5390
page
TEST INSTRUMENT SAFETY inside front cover |
FEATURES |
CONTROLS AND INDICATORS |
OPERATING INSTRUCTIONS |
POWER ON / OFF |
PROGRAMMABLE POWERUP OPTIONS |
FUNCTION SELECTION |
RANGE SELECTION |
ANALOG BARGRAPH, ZOOM |
BACKLIGHT |
HOLD & AUTO MEM MODE (AUTOSTORE) |
RELATIVE MEASUREMENTS |
PEAK HOLD MEASUREMENTS |
SURV or MIN / MAX / AVG MODE |
VOLTAGE MEASUREMENTS |
CURRENT MEASUREMENTS |
RESISTANCE MEASUREMENTS |
CONTINUITY MEASUREMENTS |
DIODE TESTING |
2
The B+K Precision Model 5390 is a highly accurate handheld digital multimeter with a multitude of highly advanced features. It is capable of taking the following measurements:
Model 5380 has the same accuracy as the 5390 and most of the measurement capabilities. It does not include the 5390's power disturbance indicator, zoom bargraph, pulse width, duty cycle, event counting, and VA measurement. It measures dB across 600 Ω only.
Model 5370 has the same accuracy as the 5390 for all functions except dc volts. It includes all of the measurement capabilities as the 5390 except for pulse width, event counting, and VA measurement. It also is capable of temperature measurement within 0.5°C using a platinum probe.
Model 5360 has the capability of taking many of the same measurements as the 5380, but at a lower accuracy. Its measurement capabilities do not include dBm, but does include duty cycle.
All models include 50,000 count resolution, true rms (ac or ac + dc) built-in RS-232 capability, standard Certificate of Calibration, ASYC®II (Advanced Safety Concept) design with IEC1010, IP677, and MIL-T-28800 compliance, a splash-proof sealed case, and a three-year warranty.
The standard SECURX test lead retainers, designed to work with shouldered test leads, ensure that the test leads will not be pulled out accidentally.
The display features clearly legible 14 mm (0.55 inch) high figures. All models except 5360 feature a backlit display.
Refer to the SPECIFICATION section of this manual for complete specifications for each of the models.
Refer to the ACCESSORIES section of this manual for a description of each of the standard and optional accessories for the multimeters.
Figures 1 through 4 illustrate each of the multimeters. Refer to the illustrations while reviewing the following descriptions of each of the controls.
Throughout this section of the manual, instructions include short and long presses of buttons on the multimeter. A short press means less than one second and a long press means greater than one second.
Fig. 4 Model 5360 Controls
between capacitance and diode testing; toggles between ohms and continuity checking; toggles between DC and AC+DC voltage measurements; or toggles between AC voltage and dB/Hz measurements. Also used to turn on the multimeter after an automatic shutdown.
16. ZOOM Button (5370 and 5390 only). Expands the scale of the 34 segment analog bargraph by a factor of five for positive measurements and a factor of 11 for bipolar measurements.
Fig. 5 VAC Position
Fig. 11 °C Position
Figure 12 shows all the annunciators on the LCD display. Refer to this illustration while reviewing the following descriptions.
Use of test equipment may expose the operator to electric shock hazards. Observe all instructions contained in the TEST INSTRU-MENT SAFETY section of this manual before using this instrument.
The TEST INSTRUMENT SAFETY section of this manual lists maximum voltage and current input limits which must be observed. Failure to adhere to these limits may result in damage to the INSTRUMENT.
Throughout this section of the manual, instructions include short and long presses of buttons on the multimeter. A short press means less than one second and a long press means greater than one second.
Turn the unit on by rotating the Function Switch clockwise to the desired function. All the LCD display segments come on for a few seconds, then only the appropriate annunciators for the selected function stay on.
Turn the unit off by rotating the Function Switch to the OFF position. The AUTO POWER OFF feature prolongs battery life by automatically shutting off the unit after approximately half an hour if no switch is operated. To turn on the unit after an AUTO POWER OFF, press the yellow SEL/ON button.
The AUTO POWER OFF feature is disabled when the unit is set to the SURV , PK +/- , pulse counting, and data print modes.
For user safety, the AUTO POWER OFF is also disabled when a dangerous voltage is present at the input jacks. The A indicator illuminates to signify the condition.
There are several operating modes that can be selected upon powerup. Typically, these modes are infrequently changed and are stored in nonvolatile memory so that the selection is retained until reprogrammed.
1. 50 Hz or 60 Hz rejection.
Hold down the HOLD button and set the rotary Function Switch to the appropriate position. The selection reverses the previous setting, is displayed for two seconds, and is stored in non-volatile memory.
2. 10MΩ / 1000M Ω input impedance for the mV range. (5370, 5380 and 5390 only).
Hold down the RANGE button and set the rotary Function Switch to the appropriate position. The selection reverses the previous setting, is displayed for two seconds, and is stored in non-volatile memory.
3. 5,000-count low-resolution mode.
Hold down the REL button and set the rotary Function Switch to the appropriate position. The display shows 5,000 for two seconds.
The low resolution mode is not stored in nonvolatile memory. If the unit is turned on after an AUTO POWER OFF with the SEL/ON button, the unit powers up in the normal 50,000 counthigh-resolution mode.
Hold down the PK +/- button and set the rotary Function Switch to the appropriate position. The display shows the RS232 indicator for two seconds. The RS232 mode is not stored in non-volatile memory.
The rotary Function Switch selects the primary function such as VAC, mVDC, VDC, Ω, –I–. All the Function Switch positions are dual or multifunctional with the primary function for each position indicated by WHITE labeling. The secondary functions for each position are indicated with YELLOW labeling, and are accessed with a press of the yellow SEL/ON button. For example, when the rotary Function is set to the 10ADC position, the DC current function is selected. A press of the yellow SEL/ON button puts the unit into the 10A AC (RMS) position, ready to measure AC cur-
rents with capacitive coupling. The AC annunciator illuminates on the LCD display. A second press of the SEL/ON button puts the unit into the 10A AC+DC (TRMS) position, ready to measure AC currents with direct coupling. The AC+DC annunciator illuminates on the LCD display. A third press of the SEL/ON button returns the unit back to the primary 10ADC function.
The 34 segment analog bargraph feature is activated for measurements of voltage, current, resistance, frequency, and capacitance. Its update speed of 20 measurements/second is 10 times that of the digital display. This
makes it suitable for measuring coarse adjustments of these parameters or indicating the direction of change of a varying quantity.
The ZOOM feature (5370 and 5390 only) expands the scale of bargraph around the measured digital reading to greatly enhance viewing fast changes. A short press of the ZOOM button increases this scale. Subsequent short presses keep expanding this scale until it returns back to its original full scale value. A long press of the ZOOM button at any time returns the bargraph back to full scale.
LCD displays work by reflecting light. As less ambient light is available, there is less apparent contrast between the LCD digits and the background, and it becomes more difficult to view the display. The backlight feature is useful in low ambient light conditions. A short press of the backlight (35) button turns on the display backlight. A second short press immediately turns off the backlight. If the multimeter is not being used, the backlight turns off automatically after about 30 seconds.
The HOLD button allows the user to freeze the current display or enable the AUTO MEM (Autostore) mode. A short press freezes the current display. A second short press releases the display.
The Autostore mode is used to obtain a stable voltage measurement when the user cannot look at the display and the measurement point simultaneously. Autostore is accessible in the VDC, mVDC, or VAC positions of the Function Switch.
When the REL button is pressed, the present reading becomes the zero reference and all subsequent readings are displayed relative to this value. A second short press disables the mode and returns the unit back to its normal mode. When in the REL mode, a long press displays the reference being used. The reference value may then be adjusted using the SEL/ON and RANGE buttons. The SEL/ON button is used to select the digits and the sign, while the RANGE button is used to increment the selected digit. A short press of the REL button returns the unit to the relative measurement mode. Another short press returns the unit back to its normal mode.
Fast (<1 ms) positive and negative peak measurements can be made when the rotary Function Switch is set to VDC , mVDC , mADC , and 10ADC . A short press of the PK +/- button puts the unit into the positive peak detection mode, and changes the unit from the AUTO ranging mode to the MANUAL ranging mode, locking the unit on the range. If the potential peak is greater than full scale of the selected range, use the RANGE button to set the unit to the next higher range, or until an on-scale measurement is possible. The A annunciator of the LCD display confirms the selection. If a negative peak detection is anticipated, press the PK +/- button a second time. The V annunciator of the LCD display confirms the selection. To return the unit back to its normal mode, press the PK +/- button again.
The surveillance mode is useful for recording the variation of the quantity (voltage, current, resistance, capacitance, etc.) being measured by the unit. Activation of the SURV mode locks the unit into its present range and deactivates the AUTO POWER OFF feature. If the unit is in the AUTO ranging mode, it changes to the MANUAL ranging mode. A long press of the SURV button enables or disables the recording mode. Short presses display current values of MIN, MAX, or AVG measurements. The annunciators flash with each selection.
The measurements are stored even when the SURV mode is disabled or the unit power is shut OFF. Short presses of the SURV button allow the MIN, MAX, or AVG values to be viewed. A long press enables the SURV mode and clears the previous values.
The SECURX adapter for the input jacks locks the test leads securely so that they cannot accidentally be pulled out. To release a test lead, push the SECURX locking bar toward the LCD display and hold it while the test leads are pulled out.
For current measurements, the meter must be connected in series with the load. If incorrectly connected (in parallel with the load), the meter presents a very low impedance (almost a short), which may blow the fuse or damage the meter or equipment under test.
To ensure operator safety, never connect leads to a live circuit when taking resistance measurements.
To ensure operator safety, never connect leads to a live circuit when taking resistance measurements.
Discharge all capacitors before taking any measurements. Damage to the instrument and injury to the operator could occur. To ensure operator safety, never connect the leads to a live circuit.
USEFUL CONVERSIONS | |||||
---|---|---|---|---|---|
pF | nF | μF | |||
1,000 | 1.0 | 0.001 | |||
10,000 | 10.0 | 0.01 | |||
100,000 | 100.0 | 0.1 | |||
1,000,000 | 1,000.0 | 1.0 | |||
10,000.0 | 10.0 | ||||
100,000.0 | 100.0 |
pF = picofarads (10-12)
nF = nanofarads (10-9)
µF = microfarads (10-6
Frequency measurement may be made when taking voltage measurements for all multimeters, and for the 5390 and 5370, while taking current measurements as well.
Connect the red test lead to the appropriate current input jack if the Function Switch is set to mADC or 10ADC on the models 5370 and 5390.
is displayed, press the yellow SEL/ON button once. The frequency is then shown on the display.
4. To measure frequency with models 5370 and 5390, follow the procedure for taking voltage or current measurements. While the measurement is displayed, press the Hz button once. The frequency is then shown on the display.
The reference impedance for dBm and resistive power measurements is selectable from 1 Ω to 9999 Ω for models 5370 and 5390.
The reference impedance is fixed at 600 Ω for the 5380, and is adjustable for the 5370 and 5390. Refer to SELECTING IMPEDANCE to set the impedance for models 5370 and 5390.
Refer to SELECTING IMPEDANCE to set the reference impedance before performing a resistive power measurement.
Duty cycle measurements can be taken while voltage measurements are taken for all meters, and while current measurements are taken using the 5390 and 5370.
To measure % positive duty cycle in the model 5360, set the Function Switch to the VAC position, and then press the yellow SEL/ON button twice. To measure % negative duty cycle, press the yellow SEL/ON button a third time. To measure % positive duty cycle in the models 5370 and 5390, set the rotary Function Switch to VAC, VDC, mVDC, mADC, or 10ADC, then press the Hz button twice. To measure % negative duty cycle, press the Hz button a third time.
To perform duty cycle measurements during current measurements (5370 and 5390 only), connect the red test lead to the appropriate current input jack and the black test lead to the COM jack,
or
to perform duty cycle measurements during voltage measurements, connect the red test lead to the VΩ+II → jack and the black test lead to the COM jack.
Pulse width measurements can be taken while taking voltage or current measurements.
or
to perform pulse count measurements during voltage measurements, connect the red test lead to the VΩ+ → jack and the black test lead to the COM jack.
3. Connect the test leads to points of measurement.
4. Read positive or negative pulse width from the display. The A or V annunciators of the LCD indicate the polarity.
Pulse count measurements can be taken while taking voltage or current measurements. This is a totalized pulse count from the time of activation until the time of reading.
to perform pulse count measurements during voltage measurements, connect the red test lead to the VΩ+II → jack and the black test lead to the COM jack.
The multimeter requires no maintenance other than replacing blown fuses or weak or dead batteries. The multimeter monitors the condition of the two fuses and the battery.
The display shows FUSE 1 if fuse F1 (0.63 A) is blown. The display shows FUSE 2 if fuse F2 (10 A) is blown. If both fuses are blown, the display shows FUSES.
Fuse F1 is tested only when the Function Switch is in the µA/mA position.
The display shows BAT when the battery is low. When the BAT annunciator first appears, the battery has approximately 10% (50 hours) of life remaining.
Use the following procedure to open the multimeter case and replace a blown fuse or the battery.
Fig. 14 Opening the Case
Because of the specialized skills and test equipment required for instrument repair and calibration, many customers prefer to rely upon B+K PRECISION for this service. We maintain a network of B+K PRECISION authorized service agencies for this purpose. To use this service, even if the instrument is no longer under warranty, follow the instructions given in the WARRANTY SERVICE INSTRUCTIONS portion of this manual. There is a nominal charge for instruments out of warranty.
DCV accuracy from 21°C to 25°C (5390 and 5380)
Accu | Innut | |||||
---|---|---|---|---|---|---|
Ranges | 5390 | 5380 | 5370 | 5360 | Resolution | Impedance |
500 mV | 0.025%R+2D | 0.025%R+2D | 0.05%R+2D | 0.1%R+2D | 10 µV |
10 MΩ or
1 GΩ * |
5 V | 0.025%R+2D | 0.025%R+2D | 0.05%R+2D | 0.1%R+2D | 100 µV | 11 MΩ |
50 V | 0.025%R+2D | 0.025%R+2D | 0.05%R+2D | 0.1%R+2D | 1 mV | 10 MΩ |
500 V | 0.025%R+2D | 0.025%R+2D | 0.05%R+2D | 0.2%R+2D | 10 mV | 10 MΩ |
1000 V | 0.025%R+2D | 0.025%R+2D | 0.05%R+2D | 0.3%R+2D | 100 mV | 10 M Ω |
* (10 MO only on 5360)
Normal Mode Rejection:
±1100 V DC+AC peak, beeper sounds if exceeded >60 dB, selectable 50 Hz and 60 Hz Common Mode Rejection: >120 dB, selectable 50 Hz and 60 Hz
Accuracy at greater than 10% of range, 50,000 counts at Full Scale:
Ranges | DC to 1 kHz |
1 kHz to
4 kHz |
4 kHz to
10 kHz |
10 kHz to
30 kHz |
30 kHz to
50 kHz |
Resolution |
Input
Impedance |
---|---|---|---|---|---|---|---|
500 mV | 0.3%R+30D | 1%R+30D | 5%R+30D | Not s | pecified | 10 μV |
10 MΩ /100 pF
or 1 GΩ |
5 V | 0.3%R+30D | 1%R+30D | 2%R+30D | 2%R+30D | 3%R+30D | 100 µV | 11 MΩ/100 pF |
50 V | 0.3%R+30D | 1%R+30D | 2%R+30D | 2%R+30D | 3%R+30D | 1 mV | 10 MΩ/100 pF |
500 V | 0.3%R+30D | 1%R+30D | 2%R+30D | 2%R∔30D | 3%R+30D | 10 mV | 10 MΩ/100 pF |
750 V | 0.3%R+30D | Not s | pecified | 100 mV | 10 MΩ /100 pF |
Crest Factor: | 6 |
---|---|
Accuracy Factor: | 0.2% for crest factor of 2 to 3 |
- | 0.5% for crest factor of 3 to 6 |
Overvoltage Protection: | ±1100V DC+AC peak, beeper sounds if exceeded |
Transient Protection: | 6 kV for 10 μs |
Common Mode Rejection: | >80 dB, selectable 50 Hz or 60 Hz |
Accuracy at greater than 10% of range, 5,000 counts at Full Scale:
Ranges |
DC to
1 kHz |
1 kHz to
5 kHz |
5 kHz to
10 kHz |
10 kHz to
20 kHz |
20 kHz to
30 kHz |
Resolution |
Input
Impedance |
---|---|---|---|---|---|---|---|
500 mV | 1%R+3D | 4%R+3D | Not specifie | d | 100 μV | 10 MΩ /100 pF | |
5 V | 1%R+3D | 1%R+3D | 2%R+3D | 2%R+3D | 3%R+3D | 1 mV | 11 M Ω/100 pF |
50 V | 1%R+3D | 1%R+3D | 2%R+3D | 2%R+3D | 3%R+3D | 10 mV | 10 MΩ/100 pF |
500 V | 1%R+3D | 1%R+3D | 2%R+3D | 2%R+3D | 3%R+3D | 100 mV | 10 MΩ/100 pF |
750 V | 1%R+3D | Not s | pecified | 1 V | 10 MΩ/100 pF |
Crest Factor:
racy Factor: |
1.0% for crest factor of 1.5 to 2
4.0% for crest factor of 2 to 3. |
---|---|
voltage Protection: | ±1100 V DC+AC peak, beeper sounds if exceeded |
ient Protection: | 6 kV for 10 μs |
non Mode Rejection: | 80 dB, selectable 50 Hz or 60 Hz |
Accur | acy | Max Burden | ||
---|---|---|---|---|
Ranges | 5390, 5380, 5370 | 5360 | Resolution | Voltage |
500 µA | 0.2%R+5D | Not Available | 10 nA | 700 mV |
5 mA | 0.2%R+2D | 0.2%R+2D | 100 nA | 700 mV |
50 mA | 0.05%R+2D | 0.2%R+2D | 1 μA | 700 mV |
500 mA | 0.2%R+2D | 0.2%R+2D | 10 µA | 1.5 V |
10 A | 0.5%R+2D | 0.5%R+2D | 1 mA | 500 mV |
erload Protection: 630 mA, 600 V fuse for mA ranges, interrupts 1500 A 10 A, 600 V fuse for 10 A range, interrupts 50,000 A Display indicates blown fuse F1, F2 or FUSES
Models 5390, 5380, and 5370.
Accuracy at greater than 10% of range, 50,000 counts at Full Scale
Ranges | DC to 1 kHz |
1 kHz to
10 kHz |
10 kHz to
20 kHz |
20 kHz to
30 kHz |
Resolution |
Max Burden
Voltage |
---|---|---|---|---|---|---|
500 µA | 0.75%R+50D | 0.75%R+50D | 1.0%R+50D | 3.0%R+50D | 10 nA | 700 mV |
5 mA | 0.6%R+50D | 0.6%R+50D | 1.0%R+50D | 2.0%R+50D | 100 nA | 700 mV |
50 mA | 0.6%R+50D | 0.6%R+50D | 1.0%R+50D | 2.0%R+50D | 1 µA | 700 mV |
500 mA | 0.7%R+50D | 0.7%R+50D | Not s | pecified | 10 µA | 1.5 V |
10 A | 1% of FS | Not specified | 1 mA | 500 mV |
Crest Factor:
Accuracy Factor: 0.2% for crest fact
0.5% for crest factor of 3 to
Overload Protection: 630 mA, 600 V fuse for mA ranges, interrupts 1500 A 10 A, 600 V fuse for 10 A range, interrupts 50,000 A Display indicates blown fuse F1, F2, or FUSES
Model 5360. Accuracy at greater than 10% of range, 5,000 counts at Full Scale
Ranges | DC to 5 kHz | Resolution | Max Burden Voltage |
---|---|---|---|
5 mA | 1.0%R+3D | 1 μA | 700 mV |
50 mA | 1.0%R+3D | 10 µA | 700 mV |
500 mA | 1.0%R+3D | 100 µA | 1.5 V |
10 A | 1% of FS | 10 mA | 500 mV |
rest Factor:
Factor: 1.0% for crest factor of 1.5 to 2 4.0% for crest factor of 2 to 3
Overload Protection: 630 mA, 600 V fuse for mA ranges, interrupts 1500 A 10 A, 600 V fuse for 10 A range, interrupts 50,000 A Display indicates blown fuse F1, F2, or FUSES
Ranges |
Accuracy
5390, 5380, 5370 |
Accuracy
5360 |
Resolution | Test Current |
---|---|---|---|---|
500 Ω | 0.07%R+5D | 0.1%R+5D | 10 mΩ | 1 mA |
5kΩ | 0.07%R+2D | 0.1%R+2D | 100 mΩ | 100 µA |
50 k Ω | 0.07%R+2D | 0.1%R+2D | 1Ω | 10 µA |
500 k Ω | 0.07%R+2D | 0.1%R+2D | 10 Ω | 1 µA |
5 MΩ | 0.3%R+2D | 0.3%R+2D | 100 Ω | 100 nA |
50 MΩ | 1.0%R+2D | 1.0%R+2D | 1 kΩ | 10 nA |
Max Open Circuit Voltage: Overload Protection:
600V DC+AC peak
Threshold:10 Ω to 20 ΩResponse Time:1 msOverload Protection:600 V DC+AC peak
Measures forward voltage drop of diode 0 to 1.999V Test current: 1 mA ±20% Overload Protection: 600V DC+AC peak
Accuracy, 5,000 counts at Full Scale:
Accuracy | Max | ||||
---|---|---|---|---|---|
Ranges | 5390,5380,5370 5360 | Resolution |
Measurement
Current |
Measurement
Time |
|
50 nF | 1.0%R+2D | 1.0%R+2D | 10 pF | 100 nA | 0.5 sec. |
500 nF | 1.0%R+2D | 1.0%R+2D | 100 pF | 1 μA | 0.5 sec. |
5 µF | 1.0%R+2D | 1.0%R+2D | 1 nF | 10 µA | 0.5 sec. |
50 µF | 1.0%R+2D | 1.0%R+2D | 10 nF | 100 µA | 0.5 sec. |
500 µF | 1.0%R+2D | 1.0%R+2D | 100 nF | 1 mA | 1.5 sec. |
5000 µF | 1.0%R+2D | 2.0%R+2D | 1 µF | 1 mA | 3 sec./ mF |
50 mF | 1.0%R+2D | 2.0%R+2D | 10 µF | 1 mA | 3 sec./ mF |
Overload Protec
tion:
600V DC+AC peak
equency | Range: | |
---|---|---|
curacy: | ||
A. |
0.6 Hz to 500 kHz ±(0.03% rdg + 1 dgt)
nsitivity:
Range |
500 mV
(5390, 5370) |
5 V to 500 V
500 µA to 500 mA |
750 V |
10 A
(5390, 5370) |
---|---|---|---|---|
0.6 Hz to 5 kHz | 2% of FS | 2% of FS | 100 V | 2 A |
5 kHz to 50 kHz | 5% of FS | 5% of FS | 250 V | |
50 kHz to 500 kHz | Not Specified | 10% of FS | Not | Specified |
Selection:Peak + or Peak - (surge or sag)Acquisition Time:<1 ms</td>Applicable Modes:VDC, mVDC, mADC, 10ADCAccuracy:For pulse >1ms, (1% reading + 50 digits)
Measures value relative to a reference from 1 to 9999 Ω. Factory preset for reference is 600 Ω. Refer to dBm measurement procedure to set impedance. Accuracy in dB: 0.09 x VAC accuracy in percent
Measurement range: 10 mVAC to 750 VAC
Converts VAC to dBm across 600 ohms, 0 dBm = 0.7746 mVAC Accuracy in dB: 0.09 x VAC accuracy in percent Resolution: ±0.01 dB Measurement range: 10 mVAC to 750 VAC
Modes: | Selectable %+, %- |
---|---|
Resolution: | 0.01% |
Minimum pulse width: | 2 μs |
Maximum pulse period: | 0.8 s |
Minimum pulse period: | 100 µs |
Accuracy: | Same as FREQUENCY mode |
Range: | 0 to 99999 (totalize) |
---|---|
Minimum pulse duration: | 2 μs |
Reset: | Relative key |
Sensitivity: | 3% of range (excludes 750 V, 10 A ranges) |
Modes: | Selectable + or |
---|---|
Resolution: | 10 µs |
Minimum pulse width: | 20 µs |
Maximum pulse period: | 12.5 s |
Accuracy: | ±(0.01% + 10 μs) |
Sensitivity: | 3% of range (excludes 750 V, 10 A ranges) |
Display indicates Power Disturbance for transient voltage >80V, or high frequency (1 kHz to 100 kHz) Harmonic Disturbance >25 V
Using Model TP-5300 Pl | atinum Temperature Probe | |
---|---|---|
Temperature Range: | -200°C to +800°C | |
Resolution: | 0.1°C | |
Accuracy: | -125 C to +75 C: | 0.5% |
Display: | °C or °F, selectable | 1.00 |
RS-232 capability built-in Requires AK-5300 or AK-5350 hardware/software kit for interconnection to computer or printer.
GENERAL SPECIFI | CATIONS | Auto Power Off Override: | Does not shut off with live circuit, RS-232 active or |
---|---|---|---|
Display: |
50,000 count LCD, 14-mm digit height, 34-segment
analog bargraph (with backlight for models 5390, 5380, and 5370). Annunciators for function and unit of |
Dimensions (HxWxD): |
Peak Hold
1-5/8" x 3-1/4" x 7-1/2" (40 mm x 82 mm x 189 mm) |
measurement. | Weight: | 14 oz. | |
Measurement Rate:
Environmental: |
2 per second, 20 per second for bargraph. | Safety: |
IEC1010-1 class 2, 600V cat III, 1000V cat II
Live circuit indication above 24 V |
Operation: 0° to 50°C
0 to 80% RH to 40°C |
6 kV transient protection
600 V high energy fuses on all current ranges |
||
Limited Operation: |
0 to 70% RH to 50°C
−10°C to +60°C, <80% RH |
Ruggedness:
Shock & Vibration: |
Sealed, watertight case, 1P677
MIL-T-28800 |
Storage: | -40°C to + 70°C without battery | Case Material: | Flame retardant, self extinguishing |
Power Requirement: | 9V battery | EMC: | Emission EN55011, class B, susceptibility |
Battery Life: | 500 hours | IEC801-4, level 3, radiation IEC 801-3 (10V/M) level | |
Low Battery Indication: | BAT displayed | Calibration: | One year, recalibration by serial interface without |
Auto Power Off: | 30 minutes | opening case |
Battery, Test Leads, Instruction Manual
The following accessories are available for each of the multimeters:
The models 5390, 5380, 5370, and 5360 have built-in RS-232 capability. However, special hardware is required to complete the interconnection to a computer. The models AK-5300 and AK-5350 include the special hardware required.
This kit contains the hardware and software needed to recalibrate and to document the calibration of your multimeter.
This kit contains the hardware and software needed to download data from your multimeter to your PC, and to control the multimeter from the PC.
This accessory adds calibration capability to the Model AK-5350 Data Acquisition Hardware and Software Kit.
This accessory adds data acquisition capability to the Model AK-5300 Calibration Hardware and Software Kit.
Provides a convenient and safe means for storing and transporting the multimeters.
Platinum temperature probe designed for use with Model 5370 Multimeter.
Replacement test leads must have a shoulder for the SECURX test lead retainer to work properly. B+K Precision Models TL-100, TL-110 and TL-120 test leads have the necessary shoulders to actuate this important safety feature.
B+K Precision offers a wide variety of test lead tips to make appropriate connections in most applications. Such tips include alligator clips, spring hooks, spade lugs, surface mount device tips, etc.