Blancett B110C-375, B111C-160, B110C-875, B110C-375-1/2, B111C-115 User manual

...
Page 1
Turbine Flow Meter
1100 Series Explosion-proof Turbine Meter
TUR-UM-00015-EN-06 (November 2022)
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
User Manual
Page 2
1100 Series Explosion-proof Turbine Meter
CONTENTS
Operating Principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Specications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Pressure Drop in Water . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Operational Startup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Turbine Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Turbine Replacement Kit Part Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Turbine Assembly Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
New Turbine Kit Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Part Number Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Troubleshooting Guide. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
Page ii November 2022TUR-UM-00015-EN-06
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 3
Introduction
Turbine Rotor
Signal
INTRODUCTION
Designed to withstand the demands of the most rigorous flow measurement applications, the Model 1100 Explosion-proof turbine flow meter is reliable, rugged and cost effective. Originally developed for the secondary oil recovery market, the Model 1100 Explosion-proof meter is an ideal meter for liquid flow measurement on or off the oil field.
The meter features a rugged 316 stainless steel housing and rotor support assemblies, CD4MCU stainless steel rotor, and abrasive-resistant tungsten carbide rotor, shaft, and journal bearings. The Model 1100 Explosion-proof meter maintains measurement accuracy and mechanical integrity in the corrosive and abrasive fluids commonly found in oil field water flood project and many industrial applications.
OPERATING PRINCIPLE
Fluid entering the meter passes through the inlet flow straightener, which reduces its turbulent flow pattern and improves the fluid’s velocity profile. Fluid then passes through the turbine, causing it to rotate at a speed proportional to the fluid velocity. As each turbine blade passes through the magnetic field, the blade generates an AC voltage pulse in the pickup coil at the base of the magnetic pickup (see Figure 1). These pulses produce an output frequency proportional to the volumetric flow through the meter. The output frequency represents flow rate and/or totalization of fluid passing through the turbine flow meter.
Magnetic Pickup
or
Other Frequency
Output Device
Figure 1: Schematic illustration of electric signal generated by rotor movement
*Rotor and Rotor Shaft
*Upstream
Rotor Support
*Retaining Ring
*Thrust Ball
NOTE: *Indicates parts supplied in repair kits.
Figure 2: Typical cross-section of B110C-375…B111C-121 turbine flow meter
Output
Magnetic Pickup (not included)
Conduit Adaptor
Meter Body
Retaining Ring
Groove
Retaining Pins
*Downstream Rotor Support
*Bearing
Page 3 November 2022 TUR-UM-00015-EN-06
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 4
Specications
SPECIFICATIONS
Body 316 stainless steel Rotor CD4MCU stainless steel
Materials of Construction
Operating Temperature
Pressure Rating*
End Connections NPT, BSP, Victaulic, Flange, and Hose Barbed Turndown Ratio 10:1
Accuracy
Repeatability ±0.1% Calibration Water (NIST traceable calibration) Pickup Not included
Certifications
* Consult factory for pressure ratings for anged meters.
Bearings Rotor Shaft Tungsten carbide Rotor Support 316 stainless steel
–150…350° F (–101…177° C) The meter should not be subjected to temperatures above 350° F (177° C), or below –150° F (–101° C) or the freezing point of the metered liquid.
5000 psi for all NPT meters up to 2 in. 3000 psi for 3 in. and 4 in. Grayloc meters. 800 psi for 3…10 in. grooved end meters.
±1% of reading for 7/8 in. and larger meters. ±1% of reading over the upper 70% of the measuring range for 3/8 in., 1/2 in., and 3/4 in.
Met Labs File No. E112860 (for explosion proof models only)
- UL1203, Fourth Edition: Explosion-Proof and Dust-Ignition-Proof Electrical Equipment
for Use in Hazardous (Classified) Locations
- CSA C22.2 No. 30-M1986, Third Edition: Explosion-Proof Enclosures for Use in Class I Hazardous Locations
INSTALLATION
PRESSURE IN EXCESS OF ALLOWABLE RATING MAY CAUSE THE HOUSING TO BURST AND CAUSE SERIOUS PERSONAL INJURY.
AVERTISSEMENT
LA PRESSION AU-DESSUS DE L’ESTIMATION PERMISE PEUT FAIRE ÉCLATER ET CAUSER LE LOGEMENT LE DOMMAGE CORPOREL SÉRIEUX.
1. Check the internals of the ow meter for any foreign material. Make sure the turbine rotor spins freely prior to installation. Also, check uid lines and remove any debris found.
2. Install the ow meter with the ow arrow, etched on the exterior of the meter body, pointing in the direction of uid ow. Though the meter is designed to function in any position, where possible, install it horizontally with the conduit adapter facing upward.
3. Thread a magnetic pickup into the conduit adapter completely nger tight without forcing. Secure with a lock nut, if supplied.
4. Install conduit or other ttings suitable for the installation area onto the conduit adapter hub on the ow meter.
Page 4 November 2022TUR-UM-00015-EN-06
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 5
Installation
All Badger Meter Model 1100 Explosion-proof turbine meters use stainless steel and tungsten carbide construction materials. Make sure the operating fluid is compatible with these materials. Incompatible fluids can cause deterioration of internal components and cause a reduction in meter accuracy.
The measured liquid should be free of any large particles that may inhibit rotation of the turbine blades. If particles are present, install a mesh strainer upstream before operating the flow meter. See Table 1 for strainer recommendations.
Part Number Strainer Mesh Clearance Filter Size
B110C-375 60 0.0092 in. 260 μm B110C-500 60 0.0092 in. 260 μm B110C-750 60 0.0092 in. 260 μm B110C-875 60 0.0092 in. 260 μm B111C-110 60 0.0092 in. 260 μm B111C-115 20 0.0340 in. 0.86 mm B111C-120 10 0.0650 in. 1.6 mm B111C-121 20 0.0340 in. 0.86 mm B111C-130 8 0.0900 in. 2.3 mm B111C-140 10 0.0650 in. 1.6 mm B111C-160 4 0.1875 in. 4.8 mm B111C-180 8 0.0900 in. 2.3 mm B111C-200 4 0.1875 in. 4.8 mm
Table 1: Strainer mesh installation details
The preferred plumbing setup is one containing a bypass line (see Figure 3) that allows meter inspection and repair without interrupting flow. If a bypass line is not used, it is important that all control valves be located downstream of the flow meter (see Figure 4).
STRIKING AN EMPTY METER WITH HIGH VELOCITY FLOW STREAM CAN CAUSES DAMAGE.
ATTENTION
DES DOMMAGES PEUVENT ÊTRE PROVOQUÉS EN FRAPPANT UN MÈTRE VIDE AVEC UN JET D’ÉCOULEMENT DE VITESSE ÉLEVÉE.
This is true with any restriction in the flow line that may cause the liquid to flash. If necessary, install air eliminators to ensure that the meter is not incorrectly measuring entrained air or gas.
Badger Meter recommends installation of a minimum length, equal to ten (10) pipe diameters of straight pipe on the upstream side and five (5) diameters on the downstream side of the flow meter. Otherwise, meter accuracy may be affected. Piping should be the same size as the meter bore or threaded port size.
Severe pulsation and mechanical vibration affect accuracy and shorten the life of the meter. If this condition is present, consider using a flow meter possessing superior resistance to pulsation and vibration like the Badger Meter QuicSert. Do not locate the flow meter or connection cable close to electric motors, transformers, sparking devices, high voltage lines, or place connecting cable in conduit with wires furnishing power for such devices. These devices can induce false signals in the flow meter coil or cable, causing the meter to read inaccurately.
If problems arise with the flow meter, consult the on page 11. If further problems arise, consult the factory. If damaged replace the internal components of the turbine flow meter with a turbine meter repair kit available from
Badger Meter. For information pertaining to the repair kits, see Turbine Replacement on page 8.
Page 5 November 2022 TUR-UM-00015-EN-06
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 6
Installation
Isolation Valve
21
Electronic
Flow Monitor
Isolation and
Flow Rate
Control Valve
Electronic
Flow Monitor
10 Pipe
Diameters
10 Pipe
Diameters
Figure 3: Meter installation with a bypass line
Turbine
Flow Meter
Figure 4: Meter installation without a bypass line
Turbine
Flow Meter
Bypass Valve
21
5 Pipe
Diameters
5 Pipe
Diameters
Control Valve
Page 6 November 2022TUR-UM-00015-EN-06
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 7
PRESSURE DROP IN WATER
Pressure Drop In Water
30
Pressure Drop vs Flow Rate
2
Pressure Drop vs Flow Rate
1
10
3/4”
3/8”
3”
1/2”
3/4”
7/8”
1”
1
3/8”
1-1/2”
2”
6”
4”
10”
8”
0.1
Pressure Drop Bar
1/2”
1”
7/8”
1-1/2”
2”
4”
3”
8”
6”
10”
Pressure Drop PSID
0.01
0.1
0.5
(17)
(34)
1
10
(343)
100
(3429)
1000
(34,286)
8000
(274,236)
Flow Rate GPM (BPD)
0.005 1
(1.44)
10
(14.4)
Flow Rate LPM (m³/Day)
100
(144)
1000
(1440)
10,000
(14,400)
30,000
(57,600)
Figure 5: English units pressure drops Figure 6: Metric units pressure drops
OPERATIONAL STARTUP
Follow these steps when installing and starting the meter.
MAKE SURE TO SHUT OFF FLUID FLOW AND RELEASE PRESSURE IN THE LINE BEFORE ATTEMPTING TO INSTALL THE METER IN AN EXISTING SYSTEM.
AVERTISSEMENT
ASSUREZ-VOUS QUE LE FLUX DE FLUIDE A ÉTÉ COUPÉ ET DE LA PRESSION DANS LA LIGNE A ÉTÉ LIBÉRÉE AVANT D’ESSAYER D’INSTALLER LE MÈTRE DANS UN SYSTÈME ACTUEL.
After meter installation, close the isolation valves and open the bypass valve. Allow liquid to flow through the bypass valve for sufficient time to eliminate any air or gas in the flow line.
HIGH VELOCITY AIR OR GAS MAY DAMAGE THE INTERNAL COMPONENTS OF THE METER.
ATTENTION
DES DOMMAGES PEUVENT ÊTRE PROVOQUÉS EN FRAPPANT UN MÈTRE VIDE AVEC UN JET D’ÉCOULEMENT DE VITESSE ÉLEVÉE.
1. Open the upstream isolating valve slowly to eliminate hydraulic shock while charging the meter with the liquid. Open the valve to full open.
2. Open downstream isolating valve to permit meter to operate.
3. Close the bypass valve to a full closed position.
4. Adjust the downstream valve to provide the required ow rate through the meter.
OTE: N If necessary, use the downstream valve as a control valve.
Page 7 November 2022 TUR-UM-00015-EN-06
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 8
Turbine Replacement
TURBINE REPLACEMENT
The Model 1100 Explosion-proof turbine flow meter uses wear-resistant moving parts to provide troublefree operation and long service life. Designed for easy field service of a damaged flow meter, Model 1100 Explosion-proof repair kits replace only the internal parts, rather than replacing the entire flow meter. Repair parts use stainless steel alloys and tungsten carbide construction materials.
Each repair kit is factory calibrated to ensure accuracy throughout the entire flow range. Each kit is complete and includes a new K-factor, which is the calibrated number of pulses generated by each gallon of liquid. Recalibration of the monitor or other electronics uses the K-factor to provide accurate output data.
OTE: N If the meter repair kit part number ends in NCC (no calibration), it was not factory calibrated. For these repair kits, use
the nominal K-factor.
Turbine Replacement Kit Part Number
Flow Meter Size Replacement Kit Fits Meter Part Number Repair Kit Part Number
3/8 in. B110C-375, B110C-375-1/2 B251-102
1/2 in. B110C-500, B110C-500-1/2 B251-105
3/4 in. B110C-750, B110C-750-1/2 B251-108
7/8 in. B110C-875 B251-109
1 in. B110C-110 B251-112
1-1/2 in. B110C-115 B251-116
2 in. (Low) B110C-121 B251-116
2 in. B110C-120 B251-120
3 in. B111C-130 B251-131
4 in. B111C-140 B251-141
6 in. B111C-160 B251-161
8 in. B111C-180 B251-181
10 in. B111C-200 B251-200
Table 2: Repair kit part numbers
Turbine Assembly Removal
HIGH-PRESSURE LEAKS ARE DANGEROUS AND MAY CAUSE PERSONAL INJURY. MAKE SURE TO SHUT OFF FLUID FLOW AND RELEASE RESIDUAL PRESSURE IN THE LINE BEFORE ATTEMPTING TO REMOVE THE METER.
AVERTISSEMENT
LES FUITES À HAUTE PRESSION SONT DANGEREUSES ET PEUVENT CAUSER LE DOMMAGE CORPOREL. ASSUREZ-VOUS QUE LE FLUX DE FLUIDE A ÉTÉ COUPÉ ET DE LA PRESSION DANS LA LIGNE A ÉTÉ LIBÉRÉE AVANT D’ESSAYER D’ENLEVER LE MÈTRE.
Page 8 November 2022TUR-UM-00015-EN-06
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 9
Turbine Replacement
Disassembly
1. Refer to Figure 7, Figure 8 and Figure 9 for relative positions of repair kit components.
2. Remove the magnetic pickup from the meter body to avoid damage during repair.
3. Remove the retaining ring from one end of the meter.
4. Remove the rotor support from the body. If the rotor support is jammed in the body, use a pair of pliers or vise-grips to break the rotor support free.
5. The rotor may also be removed at this time.
OTE: N 4 in. and larger meters have two retaining rings (one on either side of the rotor) that require removal before the rotor
can be removed (see Figure 9).
6. Remove the retaining ring from the opposite side of the meter.
7. Remove the second rotor support.
Magnetic
Pickup
(Not Included)
Flow Direction
Arrow
Flow Direction
Upstream
Rotor Support
Retaining Ring
Upstream
Rotor Support
Retaining Ring
Rotor
Assembly
Retaining Ring
Groove
Flow Direction
Arrow
Flow Meter
Body
FLOW
Retaining Ring
Downstream
Rotor Support
Flow Direction
Arrow
Groove
Flow Direction
Arrow
Figure 7: Component positions for B110C-375…B111C-115 and B111C-121
Magnetic
Pickup
(Not Included)
Body
Retaining Ring
Groove
FLOW
Flow Direction
Arrow
Flow Direction
Arrow
Arrow
Flow Direction
Arrow
Retaining Ring
Groove
Rotor
Assembly
Figure 8: Component positions for B111C-120 and B111C-130
Retaining RingFlow Meter
Retaining Ring
Downstream
Rotor Support
Flow Direction
Arrow
Upstream
Rotor Support
Flow Direction
Arrow
Retaining Ring
Retaining Ring
Grooves
Rotor
Assembly
Retaining Ring Retaining Ring
Magnetic
Pickup
(Not Included)
Flow Meter
Body
FLOW
Flow Direction
Arrow
Retaining Ring
Grooves
Downstream
Rotor Support
Flow Direction
Arrow
Retaining Ring
Figure 9: Component positions for B111C-140 and B111C-200
Page 9 November 2022 TUR-UM-00015-EN-06
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 10
Turbine Replacement
New Turbine Kit Installation
MPORTANTI
Before reassembly, note that an arrow is cast or engraved on each component. The arrow indicates the primary flow direction. When reassembled, the arrowheads must point in the direction of the fluid flow. The arrows must also be oriented in the up position on both rotor supports. The magnetic pickup side of the body signifies the up position. Performance of repair kit calibration is in the up position. Reinstallation of the repair kit in the up position ensures continuation of accurate measurements. Figure 7, Figure 8, and
Figure 9 show the proper alignment and orientation of the repair kits.
OTE: N Fractional size (3/8 in., 1/2 in. and 3/4 in.) rotors do not contain a cast or engraved arrow. However, a colored cap on
the downstream side of the rotor shaft indicates flow direction. Remove this cap before assembly, noting flow direction.
1. Install one of the rotor supports into the body bore, noting the orientation of the arrow.
2. Secure a retaining ring in the groove provided. Check for complete installation of retaining rings in each groove.
OTE: N 4 in. and larger meters have a retaining ring at both ends of the rotor (see Figure 9).
3. Insert the rotor and second rotor support in the opposite side of the body, noting the orientation of the arrow.
4. Secure the second retaining ring in the opposite groove, using the same procedure as in step 2 above.
EXCESS AIR PRESSURE MAY DAMAGE THE ROTOR AND BEARINGS BY OVER SPINNING.
ATTENTION
LA PRESSION ATMOSPHÉRIQUE EXCESSIVE PEUT ENDOMMAGER LE ROTOR ET LES ROULEMENTS PRÈS AU-DESSUS DE LA ROTATION.
5. Check the meter by lightly pung air through the assembly. If the rotor does not turn freely, disassemble the meter and remove anything that might obstruct movement of the rotor.
NOTE: At this time, electronics require recalibration. Refer to the display's user manual. If there are any questions on
recalibration, contact us or the manufacturer of the associated electronics.
6. Install the magnetic pickup.
Page 10 November 2022TUR-UM-00015-EN-06
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 11
PART NUMBER INFORMATION
Part Number Information
Part
Number
B110C-375-½
B110C-500-½
B110C-750-½
B110C-875
B111C-110
B111C-115
B111C-121
B111C-120
B111C-130
B111C-140
B111C-160
B111C-180
B111C-200
Bore Size
3/8 in.
(9.5 mm)
1/2 in.
(12.7 mm)
3/4 in.
(19.1 mm)
7/8 in.
(22.2 mm)
1 in.
(25.4 mm)
1-1/2 in.
(38.1 mm)
1-1/2 in.
(38.1 mm)
2 in.
(50.8 mm)
3 in.
(76.2 mm)
4 in.
(101.6 mm)
6 in.
(152.4 mm)
8 in.
(203.2 mm)
10 in.
(254 mm)
End Connec-
tions
1/2 in. male NPT 5000
1/2 in. male NPT 5000
1/2 in. male NPT 5000
1 in. male NPT 5000
1 in. male NPT 5000
1-1/2 in. male
NPT
2 in. male NPT 5000
2 in. female NPT 5000
3 in. grooved
end
4 in. grooved
end
6 in. grooved
end
8 in. grooved
end
10 in. grooved
end
Max.
PSI
5000
800
800
800
800
800
Flow Ranges
gpm (lpm) bpd m3/d
0.6…3
(2.3…11.4)
0.75…7.5
(2.8…28.4)
2…15
(7.6…56.7)
3…30
(11.4…113.6)
5…50
(18.9…189.3)
15…180
(56.8…681.4)
15…180
(56.8…681.4)
40…400
(151.4…1514.2)
60…600
(227.1…2271.2)
100…1200
(378.5…4542.5)
200…2500
(757.1…9463.5)
350…3500
(1324.9…13248.9)
500…5000
(1892.7…18927.1)
20…100 3.3…16 60 18000 1
25…250 4.1…41 60 13000 1
68…515 10.9…81.75 60 3300 1
100…1000 16…160 60 3100 2
170…1700 27.25…272.5 40 870 2
515…6000 82…981 20 330 5
515…6000 82…981 20 330 6
1300…13000 218…2180 20 52 14
2100…21000 327…3270 10 57 15
3400…41000 545…6540 10 29 20
6800…86000 1090…13626 4 7 46
12000…120,000 1363…19076 4 3 56
17000…171,000 2725…27252 4 1.6 80
Strainer
Mesh
Approx. K-factor
Pulse/Gal
Meter
Weight
(lb)
End to End
Length
3.00 in.
(76.20 mm)
3.00 in.
(76.20 mm)
3.00 in.
(76.20 mm)
4.00 in.
(101.60 mm)
4.00 in.
(101.60 mm)
6.00 in.
(152.40 mm)
6.00 in.
(152.40 mm)
10.00 in.
(245.00 mm)
12.50 in.
(317.50 mm)
12.00 in.
(304.80 mm)
12.00 in.
(304.80 mm)
12.00 in.
(304.80 mm)
12.00 in.
(304.80 mm)
Page 11 November 2022 TUR-UM-00015-EN-06
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 12
1100 Series Explosion-proof Turbine Meter
Control. Manage. Optimize.
Blancett is a registered trademarks of Badger Meter, Inc. Other trademarks appearing in this document are the property of their respective entities. Due to continuous research, product improvements and enhancements, Badger Meter reserves the right to change product or system speci
cations without notice, except to the extent an outstanding
contractual obligation exists. © 2022 Badger Meter, Inc. All rights reserved.
TROUBLESHOOTING GUIDE
Issue Possible Cause Remedy
Meter indicates higher than actual flow rate.
Meter indicates lower than actual flow rate.
Erratic system indication, meter alone works well (remote monitor application only).
Indicator shows flow when shut off.
No flow indication. Full or partial open position.
Erratic indication at low flow, good indication at high flow.
Cavitation. Debris on rotor support. Build up of foreign material on meter bore. Gas in liquid.
Debris on rotor. Worn bearing. Viscosity higher than calibrated.
Ground loop in shielding.
Mechanical vibration causes rotor to oscil­late without turning.
Fluid shock, full flow into dry meter or im­pact caused bearing separation or broken rotor shaft.
Rotor has foreign material wrapped around it.
Increase back pressure. Clean meter. Clean meter. Install gas eliminator ahead of meter.
Clean meter and add filter. Clean meter and add filter. Recalibrate monitor.
Ground shield one place only. Look for internal electronic instrument ground. Reroute cables away from electrical noise.
Isolate meter.
Rebuild meter with repair kit and recali­brate monitor. Move to location where meter is full on startup or add down­stream flow control valve.
Clean meter and add filter.
No flow indication. Faulty pickup. Replace pickup.
System works perfect, except indi­cates lower flow over entire range.
Bypass flow, leak.
Repair or replace bypass valves, or faulty solenoid valves.
Meter indicating high flow, up­stream piping at meter smaller
Fluid jet impingement on rotor. Change piping.
than meter bore.
Meter indicating low flow, up­stream piping at meter smaller than meter bore.
Viscosity lower than calibrated.
Change temperature, change fluid or recalibrate meter.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com