Model 2530 25 MHz, 250 MSa/s
Model 2532 40 MHz, 500 MSa/s
General Safety Summary
Review the following safety precautions to avoid injury and prevent damage to this product or any
products connected to it. To avoid potential hazards, use this product only as specified.
Only qualified personnel should perform service procedures.
To Avoid Fire or Personal Injury
Use Proper Power Cord. Use only the power cord specified for this product and certified for the country
of use.
Connect and Disconnect Properly. Do not connect or disconnect probes or test leads while they are
connected to a voltage source.
Ground the Product. This product is grounded through the grounding conductor of the power cord. To
avoid electric shock, the grounding conductor must be connected to earth ground. Before making
connections to the input or output terminals of the product, ensure that the product is properly grounded.
Connect the Probe Properly. The probe ground lead is at ground potential. Do not connect the ground
lead to an elevated voltage.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings and marking on the
product. Consult the product manual for further ratings information before making connections to the
product.
Do Not Operate Without Covers. Do not operate this product with covers or panels removed.
Avoid Exposed Circuitry. Do not touch exposed connections and components when power is present.
Do Not Operate With Suspected Failures. If you suspect there is damage to this product, have it
inspected by qualified service personnel.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
Symbols and Terms on the Product.
Protective Ground (Earth) Terminal.
CAUTION. Refer to user manual. Caution indicates a hazard to property including this product
2
Introduction
Description
The 2530 & 2532 Digital Storage Oscilloscope is a light-weight benchtop oscilloscope for viewing
waveforms and taking measurements. The 2530/2532 is an ideal education and training tool and also
well suited for applications in service and repair.
Key Features
40 MHz bandwidth, 500MSa/s sample rate with color display (model 2532)
25 MHz bandwidth, 250MSa/s sample rate with monochrome display (model 2530)
4000 point record length for each channel
One touch automatic setup for ease of use (Auto)
Save/Recall setup and waveform data
Capture, save and analyze waveform data and measurement results with the included EasyScope
Application Software
Cursors with readouts
Eleven automatic measurements
Waveform averaging and peak detection
Math functions: +, -, / and × operations
Fast Fourier Transformation (FFT)
Extensive Trigger capabilities: Pulse Width trigger, Video trigger capability with line-selectable
triggering, External trigger
Variable persistence display
Built in trigger frequency counter
Appendix C: Sample Rate versus Timebase Setting ..................... 66
Appendix D: Daily Maintenance and Cleaning ............. 66
Appendix E: Index ........................................................................68
Appendix F: Service and Warranty Information ...........................69
5
Chapter 1 Getting Started
This Chapter describes the following:
◆ Installation
◆ Basic functional check
◆ Probe check and compensation
◆ Setting of probe attenuation factor
◆ Self calibration procedure
1.1 Installation
Inspection
Inspect the shipping container and the product for damage. In case of any damage, notify your authorized
B+K Precision distributor. Make sure all the accessories as described in the specification chapter are
included.
Power Source
Connect your instrument to an AC power outlet delivering 100 – 240V, 47 – 440 Hz. Use the power cord
provided or a comparable cord with an Earth ground connector.
Security Loop
The instrument provides a security loop located in the rear panel of the instrument. Use a security cable
or a security lock to secure your instrument to your location.
1.2 Functional check
Perform this check to make sure your instrument is operating correctly.
1. Power On the oscilloscope.
Press the “DEFAULT SETUP” button. Note that the probe default attenuation is 10X.
6
ON OFF
P
/
button
Figure 1- 2
2. Set the switch to 10X on the included probe and connect it to channel 1 on the oscilloscope. To do
this, align the slot in the probe connector with the key on the CH 1 BNC, push to connect, and twist to
the right to lock the probe in place. Connect the probe tip and reference lead to the PROBE COMP
connector.
PROBE COM
CH1
Figure 1-3
3. Press the “AUTO” button. After a few seconds, you should see a square wave signal with a 1KHz
frequency and about 3V peak-peak amplitude
Figure 1- 4
4 Press the “CH1” button two times to remove the signal from CH1. Repeat steps 2) and 3) for CH2.
7
1.3 Probe handling
1.3.1 Probe Safety
A guard around the probebody provides a finger barrier for protection from electric shock.
Figure 1-5
Connect the probe to the oscilloscope and connect the ground terminal to ground before you take any
measurements.
WARNING
● To avoid electric shock when using the probe, keep fingers behind the
● To avoid electric shock while using the probe, do not touch metallic
guard on the probe body.
portions of the probe head while it is connected to a voltage source.
1.3.2 Probe Compensation
Perform this adjustment to match your probe to the input channel. This should be done whenever you
attach a probe to any input channel for the first time.
8
Figure 1-6
1. Set the Probe option attenuation in the channel menu to 10X. Set the switch to 10X on the probe and
connect the probe to channel 1 on the oscilloscope. If you use the probe hook-tip, ensure a proper
connection by firmly inserting the tip onto the probe.
2. Attach the probe tip to the PROBE COMP~3V connector and the reference lead to the PROBE
COMP Ground connector. Display the channel and then push the“AUTOSET” button.
3. Check the shape of the displayed waveform.
Over compensated Correctly Compensated Under compensated
Figure1-7
4. If necessary, use a non-metallic tool to adjust the trimmer capacitor on the probe for to resemble the
Correctly Compensated wave shape as much as possible.
Repeat as necessary.
1.3.3 Probe Attenuation Setting
Probes are available with various attenuation factors which affect the vertical scale of the signal. If you
change the Attenuation switch on the probe, you also need to change the attenuation setting on the
oscilloscope accordingly.
Note. The default setting for the Probe option is 10X.
Be sure that the Attenuation switch on the probe matches the Probe option in the oscilloscope. Switch
9
settings are 1X and 10X.
Note. When the Attenuation switch is set to 1X, the probe limits the bandwidth
of the oscilloscope to 6MHz. To use the full bandwidth of the oscilloscope, be
sure to set the switch to 10X.
1.4 Self Calibration
The self calibration routine lets you optimize the oscilloscope’s signal path for maximum measurement
accuracy. You can run the routine at any time but should always run the routine if the ambient temperature
changes by 5℃ or more.
To compensate the signal path, disconnect any probes or cables from the front-panel input connectors.
Then, push the UTILITY button, select the “Do Self Cal” option and follow the directions on the screen.
10
Chapter 2 Main Control sections overview
The front panel is divided into easy-to-use functional areas. This chapter provides you with a quick
overview of the main controls.
Figure 1- 1 Front Panel Controls
To use your oscilloscope effectively, you need to familiarize yourself with the following control sections:
◆ Vertical System
◆ Horizontal System
◆ Trigger System
◆ Acquisition System
◆ Display System
◆ Measurement System
◆ Utility Menu
◆ Store/Recall Menu
2.1 Vertical System
Each channel has a vertical control menu which appears after pressing the CH1 or CH2 button. The
vertical controls are used to size and position the signals and to turn the waveform on or off. Below is a
figure of the vertical system controls.
11
`
Figure 2-1
Table 2-1 Vertical Control menu
Menu Settings Explanation
DC
Coupling
BW limit
Volts/Div
Probe
Invert
Vertical POSITION knob :Use the VERTICAL POSITION knobs to move waveforms up or down on
AC
GND
On/Off Limits the bandwidth to 20 MHz, in order to reduce display nose and to filter out
Coarse
Fine
1X
10X
100X
1000X
on
off
Both AC and DC components of the input signal are passed on to the oscilloscope
The DC component is blocked, only the AC component of the input signal is passed
through. Signals below 10Hz are attenuated
The waveform is disconnected from the oscilloscope
unwanted high frequency components
Selects the resolution of the Volts/Div knob. Coarse defines a 1-2-5 sequence.
Fine changes the resolution to small steps between the coarse settings
Set to match the type of probe you are using to ensure correct vertical readouts
Waveform inverted
Invert function is off, original waveform is displayed.
the screen.
Vol t s /D iv kn ob :Use the VOLTS/DIV knobs to control how the oscilloscope amplifies or attenuates the
source signal. When you turn the VOLTS/DIV knob, the oscilloscope increases or decreases the vertical
size of the waveform on the screen with respect to the ground level. Pressthe“Volt/div” knob to toggle
between “Coarse” and“ Fine” adjustment of the Volt/Div setting.
GND Coupling. Use GND coupling to display a zero-volt waveform. Internally, the channel input is
connected to a zero-volt reference level.
Fine Resolution:The vertical scale readout displays the actual Volts/Div setting while in the fine
resolution setting. Changing the setting to coarse does not change the vertical scale until the VOLTS/DIV
control is adjusted.
12
Remove the Waveform from the display: Push the CH1 or CH2 button to display the respective vertical
menu. Push CH1 or CH2 button again to remove the waveform.
Note:
●If DC coupling is active,you can quickly measure the DC component of the signal by
simply noting its distance from the ground symbol.
●If the channel is set to AC coupling,the DC component of the signal is blocked allowing
you to use greater sensitivity to display the AC component of the signal.
2.2 Horizontal System
The horizontal controls change the horizontal scale and position of waveforms. The horizontal center of
the screen is the time reference for waveforms. Changing the horizontal scale causes the waveform to
expand or contract around the screen center. The horizontal position knob changes the position of the
trigger point relative to the center of the screen. Below is a figure of the front panel horizontal systems
controls.
Figure 2- 2
Table 2- 2 Horizontal Control Menu:
Menu Explanation
Main Display the waveform
Win Zone Define a window (zoom window) by adjusting the cursor position using the
Horizontal control knobs
Window Expand the waveform defined in Win Zone to full screen width
Holdoff Using the Adjust knob to adjust the holdoff time
Holdoff Reset Reset the holdoff time to 100ns
The center axis of the vertical scale is GND level. The readout on the top right corner, shows the
horizontal position. “M” stands for main time base,“W” stands for window time base (Zoom function,
horizontal menu)。The arrow on the top center scale shows the vertical position.
13
Figure 2-3
Horizontal POSITION Knob: Used to control the position of the trigger relative to the center of the
screen. When the trigger point (downward arrow) moves outside the acquired waveform, (reaches the left
or right edge of the screen) the instrument switches automatically to delayed sweep, indicated by the arrow
pointing left or right instead of downwards. In delayed sweep mode, you can use the horizontal Time/Div
knob to acquire more detail around the center of the screen.
Time/Div Knob: Used to change the horizontal time scale to magnify or compress the waveform. If
waveform acquisition is stopped (using the RUN/STOP or SINGLE button), the SEC/DIV control
expands or compresses the waveform;Pressing the Time/Div knob resets the horizontal position to
zero.
Window Zone: Define the zoom area of a waveform section. The Window time base setting cannot
be set to a value slower than the main time base setting.
Window:Press to expand the zoom area to cover the whole screen.
Holdoff:You can use Holdoff to produce a stable display of complex waveforms. During the
holdoff time the oscilloscope will not trigger until the holdoff time has expired. For a pulse train,
you can adjust the holdoff time so the oscilloscope triggers only on the first pulse in the train.
Trigger position
Figure 2-4
Holdoff time
14
Trigger level
Note:
• When you change between the Main, Window Zone, and Window views,
the oscilloscope erases any waveform which was saved on the screen
using persistence.
• Use holdoff to help stabilize the display of aperiodic waveforms.
2.3 Trigger System
The trigger determines when the oscilloscope starts to acquire data and display a waveform. The
oscilloscope acquires data while waiting for the trigger condition to occur. After it detects a trigger, the
oscilloscope continues to acquire enough data so that it can draw the waveform on the display. Below are
the trigger section controls:
Figure 2-5
“TRIG MENU” Button:press the “TRIG MENU” Button to display the TRIG MENU
Figure 2-6
15
LEVEL Knob:Set the trigger level
SET TO 50% Button:Use the SET TO 50% button to stabilize a waveform quickly. The
oscilloscope automatically sets the Trigger Level about halfway between the minimum and maximum
voltage levels. This is useful when you connect a signal to the EXT TRIG BNC and set the trigger
source to Ext or Ext/5.
FORCE Button:Use the FORCE TRIG button to complete the current waveform acquisition
regardless of whether or not the oscilloscope detects a trigger. This is useful for SINGLE acquisitions
and Normal trigger mode.
2.3.1 Signal Source
You can use the Trigger Source options to select the signal that the oscilloscope uses as a trigger. The
source can be any signal connected to the CH1 or CH2 connectors, the EXT TRIG connector or the AC
power line (available only with Edge trigger).
2.3.2 Trigger Type
The 2530 series supports three types of Trigger signals:Edge,Video and Pulse Trigger. Edge trigger can be
used with analog and digital circuits. An edge trigger occurs when the trigger input passes through a
specified voltage level with the specified slope. Pulse trigger is used to find pulses with certain pulse
widths. Video is used to trigger on fields or lines for standard video waveforms.
16
Table 2-3 Edge Trigger Menu :
Menu Setting Explanation
Type Edge
CH1
CH2
EXT
Source
EXT/5
AC Line
Rising
Slope
Falling
Auto
Mode
Normal
With Edge selected, the rising or falling edge of the input signal is
used for the trigger
Triggers on CH1 or CH2 regardless of whether or not the waveform
is displayed。
Uses the signal connected to the EXT TRIG front-panel BNC as
trigger source. Trigger level range is -2.4V to +2.4V
Same as EXT but attenuates the applied trigger signal by a factor of
five. The trigger level range is extended to +12V to -12V.
Uses power line as the trigger source. Trigger coupling is set to DC
and the trigger level is 0 volts.
Trigger to the rising or falling edge
Use the Auto Mode to let acquisition free-run in the absence of a
valid trigger condition.. This mode also supports an untriggered,
scanning waveform at 100 ms/div or slower time base settings
Use this mode when you want to see only valid triggered
waveforms. When you use this mode, the oscilloscope does not
display a waveform until after the first trigger
Coupling
Single
DC Passes all components of the signal
AC Blocks DC components and attenuates signals below 10 Hz
HF Reject Attenuates the high-frequency components above 80 kHz
LF Reject
When you want the oscilloscope to acquire a single waveform,
press the “SINGLE ”button
Blocks the DC component and attenuates the low-frequency
components below 300 kHz
17
Table 2-4 Video Trigger Menu:
Menu Setting Explanation
Type Video When you select Video Trigger mode, set coupling to AC. Now
you can trigger from a NTSC or PAL/SECAM video signal.
CH1
CH2
EXT
EXT/5
Polarity
Sync
Standard
Table 2-5 Pulse Trigger Menu:
A pulse trigger occurs when a pulse is found in a waveform that matches the pulse definition. The When
and Set Width menu buttons control the pulse definition.
Menu Setting Explanation
(Normal)
(Inverted)
Line Num
All lines
Odd field
Even Field
NTSC
Pal/Secam
Select the input source to be the trigger signal. Source
Ext and Ext/5 use the signal applied to the EXT TRIG connector as the
source.
Triggers on the negative edge of the sync pulse
Triggers on the positive edge of the sync pulse
Select on which part of the video signal you want to trigger
Select the video standard for sync and line number count
Type
Source
When
Pulse Select the pulse to trigger the
pulse match the trigger condition
CH1
CH2
EXT
EXT/5
(Positive pulse width less than pulse width setting)
(Positive pulse width larger than pulse width setting)
(Positive pulse width equal to pulse width setting)
(Negative pulse width less than pulse width setting)
(Negative pulse width larger than pulse width setting)
18
Select how to compare the trigger
pulse relative to the value
selected in the Pulse Width
option
(Negative pulse width equal to pulse width setting)
set width
Mode
Coupling
Hint:Use Pulse Width triggering to trigger on aberrant pulses.
20.0ns~10.0s
Auto
Normal
single
DC
AC
HF Reject
LF Reject
Adjust the pulse width with the
ADJUST knob
Select the type of triggering;
Normal mode is best for most
Pulse Width trigger applications
2.3.3 Coupling
Use “Coupling” to select which component of the signal to pass on to the trigger circuit. This parameter
only applies to Edge or Pulse trigger mode, for Video trigger mode coupling is always set to AC.
2.3.4 Position
The horizontal position control establishes the time between the trigger position and the screen center. You
can adjust the HORIZONTAL POSITION control to view waveform data before the occurrence of the
trigger, after the trigger, or any point in between. When you change the horizontal position of a waveform,
you are changing the time between the trigger and the actual center of the display. (This appears to move
the waveform to the right or left on the display.). Use a pretrigger setting to acquire waveforms leading up
to a trigger event. Use a post trigger setting when you want to acquire waveforms that follow an event.
When the trigger point (downward arrow) moves outside the acquired waveform, (reaches the left or right
edge of the screen) the instrument switches automatically to delayed sweep, indicated by the arrow
pointing left or right instead of downwards. In delayed sweep mode, you can use the horizontal time/Div
knob to acquire more detail around the center of the screen.
2.3.5 Slope & Level
The Slope and Level controls help to define the trigger. The Slope option (Edge trigger type only)
determines whether the oscilloscope finds the trigger point on the rising or the falling edge of a signal. The
TRIGGER LEVEL knob controls where on the edge the trigger point occurs.
19
T
r
i
g
g
e
r
l
e
v
e
l
c
a
n
b
e
a
d
j
u
s
t
e
d
v
e
r
t
i
c
a
l
l
y
Rising edge
Falling edge
Figure 2-7
Note:
• Press the SINGLE button when you want the oscilloscope to acquire a single waveform
• Trigger coupling affects only the signal passed to the trigger system. It does not affect the bandwidth
• Normal Polarity Sync triggers always occur on negative-going horizontal sync pulses. If the video
or coupling of the signal displayed on the screen.
waveform has positive-going horizontal sync pulses, use the Inverted Polarity selection.
2.4 Acquisition system
When you acquire a signal, the oscilloscope digitizes the signal and displays the resulting waveform on the
screen. The acquisition mode defines how the signal is digitized, and the time base setting affects the time
span and level of detail in the acquisition. The sampling rate adjusts automatically to a value suitable for
each selected Time/Div scaling factor. Refer to Appendix C for more detailed information.
Push the ACQUIRE button to access the Acquisition menu
Figure 2-8
Acquire button
20
Table 2- 6 Acquire Menu:
Menu Set up Explanation
Sampling Use to acquire and accurately display most waveforms
Peak Detect Use to detect glitches and decrease the possibility of aliasing.
Average Select this mode to reduce the displayed random noise.
Averages 4
16
32
64
128
256
Mode Equ time
Real time
The 3 acquisition modes Sample, Peak Detect and Average:
Specify number of averages
Select realtime or equivalent sampling
Sampling: In this acquisition mode, the oscilloscope samples the signal in evenly spaced intervals to
reconstruct the waveform. This mode accurately represents the signal most of the time.
Advantage
Disadvantage
samples. This can result in aliasing and may cause narrow pulses to be missed. In these cases, you
should use the Peak Detect mode to acquire data.
: You can use this mode to reduce random noise.
: This mode does not capture rapid variations in the signal that may occur between
Figure 2-9 Sampling mode
Peak Detect: Peak Detect finds the maximum and minimum values of a signal.
Advantage
otherwise been missed in Sample mode.
Disadvantage
: In this way, the oscilloscope can acquire and display narrow pulses, which may have
: Noise will appear to be higher in this mode.
21
Loading...
+ 49 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.