Digital Sound Level Meter
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

CONTENTS
Title P age
I. SAFETY INFORMATION………..……….………….. 1
Environment conditions…………………………….....1
Maintenance & Clearing……………………………….1
Safety symbols…………………………………………
1
II. GENERAL DESCRIPTION……………………..……..
2
III. SPECIFICATIONS..…………………..……………….. 2
IV. NAME AND FUNCTIONS..………….……………….. 5
V. CLOCK & INTV SETUP……….……………….……… 6
DataLogger……………………………………………..
6
Clock Setup……………………………………………...6
Recording Interval Setup………………………………7
Auto Power Off…………………………………………
.7
VI. CALIBRATION PROCEDURES….………………….. 7
VII . MEASUREMENT PREPARATION………….………. 8
VIII. OPERATING PRECAUTIONS………………………. 8
IX. MEASUREMENT………….………………………..… 9
X. Setup TestLink AK-73X (Sound Level Meter)
RS232 interface software……………………..….…...9
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

I. SAFETY INFORMATION
Read t he f ollowing s afety information c arefully b efore at tempting t o oper ate or
service the meter.
Use the meter only as specified in this manual; otherwise, the protection provided
by the meter may be impaired.
Environment conditions
Altitude up to 2000 meters
Relatively humidity 90% max.
Operation Ambient 0° ~ 40°C
Maintenance & Clearing
Repairs or s ervicing n ot c overed in t his m anual s hould o nly be p erformed b y
qualified personnel.
Periodically wipe the case with a dry cloth. Do not use abrasives or solvents on
this instruments.
Safety symbols
Comply with EMC
When servicing, use only specified replacement parts.
II. GENERAL DESCRIPTION
Thank you for using our Data Logger Sound Level Meter. To ensure that you can
get the most from it, we recommend that you read and follow the manual carefully
before use.
This unit conforms to the IEC651 type2, ANSI S1.4 Type2 for Sound Level Meters.
This D ata Lo gger Sound Level Meter i nternal m emory c an k eep up t o 32 000
records.
(Note 1.)
It uses RS232 interface to perform bi-directional communication with PC.
Note1 (model 735):
Press “ REC” t o begi n r ecording, press “REC” again to stop recording. The dat a w ill be l oaded i nto
the memory until it is full.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

III. SPECIFICATIONS
Standard applied: IEC651 Type2, ANSI S1.4 Type2.
Frequency range: 31.5Hz ~ 8KHz
Measuring level range: 30 ~ 130dB
Frequency weighting: A / C
Microphone: 1/2 inch electret condenser microphone
Display: LCD
Digital display: 4 digits
Resolution: 0.1dB
Display Update: 0.5 sec.
Analog display: 50 segment bargraph
Resolution: 1dB
Display Update: 50 mS
Time weighting: FAST (125mS), SLOW (1 sec.)
Level ranges: Lo: 30 – 80 dB
Med: 50 – 100 dB
Hi: 80 – 130 dB
Auto: 30 – 130 dB
Accuracy: ±1.5dB (under reference conditions @ 94dB, 1KHz)
Dynamic range: 100 dB (50 dB model 732A)
Alarm function: “OVER” is when input is more than the upper limit of range.
“ UNDER ” is when input is less than the lower limit of range.
MAX/MIN hold: Holds readings the Maximum and Minimum Value.
AC output: 1 Vrms at FS (full scale).
Output impedance: Approx. 100Ω
FS: means the upper limit of each level range.
DC output: 10mV / dB, output impedance approx. 1KΩ
Power supply: One 9V battery, 006P or IEC 6F22 or NEDA 1604.
Power life: About 50hrs (alkaline battery)
AC adapter: Voltage 9 VDC, center pin negative, (8-15VDC Max)
Supply current : > 30mADC Socket : pin → Ground
Casing → Positive E xternal Diameter → 3.5mm
Internal Diameter → 1.35mm
Electromagnetic Compatibility:
RF field = 3V/m
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Total accuracy = specified accuracy + 0.5dB
Operation temperature: 0 to 40°C (32° to 104°F)
Operation humidity: 10° to 90%RH
Storage temperature: -10° to 60°C (14° to 140°F)
Storage humidity: 10 to 75%RH
Dimensions (L x W x H): 10.8 × 2.5 × 1.2inch (275 × 64 × 30mm)
Weight: 285g (including battery)
Accessories: 9V battery, carrying case, Screwdriver, Instruction manual.
Windscreen, 3.5 ψ plug, software & RS-232 cable (optional
model 732A).
IV. NAME AND FUNCTIONS
OPEN
MANUAL FOR SAFETY
NEDA 1604 6F22 006P
PLEASE READ
9V BATTERY
2
1
3
4
5
7
6
9
RS-232
CAL
OUTPUT
DC 9V
SETUP
POWER-UP OPTIONS
30Lo =
Med=
Auto=
Hi =
50
30
80
REC
MAX
MIN
CLOCK
80 dB
100 dB
130 dB
130 dB
LEVEL
SLOW
INTV
FAST
C
A
8
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

1
Windscreen
If you operate at wind speed over 10m/sec, please put protective accessories
in front of the microphone.
2
Display
SYMBOL FUNCTION
LCD 4 digits
MAX Max imum indication
MIN Mi nimum indication
Over range
Under range
FAST F ast response
SLOW Sl ow response
dBA A-Weighting
dBC C -Weighting
Range Indicate
Low-Battery
Under range 20dB
Following items for model 735
AUTO A uto Level range selective
REC
Recording Datalogger
FULL
Memory full
Auto Power OFF active
3
Power & Backlight button
The
○I key turns the sound level meter ON or OFF and backlight ON & OFF.
Press it once to turn on the sound level meter.
Press it again for moment to turn ON or OFF backlight.
Press and hold this button 3 seconds to turn the unit OFF.
Note: When the user powers it on, the LCD will show how much memory space
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com