Section
II
Operation
2-15. To
perform a preliminary
operational
checkout,
proceed as
follows:
a.
Set the LINE
switch
to
the ON position.
The LED
annunciator
directly
above the switch
should light.
b.
Immediately at
power-on
all
the LED
display digits,
the
minus
sign,
and all annunciators
should
light for
ap-
proximately 1 second,
blank for 0.5 second,
and
then
commence
normal indications. Their failure
to
exhibit this
sequence
is an
indication
of instrument malfunction.
c
.
Connect the correct
power sensor
to
the front-panel
SENSOR
connector.
d.
With
no
signal
connected to the
sensor, press the
PWR
key
and
then
press
the ZERO
key. A “good zero” is
indicated by equal positive and negative
excursions
of the
display,
and
by
the
minus
sign flashing on
and off. If
necessary,
press
the
ZERO
key
a
second
time
to
obtain
a
good zero.
NOTE
Always have
the
instrument
in
the power
dis-
play mode before executing
a zero.
This
assures
that
the
quality
of the zero correction can
be
monitored
by
observing
the
fluctuations
of the
display around zero.
If
the
instrument
is
in the
dBm
mode, the error
signal LO
appears on
the
display for signals that
are less than
-70
dBm,
and
the
quality
of
the
zero therefore cannot
be
observed.
e.
Connect
any suitable signal source
of frequency
greater
than
200 kHzto
the
sensor
input and
verify
a proper
display of
power.
f.
Press
the
dBm
key
and
verify
an equivalent dBm
display.
g.
Press
the
REF key
to
enter
the
dBm level
displayed in
(f)
above,
as
the
reference
for
all subsequent dBr
displays,
h.
Press
the dBr key
and verify
that the
dB display has
been
offset
by the
reference established in
(g),
above.
i. Press
the dBm key, check the
displayed reading,
and
then
rotate
the CAL
FACTOR
control by
a given
amount.
Verify
that the
dBm
display
changes by
that
amount.
j.
Reset
the
CAL FACTOR
control
to 0.
k.
This concludes
the preliminary
operational
checkout.
To check
the
instrument
for
minimum performance
stand-
ards,
refer
to paragraph
2-28.
2-16.
OPERATING PROCEDURES.
2-17.
Programming
Measurement Parameters.
a. General.
Measurement
parameters are
entered
into
the
microprocessor through
the front-panel controls.
When the instrument is turned
on, the
PWR
key
is
auto-
matically
activated and no dBm
offset
is stored for
dBr
measurements.
b.
Mode Selection.
Two mode keys
select
either
power
units
or
dBm. When the
PWR
key
is
pressed,
measured
power
levels
are displayed in mW,
juW,
or nW;
the
annuncia-
tors
associated
with
the
LED
display
indicate
the
appro-
priate
unit.When the
dBm
key is
pressed,
measurement
values
are displayed in
terms
of
dBm
(0
dBm
is
equivalent
to
1
mW
across 50 ohms).
An
annunciator
associated
with
the LED display indicates when
the dBm mode
is
selected.
c.
Two mode keys select a dB display offset
by a refer-
ence value.
Any displayed dBm reading
may
be stored as
the
reference
by pressing
the REF
key.
When
the dBr
key
is
then
pressed,
all
future dB
readings
will
be offset by
that
reference. As an example, if the instrument
is
displaying
a
reading
of- 10.00
dBm,
pressing the
REF
key
will
store
this
value. When
the d
Br key
is pressed,
assuming
no change in
the input
power level to
the
sensor, the
display will indicate
0.00
dBr
[-
10.00
dBm
-
(-
10.00
dBm)].
NOTE
Since
the
difference
of two dB values
is
the
same as
the
ratio
of
the
two equivalent
power
values, offset measurements provide
ratio
ca-
pabilities. Additionally, an offset
dB display
is
a
convenient
means for normalizing
future
readings to the reference value.
d.
Range
Selection.
Range selection
is entirely
auto-
matic and
always results in
the
best possible
display
resolu-
tion.
e. CAL FACTOR Selection.
The
sensors
used
with the
instrument are frequency-sensitive; that
is,
with a
constant
input power
level applied, their output signal
level
does
not
remain
constant
as
the
measurement frequency
is
changed
.
Each
sensor is marked
with
its
required
corrections
as
a
function of frequency. The front-panel CAL
FACTOR
rotary
control provides the means
for
entering
these
correc-
tions into
the
microprocessor. As
an
example,
if a
sensor
has
a CAL
FACTOR
of-0.35
dB at a particular
frequency,
setting
the
CAL FACTOR
control to -0.35 dB
will
correct
the
display
value
for
that frequency
2
-
18 .
Zeroing the Instrument. For greatest
accuracy,
espe-
cially on the most-sensitive
ranges, the
instrument
must
be
1-5