Blancett B2800 Operating Manual

Page 1
Flow Monitor
B2800XP Explosion Proof Flow Monitor Standard
DSY-PM-00277-EN-03 (March 2017)
User Manual
Page 2
Flow Monitor, B2800XP Explosion Proof Flow Monitor Standard
CONTENTS
Scope of This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Unpacking and Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Terminology and Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Explosion Proof Enclosure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Installation Kits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Operating the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Modes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Programming. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Programming Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Run Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Additional Scaling Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Additional Input Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Battery Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Troubleshooting Guide. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Specications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Part Number Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Page ii March 2017
Page 3

Scope of This Manual

SCOPE OF THIS MANUAL
This manual is intended to help you get the B2800 flow monitor up and running quickly.
MPORTANTI
Read this manual carefully before attempting any installation or operation. Keep the manual accessible for future reference.

UNPACKING AND INSPECTION

Upon opening the shipping container, visually inspect the product and applicable accessories for any physical damage such as scratches, loose or broken parts, or any other sign of damage that may have occurred during shipment.
OTE:N If damage is found, request an inspection by the carrier’s agent within 48 hours of delivery and file a claim with the
carrier. A claim for equipment damage in transit is the sole responsibility of the purchaser.

SAFETY

Terminology and Symbols

Indicates a hazardous situation, which, if not avoided, is estimated to be capable of causing death or serious personal injury.
Indicates a hazardous situation, which, if not avoided, could result in severe personal injury or death.
Indicates a hazardous situation, which, if not avoided, is estimated to be capable of causing minor or moderate personal injury or damage to property.

Considerations

The installation of the B2800 monitor must comply with all applicable federal, state, and local rules, regulations, and codes.
EXPLOSION HAZARD - SUBSTITUTION OF COMPONENTS MAY IMPAIR SUITABILITY FOR CLASS I, DIVISION 2.
AVERTISSMENT
RISQUE D’EXPLOSION - LA SUBSTITUTION DE COMPOSANTS PEUT RENDRE CEMATÉRIEL INACCCEPTABLE POUR LES EMPLACEMENTS DE CLASSE I, DIVISION 2.
DO NOT CONNECT OR DISCONNECT EITHER POWER OR OUTPUTS UNLESS THE AREA IS KNOWN TO BE NON­HAZARDOUS.
AVERTISSMENT
RISQUE D’EXPLOSION. NE PAS DÉBRANCHER TANT QUE LE CIRCUIT EST SOUSTENSION, À MOINS QU’LL NE S’AGISSE D’UN EMPLACEMENT NON DANGEREUX.
MPORTANTI
Not following instructions properly may impair safety of equipment and/or personnel.
Page 3 March 2017 DSY-PM-00277-EN-03
Page 4

Introduction

INTRODUCTION
The B2800XP flow monitor is a state-of-the-art digital signal processing flow monitor designed to provide the user with exceptional flexibility at a very affordable price. Though designed for use with Blancett flow meters, this display can be used with almost any flow meter producing a low amplitude AC output or contact closure signals.
The B2800XP flow monitor is capable of accepting a low level frequency input for calculating flow rate and total. These calculations can then be displayed in the desired units of measurement. The monitor’s large eight digit by 0.50 inch (12.7 mm) numeric liquid crystal display makes extended range viewing practical. The second eight digit by 0.25 inch (6.35 mm) alphanumeric display provides for selectable units viewing in run mode and prompts for variables in program mode. The display can show rate, total or alternate between both rate and total.
MENU ENTER
Programming
Buttons

EXPLOSION PROOF ENCLOSURE

The EIH Instrument enclosure is designed to house instrumentation and control equipment as well as act as a conduit outlet body in hazardous, abusive and wet locations.
The EIH enclosure is approved by Underwriters Laboratories Inc., Canadian Standards Association, Factory Mutual and CENELEC for use in Class I, Groups B*, C** and D; Class II, Groups E, F and G; and Class II hazardous (classified) locations as defined by the National Electrical Code and Canadian Electrical Code. It is also NEMA/UL/CSA Type 4 and IP66 rated for watertight applications.
*With conduit seals installed within 18 inches of enclosure. **Unsealed conduit lengths must not exceed five feet (152 cm).
Page 4 March 2017 DSY-PM-00277-EN-03
Page 5

Installation

INSTALLATION
ELECTRICAL POWER MUST BE TURNED OFF BEFORE AND DURING INSTALLATION AND MAINTENANCE.
1. EIH Instrument enclosures are furnished with 3/4 inch NPT oset through feed cast hubs for conduit entries. (Use Cooper Crouse-Hinds RE21-SA to reduce to 1/2 inch hubs.)
2. Secure the enclosure to the conduit system. If the enclosure has mounting feet, select a mounting location that will provide sucient strength and rigidity to support the enclosure as well as the enclosed device and wiring.
SELECT A MOUNTING LOCATION SO THAT THE ENCLOSURE WILL NOT BE SUBJECTED TO IMPACT BY HEAVY OBJECTS. IMPACTS CAN DAMAGE ENCLOSED DEVICES OR GLASS LENS.
3. Install Cooper Crouse-Hinds EYS sealing ttings required by Section 501-5 and/or 502-5 of the National Electrical Code and Section 18 of the Canadian Electrical Code or any other applicable local codes and when enclosure is installed in Class I Group B hazardous locations. (For CSA Group C applications, unsealed conduit lengths must not exceed ve feet or 152 cm.)
THE HAZARDOUS LOCATION INFORMATION SPECIFYING CLASS AND GROUP LISTING OF EACH INSTRUMENT ENCLOSURE IS MARKED ON THE NAMEPLATE OF EACH ENCLOSURE.
ALL UNUSED CONDUIT OPENINGS MUST BE PLUGGED. PLUG UNUSED CONDUIT OPENINGS WITH COOPER CROUSE­HINDS PLG2. PLUGS MUST BE A MINIMUM OF 1⁄8 INCH THICK AND ENGAGE A MINIMUM OF FIVE FULL THREADS.
4. Unthread instrument (and power side) covers and carefully set aside to prevent damage to the cover threads and glass lens (when glass lens cover is used).
5. Pull wires into enclosure making certain they are long enough to make the required connections and to remove the instrument or power supply if servicing is required. Install instrument and power supply, if applicable, and make all electrical connections.
OTE:N When installing device, be sure to check instrument dimensions to avoid interference with clamping ring on glass
lens and the cover on standard units.
6. Test wiring for correctness by checking continuity and also check for unwanted grounds with insulator resistance tester. Make sure test equipment being used will not damage instrument to be housed in the EIH Instrument enclosure.
7. Carefully re-thread cover to enclosure housing. Tighten cover until cover ange contacts body face.
USE CARE TO PREVENT DIRT, GRIT OR OTHER FOREIGN MATERIAL FROM LODGING ON THREADS. IF SUCH MATERIAL SETTLES ON THREADS, CLEAN WITH KEROSENE OR STODDARD SOLVENT*, THEN RE-LUBRICATE WITH COOPER CROUSE-HINDS TYPE STL THREAD LUBRICANT.
8. Tighten cover set screws to prevent cover from loosening under vibration.
* TO AVOID THE POSSIBILITY OF AN EXPLOSION, OXIDATION AND CORROSION, DO NOT USE GASOLINE OR SIMILAR SOLVENT.
TO MAINTAIN THE EXPLOSION PROOF INTEGRITY OF THE ENCLOSURE WITH A SCREW IN A TAPPED MOUNTING PAD HOLE, THERE MUST BE A MINIMUM OF 1⁄16 INCH OF MATERIAL BETWEEN THE DRILL POINT AND THE BACK WALL. IF FOR ANY REASON A SCREW WILL NOT BE THREADED INTO THE DRILLED HOLE, A MINIMUM OF 1⁄8 INCH OF MATERIAL MUST REMAIN BETWEEN THE DRILL POINT AND THE BACK WALL.
Page 5 March 2017 DSY-PM-00277-EN-03
Page 6

Installation Kits

INSTALLATION KITS
To facilitate installation of the explosion proof B2800XP monitor, two sizes of installation kits are available.
Hub Size Kit Number Meter Sizes
1/2 in. NPT Hub B280-742 3/8 in., 1/2 in., 3/4 in. with 1/2 in. NPT End Fittings
1 in. NPT Hub B280-737 All Sizes with 1 in. NPT End Fittings and Larger
Installation Kit
P/N: B280-737
Installation Kit
P/N: B280-742
Page 6 March 2017 DSY-PM-00277-EN-03
Page 7

Operating the Monitor

OPERATING THE MONITOR
The monitor has two modes of operation referred to as the RUN mode and the PROGRAM mode. Both the RUN mode and the PROGRAM mode display screen enunciators confirm the state of the monitor. A quick glance at the lower left corner of the LCD screen will confirm operating status. Normal operation will be in the RUN mode. To access the PROGRAM mode, press
MENU until the first programming screen is displayed. After programming the display with the necessary information, turn on the lock out feature to prevent unauthorized access or changing the meter’s setup parameters.
MENU ENTER

Buttons

Modes

MENU Switches to PROGRAM mode
UP Scrolls forward through the parameter options and increments numeric variables
RIGHT
ENTER
RUN Normal operating mode
PROGRAM Used to program parameters in the display
Scrolls backward through the parameter options and moves the active digit to the right
Saves programming information, advances to the next programming parameter, and used in the reset process
Page 7 March 2017 DSY-PM-00277-EN-03
Page 8

Programming

PROGRAMMING
Each turbine flow meter is shipped with either a K-factor value or frequency data. If frequency data is provided, the data must be converted to a K-factor before programming the monitor. K-factor information, when supplied, can usually be found on the neck of the flow meter or stamped on the flow meter body. The K-factor represents the number of pulses per unit of volume. The K-factor is needed to program the monitor readout.

Programming Mode

IO/SETUP
METER
K Factor Units
K FACTOR
Units
DSP
Pulse Output
PASSWORD
RST PSWD
GPM/GAL
PUL/GAL PUL/FT3PUL/LTRPUL/M3
LPM/LIT M3PH/M3
BOTH TOTA L RAT E TEST
PULS ON PULS OFF
M3PD/M3
BPD/BBL
Enter Program Mode
Press MENU to enter the programming mode. The mode indicator changes from RUN to PROGRAM.
Select Meter Size
At the METER prompt, press UP or RIGHT to scroll to the bore size of your meter. Press ENTER to save and advance to the K-factor Units parameter.
OTE:N The meter connection size and the bore size are different. For example, many of the one inch NPT turbines have bore
sizes that range from 3/8…1 in. Be sure to use the correct bore size or the meter will report incorrect flows and totals.
Enter The K-Factor Unit
Press UP or RIGHT to scroll to the K-factor unit of the meter. Enter the unit that the meter was calibrated in.
Example
If the meter was calibrated in US gallons, use PUL/GAL.
Press ENTER to save the K-factor unit and advance to the K FACTOR parameter.
OTE:N Unless otherwise specified, Blancett turbine flow meters are supplied with K-factors measured in pulses per gallon
(PUL/GAL) which will automatically convert to your desired units of measure.
Enter The Meter's K-Factor
OTE:N The K-factor supplied with the meter or calculated from calibration data is needed to set this parameter.
To change the K-factor value, press RIGHT to select the digit that you wish to change. Press UP to increment the digit until it matches the meter’s K-factor. Repeat this process until all K-factor digits have been entered. Press ENTER to save the K-factor and advance to RATE/TOTAL.
Page 8 March 2017 DSY-PM-00277-EN-03
Page 9
Programming
Select the Units of Measure
The monitor is programmed with five common rate/total unit options. The monitor shows the rate/total unit that the display is currently set for. If the current selection is correct, press ENTER once to advance to the DISPLAY FUNCTION parameter. To change the unit, press UP or RIGHT to scroll to the correct rate unit. Press ENTER to save and advance to the DISPLAY FUNCTION parameter.
Selection Rate Total
GPM/GAL Gallons per minute Gallons
LPM/LIT Liters per minute Liters M3PH/M3 Cubic meters per hour Cubic meters M3PD/M3 Cubic meters per day Cubic meters
BPD/BBL Oil barrels per day Oil barrels
Select the Display Function
The monitor displays RATE, TOTAL, alternates between BOTH rate and total or TEST. If the current selection is correct, press ENTER to advance to the next parameter. To change to an alternate display mode, press UP or RIGHT to scroll to the correct display mode and press ENTER to save and advance to the TOTALIZER PULSE OUTPUT parameter.
The TEST function acts like a frequency counter and displays the raw input frequency being supplied to the frequency input terminals. This is very useful when troubleshooting flow problems.
Totalizer Pulse Output
The pulse output parameter can be enabled or disabled. When enabled this output generates 20 mS duration pulse for every time the least significant digit of the totalizer increments. The amplitude of the pulse is dependent on the voltage level of the supply connected to the pulse output and is limited to a maximum 30V DC. Press UP or RIGHT to scroll to ON or OFF. Press
ENTER to save and advance to the PASSWORD parameter.
Password
The PASSWORD setting restricts access to PROGRAM mode. Initially, the password is set to all zeros and any user can modify the parameter settings in PROGRAM mode. To change the password, enter any four digit code. Press UP to increment the digit, and RIGHT to advance to the next digit. Press ENTER to store the password and advance to RST PSWD. The new password is now required to enter PROGRAM mode. With this password set, any user is able to reset the stored totals on the monitor.
Reset Password
The RST PSWD restricts resetting the totals on the monitor. The PASSWORD must also be set to restrict the total reset. Initially, the password is set to all zeros and any user can reset the stored totals on the monitor. To change the password, enter any four digit code. Press UP to increment the digit, and RIGHT to advance to the next digit. After entering all digits, press ENTER to store the password and return to RUN mode. The reset password is now required to reset the totals on the monitor.

Run Mode

Reset Total
To reset the monitor total display, in RUN mode press MENU and ENTER simultaneously. If the RST PSWD has been set to something other than the default, you must enter the new password before the monitor will reset. The monitor flashes TOTAL RST and returns to RUN mode at the conclusion of the reset procedure.
Store Total
The current total can be manually stored in the monitor’s flash memory. Use this feature prior to replacing the battery. Press and hold ENTER for 2 seconds. The display responds with a flashing TOTALSVD and then returns to RUN mode.
Automatic Store Total
The monitor is equipped with a store total feature that works automatically, saving the current total to flash memory. The frequency of saves depends on the power supply option chosen.
Battery Powered: Once per hour and just before a low battery condition turns the unit o.
Loop Powered: Once every ten minutes.
Page 9 March 2017 DSY-PM-00277-EN-03
Page 10

Additional Scaling Parameters

ADDITIONAL SCALING PARAMETERS
OTE:N The programming instructions below are only available for loop powered units. Battery powered units do not include
these programming parameters.

Flow 4 mA Setting

Zero is the default flow rate at the 4 mA setting. If the current selection is correct, press ENTER to advance to the Flow 20 mA parameter. To change the setting, press UP to increment the flashing digit and press RIGHT to move to the next digit. Press
ENTER to save and advance to the Flow 20 mA parameter.

Flow 20 mA Setting

The turbine meter’s maximum flow rate is the default value for the 20 mA setting. If the current selection is correct, press ENTER to advance to the 4…20 mA Calibration parameter. To change the setting, press UP to increment the flashing digit and press RIGHT to move to the next digit. Press ENTER to save and advance to the 4…20 mA Calibration parameter.

4…20 mA Calibration

The 4…20 mA Calibration parameter is the fine adjustment of the 4…20 mA output. The 4 mA setting is typically between 35…50. To set the 4 mA value, connect an ammeter in series with the loop power supply. At the 4MA OUT prompt, press UP to increment and RIGHT to decrement the 4 mA value to until a steady 4 mA reading is reached on the ammeter. Press ENTER to lock in this value and advance to 20MA OUT. The 20 mA adjustment is performed using the same procedure as the 4 mA adjustment.

4…20 mA Test

The monitor contains a diagnostic routine that allows the simulation of mA values between 4…20 to check output tracking. At the 4-20TEST prompt press UP to increment the simulated mA output or RIGHT to decrement the value in increments of 1 mA. The ammeter should track the simulated mA output. If a 4…20 mA test in not necessary, press ENTER once to escape the testing at any time.
m
A
0 0
MEM
CANCEL SAVE NO
Hz % ms RANGE
nS
mV
ac+dc
V
ac+dc
OFF
mA µA
10A MAX
FUSED
DC
MANUAL
TB2 Reset
W
°F
°C
A
mA
µA
A
mA
ac+dc
µA
ac+dc
VIEW MEM CLEAR MEM
TEMPERATURE
W
COM
V
CAT III
400mA
1000V
FUSED
10…30V DC
1…5V DC
Pulse Out Freq. In4-20 mA
+-+-+
Pulse
Inuput
-
Mag Input
%
AutoHOLD FAST MIN MX LOGGING YES
HOLD MIN MAX REL
SETUP
mV
dB
V
dB
A
Page 10 March 2017 DSY-PM-00277-EN-03
Page 11

Additional Input Options

ADDITIONAL INPUT OPTIONS
The Blancett flow monitor is capable of receiving magnetic pickup input or a contact closure input. Since most Blancett flow meters use a magnetic pickup, the monitor is shipped configured for magnetic pickup input. To change to a contact closure input, remove JP2 from the bottom two pins and jumper them to the top two pins. See image below.
D
1.5 VOLTS
CAUTION DO NOT CONNECT IMPROPERLY
CHANGE OR DISPOSE OF IN FIRE. BATTERY
MAY EXPLODE OR LEAK. MADE IN U.S.A.
External
Totalizer Reset
ALKALINE BATTERY
External
Totalizer Reset
(Loop Supply Voltage - 5)
=
MAX
TB2 Reset
Pulse Out Freq. In4-20 mA
+-+-+
0.02
Pulse Inuput
-
Mag Input
20 mS Pulse
30V DC (Max)
TB2 Reset
Pulse Out Freq. In4-20 mA
+-+-+
10k
N/C N/C
Pulse Inuput
-
Mag Input
20 mS Pulse
10…30V DC
1…5V DC
30V DC (Max)
250 Typ
10k
O
Battery Power Loop Power

BATTERY REPLACEMENT

Battery powered monitors use a single 1.5V, D size alkaline battery. When replacement is necessary, use a clean fresh battery for continued trouble-free operation. Save the total to memory before removing the battery.
DO NOT OPEN EXPLOSION PROOF ENCLOSURE UNLESS THE AREA IS KNOWN TO BE FREE OF HAZARDS. FAILURE TO MAKE THE AREA SAFE BEFORE OPENING THE ENCLOSURE CAN RESULT IN A HAZARDOUS SITUATION WITH A POTENTIAL FOR INJURY.
1. Carefully unscrew the enclosure cover to access the circuit board.
2. Remove the four screws securing the circuit board to the enclosure.
3. Lay the circuit board to the side being careful not to pull any wires from their connections.
4. Clip the battery retaining wire/strap and remove the battery.
5. Replace the battery, being sure to observe the proper polarity and install a new retaining strap or wire.
6. Reassemble the monitor, reversing the disassembly process.
Page 11 March 2017 DSY-PM-00277-EN-03
Page 12

Maintenance

MAINTENANCE
ALWAYS DISCONNECT PRIMARY POWER SOURCE BEFORE OPENING ENCLOSURE FOR INSPECTION OR SERVICE.
1. Create a schedule for maintenance determined by the environment and frequency of use. Inspect the monitor at least once a year.
2. Perform visual, electrical and mechanical checks on all components on a regular basis.
a. Visually check for undue heating evidenced by discoloration of wires or other components, damaged or worn parts, or
leakage evidenced by water or corrosion in the interior.
b. Electrically check to make sure that all connections are clean and tight, and that the device is operating correctly.
Page 12 March 2017 DSY-PM-00277-EN-03
Page 13

TROUBLESHOOTING GUIDE

Issue Remedy
• Battery Powered Version: Check battery voltage. Should be 1.5V DC. Replace if low or bad.
No LCD display
No rate or total
displayed
Flow rate display
interprets reading
constantly
• Loop Powered Version: Check for current flow in the loop. Check polarity of the current loop connections for proper orientation.
• Check connection from meter pickup to display input terminals.
• Check turbine meter rotor for debris. Rotor should spin freely.
• Check programming of flow monitor.
• Check to see that the minimum flow rate is being met for the current meter in use. Otherwise, the flow meter will not accurately send pulses to the flow monitor.
• This is usually an indication of external noise. Keep all AC wires separate from DC wires.
• Check for large motors close to the meter pickup.
• Check for radio antenna in close proximity.
• Try disconnecting the pickup from the monitor pig tail. This should stop the noise. If not, then try reorientating the meter to a new location.
Troubleshooting Guide
Flow rate indicator
bounces
Meter Size Default K-Factor Lower Limit Upper Limit
3⁄8 in. (9.53 mm) 20,000 16,000 24,000
1/2 in. (12.7 mm) 13,000 10,400 15,600
3/4 in. (19.05 mm) 2750 2200 3300
7⁄8 in. (22.23 mm) 2686 2148 3223
1 in. (25.4 mm) 870.0 686.0 1044
1-1/2 in. (38.1 mm) 330.0 264.0 396.0
2 in. (50.8 mm) 52.0 41.6 62.0
3 in. (76.2 mm) 57.0 45.6 68.0 4 in. (101.6 mm) 29.0 23.2 35.0 6 in. (152.4 mm) 7.0 5.6 8.0 8 in. (203.2 mm) 3.0 2.4 4.0
10 in. (254 mm) 1.6 1.3 2.0
• This usually indicates a weak signal. Replace pickup and/or check all connections.
• Check K-factor.
Default K-Factor Values

Replacement Parts

Component Part Number
Enclosure B280635
Battery B280601
Battery Holder B280634
Battery Tie Wrap B228036
Cap Plug 3/8 inch B118236
Desiccant Pouch B220141
Pickup Cable B222-121
Battery Mount Plate B280618
Adapter Bridge Plate B280677
Hex Standoffs B280667
Meter Mount Kit - For 1 inch Hub B280-737
Meter Mount Kit - For 1/2 inch Hub B280-742
Page 13 March 2017 DSY-PM-00277-EN-03
Page 14
Specications

SPECIFICATIONS

Power Supply Options
Power Consumption
Alphanumeric Rate and Total Display
Pulsed Output Signal
Magnetic Pickup Inputs
Temperature Drift 50 ppm / °C maximum
Optional Analog Output (Loop Powered Version)
Environmental
Certification
Battery Powered One D size, 1.5V alkaline battery
Loop Powered 4…20 mA loop power
Battery Powered Less than 1 mA @ 1.5V DC
Loop Powered 25 mA (maximum)
Eight digit, 0.5 inch high numeric display
Eight character, 0.25 inch high alphanumeric display
Type
Max Voltage 30V DC
Pulse Type Opto-Isolated open collector transistor
Pulse Width ON State 20 mS / maximum pulse rate 20 Hz
Current (ON State) 0.9V drop @ 5.0 mA or 0.7V drop @ 0.1 A
Frequency Range 0…3500 Hz
Trigger Sensitivity 30 mV p-p
Over Voltage Protected ±30V DC
Frequency Measurement Accuracy
Type 4…20 mA current loop
Resolution 1:4000
Transient Over Voltages Category 3, in accordance with IEC664
Pollution Degree 2, in accordance with IEC664
Mounting Classification NEMA/UL/CSA Type 4 (IP66)
Operating Temperature –22…158° F (–30…70° C)
Humidity 0…90% non-condensing
CSA Ordinary Locations C22.2 No. 1010-1 for Canada; ANSI/ISA S82.02 for US
CSA Hazardous Locations
Outputs one pulse for each increment of the least significant totalizer digit
±0.1%
Class I, Div 1 Groups B,C, D; Class II, Div 1 Groups E, F, G; Class III; Type 4X; T6 @ 70° F; C22.2 No. 30; C22.2 No. 25
UL 698, UL 1203
Page 14 March 2017 DSY-PM-00277-EN-03
Page 15

DIMENSIONS

B 28 S X X - X X
Dimensions
B
A
MENU ENTER
C
A B C D E
5.1 in.
(19.31 mm)
5.25 in.
(19.87 mm)

PART NUMBER INFORMATION

Program Level
S -
Power Option
B - Battery L - Loop
D
2.6 in.
(9.84 mm)
Mounting Style
X - Explosion proof
5/16 in.
(1.19 mm)
Units of Measure AB - Gallons ED - Barrels HB - Liters ID - Cubic Meters
E
¾ NPT
4.51 in.
(17.07 mm)
Page 15 March 2017 DSY-PM-00277-EN-03
Page 16
Flow Monitor, B2800XP Explosion Proof Flow Monitor Standard
Control. Manage. Optimize.
BLANCETT is a registered trademark of Badger Meter, Inc. Other trademarks appearing in this document are the property of their respective entities. Due to continuous research, product improvements and enhancements, Badger Meter reserves the right to change product or system specications without notice, except to the extent an outstanding contractual obligation exists. © 2017 Badger Meter, Inc. All rights reserved.
www.badgermeter.com
The Americas | Badger Meter | 4545 West Brown Deer Rd | PO Box 245036 | Milwaukee, WI 53224-9536 | 800-876-3837 | 414-355-0400 México | Badger Meter de las Americas, S.A. de C.V. | Pedro Luis Ogazón N°32 | Esq. Angelina N°24 | Colonia Guadalupe Inn | CP 01050 | México, DF | México | +52-55-5662-0882 Europe, Eastern Europe Branch Oce (for Poland, Latvia, Lithuania, Estonia, Ukraine, Belarus) | Badger M eter Europe | ul. Korfantego 6 | 44-193 Knurów | Poland | +48-32-236-8787 Europe, Middle East and Africa | Badger Meter Europa GmbH | Nurtinger Str 76 | 72639 Neuen | Germany | +49-7025-9208-0 Europe, Middle East Branch Oce | Badger Meter Europe | PO Box 341442 | Dubai Silicon Oasis, Head Quarter Building, Wing C, O ce #C209 | Dubai / UAE | +971-4-371 2503 Slovakia | Badger Meter Slovakia s.r.o. | Raciansk a 109/B | 831 02 Bratislava, Slovakia | +421-2-44 63 83 01 Asia Pacic | Badger Meter | 80 Marine Parade Rd | 21-06 Parkway Parade | Singapore 449269 | +65-63464836 China | Badger Meter | 7-1202 | 99 Hangzhong Road | Minhang District | Shanghai | China 201101 | +86-21-5763 5412 Switzerland | Badger Meter Swiss AG | Mittelholzerstrasse 8 | 3006 Bern | Switzerland | +41-31-932 01 11 Legacy Document Number: 02-DSY-PM-00114
Loading...