Measured Catheter Tip temperature is either less than
15°C or greater than 95°C. (This code may also be dis-
played when a catheter is disconnected from the pod.)
• Press RF Power button to clear the Diagnostic
code.
• Ensure that the catheter is in-vivo before connecting the extension cable.
• Verify normal body temperature reading from the
in-vivo catheter.
• Restart the Maestro 4000 using main power
switch.
• Ensure that the POD is not covered, has adequate
ventilation and is not exposed to sources of either
heat or cold.
RF delivery was attempted when the measured temperature is greater than either 43°C or the temperature
set point (whichever is lower), within 30 seconds of
the previous RF delivery.
• Press RF Power button to clear the Diagnostic
code.
• Ensure that the catheter is in-vivo before connecting the extension cable.
• Verify normal body temperature reading from the
in-vivo catheter.
• Verify that an appropriate Temperature set point,
Power set point and Control Mode are selected for
the desired catheter.
• If temperature changes with agitation, replace
suspected components one at a time.
• Restart the Maestro 4000 using the main power
switch.
• Ensure that the POD is not covered, has adequate
ventilation and is not exposed to sources of either
heat or cold.
The Maestro 4000 detects an issue with the pod
connection.
• Press RF Power button to clear the Diagnostic code.
• Disconnect the catheter extension cable from the POD.
• Power down the Maestro 4000 using the main
power switch.
• Verify that the POD Cable is connected correctly
to the front of the Maestro 4000 Patient Isolated
Connector.
• Ensure that there are no kinks or tight loops in the
cable. Switch on the Maestro 4000 using main
power switch and allow Self Test to complete.
Once Self Test is complete and “Finding (cont...)
D06-TEMP OUT OF RANGE [15C-95C]
D07-TEMPERATURE TOO HIGH
D08-CHECK POD
Cause:
Action:
Cause:
Action:
• (Cont...) Catheter” is displayed on the screen,
connect the catheter extension cable to the POD.
The Maestro 4000 should then return to a Ready/
Standby state, displaying both real-time temperature and impedance. If this does not resolve the
issue, replace the POD.
Generator doesn’t recognize catheter ID.
• Press RF Power button to clear the Diagnostic
code.
• Replace catheter cable.
• If error recurs replace catheter.
• If error recurs replace pod.
There is a communications problem between the
Remote and Maestro 4000 or the Remote was disconnected or connected during RF.
• Check cable and connections between the Maestro
4000 and Remote, tightening the thumb screws on
the connectors. Replace cable or Remote if damage is suspected.
A diagnostic condition occurred on the MetriQ or there
is a communication error between the Maestro 4000
and the MetriQ.
• Resolve problem on pump and clear the diagnostic condition there, then press RF Power to clear
the error on the Maestro 4000.
• Check the communications cable connections
between the Maestro 4000 and MetriQ. Tighten
thumb screws on the connectors.
• Replace the communications cable if damage is
suspected.
D08-CHECK POD (Cont.)
D10-INVALID CATHETER ID
D11-CHECK REMOTE
D12-CHECK PUMP
Cause:
Action:
Cause:
Action:
Cause:
Action:
Cause:
Action:
All ci ted trademarks are t he property of the ir respec tive ow ners. C AUTION: The law rest ricts t hese devices t o sales by or on the
order o f a physician. Indica tions, co ntrain dications , warnings and instructions for use can be found in the product l abeling
supplied with e ach device. Information for the use onl y in countri es with applicable h ealth au thority produ ct regis trations .
EP-2 5260 3-AA J UN20 14 Print ed in Germany by medical vision.
www.bostonscientific.com
© 2014 Boston Scientific Corporation
or its affiliates. All rights reserved.
DINEP2210EA