2REPLACEMENT PARTS
58Battery - large 24 V lead
64Hour counter - 24V
65Flexible tube - inlet (2.5")
66WAGO - 6-pole angled
67WAGO - 2-pole angled
68Reference pressure - valve
70Reference pressure - reducer sleeve
71Mains inlet
72Touch protection
73Air-filter box
74Air-hose connection - inhale
75Chassis pinning device XLR
76Fuse holder
77Screen guard - 60x60
78Filter holder
79Contact housing - black
80Contact housing - red
82Marking point - black
83Marking point - red
84Motor
85Motor connection
86Angle bracket- reading prong (column plate)
87Air-hose connection - drive pack
88Bellows side unit - fixed
89Angle element - limit position (plastic)
90Non-return valve
91Opto switch - limit position
92Opto switch - column plate
93Column plate
94Bellows
95WAGO - 8-pole angled
96Nipple - LCN-M5-PK3
97O-ring - air-hose connection drive pack
98Bellows side unit - moveable
99Nipple - SCN-M10-PK4
100Guide - large
101Guide - small
102Push-button board - on / off
103Push-button board - battery test
104Scale - graduated
105O-ring - air-hose connection inlet
106Panel - potentiometer panel English text
107Ribbon cable
108Instrument
Doc. No.: 1523EN Issue: N-1BREAS MEDICAL4 – 11
Page 27
REPLACEMENT PARTSService Manual PV 501/501-2
109Magnetic valve
110Cyl, isolator
111Decal - "ON / OFF"
112Decal - "PATIENT PRESSURE"
113Flexible tube - out (5")
114Angle pipe - out
115Pivot
116Arrow - large guide
117Arrow - small guide
118Cover - large guide
119Cover - small guide
121Nipple - CN-M5-PK3
122Rear panel - English text
123Insulation - side unit left
124Insulation - side unit right
125Insulation - bottom
126Insulation - hood
127Panel - patient air English text
128Mains cable
130Packing - bellows side unit
134Decal - "EIN / AUS"
135Decal - "BEATMUNGSDRÜCK"
136Rear panel - German text
137Panel - potentiometer panel German text
138Panel - patient air German text
139Battery pack, complete
140Text plate (IT)
141Panel, pot. panel (IT)
142Panel, patient air (IT)
143Text plate (SP)
144Panel, pot. panel (SP)
145Panel, patient air (SP)
146End cover for motor console, new version
147Motor console, new version
148Motor coupling, Rotex
149Bellows end cover fixed, new version
150Bellows end cover moving, new version
156Spacer sleeve, moving end cover
157Opto-switch, Optek
158Seal
160Shrink tubing - end position
184115/230 switch, PV501-2
215Rubber foot
220Hook - WAGO-terminal block
4 – 12BREAS MEDICALDoc. No.: 1523EN Issue: N-1
Page 28
Service Manual PV 501/501-2REPLACEMENT PARTS
248Filter, air (5 pcs/packet)
287Patient circuit (L-C)
378Motor (BEI)
381Decal "Pression de service" Fr
382Decal "Reglage/Arret" Fr
383Panel, pot. panel (Fr)
384Panel, patient air (Fr)
385Text plate (Fr)
386Decal "PATIENTTRYK" DK
387Decal "TIL/FRA" DK
388Panel, pot. panel (Dk)
389Text plate (DK)
390Decal "PATIENT DRUK" NL
391Decal "TIL/FRA" NL
392Panel, pot. panel (NL)
393Panel, patient air (NL)
394Text plate (NL)
395Bearing sleeve (PUR drive screw)
396Bearing sleeve for bearing (drive screw)
397Flange (PUR drive package)'
398Flange (drive package)
458Membrane, check valve
515Serial No. decal with EAN code
516Filter for machine ventilation
600Tape - potentiometer box
601Cellular rubber border
602Air-hose 4 x 6 - frosted
604Fuse TT 315mA
605Fuse - T160mA
606Fuse - F3,15A
611M3 x 8 with washer
612M3 - tooth nut
613Heat conducting paste - HTC10S
614Shrink tubing - 6.4 mm
615Shrink tubing- 3.2 mm
616Shrink tubing - 2.4 mm
617Shrink tubing - 12.7 mm
618Shrink tubing- 1.6 mm
632Clips for ribbon cable
633Strain relief - 4 mm
634Strain relief - 3.5 mm
635Cable tie brace
636Cable tie- 8"
637Cable tie- 4"
Doc. No.: 1523EN Issue: N-1BREAS MEDICAL4 – 13
Page 29
REPLACEMENT PARTSService Manual PV 501/501-2
638Cable tie- 14"
6395 mm plate spacing
640Y-connection
641Blue Festo tubing
642Mains cable straps
643Black screw 201 - RXS 3.5 x 19
644Clear tubing 9 x 12
645Clear tubing 4 x 7
646Clear tubing 3 x 6
647Clear tubing 3 x 5
648SPAX 4 x 20
649SPAX 3 x 12
650SGA 22
651Seeger - AK-26 , or Sgh 32
652M6 x 25 - bolt
653M6 x 22 - washer
654M6 - flat washer
655M6 - locking nut
656M5 x 80 F
657M5 x 50 F
658M5 x 20 socket head cap screw
659M5 x 16 F
660M5 x 16
661M5 - toothed plate connector
662M5 - flat washer
663M5 - nut
664M5 - locking nut
665M4 x 8 F
666M4 x 8
667M4 x 45
668M4 x 30
669M4 x 20
670M4 x 16 F
671M4 x 16
672M4 x 12
673M4 x 10 F
674M4 - wing nut
675M4 - toothed plate connector
676M4 - flat washer
677M4 - nut
678M4 - locking nut
679M3 x 8
680M3 x 6 F
681M3 x 6
4 – 14BREAS MEDICALDoc. No.: 1523EN Issue: N-1
Page 30
Service Manual PV 501/501-2REPLACEMENT PARTS
682M3 x 16 F
683M3 x 12 black socket head cap screw
684M3 x 12
685M3 x 10 F
686M3 x 10
687M3 - toothed plate connector
688M3 - flat plate
689M3 - nut
690M3 - threaded bar
691M2 x 10
692M10 - toothed plate connector
693KFXS 2.9 x 9.5 F
694Fibre washer
699Hook - for WAGO-terminal block
700Packing - battery base
701Plastic spacing - 4 x 15
702Plastic spacing - 4 x 10
703Plastic spacing - 3 x 10
704Blind plug - angle bracket hole
705Double-sided tape
706Mains cable discharger
708O-ring - potentiometer board
709Fuse - 1.25 A
711Cable tie - mini
713M4 x 6 socket head cap screw
714IC - MDA
716Pressure sensor
723Magnetic strip (0.01 m)
727Cable tie, driller
740Screw MFZ steel M4*10 Black, chrome Poz
741M3 x 6 F - black, PV501-2
753O-ring, check valve
795Metal cable tie
796Washer, support washer (drive unit)
797Sliding bearing sleeve (drive unit)
798Sliding bearing sleeve (PUR drive unit)
799Spacer sleeve (drive unit)
800Seal (drive unit)
1141Drive unit complete (BEI)
1145Rubber washer for transformer
1146Metal washer for transformer
1410Transformer, PV501-2
1636Opening front, PV501-2
1720Lock/keys, PV501-2
7.Magnetic valve: normally not powered see fig. 2.
8.Pressure sensor
Doc. No.: 1523EN Issue: N-1BREAS MEDICAL5 – 1
Page 33
FUNCTIONAL DIAGRAMSService Manual PV 501/501-2
5.2Functional block diagram
5 – 2BREAS MEDICALDoc. No.: 1523EN Issue: N-1
Page 34
Service Manual PV 501/501-2DISMANTLING AND ASSEMBLING THE PV 501/PV501-2
6DISMANTLING AND ASSEMBLING THE PV 501/PV501-2
6.1Opening the hood
•Remove the mains power cable.
•Loosen the three screws at the top of the rear panel and the three screws at the
front, see figure below.
ENTREE AIR PATIENTTEMPS UTILISATION
Changer après 500 heures de
fonctionnement ou quand nécess aire.
Afin d'autoriser la ch arge des batteries int ernes,
le PV 501 doit rester branché sur courant secteur au
moins 24 heures par mois.
NÄT
SÄKRING
T 315 mAL
PERSONVENTILATOR
NC
MAX 24 V. 1A
TRYCKALARM
0,1V/cm H O
REFROIDISSEMENT MACHINE
SÄKRING
F3.15 AL
2
EXT.BATTERI
24 VDC
DRIFT
NÄT/LADDNING
EXTERNT BATTERI
INTERNT BATTERI
TRIGGER
Från
Från
FrånFrån
TRIGGER
cm HO
2
LÅGT TRYCK
HÖGT TRYCKI/ EFREKVENS
cm HO
cm HO
cm HO
cm HOcm HO
2222
cm HOcm HO
2222
Förhållande
Förhållande
FörhållandeFörhållande
LÅGT TRYCK
HÖGT TRYCK
LÅG BATTERINIVÅ
FUNKTIONSFEL
TIDALVOLYM
Liter
Andetag/minut
Andetag/minutcm HO
Andetag/minutAndetag/minut
Felinställd
Felinställd
FelinställdFelinställd
PATIEN TTRY CKALARM
PATIENT-
DRIFT
LUFT
TILL
FRÅN
EXP.-
VENTIL
Fig. 6-1The screw positions that hold the hood
•Lift the hood slightl y upwards at t he rear so that t he a ngled par t of the hood cle ars
the rear panel. Move the hood forward approx. 20 mm and angle the hood up. Disconnect the hose from the patient air outlet.
Fig. 6-2Opening the hood
•Lay the hood down flat against the work surf ace. It is not necessary to loosen the
other tubes and cables to the front panel.
•Disconnect the white patient inlet air hose from the end cover of the bellows.
Doc. No.: 1523EN Issue: N-1BREAS MEDICAL6 – 1
Page 35
DISMANTLING AND ASSEMBLING THE PV 501/PV501-2Service Manual PV 501/501-2
6.2Replacing pushbutton boards
•Disconnect the ribbon cable (1) from the CPU board.
•Cut the cable tie (2) holding the green air tube (see figur e).
•Remove the four nuts that hold the t wo boards (indicated by ar rows) and move the
black ground wire terminal (3) to one side.
•Install the new boards in the reverse order.
1
J1
2
3
Fig. 6-3Removing the On/Off board
Installing a new ON/OFF board with 2 sec delay in units with a Serial No.
lower than 5925
Connect a wire from new board to J1.
Place “ON/OFF 2 sec delay” sticker above on/off switches on PV501.
NOTE! Disregard the text "CUT WIRE WHEN USING CABLE WITH CPU-CARD
REV.B OR OLDER" on the On/Off board. Do not cut any wire on the On/Off
board if it is Rev C.
6 – 2BREAS MEDICALDoc. No.: 1523EN Issue: N-1
Page 36
Service Manual PV 501/501-2DISMANTLING AND ASSEMBLING THE PV 501/PV501-2
6.3Replacing the CPU board
•Disconnect the three ribbon cable connectors (indicated by arrows in the figure
below).
Fig. 6-4Removing the CPU board
•Disconnect the air hose from the magnetic val ve. NOTE! Exercise care. Do not
pull too hard towards the valve without providing a counterhold for the valve. It
may be better to cut the hose.
•Disconnect the lower hose from the Y-connector.
Fig. 6-5Removing the CPU board
•Unscrew the 6 nuts that hold the board.
Doc. No.: 1523EN Issue: N-1BREAS MEDICAL6 – 3
Page 37
DISMANTLING AND ASSEMBLING THE PV 501/PV501-2Service Manual PV 501/501-2
Fig. 6-6Removing the CPU board
•Lift out the board.
•Reassemble in reverse order.
6.4Replacing the potentiometer board
•Remove the six setting knobs by first removing the plastic cap for the knob and
then, for the two large knobs unscrew their retaining nuts, and for the smaller
knobs unscrew the screws.
AC/CHARGING
EXTERNAL BATTERY
INTERNAL BATTERY
TRIGGER
Off
Off
OffOff
LOW PRESSURE
cm HO
cm HOcm HO
2222
POWER
TRIGGER
cmH O
2
HIGH PRESSURE
cm HO
cm HO
cm HOcm HO
LOW PRESSURE
HIGH PRESSURE
LOW BATTERY
FUNCTION FAILURE
TIDAL VOLU ME
Signal
Signal
SignalSignal
2222
Litre
I/EBREATH RATE
Ratio
Ratio
RatioRatio
Setting error
Setting error
Setting errorSetting error
BPM
BPMcm HO
BPMBPM
PATIENT PRESSUREALARMS
EXHALATION
VALVE
PATIEN T
AIR
Fig. 6-7Removing the setting knobs
•Disconnect the ribbon cable from the CPU board.
•Disconnect both the air tubes from the magnetic valve.
POWER
ON
OFF
Fig. 6-8Removing the Potentiometer board
•Unscrew the 4 nuts holding the board.
6 – 4BREAS MEDICALDoc. No.: 1523EN Issue: N-1
Page 38
Service Manual PV 501/501-2DISMANTLING AND ASSEMBLING THE PV 501/PV501-2
Fig. 6-9Removing the Potentiometer board
•Lift out the board.
•Reassemble in the reverse order.
6.5Opening the rear panel
•Open the hood, if not already opened.
•Remove the three screws at the bottom of the rear panel and carefully lower it
down onto the work surface.
•Release the two securing locks for the ribbon cable connector and disconnect it.
•Disconnect the air inlet h ose from the motor unit.
Fig. 6-10Opening the rear panel
•To remove the cross bar for the side panel s, first pull it up a little and then turn it so
that the slots disengage. Loosen the rod from one side panel at a time.
6.6Replacing the MDA board
•Open the hood and rear panel.
•Disconnect the CN6, CN7, CN8, CN9, CN10 connectors.
Doc. No.: 1523EN Issue: N-1BREAS MEDICAL6 – 5
Page 39
DISMANTLING AND ASSEMBLING THE PV 501/PV501-2Service Manual PV 501/501-2
Fig. 6-11Removing the MDA board
•Unscrew the 6 M3 and the 4 M8(on PV501) nuts while holding the board.
Fig. 6-12Removing the MDA board
•Lift out the board.
•Reassemble in reverse order.
6.7Removing the motor unit
•Open the hood and rear panel.
(only PV501)
6 – 6BREAS MEDICALDoc. No.: 1523EN Issue: N-1
Page 40
Service Manual PV 501/501-2DISMANTLING AND ASSEMBLING THE PV 501/PV501-2
12
Fig. 6-13Removing the motor unit
•Remove the cross bar (1 in figure).
•Disconnect the air hoses from the bellows end cover.
•Loosen the encoder c able connecto r (2) from the MDA circuit board inside the rear
panel.
•Loosen the motor cable connector (3) from the MDA circu it board inside the rear
panel.
•Disconnect the red tube (4) to the exhalation valve on the motor unit.
•Loosen the four nuts (M4) ( indicated by the four bol d arrows in the figu re) from th e
motor unit's base plate.
•Lift the motor unit straight up.
6.8Reassembling the motor unit
•Before reassembling the motor unit, check if the internal batteries need to be
replaced as this is best done with the motor unit removed, see 3.8. 13.
•Mount the motor unit and fasten it with the four nuts (M4) at the bottom of the
motor unit.
4
3
•Connect the encoder cable connector to the MDA circuit board.
•Connect the motor cable connector to the MDA circuit board.
•Connect the red tube to the exhalation valve on the motor unit.
•Start the ventilator.
•Check that everything functions normally.
6.9Assembling the hood and rear panel
•Fit the cross bar between the side panels.
Doc. No.: 1523EN Issue: N-1BREAS MEDICAL6 – 7
Page 41
DISMANTLING AND ASSEMBLING THE PV 501/PV501-2Service Manual PV 501/501-2
•Raise the rear panel and fasten it with the lower screws. Check that no cables or
hoses are pinched. Do not forget to connect the ribbon cable.
•Fit the air inlet hose between the rear panel and motor unit.
•Move the hood up so that the patient outlet ai r hose can be fit ted to the motor unit .
•Check that the ribbon cable does not lay against the bellows. If necessary , press
the ribbon cable down at the back and front.
•Insert the screws in the front panel but do not tighten yet.
•Tighten the top screws for the rear panel.
•Tighten the screws on the front panel at the same time as the hood is pressed
downwards.
•Check that everything is assembled correctly.
6 – 8BREAS MEDICALDoc. No.: 1523EN Issue: N-1
Page 42
Service Manual PV 501/501-2MOTOR UNIT
7MOTOR UNIT
7.1Construction
12345,6 7
18
17
16
15
14
13
8
9
12
Fig. 7-1Main parts of the motor unit
Pos. No.
(1)Motor bracket spacer bar
(2)Angle element-limit position
(3)Bellows clamp (steel hose clamp)
(4)Bellows
(5)Patient air inlet, and check valve
(6)Patient air outlet, and check valve
(7)Motor
(8)Motor connection
(9)Air nipple
(10)Rubber suspension for ball bearings
(11)Ball bearings
(12)Motor bracket
(13)Ball screw complete with bearing
(14)Oil bronze bearing with rubber suspension
(15)Slotted disc
(16)Opto-switch (encoder)
(17)Opto-switch (limit position sensor)
(18)Side plate
Description
10,11
See also exploded diagrams in Chapter 4.
Doc. No.: 1523EN Issue: N-1BREAS MEDICAL7 – 1
Page 43
MOTOR UNITService Manual PV 501/501-2
7.2Lubricating the ball screw
•Turn the ball screw with the fingers, this is easiest done at the screw/motor connection(1 in figure). Keep turning until the bellows is completely compressed.
•Wipe off the old grease from the ball screw (2) if necessary e.g. if it is dirty.
2
1
Fig. 7-2Lubricating the ball screw
•Apply a thin layer of grease over the entire length of the screw. Use only BREAS
grease type 283 AZ, part no. 000 557.
•Run the motor a couple of cycles to work the grease into the unit. Remove any
excess grease.
•If necessary, repeat greasing once more.
7 – 2BREAS MEDICALDoc. No.: 1523EN Issue: N-1
Page 44
Service Manual PV 501/501-2MOTOR UNIT
7.3Replacing the membranes in the check valves
•The membranes must be changed at each maintenance service.
Material required:
Service kit for check valves part. no: 002 123, which contains: 2 membranes and
2 O-rings.
3
3
(8x)
2
1
1
Fig. 7-3Replacing the check valves
•Remove the hoses (1).
•Remove the four screws (2) (M3x8) holding each hose connection.
•Remove the old membranes together with the valve seats and O-rings (3).
•Wipe the valve seats in the end cover clean using a damp cloth.
•Carefully assemble the new membranes.
•Make sure that the new membranes lay flat against their seats.
•Fit new O-rings.
•Fit the hose connections.
•Check the function of the check valves by creating a negative pressure in the inlet
hose and a positive pressure in the outlet hose. No air should be able to pass. (Do
not forget to plug the tube connector for the exh. valve).
7.4Hoses and bellows, leakage check
Check the function of the outlet check valves by creating a positive pressure in
the exhalation hose (do not forget to block the tube nipple).
Plug the hose for air outlet and the small tube connector. Create a pressure in the
inlet hose, approx. 40 cm H
for the pressure to drop from 40 to 20 cm H
not, check for leakage or replace the motor unit.
Doc. No.: 1523EN Issue: N-1BREAS MEDICAL7 – 3
O. Check that no leakage is heard. Measure the time
2
O, it should take at least 10 sec. If
2
Page 45
MOTOR UNITService Manual PV 501/501-2
7.5Wiring diagram, Motor unit
8 POLE WAGO CONNECTOR
RED
BLACK
WHITE
BLACK
GP1L03
TO CN7 ON MDA BOARD
8-POLE WAGO CONNECTOR
OPB930W51
WHITE
BLUE
RED
BLACK
GREEN
MOTOR BEI
YELLOW
GRAY
ORANGE
BLUE
BROWN
GREEN
BLACK
RED
MOTOR MAXON
TO CN10 ON THE MDA BOARD
GREEN
BLUE
RED/GRAY
BLACK/GRAY
WHITE/GRAY
RED
BLACK
WHITE
8-POLIG WAGOKONTAKT
TILL CN10 PÅ MDA-KORTET
7 – 4BREAS MEDICALDoc. No.: 1523EN Issue: N-1
Page 46
Service Manual PV 501/501-2MOTOR UNIT
Doc. No.: 1523EN Issue: N-1BREAS MEDICAL7 – 5
Page 47
MOTOR UNITService Manual PV 501/501-2
7 – 6BREAS MEDICALDoc. No.: 1523EN Issue: N-1
Page 48
Service Manual PV 501/501-2MOTOR UNIT
Doc. No.: 1523EN Issue: N-1BREAS MEDICAL7 – 7
Page 49
Service Manual PV 501/501-2ELECTRONICS
8ELECTRONICS
8.1Function, Construction
The electronics, mechanics, pneumatics and opto-electronics are fully integrated in the
PV 501/PV 501-2 ventilators.
To understand the function of the electronics included in the Breas PV 501 you must be
able to operate the ventilator, have studied the air flow diagram and acquainted yourself
with its mechanical construction.
The following block diagram (PV6T1) outlines how the electronic system is constructed
and how it is connected to the other components.
Fig. 8-1Functional Diagram
Doc. No.: 1523EN Issue: N-1BREAS MEDICAL8 – 1
Page 50
ELECTRONICSService Manual PV 501/501-2
8.2PV 501/PV 501-2's circuit boards
The PV 501/PV 501-2’s electronics comprises 5 circuit boards:
2
3
4
1
Fig. 8-2Location of the ciruit boards
Pos. No.
1Pushbutton board “On/Off buttons”
1Pushbutton board “Mute/Battery test buttons”
2CPU board “front panel”
3Pot. board “front inside opening”
4MDA-board “rear panel”
Location:
8 – 2BREAS MEDICALDoc. No.: 1523EN Issue: N-1
Page 51
Service Manual PV 501/501-2ELECTRONICS
8.3Pushbutton board
Fig. 8-3Pushbutton board
The membrane switches for ON, OFF, Alarm mute and Battery test are located on the
pushbutton board together with LEDs that indicate that the ventilator is switched ON and
that Alarm mute has been activated.
8.4Potentiometer board
PZ1
Front viewRear view
Fig. 8-4Potentiometer board
The potentiometers (P1 - P6) are located here. They are used for setting the volume, I/
E, frequency, trigger, high pressure, low pressure and alarm volume. The alarm buzzer
PZ1 is also located here, along with the MV2 magnet valve which is normally powered,
but switches if the high pressure alarm is activated. The exhalation valve then opens as
a safety precaution. See the air flow diagram.
MV2
Doc. No.: 1523EN Issue: N-1BREAS MEDICAL8 – 3
Page 52
ELECTRONICSService Manual PV 501/501-2
8.5MDA board
The MDA board handles the voltage supply, motor control and the charging of the internal battery.
R 120
U 32
F 1
R 127
Fig. 8-5MDA board
The MDA board is supplied with 24 V AC via CN5 from the toroidal core transformer.
The current is rectified in B1 and supplies the voltage regulators U36 and U39 with voltage. At power on, the other regulators are supplied via the K2 relay.
The motor is a brushless DC motor. The U34 and U35 circuits produce current and voltage for the motor’s three windings causing it to rotate. The maximum motor current is
limited by R120. To keep track of the number of rotations and how fast the motor is running, a pulse sensor has been mounted on the ball screw, which is an extension of the
motor shaft. (The pulse sensor consists of a slotted disk with 96 slots and an optoswitch).
The U32 microprocessor receives information from the pulse sensor, and regulates the
velocity and the number of rotations the motor is running at. The home position of the
moving end of the bellows unit is also registered by U32 with the help of an opto-switch.
The internal batteries are connected to the MDA board. Fuse F1 protects these batter-
8 – 4BREAS MEDICALDoc. No.: 1523EN Issue: N-1
Page 53
Service Manual PV 501/501-2ELECTRONICS
ies.
The internal battery (lead) charger consists of a voltage regulator U36, where the
charge voltage is regulated by R127 and a current limiter equipped with transistors Q25
and Q27. When the battery voltage drops below 22.7 V during operation, the low battery
level alarm is activated. If the battery level reaches 30 VDC, a thyristor connects the battery to ground until fuse F1 blows. This prevents the battery from creating explosive gas.
The ventilator stops and sounds an alarm when the voltage drops below 22.0 V and the
bi-stable relay K5 cuts out. The ventilator must then be connected to the mains in order
for K5 to switch back again, and for the batteries to recharge.
When the external battery voltage drops below 22.5 V during operation, the ventilator
switches over to internal battery power and gives a short warning alarm.
8.5.1Checking the pressure transducer offset and gain
Connect a voltmeter to the pressure outlet on the rear panel, see Op. manual. Range: 02 VDC. Set the frequency to 9 breaths/min and start the ventilator. Wait approx. 10 min
for the pressure to stabilise. Adjust the offset between inhalations to 1,00-1,01 volt with
the trim pot R21 marked OFFS. Do not connect a "counter pressure" to the patient air
outlet. The trigger's offset reference automatically follows the adjustment of the pressure
transducers offset level in order to always calculate the adjustment of the trigger setting.
The R24 potentiometer is used to adjust the gain which should be 4.0 V at 30 mbar.
Note! If it was necessary to adjust the gain, then re-check the offset again.
R 21
R 24
Fig. 8-6Adjusting the offset and gain
8.5.2Checking the instrument accuracy
Check, using a manometer connected to the exh. valve connection, that the indicating
instrument gives the correct indication and that the needle moves "softly".
NOTE! Make small pressure changes, the indicating instrument should be
"slower" compared with the manometer.
Doc. No.: 1523EN Issue: N-1BREAS MEDICAL8 – 5
Page 54
ELECTRONICSService Manual PV 501/501-2
8.6CPU board
MV1
G1
BT1
BT2
R 21
R 24
U 3
U 6
Fig. 8-7CPU board
The “master” microprocessor U3 and its software capsule EPROM U6 are located on
the CPU board together with the pressure gauge, the pressure sensor, the LEDs indicating the voltage source as well as the alarm, alarm battery and magnetic valve MV1.
U3 takes a reading from the potentiometers on the potentiometer board for frequency,
volume, I/E, trigger and alarm limits. It also reads the “alarm mute” button, patient pressure and the voltages for internal battery DC, external DC, alarm battery DC and the DC
generated by the mains voltage.
U3 then controls and monitors the ventilator including the LEDs at the front, the form of
breathing cycle and all the alarms. Besides the pressure and voltage monitoring alarms,
there is also a system for monitoring communication between U3 and U32, e.g. that
inhalation occurs within a specified time, and that failure to do so causes an alarm.
Many of these alarms come in the form of a combination of the LEDs in the front panel
lighting. For more information, see Chapter 9 Fault tracing.
The display instrument for patient pressure is a moving coil instrument. It operates quite
slowly (having a certain inertia) which means the display is slow to react when very fast
changes in pressure occur. When servicing and fault tracing it is advisable to use an
analogue manometer.
The G1 pressure sensor is a semiconductor and the signal emitted is amplified by U14.
The R24 potentiometer adjusts the amplification and R21 adjusts the offset.
The magnetic valve MV1 is used to connect the two outlets on the pressure sensor with
reference pressure outlet on the rear panel(or front panel from serial no 5204). When
the trigger is activated, a zero pressure calibration is implemented between each inhalation for the first five minutes, and then once every minute with the help of MV1. See the
pneumatic diagram.
The BT1 & BT2 alarm batteries are of the Ni/CD or Ni/Mh type and are monitored by U3.
If are dead or out of function, the ventilator would not start without sounding an alarm
and displaying an error code, see Chapter 9. The batteries are charged as soon as the
ventilator is connected to mains supply.
8 – 6BREAS MEDICALDoc. No.: 1523EN Issue: N-1
Page 55
Service Manual PV 501/501-2ELECTRONICS
8.7Terminal diagram, test points
Fig. 8-8Test points
Doc. No.: 1523EN Issue: N-1BREAS MEDICAL8 – 7
Page 56
ELECTRONICSService Manual PV 501/501-2
MDA board, PV 501
Tes t p ointVo l t a g e n am eStatusMeasured valueAdj. pot.
TP 1GND
TP 2Unregulatedon24-38 VDC–
TP 3Int. Pb battery chargeoff, mains28.2 VDCR127
TP 424 Von24 VDC–
TP 512 Von12 VDC–
TP 65 Von5 VDC–
MDA board, PV 501-2
Tes t p ointVo l t a g e n am eStatusMeasured valueAdj. pot.
TC 1:1Internal battery raw28V–
TC 1:2Int. Pb battery chargeoff, mains28.2 VDC–
TC 1:3DC transf. after relay K6on30-40 VDC–
TC 1:4DC transf. before relay K6on30-40 VDC–
TC 1:5Over current limiton0-7 VDC–
TC 1:624 V stab (+24)on24 VDC–
TC 1:7Ground via 10k pressure -on0 VDC–
TC 1:8Pressure +on0-8 VDC–
TC 1:9Direction motoron0-5 VDC–
TC 1:10Alarm relayon0-24 VDC–
TC 1:11Encoder signal (count)on0-5 VDC–
TC 1:12Homeposition motoruniton0-5 VDC–
TC 1:135 V stab (+5s)on5 VDC–
TC 1:14External battery (Extdc)on24 VDC–
TC 1:15Gnd–––
TC 1:16Gnd–––
8 – 8BREAS MEDICALDoc. No.: 1523EN Issue: N-1
Page 57
Service Manual PV 501/501-2ELECTRONICS
CPU board, PV 501/PV501-2
Test point Vo ltage nam eStatusMeasured valueAdj. pot.
The motor unit’s electronic components consists of a motor and two opto-switches.
OPB930W51
Fig. 8-9Measuring points on the motor unit
The above diagram shows how the function of the opto-switch can be measured.
Opto-switch limitConnect a DMM (digital multi meter) between pins 3 &
4.
Start the ventilator.
At limit position: approx. 5 V
Non limit position: approx. 0.7 V
Opto-switch pulse sensorConnect an oscilloscope between pins 6 & 8 on PV 501
or on TC1 between pin 11 & 16 on PV 501-2.
Set the ventilator as follows:
Tidal volume: 1.2 l
Frequency:12
I/E1:2
Trigger -1
Set the oscilloscope to 1 V/div. and 0.2 ms/div.
Start the ventilator.
8 – 10BREAS MEDICALDoc. No.: 1523EN Issue: N-1
Page 59
Service Manual PV 501/501-2ELECTRONICS
8.9Internal Battery test
The batteries should be fully charged or charged for at least 12 hours with the ventilator
turned off before this test is carried out. The batteries, when fully charged, should operate the PV 501/501-2 for at least 2 hours, if not the batteries must be replaced.
Adjust the settings to: Tidal volume:0,8L
Frequency:14
I/E:1:2
Connect a 1 litre test lung to the exhalation valve.
Note the time on the running time meter.
Turn the ventilator on, check that the LED for "Internal battery" is flashing.
If necessary adjust the tidal volume to allow the patient pressure to reach approx.
20 cm H
Let the ventilator run until the battery charge drops too low. Check that alarm for "Low
battery level" is given for 15 minutes before the ventilator switches off.
Check that the alarm continues to sound 5 minutes after the ventilator has switched off.
O.
2
With the ventilator turned off, charge the battery for at least 12 hours. Check that the
batteries are fully charged by pressing "Batt. test" (with the ventilator running and the
mains disconnected). The needle should indicate in the upper part of the green sector.
8.10Replacing the internal power batteries (lead acid cells)
The internal power batteries must be replaced:
– If the ventilator is often run from the internal batteries, they should be replaced
every 24 months or when required.
– If the batteries, when fully charged, fail to run the ventilator for at least 2 hours.
•Remove the motor unit by loosening the four nuts holding the base plate.
•Disconnect the connector to the encoders on the rear panel.
•Loosen the motor cable connector.
•Loosen the thin red tube for the exhalation valve.
•Lift the motor unit straight up.
•Remove the two nuts for the battery fastening bar. Only disconnect the two cables
from the battery that are connected to the circuit board.
•Lift out the batteries.
•Install the new batteries in place and the fastening bar.
•Connect the wires from the circuit board.
•Install the motor unit.
•Make a test start and check that everything appears to function normally. Turn off
the ventilator.
•Charge the batteries for at least 12 hours and then perform a battery test, see
Section 8.9.
Doc. No.: 1523EN Issue: N-1BREAS MEDICAL8 – 11
Page 60
ELECTRONICSService Manual PV 501/501-2
8.11Replacing the alarm batteries (Ni-Cd)
The alarm batteries are contained on the CPU board in the hood (see figure). The internal batteries normally provide power for the alarm function, but should they become discharged, the alarm battery will provide back-up power. The alarm battery must be
replaced every 5th year counted from date of delivery or when it was last replaced.
Alarm batteries
BT1
BT2
Fig. 8-10Replacing the alarm batteries
To replace the alarm batteries:
•Open the connector locks and disconnect the three ribbon cables from the CPU
board.
•Mark the hoses A, B and C using a marker pen, then disconnect these.
•Remove the five nuts that hold the board.
•Remove the board.
•Cut the two "pins" on battery marked "Y", closest to the edge board. Unsolder the
"Y"-battery from the minus pole.
•Cut the other battery pin.
•Remove the old battery pins.
•Assemble the new batteries and solder these in place.
•Reassemble the circuit board, hoses, cable ties and ribbon cables.
•Make a test start. PV 501 checks at each start cycle the condition of the alarm batteries. If there is a fault with the alarm battery, the ventilator will not start and the
LEDs for "FUNCTION FAULT" and "TRIGGER" will light. If this happens after
replacing the battery, let the ventilator remain connected to the mains for a few
hours for the battery to be charged. Thereafter, make a new test start.
8 – 12BREAS MEDICALDoc. No.: 1523EN Issue: N-1
Page 61
Service Manual PV 501/501-2ELECTRONICS
8.12Electrical safety
Electrical safety measurements should be performed according to IEC 601, but the isolation resistance can be measured instead of the current test as according to the stipulated requirements in the standard.
The measurements can be performed with an automatic electricity safety tester. All tests
shall be performed according to class II type BF.
8.12.1 Mains
Register the mains value. As the current is directly dependant of the mains outlet being
used it is necessary to register the mains at each maintenance. This allows values registered from the same machine to be compared at different occasions.
8.12.2 Isolation
The isolation resistance is measured with 500 VDC current source. Preferably, the plus
is connected to both mains pins and minus to the exterior or patient connections. The
measurement performed at the delivery check gives the reference value that maintenance measurements are compared with. If this has not be done, the value of the isolation resistance should be > 20M .
8.12.3 Leakage current
Leakage current is registered from various parts of the machine through a RC-circuit to
earth. Measurements are made in normal cases (NC) and at single fault condition
(SFC). Reverse voltage polarity mains and note the lowest value.
Leakage to the earth should not exceed the stated limit value.
8.12.4 Leakage current from the cabinet
Leakage current from the cabinet.
Measure at an unvarnished part, i.e. a screw head.
Limit value:NC< 0,1mA
SFC< 0,5mA
Break neutral for SFC.
8.12.5 Patient leakage
Measured between patient connection and earth.
Limit value:NC. < 0,1mA
SFC.< 0,5mA
Break neutral for SFC.
8.12.6 Leakage current at mains voltage at patient connected part
This test should only be performed with the automatic electrical safety tester. See the
testing security instruction.
The opto-switch is not registering home position.
Internal air-hose blocked.
Gauge broken.
Check the air-hose, mask and
exhalation valve for leakage.
Run a leakage check.
Replace filter.
Check the function of the optoswitch.
If the voltage levels are correct
replace MDA board.
If voltage for pins 1 & 2 in CN7 is
> 1.1 V replace the motor unit.
Check the air-hose and nipples.
Using a voltmeter connected
to pressure output in the rear
panel, check to see if a change
in pressure results in a change
in voltage.
P=0 cm H
30cm H
O gives 1V,
2
O gives 4V.
2
If OK, replace the gauge.
If not OK, replace CPU board.
SEE
CH.
2
3
3
8
5
8
3 – 2BREAS MEDICALDoc. No.: 1523EN Issue: N-1
Page 100
Service Manual PV 501/501-2FAULT TRACING
9.2Error codes
Should there be a fault in the function of the PV 501 an error code is created and
stored in the PV 501's memory. At the same time, the LED for ”Function failure ”
flashes and the cause of the error is indicated by a combination of LEDs lighting
as shown in the table below. Note! The LED for Charging is not used
bination.
The PV 501's memory allows a maximum of 5 error codes to be stored. Should a
sixth error occur, the earliest error code stored will be deleted.
in any com-
9.3Recalling error codes
With the ventilator switched off, press and hold the Alarm mute button and start
the PV 501. This procedure will provide electrical power to the PV 501, but the
motor unit is disabled. All the LED indicators (except Mains/Charging) will briefly
flash.
Press the Alarm mute button again to see the first (latest) error code (an audible
signal will be heard consisting of one beep). Press again to see the second error
code (an audible signal consisting of two beeps is given to indicate the second
error code). As the contents of the memory is read the audible alarm will beep the
number of times corresponding to the error code number (1 through 5). When the
last error code is reached all the LED indicators (except Mains/Charging) will light
and the LED in the Alarm mute button will light.
To see the error codes again, the PV 501 must be switched off and then restarted
as described above.
9.4Erasing the error code memory
Step through all the error codes stored, as described above, until the last code is
displayed. If there less than five errors stored, continue to press the alarm mute
button until five beeps are heard. The memory can then be erased by pressing
and holding the Alarm mute button.
All the LEDs will light to indicate the memory has been deleted.
Doc. No.: 1523EN Issue: N-1BREAS MEDICAL3 – 3
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.