Normal use of test equipment exposes you to a certain amount of danger from electrical shock because
testing must often be performed where exposed high voltage is present. An electrical shock causing
10milliampsofcurrentto pass throughtheheart will stop most human heartbeats. Voltageas low as 35 volts
dc or ac rms should be considered dangerous and hazardous since it can produce a lethal current under
certain conditions. Higher voltage poses an even greater threat because such voltage can more easily
produce a lethal current. Your normal work habits should include all accepted practices that will prevent
contact with exposed high voltage, and that will steer current away from your heart in case of accidental
contact with a high voltage. You will significantly reduce the risk factor if you know and observe the
following safety precautions:
1. Don’t expose high voltage needlessly in the equipment under test. Remove housings and covers only when necessary.
Turn off equipment while making test connections in high-voltage circuits. Discharge high-voltage capacitors after
removing power.
2. If possible, familiarize yourself with the equipment being tested and the location of its high voltage points. However,
remember that high voltage may appear at unexpected points in defective equipment.
3. Use an insulated floor material or a large, insulated floor mat to stand on, and an insulated work surface on which to
place equipment; make certain such surfaces are not damp or wet.
4. Use the time-proven “one hand in the pocket” technique while handling an instrument probe. Be particularly careful to
avoid contacting a nearby metal object that could provide a good ground return path.
5. When using a probe, touch only the insulated portion. Never touch the exposed tip portion.
6. When testing ac powered equipment, remember that ac line voltage is usually present on some power input circuits such
as the on-off switch, fuses, power transformer, etc. any time the equipment is connected to an ac outlet, even if the
equipment is turned off.
7. Some equipment with a two-wire ac power cord, including some with polarized power plugs, is the “hot chassis” type.
This includes most recent television receivers and audio equipment. A plastic or wooden cabinet insulates the chassis
toprotect the customer. When the cabinet is removed for servicing, a serious shock hazard exists if the chassis is touched.
Not only does this present a dangerous shock hazard, but damage to test instruments or the equipment under test may
result from connecting the ground lead of most test instruments (including this oscilloscope) to a “hot chassis”. To make
measurements in “hot chassis” equipment, always connect an isolation transformer between the ac outlet and the
equipment under test. The B+K Precision Model TR-110 or 1604A Isolation Transformer, or Model 1653A or 1655A
AC Power Supply is suitable for most applications. To be on the safe side, treat all two wire ac powered equipment as
“hot chassis” unless you are sure it has an isolated chassis or an earth ground chassis.
8. Never work alone. Someone should be nearby to render aid if necessary. Training in CPR (cardio-pulmonary
resuscitation) first aid is highly recommended.
This symbol on oscilloscope means “refer to instruction manual
for further precautionary information”. This symbol appears in the
manual where the corresponding information is given.
®
+
1031 Segovia Circle, Placentia, CA 92870
TABLE OF CONTENTS
PagePage
TEST INSTRUMENT SAFETY . . . . . . inside front cover
B+K Precision’s Model 2522B is one of the lowest cost
digital storage oscilloscopes in the industry, yet it includes
all the basic features needed by most technicians and engineers. Unlike other digital oscilloscopes, it has a full analog
modewithinfiniteresolution.Digitalstoragemodesinclude
refresh, roll, and single sweep. Equivalent time storage
techniques allow storage of repetitive waveforms up to
20 MHz. This oscilloscope is built by and backed by B+K
Precision, a company that has been selling reliable, durable,
value priced test instruments for over 50 years.
DIGITAL STORAGE FEATURES
High Resolution
2048 samples across the width of the screen (200
samplesperdivision)assures good resolution ofstored
waveforms. 1024 samples taken at sweep rates of
10 µs/div to 0.1 µs/div.
High Sampling Rate
Sampling Rate up to 10 Ms/s (Megasamples per second) for storing one-time events down to 50 nS.
Slow Event Display
X100 time base selections extend sweep time to 50
seconds/division. Excellent for displaying slow speed
events.
Equivalent Time Sampling
Allows repetitive waveforms up to 20 MHz to be
stored.
Pretrigger Capture
Permits viewing activity before the trigger occurs.
Selectable 0%, 25%, 50%, or 75% pre-trigger view.
Plotter Output
Analog output plus TTL Pen-Down control for strip
chart recorder or X-Y plotter.
CRT FEATURES
Rectangular CRT
Rectangular CRTwith large 8 x 10 centimeterviewing
area.
Convenience
Tracerotation electrically adjustable from front panel.
0%, 10%, 90%, and 100% markers for rise time measurements.
DUAL TRACEFEATURES
Dual Trace
Model 2522B has two vertical input channels for displaying two waveforms simultaneously. Selectable
single trace (either CH 1 or CH 2) or dual trace.
Alternate or chop sweep selectable at all sweep rates.
Sum and Difference Capability
Permits algebraic addition or subtraction of channel 1
and channel 2 waveforms, displayed as a single trace.
Useful for differential voltage and distortion measurements.
HIGH FREQUENCY FEATURES
Wide Bandwidth
Conservatively-rated −3 dB bandwidth is dc to 20
MHz.
Fast Rise Time
Rise time is less than 18 nS.
Fast Sweep
Maximum sweep speed of 10 nS/div (with X10 MAG)
assures high frequencies and short-duration pulses are
displayed with high resolution.
VERTICAL FEATURES
High Sensitivity
5 mV/div sensitivityfor full bandwidth. High-sensitivity 1 mV/div and 2 mV/div using PULL X5 gain
control.
Calibrated VoltageMeasurements
Accurate voltage measurements (±3%) on 10 calibrated ranges from 5 mV/div to 5 V/ div. Vertical gain
fully adjustable between calibrated ranges.
SWEEP FEATURES
Calibrated Time Measurements
Accurate (±3%) time measurements. The main sweep
has 21 calibrated ranges from 0.5 s/div to 0.1 µS/div.
Sweep time is fully adjustable between calibrated
ranges.
X10 Sweep Magnification
Allows closer examination of waveforms, increases
maximum sweep rate to 10 nS/div.
3
FEATURES
TRIGGERING FEATURES
Two Trigger Modes
Selectable normal (triggered) or automatic sweep
modes.
Triggered Sweep
Sweep remains at rest unless adequate trigger signal is
applied. Fully adjustable trigger level and (+) or (−)
slope.
AUTO Sweep
Selectable AUTO sweep provides sweep without trigger input, automatically reverts to triggered sweep
operation when adequate trigger is applied.
Five TriggerSources
Five trigger source selections, including CH 1, CH 2,
alternate, EXT, and LINE.
Video Sync
Frame(TV V) or Line (TV H) triggering selectablefor
observingcomposite video waveforms.TV-Hposition
can also be used as low frequency reject and TV-V
position can be used as high frequency reject.
Variable Holdoff
Trigger inhibit period after end of sweep adjustable.
Permits stable observation of complex pulse trains.
OTHER FEATURES
X−Y Operation
Channel 1 can be applied as horizontal deflection
(X-axis) while channel 2 provides vertical deflection
(Y-axis).
Built-in Probe Adjust Square Wave
A 0.5 V p-p, 1 kHz square wave generator permits
probe compensation adjustment.
Channel 2 (Y) Output
A buffered 50Ω output of the channel 2 signal is
available at the rear panel for driving a frequency
counter or other instruments. The output is 50 mV/div
(nominal) into 50Ω.
Supplied With TwoProbes
4
SPECIFICATIONS
CRT
Type: 6-inch rectangular with integral graticule, P31
phosphor.
DisplayArea:8 x 10 div (1 div = 1 cm).
AcceleratingVoltage:2kV.
Phosphor:P31.
Trace Rotation: Electrical, front panel adjustable.
VERTICAL AMPLIFIERS (CH 1 and CH 2)
Sensitivity: 5 mV/div to 5 V/div, 1 mv/div to 1 V/div at
X5 MAG.
Attenuator:10 calibrated steps in 1-2-5 sequence.Vernier
control provides fully adjustable sensitivity between
steps; range 1/1 to at least 1/3.
Accuracy:±3%, 5 mV to 5 V/div; 5%, at X5 MAG.
Input Resistance:1MΩ±2%.
Input Capacitance: 25 pF ±10 pF.
Frequency Response:
5 mV/div to 5 V/div:
DC to 20 MHz (−3 dB).
X5 MAG:
DC to 10 MHz (−3 dB).
Rise Time:
18 nS; 35 nS at X5 MAG.
Overshoot:Less than 5%.
Operating Modes:
CH 1:CH 1, single trace.
CH 2:CH 2, single trace.
DUAL:CH 1 and CH 2, dual trace.
Alternate or Chop selectable at
any sweep rate.
ADD:Algebraic sum of CH 1 + CH 2.
Chop Frequency:Approximately 500 kHz.
PolarityReversal:CH 2 invert.
MaximumInput Voltage:400 V (dc + ac peak).
HORIZONTAL AMPLIFIER
(Input through channel 1 input)
X−Y mode:
CH 1 = X axis.
CH 2 = Y axis.
Sensitivity: Same as verticalchannel 2.
Input Impedance: Same as vertical channel 2.
Frequency Response:
DC to 1 MHz (−3 dB).
X-Y Phase Difference: 3° or less at 50 kHz.
MaximumInputVoltage: Same as vertical channel 1.
SWEEP SYSTEM
Time Base: 0.1 µS/div to 0.5 S/div in 1-2-5 sequence,
21 steps. Vernier control providesfully adjustable
sweep time between steps.
Accuracy:±3%, except ±20% on 0.1 µS/div.
Sweep Magnification: X10 ±10%.
Holdoff: Continuously adjustable for main time base from
NORM to 5 times normal.
TRIGGERING
Trigger Modes:
AUTO (free run), NORM, TV-V, TV-H.
Trigger Source:
CH 1, CH 2, Alternate, EXT,LINE.
Slope:
(+) or (–).
Trigger Coupling:
AUTO:Sweep free-runs in absence of
suitable trigger signal.
NORM:Sweep triggered only by adequate
trigger signal.
TV-V:Video vertical sync pulses are
selected. Also usable for high
frequency reject.
TV-H:Video horizontal sync pulses are
selected. Also usable for low
frequency reject.
5
SPECIFICATIONS
Trigger Sensitivity:
AC:1.0 div (internal)
≥0.5 Vp-p (external)
30 Hz – 30 MHz
TV-V:1.0 div (internal)
≥0.5 Vp-p (external)
20 Hz – 30 kHz
TV-H:1.0 div (internal)
≥0.5 Vp-p (external)
3 kHz – 100 kHz
MaximumExternalTriggerVoltage: 300 V (dc + ac
peak).
DIGITAL STORAGE FACILITIES
Storage Word Size:
2048 3 8 bits/channel (2 k/channel with direct sampling, 1 k/channel with equivalent time sampling).
Vertical Resolution:
8 bit (1 in 256) approximately 25 steps/div.
Horizontal Resolution:
1 in 2048, approximately 200 samples/div.
Sampling Rate:
10 M samples/sec to 4 samples/sec, reduced in proportion to time base. Direct sampling at time base settings
of 20 µs/div and slower, equivalent time sampling at
time base settings of 10 µs/div and faster.
Time Base Expander:
For storage of slow time events, time base steps
10 ms/divandslowerhaveselectable 1/1 or 1/100 rate.
1/100 rate expands time base from 1 sec/div to 50
sec/div in 1-2-5 sequence.
Equivqlent Time Sampling Bandwidth:
20 MHz for repetitive waveforms.
Dot Joining:
Linear interpolation between samples.
Digital Display Modes:
Roll:
Stored data and display updated continually.
Refresh:
Storeddataanddisplay updated by triggeredsweep.
Hold:
Freezes channel 1 and channel 2 data immediately.
Save CH 2:
Freezes channel 2 data immediately.
Pretrigger Storage:
Available in single sweepmode, switchable to 0%,
25%, 50%, or 75%.
PLOT OUTPUT
Analog output of the stored display.
CH 1 and CH 2 Outputs:
Selected by PLOT switch. Output via CH1 OUTPUT
and CH2 OUTPUT jacks on rear panel. Amplitude
0.2 V/div (2 V maximum).
Output Sweep Rate:
Output sweep rate is 1/10 of TIME/DIV setting (and
1/100 switch when applicable).
Cal/ProbeCompensation Voltage:0.5 V p-p ±3% square
wave,1 kHz nominal.
CH 2 (Y) Output:
Output Voltage: 50 mV/div (nominal into 50 ohm
load).
Output Impedance: Approximately 50 ohms.
Frequency Response:20 Hz to 20 MHz, −3 dB.
Power Requirements: 100–130 VAC or 200–260 VAC,
50/60 Hz, 55 watts.
Dimensions (H 3 W3 D):
5.2″ 3 12.8″ 3 15.7″
(132 3 324 3 398 mm).
Weight: 18.7 lbs (8.5 kg).
Environment:
Within Specified Accuracy: +10° to +35° C, 10–80%
relative humidity.
Full Operation: 0° to +50° C, 10–80% relative
humidity.
Storage: −30° to +70° C, 10–90% relative humidity.
ACCESSORIES SUPPLIED:
Two Switchable X1/X10 Probes.
Instruction Manual.
AC Line Cord.
6
CONTROLS AND INDICATORS
14
13
5
3
1
4
2
26
25
24
23
2221
20
19
37
38
30
29
31
32
36
33
34
35
9
Fig. 1. Model 2522B Controls and Indicators.
GENERAL FUNCTION CONTROLS
1. ON Indicator. Lights when oscilloscope is “on”.
2. POWER Pushbutton. Turns oscilloscope “on” and
“off”.
3. INTENSITY Control. Adjusts brightness of trace.
4. TRACE ROTATION Control. Adjusts to maintain
trace at a horizontal position.
5. FOCUS Control. Adjusts trace focus.
6. GNDTerminal. Oscilloscope chassis ground
jack, and earth ground via three-wire ac power cord.
7. CAL Terminal. Terminal provides 0.5 V p-p, 1 kHz
(nominal) square wave signal. This signal is useful for
checking probe compensation adjustment, as well as
providing a rough check of vertical calibration.
VERTICAL CONTROLS
8. VERTical MODE Switch. Selects vertical display
mode. Four-position lever switch with the following
positions:
CH1:
Displays the channel 1 signal by itself.
CH2/X-Y:
CH2: displays the channel 2 signal by itself.
10
11
12
6
28
7
8
15
16
17
2718
X-Y: used in conjunction with the X-Y control and
Trigger SOURCE switch to enable X-Y display
mode.
DUAL:
Displays the channel 1 and channel 2 signals simultaneously. Dual-trace mode may be either alternate
orchoppedsweep;see the description underCHan-
nel 1 POSition/PULL CHOP control.
ADD:
The inputs from channel 1 and channel 2 are
summed and displayed as a single signal. If the
Channel 2POSition/PULL INVert control is
pulled out, the input from channel 2 is subtracted
from channel 1 and the difference is displayed as a
single signal.
9. CH1 AC-GND-DC Switch. Three-position lever
switch with the following positions:
AC:
Channel 1 input signal is capacitively coupled; dc
component is blocked.
GND:
Opens signal path and grounds input to vertical
amplifier. This provides a zero-volt base line, the
position of which can be used as a reference when
performing dc measurements.
7
CONTROLS AND INDICATORS
4142
CH1
+
OUTPUT
++
SELECTOR LINE VOLTAGE FUSE
LINE VOLTAGE SLELCTOR
110
++
++
+
40
Fig. 2. Rear Panel Controls and Indicators.
DC:
Direct coupling of channel 1 input signal; both ac
and dc components of signal produce vertical de-
flection.
10. CH1 (X) Input Jack. Vertical input for channel 1.
for channel 1. Provides step adjustment of vertical
sensitivity. When channel 1 VARiable control is set
to CAL, vertical sensitivity is calibrated in 10 steps
from 5 mV/div to 5 V/div in a 1-2-5 sequence. When
the X-Y mode of operation is selected, this control
provides step adjustment of X-axis sensitivity.
12. CH1 VARiable/PULL X5 MAG Control:
VARiable:
Rotation provides vernier adjustment of channel 1
vertical sensitivity. In the fully-clockwise (CAL)
position, the vertical attenuator is calibrated. Counterclockwise rotation decreases gain sensitivity. In
X-Y operation, this control becomes the vernier
X-axis sensitivity control.
PULL X5 MAG:
When pulled out, increases vertical sensitivity by a
factor of five. Effectivelyprovides two extra sensitivity settings: 2 mV/div and 1 mV/div. In X-Y
mode, increases X-sensitivity by a factor of five.
13. CH1POSition/PULL CHOP Control:
POSition:
Adjusts vertical position of channel 1 trace.
4343
CH2
OUTPUT
PLOT OUTPUT
PEN
DOWN
(ANALOG)
Y
OUTPUT
+
WARNING
+
39
PULL CHOP:
When this switch is pulled out in the dual-trace
mode, the channel 1 and channel 2 sweeps are
chopped and displayed simultaneously (normally
used at slower sweep speeds). When it is pushed in,
the two sweeps are alternately displayed, one after
the other (normally used at higher sweep speeds).
14. CH2POSition/PULL INVert Control:
POSition:
Adjusts vertical position of channel 2 trace. In X-Y
operation, rotation adjusts vertical position of X-Y
display.
PULL INVert:
When pushed in, the polarity of the channel 2 signal
is normal. When pulled out, the polarity of the
channel 2 signal is reversed, thus inverting the
waveform.
15. CH2 VOLTS/DIV Control. Vertical attenuator for
channel 2. Provides step adjustment of vertical sensitivity. When channel 2 VARiable control is set to
CAL, vertical sensitivity is calibrated in 10 steps from
5 mV/div to 5 V/div in a 1-2-5 sequence. When the
X-Y mode of operation is selected, this control provides step adjustment of Y-axis sensitivity.
16. CH2 VARiable/PULL X5 MAG Control:
VARiable:
Rotation provides vernier adjustment of channel 2
vertical sensitivity. In the fully-clockwise (CAL)
position, the vertical attenuator is calibrated. Counterclockwise rotation decreases gain sensitivity. In
X-Y operation, this control becomes the vernier
Y-axis sensitivity control.
SERIALNO’
8
CONTROLS AND INDICATORS
PULL X5 MAG:
When pulled out, increases vertical sensitivity by a
factor of five. Effectively provides two extra sensitivity settings: 2 mV/div and 1 mV/div. In X-Y
mode, increases Y-sensitivity bya factor of five.
17. CH2 (Y) Input Jack. Vertical input for channel 2.
Y-axis input for X-Y operation.
18. CH2 AC-GND-DC Switch. Three-position lever
switch with the following positions:
AC:
Channel 2 input signal is capacitively coupled; dc
component is blocked.
GND:
Opens signal path and grounds input to vertical
amplifier. This provides a zero-volt base line, the
position of which can be used as a reference when
performing dc measurements.
DC:
Direct coupling of channel 2 input signal; both ac
and dc components of signal produce vertical deflection.
HORIZONTAL CONTROLS
19. Time Base TIME/DIV Control. Provides step selection of sweep rate for the time base. When the VARi-able Sweep control is set to CAL, sweep rate is
calibrated. This control has 21 steps, from 0.1 µS/div
to 0.5 S/div, in a 1-2-5 sequence.
20. VARiable Sweep Control. Rotation of control is vernier adjustment for sweep rate. In fully clockwise
(CAL) position, sweep rate is calibrated.
21.POSition/PULL X10 MAG Control.
POSition:
Horizontal (X) position control.
PULL X10 MAG:
Selects ten times sweep magnification when pulled
out, normal when pushed in. Increases maximum
sweep rate to 10 nS/div.
22. X-Y Switch. Used with the VERTical MODE switch
andTriggerSOURCEswitch to select X-Y operating
mode.The channel 1 input becomes the X-axis and the
channel 2 input becomes the Y-axis. Trigger source
and coupling are disabled in this mode.
TRIGGERING CONTROLS
23. HOLDOFF/PULL NORM TRIG Control.
HOLDOFF:
Rotation adjusts holdoff time (trigger inhibit period
beyond sweep duration). When control is rotated
fully counterclockwise, the holdoff period is MIN-
inum (normal). The holdoff period increases progressively with clockwise rotation.
PULL NORM TRIG:
When pushed in, selects AUTOmatic triggering
mode. In this mode, the oscilloscope generates
sweep (free runs) in absence of an adequate trigger;
itautomaticallyrevertstotriggered sweepoperation
when an adequate trigger signal is present. On the
Model 2522B, automatic triggering is applicable to
both the main sweep and delayed sweep.
When pulled out, selects NORMal triggered sweep
operation. A sweep is generated only when an adequate trigger signal is present.
24. Trigger SOURCE Switch. Selects source of sweep
trigger. Four-position lever switch with the following
positions:
CH1
Causes the channel 1 input signal to become the
sweep trigger, regardless of the VERTical MODE
switch setting.
CH2:
The channel 2 signal becomes the sweep trigger,
regardless of the VERTicalMODE switch setting.
ALTernate:
Selects alternate triggering, used in dual-trace
mode,permitseachwaveformviewed to become its
own trigger source. Must be used with alternate
sweep, cannot be used with chop sweep.
EXT:
Signal from EXTernal TRIGger jack becomes
sweep trigger.
25. Trigger COUPLING Switch. Selects trigger cou-
pling. Four-position lever switch with the following
positions:
AC:
Trigger signal is capacitively coupled. Used for
most waveforms except video.
TV-V:
Used for triggering from television vertical sync
pulses. Also serves as lo-pass/dc (high frequency
reject) trigger coupling.
TV-H:
Used for triggering from television horizontal sync
pulses.Alsoserves as hi-pass(lowfrequency reject)
trigger coupling.
LINE:
Signal derived from input line voltage (50/60 Hz)
becomes trigger.
Trigger level adjustment; determines the point on
the triggering waveform where the sweep is triggered. Rotation in the (–) direction (counterclockwise) selects more negative triggering point;
rotation in the (+) direction (clockwise) selects
more positive triggering point.
9
CONTROLS AND INDICATORS
PULL (—) SLOPE:
Two-position push-pull switch. The “in” position
selectsapositive-goingslope and the “out” position
selectsanegative-goingslopeas triggering pointfor
main sweep.
27. EXTernal TRIGger Jack. External trigger input for
single- and dual-trace operation.
DIGITAL STORAGE CONTROLS
28. STORAGE/ANALOG Pushbutton. When released,
scope is in ANALOG mode. When engaged, scope is
in digital STORAGE mode. In ANALOG mode, all
digital mode controls (29–38) are disabled. In STOR-AGE mode, all digital mode controls (29–38) are
enabled.
In STORAGE mode with all digital mode controls
(29–38) released, scope is in REFRESH mode of
storage, in which it updates the stored waveform each
time an adequate trigger signal is received (if the
previous update has been completed). REFRESH
mode is applicable at all TIME/DIV settings.
29. SAVE CH2 Pushbutton. Engaging this switch
freezes and stores the channel 2 trace immediately.
The channel 2 display and memory cannot be updated
until this switch is released.
30. SAVE ALL Pushbutton. Eng aging this switch
freezes and stores the channel 1 and channel 2 traces
immediately. The display and memory cannot be updated until this switch is released.
31. SLOW X100 Pushbutton. Engaging this switch expands the time base settings of 10 ms/div to 0.5 S/div
by a factor of 100 to become settings of 1 sec/div to
50 sec/div. Time base settings are normal when this
switch is released.
32. ROLL Pushbutton. Engaging this switch selects the
ROLL mode of operation, in which the trace moves
across the CRT from right to left like a strip chart
recorder(oppositeof conventional oscilloscope operation) and the display is continuously updated. This
update takes place even in the absence of any trigger
signal. The rolling speed equals the TIME/DIV setting (and SLOW X100 if engaged). ROLL mode is
applicable only at time base settings of 10 ms/div to
0.5 S/div.
33. SINGLE Pushbutton. Engaging this switch enables
the single sweep mode of digital storage operation, in
which the memory is filled in a single sweep and
continuously displayed. Single sweep mode is applicable for all “real time” settings of the TIME/DIV
control, that is, 20 µs/div and slower.
34. RESET Pushbutton. With the SINGLE switch engaged, this switch readies the scope for single sweep
operation.Asuitable triggersignalarrivingafter pressing the RESET switch initiates the single sweep.
35. READY Indicator. Lights when the RESET switch
readies the scope for single sweep, and goes out when
trigger signal occurs.
36. PRE-TRIGGER Pushbuttons. Selects post-trigger
or pre-trigger storage conditions. In post-trigger condition (0% pre-trigger), the entire stored waveform
occurs after the trigger. The trigger point is at the
extreme left of the display, as in a conventional analog
oscilloscope display. In pre-trigger conditions, a portion of the waveform occurring before the trigger is
displayed. Pre-trigger selections of 25%, 50%, and
75% are selectable. For 0% pre-trigger, release both
PRE-TRIGGER pushbuttons. For 25% pre-trigger,
engage the left pushbutton. For 50% pre-trigger, engagetheright pushbutton. For75%pre-trigger,engage
both pushbuttons. Pre-trigger operation is applicable
only to single sweep operation, not to ROLL or
REFRESH modes.
37. PLOT Pushbutton. When engaged, this switch acti-vates the PLOT mode. Output is available at the CH1
OUTPUT, CH2 OUTPUT, and PEN DOWN jacks
on the rear of the scope. At the appropriate time, the
PEN DOWN output goes TTL low to permit plotting
on an analog plotter or strip chartrecorder.ThePLOT
mode is applicable only when the SAVE ALL pushbutton is engaged.
38. PEN DOWN Indicator. In the PLOT mode, this
indicator lights when the PEN DOWN output on the
rear panel is TTL low.
REAR PANEL CONTROLS (see Fig. 2)
39. Fuse Holder/Line Voltage Selector. Contains fuse
and selects line voltage.
40. Power Cord Receptacle.
41. CH1 OUTPUT Jack. Supplies channel 1 digital stor-
age contents for use with an analog plotter.
42. CH2 OUTPUT Jack. Supplies channel 2 digital storage contents for use with an analog plotter.
43. PEN DOWN Output Jack. TTL open collector output to raise and lower plotter pen. Plotter information
is output at the CH1 OUTPUT and CH2 OUTPUT
jacks, starting at each positive and negative transition
ofPENDOWN.TheTIME/DIV switch sets the period
of this output, except plotting is at 10x the TIME/DIV
setting.
44. Y OUTPUT Jack. Output terminal where sample of
channel 2 analog signal is available. Amplitude of
output is 50 millivolts per division of vertical deflection seen on CRT when terminated into 50 ohms.
Output impedance is 50 ohms.
10
OPERATING INSTRUCTIONS
SAFETY PRECAUTIONS
WARNING
The following precautions must be observed to help prevent electric shock.
1. When the oscilloscope is used to make measurements
in equipment that contains high voltage, there is always acertainamountofdangerfromelectricalshock.
The person using the oscilloscope in such conditions
should be a qualified electronics technician or otherwise trained and qualified to work in such circumstances. Observe the TEST INSTRUMENT SAFETY
recommendations listed on the inside front cover of
this manual.
2. Donotoperatethisoscilloscopewith the case removed
unless you are a qualified service technician. High
voltage up to 2100 volts is present when the unit is
operating with the case removed.
3. The ground wire of the 3-wire ac power plug places
the chassis and housing of the oscilloscope at earth
ground. Use only a 3-wire outlet, and do not attempt
to defeat the ground wire connection or float the oscilloscope; to do so may pose a great safety hazard.
4. Specialprecautionsarerequiredtomeasureorobserve
line voltage waveforms with any oscilloscope. Use the
following procedure:
a. Do not connect the ground clip of the probe to
either side of the line. The clip is already at earth
ground and touching it to the hot side of the line
may “weld” or “disintegrate” the probe tip and
cause possible injury, plus possible damage to the
scope or probe.
b. Insert the probe tip into one side of the line voltage
receptacle, then the other. One side of the receptacleshouldbe“hot” and produce thewaveform.The
other side of the receptacle is the ac return and no
waveform should result.
EQUIPMENT PROTECTION
PRECAUTIONS
Thefollowing precautionswillhelpavoid
damage to the oscilloscope.
1. Never allow a small spot of high brilliance to remain
stationary on the screen for more than a few seconds.
The screen may become permanently burned. A spot
will occur when the scope is set up for X−Y operation
and no signal is applied. Either reduce the intensity so
the spot is barely visible, apply signal, orswitch back
to normal sweep operation. It is also advisable to use
low intensity with AUTO triggering and no signal
applied for long periods. A high intensity trace at the
same position could cause a line to become permanently burned onto the screen.
2. Do not obstruct the ventilating holes in the case, as this
will increase the scope’s internal temperature.
3. Excessive voltage applied to the input jacks may damage the oscilloscope. The maximum ratings of the
inputs are as follows:
CH 1 and CH 2:
400 V dc + ac peak.
EXT TRIG:
300 V dc + ac peak.
Z-AXIS INPUT:
30 V ( dc and ac peak).
4. Always connect a cable from the ground terminal of
the oscilloscope to the chassis of the equipment under
test. Without this precaution, the entire current for the
equipment under test may be drawn through the probe
clip leads under certain circumstances. Such conditionscouldalsoposea safety hazard, which theground
cable will prevent.
5. The probe ground clips are at oscilloscope and earth
ground and should be connected only to the earth
ground or isolated common of the equipment under
test. To measure with respect to any point other than
the common, use CH 2 – CH 1 subtract operation
(ADD mode and invert channel 2), with the channel 2
probe to the point of measurement and the channel 1
probe to the point of reference. Use this method even
if the reference point is a dc voltage with no signal.
11
OPERATING INSTRUCTIONS
OPERATING TIPS
The following recommendations will help obtain the best
performance from the oscilloscope.
1. Always use the probe ground clips for best results,
attached to a circuit ground point near the point of
measurement.Donotrelysolelyon an external ground
wire in lieu of the probe ground clips as undesired
signals may be introduced.
2. Avoid the following operating conditions:
a. Direct sunlight.
b. High temperature and humidity.
c. Mechanical vibration.
d. Electrical noiseand strong magnetic fields, such as
near large motors, power supplies, transformers,
etc.
3. Occasionally check trace rotation, probe compensation,andcalibrationaccuracy of theoscilloscopeusing
the procedures found in the MAINTENANCE section
of this manual.
4. Terminate the output of a signal generator into its
characteristic impedance to minimize ringing, especially if the signal has fast edges such as square waves
or pulses. For example, the typical 50 Ω output of a
square wave generator should be terminated into an
external 50 Ω terminating load and connected to the
oscilloscope with 50 Ω coaxial cable.
5. Probe compensation adjustment matches the probe to
the input of the scope. For best results, compensation
should be adjusted initially, then the same probe alwaysusedwiththesamechannel. Probe compensation
should be readjusted when a probe from a different
oscilloscope is used.
INITIAL STARTING PROCEDURE
Until you familiarize yourself with the use of all controls,
the settings given here can be used as a reference point to
obtain a trace on the CRT in preparation for waveform
observation.
1. Set these controls as follows:
VERTical MODE to CH1.
CH1AC/GND/DC to GND.
Select AUTO triggering (HOLD OFF pushed in)
Trigger COUPLING to AC.
Trigger SOURCE to CH1.
All POSition controls and INTENSITY control
centered (pointers facing up).
Time Base control to 1 mS/div.
2. Press the red POWER pushbutton.
3. A trace should appear on the CRT. Adjust the trace
brightness with the INTENSITY control, and the
trace sharpness with the FOCUS control.
SINGLE TRACEDISPLAY
Either channel 1 or channel 2 maybe used for single-trace
operation. To observe a waveform on channel 1:
1. Perform the steps of the “Initial Starting Procedure”.
2. Connect the probe to the CH 1 (X) input jack.
3. Connect the probe ground clip to the chassis or common of the equipment under test. Connect the probe
tip to the point of measurement.
4. Move the CH1 AC/GND/DC switch out of the GND
position to either DC or AC.
5. If no waveforms appear, increase the sensitivity by
turning the CH 1 VOLTS/DIV control clockwise to a
position that gives 2 to 6 divisions vertical deflection.
6. Position the waveform vertically as desired using the
CH1 POSition control.
7. Thedisplayonthe CRT may be unsynchronized.Refer
to the “Triggering” paragraphs in this section for procedures on setting triggering and sweep time controls
to obtain a stable display showing the desired number
of waveforms.
DUAL TRACE DISPLAY
In observing simultaneous waveforms on channel 1 and
2, the waveforms are usually related in frequency, or one of
the waveforms is synchronized to the other, although the
basic frequencies are different. To observe two such related
waveforms simultaneously, perform the following:
1. Connect probes to both the CH 1 (X) and CH 2 (Y)
input jacks.
2. Connect the ground clips of the probes to the chassis
or common of the equipment under test. Connect the
tips of the probes to the two points in the circuit where
waveforms are to be measured.
3. To view both waveforms simultaneously, set the
VERTical MODE switch to DUAL and select either
ALT (alternate) or CHOP with the PULL CHOP
switch.
4. In the ALT sweep mode (PULL CHOP switch
pushed in), one sweep displays the channel 1 signal
and the next sweep displays the channel 2 signal in an
alternating sequence. Alternate sweep is normally
used for viewing high-frequency or high-speed waveforms at sweep times of 1 ms/div and faster, but may
be selected at any sweep time.
12
OPERATING INSTRUCTIONS
5. In the CHOP sweep mode (PULL CHOP switch
pulled out), the sweep is chopped (switched) between
channel 1 and channel 2. Using CHOP, one channel
does not have to “wait” for a complete swept display
of the other channel. Therefore, portions of both channel’swaveformsaredisplayedwith the phase relationship between the two waveforms unaltered. Chop
sweep is normally used for low-frequency or lowspeed waveforms at sweep times of 1 ms/div and
slower; or where the phase relationship between channel 1 and channel 2 requires measurement.
If chop sweep is used at sweep times of 0.2 ms/div and
faster, the chop rate becomes a significant portion of
the sweep and may become visible in the displayed
waveform. However, you may select chop sweep at
any sweep time for special applications.
6. Adjust the channel 1 and 2
▲
POSition controls to
▼
place the channel 1 trace above the channel 2 trace.
7. Set the CH 1 and CH 2 VOLTS/DIV controls to a
position that gives 2 to 3 divisions of vertical deflection for each trace. If the display on the screen is
unsynchronized, refer to the “Triggering” paragraphs
in this section of the manual for procedures for setting
triggering and sweep time controls to obtain a stable
display showing the desired number of waveforms.
8. WhentheVERTicalMODEswitchissettoADD,the
algebraic sum of CH 1 + CH 2 is displayed as a single
trace. When the PULL INV switch is pulled out, the
algebraic difference of CH 1 – CH 2 is displayed.
9. Iftwowaveformshavenophaseorfrequencyrelationship, there is seldom reason to observe both waveforms simultaneously. However, these oscilloscopes
do permit the simultaneous viewing of two such unrelated waveforms, using alternate triggering. Refer to
the paragraphs on “Triggering — Trigger SOURCE
Switch” for details on alternate triggering.
TRIGGERING
The Model 2522B Oscilloscope provides versatility in
sync triggering for ability to obtain a stable, jitter-free display in single-trace, or dual-trace operation. The proper
settings depend upon the type of waveforms being observed
and the type of measurement desired. An explanation of the
various controls which affect synchronization is given to
help you select the proper setting over a wide range of
conditions.
AUTO or NORM Triggering
1. IntheAUTO mode (PULL NORMTRIGpushedin),
automatic sweep operation is selected. In automatic
sweep operation, the sweep generator free-runs to
generate a sweep without a trigger signal. However, it
automatically switches to triggered sweep operation if
an acceptable trigger source signal is present. The
AUTO mode is handy when first setting up the scope
to observe a waveform; it provides sweep for wave-
form observation until other controls can be properly
set. Once the controls are set, operation is often
switched back to the normal triggering mode, since it
is more sensitive. Automatic sweep must be used for
dc measurements and signals of such low amplitude
that they will not trigger the sweep.
2. In the NORM mode (PULL NORM TRIG pulled
out),normaltriggered sweep operationisselected.The
sweep remains at rest until the selected trigger source
signal crosses the threshold level set by the TRIGLEVEL control. The trigger causes one sweep to be
generated, after which the sweep again remains at rest
until triggered. In the normal triggering mode, there
will be no trace unless an adequate trigger signal is
present. In the ALT VERTICAL MODE of dualtrace operation with the SOURCE switch also set to
ALT, there will be no trace unless both channel 1 and
channel 2 signals are adequate for triggering. Typically,signalsthat produce even onedivisionofvertical
deflection are adequate for normal triggered sweep
operation.
Trigger COUPLING Switch
1. The AC position is used for most waveforms except
video.Thetriggersignal is capacitivelycoupled.Thus,
it blocks the dc component and references the average
of the “changing” portion of the waveform.
2. The TV H and TV V positions are primarily for
viewing composite video waveforms. Horizontal sync
pulses are selected as trigger when the trigger COU-PLINGswitchissettothe TV H position, and vertical
sync pulses are selected as trigger when the trigger
COUPLING switch is set to the TV V position. The
TV H and TV V positions may also be used as low
frequency reject and high frequency reject coupling,
respectively. Additional procedures for observing video
waveforms are given later in this section of the manual.
Trigger SOURCE Switch
The trigger SOURCE switch (CH 1, CH 2, etc.) selects
the signal to be used as the sync trigger.
1. If the SOURCE switch is set to CH 1 (or CH 2) the
channel 1 (or channel 2) signal becomes the trigger
source regardless of the VERTICAL MODE selection.CH1,orCH2areoftenused as the trigger source
for phase or timing comparison measurements.
2. By setting the SOURCE switch to ALT, alternating
triggering mode is activated. In this mode, the trigger
source alternates between CH 1 and CH 2 with each
sweep. This is convenient for checking amplitudes,
waveshape, or waveform period measurements, and
even permits simultaneous observation of two waveforms which are not related in frequency or period.
However,thissetting is notsuitablefor phaseortiming
comparison measurements. For such measurements,
both traces must be triggered by the same sync signal.
Alternate triggering can only be used in dual-trace
13
OPERATING INSTRUCTIONS
mode (VERT MODE set to DUAL), and with alternate sweep only (PULL CHOP not engaged).
3. In the EXT position, the signal applied to the EXTTRIG jack becomes the trigger source. This signal
must have a timing relationship to the displayed waveforms for a synchronized display.
4. In the LINE position of the COUPLING switch,
triggering is derived from the input line voltage (50/60
Hz) and the trigger SOURCE switch is disabled. This
is useful for measurements that are related to line
frequency.
TRIG LEVEL/PULL (–) SLOPE Control
(Refer to Fig. 3)
A sweep trigger is developed when the trigger source
signalcrossesapresetthreshold level. Rotation oftheTRIGLEVEL control varies the threshold level. In the + direction
(clockwise), the triggering threshold shifts to a more positive value, and in the − direction (counterclockwise), the
triggering threshold shifts to a more negative value. When
the control is centered, the threshold level is set at the
approximate average of the signal used as the triggering
source. Proper adjustment of this control usually synchronizes the display.
The TRIG LEVEL control adjusts the start of the sweep
to almost any desired point on a waveform. On sine wave
signals, the phase at which sweep begins is variable. Note
that if the TRIG LEVEL control is rotated toward its
extreme + or − setting, no sweep will be developed in the
normal trigger mode because the triggering threshold exceeds the peak amplitude of the sync signal.
When the PULL (–) SLOPE control is set to the + (“in”)
position, the sweep is developed from the trigger source
waveform as it crosses a threshold level in a positive-going
direction. When the PULL (–) SLOPE control is set to the
− (“out”) position, a sweep trigger is developed from the
trigger source waveform as it crosses the threshold level in
a negative-going direction.
TIME BASE Control
Set the Time Base TIME/DIV control to display the
desired number of cycles of the waveform. If there are too
manycyclesdisplayedforgoodresolution,switchtoafaster
sweep time. If only a line is displayed, try a slower sweep
time.Whenthesweep time is fasterthanthe waveformbeing
observed,onlypart of it willbedisplayed,which may appear
as a straight line for a square wave or pulse waveform.
HOLDOFF Control
(Refer to Fig. 4)
A “holdoff” period occurs immediately after the completion of each sweep, and is a period during which triggering
of the next sweep is inhibited. The normal holdoff period
varies with sweep rate, but is adequate to assure complete
retrace and stabilization before the next sweep trigger is
permitted. The HOLDOFF control allows this period to be
extended by a variable amount if desired.
This control is usually set to the MIN position (fully
counterclockwise) because no additional holdoff period is
necessary. The HOLDOFF control is useful when a complex series of pulses appear periodically such as in Fig. 4B.
Improper sync may produce a double image as in Fig. 4A.
Such a display could be synchronized with the VARSWEEP control, but this is impractical because time measurements are then uncalibrated. An alternate method of
synchronizing the display is with the HOLDOFF control.
The sweep speed remains the same, but the triggering of the
next sweep is “held off” for the duration selected by the
HOLDOFF control. Turn the HOLDOFF control clockwise from the MIN position until the sweep starts at the
same point of the waveform each time.
Slope “–” Range
Slope “+” Range
+
Level
–
Fig. 3. Function of Slope and Level Controls.
A. Holdoff not used
B. Holdoff used
Fig. 4. Use of HOLDOFF Control.
14
OPERATING INSTRUCTIONS
MAGNIFIED SWEEP OPERATION
Since merely shortening the sweep time to magnify a
portion of an observed waveform can result in the desired
portion disappearing off the screen, magnified display
should be performed using magnified sweep.
Using thePOSition control, move the desired portion
of waveformto the center of the CRT.Pull out the PULL X10
knob to magnify the display ten times. Forthis type of display,
the sweep time is the Time Base TIME/DIV control setting
dividedby 10. Rotation of thePOSition control can then be
used to select the desired portion of the waveforms.
X−Y OPERATION
X−Y operation permits the oscilloscope to perform many
measurements not possible with conventional sweep operation. The CRT display becomes an electronic graph of two
instantaneous voltages. The display may be a direct comparison of the two voltages such as stereoscope display of
stereo signal outputs. However, theX−Y mode can be used
to graph almost any dynamic characteristic if a transducer is
used to change the characteristic (frequency, temperature,
velocity, etc.) into a voltage. One common application is frequencyresponse measurements, wheretheY axis correspondsto
signal amplitude and the X axis corresponds to frequency .
1. Depress the X−Y switch and set the Trigger Source
and VERTical MODE switches to X−Y.
2. In this mode, channel 1 becomes the X axis input and
channel 2 becomes the Y axis input. The X and Y
positionsarenowadjustedusingthePOSitionand
the channel 2POSition controls, respectively.
3. Adjust the amount of vertical (Y axis) deflection with
the CH 2 VOLTS/DIV and VARIABLE controls.
4. Adjust the amount of horizontal (X axis) deflection
with the CH 1 VOLTS/DIV and VARIABLE controls.
VIDEO SIGNAL OBSERVATION
Setting the COUPLING switch to the TV-H or TV-V
position permits selection of horizontal or vertical sync
pulses for sweep triggering when viewing composite video
waveforms.
When the TV-H mode is selected, horizontal sync pulses
are selected as triggers to permit viewing of horizontal lines
of video. A sweep time of about 10 µs/div is appropriate for
displaying lines of video. The VAR SWEEP control can be
set to display the exact number of waveforms desired.
When the TV-Vmode is selected,vertical sync pulses are
selected as triggers to permit viewing of vertical fields and
frames of video. A sweep time of 2 ms/divis appropriate for
viewing fields of video and 5 ms/div for complete frames
(two interlaced fields) of video.
At most points of measurement, a composite video signal
is of the (−) polarity,that is, the sync pulses are negative and
the video is positive. In this case, use (− ) SLOPE. If the
waveform is taken at a circuit point where the video waveform is inverted, the sync pulses are positive and the video
is negative. In this case, use (+) SLOPE.
DIGITAL STORAGE OPERATION
Digitizing Repetitive Waveforms
While the most powerful feature of a digital storage
oscilloscope (DSO) is its ability to capture one-time events,
a DSO can also digitize conventionalrepetitivewaveforms,
such as those observed on a standard analog scope. An
analog/digital unit such as the Model 2522B allows the user
to set up sensitivity, sweep, and triggering in a familiar
analog setting, and then switch to the digital mode. This is
a good way to gain familiarity with the digital features of
the instrument. You may wish to pursue this method as
described below in the “Refresh Mode” paragraph.
Also,some repetitivesignalscan beviewedandmeasured
much more effectively in the digital store mode than in the
analog mode. One example is slow signals below 60 Hz.
Signalsthatappear asaflickeringwaveformorjust amoving
dot on the screen in the analog mode are displayed as a
bright, non-flickering, easily viewed entire waveform in the
digital store mode.
The digital store mode also improves viewing and measurementofrepetitivesignalswithlowtrigger repetition rates
relative to the sweep rate. In the analog mode, such waveformsmaybetoofaintfor viewing. Inthedigitalstoremode,
the display is equivalent to a CRT with infinite persistence.
The waveform can be easily viewed.
Refresh Mode
In the refresh mode, the waveform is digitized and displayed on the screen. After the complete waveform is digitized, the next trigger causes the stored waveform to be
replaced (refreshed) by a new waveform.
1. Set up the oscilloscope to view a periodic waveform
in the ANALOG mode. A djust the trigger,
VOLTS/DIV, sweep TIME/DIV, and vertical position controls as desired.
2. Initially, disengage all of the digital storage control
switches (the group of switches beneath the CRT).
3. Select digital storage operation by setting the ANA-LOG/STORAGE switch to STORAGE (engaged).
When all digital storage control switches are released,
the “refresh” mode is selected. The waveform should
appear on the display, relatively unchanged from the
previously displayed analog version. In the “refresh”
mode, the display is continually updated as long as a
suitable trigger signal remains present.
15
OPERATING INSTRUCTIONS
4. Once a waveform is digitized, it can be stored in
long-term memory by pressing the SAVE ALL or
SAVE CH2 switches. The SAVE CH2 switch immediately stores the channel 2 waveform. The SAVEALL switch immediately freezes the display and
stores both waveforms. Once a waveform is stored by
engagingeitheroftheseswitches, it will be storeduntil
the switch is disengaged or the power is turned off.
NOTES
Anyof theoperatingmodespreviouslydiscussed in analog operation (e.g. DUAL,
ADD, etc.) can be used in digital mode,
except for X–Y operation.
Time base settings of 10 µs/div and faster
result in a display that is acquired through
“equivalent time sampling”. This process
develops the digitized image over many
cycles of the repetitive signal; therefore,
these higher sweep speeds cannot be used
for capture of one-time events. Equivalent
time sampling is discussed in detail in Appendix II, “Unique Characteristics of Digital Storage Oscilloscopes”.
Digitizing One-Time Events
One of the most powerful features of a digital storage
oscilloscope (DSO) is its ability to capture one-time events.
To capture one-time events, single-sweep operation is employed. This is done through the use of the SINGLE switch.
When pushed, this switch releases the REFRESH or ROLL
mode if previously engaged, and readies the digital storage
circuit to receivea trigger signal — presumably the event to
be captured or some other time-related occurrence. When
the event arrives, it is stored in the memory and displayed.
The procedure is as follows:
1. Set the oscilloscope to run in analog mode (ANA-
LOG/DIGITAL switch to ANALOG). Select
NORM triggering, and adjust the TRIG LEVEL
control so that the unit triggers on the event to be
captured.
2. Set the oscilloscope to digital (STORAGE) mode by
engaging the ANALOG/STORAGE switch.
3. Initially, set the PRE TRIG switch to 50%. A more
detaileddiscussion of pre- and post-triggering is given
later in this section of the manual.
4. Press the RESET switch. The READY indicator will
momentarily light as the scope awaits the arrival of the
trigger signal. When that trigger occurs, the READY
indicator will go off, and the event being monitored
will roll to the center of the display and stop.
NOTES
Depending on the sweep TIME/DIV setting, the anticipated event may roll to the
center of the screen very rapidly, or very
slowly, after the READY indicator goes
out.
Single-sweepmodecanbeused in the presence of very rapidly occurring events, even
continuous waveforms if desired. In that
case, the period between pushing the RESET switch and the arrival of the trigger
may be very short or almost instantaneous.
As a result, the READY indicator may
light for only a veryshort time, perhaps not
being visible at all.
PushingtheSINGLE switch automatically
disengages both the REFresh and ROLL
modes, which are continuous modes.
5. Once the waveform is captured, it can be stored in
long-term memory through the use of the SAVE CH2
or SAVE ALL switches.
Pre- and Post-Trigger Capture
Another powerful feature of a DSO is is ability to display
“pre-trigger” information, that is, events occurring before
the arrival of a trigger event. On the Model 2522B, pre-trigger operation is available in single-sweep mode by setting
the PRE TRIG switches. As shown in the above section on
“Digitizing One-Time Events”, when 50% pre-trigger operation is selected, the event to be captured is situated in the
centerofthe memory (roughlythecenterof thedisplay)after
storage. The waveform to the left of center represents activity which occurred before that trigger arrived. Pre-Trigger
selection of 25% and 75% are also available.
If the PRE-TRIG switches are released (0% pretrigger),
then no pre-trigger information is stored, and the trigger
event rolls to the extreme left of the display. In this case, all
the information displayed on the screen represents activity
after the trigger event. You may wish to use post-trigger
operation first to observe where the trigger is occurring on
the waveform of interest, and then switch to pre-trigger
mode.
Roll Mode
In this mode of operation, the waveform rolls across the
CRTfrom right to left (asopposedtostandardoscilloscopes,
which have the trace moving from left to right) in the same
manner as most strip chart recorders. It is most commonly
used for viewing very slow events.
1. Setupthe oscilloscopeinanalog mode sothatthe event
to be observed is properly positioned on the display.
you may wish to use AUTO triggering so that the
scope continues to draw a trace even if the event is
especially slow.
2. S witch to digital mode (ANALOG/STORAGE
switchtoSTORAGE),anddepress the ROLLswitch.
Select a TIME/DIV setting that produces a roll at the
desired speed. As the sweep speed is decreased, the
waveform will move across the screen more slowly
and the Roll feature will become more apparent. It can
also be slowed by a factor of 100 on some ranges; see
the discussion in the next section.
16
OPERATING INSTRUCTIONS
3. The rolling display can be frozen at any time by
pressing the SAVE CH2 or SAVE ALL switches (as
discussed in the section on “Digitizing Repetitive
Waveforms”).
NOTE
ROLL mode cannot be used on sweep
speeds greater than 10 ms/div.
Expanded Sweep Settings—SLOW X100 Mode
In digital mode, the 10 ms/div to 0.5 s/div TIME/DIV
ranges can be expanded by a factor of 100 by depressing the
SLOW X100 switch. For example, the 10 ms/div setting
becomes 1000 ms (1 sec)/div when this switch is engaged.
This time base expansion is extremely useful for observing
very slow events. With it, the scope is capable of recording
an event up to 500 seconds in duration (.5 sec/div x100
becomes 50 sec/div).
Using Plotter Output
The Model 2522B oscilloscope provides facilities for
drivingananalogplotter.The following instructions explain
how to output a waveform; however, since plotters vary
greatly, only general instructions have been given. Consult
the manual for the particular plotter for more specific instructions. Also, it is recommended that you read this entire
procedure first, because some steps may need to be performed in a different sequence, depending on the plotter.
1. Initially, se the PLOT switch to “off” (released).
2. Useany of the digital modes describedabove to obtain
a digitized display to be plotted. Once the desired
display is obtained, freeze it using the SAVE ALL
switch.
3. Connect the plotter’s Y (vertical) input to either the
CH1 or CH2 OUTPUT jack on the rear panel of the
oscilloscope,dependingonwhichchanneldisplayyou
wish to plot.
4. Connect the plotter’s pen lift control input to the PENDOWN output jack on the rear of the scope.
5. Momentarily disable external up/down control of the
plotter pen; that is, set the pen to be permanently lifted
from the paper surface. This will allow you to exercise
initial “trial and error” sweeps without using up excessive ink. Consult the manual for the particular plotter
for details.
6. Engage the PLOT switch. This will gate a TTL-level
square wave through to the PEN DOWN output jack.
The repetition rate of this square wave is controlled by
the front panel TIME/DIV control. The relationship
is as follows. Each low or high portion of the square
wave represents the output of one screen, i.e. ten
divisions of waveform (10.24 divisions to be more
exact). This output takes place at a rate that is 1/10th
that of the TIME/DIV setting. For example, if the
TIME/DIV setting is 10 ms/div, the plotter output
sends out waveform information at a rate of
100 ms/div. For the full 10.24 divisions of screen, this
means a total interval of 1.024 seconds. Thus, the
square wave has low and high portions that are each
1.024secondslong.The PEN DOWN indicator on the
front panel of the oscilloscope lights during each
portion.
low
NOTE
Though many plotters use a TTL-low level
as the “pen down” signal, and technically
only require plotting information during
that low portion, this oscilloscope outputs
a full screen of data on both the low and
high intervals of the square wave. This
discussionassumes aplotterusingTTL low
as the “pen down” command.
When the PLOT switch is released, the
PEN DOWN output is at a constant TTL
high.
Engaging the SLOW X100 switch slows
the plot output by a factor of 100 on the
10 ms/div to 0.5 S/div TIME/DIV ranges.
7. With the plotter’s pen control input still disabled,
observe the PEN DOWN indicator on the scope to
determine the repetition rate of the square wave. Adjust as desired using the TIME/DIV control. It will be
necessary to use the plotter’s own time base to move
its pen in the horizontal (X) direction. If possible, set
that movement to begin on the high-to-low transition
of the PEN DOWN signal. Through trial and error,
with the pen still up, run the plotter horizontally to
determine the speed that produces the desired plot
width during the PEN DOWN interval as observed on
the LED. Also adjust the vertical deflection and offset
for the desired height.
8. Finally, during the high portion of the square wave
(PEN DOWN indicator off), re-enable external pen
control on the plotter. At the next transition of PENDOWN to low, the plotter should produce the desired
results.
NOTE
If you are unable to trigger the horizontal
plotter sweep from the PEN DOWN signal, you may have to enable that sweep
manually using the PEN DOWNindicator
as a timing guide.
APPLICATIONS GUIDEBOOK
B+K Precision offers a “Guidebook to Oscilloscopes”
which describes numerous applications for this instrument
and important considerations about probes. It includes a
glossary of oscilloscope terminology and an understanding
of how oscilloscopes operate. It may be downloaded free of
charge from our Website, www.bkprecision.com.
17
MAINTENANCE
n
n
n
WARNING
The following instructions are for use by
qualifiedservicepersonnelonly. Toavoid
electricalshock,donot perform any servicing other than contained in the operating instructions unless you are qualified
to do so.
Highvoltageup to 2000 V is presentwhen
covers are removed and the unit is operating. Remember that high voltage may
be retained indefinitely on high voltage
capacitors. Also remember that ac line
voltage is present on line voltage input
circuits any time the instrument is
plugged into an ac outlet, even if turned
off. Unplug the oscilloscope and discharge high voltage capacitors before
performing service procedures.
FUSE REPLACEMENT
If the fuse blows, the “ON” indicator will not light and the
oscilloscope will not operate. The fuse should not normally
open unless a problem has developed in the unit. Try to
determine and correct the cause of the blown fuse, then
replace only with the correct value fuse. For 110/125 V line
voltage operation, use an 800 mA, 250 V fuse. For
220/240 V line voltageoperation,usea600mA,250 V fuse.
The fuse is located on the rear panel adjacent to the power
cord receptacle.
Remove the fuseholder assembly as follows:
1. Unplug the power cord from rear of scope.
2. Insert a small screwdriver in fuseholder slot (located
between fuseholder and receptacle). Pry fuseholder
away from receptacle.
3. When reinstalling fuseholder, be sure that the fuse is
installedso that the correct line voltage is selected(see
LINE VOLTAGE SELECTION).
LINE VOLTAGE SELECTION
To select the desired line voltage, simply insert the fuse
and fuse holder so that the appropriate voltage is pointed to
by the arrow. Be sure to use the proper value fuse (see label
on rear panel).
PERIODIC ADJUSTMENTS
Probe compensation and trace rotation adjustments
should be checked periodically and adjusted if required.
These procedures are given below.
Probe Compensation
1. Connect probes to CH 1 and CH 2 input jacks. Per-
form procedure for each probe, one probe at a time.
2. Set the probe to X10 (compensation adjustment is not
possible in the X1 position).
3. Touch tip of probe to CAL terminal.
4. Adjust oscilloscope controls to display 3 or 4 cycles of
CAL square wave at 5 or 6 divisions amplitude.
5. Adjust compensation trimmer on probe for optimum
square wave (minimum overshoot, rounding off, and
tilt). Refer to Fig. 5.
Correct
Compensatio
Over
Compensatio
Insufficient
Compensatio
Fig. 5. Probe Compensation Adjustment.
Trace Rotation Adjustment
1. Set oscilloscope controls for a single trace display in
CH 1 mode, and with the channel 1 AC-GND-DC
switch set to GND.
2. Use the channel 1 POSition control to position the
trace over the center horizontal line on the graticule
scale. The trace should be exactly parallel with the
horizontal line.
3. Usethe TRACE ROTATION adjustmentonthefront
panel to eliminate any trace tilt.
18
MAINTENANCE
CALIBRATION CHECK
A general check of calibration accuracy may be made by
displaying the output of the CAL terminal on the screen.
This terminal provides a square wave of 0.5 V p-p. This
signal should produce a displayed waveform amplitude of
fivedivisionsat 0.1 V/div sensitivity for both channel 1 and
2 (with probes set for direct). With probes set for X10, there
should be five divisions amplitude at 10 mV/div sensitivity.
The VARIABLE controls must be set to CAL during this
check.
NOTE
The CAL signal should be used only as a
general check of calibration accuracy,not
as a signal source for performing recalibration adjustments; a voltage standard
calibrated at several steps and of 0.3% or
better accuracy is required for calibration
adjustments.
The CAL signal should not be used as a
time base standard.
INSTRUMENT REPAIR SERVICE
Because of the specialized skills and test equipment re-
quired for instrument repair and calibration, many customers prefer to rely upon B+K Precision for this service. To
use this service, even if the oscilloscope is no longer under
warranty, follow the instructions given in the SERVICE
INFORMATION portion of this manual. There is a flat rate
charge for instruments out of warranty.
19
APPENDIX I
IMPORTANT CONSIDERATIONS FOR RISE TIME
AND FALL TIME MEASUREMENTS
Error in Observed Measurement
The observed rise time (or fall time) as seen on the CRT
is actually the cascaded rise time of the pulse being measured and the oscilloscope’sownrisetime.Thetwo rise times
are combined in square law addition as follows:
T
observed
=
2
(T) +(T)
pulse
scope
2
The effect of the oscilloscope’s rise time is almost negligible when its rise time is at least 3 times as fast as that of
the pulse being measured. Thus, slower rise times may be
measured directly from the CRT. However, for faster rise
time pulses, an error is introduced that increases progressively as the pulse rise time approaches that of the oscilloscope. Accurate measurements can still be obtained by
calculation as described below.
Direct Measurements
The Model 2522B has a rated rise time of 18 ns. Thus,
pulse rise times of about 54 ns or greater can be measured
directly. Most fast rise times are measured at the fastest
sweep speed and using X10 magnification. This sweep rate
is 10 ns/div. A rise time of less than about five divisions at
this sweep speed should be calculated.
Calculated Measurements
For observed rise times of less than 54 ns, the pulse rise
time should be caluclated to eliminate the error introduced
by the cascaded oscilloscope rise time. Calculate pulse rise
time as follows:
T
pulse
observed
2
=
(T) +(T)
scope
2
Limits of Measurement
Measurements of pulse rise times that are faster than the
scope’srated rise time are not recommended because a very
small reading error introduces significant error into the
calculation. This limit is reached when the “observed” rise
time is about 1.3 times greater than the scope’s rated rise
time, about 23 ns minimum for the Model 2522B.
Probe Considerations
Forfastrise time measurementswhichapproachthelimits
of measurement, direct connection via 50 Ω coaxial cable
and 50 Ω termination is recommended where possible.
When a probe is used, its rise time is also cascaded in square
law addition. Thus the probe rating should be considerably
faster than the oscilloscope if it is to be disregarded in the
measurement.
20
APPENDIX II
UNIQUE CHARACTERISTICS OF DIGITAL STORAGE OSCILLSCOPES
Digital Storage Oscilloscopes (DSO’s) use a digital sampling technique to convert analog input signals to a series of
digital words that can be stored in memory. Since digital
sampling has disadvantages as well as advantages, it is
important to be aware of these unique characteristics of
DSO’s.
obviously incorrect. This occurs because the DSO is
taking one sample every 0.1 ms and a 10 kHz signal
hasonecycle every 0.1ms.Whatis actually happening
is that the frequency is off (not perfectly 10 kHz) by
just enough to cause the DSO to take one sample at a
slightly different place on each cycle of the waveform.
ALIASING
This DSO uses Real Time Sampling when sweep
TIME/DIV settings of 50 sec/div to 20 µs/div are selected.
Real TimeSampling simply means that samples of the input
signals are taken at equal spaces (e.g., every 0.25 ms when
the 50 ms/div range is selected). With Real Time Sampling,
a phenomena called “Aliasing” can occur when the input
signalisnotsampledoftenenough.Thiscausesthe digitized
signal to appear to be of a lower frequency than that of the
input signal. Unless you have an idea what the input signal
is supposed to look like, you will usually be unaware that
Aliasing is occurring.
Aliasing Example
To see what actually occurs when a Digital Storage Oscilloscope is Aliasing, perform the following example.
1. Apply a 10 kHz signal to the input jack and set the
sweepTIME/DIVcontrolto 50 µs/div. You shouldsee
about 5 cycles of the waveform on the display. Since
the DSO samples the input waveform 200 times per
division, each cycle is sampled 400 times.
2. Nowchangethesweep TIME/DIV control to 2ms/div.
The display should look crowded. Because the DSO
takes 200 samples per division, the sample points are
10 µs apart. Since the input signal is at a frequency of
10 kHz, it is being sampled 10 times per cycle. The
resulting display is too crowded to be useful, however,
it is not incorrect (it is very similar to what you would
see on a conventional analog oscilloscope).
3. Change the setting of the sweep TIME/DIV control to
20 ms/div. Vary the frequency a slight amount (until
the display is readable) to obtain as few cycles as
possible on the CRT. If you were to calculate the
frequency of the signal from the display, you would
come up with a much lower frequency than that of the
actual frequency of the signal at the input jack. As an
example, if three cycles are displayed, the calculated
frequency would be approximately 15 Hz. This is
Avoiding Aliasing
Aliasing is not limited to the above example. This phenomenon can occur anytime that at least two samples per
cycle are not taken (wheneverthe sweep TIME/DIV setting
is much too slow for the waveform being applied to the
input). Whenever the frequency of the signal is unknown,
always begin with the fastest real-time sweep speed
(20 µs/div) or by viewing the waveform in the analog mode
of operation first.
NOTE
Viewing one-time events or glitches is not
possible when sweep TIME/DIV settings
higher than 20 µs/div are selected. Viewing
one-timeeventsposesnoproblem withAliasingbecauseAliasing can occur only with
repetitive waveforms.
EQUIVALENT TIME SAMPLING
This oscilloscope uses a sampling method called Equivalent Time Sampling when sweep speeds higher than
20 µs/div are selected. This method permits viewing of
repetitive waveforms to 20 MHz, although the maximum
digital sampling rate is 10 Msamples/sec. When the Equivalent Time Sampling mode is activated, one sample is taken
during each cycle. Of course if one sample is taken during
eachcycleat the trigger point (the samepointoneachcycle),
only a flat trace would be produced. Therefore, it is necessary to take each sample further (in time) from the trigger
point than the last sample. This incremental delay is determined by the sweep TIME/DIV control setting. Because
1024 (1 k) samples are needed to fill the display, the oscilloscope must sample 1024 cycles of the waveform.
Therefore, only repetitive waveforms should be observed
in this mode. Irregularities that are present on an otherwise
repetitive waveform are not likely to show up when the
Equivalent Time Samplling method is used. With only one
sample being taken during each cycle, it is very likely that
glitches and other irregularities will be skipped over.
21
SERVICE INFORMATION
Warranty Service: Please return the product in the original packaging with dated proof of purchase to the address below.
Clearly state in writing the performance problem and return any leads, connectors, and accessories that you are using with the
device.
Non-Warranty Service: Return the product in the original packaging to the address below. Clearly state in writing the
performance problem and return any leads, connectors, and accessories that you are using with the device. Customers not on
open account must include payment in the form of a money order or credit card. For the most current flat-rate repair charges,
contact the factory before shipping the product.
Return all merchandise to B+K PrecisionCorp.withpre-paid shipping. The flat-rate repair chargeincludesreturn shipping
to locations in North America. For overnight shipments and non-North America shipping fees, contact B+K PrecisionCorp.
IMPORTANT
Includewiththe instrument your completereturnshipping address, contactname,phone number,and description ofproblem.
Service Department
B+K Precision Corp.
1031 Segovia Circle
Placentia, CA 92870
Phone (714) 237-9220
Fax (714) 237-9214
E-mail: service@bkprecision.com
22
LIMITED WARRANTY
B+K Precision Corp. warrants to the original purchaser that its product, and the component parts thereof, will be free from
defects in workmanship and materials for a period of three years from the date of purchase.
B+K Precision Corp. will, without charge, repair or replace, at its option, defective product or component parts upon
delivery to an authorized B+K Precision service contractor or the factory service department, accompanied by proof of the
purchase date in the form of a sales receipt.
To obtain warranty coverage in the U.S.A., this product must be registered by completing and mailing the enclosed warranty
registration card to B+K Precision, 1031 Segovia Circle, Placentia, CA 92870 within fifteen (15) days from the date of
purchase.
Exclusions: This warranty does not apply in the event of misuse or abuse of the product or as a result of unauthorized
alterations or repairs. It is void if the serial number is altered, defaced or removed.
B+K Precision Corp. shall not be liable for any consequential damages, including without limitation damages resulting
from loss of use. Some states do not allow limitation of incidental or consequential damages, so the above limitation or
exclusion may not apply to you.
This warranty gives you specific rights and you may also have other rights which vary from state to state.
23
INFORMATION
One of the best tutorials on oscilloscopes in the industry. Valuable to those with little knowledge
of oscilloscopes as well as the experienced technician or engineer who wishes
to refresh their memory or explore new uses for oscilloscopes.