CAUTION- Investigational Device. Limited by Federal (or United States) law to investigational use.
EP Technologies, Inc.
2710 Orchard Parkway
San Jose, CA 95134
Phone: 408.895.3500P/N 13754-901 EAW Ver AA 6/01
Manual Conventions
Preface
The following naming conventions are used in the EPT-1000 XP
System Operator’s Manual:
•The EPT-1000 XP
complete description of the System’s components, refer to Section 1, System
Description.
• The EPT-1000 XPTM Cardiac Ablation Controller is referred to as the “Controller.”
• The High Power Automatic Personality Module is referred to as the “XP APM.”
The Catheters are referred to as follows:
• Steerable Ablation Catheter with temperature measuring capability: “Catheter(T°).”
• Steerable Ablation Catheter without temperature measuring capability:
“Catheter(Non-T).”
•Catheter information pertaining to both Catheter styles: “Catheter.”
General Information
The use of all components and accessories of the EPT-1000 XPTM Cardiac Ablation
System is fully described in this Manual, except for the Steerable Ablation Catheters.
This Manual provides a description of the Controller, its controls and displays, and a
sequence for its operation. In addition, it provides information relevant to the operation of
the XP APM with the EPT-1000 XPTM only, as well as other information of importance
to the user.
TM
Cardiac Ablation
TM
Cardiac Ablation System is referred to as the “System.” For a
!
Precaution
Operator’s Manual. It is important that the equipment’s operating instructions be read,
understood and followed. For future reference, retain this Manual in a convenient,
readily accessible place.
Do not attempt to operate the System before thoroughly reading this
The EPT-1000 XPTM Cardiac Ablation System (see Figure 1-1) consists of two principal
subsystems:
• EPT Model #800XP Cardiac Ablation Controller (RF Generator)
• Catheter(T°) or Catheter(Non-T) with Quick Connect Instrument Cables for
connection to the EPT Model #822T High Power Automatic Personality Module
"
Two accessories are provided with the Controller:
• EPT Model 822T High Power Automatic Personality Module (XP APM)
• EPT Model 840 Footswitch
For system operation and power level access greater than 50 W/1 A, it is mandatory to
use the XP APM. If accessing up to 100 W/2 A power levels, both an XP APM and a
special high power catheter must be used. The Footswitch usage is optional as the
Controller can be operated directly from its front panel controls. The EPT-1000 XP
Cardiac Ablation System is operated in conjunction with commercially available external
Disposable Indifferent (Dispersive) Patch (DIP) Electrodes, which are in compliance with
ANSI/AAMI Standard HF-18.
"
When using special high power catheters with the EPT-1000 XPTM Controller, two
Note
DIP Electrodes are required.
The Catheter delivers RF power in a monopolar mode between its distal electrode and the
large DIP Electrode(s). Detailed information regarding the Catheter is contained in the
accompanying manual titled, Blazer II XP (High Power) Cardiac Ablation Catheter
Instructions for Use.
Refer to Figure 1-2 for cable configuration.
Note
TM
1-1
System Features
The EPT-1000 XPTM Cardiac Ablation System features the following:
•The Controller produces a continuous unmodulated radio frequency (RF) output at
500 kHz.
•The front panel displays the actual power output, tissue impedance, and if a
Catheter(T°) is connected to the Controller, catheter tip temperature.
•The Controller measures RMS voltage, RMS current, and actual power output by
taking the average value of the product of voltage and current. (This reflects the
effective heating power delivered to the tissue from the Catheter electrode tip.
Impedance is calculated as RMS voltage divided by RMS current.)
•The amount and duration of RF power delivery is user-selectable, and if a
Catheter(T°) is connected, the desired catheter tip temperature or maximum catheter
tip temperature is also user-selectable.
•The number of RF power deliveries is automatically counted for the physician.
Operator’s Manual
•The XP APM provides RF filtering to allow continuous electrogram recording during
RF power delivery via the catheter tip electrode.
•The XP APM allows for safe use of either standard or special high power EPT
ablation catheters.
Safety Features
The System’s built-in safety features include automatic shutoff for the following
conditions:
• If accessing up to 50 W of RF power and measured tissue impedance is less than 50
Ω.
• If accessing up to 100 W of RF power and measured tissue impedance is less than 25
Ω.
• If measured tissue impedance exceeds 300 Ω.
• If maximum voltage or current limits are exceeded.
• The Controller shuts off if the measured temperature exceeds 85°C for greater than 1
second, or if measured temperature exceeds the upper temperature limit selected in
the Power Control mode. This temperature is not the maximum tissue temperature.
The recorded temperature may be up to 15°C different from the maximum tissue
temperature. Do not set the temperature greater than 80°C.
1-2
System Description
•The Controller shuts off if the measured temperature exceeds the user-selected
temperature setpoint by 5°C for greater than 4 seconds. This temperature is not the
maximum tissue temperature. Do not set the temperature greater than 80°C.
•If greater than 1 A of current is delivered to either of the two DIP Electrodes
connected to the XP APM, the Controller shuts off.
Control Modes
The Controller is capable of operating in two “control modes” (Power Control or
Temperature Control).
Power Control
The Power Control mode is compatible with either Catheter type and allows selection of
the quantity of RF power that is delivered by the Catheter. If using a Catheter(T°), the
Controller delivers the set power unless the upper temperature limit is exceeded, which
results in a reduction in the amount of power delivered.
Temperature Control
The Temperature Control mode is only compatible with Catheter(T°) and allows selection
of a desired catheter tip temperature.
The maximum amount of power to be delivered by the Catheter must be selected for both
control modes (Power Control or Temperature Control). If insufficient power is selected
for the Temperature Control mode, the desired catheter tip temperature (temperature
setpoint) may not be achieved. This may indicate that the Catheter should be repositioned
to achieve better tip contact or a higher power setpoint should be selected.
"
When using the EPT-1000 XPTM Cardiac Ablation Controller in the Temperature
Note
Control mode, ALWAYS ensure that the TEMP LED on the Control Panel is
illuminated.
The following table defines the power delivery protocol for both the Power Control and
Temperature Control modes for the Controller with XP APM connected.
Table 1-1: Power Delivery Protocol
Catheter TypeMaximum
Power Setting
(W)
Maximum
Current
(A)
Catheter(Non-T)501
Catheter(T°)501
Special High Power Catheter(T°)1002
1-3
Figure 1-1: System Installation Diagram
Operator’s Manual
1-4
System Description
Figure 1-2: Cable Configuration for Q ui ck Connect XP APM
CARDIAC
ABLATION
EPT-1000 XP
CONTROLLER
(10ft.)
Model 654
Model 822T
(Optional)
APM Extension Cable
(Switch Box)
Quick Connect XP APM
EGM
EQUIPMENT
Model 653
APM to EGM
Instrument Cable
Quick Connect 6-Pole 3ft.
Model 651
Quick Connect 9-Pole 3 ft.
Thermistor Instrument Cable
9-PINQCOUT
WITH
QUICK
CONNECT
CATHETER
THERMISTOR
Model 652
Adapter Cable
Quick Connect 6-Pole to 9-Pole 3ft.
6-PINQCOUT
WITH
QUICK
CONNECT
CATHETER
NONTHERMISTOR
Model 703
Quadripolar
Adapter Cable
Quick Connect 9-Pole
PINS
OUT
PIN
CATHETER
CONNECTS
THERMISTOR/
NONTHERMISTOR
1-5
2. Indications/Contraindications
Indications for Use
The System is indicated for interruption of accessory atrioventricular conduction
pathways associated with tachycardia, for treatment of AV nodal reentrant tachycardia,
for treatment of atrial flutter tachycardia, and for creation of complete AV block in
patients with a rapid ventricular response to an atrial arrhythmia – typically chronic, drug
refractory atrial fibrillation.
Contraindications
• Use of the device is contraindicated for patients with active systemic infection.
• Use of the device in patients with implanted permanent leads of an implantable
device of any kind.
• Patients with implanted mitral/tricuspid artificial valve.
• Pregnant women.
• Presence of tumors or clots in atrium and/or ventricules.
• The retrograde transaortic approach is contraindicated in patients with aortic valve
replacement.
#Note: Refer to individual Directions for Use for the Catheter and ground pads
contraindications.
2-1
3. Warnings, Precautions, and Adverse Reactions
Warnings
Before operating the System, read these warnings carefully:
•The displayed temperature is not the maximum tissue temperature. Do not set the
maximum temperature higher than 80°C.
•Use Lower Power First - When first delivering RF energy, begin by using a low
power setting (i.e., 50W). If the created lesion is unsuccessful or inadequate,
incrementally increase the power output with successive ablation attempts to
minimize the potential for thrombus formation and /or inadvertent damage to cardiac
tissues.
•Catheter ablation procedures present the potential for significant X-ray exposure,
which can result in acute radiation injury as well as increased risk for somatic and
genetic effects, to both patients and laboratory staff due to the X-ray beam intensity
and duration of the fluoroscopic imaging. Catheter ablation should only be performed
after adequate attention has been given to the potential radiation exposure associated
with the procedure, and steps taken to minimize this exposure. Careful consideration
must therefore be given for the use of the device in pregnant women.
•Patients undergoing AV nodal modification or ablation of septal accessory pathways
are at risk for inadvertent AV block. It is advisable to use lower initial power in such
patients and to monitor anterior conduction closely during RF power delivery.
•Pacemakers and implantable cardioverter/defibrillators can be adversely affected by
RF signals. It is important to:
a) Retain temporary external sources of pacing available during ablation.
b) Reprogram the pacing system temporarily to minimum output or 000 mode to
reduce risk of inappropriate pacing.
c) Exercise extreme caution during ablation when in close proximity to atrial or
ventricular permanent pacing leads.
d) Perform complete pacing system analysis on all patients after ablation.
•Implanted cardioverter/defibrillators should be deactivated during delivery of RF
power.
•During a transaortic approach, adequate fluoroscopic visualization is necessary to
avoid placement of the ablation catheter within the coronary vasculature. Catheter
placement and RF power application within the coronary artery has been associated
with myocardial infarction and death.
3-1
Operator’s Manual
•Patients undergoing left-sided ablation procedures should be closely monitored
during the post-ablation period for clinical manifestations of infarction.
•The Steerable Ablation Catheter is intended for single patient use only. Do not
reprocess or reuse. Reuse can cause patient injury and/or the communication of
infectious disease(s) from one patient to another.
•Catheter entrapment within the heart or blood vessels is a possible complication of
cardiac ablation procedures. The potential for catheter entrapment may be increased
when the Catheter is positioned in the vicinity of the chordae tendinae. The occurrence
of this complication may necessitate surgical intervention and/or repair of injured
tissues.
•The possibility of skin burns to the patient may exist whenever a special high power
catheter is utilized.
•The use of catheters or cables with unprotected male pin connectors presents a risk of
electrical hazard. Inadvertent attachment of pin connectors to power supply sockets or
connectors could result in electrocution of the patient or operator. Misconnection of
the pins could also lead to inappropriate delivery of RF current through a band
electrode. The users of components with unprotected male pin connectors must
exercise extreme caution during device set-up to prevent patient or operator injury.
Unused pins should be secured so that they do not inadvertantly touch other equipment
or surfaces. Never insert pins into mains outlets or into any equipment other than
systems providing patient electrical isolation in accordance with I EC-601.
•The APM output labeled recorder must only be connected to a medically isolated
recorder.
•Grounding reliability can only be acheived when the power supply cord is connected to
a receptacle marked Hospital Only or Hospital Grade.
Precautions
Observe these precautions, before using the System:
•Careful catheter manipulation must be performed in order to avoid cardiac damage,
perforation, or tamponade. Catheter advancement should be done under fluoroscopic
guidance. Do not use excessive force to advance or withdraw the Catheter when
resistance is encountered.
• The sterile packaging should be inspected prior to use.
• When using special high power catheters, it is required that two Disposable
Indifferent Patch Electrodes satisfying the requirements of ANSI/AAMI Standard
HF-18 be used as the ablation return electrodes or skin burns may result.
• It is recommended not to exceed thirty (30) RF power applications per Catheter.
• The EPT-1000 XPTM Cardiac Ablation Controller is intended for use with EP
Technologies Steerable Ablation Catheters and accessories only.
3-2
Warnings, Precautions, and Adverse Reactions
•Do not attempt to operate the EPT Cardiac Ablation System before thoroughly
reading the EPT-1000 Cardiac Ablation Controller Operator's Manual.
•The IMPEDANCE Display of the Controller should be continuously monitored
during RF power delivery. If a sudden rise in impedance is noted, power delivery
should be discontinued. The Catheter should be removed and the distal tip of the
Catheter cleaned to eliminate any coagulum.
•Excessive bending or kinking of the catheter shaft may damage internal wires.
Manual prebending of the distal curve can damage the steering mechanism and may
cause patient injury.
•Cardiac ablation procedures should be performed only by physicians thoroughly
trained in the techniques of RF Powered Catheter Ablation in a fully-equipped
electrophysiology laboratory.
•Adequate filtering must be used to allow continuous monitoring of the surface
electrocardiogram (ECG) during RF power applications.
•The long-term risks of protracted fluoroscopy have not been established. Careful
consideration must therefore be given for the use of the device in prepubescent
children.
•The long-term risks of lesions created by RF ablation have not been established. In
particular, any long-term effects of lesions in proximity to the specialized conduction
system or coronary vasculature are unknown. Furthermore, the risk/benefit in
asymptomatic patients has not been studied.
•Read and follow the Disposable Indifferent (Dispersive) Patch (DIP) Electrode
manufacturer’s instructions for use; the use of DIP Electrodes which meet or exceed
ANSI/AAMI HF-18 requirements is required.
•Placement of a DIP electrode on the thigh could be associated with higher
impedance, which could result in automatic RF power shut-off.
•The Controller is capable of delivering significant electrical power. Patient or
operator injury can result from improper handling of the Catheter and DIP Electrode,
particularly when operating the device. During power delivery, the patient should not
be allowed to come in contact with grounded metal surfaces.
•Apparent low power output or failure of the equipment to function correctly at
normal settings may indicate faulty application of the DIP Electrode or failure of an
electrical lead. Do not increase power before checking for obvious defects or
misapplication.
•The risk of igniting flammable gases or other materials is inherent in the application
of RF power. Precautions must be taken to restrict flammable materials from the
ablation site.
•Electromagnetic interference (EMI) produced by the Controller during the delivery of
RF power may adversely affect the performance of other equipment.
3-3
• Regularly inspect and test re-usable cables and accessories.
• EP Technologies relies on the physician to determine, assess and communicate to the
individual patient all foreseeable risks of the cardiac ablation procedure.
Potential Adverse Reactions
The risks involved in RF ablation for the treatment of conditions outlined in the
Indications for Use include those associated with all cardiac catheter ablation procedures.
These potential adverse events may differ in severity from mild to severe. Additionally,
these complications may necessitate additional medical treatment including surgical
intervention. Complications may include the following: death, perforation or tamponade,
stroke, myocardial infarction, spasm of the coronary artery, endocarditis, pericarditis,
pericardial/pleural effusion, pleurisy, hemoptysis, hemorrhage, pulmonary edema,
embolus, vessel thrombus, valve damage, infection, catheter entrapment, skin burns,
pseudoaneurysm, cardiac/respiratory arrest, sinoatrial (SA) node damage, atrioventricular
(AV) node damage, phrenic nerve damage, aspiration, atrial septal defect, ruptured
chordae tendineae, femoral artery thrombus, aortic puncture, vasovagal reaction, allergic
reaction, vascular and heart damage, radiation exposure, hypotension and
hematoma/bruising at the insertion site. In addition, abnormal heart rhythms
(bradycardia, tachycardia, or heart block) requiring drugs, and/or electric shock, and/or
implantation of a permanent pacemaker can occur.
Operator’s Manual
In the presence of anticoagulation there may be an increase risk of bleeding from all
causes.
Chest discomfort, particularly during the radiofrequency power delivery, may necessitate
the added risks associated with sedatives or general anesthesia.
3-4
4. Unpacking the Controller
Unpacking
The System’s shipping carton contains all of the components identified below:
$ 1 - Controller
$ 1 - Footswitch
$ 1 - XP APM with Quick Connect Connectors
$ 2 - EPT-1000 XP
$ 1 - Interconnect Cable for XP APM to EGM Equipment
$ 1 - Quick Connect Temperature Instrument Cable
$ 1 - Quick Connect Non-Temperature Instrument Cable
$ 1 - Catheter Pin Connect Adapter Cable
TM
Cardiac Ablation System Operator’s Manuals
To unpack the System, following these steps:
1. Unpack the Controller and accessories carefully and inspect for damage.
2. Notify the carrier immediately if the shipment carton is damaged.
3. Verify that the items listed above are received.
% Call EPT for Service
Number: 408.895.3500).
4. Read Section 6, Operational Sequence in this Manual very carefully and follow the
installation and setup instructions.
Repackaging the Controller
When the Controller requires service or transfer to other location, use the original shipping
carton and packing materials to repack and ship the Controller.
For shipping, disconnect the XP APM and Footswitch. Place all the components into the
locations reserved for these units in the c ar ton.
If there are any discrepancies, notify EPT (Telephone
4-1
5. Controls and Displays
Front Panel
The operator controls for the EPT-1000 XPTM Cardiac Ablation System are located on the
front panel of the Controller. The buttons on the Controller’s front panel allow the user to
control the operation of the Controller. The lights/LEDs on the front panel show the
Controller’s status and/or report error conditions.
This section describes the operator controls on the front panel shown in Figure 5-1.
Figure 5-1: Controller Front Panel
Isolated Patient
Connector
POWER
ISOLATED PATIENT
CONNECTOR
Power
Panel
(W)
POWERTEMP
CONTROL
Control
Panel
Temperature
TEMPERATURE
CALIBRATION
Calibration
Panel
Panel
(°C)
Impedance
IMPEDANCE
COUNTERCLEAR
Counter
Panel
Panel
CARDIAC ABLATION CONTROLLER
TIME(S)
()
Mains
Display
Time
Panel
EPT-1000 XP
RF POWER
CONTROL
MAINS
RF Power
Control Panel
5-1
CONTROL Panel
l
CONTROL Indicator LEDs
The control mode of the Controller is indicated by either of two green LEDs (POWER or
TEMP) on the CONTROL Panel (see Figure 5-2). When the Controller is initially
powered on, one of the green LEDs displays the control mode which depends on the type
of Catheter connected to the Controller.
If a Catheter(T°), the Controller automatically switches to the Temperature Control mode
when initially powered on, as indicated by the flashing green TEMP LED. If it is
desirable to operate the Controller in the Power Control mode while using a Catheter(T°),
pressing the CONTROL Button (see Figure 5-2) changes the control mode to Power
Control, as indicated by the flashing green POWER LED.
If a Catheter(Non-T) or no Catheter is connected, then the Controller automatically
switches to the Power Control mode when initially powered on, as indicated by the
flashing green POWER LED.
Operator’s Manual
"
POWER, TIME, and TEMPERATURE setpoints all reset to default values
Note
whenever a Catheter(T°) is disconnected and reconnected, or when a switch is made
between control modes (Power Control to Temperature Control or vice versa). During RF
power delivery, the control mode cannot be changed.
Figure 5-2: CONTROL Panel Indicator LEDs and Butt on
Control
Panel
CONTROL
Power
LED
POWERTEMP
Temp
LED
Contro
Button
5-2
Controls and Displays
CONTROL Button
The Controller is capable of operating in Power Control or Temperature Control mode.
Press the CONTROL Button to toggle (or change) to a new control mode, as indicated by
a flashing LED.
The Power Control mode is compatible with Catheter(T°) and Catheter(Non-T) and
allows selection of the quantity of RF power that is delivered by the Steerable Ablation
Catheter (see Table 5-1). It also allows selection of an upper temperature limit from 55oC
to 85oC (the default value).
The Temperature Control mode is only compatible with Catheter(T°) and allows selection
of a desired catheter tip temperature. The maximum amount of power to be delivered by
the Catheter, from a value of 0 W to a maximum of either 50 or 100 W depending on
catheter type, must also be selected in the Temperature Control mode (see Table 5-1).
If insufficient power is selected, the desired catheter tip temperature may not be achieved.
Catheters are designed to transmit three distinct quantities of RF power, based on their
electrode size, internal construction and intended use. The maximum amount of power
the Catheters can access is limited by the control mode and catheter type as follows:
Table 5-1: Maximum Amount of Power Output for Catheters
Catheter TypePower Control (W)Temperature Control (W)
Control Modes
Catheter(Non-T)50N/A
Catheter(T°)5050
Special High Power
100100
Catheter(T°)
5-3
POWER Panel
POWER Display
The POWER Display (see Figure 5-3) shows the RF power output in Watts. The available
power output ranges are dependent on the catheter type and the control mode. Ranges are
from a value of 0 Watt to the maximum amount of power output (see Table 5-1).
When the Controller is initially powered on, the POWER Display indicates the default
power setpoint of 0 Watt. After the user sets the power setpoint and presses the RF
POWER CONTROL Button, the display changes from the setpoint value and begins
displaying the actual power output the Catheter delivers to the tissue.
In the Power Control mode, the Controller adjusts its output to maintain the measured
power at the power setpoint, and in the Temperature Control mode, the Controller adjusts
its output to maintain the measured temperature at the temperature setpoint. In the
Temperature Control mode, measured power may be lower than the power setpoint value,
depending on the measured catheter tip temperature and/or tissue impedance. When RF
power is discontinued (DONE operational mode), the last measured value of RF power
delivered remains on the display, flashing at 1-second intervals.
Operator’s Manual
Figure 5-3: POWER Display and Buttons
POWER
Power
Panel
Power
Increase
Button
(W)
Power
Display
Power
Decrease
Button
5-4
Controls and Displays
POWER &
&/'''' Buttons
&&
The POWER &/' (increase/decrease) Buttons select the desired level of delivered RF
power for the Power Control mode, or the maximum amount of RF power available for
the Temperature Control mode.
For instructions on how to select the level of delivered RF power, see the section
Selecting the Power Output on page 6-9.
"
Access to greater than 50 W of power requires the use of appropriately coded
Note
(special high power) catheters and the XP APM.
TEMPERATURE Panel
TEMPERATURE Display
The TEMPERATURE Display (see Figure 5-4) shows the temperature setpoint (desired
catheter tip temperature) in degrees Centigrade. This feature is only available if a
Catheter(T°) is connected to the Controller.
Figure 5-4: TEMPERATURE Display and Buttons
Temperature
Panel
Temperature
Increase
Button
TEMPERATURE
( C)
Temperature
Display
Temperature
Decrease
Button
5-5
Operator’s Manual
Table 5-2 lists all possible temperature display readings under specific operating
conditions. Refer to the instructions specified in the table for additional information.
Table 5-2: Summary of TEMPERATURE Display Readings
instructions, display indicates desired catheter tip
temperature. (Refer to instructions on page
6-10, Selecting Temperature When Using theCatheter(T°) in TEMPERATURE Control Mode.)
• In ON or CALIBRATE operational modes,
display indicates measured catheter tip
temperature.
• In ON operational mode:
− If the temperature is within 20°C – 120° C
range, the measured temperature displays.
− If the measured temperature is below 20°C, the
message “LO” displays.
− If the measured temperature is above 120°C,
the message “HI” displays.
• In DONE operational mode: When RF power is
discontinued, the last measured value of
temperature is shown on the display, flashing at
1-second intervals.
Catheter(T°)Power Control
Mode
• In READY mode, display reading is 85°C.
• In ON or CALIBRATE operational modes,
display indicates measured catheter tip
temperature.
• In ON operational mode:
− If the temperature is within 20°C – 120° C
range, the measured temperature displays.
− If the measured temperature is below 20°C, the
message “LO” displays.
− If the measured temperature is above 120°C,
the message “HI” displays.
• In READY mode, upper temperature limit can be
selected. After setting temperature setpoint per
instructions, display indicates desired catheter tip
temperature. (Refer to instructions on page
6-11, Selecting Temperature When Using theCatheter(T°) in POWER Control Mode.)
When RF power is discontinued, the last
•
measured value of temperature is shown on the
5-6
Controls and Displays
display,
Catheter(Non-T)*N/A
* Temperature Control is not accessible for this Catheter.
In Power Control mode only, when the measured catheter tip temperature
Note
• Display shows three dashes (---)
flashing at 1-second intervals.
decreases to a value that is below the upper temperature limit, the Controller resumes
delivery of RF power at a value that is 50% of the power which caused the upper
temperature limit to be exceeded. If the upper temperature limit is exceeded a second
time, the Controller immediately ceases delivery of RF power and display an error code
indicating that the upper temperature limit has been exceeded a second time. It is
recommended that if this error condition arises, the power level should be decreased to
prevent the upper temperature limit from being exceeded. The Controller can be placed
into the READY mode by depressing the RF POWER CONTROL Button, which allows
adjustment of Controller parameters such as RF power, upper temperature limit, etc.
&
&/'
TEMPERATURE
&&
'
Buttons
''
The TEMPERATURE &/' (increase/decrease) Buttons select the temperature that the
Contro ller attemp ts to achieve during RF power delivery. In Temperature Control mode,
the Controller automatically adjusts power, within the user-selected upper power limit, to
achieve the desired catheter tip temperature. In Power Control mode, the Controller
automatically decreases RF power to the Catheter if this value of user-selected upper
temperature limit is exceeded.
For instructions on how to select the temperature setpoint, see the section Selecting the
Desired Catheter Tip Temperature on page 6-10.
IMPEDANCE Panel
IMPEDANCE Display
The IMPEDANCE Display (see Figure 5-5) shows the measured impedance value in Ω
(1 Ω resolution). When the Controller is initially powered on, the display shows three
dashes “---”, indicating that no measurement is being made. As soon as RF power is
delivered by pressing the RF POWER CONTROL Button or the Footswitch, the
Controller begins to measure impedance. Because this display reflects the actual tissue
impedance, the value shown may fluctuate slightly, depending on the quality of the
electrode/tissue interface, but should remain relatively stable throughout power deliveries.
For a summary of the impedance ranges, see Table 6-2 in section Turning the RF Power
Delivery “ON”
5-7
Operator’s Manual
When RF power is discontinued, the last measured value of impedance remains on the
display, flashing at one second intervals. If the Controller is placed into the
CALIBRATION mode (by depressing the CALIBRATION Button) and if a Catheter has
been positioned within the heart, then the Controller delivers 1 Joule of energy which
allows a single measurement of impedance, and continuously displays the value of
impedance measured. When either the CALIBRATION or RF POWER CONTROL
Buttons are depressed, the Controller returns to the READY mode and the IMPEDANCE
Display returns to three dashes “---”.
Figure 5-5: IMPEDANCE Display
Impedance
Panel
IMPEDANCE
()
Impedance
Display
5-8
Controls and Displays
e
TIME Pa nel
TIME Display
The TIME Display (see Figure 5-6) shows the user-selected duration (in seconds) of RF
power delivery to the Catheter. When the Controller is initially powered on, the TIME
Display reads zero. After the user sets the duration and presses the RF POWER
CONTROL Button, the displayed value changes to zero and initiates counting upward at
1-second intervals until the preset time limit is reached (or the RF POWER CONTROL
Button is actuated a second time or the Footswitch is released), stopping RF power
delivery. If RF power delivery is stopped before the preset maximum duration has been
reached, actual duration of RF power delivery is displayed, flashing at 1-second intervals.
&
&/''''
TIME
&&
The POWER &/' (increase/decrease) Buttons select the maximum durati on of each RF
power delivery. The range of times available is 1 to 120 seconds in increments of one.
Buttons
"
These buttons are inoperative during RF power delivery.
Note
For instructions on how to select the duration of RF power delivery, see the section
Setting the Time Control on page 6-11.
Figure 5-6: TIME Display and Buttons
Time
Panel
Time
Increase
Button
TIME
(S)
Time
Display
Time
Decreas
Button
5-9
COUNTER Panel
COUNTER Display
The COUNTER Display (see Figure 5-7) automatically counts and displays each delivery
of RF power. When the Controller is initially power on, the display clears and reads zero.
The range of values available is 0 to 99 in increments of one.
CLEAR (Counter) Button
The CLEAR Button clears the COUNTER Display (Figure 5-7) to zero and allows the
user to keep an accurate count of the number of RF power deliveries to the patient. Press
this button before starting initial RF power application on a patient.
Figure 5-7: COUNTER Display and Clear Button
Operator’s Manual
Counter
Panel
Counter
Display
RF POWER CONTROL Panel
RF POWER CONTROL Button/Light
The RF POWER CONTROL Button (see Figure 5-8) controls the three operational
modes of the Controller:
• READY mode – This mode allows the ablation parameters to be adjusted.
• ON mode – This mode is the Controllers’ state during the delivery of RF power.
• DONE mode – This mode defines the system state after cessation of RF power
delivery.
"
This button can also be used for exiting the CALIBRATION or Error
Note
modes and returning the Controller to the READY mode.
COUNTERCLEAR
Clear
Button
When the Controller is initially powered on, the RF POWER CONTROL Light flashes,
indicating the Controller is in the READY mode and prepared for RF power delivery.
Press the yellow RF POWER CONTROL Button once to place the Controller into the ON
mode and allow the Catheter to transfer RF power to the tissue. To alert the user, this
button is backlit with a light which is ON during the delivery of RF power (RF Power ON
mode), and OFF when RF power delivery stops (DONE mode).
5-10
Controls and Displays
If the Controller is in the Temperature Control mode, a test of the temperature sensor
functionality is made prior to delivery of RF power. If a faulty temperature sensor is
detected by the Controller, which can be caused by the temperature sensor indicating a
temperature outside the operating range of 31o C to 43o C, or a short or open in the
temperature measurement circuit, the Controller does not enter the RF Power ON mode.
It enters an Error mode, indicating a faulty temperature measuring circuit has been
detected or the temperature is outside the operating range. To exit the Error mode,
depress the RF POWER CONTROL Button, which returns the Controller to the READY
mode.
Figure 5-8: RF POWER CONTROL Panel and Button/Light
RF Power
Control
Panel
CALIBRATION Panel
CALIBRATION Button
Pressing the CALIBRATION Button (see Figure 5-9) places the Controller into the
CALIBRATION mode which performs the following functions:
•Approximately 1 Joule of energy is delivered to the tissue which is used to make a
single measurement of tissue impedance. This value is continuously displayed on the
front panel.
•The catheter tip temperature is continuously monitored and displayed on the front
panel. This allows assessment of the temperature measuring function of the
Catheter(T°). If this value differs significantly from the normal body temperature of
37o C, the Catheter(T°) should be removed and replaced with another Catheter(T°).
RF Power
Control
Button/Light
RF POWER
CONTROL
•A zero to one volt square-wave signal representing a minimum-to-full-scale signal is
transmitted from each of the rear panel analog outputs (for calibration of an optional
recorder). This square-wave calibration signal is transmitted continuously to the rear
panel analog output ports.
5-11
Operator’s Manual
All of the above functions continue to operate as described above until either the
CALIBRATION or RF POWER CONTROL Buttons are pressed again. Pressing either of
these buttons causes the Controller to return to the READY mode. The minimum-to-fullscale values for the individual analog outputs are shown in Table 5-3.
Table 5-3: Individual Analog Outputs
Power0.0 - 1.0 volt corresponds to 0 -100 W
Impedance
Temperature
The CALIBRATION LED illuminates when a calibration signal is transmitted to the
analog output ports. The calibration signal automatically discontinues if the RF POWER
CONTROL Button or the Footswitch is pressed.
Figure 5-9: CALIBRATION Panel and Button/LED
Calibration
Calibration
Isolated Patient Connector
Panel
Button
0.0 - 1.0 volt corresponds to 0 - 300 Ω
0.0 - 1.0 volt corresponds to 20o - 120°C
CALIBRATION
Calibration
LED
The ISOLATED PATIENT CONNECTOR (see Figure 5-1) provides for connection of
the XP APM to the Controller. The connector has a screw-on/locking mechanism for
secure cable installation. The XP APM Patient Cable must be installed in the ISOLATED
PATIENT CONNECTOR before the Controller delivers any RF power. This cable uses a
circular connector that is keyed for proper alignment.
MAINS (Power “ON”) Display
The MAINS Display (see Figure 5-1) indicates that the power swit ch (see Figure 5-10)
located on the rear panel is ON (“1” position) and the Controller is plugged into an
electrical outlet.
5-12
Controls and Displays
Rear Panel
This section describes the output ports and power controls on the rear panel shown in
Figure 5-10.
Figure 5-10: Controller Rear Panel
Footswitch
Cable
Connector
Analog
Outputs
ANALOGOUTPUT
RF POWER
OUTPUT POWER 150W
OPERATING FREQUENCY 500KHz
Ω
IMPEDANCE TEMPERATURE SPAREVOLUME
FOOT SWITCHRS-232
Serial
Port
Volume
Control
Knob
SPEAKER
CAUTION: DO NOT OPEN
ELECTRICAL SHOCK HAZARD
Mains
Power
Switch
CAUTION: THE TOTAL SYSTEM
CHASSIS RISK CURRENT SHOULD NOT
A
EXCEED 100
µµµµ
ATTENTION: LE COURANT DE TISQUE
TOTAL'ADMISSIBLE AU CHASSIS NE
DOIT PAS DEPASSER 100
MAINS
MAIN
FUSES
2X
T3A 250V
T1,6A 250V
WARNING
RISK OF FIRE.
REPLACE FUSE
AS MARKED.
100-120V
3A 50/60 Hz
220-240V
1,6A 50/60 Hz
Ground
Stud
Fuse
Holder
A.
µµµµ
AUX
AUX
2X
FUSES
WARNING
RISK OF FIRE.
REPLACE FUSE
AS MARKED.
100-120V
1A 50/60 Hz
220-240V
500mA 50/60 Hz
T1A 250V
T500mA 250V
Auxiliary
Power
Outlet
Fuse
Holders(2)
Power
Cord
POWER Switch
The power switch (rocker switch) on the rear panel powers the Controller ON (“1”
position) or OFF (“0” position).
Analog Outputs
5-13
Operator’s Manual
There are 4 analog output ports (see Figure 5-10), which are labeled as follows: RF
POWER, IMPEDANCE, TEMPERATURE, and SPARE.
Power Analog Output
RF POWER – This output port provides direct connection to an electrically isolated
general-purpose recorder or monitor, satisfying the requirements of IEC-601, using a
standard BNC connector to facilitate recording of the RF power output during delivery of
RF power.
Impedance Analog Output
IMPEDANCE – This output port provides direct connection to an electrically isolated
general-purpose recorder or monitor, satisfying the requirements of IEC-601, using a
standard BNC connector to facilitate recording of the tissue impedance during delivery of
RF power.
Temperature Analog Output
TEMPERATURE – This output port is active only when a Catheter(T°) is in use. It
provides direct connection to an electrically isolated general-purpose recorder or monitor,
satisfying the requirements of IEC-601, using a standard BNC connector to facilitate
recording of the Catheter(T°) measured catheter tip temperature during delivery of RF
power.
Spare
SPARE – This output port is not in use at this time.
Power Cord
The Power Cord is “Hospital Grade” and must be installed at an AC electrical wall outlet
designated “Hospital Grade” or “Hospital Only.”
MAINS Fuse Holder
Housing for AC fuse. (See Controller rear panel for appropriate type of fuse.)
Fuse Holders (2)
Housing for AC fuses. (See Controller rear panel for appropriate type of fuse.)
Footswitch Cable Connect or
This connector is polarized for connecting the Footswitch Cable.
5-14
Controls and Displays
Volume Control Knob
This knob provides the user with volume control for the audio signal that accompanies
RF power delivery. If no audio tone is heard, check the knob to determine if the volume
level has been adjusted too low.
Serial Port
This port is labeled “RS-232.” It provides for computer interface to EPT approved
accessory computers only.
Auxiliary Power Outlet
This outlet is labeled “Auxiliary Power Outlet.” It provides for future optional EP
Technologies’ approved accessories only.
Ground Stud
Protective Earth Ground stud (chassis ground).
5-15
Accessories
Footswitch
An easy-to-use Footswitch (see Figure 1-1) can alternatively be used to provide ON/OFF
control of the RF power delivery. The connecting cable’s 10-ft length allows the user to
stand at the catheterization table near the patient without requiring another person for
starting/stopping RF power delivery. The Footswitch functions similarly to the RF
POWER CONTROL Button except that the user must continuously hold the Footswitch
down for RF power to be delivered. The RF power delivery is immediately terminated
when the user’s foot is lifted off the Footswitch. Use only the EPT Model 840
Footswitch, which is rated IPX8 (continuous immersion).
High Power Automatic Personality Module (XP APM)
The XP APM provides RF filtering to allow continuous electrogram recording during RF
power delivery via the catheter tip electrode. Additionally, the XP APM allows for safe use
of either standard or special high power EPT ablation catheters. The XP APM passes RF
energy (at 500 kHz) from the Controller to the patient via the catheter and one or two DIP
Electrodes. The maximum power delivered depends on the configuration of the attached
catheter. The XP APM detects the presence of a special hi gh power catheter and enables
the Controller to deliver RF power up to 100 W. If a standard catheter is detected, the
Controller is not able to deliver power greater than 50 W.
Operator’s Manual
The Module connections are as follows (see Figure 5-11):
•One 9-pin Quick Connect Connector labeled “CATHETER” for attaching the Quick
Connect Instrument Cable to the catheter.
•One 6-pin Recorder Connector labeled “RECORDER” for attaching the XP
APM/EGM Interconnect Cable to the recording equipment.
•Two male 2-pin connectors labeled “INDIFFERENT ELECTRODE” for attaching
the two DIP Electrodes.
"
If greater than 1 A of current is delivered to either of the two DIP
Note
Electrodes, the System shuts down.
For additional information on the XP APM, refer to Table 8-1: System Specifications in
Section 8.
5-16
Controls and Displays
Figure 5-11: High Power Automatic Personality Module with Quick Connect
EP Technologies, Inc.
CARDIAC ABLATION SYSTEM
XP AUTOMATIC PERSONALITY MODULE
FOR USE WITH EP TECHNOLOGIES
CARDIAC ABLATION SYSTEM ONLY!
Top View
!
EPT-1000 XP APM
RECORDER
RECORDER
INDIFFERENT
INDIFFERENTELECTRODES
ELECTRODES
X 2
Front Panel
CATHETER
Side Panel
CATHETER
5-17
6. Operational Sequence
Initial Installation
Follow the instructions in this section carefully to prepare the System for operation:
1. Connect the Controller Power Cord plug into a properly grounded AC electrical
outlet designated “Hospital Grade” or “Hospital Only.”
!
Precaution
Never use an outlet without a grounding connection.
2. Position the Controller for easy viewing of its front panel displays.
3. Install the XP APM Patient Cable to the ISOLATED PATIENT CONNECTOR
socket on the Controller front panel by carefully lining up the connector pins with the
socket. Push the connector firmly into the socket; then secure in place by rotating the
locking ring in a clockwise direction.
"
Do not twist the XP APM Patient Cable connector while inserting or
Note
removing it from the Controller ISOLATED PATIENT CONNECTOR socket or
the connector pins may be damaged.
Figure 6-1: Connecting the XP APM to the Controller Front Panel
POWER
ISOLATED PATIENT
CONNECTOR
(W)
CONTROL
TEMPERATURE
CALIBRATION
(°C)
IMPEDANCE
COUNTER CLEAR
TIME
(
)
ΩΩΩΩ
RF POWER
CONTROL
(S)
MAINS
XP APM
Patient
Cable
INDIFFERENT ELECTRODES
X 2
CATHETER
XP APM
6-1
Operator’s Manual
4. To disconnect the XP APM from the Controller, rotate the locking ring counter
clockwise to unlock; then grasp the connector and gently pull it out of the socket.
"
Do not twist the XP APM Patient Cable connector while inserting or
Note
removing it from the Controller ISOLATED PATIENT CONNECTOR socket or
the connector pins may be damaged.
!
Precaution
Never disconnect the XP APM by pulling on the cable.
5. For Optional Patient Monitor Equipment Installation only, follow these instructions:
a) Use the XP APM/EGM Interconnect Cable to connect the XP APM to optional
recording equipment (see Figure 6-2).
b) Plug in the Quick Connect fitting from the XP APM/EGM Interconnect Cable
into the connector located on the XP APM side panel labeled “RECORDER” and
plug the opposite end of this cable into isolated recorder inputs.
"
Ensure that the pins are not bent or broken and that the monitoring
Note
connectors are securely in place. Unused pins should be secured so that they
do not inadvertantly touch other equipment or surfaces. Never insert pins
into mains outlets or into any equipment other than systems providing patient
electrical isolation in accordance with IEC-601.
Figure 6-2: Connecting the XP APM to the Recording Equipment
XP APM
Recorder
INDIFFERENT ELECTRODES
X 2
CATHETER
6-2
Controls and Displays
6. Position the XP APM near the table where the procedure is to be performed. The
System may only be connected to recording systems providing patient electrical
isolation in accordance with IEC-601.
7. Install the Catheter Quick Connect Instrument Cable fitting into the XP APM front
panel fitting labeled “CATHETER.”
Figure 6-3: Connecting the Catheter to t he XP APM Front Panel
INDIFFERENT ELECTRODES
X 2
EP Technologies, Inc.
CATHETER
Quick Connnect
Instrument Cable
Catheter
8. Install a DIP Electrode (or 2 DIP electrodes, if using a special high power catheter)
into either of the two the INDIFFERENT ELECTRODE receptacles located on the
XP APM front panel (see Figure 6-4) as follows:
a) Read the manufacturer’s manual before installing the DIP Electrode pads.
b) Gently push each DIP Electrode fitting straight and firmly into place.
c) To disconnect a DIP Electrode, grasp the DIP Electrode fitting and gently pull it
out from the XP APM Indifferent Electrode receptacle.
!
Precaution
Do not disconnect a DIP Electrode Connector by pulling on its
cable.
d) Before use, it is important to check the DIP Electrode(s) for damage to the sealed
foil package. Exposure to air, due to a damaged package, could cause the DIP
Electrode(s) to become dry and limit their grounding capability.
e) Be sure the pads are moist and sticky to the touch before placing on the patient.
!
Precaution
Do not attempt to relocate the patient grounding pad after initial
application. Electrode gel is NOT required and should NOT be used.
6-3
Operator’s Manual
)
f) The DIP Electrode(s) should be placed on a well-vascularized, convex skin
surface that is in close proximity to the ablation site (left upper quadrant of the
back is suggested unless the patient’s scapula is especially prominent or patient is
extremely thin). Other possible locations are the upper arm or left flank area.
Avoid scar tissue, bony prominence, adipose tissue, or any areas where fluid may
pool. Shave, clean, and dry the application site as needed. Check for wrinkles or
folds when applying the pad as these decrease conductivity.
"
RF output ceases to the patient if more than 1 A flows through either of the
Note
DIP Electrodes.
Figure 6-4: Connecting the Disposable Indi fferent Patch Electrodes
CATHETER
Polarity
Unimportant
INDIFFERENT ELECTRODES
X 2
DIP Electrodes (2
6-4
Controls and Displays
9. If using the Footswitch, install its cable connector into the FOOTSWITCH Cable
Connector on the Controller rear panel (see Figure 6-5). Carefully push the cable
connector into the FOOTSWITCH connector until it is seated firmly in place.
"
Do not twist or bend the FOOTSWITCH connector during insertion as this
Note
action might cause damage to the pins.
10. Position the Footswitch to allow easy access by the physician.
Figure 6-5: Connecting the Footswitch to the Controller Rear Panel
CAUTION: THE TOTAL SYSTEM
CHASSIS RISK CURRENT SHOULD NOT
EXCEED 100
A
µµµµ
ATTENTION: LE COURANT DE TISQUE
TOTAL'ADMISSIBLE AU CHASSIS NE
DOIT PAS DEPASSER 100
MAIN
FUSES
2X
T1,6A 250V
WARNING
RISK OF FIRE.
REPLACE FUSE
AS MARKED.
100-120V
3A 50/60 Hz
220-240V
1,6A 50/60 Hz
A.
µµµµ
MAINS
T3A 250V
Footswitch
Cable
Connector
ANALOGOUTPUTSPEAKER
Ω
RF POWER
IMPEDANCE TEMPERATURE SPARE
FOOT SWITCH
OUTPUT POWER 150W
OPERATING FREQUENCY 500KHz
RS-232
CAUTION: DO NOT OPEN
ELECTRICAL SHOCK HAZARD
VOLUME
AUX
AUX
FUSES
WARNING
RISK OF FIRE.
REPLACE FUSE
AS MARKED.
100-120V
1A 50/60 Hz
220-240V
500mA 5 0 / 60 Hz
2X
T500mA 250V
T1A 250V
11. The System is now ready to record electrograms prior to and during delivery of RF
power.
Turning the Controller “ON”
1. Turn the Controller ON by pressing the power switch (to the “1” position) located in
the power input module on the Controller rear panel (see Figure 6-6).
The Controller automatically initiates a self-test procedure which is indicated by the
illuminated front panel displays and continuous audio tone generated for
approximately 2 seconds. If no system malfunction is detected, the Controller
changes to the READY Mode.
"
The connection between the XP APM and Controller is not required to
Note
complete the initial self-test.
Footswitch
6-5
Operator’s Manual
2. If a system malfunction is detected during self-test, all front panel displays remains
lit and the Controller does not operate. To clear any malfunctions found during selftest, the Controller must be powered OFF, then back ON. The self-test repeats.
%
Call EPT for Service
A second self-test failure is indicative of a Controller
malfunction and should be referred to EPT for service (Telephone Number:
408.895.3500). The Controller does NOT operate unless the initial self-test has
been successfully completed.
Figure 6-6: Turning the Controller to the ON Positi on
MAINS
Press Power Switch to "1" position.
Power Switch
Power Input
Module
READY Mode
The Ready mode is designed for setting the desired parameters for RF power delivery. It
is automatically initiated after completion of a successful Controller self-test. Ready
mode is indicated by a flashing RF POWER CONTROL Button on the Controller front
panel; TEMPERATURE, IMPEDANCE, TIME, and COUNTER Displays are
continuously lit. Values displayed when the Controller is first initialized are zero in the
POWER, TIME, and COUNTER Displays and dashes (---) in the IMPEDANCE Display.
(Refer to Figure 5-1 for the location of displays and buttons.)
Controller in Temperature Control Mode - Catheter(T°) Connected
If a Catheter(T°) is connected, the TEMPERATURE Display shows one of two values
depending on the control mode. When the Controller is initially powered on and a
Catheter(T°) is connected, the TEMPERATURE Display indicates 30°C, since the
Controller is automatically entered into the Temperature Control mode. The user must
enter the desired catheter tip temperature, the maximum power that the Controller can
deliver to achieve the desired temperature, and the ablation time before RF power
delivery.
6-6
Controls and Displays
r
r
Controller in Power Control Mode - Cathet er(T°) Connected
If the control mode is manually changed to Power Control, by pressing the RF POWER
CONTROL Button, the TEMPERATURE Display indicates 85o C during the READY
mode of the Controller, indicating the upper temperature limit at which the Controller
continues to deliver set power.
Controller in ON or CALIBRATE Operational Modes - Cat heter(T°) Connected
When the Controller is in either the ON or CALIBRATE operational modes, the
TEMPERATURE Display indicates “LO” if the measured temperature is less than 20°C,
“HI” if temperature is greater than 120°C; otherwise, it shows the measured temperature
(see Figure 6-7).
If a Catheter(Non-T) or no Catheter is connected, the display shows “---” (Figure 6-8).
Figure 6-8: Ready Mode, Initializati on - Cat heter(Non-T) Connected
TEMPERATURE
( C)
Temperatu
Display
6-7
Calibrating the Controller Analog Outputs for the Recorder
Each time the Controller is powered on, the analog outputs (see Figure 5-10) should be
placed in the CALIBRATION mode to verify that the optional strip chart recorder and/or
monitor are calibrated correctly, if used.
To perform calibration of the Controller analog outputs, follow these steps:
1. Check that all connections to the analog outputs are secure.
2. Ensure that the recorder/monitor is powered on and functioning properly.
3. Ensure the recorder satisfies the requirements of IEC-601 prior to connection to
Controller analog outputs.
4. To calibrate with the analog outputs, press the CALIBRATION Button (Figure 6-9)
at the Controller front panel. This generates a minimum-to-full-scale (0.0 to 1.0 volt)
continuous square-wave calibration signal which is sent continuously and
simultaneously to each of the analog outputs (until the CALIBRATION Button is
pressed again or until the RF POWER CONTROL Button is pressed).
Operator’s Manual
Figure 6-9: Calibrating the Controll er Analog Outputs
Calibration
Button
CALIBRATION
COUNTER CLEAR
RF POWER
CONTROL
MAINS
5. With the C ontroller in the READY mode, analog output calibration can be repeated
or checked at any time by repeating the steps identified in steps 1 through 3 above.
6. When RF power is delivered, the analog signals corresponds to the Controller’s
measured power, temperature, and impedance values.
Selecting the Control Mode
The Controller is capable of operating in Power Control or Temperature Control mode.
Control mode selection is made via the CONTROL Button. (For complete description of
the CONTROL Button, see CONTROL Panel in Section 1.)
6-8
Controls and Displays
To select the proper control mode, follow these steps:
1. Ensure that the Controller is in the READY mode.
"
Note
The control mode can only be changed when the Controller is in the
READY mode.
2. Press the CONTROL Button to select the appropriate control mode. The CONTROL
Indicator LEDs display the selected mode.
Selecting the Power Output
To deliver RF power, it is necessary to select a Controller power output. This is true for
both Temperature Control mode and Power Control mode.
" Note Use Lower Power First - When first delivering RF energy, begin by using a low
power setting (i.e., 50W). If the created lesion is unsuccessful or inadequate,
incrementally increase the power output with successive ablation attempts to minimize
the potential for thrombus formation and/or inadvertent damage to cardiac tissues.
Controller in Temperature Control Mode - Catheter(T°) Connected
The Controller is automatically entered into the Temperature Control mode and the RF
power setpoint is the maximum amount of power that the Controller can access to achieve
the desired catheter tip temperature. If the RF power setpoint is equal to or greater than the
amount of power necessary to achieve the desired catheter tip temperature, then the
Controller supplies the appropriate amount of power to achieve the temperature setpoint.
The Controller only supplies as much RF power as necessary to achieve the desired catheter
tip temperature.
If the RF power setpoint is less than the amount of power required to achieve the desired
catheter tip temperature, then the Controller may not be capable of achieving the desired
catheter tip temperature and thus the measured temp erature is less than the desired cathet er
tip temperature.
Controller in Power Control Mode - Cathet er (Non-T) or Catheter(T°) Connected
If a Catheter(Non-T) or Catheter(T°) is connected, the Controller is manually placed into
the Power Control mode, then the RF power setpoint corresponds to the actual amount of
power that is delivered to the tissue, unless the upper temperature limit is exceeded, in
which case, the Controller reduces and/or ceases delivery of set power.
Controller in Power Control Mode - Cathet er(T°) Connected
If using a Catheter(T°) with the Controller in the Power Control mode, the
TEMPERATURE Display shows 85oC when the Controller is in the READY mode and
displays measured temperature when the Controller is in the ON, CALIBRATE, and DONE
operational modes.
6-9
Operator’s Manual
Selecting the RF Power Setpoint
To set the RF power setpoint (in W), follow these steps:
1. Press the appropriate POWER (&/') Butt on to increase or decrease the RF power
setpoint by 1 W.
2. To scroll rapidly to the desired power setpoint, depress and hold the appropriate
POWER (&/') Button down.
3. Release the POWER (&/') Button when the POWER Display shows the
appropriate RF power setpoint.
4. During RF power delivery (RF Power ON mode), the power setpoint can be adjusted
in 1-W increments to provide better control during the procedure by pressing the
appropriate POWER (&/') Button. When the button is released, the POWER
Display returns to showing the current value of measured power.
•Holding the POWER & Button down while RF power is being delivered (RF
Power ON mode) does NOT cause the power setpoint to scroll as it does when
the Controller is in the READY mode.
•Holding the POWER ' Button down while RF power is being delivered (RF
Power ON mode) causes the power setpoint to scroll at a greater rate than it does
when the Controller is in the READY mode.
Selecting the Desired Catheter Tip Temperature
For ablation procedures involving automatic temperature control, it is recommended that
the operator initially select a lower set temperature range (from 55°C to 70°C). The
operator should consider a lower initial set temperature to avoid excessive thermal
damage to tissue, or collateral damage to adjacent tissue not intended for ablation,
particularly in areas where high blood flow and correspondingly high convective cooling
may be present.
The displayed catheter tip temperature may be significantly lower than the
Note:
maximum tissue temperature, which typically occurs at a depth of a few millimeters. The
displayed catheter tip temperature is measured at the electrode-tissue interface and may
be influenced by the degree of tissue contact and the convective cooling effects of blood
flow. Do not set the maximum temperature higher than 80ºC.
Selecting Temperature When Using the Catheter(T°) in TEMPERATURE Control Mode
If a Catheter(T°) is connected to the Controller, a desired value of catheter tip temperature
(temperature setpoint) can be selected. The Controller automatically adjusts power,
within a user-selected upper power limit, to achieve the desired catheter tip temperature,
in the Temperature Control mode.
To select the temperature setpoint for the Controller, follow these steps:
1. Ensure that the Controller is in READY mode.
6-10
Controls and Displays
2. Press the appropriate TEMPERATURE (&/') Button to increase or decrease the
temperature setpoint by 1°C.
3. To scroll rapidly to the desired temperature setpoint, depress and hold the appropriate
TEMPERATURE (&/') Button down.
4. Release the TEMPERATURE (&/') Button when the TEMPERATURE Display
shows the appropriate temperature setpoint.
The catheter tip temperature selection range is 30° to 80°C in increments of 1°C,
although the Controller cannot achieve a temperature which is less than the ambient
blood temperature (normally 37°C).
5. During RF power delivery (RF Power ON mode), the temperature setpoint can be
increased or decreased by pressing the appropriate TEMPERATURE (&/') Button.
When the TEMPERATURE (&/') Button is depressed, the display shows the new
temperature setpoint. When the button is released, the display shows the present
value of catheter tip temperature. Depressing and holding the TEMPERATURE
(&/') Button down while RF power is being delivered does NOT cause the
temperature setpoint to scroll up, but allows it to scroll down at a slightly greater rate
than it does when the Controller is in the READY mode.
Selecting Temperature When Using the Catheter(T°) in POWER Control Mode
If a Catheter(T°) is connected to the Controller, an upper temperature limit can be
selected. The Controller automatically decreases RF power to the Catheter if this value of
temperature is exceeded in the Power Control mode.
To select the temperature setpoint for the Controller, follow these steps:
1. Ensure that the Controller is in READY mode.
2. Press the appropriate TEMPERATURE (&/') Button to increase or decrease the
upper temperature limit by 1°C during RF power delivery.
3. To scroll rapidly to the desired temperature setpoint, depress and hold the appropriate
TEMPERATURE (&/') Button down. (The temperature value shown in the
TEMPERATURE Display is the maximum catheter tip temperature at which the
Controller attempts to deliver RF power for.)
4. Release the TEMPERATURE (&/') Button when the TEMPERATURE Display
shows the appropriate temperature setpoint.
The catheter tip temperature limit selection range is 55° to 85°C in increments of
1°C.
Setting the Time Control
The TIME Display, TIME & (increase) Button, and TIME ' (decrease) Button allow the
user to select the duration (in seconds) that RF power is delivered to the Catheter.
6-11
Operator’s Manual
To select the duration for RF power delivery, follow these steps:
1. Ensure that the Controller is in READY mode.
2. Press the appropriate TIME (&/') Button to increase or decrease the RF power
output duration (up to a maximum of 120 seconds).
3. To scroll rapidly to the desired duration, depress and hold the appropriate TIME
(&/') Button down.
4. Release the TIME (&/') Button when the TIME Display shows the appropriate
duration. (When the selected value is programmed into the Controller control
circuit.)
"
The value can be changed at any time during the READY mode by
Note
depressing the TIME & Button (increase) or TIME ' Button (decrease).
Clearing the Counter
"
Important
1. To clear the counter to zero, depress the CLEAR Button.
Always clear the Counter between patients.
6-12
Controls and Displays
f
Adjusting the Audio Control
The user may adjust the volume of the audio signal when the Controller is initially powered
on or during a clinical procedure.
1. Adjust the volume of the audio signal by rotating the Volume Control Knob on the
Controller rear panel (see Figure 6-10).
2. If desired during a clinical procedure, the Volume Control Knob may be rotated fully
counter-clockwise to its lowest setting
"
At the minimum setting, the tone may NOT be audible during RF power
Note
delivery.
Figure 6-10: Adjusting Audio Control
ANALOGOUTPUT
RF POWER
IMPEDANCE TEMPERATURE SPAREVOLUME
FOOT SWITCHRS-232
SPEAKER
Volume
Control
Knob
VOLUME
VOLUME
To increase the volume of
the audio signal, turn the
Volume Control Knob
clockwise.
To decrease the volume o
the audio signal, turn the
Volume Control Knob
counter-clockwise.
6-13
Turning the RF Power Delivery “ON”
Operator’s Manual
"
Important
Do not continue with this procedure unless all preceding procedures in this
chapter have been completed.
To deliver RF power to the Catheter, follow these steps:
1. Ensure that all connections are secure and correct.
2. Verify that all aforementioned selections have been made.
3. To begin RF power delivery, press the RF POWER CONTROL Button once or hold
the Footswitch down. The POWER Display shows the RF power delivered to the
Catheter (in W).
The Audible tone sounds during RF power delivery.
"
If the Controller is in the Temperature Control mode and the measured
Note
temperature is outside the range of 31oC to 43oC, the Controller does not deliver
RF power and an E07 error code message displays.
When RF power is delivered to the Catheter, the displays function as follows:
RF POWER CONTROL BUTTON/Light
This light illuminates and remains lit until RF power delivery is discontinued. (For
instructions on how to discontinue RF power delivery, see the section, Turning the RFPower Delivery “OFF” on page 6-19.
POWER Display
This display shows (in W) RF power delivered to the Catheter. The value may fluctuate
slightly due to changes in tissue impedance.
6-14
Controls and Displays
TEMPERATURE Display
The readings on this display depends on the current control mode of the Controller
(Power Control and Temperature Control), the operational mode, and the type of catheter
connected. Table 6-1 summarizes the readings on the display under specified operating
conditions.
modes, display indicates measured catheter tip
temperature.
• If the measured temperature is outside the range
of 31°C to 43°C, the Controller does not deliver
RF power.
• In ON operational mode:
− If the temperature is within 20°C – 120° C
range, the measured temperature displays.
− If the measured temperature is below 20°C, the
message “LO” displays.
− If the measured temperature is above
120°C, the message “HI” displays.
• In READY mode, display reading is 85°C.
• In ON, CALIBRATE, or DONE operational
modes, display indicates measured catheter tip
temperature.
Catheter(Non-T)N/A
• In ON operational mode:
− If the temperature is within 20°C – 120° C
range, the measured temperature displays.
− If the measured temperature is below 20°C, the
message “LO” displays.
− If the measured temperature is above 120°C,
the message “HI” displays.
• Display shows three dashes (---)
6-15
Operator’s Manual
IMPEDANCE Display
This display shows measured impedance. The displayed readings may fluctuate slightly
due to variation in the stability of electrode/tissue contact as the heart beats. If the
measured value demonstrates a significant fluctuation, the Catheter should be
repositioned. Alternatively, slight pressure on the Catheter shaft may minimize Catheter
motion. Table 6-2 summarizes the readings on the display by impedance range.
Table 6-2: Impedance Display Readings
Impedance
Values
Typical Impedance
80 to 150
Low Impedance
Ω
1
< 50 Ω
< 25 Ω
High Impedance
2
> 300 Ω
Operating ConditionController ActivityDisplay
Reading
Normal operating
condition
Normal RF power
delivery
Display shows
measured
impedance
If accessing up to 50 W
of RF power
If accessing up to 100 W
of RF power
N/AController automatically
Controller automatically
shuts off RF power
Controller automatically
shuts off RF power
shuts off RF power
Display flashes
“LO”
Display flashes
“LO”
Display flashes
between
measured
impedance and
“HI”
1. Low impedance limit depends on type of catheter used in the ablation procedure.
2. High Impedance indicates abnormal operation. This condition also may occur if > 1 A is detected by the XP
APM from either DIP Electrode. Refer to the following procedure, Correcting a High Impedance Condition.
6-16
Controls and Displays
Correcting a High Impedance Condition
If a high impedance value is detected during RF power delivery, follow these steps:
1. Ensure that the connection between the Catheter and XP APM is correct and secure.
2. Ensure that the connect ion between each DIP Electrode and XP APM is correct and
secure.
3. Ensure that each DIP Electrode is properly applied.
4. If an improper cable connection is not the cause of the high impedance value, then a
build-up of coagulum on the Catheter tip may be the cause, and RF power output
should be discontinued.
5. To discontinue RF power delivery, refer to the instructions in section, Turning the
RF Power Delivery “OFF” on page 6-19.
TIME Display
This display shows the user-selected value and elapsed time as follows:
•When the RF POWER CONTROL Button is pressed or the Footswitch is depressed,
this display changes from the user-selected time limit to 0 seconds and begins to
count up to the user-selected value.
•When the preset duration has elapsed, the RF power automatically shuts “OFF.” The
user-selected value remains flashing on the TIME Display.
•If the RF POWER CONTROL Button is pressed during delivery of RF power or the
Footswitch is released, the RF power output immediately turns “OFF.” The actual
elapsed time in seconds remains flashing on the TIME Display.
COUNTER Display
This display increases by 1 count with each RF power delivery. This value resets to zero
at any time when the Controller is in the READY mode.
CONTROL Display
This display indicates the Controller’s current control mode (Power Control or
Temperature Control). The green LEDs indicate when the Controller is either in the
Power Control or Temperature Control modes, respectively.
6-17
Directions for Use
The Steerable Ablation Catheter is usually inserted into a vein or artery and is then
positioned into the appropriate chamber of the heart under fluoroscopic guidance. A
transseptal approach may be used. Precise placement of the catheter prior to ablat ion is
accompanied by endocardial mapping using the tip and/or ring electrodes. Once
appropriate positioning has been achieved, RF power is delivered via the EPT-1000 XP
Cardiac Ablation Controller resulting in the ablation of the targeted cardiac tissue.
Temperature can be monitored during the ablation procedure through the use of an EPT
temperature sensing catheter.
1. Pri or to insertion of the St eerable Ablation Catheter, prepare the ent ry site according
to standard aseptic technique practices.
2. Insert the catheter percutaneously into the appropriate artery or vein by the Seldinger
technique using an 8 French hemostatic introducer sheath.
3. Once inside the vessel, the catheter tip can be deflected as necessary to facilitate
advancement into the selected heart chamber. The degree of tip deflection is
controlled by the Steering Lever on the catheter handle. If the Steering Lever is
pushed forward from its neutral position, the tip curves proportionately up to a
maximum of 270 degrees in one direction depending upon the curve option selected.
Pulling the Steering Lever back causes the tip to deflect in the opposite direction. To
prevent over stressing the tip, the Steering Lever movement is limited by the handle
design.
Operator’s Manual
TM
4. When crossing the aortic valve with the ablation catheter, it is recommended that the
catheter tip be deflected to resemble a “pigtail” curve to avoid damage to the valve
leaflets.
5. The catheter curve can be straightened completely and deflected in the opposite
direction against cardiac tissue, facilitating stability during ablation.
6. The adjustable Tension Control Lever may be tightened to retain the tip in the
desired curvature or to increase steering resistance. The catheter is shipped with the
Tension Control Lever in the “(-)” position, which is the minimum tension
adjustment. In this position, the catheter steers freely and does not hold a preset
curve. Rotation of the Tension Control Wheel in the clockwise direction increase
tension. In the “(+)” position, maximum tension is achieved. The catheter should not
be steered in the maximum “(+)” position.
7. Connect the XP APM to the ISOLATED PATIENT CONNECTOR located on the
Controller’s front panel using the attached patient cable. Be sure to carefully follow
the instructions in the Operator’s Manual to connect the XP APM.
6-18
Controls and Displays
8. Connect the Steerable Ablation Catheter to the XP APM by plugging the Quick
Connect Instrument Cable.
9. When the ablation site has been accessed and the tip of the catheter is in contact
against the endocardial surface, intracardiac electrogram signals may be obtained.
Unipolar electrograms can be obtained between the distal tip electrode and any
commercially available low impedance ECG electrode. Bipolar electrogram
recordings can be recorded between the distal tip electrode and any ring electrode or
between any two ring electrodes.
10. When the arrhythmogenic site has been located, the same catheter can be used
therapeutically in the “Ablate” mode to deliver RF energy. RF power is delivered to
the tissue via the distal tip (ablation) electrode which results in thermal necrosis
(ablation) of the arrhythmogenic tissue.
11. The Thermistor Steerable Ablation Catheter differs from the Steerable Ablation
Catheter in that the distal tip electrode incorporates a th ermistor temp erature sensing
component and has associated electrical leads designed to monitor and/or control
temperature at the endocardial surface during the ablation procedure.
Turning the RF Power Delivery “OFF”
RF power delivery ceases by user intervention or certain operating conditions listing in
the next section, Possible Causes of RF Power Delivery Interruption.
To discontinue RF power delivery during operation (RF Power Done mode), the user may
either:
• Press the RF POWER CONTROL Button or
• Release the Footswitch.
Possible Causes of RF Power Delivery Interruption
RF power delivery can cease due to other operating conditions other than user
intervention. RF power ceases when:
• User-selected time limit has been reached.
• XP APM detects greater than 1 A in either of the two DIP Electrodes.
• Impedance is less than 50 Ω when accessing up to 50 W of RF power.
• Impedance is less than 25 Ω when accessing up to 100 W of RF power.
• Impedance is greater than 300 Ω.
• Temperature exceeds setpoint by 5o C for greater than four seconds (Controller in
Temperature Control mode).
•Temperature exceeds 85o C for greater than one second (Controller in Power Control
or Temperature Control mode with a Catheter(T°) connected).
6-19
Operator’s Manual
• Temperature is less than 20o C for greater than one second.
• Temperature exceeds temperature setpoint in Power Control mode (e.g., If
temperature is set at 60o C and measured temperature exceeds 60o C, RF power shuts
off).
•Upper temperature limit is exceeded twice during a single RF power delivery in
Power Control mode.
•System operation errors are detected.
Display Readings During Interruption of RF Pow er Del ivery
When RF power delivery ceases for any reason other than an out-of-range impedance
value or an error, the last value appearing on each display remains flashing and the RF
POWER CONTROL Light turns “OFF.”
If RF power delivery ceases due to an out-of-range impedance value, the l ast measured
impedance value remains flashing on the IMPEDANCE Display, alternating with the
appropriate “HI” or “LO” message.
If RF power delivery ceases due to an error, the last measured impedance value remains
flashing on the IMPEDANCE Display, alternating with the appropriate error code. (See
Table 8-3 for a listing of error codes.)
In any of the above cases, when either the RF POWER CONTROL Button or the
Footswitch is depressed again, the flashing display values change immediately to the
previous user-selected values and the RF POWER CONTROL Light begins flashing.
This signals that the Controller is ready for additional RF power deliveries. This is true
unless RF power delivery, while in the Power Control mode, is interrupted for exceeding
the upper temperature limit a second time. In this case, the set power is 50% of the value
of RF power which causes the upper temperature limit to be exceeded.
6-20
7. Service and Maintenance
The Controller requires no routine service or maintenance. If the Controller fails to
operate when plugged into a proper AC power receptacle and the power switch is t urned
on, check the fuse (refer to Replacing the Fuses in this chapter).
The Controller contains no user-serviceable parts; disassembly and attempted repair by
unqualified personnel may create a hazardous condition and voids the warranty.
% Call EPT for Service
Number: 408.895.3500).
!
Precaution
in personal injury and/or damage to the Controller.
Cleaning/Disinfecting
The outer surfaces of the Controller and its accessories may be cleaned with a mild soapy
solution. If disinfecting is required, isopropyl alcohol may be used to clean the outer
surfaces.
!
Precaution
or abrasive cleaners.
Replacing the Fuses
Replacing the Main Fuse
1. Before replaci ng a fuse in the Controller, disconnect the mains power cord from t he
Controller.
Do not remove the cover of the Controller. Removing the cover may result
Do not immerse the Controller or its accessories in any liquid. Avoid caustic
If a second failure occurs, notify EPT for service (Telephone
2. Repl ace the fuse with another of the same type and rating. Refer to the fuse label on
the rear panel (see Figure 5-10).
3. Pull the fuse holder out of the power entry module.
4. Use a slotted screwdriver to assist in removing the fuse holder.
5. Insert the new fuse in the fuse holder and reinsert it in the power input module.
"
When replacing the fuse holder, ensure the fuse holder is inserted in the
Note
correct orientation for the operational voltage level.
7-1
Operator’s Manual
Auxiliary Power Outlet Fuses
1. Before replaci ng a fuse in the Controller, disconnect the mains power cord from t he
Controller.
2. Repl ace the fuse with another of the same type and rating. Refer to the fuse label on
the rear panel (see Figure 5-10).
3. Use a slotted screwdriver to rotate the fuse holder counter-clockwise 1/8th of a turn.
The fuse holder “pops” out approximately 6 mm (1/4-inch).
4. Remove the fuse holder and the fuse.
5. Insert the new fuse in the fuse holder and reinsert it into the housing.
% Call EPT for Service
please contact EPT for instructions on returning the Controller to EP
Technologies for repair (Telephone Number: 408.895.3500).
If there appears to be a problem with the Controller,
7-2
8. Product Specifications
General Specifications
Table 8-1 lists the specifications for the System (Controller and XP APM).
Table 8-1: System Specifications
DescriptionSpecification
Power Specifications
Line Power100 - 120 VAC (Model #800XP)
Current Rating
Fuse Rating3 A, 50/60 Hz (100 - 120 VAC)
Auxiliary Power100 - 120 VAC (Model #800XP)
Current Rating
Fuse Rating1 A, 50/60 Hz (100 - 120 VAC)
Power Length Cord10 ft
Patient Cable to High Power
Automatic Personality Module
Length
Connector
High Power Automatic Personality Module
Dimensions
Height
Width
Depth
Weight
10 ft
“Multi-pin” Polarized Connector
1.8”(4.6 cm)
5.9” (15.0 cm)
3.1”(7.9 cm)
2.2 lb (1.0 kg)
8-1
Operator’s Manual
Table 8-1: System Specifications, Continued
DescriptionSpecification
High Power Automatic Personality Module, Continued
Recorder ConnectorsQuick Connect Connector
Catheter ConnectorsQuick Connect Connector
Indifferent Electrode ConnectorsStandard male 2-pin for commercial pads
Recorder Filters
Low Pass FiltersReferenced to the INDIFFERENT ELECTRODE
Low Frequency Cutoff
7 kHz, -3dB at 7 kHz ± 2 kHz
High Power Disconnect MechanismRF output ceases to the patient if more than
1 A flows in either INDIFFERENT
ELECTRODE.
Controller
RF Power Outputs
• 50 W maximum into a minimum impedance of
50 Ω, within a 1 A current limit
• 100 W maximum into a minimum impedance
of 25 Ω, within a 2 A current limit
Maximum RF power output based on catheter
type used and control mode.
Impedance
• Measures 50 Ω and above, if accessing up to
50 W of RF power
• Measures 25 Ω and above, if accessing up to
100 W of RF power
• Measures up to 999 Ω (Controller
automatically shuts off RF power at 300 Ω)
• Displays “LO” or “HI” outside the 25 Ω to
300 Ω range
8-2
Controls and Displays
Controller, Continued
Temperature
Table 8-1: System Specifications, Continued
DescriptionSpecification
With Catheter(T°) Connected and
Controller in Temperature Control Mode
With Catheter(T°) Connected and
Controller in Power Control Mode
With Catheter(Non-T) ConnectedController displays “---”
Normal Operating Condition:
With Controller in the Temperature
Control or Power Control Modes
Outside of Normal Operating RangeController displays “LO” or “HI”
Dimensions
Height
Width
Depth
Weight
Allows selection of desired catheter tip
temperature (temperature setpoint) within the
range of 30° to 80°C.
Allows selection of upper temperature limit or
maximum temperature at which the Controller
continues to deliver the setpower.
Controller measures 20° to 120°C
8” (20 cm)
13”(33 cm)
21”(53 cm)
23 lb.(10.4 kg)
Time0 to 120 seconds in increments of 1
second.
Counter0 to 99 RF power deliveries
Low-Frequency Leakage (50/60 Hz);
Source Current, Patient Leads, All Outputs
Tied Together
Normal Polarity, Intact Chassis
Ground
Normal Polarity, Ground Open
Reverse Polarity, Ground Open
Sink Current, at 120 V Applied, All
Inputs
Chassis Source Current, Ground Open
<10 µA
<10 µA
<10 µA
<10 µA
<100 µA
8-3
Operator’s Manual
Table 8-1: System Specifications, Continued
DescriptionSpecification
Back Panel Analog OutputsStandard BNC connections to output
signals representing RF power,
impedance, and tip temperature.
Calibration Mode
Power Output0 - 1 volt corresponds to 0 - 100 W,
continuous square wave
Impedance
Temperature
0 - 1 volt corresponds to 0 - 300 Ω,
continuous
square wave
0 - 1 volt corresponds to 20 - 120°C,
continuous square wave
8-4
Controls and Displays
Environmental Specifications
Table 8-2 lists the environmental specifications for the System.
Table 8-2: Environmental Specifications
DescriptionSpecification
Storage
Temperature
-40°C to 70°C
The unit should be gradually returned to the
operating temperature range before use and
stabilized for one hour before operation.
Relative Humidity10% to 100%, non-condensing
Atmospheric pressure500 to 1060 millibar
Operating
Temperature
10°C to 40°C
Relative humidity30% to 75%, non condensing
Atmospheric pressure700 to 1060 millibar
8-5
Power Delivery
(
)
As an enhanced patient safety feature, the output voltage and current of the System has
been limited. Under normal circumstances (normal patient impedances), the maximum
output power of the System is either 50 or 100 W, depending on catheter type and control
mode. In the event of unusually high or low tissue impedance, the System limits the
maximum power. A graph which depicts the maximum output power of the System as a
function of tissue impedance is shown in Figure 8-1.
Operator’s Manual
"
Only the maximum power is affected. If the power setting is less than the
Note
maximum power available, the System delivers the set power value.
Table 8-3 defines the error codes for the EP Technologies Cardiac Ablation System.
CodeDescription
E01Apparent power (Vrms * Irms) not greater or equal to real power
E02Apparent power (Vrms * Irms) measured greater than the setpoint
Table 8-3: Error Codes
E03
E04
E05
E06System Operation Error
E07An attempt to enter RF Power On mode with a measured temperature
E08Upper temperature limit (cutoff) exceeded for more than 4.0 seconds.
E09Upper temperature limit (cutoff) exceeded for the second time.
E01-E03 Error Codes
If error codes E01-E03 are observed, it is recommended that the ablation be attempted
again.
% Call EPT for Service
(Telephone Number: 408.895.3500).
Limit exceeded: V 〉 Vmax or I 〉 Imax or P 〉 Pmax
Temperature is greater than setpoint by at least 5°C for greater than
four seconds
Temperature greater than 85°C or less than 20°C for greater than
second
outside the range of 31oC to 43oC.
If the error code displays again, notify EPT for service
E06 Error Code
If an E06 error code message occurs, turn the Controller “OFF.” Wait approximately five
seconds before turning the Controller “ON.”
% Call EPT for Service
procedure, or if the E06 error code message occurs again following successful
completion of the self-test procedure, notify EPT for service (Telephone Number:
408.895.3500).
If the Controller does not successfully pass the self-test
8-7
Figure 8-2: Labeling Symbols
Operator’s Manual
Power ON Power OFF
IncreaseDecreaseProtective
VariabilityClock:
Non-Ionizing
Radiation
Time Switch:
Timer
LoudspeakerFuseCounter
Temperature at
Distal Electrode
Degree C
Earth Ground
Dangerous
Voltage
Alternating
Current
Defibrillator-Proof
Type Equipment
Attention, Consult
Accompanying
Documents
000
M
CalibrationClearIndifferent
Explosion
Hazard
(Dispersive)
Patch Electrode
Equipotentiality
Radio Frequency
Power (On/Off)
AP
8-8
9. Limited Warranty and Disclaimer
Limited Warranties
Boston Scientific Corporation (BSC) warrants that reasonable care has been used in the design
and manufacture of this instrument. This warranty is in lieu of and excludes all other warranties
not expressly set forth herein, whether expressed or implied by operation of law or otherwise,
including, but not limited to, any implied warranties of merchantability or fitness. Handling,
storage, cleaning and storage, cleaning and sterilization of this instrument as well as other factors
relating to the patient, diagnosis, treatment, surgical procedures, and other matters beyond BSC's
control directly affect the instrument and the results obtained from its use. BSC's obligation
under this warranty is limited to the repair or replacement of this instrument and BSC shall not
be liable for any incidental or consequential loss, damage, or expense directly arising from the
use of this instrument. BSC neither assumes, not authorizes any other person to assume for it, any
other or additional liability or responsibility in connection with this instrument. BSC assumes no
liability with respect to instruments reused, reprocessed or resterilized and makes no warranties,
expressed or implied, including but not limited to merchantability or fitness for intended use,
with respect to such instrument.
Disclaimer and Exclusion of Other Warranties
There are no warranties of any kind which extend beyond the description of the
warranties above. EPT disclaims and excludes all warranties, whether expressed or
implied, of merchantability of fitness for a particular use or purpose.
Limitation of Liability for Damages
In any claim or lawsuit for damages arising from alleged breach of warranty, breach of
contract, negligence, product liability or any other legal or equitable theory, the buyer
specifically agrees that EPT shall not be liable for damages for loss of profits or revenues,
loss of use of the product, loss of facilities or services, any downtime costs, or for claims
of buyer’s customers for any such damages. EPT’s sole liability for damages shall be
limited to the cost to buyer of the specified goods sold by EPT to buyer which give rise to
the claim for liability.
The buyer’s use of this product shall be deemed acceptance of the terms and conditions of
these limited warranties, exclusions, disclaimers and limitations of liability for money
damages.
U.S.5195968; U.S.5254088; U.S.5257451; U.S.5336182;. U.S.5363861; U.S.5364351; U.S.5363874; U.S.5273535; U.S.5358478;
U.S.5395327, U.S.5456682; U.S.4566682; U.S.5531686; U.S.5524337; U.S.5309910; U.S.5313943; U.S.5398683; U.S.5471982;
U.S.5549108; U.S.5489275; U.S.5328467; other U.S. patents pending; foreign counterparts also pending.
EP TECHNOLOGIES, INC.
2710 Orchard Parkway
San Jose, CA 95134
9-1
Index
ablation procedures
temperature control, 6-10temperature sensing catheter, 6-18warnings, 3-1
accessories
footswitch, 1-1, 5-16XP APM, 1-1
accessory atrioventricular conduction
pathways
indications for use, 2-1adverse reactions, 3-4analog outputs
description, 5-14atrial flutter
indications for use, 2-1audio control. See volume control knobauxiliary power outlet
description, 5-15AV block
indications for use, 2-1AV nodal reentrant tachycardia