B&K Precision 5492BGPIB Datasheet

Data Sheet
True RMS Bench Multimeter
5492B
True RMS Bench Multimeter
The B&K Precision model 5492B is a versatile 5½-digit, 120,000-count bench multimeter suitable for applications in education, service repair, and manufacturing. The instrument enhances your productivity with built-in math functions and USB connectivity. Rel, Max/Min, dBm, dB, %, and Hold functions provide educators with a convenient tool to teach basic math concepts. Additionally, the 5492B offers powerful features not commonly found in other 5½- digit DMMs such as advanced triggering, buffer storage operation, and a GPIB interface option.
Increase Productivity with PC Connectivity
The 5492B is programmable via USB, RS232, and GPIB (5492BGPIB only) interface using industry standard SCPI commands. Users can remotely control and configure the instrument from a PC and retrieve measurement results for further analysis. The multimeter can also be remotely controlled using application software (via USB & RS232), which supports front panel emulation and data logging of measurement results.
The 5492B measures volts and amps with great accuracy and stability, with basic VDC accuracy of 0.01%. The meter is also capable of measur­ing frequency, period, resistance, continuity, and performing diode tests. Readings can be taken at a maximum rate of 57 readings/sec and stored to an internal 512-byte buffer for post acquisition statistical analysis.
The 5492B was designed for cost-conscious users requiring an accurate meter with a broad range of features offered at a value price.
Features & benefits
5½-digit, 120,000-count display resolution
0.01% basic VDC accuracy
Advanced triggering options such as external,
bus and event trigger
Two and four-wire resistance measurement up
to 120 M
AC voltage and AC current measurement over
wide frequency range (ACV 100 kHz/ACI 10
kHz)
AC (RMS) and DC current measurements up
to 12 A
Limit mode for Pass/Fail testing
Save/recall up to 10 instrument settings
Built-in math functions: Rel, Max/Min, dBm,
dB, %, Hold, Compare
CATI (1000 V)/CATII (300 V) protection
USB (Virtual Com) and RS232 interface
SCPI compatible
Technical data subject to change © B&K Precision Corp. 2014
ww w .bk pre c ision .c om
True RMS Bench Multimeter
odel 5492B
M
Versatile tools
Limit Operation
The limit operation lets you set and control the values that determine a HI / IN / LO status of subsequent measurements. The meter can be configured to emit an audible alarm when readings are outside of the configured limit.
Advanced Triggering
A variety of advanced trigger options are available: Manual Trigger via a front panel button press, Bus Trigger sending a *TRG command via PC interface, External Trigger using the EXT TRIG input terminal on the rear panel. Additionally, a trigger delay can be programmed after an event detection, and a VM Comp output is available to indicate the instrument’s completion of a present measurement.
Easy operation
Front panel
right VFD display
B
nput banana terminals
I
mA Fuse socketInterface keypadPower button
Fast Buffering with Built-in Statistics
The 5492B provides an internal buffer capable of storing up to 512 readings. Buffer storage operation can be enabled via front panel or remote command. The buffer fills up with the requested number of readings at a rate of up to 57 readings/sec and then stops. The instrument automatically calculates statistical information based on the stored readings such as minimum, maximum, or average reading, and standard deviation, which can then be retrieved along with the readings via the front panel or remote command.
Rear panel
Serial number label
Power line
fuse holder
VM Comp
(Voltmeter complete) BNC output terminal
GPIB (IEEE-488) interface
(model 5492BGPIB)
External
Trigger BNC
input terminal
RS-232
(Serial) interface
USB interface
Chassis ground
screw terminal
2 ww w.b k pr ec is ion .c om
True RMS Bench Multimeter
odel 5492B
M
Specifications
pecifications are based on the following conditions:
S
One year calibration cycle
Operating temperature between 18 °C to 28 °C
Accuracy expressed as: ± (% of reading + % of range) after 30 minutes warm-up and valid for 10 PLC (Slow)
Temperature coefficient: add ± [0.1% × (applicable accuracy)/°C] for 0 °C to 18 °C and 28 °C to 40 °C
Relative humidity: Up to 80% RH for 0 °C to 28 °C (75% RH for 10 M and above ranges
for resistance measurement), up to 70% RH for 28 °C to 40 °C
DC Voltage
Resolution, Full Scale Reading and Accuracy ± (% of reading + % of range), 23 °C ± 5 °C
Ra ng e Re solu tio n
20.000 mV
1
µV
1
Ful l S c al e
Re a d ing
19.999
1
1.20000 V 10 µV 1.19999 0.01+0.004
2.0000 V
1
20.000 V 1 mV 119.999 0.01+0.004 10 M ±1%
1
000.00 V
1
1) under REL status
(
00 µV
1
0 mV
1
1.9999
1
010.00
1
(2)
(2) 1% overrange (1010 V) is readable at 1000 V range
Input protection: 1000 V
Ac c urac y
(1 y ear)
.02+0.008
0
.01+0.004
0
.01+0.004
0
(1)
1)
(
Ty p ica l In p ut
Im pe da nce
10 G
> >10 G
10 G
>
0 M ±1%
1
AC Voltage (True RMS, AC coupled)
Resolution, Full Scale Reading and Accuracy ± (% of reading + % of range), 23 °C ± 5 °C
Ra nge Re sol u tio n
Fu l l Sca le
Re adi n g
10 -20 Hz 20 -50 H z 50 -20 kHz 20 -50 kHz 50 -10 0 k Hz
Acc urac y (1 year)
120.000 mV 1 µV 119,999 1.50+0.100 0.50+0.100 0.10+0.100 0.30+0.150 1.0+0.150
1.20000 V 10 µV 1.19999 1.50+0.100 0.50+0.100 0.10+0.100 0.30+0.150 1.0+0.100
12.0000 V 100 µV 11.9999 1.50+0.100 0.50+0.100 0.10+0.100 0.30+0.150 1.0+0.100
120.000 V 1 mV 119.999 1.50+0.100 0.50+0.100 0.10+0.100 0.30+0.150 1.0+0.100
750.00 V 10 mV 757.50
(1) Specifications are for sine wave inputs >5% of range (2) 3×10
7
Volt-Hz for 750 V range
3)
(
1.50+0.100 0.50+0.100 0.10+0.100 0.30+0.150
(3) 1% overrange (757.5V) is readable at 750 V range
Measurement method: AC coupled true RMS, measuring the AC component only Maximum Crest factor: 3.0 at full scale Input impedance: 1 M ± 2% in parallel with capacitance <100 pF Input protection: 750 Vrms, 500 Vdc
(1 )
2)
(
1.0+0.100
DC Current
Resolution, Full Scale Reading and Accuracy ± (% of reading + % of range), 23 °C ± 5 °C
Ra nge Re solu tio n
Fu l l Sca le
Re adi n g
12.0000 mA 0.1 µA 11.9999 0.05+0.008
120.000 mA 1 µA 119.999 0.05+0.004
1.20000 A 10 µA 1.19999 0.10+0.004
12.0000 A
(3)
100 µA 11.9999 0.25+0.004
Ac c urac y
(1 y ear )
Bur d en Vo lta ge & Sh u n t Re sis t o r
(2)
<0.15 V / 10.1 Ω
(2)
<1.5 V / 10.1 Ω
<0.3 V / 0.1 Ω
<0.15 V / 10 mΩ
(1) Typical voltage across the input terminals at full scale reading (2) Use REL function (3) In 12 A range, >10~12 ADC is readable for 20 seconds maximum
Input protection: Externally accessible 2A / 250V fuse protecting the low input current terminal (LO). Internally accessible 20A / 250V fuse protecting the high current input terminal (10A)
(1)
1)
(
3 ww w.b k pr ec is ion .c om
True RMS Bench Multimeter
odel 5492B
M
Specifications (cont.)
AC Current (True RMS, AC Coupled)
esolution, Full Scale Reading and Accuracy ± (% of reading + % of range), 23 °C ± 5 °C
R
Ra nge R eso lut ion
2.0000 mA
1
.20000 A
1
2)
(
12.0000 A
(1) Typical voltage across the input terminals at full scale reading (2) In 12 A range,>10-12 A AC is readable for 20 seconds maximum
3) Specifications are for sine wave inputs >5% of range
(
easurement method: AC coupled true RMS, measuring the AC component only
M
nput protection: Externally accessible 2 A / 250 V fuse protecting the low input current terminal (LO). Internally accessible 20 A / 250 V fuse protecting the
I high current input terminal (10A)
ax. crest factor: 3.0 at full scale
M
.1 µA
0
0 µA
1
100 µA 11.9999 <0.15 V / 10 m 1.0+0.080 0.50+0.080 0.25+0.080 2.0+0.080
Ful l S ca le
Re a ding
1.9999
1
.19999
1
Bu rde n Volta g e & Shun t R esis t o r
0.15 V / 10.1
<
0.3 V / 0.1
<
Resistance (2-Wire and 4-Wire)
esolution, Full Scale Reading and Accuracy ± (% of reading + % of range), 23 °C ± 5 °C
R
(1)
Ra n g e
120.000 1 m 119.999 1 mA 0.05 +0.008
1.20000 k 10 m 1.19999 1 mA 0.03 +0.004
12.0000 k 100 m 11.9999 100 µA 0.03 +0.004
120.000 k 1 119.999 10 µA 0.03 +0.004
1.20000 M 10 1.19999 10 µA 0.03 +0.004
12.0000 M 100 11.9999 7.0/(10M+Rx) µA 0.10 +0.004
120.000 M 1 k 119.999 7.0/(10M+Rx) µA 0.50 +0.008
(1) In order to eliminate the noise interference, which might be induced to the test leads, it is recommended to use a shielded test cable for measuring resistance above 120 k (2) Using REL function
Input protection: 1000 Vdc or 750 Vac for all ranges Note: Rx is the measured resistance value
Re sol u tio n Full Sc al e Re adi n g Tes t C urr e n t
(1 )
10 -20 H z 20 -50 H z 50-2 kHz 2- 20 kH z
.0+0.080
1
.0+0.080
1
Accu r ac y
(1 yea r )
Ac curac y (1 yea r)
.50+0.080
0
.50+0.080
0
2)
(
2)
(
(2)
.25+0.080
0
.25+0.080
0
(3)
.0+0.080
2
.0+0.080
2
Continuity
Resolution, Full Scale Reading and Accuracy ± (% of reading + % of range), 23 °C ± 5 °C
Ra nge Re solu t ion
1 k 100 m 999.9 1 mA 0.1%+0.02%
For 0.1 PLC (Fast) Input protection: 1000 Vdc or 750 Vac
Ful l S c al e
Re a d ing
Tes t C urr e n t
Accu r ac y
(1 yea r )
Diode
Resolution, Full Scale Reading and Accuracy ± (% of reading + % of range), 23 °C ± 5 °C
Ran g e Resol uti o n
3.0000 V 100 µV 2.9999 1 mA 0.030+0.020
10.0000 V 100 µV 10.0000 100 µA 0.030+0.020
10.0000 V 100 µV 10.0000 10 µA 0.030+0.020
For 1 PLC (Medium) Input protection: 1000 Vdc or 750 Vac for all ranges
Fu l l Sca le
Rea d ing
Te st C u r ren t
4 ww w.b k pr ec is ion .c om
Ac c urac y
(1 y ear)
True RMS Bench Multimeter
odel 5492B
M
Specifications (cont.)
Frequency
esolution, Full Scale Reading and Accuracy ± (% of reading + % of range), 23 °C ± 5 °C
R
ACV
Ra nge
00 mV
1
750 V
(1) If testing frequency is greater than 1MHz, it will be displayed but no specified accuracy is guaranteed. (2) Specified accuracy at input >5 % of range.
Gate Time at 1 Sec. Input protection: 750 Vrms or 500 Vdc
Fr e quen c y R ang e Res o luti on
5 ~ 10 Hz 10 µHz 9.99999 0.05 200 mVrms
o
t
10 ~ 100Hz 100 µHz 99.9999 0.01 40 mVrms 00 ~ 100 kHz
1
100 k ~ 1 MHz
(
1 Hz 999.999 0.005 100 mVrms
mHz
1
1)
Period
Resolution, Full Scale Reading and Accuracy ± (% of reading + % of range), 23 °C ± 5 °C
AC V Rang e Pe rio d Ran ge R e s olut i on
(1)
~ 10 µs
00 mV
1
to
750 V
(1) If testing frequency is greater than 1MHz, it will be displayed but no specified accuracy is guaranteed. (2) Specified accuracy at input >5 % of range.
Gate Time at 1 Sec. Input protection: 750 Vrms or 500 Vdc
1
10 µs ~ 10 ms 0.1 ns 9.99999 0.005 40 mVrms
10 ms ~ 100 ms 0.1 µs 99.9999 0.01 40 mVrms
100 ms ~ 200 ms 1 µs 199.999 0.05 200 mVrms
.01 ns
0
Fu l l Sca le
Re adi n g
99.999
9
Fu l l Sca le
Re adi n g
.99999
9
Ac c urac y
0
Ac c urac y
.005
0
.005
In p ut Sensi tiv i ty
(2 )
(S ine Wav e )
Inp ut Sens i tiv ity
(2 )
0 mVrms
4
(S i n e Wa ve)
00 mVrms
1
Supplemental
Environmental
Op erati n g Env iro n men t
St ora ge En viro n m ent
Hu mid i ty
General
Re m o te Inte r faces
Me a surem e nt Ca teg o ry
Ma th Func tio n s
AC Inp ut
Po we r Co nsu m ptio n
Sa fet y & EMC
Di m ensi ons (W× H×D)
Ne t Wei g h t
Inc lud ed Acce s sor ies
0 °C to 40 °C, 32 ºF to 104 ºF
-40 °C to 70 °C -40 ºF to 158 ºF
90 %RH
RS232, USB (virtual COM) GPIB (model 5492BGPIB only)
CAT II 300V, CAT I 1000V, pollution degree 2
Rel, Max/Min/Average/StdDev (for stored readings), dB, dBm, Limit Test, % and mX+b
110/220 V ± 10%, 50/60 Hz ± 5%
20 VA
EN 61010-1:2001, EN61326-1:2006 and
EN61326-2-1:2006
8.85” x 3.93 “ x 13.97”
(225 mm×100 mm×355 mm)
5.51 lbs (2.5 kg)
Three-Year Warranty
Test leads, AC power cord, spare fuse, user manual, USB cable, calibration certificate
and test report