The following safety precautions apply to both operating and maintenance personnel and must be followed
during all phases of operation, service, and repair of this instrument.
Before applying power to this instrument:
• Read and understand the safety and operational information in this manual.
• Apply all the listed safety precautions.
• Verify that the voltage selector at the line power cord input is set to the correct line voltage. Operating
the instrument at an incorrect line voltage will void the warranty.
• Make all connections to the instrument before applying power.
• Do not operate the instrument in ways not specied by this manual or by B&K Precision.
Failure to comply with these precautions or with warnings elsewhere in this manual violates the safety standards
of design, manufacture, and intended use of the instrument. B&K Precision assumes no liability for a customer’s
failure to comply with these requirements.
Category rating
The IEC 61010 standard denes safety category ratings that specify the amount of electrical energy available
and the voltage impulses that may occur on electrical conductors associated with these category ratings. The
category rating is a Roman numeral of I, II, III, or IV. This rating is also accompanied by a maximum voltage of
the circuit to be tested, which denes the voltage impulses expected and required insulation clearances. These
categories are:
Category I (CAT I): Measurement instruments whose measurement inputs are not intended to be connected
to the mains supply. The voltages in the environment are typically derived from a limited-energy
transformer or a battery.
Category II (CAT II): Measurement instruments whose measurement inputs are meant to be connected to the
mains supply at a standard wall outlet or similar sources. Example measurement environments are
portable tools and household appliances.
Category III (CAT III): Measurement instruments whose measurement inputs are meant to be connected to
the mains installation of a building. Examples are measurements inside a building’s circuit breaker panel or
the wiring of permanently-installed motors.
Category IV (CAT IV): Measurement instruments whose measurement inputs are meant to be connected to
the primary power entering a building or other outdoor wiring.
i
Do not use this instrument in an electrical environment with a higher category rating than what is specied in
this manual for this instrument.
You must ensure that each accessory you use with this instrument has a category rating equal to or higher than
the instrument’s category rating to maintain the instrument’s category rating. Failure to do so will lower the
category rating of the measuring system.
Electrical Power
This instrument is intended to be powered from a CATEGORY II mains power environment. The mains power
should be 115 V RMS or 230 V RMS. Use only the power cord supplied with the instrument and ensure it is
appropriate for your country of use.
Ground the Instrument
ii
To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical safety ground.
This instrument is grounded through the ground conductor of the supplied, three-conductor AC line power
cable. The power cable must be plugged into an approved three-conductor electrical outlet. The power jack and
mating plug of the power cable meet IEC safety standards.
Do not alter or defeat the ground connection. Without the safety ground connection, all accessible conductive
parts (including control knobs) may provide an electric shock. Failure to use a properly-grounded approved
outlet and the recommended three-conductor AC line power cable may result in injury or death.
Unless otherwise stated, a ground connection on the instrument’s front or rear panel is for a reference of
potential only and is not to be used as a safety ground. Do not operate in an explosive or ammable atmosphere.
Do not operate the instrument in the presence of ammable gases or vapors, fumes, or nely-divided particulates.
The instrument is designed to be used in oce-type indoor environments. Do not operate the instrument
• In the presence of noxious, corrosive, or ammable fumes, gases, vapors, chemicals, or nely-divided
particulates.
• In relative humidity conditions outside the instrument’s specications.
• In environments where there is a danger of any liquid being spilled on the instrument or where any liquid
can condense on the instrument.
• In air temperatures exceeding the specied operating temperatures.
• In atmospheric pressures outside the specied altitude limits or where the surrounding gas is not air.
• In environments with restricted cooling air ow, even if the air temperatures are within specications.
• In direct sunlight.
This instrument is intended to be used in an indoor pollution degree 2 environment. The operating temperature
range is 0◦C to 40◦C and 20% to 80% relative humidity, with no condensation allowed. Measurements made by
this instrument may be outside specications if the instrument is used in non-oce-type environments. Such
environments may include rapid temperature or humidity changes, sunlight, vibration and/or mechanical shocks,
acoustic noise, electrical noise, strong electric elds, or strong magnetic elds.
Do not operate instrument if damaged
If the instrument is damaged, appears to be damaged, or if any liquid, chemical, or other material gets on or
inside the instrument, remove the instrument’s power cord, remove the instrument from service, label it as not
to be operated, and return the instrument to B&K Precision for repair. Notify B&K Precision of the nature of
any contamination of the instrument.
iii
Clean the instrument only as instructed
Do not clean the instrument, its switches, or its terminals with contact cleaners, abrasives, lubricants, solvents,
acids/bases, or other such chemicals. Clean the instrument only with a clean dry lint-free cloth or as instructed
in this manual. Not for critical applications
This instrument is not authorized for use in contact with the human body or for use as a component in a
life-support device or system.
Do not touch live circuits
Instrument covers must not be removed by operating personnel. Component replacement and internal
adjustments must be made by qualied service-trained maintenance personnel who are aware of the hazards
involved when the instrument’s covers and shields are removed. Under certain conditions, even with the power
cord removed, dangerous voltages may exist when the covers are removed. To avoid injuries, always disconnect
the power cord from the instrument, disconnect all other connections (for example, test leads, computer
interface cables, etc.), discharge all circuits, and verify there are no hazardous voltages present on any
conductors by measurements with a properly-operating voltage-sensing device before touching any internal parts.
Verify the voltage-sensing device is working properly before and after making the measurements by testing with
known-operating voltage sources and test for both DC and AC voltages. Do not attempt any service or
adjustment unless another person capable of rendering rst aid and resuscitation is present.
Do not insert any object into an instrument’s ventilation openings or other openings.
Hazardous voltages may be present in unexpected locations in circuitry being tested when a fault condition in
the circuit exists.
Fuse replacement must be done by qualied service-trained maintenance personnel who are aware of the
instrument’s fuse requirements and safe replacement procedures. Disconnect the instrument from the power line
before replacing fuses. Replace fuses only with new fuses of the fuse types, voltage ratings, and current ratings
specied in this manual or on the back of the instrument. Failure to do so may damage the instrument, lead to
a safety hazard, or cause a re. Failure to use the specied fuses will void the warranty.
Servicing
iv
Do not substitute parts that are not approved by B&K Precision or modify this instrument. Return the
instrument to B&K Precision for service and repair to ensure that safety and performance features are
maintained.
For continued safe use of the instrument
• Do not place heavy objects on the instrument.
• Do not obstruct cooling air ow to the instrument.
• Do not place a hot soldering iron on the instrument.
• Do not pull the instrument with the power cord, connected probe, or connected test lead.
• Do not move the instrument when a probe is connected to a circuit being tested.
Safety Symbols
v
DANGER indicates a hazardous situation which, if not avoided, will
result in death or serious injury.
WARNING indicates a hazardous situation which, if not avoided,
could result in death or serious injury
CAUTION indicates a hazardous situation which, if not avoided, will
result in minor or moderate injury
A Caution. Refer to the text near the symbol.
Electric Shock hazard
Alternating current (AC)
Chassis ground
Earth ground
On (Power). This is the In position of the power switch when instrument is ON.
O (Power). This is the Out position of the power switch when
instrument is OFF.
NOTICE is used to address practices not related to physical injury.
Note: The contents of this manual and included specications are subject to change go to for the latest version.
The BK5335B power meter measures AC and DC inputs up to 600Vrms and 20Arms from DC to 100kHz. It
measures voltage, current, power, frequency, power factor, phase and harmonic parameters up to the 50th order.
Remote control is available over USB, GPIB, RS232 and LAN communication interfaces. Voltage and current
measurement precision is nominally 0.1%. See Figure 1.1 for a view and details of the front panel.
1.1Features
• 4.3-inch color LCD
• Congurable display of measurements in 3 formats
• Input range: 600Vrms/20Arms from DC to 100kHz
• Voltage, current, power, harmonics and other parameters are measured at the same time simultaneously
• Up to 0.1% voltage and current measurement accuracy
• Harmonic measurement up to the 50th-order harmonic
• Screen captures are saved to USB via the front panel connector
• Integration measurement of power produced or consumed
• Frequency measurement
• Remote control via USB, GPIB, RS232 and LAN communication interfaces
1
CHAPTER 1. GENERAL INFORMATION2
1.2Front Panel
Figure 1.1: Front Panel
Power ButtonRotary Knob
4.3” LCDArrow Keys
USBSoft-keys
Main FunctionsSoft-keys
Menu, Enter, ESCPrint (Screen Capture)
KeyName and function
Waveform Display key: press to view waveforms. See Chapter 6
Harmonic Measurement key: For viewing harmonic measurements. See Chapter 7
Basic Measurement key: The normal measurement screen. See Chapter 4
Integral Measurement key: Measurements over time. See Chapter 8
Menu key, Press to view and congure settings. See Chapter 2
Enter key: Press to commit a setting or value.
Image Save key: press to save a hardcopy of the screen to the USB host port on the front panel.
CHAPTER 1. GENERAL INFORMATION3
1.3Rear Panel Summary
Figure 1.2: Rear View
GPIB ConnectorAC Line
LAN ConnectorExternal Current Sensor
USB ConnectorVoltage Input
RS-232 ConnectorCurrent Shunt Input
Synchronization BNC
1.4Power Line Connection
Input power requirements:
AC Voltage 110V ±10% or 220V ±10%
Frequency 47Hz – 63Hz
1.4.1Fuse Requirements
An AC input fuse is necessary when powering the instrument. Refer to Table 1.1 for the fuse requirements.
2. Locate the fuse box in the rear panel, beneath the AC power socket. Figure 1.3
3. With a small at blade screwdriver, insert the blade into the fuse box slit to pull and slide out the fuse box.
4. Pull out the fuse inside to check and/or replace with the appropriate fuse for the line voltage used.
5. Insert the fuse in the same location.
1.4.2Power-up Sequence
Connect AC power cord to the AC receptacle in the rear panel and press the power switch to the (ON) position
to turn ON the instrument. It will display the BIOS version then run through a self-test procedure, Figure 2.9.
Chapter 2
Menu
Conguration of system-wide settings is done from the “Menu”. Pressing thebutton enters the
conguration/system menu, Figure 2.1.
From the main menu, the soft-keys at the bottom of the screen provide access to the conguration screens. See
Table 2.1
2.1MENU > SYSTEM menus
From the system menu, system-wide functions are set and viewed. Figure 2.2, details such as the LAN MAC
address, date, screen brightness, system initialization (settings reset), and self test are accessed here. See
Table 2.2
2.1.1System Information Menu
Information about the device is listed in this menu. The serial number listed is that used in production at the
factory and does not match the label. When contacting B&K, use the serial number on the label.
Figure 2.1: Conguration/System Menu
5
CHAPTER 2. MENU6
FunctionDescription
SYSTEMView and change overall system information and parameters
SETUPDevice measurement and measurement conguration settings
CAL ZEROZero the meter
RATIO SETChange the overall scaling factor for measurements
Table 2.1: Main “Menu” functions
FunctionDescription
SYSTEM INFOSystem details, Model, Serial, etc…
COMM CONFIGSetup the communication interfaces
SYSTEM CONFIGSet date, time, beep and brightness
SELF TESTPerform a self diagnostic
INITIALReset device settings
Table 2.2: System Menu sub-menus
Figure 2.2: System Info Screen
2.1.2Communication Conguration Menu
Available remote interfaces are RS-232, USBTMC, GPIB and LAN. For details about commands and use of the
remote interfaces, see the programming manual. The manual is found on the product page of the 5335B at
www.bkprecision.com
2.1.2.1 RS-232
This interface support 6 common baudrates: 4800, 9600, 19200, 38400, 57600, 115200. Use the arrow keys to
navigate the menu screen and the soft-keys to select the desired baudrate. The number of data bits, parity and
stop bits are preset and unchangeable. (8 bits, no parity, and 1 stop bit, so called N-8-1). See Figure 2.3.
CHAPTER 2. MENU7
Figure 2.3: COMM Conguration - RS-232
Figure 2.4: COMM Conguration - USB
2.1.2.2 USB
This unit operates as a USBTMC (USB Test and Measurement Class) device. There are no conguration
options for this interface. Figure 2.4
2.1.2.3 GPIB
The GPIB interface may be congured to addresses 1 through 30. Use the arrow keys to select the numeric eld
beside “GPIB Address” and use the rotary knob to change the value. The user may select each digit by the
arrow keys as well. Figure 2.5
CHAPTER 2. MENU8
Figure 2.5: COMM Conguration - GPIB
Figure 2.6: COMM Conguration - LAN (DHCP)
2.1.2.4 LAN
The LAN (Ethernet) interface may be congured as either DHCP or static. Use the arrow keys to select the IP
Mode eld. When DHCP is selected, the current IP, subnet, and gateway are shown on screen (Figure 2.6).
When “MANU” is selected (static mode) (Figure 2.7), the screen shows elds for setting the IP, subnet, and
gateway. Using the arrow keys and the rotary knob allows the user to move between elds and digits, and use
the rotary knob to change the value.
2.1.3System Conguration Menu
From this menu, setting the date, time, brightness, and the state of the beep function is set. Use the arrow keys
to navigate the elds, and the rotary knob to change the value. For the beep function, the soft keys are used to
change the value. When settings are changed, press thekey to commit the changes. See Figure 2.8.
CHAPTER 2. MENU9
Figure 2.7: COMM Conguration - LAN (Manual (static))
Figure 2.8: System Cong Menu
2.1.4Self Test
The unit has a built in self test. This function tests the LEDs, the screen and some of the internal electronics,
Figure 2.9. When the test is running, it cycles through colors on the LCD, and lights up the individual LEDs
that illuminate the front panel buttons. When complete, the results are displayed on screen, DONE for the LCD
and LEDs, and OK for the internal electronic tests.
From this screen, a test of each button is also available. A representation of all the buttons appears on screen
after selecting the “KEY TEST” key, pressing the individual keys will highlight the text on screen indicating the
button is working. Press theto exit the test.
2.1.5System Initialize Menu
System initialization will bring the manifold functions to their initial states. By navigating the functions listed
on-screen and either selecting the check mark or “x” via the soft-keys, sets whether the function is initialized.
CHAPTER 2. MENU10
Figure 2.9: System Self Test
Figure 2.10: System Initialize Menu
2.2MENU > SETUP menus
From the “MENU” function, the next page is the “SETUP” soft-key (Figure 2.11). This screen essentially gives
a summary of the settings controlling measurements.
2.2.1Averaging Setup
When measuring low frequency signals where measurements begin to become unstable, averaging may be useful.
To enable averaging enter the “Average Setup” menu (Menu>Setup>Average Set).
CHAPTER 2. MENU11
FieldDescription
AverageListing of the average mode settings.
Sync SourceThe measurement synchronization signal source.
Line FilterThe state of the line lter.
Update rateHow often to measure
Freq FilterThe state of the frequency lter.
Crest FactorThe Crest Factor setting state
ExSensor1External Sensor 1 state (on/o)
ExSensor2External Sensor 2 state (on/o)
U_RangeThe voltage range setting.
Rate(V/A)External Sensor 1’s scale value
Rate(mV/A)External Sensor 2’s scale value
I_RangeThe current range setting.
Figure 2.11: Setup Information Screen
CHAPTER 2. MENU12
State The state of the average function, On or O.
Type Exp or Line. Exp enables index averaging, and Line enables linear averaging.
Tcontrol Sets averaging to being either a rolling of repeating average.
Count EXP averaging attenuation constant, or LINE average count.
2.2.1.1 EXP - Index averaging
Dn= D
n−1
Mn− D
+
n−1
k
(2.1)
Dn The value displayed after the nth index averaging (D1, the value displayed after the rst averaging, is equal
to M1)
Dn-1 the value displayed after the (n-1)th index averaging
Mn the nth measured data.
K attenuation constant (1-64)
2.2.1.2 LINE - Linear averaging
Dn=
M
n−(M−1)
+ . . . + M
m
n−2
+ M
n−1
+ M
n
(2.2)
Dn The value displayed after linear averaging of m values from the n − (m − 1)thto nthvalue
M Measured value
m The number of values in the average set (1-64)
When m is divisible by n, the calculated value is the moving average; otherwise calculated value is the repeated
average.
NOTE: When index averaging is set, averaging is implemented under the harmonic measurement function.
When linear averaging is set, averaging can only be implemented in the conventional measurement function and
this mode is not applicable to the harmonic measurement function.
2.2.2External Sensor Setup
Figure 2.12 accesses via Menu>Setup>Ext Sen Set contains settings used with external sensors. The state
(on/o) and gain (V/A (mA)) are set here.
2.2.3Other settings: Filter and Crest Factor
Congure the measurement lter, synchronization signal source and crest factor from the “Other Set” menu.
Access this menu via Menu>SET UP>OTHER SET. The “Other Set” menu is shown in Figure 2.13.
Loading...
+ 44 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.