1. Safety
Always observe the following safety precautions:
• Use only as specified by the operating instructions or the intrinsic
protection may be impaired. After transport or storage in humid
conditions, dry out the unit before connecting it to the supply voltage.
During drying out the intrinsic protection may be impaired.
• Connect only to a power supply with a voltage corresponding to that on
the serial number label.
• Connect only to a power supply that provides a safety ground terminal.
• Before moving, disconnect at the power supply socket.
• Do not check the temperature by touch, but instead use the thermometer.
• To reduce the risk of eye injury during high temperature operation, use
safety goggles or spectacles.
• Do not touch surfaces that become hot during high temperature operation.
• Insure that the operating temperature is less than the maximum operating
temperature of your sample material.
• Insure that the power switch is easily accessible during use.
• If liquid is spilled inside the unit, disconnect it from the power supply and
have it checked by a competent person.
• It is the user’s responsibility to carry out appropriate decontamination if
hazardous material is spilled on or inside the equipment.
• The responsible body shall be made aware that, if the equipment is used
in a manner not specified by the manufacturer, the protection provided by
the equipment may be impaired.
2
2. Product Information
132000 Incubator
• Operating Instructions
Two Shelves
Eight Shelf Brackets
The Boekel series of Incubators are designed for laboratories and clinics that
require contemporary styling, accuracy and economy in a basic incubator package.
Boekel’s 132000 Incubator provides excellent incubator performance at the most
economical cost. This Incubator uses a dial controlled thermostat for temperature
control and provides a top enter ferrule entry point for thermometer insertion. The
132000 incubator has a chamber volume of 0.8 cu. ft. (0.02 cu. m) and a chamber
temperature range of ambient +5˚C to 60˚C. The incubator comes with two
movable shelves. Additional shelving may be purchased.
3. Assembly
3.1 Unpacking
Remove packing materials carefully, and retain for future shipment or storage of
the unit. Inspect for damage. Report all shipping damage to the carrier
immediately. Shipping damage is covered by the carrier and repair/replacement
for shipping damages must be coordinated through the carrier. Complete and
return the Warranty Registration Card. Pack should contain:
•
•
•
3.2 Installation
Place the Incubator on a flat and stable surface, preferably away from drafts.
Insure that the surface on which the unit is placed will withstand the radiated heat
produced by typical laboratory incubators. Plug power cord into a power supply
that matches the voltage listed on the serial number label on the rear of the unit.
Insert the wire rack shelves at the desired heights. Refer to the sketch below for
proper shelf bracket installation for the Incubator.
3
Power
Switch
OI
Heater
Lamp
MODEL 132000
HEAT
0
1
2
3
4
5
6
7
8
9
10
Temperature
Adjust Knob
132000 SHELF INSTALLATION DIAGRAM
STEP #1
SHELF
BRACKET
INNER
WALL
STEP #2
STEP #3
SHELF
INNER
WALL
4. Operation
4.1 Controls and Indicator Lamps (see Figure below)
The power switch controls power to the unit.
The heater lamp indicates when the heater is operating.
The temperature adjust knob is used to control the chamber temperature.
90°
INNER
WALL
INNER
WALL
4
4.2 Setting the Temperatur e
Use the temperature adjust knob to set the desired temperature. The knob is
rotated clockwise to increase the chamber temperature and counter clockwise to
decrease the chamber temperature.
4.3 Operating the Incubator
To turn the incubator on, switch the power switch to the on (I) position. Set the
desired chamber temperature. The heater lamp will illuminate and remain lit until
the chamber reaches the set temperature. After reaching the set temperature,
the heater lamp will cycle on and off with heater operation. Allow a 60 minute
chamber temperature equilibration when starting the incubator from a cold start.
Allow at least 15 minutes for the re-equilibration when changing temperatures.
To turn the incubator off, switch the power switch to the off (O) position or unplug
the unit. If the temperature adjust knob is not moved when the incubator is
powered down, the incubator will return to its previously set temperature upon
restart.
5. Accessories
5.1 Removable Shelf – Catalog Number C1902053
5.2 Shelf Bracket – Catalog Number A1903208
5
6. Fault Diagnosis
Symptom
Possible Cause
Action Required
1. Unit does not operate
a. Unit not switched on
a. Switch on
2. Chamber temperature
a. Actual temperature is
a. Check set temperature
3. Temperature contin ues
a. Actual temperature is
a. Check set temperature
does not rise when
expected
to rise when not
expected
b. Unit not plugged into
power supply
c. Power supply failure
higher than set
temperature
b. Temperature control
circuit fault
lower than set
temperature
b. Temperature control
circuit fault
b. Plug in, switch on
c. Check that other
electrical appliances on
the same circuit are
working
b. Have unit checked by
competent person
b. Have unit checked by
competent person
7. Technical Specifications
This equipment is intended for indoor use and will meet its performance figures within
the ambient temperature range of 10°C to 35°C, with maximum relative humidity of 80%
(non-condensing). Installation Category II (transient voltages). Pollution Degree 2 in
accordance with IEC 664. Suitable for operation at altitudes of up to 6500 feet.
Specifications:
132000 Incubator
Temperature Range: (Ambient +5°C) to 60°C
Supply Voltage Range: 120V +/- 10%, 50/60 Hz
230V +/- 10%, 50/60 Hz
Power Rating: Model 132000: 90W Model 132000-2: 90W
6
8. Maintenance and Service
All Boekel laboratory products are designed to comply with IEC1010-1. No routine
maintenance is required. There are no user serviceable parts in this product.
8.1 Cleaning
Disengage power cord prior to cleaning. If a spill occurs, use appropriate clean
up procedures as required for radiation or biohazard control. The outer casing
may be cleaned with water and a damp cloth. Do not submerge or immerse the
Incubator in water. Before using any cleaning or decontamination method, users
should check that the proposed method would not damage the equipment.
9. Warranty
When used in laboratory conditions and according to these operating instructions,
Boekel warrants this product to be free of defective material and workmanship for a
period of two years from the date of manufacture. The liability of Boekel for any
defective equipment during the warranty period shall be limited to the repair of
such equipment or replacement thereof without charge for parts or labor. Boekel
Scientific’s warranty provided here under shall be null and void and without further
force or effect if there is any: (1) Repair made to the product by a party other than
Boekel Scientific, or its authorized service representative. (2) Misuse (including
use inconsistent with written operating instructions for the product), mishandling,
contamination, overheating, modifications or alteration of the product by any
customer or third party. (3) Use or replacement parts that are obtained from a
party who is not an authorized distributor of Boekel Scientific.
10. Service
It is required to obtain a Returned Material Authorization (RMA) number before any
Boekel products are returned for any reason. A Decontamination Certificate must
be completed, signed by the user, and returned to Boekel Scientific prior to
receiving the RMA number. Please be sure to mark the outside of the returned
goods package with this RMA number to ensure prompt handling.
Boekel Scientific
855 Pennsylvania Boul evard
Feasterville, PA 19053
Phone: (215) 396-8200 or (800) 336-6929
Fax: (215) 396-8264
e-mail: boekel-info@boekelsci.com
7
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.