Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
TEST INSTRUMENT SAFETY
WARNING
Normal use of test equipment exposes you to a certain amount of danger from electrical shock because testing must sometimes be
performed where exposed high voltage is present. An electrical shock causing 10 milliamps of current to pass through the heart will
stop most human heartbeats. Voltage as low as 35 volts dc or ac rms should be considered dangerous and hazardous since it can
produce a lethal current under certain conditions. Higher voltages are even more dangerous. Your normal work habits should
include all accepted practices to prevent contact with exposed high voltage, and to steer current away from your heart in case of
accidental contact with a high voltage. Observe the following safety precautions:
1. There is little danger of electrical shock from the dc output of this power supply. However, there are several other possible test
conditions using this power supply that can create a high voltage shock hazard:
a. If the equipment under test is the “hot chassis” type, a serious shock hazard exists unless the equipment is unplugged (just
turning off the equipment does not remove the hazard), or an isolation transformer is used.
b. If the equipment under test is “powered up” (and that equipment uses high voltage in any of its circuits), the power supply
outputs may be floated to the potential at the point of connection. Remember that high voltage may appear at unexpected
points in defective equipment. Do not float the power supply output to more than 100 volts peak with respect to chassis or
earth ground.
c. If the equipment under test is “off” (and that equipment uses high voltage in any of its circuits under normal operation),
discharge high-voltage capacitors before making connections or tests. Some circuits retain high voltage long after the
equipment is turned off.
2. Use only a polarized 3-wire ac outlet. This assures that the power supply chassis, case, and ground terminal are connected to a
good earth ground and reduces danger from electrical shock.
3. Don’t expose high voltage needlessly. Remove housings and covers only when necessary. Turn off equipment while making
test connections in high-voltage circuits. Discharge high-voltage capacitors after removing power.
(continued on inside back cover)
Instruction Manual
Triple Output
DC POWER SUPPLY
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
The B & K-Precision Models 1651A & 1652 Triple Output
DC Power Supplies are high quality, general purpose dc power
sources. They provide two supplies with a 0-24 volt dc output
and one with a fixed 5 volt dc output. The 0-24V supplies are
adjustable and are capable of current output of 0-0.5 amp. The
fixed 5V supply has a current output of 0-4 amps, allowing it
to handle extensive digital logic circuitry. Two panel mounted
meters can simultaneously monitor the output current and
output voltage of either of the 0-24V supplies.
The two 0-24 volt supplies can be operated independently or
in one of two tracking modes. In the series tracking mode, the
“B” supply tracks the voltage of the “A” supply. In the series
tracking mode the “A” and “B” supplies are connected in
series, allowing a single output of 0-48V at up to 0.5 amps. In
the parallel tracking mode, the two supplies are connected in
parallel, allowing a single 0-24V output at up to 1 amp.
Both 0-24 volt supplies may be used in constant voltage or
constant current applications. The crossover from constant
voltage to constant current modes is smooth and automatic.
LED’s indicate the Constant Current mode of operation. In
constant voltage applications, a current limit may be preset.
When load variations cause the current to reach the preset
limit, the unit then regulates output current rather than output
voltage. Current limits are adjustable from 6% to 100% of
maximum. In constant current applications, the maximum
voltage may be preset. When load variations cause current to
drop below the regulated value, the unit reverts to regulated
voltage operation at the preset value.
The fixed 5V supply is ideal for powering digital logic
circuitry. The 0-4 amp capacity allows the supply to be used
for large circuits. Built-in overload protection automatically
limits the current output to a maximum of 4 amps. An LED
indicator lights when the supply is overloaded.
These supplies exhibits excellent regulation and low ripple
characteristics. The circuit design incorporates a pre-regulator,
which greatly reduces internal power dissipation at low output
voltages.
Reverse polarity protection prevents accidental damage to
the power supply from improper connection to an external
voltage, and current limiting protects the equipment being
powered, as well as the power supply.
The output is isolated from chassis and earth ground, which
permits full flexibility of connections. When needed, the (+) or
(-) polarity may be strapped to ground, or either polarity may
be floated to an external voltage. Additionally, the two “main”
volt supplies can be used as a “split supply” with two positive
voltages and a common negative, two negative voltages and a
common positive, or one positive, one negative, and a
common. All of these configurations can be used with either
matching (tracking) or differing (independent) voltages.
1
INTRODUCTION
The features and versatility of these units, especially the
triple output and tracking features, make them an ideal
general purpose power supply for engineering lab
applications. They can serve as a single or multi-voltage
power source, including the bias supply, for breadboard and
prototype circuits and equipment. They can provide single or
simultaneously varying voltages for circuit evaluation. They
can provide tracking (+) and (-) voltages for evaluating
differential amplifiers. They may be used as a battery
eliminator, or to power individual circuit boards or cards
while removed from the system. Their output can be
evaluated while powering a breadboard or prototype circuit to
determine the circuit’s power supply requirements. Their
laboratory quality specifications will meet most engineering
laboratory requirements.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
The same features that make the Model 1651A and 1652 a
good choice for an engineering lab also make them a good choice
for most other solid state electronic applications. These
applications include service shops; industrial production testing
of components, assemblies, and complete equipment; for school
laboratories, and home use by electronic hobbyists.
2
FEATURES
TRIPLE OUTPUT
Operates as three separate power supplies. Each has floating
output and is completely isolated from the other two.
ONE FIXED 5V SUPPLY
0-to-4 amp f
digital logic circuitry. Adequate current capacity for extensive
circuitry.
TWO 0-24 VOLT SUPPLIES
“A” and “B” supplies are continuously variable over 0-to-24
volt range. Each supply 0.5 amp current capacity.
UNIQUE TRACKING FEATURE
The two 0-to-24 V supplies can be operated so that the “B”
supply tracks the “A” supply. Outputs can be strapped for two
positive voltages with a common negative, two negative
voltages with a common positive, or one positive and one
negative with a neutral common.
SINGLE 0-48V SUPPLY
Series tracking feature allows use of “A” and “B” supplies as
one 0-to-48 V, 0.5 amp supply.
ixed 5 volt supply is ideal for use with most
SINGLE 0-24V 1 AMP SUPPLY
Parallel tracking feature allows use of “A” and “B”
supplies as a 0-to-24 V supply with a 1 amp current capacity
(through “A” output terminals).
CONSTANT VOLTAGE OR CONSTANT CURRENT
The “A” and “B” supplies provide regulated dc voltage
output or regulated dc current output. Crossover is smooth
and automatic.
METERING
Two, easy-to-read meters monitor output voltage and
output current of the “A” and “B” supplies. Use of two
meters allows simultaneous current and voltage metering
when using “A” and “B” supplies in tracking or independent
operation.
LABORATORY QUALITY
Excellent regulation, low ripple.
LED INDICATORS
Identify mode of operation.
PRE-REGULATOR
Limits internal dissipation for higher reliability and
efficiency.
3
FEATURES
ISOLATED OUTPUT
Either polarity may be floated or grounded
OVERLOAD PROTECTION
Fully adjustable current limiting (from 6% to 100%
of maximum output current) for “A” and “B” supplies
protects circuit under test and the power supply
REVERSE POLARITY PROTECTION
Prevents damage to power supply from external voltage
of reverse polarity.
HOOK-UP CABLES
Supplied with three sets of red and black hook-up leads.
4
SPECIFICATIONS
“A” AND “B” SUPPLIES
Output Voltage Range:
0V to 24VDC (0 ± 100mV)
Output Current Limit Range:
0 to 0.5A.
Constant Voltage Operation:
Voltage Regulation:
Load: ≤0.01% + 3mV
Line (108 – 132V): ≤0.01% + 3mV
Ripple Noise: ≤2mVrms (5Hz to 1MHz)
Recovery Time: ≤100us typical.
Temp. Coefficient
0°C to 40°C ≤300ppm/°C
Tracking Error:
No Load: ≤0.2% + 20mV
Full Load: ≤0.2% + 100mV
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Constant Current Operation:
Adjustable Current Limits: ≤30mA to ≥500mA
Current Regulation:
Load: ≤0.2% + 6mA.
Line (108 – 132V): ≤0.2% + 3mA.
Ripple Current and Noise: ≤3mA rms.
Metering (“A” & “B” only):
Voltmeter:
Range: 0 to 25V
Accuracy: ≤2.5% of Full Scale
±2% + 2 digits
(1652)
Ampmeter:
Range: 0 to 600mA
Accuracy: ≤2.5% of Full Scale
Line Regulation 108 – 132 V: ≤5mV
Ripple And Noise: ≤5mV rms (5 Hz to 1
MHz)
Overvoltage Protection
Threshold: 5.7 to 6.5 V
GENERAL
Power Requirements: 100/120/220/240VAC ±10%,
50/60Hz.
Power Consumption: 165VA.
Protection: Reverse polarity, overvoltage
and current limiting.
NOTE: Specifications and information are subject to change without notice.
Temperature Range
& Humidity:
Operation: 0°C to 40°C <80% R.H.
Storage: -20°C to 60°C <70%
R.H.
Dimensions (HxWxD): 4.5 x 11.75 x 10.375”
(114 x 298 x 264 mm)
Weight: 4.8 kg (10.6 lbs)
Accessories Supplied: Hook-Up Cables, 3 pair
(Black & Red).
Power Cord.
Spare Fuse.
Instruction Manual
6
CONTROLS AND INDI
CATORS
1. POWER Switch. Turns power on and off.
2. Power ON light. Red LED lights to indicate a power on
condition.
3. GND Terminal (Green). Earth and Chassis Ground.
4. A/B Metering Switch. Selects simultaneous Voltage &
Current metering for the “A” or “B” supplies. When in the
A position, the V and mA meters are connected to the “A”
supply. When in the B position, the V and mA meters are
connected to the “B” supply.
5. V Meter. Indicates voltage on the “A” or “B” supply
depending on the position of the A/B Metering switch.
6. mA Meter. Indicates current on the "A" or "B" supply
depending on the position of the A/B Metering switch.
7. Zero Adjusts. Mechanical zero adjusts for the V & mA
Meters thru the front panel with a slotted screwdriver.
8. TRACKING/INDEPENDENT Mode Switch: Three
Position switch that selects INDEPENDENT mode,
PARALLEL TRACKING mode, or SERIES TRACKING
mode of the “A” and “B” supplies as follows:
a.When the switch is in the right position, the unit is
in the INDEPENDENT mode and the “A” and “B”
power supplies are completely independent from
one another.
b.When the switch is in the middle position, the
unit is in the PARALLEL TRACKING mode. In
this mode the “A” and “B” supplies are wired
together in parallel and both the maximum
current and voltage are set using the “A”
controls. The “A” and “B” outputs can be used
as two individual (but tracking) power supplies
or just the “A” output can be used as a 0-to-24
volt supply with a 1 A capability.
c.When the switch is in the left position, the unit is
in the SERIES TRACKING mode. In this mode,
maximum voltage of both supplies is set using
the “A” VOLTAGE controls (voltage at output
terminals of the “B” supply tracks the voltage at
the output terminals of the “A” supply). Also, in
this mode of operation the positive terminal (red)
of the “B” supply is internally connected to the
negative terminal (black) of the “A” supply.
This allows the two supplies to be used as one 0to-48 volt supply.
“A” SUPPLY CONTROLS AND INDICATORS
9. VOLTAGE Control. Adjusts the output voltage of the
“A” supply. Also functions as the adjustment control for
the maximum output voltage of the “B” supply when either
parallel or series tracking mode is selected. Voltage can be
read from the V Meter when the A Metering mode is
selected.
7
CONTROLS AND INDICATORS
Figure 1. Front panel controls and indicators
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
8
CONTROLS AND INDICATORS
10. CURRENT Control. Adjusts current limit of “A”
supply in constant voltage mode. Adjusts constant
current value of “A” supply in constant current mode.
Adjusts the constant current value of the “B” supply
when either SERIES or PARALLEL TRACKING is
selected. Current can be read from the mA Meter
when the A Metering mode is selected.
11. “+” Terminal (Red). Positive polarity
for the “A” supply. Also serves as the positive polarity
terminal for 0-to-48 V, 1A parallel tracking and 0-to48 V, 0.5A series tracking operation.
12. “-” Terminal (Black). Negative polarity output
terminal for the “A” supply. Also serves as the
negative polarity terminal for 0-to-24 V, 1 A parallel
tracking operation. In series tracking operation, this
terminal is internally tied to the (+) positive terminal
of the “B” supply.
13. CONSTANT CURRENT Indicator.
a.Red LED lights when “A” supply is in the
Constant Current mode. The Power Suppl
regulates the output current at the value set by the
“A” CURRENT control. In the Parallel Tracking
mode, when this indicator is lit, both the “A” and
“B” supplies are in the Constant Current mode.
b.When the LED is off, the “A” supply is in the
Constant Voltage mode. The Power Supply
regulates the output voltage at the value set by the
“A” VOLTAGE controls. In either the Series or
Parallel Tracking mode, when this indicator is off,
both the A”A and “B” supplies are in the Constant
Voltage mode.
output terminal
“B” SUPPLY CONTROLS AND INDICATORS
14. VOLTAGE Control. Adjusts the output voltage of the
“B” supply when the INDEPENDENT mode is selected.
Voltage can be read on the V Meter when the B Metering
mode is selected. Control is disabled when TRACKING
mode is selected.
15. CURRENT Control. Adjusts current limit of “B” supply
in constant voltage mode. Adjusts constant current value
of “B” supply in constant current mode. Current can be
read from the mA Meter when the current mA Metering
mode is selected. Control is disabled when TRACKING
mode is selected.
16. “+” Terminal (Red). Positive polarity output terminal
for the “B” supply. In series tracking operation, this
terminal is connected to the negative terminal of the “A”
supply.
17. “-” Terminal (Black). Negative polarity output
terminal for the “B” supply. Also serves as the negative
polarity terminal for 0-to-48 V series tracking operation.
y
18. CONSTANT CURRENT Indicator
a.Red LED lights when “B” supply is in the
Constant Current mode. The Power Supply
regulates the output current at the value set by
the “B” CURRENT control when in the
Independent mode.
21. OVERLOAD Indicator. Lights when load on FIXED 5
Volt supply becomes too large.
REAR PANEL CONTROLS
22. LINE VOLTAGE SELECT Switches. Combination
settings allow Universal Power Operation;
23. Line Cord Receptacle. Fuse.
10
CONTROLS AND INDICATORS
Figure 2. Rear panel controls.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
11
OPERATING INSTRUCTIONS
SAFETY PRECAUTIONS
CAUTION
Avoid contacting the heat sink at the rear of the power
supply. When the unit is providing large amounts of
current at any or all of its outputs, the heat sink can
become very hot. Contacting the heat sink when it is hot
could result in skin burns or damage to the equipment in
contact with them.
Use only a polarized 3-wire ac outlet. This assures that
the power supply chassis, case, and ground terminal are
connected to a good earth ground and reduces danger
from electrical shock.
There may be great danger of electrical shock if the
power supply output is connected to an external high
voltage. Some equipment being powered may contain
high voltage and present a shock hazard. Observe
caution. If the power supply output is floated (referenced
to a voltage rather than earth ground) turn off the power
supply and the equipment under test when making
connections. Never float the power supply to a potential
greater than 100 volts peak with respect to earth ground.
EQUIPMENT PRECAUTIONS
Avoid using the power supply in ambient temperatures
above +40° C. Always allow sufficient air space around
the heat sink at the rear of the power supply for effective
radiation to prevent internal heat build-up.
Although the power supply is protected against reverse
polarity damage, the circuit being powered may not
include such protection. Always carefully observe polarity;
incorrect polarity may damage the equipment under test.
Do not exceed the voltage rating of the circuit being
powered. Many transistors and integrated circuits will not
withstand voltage of 30 volts.
There is no need to worry about voltage spikes or
overshoot damaging the equipment under test. The voltage
between the output terminals of the power supply never
exceeds the preset value as the POWER switch is turned
on or off.
INDEPENDENT USE OF “A” OR “B” SUPPLY
The “A” and “B” supplies each provide a 0-to-24 volt
output at up to 0.5 amps. This procedure covers the use of
the “A” and “B” supplies only when they are used
independently from one another. When used in the
INDEPendent operating mode, the operating controls of
the two power supplies are completely independent and
either supply can be used individually or both can be used
simultaneously. Basic operation is covered here. Several
variations are covered in the APPLICATIONS section of
this manual.
12
OPERATING INSTRUCTIONS
Hook-up
3. Set the INDEPENDENT/TRACKING mode switch to
the right position so that the power supply is in the
INDEPendent operating mode.
4. Set the A/B Metering selection switch to the A (up)
position to monitor the “A” supply.
5. Turn off the power supply and the equipment to be
powered during hook-up.
6. Connect the positive polarity of the device being
powered to the red (+) terminal of the power supply.
7. Connect the negative polarity of the device being
powered to the black (-) terminal of the power supply.
8. Fig. 3 illustrates the grounding possibilities when used
in the INDEPendent mode.
a. If the negative polarity of the equipment or circuit
being powered is also the chassis or common, it may
be grounded to earth by strapping the black (-)
terminal to the green (GND) terminal as shown in Fig.
3A.
b. Similarly, the positive polarity can be grounded by
strapping the red (+) terminal to the green (GND)
terminal as shown in Fig. 3B.
c. If an earth ground reference is not required, the
configuration of Fig. 3C may be used. The scheme in
Fig. 3C should also be used where it is not known
whether the chassis is common with either the
positive or negative polarity.
d. If the chassis or common of the equipment being
powered is separate from both the positive and
negative polarity power inputs, use the connection
shown in Fig. 3D.
1. Observe proper polarity. If the circuit being powered
is not equipped with reverse polarity protection,
damage to the circuit can result from reverse polarity.
Use color coded hook-up leads, for convenience in
identifying polarity, red for (+) and black for (-).
2. Make sure that the hook-up leads offer sufficient
current capability and low resistance between the
power supply and the circuits being powered. The
hook
-up leads supplied wi
rated for 4 amp.
Typical Constant Voltage Operation
1. Before connecting the device to be powered to the
power supply, determine the maximum safe load
current for the device to be powered and set the
current limit value (see “Setting Current Limit”
procedure in this section).
2. Set the A/B Meter selection switch to the A (up)
position to monitor the “A” supply.
3. Set VOLTAGE control to minimum (fully
counterclockwise).
4. Turn off power supply and connect it to the device
to be powered (see “Hook-Up” procedure in this
section).
5. Turn on POWER switch. The CONSTANTCURRENT indicator should not light.
6. Increase the VOLTAGE setting until the Volt meter
reads the desired value.
7. The load current is read directly on the mA meter.
13
th the power supply are
OPERATING INSTRUCTIONS
Figure 3. Independent operation grounding possibilities (sheet 1 of 2)
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
14
OPERATING INSTRUCTIONS
Figure 3. Independent operation grounding possibilities (sheet 2 of 2)
15
OPERATING INSTRUCTIONS
8. If the load current exceeds the preset current limit, the
CONSTANT CURRENT indicator will light. In this case, the
power supply automatically switches to the constant current
mode and further rotation of the VOLTAGE control will not
increase the output voltage.
Setting Current Limit
1. Determine the maximum safe current for the device to be
powered.
2. Temporarily short the (+) and (-) terminals of the power
supply together with a test lead.
3. Rotate the VOLTAGE control away from zero sufficiently
for the Constant Current indicator to light.
4. Adjust the CURRENT control for the desired current limit.
Read the current value on the mA meter.
5. The current limit (overload protection) has now been
preset. Do not change the CURRENT control setting after
this step.
6. Remove the short between the (+) and (-) terminals and
hook up for constant voltage operation.
Typical Constant Current Operation
1. Before connecting the device to be powered to the power
supply, determine the maximum safe voltage to be applied,
set the A/B Meter s election switch to the A (up) position, and
set the VOLTAGE control to obtain that voltage reading on the
Volt meter.
2. Determine the desired constant current value.
3. Set the CURRENT control to minimum (fully
counterclockwise).
Figure 4. Typical constant voltage operation
4. Turn off the power supply and connect it to the device to
be powered.
5. Turn on the power supply. The CONSTANT CURRENT
indicator should light.
6. The current can be read directly on the mA meter.
16
Figure 5. Setting Current limit.
7. Increase the CURRENT control setting until the desired
constant current value is read on the display, or set the current
limit in advance (before connecting the load) as prescribed earlier
in the “Setting Current Limit” procedure.
17
OPERATING INSTRUCTIONS
Figure 6. Typical constant current operation.
8. If the load current drops below the constant current value,
the CONSTANT CURRENT indicator will go off. In this
case, the power supply automatically switches to the constant
voltage mode, and further rotation of the CURRENT control
will not increase the output current.
OPERATING INSTRUCTIONS
Constant Voltage/Constant Current Characteristic
The working characteristic of this power supply is called
a constant voltage/constant current automatic crossover
type. This permits continuous transition from constant
current to constant voltage modes in response to the load
change. The intersection of constant voltage and constant
current modes is called the crossover point. Fig. 7 shows
the relationship between this crossover point and the load.
For example, if the load is such that the power supply is
operating in the constant voltage mode, a regulated output
voltage is provided. The output voltage remains constant as
the load increases, up until the point where the preset
current limit is reached. At that point, the output current
becomes constant and the output voltage drops in
proportion to further increases in load. The crossover point
is indicated by the front panel LED indicators. The
crossover point is reached when the CONSTANT CURRENT indicator comes on.
Similarly, crossover from the constant current to the
constant voltage mode automatically occurs from a
decrease in load. A good example of this would be seen
when charging a 12-volt battery. Initially, the open circuit
voltage of the power supply may be preset for 13.8 volts. A
low battery will place a heavy load on the supply and it will
operate in the constant current mode, which may be
adjusted for a 0.5 amp charging rate. As the battery
becomes charged, and its voltage approaches 13.8 volts, its
load decreases to the point where it no longer demands the
full 0.5 amp charging rate. This is the crossover point
where the power supply goes into the constant voltage
mode.
Figure 7. Constant voltage/constant current characteristic.
SERIES TRACKING OPERATION
When the series tracking mode of operation is selected, the
positive (red) terminal of the “B” supply output is internally
connected to the negative (black) terminal of the “A” supply.
This allows the power supply to be used as a single 0-to-48
volt power supply simply by using the negative (black)
terminal of the “B” supply and the positive (red) terminal of
18
the “A” supply.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
OPERATING INSTRUCTIONS
Figure 8. Series tracking (0-48 V) operation grounding possibilities (sheet 1 of 3).
19
OPERATING INSTRUCTIONS
Figure 8. Series tracking (0-48 V) operation grounding possibilities (sheet 2 of 3)
20
In the series tracking mode, the maximum output voltage of
both the “A” and “B” supplies can be simultaneously varied
with one control. The maximum “B” supply voltage can be
set to the same value of the “A” supply by using the “A”
VOLTAGE control.
The output voltage (across the two supplies) is actually
double the Volt meter reading. The actual output current
would be the value read from the mA meter (since the two
supplies are wired in series, current flowing through each
supply must be equal).
1. Set the power supplies to the TRACKING SERIES mode by
setting the TRACKING/INDEPENDENT switch to the
SERIES (left) position.
2. Set the A/B Metering switch to the A (up) position.
3. The “B” VOLTAGE AND CURRENT controls are
disabled; turn both to their minimum positions. The
maximum current is set using the “A” CURRENT control.
Follow the instructions for “Setting Current Limit”
(INDEPENDENT USE OF “A” OR “B” SUPPLY section of
this manual) using the “A” CURRENT control.
4. Adjust the output voltage to the desired level using the “A”
VOLTAGE control (remember that the actual output voltage
is double the reading on the Volt meter).
5. Turn off the power supply and the equipmen
during hook-up.
6. Connect the positive polarity of the device being powered to
the red (+) terminal of the “A” power supply.
7. Connect the negative polarity of the device being powered
to the black (-) terminal of the “B” power supply.
t to be powered
21
OPERATING INSTRUCTIONS
Figure 8. Series tracking (0-to-48 V) operation
grounding possibilities (sheet 3 of 3).
8. Fig. 8 illustrates the grounding possibilities when the
unit is used as a 0-to-48 \volt supply.
a. If the negative polarity of the equipment or circuit
being powered is also the chassis or common, it
may be grounded to earth by connecting the black
(-) terminal of the “B” supply to the green (GND)
terminal as shown in Fig. 8A.
OPERATING INSTRUCTIONS
b. Similarly, the positive polarity can be grounded by
strapping the red (+) terminal of the “A” supply to
the green (GND) terminal as shown in Fig. 8B.
c. If “split supply” operation is desired, a positive and
negative voltage with a center ground can be
achieved by strapping the black (-) terminal of the
“A” supply to the green (GND) as shown in Fig. 8C.
See the APPLICATIONS section of this manual for
more information on “split supply” operation.
d. If an earth ground reference is not required, the
configuration of Fig. 8D may be used. The scheme
in Fig. 8D should also be used where it is not known
whether the chassis is common with either the
positive or negative polarity.
e. If the chassis or common of the equipment being
powered is separate from both the positive and
negative polarity power inputs, use the connection
shown in Fig. 8E.
9.Observe proper polarity. If the circuit being powered is
not equipped with reverse polarity protection, damage
to the circuit can result from reverse polarity. Use
color coded hook-up leads, for convenience in
identifying polarity, red for (+) and black for (-).
10. Make sure that the hook-up leads offer sufficient
current capability and low resistance between the
power supply and the circuits being powered. The
hook-up leads supplied with the power supply are rated
for 4 amps.
PARALLEL TRACKING OPERATION
In the parallel tracking mode of operation, both supplies
are strapped together (in parallel). This allows for a 0-24 V
supply with a 1 amp current capability. Only the “A”
output terminals are used for parallel tracking operation. In
the parallel tracking mode, the “B” supply output voltage
and current track the “A” supply output voltage and
current.
1. Set the power supplies t TRACKING PARALLEL mode
by setting the TRACKING/INDEPENDENT switch to
the PARALLEL (middle) position.
2. Set the A/B Metering switch to the A (up) position.
Output voltage will be read from the Volt meter. Output
current is exactly DOUBLE the value read from the mA
meter (because
of current).
3. The “B” VOLTAGE and CURRENT controls are
disabled; turn both to their minimum positions. The
maximum current and voltage are set using the “A”
controls. Using the “A” supply output jacks, follow the
instructions for “Setting Current Limit”
(INDEPENDENT USE OF “A” OR “B” SUPPLY
paragraph of this section). Remember that the actual
current output at the “A” supply output jacks is double
the reading on the mA meter.
4. Adjust the output voltage to the desired level using the
“A” VOLTAGE control.
5. Turn off the power supply and the equipment to be
powered during hook-up.
6. Connect the positive polarity of the device being
powered to the red (+) terminal of the “A” power supply.
7. Connect the negative polarity of the device being
powered to the black (-) terminal of the “A” power
supply.
8. Fig. 9 illustrates the grounding possibilities when used in
22
the TRACKING PARALLEL mode.
each supply is providing the same amount
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
a. If the negative polarity of the equipment or circuit
being powered is also the chassis or common, it may
be grounded to earth by strapping the black (-)
terminal to the green (GND) terminal as shown in Fig.
9A.
b. Similarly, the positive polarity can be grounded by
strapping the red (+) terminal to the green (GN D)
terminal as shown in Fig. 9B.
c. If an earth ground reference is not required, the
configuration of Fig. 9C may be used. The scheme in
Fig. 9C should also be used where it is not known
whether the chassis is common with either the positive
or negative polarity.
d. If the chassis or common of the equipment being
powered is separate from both the positive and
negative polarity power inputs, use the connection
shown in Fig. 9D.
9.Observe proper polarity. If the circuit being powered is not
equipped with reverse polarity protection, damage to the
circuit can result from reverse polarity. Use color coded
hook-up leads, for convenience in identifying polarity, red
for (+) and black for (-).
10. Make sure that the hook-up leads offer sufficient current
capability and low resistance between the power supply
and the circuits being powered. The hook-up leads
supplied with the power supply are rated for 4 amps.
FIXED 5 V POWER SUPPLY OPERATION
The FIXED 5 V supply provides a 4 amp current capacity.
The supply is ideal for use with TTL circuits.
1. Turn off the power supply and the equipment to be powered
during hook-up.
2. Connect the positive polarity of the device being powered to
the red (+) terminal of the FIXED 5 V supply.
3. Connect the negative polarity of the device being powered
to the black (-) terminal of the FIXED 5 V supply.
4. Fig. 10 illustrates the grounding possibilities of the FIXED
5 V supply.
a. If the negative polarity of the equipment or circuit
being powered is also the chassis or common, it may be
grounded to earth by connecting the black (-) terminal
to the green (GND) terminal as shown in Fig. 10A.
b. Similarly, the positive polarity can be grounded by
connecting a jumper between the red (+) terminal and
either green (GND) terminal as shown in Fig. 10B.
c. If an earth ground reference is not required, the
configuration of Fig. 10C may be used. The scheme in
Fig. 10C should also be used where it is not known
whether the chassis is common with either the
positive or negative polarity.
d. If the chassis or common of the equipment being
powered is separate from both the positive and
negative polarity power inputs, use the connection
shown in Fig. 10D.
5. Observe proper polarity. If the circuit being powered is not
equipped with reverse polarity protection, damage to the
circuit can result from reverse polarity. Use color coded
hook-up leads, such as the sets supplied with the power
supply, for convenience in identifying polarity, red for (+)
and black for (-).
25
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
OPERATING INSTRUCTIONS
Figure 10. Grounding possibilities for fixed 5 V power supply (sheet 1 of 2).
26
OPERATING INSTRUCTIONS
Figure 10. Grounding possibilities for fixed 5 V power supply (sheet 2 of 2).
27
OPERATING INSTRUCTIONS
6. Make sure that the hook-up leads offer sufficient current
capability and low resistance between the power supply
and the circuits being powered. The hook-up leads
supplied with the power supply are rated for 4 amps.
7. If the red OVERLOAD indicator lights, too much load
has been placed on the supply. This will cause voltage
and current to drop and prevent proper operation of the
FIXED 5 V supply. To correct this situation, the load
on the supply much be decreased so that no more than 4
amps of current are required.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
NOTE
If decreasing the load does not cause the
overload indicator to turn off, the
overvoltage protection circuitry has turned
on. In order to return the supply to normal
operation, the output voltage must be
decreased (or the external voltage source
must be removed) and the power must be
momentarily shut off.
28
APPLICATIONS
GENERAL
The Model 1651A power supply has a very wide variety of
applications in electrical and electronics servicing, engineering
laboratories, manufacturing and testing facilities, schools, and
home hobbying. The “A” and “B” power supply outputs are
fully adjustable from 0-to-24 volts and 0-to-0.5 amps and the
FIXED 5 V supply has a current capability of 0-to-4 amps.
This flexibility makes it suitable for most applications requiring
a dc power source.
ELECTRONICS SERVICING
Most electronics troubleshooting and repair is performed on a
test bench. This power supply can provide the dc power source
to operate a module or circuit board on the test bench when it is
removed from its parent equipment. It can be used to power
portable, battery-operated equipment and check the effect of
low battery voltage. It can power some vehicular equipment
such as tape players, auto sound systems, CB radios, etc. on the
test bench. Parallel tracking supplies up to 1 amp.
Most automobiles and other vehicles use 12-volt electrical
systems. Although the electrical system is normally referred to
as a 12-volt system, actual battery voltage when fully charged
is approximately 13.8 volts. The power supply may be set to
13.8 volts for servicing equipment from vehicles with 12-volt
electrical systems. Some trucks use a 24-volt electrical system;
bench testing of equipment from these systems should be
performed at approximately 28 volts.
Some servicing applications require the injection of a
iabl
e dc voltage for certain tests, such as checking the
var
effect of AGC bias in a television receiver. This requires an
isolated dc power supply, such as the Model 1651A. The
equipment being tested may contain its own power supply
and operate from ac power. A dc voltage may already be
present in the circuit. One polarity of the power supply
output is floated to an appropriate point in the circuit, such as
the emitter of a transistor. The other polarity of the power
supply output is then applied to another point in the circuit,
such as the base of that transistor. Varying the power supply
voltage then varies the dc bias on the stage, and the effects
may be noted. A series limiting resistor is often used to
protect the circuits from overdissipation.
ELECTRONICS MANUFACTURING
In electronics manufacturing facilities, the power supply is
often used as a dc power source while testing and adjusting
modules, subassemblies, and complete units in the
production and assembly area or in the quality control area.
The instrument can be used in incoming inspection as a dc
power source for testing purchased components and
subassemblies.
This power supply is particularly well suited for
manufacturing applications because of its ease of operation
and its continuous duty rating. When load current or total
power dissipation are among the main characteristics to be
measured, the total load current and voltage are
simultaneously displayed on the panel meters. The current
limit can be set so that all units which do not meet the load
current specification will cause the CONSTANT
29
CURRENT indicator to light, and the unit can be rejected.
APPLICATIONS
ELECTRONICS DESIGN LAB
The technician or engineer working in an engineering
laboratory requires a dc power supply to power breadboard and
prototype circuits. This power supply is ideal because it
monitors output current and voltage, limits current to protect
the circuit, is adjustable over a wide range, and has excellent
regulation and very low ripple.
Use of the instrument in an engineering laboratory is very
similar to that described for servicing electronics equipment
and modules, except that lower currents may be prevalent when
powering individual circuits. The current limiting feature is
very valuable in this application because it can protect
unproven circuits from damage.
ELECTRONICS EDUCATION
The student in an electronics curriculum may use the power
supply for powering equipment and circuits as previously
described for all other applications. In addition, the power
supply can be used in the classroom laboratory to conduct
experiments in fundamental electronics. In learning Ohm’s law,
for example, the relationships of resistance, current, and voltage
are easily demonstrated by the use of a power supply.
BATTERY CHARGING
The power supply can be used as a battery charger to restore
the charge in rechargeable batteries such as lead-acid, nickelcadmium, and some alkaline types. Refer to the battery
manufacturer’s charging specifications for proper voltage and
current settings.
Charging information is sometimes printed on the batteries.
Battery charging, at least initially, requires the constant
current mode of operation. Before connecting the power
supply to the battery, preset the VOLTAGE controls to the
fully charged terminal voltage specified by the battery
manufacturer. Turn off the power supply while connecting
the battery. Observe proper polarity and connect as for
constant current operation. Adjust the CURRENT control
for the maximum charging current specified by the battery
manufacturer (If the maximum charging current is greater
than the power supply’s maximum load current, set the
CURRENT control to maximum). The CONSTANT
CURRENT indicator will light and the battery will charge at
the preset current limit. As the battery approaches full
charge, its terminal voltage will approach that of the power
supply output and the charging current will taper off. The
power supply may automatically switch to constant voltage
operation. When this occurs, the power supply will continue
to provide a trickle charge.
SPLIT SUPPLY
Frequently, “split power supplies” are required for
amplifiers and other electronic circuits. The Model 1651A is
ideally suited for “split power supply” operation. This supply
can be configured to provide two positive voltages with a
common negative, two negative voltages with a common
positive, or one positive and one negative with a common
ground. In addition, each of these configurations can be
obtained with identical or differing voltages.
30
Two Identical Positive Voltages With a Common
Negative (Refer To Fig. 11)
Some electronic equipment requires two identical positive
voltages with a common negative. A good example of this
would be a digital car clock where there are two +12 volt
inputs and a common negative. Using both supplies in the
parallel tracking mode would provide the simplest hook-up
and operation. This type of “split supply” operation is
obtained as follows:
1. Select the TRACKING PARALLEL operating mode and
set the A/B Metering to monitor the “A” supply.
2. Set the desired voltage and maximum current using the
“A” VOLTAGE and CURRENT controls.
3. Connect a ground wire between the “A” supply’s
negative terminal and the GND (green) terminal.
4. Turn off the power supply and the equipment to be
powered during hook-up.
5. Connect the positive polarity inputs of the circuit to be
powered to the positive (red) terminals of the supplies
and connect the common negative input of the circuit to
be powered to the “A” supply’s negative (black) or GND
(green) terminal.
APPLICATIONS
31
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Fig. 11. Typical Hook-Up Using Two Identical
Positive Voltages and a Common Negative.
APPLICATIONS
Two Differing Positive Voltages With a Common
Negative (Refer To Fig. 12)
Many electronic circuits require two different positive
voltages with a common negative. A typical example of this
would be a device that uses both TTL (+5 V) and analog
(typically +15 V) circuitry. Using both supplies, two differing
positive voltages with a common negative are obtained as
follows:
1. Select the INDEPendent operating mode and set the A/B
Metering switch to monitor the “A” supply.
2. Set the desired voltage and maximum current for the “A”
supply using the “A” VOLTAGE and CURRENT
controls.
3. Set the A/B Metering switch to monitor the “B” supply.
4. Set the desired voltage and maximum current for the “B”
supply using the “B” VOLTAGE and CURRENT
controls.
5. Connect the ground straps between each supplies’
negative terminal and the GND (green) terminal.
6. Turn off the power supply and the equipment to be
powered during hook-up.
7. Connect the positive polarity inputs of the circuit to be
powered to the positive (red) terminal of the supply.
Connect the common negative input of the circuit to be
powered to either the supply’s negative (black) or GND
(green) terminal.
NOTE
The example in Fig. 12 uses the “B”
supply to provide the +5V, so that
current can be monitored on the mA
32
Fig. 12. Typical Hook-Up Using Two Differing
Positive Voltages and a Common Negative.
meter. If the current requirements of
the +5 circuits exceed 0.5A, then the
FIXED 5 V output should be used.
Two Identical Negative Voltages With a Common
Positive (Refer To Fig. 13)
When the same negative voltage is required at two points
in the same circuit and a common positive is needed,
perform the following:
1. Select the TRACKING PARALLEL operating mode
and set the A/B Metering switch to monitor the “A”
supply.
2. Set the desired voltage and maximum current using the“A” VOLTAGE and CURRENT controls.
3. Connect a ground wire between the “A” supply positive
terminal and the GND (green) terminal.
4. Turn off the power supply and the equipment to be
powered during hook-up.
5. Connect the negative polarity inputs of the circuit to be
powered to the negative (black) terminals of the
supplies. Connect the common positive input of the
circuit to be powered to the “A” supply’s positive (red)
or GND (green) terminal.
APPLICATIONS
Fig. 13. Typical Hook-Up Using Two Identical
Negative Voltages and a Common Positive.
33
APPLICATIONS
Two Differing Negative Voltages With a Positive
Common (Refer To Fig. 14)
Using both supplies, two differing negative voltages with a
common positive are obtained as follows:
1. Select the INDEPendent operating mode and set the A/B
Metering switch to monitor the “A” supply.
2. Set the desired voltage and maximum current for the “A”
supply using the “A” VOLTAGE and CURRENT
controls.
3. Set the A/B Metering switch to monitor the “B” supply.
4. Set the desired voltage and maximum current for the “B”
supply using the “B” VOLTAGE and CURRENT
controls.
5. Connect the ground wires between each supplies’
positive terminal and GND (green) terminal.
6. Turn off the power supply and the equipment to be
powered during hook-up.
7. Connect the negative polarity inputs of the circuit to be
powered to the negative (black) terminals of the supplies.
8. Connect the common positive input of the circuit to be
powered to either supplies’ positive (red) or the GND
(green) terminal.
34
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Fig. 14. Typical Hook-Up Using Two Different
Negative Voltages and a Common Positive.
Identical Positive and Negative Voltages With a
Separate Common (Refer To Fig. 15)
Another typical “split supply” application is when a circuit
uses operational amplifiers (op-amps). Typically, identical
positive and negative voltages are required to power op-amp
circuits. Using both supplies and the series tracking mode of
operation, identical positive and negative voltages with a
separate common are obtained as follows:
1. Select the Tracking Series operating mode and set A/B
Metering switch to monitor the “A” supply.
2. Set the desired voltage using the “A” VOLTAGE
controls.
3. Connect a ground wire between the “A” supply negative
terminal and the GND (green) terminal.
4. Turn off the power supply and the equipment to be
powered during hook-up.
5. Connect the positive polarity input of the circuit to be
powered to the positive (red) terminal of the “A” supply
and connect the negative polarity of the circuit to the
negative terminal of the “B” supply. Connect the circuit
ground to the ground terminal of the “A” supply, the
positive terminal of the “B” supply, or the GND (green)
terminal.
APPLICATIONS
35
Fig. 15. Typical Hook-Up Using Identical Positive
and Negative Voltages with a Separate Common.
APPLICATIONS
Differing Positive and Negative Voltages With a
Separate Common (Refer To Fig. 16)
Using both supplies in the independent mode of operation,
different positive and negative voltages with a separate
common are obtained as follows:
1. Select the INDEPENDENT operating mode and set the
A/B Metering switch to monitor the “A” supply.
2. Set the desired voltage and maximum current on the “A”
supply using the “A” VOLTAGE and CURRENT
controls.
3. Set the A/B Metering switch to monitor the “B” supply.
4. Set the desired voltage and maximum current on the “B”
supply using the “B” VOLTAGE and CURRENT
controls.
5. Connect one ground wire between the negative terminal
of the “A” supply to the positive terminal of the “B”
supply and another ground wire between the positive
terminal of the “B” supply and the GND (green)
terminal.
6. Turn off the power supply and the equipment to be
powered during hook-up.
7. Connect the positive polarity input of the circuit to be
powered to the positive (red) terminal of the “A” supply
and connect the negative polarity of the circuit to the
negative terminal of the “B” supply. Connect the circuit
ground to the negative terminal of the “A” supply or the
GND (green) terminal.
Fig. 16. Typical Hook-Up Using Different Positive
and Negative Voltages and a Separate Common.
36
MAINTENANCE
WARNING
The following instructions are for
use by qualified personnel only.
To avoid electrical shock, do not
perform any servicing other than
contained in the operating
instructions unless you are
qualified to do so.
FUSE REPLACEMENT
If the fuse blows, the LED indicator will not light and
the power supply will not operate. The fuse should not
normally open unless a problem has developed in the unit.
Try to determine and correct the cause of the blown fuse,
then replace only with a fuse of the correct rating. For
110 or 120V operation a 1.6A, 250 V, 3AG fuse should
be used and for 220 or 240V operation a 0.75A, 250V,
3AG fuse should be used. The fuse is located on the rear
panel (see Fig. 2).
Figure 17. Line voltage conversion switch,
determined by fuse holder arrow position.
37
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
38
39
TEST INSTRUMENT SAFETY
(continued from inside front cover)
4.If possible, familiarize yourself with the equipment being tested and the location of its high voltage points. However,
remember that high voltage may appear at unexpected points in defective equipment.
5.Use an insulated floor material or a large, insulated floor mat to stand on, and an insulated work surface on which to place
equipment; and make certain such surfaces are not damp or wet.
6.Use the time-proven “one hand in the pocket” technique while handling an instrument probe. Be particularly careful to
avoid contacting a nearby metal object that could provide a good ground return path.
7.When testing ac powered equipment, remember that the ac line voltage is usually present on some power input circuits such
as the on-off switch, fuses, power transformer, etc. any time the equipment is connected to an ac outlet, even if the
equipment is turned off.
8.Some equipment with a two-wire ac power cord, including some with polarized power plugs, is the “hot chassis” type. This includes most
recent television receivers and audio equipment. A plastic or wooden cabinet insulates the chassis to protect the customer. When the
cabinet is removed for servicing, a serious shock hazard exists if the chassis is touched. Not only does this present a dangerous shock
hazard, but damage to test instruments or the equipment under test may result from connecting an earth ground lead of a test instrument to
a “hot chassis”. To make measurements in “hot chassis” equipment, always connect an isolation transformer between the ac outlet and the
equipment under test. The B+K Precision Model TR-110 or 1604 Isolation Transformer, or Model 1653 or 1655 AC Power Supply is
suitable for most applications. To be on the safe side, treat all two-wire ac powered equipment as “hot chassis” unless you are sure it has
an isolated chassis or an earth ground chassis.
9.B+K Precision products are not authorized for use in any application involving direct contact between our product and the
human body, or for use as a critical component in a life support device or system. Here, “direct contact” refers to any
connection from or to our equipment via any cabling or switching means. A ”critical component” is any component of a life
support device or system whose failure to perform can be reasonably expected to cause failure of that device or system, or
to affect its safety or effectiveness.
10. Never work alone. Someone should be nearby to render aid if necessary. Training in CPR (cardio-pulmonary resuscitation)
first aid is highly recommended.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.