Boston Scientific L211, L231, L310, L311, L331 Technical Manual

...
MRI TECHNICAL GUIDE
IMAGEREADY™MR CONDITIONAL PACING SYSTEM
REF L110, L111, L131, L210, L211, L231, L310, L311, L331, 4456, 4457, 4458, 4459, 4469, 4470, 4471, 4472, 4473, 4474, 4479, 4480, 6220, 6221, 6402, 7145, 7731, 7732, 7735, 7736, 7740, 7741, 7742
ABOUT THIS MANUAL
This manual is intended for use by physicians and other health care professionals (HCPs) involved in managing patients with an ImageReady MR Conditional Pacing System, as well as radiologists and other HCPs involved in performing magnetic resonance imaging (MRI) scans on such patients.
NOTE: For the purposes of this Technical Guide, MRI is used as a general term and encompasses all MR-based clinical imaging activities. In addition, information in this guide applies only to
Read this manual in its entirety before scanning patients who are implanted with an ImageReady MR Conditional Pacing System.
This manual contains:
Information about ImageReady MR Conditional Pacing Systems
Information about ImageReady Pacing System patients who can and cannot undergo an MRI scan and the Conditions of Use that must be met in order for an MRI scan to be performed
Instructions for carrying out an MRI scan on ImageReady Pacing System patients
How to use this manual:
1
H MRI (Proton MRI) scanners.
1. Refer to the patient’s records to locate model numbers for all components of the patient’s
implanted system.
2. Refer to "System Configuration for 1.5 T" on page 1-3 and "System Configuration for 3 T" on
page 1-3 to determine if all components of the patient’s implanted system are found within the tables. If any of the components cannot be found within the tables, the system is not an ImageReady MR Conditional Pacing System.
NOTE: Multiple Boston Scientific ImageReady MRI Technical Guides are available based on therapy type, for example, a pacing system versus a defibrillation system. If a particular pulse generator model is not represented in this manual, refer to the other Boston Scientific ImageReady MRI Technical Guides. If a particular model is not represented in any Boston Scientific ImageReady MRI Technical Guide, the patient’s implanted system is not an ImageReady MR Conditional system.
Refer to the Physician’s Technical Manual, Reference Guide, Leads Manual, Clinician Manual, or Programmer Operator’s Manual for detailed information about non-MRI aspects of implantation, features, programming, and use of the components of the Pacing System.
The following are trademarks of Boston Scientific Corporation or its affiliates:
ACCOLADE, ESSENTIO, FINELINE, IMAGEREADY, INGEVITY, LATITUDE, PaceSafe,
PROPONENT, ZOOM, ZOOMVIEW.
TABLE OF CONTENTS
INTRODUCTION TO MR CONDITIONAL PACING ..............................................................1-1
CHAPTER 1
System Description ............................................................................................................1-2
Valid Combinations of Pulse Generators and Leads to Use in 1.5 Tesla and 3 Tesla
Environments ..............................................................................................................1-2
System Configuration for 1.5 T ......................................................................................1-3
System Configuration for 3 T .........................................................................................1-3
MRI Conditions of Use........................................................................................................1-4
Cardiology ...................................................................................................................1-4
Radiology ....................................................................................................................1-4
Conditions for Scanning......................................................................................................1-5
MRI Protection Mode........................................................................................................ 1-11
MRI Basic Concepts......................................................................................................... 1-11
X-ray or Fluoroscopic Identification.................................................................................... 1-12
MR Conditional Pacing System Warnings and Precautions.................................................. 1-12
General .....................................................................................................................1-12
Programming Considerations ...................................................................................... 1-13
Safety Mode .............................................................................................................. 1-13
MRI Site Zone III Exclusions ....................................................................................... 1-14
Precautions ...............................................................................................................1-14
Potential Adverse Events.................................................................................................. 1-14
MRI SCAN PROCEDURE PROTOCOL...............................................................................2-1
CHAPTER 2
Patient Flow.......................................................................................................................2-2
MRI Protection Mode General Information............................................................................2-2
Pre-Scan Activities .............................................................................................................2-4
1. Programming the Pulse Generator for a Scan .............................................................2-5
2. Confirming MRI Scanner Settings and Configuration ................................................. 2-13
3. Preparing the Patient for the Scan ........................................................................... 2-13
During the Scan ............................................................................................................... 2-14
After the Scan.................................................................................................................. 2-14
CARDIOLOGY CHECKLIST FOR THE IMAGEREADY PACING SYSTEM .......................... A-1
APPENDIX A
RADIOLOGY CHECKLIST FOR THE IMAGEREADY PACING SYSTEM ............................. B-1
APPENDIX B
IMAGEREADY PACING SYSTEM COMPONENTS FOR 1.5 T AND 3 T............................... C-1
APPENDIX C
SYMBOLS ON PACKAGING............................................................................................. D-1
APPENDIX D
INTRODUCTION TO MR CONDITIONAL PACING
CHAPTER 1
This chapter contains the following topics:
“System Description” on page 1-2
“MRI Conditions of Use” on page 1-4
“Conditions for Scanning” on page 1-5
“MRI Protection Mode” on page 1-11
“MRI Basic Concepts” on page 1-11
“X-ray or Fluoroscopic Identification” on page 1-12
“MR Conditional Pacing System Warnings and Precautions” on page 1-12
“Potential Adverse Events” on page 1-14
1-1
1-2
Introduction to MR Conditional Pacing

System Description

SYSTEM DESCRIPTION
An ImageReady MR Conditional Pacing System consists of specific Boston Scientific model components including pulse generators, leads, accessories, the Programmer/Recorder/Monitor (PRM), and the PRM Software Application. For the model numbers of MR Conditional Pacing System components, see Table 1–2 System Configuration for 1.5 T on page 1-3 and Table 1–3 System Configuration for 3 T on page 1-3.
The ImageReady MR Conditional Pacing Systems were created specifically as a system for use with MRI scans performed under the Conditions of Use described in this Technical Guide. The pulse generator design has minimized use of ferromagnetic materials, which can interact with the fields generated during a typical MRI scan, and the circuits have been designed to tolerate voltages that may be induced during scans. Any part of the body may be imaged. Boston Scientific MR Conditional pulse generators and leads, when used together, have mitigated risks associated with MRI scans as compared to conventional pulse generators and leads. The implanted system, as opposed to its constituent parts, is determined to have the status of MR Conditional as described in ASTM F2503:2008. Additionally, an MRI Protection Mode has been created for use during the scan. MRI Protection Mode modifies the behavior of the pulse generator and has been designed to accommodate the MRI scanner electromagnetic environment. A Time-out feature can be programmed to allow automatic exit from MRI Protection Mode after a set number of hours chosen by the user. These features have been tested to verify the effectiveness of the designs. Other MRI-related risks are further reduced by adherence to the conditions for scanning specified in this Technical Guide.
For additional information, see the Boston Scientific Website at http://www.bostonscientific.com/ imageready.
Only specific combinations of pulse generators and leads constitute an ImageReady Pacing System. Consult the following tables to distinguish between combinations that are valid for use with only 1.5 T scanners and combinations that are valid for use with both 1.5 T and 3 T
scanners.

Valid Combinations of Pulse Generators and Leads to Use in 1.5 Tesla and 3 Tesla Environments

Table 1–1. Valid Combinations of Pulse Generators and Leads to Use in 1.5 T and 3 T Environments
Combination of one
INGEVITY MRI Lead and one
FINELINE II Lead
1.5 T scanner only. 3 T scanner not allowed.
Normal Operating Mode only.
ESSENTIO MRI Pulse Generator PROPONENT MRI Pulse Generator ACCOLADE MRI Pulse Generator
INGEVITY MRI Leads
only
1.5 T or 3 T scanner allowed.
Normal Operating Mode or First Level Controlled Operating Mode.
FINELINE II Leads only
1.5 T scanner only. 3 T scanner not allowed.
Normal Operating Mode only.
Refer to Table 1–2 System Configuration for 1.5 T on page 1-3 and Table 1–3 System Configuration for 3 T on page 1-3 for a complete list of the model numbers of MR Conditional Pacing System components.
Refer to "MRI Conditions of Use" on page 1-4 for the entire set of MRI Conditions of Use.
WARNING: Only the combination of INGEVITY MRI lead(s) with an ESSENTIO MRI, PROPONENT MRI, or ACCOLADE MRI pulse generator is valid to use with either 1.5 T or 3 T scanners. All other allowable combinations of Boston Scientific MR Conditional system components must use only 1.5 T scanners.
NOTE: 3 T MRI scanners should be operated in Circularly Polarized mode or Dual Channel Multi-Transmission mode since ImageReady Pacing Systems have been evaluated against 3 T MRI scanners during Circularly Polarized and Dual Channel Multi-Transmission RF operation modes. Multi-Transmission RF operation with greater than two transmit channels has not been evaluated.

System Configuration for 1.5 T

Table 1–2. System Configuration for 1.5 T
Component Model Number(s) MR Status
Pulse Generators
ESSENTIO MRI Pulse Generator L110, L111, L131 MR Conditional
PROPONENT MRI Pulse Generator L210, L211, L231 MR Conditional
ACCOLADE MRI Pulse Generator L310, L311, L331 MR Conditional
Leads and Accessories
FINELINE II Sterox Pacing Lead 4456, 4457, 4458, 4459, 4479, 4480 MR Conditional
Introduction to MR Conditional Pacing
System Configuration for 1.5 T
FINELINE II Sterox / Sterox EZ Leads
1-3
FINELINE II Sterox EZ Pacing Lead 4469, 4470, 4471, 4472, 4473, 4474 MR Conditional
Suture Sleeve for FINELINE II leads 6220, 6221 MR Conditional
IS-1 Lead Port Plug 7145 MR Conditional
INGEVITY MRI Pacing Lead (Tined Fixation) 7731, 7732, 7735, 7736 MR Conditional
INGEVITY MRI Pacing Lead (Extendable/ Retractable Fixation)
Suture Sleeve for INGEVITY MRI leads 6402 MR Conditional
IS-1 Lead Port Plug 7145 MR Conditional
ZOOM LATITUDE Programmer/Recorder/Monitor (PRM) and PRM Software Application
ZOOM LATITUDE PRM 3120 MR Unsafe
ZOOM LATITUDE PRM Software App. 2869 N/A
a. See PRM is MR Unsafe Warning regarding the PRM.

System Configuration for 3 T

Table 1–3. System Configuration for 3 T
Component Model Number(s) MR Status
Pulse Generators
ESSENTIO MRI Pulse Generator L110, L111, L131 MR Conditional
PROPONENT MRI Pulse Generator L210, L211, L231 MR Conditional
ACCOLADE MRI Pulse Generator L310, L311, L331 MR Conditional
Leads and Accessories
INGEVITY MRI Pacing Lead (Tined Fixation) 7731, 7732, 7735, 7736 MR Conditional
INGEVITY MRI Leads
7740, 7741, 7742 MR Conditional
a
INGEVITY MRI Leads
INGEVITY MRI Pacing Lead (Extendable/ Retractable Fixation)
Suture Sleeve for INGEVITY MRI leads 6402 MR Conditional
7740, 7741, 7742 MR Conditional
1-4
Introduction to MR Conditional Pacing

MRI Conditions of Use

Table 1–3. System Configuration for 3 T (continued)
Component Model Number(s) MR Status
IS-1 Lead Port Plug 7145 MR Conditional
ZOOM LATITUDE Programmer/Recorder/Monitor (PRM) and PRM Software Application
ZOOM LATITUDE PRM 3120 MR Unsafe
ZOOM LATITUDE PRM Software App. 2869 N/A
a. See PRM is MR Unsafe Warning regarding the PRM.
MRI CONDITIONS OF USE
The following Conditions of Use must be met in order for a patient with an ImageReady Pacing System to undergo an MRI scan. Adherence to the Conditions of Use must be verified prior to each scan to ensure that the most up-to-date information has been used to assess the patient’s eligibility and readiness for an MR Conditional scan.

Cardiology

1. Patient is implanted with an ImageReady MR Conditional Pacing System1(see "System
Description" on page 1-2)
a
WARNING: Only the combination of INGEVITY MRI lead(s) with an ESSENTIO MRI, PROPONENT MRI, or ACCOLADE MRI pulse generator is valid to use with either 1.5 T or 3 T scanners. All other allowable combinations of Boston Scientific MR Conditional system components must use only 1.5 T scanners.
2. Pulse generator in MRI Protection Mode during scan
3. Bipolar pacing operation or pacing off
4. Patient does not have elevated body temperature or compromised thermoregulation at time
of scan
5. Pulse generator implant location restricted to left or right pectoral region
6. At least six (6) weeks have elapsed since implantation and/or any lead revision or surgical
modification of the MR Conditional Pacing System
7. No cardiac-related implanted devices, components, or accessories present other than an
ImageReady MR Conditional Pacing System (see "System Description" on page 1-2)
8. Pacing threshold ≤ 2.0 V in pace-dependent patients
9. No abandoned leads or pulse generators
10. No evidence of a fractured lead or compromised pulse generator-lead system integrity

Radiology

1. MRI magnet strength of 1.5 T or 3 T
a. MRI magnet strength of 1.5 T (See Table 1–2 System Configuration for 1.5 T on page 1-3
to determine which pulse generators and leads are valid for use with 1.5 T magnets.)
1. Defined as a Boston Scientific MR Conditional pulse generator and lead(s), with all ports occupied by a lead or port
plug.
Introduction to MR Conditional Pacing

Conditions for Scanning

1-5
• Radio frequency (RF) field of approximately 64 MHz
• Spatial gradient no greater than 50 T/m (5,000 G/cm)
b. MRI magnet strength of 3 T (See Table 1–3 System Configuration for 3 T on page 1-3 to
determine which pulse generators and leads are valid for use with 3 T magnets.)
• RF field of approximately 128 MHz
• Spatial gradient no greater than 50 T/m (5,000 G/cm)
WARNING: Only the combination of INGEVITY MRI lead(s) with an ESSENTIO MRI, PROPONENT MRI, or ACCOLADE MRI pulse generator is valid to use with either 1.5 T or 3 T scanners. All other allowable combinations of Boston Scientific MR Conditional system components must use only 1.5 T scanners.
2. Horizontal,
1
H proton, closed bore scanners only
3. Specific Absorption Rate (SAR) limits:
a. For an ImageReady Pacing System with FINELINE II leads only or with one FINELINE II
lead and one INGEVITY MRI lead (see "Valid Combinations of Pulse Generators and Leads to Use in 1.5 Tesla and 3 Tesla Environments" on page 1-2), SAR limits for Normal Operating Mode
2
must be observed for the entire active scan session as follows:
• Whole body averaged, ≤ 2.0 watts/kilogram (W/kg)
• Head, ≤ 3.2 W/kg
b. For an ImageReady Pacing System with INGEVITY MRI leads (see "Valid Combinations
of Pulse Generators and Leads to Use in 1.5 Tesla and 3 Tesla Environments" on page 1-2), SAR limits for Normal Operating Mode
4
Mode
must be observed for the entire active scan session as follows:
3
or for First Level Controlled Operating
• Whole body averaged, ≤ 4.0 W/kg
• Head, ≤ 3.2 W/kg
4. Gradient Field limits: Maximum specified gradient slew rate ≤ 200 T/m/s per axis
5. No local transmit-only coils or local transmit/receive coils placed directly over the pacing system; the use of receive-only coils is not restricted
6. Patient in supine or prone position only
7. The patient must be monitored during the MRI scan by pulse oximetry and/or electrocardiography (ECG)
Refer to Table 1–4 Cardiology Conditions/Patient Conditions on page 1-6 and Table 1–5 Radiology Conditions on page 1-8 for additional information about the Conditions of Use.
CONDITIONS FOR SCANNING
Table 1–4 Cardiology Conditions/Patient Conditions on page 1-6 summarizes the Cardiology Conditions/Patient-related Conditions of Use that must be met in order for an MR Conditional
2. As defined in IEC 60601-2-33, 201.3.224, 3rd Edition.
3. As defined in IEC 60601-2-33, 201.3.224, 3rd Edition.
4. As defined in IEC 60601-2-33, 201.3.208, 3rd Edition.
1-6
Introduction to MR Conditional Pacing
Conditions for Scanning
scan to be performed. For each condition or requirement, actions to determine eligibility, the potential clinical consequences of failing to meet the condition(s), and the patient population most impacted by failure to meet the condition(s) are listed.
Table 1–4. Cardiology Conditions/Patient Conditions
Condition for Scanning
(Rationale)
1. Patient is implanted with an ImageReady MR Conditional Pacing System (defined as a Boston Scientific MR Conditional pulse generator and lead(s), with all ports occupied by a lead or port plug).
For model numbers of MR Conditional components, and to identify an appropriate combination, refer to "System Description" on page 1-2 in this Guide, http://www. bostonscientific.com/ imageready, or Boston Scientific Technical Services.
Actions If Condition is Not Met
Check patient records.
Interrogate device. (Pulse generator model number is provided on PRM screen and MRI Protection Settings Report.)
Check patient ID card.
Check model numbers in "System Description" on page 1-2 of this Guide, at
http://www.bostonscientific. com/imageready, or by
contacting Boston Scientific Technical Services.
Confirm with physician responsible for managing the patient’s Pacing System.
Check X-ray or fluoroscopic images of pulse generator to assist identification (see "X-ray or Fluoroscopic Identification" on page 1-12).
Potential Clinical Consequences
Arrhythmia induction
Clinically significant pacing threshold changes and sensing changes as a result of heating at the lead/tissue interface
Damage to pulse generator and/or lead
Erratic pulse generator behavior
Inappropriate pacing, inhibition of pacing, or irregular intermittent capture or pacing, possibly resulting in pre-syncope or syncope
Physical movement of pulse generator and/or leads
Pocket discomfort due to pulse generator heating
Risk is Highest for
Pacing-dependent patients
Patients prone to sustained arrhythmias
Patients with high capture thresholds
The appropriate Boston Scientific MR Conditional pulse generator and Boston Scientific MR Conditional lead(s) must be used together to obtain the intended risk reduction needed for MR Conditional scans.
Another manufacturer's MR Conditional pulse generator combined with a Boston Scientific MR Conditional lead (or vice versa) do not constitute an MR Conditional System, because the components were not designed to work together in the MRI environment.
WARNING: Only the combination of INGEVITY MRI lead(s) with an ESSENTIO MRI, PROPONENT MRI, or ACCOLADE MRI pulse
generator is valid to use with either 1.5 T or 3 T scanners. All other allowable combinations of Boston Scientific MR Conditional system
components must use only 1.5 T scanners.
2. Pulse generator in MRI Protection Mode during scan.
Effects of RF or gradient fields create the potential for oversensing, and/or induced voltages in the pulse generator. MRI Protection Mode is designed to mitigate these effects.
3. Bipolar pacing operation or pacing off.
Unipolar lead configurations increase the risk of induced voltages in the lead system. Bipolar ventricular pacing operation is required to support Safety Core operation, if Safety Core is entered from MRI Protection Mode.
Program the pulse generator to MRI Protection Mode using the PRM.
Confirm that pacing lead configuration is bipolar. If unipolar, program to bipolar.
Unipolar configuration of the lead used for pacing in MRI Protection Mode will prevent entry into MRI Protection Mode.
Arrhythmia induction
Inappropriate pacing, inhibition of pacing, or irregular intermittent capture or pacing, possibly resulting in pre-syncope or syncope
Arrhythmia induction
Inappropriate pacing, inhibition of pacing, or irregular intermittent capture or pacing, possibly resulting in pre-syncope or syncope
Pacing will not occur if Safety Core is entered.
Pacing-dependent patients
Patients prone to sustained arrhythmias
Pacing-dependent patients
Patients prone to sustained arrhythmias
4. Patient does not have an elevated body temperature or compromised thermoregulation at time of MRI scan.
Check patient’s temperature prior to scan.
Clinically significant pacing threshold changes and sensing changes as a
Patients with high capture thresholds
Table 1–4. Cardiology Conditions/Patient Conditions (continued)
Introduction to MR Conditional Pacing
Conditions for Scanning
1-7
Condition for Scanning
(Rationale)
Pre-existing elevated temperature is additive with any scan-induced heating.
5. Pulse generator implant location restricted to left or right pectoral region.
Lead trajectories associated with non-pectoral implant locations pose risks for heating, inappropriate stimulation, and arrhythmia induction.
6. At least six weeks have elapsed since implantation and/ or any lead revision or surgical modification of the MR Conditional Pacing System.
A six-week period of healing allows for formation of scar tissue and capsule maturation, which reduce the impact of heating, vibration, and movement potentially caused by the magnetic fields of the MRI scanner.
Actions If Condition is Not Met
Check patient records.
Check by physical exam or X-ray.
Check patient records and/ or patient ID card.
Check PRM data for the user-entered Implant Date, if available.
Upon user request to enter MRI Protection Mode, the PRM provides an attention message if the calculated time since exit from Storage Mode is less than or equal to six weeks. (Check
that the PRM is set with the correct time and date to ensure accuracy.)
Potential Clinical Consequences
result of heating at the lead/tissue interface
Pocket discomfort due to pulse generator heating
Arrhythmia induction
Clinically significant pacing threshold changes and sensing changes as a result of heating at the lead/tissue interface
Inappropriate pacing, inhibition of pacing, or irregular intermittent capture or pacing, possibly resulting in pre-syncope or syncope
Physical movement of pulse generator in pocket
Pocket discomfort due to pulse generator heating
Arrhythmia induction
Clinically significant pacing threshold changes and sensing changes as a result of heating at the lead/tissue interface
Increased rate of lead dislodgement due to incomplete capsule maturation
Physical movement of pulse generator in pocket
Risk is Highest for
Pacing-dependent patients
Patients prone to sustained arrhythmias
Patients with high capture thresholds
Patients prone to sustained arrhythmias
Patients with high capture thresholds
7. No cardiac-related implanted devices, components, or accessories present other than an ImageReady MR Conditional Pacing System (see "System Description" on page 1-2).
The presence of other cardiac implants or accessories such as
lead adaptors, extenders, or abandoned leads or pulse generators may significantly
reduce the effectiveness of an ImageReady MR Conditional Pacing System in reducing risks of MRI scanning.
8. Pacing threshold ≤ 2.0 V in pace-dependent patients.
Pulse generator pulse amplitude in MRI Protection Mode is set to 5.0 V, providing a minimum two-fold safety margin
Check patient records.
Confirm with physician responsible for managing the patient’s Pacing System.
Check X-rays.
Check model numbers in this Guide ("System Description" on page 1-2) or at http://www.
bostonscientific.com/ imageready.
For model numbers of MR Conditional components, and to identify an appropriate combination, refer to "System Description" on page 1-2 in this Guide, http://www. bostonscientific.com/ imageready, or Boston Scientific Technical Services.
Check patient records for most recent pacing threshold values or run a pacing threshold test.
The device will check the most recently recorded pacing
Arrhythmia induction
Clinically significant pacing threshold changes and sensing changes as a result of heating at the lead/tissue interface
Damage to pulse generator, lead, or connection
Physical movement of pulse generator and/or leads
Pocket discomfort due to pulse generator heating
Inappropriate pacing, inhibition of pacing, or irregular intermittent capture or pacing, possibly resulting in pre-syncope or syncope
Pacing-dependent patients
Patients prone to sustained arrhythmias
Patients with high capture thresholds
Pacing-dependent patients
1-8
Table 1–4. Cardiology Conditions/Patient Conditions (continued)
Introduction to MR Conditional Pacing
Conditions for Scanning
Condition for Scanning
(Rationale)
for patients with a pacing threshold ≤ 2.0 V plus an additional 1.0 V to counteract gradient-induced pace pulse offsets.
9. No abandoned leads or pulse generators.
The presence of abandoned leads or pulse generators may significantly reduce the effectiveness of the ImageReady MR Conditional Pacing System in reducing risks of MRI scanning.
10. No evidence of a fractured lead or compromised pulse generator-lead system integrity.
Lead impedance values within the programmed normal range. No record or evidence of damage to pulse generator seal plug and front lead sealing rings.
Abnormal lead impedance values may indicate a short or open circuit in the lead system. This could result in abnormal conductive trajectories and induced voltages. Broken conductors in the lead system could result in increased potential for heating at the lead tip. A damaged seal plug or front lead sealing ring could promote an alternate current flow path during MRI scanning.
Actions If Condition is Not Met
threshold testing results for each chamber when MRI Protection Mode is programmed and provides an attention message on the PRM screen if > 2.0 V.
Check patient records.
Confirm with physician responsible for managing the patient's Pacing System.
Check X-rays.
Check patient records for most recent lead impedance values.
Review Daily Measurements on the Leads Status Summary Screen to verify stability over time of pace impedance, pace threshold, and intrinsic amplitude values.
Check patient records from implant procedure.
Check patient records for a history of noise on EGMs.
The device measures lead impedances upon user request to enter MRI Protection Mode and provides an attention message on the PRM screen if the values are out of the programmed normal range.
A history of noise on EGMs could be indicative of a damaged seal plug or front lead sealing rings.
Potential Clinical Consequences
Arrhythmia induction
Clinically significant pacing threshold changes and sensing changes as a result of heating at the lead/tissue interface
Damage to pulse generator, lead, or connection
Pocket discomfort due to pulse generator heating
Arrhythmia induction
Clinically significant pacing threshold changes and sensing changes as a result of heating at the lead/tissue interface
Inappropriate pacing, inhibition of pacing, or irregular intermittent capture or pacing, possibly resulting in pre-syncope or syncope
Risk is Highest for
Pacing-dependent patients
Patients prone to sustained arrhythmias
Patients with high capture thresholds
Pacing-dependent patients
Patients prone to sustained arrhythmias
Patients with high capture thresholds
Table 1–5 Radiology Conditions on page 1-8 summarizes the Radiology-related Conditions of Use that must be met in order for an MR Conditional MRI scan to be performed. For each condition or requirement, actions to determine eligibility, the potential clinical consequences of failing to meet the condition(s), and the patient population most impacted by failure to meet the condition(s) are listed.
Table 1–5. Radiology Conditions
Condition for Scanning (Rationale) Actions If Condition is Not Met
1. MRI magnet strength of 1.5 T or 3 T (see a and b below).
1a. MRI magnet strength of 1.5 T:
Check technical specifications of MRI scanner.
Potential Clinical Consequences
Arrhythmia induction Pacing-dependent
Risk is Highest for
patients
Introduction to MR Conditional Pacing
Conditions for Scanning
Table 1–5. Radiology Conditions (continued)
Condition for Scanning (Rationale) Actions If Condition is Not Met
Potential Clinical
Risk is Highest for
Consequences
1-9
RF field of approximately 64 MHz
Spatial gradient no greater than 50 T/ m (5,000 G/cm)
Refer to "System Description" on page 1­2 to determine which components are valid for use with 1.5 T magnets.
Clinically significant pacing threshold changes and sensing changes as a result of heating at the lead/ tissue interface
Patients prone to sustained arrhythmias
Patients with high capture thresholds
Damage to pulse generator, lead, or connection
Physical movement of pulse generator and/or leads
Pocket discomfort due to pulse generator heating
The ImageReady Pacing Systems mitigate hazards associated with 1.5 T or 3 T MRI scanning. System response to other magnet strengths has not been evaluated. Refer to "System Description" on page 1-2 to determine which components are valid for use with 1.5 T scanners.
System response to spatial gradients greater than 50 T/m (5,000 G/cm) has not been evaluated.
WARNING: Only the combination of INGEVITY MRI lead(s) with an ESSENTIO MRI, PROPONENT MRI, or ACCOLADE MRI pulse
generator is valid to use with either 1.5 T or 3 T scanners. All other allowable combinations of Boston Scientific MR Conditional system
components must use only 1.5 T scanners.
1b. MRI magnet strength of 3 T:
RF field of approximately 128 MHz
Spatial gradient no greater than 50 T/ m (5,000 G/cm)
Check technical specifications of MRI scanner.
Refer to "System Description" on page 1­2 to determine which components are valid for use with 3 T magnets.
Arrhythmia induction
Clinically significant pacing threshold changes and sensing changes as a result of heating at the lead/ tissue interface
Damage to pulse generator, lead, or
Pacing-dependent patients
Patients prone to sustained arrhythmias
Patients with high capture thresholds
connection
Physical movement of pulse generator and/or leads
Pocket discomfort due to pulse generator heating
The ImageReady Pacing Systems mitigate hazards associated with 1.5 T or 3 T MRI scanning. System response to other magnet strengths has not been evaluated. Refer to "System Description" on page 1-2 to determine which components are valid for use with 3 T scanners.
System response to spatial gradients greater than 50 T/m (5,000 G/cm) has not been evaluated.
WARNING: Only the combination of INGEVITY MRI lead(s) with an ESSENTIO MRI, PROPONENT MRI, or ACCOLADE MRI pulse
generator is valid to use with either 1.5 T or 3 T scanners. All other allowable combinations of Boston Scientific MR Conditional system
components must use only 1.5 T scanners.
2. Horizontal,
1
H proton, closed bore
scanners only.
The ImageReady Pacing Systems mitigate hazards associated with horizontal closed bore scanners.
Check technical specifications of MRI scanner.
Arrhythmia induction
Clinically significant pacing threshold changes and sensing changes as a result of heating at the lead/ tissue interface
Pacing-dependent patients
Patients prone to sustained arrhythmias
Patients with high capture thresholds
Inappropriate pacing, inhibition of pacing, or irregular intermittent capture or pacing,
1-10
Introduction to MR Conditional Pacing
Conditions for Scanning
Table 1–5. Radiology Conditions (continued)
Condition for Scanning (Rationale) Actions If Condition is Not Met
Potential Clinical Consequences
possibly resulting in pre-syncope or syncope
Damage to pulse generator, lead, or connection
Physical movement of pulse generator and/or leads
Pocket discomfort due to pulse generator heating
3. Specific Absorption Rate (SAR) limits (see a and b below).
Risk is Highest for
3a. SAR limits for Normal Operating Mode must be observed for the entire active scan session with an ImageReady Pacing System with FINELINE II leads only or with a combination of one INGEVITY lead and one FINELINE II lead.
Whole body averaged, ≤ 2.0 W/kg
Head, ≤ 3.2 W/kg
Ensure MRI scanner is operated in Normal Operating Mode (NOT in First Level Controlled Operating Mode).
Clinically significant pacing threshold changes and sensing changes as a result of heating at the lead/ tissue interface
Inappropriate pacing, inhibition of pacing, or irregular intermittent pacing, possibly
Pacing-dependent patients
Patients with high capture thresholds
resulting in pre-syncope or syncope
An ImageReady Pacing System with FINELINE II leads only or with a combination of one INGEVITY lead and one FINELINE II lead mitigates hazards associated with Normal Operating Mode. System response to other scanner settings has not been evaluated.
3b. SAR limits for Normal Operating Mode or First Level Controlled Operating Mode must be observed for the entire active scan session with an ImageReady Pacing System with INGEVITY MRI leads only.
Whole body averaged, ≤ 4.0 W/kg
Head, ≤ 3.2 W/kg
Ensure MRI scanner is operated in Normal Operating Mode or First Level Controlled Operating Mode.
Clinically significant pacing threshold changes and sensing changes as a result of heating at the lead/ tissue interface
Inappropriate pacing, inhibition of pacing, or irregular intermittent
Pacing-dependent patients
Patients with high capture thresholds
pacing, possibly resulting in pre-syncope or syncope
An ImageReady Pacing System with INGEVITY MRI leads mitigates hazards associated with Normal Operating Mode or First Level Controlled Operating Mode. System response to other scanner settings has not been evaluated.
4. Gradient Field limits: Maximum specified gradient slew rate ≤ 200 T/m/s per axis.
System response to gradient slew rates higher than 200 T/m/s per axis has not been evaluated.
5. No local transmit-only coils or local transmit/receive coils placed directly over the pacing system; the use of receive-only coils is not restricted.
Check technical specifications of MRI scanner.
Ensure no local transmit-only or transmit/receive coils are placed directly over the pacing system.
Arrhythmia induction
Inappropriate pacing, inhibition of pacing, or irregular intermittent capture or pacing, possibly resulting in pre-syncope or syncope
Damage to pulse generator, lead, or connection
Pocket discomfort due to pulse generator heating
Arrhythmia induction
Clinically significant pacing threshold changes and sensing changes as a result of
Pacing-dependent patients
Patients prone to sustained arrhythmias
Pacing-dependent patients
Patients prone to sustained arrhythmias
Introduction to MR Conditional Pacing
Table 1–5. Radiology Conditions (continued)
Condition for Scanning (Rationale) Actions If Condition is Not Met
Potential Clinical Consequences

MRI Protection Mode

Risk is Highest for
1-11
There are no restrictions for positioning the pacing system within the integrated body coil of the MRI scanner. System response to local transmit-only or transmit/receive coils placed directly over the pacing system has not been evaluated.
6. The patient must be in a supine or prone position during the scan.
An ImageReady Pacing System mitigates hazards associated with a patient position of supine or prone. System response to other patient positions has not been evaluated.
Ensure patient is in the correct position during scan.
heating at the lead/ tissue interface
Erratic pulse generator behavior
Inappropriate pacing, inhibition of pacing, or irregular intermittent capture or pacing, possibly resulting in pre-syncope or syncope
Pocket discomfort due to pulse generator heating
Arrhythmia induction
Clinically significant pacing threshold changes and sensing changes as a result of heating at the lead/ tissue interface
Inappropriate pacing, inhibition of pacing, or irregular intermittent capture or pacing, possibly resulting in pre-syncope or syncope
Physical movement of pulse generator and/or leads
Pocket discomfort due to pulse generator heating
Patients with high capture thresholds
Pacing-dependent patients
Patients prone to sustained arrhythmias
Patients with high capture thresholds
7. The patient must be monitored during the MRI scan by pulse oximetry and/or electrocardiography (ECG).
MRI PROTECTION MODE
In preparation for an MRI scan, the pulse generator must be programmed into MRI Protection Mode using the programmer. MRI Protection Mode modifies certain pulse generator functions in order to mitigate risks associated with exposing the ImageReady MR Conditional System to the MRI environment. For a list of features and functions that are suspended in MRI Protection Mode, see "MRI Protection Mode General Information" on page 2-2.

MRI BASIC CONCEPTS

MRI is a diagnostic tool that uses three types of magnetic and electromagnetic fields to image soft tissue in the body:
A static magnetic field generated by a superconducting electromagnet coil, 1.5 T or 3 T in strength.
Gradient magnetic fields of much lower intensity, but with high rates of change over time. Three sets of gradient coils are used to create the gradient fields.
Ensure patient is being monitored during scan.
Lack of patient monitoring could result in failure to detect potentially dangerous changes in the patient’s cardiac or hemodynamic function.
All patients
1-12
Introduction to MR Conditional Pacing

X-ray or Fluoroscopic Identification

A pulsed radio frequency (RF) field produced by transmission RF coils (approximately 64 MHz for 1.5 T and 128 MHz for 3 T).
These fields may create physical forces or electrical currents that can affect the functioning of active implantable medical devices (AIMDs) such as pulse generators and leads. Therefore, only patients implanted with an MR Conditional system are eligible to be scanned. Furthermore, by complying with the MRI Conditions of Use, outlined in this Technical Guide ("MRI Conditions of Use" on page 1-4), ImageReady MR Conditional System patients can undergo MRI scans with risks mitigated to the best current standard of care.
X-RAY OR FLUOROSCOPIC IDENTIFICATION
The primary means of identification of the implanted device and lead(s) should be the patient's medical files. To provide additional means of identification of the device, images of the pulse generators as visible under X-ray or fluoroscopy are shown in Table 1–6 Pulse Generator Images on page 1-12.
Table 1–6. Pulse Generator Images
Pulse Generator Images
Pulse Generators X-Ray Identifier Approximate Location on Device (1)
ESSENTIO MRI PROPONENT MRI ACCOLADE MRI

MR CONDITIONAL PACING SYSTEM WARNINGS AND PRECAUTIONS

General

WARNING: Unless all of the MRI Conditions of Use ("MRI Conditions of Use" on page 1-4) are
met, MRI scanning of the patient does not meet MR Conditional requirements for the implanted system, and significant harm to or death of the patient and/or damage to the implanted system may result.
For potential adverse events applicable when the Conditions of Use are met or not met, see "Potential Adverse Events" on page 1-14.
NOTE: Table 1–4 Cardiology Conditions/Patient Conditions on page 1-6 and Table 1–5 Radiology Conditions on page 1-8 provide information on the nature of the increased risk(s) associated with the failure to meet each Condition of Use. This information is intended to assist in performing a risk/benefit analysis to decide whether or not to scan a patient who does not meet all the stated criteria for MR Conditional status. Alternatives including other imaging methods may also be considered.
WARNING: Only the combination of INGEVITY MRI lead(s) with an ESSENTIO MRI, PROPONENT MRI, or ACCOLADE MRI pulse generator is valid to use with either 1.5 T or 3 T scanners. All other allowable combinations of Boston Scientific MR Conditional system components must use only 1.5 T scanners.
WARNING: Ensure that an external defibrillator and medical personnel skilled in cardio­pulmonary resuscitation (CPR) are present during the MRI scan should the patient require external rescue.
WARNING: MRI scanning after Explant status has been reached may lead to premature battery depletion, a shortened device replacement window, or sudden loss of pacing. After performing an MRI scan on a device that has reached Explant status, verify pulse generator function and schedule device replacement.
WARNING: When the Time-out parameter is programmed to a value other than Off, the patient must be out of the scanner before the time programmed elapses. Otherwise, the patient will no longer meet Conditions of Use ("MRI Conditions of Use" on page 1-4).

Programming Considerations

WARNING: If the MRI Protection Time-out value of Off is combined with a Pacing Mode of Off,
the patient will not receive pacing until the pulse generator is programmed out of MRI Protection Mode and back to normal operation. It is recommended to have the Programmer/Recorder/ Monitor (PRM) near the MRI room in case the patient develops the urgent need for pacing. Patients with the following conditions may have increased risk of developing transient pacing­dependence:
Introduction to MR Conditional Pacing
Programming Considerations
1-13
A history of syncope related to bradyarrhythmia
A history of syncope of unknown etiology
Sinus pauses (Pause > 2 s), AV block permanently or intermittently
At risk for intermittent AV block (for example, those with progressive AV block, or a history of unexplained syncope)
At risk for trifascicular block (alternating bundle branch block or PR interval > 200 ms with LBBB or other bifascicular block)
WARNING: Use caution when programming the MRI Protection Mode pacing amplitude for pacing-dependent patients who have high pacing thresholds (> 2.0 V). Programming pacing amplitude below 5.0 V is provided as an option in case of extracardiac stimulation (for example, diaphragmatic stimulation for RV pacing). If pacing amplitude is programmed below 5.0 V, an appropriate safety margin (2X the pacing threshold + 1.0 V) should be maintained. An inadequate safety margin may result in loss of capture.
WARNING: Exit MRI Protection Mode after MRI scanning is completed. If the MRI Protection Time-out value of Off is selected, the pulse generator will remain permanently in the MRI Protection Mode until it is programmed otherwise. Prolonged use of the MRI Protection Mode (such as may occur when the Time-out feature is programmed to Off) may increase the rate of battery depletion. In addition, prolonged exposure of a patient to the XOO mode chosen may be deleterious to the patient’s health.

Safety Mode

WARNING: Do not perform an MRI scan on a patient whose device has entered Safety Mode.
Safety Mode pacing is VVI unipolar, which, in the MRI environment, subjects the patient to increased risk of arrhythmia induction, inappropriate pacing, inhibition of pacing, or irregular intermittent capture or pacing.
WARNING: If the pulse generator enters Safety Mode from MRI Protection Mode, backup pacing will not occur in the following scenarios:
if a functional bipolar ventricular pacing lead is not present
1-14
Introduction to MR Conditional Pacing

MRI Site Zone III Exclusions

if the Pacing Mode under MRI Protection Mode settings is programmed to Off; the pulse generator will continue permanently with the Pacing Mode programmed to Off, and the patient will not receive pacing therapy until the pulse generator is replaced
MRI Site Zone III Exclusions
WARNING: The Programmer/Recorder/Monitor (PRM) is MR Unsafe and must remain outside
the MRI site Zone III (and higher) as defined by the American College of Radiology Guidance Document for Safe MR Practices MRI scanner room, the control room, or the MRI site Zone III or IV areas.
5
. Under no circumstances should the PRM be brought into the

Precautions

WARNING: Implant of the system cannot be performed in an MRI site Zone III (and higher) as
defined by the American College of Radiology Guidance Document for Safe MR Practices
6
. Some of the accessories packaged with pulse generators and leads, including the torque wrench and stylet wires, are not MR Conditional and should not be brought into the MRI scanner room, the control room, or the MRI site Zone III or IV areas.
CAUTION: The physician choosing MRI Protection Mode parameter values will need to exercise professional judgment to determine an individual patient’s ability to tolerate the pacing parameters required for MR Conditional scanning, in conjunction with the physical conditions required during a scan (for example, prolonged time in a supine position).
CAUTION: If the MR Conditional Pacing System enters Safety Core Operation during MRI Protection Mode and the pacing mode was set to a value other than Off, MRI Protection Mode pacing will be automatically switched to VOO mode, RV bipolar configuration (sensing and pacing), 5.0 V pace pulse amplitude, 1.0 ms pulse width, and 72.5 min
-1
pacing rate as the safety
mode.
CAUTION: The presence of the implanted Pacing System may cause MRI image artifacts (see "3. Preparing the Patient for the Scan" on page 2-13).
NOTE: All normal risks associated with an MRI procedure apply to MRI scans with the MR Conditional Pacing System. Consult MRI scanner documentation for a complete list of risks associated with MRI scanning.
NOTE: Other implanted devices or patient conditions may cause a patient to be ineligible for an MRI scan, independent of the status of the patient’s ImageReady MR Conditional Pacing System.

POTENTIAL ADVERSE EVENTS

Potential adverse events differ depending on whether the MRI Conditions of Use ("MRI Conditions of Use" on page 1-4) are met. For a complete list of potential adverse events, refer to the Physician's Technical Manual for the pulse generator.
MRI scanning of patients when the Conditions of Use are met could result in the following potential adverse events:
Arrhythmia induction
Bradycardia
Patient death
5. Kanal E, et al., American Journal of Roentgenology 188:1447-74, 2007.
6. Kanal E, et al., American Journal of Roentgenology 188:1447-74, 2007.
Introduction to MR Conditional Pacing
Potential Adverse Events
1-15
Patient discomfort due to slight movement or heating of the device
Side effects of MRI Protection Mode pacing at elevated fixed rate and increased output including reduced exercise capacity, acceleration of heart failure, and competitive pacing/ arrhythmia induction
Syncope
MRI scanning of patients when the Conditions of Use are NOT met could result in the following potential adverse events:
Arrhythmia induction
Bradycardia
Damage to the pulse generator and/or leads
Erratic pulse generator behavior
Inappropriate pacing, inhibition of pacing, failure to pace
Increased rate of lead dislodgement (within six weeks of implant or revision of system)
Irregular or intermittent capture or pacing
Pacing threshold changes
Patient death
Patient discomfort due to movement or heating of the device
Physical movement of pulse generator and/or leads
Sensing changes
Syncope
1-16
Introduction to MR Conditional Pacing
Potential Adverse Events
MRI SCAN PROCEDURE PROTOCOL
CHAPTER 2
This chapter contains the following topics:
“Patient Flow” on page 2-2
“MRI Protection Mode General Information” on page 2-2
“Pre-Scan Activities” on page 2-4
“During the Scan” on page 2-14
“After the Scan” on page 2-14
2-1
2-2
MRI Scan Procedure Protocol

Patient Flow

Before proceeding with this MRI scan procedure protocol, verify that the patient and the MRI scanner meet the MRI Conditions of Use ("MRI Conditions of Use" on page 1-4). This verification must be performed prior to each scan to ensure that the most up-to-date information has been used to assess the patient’s eligibility and readiness for an MR Conditional scan.
WARNING: Unless all of the MRI Conditions of Use ("MRI Conditions of Use" on page 1-4) are met, MRI scanning of the patient does not meet MR Conditional requirements for the implanted system, and significant harm to or death of the patient and/or damage to the implanted system may result.
For potential adverse events applicable when the Conditions of Use are met or not met, see "Potential Adverse Events" on page 1-14.
NOTE: Table 1–4 Cardiology Conditions/Patient Conditions on page 1-6 and Table 1–5 Radiology Conditions on page 1-8 provide information on the nature of the increased risk(s) associated with the failure to meet each Condition of Use. This information is intended to assist in performing a risk/benefit analysis to decide whether or not to scan a patient who does not meet all the stated criteria for MR Conditional status. Alternatives including other imaging methods may also be considered.
PATIENT FLOW
A sample patient flow sequence for an ImageReady Pacing System patient who needs an MRI scan is described below. For a more detailed description of the programming and scanning procedure, see this chapter.
1. MRI recommended to patient by specialist (for example, orthopedist or oncologist).
2. Patient or specialist or radiologist contacts the electrophysiologist/cardiologist who manages the patient’s MR Conditional Pacing System.
3. Electrophysiologist/cardiologist determines patient eligibility for scan per the information in this Technical Guide.
4. If the patient is eligible, the PRM is used to put the pulse generator in MRI Protection Mode as close in time to the scan as reasonable. The MRI Protection Settings Report is printed, placed in the patient's file, and provided to radiology personnel. The report documents MRI Protection Mode settings and details. If the Time-out feature is used, the report includes the exact time and date when MRI Protection Mode will expire.
5. The model number of each lead implanted in the patient is identified, and this information is communicated to the HCPs involved in performing the MRI scan.
6. The radiologist checks the patient file and/or printed report. If the Time-out feature is used, the radiologist verifies that adequate time remains to complete the scan.
7. Patient undergoes scan according to the conditions of use described in this Technical Guide.
8. After the scan, manually exit MRI Protection Mode using the PRM to return the pulse generator to pre-MRI operation. Perform follow-up testing of the implanted system.

MRI PROTECTION MODE GENERAL INFORMATION

Prior to the patient undergoing an MRI scan, an ImageReady MR Conditional Pacing System must be programmed to the MRI Protection Mode using the PRM (see Table 2–1 MRI Protection Parameters on page 2-4). In MRI Protection Mode:
MRI Protection Mode General Information
MRI Scan Procedure Protocol
2-3
Pacing mode options include asynchronous pacing or no pacing (DOO, AOO, VOO, or Off). Off should only be used if the patient is judged to be clinically capable of receiving no pacing during the time the pulse generator will be in MRI Protection Mode, including during the scan.
The programmed pacing mode prior to entry into MRI Protection Mode determines the default MRI Protection pacing mode. For example, if MRI Protection Mode is entered from DDD(R), the pacing mode will be DOO. Any of the other pacing mode options may then be selected.
WARNING: The Programmer/Recorder/Monitor (PRM) is MR Unsafe and must remain outside the MRI site Zone III (and higher) as defined by the American College of Radiology Guidance Document for Safe MR Practices
1
. Under no circumstances should the PRM be
brought into the MRI scanner room, the control room, or the MRI site Zone III or IV areas.
WARNING: If the pulse generator enters Safety Mode from MRI Protection Mode, backup pacing will not occur in the following scenarios:
if a functional bipolar ventricular pacing lead is not present
if the Pacing Mode under MRI Protection Mode settings is programmed to Off; the pulse
generator will continue permanently with the Pacing Mode programmed to Off, and the patient will not receive pacing therapy until the pulse generator is replaced
The Lower Rate Limit is nominally set to 20 min
-1
above the starting LRL, and is
programmable in normal increments. For both the nominal setting based on the LRL and the
-1
programmable setting, the maximum value is 100 min
.
Atrial pulse amplitude and ventricular pulse amplitude are nominally set to 5.0 V and are programmable in normal increments between 2.0 V and 5.0 V.
WARNING: Use caution when programming the MRI Protection Mode pacing amplitude for pacing-dependent patients who have high pacing thresholds (> 2.0 V). Programming pacing amplitude below 5.0 V is provided as an option in case of extracardiac stimulation (for example, diaphragmatic stimulation for RV pacing). If pacing amplitude is programmed below
5.0 V, an appropriate safety margin (2X the pacing threshold + 1.0 V) should be maintained. An inadequate safety margin may result in loss of capture.
AV Delay is fixed at 100 ms
Pulse width is fixed at 1.0 ms for both chambers
A Time-out feature is nominally set to 24 hours, with programmable values of Off, 12, 24, and 48 hours
WARNING: Exit MRI Protection Mode after MRI scanning is completed. If the MRI Protection Time-out value of Off is selected, the pulse generator will remain permanently in the MRI Protection Mode until it is programmed otherwise. Prolonged use of the MRI Protection Mode (such as may occur when the Time-out feature is programmed to Off) may increase the rate of battery depletion. In addition, prolonged exposure of a patient to the XOO mode chosen may be deleterious to the patient’s health.
NOTE: Twenty-four hours in MRI Protection Mode (with pacing on) reduces pulse generator longevity by approximately 5 days.
CAUTION: The physician choosing MRI Protection Mode parameter values will need to exercise professional judgment to determine an individual patient’s ability to tolerate the pacing parameters required for MR Conditional scanning, in conjunction with the physical conditions required during a scan (for example, prolonged time in a supine position).
1. Kanal E, et al., American Journal of Roentgenology 188:1447-74, 2007.
2-4
MRI Scan Procedure Protocol

Pre-Scan Activities

The following features and functions are suspended in MRI Protection Mode:
• PaceSafe RV automatic capture
• PaceSafe RA automatic threshold
• Cardiac sensing
• Daily diagnostics (lead impedance, intrinsic amplitude, pace threshold)
• Motion and respiratory sensors
• Magnet detection
• RF telemetry
• Battery voltage monitoring
The following device conditions will preclude the user from having the option to enter MRI Protection Mode (see the Reference Guide for the pulse generator for additional information about these conditions):
• Battery capacity status is Depleted
• Pulse generator is in Storage Mode
• Pulse generator is in Electrocautery Mode
• Pulse generator is in Safety Core operation (Safety Mode)
• Diagnostic test is in progress
• EP test is in progress
WARNING: MRI scanning after Explant status has been reached may lead to premature battery depletion, a shortened device replacement window, or sudden loss of pacing. After performing an MRI scan on a device that has reached Explant status, verify pulse generator function and schedule device replacement.
WARNING: Do not perform an MRI scan on a patient whose device has entered Safety Mode. Safety Mode pacing is VVI unipolar, which, in the MRI environment, subjects the patient to increased risk of arrhythmia induction, inappropriate pacing, inhibition of pacing, or irregular intermittent capture or pacing.
Table 2–1. MRI Protection Parameters
Parameter Programmable Values Nominal
MRI Brady Mode Off; VOO; AOO; DOO DOO for DDD(R), DDI(R), or DOO normal
MRI Lower Rate Limit (LRL) ( min
MRI Atrial Amplitude (V) 2.0; 2.1; …; 3.5; 4.0; …; 5.0 5.0 (Tolerance ± 15% or ± 100 mV,
MRI Ventricular Amplitude (V) 2.0; 2.1; …; 3.5; 4.0; …; 5.0 5.0 (Tolerance ± 15% or ± 100 mV,
-1
) 30; 35; ...; 100 20 min
Brady modes; VOO for VDD(R), VVI(R), or VOO normal Brady modes; AOO for AAI(R) or AOO normal Brady Mode; Off for Normal Brady Mode Off
-1
above the normal mode LRL
whichever is greater)
whichever is greater)
a
a
MRI Protection Time-out (hours) Off; 12; 24; 48 24
a. During the transition into the MRI Protection Mode, it may take up to 6 cardiac pacing cycles for the pace amplitude to meet the specified tolerance range.
PRE-SCAN ACTIVITIES
Three activities are required before the MRI scan takes place:
1. Prepare the pulse generator for the scan by programming into MRI Protection Mode ("1. Programming the Pulse Generator for a Scan" on page 2-5)
2. Confirm the MRI scanner settings and configurations ("2. Confirming MRI Scanner Settings and Configuration" on page 2-13)
3. Prepare the patient for the scan ("3. Preparing the Patient for the Scan" on page 2-13)

1. Programming the Pulse Generator for a Scan

Use the PRM to program pulse generator entry into MRI Protection Mode.
NOTE: Maintain access to the programmer wand, as RF telemetry becomes unavailable during the process of entering MRI Protection Mode.
From the Main screen, use the Device Mode button to enable MRI Protection Mode.
The user chooses whether to Cancel Changes or Apply Changes to proceed with entry into MRI Protection Mode (Figure 2–1 Change Device Mode dialog on page 2-5).
1. Programming the Pulse Generator for a Scan
MRI Scan Procedure Protocol
2-5
Figure 2–1. Change Device Mode dialog
Certain conditions in the pulse generator and/or system will cause a user request to enter MRI Protection Mode to be rejected. These include:
A ventricular episode as detected and recognized by the pulse generator is in progress
Magnet presence is detected by magnet sensor
Pulse generator is in STAT PACE mode
Unipolar pacing configuration in chamber(s) where pacing will occur in MRI Protection Mode
If one or more of these conditions are present, a dialog box will appear describing the condition, and MRI Protection Mode cannot be entered. For example, see Figure 2–2 Episode in progress attention message on page 2-6.
2-6
MRI Scan Procedure Protocol
1. Programming the Pulse Generator for a Scan
Figure 2–2. Episode in progress attention message
In addition to the above-listed conditions that prevent entry into MRI Protection Mode, two other conditions of use are assessed by the PRM upon a request to enter MRI Protection Mode: lead impedance and time since implant.
Lead Impedance
A user request to enter the MRI Protection Mode triggers a lead impedance test in all chambers. If the lead impedance values obtained from this testing are outside the programmed normal range, the PRM provides a dialog box recommending a review of the associated risks if the user chooses to proceed (see Table 1–4 Cardiology Conditions/Patient Conditions on page 1-6). The dialog provides the option of activating MRI Protection Mode in the presence of these conditions, or cancelling entry into MRI Protection Mode. The dialog box that appears in the case of an out-of-range lead impedance value is shown in Figure 2–3 Lead impedance out of range attention message on page 2-6.
Figure 2–3. Lead impedance out of range attention message
Time Since Implant
The PRM also determines the time since implant, calculated based on the date at which the pulse generator was taken out of Storage Mode.
NOTE: If the PRM clock is not set to the correct time and date, this determination may not be accurate.
If the calculated time since exit from Storage Mode is less than 6 weeks, the PRM provides a dialog box recommending a review of the associated risks if the user chooses to proceed (see Table 1–4 Cardiology Conditions/Patient Conditions on page 1-6). The dialog provides the option of activating MRI Protection Mode in the presence of these conditions, or cancelling entry into MRI Protection Mode.
Upon continuing with entry into MRI Protection Mode, the MRI Protection Checklist screen is displayed (Figure 2–4 MRI Protection Checklist on page 2-7). The Checklist summarizes the conditions that must be met at the time of scanning in order for a patient to be eligible for an MR
1. Programming the Pulse Generator for a Scan
MRI Scan Procedure Protocol
2-7
Conditional scan. Re-verification is required before every scan to guard against the possibility that changes in the system or patient occurred subsequent to the original pulse generator/system implant. These conditions are described in greater detail in Table 1–4 Cardiology Conditions/ Patient Conditions on page 1-6 and Table 1–5 Radiology Conditions on page 1-8.
Figure 2–4. MRI Protection Checklist
If the Conditions of Use as described in this manual are not met, the Cancel button is selected to return to normal system operation, and the patient does not undergo an MRI scan.
If the Conditions of Use are met, or if the Conditions of Use are not met, but the user elects to continue with MRI Protection Mode after reviewing the risks of proceeding (see Table 1–4 Cardiology Conditions/Patient Conditions on page 1-6 and Table 1–5 Radiology Conditions on page 1-8 for additional information about risks), the Continue with MRI Protection button is selected. As a result, the Program MRI Protection screen appears (Figure 2–5 Program MRI Protection dialog on page 2-9).
Use the dialog boxes to set the:
Pacing mode (DOO, VOO, AOO, Off)
Lower rate limit (nominally set to 20 min
increments to a maximum value 100 min
-1
above normal mode LRL, programmable in normal
-1
)
NOTE: Because MRI Protection Mode pacing is asynchronous, when setting the lower rate limit, consider the patient’s intrinsic rate to avoid competitive pacing.
Atrial amplitude (nominally set to 5.0 V, programmable in normal increments from 2.0 V to 5.0
V)
Ventricular amplitude (nominally set to 5.0 V, programmable in normal increments from 2.0 V
to 5.0 V)
2-8
MRI Scan Procedure Protocol
1. Programming the Pulse Generator for a Scan
WARNING: Use caution when programming the MRI Protection Mode pacing amplitude for
pacing-dependent patients who have high pacing thresholds (> 2.0 V). Programming pacing amplitude below 5.0 V is provided as an option in case of extracardiac stimulation (for example, diaphragmatic stimulation for RV pacing). If pacing amplitude is programmed below
5.0 V, an appropriate safety margin (2X the pacing threshold + 1.0 V) should be maintained. An inadequate safety margin may result in loss of capture.
MRI Protection Time-out (nominally set to 24 hours, programmable values of Off, 12, 24, and 48 hours)
The MRI Protection Mode Time-out function allows the user to choose the length of time the pulse generator remains in MRI Protection Mode. Check that the programmer clock is set to the correct time and date to ensure the accuracy of the projected expiration time (displayed on the screen and on the printed MRI Protection Settings Report). When the programmed time has elapsed, the pulse generator automatically exits MRI Protection Mode and returns to the previously programmed settings.
WARNING: When the Time-out parameter is programmed to a value other than Off, the patient must be out of the scanner before the time programmed elapses. Otherwise, the patient will no longer meet Conditions of Use ("MRI Conditions of Use" on page 1-4).
NOTE: Any subsequent interrogation by a PRM while the device is still in MRI Protection Mode will reset the Time-out feature to the start of the initially selected time period.
WARNING: If the MRI Protection Time-out value of Off is combined with a Pacing Mode of Off, the patient will not receive pacing until the pulse generator is programmed out of MRI Protection Mode and back to normal operation. It is recommended to have the Programmer/Recorder/ Monitor (PRM) near the MRI room in case the patient develops the urgent need for pacing. Patients with the following conditions may have increased risk of developing transient pacing­dependence:
A history of syncope related to bradyarrhythmia
A history of syncope of unknown etiology
Sinus pauses (Pause > 2 s), AV block permanently or intermittently
At risk for intermittent AV block (for example, those with progressive AV block, or a history of unexplained syncope)
At risk for trifascicular block (alternating bundle branch block or PR interval > 200 ms with LBBB or other bifascicular block)
WARNING: If the pulse generator enters Safety Mode from MRI Protection Mode, backup pacing will not occur in the following scenarios:
if a functional bipolar ventricular pacing lead is not present
if the Pacing Mode under MRI Protection Mode settings is programmed to Off; the pulse generator will continue permanently with the Pacing Mode programmed to Off, and the patient will not receive pacing therapy until the pulse generator is replaced
WARNING: Exit MRI Protection Mode after MRI scanning is completed. If the MRI Protection Time-out value of Off is selected, the pulse generator will remain permanently in the MRI Protection Mode until it is programmed otherwise. Prolonged use of the MRI Protection Mode (such as may occur when the Time-out feature is programmed to Off) may increase the rate of battery depletion. In addition, prolonged exposure of a patient to the XOO mode chosen may be deleterious to the patient’s health.
1. Programming the Pulse Generator for a Scan
MRI Scan Procedure Protocol
2-9
CAUTION: The physician choosing MRI Protection Mode parameter values will need to
exercise professional judgment to determine an individual patient’s ability to tolerate the pacing parameters required for MR Conditional scanning, in conjunction with the physical conditions required during a scan (for example, prolonged time in a supine position).
CAUTION: If the MR Conditional Pacing System enters Safety Core Operation during MRI Protection Mode and the pacing mode was set to a value other than Off, MRI Protection Mode pacing will be automatically switched to VOO mode, RV bipolar configuration (sensing and pacing), 5.0 V pace pulse amplitude, 1.0 ms pulse width, and 72.5 min
-1
pacing rate as the safety
mode.
Figure 2–5. Program MRI Protection dialog
NOTE: Use of the wand is necessary to complete entry into MRI Protection Mode. Keep the wand in place until receiving confirmation that MRI Protection Mode is programmed. Wanded communication is also required for manual cancellation of MRI Protection Mode (see Manual Exit from MRI Protection Mode in "After the Scan" on page 2-14).
After the values are chosen, the Program MRI Protection button is selected. Selection of the Program MRI Protection button triggers two additional tests: Previous Pacing Threshold and Pacing Lead Configuration. If the results indicate that the Previous Pacing Threshold is less than or equal to 2.0 V and the Pacing Lead Configuration is bipolar, the device enters MRI Protection Mode and the MRI Protection Mode Programmed screen (Figure 2–7 MRI Protection Mode Programmed dialog on page 2-10) appears. The two tests are described below.
Previous Pacing Threshold
The most recently recorded pacing threshold test results (whether from a commanded or automatic test) are used by the programmer to determine if pacing thresholds are less than or equal to 2.0 V, a Condition of Use applicable to pace-dependent patients. Thresholds greater than 2.0 V may result in an insufficient safety margin and failure to capture in MRI Protection Mode (see Table 1–4 Cardiology Conditions/Patient Conditions on page 1-6). If the threshold is greater than 2.0 V, an attention message appears on the PRM screen advising the user to review the risks of proceeding (Figure 2–6 Pacing threshold greater than 2.0 V attention message on page 2-10). Running these tests prior to programming the device to MRI Protection Mode will ensure that the most up-to-date information is used to determine whether this Condition of Use is satisfied.
NOTE: Threshold values available for leads that are not enabled for Daily Measurements will only be as current as the date of the last commanded test. Lack of a pace threshold attention message when MRI Protection Mode is programmed does not mean that all leads have threshold values of 2.0 V or lower.
2-10
MRI Scan Procedure Protocol
1. Programming the Pulse Generator for a Scan
WARNING: Use caution when programming the MRI Protection Mode pacing amplitude for
pacing-dependent patients who have high pacing thresholds (> 2.0 V). Programming pacing amplitude below 5.0 V is provided as an option in case of extracardiac stimulation (for example, diaphragmatic stimulation for RV pacing). If pacing amplitude is programmed below 5.0 V, an appropriate safety margin (2X the pacing threshold + 1.0 V) should be maintained. An inadequate safety margin may result in loss of capture.
Figure 2–6. Pacing threshold greater than 2.0 V attention message
Pacing Lead Configuration
Upon programming parameters for MRI Protection Mode, the device also checks the pacing lead configuration to confirm that it is bipolar in chambers where pacing will occur in MRI Protection Mode. If the lead(s) to be used for pacing while in MRI Protection Mode are programmed to a unipolar pacing configuration, entry into MRI Protection Mode is denied, since the device does not meet the Condition of Use related to bipolar pacing (see Table 1–4 Cardiology Conditions/ Patient Conditions on page 1-6). In order to proceed, program any lead that will be used to pace in the MRI Protection Mode to bipolar or choose a pacing mode of Off.
If threshold tests are within range and the pacing configuration is bipolar in chambers where pacing will occur in MRI Protection Mode, or if the user elects to continue with MRI Protection Mode after reviewing the risks of proceeding in the presence of pacing thresholds greater than
2.0 V, the following screen appears, indicating that the device has successfully been
programmed into MRI Protection Mode at the settings indicated (Figure 2–7 MRI Protection Mode Programmed dialog on page 2-10).
Figure 2–7. MRI Protection Mode Programmed dialog
1. Programming the Pulse Generator for a Scan
MRI Scan Procedure Protocol
2-11
To exit MRI Protection Mode manually, select the Cancel MRI Protection button (see Manual Exit from MRI Protection Mode in "After the Scan" on page 2-14). If necessary, STAT PACE or DIVERT THERAPY can also be used to exit MRI Protection Mode and return the pulse generator to previously programmed settings (DIVERT THERAPY) or initiate STAT PACE pacing parameters (see the pulse generator Reference Guide for more information about STAT PACE).
Once MRI Protection Mode has successfully been programmed, print a copy of the MRI Protection Settings Report by selecting the Print Settings button on the MRI Protection Mode Programmed screen. The report lists the settings in operation during MRI Protection Mode. If the Time-out feature is used, the report includes the time and date when MRI Protection Mode will expire, returning the pulse generator to the pre-MRI Protection Mode settings.
The printed report can be placed in the patient's file and used by radiology personnel, for example, to confirm that sufficient time remains to complete the MRI scan. A sample Settings Report and checklist printout is shown in Figure 2–8 Sample settings report and checklist printout on page 2-12.
Ensure that the HCPs involved in performing the MRI scan have received the model numbers of the pulse generator and lead(s) implanted in the patient.
WARNING: When the Time-out parameter is programmed to a value other than Off, the patient must be out of the scanner before the time programmed elapses. Otherwise, the patient will no longer meet Conditions of Use ("MRI Conditions of Use" on page 1-4).
2-12
MRI Scan Procedure Protocol
1. Programming the Pulse Generator for a Scan
[1] If MRI Protection Time-out is displayed as “Off”, the pulse generator remains in MRI Protection Mode until manually reprogrammed; [2] Twenty-four hour time format is used; [3] Column indicates date measurement was taken
Figure 2–8. Sample settings report and checklist printout
The End Session button will end the current programmer session with the pulse generator remaining in MRI Protection Mode (Figure 2–9 End Session Confirmation dialog on page 2-13).

2. Confirming MRI Scanner Settings and Configuration

Figure 2–9. End Session Confirmation dialog
2. Confirming MRI Scanner Settings and Configuration
Ensure that the MRI scanner equipment meets the "MRI Conditions of Use" on page 1-4. Refer to Table 2–2 Valid Combinations of Pulse Generators and Leads to Use in 1.5 T and 3 T Environments on page 2-13 for component combinations.
MRI Scan Procedure Protocol
2-13
Valid Combinations of Pulse Generators and Leads to Use in 1.5 T and 3 T Environments
Table 2–2. Valid Combinations of Pulse Generators and Leads to Use in 1.5 T and 3 T Environments
Combination of one
INGEVITY MRI Lead and one
FINELINE II Lead
1.5 T scanner only. 3 T scanner not allowed.
Normal Operating Mode only.
ESSENTIO MRI Pulse Generator PROPONENT MRI Pulse Generator ACCOLADE MRI Pulse Generator
INGEVITY MRI Leads
only
1.5 T or 3 T scanner allowed.
Normal Operating Mode or First Level Controlled Operating Mode.
FINELINE II Leads only
1.5 T scanner only. 3 T scanner not allowed.
Normal Operating Mode only.
Refer to Table 1–2 System Configuration for 1.5 T on page 1-3 and Table 1–3 System Configuration for 3 T on page 1-3 for a complete list of the model numbers of MR Conditional Pacing System components.
Refer to "MRI Conditions of Use" on page 1-4 for the entire set of MRI Conditions of Use.
WARNING: Only the combination of INGEVITY MRI lead(s) with an ESSENTIO MRI, PROPONENT MRI, or ACCOLADE MRI pulse generator is valid to use with either 1.5 T or 3 T scanners. All other allowable combinations of Boston Scientific MR Conditional system components must use only 1.5 T scanners.
NOTE: 3 T MRI scanners should be operated in Circularly Polarized mode or Dual Channel Multi-Transmission mode since ImageReady Pacing Systems have been evaluated against 3 T MRI scanners during Circularly Polarized and Dual Channel Multi-Transmission RF operation modes. Multi-Transmission RF operation with greater than two transmit channels has not been evaluated.

3. Preparing the Patient for the Scan

The patient must not have an elevated temperature or compromised thermoregulation. Patient position within the bore must be prone or supine, and the appropriate monitoring system must be put in place (pulse oximetry and/or ECG). See "MRI Conditions of Use" on page 1-4.
If the MRI Protection Mode Time-out feature is being used, be sure to note the time at which the pulse generator is scheduled to exit MRI Protection Mode. Refer to Figure 2–8 Sample settings report and checklist printout on page 2-12.
2-14
MRI Scan Procedure Protocol

During the Scan

NOTE: If the time remaining is not sufficient for the patient to undergo the MRI scan, re­interrogation of the device will reset the Time-out value to the start of the originally programmed timer setting.
WARNING: When the Time-out parameter is programmed to a value other than Off, the patient must be out of the scanner before the time programmed elapses. Otherwise, the patient will no longer meet Conditions of Use ("MRI Conditions of Use" on page 1-4).
Image distortion must be considered when planning an MRI scan, and when interpreting MRI images of fields containing the pulse generator and/or leads. Pulse generator artifacts extend beyond the margin of the device in all directions. Only minor artifacts are present around the pacing lead. Some artifacts include moderate spatial distortion beyond the boundaries of the visible pulse generator artifact. Gradient Recalled Echo artifacts are generally larger and more prone to have accompanying spatial distortion than Spin Echo artifacts.
DURING THE SCAN
Patient Monitoring
Normal voice and visual contact, as well as pulse oximetry and/or ECG, must be monitored for the duration of the scan.
WARNING: Ensure that an external defibrillator and medical personnel skilled in cardio­pulmonary resuscitation (CPR) are present during the MRI scan should the patient require external rescue.
WARNING: If the MRI Protection Time-out value of Off is combined with a Pacing Mode of Off, the patient will not receive pacing until the pulse generator is programmed out of MRI Protection Mode and back to normal operation. It is recommended to have the Programmer/Recorder/ Monitor (PRM) near the MRI room in case the patient develops the urgent need for pacing. Patients with the following conditions may have increased risk of developing transient pacing­dependence:
A history of syncope related to bradyarrhythmia
A history of syncope of unknown etiology
Sinus pauses (Pause > 2 s), AV block permanently or intermittently
At risk for intermittent AV block (for example, those with progressive AV block, or a history of unexplained syncope)
At risk for trifascicular block (alternating bundle branch block or PR interval > 200 ms with LBBB or other bifascicular block)
WARNING: The Programmer/Recorder/Monitor (PRM) is MR Unsafe and must remain outside the MRI site Zone III (and higher) as defined by the American College of Radiology Guidance Document for Safe MR Practices
2
. Under no circumstances should the PRM be brought into the
MRI scanner room, the control room, or the MRI site Zone III or IV areas.

AFTER THE SCAN

WARNING: Exit MRI Protection Mode after MRI scanning is completed. If the MRI Protection
Time-out value of Off is selected, the pulse generator will remain permanently in the MRI Protection Mode until it is programmed otherwise. Prolonged use of the MRI Protection Mode (such as may occur when the Time-out feature is programmed to Off) may increase the rate of
2. Kanal E, et al., American Journal of Roentgenology 188:1447-74, 2007.
MRI Scan Procedure Protocol
After the Scan
2-15
battery depletion. In addition, prolonged exposure of a patient to the XOO mode chosen may be deleterious to the patient’s health.
Manually exit MRI Protection Mode using the PRM after the scan (see Manual Exit from MRI Protection Mode). After exit from MRI Protection Mode, check system integrity by running lead impedance, pacing threshold, and intrinsic amplitude tests.
Upon exit from MRI Protection Mode, all parameters are immediately restored to pre-MRI Protection Mode values with two exceptions. If PaceSafe Automatic Capture (RVAC) was programmed on, this feature enters suspension upon entry of the device into MRI Protection Mode. Upon exit from MRI Protection Mode, RV pace amplitude is set to two times the last capture threshold determined by the RVAC feature before it entered suspension (output limited to between 3.5 V and 5.0 V). After the next scheduled autothreshold test runs (within the next 21 hours) and is successful, the RV pace amplitude is set to the new capture threshold plus 0.5 V. This behavior was designed to provide a safety margin against loss of capture during the transient period between MRI completion and full body recovery from effects of the scanner electromagnetic fields.
For details about the PaceSafe Automatic Capture feature, see the Reference Guide for the pulse generator.
Restoration of function of the Minute Ventilation sensor is also delayed upon exit from MRI Protection Mode. If MV is programmed to On or Passive at the time of entry into MRI Protection Mode, upon exit from the mode, an automatic six-hour calibration of the sensor will begin. MV­driven rate response is not available during this calibration period. If MV-driven rate response is desired sooner, a manual calibration can be performed. Manual calibration is completed in five minutes or less. For additional information about MV calibration, see the Reference Guide for the pulse generator.
Time-out (automatic) Exit from MRI Protection Mode
If the MRI Protection Mode Time-out parameter was programmed to a value other than Off, the pulse generator will exit MRI Protection Mode automatically after the selected number of hours, and the system will return to previously programmed settings (for details about the resumption of PaceSafe Automatic Capture and Minute Ventilation, see "After the Scan" on page 2-14).
Manual Exit from MRI Protection Mode
Manually exit MRI Protection Mode using the PRM after the scan.
Do not leave the pulse generator in MRI Protection Mode any longer than necessary following the scan. MRI Protection Mode pacing is performed at a fixed, elevated rate. Some patients may experience side effects during protracted pacing in this mode, including decreased exercise capacity, acceleration of heart failure, and proarrhythmia.
Interrogate the pulse generator using the wand (RF telemetry is disabled in MRI Protection Mode)
Select the Cancel MRI Protection Mode button from the MRI Protection Mode Programmed screen (Figure 2–10 MRI Protection Mode Programmed (Cancel MRI Protection) on page 2-
16)
2-16
MRI Scan Procedure Protocol
After the Scan
Figure 2–10. MRI Protection Mode Programmed (Cancel MRI Protection)
Following user-initiated cancellation of MRI Protection Mode, the PRM will automatically navigate to the Lead Tests screen and prompt the user to perform the following lead tests (Figure 2–11 MRI Protection Cancelled dialog on page 2-16):
Lead impedance
Pacing threshold
Intrinsic amplitude
These tests may be performed subsequent to automatic (Time-out) exit from MRI Protection Mode as well.
Figure 2–11. MRI Protection Cancelled dialog
On exit from MRI Protection Mode, a summary report of the MRI is stored as an MRI episode and can be printed as an episode report using the PRM printer. The MRI Protection episode can also be accessed and viewed via the Arrhythmia Logbook. A sample stored event detail printout is shown in Figure 2–12 Sample stored event detail printout on page 2-17.
The MRI episode can also be viewed on the Arrhythmia Logbook via remote patient monitoring (if available).
Figure 2–12. Sample stored event detail printout
MRI Scan Procedure Protocol
After the Scan
2-17
2-18
MRI Scan Procedure Protocol
After the Scan
A-1
CARDIOLOGY CHECKLIST FOR THE IMAGEREADY PACING SYSTEM
APPENDIX A
This appendix is provided for convenience. Refer to the remainder of this Technical Guide for the full list of Warnings and Precautions and complete instructions for using the ImageReady Pacing System.
Conditions of Use – Cardiology Scanning Procedure
The following Conditions of Use must be met in order for a patient with an ImageReady Pacing System to undergo an MRI scan
Patient is implanted with an ImageReady MR Conditional Pacing System (see "ImageReady Pacing System Components for 1.5 T and 3 T" on page C-1)
Pulse generator in MRI Protection Mode during scan
Bipolar pacing operation or pacing off
Patient does not have elevated body temperature or compromised thermoregulation at time of scan
Pulse generator implant location restricted to left or right pectoral region
At least six (6) weeks have elapsed since implantation and/or any lead revision or surgical modification of the MR Conditional Pacing System
No cardiac-related implanted devices, components, or accessories present other than an ImageReady MR Conditional Pacing System
Pacing threshold ≤ 2.0 V in pace-dependent patients
No abandoned leads or pulse generators
No evidence of a fractured lead or compromised pulse generator-lead system integrity
Pre-scan
1. Ensure patient meets all Cardiology Conditions of Use for MRI scanning (see left column).
2. Ensure that the HCPs involved in performing the MRI scan have received the model numbers of the pulse generator and lead(s) implanted in the patient.
3. As close to the start of the scan as possible, program the pulse generator into MRI Protection Mode.
4. Print the MRI Protection Settings Report, place it in the patient’s file, and provide to radiology personnel.
The report documents MRI Protection Mode settings and details. If the Time-out feature is used, the report includes the exact time and date when MRI Protection Mode will expire
During Scan
5. Ensure the patient is continuously monitored by pulse
oximetry and electrocardiography (ECG), with backup therapy available (external rescue), while the device is in MRI Protection Mode.
After Scan
6. Manually exit MRI Protection Mode using the PRM to
return the pulse generator to pre-MRI operation. Perform follow-up testing of the implanted system.
WARNING: Unless all of the MRI Conditions of Use are met, MRI scanning of the patient does not meet MR Conditional requirements for the implanted system, and significant harm to or death of the patient and/or damage to the implanted system may result.
WARNING: Only the combination of INGEVITY MRI lead(s) with an ESSENTIO MRI, PROPONENT MRI, or ACCOLADE MRI pulse generator is valid to use with either 1.5 T or 3 T scanners. All other allowable combinations of Boston Scientific MR Conditional system components must use only 1.5 T scanners.
WARNING: The Programmer/Recorder/Monitor (PRM) is MR Unsafe and must remain outside the MRI site Zone III (and higher) as defined by the American College of Radiology Guidance Document for Safe MR Practices
1
. Under no circumstances should the PRM be brought into the
MRI scanner room, the control room, or the MRI site Zone III or IV areas.
1. Kanal E, et al., American Journal of Roentgenology 188:1447-74, 2007.
A-2
CARDIOLOGY CHECKLIST FOR THE IMAGEREADY PACING SYSTEM
RADIOLOGY CHECKLIST FOR THE IMAGEREADY PACING SYSTEM
APPENDIX B
This appendix is provided for convenience. Refer to the remainder of this Technical Guide for the full list of Warnings and Precautions and complete instructions for using the ImageReady Pacing System.
Conditions of Use – Radiology Scanning Procedure
B-1
The following Conditions of Use must be met in order for a patient with an ImageReady Pacing System to undergo an MRI scan.
MRI magnet strength of 1.5 T or 3 T (see "ImageReady Pacing System Components for 1.5 T and 3 T" on page C-1)
1.5 T
Radio frequency (RF) field of approximately 64 MHz
Spatial gradient no greater than 50 T/m (5,000 G/cm)
3 T
RF field of approximately 128 MHz
Spatial gradient no greater than 50 T/m (5,000 G/cm)
Horizontal,1H proton, closed bore scanners only
Specific Absorption Rate (SAR) limits:
For FINELINE II leads only or for one FINELINE II lead and one INGEVITY MRI lead, SAR limits for Normal Operating Mode entire active scan session as follows:
Whole body averaged, ≤ 2.0 watts/kilogram (W/
kg)
Head, ≤ 3.2 W/kg
For INGEVITY MRI leads only, SAR limits for Normal Operating Mode Operating Mode active scan session as follows:
Whole body averaged, ≤ 4.0 W/kg
Head, ≤ 3.2 W/kg
a
must be observed for the
a
or for First Level Controlled
b
must be observed for the entire
Pre-scan
1. Ensure Cardiology has cleared the patient for scanning eligibility based on the Cardiology MRI Conditions of Use (see "Cardiology Checklist for the ImageReady Pacing System" on page A-1) and has provided the model numbers of the pulse generator and lead(s) implanted in the patient.
2. Ensure patient meets all Radiology Conditions of Use for MRI scanning (see left column).
3. Refer to the MRI Protection Settings Report to confirm that the patient’s device is in MRI Protection Mode. If the Time-out feature is used, the report includes the exact time and date when MRI Protection Mode will expire.
Verify that adequate time remains to complete the scan.
During Scan
4. Ensure the patient is continuously monitored by pulse oximetry and electrocardiography (ECG), with backup therapy available (external rescue), while the device is in MRI Protection Mode.
After Scan
5. Manually exit MRI Protection Mode using the PRM to return the pulse generator to pre-MRI operation. Perform follow-up testing of the implanted system.
Gradient Field limits: Maximum specified gradient slew
rate ≤ 200 T/m/s per axis
No local transmit-only coils or local transmit/receive coils placed directly over the pacing system; the use of receive-only coils is not restricted
Patient in supine or prone position only
The patient must be monitored during the MRI scan by pulse oximetry and/or electrocardiography (ECG)
a. As defined in IEC 60601-2-33, 201.3.224, 3rd Edition. b. As defined in IEC 60601-2-33, 201.3.208, 3rd Edition.
WARNING: Unless all of the MRI Conditions of Use are met, MRI scanning of the patient does not meet MR Conditional requirements for the implanted system, and significant harm to or death of the patient and/or damage to the implanted system may result.
WARNING: Only the combination of INGEVITY MRI lead(s) with an ESSENTIO MRI, PROPONENT MRI, or ACCOLADE MRI pulse generator is valid to use with either 1.5 T or 3 T
B-2
RADIOLOGY CHECKLIST FOR THE IMAGEREADY PACING SYSTEM
scanners. All other allowable combinations of Boston Scientific MR Conditional system components must use only 1.5 T scanners.
WARNING: The Programmer/Recorder/Monitor (PRM) is MR Unsafe and must remain outside the MRI site Zone III (and higher) as defined by the American College of Radiology Guidance Document for Safe MR Practices
1
. Under no circumstances should the PRM be brought into the
MRI scanner room, the control room, or the MRI site Zone III or IV areas.
CAUTION: The presence of the implanted Pacing System may cause MRI image artifacts (see "3. Preparing the Patient for the Scan" on page 2-13).
NOTE: All normal risks associated with an MRI procedure apply to MRI scans with the MR Conditional Pacing System. Consult MRI scanner documentation for a complete list of risks associated with MRI scanning.
NOTE: Other implanted devices or patient conditions may cause a patient to be ineligible for an MRI scan, independent of the status of the patient’s ImageReady MR Conditional Pacing System.
NOTE: 3 T MRI scanners should be operated in Circularly Polarized mode or Dual Channel Multi-Transmission mode since ImageReady Pacing Systems have been evaluated against 3 T MRI scanners during Circularly Polarized and Dual Channel Multi-Transmission RF operation modes. Multi-Transmission RF operation with greater than two transmit channels has not been evaluated.
1. Kanal E, et al., American Journal of Roentgenology 188:1447-74, 2007.
IMAGEREADY PACING SYSTEM COMPONENTS FOR 1.5 T AND 3 T
APPENDIX C
Only specific combinations of pulse generators and leads constitute an ImageReady Pacing System. Consult the following tables to determine which combinations are valid for use with only
1.5 T scanners and which are valid for use with both 1.5 T and 3 T scanners.
Table C–1. Valid Combinations of Pulse Generators and Leads to Use in 1.5 T and 3 T Environments
INGEVITY MRI Leads
ESSENTIO MRI Pulse Generator PROPONENT MRI Pulse Generator ACCOLADE MRI Pulse Generator
Table C–2. ImageReady MR Conditional Pacing System Components for 1.5 T and 3 T
1.5 T or 3 T scanner allowed.
Normal Operating Mode or First Level Controlled Operating Mode.
only
FINELINE II Leads only
1.5 T scanner only. 3 T scanner not allowed.
Normal Operating Mode only.
Combination of one
INGEVITY MRI Lead and one
FINELINE II Lead
1.5 T scanner only. 3 T scanner not allowed.
Normal Operating Mode only.
C-1
Component Model Number(s) MR Status Valid
Pulse Generators
ESSENTIO MRI L110, L111, L131 MR Conditional
PROPONENT MRI L210, L211, L231 MR Conditional
ACCOLADE MRI L310, L311, L331 MR Conditional
Leads and Accessories
FINELINE II Sterox Pacing Lead 4456, 4457, 4458,
FINELINE II Sterox EZ Pacing Lead 4469, 4470, 4471,
Suture Sleeve for FINELINE II leads 6220, 6221 MR Conditional
INGEVITY MRI Pacing Lead (Tined Fixation) 7731, 7732, 7735,
INGEVITY MRI Pacing Lead (Extendable/ Retractable Fixation)
Suture Sleeve for INGEVITY MRI leads 6402 MR Conditional
IS-1 Lead Port Plug 7145 MR Conditional
ZOOM LATITUDE Programmer/Recorder/Monitor (PRM) and PRM Software Application
ZOOM LATITUDE PRM 3120 MR Unsafe
ZOOM LATITUDE PRM Software App. 2869 N/A
4459, 4479, 4480
4472, 4473, 4474
7736
7740, 7741, 7742 MR Conditional
MR Conditional
MR Conditional
MR Conditional
a
Combinations
For valid
combinations for 1.5
T and 3 T see the “Valid Combinations of Pulse Generators
and Leads to Use in
1.5 T and 3 T
Environments” table
above.
a. See PRM is MR Unsafe Warning regarding the PRM.
C-2
IMAGEREADY PACING SYSTEM COMPONENTS FOR 1.5 T and 3 T
SYMBOLS ON PACKAGING
APPENDIX D
The following symbols may be used on packaging and labeling.
Table D–1. Symbols on Packaging
Symbol Description
D-1
MR Conditional
Reference number
D-2
Symbols on Packaging
INDEX
A
Abandoned leads or pulse generators 1-4, 1-7–1-8 ACCOLADE MRI 1-2–1-3, 1-12, 2-13 Active implantable medical devices (AIMDs) 1-11 Arrhythmia Logbook 2-16
B
Battery capacity status 2-4 Bipolar pacing configuration 1-4, 1-6, 2-10
C
Cardiology Checklist A-1 Closed bore 1-5, 1-9, 2-13 Coils 1-11
receive-only 1-5, 1-10 transmit-only 1-5, 1-10 transmit/receive 1-5, 1-10
M
Magnet sensor 2-5 Minute Ventilation 2-15 Models for use with 1.5 T 1-3 Models for use with 3 T 1-3 MR Unsafe 1-3 MRI magnet strength
1.5 T 1-2
1.5 Tesla 1-2–1-4, 1-8, 1-11, 2-13 3 T 1-2
3 Tesla 1-2–1-3, 1-5, 1-8–1-9, 1-11, 2-13 MRI Protection Checklist 2-6–2-7 MRI Protection episode 2-16 MRI Protection Mode 1-4, 1-11, 2-5, 2-15
automatic exit 2-15
conditions preventing entry 2-4–2-5
default pacing mode 2-3
entry into 2-5
manual exit 2-9, 2-11, 2-15
risks when Conditions of Use are not met 2-7, 2-9–
2-10 suspended features and functions 2-4 Time-out feature 1-2, 2-2–2-4, 2-8, 2-11, 2-13, 2-
16
MRI Protection Settings Report 1-6, 2-2, 2-8, 2-11
E
Electrocautery Mode 2-4 ESSENTIO MRI 1-2–1-3, 1-12, 2-13
F
FINELINE II 1-2, 1-5, 1-10 First level controlled operating mode 1-2, 1-5, 1-10,
2-13
Fractured lead 1-4, 1-8
I
Image distortion 2-14 ImageReady MR Conditional Pacing System 1-2, 1-
4, 1-6–1-8 INGEVITY MRI 1-2–1-3, 1-5, 1-10, 1-12, 2-13 Intrinsic amplitude 1-8, 2-4, 2-15–2-16
L
LATITUDE 1-3 Lead impedance 1-8, 2-4, 2-6, 2-15–2-16 Leads
FINELINE II 1-2, 1-5, 1-10 INGEVITY MRI 1-2–1-3, 1-5, 1-10, 1-12, 2-13
N
Normal operating mode 1-2, 1-5, 1-10, 2-13
O
Operating mode
first level controlled 1-2, 1-5, 1-10, 2-13 normal 1-2, 1-5, 1-10, 2-13
P
Pace-dependent patients 1-4, 1-7, 2-9 PaceSafe Automatic Capture 2-15 Pacing threshold 1-4, 1-7, 2-10, 2-15–2-16 Pacing threshold changes 1-6, 1-15 Patient position 1-5, 1-11, 2-13 Previous pacing threshold 2-9 PRM 1-2–1-3 Programmer wand 2-5, 2-9, 2-15 PROPONENT MRI 1-2–1-3, 1-12, 2-13 Pulse generators 1-3
ACCOLADE MRI 1-2–1-3, 1-12, 2-13 ESSENTIO MRI 1-2–1-3, 1-12, 2-13 PROPONENT MRI 1-2–1-3, 1-12, 2-13
Pulse oximetry 1-5, 1-11, 2-13–2-14
Q
Quick Reference Guide C-1
R
Radiology Checklist B-1 Receive-only coils 1-5, 1-10 RF telemetry 2-4–2-5, 2-15
S
Safety Core operation 1-6, 2-4 SAR limits 1-5, 1-10 Six weeks since implant 1-4, 1-7, 1-15 Specific Absorption Rate (SAR) limits 1-5, 1-10 STAT PACE mode 2-5, 2-11 Storage Mode 1-7, 2-4, 2-6 System integrity 2-15
compromised 1-4, 1-8
T
Tesla
1.5 T 1-2–1-4, 1-8, 1-11, 2-13
3 T 1-2–1-3, 1-5, 1-8–1-9, 1-11, 2-13 Time since implant 2-6 Transmit-only coils 1-5, 1-10 Transmit/receive coils 1-5, 1-10
U
Unipolar pacing configuration 2-5, 2-10
V
Valid combinations 1-2, 2-13 Ventricular episode 2-5
Z
ZOOM LATITUDE 1-3
Cardiac Pacemakers Incorporated a wholly owned subsidiary of Guidant Corporation a wholly owned subsidiary of Boston Scientific Corporation 4100 Hamline Avenue North St. Paul, MN 55112-5798 USA
www.bostonscientific.com
1.800.CARDIAC (227.3422)
+1.651.582.4000
©
2017 Boston Scientific Corporation or its affiliates.
All rights reserved.
359486-005 EN Canada 2017-08
*359486-005*
Loading...