Review the following safety precautions to avoid injury and prevent
damage to this product or any products connected to it. To avoid potential
hazards, use this product only as specified.
Only qualified personnel should perform service procedures.
To Avoid Fire or Personal Injury
Use Proper Power Cord. Use only the power cord specified for this
product and certified for the country of use.
Connect and Disconnect Properly. Do not connect or disconnect probes
or test leads while they are connected to a voltage source.
Ground the Product. This product is grounded through the grounding
conductor of the power cord. To avoid electric shock, the grounding
conductor must be connected to earth ground. Before making connections
to the input or output terminals of the product, ensure that the product is
properly grounded.
Connect the Probe Properly. The probe ground lead is at ground
potential. Do not connect the ground lead to an elevated voltage.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe
all ratings and marking on the pro d uct . Co nsult the product manual for
further ratings information before making connections to the product.
Version 1.03
Do Not Operate Without Covers. Do not operate this product with
covers or panels removed.
Avoid Exposed Circuitry. Do not touch exposed connections and
components when power is present.
Do Not Operate With Suspected Failures. If you suspect there is
damage to this product, have it inspected by qualified service personnel.
GeneralSafetySummary
GeneralSafetySummary
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
Symbols and Terms on the Product.
Protective Ground (Earth) Terminal.
CAUTION. Refer to user manual. Caution indicates a hazard
to property including this product
Make Simple Measurements..........................................129
5
6
Getting Started
Getting Started
1. Getting Started
Inspect package content
Inspect the shipping container for damage. If the shipping
container and/or packaging material are damaged, it should
be kept until the content of the shipment has been checked
for completeness and the oscilloscope has been checked
mechanically and electrically.
Verify that you received the following items:
Oscilloscope
Two high performance 150MHz oscilloscope probes
Power cord
User Manual
Test report
Comsoft Installation disk
USB data cable
Front Panel
This section provides an introduction to the front panel of the
model 2534 and 2540 series oscilloscopes. Generally, you
set up the front panel controls first and then make your
measurements.
The keys or knobs on the front panel bring up softkey menus
on the display that provide further access to the
oscilloscope’s features. The entry knob
select or change values. Five softkeys are located along the
right side of the display screen.
The following figure shows the front panel of the 2534/2540
series oscilloscopes.
is usually used to
If the content is incomplete or if there is any mechanical
damage or defect, please contact your authorized distributor.
7
8
Front panel
Getting Started
Getting Started
1. PRINT key
Press this key to print the current display to a USB mass
storage device.
2. AUTO key
When you press the AUTO key, the oscilloscope will
quickly determine which channels are active and
automatically scale the display to show the signals on
these channels.
3. Entry knob
The Entry knob is used to select items from menus and
input values. Its function changes when a different menu
is displayed. The curved arrow symbol
the Entry knob lights up when the Entry knob is active
and can be used to select a value.
4. Utility and Save/Load Menu
UTILITY
Access the system utility functions, such as Language
to the left of
Perform automated measurements of waveform
parameters.
ACQUIRE
The ACQUIRE menu lets you set the oscilloscope to
acquire waveforms in Normal, Peak Detect, or Average
mode and lets you select Real Time or Equivalent
sampling.
CURSOR
Press the CURSOR key to activate the cursors for
making custom voltage or time measurements on scope
signals.
DISPLAY
You can change the appearance of waveforms and the
display screen, select the color schemes, and adjust the
contrast, etc.
6. RUN control keys
Setup, I/O Setup, and Print Setup etc.
SAVE/LOAD
You can save your current setup and waveform data to
the oscilloscope’s internal memory or to a USB mass
storage device and retrieve the setup or waveform later.
5. Measure and Waveform Menu
MEASURE
9
10
The RUN/STOP key will be green when the oscilloscope
is waiting for a trigger event. When the trigger mode is
set to Normal, the display will not update until a trigger
event occurs. If the trigger mode is set to Auto, the
oscilloscope looks for a trigger; if no trigger is found, it
will be triggered automatically and the input signals will
be displayed.
Getting Started
Getting Started
Press the RUN/STOP key again to stop acquiring data.
The RUN/STOP key will be red. Now you may examine
the acquired waveform by panning and zooming.
Press the SINGLE key to acquire a single waveform
trace. The key will remain orange until the oscilloscope is
triggered. After the trigger event, the waveform is
displayed and triggering is disabled until re-armed by
pressing the SINGLE key again.
7. Vertical controls
You can use the vertical position control knob to move
the waveforms up or down on the display. There is one
vertical position control knob for each channel.
You can press the channel key CH1 or CH2 to switch the
channel on or off or access the associated channel menu
via the softkeys. There is one channel on/off key for each
channel.
contract or expand relative to the ground reference level.
There is one vertical scale control knob for each channel.
8. Horizontal controls
When the oscilloscope is in run mode, the horizontal
position control knob lets you set the acquisition window
relative to the trigger point. When acquisition is stopped,
you can turn this knob to pan through the stored
waveform data horizontally. This lets you see the
captured waveform before or after the trigger.
Press the horizontal MENU key to access the menu
where you can activate a delayed sweep or select X-Y or
Roll modes.
Turn the horizontal sweep rate control knob to adjust the
sweep speed. This changes the time base on the
display. When adjusted after the waveform has been
acquired and the oscilloscope is stopped, this has the
You can press the MATH key to access the FFT (Fast
Fourier Transform), multiply, subtract, and add functions.
You can press the REF key to save or load a reference
waveform from the internal memory or an external USB
mass storage device.
You can use the vertical scale control knob to change the
vertical scale of a waveform. The waveform display will
11
effect of stretching or compressing the waveform
horizontally.
9. Trigger controls
These controls are used to control how the oscilloscope
triggers to capture waveforms.
10. Probe compensation terminals
12
Getting Started
Getting Started
Use these two probe compensation terminals to match
each probe’s characteristics to the oscilloscope channel
to which it is connected.
11. External trigger input
This is the external trigger input BNC connector.
12. Channel input BNC
This is the channel’s input BNC connector. Connect the
oscilloscope probe or BNC cable to the BNC Connector.
13. MENU On/Off key
Press this key to toggle the menu display on and off.
14. Softkeys
Five menu softkeys are used to select the control and
parameter setting functions. Each softkey has a label
along the right side of the screen.
15. Power switch
Press once to turn power on, press again to turn power
off.
16. LCD display
The 320*240 matrix (5.7 inch) LCD displays captured
waveforms, setup information, measurement results, and
softkeys for setting up parameters.
Rear panel
3
1. Line input receptacle
AC power cord receptacle. Attach to an AC power line
with a safety ground.
2. USB connector
The USB device connector can be connected to a
computer.
3. Extension port
Connect optional extension communication modules.
Contact B&K Precision for availability.
Rear panel
21
13
14
Getting Started
Getting Started
Interpreting the display
The oscilloscope display contains channel acquisition
information, setup information, measurement results, and soft
keys for setting up measurement parameters.
record line color corresponds to the selected waveform
color.
5. The trigger position icon shows the trigger location within
the record.
6. The trigger position icon shows the trigger location in the
current displayed waveforms.
7. This readout shows the frequency of the trigger signal.
8. The trigger status readout shows the trigger status.
9. These are soft keys which allow you to set up additional
parameters.
10. The display area contains the displayed waveforms,
channel identifiers, and trigger and ground level
indicators. The channel information is shown in the same
color as the channel’s waveform.
11. This readout shows the delay setting or the trigger
location within the record, trigger source, trigger type,
and trigger level.
Interpreting the display
1. Readout shows the real time clock.
2. The USB icon displays when a USB device is connected.
3. Acquisition status readout shows RUN, STOP, WAIT, or
ROLL.
4. The square brackets show the location of the current
display window within the recorded waveform. The
15
12. The horizontal readout shows the Main or Delayed time
base setting.
13. The channel readouts show the scale factor, coupling,
bandwidth limit, digital filter, and invert status.
14. Waveform baseline icons show the zero-volt levels of the
waveforms. The icon colors correspond to the waveform
colors.
16
Basic Operation
Basic Operation
3. Connect probes to all other oscilloscope channels.
Repeat the procedure for each channel. This matches
2. Basic Operation
Probe Compensation
Perform this adjustment to match your probe to the input
channel. This should be done whenever you attach a passive
probe for the first time to any input channel. A poorly
compensated probe can introduce measurement errors.
1. Connect the oscilloscope probe to channel 1. Connect
the probe tip to the 3Vp-p@1kHz terminal and connect
the probe’s ground lead to the chassis ground terminal,
then press the AUTO key.
2. Use a nonmetallic tool to adjust the trimmer capacitor on
the probe for the flattest pulse possible. The trimmer
capacitor is located on the probe BNC connector.
Correctly compensated
each probe to each channel.
Using Autoset
The 2534 & 2540 series digital storage oscilloscopes provide
the Autoset function which sets the vertical, horizontal, and
trigger controls automatically.
Autoset function finds, turns on, and scales any channel with
a repetitive waveform that has a frequency of at least 50 Hz,
a duty cycle greater than 0.5% and an amplitude of at least
10 mV peak-to-peak. Any channels that do not meet these
requirements are turned off.
When you are using more than one channel, the Autoset
function sets the vertical controls for each channel and uses
the lowest-numbered active channel to set the horizontal and
trigger controls.
To configure the oscilloscope quickly and automatically,
Over compensated
Under compensated
press the AUTO key to display the connected signals that are
17
active.
To configure the oscilloscope to display multiple cycles,
press the Multi-Cycle softkey in the AUTO menu.
To configure the oscilloscope to display a single cycle, press
the Single Cycle softkey in the AUTO menu.
18
Basic Operation
Basic Operation
To undo the effects of Autoset, press the Undo Autoset
softkey in the AUTO menu before pressing any other key.
This is useful when you have unintentionally pressed the
AUTO key or do not like the settings Autoset has selected
and want to return to your previous settings.
Vertical Controls
Au
to set oscilloscope channel 1
Vertical controls
Vertical Position Control (CH1, CH2)
Turn the small vertical position knob above the channel key
to move the channel’s waveform and its ground level icon (
up or down on the display. The voltage value momentarily
displayed in the bottom left portion of the display represents
19
20
)
Basic Operation
Basic Operation
the voltage difference between the vertical center of the
display and the ground level(
Press the small vertical position knob above the channel key
to bring the channel’s waveform and its ground level icon (
directly back to the vertical center of the display.
Channel key CH1, CH2, MATH, REF
Press the channel key from the front panel to display the
channel’s menu. Pressing it again turns the display of the
channel on or off. The channel is displayed when the key is
illuminated.
You must be viewing the menu for a channel before you can
turn it off. For example, if CH1 and CH2 are both displayed
and the CH2 menu is now displayed. In order to turn CH1 off,
you should press the CH1 key first and CH1 menu will be
displayed; then press CH1 key again to turn off CH1.
Vertical Scale Control (CH1, CH2)
).
)
size. When coarse is selected, the vertical scale knob
changes the channel scale in a 1-2-5 step sequence.
CH1, CH2 Menu
Press the channel key CH1 to display the channel’s menu
and turn the display of the channel on.
Channel Coupling
Press the channel key CH1, then press the Coupling softkey
to select AC coupling mode.
AC coupling places a high pass filter in series with the input
waveform that blocks the DC component of the input signal.
AC coupling is useful for viewing waveforms with DC offsets.
Turn the large vertical scale knob below the channel key to
set the scale factor for the channel. The vertical scale knob
changes the channel scale in a 1-2-5 step sequence. The
scale factor value is displayed in the bottom left portion of the
display.
Press the large vertical scale knob to toggle between Fine
and Coarse. When fine is selected, you can change the
channel’s vertical sensitivity in smaller steps. This is useful
when you want to scale the displayed waveform to an exact
21
22
AC Coupling
Basic Operation
Basic Operation
Press the channel key CH1, then press the Coupling softkey
to select DC coupling mode.
DC coupling passes both AC and DC components of the
input signal. DC coupling is useful for viewing low frequency
waveforms that do not have large DC offsets.
DC Coupling
Press the channel key CH1, then press the Coupling softkey
to select GND coupling mode.
GND mode blocks both AC and DC components of the input
signal by grounding the channel amplifier’s input.
GND Coupling
23
24
Basic Operation
Basic Operation
Bandwidth Limit
When Bandwidth limit is on, the maximum bandwidth for the
channel is approximately 20 MHz. For waveforms with
frequencies below this, turning bandwidth limit on removes
unwanted high frequency noise from the waveform. The
bandwidth limit also limits the trigger signal path’s bandwidth
of any channel that has BW Limit turned on.
Press the channel key CH1, then press the BW Limit softkey
to turn the bandwidth limit off for the selected channel 1. BW
Limit off mode passes both the high and low frequency
components.
Press the channel key CH1, then press the BW Limit softkey
to turn the bandwidth limit on for the selected channel 1. BW
Limit on mode blocks the high frequency components over
20MHz.
BW Limit on
BW Limit off
25
26
Basic Operation
Basic Operation
Probe Attenuation Setting
Probes are available with various attenuation factors which
affect the vertical scale of the signal. You can manually select
the factor that matches the attenuation of your probe. This
allows you to see the correct measured values for
waveforms.
For example, to match a probe set to 1X connected to CH1,
press the channel key CH1, and then press the Probe
softkey and select 1X.
Digital Filter
Press the channel key CH1, then press the Digital Filter
softkey to display the FILTER-CH1 menu. Four kinds of filter
types are available:
Press the Upper Limit or Lower Limit softkey and then
adjust the Entry knob
range for the filter.
Low pass filter
High pass filter
Band pass filter
Notch filter
to set the high and low frequency
Set Probe Attenuation Factor to 1X
Digital Filter is off
27
28
Basic Operation
Basic Operation
Vertical Scale
Turn the large vertical scale knob below the channel key to
set the scale factor for the channel. The channel scale factor
value is displayed in the bottom left portion of the display.
Press CH1→ More 1/2→Volts Scale to select Coarse or
Fine adjustment. You can also press the large vertical scale
knob to toggle between Coarse and Fine. When Coarse is
selected, the vertical scale knob changes the channel scale
in a 1-2-5 step sequence. When Fine is selected, the vertical
Digital Filter is on
scale knob changes the channel scale using a smaller step
size.
Fine Vertical Scale
29
30
Basic Operation
Vertical Invert
Press CH1→ More 1/2→Invert to turn Invert on or off.
When Invert is turned on, the voltage values of the displayed
waveform are inverted. Invert affects how a channel is
displayed, but does not affect triggering. If the oscilloscope is
set to trigger on a rising edge, it remains set to trigger on a
same edge after the channel is inverted. If the signal is
represented as f(t), then inverting the signal displays –f(t).
Inverting a channel will also change the result of any function
Basic Operation
selected in the MATH menu or any measurement.
Vertical Invert off
Vertical Invert on
31
32
Basic Operation
Basic Operation
MATHFunctions
Dual Waveform Calculation
Press the MATH channel key to turn on the MATH menu
page1/2.
Softkey Options Description
A+B Add A and B
A-B Subtract B from A
Operate
Source A
Source B
Invert
More 1/2
A×B
FFT Access FFT menu
CH1 Select CH1 as Source A
CH2 Select CH2 as Source A
CH1 Select CH1 as Source B
CH2 Select CH2 as Source B
ON Invert ON
OFF Invert OFF
---- Select page 2/2
Multiply A by B
Press the softkey More ½ to display the MATH menu page
2/2.
Softkey Options Description
A+B Add A and B
A-B Subtract B from A
Operate
Display
Mode
More 2/2
A×B
FFT Access FFT menu
Split
Screen
Full
Screen
---- Select page 1/2
Multiply A by B
Vertical scale control
Vertical position control
Split the display into Main
and Math sections
Display Math waveform
in full screen
33
34
Basic Operation
Basic Operation
For example, we select the A+B math function, select CH1 as
the Source A, and select CH2 as the Source B, then we will
see the math waveform displayed in red as follows:
Math A+B
FFT Spectrum Analysis
You can use the FFT function to measure, for example, the
harmonic content and distortion in systems, to characterize
noise in DC power supplies, and to analyze vibration.
Press the MATH key to turn on the MATH menu page 1/2,
and then press the Operate softkey to select FFT. The FFT
menu page 1/2 will be displayed.
Softkey Options Description
A+B Add A and B
A-B Subtract B from A
Operate
A×B
Multiply A by B
FFT Enter FFT menu
CH1 Select CH1 for FFT
Source
CH2 Select CH2 for FFT
Hanning Use Hanning window
Hamming Use Hamming window
Blackman Use Blackman window
Window
Flattop Use Flattop window
Use Rectangular
Rectangular
window
Vertical scale in dBV
dBV RMS
Scale
RMS
V RMS Vertical scale in V RMS
More 1/2
---- Select page 2/2
35
36
Basic Operation
Basic Operation
Press the softkey More 1/2 to display the FFT menu page
2/2.
Softkey Options Description
A+B Add A and B
A-B Subtract B from A
Operate
Display
Mode
More 2/2
A×B
FFT Enter FFT menu
Split
Screen
Full
Screen
---- Select page 1/2
Multiply A by B
Vertical scale control
Vertical position control
Split the display into
Main and Math sections
Display Math waveform
in full screen
For example, we select CH1 as the source for the FFT, select
the rectangular window, set the vertical scale to dBV RMS,
and then we will get an FFT waveform similar to the following
figure. We can also measure the amplitude and frequency of
FFT points with the manual cursors ( see chapter application
examples, example “measure Frequency and Amplitude of a
FFT waveform”)
FFT Spectrum Analysis
37
38
Basic Operation
Basic Operation
REF Function
You might make measurement on a known good system,
save the result to internal memory or to a USB mass storage
device, then make the same measurement on a test system
and recall the reference waveform to see the difference.
Press the REF channel key to turn on the REF menu page
1/2.
Softkey Options Description
Source
Invert
More
1/2
CH1 Save CH1 as reference
CH2 Save CH2 as reference
REF vertical scale control
REF vertical position control
ON REF invert ON
OFF REF invert OFF
---- Select page 2/2
Press the softkey More 1/2 to display REF menu page 2/2.
Softkey Options Description
Internal
Storage
External
Storage
More
2/2
Press REF channel key to turn on the REF menu page 1/2
and load the latest saved reference waveform at the same
time from the internal memory.
You can use the horizontal position and scale control knob to
change the time base of the reference waveform.
INTERNAL
menu
EXTERNAL
menu
---- Select page 1/2
Save the reference
waveform to
internal memory.
Save the reference
waveform to a USB
mass storage
device.
Press
to change the vertical scale or position of the reference
waveform.
39
40
or softkey and turn the Entry knob
Basic Operation
Basic Operation
Press REF→ Internal Storage →Save to save the waveform
of the Source channel as the reference waveform to internal
memory.
Save a Reference waveform
Note: The reference waveform function is unavailable
Horizontal Controls
Use the horizontal controls to adjust the time base, adjust the
trigger location, and to examine waveform details more
closely.
when X-Y mode is selected.
Horizontal Controls
41
42
Basic Operation
Basic Operation
Horizontal Position Control
When the oscilloscope is in run mode, this control lets you set
the acquisition window relative to the trigger point. When
acquisition is stopped, you can turn this knob to pan through
the data horizontally. This lets you see the captured
waveform before the trigger or after the trigger.
The trigger position is marked with the letter “T” at the top of
the graticule and also in the waveform record icon at the top
of the screen.
The small inverted triangle is the time reference indicator.
When you change the horizontal scale, the waveforms
contract or expand about this point.
Press the horizontal position control knob to set the time
delay to zero. Then the trigger position indicator (
overlays the time reference indicator().
Note: The horizontal position control is unavailable when
X-Y horizontal mode is selected.
)
Horizontal MENU key
Press the horizontal MENU key to display the HORIZONTAL
menu. This menu lets you select the horizontal mode: Main,
Delayed, Roll, or X-Y, and set the Holdoff time.
Press the horizontal MENU key to display the HORIZONTAL
menu page 1/2.
Softkey Options Description
√ Main mode is ON
Main
---- Main mode is OFF
√Delayed mode is ON
Delayed
---- Delayed mode is OFF
√X-Y mode is ON
X-Y
---- X-Y mode is OFF
√Roll mode is ON
Roll
---- Roll mode is OFF
-More-
---- Select page 2/2
1/2
Horizontal Scale Control
Use the horizontal scale control to adjust the time base. The
scale expands or contracts around the center of the screen.
The horizontal scale factor can be set in a 1-2.5-5 sequence
(2534) or in a 1-2-5 sequence (2540/2542).
Press the horizontal scale control knob to toggle between
Main and Delayed horizontal mode.
43
44
Basic Operation
Basic Operation
Press the softkey More 1/2 to display the HORIZONTAL
menu page 2/2.
Softkey
Holdoff
Holdoff
Reset
Trig-Offset
Reset
-More-
2/2
Main Horizontal Mode
The Main horizontal mode is the normal viewing mode for the
Options Description
Set the holdoff time
between two triggers.
Reset the holdoff
----
----
---- Select page 1/2
time to the default
value of 100 ns.
Reset the delay time
to zero.
Press the horizontal MENU key and then press the Main
softkey to select the main horizontal mode.
oscilloscope. When acquisition is stopped, you can use the
horizontal controls to pan and zoom the waveform. When the
oscilloscope is operating in Main mode, use the horizontal
scale knob to change horizontal scale factor and use the
horizontal position knob to set the delay time. When the
oscilloscope is stopped, use the horizontal control knobs to
pan and zoom the waveform. The time base (second/division)
value is displayed at the bottom of the screen.
45
Main Horizontal Mode
1. The square brackets show the location of current display
window within the record.
2. Trigger position within the record.
3. Trigger position on the current waveform display window.
4. Readout shows the delay time or the trigger location
within the record relative to the time reference point ().
46
Basic Operation
Delayed Horizontal Mode
Delayed horizontal mode is an expanded version of main
mode. When Delayed mode is selected, the display divides in
half. The top half of the display shows the normal waveform
and bottom half displays the delayed waveform.
The Delayed waveform is a magnified portion of the normal
waveform. You can use the delayed waveform to locate and
horizontally expand part of the normal waveform for a more
detailed analysis of the waveform.
The area of the normal display that is expanded is marked on
each end of the trace with a colored area. The uncolored
Basic Operation
area shows what portion of the normal waveform is expanded
in the lower half.
To change the time base for the delayed window, turn the
horizontal scale knob. As you turn the knob, the time base for
the delayed window is displayed just above the main time
base.
To change the time base for the normal window, press the
Main softkey, then turn the horizontal scale control knob.
To view the following figure, connect a triangle signal to CH1,
press the horizontal MENU key and then press the Delayed
softkey to enter the Delayed mode. You can also press the
horizontal scale control knob to toggle between Main and
Delayed mode directly.
47
48
Delayed Horizontal Mode
Basic Operation
X-Y Horizontal Mode
X-Y mode changes the display from a volts-versus-time
display to a volts-versus-volts display. The time base is
turned off. The CH1 signal is plotted on the X axis and CH2
signal is plotted on the Y axis.
You can use X-Y mode to compare the frequency and phase
relationships between two signals. X-Y mode can also be
used with voltage transducers to display, for example, strain
Basic Operation
versus displacement, flow versus pressure, volts versus
current, or voltage versus frequency.
In order to get a better view of the waveform, the proper
vertical scale should be selected before entering the X-Y
mode.
Example: Use the X-Y mode to compare two signals with
the
same frequency and different phase. Connect the two signals
to CH1 and CH2 respectively. Press horizontal MENU key
and then X-Y softkey to select X-Y mode. The displayed
ellipse (a Lissajous figure) indicates the phase relationship of
the two signals.
X-Y Horizontal Mode
49
50
Basic Operation
Roll Horizontal Mode
Roll mode causes the waveform to move slowly across the
screen from right to left. It only operates on time base
settings of 500 ms/div or slower. If the current time base
setting is faster than the 500 ms/div limit, it will be set to
500ms/div when Roll mode is selected.
In Roll mode there is no trigger. The fixed reference point on
the screen is the right edge of the screen and refers to the
current moment in time. Events that have occurred are
scrolled to the left of the reference point. Since there is no
trigger, no pre-trigger information is available.
If you would like to pause the display in Roll mode, press the
SINGLE key. To clear the display and restart an acquisition
in Roll mode, press the SINGLE key again.
Basic Operation
Use Roll mode on low-frequency waveforms to yield a display
much like a strip chart recorder. The waveform will “roll”
across the display.
Press the horizontal MENU key and then press the Roll
softkey to select the Roll mode. The waveform moves slowly
across the screen from right to left.
51
52
Roll Horizontal Mode
Basic Operation
Holdoff Function
Holdoff sets the amount of time that the oscilloscope will wait
before rearming the trigger circuit. You can use the holdoff
function to stabilize the display of complex waveforms.
With the holdoff function, you can synchronize triggers. The
oscilloscope will trigger on one edge of the waveform, and
ignore further edges until the holdoff time is up. The
oscilloscope will then rearm the trigger circuit to wait for the
next edge trigger. This allows the oscilloscope to trigger on a
repeating pattern in a complex waveform.
Turn the Entry knob to increase or decrease the trigger hold
off time shown in the Holdoff softkey.
To get a stable trigger on the pulse burst shown on the
screen, set the holdoff time to be slightly less than the period
of the pulse burst.
Basic Operation
Holdoff Function
53
54
Basic Operation
Basic Operation
Trigger Controls
The trigger controls determine when the oscilloscope starts
to acquire and display the waveform. When a trigger is found,
the oscilloscope will acquire sufficent data to display the
waveform. The trigger controls are functional when the
oscilloscope is in Main or Delayed horizontal mode.
Trigger Control MENU key
Press the trigger control MENU key to show the TRIGGER
menu and then press the Type softkey to select Edge, Pulse
or Video.
Set to 50% key
Press the 50% key to set the trigger level to the 50%
amplitude level of the trigger source waveform.
Force Trigger key and secondary Local function
Press the FORCE key to force an immediate trigger event,
even in the absence of a signal. This function is useful in the
following situations.
If you do not see a waveform on the screen when using
Normal trigger mode, press the FORCE key to see if the
signal is on the screen (if not, you might have to adjust the
vertical position).
After you press the SINGLE key to set up for a single shot
Trigger Controls
55
acquisition, you can press the FORCE key to do a practice
acquisition to verify the control settings.
When the instrument is in remote mode, pressing this button
will revert the instrument to local mode. This function is
useful to set the DSO back to local mode after exiting the
Comsoft PC application.
56
Basic Operation
Basic Operation
Trigger Level Control
Use the trigger level control knob to adjust the trigger level.
This is the voltage level on the trigger waveform that causes
a trigger event. When you change the trigger level, a
horizontal line temporarily appears to show you the trigger
level on the screen. After the line disappears, the trigger level
is marked with a small left arrow.
Auto and Normal Trigger Modes
Press the trigger MENU key to display the TRIGGER menu
and press the Mode softkey to select Auto or Normal trigger
mode.
Note: The Auto and Normal trigger mode are unavailable
when the Video trigger type is selected.
Auto mode
Use the auto trigger mode for signals other than low
repetitive-rate signals and for unknown signal levels. To
the pre-trigger buffer will contain the events that occurred just
before the trigger. If no trigger is found, the oscilloscope
generates a trigger and displays the data as though a trigger
had occurred. In this case, the background of the Auto
indicator at the top of the display will flash, indicating that the
oscilloscope auto triggered.
When you press the SINGLE key, the oscilloscope will fill the
pre-trigger buffer and continue to flow data through the
pre-trigger buffer until the Auto trigger overrides the
searching and forces a trigger. At the end of the trace, the
oscilloscope will stop and display the results.
Normal mode
Use Normal trigger mode for low repetitive-rate signals or
when Auto trigger is not required.
In Normal mode the oscilloscope must fill the pre-trigger
buffer with data before it will begin searching for a trigger
display a DC signal, you must use Auto trigger mode since
there is no edge to trigger on.
When you press the RUN/STOP key to start acquiring a
waveform, the oscilloscope first fills the pre-trigger buffer. It
then starts to search for a trigger after the pre-trigger buffer is
filled and continues to flow data through this buffer while it
searches for the trigger. While searching for the trigger, if the
oscilloscope overflows the pre-trigger buffer, the first data put
into the buffer is the first pushed out. When a trigger is found,
57
event. While searching for the trigger, if the oscilloscope
overflows the pre-trigger buffer; the first data put into the
buffer is the first pushed out.
When the trigger event is found, the oscilloscope will fill the
post-trigger buffer and display the results. If the acquisition
was initiated by RUN/STOP, the process repeats. If the
acquisition was initiated by SINGLE, then the acquisition
stops.
58
Basic Operation
Basic Operation
In either Auto or Normal mode, the trigger may be missed.
This is because the oscilloscope will not recognize a trigger
event until the pre-trigger buffer is full.
Video Trigger
Choose video triggering to trigger on the odd fields, even
fields, or on all the lines of a NTSC or PAL/SECAM video
signal.
Press the trigger control MENU key to display the TRIGGER
menu, then press the Type softkey to select Video trigger.
Softkey Options Description
Video Video triggering
Type
Source
Polarity
Edge Edge triggering
Pulse Pulse width triggering
CH1 Trigger on CH1
CH2 Trigger on CH2
EXT Trigger on EXT
EXT/5 Trigger on EXT/5
Alternating
Odd Field Trigger on odd fields
Even Field Trigger on even fields
Trigger on CH1 and
CH2 alternately
Positive polarity
Negative polarity
Sync
Standard
59
60
All Lines Trigger on all lines
Trigger on a specific
Line #
line
Trigger on NTSC
NTSC
signal
Trigger on PAL or
PAL/SECAM
SECAM signal
Basic Operation
Basic Operation
The following figures show the video waveforms triggered on
odd fields and line 6.
Trigger on odd fields
Note: The trigger level control is unavailable when Video
type is selected.
Edge Trigger
Use edge triggering to trigger on the rising or falling edge of
the input signal at the trigger threshold.
Trigger on specific line 6
61
62
Basic Operation
Basic Operation
Press the trigger control MENU key to display the TRIGGER
Softkey Options Description
Video Video triggering
Type
Source
Slope
Mode
Edge Edge triggering
Pulse Pulse width triggering
CH1 Trigger on CH1
CH2 Trigger on CH2
EXT Trigger on EXT
EXT/5 Trigger on EXT/5
AC Line
Alternating
Auto Trigger even without a valid event.
Normal Trigger only on a valid event
Trigger on AC line signal
Trigger on CH1 and CH2
alternately
Rising edge of a signal
Falling edge of a signal
Pulse Width Trigger
Pulse width triggering sets the oscilloscope to trigger on a
positive or negative pulse of a specified width from 20ns to
10s.
AC AC coupling
DC DC coupling
Coupling
LF
Reject
HF
Reject
menu, then press the Type softkey to select Edge trigger.
Reject low frequencies
Reject high frequencies
63
64
Basic Operation
Basic Operation
Press the trigger control MENU key to display the TRIGGER
menu, then press the Type softkey to select Pulse trigger.
Softkey Options Description
Video Video triggering
Type
Edge Edge triggering
Pulse Pulse width triggering
CH1 Trigger on CH1
CH2 Trigger on CH2
EXT Trigger on EXT
Source
EXT/5 Trigger on EXT/5
Alternating
CH1 and CH2
alternately
Positive greater than
Positive equal
Positive within
Pulse
Mode
Positive less than
Negative greater than
Negative equal
Negative within
Negative less than
Pulse
Set the pulse width
Setup
Press the trigger control MENU key to display the TRIGGER
menu, press the Type softkey to select Pulse trigger, and
then press the More 1/2 softkey to display TRIGGER menu
page 2/2.
Softkey Options Description
Video Video triggering
Type
Edge Edge triggering
Pulse Pulse width triggering
Auto
Trigger even without a
Mode
valid event.
Normal
Trigger only on a valid
event
AC AC coupling
DC DC coupling
LF
Coupling
Reject low frequencies
Reject
HF
Reject high frequencies
Reject
More
---- Select page 1/2
2/2
More 1/2
---- Select page 2/2
65
66
Basic Operation
Basic Operation
ACQUIRE Menu
Press the ACQUIRE menu key to show the ACQUIRE menu.
ACQUIRE Menu key
Normal acquisition mode yields the best display for most
waveforms.
Average mode lets you average multiple waveforms to
reduce noise and increase resolution.
Softkey Options Description
Normal Normal acquisition.
Mode
Sampling
Record
Average
Peak Detect
Equivalent Equivalent sampling.
Real Time Real time sampling.
---- Select Record menu.
Average the
waveforms.
Peak detect
acquisition.
Peak Detect mode should be used to display narrow pulses
that occur infrequently. It’s useful when looking for very
narow pulses at very slow time base.
Equivalent sampling mode is useful to display high
frequency repetitive signals.
Real Time sampling mode is useful to capture single-shot
signals.
Press the Mode softkey to select the Normal mode.
67
68
Basic Operation
Basic Operation
Press the Mode softkey to select the Average mode.
Softkey Options Description
Normal Normal acquisition.
Mode
Averages
Sampling
Record
Press the Mode softkey to select the Peak Detect mode.
Softkey Options Description
Average
Peak
Detect
Equivalent Equivalent sampling.
Real Time Real time sampling.
---- Select Record menu.
Averages the
waveforms.
Peak detect
acquisition.
Set the average
number to 2, 4, 8, 16,
32, 64, 128, or 256.
Record
Connect a sine wave signal to the CH1 channel, press
ACQUIRE→Mode to select Average mode. Turn the Entry
knob to set the number of averages to 16. The following two
figures show the difference between Normal acquisition and
Average acquisition.
---- Select Record menu.
Normal Normal acquisition.
Mode
Average
Peak
Detect
Equivalent Equivalent sampling.
Sampling
Real Time Real time sampling.
Averages the
waveforms.
Peak detect
acquisition.
69
70
Random noise on the displayed waveform
Basic Operation
Basic Operation
Waveform recorder
The waveform recorder can record input waveforms in a
sequence of frames originating from channel 1 or channel 2,
with a maximum acquisition depth of 1000 frames of 4k data
points each. The time interval between each frame is
adjustable from 1ms – 1000s. The waveform recorder lets
you monitor and analyze long term signal behavior by
recording data continuously over long periods of time then
playing it back for post acquisition analysis.
Press ACQUIRE→Record to show the RECORD menu.
Softkey Options Description
Average count of 16 selected to reduce random noise
Mode
Source
71
72
Record Record waveform.
Play
Back
Save
/Recall
OFF Exit Record function.
CH1 Record CH1 channel.
CH2 Record CH2 channel.
Pass/Fail
Out
Play back a recorded
waveform.
Save/Recall from
internal or external
memory.
Record waveform
from Pass/Fail mask
test
Basic Operation
Basic Operation
Set the time interval
Interval
End
Frame
Operate
between captured
frames. <1ms –
1000s>
Set the number of
frames you want to
record
Start recording.
Stop recording
Play back the record waveform sequence (menu 1 of 2
Press ACQUIRE→Record to show the RECORD menu.
Press the Mode softkey to select the Play Back function.
Softkey Options Description
Record Record the waveform
Play
Back
Mode
Save
/Recall
OFF Exit Record function
Operate
Play
Mode
Current
Frame
More
---- Select menu page 2/2
1/2
Press the More ½ button to show RECORD menu page 2
Play back a recorded
waveform.
Save/Recall from
internal or external
memory.
Start Playback
Stop Playback
Continuous play mode
One time play mode
Select a specific
frame within the frame
sequence and display
it on screen
73
74
Basic Operation
Basic Operation
Play back the record waveform sequence (menu 2 of 2)
Press ACQUIRE → Record to show the RECORD menu.
Softkey Options Description
Interval
Start
Set the playback interval
between two frames
Set the start frame to
Frame
End
Frame
ON
Msg
Display
OFF
More
---- Select menu page 1/2
2/2
playback
Set the end frame to
playback
Turn recorder message
display on
Turn recorder message
display off
Save/Recall the recorded waveform
Press ACQUIRE→Record to show the RECORD menu.
Press the Mode softkey to select the Save/Recall function.
Softkey Options Description
Record Record the waveform.
Play
back
Mode
Save
/Recall
OFF Exit Record function
Start
Play back a recorded
waveform.
Save/Recall from
internal or external
memory.
Set the first frame to
Frame
End
Frame
Internal
----
Storage
External
----
Storage
be saved.
Set the last frame to
be saved
Save/Recall from
internal memory.
Save/Recall from
external memory.
75
76
Basic Operation
Basic Operation
Exit the waveform recorder
Press the Mode softkey to select the OFF option and return
to the ACQUIRE menu.
Softkey Options Description
Record Record the waveform
Play back Play back the record
Save/Recall from
internal or external
memory
Return to ACQUIRE
menu
Mode
Save
/Recall
OFF Exit Record function
----
UTILITY Menu
Press the UTILITY menu key to show the UTILITY menu.
UTILITY Menu key
77
78
Basic Operation
Basic Operation
Press the UTILITY key to display the UTILITY menu page
1/2.
Softkey Options Description
I/O Setup
Print
Setup
System
Setup
Language
More 1/2
Press the More 1/2 softkey to display the UTILITY menu
page 2/2.
Softkey Options Description
Service
Pass/Fail
----
----
----
English default
---- Select menu page 2/2
----
----
Select I/O SETUP menu
Select PRINT menu
Select SYSTEM menu
Select Service menu
Select PASS/FAIL menu
I/O Setup
Press UTILITY → I/O Setup to display the I/O SETUP menu.
Softkey Options Description
USB SlaveSelect USB
Type
Note: An optional expansion module is required to
remotely control the instrument via RS232 and GPIB
interface. Contact B&K Precision Corp. for details and
availability.
RS232C Select RS232C
GPIB Select GPIB
Return to the
----
UTILITY menu
Self-Cal
More 2/2
RUN/STOP
AUTO
---- Select menu page 1/2
Start self-calibration
Exit self-calibration
79
80
Basic Operation
Print Setup
Press UTILITY→Print Setup to display the PRINT menu.
Press the Print to softkey to select File.
Softkey Options Description
Print to
File Type
Connect a USB mass storage device to the USB host
connector on the front or rear panel.
Press the File Type softkey to select the file format you want.
BMP is a bitmap of the screen and CSV are
comma-separated values. CSV format can be imported into
spreadsheets such as Excel.
Press the PRINT key to save the file to the USB mass
File Print to file
BMP BMP file format
CSV CSV file format
Return to the UTILITY
---menu
Basic Operation
System Setup
Press UTILITY→System Setup to display the SYSTEM
menu.
Softkey Options Description
Key
Sound
Alarm
Sound
Counter
Set Date
&Time
Note: With Alarm set to “On”, you will hear a beep under the
following conditions: A control reached its limit or an
invalid setting or configuration was encountered, in
which case you will also see a warning message
displayed on screen (example: Auto button is
pressed without detecting a signal).
ON Frequency counter on
OFF Frequency counter off
----
----
Key press sound on
Key press sound off
Alarm sound on
Alarm sound off
Select the DATE&TIME
menu
Return to the UTILITY
menu
storage device.
81
82
Basic Operation
Basic Operation
Press the Set Date&Time softkey to display the
DATE&TIME menu.
Softkey Options Description
Display
Hour
Min
Month
Day
Year
OK! Enter
ON Date & time display on
OFF Date & time display off
Set hour
Set minute
Set month
Set day
Set year
---- Apply the time and date
Service
Press UTILITY → Service to display the Service menu.
Softkey OptionsDescription
Display system
System
Information
Screen
Test
Key Test
----
---- Test the LCD screen.
----
----
information: Model,
Serial number,
Software version,
Installed modules.
Check keys and
controls operation.
Return to the UTILITY
menu.
83
84
Basic Operation
Basic Operation
Press UTILITY→Service to display the Service menu, then
press the System Info softkey to display the system
information, such as Model, Serial number, Power up times,
Software version and a list of installed modules.
System Information
Pass/Fail Mask test
The Mask Test function monitors waveform changes by. If
the input signal is within the mask threshold, a PASS verdict
will be generated. If the input signal lies outside the threshold,
a FAIL verdict will be generated.
For step by step setup instructions, refer to the Application
examples in chapter 3, section Pass/Fail measurement.
Press UTILITY→Pass/Fail to display the PASS/FAIL menu
1/2.
Softkey OptionsDescription
Enable
Test
Source
Operate
ON Turn mask function on
OFF Turn mask function off
CH1 Source signal is CH1
CH2 Source signal is CH2
Start Pass/Fail test
Stop Pass/Fail test
Setup Mask
More 1/2
85
86
----
---- Display the menu 2/2
Set up the
specifications
Basic Operation
Basic Operation
Press More 1/2 to display the PASS/FAIL menu 2/2.
Softkey Options Description
Msg
Display
Output
Stop on
Output
ON Display P/F information
OFF
PASS Indicate Pass condition
PASS+
FAIL Indicate Fail condition
FAIL+
ON Stop sampling on output
OFF
Turn Display P/F
information off
Output to screen and
sound alarm on Pass
condition
Output to screen and
sound alarm on Fail
condition
Continue sampling on
output
Press UTILITY→Pass/Fail→Setup Mask to display the
MASK menu 1/2.
Softkey Options Description
X Mask
Y Mask
Create
Mask
More
1/2
----
----
----
Set the mask’s
horizontal failure margin
Set the mask’s vertical
failure margin
Create the P/F mask
based on the current
waveform and the
above failure margins
Return to the
PASS/FAIL menu
Display the menu 2/2
More 2/2
Note:The Pass/Fail function is not available when X-Y
mode is selected.
----
----
Return to the UTILITY
menu
Display the menu page
1/2
87
88
Basic Operation
Basic Operation
Press More 1/2 to display the MASK menu 2/2.
Softkey Options Description
Internal
Storage
External
Storage
More 2/2 ----
----
----
----
Store the PASS/FAIL
tolerance mask to
internal memory.
Store the PASS/FAIL
tolerance mask to
external USB mass
storage device.
Return to the
PASS/FAIL menu.
Display the menu page
½.
Self-Calibration
To maximize the measurement accuracy, it is recoomended
to perform a self-calibration.
Self-calibration uses internally-generated signals to calibrate
circuits that affect channel scale, offset, and trigger
parameters. Disconnect all inputs and allow the oscilloscope
to warm up for at least 30 minutes before performing a
self-calibration.
Press UTILITY→Self-Cal to display the self-calibration
page. Press the AUTO key to exit the self-calibration page.
Press the RUN key to start the self-calibration.
Self Calibration
89
90
Basic Operation
Basic Operation
Note: Warm up the oscilloscope for at least 30 minutes
before performing self-calibration.
MEASURE Menu
MEASURE Menu key
Press the MEASURE key to display the MEASURE menu.
Softkey Options Description
Source
Voltage
Time
Clear
Measure
All
CH1 Measure CH1
CH2 Measure CH2
Select the Voltage
---measurement menu.
Select the Time
---measurement menu.
Turn off the
---measurement readouts.
Display all
ON
measurements.
Turn off all
OFF
measurements.
91
92
Basic Operation
Basic Operation
Voltage Measurements
Press MEASURE→Voltage to display the VOLTAGE menu
page 1/4.
Softkey Options Description
The Peak-Peak value is
Peak-Peak
Amplitude
Max
Min
----
----
----
----
the difference between
maximum and minimum
values.
The Amplitude value is
the difference between
its High and Low values.
Max is the maximum
voltage in the waveform
display.
Min is the minimum
voltage in the waveform
Press the More 1/4 softkey to display the VOLTAGE menu
page 2/4.
Softkey Options Description
Measures the flattop
High
Low
Average
----
----
----
voltage (average value)
of the upper part of the
waveform.
Measures the flat base
voltage (average value)
of the upper part of the
waveform.
Average value is the sum
of the samples divided by
the number of samples
over the entire waveform.
RMS value is the true
display
More 1/4
---- Display menu page 2/4
93
94
RMS
More
2/4
----
---- Display menu page ¾.
Root Mean Square
voltage over the entire
waveform.
Basic Operation
Basic Operation
Press the More 2/4 softkey to display the VOLTAGE menu
page 3/4.
Softkey Options Description
The Cycle Avg value is
Cycle Avg
Cycle
RMS
Overshoot
----
----
----
the sum of the samples
divided by the number of
samples over one period.
The Cycle RMS value is
the true Root Mean
Square voltage over one
period.
The Overshoot value is
the distortion that follows
a major edge transition
expressed as a
percentage of amplitude.
Press the More 3/4 softkey to display the VOLTAGE menu
page 4/4.
Softkey Options Description
More 4/4
----
---- Display menu page 1/4
Return to the MEASURE
menu
The Preshoot value is the
distortion that precedes a
Preshoot
More 3/4
----
---- Display menu page 4/4.
major edge transition
expressed as a
percentage of amplitude.
95
96
Basic Operation
Basic Operation
Time Measurements
Press MEASURE→Time to display the TIME menu page
1/5.
Softkey Options Description
Frequency is defined as
Frequency
Period
Rise Time
----
----
----
1/period of the first
cycle.
Period is the time period
of the first complete
waveform cycle.
Rise Time is the time
that the first
positive-going edge
takes to rise from 10% to
90% of its amplitude.
Fall Time is the time that
the first negative-going
Press More 1/5 softkey to display the TIME menu page 2/5.
Softkey Options Description
+Width is the time
between the 50%
amplitude points of the
first positive pulse.
-Width is the time
between the 50%
amplitude points of the
first negative pulse.
+Duty is the ratio of the
first positive width to its
period, expressed as a
percentage.
-Duty is the ratio of the
first negative width to its
period, expressed as a
+Width
-Width
+Duty
-Duty
----
----
----
----
Fall Time
More 1/5
----
---- Display menu page 2/5.
edge takes to fall from
90% to 10% of its
amplitude.
97
98
More 2/5
---- Display menu page 3/5.
percentage.
Basic Operation
Basic Operation
Press the More 2/5 softkey to display the TIME menu page
3/5.
Softkey Options Description
Measures the delay
between 2 waveforms
Delay
Delay
----
----
using the first rising edge
of each channel as
reference.
Measures the delay
between 2 waveforms
using the first falling edge
of each channel as
reference.
Measures the delay
between 2 waveforms
using the time between
Measures the delay
between 2 waveforms
Delay
More 3/5
----
---- Display menu page 4/5.
using the time between
the first falling edge of
CH1 and the first rising
edge of CH2.
Delay
----
the first rising edge of
CH1 and the first falling
edge of CH2 (50%
amplitude point.
99
100
Basic Operation
Basic Operation
Press the More 3/5 softkey to display the TIME menu page
4/5.
Softkey Options Description
Phase 1→2 is the time in
degrees between
corresponding points from
waveforms 1 to waveform
2. [See note]
Phase 2→1 is the time in
degrees between
corresponding points from
waveforms 1 to waveform
2. [See note]
X at Max is the X axis value
(i.e., time) referred to the
trigger point at the first
Phase
----
1→2
Phase
----
2→1
X at Min is the X axis value
(i.e., time) referred to the
trigger point at the first
X at Min
More 4/5
Note: time in degrees between the two points is calculated
by measuring the time between the two points, dividing the
time by the waveform’s period, then multiplying the fraction
by 360 degrees.
Press the More 4/5 softkey to display the TIME menu page
5/5.
Softkey Options Description
----
---- Display menu page 5/5
displayed occurrence of the
waveform Minimum,
starting from the left side of
the display, to the following
waveforms minimum
X at Max
----
displayed occurrence of the
waveform Maximum,
starting from the left side of
the display to the following
waveform maximum.
101
102
More 5/5
----
---- Display menu page 1/5
Return to the MEASURE
menu
Basic Operation
Basic Operation
Automatic Measurement Procedure
Press MEASURE→Measure All to turn on Auto
Measurements. Up to 20 measurements of the current
channel are displayed at the center of the screen.
Press the Clear softkey to clear all displayed measurement
parameter.
Note: Up to three parameters can be displayed at the
same time on the bottom of the display. When
three parameters are already displayed, the first
parameter will be pushed out of the display
window when a new parameter is chosen for
display.
Note: “****” will be displayed when a parameter cannot
be measured correctly.
Press the Measure All softkey again to turn off Auto
Measurements.
Press MEASURE→Voltage to display the VOLTAGE menu
or press MEASURE → Time to display the TIME menu.
Press the softkey of the voltage or time parameters you want
to measure.
The selected parameter will be displayed on the bottom of
the display.
103
104
Basic Operation
Basic Operation
Measurement definitions
The following figure shows the voltage measurement points.
The following figure shows the time measurement points.
SAVE/LOAD Menu
SAVE/LOAD MENU key
Press the SAVE/LOAD key to display the SAVE/LOAD
menu.
Softkey Options Description
Internal
----
Storage
Display the INTERNAL
menu.
External
Storage
Factory
105
106
----
----
Display the EXTERNAL
menu.
Set the instrument to the
factory default
configuration.
Basic Operation
Basic Operation
Internal Storage
Press SAVE/LOAD→Internal Storage→Storage type to
display the INTERNAL menu and select Waveforms storage
type.
Softkey Options Description
Storage
type
Waveforms
Setups Setup file format
Waveform file format
Select a waveform file from
Wavexx
Wave01 to Wave10. Date
mm/dd/yy
and time of the current file
hh/mm
are displayed.
Save the display to the
Save
---current waveform file.
Load the current waveform
Load
---file.
Return to the SAVE/LOAD
---menu.
Press SAVE/LOAD→Internal Storage→Storage type to
display the INTERNAL menu and select Setups storage type.
Softkey Options Description
Storage
type
Waveforms
Setups Setup file format
Waveform file format
Select a setup file from
Setupxx
Setup01 to Setup10. The
mm/dd/yy
date and time of the current
hh/mm
file are displayed.
Save the current
Save
----
configuration to the
selected setup file.
Load from the selected
Load
---setup file.
Return to the SAVE/LOAD
---menu.
107
108
Basic Operation
Basic Operation
External Storage
Press SAVE/LOAD→External Storage to display the
EXTERNAL menu.
Softkey Options Description
New
Rename
Load
Delete
----
----
---- Load the current file.
----
----
Create a new file or folder in
the external memory.
Rename the current file or
folder.
Delete the current file or
folder.
Return to the SAVE/LOAD
menu
Press SAVE/LOAD→External Storage→New to display the
New menu.
Softkey OptionsDescription
New File
New
Folder
----
----
----
Display the New File menu.
Display the New Folder
menu.
Return to the EXTERNAL
menu
Note: The above External Storage menu is only available
if a USB flashdrive is connected to the front panel
USB host port.
109
110
Basic Operation
Basic Operation
Press SAVE/LOAD→External Storage→New→New File to
display the New File menu.
Softkey Options Description
Setups Save as setup files
Save as
Enter
Character
Delete
Character
Save
Note: The maximum length of a file name is 8 characters.
Press Enter Character to select a character
position in the file name. Turn the entry knob to
select a character. Press Delete Character to delete
Waveforms
BMP Save as BMP files
CSV Save as CSV files
----
----
---- Save the new file.
----
Save as waveform files
Enter the selected
character and go to the
next character position.
Delete the selected
character.
Return to the New menu.
Press SAVE/LOAD→External Storage→New→New Foler
to display the New Folder menu.
Softkey Options Description
Enter
Character
Delete
Character
Save
----
----
---- Save the new folder.
----
Enter the selected
character and go to the
next character position.
Delete the selected
character.
Return to the New
menu.
the selected character. Press Enter Character to
enter the selected character and go to the next
character position.
111
112
Basic Operation
Basic Operation
Press SAVE/LOAD→External Storage→Rename to display
the Rename menu.
Softkey Options Description
Enter
Character
Delete
Character
OK
----
----
----
----
Enter the selected
character and go to the
next character position.
Delete the selected
character.
Rename the selected file
or folder.
Return to the
EXTERNAL menu.
Press SAVE/LOAD→External Storage→Delete to display
the Delete menu.
Softkey Options Description
OK
Cancel
----
----
----
Confirm the deletion of
the selected file or folder.
Cancel the deletion
operation.
Return to the
EXTERNAL menu.
113
114
Basic Operation
Basic Operation
Firmware Update
The oscilloscope firmware can be conveniently updated via
the front panel USB host interface.
Press SAVE/LOAD→External Storage to display the
EXTERNAL menu.
Turn the entry knob to select the correct update file. The file
2042A.UPT is selected as shown in the following figure.
Finally, the message “Restart to complete updating” will
be displayed to remind you to restart (power cycle) the
instrument.
If the software update failed, repeat the above procedures to
update again.
Note: The default file extension of the update file is ”.upt”.
Select the correct update file according to the
model of the oscilloscope. The error message
“Incompatible file ” will be displayed if the file is
not the correct one for your oscilloscope.
Note: Power to the oscilloscope must not be turned off
during the updating process. If this happens, you
will have to return the instrument to the factory for
service.
Press the Load softkey to start the update operation. A
loading progress bar and then an updating progress bar will
be displayed and indicate the progress of the update
operation.
115
116
Basic Operation
Basic Operation
CURSOR Menu
You can measure waveform parameters using the cursors.
Cursors are horizontal and vertical markers that indicate
X-axis values (usually time) and Y-axis (usually voltage) on a
selected waveform . The position of the cursors can be
moved by turning the Entry knob.
Cursor Menu key
The oscilloscope provides three kinds of cursor
measurement modes: Manual, Auto and Track.
Manual Cursor Mode
In manual cursor mode, you can move the cursors to
measure the voltage or time on the selected waveform.
Press CURSOR→Mode to display the CURSOR menu and
select the Manual mode. Press the Type softkey to select
Voltage measurement.
Softkey Options Description
Manual Manual cursor measurement
Mode
Source
Type
Y1--
Y2--
∆Y
Auto Auto cursor measurement
Track Track cursor measurement
CH1 Measure CH1
CH2 Measure CH1
MATH Measure MATH
Voltage Measure voltage value
Time Measure time value
Press this softkey to active
Y1, Y2, or both Y1 and Y2
cursors for adjustment.
Current voltage values for Y1
and Y2 are displayed in the
softkey
The difference value between
----
the Y1 and Y2 cursor voltage
values.
117
118
Basic Operation
Basic Operation
Press CURSOR→Mode to display the CURSOR menu and
select the Manual mode. Press the Type softkey to select
Time measurement.
Softkey Options Description
Mode
Source
Type
X1--
X2--
Manual Manual cursor measurement
Auto Auto cursor measurement
Track Track cursor measurement
CH1 Measure CH1
CH2 Measure CH1
MATH Measure MATH
Voltage Measure voltage value
Time Measure time value
Press this softkey to select
X1, X2, or both X1 and X2
cursors for adjustment.
Current time values for X1
Track Cursor Mode
In the track mode, the screen displays two cross hair cursors.
The cross hair of the cursor is positioned on the waveform
automatically. You can adjust the selected cursor's horizontal
position on the waveform by turning the entry knob.
The oscilloscope displays the values of the coordinates on
the screen, either in the softkey area or on the top right
corner in case the menu is off.
and X2 are displayed in the
softkey
∆X is the time difference
value between the X1 and
∆X
----
1/∆X
X2 cursors.
1/∆X is the frequency
“between” the X1 and X2
cursors.
119
120
Basic Operation
Basic Operation
Press CURSOR→Mode to display the CURSOR menu and
select the Track mode.
Softkey Options Description
Manual Manual cursor measurement
Mode
Cursor A
Cursor B
Ax--
Ay--
Auto Auto cursor measurement
Track Track cursor measurement
CH1 Track CH1 with Cursor A
CH2 Track CH2 with Cursor A
None Turn off Cursor A
CH1 Track CH1 with Cursor B
CH2 Track CH2 with Cursor B
None Turn off Cursor B
Press this softkey to select
cursor A for adjustment. The
current X, Y axis values for
the tracking point of cursor A
Auto Cursor Mode
The Auto mode cursors are displayed only when the auto
measurement function is enabled. The oscilloscope displays
the auto cursors corresponding to the latest auto
measurement parameter. The Auto cursors will not be
displayed when no auto measurement parameter is selected.
are displayed in the softkey.
Press this softkey to select
Bx--
By--
cursor B for adjustment. The
current X, Y axis values for
the tracking point of Cursor B
are displayed in the softkey.
121
122
Basic Operation
Basic Operation
DISPLAY Menu
Display Menu key
Press the DISPLAY menu key to display the DISPLAY menu
page 1/2.
Softkey Options Description
Vector mode draws a line
Vector
Type
Dots
Grid
Contrast
between adjacent sample
points in the waveform.
Dot mode displays only
the sample points.
Display both grids and
axes.
Turn off the axes.
Turn off the grids.
Turn off both grids and
axes.
Adjust the display contrast
and brightness.
Color
Setup
More 1/2
123
124
---- Select the color scheme.
---- Display menu page 2/2.
Basic Operation
Basic Operation
Press More 1/2 softkey to display the DISPLAY menu page
2/2.
Softkey Options Description
The scope updates
the waveform without
ON
Persist
erasing the
previously-displayed
waveforms.
Turn off the
OFF
persistence function
Press the softkey to
Clear
Persistence
----
erase the previous
sample points.
Select the color
Normal
display.
Waveforms
Select the
Monochrome
monochrome display.
Display menu page
More 2/2
---1/2.
RUN Controls
Run controls
Press the SINGLE key to execute a single-shot acquisition.
The key will illuminate in orange until the oscilloscope is
triggered. When the oscilloscope is triggered, the waveform
will be acquired and displayed, after which the oscilloscope
will ignore any further triggers.
The RUN/STOP key is used to manually start/stop the
oscilloscopes acquisition system. In run mode, the
oscilloscope will start looking for a trigger and the
125
126
Basic Operation
RUN/STOP key be green. When the trigger mode is set to
Normal mode, the display will not update until a trigger is
found. If the trigger mode is set to Auto mode, the
oscilloscope looks for a trigger, and if no trigger is found, it
will be triggered automatically and the input signal(s) will be
displayed immediately.
Press the RUN/STOP key again to stop acquiring data. The
RUN/STOP key will be a red color. Now you can pan across
and zoom in on the acquired waveform.
Basic Operation
127
128
Application Examples
Application Examples
Press the MEASURE key to display the MEASURE
menu.
3. Application Examples
This section presents 7 typical application examples. These
examples highlight the features of the oscilloscope and give
you ideas on how to solve your own test problems.
Make Simple Measurements
You need to measure the amplitude and frequency of an
unknown signal on CH1.
Perform following steps to quickly display the signal.
Connect the channel 1 probe to the unknown signal.
Press the AUTO key.
The oscilloscope automatically sets the vertical, horizontal,
and trigger controls. You can adjust any of these controls
manually if you need to optimize the display of the waveform.
When you are using both CH1 and CH2 channels, the
Autoset function sets the vertical controls for each channel
Press the Voltage softkey to display the VOLTAGE
menu.
Press the Amplitude softkey to measure the amplitude.
The amplitude value will be displayed at the bottom of
the screen.
Press the MEASURE key again to display the
MEASURE menu.
Press the Time softkey to display the TIME menu.
Press the Frequency softkey to measure the frequency.
The frequency value will be displayed at the bottom of
the screen to the right of the voltage value.
and uses the CH1 channel to set the horizontal and trigger
controls.
The oscilloscope can automatically display most signals.
Perform following steps to measure signal amplitude and
frequency.
129
130
Application Examples
Application Examples
Capture a Single-Shot Signal
A digital storage oscilloscope can easily capture an
unrepeated signal:
Connect the channel 1 probe to the unknown signal.
Press the trigger MENU key to display the TRIGGER
menu.
Press the Source softkey to select CH1.
Press the Mode softkey to select the Auto trigger mode.
Set the vertical and horizontal controls to what you
expect to be appropriate for the signal.
Press the Type softkey from the TRIGGER menu page
1/2 to select the pulse trigger type.
Press the More 1/2 sofkey to display the TRIGGER
menu page 2/2.
When trigger condition is met, the captured waveform is
displayed, the SINGLE key is not lit, and the RUN/STOP
key is illuminated in red.
Press the Mode softkey to select the normal trigger
mode.
Press the More 2/2 sofkey to display the TRIGGER
menu page 1/2.
Press the Pulse Mode softkey to select (positive
less than).
Rotate the entry knob (
Press the SINGLE key to start the search for the trigger
event. The SINGLE key is illuminated in orange
) to set up the pulse width.
131
132
Application Examples
Application Examples
Reduce the Random Noise on a Signal
If the test signal is noisy, you can set up the oscilloscope to
reduce the noise on the displayed waveform. First, you
stablize the displayed waveform by removing the noise from
the trigger path. Second, you reduce the noise on the
displayed waveform.
Connect a signal to the oscilloscope. Press the AUTO
key to display the signal.
Press the Trigger MENU key to display the TRIGGER
menu.
Press the Type softkey to select Edge trigger type.
Press the Coupling to select HF Reject or LF Reject
coupling mode to reduce the noise from the trigger
channel.
Press the ACQUIRE key to display the ACQUIRE menu.
Press the Mode softkey to select Average mode.
Rotate the entry knob (
that best eliminates the noise from the displayed
waveform.
) to set the number of averages
133
134
Application Examples
Trigger on a Video Signal
The video trigger can be used to capture standard video
signals. The trigger circuit detects the vertical and horizontal
intervals of the waveform and produces triggers based on the
video trigger settings you have selected.
Trigger on Odd or Even Fields of the Video Signal
Press the Trigger MENU key to display the TRIGGER
menu.
Press the Type softkey to select the Video trigger mode.
Application Examples
Press the Source softkey to select CH1 (this assumes
the signal is on channel 1).
Press the Polarity softkey to select negative polarity.
Press the Sync softkey to select Odd Field or Even
Field.
135
Trigger on a Specific Line or All Lines of the Video Signal
Press the Trigger MENU key to display the TRIGGER
menu.
Press the Type softkey to select the Video trigger mode.
Press the Source softkey to select CH1.
Press the Polarity softkey to select negative polarity.
Press the Sync softkey to select Line # or All Lines.
136
Application Examples
Application Examples
PASS/FAIL Measurement
The oscilloscope measures and compares the input signal
with a predefined Pass/Fail mask. If the input signal is within
the mask threshold, a PASS verdict will be generated. If the
input signal lies outside the threshold, a FAIL verdict will be
generated.
Perform the following steps to make a PASS/FAIL
measurement.
Press the UTILITY key to display the UTILITY menu
page 1/2.
Press the More 1/2 softkey to display the UTILITY menu
page 2/2.
Press the Pass/Fail softkey to display the PASS/FAIL
menu.
Press the Enable Test softkey to turn on the PASS/FAIL
measurement.
Press the Setup Mask softkey to display the MASK
menu.
Press the X Mask softkey and then rotate the entry knob
to setup the horizontal threshold.
Press the Y Mask softkey and then rotate the entry knob
to setup the vertical threshold.
Press the Create Mask softkey to update the thresholds.
Press the
137
138
softkey to return to the PASS/FAIL menu.
Application Examples
Application Examples
Press the More 1/2 softkey to display the PASS/FAIL
menu page 2/2.
Press the Msg Display softkey to display the Pass/Fail
measurement results on the top left corner of the screen.
Press the Output softkey to set how the results will be
indicated.
Press the More 2/2 to display the PASS/FAIL menu
page 1/2.
Press the Operate softkey to start the PASS/FAIL
measurement.
Waveform Recorder
The waveform recorder lets you monitor and analyze long
term signal behavior by recording data continuously over long
periods of time then playing it back for post acquisition
analysis. Data is recorded in a sequence of up to 1000
frames of 4k data points each and the time interval between
each frame is adjustable from 1ms – 1000s.
Perform the following steps to record waveforms.
Press the ACQUIRE key to display the ACQUIRE menu.
Press the RECORD softkey to display the RECORD
menu.
Press the Mode softkey to select Record mode.
Press the Source softkey to select the source channel
CH1.
Press the Operate key to start recording. The total
recorded frame count is displayed on the top left screen.
139
140
Application Examples
Application Examples
Press the Operate softkey to playback the waveform.
Perform the following steps to playback the waveforms.
Press the ACQUIRE key to display the ACQUIRE menu.
Press the RECORD softkey to display the RECORD
menu.
Press the Mode softkey to select Play back mode.
Press the Play Mode softkey to select or
mode.
Press the Start Framesoftkey and turn the entry knob to
set the start frame.
Press the End Frame softkey and turn the entry knob to
set the end frame.
Press the Interval softkey and turn the entry knob to set
the interval time.
141
142
Application Examples
Application Examples
Perforxm the following steps to save or recall the waveform
recorded.
Press the ACQUIRE key to display the ACQUIRE menu.
Press the RECORD softkey to display the RECORD
menu.
Press the Mode softkey to select Save/Recall mode.
Press the Start Framesoftkey and turn the entry knob to
set the start frame.
Press the End Frame softkey and turn the entry knob to
set the end frame.
Press the Internal Storage softkey to save or load the
recorded waveform from the internal memory.
Cursor Measurements
You can use the cursors to quickly take time and voltage
measurements on a waveform. You can use the cursors to
measure the amplitude and frequency of a FFT waveform.
You can also use the cursors to measure the phase
difference between two signals with the same frequency
when X-Y horizontal mode is selected.
Measure the time and amplitude characteristics of a
waveform
Perform the following steps to take time and frequency
measurements.
Press the CURSOR key to display the CURSOR menu.
Press the Mode softkey to select the Manual mode.
Press the Type softkey to select the Time type.
Press the X1--/X2— softkey or press the entry knob
to select the X1 cursor.
Rotate the entry knob
Press the X1--/X2— softkey or press the entry knob
to select the X2 cursor.
Rotate the entry knob
ΔX and 1/ΔX are displayed in the softkey area. ΔX is the
time difference between X1 and X2; 1/ΔX is the
frequency “between” X1 and X2.
143
144
to move the X1 cursor.
to move the X2 cursor.
Application Examples
Application Examples
Perform the following steps to make a voltage measurement
on a waveform.
Press the CURSOR key to display the CURSOR menu.
Press the Mode softkey to select the Manual mode.
Press the Type softkey to select the Voltage type.
Press the Y1--/Y2—softkey or press the entry knob
to select the Y1 cursor.
Rotate the entry knob
Press the Y1--/Y2—softkey or press the entry knob
to select the Y2 cursor.
Rotate the entry knob
The ΔY displayed in the softkey area is the voltage
difference between Y1 and Y2.
to move the Y1 cursor.
to move the Y2 cursor.
Measure the frequency and amplitude of an FFT signal
Perform the following steps to take frequency measurements.
Press the MATH key to display the Math menu.
Press the Operate softkey to select FFT and display the
FFT menu.
Press the CURSOR key to display the CURSOR menu.
Press the Mode softkey to select the Manual mode.
Press the Source softkey to select FFT.
Press the Type softkey to select the Time type.
Press the X1--/X2— softkey or press the entry knob
to select the X1 cursor.
Rotate the entry knob
Press the X1--/X2— softkey or press the entry knob
to select the X2 cursor.
Rotate the entry knob
The ΔX displayed in the softkey area is the frequency
to move the X1 cursor.
to move the X2 cursor.
difference between X1 and X2. 1/ΔX is the time
difference between X1 and X2.
Perform the following steps to take amplitude measurements
on an FFT signal
Press the MATH key to display the Math menu.
Press the Operate softkey to select FFT and display the
FFT menu.
145
146
Application Examples
Application Examples
Press the CURSOR key to display the CURSOR menu.
Press the Mode softkey to select the Manual mode.
Press the Source softkey to select FFT.
Press the Type softkey to select the Voltage type.
Press the Y1--/Y2—softkey or press the entry knob
to select the Y1 cursor.
Rotate the entry knob
Press the Y1--/Y2—softkey or press the entry knob
to select the Y2 cursor.
Rotate the entry knob
The ΔY displayed in the softkey area is the voltage
difference between Y1 and Y2.
to move the Y1 cursor.
to move the Y2 cursor.
Measure the Phase Difference Between Two Signals of
the Same Frequency under X-Y Display Mode.
Connect a sine wave signal to channel 1 and a sine
wave signal of the same frequency but different phase to
channel 2.
Press the horizontal MENU key to display the
Horizontal menu.
Press the X-Y softkey to select the X-Y display mode.
Center the signal on the display with the vertical position
knob of each channel.
Use the vertical scale control knob of each channel to
expand the signal for convenient view.
Press the CURSOR key to display the CURSOR menu.
Press the Mode softkey to select the Manual mode.
Press the Source softkey to select CH2.
Press the Type softkey to select the Voltage type.
Press the Y1--/Y2— softkey or press the entry knob
to select the Y1 cursor.
Rotate the entry knob
top of the signal.
Press the Y1--/Y2— softkey or press the entry knob
to select the Y2 cursor.
Rotate the entry knob
bottom of the signal.
147
148
to move the Y1 cursor to the
to move the Y2 cursor to the
Application Examples
The ΔY displayed in the softkey area is the voltage
difference D (or 2B) between Y1 and Y2.
Press the Y1--/Y2—softkey or press the entry knob
to select the Y1 cursor.
Rotate the entry knob
upper intersection of the signal and the Y axis.
Press the Y1--/Y2—softkey or press the entry knob
to select the Y2 cursor.
Rotate the entry knob
lower intersection of the signal and the Y axis.
The ΔY displayed in the softkey area is the voltage
difference C (or 2A) between Y1 and Y2.
Calculate the phase difference using the formulas
C
arcsin
or
D
to move the Y1 cursor to the
to move the Y2 cursor to the
A
arcsin
.
B
149
System Message and General Problems
System Message and General Problems
Empty storage memory: This message will be displayed
when you try to load a file which does not exist from the
4. System Message and
General Problems
System Message
Function is not available: The control knob, key, or softkey
is not available under a specific operating condition.
The control is at its limit: This message will be displayed
when the maximum or minimum value is reached by turning
the Entry knob, Vertical Control knobs, Horizontal Control
knobs, or Trigger Level knob.
Total is at its maximun: This message will be displayed
when the maximum value of Total count for PASS/FAIL is
reached.
Record is completed: This message will be displayed when
the number of waveforms (set in the End Frame softkey)
have been recorded.
internal memory.
Unrecognized file: This message will be displayed when
you try to load a file which cannot be recognized by the
oscilloscope.
Update failed: This message will be displayed when a
software update has failed.
No record data: This message will be displayed when you
try to save a record file without record data.
Record is failed: This message will be displayed when
waveform record is failed.
Fatory setup is recalled: This message will be displayed
when the default factory configuration is recalled.
No signal is found: This message will be displayed when
you press the AUTO key without any signal connected to
either channel.
Invalid data: This message will be displayed when you try to
save a *.CSV file without any valid CSV data.
No external memory: This message will be displayed when
you try to save a file to an external mass storage device
which has not been installed.
Save error: This message will be displayed when a failure to
save a file to the internal or external memory occurs.
151
Load finished: This message will be displayed when a file
has been successfully loaded from the internal or external
memory.
Save finished: This message will be displayed when a file
has been successfully saved to the internal or external
memory.
152
System Message and General Problems
System Message and General Problems
Incompatible file: This message will be displayed when the
update software is not compatible with the model type.
Load error: This message will be displayed when you fail to
load a file from the internal or external memory.
Restart to complete updating: This message will be
displayed to let you restart (power cycle) the oscilloscope
when the software update is successfully finished.
USB device is installed: This message will be displayed
when a USB device is connected and recognized by the
oscilloscope.
USB device is removed: This message will be displayed
when a USB device is removed from the oscilloscope.
Print finished: This message will be displayed when the
current waveform is printed successfully.
Print failed: This message will be displayed when the current
waveform is not printed successfully.
USB host error: This message will be displayed when the
USB host control circuit is not working correctly.
Setup finished: This message will be displayed when the
General Problems
If there is no display on the screen:
Check that the power cord is connected to the
oscilloscope and to a live power source.
Check that the power switch is on.
Check that the display contrast is adjusted properly.
Contact B+K Precision if there is still no display.
If there is no waveform displayed:
Check that the oscilloscope probe id properly connected
to the BNC connector and that the probe clips make
good contact with the probe lead wires.
Check that the probe clips are securely connected to
points in the circuit under test and that the probe ground
clip is connected to the circuit ground.
Check that the circuit under test is powered on.
Press the AUTO key again.
date and time are set successfully.
Setup failed: This message will be displayed when the date
and time are not set successfully.
153
If the waveform display is not stable:
Check that the trigger source channel is actually the
channel to which the signal is connected.
Check that the proper trigger type is selected. The video
type is only used to trigger on a video signal. The proper
trigger type is essential for a stable display.
154
System Message and General Problems
Try using HF Reject or LF Reject to reduce the noise of
the trigger signal.
If the amplitude is not identical with the actual voltage:
Check that the attenuation factor of the probe
corresponds to the attenuation factor set in the channel
menu.
155
Specifications
Specifications
2 mV/div to 5 mV/div range:
5. Specifications and
Characteristics
Characteristics
The characteristics given below are typical performance
values and are not warranted. Characteristics are valid after
a 30 minute warm-up time and within ±5 ℃ of the last
self-calibration temperature.
Vertical system
Scope channels 2 channels plus external trigger input.
60 MHz: 2534, 2540
Bandwidth
Calculated rise time
(=0.35/bandwidth)
100 MHz: 2542
5.83 ns: 2534, 2540
3.50 ns: 2542
DC Measurement
Accuracy
Position range ±8 divisions from the center of the screen
Attenuation factor ×1, ×10, × 100, × 1000
Channel common
mode rejection
Lower frequency
limit, AC coupled
Channel to
channel crosstalk
Input Impedance
Maximum input CAT II, 400Vpk
Differential delay
±(4% × reading + 0.1 × V/div + 0.5 mV)
10 mV/div to 5 V/div range:
±(3% × reading + 0.1 × V/div + 1.0 mV)
100:1 at 60Hz
20:1 at 10MHz
≤5Hz at BNC
≤1Hz when using a 10X passive probe
≥100:1 at 1MHz
≥100:1 at 10MHz
Source CH1, CH2, EXT, EXT/5, AC Line, Alternating.
Modes Auto, Normal, Single
Coupling DC, AC, LF-Reject, HF-Reject
Type Edge, Pulse, Video
Trigger sensitivity,
Edge Trigger Type,
DC coupling
Trigger sensitivity,
Edge Trigger Type,
AC coupling
Trigger sensitivity,
Edge Trigger Type,
LF REJ
Trigger sensitivity,
Edge Trigger Type,
HF REJ
CH1,CH2: 1div from DC to 10MHz
EXT: 100mV from DC to 10MHz
200mV from 10MHz to full Bandwidth
EXT/5: 500mV from DC to full Bandwidth
Same as DC coupling at 50 Hz and above
Same as the DC coupling limits for frequencies
above 100kHz, attenuates signals below 8kHz
Same as the DC coupling limits from DC to 10kHz,
attenuates signals above 150kHz
Time measurement
Math
Cursors Manual, Automatic and Track
Counter
-Duty, Rise time, Fall time, Delay, Phase, X at
MAX, X at MIN
CH1-CH2, CH1+CH2, CH1×CH2, FFT (2k
points)
Built-in 5-digit frequency counter. Count up to
the oscilloscope’s maximum bandwidth.
159
Trigger sensitivity,
Video Trigger Type
Signal formates
and Field Rates,
Video Trigger Type
Trigger level range ±8 divisions from screen center
160
Internal: Pk to Pk amplitude of 2 divisions
EXT: 400 mV
EXT/5: 2V
Supports NTSC, PAL, and SECAM broadcast
systems for any field or any line
Specifications
EXT: ±1.6V
EXT/5: ±8V
Holdoff Range 100ns to 1.5s
Trigger Level
Accuracy
Input impedance
Maximum input CAT II, 400V
SET LEVEL TO 50%
Pulse Width
Trigger mode
Pulse Width Range 20ns to 10s
Internal: ±0.3 div×volts/div
1MΩ±1%||19pF
Operates with input signal ≥50 Hz.
Trigger when Less than, Greater than, Equal
Positive pulse , Negative pulse
@1MΩ
pk
Storage and IO
Internal memory
File format
Standard ports
10 setups and waveforms can be saved and
recalled internally.
Setup file (*.STP), Waveform file (*.WFM), BMP
file (*.BMP), CSV file (*.CSV). Files can be
saved with real time stamp (date, time)
USB host
USB device
Acquisition system
Max. real-time
sample rate
Max. equivalent
sample rate
Memory Depth 4000 points
Vertical resolution 8 bits
Sample mode Normal, Average, Peak Detect
Autoset
400 MSa/s: 2534
1 GSa/s: 2540, 2542
40 GSa/s: 2534
50 GSa/s: 2540, 2542
Finds and displays all active channels. Sets edge
trigger mode on lowest channel, sets vertical
sensitivity on for both channels and time base to
display one or five periods. Requires minimum
voltage > 10 mVpp, 0.5% duty and minimum
frequency > 50 Hz.
161
162
Specifications
Display system
Display 5.7 inch (145cm) diagonal STN LCD.
Resolution 240 vertical by 320 horizontal pixels
Color 256 VGA colours
Brightness Adjustable
Language Multi language user interface
Menu ON:
8 vertical by 10 horizontal divisions
or 200 vertical by 250 horizontal pixels
Display area
Menu OFF:
8 vertical by 12 horizontal divisions
or 200 vertical by 300 horizontal pixels
Display mode Vector, Dots
Interpolation Sinx/x, Linear
Persistence Off, Infinite persistence
Power and environmental requirments
Line voltage 99V to 242VAC
Line frequency 47Hz to 440Hz
Power consumption Less than 50VA
Operating temperature
Non-operating temperature
Humidity
Operating altitude
Non-operating altitude
0C to40C
-20C to 55C
≤90% relative humidity
≤3000m
≤15000m
Physical size and Weight
Instrument height 147 mm
Instrument width 310 mm
Instrument depth 269 mm
Compartment height 57 mm
Compartment width 225 mm
Compartment depth 189 mm
Net weight Approximately 3.6kg
Calibration interval
Recommended calibration interval One year
163
164
Appendix A: Service and
Warranty Information
ServiceInformation
Warranty Service: Please return the product in the original packaging
with proof of purchase to the address below. Clearly state in writing the
performance problem and return any leads, probes, connectors and
accessories that you are using with the device.
Non-Warranty Service: Return the product in the original packaging to
the address below. Clearly state in writing the performance problem and
return any leads, probes, connectors and accessories that you are using
with the device. Customers not on open account must include payment in
the form of a money order or credit card. For the most current repair
charges please visit www.bkprecision.com
Return all merchandise to B&K Precision Corp. with pre-paid shipping.
The flat-rate repair charge for Non-Warranty Service does not include
return shipping. Return shipping to locations in North American is
included for Warranty Service. For overnight shipments and non-North
American shipping fees please contact B&K Precision Corp.
Include with the returned instrument your complete return shipping
address, contact name, phone number and description of problem.
and click on “service/repair”.
LimitedTwo‐YearWarranty
B&K Precision Corp. warrants to the original purchaser that its products
and the component parts thereof, will be free from defects in
workmanship and materials for a period of 2 years from date of purchase.
B&K Precision Corp. will, without charge, repair or replace, at its option,
defective product or component parts. Returned product must be
accompanied by proof of the purchase date in the form of a sales receipt.
To obtain warranty coverage in the U.S.A., this product must be
registered by completing a warranty registration form on
www.bkprecision.com
Exclusions: This warranty does not apply in the event of misuse or
abuse of the product or as a result of unauthorized alterations or
repairs. The warranty is void if the serial number is altered, defaced
or removed.
B&K Precision Corp. shall not be liable for any consequential damages,
including without limitation damages resulti ng f rom loss of use. Some
states do not allow limitations of incidental or consequential damages. So
the above limitation or exclusion may not apply to you.
This warranty gives you speci fic rights and you may have other rights,
which vary from state-to-state.