Boonton 4210 Instruction Manual

ELECTRONICS
CORPORATION
P/N
98100902
791
ROUTE
10,
NJ
07859
TELEPHONE:
201
584-1077
TWX:
710-986-821
5
'***
'
Printed in U.S A.
8/8 5
IV
SECTION
I
INTRODUCTION
Section
I
Introduction
1-1.
SCOPE OF
MANUAL.
1
-
2
.
This
instruction manual
provides descriptive
data,
operating instructions, theory
of
operation,
maintenance
instructions,
and a
parts
list
for
the
Model 4210 R.
F.
Micro
wattmeter. (See Figure 1-1.)
SAFETY NOTICE
Although this
instrument has
been
designed
in
accordance
with
international
safety
standards,
general safety precau-
tions must
be
observed
during all phases
of
operation and
maintenance
of
the instrument. Failure to
comply
with
the
precautions
listed in the Safety Summary
located
in the
front of this
manual
could
result in serious injury
or
death.
Service
and
adjustment
should
be
performed only
by
qualified service
personnel.
1-3. PURPOSE
OF
EQUIPMENT.
1
-
4
.
The
instrument
is
a
microprocessor-controlled,
solid-
state
unit
that
features
ease
of
operation, high
sensitivity,
low
input s.w.r., and low
noise.
It measures
r.f. power
levels
from
1
nW (-60 dBm)
to 100
mW
(+20
dBm). The
cali-
brated
frequency range
extends from
0.2
MHz to
18
GHz,
depending
upon
the
accessory
sensor used; useful
response
for
relative measurements
is
obtained
from
20 kHz
to
approximately 20
GHz.
Representative
uses of the
instru-
ment
include:
a.
Low-power
transmitter, signal generator, and oscilla-
tor
measurements
b.
S.W.R. and return-loss measurements
with direc-
tional
couplers
and
slotted
lines
c.
Gain
and insertion-loss measurements
d.
R.F.
attenuation
and
s.w.r.
measurements
e.
Antenna
measurements
1-5.
DESCRIPTION
OF
EQUIPMENT.
1
-
6
.
The instrument is
packaged
as
a
compact
bench
unit.
When
operated with Boonton Series
4210-4/
5 sensors,
the
instrument
displays r.f.
power
by
measuring the voltage
across
a precision,
non-inductive resistor
in the
sensor
head
with
specially selected diodes. Panel
indications
are
cali-
brated
in
terms
of power according
to the
relationship
P
=
E
2
/R. This
detection
system
has important performance
advantages
over
power
meters that use
bolometer
or
thermo-electric
detection. The instrument
sensitivity
of
1
nanowatt
(-60
dBm) is
orders
of
magnitude better;
temper-
ature
stability of
better than
0.001
dB/°C supports
this
sensitivity;
and
burnout levels exceed
300
milliwatts.
average power
for all
types of
waveforms. Above
these
levels,
the instrument displays,
by
means
of
internal
shap-
ing,
the
true
average
power of
c.w.
signals.
If
the
r.f.signal
is
gated or
amplitude
modulated, the indicated
power
may
not
be the
true
average
power.
The signal
may be
attenu-
ated
to bring it
within the r.m.s. region of
measurement.
Alternatively,
the
instrument
can be
used with thermal
sensors
Model 4210-7E/8E;
in this case,
the
instrument
mea-
sures and
displays
the true
average
power for
all
waveforms.
1-8
The
outstanding
design
features
are:
a.
Wide
Frequency
Range. The
calibrated
frequency
range
of the
instrument
is
determined
by
the
sensor used
with the instrument.
The
instrument
is normally
ordered
with
one
of the
following
sensors: / ,
FREQUENCY RANGE”"
60
/
+
to
dDnn
Model
(Impedance) Power
Range
42
10-4
A
4210-4B
42
1
CMC
4210-4E
42
1
0-5
B
4210-5E
4210-7E
421
0-8
E
200 kHz
to
7
GHz
(50
ohms)
200
kHz
to 12.4
GHz
(50
ohms)
200 kHz
to
I GHz
(75
ohms)
200
kHz
to 18 GHz
(50
ohms)
200
kHz to
12.4 GHz
(50
ohms)
200 kHz
to 18
GHz
(50
ohms)
10
MHz
to 18 GHz
(50
ohms)
10
MHz to
18
GHz
(50
ohms)
1
nW
to
10
mW
1 nW
to
10
mW
1 nW
to 10 mW
1
nW
to
10
mW
10nWto 100
mW
10
nW
to 100 mW
1
jaW
to
10
mW
IOjuW
to
100
mW
b. Wide
Power Range.
Depending
on
the selected
sen-
sor,
the
instrument
will measure
r.f. power from 1
nW
up
to
100
mW.
Temporary
overloads
up to
300
mW
with
Series
42104
sensors,
and up
to
2 W with
Series
4210-5
sensors,
will
do
no permanent harm
to
the
instrument
or the
sensor.
When
measuring
pulsed signals,
the power
indications
are
accurate
up to
20 microwatts peak
power
(up to
200
mi-
crowatts
with Series
4210-5
sensors).
External
attenuators
may
be
used to extend
the
measurement range
of
the
instrument.
c.
Low Noise. The
instrument has been designed
and
constructed to
minimize
noise from
all
sources.
The
sensor
cable
is
of
a
special low-noise
design; vigorous
flexing
causes
only
momentary
minor excursions
of
the display
on
the
most
sensitive
range
of the
instrument. The
sensors
are
insensitive to
shock and
vibration; even
sharp
tapping
on
the
sensor
barrel causes
no visible
excursions on any
range.
Internal signal
amplification
occurs at
approximately
94
Hz,
thereby reducing
susceptibility
to
50-
or 60-Hzfields.
A
low-noise solid-state
chopper
is
used.
1
-
7.
Diode
sensors are
r.m.s. -responding
for
low-power
levels
(below
20
microwatts
for Series 42104 sensors and
200
microwatts
for Series
4210-5
sensors). At these low-
power
levels, the
instrument measures and displays true
d.
LED
Display.
Measured power levels are
displayed
by
a
4-digit,
LED-type
readout with decimal
points
and
minus sign.
Annunciators associated
with the
LED
display
indicate
the
units
of measurement.
The result is
a clear,
1-1
Section
I
Introduction
unambiguous
readout that
minimizes the
possibility of
misinterpretation.
e.
Analog
Indications. A
front-panel
analog meter
pro-
vides
relative power
indications for
peaking or nulling
applications.
The
display
is proportional
to
power on
each
range
(PWR
mode) or
to
dB
over the
entire range (dB
mode).
f.
Pushbutton
Measurement-Mode Selection.
A
choice
of
measurement
modes is
available
to
the operator. Indica-
tions
in
terms of
power or of
dBm can be selected
by
pressing
the
appropriate
front-panel
key
switch.
A dB ref-
erence
level
equal
to
the
last
displayed dBm value can be
entered
through the
keyboard
REF switch,
and
a display
mode
can
be
selected to
indicate
power levels in dBr relative
to
that
dB reference level.
g.
Automatic
Ranging.
Autoranging under control of
the
microprocessor
eliminates the
need for
manual
ranging.
Applications of
power
levels that exceed
the
maximum
capability
of the
instrument
result in an
error indication on
the
LED
display.
h.
Automatic
Zeroing.
An
automatic
zeroing
circuit
eliminates
the
need
for tedious,
often
inaccurate,
manual
zeroing.
With
zero input
to
the
sensor,
pressing
a front-
panel
ZERO key
switch
directs
the microprocessor to
compute
and
store
zero
corrections
for each
range, and the
instrument is
thereafter
corrected
on each range
in
accor-
dance
with
the
stored data.
This
method
is
considerably
simpler,
faster,
and
more
accurate
than
manual
zeroing.
i. Sensor
Compensation.
Calibration factors in
dB
are
selected
by
a
front-panel
rotary
control calibrated in
0. 1 dB
stepsfrom
1 . 10 to -1
.10
dB. The
sensoritself
is
marked with
the
appropriate
calibration factors
as a function of
frequency.
j.
Solid-state Chopper. Signal
amplification in
the in-
strument
occurs at approximately
94
Hz.
Input
signals
from the sensor
are
converted into a
94-Hz signal by a
solid-state, low-level input modulator
(chopper),
which
rep-
resents a distinct improvement over
electromechanical
choppers. Extended service
life
is
assured through
the
elim-
ination
of
contact wear, contamination, and
other
prob-
lems associated with electromechanical
choppers.
k.
Signature-Analysis
Maintenance.
Connection
facili-
ties
to
permit
signature-analysis maintenance
are incorpo-
rated. Digital
circuit
troubles
can
be localized rapidly
and
accurately
using the
signature-analysis maintenance
tech-
nique,
thereby reducing instrument down-time. Anadapter
(P/N
950028)
is available from Boonton Electronics
Cor-
poration
for
signature-analysis maintenance.
1-9.
ACCESSORIES.
1-10.
The basic
instrument is supplied complete
with
power
cord, sensor cable and
sensor.
It is designed
to
operate
with any Boonton Electronics Corporation
Series
4210 sensor,
the
characteristics of
which are
listed
in
Table
1-1.
The
calibration data for
the
particular
sensor ordered
are
written
into microprocessor memory
before
shipment
of
the instrument.
1-11.
Inquiries
regarding special applications of
the
in-
strument
to
specific
customer requirements
are
invited.
Direct such
inquiries
to
the Applications
Engineering
De-
partment
of Boonton Electronics Corporation.
1-12.
SPECIFICATIONS.
1-13.
Pertinent performance specifications for
the
instru-
ment are listed
in Table
1-1.
1-14.
OUTLINE
DIMENSIONS.
1-15.
Outline
dimensions
of the
instrument are
shown
in
Figure
1-4.
TABLE
1-1.
PERFORMANCE
SPECIFICATIONS
Parameter
Specifications
FREQUENCY
RANGE:
200
kHz
to
18
GHz (depending
on
power
sensor
used)
MODES:
Power
Display in
nW,
juW,
or
mW
dBm
Display in
dB relative
to
1
mW
dBr
Display
in
dB
relative
to the previous
dBm display
POWER
RANGE:
With all
4210-4
series
sensors
70 dB dynamic range
with
7
full-scale
ranges
of-50,-40,-30,-20,-10,
0
and
+
10
dBm
(10
nW
to
10 mW
f.s.)
With
all
4210-5
series
sensors
70
dB
dynamic
range
with
7 full-scale
ranges
of-
40,
-30,
-20,-
10,
0,
+
10
and
+20
dBm
(100
nW
to
100
mW
f.s.)
1-2
Section
I
Introduction
TABLE
1-1.
PERFORMANCE
SPECIFICATIONS
(cont.)
Parameter
Specifications
With
4210-7E
sensor
With
4210-8E
sensor
Ranging
40 dB
dynamic
range with
4 full-scale ranges
of-
20,-
10,
0,
and
+
10
dBm
(10/xWto
lOmWf.s.)
40
dB
dynamic
range with
4
full-scale
ranges of
-10,
0,
+10,
and
+20
dBm
(100
to
100
mW
f.s.)
Automatic
INSTRUMENTATION ACCURACY:
(Uncertainties include
instrumentation, non-linearity, range
zero correc-
tions, and
noise).
Zero
Zero
drift
Temperature
effect
GENERAL:
Cal.
Factor
Waveform
Response
Modes
Ranges
All
But
Lowest
(All
Sensors)
Lowest
(4210-4
&-5
Sensors)
Lowest
(4210-7E Sensor)
PWR
MODE
±
1
.5%
rdg. ±0.1%
f.s ±2%
rdg.
±
1.5%
f.s.
±
1%
rdg.
+
3%
f.s.
dB
Modes
+
0 07 HR ±
dB
f.s.
-
dB
rdg
±0 15dR±
dB
f
.,-dB
rd
,
±n.l dR±
dB
f.s.
-
dB
rde.
250
15 10
Automatic,
operated
by front-panel
key
switch
With all
4210-4
series
sensors:
1
riW/h,
max.,
on
10
nW range
With all
4210-5
series sensors:
10
nW/h, max.,
on
100
nW
range
With 4210-7E sensor:
1
/xW/h, max., on
10
(JW
range
With
4210-8E
sensor:
1
/xW/h,
max.,
on
100
jliW
range
Range Instrument
21°
C
to
25°
C 0
dB
18°
C
to
30°
C
0
dB
10°
C
to
40°
C
±0.2 dB
Sensor
0
dB
±0.1
dB
±0.2
dB
Range
1.10 dB
to
-
1.10
dB,
entered via
calibrated front-panel
control
With
all
4210^1
series sensors: True average power
to
20
/xW;
above
20
/uW,
average
power of sine wave
With
all
4210-5
series
sensors:
True
average
power
to
200
/xW;
above
200/xW,
average power of
sine
wave
Measurement
Speed
With 4210-7E/8E sensor:
True
average
power
With
4210^4
and
-5
sensors:
Typically 1
.0s
to
2.5s.
Worst
case
below
-40dBm,
8s
for
increasing and
27s
for
decreasing levels
Display
With
4210-7E
and
4210-8E sensors:
Typically
2s
to
5s for increasing
levels
and
2s
to
23s
for
decreasing
levels
4-digit LED. 3-1
/
2 digit display of power, 4-digit display
of
dB
with 0.01
dB
resolution
Annunciators
Error
Indication
Power
Consumption
Weight
Size
Auxiliary analog
display,
uncalibrated, proportional
to
power
(PWR
mode) or dB
(dB mode)
LED display of
nW, /xW,
mW,
dBm
and
dBr
Display
shows
overrange
condition
100,
120,
220,
240
V, 50
-
400
Hz,
18
VA
2.27 kg
(5
lbs.), approx.
114
mm
high x
216 wide
x
257 deep
(4.5
in.
x
8.5
x
10.1)
1-3
Section
I
Introduction
TABLE
1-2.
SENSOR
CHARACTERISTICS
S«fiaor
Typo
Full-wave dfode
Thermocouple
type
Model
4210-4A
4210-4B
4210-4C 4210-4E
4210-5B 4210-5E
4210-7E
4210-8E
Input
son
so
n
75 £1
50fi
so
n
5012
so
n
Frequency
Range
200kHz-7GHz 200kHz-12.4
GHz
200kHz-1
GHz
200kHz-18GHz
200kHz-12.4GHz
200kHz-18GHz
10MHz-18GHz
Power
Range
1
nW
to
10
mW
10
nW to
100
mW
1/jW to 10
mW
10 pW
to
10
mW
Sun of
Calibration
Factor
Uncertainty
1%,
0.2-300
MHz
1,3%,300MHz-2GHz
3.0%,
2G
Hz-4
GHz
3.5%,
4GHz-7GHz
1%,
0.2-300MHZ
1.3%,
300MHz-2GHz
3.0%,
2G
Hz-4 GHz
3.5%,
4GHz-8GHz
4.0%,
dGHz-IOGHz
4.5%,
10GHz-12GHz
6.0%,
12GHz-18GHz
Add
0.05
dB/mW
above
4 GHz
1%,
0.2-300
MHz
1.3%,
300MHx-2GHx
3.0%,
2
GHz-4GHz
3.5%,
4
GHz-8 GHz
4.0%, 8GHz-l0GHz
4.5%, 10GHz-12GHz
6.0%,
12GHz-18GHz
Add 0.005
dB/mW above 4
GHz
1%, 10-300MHz
1.3“/*,
300MHz-2GHz
3.0%,
2GHz-4GHz
3.5%,
4GHz-8GHz
4.0%,
8
GHz-
10 GHz
4.5%, 10GHz-12GHz
6,0%,
12GHz-18GHz
Max. SWR 1.12,
200kHz-2GHz
1.2,
2GHz-4GHz
1.4,
4GHz-7GHz
1.12,
200kHz-2GHz
1.2,
2G
Hz-4 GHz
1.4,
4GHz-1
1GHz
1.6,
11
GHz-12.4
GHz
1.18,
20
0k
Hz-1
GHz
1.3,
200kHz-4GHz
1.5,
4GHz-10GHz
1.7,
10GHz-18GHz
1.07,
200kHz-1 GHz
1.10,
1GHz-2GHz
1.12,
2GHz-4GHz
1.18,
4GHz-12.4GHz
1.07,
200kHz-1GHz
1.10,
1GHz-2GHz
1.12,
2G
Hz-1 GHz
1.18,
4GHZ-12.4GHZ
1.28,
12.4GHz-18GHz
1.5,
10
MHz-15MHz
1.35,
15MHz-10GHz
1.6,
10GHz-18GHz
|
Max
Average
Power
10 mW {+10
dBm)
100
mW (+20
dBm)
10
mW
(+10 dBm)
100 mW
(+20 dBm}
|
Overload
Bating
300
mW
(+25
dBm)
2W
(+33
dBm)
30 mW
(+14
dBm)’
200 mW (+23 dBm)
|
RF
Connector Precision Type
N
male
Calibration
Factor
Individually calibrated
at up to
9 frequencies,
depending
on sensor.
1-4
Section
I
Introduction
MAXIMUM RESPONSE TIME
IN
SECONDS
WITH
4210-4/5
SENSORS
LI
(dBm)
L2 (dBm)
+
10 0
-10
-20
-30
-40
-50 -60
+io-
0
-10
-20
-30
-40
-50
-60
1.5
1.5
2.5
2.5
1.75
2.5
2.5
2.5
2.5
2.5
2.5
2.5
2.5
INCREASING
LEVEL
LI TO
L2
NOTE
For
Series
5 sensors
increase
levels
by 10
dB.
MAXIMUM RESPONSE
TIME
IN
SECONDS
WITH
4210-4/5
SENSORS
LI
(dBm)
L2
(dBm)
+
10
0
-10
-20
-30
-40
-50 -60
+
10
-
0
-
-
10
-
-
20
-
-30-
-40-
-50-
-60-
1.5
1.5
1.75
1.5
1.75
1.25
27
18
15
10
10
11
DECREASING
LEVEL
LI to
L2
NOTE
For
Series
5 sensors
increase
levels by 10
dB.
Figure
1-2
Response-Time Chart
for
Series
42 10
A/
5
Sensors
1-5
Section
I
Introduction
MAXIMUM
RESPONSE
TIME
IN SECONDS
WITH 42I0-7E/8E
SENSORS
LI dBm
L2dBm
INCREASING
LEVEL
LI to L2
NOTE
For Series 8E
sensors increase
levels by 10 dB.
MAXIMUM
RESPONSE
TIME IN
SECONDS
WITH
4210-7E/8E
SENSORS
L2
dBm
Figure
1-3
Response-Time
Chart
for
Instrument and 4210-7E/8E Sensor
1-6
Figure
1-4
Outline
Dimensions
1-7
Section
II
Operation
SECTION
II
OPERATION
2-1.
GENERAL.
2-2.
This section
provides instructions for installation
and
operation of
the
instrument. Although the design
of the
instrument
emphasizes ease
of
operation,
it
is recom-
mended
that the operator familiarize himself
thoroughly
with
the
material in
this section before attempting to
oper-
ate
the instrument;
otherwise,
the full capabilities
of
the
instrument
may
not be realized in
use.
2-3.
INSTALLATION.
2-4.
Unpacking. The
instrument is
shipped
complete
with
sensor,
and is
ready
for
use upon receipt. Packaging
details
are
shown
in Figure
2-1.
Unpack the
instrument carefully,
and inspect it for
any signs of
shipping damage.
Should
any
damage be
noted, notify
the carrier
and the
factory
immediately.
NOTE
Save the
packing material and container
for
possible
use
in
reshipment of
the instrument.
2-5.
Mounting.
For
bench
use,
choose
a clean,
sturdy,
uncluttered
surface.
See
Figure
1-4
for
space
requirements.
2-6.
Power
Requirements.
The instrument
has a tapped
power transformer which
permits operation
from
100,
120,
220,
or
240
volt
±
10%,
50 to 400 Hz, single-phase a.c.
power
sources.
Power consumption is
approximately
18
volt-amperes
at
60 Hz.
Section
II
Operation
2-7. Cable
Connections.
A
line
cord
and sensor
cable
are
supplied
with
the instrument.
Any other
cables
required
must
be
supplied by
the
user.
Cable
connections
that may
be
required
are:
a.
Sensors.
The
sensor cable
supplied
with
the
basic
instrument
connects
directly
to
the
front-panel
SENSOR
connector,
and
the
sensor
that is to
be used
for power
measurements
connects
directly
to
the
other
end
of the
sensor
cable.
Although the
sensors are
insulated
against
extreme
temperature variations,
it
is
advisable to
locate the
sensor
away
from heat
sources
when using
the most-
sensitive
ranges
of the
instrument.
If the
instrument
is
to
be
used
to
measure
the
output of
equipment
that
generates
heat
significantly
above
the
ambient
temperature,
a short
length
of
coaxial cable or
solid line
having the
same
charac-
teristic
impedance
as
the
sensor
may
be
used between the
sensor
and the
equipment
undergoing test
to allow
heat to
dissipate
before
reaching the
sensor. If such
a cable is used
,
the
length
must
be kept
as
short as
possible for
operation
at
the
high
end
of
the
frequency range;
cable
losses
and
an
increase in
s.w.r.
will tend to
degrade measurement
accuracy.
2-8.
OPERATING CONTROLS,
INDICATORS,
AND
CONNECTORS.
2-9.
Controls,
indicators,
and connectors
used
during
op-
eration
of
the
instrument are
shown
in Figures
2-2
and
2-3.
Table
2-1
lists their
functions.
2-10.
SENSOR
CALIBRATION DATA.
2-11.
Low-frequency
calibration
corrections for sensors
ordered with
the
instrument are
written into microproces-
sor
memoryat thefactory
before
shipmentof
the instrument
and sensor.
High-frequency
calibration
data appear
on
the
chart on the sensor barrel and
the
proper
dB
correction
is
entered using
the
CAL FACTOR
dB control.
For the
procedure
for
determining
the
calibration
factor, refer
to
paragraph
4-21.
The field
replacement
of any
sensor with
TABLE
2-1.
OPERATING
CONTROLS, INDICATORS,
AND
CONNECTORS
Control,
Indicator,
or
Connector
Figure
and
Index
No. Function
METER
2-2, 1
Provides relative
indication of power
or
dB for
peaking
and
nulling operations.
LED
display
2-2,2 Four-digit LED
display
with
minus
sign and
decimal
points;
provides
numerical
indication of measured
power,
dBm, or
dBr.
mW,
/AV,
and nW
annunciators
2-2,
3
Indicate units of power
when instrument
is operating in
power mode.
dBm
and
dBr
annunciators
2-2,4 Indicate
mode of dB operation.
CAL
FACTOR
control
2-2,
5
Provides
means of
entering
calibration
factors
in dB.
LINE
switch
2-2,6
Provides
means for switching
a.c. line
power on and off.
REF
key
2-2,
7 Provides
means
for
storing
a
displayed dBm reading
as
the
reference
for future dBr
measurements.
dBr
key
2-2,8
Provides
means for displaying dB measurements
offset by
the
selected
reference (see 2-2,
7).
PWR/dBm
keys
2-2,9 Provide means
for
selecting
power or
dBm
mode
of display.
ZERO key
2-2,
10
Provides means
for generating and storing zero
corrections
for
all
ranges
with
zero
input
to
sensor.
SENSOR
connection
2-2,
1
1
Provides
means for
connecting sensor.
Power
connection
2-3,
1 Provides
connection
for a.c.
power
cord.
FI
fuse
2-3,2
Accepts 0.3
A
(100/120
V
a.c.)
or
0.2
A
(220/240
V a.c.) fuses.
A.C.
power
switch
2-3,
3
Provides
means for selecting
a.c.
power
of
100,
120,
220,
or
240
V.
Fuse
chart
2-3,4 Describes
required fuse value
for various a.c.
line
voltages.
2-2
Section
II
Operation
Section II
Operation
another of
the
same
type requires only
a simple recalibra-
tionprocedure (see
Section
IV).
If
a
sensor
is replaced
with
one
of
a
different
type,
a
replacement
PROM
is also
re-
quired
to
reconfigure the
4210
to
the
new
sensor.
a. General.
Power
sensors
used
with
the
Model
4210
can be
thought
of as
four-part items: the power
sensor
itself,
an
EPROM,
and two
resistors.
If
you
change the
type
of sensor
from
that
with
which the
4210 was
originally
supplied,
it may
be necessary to
change
the EPROM and
the
resistors
as
well. The
following
table
gives
the part
number for
ordering a new
sensor;
it
is
in fact the
part
number
for a kit that
includes the
sensor,
an
EPROM
and
two
resistors.
In
some
cases
you
may
receive
an EPROM with the
same
part
number
as that
of the EPROM
already installed in the
instrument.
However, since the new one
may be an updated
version,
it
is
preferable to install
the
new EPROM.
Note
that even
if neither the EPROM nor the
resistors
need
be
changed, it will
always
be
necessary to recalibrate the
4210
when
changing sensors.
Model
4210 Power-Sensor
Configuration
Part
No.
Sensor
Type
EPROM
Resistors
95
101801A
42
10-4
A
53427600A 341325000
95
101901A
4210-4B
53427600A 341325000
95102001A
4210-4C
53427600A
341325000
95
102101A
4210-4E
53427600A
341325000
95
102201A
4210-5B 53428300A
341325000
95
102301A
4210-5E
53428300A
341325000
95102401
A
4210-7E*
53429200A
**
95102501A
4210-8E*
53430700A
**
534388000
**
534389000
**
*
Add
two
(2)
(343729000)
20M
5
%
1/4
W
resistors to
the
Chopper
Board.
**
A
selected
resistor
value between 4.32K
to
8.25K
must
be
used.
b.
Installation.
(1)
Remove line
power
from
the
4210.
(2)
Remove
the instrument’s top
cover.
(3)
Replace
U29 (EPROM)
in
the
instrument
with
the EPROM
supplied
in the
kit. Note that if
you are
replacing a sensor with
one of the same type,
you
do
not
have
to
replace
U29
but, as noted earlier,
it
may
be
desirable
to
do so.
(4)
Use
the
Overlay
drawing on
page
6-7
to help
locate resistors
R21
and
R23.
(They
are
at
the
very
bottom of
the
drawing, just to
the
right
of
center.)
Replace
these
resistors
with the ones
that are
supplied
in the
kit: 1.8
k-ohm for
diode-type
sensors;
and
7.5
k-ohm
for
thermal
sensors.
Note that
if
you
are
replacing a
sensor with one of
the
same type,
there is
no need
to
change
R21 and
R23.
(5)
If parts
have
been
replaced
or
if you
are changing
sensors
(even
though
the sensors are the same type),
the Model 4210 must
be
recalibrated.
Refer to
page
4-2,4-14.
2-12.
POWER
APPLICATION.
2-13.
The basic instrument
is
designed
for
operation
from
a.c.
line
voltages
of
100,
120, 220,
or
240,
50 to
400
Hz,
single phase. To
apply
a.c. power, proceed
as
follows:
a.
Determine
the
line
voltage at the a.c.
power
output
receptacle.
b.
Position the
two voltage-selector
slide
switches
on the
rear
panel to
agree
with
the a.c.
line voltage.
c. Check
the rating
of
the
fuse
in the rear-panel
fuse-
holder.
For
100-
or 1 20-volt
operation, the fuse
should
be a
0.3-ampere,
MDL Slo-Blo
type;
for
220-
or 240-volt opera-
tion, it
should
be
a
0.2-ampere, MDL
Slo-Blo type.
If
the
ratingofthe
fuse
is
incorrect, install
a
fuse of
the
required
rating in
the
fuseholder.
WARNING
|
The
instrument
is
designed
to
operate
from
a
3-terminal
(one ground) a.c.
power
receptacle.
If
only
a
2-terminal a.c.
power receptacle
is
available,
use
a 3-prong-to-2 prong
adapter.
Connect
the
ground
wire of the
adapter
to
the
power
receptacle
ground,
to
eliminate
a
poten-
tial
shock
hazard
to
the operator.
d.
Connect the power
cord
between the
a.c.
power
con-
nector
on the rear
panel
of
the
instrument
and the
a.c.
power receptacle
(with adapter,
if necessary).
2-14.
PRELIMINARY
CHECK OF
INSTRUMENT.
NOTE
The
following checkout
procedure is
intended
merely
to demonstrate that
the major
circuits
of
the
instrument are operating,
before the in-
strument
is
placed
in
service.
For a
detailed
check of the
instrument
against
performance
specifications,
refer
to
paragraph
2-28.
2-4
Section
II
Operation
2-15. To
perform a preliminary
operational
checkout,
proceed as
follows:
a.
Set the LINE
switch
to
the ON position.
The LED
annunciator
directly
above the switch
should light.
b.
Immediately at
power-on
all
the LED
display digits,
the
minus
sign,
and all annunciators
should
light for
ap-
proximately 1 second,
blank for 0.5 second,
and
then
commence
normal indications. Their failure
to
exhibit this
sequence
is an
indication
of instrument malfunction.
c
.
Connect the correct
power sensor
to
the front-panel
SENSOR
connector.
d.
With
no
signal
connected to the
sensor, press the
PWR
key
and
then
press
the ZERO
key. A “good zero” is
indicated by equal positive and negative
excursions
of the
display,
and
by
the
minus
sign flashing on
and off. If
necessary,
press
the
ZERO
key
a
second
time
to
obtain
a
good zero.
NOTE
Always have
the
instrument
in
the power
dis-
play mode before executing
a zero.
This
assures
that
the
quality
of the zero correction can
be
monitored
by
observing
the
fluctuations
of the
display around zero.
If
the
instrument
is
in the
dBm
mode, the error
signal LO
appears on
the
display for signals that
are less than
-70
dBm,
and
the
quality
of
the
zero therefore cannot
be
observed.
e.
Connect
any suitable signal source
of frequency
greater
than
200 kHzto
the
sensor
input and
verify
a proper
display of
power.
f.
Press
the
dBm
key
and
verify
an equivalent dBm
display.
g.
Press
the
REF key
to
enter
the
dBm level
displayed in
(f)
above,
as
the
reference
for
all subsequent dBr
displays,
h.
Press
the dBr key
and verify
that the
dB display has
been
offset
by the
reference established in
(g),
above.
i. Press
the dBm key, check the
displayed reading,
and
then
rotate
the CAL
FACTOR
control by
a given
amount.
Verify
that the
dBm
display
changes by
that
amount.
j.
Reset
the
CAL FACTOR
control
to 0.
k.
This concludes
the preliminary
operational
checkout.
To check
the
instrument
for
minimum performance
stand-
ards,
refer
to paragraph
2-28.
2-16.
OPERATING PROCEDURES.
2-17.
Programming
Measurement Parameters.
a. General.
Measurement
parameters are
entered
into
the
microprocessor through
the front-panel controls.
When the instrument is turned
on, the
PWR
key
is
auto-
matically
activated and no dBm
offset
is stored for
dBr
measurements.
b.
Mode Selection.
Two mode keys
select
either
power
units
or
dBm. When the
PWR
key
is
pressed,
measured
power
levels
are displayed in mW,
juW,
or nW;
the
annuncia-
tors
associated
with
the
LED
display
indicate
the
appro-
priate
unit.When the
dBm
key is
pressed,
measurement
values
are displayed in
terms
of
dBm
(0
dBm
is
equivalent
to
1
mW
across 50 ohms).
An
annunciator
associated
with
the LED display indicates when
the dBm mode
is
selected.
c.
Two mode keys select a dB display offset
by a refer-
ence value.
Any displayed dBm reading
may
be stored as
the
reference
by pressing
the REF
key.
When
the dBr
key
is
then
pressed,
all
future dB
readings
will
be offset by
that
reference. As an example, if the instrument
is
displaying
a
reading
of- 10.00
dBm,
pressing the
REF
key
will
store
this
value. When
the d
Br key
is pressed,
assuming
no change in
the input
power level to
the
sensor, the
display will indicate
0.00
dBr
[-
10.00
dBm
-
(-
10.00
dBm)].
NOTE
Since
the
difference
of two dB values
is
the
same as
the
ratio
of
the
two equivalent
power
values, offset measurements provide
ratio
ca-
pabilities. Additionally, an offset
dB display
is
a
convenient
means for normalizing
future
readings to the reference value.
d.
Range
Selection.
Range selection
is entirely
auto-
matic and
always results in
the
best possible
display
resolu-
tion.
e. CAL FACTOR Selection.
The
sensors
used
with the
instrument are frequency-sensitive; that
is,
with a
constant
input power
level applied, their output signal
level
does
not
remain
constant
as
the
measurement frequency
is
changed
.
Each
sensor is marked
with
its
required
corrections
as
a
function of frequency. The front-panel CAL
FACTOR
rotary
control provides the means
for
entering
these
correc-
tions into
the
microprocessor. As
an
example,
if a
sensor
has
a CAL
FACTOR
of-0.35
dB at a particular
frequency,
setting
the
CAL FACTOR
control to -0.35 dB
will
correct
the
display
value
for
that frequency
2
-
18 .
Zeroing the Instrument. For greatest
accuracy,
espe-
cially on the most-sensitive
ranges, the
instrument
must
be
1-5
Section
II
Operation
zeroed.
To
eliminate
tedious
and less-accurate
manual
zeroing,
the
instrument
incorporates
an
automatic zeroing
capability.
When
automatic zeroing
is
initiated, the mi-
croprocessor
reads
and
averages
the
offset and stores the
zero
correction
required on the
most sensitive range,
as
well
as
scaled
corrections required
for
the
other ranges.
The
instrument
then
applies
the proper
zero
correction for
the
range
in use
for
all
subsequent
measurements.
During in-
strument
warmup, or
when
used in
an
environment with
varying
ambient
temperatures,
the instrument should
be
zeroed
frequently
if
measurements
are
being made
on the
lowest
ranges.
To
zero
the instrument,
proceed
as
follows:
{
CAUTION
:
Never
press the
ZERO
key
with
a signal applied
to
the
sensor; to
do
so
will
result
in large,
erroneous
zero
corrections and
grossly
inaccu-
rate
measurements.
a.
Remove
all
input power to
the sensor.
NOTE
If
the instrument
is
subjected to strong
r.f.
fields,
shield
the sensor
in
accordance
with the
instructions
in
paragraph
2-27.
b. If
the
instrument is
in the
power mode,
press
the
ZERO key.
If
the instrument is
in
the dBm or dBr
mode,
first
press
the
PWR
key, then press
the ZERO
key.
A
“good
zero”
is
indicated
by
equal
positive
and
negative excursions
of
the
display and
by the
minus sign flashing
on and
off.
If
necessary,
press
the
ZERO key
a
second
time
to
obtain
a
good zero.
2-19. Making
Absolute
Power
Measurements.
Once the
instrument
is
zeroed,
it
is
ready for
power-level measure-
ments.
Merely
connect
the sensor
to the
source
whose
power
level
is to
be
measured and press the PWR
key,
for
display
in power
units, or
press
the
dBm key for
display in
dBm
units.
2-20.
Making
Relative or Offset
Power Measurements.
If
the
instrument
is
in the
dBm
mode of
operation,
any dis-
played
dBm
value
may be
used
as a reference
for subse-
quent
relative
power
measurements,
as
follows:
a.
Press
the REF key. This
stores the
current dBm
value
as
a
reference.
b. Press
the
dBr
key. The
dBr
annunciator
will light,
indicating
that the
instrument is displaying dB
values rela-
tive
to the
stored
reference. The
stored reference
remains
unchanged
for
all subsequent measurements
until
a new
reference
is
entered
or
the
instrument
is turned off.
2-21.
MAKING
MEASUREMENTS.
2-22.
Low-Level
Measurements. The instrument
will
pro-
vide reliable,
reproducible
measurements
of
c.w.,
a.m.,
and
f.m.
power
levels
as
low
as
1
nW
(-60
dBm).
It
can also be
used
for
pulse
measurements
but
with
slightly
decreased
accuracy
(±1
dB).
Peak
power levels
for
pulse measure-
ments should
not
exceed
200
fJSM
(20
fJW
for Series
4210-4
sensors);
above this level
the sensor enters the region where
it
operates out of
the
square-law
region,
and accuracy
at
such
signal levels
is
correct
for c.w.
and f.m. only.
2-23.
High-Level
Measurements.
Zeroing of the
instru-
ment is not
critical
when
making high-level
measurements
(10
fx
W
to 100 mW).
C.W. and
f.m.
power
measurements
can
be
obtained within the specified
accuracy up
to
100
mW; accuracy
cannot
be guaranteed for
pulse power mea-
surements with instantaneous peaks
exceeding
350
juW
(35
iuW
for Series
4210-4
sensors).
2-24.
High-Frequency
Measurements.
At
frequencies
above
1
GHz,
the
appropriate
sensor-calibration
factor
must
be entered by
the
CAL
FACTOR control
if
the
specified accuracy
of the
instrument
is
to be
realized. (Refer
to paragraph
2-17e.)
NOTE
Model
4210-4
A,
4210-4B,
4210-4E, 4210-5B,
4210-5
E and 42
1
0-7E
sensors
are calibrated for
use
with
a 50-ohm source; model
4210-4C
sen-
sors
are calibrated
for use with a
75-ohm
source.
Impedance
mismatch results
in
in-
creased
s.w.r.,
which affects measurement
ac-
curacy. (Refer to paragraph
4-30
and Figure
4-7.)
This effect can be
reduced
by
inserting
a
low-s.w.r. attenuator (s.w.r.
less than
1.10)
ora
low-loss
tuner
between
the source
and
the
sensor.
2-25.
Temperature
Effects.
Specified instrument accura-
cies
apply over an
ambient temperature range
of
21°
C to
25°
C. Operation outside this temperature range
causes
some additional error.
Figure
2-4
shows typical tempera-
ture characteristics
of
a
Series
4210 sensor,
and
Figure
2-5
shows
typical
temperature
characteristics
of the
instrument
and sensor
combined.
NOTE
For
best zero
stability of the
instrument,
allow
the
instrument and
sensor
to
reach
a stable
temperature.
2-26.
S.W.R.
Measurements.
The
high
upper-fre-
quency
limit
and
sensitivity
of the
instrument
facilitates
s.w.r.
measurements with a
slotted line.
S.W.R.
measure-
2-6
Section
II
Operation
ments
require
only comparative,
rather
than
absolute,
measurement
values;
therefore, the
instrument may
be
used
up to 20
GHz
with
a model
4210-4E sensor. The front-panel
meter
is
especially useful for
rapid determination
of maxi-
mum
and
minimum
power
points.
S.W.R.
is
determined
by
measuring
the
dB
difference
between
a
maximum
and
a
minimum
voltage
point on
a slotted
line
and converting
this
difference to
s.w.r.
An adapter, usually available from
the
slotted-line
manufacturer,
is
required to couple
the
sensor
to
the
slotted
line. To
make
slotted-line s.w.r. measure-
ments,
proceed as
follows:
a.
Connect
the sensor
to
the sliding carriage,
using
a
suitable
adapter.
b.
Ascertain that
the
signal source
is
turned off;
then,
zero
the
instrument.
c.
Turn
on
the
signal source
and
slide the carriage
along
the slotted
line
until a point
of
maximum
indication
is
located.
Adjust
the source signal level
and the
probe
setting
for
the
least
coupling that yields
a-41
dBm reading at
the
maximum
point.
(The
incident power should be
at
least
0 dBm.)
d.
Slide
the carriage along
the slotted
line until
a min-
imum
indication
is
located. Read the level
at this
point.
Subtract the
measured
level at the
minimum
point from
that at
the
maximum point,
ignoring signs.
Convert the
resultant
AdB into
s.w.r.,
either
through
use
of
the
s.w.r.
conversion
chart (Figure
2-6)
or by computation.
S.W.R.
is
the
antilog,
base
10,
of
AdB/
20.
2-27. Shielding
Recommendations.
If the instrument
is
subjected
to
strong r.f. fields, accurate zeroing
may
be
difficult
unless
the sensor is shielded during the
zeroing
operation.
The
simplest method
of
shielding
is
to
connect
the
sensor to
the device
whose power level
is
to
be
mea-
sured,
first
making sure
that the device
is
turned
off; how-
ever, in
some
instances, the device may
act
as
an
antenna
and
introduce
additional noise voltage into
the sensor.
If
this
happens,
disconnect the
sensor
from the
device,
stand
the
sensor,
end down, on
a copper plate, and
hold it
down
firmly
so that the
rim
of
the
sensor
connector
makes
good
contact with
the copper
plate at
all points.
Alternatively,
wrap a
piece of
thin copper
foil
around
the body
of
the
connector
body,
and
crimp
the
foil around the
open end
of
the
connector,
making certain that the
center pin of
the
connector is
not shorted.
If
frequent zeroing
in strong noise
fields is
necessary,
construct an adapter,
using
a
Type
N
connector
permanently fitted with
a
copper
foil
shield.
2-28. MINIMUM
PERFORMANCE
STANDARDS.
2-29.
Test
Equipment
Required. For minimum
perfor-
mance testing
of
the
instrument, a precision
adjustable
power
source, such as
the Boonton Model 25A
Power
Meter
Calibrator,
is
required.
NOTE
A 1 MHz
power source,
such as
the
Boonton
Model
25A,
is
recommended
for testing
the
instrument and its
sensors,
except
when
the
instrument is fitted
with
a
4210-7E sensor.
In
that case, an
equivalent 50
MHz
power
source
is
required.
2-30.
Preliminary
Setup.
a. Turn on
the instrument and
the
adjustable
power
source
and allow a warmup period
of at least
one
hour. The
ambient temperature should be
21°
C
(70°
F) to
25°
C
(77°
F).
b. Set
the adjustable
power
source
to zero
and
connect
the sensor between
the
power
source and the
front-panel
SENSOR
connector
of the
Model
4210,
using
the
sensor
cable.
c.
If
the
instrument
is
in its power
mode,
as indicated
by
the nW,
ynW,
or
mW
annunciators, press the
ZERO
key
of
the instrument.Ifthe instrument is
in
either dBm
or
dBr
modes, as indicated
by
those annunciators,
first
press the
PWR
key
followed by
the
ZERO key.
A
good
zero
is
indicated
by
equal excursions of
the
display
in
both
the
positive and
negative
directions,
with the
minus
sign
flash-
ing on and off.
Repeat
the zeroing process, if
necessary,
to
obtain
a good zero.
2-31.
Ranging
Test. To
check the
autoranging
function
of
the
instrument,
set
the
output
level
of
the adjustable
power
source
to
each of the values
listed below.
If
the instrumentis
ranging properly,
the
indications
will
be approximately
as
shown,
with
correct
decimal point and
annunciators.
Indications
with
Sensor
Power
4210-4
4210-5
4210-7E
Annunci-
Source
ators
10
nW
10.00
nW
100 nW 100.0
100.0
nW
1
mW
1.000
1.000
fJL
W
io
mW
10.00
10.00
10.00
100
mW
100.0
100.0
100.0
mW
1
mW
1.000
1.000
1.000 mW
10 mW
10.00 10.00
10.00 mW
100 mW
100.0
mW
2-32.
dBm Mode.
When
the instrument
is
turned
on
it
always reverts
to
the power
mode. To
check the
dBm
mode
of
the
instrument,
proceed as follows:
2-7
Loading...
+ 39 hidden pages