The following general safety precautions must be observed during all phases of operation and maintenance
of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this
manual violates safety standards of design. manufacture. and intended use of the instrument. Boonton
Electronics assumes no liability for the customer’s failure to comply with these requirements.
THE INSTRUMENT MUST BE GROUNDED
To minimize shock hazard the instrument chassis and cabinet must be connetted to an electrical ground.
The instrument is equipped with a three conductor. three prong a.c. power cable. The power cable must
either be plugged into an approved three-contact electrical outlet or used with a three-contact to a
two-contact adapter with the (green) grounding wire firmly connected toan electrical ground at the power
outlet.
DO NOT OPERATE THE INSTRUMENT IN AN EXPLOSIVE ATMOSPHERE.
Do not operate the instrument in the presence of flammable gases or fumes.
KEEP AWAY FROM LIVE CIRCUITS.
Operating personnel must not remove instrument covers. Component replacement and internal adjust-
ments must be made by quailfied maintenance personnel. Do not replace components with the power cable
connected. Under certain conditions dangerous voltages may exist even though the power cable was
removed. therefore: always disconnect power and discharge circuits before touching them.
DO NOT SERVICE OR ADJUST ALONE.
Do not attempt internal service or adjustment unless another person, capable of rendering first aid and
resuscitation, is present.
DO NOT SUBSTITUTE PARTS OR MODIFY INSTRUMENT.
Do not install substitute parts or perform any unauthorized modification of the insturment. Return the
instrument to Boonton Electronics for repair to ensure that the safety features are maintained.
SAFETY SYMBOLS.
This safety requirement symbol (located on the rear panel) has been adopted by the
International Electrotechnical Commission. Document 66 (Central Office) 3, Para-
graph 5.3. which directs that and instrument be so labeled if. for the correct use of the
instrument. it is necessary to reler to the instruction manual. In this case it is
1
8
n
[WARNlNC)
recommended that reference be made to the instruction manual when connecting the
instrument to the proper power source. Verify that the correct fuse is installed for the
power available. and that the switch on the rear panel is set to the applicable
operating voltage.
The CAUTION sign denotes a hazard. It calls attention to an operation procedure.
practice. or the like. which. if not correctly performed or adhered to. could result in
damage to or destruction of part or all of the equipment. Do not proceed beyond a
CAUTION sign until the indicated conditions are fully understood and met.
The WARNING sign denotesa hazard. It calls attention to an operation procedure.
practice. or the like. which, if not correctly performed or adhered to. could result in
injury or loss of life. Do not proceed beyond a WARNING sign until the indicated
3-10. Function Selection.
3-11. Data Entry Operation.
3-13. Analyzer Measurement Description.
3-14. Analyzer Input Description.
3-15. Frequency Measurement Function.
3-16. Frequency Measurement Display Units.
3-17. Special Frequency Measurement Modes.
TABLE l-l. PERFORMANCE SPECIFICATIONS.
TABLE l-l. PERFORMANCE SPECIFICATIONS CObiINiJED: :
TABLE l-l. PERFORMANCE SPECIFICATIONS CONTINUED.
TABLE 2-l. INITIAL CONDITIONS.
TABLE 3-l. CONTROLS, INDICATORS AND dOiJN&‘ORS.
TABLE 3-2. FUNCTION DISPLAY AND DATA ENTRY UNITS.
TABLE 3-3. VALID FUNCTION ARGUMENT RANGE.
TABLE3-4.INPUTLEVELRANGE.S.
TABLE 3-5. DISTORTION AND SINAD RANGES.
TABLE3-6.OUTPUTLEVELRANGES.
TABLE 3-7. SPECIAL FUNCTIONS.
TABLE 3-8. OPTION SWITCH A4Sl.
TABLE 3-9. ERROR CODES.
TABLE 3-10. IEEE488 BUS MNEMONICS.
TABLE 3-11. TALK FUNCTION (TF) DECODING.
TABLE 3-12. BUS COMMAND RESPONSES.
TABLE 5-1. RECOMMENDED TEST EQUIPMBNT.’
TABLE 5-2. OPTIONAL FILTERS.
TABLE 5-3. ANALYZER DC LEVEL ACCURACY TEST RECORD.
TABLE 5-4. ANALYZER AC LEVEL ACCURACY TEST RECORD.
TABLE 5-5. ANALYZER AC LEVEL FLATNESS TEST RECORD.
TABLE 5-6. SOURCE IMPEDANCE ACCURACY TEST RECORD.
TABLE 5-7. SOURCE LEVEL ACCURACY TEST RECORD.
TABLE 5-8. SOURCE LEVEL FLATNESS TEST RECORD.
TABLE 5-9. ANALYZER LOW LEVEL AC ACCURACY TEST RECORD.
TABLE 5-10. SOURCE FREQUENCY ACCURACY TEST RECORD.
TABLE5-11. ANALYZER FREQUENCY ACCURACY TESTRECORD: : :
TABLE 5-12. FILTER ACCURACY TEST RECORD.
TABLE 5-13. RESIDUAL DISTORTION TEST RECORD. : : :
TABLE 5-14. RESIDUAL SIGNAL-TO-NOISE TEST RECORD.
TABLE5-15. COMMON MODE REJECTION RATIO TEST RECORD. : : : : : : . .
TABLE 5-16. CCITT FILTER ACCURACY TEST RECORD.
TABLE 5-17. CCIR FILTER ACCURACY TEST RECORD.
TABLE 5-18. A WEIGHTING FILTER ACCURACY TEST RECORD.
TABLE 5-19. B WEIGHTING FILTER ACCURACY TEST RECORD.
TABLE 5-20, C WEIGHTING FILTER ACCURACY TEST RECORD. ......... 5-20
TABLE 5-21. C-MESSAGE FILTER ACCURACY TEST RECORD.
TABLE 5-22. LIST OF ADJUSTMENTS.
TABLE5-23. DIAGNOSTICERROR CODEdE&kIiTiON. .
TABLE 6-l. MANUFACTURER’S FEDERAL SUPPLY CODE NUMBERS.
TABLE 6-2. REPLACEABLE PARTS LIST.
TABLE 7-l. LIST OF SCHEMATICS.
Figure 7-43. C-MESSAGE Board AlA Schematic. .................
7-39
7-40
7-40
7-41
7-41
7-42
7-42
VI
MODEL 1121 AUDIO ANALYZER
General Information
section 1
SECTION I
GENERAL INFORMATION
l-l.
INTRODUCTION.
1-2. This instruction manual
provides installation, operating
and maintenance instructions, theory of operation,
schematics and parts lists for the Model 1121 Audio
Wr-
13. DESCRIPTION.
l-4. The Model ll21 is a versatile, precision, solid-state
instrument with features and performance characteristics
especially suited to laboratory and industrial applications.
Human engineering considerations have been emphasized in
both the mechanical and electrical design of the Model 1121.
The result is an audio analyzer that is easy and convenient to
use. Among the outstanding features are:
Versatile Audio Analyzer.
a.
Advanced generation and measurement techniques enable the Model 1121 to
provide fast, accurate measurements. Measurement modes
include frequency, AC or DC level, distortion, SINAD, signal-to-noise and full ratiometric capability. The precision
audio source provides accurate, low distortion signals over
wide frequency and level ranges. Demanding applications are
satisfied by the high output power capability and selectable
output impedance.
b.
Ultra-low Distortion
. The mark of quality for any
audio analyzer is low residual distortion and noise. The total
harmonic distortion of the Model 1121 is specified as less
than 0.01% with typical performance an order of magnitude
better than specified.
C.
Versatile
Source
Output.
The Model 1121 fea-
tures wide control over its output configurations. Source
impedance can be set to SO, 150, or 600 ohms in either a
floating or single-ended co&guration. All source impedan-
ces are available at the same set of output connectors,
eliminating the need to multiplex separate SO ohm and high
impedance outputs.
d.
Balanced Input.
The Model 1121 has a fully dif-
ferential/balanced input for testing bridged amplifiers and
power supplies.
IEEE-488 bus status is also presented.
f.
Full Range
of
Filter Selections.
The Model
1121 provides a wide range of filter selections and weighting
characteristics for industry-standard audio measurements.
6
Sweep
Operation.
Frequency or level can be
swept in user-sehed linear or logarithmic steps over any
portion of the range. The selected analyzer measurement
provides the Y axis information. Rear panel X axis, Y axis
and PEN outputs are provided for plotter application.
h.
Instrument Setup Memory.
Up to 99 front panel
setups containing all data required to configure the instrument to a previous operating mode can be stored in non-
volatile memory for future recall. The last valid instrument
setup before power interruption is also saved automatically
and restored when power is resumed.
i.
IEEE-488 lnterfa~e
BUS.
All instrument functions
are programmable except line on/off. Annunciators to the left
of the BUS/PRGM display window show the status of bus
activity. The 1121 is designed to interface easily with controllers currently in use. A versatile free-form number entry
system is used so that the 1121 will accept any conceivable
valid number string. Triggering may be performed in immediate or wait modes. There are six talk modes which can be
addressed in either the remote or local state. The 1121 also
provides a choice of several end-of-string terminators. Service-request (SRQ) can be asserted on errors or using the
front panel SRQ key and the LCL4INK key will force return
to local control when using the bus as long as a lockout
message has not been sent.
l-5.
ACCESSORIES.
1-6. The available accessories are listed in Table l-l. The AC
power cord, spare input, output and line fuses are supplied
with the instrument:
1-7.
OPTIONS.
18. The available options are listed in Table 1-1.
e.
Separate Displays of All Functions.
The
Model 1121 has 3 separate display windows to simultaneously
present analyzer measurements, source settings and program
number or bus address information. Continuous display of
1-9.
SPECIFICATIONS.
l-10. Performance specilications for the Model 1121 Audio
Analyzer are listed in Table l-l.
l-l
section 1
TABLE l-l. PERFORMANCE SPECIFICATIONS.
SYSTEM SPECIFICATIONS
General Information
SigMl-tddW
Fundamental Frequency Range:
10 Hz to 100 kHz usable to 140 kHz
Display Range 0.00
to 140.00 dB
Afcurrrcy: +,ldB
LnputVoUageRangez
Residual Noise (the greater of):
5OmVto3OOV
85dBorlO/tV,8OkHzBW
85dBor#)~V;22OkHzBW
85dBor4O/tV;5OOkHzBW
SOURCE SPECIFICATIONS
Frequency
Ranges lOHzto14OkHz
Resolution:
0.001 Hz; 10.000 to 199.999 Hz
0.01 Hz; 200.00 to 1999.99 Hz
0.1 Hz; 2.oooO to 19.9999 kHz
1.0 Hz; 2O.ooO to 140.000 kHz
Accuracy: 10 ppk~ + timebase accuracy + 1 count
Level
FE&M& 16.000 V rms, open circuit
16.CNM V, 3.000 t, 300.0 mV, 30.00 mV
Re!sdution:
0.01 mV, 0.00 to 30.00 mV
0.1 mV, 30.0 to 300.0 mV
lmV,3OOto3000mV
5
mV, 3.ooO to 16.000 V
Accuracy (settiqp fiwm Oh0 mV to 16.000 V):
2 05 % of setting + 0.05 % of range;
10Hzto5OkHz
2 1.0 % of setting + 0.05 % of range;
5OkHzto1OOkHz
2 1.5 % of setting + 0.1% of range;
1OOkHzto14OkHz
Distottion
Residual Distortion and Noise (the greater of):
-80dBor10~V,10Hzto20)8080zBW
-74dBorm~V;lOHzto50IrHz,~LHz’BW
-7OdBor4O~V,1OHzto5OkHq5OOkHzBW
-65dBor4O~V,5Oto1OOkHz,5OOkHzBW
Flatness, ref 1 kH2
(030 mV to 8 V into 50 ohms):
+ 0.5 %; 10 Hz to 50 kHz
2 1.0 %; 10 Hz to 100 kHz
f 1.5 %; 10 Hz to 140 kHz
output
Impedance:
5oohms+2%
lxlohms&l%
6OOohmsZkl%
Distortion and Noise (the greater of):
0.01% (-80 dB) or 10 p V,
10 Hz to 20 kHz, 80 kHz bandwidth
0.02 % (-74
dB) or 20 p V,
10 Hz to 50 kHz, 220 kHz bandwidth
0.032 % (-70
dB) or 35 V,
p
10 Hz to 50 kHz, 500 kHz bandwidth
0.056 % (-65 dB) or 35 /.f v;
50 to 100 kHz, 500 kHz bandwidth
0.1% (-60 dB) or 35pV;
100 to 140 kHz, 500 kHz bandwidth
Output Power (50 ohm source):
31.07 dBm (8.00 V) into Xl ohm load
29.82
dBm (12.00 V) into 150 ohm load
25.60 dBm (14.76 V) into 600 ohm load
ANALYZER SPECIFICATIONS
Frequency Meesurement
Rangez 5Hzto2OOkHz
Sensitivityz
5 mV in the Frequency mode 0.01 Hz; 200.00 to 1999.99 Hz
50 mV in the Distortion and SINAD modes 0.1 Hz; 2.0000 to 19.9999 kHz
954018 Single binding post to BNC (M)
954019 BNC (F) to phono plug
954020 Phono jack to BNC (M)
954021 Two conductor shielded balanced line, 36 ”
954022 XLR Audio conncctor to three banana
P’ W
options:
-01 Rear panel input and output
-11 400 Hz high-pass filter
-12 CClTl’ band-pass filter
-l3 CCIR band-pass filter
-15 A weighting filter
-16 B weighting filter
-17 C weighting filter
-18 Audio band-pass filter
-19 C-Message band-pass filter
Remote Interlace:
IEEE--1978. Implements AHl, SHl, T6, TEO,
LA, LEO, SRl, RLl, PPO, DCl, DTl, CO, and El
l-4
- -..---__
“--
General Information
,
---------.-
17.34 -
‘
-
I
l&SO 10.00
A
Figure l-l. Outline Dimensions.
l-5
c’
section 1
General Information
1-6
Installation
Section 2
SECTION II
INSTALLATION
2-1.
INTRODUCTION.
2-2. This sqction umtains the unpackiq mount& power
requirements, cable connections, and preliminary check-out
instructions for the Model 1121 Audio Analyzer.
2-3. UNPACKING.
24. The instrument is shipped complete and is ready to use
upon receipt. Unpack the instrument from its shipping container and inspect for damage that may have occurred during
shipment. Refer to Figure 2-l.
NOTE
Save rhe packing mate&al and conkainer for possible
use in teshipmenr of the instrumenl.
2-5. MOUNTING.
24. For bench mounting choose a clean, sturdy, uncluttered
mounting surface. For rack mounting, an accessory kit is
available which provides mounting ears. The rack mounting
kit contains the required hardware and instructions.
2-7. POWER REQUIREMENTS.
2%. The instrument has a tapped power transformer and two
line voltage selection switches which permit operation from
100,120,220, or 240 volt + - 10 %, 50 to 400 Hz, single phase
AC power sources.
CAUTION
Always make certain that the line voltage selection
swikhes rue set to rhe correcl positions coresponding
to the voltage of
of
rhe conzct tadng is installed befoe connecting the
instmmenf to any AC power source.
2-9. Set the rear panel line voltage selector switches to the
appropriate positions as indicated in the Line Voltage Select
the
ACpower source, and that a fuse
Line
Voltage
Select Chart
Chart and check that the line fuse is correct for the selected
power source.
2-10.
CABLE CONNECTIONS.
2-11. Front panel connector:
INPUT.
conne&ors and chassis ground allow connection of external
audio signals for analysis. The input impedance is 100 k ohms
either side to ground. The LOW terminal is connected to
chassis ground in the non-floating mode.
OUTPUT. Source output HIGH and LOW BNC type
connectors and chassis ground allow connection to external
devices and components. The output impedance is selected
using special functions 75,76, and 77. The LOW terminal is
connected to chassis ground in the non-floating mode.
2-E Rear panel connectors:
MONITOR.
tor provides a scaled output of the input signal in the level,
frequency and signal-to-noise measurement modes and a
scaled output of the input signal with the fundamental
removed in the distortion and SINAD measurement modes.
The output impedance is 600 ohms.
X
a means of connecting to an external 10 MHz counter reference. The external reference is automatically selected when
a ?TL level signal is present.
SYNC. The SYNC BNC type output connector provides
a lTL compatible signal relative to the source frequency
setting.
XAXIS.The X AXIS BNC type output connector
provides a0 to 5 volt d. c. level relative to the sweep frequency
or level in the sweep mode. The output impedance is loo0
Ohms.
Y AXIS. The Y AXIS BNC type output connector
provides a 0 to 5 volt d. c. level relative to the analyzer
measurement and entered HIGH and LOW plot limits. The
output impedance is loo0 ohms.
Analyzer Input HIGH and LOW BNC type
The MONITOR BNC type output connec-
CLK.
The X CLK BNC type input connector provides
PEN.
The PEN BNC type output connector provides a
‘ITL compatible signal to control the pen of an external
2-l
Section 2
InstaJ.Jation
recorder.
2-13. PRELMNARY CHECK.
2-14. The preJhinary check verifies that the Model 1121 is
operational and should be performed before the instrument
Section 2
KYJ3D legend illuminated. The ANALYZER display wiiJ
contain the D message for one level measurement cycle.
The initiaJize sequence resets aJJ functions and operating
modes of the Model 1121 to the initia.Jized values and conditions listed in Table 2-1.
InstaJJation
Operation
SECTION III
OPERATION
3-1. INTRoDUCTloN.
33. This section contains the operating instructions for the
Model 1121 Audio Analyzer.
3-3. OPERAnNG CONTFlOLS, INDICATORS AND
CONNECTIONS.
3-4. The controls, indiators and eonneetors used during the
operation of the instrument are listed in Table 3-l and shown
in Fies 3-l and 3-2.
3-S. OPERATlNG INSTRUCTlONS.
34. The operating instructions for the Model 1121 are
divided into sections of Initial Conditions, Local Operation
and Remote Operation.
3.7. INlTlAL CONDITIONS.
3-8. Initialize the instrument as follows:
a. Connect the power cord to the instrument and to the
desired power source. Refer to paragraph 2-7 for proper
power application.
b. Set the front panel power switch to ON.
C.
Depress the LCIANIT key.
d. The SOURCE display will contain the instrument
firmware identification number and the other displays will
contain dashes for a period of about two seconds. The
SOURCE display will then contain 1OCNI.00 Hz with the
KYBD legend ill~ted and the ANALYZER display will
change to contain the 0 message for one level measurement cycle.
33. LOCAL OPERATION.
3-10. Function Selection. The DATA ENTRY keypad is
common to all functions of the Model 1121. The KYBD
legend determines the active display window to which the
DATA ENTRY keypad is dedicated at any given time. To
select a functioa simply depress the fun&ion key desired. The
results will be the LED of the function key will be illuminated,
the current value of the selected function will be displayed in
the window above the key, and the KYBD legend will be
illuminated in the display window. The DATA ENTRY
keypad is dedicated to the selected Bmetion and any unit
selection or number entq wig appear in the active display
window. When selecting measurement functions the [
message may appear to indicate that a measurement cannot
be displayed instantly for any of five possible reasons:
The first measurement cycle is in progress and can-
notke displayed
2. The measurements’
are not met, for example, frequency measurements cannot be
made if the input level is too low.
3. The input level is over-range.
The input signal is changing faster than the analyzer
4.
can respond.
5, The notch filter is in the process of beii tuned to
the fundament frequency in the distortion or SfNAD modes.
3-11. Datr Entry Operation. Once a function has been
selected, new values maybe entered with the DATA ENTRY
keypad. To enter data simply depress the desired digit keys
followed by the appropriate unit key or ENTER key. During
digit selection a (‘) mark will appear in the display to the left
of the first digit selected to indicate the number in the display
is in the process of being entered. No action is taken until the
unit or ENTER key is depressed. The unit keys can also be
used aside from number entry to select display modes. For
example, to change the level measurement displayed in mV
to logarithmic units in dBV, simply select the analyzer
LEVEL key and depress the dB key in the DATA ENTRY
keypad. The display program will calculate and display the
logarithmic value. The ENTER key serves a dual function as
a dimensionless unit key for SPCL, ADRS, and PRGM num-
ber entry and also as a default unit terminator of V, %, and
Hz for functions where more than one unit can be selected.
342.
Many
of the Model 1121 functions have multiple display and entry modes. Listed in table 3-2, Function Display
And Data Entry Units, are the display legends which can be
active for each function along with the unit keys in the DATA
ENTRY keypad which select the available display modes.
Argument entry ranges for all the Model 1121 functions are
described in Table 3-3, Valid Function Argument Range.
Number entry out of range of the sehxted functions wig result
in an error displayed in the SOURCE display window. Errors
minimum signal requirements
3-1
Section 3 Operation
. . . ..I‘.
Figure 3-1. Model 1121, Front View.
3-2
Figure 3-2. Model 1121, Rear View.
.A
_” .._ ..-. --l-lll_
Operation
section 3
TARIB X-1.
-_--- -. ---.--___ -,-.---.-- -~~-
or Connector
ANALYZER display
SOURCE d&p&y
BUS’PRGM display
ANALYZER keys
C0NTROL.S. lNDICATORS AND
Control,
Indicator,
FREQW
-IJLey
DIST key
Figure
and Index
No.
CONNECl-ORS.
Function
Displays frequency, AC and DC level, distortion, SINAD, S/N and ratio
measurements. (6 digit LED display)
Displays source frequency and level, frequency and level step sizes,
special function, start, stop, low and high sweep values. Alternately
displays error codes and messages. (8 digit LED display)
Displays current program number or IEEE-488 bus address.
(2 digit LED display)
Selects the following active analyzer functions.
Displayed in Hz or kHz with GATE indication.
Displayed in V, mV, dBV or dBm.
Displayed in %, dB, dFJV, dBm, V, or mV. Notch tune frequency dis-
played in Hz or kHz. Notch capable of automatic or manual tuuing.
SINAD key Displayed in dB. Notch filter is tuned to the source frequency setting.
Displayed in dB. Measured by monitoring the AC level while turning the
source level on and off.
RATIO key
FILTER keys
FLOAT keys
DATA ENTRY keyPad
SOURCE keys
F=Q
W
FREQ STEP key
LEVEL key
LEVEL STEP key
Displays amplitude ratios in % or dB and frequency ratios in Hz or lrHz
Selects optional filters, 30 kHz, 80 m 220 kHz low-pass or DC level
fdters.
Selects floating or single-ended input and output connection.
Used with the function keys to enter data into the active display desig-
nated by the KYBD annunciator.
Selects the following active functions.
Allows display and entry of the source frequency in Hz or kHz units.
Allows display and entry of the frequency increment value in Hz or Wz
units for use with the step keys and the linear frequency sweep mode.
Allows display and entry of the source level in mV, V, dBV or
Allows display and entry of the level increment value in mV, V, or dEi
units for use with the step keys and the linear and log level sweep mode.
dBm units.
--
3-3
section 3
TABLE 31. CONTROLS, INDICATORS AND CONNECTORS CONTINUED.
Control, Figure
Indicator,
and Indelc
Function
or Connector No.
Allows alteration to the normal analyzer modes of operation such as:
range hold, notch tune hold, slow responding detector, and special
modes for test& troubIeshoot& and automatic calibration.
Operation
SWEEP keyis
SWEEP key
START key
STOP key
HIGH key
LINEswitcb
PROGRAM keys
PRGM key
3-1,9
3-1,lO
3-1,ll
3-1,12
Selects the following sweep parameters.
Enables the start of the sweep
Allows display and entry in Hz or kHz units of the starting sweep
frequency or in mV, V, dBV or dJ3m units of the start& sweep level.
Allows display and entry in Hz or kHz units of the ending sweep
frequency or in mV, V, dBV or dBm units of the ending sweep 1eveL
Allows display and entry of the upper Y axis scale value in units
compatible with the active analyzer measurement mode.
Allows display and entry of the lower Y axis scale value in units compatible with the active analyzer measurement mode.
Step keys increment or decrement the active source or sweep function
in frequency or level step sizes and single steps the program location
fiUlCtiOU
Switches the instrument AC power on or off.
Selects the following program functions.
Mows display and entry of the store/recall program location.
sequence
and indicates sweep in progress.
Bus status
ADRS key
LCUINIT key
SRQ
34
STOW
RCL key
key
3-1,x3
3-1,14
3-1,15
3-1,16
Stores the instrument setup at the current program location.
RecaUs the instrument setup at the current program location.
Displays the current IEEE-488 bus status; REM (remote enabled),
LSN (listener addressed), TLK (talker active), and SRQ (service request).
Allows display and entry of IEEE-488 bus address.
Causes the instrument to “go-to-local” when remote enabled otherwise
executes the initialize sequence.
Sets the IEEE-488 bus SRQ line true.
operation
TABLE 31. CONTROLS, INDICATORS AND CONNECTORS CONTINUED.
sectioll3
Control,
Indicator,
or Connector
0uTPuTcon~
INPuTcon-
optluMlconnedors
Ground connector 3-2,2
MONITORconnecW 3-2,3
SYNC connectur
XCLKconncctor
Figure
md Index
No.
3-1,17
34.18
3-2.1
33,4
3-2,s
Function
LOW and
LOW terminal is
LOW and HIGH.
The LOW terminal is
mode.
Used to replace the front panel input and output
Chassis ground connector,
Provides a scaled output of the input signal in the frequency, level and
S/N modes. Provides a scaled output of the input sigual with the fun-
damental removed in the distortion and SINAD modes.
Provides a ‘lTL compatible output relative to the source frequency.
Provides a lTL compatible input for an external 10 MHz timebase
reference. Automatic switching to external reference when present.
HIGH.
Used to umnect the source to mental devices. The
co~etcted
Used to apply an external audio signal for analysis.
co~eckd
to chassis
ground in the non-floating
to chassis ground in the non-floating
co~edors.
mode.
x~COMWt4W
YAxIscoMectur
PEN connecW
IEEE-488 connector 33,9
Fuse holder 3-2,10
I VoMJIge
sele!ctorswitches
AC coo~ector 3-2,12
3.2,6
3-2,7
3-2,s
3-2,11
Provides a 0 to 5 v DC output for plotter application.
Provides a 0 to 5 v DC output for plotter application.
Provides a TIZ compatible output for plotter pen control.
Provides a means for
cable.
AC line fuse holder.
Selects the desired line operating voltage.
AC power connector.
connecting the
standard IEEE-488 bus interface
3-5
Operation
can be cleared by depressing any key. If at any time prior to
entry a wrong digit is entered, depress the CLR key to clear
and restore the previous display.
3-W.
Analyzer Measurement Descriptioa The
1121 contains an independent distortion analyzer which can
measure frequency, AC and DC level, distortion, SINAD and
signal-to-noise. In addition, ratio measurements can be made
with all analyzer measurement modes A wide range of spe-
clal functlorGenha& the basic measurement modes without
sacrificing the simpiiiled operation of the analyzer. Standard
and optional audio filters are provided to aid in harmonic
distortion analysis and weighted noise measurements. Finally, the ability to store and recall specific measurement combinations aid in configmmg measurement applications for
manual and remote use.
3-14. Analyzer input Description.
tion of the Model 1121 can be selected for single-ended or
balanced/different operation. The input mode can be
enabled using the front panel FLQAT key or over the IEEE483 bus interface.
3-15. Frequency Measurement Function.
1121 measures wide ranges of audio frequency with high
accuracy and resolution. Microprocessor control of the
reciprocal counter results in automatic selection of frequency
ranges for maximum resolution. Measurements are refer-
enced to an internal 10 MHz timebase accurate to 0.0001%
and external reference capability is also provided.
3-16.
Frequency Measurement Display Units.
Frequency measurements can be displayed in Hz or kHz for
values above 199.999 Hz with automatic selection of Hz units
below this limit. To select the Frequency measurement function simply depress the FREQ key which illuminates both the
key’s LED and the KYBD legend in the ANALYZER display. Display units can then be selected by depressing the Hz
or kHz keys.
The input cxmligura-
The Model
Model
paragraph 3-74. A period sampling measurement technique
is employed which results ln adaptive measurement rates
optimized to the period of the dominant AC component of
the input @al. This technique yields fast settled measurements in the AC level mode while effectively filtering large
AC components in the DC level mode. Measurement
bandwidth is selectable using the standard and optional filters to reject out-of-band noise or provide industry-standard
weighting characteristics.
3-19. Level Measurement Dirpiay Units.
level measurements can be displayed in linear or logarithmic
units. Linear measurements are displayed in mV or V with V
automatically selected for levels above 750 mV and mV
automatically selected for levels below 0.300 V. Logarithmic
measurements are displayed in various forms. The default
mode uses dBV units (dR relative to l.ooO V rms). Power in
dBm units (dB relative to 1 mW) can be selected for various
impedances as described in paragraph 3-74. To select the
Level measurement function simply depress the LEVEL key
which illuminates both the key’s LED and the KYBD legend
in the ANALYZER display. The various display modes can
then be selected by depressing the appropriate units associated with the desired display mode. For example, to
select AC level in logarithmic units depress the dB key and
to return the display to linear units depress the mV or V keys.
3-20.
Special Level Measurement Modes.
function 11 is provided to preset and hold specific level
ranges to achieve faster first measurement rates and to
eliminate possible range-to-range nonlinearity. Refer to
paragraph 3-65. Special function 17 extends the measurements sampling period to provide a more consistent reading
in the presence of noise. AC and DC calibration is performed through the use of special function codes 2tI through
24. The rms, average or quasi-peak AC detector type can be
selected using special function 70, 71 or R respectively.
,Speciai functions 80 through 86 select logarlthmlc display
modes in dBV or dBm units.
AC and DC
Special
3-17. Special Frequency Measurement Modes.
Special function 11 is provided to preset and hold specific
level ranges for frequency measurement. Refer to paragraph
3-65. The frequency measurement mode can function to
input levels 14 db below the selected level range. For example
by setting the 3,ooO volt level range, measurements can be
made with signal levels as low as 600 mV.
3-18. Level Measurement Function.
measures both AC and DC voltage with high dynamic range
and selectable AC bandwidth. Resolution at full scale is 3000
counts with an additional 33 % overrange capability. The AC
rms detector is true rms responding for signals with crest
factors less than 3. Average and quasi-peak responding
detectors (rms calibrated) can also be selected. Refer to
The Model 1121
3-6
3-21. Distortion Measurement Function. The Model
1121 measures total harmonic distortion and noise over a
wide range of frequency. The notch filter is automatically
tuned to reject the fundamental frequency and pass only the
harmonic and noise content..The AC measurement technlques are similar to those used iu the level measurement
function yielding fast settled measurements. Measurement
bandwidth is selectable to reject noise while accurately
preserving harmonic components. Measurement results can
be displayed in several forms. The combination of harmonics
and noise can be displayed as an absolute level in mV, V, dBV
or dBm units or as a ratio in % or dB u&s to the total input
signal consisting of fundamental, harmonics and noise.
Operation
Section 3
3-22. Distartkm Merwrement Display Units.
Distortion measurements can be displayed in linear or
bgadhmic units. Linear ratiometric measurements are displayed in 5% while logarithmic measurements are displayed
in dB where 0.00 dB is referenced to 100.0 %. Distortion
measurements can also be displayed as an absolute level in
units of mV, V, dBV or dBm. The distortion level display
mode is useful to measure noise level in the presence of a
TABLE 3-2. FUNCTION DISPLAY AND DATA ENTRY UNITS.
holding tone. The holding tone is removed by the notch filter
and the noise level alone is measured and displayed. In
addition the frequency of the notch tilter can be displayed in
Hz or kHz units. To select the distortion measurement function simply depress the DIST key which illuminates both the
key’s LED and the KYBD legend in the ANALYZER display. The various display modes can then be selected by
depre4sing the appropriate units associated with the desiied
Function Display Unit
Legends Keys
ANALYPR GROUP:
FREQ
LEVEL
DIST
SINAD
Hz kI4z
mVVdBVdBm
mV V Hz k.Hz
%dBdBVdBm
d.B
Default Units
(ENTER Key)
mVVHzkHz V
mV V dJ3 V
mVVHzkHz %
%dB
mV V dB dB
S/N
RATIO
SOURCE GROUP:
PREQ
FREQ STEP
LEVEL
LEVEL STEP
SPCL
START
STOP
LOW
HIGH
BUWPRGM GROUP:
dJ3
%dB
HzkHz
HZkHZ
mVVdBVdBm
mV V dB
SPCL
mV V Hz kHz
dBV dBm
mVVHz kHz
dBV dBm
mV V Hz kHz
%dBdBVdBm
mVV Hz kHz
%dBdBVdBm
dB
%dB no entry
HzkHi Hz
HzkHz Hz
mV V dB
mV V dB V
ENTER dimensionless
mV V Hz kHz Hz
dB
mVVHzkHz
dB
mVVHzkHz V.
%dB
mVVHzkHz V
%dB
no entry
V
Hz
ADRS
PRGM
ADRS
PRGM
ENTER dimensionless
ENTER dimensionless
3-7
sectioll3
TABLE 3-3. VALID FUNCTION ARGUMENT RANGE.
Operation
Function
ANALYZER GROUP:
FREQ
LEVEL
DIST
SINAD
S/N
RATIO
SOURCE GROUP:
FREQ
PREQ STEP
LEVEL1
LEVEL STEP
SPCL
START’
STOP’
LOW
HIGH
Argument Entry Error
Range
OmVto3OOV
OmVto3OOV
OmVto3CMV
OtolOO%
-140 to 0.0 dB
5HztoMOkHz
OmVto3OOV
oto14OdB
no entry allowed
no entry allowed
10HztolsOkHz Set source frequency 01
OHztol5OkHz Set frequency increment 02
0 mV to 16 V Set open circuit output level (lin) 03
-140 to 24.08 dBv Set open circuit output level (log) 03
OmVto16V Set level increment (lin) 04
oto14OdB Set level increment (log) 04
See TABLE 3-7 See TABLE 3-7 05
10 Hz to 150 kHz Set sweep start frequency 06
OmVto16V
-140 to 24.08 dBv Set sweep start level (log) 06
10Hztol5OkHz Set sweep stop frequency 07
OmVto16V Set sweep stop level (lin) 07
-140 to 24.08 dsv Set sweep stop level (log) 07
-3ooto3alv set low plot limit (lin) 08
-3ooooto3oooo% Set low plot limit (lin) 08
-140 to 49.54 dB Set low plot limit (log) 08
-49.54 to 140 dB Set low plot limit (S/N, SINAD) 08
-3ooto3ookHz Set low plot limit (lin) 08
-3ooto3ooV Set high plot limit (lin) 09
-3ooooto30000% Set high plot limit (lin) 09
-140 to 49.54 dB Set high plot limit (log) 09
-49.54 to 140 dB Set high plot limit (S/N, SINAD)
-3ooto3ookHz set high plot limit (lin) 09
Action
Set input level range
Set input level range
Set input level range
Set distortion range (Iin)
Set distortion range (log)
Set notch tune frequency
Set input level range
Set SLNAD range (log)
Set sweep start level (lin) 06
No.
12
l3
14
14
14
14
15
15
20
17
09
BUS/PRGM GROUP:
ADRS oto30 Set IEEE-488 bus address 10
PRGM oto99
NOTE 1 Amplitude values can be entered and disphyed in dBm units. The alhwabk argument range for dBm values varies bawd
on the source and load impedance scttinp and is limited by the open-circuit voltage range.
3-8
Set store/recall location 11
Operation
section 3
display mode. For example, to select distortion level in
logarithmic units (dBV or dRm) depress the DIST key followed in sequence by the mV or V keys aad the dR key. The
mV or V keys select the distortion measurement to be displayed as aa absolute level aad the dR key converts the results
to logarithmic units.
3-23. Special lMsto&m
fimctioas 12 aad I3 are provided to preset aad hold specific
input level and distortion ranges. Refer to paragraph 3-65.
Special function 17 extends the measurements sampling
period to provide a more consistent reading ia the presence
of noise Using special function 14, notch filter tuaiag can be
held at specific frequencies to aid in tuning the notch filter in
the event that a stable frequency measurement caa not be
achieved or to enable notch 6lter tuning to frequencies other
than the fuadarneatal. As with all amplitude measurement
fuactioas, the rms, average, or quasi-peak detector type caa
be selected using special function 70,71 or 72, respectively.
Special functions 80 through 86 select logarithmic distortion
level display traits in dBV or dRm.
3-24. SINAD Measurement Function.
measures SINAD (signal-to-noise and distortion) in the same
manner as the distortion measurement except that the notch
filter is tuned and held at the source frequency to permit
measurements in the presence of large amounts of noise. If
an external oscillator is used, it must be tuned to within 3 %
of the source frequency setting. The AC measurement tech-
niques are similar to those used ia the level measurement
function yielding fast settled measurements. Measurement
bandwidth is selectable to reject noise while accurately
preserving harmonic components. The combination of har-
moaics aad noise is displayed as a percentage of the total
input sigaal consisting of fundamental, harmonics aad noise
3-25. SINAD Measurement Display Units.
measurements are only displayed ia dB uaits. To select the
SINAD measurement function simply depress the SINAD
key which illtmriaates both the key’s LED and the KYBD
legend in the ANALYZER display.
3-26. Special SINAD Measurement Modes. S
functions 12 aad 13 are provided to preset and hold specific
input level and SINAD ranges. Refer to paragraph 3-65.
Special function 17 extends the measurements sampling
period to provide more consistent readiags in the presence
of noise. When measuring large amounts of noise (0 to lOdI3
SINAD), the notch filter tuaing may become indeterminate
and causes the [ message to be displayed. Special
function I.5 is designed to disable the [x message ia this
circumstance. As with all amplitude measurement functions,
the rms, average, or quasi-peak detector type can be selected
using special fuactioa 70,71 or 72, respectively.
Measufement
Modes. Special
The Model 1121
SINAD
p e c i a 1
3-27. S/N Measurement Function.
measures S/N (signal-to-noise) by alternately tumiag the
source output on aad off aad displaying the ratio of the two
measurements. The
those used by the level measurement mode. Measurement
baadwidth is selectable using the filters to reject out of band
noise or provide iadustry staadard weightiag characteristics.
measurement
3-28. S/N Measurement Display Unlts.
ments are oaly displayed ia dB traits. To select the S/N
measurement function simply depress the S/N key which
ilhmiaates both the key’s LED aad the KYBD legend ia the
ANALYZER display.
The Model 1121
techniques are similar to
S/N measure-
3-29. Special SIN Meanurement Modes. Special func-
tioa 17 extends the measurements sampling period to provide
more consistent readiags ia the presence of noise. Special
fuactioas 40 through 49
delay between the signal measurement and noise measuremeat to allow time for the device under test to respond to the
change ia amplitude. As with all amplitude measurement
functions, the rms, average, or quasi-peak detector type caa
be selected usiag special functions 70,71, aad 72.
3-30.
Using the Ratio Mode.
measurement modes to be displayed as a relative value to a
previous measurement value. Ia a ratiometric measurement,
such as a flatness response, amplitude measurements at
various frequencies are displayed relative to a reference level
at a frequency of 1 kH2. Another example of a ratiometric
measurement is the measurement of the percent of AC ripple
on a DC level.
331. Ratio Measurement Display Units.
level measurements are displayed ia % or dI3 units whereas
relative frequency measurements are displayed ia Hz or kHz.
To select the ratio mode simply depress an aaalyzr measuremeat key such as LEVEL followed by the RATIO key. The
LEDs of both keys will be illuraiaated along with the KYRD
legend in the ANALYZER display. When the next measuremeat cycle is complete, the measurement value will become
the ratio reference aad the display will iadicate 100.00 %, 0.00
dB or O.tXIO Hz depending on the previous log&ear display
mode. Selection of display units can be made by depressing
either the %, dB, kHz or Hz keys. On subsequent measuremeat cycles the results wilI be displayed relative to the
origiaal ratio reference. The RATIO key is aa alternate
action key, therefore, depressing the RATIO key again will
deselect the ratio mode, extinguish the key’s LED and return
the display to the normal measurement mode. If another
measurement function is selected while the ratio mode is
active, the LED on the RATIO key will be extiagGhed but
the ratio reference is preserved for the origiaal measurement
function aad caa be reactivated by depressing the origiaal
measurement function key. The ratio mode is Gaited to only
are used to program the amount of
The Model 1121 enables all
Rat iomet ric
3-9
section 3
TABLE 3-4. INF’UT LEVEL, RANGES.
AC Level Ranges: DC Level Ranges: Distortion and SINAD
Input Level Ranges
operation
300.0 to 150.1 v
150.0 to 75.1 v
75.00 to 30.1 v
30.00 to 15.01 v
15.00 to 7.51 v
7.500 to 3.01 v
3.000 to 1.501 v
300.0 to 150.1 v
150.0 to 75.1 v
75.00 to 30.1 v
30.00 to 15.01 v
15.00 to 7.51 v
7.500 to 3.01 v
3.000 V and below
1500 to 751 mV
750.0 to 301 mV
300.0 to 150.1 mV
150.0 to 75.1 mV
75.00 to 30.1 mV
30.00 to 15.01 mV
15.00 to 7.51 mV
7.500 to 3.01 mV
3.000 to 1.501 mV
1.500 to 0.751 mV
0.7500 to 0.301 mV
0.3000 mV and below
one reference value and the old ratio reference is lost if the
ratio mode is activated in an alternate measurement function.
3-32.
Using A~ly~er Filter+. The
ternate action keys which means they are “toggled” on and off
with each key stroke. The optional filters are mutually exclusive; therefore, depressing one of the keys will cancel the
other. The same is true of the low-pass filters, only one
low-pass filter can be used at a time. The DC filter, however,
is mutually exclusive with all filters and will also be canceled
by selecting any measurement functions other than Level.
3-33. The minimum bandwidth consistent with the measure-
ment bandwidth should be used to minim& noise errors. For
example, when measuring the distortion of a 1 kHx fun-
damental tone, the 30 kHx low-pass tilter is recommended.
The DC low-pass filter is provided to attenuate all AC components and measure DC level directly. The DC low-pass
filter can only be activated in the analyxer level mode. Bandpass lilters are combiitions of high- and low-pass filters and
are used in some measurements to simulate the sensitivity of
the human ear to the audible frequency spectrum. High-pass
filter selection is used to eliminate power line harmonics
when present. The 400 Hz high-pass filter typically provides
audio filter keys are al-
300.0 to 150.1 v
150.0 to 75.1 v
75.00 to 30.1 v
30.00 to 15.01 v
15.00 to 7.51 v
7500 to 3.01 v
3.000 to 1.501 v
1500 to 751 mV
750.0 to 301 mV
300.0 to 150.1 mV
150.0 to 100.1 mV
100.0 to 50.1 mV
50.0 mV and below
more than 80 dB of attenuation at 60 Hz.
3-34. The CCIR filter is a special case since its application
depends on whether the average or quasi-peak detector is
selected. The CCIR r ecommendation 468-3 specifies the
filter should be used with the quasi-peak detector. Another
application referred to as the CCIR/ARM calls for using the
average responding meter (ARM) and applying a gain correction value of -6.6 dB. The Model 1121 wiU accommodate
both applications. The average
selected using special
readings will also be adjusted by -6.6 dB automatically when
the average detector is selected.
3-35.
Audio Oscillator
Model 1121 contains a variable frequency, low distortion
audio oscillator and a precision progranunable amplifier/attenuator. The oscillator frequency tuning uses a frequency
lock technique under microprocesso
accuracy and resolution. The output level can be varied in
very fme increments over a wide range of levels. Levels can
be set in either linear or logarithmic units to accommodate
existing test procedures and applications. The oscillator incorporates a user confiigurable sweep mode which can be
function
or
quasi-peak detector can be
71 or 72 respectively. The level
General
Descripth.
r control to yield high
The
3-10
Operation
TABLE 3-5. DISTORTION AND SINAD RANGES.
section 3
Linear:
100.0 to 50.01%
50.00 to 20.01%
20.00 to 10.01%
10.00 to 5.001%
moo to 2.001%
2.000 to 1.001%
1.000 to 0.5001 %
o.moo to 0.2001 %
0.2000 to 0.1001 %
0.1000 to 0.05001%
0.05000 to 0.02001 %
0.02000 to 0.01001%
0.01000 % and below
programmed to sweep frequency or level in logarithmic or
linear increments.
3-34. Oscillator Output Description.
figuration of the Model 1121 can be selected for single-ended
or balanced/floating operation using the FLOAT key. Special
functions 75,76, and 77 select the respective 50 ohm, IS0 ohm,
and 600 ohm source output impedance.
The output con-
Logarithmic:
0.00 to -6.01 dB
-6.02
to
-1T.97 dB
* -13.98 to -19.99 dB
-20.00 to -26.01 dB
-26.02 to -33.97 dB
-33.98 to -39.99 dB
-40.00 to 46.01 dB
46.02 to -53.97 dB
-53.98 to -59.99 dB
-60.00 to -66.01 dB
-66.02 to -73.97 dB
-73.98 to -79.99 dB
-80.00 dB and below
3-39. When a frequency setting is entered the oscillator is
coarse tuned to the setting by the control program. Thereafter, the internal frequency of the oscillator is measured and
fme adjustments are made by the control program to tune the
oscillator to within tolerance. Special function 16 disables the
frequency lock mode to enable faster frequency updates and
sweep rates where the full frequency accuracy of the oscillator is not required.
3-37. Source Frequency Display and Selection.
select the Frequency hrnction simply depress the FREQ key.
The key’s LED and the KYBD legend will be illuminated and
the SOURCE display window will contain the current frequency setting. Once the function is selected a new frequency
may be entered using the DATA ENTRY keypad. Display
units can be selected by depressing either the Hz or kHz keys.
3.38, Source Frequency Lock Mode Description.
Model 1121 source oscillator achieves high frequency accuracy and resolution through the use of a frequency lock
technique. The design of the Model 1121 enables internal
measurement of the oscillator frequency. During initialization after power is applied or when the LCIANIT key is
depressed, the oscillator is sequenced through its live frequency bands and specific frequencies settings are verified
for accuracy. Tuning errors are stored by the control program
in internal memory as calibration factors. If the oscillator fails
to function on any of the frequency bands, an error (30-34)
will be displayed as an indication of a hardware fault requir-
ing service.
To
The
3-40. Source Level Display And Selection.
the Level function simply depress the LEVEL key. The key’s
LED and the KYBD legend will be illuminated and the
LEVEL display window will contain the current level setting.
Once the function is selected a new level maybe entered using
the DATA ENTRY keypad. The level function allows display
and selection of the open-circuit output level. When a level
is selected, the programmable amplifier/attenuator is configured by the microprocessor circuits to one of the ranges
listed in Table 3-6. Special function 18 disables the automatic
range selection and holds the range that is active at the time
the special function was enabled. When the range-hold mode
is active, new levels may be entered from 0 V up to the
maximum value for the selected range. This mode provides
the abiity to set levels over a wide dynamic range without the
possible delay and transients associated with range to range
transitions. The LEVEL function can be incremented or
decremented by the value in the LEVEL STEP function.
341. Amplitude Display Units.
LEVEL, START and STOP functions may be displayed in
linear units of mV or V or as a logarithmic value in units of
dBV or dBm. Display units can be selected by depressing
Amplitude values for the
To select
3-11
Section
3
Operation
either the mV, V, or dB keys. Special functions 80 through 86
select the logarithmic display mode where 80, dRV, is the
default selection. The dBV display mode converts the linear
voltage display (open-circuit) to a logarithmic value relative
to 1,000 V rms. When a dBm mode is selected, the actual
power into the load is calculated and displayed based on the
open-circuit level and source and load impedance selections,
342.
Step Size and Step Key Operation. The keys
marked with up and down arrows are used to increment and
decrement source and program function values in the active
window designated by the KYBD legend. If an arrow key is
constantly depressed, the stepping will repeat. Functions
which can be stepped are PRGM, FREQ, START, STOP,
and LEVEL. Stepping any other function has no effect.
343. Increment/Decrement Program Number. The
Program number can be stepped in single increments. A
special Auto-recall function is provided which automatically
executes the recall operation after the program location function has been incremented or decremented with the step keys.
This function allows rapid recall of sequential program
setups by using a single key or bus mnemonic. The auto-recall
function can be enabled or disabled using the internal option
switch A4Sl-5 or special functions 7 and 8.
corder. The analyzer measurement provides the Y axis information. The Y axis scale can be set using the HIGH and LOW
plot limit functions.
347. SWEEP Key Description. The SWEEP key is an alternate action key which initiates and terminates the sweep.
The key’s LED indicates a sweep is in progress and will be
illuminated for the period of one sweep.
348. START Key Description. The START key allows
entry of either the frequency or level START value. The
START value defines the origin of the X axis which corresponds to 0 volts at the X AXIS output connector. When
the START key is depressed the SOURCE display will contain the START value and 0 volts is presented at the X AXIS
output connector as an aid in setting the offset adjustment on
most X-Y plotters.
349.
STOP Key Description. The STOP key allows entry
of either the frequency or level STOP value. The STOP value
defmes the end of the X axis which corresponds to 5 volts at
the X AXIS output connector. When the STOP key is
depressed the SOURCE display will contain the STOP value
and 5 volts is presented at the X AXIS output connector as
an aid in setting the sensitivity adjustment on most X-Y
plotters.
3-M. Increment/Decrement Frequency Values.
Frequency values in the FREQ, START, and STOP functions
can be stepped by the value in the FREQ STEP function.
Available frequency step sixes are listed in Table 3-3. If the
step size is smaller than the resolution of the selected func-
tion, the function will be stepped by the resolution value. The
frequency step size also functions as the step size for the
linear frequency sweep mode.
34. Increment/Decrement Level Values.
Amplitude
values in the LEVEL, START, and STOP functions can be
stepped by the value in the LEVEL STEP function. Level
step sixes can be linear or logarithmic values. Available level
step sixes are listed in Table 3-3. If the step size is smaller than
the resolution of the selected function, the function will be
stepped by the resolution value. Table 3-6 lists the output
level ranges and resolution. The level step size also functions
as the step size for the linear and logarithmic level sweep
modes.
346. Sweep Mode General Description. The Model
1121
provides a user programmable sweep capability which
simplifies time consuming measurements such as flatness,
distortion vs. power output and compression/expansion
linearity.
In
the sweep mode the
source
of
the
Model
1121 is
used as the stimulus and can be configured to sweep frequency or level in linear or logarithmic steps. The source provides
the X axis and pen control signals for an external plotter/re-
3-50. X AXIS Output Description. The X AXIS output of
the Model 1121 is the scaled result of the sweep frequency or
level relative to the START and STOP values. The Model
1121 can be swept in level or frequency with 4096 points of
resolution on the X axis. Selection of a level or frequency
sweep is determined by the type of START and STOP values
entered: START and STOP level entries designate a level
sweep and START and STOP frequency entries designate a
frequency sweep.
3-51. Y AXIS Output Description. The Y AXIS output of
the Model 1121 is the scaled result of the analyzer measurement value relative to the HIGH and LOW values. There are
4096 points of resolution between 0 and 5 volts on the Y axis.
Any measurement mode can be used for sweeping. Various
measurement modes and applications require greater measurement settling before a data point is generated. The Model
1121 allows for user configurable settling times using special
functions 61 through 69. The settling time is based on the
number of consecutive measurement cycles achieved before
a data point is plotted. Special function 60 disables the
analyzer measurement and the Y AXIS output and allows the
source to be used as a rapid incremental sweep generator
with X AXIS and PEN control.
3-52. The HIGH and LOW plot limits are always displayed
in the same units as the analyzer measurement mode. The Y
AXIS output will be scaled linearly or logarithmically
3-12
Loading...
+ 137 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.