The following safety precautions apply to both operating and maintenance personnel and must be followed
during all phases of operation, service, and repair of this instrument.
Before applying power to this instrument:
• Read and understand the safety and operational information in this manual.
• Apply all the listed safety precautions.
• Verify that the voltage selector at the line power cord input is set to the correct line voltage. Operating
the instrument at an incorrect line voltage will void the warranty.
• Make all connections to the instrument before applying power.
• Do not operate the instrument in ways not specied by this manual or by B&K Precision.
Failure to comply with these precautions or with warnings elsewhere in this manual violates the safety standards
of design, manufacture, and intended use of the instrument. B&K Precision assumes no liability for a customer’s
failure to comply with these requirements.
Category rating
The IEC 61010 standard denes safety category ratings that specify the amount of electrical energy available
and the voltage impulses that may occur on electrical conductors associated with these category ratings. The
category rating is a Roman numeral of I, II, III, or IV. This rating is also accompanied by a maximum voltage of
the circuit to be tested, which denes the voltage impulses expected and required insulation clearances. These
categories are:
Category I (CAT I): Measurement instruments whose measurement inputs are not intended to be connected
to the mains supply. The voltages in the environment are typically derived from a limited-energy
transformer or a battery.
Category II (CAT II): Measurement instruments whose measurement inputs are meant to be connected to the
mains supply at a standard wall outlet or similar sources. Example measurement environments are
portable tools and household appliances.
Category III (CAT III): Measurement instruments whose measurement inputs are meant to be connected to
the mains installation of a building. Examples are measurements inside a building’s circuit breaker panel or
the wiring of permanently-installed motors.
Category IV (CAT IV): Measurement instruments whose measurement inputs are meant to be connected to
the primary power entering a building or other outdoor wiring.
i
Do not use this instrument in an electrical environment with a higher category rating than what is specied in
this manual for this instrument.
You must ensure that each accessory you use with this instrument has a category rating equal to or higher than
the instrument’s category rating to maintain the instrument’s category rating. Failure to do so will lower the
category rating of the measuring system.
Electrical Power
This instrument is intended to be powered from a CATEGORY II mains power environment. The mains power
should be 115 V RMS or 230 V RMS. Use only the power cord supplied with the instrument and ensure it is
appropriate for your country of use.
Ground the Instrument
ii
To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical safety ground.
This instrument is grounded through the ground conductor of the supplied, three-conductor AC line power
cable. The power cable must be plugged into an approved three-conductor electrical outlet. The power jack and
mating plug of the power cable meet IEC safety standards.
Do not alter or defeat the ground connection. Without the safety ground connection, all accessible conductive
parts (including control knobs) may provide an electric shock. Failure to use a properly-grounded approved
outlet and the recommended three-conductor AC line power cable may result in injury or death.
Unless otherwise stated, a ground connection on the instrument’s front or rear panel is for a reference of
potential only and is not to be used as a safety ground. Do not operate in an explosive or ammable atmosphere.
Do not operate the instrument in the presence of ammable gases or vapors, fumes, or nely-divided particulates.
The instrument is designed to be used in oce-type indoor environments. Do not operate the instrument
• In the presence of noxious, corrosive, or ammable fumes, gases, vapors, chemicals, or nely-divided
particulates.
• In relative humidity conditions outside the instrument’s specications.
• In environments where there is a danger of any liquid being spilled on the instrument or where any liquid
can condense on the instrument.
• In air temperatures exceeding the specied operating temperatures.
• In atmospheric pressures outside the specied altitude limits or where the surrounding gas is not air.
• In environments with restricted cooling air ow, even if the air temperatures are within specications.
• In direct sunlight.
This instrument is intended to be used in an indoor pollution degree 2 environment. The operating temperature
range is 0◦C to 40◦C and 20% to 80% relative humidity, with no condensation allowed. Measurements made by
this instrument may be outside specications if the instrument is used in non-oce-type environments. Such
environments may include rapid temperature or humidity changes, sunlight, vibration and/or mechanical shocks,
acoustic noise, electrical noise, strong electric elds, or strong magnetic elds.
Do not operate instrument if damaged
If the instrument is damaged, appears to be damaged, or if any liquid, chemical, or other material gets on or
inside the instrument, remove the instrument’s power cord, remove the instrument from service, label it as not
to be operated, and return the instrument to B&K Precision for repair. Notify B&K Precision of the nature of
any contamination of the instrument.
iii
Clean the instrument only as instructed
Do not clean the instrument, its switches, or its terminals with contact cleaners, abrasives, lubricants, solvents,
acids/bases, or other such chemicals. Clean the instrument only with a clean dry lint-free cloth or as instructed
in this manual. Not for critical applications
This instrument is not authorized for use in contact with the human body or for use as a component in a
life-support device or system.
Do not touch live circuits
Instrument covers must not be removed by operating personnel. Component replacement and internal
adjustments must be made by qualied service-trained maintenance personnel who are aware of the hazards
involved when the instrument’s covers and shields are removed. Under certain conditions, even with the power
cord removed, dangerous voltages may exist when the covers are removed. To avoid injuries, always disconnect
the power cord from the instrument, disconnect all other connections (for example, test leads, computer
interface cables, etc.), discharge all circuits, and verify there are no hazardous voltages present on any
conductors by measurements with a properly-operating voltage-sensing device before touching any internal parts.
Verify the voltage-sensing device is working properly before and after making the measurements by testing with
known-operating voltage sources and test for both DC and AC voltages. Do not attempt any service or
adjustment unless another person capable of rendering rst aid and resuscitation is present.
Do not insert any object into an instrument’s ventilation openings or other openings.
Hazardous voltages may be present in unexpected locations in circuitry being tested when a fault condition in
the circuit exists.
Fuse replacement must be done by qualied service-trained maintenance personnel who are aware of the
instrument’s fuse requirements and safe replacement procedures. Disconnect the instrument from the power line
before replacing fuses. Replace fuses only with new fuses of the fuse types, voltage ratings, and current ratings
specied in this manual or on the back of the instrument. Failure to do so may damage the instrument, lead to
a safety hazard, or cause a re. Failure to use the specied fuses will void the warranty.
Servicing
iv
Do not substitute parts that are not approved by B&K Precision or modify this instrument. Return the
instrument to B&K Precision for service and repair to ensure that safety and performance features are
maintained.
For continued safe use of the instrument
• Do not place heavy objects on the instrument.
• Do not obstruct cooling air ow to the instrument.
• Do not place a hot soldering iron on the instrument.
• Do not pull the instrument with the power cord, connected probe, or connected test lead.
• Do not move the instrument when a probe is connected to a circuit being tested.
Safety Symbols
v
DANGER indicates a hazardous situation which, if not avoided, will
result in death or serious injury.
WARNING indicates a hazardous situation which, if not avoided,
could result in death or serious injury
CAUTION indicates a hazardous situation which, if not avoided, will
result in minor or moderate injury
A Caution. Refer to the text near the symbol.
Electric Shock hazard
Alternating current (AC)
Chassis ground
Earth ground
On (Power). This is the In position of the power switch when instrument is ON.
O (Power). This is the Out position of the power switch when
instrument is OFF.
NOTICE is used to address practices not related to physical injury.
Note: The contents of this manual and included specications are subject to change go to for the latest version.
The BK5335B power meter measures AC and DC inputs up to 600Vrms and 20Arms from DC to 100kHz. It
measures voltage, current, power, frequency, power factor, phase and harmonic parameters up to the 50th order.
Remote control is available over USB, GPIB, RS232 and LAN communication interfaces. Voltage and current
measurement precision is nominally 0.1%. See Figure 1.1 for a view and details of the front panel.
1.1Features
• 4.3-inch color LCD
• Congurable display of measurements in 3 formats
• Input range: 600Vrms/20Arms from DC to 100kHz
• Voltage, current, power, harmonics and other parameters are measured at the same time simultaneously
• Up to 0.1% voltage and current measurement accuracy
• Harmonic measurement up to the 50th-order harmonic
• Screen captures are saved to USB via the front panel connector
• Integration measurement of power produced or consumed
• Frequency measurement
• Remote control via USB, GPIB, RS232 and LAN communication interfaces
1
CHAPTER 1. GENERAL INFORMATION2
1.2Front Panel
Figure 1.1: Front Panel
Power ButtonRotary Knob
4.3” LCDArrow Keys
USBSoft-keys
Main FunctionsSoft-keys
Menu, Enter, ESCPrint (Screen Capture)
KeyName and function
Waveform Display key: press to view waveforms. See Chapter 6
Harmonic Measurement key: For viewing harmonic measurements. See Chapter 7
Basic Measurement key: The normal measurement screen. See Chapter 4
Integral Measurement key: Measurements over time. See Chapter 8
Menu key, Press to view and congure settings. See Chapter 2
Enter key: Press to commit a setting or value.
Image Save key: press to save a hardcopy of the screen to the USB host port on the front panel.
CHAPTER 1. GENERAL INFORMATION3
1.3Rear Panel Summary
Figure 1.2: Rear View
GPIB ConnectorAC Line
LAN ConnectorExternal Current Sensor
USB ConnectorVoltage Input
RS-232 ConnectorCurrent Shunt Input
Synchronization BNC
1.4Power Line Connection
Input power requirements:
AC Voltage 110V ±10% or 220V ±10%
Frequency 47Hz – 63Hz
1.4.1Fuse Requirements
An AC input fuse is necessary when powering the instrument. Refer to Table 1.1 for the fuse requirements.
2. Locate the fuse box in the rear panel, beneath the AC power socket. Figure 1.3
3. With a small at blade screwdriver, insert the blade into the fuse box slit to pull and slide out the fuse box.
4. Pull out the fuse inside to check and/or replace with the appropriate fuse for the line voltage used.
5. Insert the fuse in the same location.
1.4.2Power-up Sequence
Connect AC power cord to the AC receptacle in the rear panel and press the power switch to the (ON) position
to turn ON the instrument. It will display the BIOS version then run through a self-test procedure, Figure 2.9.
Chapter 2
Menu
Conguration of system-wide settings is done from the “Menu”. Pressing thebutton enters the
conguration/system menu, Figure 2.1.
From the main menu, the soft-keys at the bottom of the screen provide access to the conguration screens. See
Table 2.1
2.1MENU > SYSTEM menus
From the system menu, system-wide functions are set and viewed. Figure 2.2, details such as the LAN MAC
address, date, screen brightness, system initialization (settings reset), and self test are accessed here. See
Table 2.2
2.1.1System Information Menu
Information about the device is listed in this menu. The serial number listed is that used in production at the
factory and does not match the label. When contacting B&K, use the serial number on the label.
Figure 2.1: Conguration/System Menu
5
CHAPTER 2. MENU6
FunctionDescription
SYSTEMView and change overall system information and parameters
SETUPDevice measurement and measurement conguration settings
CAL ZEROZero the meter
RATIO SETChange the overall scaling factor for measurements
Table 2.1: Main “Menu” functions
FunctionDescription
SYSTEM INFOSystem details, Model, Serial, etc…
COMM CONFIGSetup the communication interfaces
SYSTEM CONFIGSet date, time, beep and brightness
SELF TESTPerform a self diagnostic
INITIALReset device settings
Table 2.2: System Menu sub-menus
Figure 2.2: System Info Screen
2.1.2Communication Conguration Menu
Available remote interfaces are RS-232, USBTMC, GPIB and LAN. For details about commands and use of the
remote interfaces, see the programming manual. The manual is found on the product page of the 5335B at
www.bkprecision.com
2.1.2.1 RS-232
This interface support 6 common baudrates: 4800, 9600, 19200, 38400, 57600, 115200. Use the arrow keys to
navigate the menu screen and the soft-keys to select the desired baudrate. The number of data bits, parity and
stop bits are preset and unchangeable. (8 bits, no parity, and 1 stop bit, so called N-8-1). See Figure 2.3.
CHAPTER 2. MENU7
Figure 2.3: COMM Conguration - RS-232
Figure 2.4: COMM Conguration - USB
2.1.2.2 USB
This unit operates as a USBTMC (USB Test and Measurement Class) device. There are no conguration
options for this interface. Figure 2.4
2.1.2.3 GPIB
The GPIB interface may be congured to addresses 1 through 30. Use the arrow keys to select the numeric eld
beside “GPIB Address” and use the rotary knob to change the value. The user may select each digit by the
arrow keys as well. Figure 2.5
CHAPTER 2. MENU8
Figure 2.5: COMM Conguration - GPIB
Figure 2.6: COMM Conguration - LAN (DHCP)
2.1.2.4 LAN
The LAN (Ethernet) interface may be congured as either DHCP or static. Use the arrow keys to select the IP
Mode eld. When DHCP is selected, the current IP, subnet, and gateway are shown on screen (Figure 2.6).
When “MANU” is selected (static mode) (Figure 2.7), the screen shows elds for setting the IP, subnet, and
gateway. Using the arrow keys and the rotary knob allows the user to move between elds and digits, and use
the rotary knob to change the value.
2.1.3System Conguration Menu
From this menu, setting the date, time, brightness, and the state of the beep function is set. Use the arrow keys
to navigate the elds, and the rotary knob to change the value. For the beep function, the soft keys are used to
change the value. When settings are changed, press thekey to commit the changes. See Figure 2.8.
CHAPTER 2. MENU9
Figure 2.7: COMM Conguration - LAN (Manual (static))
Figure 2.8: System Cong Menu
2.1.4Self Test
The unit has a built in self test. This function tests the LEDs, the screen and some of the internal electronics,
Figure 2.9. When the test is running, it cycles through colors on the LCD, and lights up the individual LEDs
that illuminate the front panel buttons. When complete, the results are displayed on screen, DONE for the LCD
and LEDs, and OK for the internal electronic tests.
From this screen, a test of each button is also available. A representation of all the buttons appears on screen
after selecting the “KEY TEST” key, pressing the individual keys will highlight the text on screen indicating the
button is working. Press theto exit the test.
2.1.5System Initialize Menu
System initialization will bring the manifold functions to their initial states. By navigating the functions listed
on-screen and either selecting the check mark or “x” via the soft-keys, sets whether the function is initialized.
CHAPTER 2. MENU10
Figure 2.9: System Self Test
Figure 2.10: System Initialize Menu
2.2MENU > SETUP menus
From the “MENU” function, the next page is the “SETUP” soft-key (Figure 2.11). This screen essentially gives
a summary of the settings controlling measurements.
2.2.1Averaging Setup
When measuring low frequency signals where measurements begin to become unstable, averaging may be useful.
To enable averaging enter the “Average Setup” menu (Menu>Setup>Average Set).
CHAPTER 2. MENU11
FieldDescription
AverageListing of the average mode settings.
Sync SourceThe measurement synchronization signal source.
Line FilterThe state of the line lter.
Update rateHow often to measure
Freq FilterThe state of the frequency lter.
Crest FactorThe Crest Factor setting state
ExSensor1External Sensor 1 state (on/o)
ExSensor2External Sensor 2 state (on/o)
U_RangeThe voltage range setting.
Rate(V/A)External Sensor 1’s scale value
Rate(mV/A)External Sensor 2’s scale value
I_RangeThe current range setting.
Figure 2.11: Setup Information Screen
CHAPTER 2. MENU12
State The state of the average function, On or O.
Type Exp or Line. Exp enables index averaging, and Line enables linear averaging.
Tcontrol Sets averaging to being either a rolling of repeating average.
Count EXP averaging attenuation constant, or LINE average count.
2.2.1.1 EXP - Index averaging
Dn= D
n−1
Mn− D
+
n−1
k
(2.1)
Dn The value displayed after the nth index averaging (D1, the value displayed after the rst averaging, is equal
to M1)
Dn-1 the value displayed after the (n-1)th index averaging
Mn the nth measured data.
K attenuation constant (1-64)
2.2.1.2 LINE - Linear averaging
Dn=
M
n−(M−1)
+ . . . + M
m
n−2
+ M
n−1
+ M
n
(2.2)
Dn The value displayed after linear averaging of m values from the n − (m − 1)thto nthvalue
M Measured value
m The number of values in the average set (1-64)
When m is divisible by n, the calculated value is the moving average; otherwise calculated value is the repeated
average.
NOTE: When index averaging is set, averaging is implemented under the harmonic measurement function.
When linear averaging is set, averaging can only be implemented in the conventional measurement function and
this mode is not applicable to the harmonic measurement function.
2.2.2External Sensor Setup
Figure 2.12 accesses via Menu>Setup>Ext Sen Set contains settings used with external sensors. The state
(on/o) and gain (V/A (mA)) are set here.
2.2.3Other settings: Filter and Crest Factor
Congure the measurement lter, synchronization signal source and crest factor from the “Other Set” menu.
Access this menu via Menu>SET UP>OTHER SET. The “Other Set” menu is shown in Figure 2.13.
CHAPTER 2. MENU13
Figure 2.12: Ratio Set Menu
2.2.3.1 Filter
Frequency lter it is inserted in the frequency measurement circuit and may aect frequency measurement. It
can be used for ltering high-frequency components of interference to make the measured frequency
parameter more accurate. When the frequency lter is switched on, the voltage or current of no more
than 200Hz can be measured. The cuto frequency is 500HZ.
Line lter it is inserted in the voltage and current measurement circuit and has direct inuence on measurement
of the voltage, current and power. When the line lter is switched on, noise and high-frequency
components from the inverter or distortion waveform can be ltered. The cuto frequency is 500HZ.
2.2.3.2 Crest factor
The crest factor is the ratio of the waveform peak to the eective value. The Measurement Conditions crest
factor of 5335B is specied as the times of the crest value which can be input under the rated input conditions.
The crest factor CF3 or CF6 can be selected in the interface “Menu > SETUP > OTHER SET”. See Chapter 4
for more details.
2.2.4Inrush Measurement Setup
Feature not currently implemented.
CHAPTER 2. MENU14
Field
Sync Source
Freq Filter
Line Filter
Crest Factor
Update Rate
Description
Select the synchronization source: U/I/OFF. The overall interval of
the signal voltage, current or data updating cycle can be selected as
the synchronization source for measurement.
Set the status of the frequency lter. When “ON” is selected, the
frequency lter is turned on. When “OFF” is selected, the frequency
lter is turned o.
Set the status of theline lter. When “ON” is selected, the line lter
is turned on. When “OFF” is selected, the line lter is turned o.
Set the crest factor: CF3/CF6
Data Updating Rate Setting key: when this key is pressed, the capture interval of the voltage, current, power and other data, i.e.
data updating rate, can be set.When the data updating rate
is increased, rapid load changes of the power system can be obtained. When the data updating rate is decreased, relative lowfrequency signals can be measured. Options of the dataupdating
rate: 0.1s/0.25s/0.5s/1s/2s/5s
Figure 2.13: Other Set Menu
CHAPTER 2. MENU15
Field
State
Trig Level(A)
Delay Time(ms)
Measure Time(s)
Description
ON/OFF - Enable or disable inrush measurement.
The current level to triggin inrush measurement
The hold o time following a trigger to start measurement
The duration of the measurement following trigger and delay
2.3CAL ZERO
Perform a zero calibration of the unit. Enter this menu via “Menu>Cal Zero”.
2.4RATIO SET
This setting has the eect of scaling the measured value of the current and voltage. For example, a 116V signal
becomes 232V when the “Voltage Ratio” is set to 2. The Ranges are not eected by this setting, so the 150V
range is still shown and used. Use the arrow keys to move between parameters and digits. Use the knob to
change the value of the selected digit. To commit the setting change, press thebutton. See Figure 2.4
CHAPTER 2. MENU16
Chapter 3
External Current Sensor
More information to be added in future versions of this documentation.
When using an external current sensor, the power meter has more current range options.
Central to setting the proper range of the instrument is the Crest Factor. Crest factor is the ratio of the peak
value of a waveform to the RMS value of the waveform. For example, a perfect sine wave crest factor RMS
value is 0.707*Peak. The crest factor is the inverse of the normalized RMS value, 0.707−1= 1.414. In many
cases, like the current waveform of a AC-DC power supply, the crest factor is a larger value. For accurate
measurement, the signal measured should stay within the measurement range of the instrument. For example, a
100V RMS signal with a crest factor of 5 has peaks of 500V. If the range selected is smaller than 500V than
inaccuracy will occur as the signal will be clipped.
4.0.1Crest factor
The Crest Factor, the ratio of the peak to the RMS value, has two settings, 6 or 3. This setting applies to both
current and voltage simultaneously, and modies the available ranges listed in Section 4.0.2. Essentially, by
knowing the RMS signal level and the peak value, the setting of this parameter is determined. For an 117V
RMS AC line signal with little distortion and a purely real load (power factor = 1), the peak value is
1.414 ∗ 117V = 165V . 1.414 is the approximate Crest Factor of a Sine wave. So, the appropriate setting will be
3 (CF3).
CrestF actor(CF ) =
As this meter also measures the DC level of the signal, this must also be accounted for when choosing the Crest
Factor. If the the signal will exceed the range of the unit (crest factor * range) then the measurement value will
be clipped and the value inaccurate. Conversely, by choosing a Crest Factor and Range combination that
accommodates the peak value of the signal, a larger crest factor than set may be measured. The trade-o is
then the accuracy, and is listed in the specications, Chapter 10.
Example (Range=150 V, CF = 6):The maximum input signal and crest factor are as follows:
V
= 100Range = 150VV
rms
So, a signal with a peak that is 9 times that of the RMS value (Crest Factor 9) is measurable using these
settings.
P eak
RM S
maxrange
= 900V
4.0.2Set Measurement Range
The appropriate measurement range (voltage and current range) must be set for accurate measurement.
18
CHAPTER 4. MEASUREMENT SETUP19
1. In the “Meter” interface Press the soft key corresponding to either “U-RANGE” or “A-RANGE”, and use
the knob or the arrow keys to select the voltage or current range desired. See Table 4.1
2. Press the “Enter” key to conrm the setting. Otherwise the instrument will automatically conrm and exit
the setting after 5 seconds of no activity.
Table 4.1: Current and Voltage Ranges per Crest Factor setting
4.1Voltage and current range
4.1.1Fixed range
Select the required range from a number of options. For the voltage range, when the crest factor is 3, the
maximum option is 600V and the minimum option is 15V. When the crest factor is 6, the maximum option is
300V and the minimum option is 7.5V. See Table 4.1
Measurement accuracy is improved by minimizing the range for a
given input signal.
4.1.2Auto Range
The measurement range adjusts automatically according to the input signal, and uses same ranges as in Fixed
range mode. See Table 4.1.
Adjustment of the range occurs when the following conditions occur:
Increasing range
• Urms or Irms exceeds 110% of the current range setting.
• When the crest factor is 3, and the peak voltage or current exceeds 330% of the range currently set.
• When the crest factor is 6, and the peak voltage or current exceeds 660% of the range currently set.
Decreasing range
• Urms or Irms is less than or equal to 30% of the current measurement range.
• The crest factor is 3. The value Upk or Ipk of the input signal is less than 300% of the range at the
following level.
• The crest factor is 6. The value Upk or Ipk of the input signal is less than 600 % of the range at the
following level.
The selected automatic range may change when the input is a random
pulse. In this case, a xed range should be used.
CHAPTER 4. MEASUREMENT SETUP20
4.2Measurement interval
The measurement interval is the time during which data is taken. The fundamental frequency of the measured
signals restricts the measurement intervals that yield meaningful results. For example, measuring 10Hz with an
interval of less than the 0.5s setting will not yield stable results. The unit needs a number of cycles to be
present to determine the frequency, and measurement intervals are not synchronized to the input.
Note: The scope timing parameters do not control the measurement interval.
• A cycle is determined by detecting zero crossings.
• Either positive, or negative going transitions of the synchronization source determine the measurement
timing. If the zero crossings do not occur within the set update interval, the entire interval is used for
calculating the various measurements. See Figure 4.1.
The choice of rising or falling transition is determined as that which gives the greatest amount of time.
• For peak value measurements (Ipk+, Upk-, etc…), the entire update interval is used. Measurements of
Crest Factor, as it is the Peak to RMS ratio, are also determined using the entire measurement update
interval.
• The input signal used as the synchronization source is congurable, and determines the synchronization
with the zero point of that input signal. The overall interval of the voltage, current or data updating cycle
may be selected as the synchronization source.
Figure 4.1: Measurement Timing
The data updating cycle refers to the cycle used for calculating sampling data of the measurement function. It is identical to the set
value of the data updating rate.
The slope refers to signal changes from low level to high level (rising
edge) or from high level to low level (descending edge).
4.3Filter and Crest Factor Setting
1. Select Menu → SET UP → OTHER SET and enter the OTHER conguration page.
2. Pressto select the parameter to be congured (blue font background), and then press the soft key
corresponding to the parameter on the right to set the required value, as shown in the Figure 4.3.
3. Press the Enter key to save the settings.
CHAPTER 4. MEASUREMENT SETUP21
Figure 4.2: Measurement Timing 2
FunctionDescription
Sync Source
Freq FilterEnable or disable the Frequency Filter.
Line FilterEnable or disable the Line Filter.
Crest FactorSet the crest factor to 3(CF3), or 6 (CF6) (See Section 4.0.1)
Update Rate
Synchronization Source, the source used to determine the measurement interval. May be set to voltage (U), current (I) or turned o.
Data Update Rate: the capture interval of the voltage, current, power
and other data can be set. When the data updating rate is increased,
rapid load changes of the power system can be measured. Conversely,
when decreased, relative low-frequency signals can be measured. Values = 0.1, 0.25, 0.5, 1, 2, and 5 seconds
Figure 4.3: Other Settings Display
4.3.1Filter
Frequency FilterIt is inserted in the frequency measurement circuit and may aect frequency measurement.
It can be used for ltering high-frequency components of interference to make the measured frequency
parameter more accurate. When the frequency lter is switched on, the voltage or current of no more than
200Hz can be measured. The cuto frequency is 500Hz.
CHAPTER 4. MEASUREMENT SETUP22
Line lterIt is inserted in the voltage and current measurement circuit and has direct inuence on
measurement of the voltage, current and power. When the line lter is switched on, noise and high-frequency
components from the inverter or distortion waveform can be ltered. The cuto frequency is 500Hz.
4.4Averaging Function
1. Select⇒ SET UP ⇒ AVERAG SET to enter the conguration page of the averaging function.
2. Pressto select the parameter to be congured (blue background), and press the soft key
corresponding to the required value. See Figure 4.4.
3. Press the Enter button to conrm the setting.
Figure 4.4: Averaging Settings
CharacterFunction description
StateEnable or disable averaging.
ModeAverage mode Line=linear or Exp=cumulative moving average
• If the mode of the averaging function is set as EXP (index averaging), the attenuation constant can be set.
• If the mode of the averaging function is set as LINE (linear averaging), the averaging times can be set.
4.4.1Index averaging
Dn= D
n−1
Mn− D
+
n−1
k
CHAPTER 4. MEASUREMENT SETUP23
D
The value displayed after the nth index averaging, (D1, the value
n
displayed after the rst averaging, is equal to M1)
D
M
The value displayed after the (n-1)th index averaging
n−1
the nth measured data.
n
kattenuation constant (1-64)
4.4.2Linear averaging
M
n−(m−1)
D
n
M
n−(m−1)
M
n−2
M
n−1
M
n
Dn=
linear average of m values from the (n − (m − 1))thto nthvalue
(n − (m − 1))thmeasured data
(n − 2)thmeasured data
(n − 1)thmeasured data
nthmeasured data
Mthe number of average values (1-64)
When m can be divided exactly by n, the calculated value is the moving average; when there is no particular
relationship between m and n, the calculated value is the repeated average.
+ . . . + M
m
n−2
+ M
n−1
+ M
n
When index averaging is set, averaging is implemented under the harmonic measurement function.
When linear averaging is set, averaging can only be implemented in the conventional measurement function and this mode is not applicable to the harmonic
measurement function.
The following measurements are subject to direct averaging:
1. Urms, Umn, Udc, Urmn, Uac, Irms, Imn, Idc, Irmn, Iac, P, S and Q.
2. Ucf, Icf, λ and WPAV are calculated by operation of the averaged Urms, Irms, P and S.
4.4.3Harmonic measurement averaging
The following measurement functions are subject to direct averaging:
1. U(k), A(k),W(k), S(k) and Q(k).
2. λ(k) is calculated via operation of the averaged W(k) and Q(k).
3. U(%r), A(%r), W(%r) , U(%f), A(%f) and W(%f) are calculated via operation of the averaged U(k),
A(k) and W(k). (k indicates the harmonic times.)
Chapter 5
Meter Display
The 5335B has 3 congurable display formats. Each format also allows for 5 dierent congured sets of
measurement. 1 large and 6 small, 4 large and 6 small, or 12 small measurements may be displayed. See
Figures 5.1, 5.2, 5.3.
5.0.1Operation steps
1. Press
2. The “View” soft-keys select the display format (1,4,12). Repeated press of the soft-key cycles through the
dierent congurations, up to 5 of them.
3. The “CONFIG” soft-key enters the conguration mode for that set of pages.
4. The “FUNC” soft-key cycles through the sets of measurements available for Power (P), Current (I), and
Voltage (U).
5. To change the measurements shown on a page, use the arrow keys and move the cursor to an on-screen
parameter, and select the desired measurement from the soft-keys on the right of the display.
6. To add or remove a page press the “INSERT PAGE” or “DELETE PAGE” respectively. 1 to 5 pages are
supported.
24
CHAPTER 5. METER DISPLAY25
Figure 5.1: 1 main measurement
Figure 5.2: 4 main measurements
CHAPTER 5. METER DISPLAY26
Figure 5.3: 12 main measurements
Chapter 6
Waveform Display Function
The 5335B power meter displays waveform representations of the sampled data measured. This function is
designed to be similar to an oscilloscope with many of the typical functions of oscilloscopes available, like
trigger, run/stop and single trigger. The capture of data is not limited to that displayed onscreen, the
measurement interval is also involved. For example, measuring a 15Hz signal with the interval set to 0.1s does
not allow for stable measurement of the frequency. The waveform is however still viewable at 100ms/div.
To enter the waveform function press thebutton.
6.0.1Trigger Status
The Auto/Stop eld in the upper right corner of the display, below the time shows the trigger status. The
meanings are as follows:
Auto Auto mode and a trigger has occured
Auto? Auto mode waiting for a trigger
Trig? Single mode waiting for a trigger
Stop Capture stopped. Reached either by pressing “stop” or completing a single mode capture
Figure 6.1: Waveform Display Details
27
CHAPTER 6. WAVEFORM DISPLAY FUNCTION28
Parameter nameParameter descriptions
V_RANGEVoltage range setting: press to set the voltage range
A_RANGECurrent range setting: press to set the current range
RUN/STOPRun/stop: press to run or stop waveform capture
SINGLE
SCOPE(U/A/UA)Select the waveform(s) to display on screen (voltage/current/voltage and current)
KNOB SEL (U/A/TL/TD/Td)
AUTO
TRIG SETTrigger setup menu. See Section 6.1
MEASURE SETMeasurement setup menu.
Single measurement: pressing this stops capture and proceeds with a single capture
of data following a trigger event.
Knob selection, the parameter adjusted by turning the knob. voltage waveform
position (U), current waveform position (I), trigger level (TL), trigger delay (TD),
time/division (T/d)
Automatic mode: the power meter will automatically set the waveform display
conguration
Table 6.1: Waveform Display Parameters
Figure 6.2: Trigger Setup Screen
6.1Trigger Setup
When you need to enable the trigger function, you should select the trigger source, trigger mode, trigger slope
and other trigger-related conguration. Detailed steps are as follows:
Operation steps
1. Pressto enter the waveform display interface.
2. In the waveform display interface Press the soft key corresponding to the “TIRG SET” parameter to enter
the trigger setting interface, as shown below.
3. Press the right soft key corresponding to the parameter to select the required trigger conguration.
Source trigger source
CHAPTER 6. WAVEFORM DISPLAY FUNCTION29
Mode trigger mode
Slope trigger slope
6.1.1External trigger input (Ext)
When the trigger source is set as Ext, input the trigger signal into the external signal input interface
(Synchronous) of the rear panel according to the following specications.
NOTE: When the voltage higher than 0-3.3V is applied on the external signal input interface (Synchronous),
the instrument may be damaged.
Minimum pulse width refers to the width of the high or low trigger level, at least 1µs.
Trigger delay time refers to the delay between the appearance of the trigger level and the response of CPU,
within (1µs + 3 sampling cycles).
Chapter 7
Harmonic Measurement Function
With the 100kHz bandwidth, the 5335B power meter can realize harmonic measurement of high speed and wide
dynamic range. The voltage, current, active power, reactive power, phase of harmonics and total harmonic
distortion (THD) factor can be tested in harmonic measurement mode. Display of harmonic parameters is either
in list or bar chart form for clear analysis of test results.
To enter this mode, select thebutton, and the initial harmonic measurement interface starts.
Parameter nameParameter descriptions
V_RANGEVoltage range setting
A_RANGECurrent range setting
RUN/HOLD
RESETClear measured values
FUNC(W/A/V)
BARDisplay the bar chart mode
LISTDisplay the list mode
SETUPEnter the setup menu. See Section 7.3
Set the instrument to either “run” and continue measuring, or “hold” to freeze the measurements
When in bar mode, rotate between viewing power, current, and voltage harmonic measurements
Table 7.1: Harmonic Measurement Parameters
7.1Bar Graph Mode
The bar chart is used for displaying the percentage of dierent harmonics. Enter this mode by pressing the
button (if not in the harmonic mode already), then select “BAR” from the on-screen soft-keys.
Harmonics from 1 to 50 may be displayed at once. From the list of harmonics, Odd, Even or All harmonic
subsets may be displayed. The bar graph shows either the Power, Current, or Voltage harmonic components.
Select the function by repeatedly pressing the “FUNC” soft-key.
On the bar graph display, individual harmonics may be selected by using the arrow keys or the rotary knob.
Figure 7.2 shows the 9th harmonic of 62Hz. Also shown highlighted in cyan is the bar and measurements
associated with that particular harmonic (frequency, harmonic content, harmonic distortion factor and phase).
30
CHAPTER 7. HARMONIC MEASUREMENT FUNCTION31
h!
Figure 7.1: Bar graph mode - Harmonics
h!
Figure 7.2: Selecting Specic Harmonics
7.2List Mode
Enter this mode by pressing thebutton (if not in the harmonic mode already), then select “LIST” from
the on-screen soft-keys. This list is used for showing the voltage, current, active power, reactive power, phase
and total harmonic distortion (THD) factor of dierent harmonics. Harmonic lists can be displayed as a
complete, odd, or even set. See Figure 7.3. From the “SETUP” menu, the list of harmonics is dened. The 1
through the 50th, odd, even, or all harmonics to display is congured here.
The list is composed 4 pages. Navigate to each page using the right and left arrow keys. The table columns are
listed in Table 7.2
st
CHAPTER 7. HARMONIC MEASUREMENT FUNCTION32
Figure 7.3: List mode - Harmonics
AbbreviationsInstruction
U(V)Voltage
I(mA)Current
P(mW)Power
Q(mvar)Reactive Power
S(mVA)Apparent Power
PF()Power Factor
ϕUI(◦)Phase dierence of k-order harmonic voltage and harmonic current
ϕUU(◦)Phase dierence of harmonic voltage U(k) and fundamental U(1)
ϕII(◦)Phase dierence of harmonic current I(k) and fundamental I(1)
U(%r) or U(%f )Voltage harmonic distortion
I(%r) or I(%f )Current harmonic distortion
P(%r) or P(%f )Power harmonic distortion
Table 7.2: Measurement Parameters
CHAPTER 7. HARMONIC MEASUREMENT FUNCTION33
7.3Setup Menu
The setup menu allows the user to set the distortion factor calculation formula, PLL source, and set of harmonics
to analyse and display. From the Harmonic Measurement function,, press the “MENU” soft-key.
Figure 7.4: Harmonic Measurement Setup Menu
7.3.1Parameters
Use the arrow keys to select the dierent parameters, and if numeric, the rotary knob to set the value, or the
soft-keys.
THD Formal Distortion factor calculation formula.
% r displaying harmonics in the form of percentage to the overall current (voltage, power) amplitude of
all harmonics.
% f displaying harmonics in the form of percentage to the fundamental wave current (voltage, power).
PLL Source Select the PLL (Phase Locked Loop) source: U/I/OFF. Used for determining the fundamental
wave cycle as the reference for analysis of harmonic orders.
Serial Harmonic sequence: whole sequence/odd sequence/even sequences
Order Max Set the harmonic analysis orders (1-50). You can specify the harmonic measurement range. These
specied analysis orders are used for calculating the value of the distortion factor.
7.3.2Distortion factor
Distortion is either calculated relative to the fundamental (%f) or to the total signal (%r). Internally, the 5335B
calculates according to the following equations:
CHAPTER 7. HARMONIC MEASUREMENT FUNCTION34
Measurement function%r%f(7.1)
Harmonic distortion factor of voltage
Harmonic distortion factor of current
Harmonic distortion factor of active power
Total harmonic distortion rate of voltage
Total harmonic distortion rate of current
Total harmonic distortion rate of active power
√
√
U(k)
U(total)
I(k)
I(total)
P (k)
P (total)
max
∑
k=2
U(total)
√
max
∑
k=2
I(total)
max
∑
k=2
P (total)
U(k)
I(k)
P (k)
U(k)
U(1)
I(k)
I(1)
P (k)
P (1)
√
max
∑
2
2
2
k=2
U(1)
√
max
∑
k=2
I(1)
√
max
∑
k=2
U(k)
I(k)
P (k)
P (1)
2
2
2
(7.2)
(7.3)
(7.4)
(7.5)
(7.6)
(7.7)
NOTE Total harmonics are calculated as follows: U(total) =
max
P (total) =
∑
P (k) . “k” indicates the harmonic order, and max indicates the upper limit of analysis orders,
k=0
√
max
∑
U(k)2, I(total) =
k=0
√
max
∑
k=0
I(k)2,
and the max is determined by the frequency of the PLL source automatically, at most 50.
7.3.3PLL source
In the harmonic mode, 5335B uses a PLL to multiply the input signal fundamental. The frequency
multiplication is used as the A/D sampling clock in the instrument in order to achieve ideally synchronous
sampling. The range of the PLL source is from 10Hz to 1.2kHz. To maximize accuracy, select the cleanest
signal source available, either the voltage or the current. This improves measurement consistency and stability,
especially in relation to harmonic analysis.
If the amplitude level of the signal input into the unit as the PLL source is small relative to the range, PLL
synchronization might fail. If the crest factor is set as 3, the amplitude level of the PLL source should be greater
than 50% of the range. If the crest factor is set as 6, the amplitude level of the PLL source should be greater
than 100% of the range.
When the fundamental wave frequency is less than 200Hz, the frequency lter should be turned on. When lower
than 440Hz and containing high-frequency components, it is also best to turn on the frequency lter. This lter
only applies to the frequency measurement circuit.
If the frequency of the PLL source changes, the correct measured value will be displayed after data are updated
several times. The frequency of the PLL source needs to be tested again via the PLL circuit in the instrument in
case of changes of the PLL source or its frequency, the correct measured value will be determined afterward.
NOTE If the analysis window determined by the fundamental frequency of the PLL source is longer than the
data updating cycle, no harmonic data will be detected. In this case, the data update cycle should be lengthened.
For example, when the fundamental wave frequency of the PLL source is 10Hz (cycle: 100ms), the analysis
window is one-wave long, and the data measurement interval is 100ms. In this case, the harmonic measurement
CHAPTER 7. HARMONIC MEASUREMENT FUNCTION35
time is approximately more than or equal to 150ms (data measurement interval and data processing time).
Therefore, select the data update rate of 250ms or more for measurement and display of harmonic data.
Chapter 8
Integral Operation Function
Access the integration function by pressing the “Integ” button. This function integrates power and current
measurements over time to generate measurements in producing (sold) or consuming (buying) directions.
The 5335B power meter can be used for integral operation of the current and power of the input unit. Technical
indicators can be calculated. In addition, the range can be switched automatically in the Buy and Sell modes
according to the input level so as to accurately complete integrate measurement.
During operation of the air-conditioner, refrigerator, induction cook and other household appliances, the working
status will change frequently, resulting in changes of the working current in a large scale. The current may
decrease of dozens of amperes to a few milliamperes. Ordinary power measuring instruments can only be used
for integral measurement within one range. If the measured value exceeds the current maximum range, the
measurement result is inaccurate. You need to manually switch the range before integral measurement. Integral
measurement of equipment subject to large current changes cannot be performed continuously. The BK5335B
power meter can automatically switch the range in the integral mode. Therefore, errors of integral measurement
caused by manual range switching can be eliminated, and the power consumption of household appliances can
be measured more accurately. See the schematic diagram below.
Figure 8.1: Range Switching
36
CHAPTER 8. INTEGRAL OPERATION FUNCTION37
Figure 8.2: Meter Display
Parameter nameParameter descriptions
V_RANGE
A_RANGE
RUN/HOLD
RESETReset.
FUNC(w/q/AV)
LARGE
(WP/WP+/WP-,
q/q+/q-, WPAV)
START
STOP
SETUPSet relevant parameters of integral measurement.
Voltage range setting: press the soft key corresponding to this parameter to set
the voltage range.
Current range setting: press the soft key corresponding to this parameter to set
the current range.
Run/hold: press the soft key corresponding to this parameter to run or hold the
integral function.
Selection of the integral function: active power integral (W), current integral (q)
and average active power integral (AV).
Selection of the item to be displayed in a amplied manner: WP (watt hour, the
sum of positive and minus watt hours), WP+ (the consumption of positive watt
hours), WP- (the negative watt hours of feedback power), q (ampere hour, the sum
of positive and negative Ampere hours), q+ (the consumption of positive Ampere
hours), q- (the negative ampere hours of feedback power) and WPAV (average
active powerintegral).
Integral Start button In the manual start mode, the integral function can be enabled
by pressing the soft key corresponding to this parameter.
Integral Stop button In the manual stop mode, the integral function can be disabled
by pressing the soft key corresponding to this parameter.
Table 8.1: Integral Measurement Soft Keys
8.1Introduction of soft keys on the interface
When the “Integ” button is pressed, the initial integral measurement interface below will appear.
Description of information of integral measurement interface:
CHAPTER 8. INTEGRAL OPERATION FUNCTION38
Character
Mode
State
Start
Stop
Ready
Time Up
Reset
Error
Time
St
Et
Timer
Function description
Display the integral start and stop mode. Start mode: MANUAL and TIME
Stop mode: MANUAL, TIME and TINTerval.
Display the current status of the integral function.
displayed when the integral function is working.
displayed when the integral function is interrupted, canceled or stopped.
displayed in the ready state of the real-time start mode.
displayed when it reaches the specied time of the integral timer.
displayed when the integral value and integral time are reset via integral resetting.
when the power supply is recovered, the integral function is stopped and the integral
result before power failure is displayed. This integral state is called Error state.
Display the integral time.
Display the set integral time.
Display the set integral ending time.
Display the xed integral time.
Table 8.2: Measurement Information
8.2Integral measurement display information
Description of measurement information:
8.2.1Integral operation
• Active power integral
In watt hour, displayed as WP (watt hours, the sum of positive and negative watt hours), WP+ (the
consumption of positive watt hours) and WP- (the negative watt hours of feedback power).
• Current integral
In q, displayed as q (ampere hours, the sum of positive and negative ampere hours), q+ (the consumption
of positive ampere hours) and q- (the negative ampere hours of feedback power)
• Average active power integral
In watt, displayed as WPAV (average active power integral)
• Integration time
Displayed as Time, in the format of hhhh:mm:ss.
8.3Specication
• Save in case of power failure
The integral result can be kept in the memory even in case of power failure during operation of the
integral function. When the power supply is recovered and the integral function is disabled, the integral
result before power failure is displayed. When the power supply is recovered and the integral is rest, the
integral function is enabled again.
• Display resolution
The maximum display resolution of the integral value is 99999. When the integral value reaches 100000,
CHAPTER 8. INTEGRAL OPERATION FUNCTION39
Function
Wiring
Measurement range
Filter
Averaging function
Synchronization
Data updating rate
Integral mode
Integral timer
Integral start
Integral stop
Integral resetting
Save operation
Hold operation
Single measurement operation
Integral resetting
ExecutableUnenforceableUnenforceable
ExecutableUnenforceableUnenforceable
ExecutableUnenforceableUnenforceable
ExecutableUnenforceableUnenforceable
ExecutableUnenforceableUnenforceable
ExecutableUnenforceableUnenforceable
ExecutableUnenforceableUnenforceable
ExecutableUnenforceableUnenforceable
ExecutableUnenforceableExecutable
UnenforceableExecutableUnenforceable
ExecutableUnenforceableExecutable
ExecutableExecutableExecutable
ExecutableExecutableExecutable
ExecutableExecutableExecutable
Table 8.3: Settings congurability per mode
Currentintegral
status
Integralinterruption
the decimal point will move automatically. For example, when 0.01mWh is added to 999.99mWh,
1.0000Wh will be displayed.
• Display in case of overow
When the integral value meets the following overow conditions, the integral function will be disabled, and
the integral time and integral value at this point will be kept.
– The integral time reaches the maximum value (10000hours).
– The integral values of WP, q and WPAV reach the maximum/minimum display integral values.
• Maximum/minimum display integral value
– Active power integral (WP): ±99999Mwh
– Current integral (q): ±99999MAh
– Average power integral (WPAV): ±99999Mw
• Restricted implementation
Settings of the following functions cannot be changed during integral operation.
NOTE In addition to the above items, other settings (such as self-testing and date/time setting) or operations
with inuence on the integral cannot be executed; otherwise, errors may be caused.
8.4Setting of Integral Measurement Conguration
You can set the start mode, stop mode, automatic zero setting, automatic calibration, watt-hour integration,
current integration and other parameters of integral measurement. Specic steps are as follows:
Operation steps
CHAPTER 8. INTEGRAL OPERATION FUNCTION40
1. Press “Integ” to enter the integral measurement interface.
2. Press the soft key corresponding to the “SETUP” parameter in the integral measurement interface to
enter the integral parameter conguration interface.
8.4.1Parameter descriptions:
Start set in the MANUAL mode or TIME mode via the right soft key.
MANUAL press the “START” soft key in the integral measurement interface in the manual start mode to
trigger the manual start of the integral function.
TIME: the date on which the integral mode is started can be set in the real-time start mode. Integration will
be started on the set date. When the time/date combination is set in the real-time start mode, integration
will not be started before the current time and date. Integration will be started when the screen is
updated once at least before the start time.
Stop: set in the MANUAL mode, TIME mode and TINTerval mode via the right soft key.
MANUAL: press the “STOP” soft key in theintegral measurement interface in the manual stop mode
tomanually stop integration. When the integration time reaches the maximum integration time (10,000
hours) or the integral value reaches the maximum/minimum display value, integration will be stopped,
and the current integral time and integral value will be kept.
TIME The date for stop of the integral mode can be set in the real-time stop mode. Integration will be stopped
on the set date. When it reaches the set time or the integral value reaches the maximum/minimum display
integral value, integration will be stopped, and the current integral time and integral value will be kept.
TINTerval the integral measurement time can be set in the set-time stop mode. When it reaches the set
ending date and time or the integral value reaches the maximum/minimum display integral value,
integration will be stopped, and the current integral time and integral value will be kept.
Auto Clear used for enabling (ON) or disabling (OFF) the function of automatic zero clearing of the integral.
On: integration is restarted when the START button is pressed to start integration. OFF: integration is
restarted based on the integral value at the end of previous integration when the START button is pressed
to start integration.
Auto Cal used for enabling (ON) or disabling (OFF) the function of automatic integral calibration. When
automatic integral calibration is enabled, zero setting is implemented. In this case, the power and current
measured just now are subject to integration.
WP type used for selecting the integral mode for positive and negative watt hours. There are four integral
modes:
Charge/Discharge: used for measuring the DC positive and negative watt hours (integration of sampling data).
Sold/Bought used for measuring the AC positive and negative watt hours (value integration in each data
updating cycle).
q type used for selecting the current integration mode. Options of the current integration mode are as follows:
rms Eective value;
mn Calibration to the average rectied value of the eective value.
dc Simple averaging;
rmn Average rectied value;
ac AC component
CHAPTER 8. INTEGRAL OPERATION FUNCTION41
8.5Integration
When the integral measurement function is enabled, you can keep the current integral information and carry out
the following operations: exit, start and stop. Specic steps are as follows:
Operation steps
1. Press “Integ” to enter the integral display interface.
2. In the integral measurement display interface Press the soft key corresponding to the “START” parameter
to start or recover integral operation. The Start status and integral operation value are displayed in the
interface.
3. When you need to keep the current integral status and integral result, you can press the soft key
corresponding to the “RUN HOLD” parameter. Then the current status of the integral result display and
communication output will be kept no matter whether integral operation is running.
4. When you need to continue to implement integral operation, double-press the soft key corresponding to
the “RUN HOLD” parameter to exit the HOLD status. The power meter will display the integral result
after exiting the HOLD status. The schematic diagram of display of the HOLD/RUN integral result is
shown below.
5. When you need to suspend integral operation, press the soft key corresponding to the “Stop” parameter to
suspend integral operation.
6. Press the soft key corresponding to the “RESET” parameter to reset the integral value and integral time.
The relationship diagram of start, suspension and resetting of the integral operation status is shown below.
Auto Clear when ON is selected.
Auto Clear when OFF is selected.
Chapter 9
Remote Operation
There are four types of communication interfaces available:USB,Ethernet,GPIB and RS232.You can choose any
one of them to communicate with a PC.
9.1RS-232 Interface
RS232 interface:use a cable with two COM interface (DB9) to connect power meter and PC. It can be activated
by menu key on the front panel.All SCPI commands are available through RS-232 programming.
• RS-232 data is a 10-bit word with one start bit and one stop bit. The number of start and stop bits is not
programmable.
• Baudrate Available baudrate:4800 9600 19200 38400 57600 115200
• RS-232 connection The RS-232 serial port can be connected to the serial port of a controller (i.e.,
personal computer) using a straight through RS-232 cable terminated with DB-9 connectiors. Do not use
a null modem cable. Table 9.1 shows the pinout for the connector.
If your computer uses a DB-25connector for the RS-232interface,you will need a cable or adapter with a DB-25
connector on one end and a DB-9 connector on the other,wired straight through(not null modem).
RS-232
9.1.1RS-232 troubleshooting
If you are having trouble communicating over the RS-232 interface,check the following: The computer and the
power meter must be congured for the same baudrate, parity, number of data bits,and ow control options.
Figure 9.1: RS-232 DE-9 Connector
42
CHAPTER 9. REMOTE OPERATION43
Pin #Signal
1NC
2TX
3RX
4NC
5GND
6NC
7CTS
8RTS
9NC
Table 9.1: RS-232 (DE-9) Pinout
Note that the eledtronic load is congured for 1 start bit and 1stop bit (these values are xed). The correct
interface cables or adapters must be used, as described under RS-232 connector. Note that even if the cable has
the proper connectors for your system,the internal wiring may be incorrect. The interface cable must be
connected to the correct serial port on your computer (COM1, COM2,etc.).
9.1.2RS-232 Communication settings
Before communication operation, please make sure that the following parameters of electronic load match that
of PC.
Baud rate 9600(4800,9600,19200,38400,57600,115200). You can enter system menu through panel to set
communication baud rate.
Data bit 8
Stop bit 1
Parity (none,even,odd)
EVEN 8 data bits with even parity
ODD eight data bits with odd parity
NONE eight data bits without parity
Recommended settings Start Bit, 8 Data Bits, Parity=None, 1 Stop Bit
9.2USB interface
Use Type A to Type B USB cables to connect the power meter.
The USB interface capabilities of the power meter are described below:
• The interface is IEEE488.2 standard USB488 interface
• The interface accepts REN_CONTROL,GO_TO_LOCAL ,LOCAL_LOCKOUT request.
• The interface accepts MsgID = TRIGGER USBTMC command message and forwards TRIGGER requests
to the function layer.
CHAPTER 9. REMOTE OPERATION44
The USB488 device capabilities of the power meter are described below:
• The device understands all mandatory SCPI commands.
• The device is SR1 capable.
• The device is RL1 capable.
• The device is DT1 capable.
9.3GPIB interface
First conncet GPIB port of power meter to GPIB card of PC. They must be sucient contact and tighten the
screws. And then set address. The address can be set from 0 to 30.
9.4LAN interface
Use a network cable to connect PC through LAN interface of the power meter.
• IP mode
MANU: manually set IP address, subnet mask, default gateway, etc.
DHCP (Dynamic Host Conguration Protocol)
DHCP refers to a protocol that temporarily assigns necessary information to PCs connected to the
Internet. If the network is provided with an available DHCP server, the server will automatically assign
information (IP address, subnet mask and default gateway) to PCs and other equipment connected to the
network. 5335B cannot use DHCP unless the network is provided with a DHCP server. Please ask your
network administrator DHCP if DHCP is available.
• IP Address
IP addresses assigned to 5335B can be set. Default address is 192.168.000.000.
• IP address refers to ID assigned by the network to each equipment (Internet or enterprise Intranet).
• IP address refers to four 32-bit values within 0 - 255 which are separated by decimal points, such as
[192.168.111.24].
• Please ask the network administrator to provide an IP address.
• Use DHCP network to automatically set the IP address.
• Subnet masksetting
The mask used when subnet website is determined from an IP address may be set. Default address is
255.255.255.255.
• Vast TCP/IP networks like Internet are often divided into several small networks, i.e. subnets. The subnet
mask is used to identify how many digits of the 32-digit values in the IP address belong to the subnet
address. The part beyond the network address is used to identify the host number of each PC connected
to the network.
• Ask your network administrator about values of subnet mask.
• Use DHCP network to automatically set the subnet mask.
• Gatewaysetting
IP address of gateway may be set to communicate with equipment in other network (default gateway).
Default address is 192.168.000.000.
CHAPTER 9. REMOTE OPERATION45
• The default gateway controls data exchange between networks and protocols to ensure smooth data
transmission.
• Ask your network administrator about values of default gateway. It is possible that the setting is
unnecessary.
Use DHCP network to automatically set the default gateway.
Chapter 10
Specications
The specications listed below are valid and specied for the following conditions:
• Warm up time of 30 minutes
• Ambient temperature - 23 ± 5
• Relative humidity 30 to 75%
◦
46
Power Meter
5335B
Specifications
Specifications are subject to the following conditions
Temperature: 23±5° C, humidity: 30 to 75% RH.
Warm-up time: 30 minutes
Model5335B
General Measurement Specifications
Voltage, Current
Basic measurements
PowerReal, Apparent, Reactive, DC, Power factor
TimeFrequency, Phase
IntegrationTotal power, Total current, Maximum power, Minimum power
Type
Harmonic measurements
RangeDC up to 50
Max. Frequency100 kHz
Input bandwidthDC, 0.5 Hz to 100 kHz
Measurement methodDigital sampling
A/D ConverterSimultaneous conversion of voltage and current inputs, Resolution: 18-bit, Maximum conversion rate: 10 μs
Line filterSelect OFF or ON (cutoff frequency at 500 Hz)
Peak (max,min)Voltage, current, or power
Input voltage continuous max.1.5 kV-peak or 1 kV-RMS, whichever is less
Input voltage transient (<1s) max.2 kV-peak or 1.5 kV-RMS, whichever is less
Input voltage common-mode max.600 Vrms
Voltage input impedance2 MΩ + 13 pF in parallel (typical)
5 mA to 200 mA range505 mΩ + 0.1 μH
Current input
0.5 A to 20 A range5 mΩ + 0.1 μH
impedance (typical)
Sensor input
5 mA to 200 mA range30 A-peak or 20 A-RMS, whichever is less
Input current continuous max.
0.5 A to 20 A range100 A-peak or 30 A-RMS, whichever is less
Sensor inputPeak value less than or equal to 5 times the rated range
5 mA to 200 mA range30 A-peak or 20 A-RMS, whichever is less
Input current transient (<1s) max.
0.5 A to 20 A range150 A-peak or 40 A-RMS, whichever is less
Sensor inputPeak value less than or equal to 10 times the rated range
Voltage Measurement Accuracy and Ranges
Ranges
DC to 1 kHz±(0.1% + 0.2% F.S.)
2
Accuracy
(line, frequency, &
digital filter set to off)
1 kHz < f ≤ 10 kHz
10 kHz < f ≤ 100 kHz
For temperature changes after zero-level
Temperature
coefficient
compensation or range change
Influence of self-generated heat caused by voltage
input (U is the voltage reading (V))
1
Input signal frequency in kHz
2
Input waveform: Sine wave crest factor: 3, common-mode voltage: 0 V, power factor: 1
Frequency filter: Turn on when measuring ≤ 200 Hz
Peak to peak, Maximum, Minimum, Average_rms, Average_rectified, DC,
Crest factor (current), Inrush (current)
Current, Voltage, Real power, Apparent power, Reactive power,
Power factor, Phase, Percentage of total (Current, Voltage, Power)
20 kΩ (50 mV to 2 V)
100 kΩ (2.5 V to 10 V)
CF=3: 15 V, 30 V, 60 V, 150 V, 300 V, 600 V
CF=6: 7.5 V, 15 V, 30 V, 75 V, 150 V, 300 V
±((0.07 f
±(0.5% + 0.5% F.S.) ± [{0.04 × (f
+ 0.02% F.S. /°C to the DC voltage accuracy
+ 0.0000001 × U
2
+ 0.0000001 × U
% + 0.0000001 × U
th
order
1
)% + 0.3% F.S.)
1
- 10)}% ]
2
% to the AC voltage accuracy
2
% F.S. to DC current accuracy
4
www.bkprecision.com
Power Meter
5335B
Specifications (cont.)
Current Measurement Accuracy and Ranges
Direct input range
External 1
Sensor input range
External 2
DC to 1 kHz±(0.1% + 0.2% F.S.)
2
Accuracy
(line, frequency, &
digital filter set to off)
1 kHz < f ≤ 10 kHz
10 kHz < f ≤ 100 kHz
2.5 to 200 mA5 μA/ °C (after zero-level compensation, or range change)
Temperature
coefficient
Power Measurement Accuracy
500 mA to 20 A500 μA/ °C (after zero-level compensation, or range change)
Apparent power (S)Voltage accuracy + current accuracy
Reactive power (Q)
Power factor (
Phase angle (Ф)
PF)
±[(PF–PF/1.0002) + abs(cosØ - cos{Ø+sin
±[abs(Ø - cos
Temperature coefficientSame as the temperature coefficient for voltage and current
Frequency Measurement Accuracy
Frequency
measurement
range
Data update interval0.1 s0.25 s0.5 s1 s2 s5 s
Measurement range 25 Hz ≤ f ≤ 100 kHz 10 Hz ≤ f ≤ 100 kHz5 Hz ≤ f ≤ 100 kHz2.5 Hz ≤ f ≤ 100 kHz1.5 Hz ≤ f ≤ 50 kHz0.5 Hz ≤ f ≤ 20 kHz
Accuracy±0.06%
Frequency filter500 Hz low-pass
1
Input signal frequency in kHz
2
Input waveform: Sine wave crest factor: 3, common-mode voltage: 0 V, power factor: 1
Frequency filter: Turn on when measuring ≤ 200 Hz
3
When power factor (PF)=0 (apparent power (S)):
±0.2% of S when 45 Hz ≤ f ≤ 66 Hz
±{(0.2+0.2×f)% of S} when 0.066 ≤ f ≤ 100 kHz
When 0<PF<1(phase angle (Ф)):
(power reading ) × [(power reading error %) + (power range %) × (power range/indicated apparent power value) + {tan
When the line filter is turned ON:
45 to 66 Hz: Add 0.3% of reading
<45 Hz: Add 1% of reading
4
Accuracy when the crest factor is set to 6, the accuracy is obtained by doubling specified accuracies
CF= 3:5 mA, 10 mA, 20 mA, 50 mA, 100 mA, 200 mA, 0.5 A, 1 A, 2 A, 5 A, 10 A, 20 A
CF= 6:2.5 mA, 5 mA, 10 mA, 25 mA, 50 mA, 100 mA, 250 mA, 0.5 A, 1 A, 2.5 A, 5 A, 10 A
CF = 3: 2.5 V, 5 V, 10 V
CF = 6: 1.25 V, 2.5 V, 5 V
CF= 3: 50 mV, 100 mV, 200 mV, 500 mV, 1 V, 2 V
CF= 6: 25 mV, 50 mV, 100 mV, 250 mV, 500 mV, 1 V
±{(0.07 f
±(0.5% + 0.5% F.S.) ± [{0.04×(f
2
% of reading to the AC current accuracies + 0.00013 × I
2
% of reading + 0.00004 × I2 mA (2.5 to 200 mA), add to the DC current accuracy specifications
Apparent power accuracy + ( 1.0004-PF
-1
(influence from the power factor when PF=0%/100)})] ± 1 digit
when voltage and current are at the measurement range rated input
-1
(PF/1.0002)) + sin-1{(influence from the power factor when PF=0%)/100}] deg ± 1 digit
when voltage and current are at the measurement range rated input
1
)% + 0.3% F.S.}
1
-10)}%]
2
% of reading + 0.004 × I2 mA (0.5 to 20 A) or
2
) - ( 1-PF2) × 100%
(CF 3 and signal <30% F.S.) or,
(CF 6 and signal <60% F.S.), and
≤ 200 Hz with frequency filter on
Ф× (influence when PF=o)%}]
5
www.bkprecision.com
Power Meter
5335B
Specifications (cont.)
Harmonic Measurement Parameters
Measurement methodPLL synchronization
Frequency rangePLL frequency source range 10 Hz to 1.2 kHz (typical)
FFT data length1024
Window functionRectangle
Fundamental frequency (Fund. freq.)10 Hz to 75 Hz75 Hz to 150 Hz150 Hz to 300 Hz300 Hz to 600 Hz600 Hz to 1200 Hz
Sample rate
(Fund. freq.) x
1024(Fund. freq.) x 512
Window width124816
Upper limit of analysis orders50321684
Harmonic Measurement Accuracy (when line filter is off)
Frequency10 Hz ≤ f < 45 Hz45 Hz ≤ f ≤ 440 Hz440 Hz < f ≤ 1 kHz1 kHz < f ≤ 2.5 kHz2.5 kHz < f ≤ 5 kHz
Voltage and current±0.15% ± 0.35% F.S.±0.15% ± 0.35% F.S.±0.20% ± 0.35% F.S.±0.80% ± 0.45% F.S.3.05% ± 0.45% F.S.
I: 2.5, 5, 10, 25, 50, 100, 250, 500 mA/div, 1 A, 2.5 A, 5 A, 10 A/div,
I: 5, 10, 20, 50, 100, 200, 500 mA/div, 1 A, 2 A, 5 A, 10 A, 20 A/div,
Vertical scale ranges
(Accuracy ±4.0%)
CF 3
CF 6
Maximum input voltage (DC+AC peak)1800 V
Maximum input current (DC+AC peak)60 A
Environmental and Safety
Temperature
Operating: 41 ºF to 104 ºF (5 °C to 40 °C)
Storage: -4 ºF to 122 ºF (-20 °C to 50 °C)
Humidity20% RH to 80% RH (non-condensing)
Electromagnetic compatibilityIEC 61326
SafetyIEC 61010-1, EN 61010-1, Measurement 600 V CAT II
General
Display4.3” TFT-LCD display, 480 x 272
Remote InterfacesUSB, GPIB, RS232, LAN
Power100 to 240 VAC, 50 / 60 Hz
Power Consumption50 VA max.
Dimensions (W x H x D)8.4” x 3.5” x 14” (214.5 mm × 88.2 mm × 354.6 mm)
Weight6.2 lbs (2.8 kg)
Standard AccessoriesGetting started manual, instruction manual (downloadable), AC power cord, USB type A-to-type B cable, certificate of calibration
(Fund. freq.) x 256
U: 7.5, 15, 30, 75, 150, 300 V/div
U: 15, 30, 60, 150, 300, 600 V/div
(Fund. freq.) x 128(Fund. freq.) x 64
Three-Year Warranty
6
v092117
www.bkprecision.com
Chapter 11
Routine Maintenance
This chapter describes general maintenance items and maintenance methods of the 5335B power meter.
11.1Self-inspection
The 5335B power meter has a self-inspection function. See “Power-on Self-Test” of Chapter 3 “Inspecting the
Instrument” of 5335B Installation Instructions for detailed steps of self-inspection.
11.2Error Information References
The 5335B power meter has a detailed error and prompt information function, so as to help the user to easily
carry out positioning and measurement during measurement and use. This section describes all error information
of the 5335B power meter as well as error causes and disposals. All prompt information is listed.
11.2.1Prompt Message List
11.2.2Error information list
Error informationError information explanation
Error description No USB peripheral is found.
usb is not detected
Save screen fail
Possible cause No USB peripheral is inserted.
Disposal Insert the U disc type USB peripheral and then copy the screen.
Error description The screen is not saved successfully.
Possible cause USB is disconnected.
Disposal Reinsert the USB peripheral.
50
CHAPTER 11. ROUTINE MAINTENANCE51
Error informationError information explanation
Start time is less than current
End time is less than current
Timer must be larger than
zero.
Harmonic open fail
Error description The start time is less than the current time.
Possible cause The integral setting is incorrect.
Disposal Reset the integral start time.
Error description The ending time is less than the current time.
Possible cause The integral setting is incorrect.
Disposal Reset the integral ending time.
Error description The set time must be larger than zero.
Possible cause The set integral time is not correct.
Disposal Reset the set integral time.
Error description The harmonic function cannot be enabled.
Possible cause Communication abnormality
Integ open fail
Integ start fail
Integ stop fail
Disposal Check the communication cable.
Error description The integral function cannot be enabled.
Possible cause Communication abnormality
Disposal Check the communication cable.
Error description The integral function cannot be started.
Possible cause Communication abnormality
Disposal Check the communication cable.
Error description The integral function cannot be stopped.
Possible cause Communication abnormality
Disposal Check the communication cable.
CHAPTER 11. ROUTINE MAINTENANCE52
Error informationError information explanation
Error description The oscilloscope function cannot be enabled.
Scope openfail
Time set fail
Cal zero fail
Possible cause Communication abnormality
Disposal Check the communication cable.
Error description Time setting fails.
Possible cause Time setting is illegal.
Disposal Reset the system time.
Error description Zero calibration fails.
Possible cause Communication abnormality
Disposal Check the communication cable.
Table 11.1: Errors
11.3Daily maintenance
Introduce basic maintenance in daily use of equipment. Such as cleaning, self-maintenance allowed to be
performed by the user, etc.
11.3.1Equipment cleaning
Use dry cloth or slightly wet cloth to gently wipe the equipment. Do not arbitrarily wipe the inside of the
instrument. Cut o the power supply before cleaning.
11.3.2Initialize
When this operation is implemented, the system settings are recovered to the factory default values.
1. Select “SYSTEM > INITIAL” in the “Menu” interface to enter the system initialization interface. See the
following gure.
2. Select the menu items to be initialized (blue symbol background) via button. Menu items with characters
## following should be initialized; ## indicates that the items will not be initialized.
3. Press the soft key corresponding to the symbol on the right of the interface to determine whether to
initialize this menu item. Or use the soft key “ ” (for initialization of all menu items) or “ ” (no
initialization for all menu items) in the lower part of the interface.
4. Press the “START” soft key to initialize system settings. Press the ESC to exit.
CHAPTER 11. ROUTINE MAINTENANCE53
Prompt informationExplanation of prompt information
Cal Zero is working!The instrument is calibrating the zero point.
Cal Zero is completed!The instrument has completed zero point calibration.
Test screenSelf-inspection of LCD screen
Test dspSelf-inspection of DSP.
Test beepSelf-inspection of the buzzer.
SelfTest is nished!The instrument has nished the self-test.
Softkey is not availableThe current soft key is not available.
Time set ok!The time setting is completed.
Meter InitializationInitialization of general measurement
Integ InitializationInitialization of integral function
Scope InitializationInitialization of the oscilloscope function
Harmonic InitializationInitialization of the harmonic function
Inrush InitializationInitialization of the inrush function
Setup InitializationInitialization of setup
System InitializationInitialization of the system
Initialization is nished!Initialization is nished.
Integ startIntegration is started.
Integ stopIntegration is stopped.
Integ time upThe integral time is up.
Maxpagehasbeen
reached.
One page should exist at
least
Not used in current stateThe current operation is not available.
Please reset rstPlease reset rst.
It displays The maximum page has been reached.
One page should exist at least.
11.4Troubleshooting
This section describes operations to be carried out by the user in case of failure of the instrument. Preparation
before contact
When the instrument fails, check the following:
• Check whether the power meter is powered up.
• Check whether the power meter is started normally.
• Check whether the instrument fuse is in good conditions.
• Check whether other connectors are in good conditions and whether cables, plugs and other parts are
connected properly.
• Check whether the system conguration of the instrument is correct during use.
• Check whether self-inspection of the instrument is successful and whether the specications and
performance are within the indicator ranges.
CHAPTER 11. ROUTINE MAINTENANCE54
• Check whether the instrument displays error information.
• Use other instruments instead of this instrument for conrmation.
Chapter 12
LIMITED THREE-YEAR WARRANTY
B&K Precision Corp. warrants to the original purchaser that its products and the component parts thereof, will
be free from defects in workmanship and materials for a period of three years from date of purchase.
B&K Precision Corp. will, without charge, repair or replace, at its option, defective product or component parts.
Returned product must be accompanied by proof of the purchase date in the form of a sales receipt.
To help us better serve you, please complete the warranty registration for your new instrument via our website
www.bkprecision.com
Exclusions: This warranty does not apply in the event of misuse or abuse of the product or as a result
of unauthorized alterations or repairs. The warranty is void if the serial number is altered, defaced or
removed.
B&K Precision Corp. shall not be liable for any consequential damages, including without limitation damages
resulting from loss of use. Some states do not allow limitations of incidental or consequential damages. So the
above limitation or exclusion may not apply to you.
This warranty gives you specic rights and you may have other rights, which vary from state-to-state.
B&K Precision Corp.
22820 Savi Ranch Parkway
Yorba Linda, CA 92887
www.bkprecision.com
714-921-9095
55
Chapter 13
Service Information
Warranty Service: Please go to the support and service section on our website at bkprecision.com to obtain an
RMA #. Return the product in the original packaging with proof of purchase to the address below. Clearly
state on the RMA the performance problem and return any leads, probes, connectors and accessories that you
are using with the device.
Non-Warranty Service: Please go to the support and service section on our website at bkprecision.com to
obtain an RMA #. Return the product in the original packaging to the address below. Clearly state on the
RMA the performance problem and return any leads, probes, connectors and accessories that you are using with
the device. Customers not on an open account must include payment in the form of a money order or credit
card. For the most current repair charges please refer to the service and support section on our website.
Return all merchandise to B&K Precision Corp. with prepaid shipping. The at-rate repair charge for
Non-Warranty Service does not include return shipping. Return shipping to locations in North America is
included for Warranty Service. For overnight shipments and non-North American shipping fees please contact
B&K Precision Corp.
B&K Precision Corp.
22820 Savi Ranch Parkway
Yorba Linda, CA 92887
bkprecision.com
714-921-9095
Include with the returned instrument your complete return shipping address, contact name, phone number and
description of problem.
Version – May 25, 2018
56
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.