The following safety precautions apply to both operating and maintenance personnel and must be followed during all
phases of operation, service, and repair of this instrument.
Before applying power to this instrument:
•
Read and understand the safety and operational information in this manual.
•
Apply all the listed safety precautions.
•
Verify that the voltage selector at the line power cord input is set to the correct line voltage. Operating the instrument
at an incorrect line voltage will void the warranty.
•
Make all connections to the instrument before applying power.
•
Do not operate the instrument in ways not specified by this manual or by B&K Precision.
Failure to comply with these precautions or with warnings elsewhere in this manual violates the safety standards of design,
manufacture, and intended use of the instrument. B&K Precision assumes no liability for a customer’s failure to comply
with these requirements.
Category rating
The IEC 61010 standard defines safety category ratings that specify the amount of electrical energy available and the
voltage impulses that may occur on electrical conductors associated with these category ratings. The category rating is
a Roman numeral of I, II, III, or IV. This rating is also accompanied by a maximum voltage of the circuit to be tested,
which defines the voltage impulses expected and required insulation clearances. These categories are:
Category I (CAT I):
Measurement instruments whose measurement inputs are not intended to be connected to the
mains supply. The voltages in the environment are typically derived from a limited-energy transformer or a battery.
Category II (CAT II):
Measurement instruments whose measurement inputs are meant to be connected to the mains
supply at a standard wall outlet or similar sources. Example measurement environments are portable
tools and household appliances.
Category III (CAT III):
Measurement instruments whose measurement inputs are meant to be connected to the mains
installation of a building. Examples are measurements inside a building’s circuit breaker panel
or the wiring of permanently-installed motors.
Category IV (CAT IV):
Measurement instruments whose measurement inputs are meant to be connected to the primary
power entering a building or other outdoor wiring.
Do not use this instrument in an electrical environment with a higher category rating than what is specified in this manual
for this instrument.
You must ensure that each accessory you use with this instrument has a category rating equal to or higher than the
instrument’s category rating to maintain the instrument’s category rating. Failure to do so will lower the category rating
of the measuring system.
This instrument is intended to be powered from a CATEGORY II mains power environment. The mains power should be
115 V RMS or 230 V RMS. Use only the power cord supplied with the instrument and ensure it is appropriate for your
country of use.
Ground the Instrument
To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical safety ground. This
instrument is grounded through the ground conductor of the supplied, three-conductor AC line power cable. The power
cable must be plugged into an approved three-conductor electrical outlet. The power jack and mating plug of the power
cable meet IEC safety standards.
Do not alter or defeat the ground connection. Without the safety ground connection, all accessible conductive parts
(including control knobs) may provide an electric shock. Failure to use a properly-grounded approved outlet and the
recommended three-conductor AC line power cable may result in injury or death.
Unless otherwise stated, a ground connection on the instrument’s front or rear panel is for a reference of potential only
and is not to be used as a safety ground. Do not operate in an explosive or flammable atmosphere.
Do not operate the instrument in the presence of flammable gases or vapors, fumes, or finely-divided particulates.
The instrument is designed to be used in office-type indoor environments. Do not operate the instrument
•
In the presence of noxious, corrosive, or flammable fumes, gases, vapors, chemicals, or finely-divided particulates.
•
In relative humidity conditions outside the instrument’s specifications.
•
In environments where there is a danger of any liquid being spilled on the instrument or where any liquid can condense
on the instrument.
•
In air temperatures exceeding the specified operating temperatures.
•
In atmospheric pressures outside the specified altitude limits or where the surrounding gas is not air.
•
In environments with restricted cooling air flow, even if the air temperatures are within specifications.
•
In direct sunlight.
This instrument is intended to be used in an indoor pollution degree 2 environment. The operating temperature range is
0∘C to 40∘C and 20% to 80% relative humidity, with no condensation allowed. Measurements made by this instrument
may be outside specifications if the instrument is used in non-office-type environments. Such environments may include
rapid temperature or humidity changes, sunlight, vibration and/or mechanical shocks, acoustic noise, electrical noise,
strong electric fields, or strong magnetic fields.
If the instrument is damaged, appears to be damaged, or if any liquid, chemical, or other material gets on or inside the
instrument, remove the instrument’s power cord, remove the instrument from service, label it as not to be operated,
and return the instrument to B&K Precision for repair. Notify B&K Precision of the nature of any contamination of the
instrument.
Clean the instrument only as instructed
Do not clean the instrument, its switches, or its terminals with contact cleaners, abrasives, lubricants, solvents, acids/bases,
or other such chemicals. Clean the instrument only with a clean dry lint-free cloth or as instructed in this manual. Not
for critical applications
This instrument is not authorized for use in contact with the human body or for use as a component in a life-support
device or system.
Do not touch live circuits
Instrument covers must not be removed by operating personnel. Component replacement and internal adjustments must
be made by qualified service-trained maintenance personnel who are aware of the hazards involved when the instrument’s
covers and shields are removed. Under certain conditions, even with the power cord removed, dangerous voltages may
exist when the covers are removed. To avoid injuries, always disconnect the power cord from the instrument, disconnect
all other connections (for example, test leads, computer interface cables, etc.), discharge all circuits, and verify there
are no hazardous voltages present on any conductors by measurements with a properly-operating voltage-sensing device
before touching any internal parts. Verify the voltage-sensing device is working properly before and after making the
measurements by testing with known-operating voltage sources and test for both DC and AC voltages. Do not attempt
any service or adjustment unless another person capable of rendering first aid and resuscitation is present.
Do not insert any object into an instrument’s ventilation openings or other openings.
Hazardous voltages may be present in unexpected locations in circuitry being tested when a fault condition in the circuit
exists.
Fuse replacement must be done by qualified service-trained maintenance personnel who are aware of the instrument’s fuse
requirements and safe replacement procedures. Disconnect the instrument from the power line before replacing fuses.
Replace fuses only with new fuses of the fuse types, voltage ratings, and current ratings specified in this manual or on
the back of the instrument. Failure to do so may damage the instrument, lead to a safety hazard, or cause a fire. Failure
to use the specified fuses will void the warranty.
Do not substitute parts that are not approved by B&K Precision or modify this instrument. Return the instrument to
B&K Precision for service and repair to ensure that safety and performance features are maintained.
For continued safe use of the instrument
•
Do not place heavy objects on the instrument.
•
Do not obstruct cooling air flow to the instrument.
•
Do not place a hot soldering iron on the instrument.
•
Do not pull the instrument with the power cord, connected probe, or connected test lead.
•
Do not move the instrument when a probe is connected to a circuit being tested.
Working Environment
Environment
This instrument is intended for indoor use and should be operated in a clean, dry environment.
Temperature
Operating: 0℃ to +40℃
Non-operation:-20℃ to +60℃
Direct sunlight, radiators, and other heat sources should be taken into account when assessing the ambient temperature.
Operating: less than 3 Km
Non-operation: less than 15 Km
Installation (overvoltage) Category
This product is powered by mains conforming to installation (overvoltage) category II.
Degree of Pollution
The oscilloscopes may be operated in environments of Pollution Degree II.
Degree of Pollution II refers to a working environment which is dry and non-conductive pollution occurs. Occasional
temporary conductivity caused by condensation is expected.
Disposal of Old Electrical & Electronic Equipment (Applicable in the European Union and other European
countries with separate collection systems)
This product is subject to Directive 2002/96/EC of the European Parliament
and the Council of the European Union on waste electrical and electronic equipment
(WEEE), and in jurisdictions adopting that Directive, is marked as being put on the
market after August 13, 2005, and should not be disposed of as unsorted municipal
waste. Please utilize your local WEEE collection facilities in the disposition of this
product and otherwise observe all applicable requirements.
Safety Symbols
Symbol
Description
indicates a hazardous situation which, if not avoided, will result in death or serious injury.
indicates a hazardous situation which, if not avoided, could result in death or serious injury
indicates a hazardous situation which, if not avoided, will result in minor or moderate injury
Refer to the text near the symbol.
Electric Shock hazard
Alternating current (AC)
Chassis ground
Earth ground
This is the In position of the power switch when instrument is ON.
This is the Out position of the power switch when instrument is OFF.
is used to address practices not related to physical injury.
The 2560B Digital Storage (DSO) and Mixed Signal Oscilloscope (MSO) Series delivers advanced features and debug
capabilities for a wide range of applications. With increasing bandwidths to 350 MHz in a 4-channel configuration, each
model offers a maximum sample rate of 2 GSa/s and a maximum memory depth of 200 Mpts. Equipped with a 10.1”
(1024 x 600) capacitive touchscreen and high waveform update rate of 120,000 wfms/s, these oscilloscopes can capture
infrequent glitches with excellent signal fidelity.
Before connecting and powering up the instrument, review the instructions in this section.
2.1 Input Power Requirements
The oscilloscope has a universal AC input that accepts line voltage and frequency input within:
Line
Voltage Range
100-120 V
100-240 V
Frequency
400 Hz
50/60 Hz
Power
80W Max
Table 2.1
Before connecting to an AC outlet or external power source, be sure that the power switch is in the OFF position and
verify that the AC power cord, including the extension line, is compatible with the rated voltage/current and that there
is sufficient circuit capacity for the power supply. Once verified, connect the cable firmly.
The included AC power cord is safety certified for this instrument operating in rated range. To
change a cable or add an extension cable, be sure that it can meet the required power
ratings for this instrument. Any misuse with wrong or unsafe cables will void the warranty.
The power cord provides a chassis ground through a third conductor. Verify that your power
outlet is of the three-conductor type with the correct pin connected to earth ground.
Complete the following steps to verify that the oscilloscope is ready for use.
2.3.1 Verify AC Input Voltage
Verify proper AC voltages are available to power the instrument.
The AC voltage range must meet the acceptable specification stated in section
Input Power Requirements.
2.3.2 Connect Power
Connect the AC power cord to the AC receptacle in the rear panel and press the power switch to turn on the instrument.
The instrument will have a boot up screen while loading, after which the main screen will be displayed.
2.3.3 Self-Test
The instrument has 3 self-test option to test the screen ,keyboard, and the LED back light.
To perform the self-test, please refer to the
Self Test
section for further instructions.
2.3.4 Self-Cal
Self option runs an internal self-calibration procedure that will check and adjust the instrument. To perform the selfcalibration, refer to the Self-Calibration section for further instructions.
2.3.5 Check Model and Firmware Version
The model and firmware version can be verified from within
the menu system.
Probes are available with various attenuation factors which affect the vertical scale of the signal. The Probe Check
function verifies that the probe attenuation option matches the attenuation of the probe.
Press CH 1 once to open the channel menu. Select the probe option that matches the attenuation of the probe.
The default setting for the Probe option is 1 X.
Verify that the attenuation switch on the probe matches the Probe option in the oscilloscope. Switch settings are 1 X
and 10 X.
Probe Compensation
Before taking any measurements using a probe, verify the compensation of the probe and adjust it to match the channel
inputs.
To match your probe to the input channel:
1.
Set the channel’s probe attenuation to 10X.
–
Press the
CH #
key corresponding to the channel the probe is connected to.
–
Use the softkeys to navigate to page 1.
–
Use the softkeys to select
Probe.
–
Use the
Intensity Adjust
knob to select 10X.
2.
Attach the probe tip to the
Compensation Signal Output Terminal 3 V(Cal
) connector and the reference lead to
the Probe Ground terminal connector.
–
Press the
Auto Setup
key to display the square wave.
3.
Check the shape of the displayed waveform.
Undercompensated Correctly Compensated
Overcompensated
Figure 2.5
Probe Compensation
4.
If necessary, adjust your probe’s compensation trimmer pot.
The control panel is designed to operate the basic functions
without having to open the software menu. Most of the front
panel controls duplicate functionality available through the
touch screen.
All the knobs on the front panel are multifunctional.
They can be pushed as well as rotated. Pushing a
knob quickly recalls a specific function, which is
indicated by the silkscreen near to the knob.
The
Control Panel
consist of 9 sections:
•
Vertical Control
•
Horizontal Control
•
Trigger Control
•
Run/Stop Key
•
Auto Setup Key
•
Analysis Keys
•
Universal Knob
•
Menu Keys
•
Other Keys
Figure 3.1 Control Panel
3.1 Vertical Control
Channel Keys
: Press the channel keys to toggle a channel
On
or
Off
, or to access the
channel’s menu. There is one channel on/off key for each analog channel.
Digital Key
: Enable/disable the digitals channels and display the waveforms.
See section Digital Channels for more details.
Math Key
: The
Math
key provides access to the math (add, subtract, etc.) waveform
func
tions.
See section
Math
for more details.
Ref Key
: Toggle the reference function on and off.
Vertical Scale Knob
: Analog channels (C1-C4), digital channels (d), math (F1-F2) and
references (Ref) share the same vertical scale knob. Turn the knob to adjust the gain
(volts/div). Push to alternate between coarse and fine adjustments. When the digital
channel is active, rotate the knob to change the selected digital channel.
Vertical Position Knob
: Analog channels (C1-C4), digital channels (D), math (F1-F2)
and references (Ref) share the same vertical position knob. Turn the knob to adjust the
DC offset or vertical position of the channel. Push to set the position to zero.
: Turn the knob to adjust the horizontal scale (time/div). The symbols
above the knob indicate that this control has the effect of spreading out or zooming in on the
waveform using the horizontal scale.
Push the horizontal scale knob to toggle between fine and coarse adjustment.
Zoom Key
: Press the
Zoom
key to split the oscilloscope display into Normal and Zoom sections.
See section Zoom for more details.
Roll Key
: Toggle the roll function on and off. At timebase settings larger than 50 ms/div, it is
recommended to set the oscilloscope to Roll mode so that the waveform is displayed in real time.
See section Roll Mode for more details.
Horizontal Position Knob
: Turn the knob marked to pan through the waveform data
horizontally. The captured waveform before the trigger (turn the knob clockwise) or after the
trigger (turn the knob counterclockwise) can be seen. Panning through the waveform when the
oscilloscope is stopped (not in Run mode) will display the waveform data from the last acquisition
taken.
Push the horizontal position knob to zero the horizontal position.
3.3 Trigger Control
Figure 3.4 Trigger Control
The Trigger controls determine how the oscilloscope triggers to capture data. These controls consist of:
Setup
: Opens the
Trigger Setup
menu. In the
Trigger Setup
menu, the trigger type, source, and options that
affect
all trigger types can be set. See section Triggers for more details.
Normal
: Sets the
Trigger Mode
to
Normal
. In
Normal Mode
acquisitions only occur when the trigger conditions are
met. Otherwise, the oscilloscope holds the last waveform on the display and waits for the next trigger.
Auto
: Sets the
Trigger Mode
to
Auto
. In
Auto Mode
an internal timer triggers the sweep after a preset timeout
period if no trigger has been found so that the oscilloscope continuously updates the display whether a trigger occurs or
not.
is used to select items from pop-up menus and to change values. The
function of the
Universal Knob
changes based upon the current menu.
Note that the curved arrow symbol below the knob illuminates whenever the
Universal
Knob
can be used to select a value. Also, note that when the
Universal Knob symbol ap-
pears, you can use the Universal Knob, to select values.
Rotating the
Universal Knob
adjusts the trace intensity. Pushing
the Universal Knob
instantly sets the intensity to 50%.
Acquire
: Access the
Acquire
menu. In the acquire menu Interpolation,
Acquire Mode, Acquisition, Memory Depth, Sequence, and Resolution can be set.
See section Acquire for more details.
Display
: Access the
Display
menu to enable persistence, adjust the display grid (graticule)
intensity, label waveforms, add an annotation, and clear the display.
See section Display for more details.
Save/Recall
: Save oscilloscope setups, screen images, waveform data, or mask files or to
recall setups, mask files or reference waveforms.
See section
Save/Recall
for more details.
Utility
: Access the
Utility
menu, to configure the oscilloscope’s I/O settings, use the file
explorer, set preferences, access the service menu, and choose other options.
AWG
: Built-in waveform generator can output arbitrary, sine, square, ramp, pulse,
DC, noise, and other 45 built-in waveforms. Modulated waveforms are available
except for arbitrary, pulse, DC, and noise waveforms.
The vertical resolution. "8-bits" is the default setting.
–
In
10-bits
mode, the vertical resolution is 4x better, while the bandwidth
is limited to about 100 MHz.
•
Touch
Interpolation
to toggle between x and
Sinc.
•
At small timebase settings, the number of original points on the screen
may be less than the number display pixels in the grid area, so
interpolation is necessary to display a continuous waveform. For
example, at 1 ns/div timebase and 2 GSa/s sample rate, the number of
original points is 20, but the grid area includes 1000 horizontal pixels. In
this case, the
oscilloscope needs to interpolate the original points by
50.
–
X
: Linear interpolation, the simplest way of interpolation, connects
two original points with a straight line.
–
Sinc
: Sin(x)/x interpolation, the original point is interpolated
according to the Nyquist reconstruction formula, which has a good
time- domain recovery effect for sine wave. But for step signals/fast
rise
times, it will introduce false overshoot due to the Gibbs phenomenon.
Figure 4.1 Acquire Menu
•
Touch
Acq Mode
to toggle between
Fast
and
Slow
acquisition.
–
Fast
is the default setting. A very high waveform update rate is provided in fast mode.
–
Slow
mode will slow down the waveform update.
•
Touch
Acquisition
to determine how to acquire and process the signal.
•
Normal
: The oscilloscope samples the signal with an equal time interval. For most waveforms, the best display
effect can be obtained using this mode.
•
Peak
: The oscilloscope acquires the maximum and minimum values of the signal within the sample interval so
that the peak (maximum – minimum) in the interval is obtained. This mode is effective to observe occasional
narrow pulses or spurs with a low sample rate, but the noise displayed is larger. In peak mode, the oscilloscope
will display all pulses with a pulse width longer than 400 ps.
•
Touch
Memory Depth
to set the maximum memory depth that is supported.
acquisition time = sample points x sample interval", setting a larger memory depth can achieve a higher sample
rate for a given time base, but more samples require more processing time, degrading the waveform update rate.
With 200 Mpts memory depth, the 2560B series can still run at full sample rate (2 GSa/s) even when set to the
10 ms/div timebase.
The memory depth here is the upper limit of the memory space allocated by the oscilloscope. The actual sample
points are related to the current timebase and may be less than memory depth. The actual sample points
information can be obtained in the timebase descriptor box (see the section "Timebase and Trigger" for details).
The maximum memory depth in single-channel mode is 2 times that of the dual-channel mode, as the following table:
Single-Channel Mode
Dual-Channel Mode
20k
10k
200k
100k 2M 1M
20M
10M
200M
100M
Table 4.1 Memory Depth
Single-channel mode (interleaving mode): Only one of C1/C2 is turned on, and only one of C3/C4 is turned on.
Dual-channel mode (non-interleaving mode): Both C1/C2 are turned on, or both C3/C4 are turned on.
Sequence mode is a fast acquisition mode, which divides the memory depth to multiple segments (up to 90,000), each
of which stores a single shot.
In sequence mode, the oscilloscope only acquires and stores data without processing and displaying, until the specified
segments are acquired. As a result, the dead time between trigger events is minimized, improving the waveform update
rate.
If sequence mode is enabled, the display will not update until all of the sequences have been acquired. The 2560B series
can achieve a minimum 2 s trigger interval in sequence mode, corresponding to a waveform update rate of 500,000
wfms/s.
Sequence mode can capture and record rare events over long time periods. The oscilloscope can capture multiple events
that satisfy the trigger conditions, ignoring the periods of no interest between adjacent events, maximizing the use of
waveform memory. Full accuracy of the acquisition timebase can be used to measure selected segments.
After the acquisition is finished, the oscilloscope will map all the segments together to the screen. To view and analyze
each frame separately, use history mode (see the section " History" for details). History mode provides timestamp labels
for each segment.
The history function can record the waveforms of the input channels before press the
Run/Stop
button. In run state,
the oscilloscope records input waveform continually; when the memory is full (reach the maximal frame), the new frames
will cover the old frames and keep the latest frames.
The oscilloscope automatically stores acquired frames. It can store up to 90,000 frames but the number may vary due to
the memory depth and timebase settings. Turn on history mode, then the stored frames can be recalled and measured.
In sequence mode, all waveforms that satisfy the trigger conditions are mapped to the display.
To view a single frame:
1.
Press the
History
button on the front panel to enable the History function.
–
The history menu shown in figure
5.1
will be displayed.
2.
Touch
History
to toggle the history function on and off.
3.
Touch
Frame No.
and use the
Universal Knob
or the virtual keypad to
specify the frame index.
4.
Touch
List
to toggle the list function on and off.
5.
Touch the navigation options on to play backward, stop, or play
forward.
–
Touch
List Time Type
to toggle the type between
Acq Time
and
Delta T.
6.
Touch
Interval Time
to set the playing interval by using the
Universal
Knob or the virtual keypad.
•
In addition to manually specifying a frame, history mode supports autoplay.
–
Touch the option to replay the waveform from the current frame
to the first.
–
Touch the option to stop replay.
–
Touch the option to replay the waveform from the current frame
to the last.
Changing the delay time moves the trigger point (solid inverted triangle
) horizontally and indicates how far it is
from the time reference point (hollow inverted triangle ). These reference points are indicated along the top of the
display grid.
Figure
6.3
shows the trigger point with the delay time set to 170 µs. The delay time number tells you how far the time
reference point is located from the trigger point. When delay time is set to zero, the delay time indicator overlays the
time reference indicator.
All events displayed left of the trigger point happened before the trigger occurred. These events are called pre-trigger
information, and they show events that led up to the trigger point.
Everything to the right of the trigger point is called post-trigger information. The amount of delay range (pre-trigger
and post-trigger information) available depends on the time/div selected and memory depth.
The horizontal position knob works (in the Normal time mode) while acquisitions are running or when they are stopped.
When running, adjusting the horizontal scale knob changes the sample rate. When stopped, adjusting the horizontal
scale knob lets you zoom into acquired data.
The
Horizontal Position Knob
has a different purpose in the
Zoom display.
See section
Zoom
for more details
6.3 Panning and Zooming Single or Stopped Acquisitions
When the oscilloscope is stopped, use the touchscreen horizontal pinch or drag gestures or use the horizontal scale
and position knobs to pan and zoom your waveform. The stopped display may contain several acquisitions worth of
information, but only the last acquisition is available for pan and zoom.
The ability to pan (move horizontally) and scale (expand or compress horizontally) an acquired waveform is important
because of the additional insight it can reveal about the captured waveform. This additional insight is often gained from
seeing the waveform at different levels of abstraction. You may want to view both the big picture and the specific little
picture details. For example, figure
6.3
demonstrates how these features bring insight on the waveform’s rise time.
6.4 Horizontal Scale (Coarse/Fine)
To change the horizontal scale knob adjustment setting:
•
Push the
Horizontal Scale Knob
to toggle between fine and coarse adjustment.
When Fine is enabled, turning the horizontal scale knob changes the time/div (displayed in the status line at the top of
the display) in smaller increments. The time/div remains fully calibrated when Fine is on.
When Fine is turned off, the Horizontal scale knob changes the time/div setting in a 1-2-5 step sequence.
Zoom is a horizontally expanded version of the normal display. When Zoom is selected, the display divides into two
windows. The top window covers about a third of the displays. This window displays the normal time/div window. The
area of the normal display that is expanded is outlined with a box and the rest of the normal display is ghosted. The box
shows the portion of the normal sweep that is expanded in the lower half.
The bottom window covers two-thirds of the screen. This window is a magnified portion of the normal time/div window.
Use Zoom to locate and horizontally expand part of the normal window for a more detailed analysis of the waveform.
To enable the Zoom function:
•
Press the
Zoom
key or push the
Horizontal Scale Knob.
–
Zoom can also be enabled using the touch screen. Press the Acquire option located on the Menu Bar. Then
press Zoom.
Figure 6.6 Zoom
To change the time/div for the Zoom window, turn the
Horizontal Scale Knob
. As you turn the knob, the zoomed
window time/div is highlighted in the status line above the waveform display area. The Horizontal scale knob controls
the size of the box.
The Horizontal position knob sets the left-to-right position of the zoom window. Negative delay values indicate you’relooking at a portion of the waveform before the trigger event, and positive values indicate you’re looking at the waveform
after the trigger event.
To change the time/div of the normal window, turn off Zoom by pressing the
Roll causes the waveform to move slowly across the screen from right to left. It only operates on the time bases settings
of 50 ms/div and slower. If the current time base settings are faster than the 50 ms/div limit, it will be set to 50 ms/div
when Roll mode is entered.
To enable the Roll mode:
•
Press the
Roll
key.
–
Roll can also be enabled using the touch screen. Press the
Acquire
option located on the Menu Bar.
Then press
Roll.
In Roll mode there is no trigger. The fixed reference point on the screen is the right edge of the screen and refers to
the current moment in time. Events that have occurred are scrolled to the left of the reference point. Since there is no
trigger, no pre-trigger information is available.
If you would like to pause the display in Roll mode press the
Single
key. To clear the display and restart an acquisition
in Roll mode, press the [Single] key again.
Use Roll mode on low-frequency waveforms to yield a display much like a strip chart recorder. It allows the waveform to
roll across the display.
The vertical scale and position knobs for each analog channel.
•
The channel keys for turning a channel on or off.
Figures
7.2
shows the Channel 1 Menu that appears after pressing the 1 channel key.
7.1 Enabling Channels
Push the channel button (1 - 4) to turn on the corresponding channel. Its channel
descriptor box and Channel menu will appear on the display. Push the same button
again
to disable the channel. The channel menu also has an on/off option.
7.1.1 Channel Enable/Disable From the Touch Screen
Touch the + button shown on figure
7.1
and then select a channel to turn it on. The
selected channel’s descriptor box and a menu box will appear on the display. Touch the
channel descriptor box and then touch the Off button to disable it.
step sequence (with a 1:1 probe attached), unless Fine Adjustment is enabled.
The analog channel’s vertical scale (Volts/div) value is displayed in the
channel’s
descriptor box.
To adjust the vertical scale:
•
Use the touchscreen vertical pinch gesture.
•
Touch the channel descriptor box to open the vertical scale/offset dialog.
•
Turn the
Vertical Scale Knob.
7.2.1 Coarse/Fine Adjustment
Push the channel’s vertical scale knob (or press the channel’s descriptor box then
uncheck Coarse) to toggle between fine and coarse adjustment of the vertical scale.
When
Fine
adjustment is selected, you can change the channel’s vertical sensitivity in
smaller increments. The channel sensitivity remains fully calibrated when
Fine
is on.
When Fine is turned off, turning the
Vertical Scale Knob
changes the channel
sensi
tivity in a 1-2-5 step sequence.
7.3 Vertical Position
The offset voltage value represents the voltage difference between the vertical center
of the display and the ground level (
) icon.
To adjust the vertical position:
•
Use the touchscreen pinch and drag gestures.
•
Touch the channel descriptor box to open the vertical scale/offset dialog.
key to open the corresponding channel menu. As shown in figures
7.2
and
7.3
the following
options can be configured:
•
Channel Coupling
•
BW Limit
•
Probe
•
Label
•
Apply To
•
Impedance
•
Unit
•
Deskew
•
Invert
•
Trace
7.4.1 Channel Coupling
Coupling changes, the channel’s input coupling to either
AC
(alternating current),
DC
(direct current), or
GND
(ground).
DC (Direct Current)
All of the input signal frequency components are passed to the display.
DC coupling is useful for viewing waveforms as low as 0 Hz that do not have large DC offsets.
If the channel is DC coupled, you can quickly measure the DC
component of the signal by simply noting its distance from the ground.
Alternating Current
The signal is capacitively coupled. DC signal components are rejected. See the datasheet for details of the cut-off
frequency.AC coupling is suitable for observing AC signals with DC offset, such as power ripple.
AC coupling is useful for viewing waveforms with large DC offsets.
AC coupling places a 10 Hz high-pass filter in series with the input waveform that removes any DC offset voltage from
the waveform.
If the channel is AC coupled, the DC component of the signal is removed,
allowing for greater sensitivity to display the AC component of the signal.
Ground
The channel is grounded by an internal switch. GND coupling is used to observe the zero offset error of the analog
channels or determine the source of noise in the waveform (from signal or from oscilloscope itself).
A drop-down menu will appear with all 3 coupling option.
–
Use the touchscreen to select the desired coupling.
Channel Coupling
is independent of
Trigger Coupling
. To change trigger coupling see
Trigger Coupling.
7.4.2 Bandwidth Limit
Full bandwidth will pass through signals with high-frequency components, but it also means that noise with high-frequency
components can pass through. When the frequency component of a signal is very low, better signal-to-noise ratios (SNR)
can be obtained by enabling one of the available bandwidth limits.
The 2560 provides two bandwidth limit options: 20 MHz and 200 MHz. The bandwidth limit effectively lowers the input
frequency response of the input to the selected limit value.
For waveforms with frequencies below the bandwidth limit, turning the bandwidth limit on removes unwanted high
frequency noise from the waveform.
To set the bandwidth limit:
1.
Press the desired channel key.
2.
In the
Channel Menu
, use the touchscreen to press the
BW Limit
option.
–
A drop-down menu will appear with three bandwidth options.
–
Use the touchscreen to select the desired bandwidth limit.
7.4.3 Probe Attenuation
This is set automatically if the oscilloscope can identify the connected probe. The 2560 provides 1X, 10X, 100X
and custom probe attenuation factor options. The custom values can be between 10−6𝑡𝑜 106. The oscilloscope will
automatically convert the vertical scale according to the current probe attenuation factor.
For example, the vertical scale of the oscilloscope under 1X attenuation is 100 mV/div, and the vertical scale will be
automatically set to 1 V/div if the probe attenuation is changed to 10X. If a standard probe with readout terminal is
connected, the oscilloscope will automatically set the probe attenuation to match the probe.
To set the probe attenuation:
1.
Press the desired channel key.
2.
In the
Channel Menu
, use the touchscreen to press the
Probe
option.
–
The
Probe
submenu will appear with four bandwidth options.
–
Use the touchscreen to select the desired probe attenuation.
Customize the label text, of the selected source. The source can be C1 - C4, F1 or F2, and RefA - RefD.
The length of the label is limited to 20 characters. The characters beyond this length will not be displayed.
When the
Display
option is set to “on”, the label will be displayed on the right side of the channel offset indicator.
To set and enable a channel’s label:
1.
Press the desired channel key.
2.
In the
Channel Menu
, use the touchscreen to press the
Label
option.
–
The
Label
submenu will appear.
•
Select a source to assign the label to the specified source.
•
Enable/disable the label using the
Display
option
•
Set the text of the label in the
Label Text
option.
Figure 7.6
Label Submenu
Label On Label Off
Figure 7.7
Label
7.4.5 Apply To
Common functions such as Trigger, Cursor, Measure, FFT, Search, Mask Test, Counter, and AWG can be quickly applied
to the selected channel using the
Apply To
option. Once a function is specified, it will switch directly to the function
menu and automatically set that channel as the source.
To apply a function:
1.
Press the desired channel key.
2.
In the
Channel Menu
, use the touchscreen to press the
Apply To
option.
–
A drop-down menu will appear listing the available functions.
–
Use the touchscreen to select the desired function.
1 MΩ: When a passive probe with high impedance is connected, the impedance must be set to 1 MΩ, otherwise the
signal will not be detected.
50 Ω: Suitable for high-frequency signals transmitted through 50 Ω coaxial cables and can minimize the amplitude
distortion caused by impedance mismatching.
7.4.7 Unit
Channel sensitivity, trigger level, measurement results, and math functions will reflect the measurement units selected.
To change the units:
1.
Press the desired channel key.
2.
In the
Channel Menu
, use the touchscreen to select either V or A.
Use Volts for a voltage probe.
Use
Amps
for a current probe.
7.4.8 Deskew
When measuring time intervals in the nanoseconds (ns) range, small differences in cable length can affect the
measure
ment. Use Deskew to remove cable-delay errors between any two channels.
For example, two coaxial cables with a 1-inch difference in length could introduce a skew of more than 100 ps. In some
scenarios (e.g. measuring the setup/hold time between clock and data), it may be necessary to compensate the skew
between channels.
The method of compensation: Probe the same signal simultaneously using two channels (including the cables or probes
that you intend to use for measurements) and adjust the deskew parameter of one channel until the waveforms of the
two channels observed on the screen coincide horizontally.
This chapter describes how to use the digital channels of a Mixed-Signal Oscilloscope (MSO).
The digital channels are standard on the 2560B MSO models and the 2560B Series DSO models have the upgrade license
option.
8.1 LP2560 Probe
The LP2560 is a logic probe designed to monitor up to 16 digital signals at once. The 16 digital channels are separated
into two groups and each group has its threshold, making it possible to simultaneously view data from different logic
families.
The 16-channel color-coded logic probe consists of
two eight-channel pods. To make contact with the
DUT, the probe connects directly to square pins or
clips to test points using the included grabbers.
With an input capacitance of only 18 pF and 100 kΩ
input impedance, the probe protects the integrity of
your signal.
The probe is included with MSO models.
Figure 8.1 LP2560 Probe
To avoid personal injury or damage to the logic probe and any
associated equipment, the following safety precautions should be noted.
The equipment shall be used only for the purposes specified by the manufacturer.
The LP2560 probe is used only for BK’s special series of oscilloscopes. Protection mechanisms can be compromised if
the way the devices connected by the LP2560 are not used for their intended purpose.
Connect and disconnect correctly
.
Excessive bending can damage the cable.
Do not use equipment in humid or explosive environment.
Only used indoors. The LP2560 is designed to be used indoors and should only be operated in a clean, dry environment.
Do not use the equipment when you suspect a problem.
Do not use the LP2560 if any parts are damaged. Maintenance work shall be performed by maintenance personnel with
appropriate qualifications. Keep the product surface clean and dry.
If necessary , turn off the power supply to the device under test.
–
Turning off power to the device under test only prevents damage that might occur if two lines are accidentally
shorted when connecting the probes. The oscilloscope can be powered on because no voltage appears at the
probes.
2.
Connect the digital probe cable to the
D0-D15
connector located on the front panel.
3.
Connect a SMD Mini-grabber test clip to one of the digital ground pins of the probe and then connect the grabber
to ground in the device under test.
–
This will improve the signal fidelity and makes sure the oscilloscope gives accurate measurements.
4.
Connect a flying lead probe to one of the digital channel pins; connect a grabber to the flying lead probe, and then
connect the grabber to the node in the circuit to be tested.
5.
Repeat step 4 until all points of interest are connected.
Figure 8.2 Connecting Digital Probes
8.3 Acquiring Digital Waveforms
To acquire digital waveforms:
•
Press the
Digital
button to open the digital channels and start acquiring waveforms.
Press the
Run/Stop
or
Single
key to run the oscilloscope, the oscilloscope examines the input voltage at each input
probe. When the trigger conditions are met the oscilloscope triggers and displays the acquisition.
For digital channels, each time the oscilloscope takes a sample it compares the input voltage to the logic threshold. If
the voltage is above the threshold, the oscilloscope stores a 1 in sample memory; otherwise, it stores a 0.
Digital data can be stored as waveform files. Horizontal cursors and most of the horizontal measurements also apply to
digital waveforms.
The threshold level determines how the input signal is evaluated. The threshold level can be set in the
Logic Setting.
The threshold you set applies to all channels within the selected D7 - D0 or D15 - D8 group. Each of the two channel
groups can be set to a different threshold if desired.
Values greater than the set threshold are high (1) and values less than the set threshold are low (0).
To set a threshold level:
1.
Press the
Digital
key to open the
Digital
menu.
2.
Use the touchscreen controls to select the
Logic Setting (D7:D0)
or
Logic
Setting (D17:D8) option.
–
The window shown in figure
8.8
will appear.
3.
Use the touchscreen controls to select the desired threshold.
–
Refer to table
8.1
for more information on the threshold voltage.
4.
If the Thresholds is set to
Custom
, use the touchscreen to select
Threshold
Figure 8.8 Logic Setting
for the channel group, then use the virtual keypad to set the logic threshold.
Logic Family Threshold Voltage
TTL
1.5 V
CMOS
1.65 V
LVCMOS 3.3 V 1.65 V
LVCMOS 2.5 V 1.25 V
Custom
Variable from -10 V to +10 V
Table 8.1 Logic Thresholds
8.7 Displaying Digital Channels as a Bus
Digital channels may be grouped and displayed as a 16-bit bus, with each value displayed at the bottom of the display
in binary, decimal, unsigned dec, or hex.
To configure and display each bus:
For triggering, certain condition can be set according to the requirement and when a waveform in the waveform meets
this condition. Digital oscilloscope, display a waveform continuously regardless of the trigger stability, but only stable
trigger can ensure stable display.
The trigger circuit ensures that every time base sweep or acquisition starts from the input signal and the user-defined
trigger condition, namely every sweep is synchronous to the acquisition and the waveforms acquired overlap to display
stable waveform.
The trigger position is movable on the display. Figure
9.1
demonstrates how the position of the trigger event determines
the reference time point and the delay setting. The acquisition memory is divided into pre-trigger and post-trigger buffers
and the boundary between them is the trigger position.
Before the trigger event arrives, the oscilloscope fills the pre-trigger buffer first, and then continuously updates it in FIFO
mode until the trigger event arrives. After the trigger event, the data fills the post-trigger buffer. When the post-trigger
buffer is full, an acquisition is completed.
Figure 9.1 Acquisition Memory
A trigger setup tells the oscilloscope when to acquire and display data. Trigger setting are based on the features of the
input signal, therefore knowledge of the signal under test is required to quickly capture the desired waveform.
Changes to the trigger setup are applied immediately. If the oscilloscope is stopped when changes are made to the trigger
setup, the oscilloscope uses the new specification when the
Run/Stop
or
Single
key is pressed. If the oscilloscope is
running when changes are made, the oscilloscope uses the new trigger definition when it starts the next acquisition.
The 2560B Series oscilloscopes allow the use of either voltage or current units for waveform measurements.
The remainder of this chapter will refer to just voltages, but it applies to current levels, too.
The 2560B Series trigger source includes four analog channels sixteen Digital channels, Ext, Ext/5, and AC line.
The trigger source is the signal that will be compared to the logical conditions you set to generate a trigger event. The
most common trigger source is the signal on one of the analog input channels, but the EXT connector on the back panel
can be used to trigger on an external signal.
When looking at waveforms that are derived from the AC line power, you’ll probably want to use the AC line as the
trigger source. For example, to measure the 120 Hz ripple in a voltage regulator circuit, you’d want to trigger on the 60
Hz AC line for a stable signal.
The trigger sources supported by each trigger type are different. Reference table
??
for more details.
Table 9.1
Trigger Source
Once a trigger type has been selected set the trigger source:
1.
Press the Setup key or use the touchscreen controls to select
Trigger > Menu...
to enter the
Trigger
menu.
2.
Use the touchscreen controls to select
Source.
–
The
Source Setting
window will appear.
–
Use the touchscreen controls to navigate the available trigger sources.
The currently selected trigger source is displayed at the upper right corner of the
Trigger dialog box as shown in figure 9.2.
The 2560B Series provides the following trigger types :
•
Edge Trigger
•
Slope Trigger
•
Pulse Trigger
•
Video Trigger
•
Window Trigger
•
Interval Trigger
•
Dropout Trigger
•
Runt Trigger
•
Pattern Trigger
•
Serial Trigger
To select a trigger type:
Figure 9.3
Trigger Types
1.
Press the Setup key or use the touchscreen controls to select
Trigger > Menu...
to enter the
Trigger
menu.
2.
Use the touchscreen controls to select
Type.
–
The
Trigger Type
window shown in figure
9.3
will appear.
–
Use the touchscreen controls to select a trigger type.
9.2.1 Edge Trigger
Edge trigger distinguishes the trigger points by seeking the specified edge (rising, falling, alter) and trigger level. Trigger
source and slope can be set in the trigger menu.
Figure 9.4 Edge Trigger Point
To set the trigger source:
Use the touchscreen controls to select
Source
. Refer to table
9.1
for more information on the available trigger sources.
Holdoff, coupling, and noise reject can be set in edge trigger, see the sections
Holdoff, Trigger Coupling
and
Noise
Reject for more details.
9.2.2 Slope Trigger
The slope trigger looks for a rising or falling transition from one level to another level in greater than or less than a
certain amount of time.
In the oscilloscope, positive slope time is defined as the time difference between the two crossing points of trigger level
A and B with the positive edge as shown in the figure below.
Figure 9.6
Slope Trigger
Configuring a Slope Trigger
To select Slope Trigger type:
1.
Press the Setup key or use the touchscreen controls to enter the
softkey to select the Lower or Upper trigger level.
–
Turn the Trigger Level Knob to adjust the position.
–
The lower trigger level cannot be higher than the upper trigger level.
6.
Use the touchscreen to select the
Limit Range.
•
<= (less than a time value)
: Trigger when the positive or negative slope time of
the input signal is lower than the specified time value.
•
>= (greater than a time value)
: Trigger when the positive or negative slope time
of the input signal is greater than the specified time value.
•
[- - . - -] (within a range of time value)
:Trigger when the positive or negative slope
time of the input signal is greater than the specified lower limit of time and lower than
the specified upper limit of time value.
Figure 9.7 Limit Range
•
- -][- - (outside a range of time value)
: Trigger when the positive or negative slope
time of the input signal is greater than the specified upper limit of time and lower than
the specified lower limit of time value.
7.
When the trigger type is slope trigger, touch the trigger descriptor box, the trigger dialog box will the upper and
lower levels. The upper/lower level can be set in the following two ways.
•
Touch the Level Upper area in the quick menu to select the upper level, and then set
the level value by the virtual keypad or the Level knob on the front panel. To set the
lower level is similar.
•
Use the Level knob on the front panel directly to set the level value. Press the knob to
switch between upper and lower level, and rotate it to set the value.
type triggers on the positive or negative pulse with a specified width.
Figure 9.9
Pulse Trigger
To select
Pulse Trigger
type:
1.
Press the Setup key or use the touchscreen controls to enter the
Trigger Menu.
2.
Use the touchscreen to select
Type .
3.
The window shown in figure
9.3
will appear.
– Use the touchscreen to
Pulse.
4. Use the touchscreen to toggle the
Polarity
between
Positive
and
negative
.
condition.
5.
Use the touchscreen to select the
Limit Range.
•
<= (less than a time value)
: Trigger when the positive or negative pulse time of the input signal is lower than
the specified time value. For example, for a positive pulse, if you set t (pulse real width) 100ns, the waveform
will trigger.
Figure 9.10 Less than a Time Value
•
>= (greater than a time value)
: Trigger when the measured pulse width is less than the specified time. For
example, for a positive pulse, if you set the pulse width < 100 ns, the oscilloscope will trigger on the following
waveform:
Figure 9.11 Greater than a Time Value
•
[- -.- -] (within a range of time value)
: Trigger when the pulse width is between the lower and upper specified
times. For example, for a positive pulse, if you set the pulse width) between 100 ns and 300 ns, the oscilloscope
will trigger on the following waveform
Video triggering can be used to capture the complicated waveforms of most standard analog video signals. The trigger
circuitry detects the vertical and horizontal interval of the waveform and produces triggers based on the video trigger
settings you have selected.
The oscilloscope supports standard video signal field or line of NTSC (National Television Standards Committee), PAL
(Phase Alternating Line) HDTV (High Definition Television) and custom video signal trigger. To select
Pulse Trigger
type:
1.
Press the Setup key or use the touchscreen controls to enter the
Trigger Menu.
2.
Use the touchscreen to select
Type .
3.
The window shown in figure
9.3
will appear.
– Use the touchscreen to
Video.
4.
Use the touchscreen to select
Standard
, select the desired video standard. (
9.2.)
Standard
Type
Sync Pulse
NTSC
Interlaced
BI-level
PAL
Interlaced
BI-level
HDTV 720P/50
Progressive
Tri-level
HDTV 720P/60
Progressive
Tri-level
HDTV 1080P/50
Progressive
Tri-level
HDTV 1080P/60
Progressive
Tri-level
HDTV 1080iP/50
Progressive
Tri-level
HDTV 1080i/50
Progressive
Tri-level
Custom
Table 9.2
Video Standards
The parameters of
Custom Video Trigger
are shown in table
9.3.
Frame Rate
25 Hz, 30 Hz, 50 Hz, 60 Hz
Of Lines 300 to 2000
PAL
1, 2, 4, 8
HDTV 720P/50
1:1, 2:1, 4:1, 8:1
Trigger Position
Line
Field
(line value)/1
1
(line value)/2
1,2,3,4,5,6,7,8
(line value)/3
1,2,3,4,5,6,7,8
(line value)/4
1,2,3,4,5,6,7,8
(line value)/5
1,2,3,4,5,6,7,8
(line value)/6
1,2,3,4,5,6,7,8
(line value)/7
1,2,3,4,5,6,7,8
(line value)/8
1,2,3,4,5,6,7,8
Table 9.3
Custom Video Trigger Parameters
Line value: The number of lines set in the Of Lines (300 2000).
In the custom video trigger type, the corresponding "Of Fields" varies with the selection of the “Interlace” ratio.
Therefore, the number of fields selected and the number of lines corresponding to each field can also be varied. If
the "Of Lines" is set to 800, the correct relationship between them is as follows:
Of Lines
Of Fields
Interlace
Trigger Line
Trigger Field
800 1
1:1
800 1
800
1, 2, 4 or 8
2:1
400
1, 1~2, 1~4, 1~8
800
1, 2, 4 or 8
4:1
200
1, 1~2, 1~4, 1~8
800
1, 2, 4 or 8
8:1
100
1, 1~2, 1~4, 1~8
Table 9.4
Parameters Relations
5.
Press the
Sync
softkey to select
Any
or
Select
trigger mode.
•
Any
: Trigger on any of the horizontal pulses.
•
Select
: Trigger on the appointed line and field you have set.
–
Press the
Line
or
Field
softkey.
–
Turn the
Universal Knob
to set the value.
Table
9.5
list the line numbers per field for each video standard.
Standard
Field 1
Filed 2 NTSC
1 to 263
1 to 262
PAL
1 to 313
1 to 312
HDTV 720P/50, HDTV 720P/60
1 to 750
HDTV 1080P/50, HDTV 1080P/60
1 to 1125
HDTV 1080iP/50, HDTV 1080i/60
1 to 563
1 to 562
Table 9.5 Line Numbers Per Field
Triggering on a Specific Line of Video
Video triggering requires greater than 1/2 division of sync amplitude with any analog channel as the trigger source.
The example below is set to trigger on field 2, line 124 using the NTSC video standard.
1.
Press the Setup key or use the touchscreen controls to enter the
Figure 9.14 Triggering on a Specific Line of Video
Use a Custom Video Trigger
Custom video triggering supports frame rate of 25Hz, 30Hz, 50Hz and 60Hz, and the line range is available from 300 to
2000. The steps below show how to set custom trigger.
1.
Press the
Setup
key on the front panel to open the trigger menu.
2.
In the trigger menu, touch the
Type
, and select
Video.
3.
Touch the
Source
and select
CH1
as the trigger source.
4.
Touch the
Standard
and select the
Custom.
5.
Touch the
Custom Setting
to open the custom setting menu, touch the
Interlace
to select the required interlace
ratio (assuming that the interlace ratio is 8:1). Then set the frame rate, select the number of lines and the number
of fields.
6.
Touch the
Sync
to select the synchronization mode for the input signal:
a.
Select the
Any
mode, and the signal can be triggered on any line that meets the trigger condition.
b.
Select the
Select
mode, then set the specified line and the specified field to trigger the signal. Assuming that the
Field
is set to 8, you can select any field from 1 to 8, and each field can choose any line from 1 to 100.
Windows trigger provides a high trigger level and a low trigger level. The instrument triggers when the input signal passes
through the high trigger level or the low trigger level.
There are two kinds of window types: Absolute and Relative. They have different trigger level adjustment methods.
Under Absolute window type, the lower and the upper trigger levels can be adjusted respectively via the Level knob.
Under Relative window type the Center value is adjusted to set the window center, and the Delta value is adjusted to set
the window range. The lower and the upper trigger levels always move together.
Figure 9.15
Window Trigger
•
If the lower and the upper trigger levels are both within the waveform amplitude range, the oscilloscope will trigger
on both rising and falling edge.
•
If the upper trigger level is within the waveform amplitude range while the lower trigger level is out of the waveform
amplitude range, the oscilloscope will trigger on rising edge only.
•
If the lower trigger level is within the waveform amplitude range while the upper trigger level is out of the waveform
amplitude range, the oscilloscope will trigger on falling edge only.
Set Window Trigger Via Absolute and Relative Window Type
To select
Window Trigger
type:
1.
Press the Setup key or use the touchscreen controls to enter the
Trigger Menu.
2.
Use the touchscreen to select
Type .
3.
The window shown in figure
9.3
will appear.
– Use the touchscreen to select
Window.
4.
Use the touchscreen to toggle the
Window Type
between
Absolute
and
Relative
. condition.
When the window trigger type is set to
Relative
, touch the trigger descriptor box. The
pop-up menu will show two user-defined parameters:
Level +/-Delta
and
Level Cen-
ter.
The above two parameters can be set in the following two ways:
1.
Select the parameter in the Level +/-Delta area of the quick menu, then set the
parameter value by the virtual keypad or the Level knob on the front panel. Setting
the center level is similar.
2.
Directly use the Level knob on the front panel. Press the knob to switch between
Level +/-Delta
and
Center Level
, and rotate it to set values.
Figure 9.16
Window
Trigger Dialog Box
Level +/-Delta
represents half of the actual window area. For example, when the value is
200 mV, it actually represents a range of ± 200 mV, which is a 400 mV window.
Holdoff, coupling, and noise reject can be set in edge trigger, see the sections
Trigger when the times difference between the neighboring rising or falling edges meets the time limit conditions
(< =, > =, [ - - . - - ], - - ][ - - ).
When the trigger condition is set as an interval between two neighboring rising edges and it is less than the set time
value, the trigger diagram is as follows:
Figure 9.17
Interval Trigger
To select
Interval Trigger
type:
1.
Press the Setup key or use the touchscreen controls to enter the
Trigger Menu.
2.
Use the touchscreen to select
Type .
3.
The window shown in figure
9.3
will appear.
– Use the touchscreen to select
Interval.
4. Use the touchscreen to toggle the
Slope
between
Rising
and
Falling
.
condition.
5.
Use the touchscreen to select the
Limit Range.
•
< = (less than a time value)
: Triggers when the positive or negative pulse time of the input signal is lower
than the specified time value.
•
> = (greater than a time value)
: Triggers when the positive or negative pulse time of the input signal is
greater than the specified time value.
•
[ - - . - - ] (within a range of time value)
: Triggers when the positive or negative pulse time of the input
signal is greater than the specified lower limit of time and lower than the specified upper limit of time value.
•
- - ] [ - - (outside a range of time value)
: Triggers when the positive or negative pulse time of the input signal
is greater than the specified upper limit of time and lower than the specified lower limit of time value.
Holdoff, coupling, and noise reject can be set in edge trigger, see the sections
Dropout trigger includes two types: edge and state.
Edge
Triggers when an edge followed by a specified time with no edges is detected. This is useful for triggering on the end of
a pulse train.
Figure 9.18
Dropout Trigger Edge
To select
Edge Dropout Trigger
type:
1.
Press the Setup key or use the touchscreen controls to enter the
Trigger Menu.
2.
Use the touchscreen to select
Type .
3.
The window shown in figure
9.3
will appear.
– Use the touchscreen to select
Dropout.
4.
Use the touchscreen to set the
Over Time Type
to
Edge .
5. Use the touchscreen to toggle the
Slope
between
Rising
and
Falling.
Trigger source, slope (rising, falling), dropout type and time value can be set in the trigger dialog box. Holdoff, coupling,
and noise reject can be set in edge trigger, see the sections
Triggers when the signal enters or leaves a voltage level and stays there for a specified time. This is useful for detecting
when a signal gets stuck at a particular level.
Figure 9.19 Dropout Trigger State
To select
State Dropout Trigger
type:
1.
Press the Setup key or use the touchscreen controls to enter the
Trigger Menu.
2.
Use the touchscreen to select
Type .
3.
The window shown in figure
9.3
will appear.
– Use the touchscreen to select
Dropout.
4.
Use the touchscreen to set the
Over Time Type
to
State .
5. Use the touchscreen to toggle the
Slope
between
Rising
and
Falling.
Trigger source, slope (rising, falling), dropout type and time value can be set in the trigger dialog box. Holdoff, coupling,
and noise reject can be set in edge trigger, see the sections
The runt trigger detects a pulse that crosses the first threshold but not the second. It can occur when a logic driver has
insufficient slew rate to reach a valid logic level in the time available.
Figure 9.20
Runt Trigger
•
A positive runt pulse crosses through a lower threshold but not an upper threshold.
•
A negative runt pulse crosses through an upper threshold but not a lower threshold.
To select
Runt Trigger
type:
1.
Press the Setup key or use the touchscreen controls to enter the
Trigger Menu.
2.
Use the touchscreen to select
Type .
3.
The window shown in figure
9.3
will appear.
– Use the touchscreen to select
Runt.
4. Use the touchscreen to toggle the
Polarity
between
Positive
and
Negative.
5.
Use the touchscreen to select the
Limit Range.
•
<= (less than a time value)
: Trigger when the positive or negative slope time of the input signal is lower than
the specified time value.
•
>= (greater than a time value)
: Trigger when the positive or negative slope time of the input signal is greater
than the specified time value.
•
[- - . - -] (within a range of time value)
:Trigger when the positive or negative slope time of the input signal
is greater than the specified lower limit of time and lower than the specified upper limit of time value.
•
- -][- - (outside a range of time value)
: Trigger when the positive or negative slope time of the input signal is
greater than the specified upper limit of time and lower than the specified lower limit of time value.
Holdoff, coupling, and noise reject can be set in edge trigger, see the sections
The Pattern trigger identifies a trigger condition by looking for a specified pattern. The pattern trigger can be expanded
to incorporate delays.
Pattern durations are evaluated using a timer. The timer starts on the last edge that makes the pattern “true”. Potential
triggers occur on the first edge that makes the pattern false, provided that the time qualifier criterion has been met.
The oscilloscope provides 4 patterns: logical AND, OR, NAND and NOR combination of the channels. Each channel
can set to low, high or invalid.
Figure 9.21
Pattern Trigger
To select
Pattern Trigger
type:
1.
Press the Setup key or use the touchscreen controls to enter the
Trigger Menu.
2.
Use the touchscreen to select
Type .
3.
The window shown in figure
9.3
will appear.
–
Use the touchscreen to select
Pattern.
4.
Use the touchscreen to select the
Logic
pattern.
5.
Press the
Source Setting
to assign one of the following to each channel:
Don’t care, High
or
Low.
–
When digital channels are turned on, the logic state of the digital channel can also be set in the source setting
dialog box.
Logic Settings of Analog Channels Logic Settings of Digital Channels
Figure 9.22
Source Setting
•
Low: Sets the pattern to low on the selected channel. A low is a voltage level that is less than the channel’s
trigger level or threshold level.
•
High sets the pattern to high on the selected channel. A high is a voltage level that is greater than the channel’s
trigger level or threshold level.
•
Don’t care sets the pattern to don’t care on the selected channel. Any channel set to don’t care is ignored and is
not used as part of the pattern.
– If all channels in the pattern are set to
Don’t care
, the oscilloscope will not trigger.
6.
Use the touchscreen to select the
Limit Range.
•
When the logic is AND or NOR, the time limit condition is available. This setting is particularly useful to filter
the hazard signals of combinational logic.
•
When the logic is
OR
or
NAND
, the time limit setting is not supported.
Holdoff, coupling, and noise reject can be set in edge trigger, see the sections
The oscilloscope’s trigger mode includes Auto, Normal, and Single. Trigger mode affects the way in which the oscilloscope
searches for the trigger.
After the oscilloscope starts running, the oscilloscope operates by first filling the pre-trigger buffer. It starts searching
for a trigger after the pre-trigger buffer is filled and continues to flow data through this buffer while it searches for the
trigger. While searching for the trigger, the oscilloscope overflows the pre-trigger buffer and the first data put into the
buffer is first pushed out (First Input First Out, FIFO).
When a trigger is found, the pre- trigger buffer contains the events that occurred just before the trigger. Then, the
oscilloscope fills the post- trigger buffer and displays the acquisition memory.
To select the trigger mode, press the key corresponding to the desired mode.
Auto
If the specified trigger conditions are not found, the triggers are forced and acquisitions are made so that signal activity
is displayed on the oscilloscope.
Auto
trigger mode is appropriate when:
•
Checking DC signals or signals with unknown levels or activity.
•
When trigger conditions occur often enough that forced triggers are unnecessary.
Normal
Triggers and acquisitions only occur when the specified trigger conditions are found. Otherwise, the oscilloscope holds
the original waveform and waits for the next trigger.
Normal
trigger mode is appropriate when:
•
Only specific events specified by the trigger settings are to be acquired.
•
Triggering on an infrequent signal from a serial bus such as I2C, SPI, CAN, LIN, etc. or another signal that arrives
in bursts.
Normal trigger mode stabilizes the display by preventing the oscilloscope from auto-triggering.
Single
The oscilloscope waits for a trigger and displays the waveform when the trigger condition is met, the acquisition is stopped
when the trigger conditions are met.
The oscilloscope can be forced to trigger by pressing the Single button twice. The trigger status in the upper right
corner of the screen will be displayed as "Stop".
9.4 Trigger Level
Trigger level and slope define the trigger point.
Figure 9.23
Trigger Point
To set the trigger level:
•
Use the touchscreen controls to open the
Trigger
dialog by pressing the trigger descriptor box.
Figure 9.24
Trigger
Descriptor Box
–
Press the
Level
box to set the trigger level by the virtual keypad. (Figure ??)
– Press
to increase and to decrease the trigger level scale.
The position of the trigger level for the analog channel is indicated by the trigger level icon (if the analog channel is
on). The value of the analog channel trigger level is displayed in the upper- right corner of the trigger descriptor box.
The example in figure
9.23
highlights the Positive edge slope at a trigger level of 100 mV.
Trigger coupling is only valid when the trigger source is set to an analog channel, EXT or EXT/5.
Figure 9.25 Trigger
Coupling
To set the trigger coupling:
1.
Press the Setup key or use the touchscreen controls to enter the
Trigger
menu.
2.
Use the touchscreen controls to select the
Coupling
option.
– The window shown in figure
9.25
will appear.
3.
Use the touchscreen controls to select the desired coupling.
Trigger coupling is separate from channel coupling.
The oscilloscope provides 4 kinds of trigger coupling modes:
DC
Allows both AC and DC components of the waveform into the trigger path.
AC
Blocks all the DC components and attenuate signals lower than 8 Hz. Use AC coupling to get a stable edge trigger when
your waveform has a large DC offset.
LF Reject
The signal is coupled through a capacitive high-pass filter network, DC is rejected and low frequencies are attenuated.
For stable triggering on medium to high-frequency signals. See the Specifications for details of the cut-off frequency.
HF Reject
Signals are DC coupled to the trigger circuit and a low-pass filter network attenuates high frequencies (used for triggering
on low frequencies). See the Specifications for details of the cut-off frequency.
Trigger holdoff can be used to add an additional, user-defined delay to the re-arming of the trigger circuit. This provides
control over how rapidly, or how often, the oscilloscope can be triggered. The oscilloscope will not trigger until the
holdoff time expires.
Use the holdoff to trigger on repetitive waveforms that have multiple edges (or other events) between waveform repetitions.
You can also use holdoff to trigger on the first edge of a burst when you know the minimum time between bursts.
9.6.1 Holdoff by Time
Holdoff time is the amount of time that the oscilloscope waits before rearming the trigger circuitry. The oscilloscope will
not trigger until the holdoff time expires.
Use the holdoff to trigger on repetitive waveforms that have multiple edges (or other events) between waveform repetitions.
You can also use holdoff to trigger on the first edge of a burst when you know the minimum time between bursts.
For example, to achieve a stable trigger on the repetitive pulses shown in the figure below set the holdoff time (t) to 200
ns < t < 400 ns.
Figure 9.26
Holdoff
9.6.2 Holdoff by Event
The holdoff event is the number of events that the oscilloscope counts before re-arming the trigger circuitry. The
oscilloscope will not trigger until the counter tracking holdoff events reaches the set value.
Finding the Repetition of a Waveform
To find the repetition of the waveform:
1.
Press the Stop key to stop data acquisition.
2.
Turn the Horizontal Position Knob and the Horizontal Scale Knob to find where the waveform repeats.
Press the Setup key or use the touchscreen controls to enter the
Trigger
menu.
2.
Use the touchscreen controls to select the
Holdoff
option.
– The window shown in figure
9.27
will appear.
3.
Use the touchscreen controls to select the desired Holdoff.
4.
Use the touchscreen controls to select
Holdoff Events
or
Holdoff Time
depending of the trigger holdoff that was selected.
Holdoff Events
Holdoff Time
Figure 9.28
Trigger Holdoff
–
The holdoff events of the oscilloscope is adjustable from 1 to 100000000
events.
–
The holdoff time of the oscilloscope is adjustable from 8 ns to 30 s.
5.
Use the virtual keypad or the
Universal knob
to adjust the number of hold-
off events or the holdoff time.
Adjusting the time scale and horizontal position will not affect the holdoff time.
9.6.3 Start Holdoff On
Start Holdoff On
defines the initial position of holdoff.
The 2560B series offers the following start holdoff on options:
Acq Start
: The initial position of holdoff is the first time point satisfying the trigger condition. In the example above,
each holdoff starts from the first rising edge of the pulse sequence.
Last Trig Time: The initial position of holdoff is the time of the last trigger. In the example above, the last trigger
position is at the second rising edge of the pulse sequence and the second holdoff starts from that point.
Press the Setup key or use the touchscreen controls to enter the
Trigger
menu.
2.
Use the touchscreen controls to select the
Start Holdoff On
option.
– The window shown in figure
9.29
will appear.
3.
Use the touchscreen controls to select the desired start holdoff on.
9.7 Noise Reject
Noise Rejection
adds additional hysteresis to the trigger circuitry. By increasing the trigger hysteresis band, the possibility
of triggering on noise is decreased, however the trigger sensitivity is also decreased. A larger signal is required to trigger
the oscilloscope when Noise Rejection is enabled.
To enable
Noise Rejection:
To set the start holdoff on:
1.
Press the Setup key or use the touchscreen controls to enter the
Trigger
menu.
2.
Use the touchscreen controls to select the
Noise Reject
option.
– Tapping
Noise Reject
will toggle between the on and off mode.
Disabled Enabled
Figure 9.30
Noise Reject
If the signal being probed is noisy, set up the oscilloscope to reduce the noise in the trigger path and on the displayed
waveform. First, stabilize the displayed waveform by removing the noise from the trigger path. Second, reduce the noise
on the displayed waveform.
•
Obtain a stable display.
•
Remove the noise from the trigger path setting Trigger coupling to LF Reject,
The 2560B Series includes a zone trigger to help isolate elusive glitches. There are
two user-defined areas: Zone1 and Zone2. The property of each zone as
intersect
or
not intersect
as an additional condition to further isolate an event.
Intersect
only
includes events that occur within the zone.
Not intersect
includes all events that
occur
outside of the zone.
To enable and the trigger zone function:
1.
Press the Setup key or use the touchscreen controls to enter the
Trigger Menu.
2.
Use the touchscreen to select
Zone .
3.
The menu shown in figure
9.32
will appear.
4.
Press the
Zone
option to toggle the zone function on or off.
9.8.1 Creating and Moving Zones
The zones can be created and moved by gestures or by setting the zone settings in the
Zone Setting Menu. The color of the zone’s outline is consistent with the color of the
specified source.(Channel 1 = Yellow, etc.)
Gestures
When the zone trigger is turned on , touch-and-hold on any position within the waveform area and draw a rectangular box, as shown in figure 9.31.
When the finger moves out of the screen, a menu pops up for selecting the zone and
setting the zone properties:
The 2560B series supports serial bus triggering and decoding on the following serial bus protocols:
I2C, SPI, UART,
CAN and LIN
and also support optional
FlexRay, CAN FD, I2S, MIL-STD-1553B, SENT and Manchester.
This chapter introduces the method of triggering and decoding these serial signals in details.
Figure 10.1
Bus Protocols
To select
Serial Trigger
type:
1.
Press the Setup key or use the touchscreen controls to enter the
Trigger Menu.
2.
Use the touchscreen to select
Type .
3.
The window shown in figure
9.3
will appear.
– Use the touchscreen to select
Serial.
4.
Use the touchscreen to select
Protocol.
–
The dropdown menu shown in figure
10.1
will appear.
5.
Select the desired bus protocol.
To select the decoding
Bus Protocol:
1.
Press the Decode key or use the touchscreen controls to enter the
Decode Menu.
2.
Use the touchscreen to enable
Bus Operation.
3.
Use the touchscreen to select
Bus Protocol.
4.
The window shown in figure
10.1
will appear with the addition of the
Manchester
protocol.
5.
Select the desired bus protocol.
10.1 I2C Trigger and Serial Decode
10.1.1 Setup for I2C Signals
To set the I2C (Inter-IC bus) signal first, connect the serial data signal (SDA) and serial clock signal (SCL) to oscilloscope.
Set the mapping relation between channels and signals and then set the threshold level of each signal.
Set the source of SCL. In the example shown in figure
10.2
, SCL is connected to C2.
5.
Set the threshold level of SCL. It is 750 mV for the LVTTL signal in this example.
6.
Set the source of SDA. In the example above, SDA is connected to C1.
7.
Set the threshold level of the SDA channel.
–
The threshold voltage level is for decoding, and it will be regarded as the trigger voltage level when the trigger
type is set to serial.
–
Threshold level line. It only appears when adjusting the threshold level.
8.
Touch
option to return previous menu.
The signal settings of decoding and triggering are independent. To
synchronize the settings between decode and trigger, perform the
Copy Setting function in the Decode menu.
1.
Touch
Protocol Copy
in the Decode menu to synchronize the
settings between the trigger and decoding configurations.
Touch Setup key or use the touchscreen controls to enter the
Trigger Menu.
2.
Use the touchscreen to select
Type .
3.
The window shown in figure
9.3
will appear.
– Use the touchscreen to select
Serial.
4.
Use the touchscreen to select
Protocol.
–
The dropdown menu shown in figure
10.1
will appear.
5.
Use the touchscreen to select
I2C
protocol.
6.
Use the touchscreen to select
Trigger Setting.
7.
Use the touchscreen to select
Conditions.
–
The dropdown menu shown in figure
10.8
will be displayed.
–
Select the desired condition.
Figure 10.8
I2C Conditions
•
If EEPROM is selected :
–
Touch
Limit Range
option to set the qualifier (
=, < or >).
–
Touch
Data1
and set its value by turning the
Universal Knob
or using the virtual keypad.
•
If 7 Addr & Data or 10 Addr & Data is selected:
–
Touch
Addr
option and turn the
Universal Knob
to select the 7- bit or 10- bit device address.
–
Touch
Data1
or
Data2
and set the value.
–
Touch
R/W bit
and select write-frame or read-frame to trigger the oscilloscope.
If the device’s address is 7-bit, the value of address is in range of 0x00 to 0x7F. If device’s address is 10-bit,
the value of address is in range of 0x00 to 0x3FF.
Setting the SPI (Serial Peripheral Interface) signal includes two steps: Connecting the CLK, MISO, MISO and CS signal
to oscilloscope, specifying the parameters of each input signal.
1.
Touch Decode key to enter the
Decode
menu.
2.
Touch
Decode
and select the desired slot (
Decode1
or
Decode2).
3.
Touch
Protocol
to display the available SPI Signals.
4.
Set CLK (clock signal):
•
Touch CLK to enter CLK menu.
•
Touch
CLK
to select the channel that is connected to the SPI clock signal.
•
Touch
Threshold
to set the SPI clock signal’s threshold voltage level by Universal Knob.
–
The threshold voltage level is for decoding, and it will be regard as the trigger voltage level when set the
trigger type to serial.
•
Touch
Edge Select
to set the oscilloscope will samples at clock signal’s rising edge or falling edge.
•
Touch
to return previous menu.
5.
Set MISO:
•
Touch
MISO
to enter the
MISO
menu.
•
Touch
MISO
to select the channel that is connected to the SPI MISO signal.
•
Touch
Threshold
softkey, then use the
Universal Knob
to set the SPI MISO signal’s threshold voltage level.
–
The threshold voltage level is for decoding, and it will be regard as the trigger voltage level when set the
trigger type to serial.
•
Touch
to return to the previous menu.
6.
Set MOSI:
•
Press the
MOSI
to enter the MOSI menu.
•
Touch
MOSI
to select the channel that is connected to the SPI MOSI signal.
•
Press the
Threshold
softkey, then use the
Universal Knob
to set the SPI MOSI signal’s threshold voltage level.
– The threshold voltage level is for decoding, and it will be regarded as the trigger voltage level when the trigger
If the time between two edges of the clock signal is less than (or equal to)
the value of timeout, the signal between the two edges is treated as a frame.
The range of clock timeout is 100 ns - 5 ms.
This setting is suitable for case where the CS signal is not connected, or the
number of oscilloscope channels is insufficient (such as 2-channel
oscillo
scopes).
Table 10.1 CS Type Parameters
Example
Connect the data, CLK and ˜CS signals of a SPI bus respectively to C1, C2, and C3. Data width = 8-bit, Bit order =
MSB, CS polarity = CS, and 12 data bytes are transmitted in one frame.
In the SPI trigger signal menu, set the source and threshold of CLK, MISO and CS signals, then copy the trigger settings
to decoding. Adjust the timebase, so that the falling edge of the CS signal is shown in the display:
When the CS type is set to Clock Timeout, the clock idle time between frames is T3, the clock period is T1, then set
the timeout to a value between T1 and T3
Figure 10.12
Example 2
If the data width is set to be greater than 8 bits (such as 16 bits), the clock idle time between 8-bit data packets T2,
and then set the timeout time to a value between T1/2+T2 and T3.
to select the channel that is connected to the RX signal.
b.
Touch first
Threshold
key to set the RX signal’s threshold voltage level by
Universal Knob
. The threshold
voltage level is for decoding, and it will be regard as the trigger voltage level when set the trigger type to serial.
6.
Set TX:
a.
Touch TX to select the channel that is connected to the TX signal.
b.
Touch second Threshold key to set the TX signal’s threshold voltage level by Universal Knob. The threshold
voltage level is for decoding, and it will be regard as the trigger voltage level when set the trigger type to serial.
7.
Touch
to return previous menu.
8.
Touch
Configure
to enter
BUS CONFIG
menu.
9.
Touch
Baud
to set baud rate.
•
The baud rate can be set as predefined value.
•
If the desired baud rate is not listed, press the
Baud
and select custom option, press the
Custom
and turn the
Universal Knob
to set the desired baud rate.
10.
Touch
Data Length
and set byte bits (5-8) by
Universal Knob.
11.
Touch
Parity Check
to set the type of parity check (
Even, Odd, Mark, Space
or
None).
12.
Touch
Stop Bit
to set the length of stop bit (1, 1.5 or 2 bits).
13.
Touch
Next Page
softkey.
14.
Touch Bit Order to select the bit order (LSB or MSB).
the operator to set the parameters, ID b is split into several bytes. For example, if the ID’s length is 11 bits,
it will be split into two bytes, a byte includes 8 bits. If “1st byte” is selected, only the 8 least significant bits
can be changed.)
•
If you select the
ID+DATA
condition:
a.
Touch
ID
bits to select the ID’s length (11 or 29 bits).
b.
Touch
Curr ID Byte
and turn the
Universal Knob
to select the byte that you want to modify.
c.
Touch
ID
and set the ID’s value by turning the
Universal Knob.
d.
Touch
Data
and set the value of the first byte by turning the
Universal Knob.
Figure 10.20 CAN Trigger
10.4.1 CAN Serial Decode
Upon completing the setup of can signal and trigger, the CAN signals can be decoded.
1.
Press
Decode
–>
Decode
. Select one of the options from the Decode1 and Decode2.
There are two steps of setting the LIN signal, connecting the signal to oscilloscope, specifying the parameters of each
input signal.
1.
Press the
Decode
key to enter the
DECODE
function menu.
2.
Touch
Decode
and select the desired slot (Decode1 or Decode2).
3.
Touch
Protocol
and then select
LIN
by turning Universal Knob.
4.
Touch
Signal
to enter the
SIGNAL
menu as shown in figure ??.
5.
Touch
Source
to select the channel that is connected to the LIN signal.
6.
Touch
Threshold
and set the LIN signal’s threshold voltage level by
Universal Knob
. The threshold voltage level is
for decoding, and it will be regard as the trigger voltage level when set the trigger type to serial.
7.
Touch
to return previous menu.
8.
Touch
Configure
to enter the
BUS CONFIG
menu.
9.
Touch
Baud
to set baud rate.
•
The baud rate can be set as predefined value.
•
If the desired baud rate is not listed, select
custom
option, press the Custom and turn the
Universal Knob
to
set the desired baud rate.
10.5.2 LIN Trigger
This section provides a brief introduction and description for the operation of the LIN trigger.
Trigger Condition
•
Break
— the oscilloscope will be triggered at the position of break field’s break delimiter.
•
ID
(Frame ID) — the oscilloscope will be triggered at the position of identifier field’s stop bit, if the value of a frame’s
ID is equal to specified value. (Note: If the data’s value is 0xXX, any data value will be matched)
•
ID + Data
(Frame ID and Data) — the oscilloscope triggers when a frame with an ID and data equal to the selected
values is detected. Use the Universal Knob to select the value for the ID, Data1 and Data2.
a.
The ID’s value is the same as set value.
b.
If you have set either Data1’s or Data2’s value, and the signal has a data is the same as that value. If you have
set both Data1’s and Data2’s value, the signal should have two consecutive data, the first data’s value is Data1,
second data value is Data2.
If the data’s value is 0xXX, any data value will be matched
•
Data Error
—the oscilloscope will be triggered when errors (such as ID check error, checksum error, sync byte field