B-K Medical Falcon 2101 Service manual

5 (2)

Service Manual

Diagnostic Ultrasound System

Falcon 2101, Falcon 2101EXL,

Hawk 2102, Hawk 2102XDI, Hawk 2102EXL, Surgical Hawk 2102XDI/EXL

This manual covers all existing versions (as of May 2003) of the 2100 series scanners.

World Headquarters: Mileparken 34 • DK-2730 Herlev • Denmark

Tel.: +45 4452 8100 • Fax +45 4452 8199 • www.bkmed.com • e-mail: support@bkmed.com

1

General Information

 

 

2

Service Notes

 

 

3

Checking Procedure

 

 

4

Trouble Shooting

 

 

5

Adjustment Procedure

 

 

6

Mechanical Parts

 

 

7

Preventative Maintenance

 

 

8

Theory of Operation

 

 

9

System Interconnections

 

 

10

 

 

 

System Falcon 2101/2101EXL & Hawk 2102/2102XDI/2102EXL/Surgical Hawk

General Information

Section 1

GENERAL INFORMATION

CONTENTS

PAGE

1

 

Safety Aspects .......................................................................................................................................

1-2

1.1

Terms and Symbols used ..................................................................................................................

1-2

1.2

WARNINGS and CAUTIONS:...........................................................................................................

1-3

2

 

System Overview ...................................................................................................................................

1-4

2.1

Front View ..........................................................................................................................................

1-4

2.2

Rear Connectors ................................................................................................................................

1-5

2.3

System Configuration .........................................................................................................................

1-6

2.4

Configuration Upgrade .......................................................................................................................

1-6

2.5

Modules and PC Boards ....................................................................................................................

1-7

3

 

Revision System ....................................................................................................................................

1-8

4

Special Tools and Equipment...............................................................................................................

1-9

4.1

Checking Procedure (Section 3) ........................................................................................................

1-9

4.2

Troubleshooting (Section 4) .............................................................................................................

1-10

4.3

Adjustment Procedure (Section 5) ...................................................................................................

1-10

4.4

Mechanical Parts (Section 6) ...........................................................................................................

1-10

4.5

Preventative Maintenance (Section 7) .............................................................................................

1-10

4.6

Electrical Safety Test........................................................................................................................

1-10

BI2102-D

1-1

General Information

System Falcon 2101/2101EXL & Hawk 2102/2102XDI/2102EXL/Surgical Hawk

1 Safety Aspects

The Falcon 2101/ Hawk 2102 System complies with IEC 60601-1 safety class I.

The Falcon 2101/ Hawk 2102 System is classified as follows:

Array Input Module:

Type B (Body)

Single-element Transducer Module:

Type B (Body)

ECG Input Module:

Type BF (Body Floating)

1.1Terms and Symbols used

Throughout this manual the following terms are used to indicate a situation where safety precautions are required:

"WARNING":

Indicates a situation involving risk of injury or loss of life to personnel or

 

 

patient.

 

 

"CAUTION":

Indicates a situation involving risk of damage to the instrument or other

 

 

equipment connected.

 

 

 

 

 

 

Symbol

Name

Description

 

 

 

 

 

 

 

 

ATTENTION

Consult ACCOMPANYING User Guide (BB0340) when this sign is

 

 

 

 

encountered on the instrument, to avoid reducing its safety

 

 

 

 

 

 

 

 

Potential

Terminal connected to the chassis. Should be connected to

 

 

 

Equalisation

corresponding terminals on other equipment to eliminate potential

 

 

 

 

differences.

 

 

 

 

 

 

 

 

Protective

Additional Protective Earth

 

 

 

Earth

 

 

 

 

 

 

 

 

 

Type CF

CF: Isolated from earth. Maximum Patient Leakage Current under:

 

 

 

 

Normal Condition 10 A, Single Fault Condition 50 A

 

 

 

 

 

 

 

 

Type BF

BF: Isolated from earth. Maximum Patient Leakage Current under:

 

 

 

 

Normal Condition 100 A, Single Fault Condition 500 A

 

 

 

 

 

 

 

 

Type BF

BF, DEFIBRILLATOR-PROOF

 

 

 

 

 

 

 

 

Type B

B: Maximum Patient Leakage Current under: Normal Condition

 

 

 

 

100 A, Single Fault Condition 500 A

 

 

 

 

 

 

 

IP57

SEALING

Dust and immersion protected according to IEC Publication 529

 

 

 

 

 

 

 

 

Stand-by

Push button for switching the scanner from stand-by to active.

 

 

 

 

(The power supply cord is the means of separation from the main

 

 

 

 

power supply.)

 

 

 

 

 

 

 

 

Off

Main power supply off

 

 

 

 

 

 

 

 

On

Main power supply on

 

 

 

 

 

 

 

 

Non-ionising

Ultrasound Scanner emits acoustic radiation

 

 

 

radiation

 

 

 

 

 

 

 

 

 

STERILE

Device is in a sterile condition

 

 

 

 

 

 

Table 1. IEC safety symbols

1-2

BI2102-D.

System Falcon 2101/2101EXL & Hawk 2102/2102XDI/2102EXL/Surgical Hawk

General Information

1.2WARNINGS and CAUTIONS:

For your own and others safety please read the following carefully:

Warnings:

Opening the instrument can expose live parts.

Any work done on the open instrument with power on must only be done by B-K Medical or their authorised representatives, who are aware of the hazards involved.

Any repair on the 2101/2102 must be followed by an electrical safety test to verify a continuous safe operation of the system.

Only the original mains cable must be used

NEVER USE EXTENSION CABLES!!

The 2101/2102 contains a Lithium battery. Under no circumstances must this battery be removed or replaced by the user as there is danger of explosion.

Personal Safety:

Be aware that there may be a risk of infection due to contaminated equipment, especially puncture guides/needles and transducers. The following precautions should be taken:

At the hospital ask the staff to sterilise transducers and puncture guides before receipt. Consoles must be disinfected as recommended in the User Guide before any repair.

When working with possible infected equipment, use gloves especially if you have open wounds or scratches.

Possible infected equipment must be sterilised before handed over to customers. Follow the recommendations in the Transducer User Guide.

Always wash your hands after working with the equipment.

If you scratch yourself on contaminated equipment you should immediately contact the hospital staff or see a doctor.

Cautions:

Ensure that the two mains voltage selector switches on the rear panel of the power supply are set to match the actual mains voltage.

Always use correct fuses.

Switch off all equipment before connecting or disconnecting their interfaces. Failure to do so could damage the equipment.

A 2101/2102 set up to operate in the 90-132 Volt main voltage range, will be damaged if connected to 230 volt.

The power supply cord is the means of separating the 2101/2102 from the main power supply.

BI2102-D

1-3

General Information

System Falcon 2101/2101EXL & Hawk 2102/2102XDI/2102EXL/Surgical Hawk

2 System Overview

2.1Front View

1.

2.

3.

4. 10.

6.

9.

8.

5.

5.

7.

Identify the following main parts.

1.Monitor

2.Keyboard / TGC / Track Ball

3.Gel bottle holder

4.Array Transducer connector(2102)/"Parking"(2101)

5.Array Transducer connector

6.Single Transducer connector (option)

7.Trolley

8.ECG input (option)

9.Floppy Disk Drive

10.CD-ROM drive (option)

1-4

BI2102-D.

System Falcon 2101/2101EXL & Hawk 2102/2102XDI/2102EXL/Surgical Hawk

General Information

2.2Rear Connectors

Monitor Connector

Light pen Connector

Audio In (2102 only)

RGB out Connector

Audio Out (2102 only)

Configuration switch

Foot switch connector

RGB input connector

Doc Trigger output

CBVS out switch (B/W – colour)

RS 232 connector B

CBVS out (B/W – Colour) Connector

5V for PCU(Palm Control unit)

CBVS In Colour Connector

RS 232 connector A

CBVS out Colour Connector

S-VHS out Connector

S-VHS in Connector

2102, 2102XDI

All

Mains Input Fuses Ground connectors

Voltage selector

Light pen Connector Configuration Switch Composite out Connector VCR out Connector VCR in Connector

Monitor Connector

AUX out Power output for monitor 2101

Power output for monitor 2101EXL & 2102

2101

All

RGB out Connector Configuration Switch RGB input Connector

CBVS out switch (B/W – colour) CBVS out (B/W – Colour) Connector CBVS out Colour Connector

CBVS In Colour Connector

S-VHS out Connector

S-VHS in Connector

Light pen Connector

Monitor Connector (VGA)

VGA out Connector

Optional PC Backend: RS232 (MTF option) 5 V Output

12 V Output (CD-ROM drive) LAN Connector (future purposes) Foot switch Connector

Mouse (Trackball) Connector Mover Connector (option)

USB connector (CD-ROM drive)

2101/2102EXL

BI2102-D

1-5

General Information

System Falcon 2101/2101EXL & Hawk 2102/2102XDI/2102EXL/Surgical Hawk

2.3System Configuration

The Falcon 2101 and Hawk 2102 exists in the below listed versions:

Falcon 2101 for use in B-, and M-mode.

Falcon 2101 EXL use in B-, and M-mode and additional 3D mode

Hawk 2102 for use in B-, Doppler-, and M-mode.

Hawk 2102 XDI for use in B- (Incl. of True Echo Harmonics), Doppler-, and M-mode.

Hawk 2102 EXL for use in B- (Incl. of True Echo Harmonics), Doppler-, and M-mode and additional 3D mode

Surgical Hawk 2102 XDI for use in B- (Incl. of True Echo Harmonics), Doppler-, and M- mode equipped with a Rack Docking System , PCU (Palm Control unit), and PIP(Picture In Picture) functionality

Surgical Hawk 2102 EXL for use in B- (Incl. of True Echo Harmonics), Doppler-, and M- mode equipped with a Rack Docking System , PCU (Palm Control unit), and PIP(Picture In Picture) functionality and additional 3D mode

2.4Configuration Upgrade

The following upgrade kits are available:

Valid for all models:

UA1254-K – ECG module upgrade kit

UA1255-K – Single Input module for mechanical transducers

UA1270-K – Palm Control Unit Kit for scanners delivered before April 2001

Hawk 2102:

UA1275-K - Harmonic upgrade kit, valid for Hawk 2102 s/n 1824377 or higher

Hawk 2102XDI:

UA1276-K – PIP upgrade kit, valid for Hawk 2102XDI

Hawk 2102EXL:

UA1290 PIP upgrade kit, valid for Hawk 2102EXL

1-6

BI2102-D.

System Falcon 2101/2101EXL & Hawk 2102/2102XDI/2102EXL/Surgical Hawk

General Information

2.5Modules and PC Boards

 

Model

2101

2101EXL

2102

2102XDI

Surgical

2102EXL &

Board/

 

 

 

 

Hawk

Surgical

subassembly

 

 

 

 

2102XDI

2102EXL

Monitor

ZV0049

ZV0065

ZV0050/512

ZV0060

ZV0060

ZV0065

Video Converter/Mixer board

 

ZH0743

ZH0691

ZH0691

ZH0699

ZH0743

Video Connector board

 

ZH0745

ZH0692

ZH0692

ZH0722

ZH0745

VGA connector board

 

ZH0744

 

 

 

ZH0744

Core board

ZD0753

ZD0771

ZD0762/673

ZD0762/673

ZD0762/673

ZD0767

Front-end board

ZE0731

ZE0731

ZE0724

ZE0724

ZE0724

ZE0724

Delay board

ZE0726

ZE0774

ZE0725

ZE0768

ZE0768

ZE0772

Doppler board

 

 

ZD0758

ZD0758

ZD0758

ZD0758

Motherboard

ZH0675/95

ZH0675/95

ZH0695

ZH0695

ZH0695

ZH0695

Power supply

ZG0341

ZG0341/48

ZG0341

ZG0341

ZG0341

ZG0341/48

Keyboard assembly

ZN0006

ZN0006

ZN0009

ZN0009

ZN0009

ZN0009

-

Potentiometer Board

ZH0676

ZH0676

ZH0676

ZH0676

ZH0676

ZH0676

-

Gain Board

ZH0678

ZH0678

ZH0678

ZH0678

ZH0678

ZH0678

-

Tracker Ball

NT0254

NT0254

NT0254

NT0254

NT0254

NT0254

Floppy Drive

UL0018

UL0018

UL0018

UL0018

UL0018

UL0018

ECG Module (UA1254 Option)

ZE0762

ZE0762

ZE0762

ZE0762

ZE0762

ZE0762

Single Module (UA1255 Option)

ZH0709

ZH0709

ZH0709

ZH0709

ZH0709

ZH0709

 

 

 

 

 

 

 

PC-Backend (6503 Option)

N/A

ZN0377

N/A

N/A

N/A

ZN0377

-

Single board PC

N/A

ZN0042

N/A

N/A

N/A

ZN0042

- Frame grabber with connect.

N/A

ZN0043

N/A

N/A

N/A

ZN0043

-

Harddisk

N/A

UL0023

N/A

N/A

N/A

UL0023

-

Power supply

N/A

ZG0345

N/A

N/A

N/A

ZG0345

-

Connection board

N/A

ZH0751

N/A

N/A

N/A

ZH0751

-

MCM base

N/A

ZH0752

N/A

N/A

N/A

ZH0752

- MCM Motor Control (option)

N/A

ZH0753

N/A

N/A

N/A

ZH0753

-

CD/RW

N/A

ZN0041

N/A

N/A

N/A

ZN0041

-

Trackball for PC-Backend

N/A

NT0262

N/A

N/A

N/A

NT0262

-

Footswitch for PC-Backend

N/A

UA1215

N/A

N/A

N/A

UA1215

2ZV0050=10” / ZV0051=12”

3ZD0762-sw. < 3.00 / ZD0767 sw. 3.00

BI2102-D

1-7

General Information

System Falcon 2101/2101EXL & Hawk 2102/2102XDI/2102EXL/Surgical Hawk

3 Revision System

All PC Boards in the 2101/2102 are described by the board name, Type number, Revision code (ID) and PCB Version.

The name of the board indicates the function of the board, for example the Delay Board. The name of the board is changed only if the function of the board changes.

Part number

Serial Number

Revision level in this case: E

| | | || |

| ||| | | |

| | |||

| ||| |

ZD0753

001 837

B-K Medical A/S

ABCDEFGHIJKLMNOPQRSTUVXY

Bar Code Label

The Part number of the board, for example ZD0753, is the order number of the board. The number is changed if the modification of the existing PCB is found to be too extensive.

The serial number of the board is a unique number used for tracking purposes.

The revision code (ABCDEFGH.......) is related to the modifications made on the board. The revision letter is marked when modifications are made. When boards are ordered from the B-K Medical stock it is important that the serial number of the scanner is stated. The serial number ensures that the correct revision is shipped.

The PCB version (number e.g. 3) is printed on the circuit board.

Information about Hardware/ Software compatibility can be found in the B-K Medical Hot Line on the support section of www.bkmed.com (requires password)

1-8

BI2102-D.

System Falcon 2101/2101EXL & Hawk 2102/2102XDI/2102EXL/Surgical Hawk

General Information

4 Special Tools and Equipment

The tools and equipment listed below does not include standard tools and commonly used equipment.

4.1Checking Procedure (Section 3)

Transducer Phantom, model 254 (WQ 0973) for checking a 2101/2102 system equipped with Arrayand Single-element Transducers.

Transducer Phantom, type 251 (WQ 0972) for checking a 2101/2102 system equipped with the Endosonic Probe 1850.

Ruler (for checking the size of the image on the monitor)

Floppy disk 3.5" (for testing the floppy disk drive)

Blank CD-R (for testing PC Back-end CD Drive – option)

Further it is recommended to bring a 150 ml bottle containing preserving fluid for the phantom(s):

Prescription for preserving fluid:

94 ml Glycerol (85%)

50 ml Rodalon (10%)

1000 ml purified Water

BI2102-D

1-9

General Information

System Falcon 2101/2101EXL & Hawk 2102/2102XDI/2102EXL/Surgical Hawk

4.2Troubleshooting (Section 4)

For troubleshooting the power supply

DVM

4.3Adjustment Procedure (Section 5)

For adjusting the monitor:

Nonmagnetic screwdriver

Ruler (for checking the size of the image on the monitor)

Colour Analyser (for 2102 and 2101EXL monitor) For adjusting the monitor friction:

4 mm Allen key

4.4Mechanical Parts (Section 6)

Static Control Service Kit, type 3M 8501 (WQ 0969) or similar. (when handling the static sensitive PCB’s).

TORX key size 10 and 20. A short TORX key size 20 is necessary in order to get easy access to some screws on the monitor assembly.

Wrench key for Video converter mounted on the Core Board or CD ROM drive for PC Back-end

4.5Preventative Maintenance (Section 7)

Equipment necessary to perform the Preventative Maintenance is the equipment used in the Checking Procedure and Electrical Safety Test.

4.6Electrical Safety Test

Testers required:

Safety Tester

HV Tester

WB 1275 HV test adaptor for type BF Transducers

High Voltage test plugs required:

WJ 0246 HV Test Plug for Mains

WJ 0287 HV Test Plug for ECG (for testing ECG Input Module)

Electrical Safety Test Record (Enclosed)

1-10

BI2102-D.

2101/2102 Electrical Safety Test Record

Equipment under test

Serial no:

Location:

 

 

 

For INSTRUCTIONS read the back of this sheet.

 

 

 

 

Test Date

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Test

Test

 

 

 

 

 

 

 

Typical

Test Limits

Select

 

 

 

 

 

 

 

 

value

 

1

Rigel self test

-

 

 

 

 

 

-

-

2

Mains

-

 

 

 

 

 

-

Nominal +/-

 

Voltage

 

 

 

 

 

 

 

 

10%

3

Insulation

Mains to Case

 

 

 

 

 

 

 

5 MΩ

Resistance

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5

Pr. gnd

Console

 

 

 

 

 

0,1

0,2 ohm

(earth terminal)

 

 

 

 

 

 

ohm

 

 

 

 

 

 

 

 

 

5

Pr. gnd

Transducer

 

 

 

 

 

0,1

0,2 ohm

connector –array

 

 

 

 

 

 

ohm

 

5

Pr. gnd

Video shelf

 

 

 

 

 

0,1

0,2 ohm

 

 

 

 

 

 

 

 

 

ohm

 

5

Pr. gnd

Single module

 

 

 

 

 

0,1

0,2 ohm

 

 

connector

 

 

 

 

 

 

ohm

 

5

Pr. gnd

Monitor

 

 

 

 

 

0,1

0,2 ohm

 

 

 

 

 

 

 

 

 

ohm

 

5

Pr. gnd

ECG

 

 

 

 

 

0,1

0,2 ohm

 

 

(Potentiometer button)

 

 

 

 

 

 

ohm

 

6

Earth

Norm.

 

 

 

 

 

 

40 µA

500 µA

 

Leakage

 

 

 

 

 

 

 

 

 

6

Earth

Rev.

 

 

 

 

 

 

80 µA

500 µA

 

Leakage

 

 

 

 

 

 

 

 

 

7

Earth

Norm.

 

 

 

 

 

 

2 µA

1000 µA

 

Leakage

 

 

 

 

 

 

 

 

 

7

Earth

Rev.

 

 

 

 

 

 

120 µA

1000 µA

 

Leakage

 

 

 

 

 

 

 

 

 

8

Enclosure

Norm.

 

 

 

 

 

 

 

100 µA

 

Leakage

 

 

 

 

 

 

 

 

 

8

Enclosure

Rev.

 

 

 

 

 

 

 

100 µA

 

Leakage

 

 

 

 

 

 

 

 

 

11

Patient

Norm.

 

 

 

 

 

 

 

100 µA

 

Leakage

 

 

 

 

 

 

 

 

 

11

Patient

Rev.

 

 

 

 

 

 

 

100 µA

 

Leakage

 

 

 

 

 

 

 

 

 

A

2.2 kV (DC)

Mains

 

 

 

 

 

 

 

No Flash over

B

2.2 kV (DC)

ECG

 

 

 

 

 

 

 

No Flash over

 

 

Signature

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1 of 2

The test complies with IEC 60601-1 regulations for medical equipment, safety class I, Type B (ECG type BF).

Important:

High Voltage testers deliver hazardous currents. Therefore these testers should only be operated by technicians who are aware of the hazards involved.

Disconnect transducers and accessories from the equipment under test. Test limits are valid only for the equipment itself without any applied part(s) connected.

The complete test must always follow immediately after a repair made on the 2101/2102 Ultrasound Scanner, and always on the fully assembled unit.

Protective GND: General remarks and settings for Test no. 5:

Tester:

Rigel 233

Cables:

Mains cable for scanner; clip/probe lead (Rigel accessories)

Basic settings:

2101/2102 mains switch ON; Rigel switch settings: Class=I, Type: B (For ECG test type BF)

Calibrate Rigel:

Connect clip/probe lead between PROBE and GND in IUT POWER socket; Press TEST and zero ohm-meter.

Connect the scanners Main cable from IUT POWER (Rigel) to power inlet on the scanner.

1.

Connect the clip/probe lead between PROBE (Rigel) and the left GND connector on the back of the scanner.

 

Press the TEST button and record the resistance. Disconnect the power cable for the monitor.

Note:

As the test current is 25 Amps. it is important that the clip/probe lead is held firmly against, or clipped to the GND

 

connector before the TEST button is pressed !

2.

Move the clip/probe to the metal on one of the array transducer connectors. Press the TEST button and record the

 

resistance.

3.

Move the clip/probe to on of the metal screws holding the video shelf on the scanner. Press the TEST button and

 

record the resistance.

4.

If single module installed: Move the clip/probe to the metal ring on the single module connector (if installed). Press

 

the TEST button and record the resistance.

5.

Move the clip/probe to on of the metal on the back of the monitor. Press the TEST button and record the resistance.

6.

If ECG installed: Set the Rigel tester to type BF. Remove the top on one of the potentiometer buttons on the ECG

 

module. Move the clip/probe to the nut on the potentiometer button. Press the TEST button and record the

 

resistance. Set the Rigel tester to type B again.

Leakage Current: General remarks and settings for test no. 6 and 7.

Test 6 measures Earth Leakage current – normal condition

Test 7 measures Earth Leakage current – single Fault Condition (supply open)

Tester:

Rigel 233

Cables:

Mains cable for scanner;

Basic settings:

2101/2102 mains switch ON; Rigel switch settings: Class = I, Type: B,

1.

Connect the mains cable for the scanner to the IUT POWER outlet (Rigel) to power inlet on the scanner

2.

Position the Rigel Normal-Reverse switch in Normal and record the current.

3.

Position the Rigel Normal-Reverse switch in Reverse and record the current.

4.

Repeat the above test for test no. 7.

High Voltage Test: General remarks and settings for A and B

Tester:

HV Insulation Tester JP15

Cables:

HV probe and GND lead (JP15 accessories)

Adaptors:

HV test plug for mains WJ0246, HV test plug for ECG Module WJ0287

Connections:

Insert the HV test plug for mains into socket for mains input, and HV test plug for ECG Module into ECG

 

socket. Connect the JP15 GND with the scanners BNC GND using the JP15 GND lead.

Basic settings:

Mains switch on scanner ON; on JP15: Connect HV probe cable with switch plug into socket "HT"/"sw". Select

 

1 A and 7,5 kV. "Volume" to mid position and HV potentiometer fully anti-clockwise.

Test A:

Insert tip of HV probe into the HV test plug for mains. Press button on HV probe handle while increasing the HV to

 

2.2 kV (DC) read from the voltmeter. Apply this voltage for maximum 5 sec. There must be no flash over nor

 

breakdowns indicated by full deflection on the A-meter.

Test B:

Insert tip of HV probe into HV test plug for ECG Module and repeat above test.

2 of 2

System Falcon 2101/2101EXL & Hawk 2102/2102XDI/2102EXL/Surgical Hawk

Service Notes

Section 2

SERVICE NOTES

Service Notes

To improve the performance of the Falcon and Hawk Ultrasound System 2101/2102, small changes in the hardware and firmware may be made. All information about these changes, and any other change or correction to this Service Manual will be released in the form of Service Notes in the Medical Hotline.

Note

All existing and future Service Notes concerning the 2101 Falcon or 2102 Hawk system should be placed in this section.

Edition Number

If any major changes are made to the 2101 Falcon or 2102 Hawk system and/or the

Service Manual, a new edition will be released.

Any new editions can be recognized by the edition number given in the footer.

The serial numbers from which each edition is valid are listed on the front page.

BI2102-D

2-1

Service Notes

System Falcon 2101/2101EXL & Hawk 2102/2102XDI/2102EXL/Surgical Hawk

This page is left blank intentionally

2-2

BI2102-D

System Falcon 2101/2101EXL & Hawk 2102/2102XDI/2102EXL/Surgical Hawk

Checking Procedure

Section 3

 

CHECKING PROCEDURE

CONTENTS

PAGE

3.1

Introduction ...................................................................................................................................................

2

3.1.1.

 

Necessary Equipment to Perform the Checking Procedure...............................................................

2

3.2.

Basic Checks ................................................................................................................................................

2

3.2.1.

 

Visual Inspection .................................................................................................................................

2

3.2.2.

 

Monitor................................................................................................................................................

3

3.2.3.

 

Before Power Up ................................................................................................................................

3

3.2.4.

 

Power Up ............................................................................................................................................

4

3.2.5.

 

Test Oscillators...................................................................................................................................

5

3.3.

Monitor Checking.......................................................................................................................................

14

3.4.

Transducer Inputs......................................................................................................................................

14

3.5.

Track Ball ....................................................................................................................................................

15

3.6.

Gain/TGC.....................................................................................................................................................

15

3.7.

Size...............................................................................................................................................................

15

3.8.

Focusing......................................................................................................................................................

16

3.9.

Frame Rate..................................................................................................................................................

16

3.10.

 

Image Processing...................................................................................................................................

16

3.11.

 

Persistence..............................................................................................................................................

16

3.12.

 

Puncture ..................................................................................................................................................

16

3.13.

 

Zoom ........................................................................................................................................................

16

3.14.

 

Measure ...................................................................................................................................................

17

3.15.

 

Image review ...........................................................................................................................................

17

3.16.

 

M-mode....................................................................................................................................................

17

3.17.

 

Split Screen.............................................................................................................................................

17

3.18.

 

Image Storage.........................................................................................................................................

17

3.19.

 

Audio/Spectral Doppler (only 2102) .....................................................................................................

18

3.20.

 

CFM (only 2102)......................................................................................................................................

18

3.21.

 

PIP (only for 2102 with the PIP option)................................................................................................

19

3.22.

 

3D system (6503 option)........................................................................................................................

20

3.22.1. Basic check – start up & signal transfer to/from scanner .................................................................

20

3.22.2.

Acquisition ........................................................................................................................................

20

3.22.3.

CD-RW drive ....................................................................................................................................

21

3.22.4.

ECRM option ....................................................................................................................................

22

3.22.5.

Pull back option ................................................................................................................................

23

3.22.6.

2050 3D scanning.............................................................................................................................

23

3.22.7.

MTF (Magnetic Tracked Freehand) option.......................................................................................

24

BI2102-D

3-1

Checking Procedure

System Falcon 2101/2101EXL & Hawk 2102/2102XDI/2102EXL/Surgical Hawk

3.1.Introduction

The Checking Procedure is to verify proper operation of the main functions of the 2101/2102 System.

The procedure should be used during installation, incoming inspection, regular maintenance and before and after repair.

The procedure is arranged in sections and must be followed from the beginning when performed.

Notes

If the Checking Procedure cannot be performed successfully or if there is a specific fault in the System, refer to Section 4, TROUBLESHOOTING or Section 5, ADJUSTMENT PROCEDURE.

The Checking Procedure does not include a check of the different transducer types.

If in doubt about the functions of the 2101/2101EXL/2102/2102XDI/2102EXL you should consult the user guide BB0340/BB1086/BB0360/BB0950/BB1075.

Important!!

Any inside repairs performed on the 2101/2102 must be followed by an Electrical Safety

Test.

3.1.1.Necessary Equipment to Perform the Checking Procedure

Transducer Phantom, model 254 (WQ 0973) for checking a 2101/2102 system equipped with Arrayand Single-element Transducers.

Transducer Phantom, type 251 (WQ 0972) for checking a 2101/2102 system equipped with the Endosonic Probe 1850.

Ruler (for checking the size of the image on the monitor)

Floppy disk 3.5" (for testing the floppy disk drive)

Array Transducer (Linear or Convex).

Sector Transducer (for systems equipped with Single-element module)

Note: If a phantom is not available the human body is a good alternative.

3.2.Basic Checks

3.2.1.Visual Inspection

1.Check the overall appearance of the 21/012102.

Especially check for scratches and stain's on the paint.

3-2

BI2102-D

System Falcon 2101/2101EXL & Hawk 2102/2102XDI/2102EXL/Surgical Hawk

Checking Procedure

3.2.2.Monitor

1.Check that the monitor is securely fastened on its base and is easy to rotate.

3.2.3.Before Power Up

1.Check that the two voltage selector switches found on the rear panel of the Power Supply are set for the actual mains voltage.

2.Check the rating of the primary fuses:

If 230 V AC:

T 6.3 A H, 250V (Time lag high breaking capacity)

If 115 V AC:

T 8 A H, 250V (Time lag high breaking capacity)

3.The 2101/2102 System complies with the IEC 60601-1 safety regulations and as such it must be connected to a mains outlet having a safety ground connection. Therefore: Check that the 2101/2102 mains cable is terminated with a three-pole plug that fits a three-pole mains outlet. For safety reasons do not use extension cables!

BI2102-D

3-3

Checking Procedure

System Falcon 2101/2101EXL & Hawk 2102/2102XDI/2102EXL/Surgical Hawk

3.2.4.Power Up

1.Disconnect all accessories from the 2101/2102 and connect any transducer available which fits the Array Input Module.

2.Switch On the 201/2102.

After a few moments the power up (boot) sequence is completed.

3.Verify that the monitor opening layouts are shown in Fig. 3.3 -1. Note that the menu shown (settings) differs from transducer type.

Fig. 3.3-1. The default display.

4.Check that the following is displayed on the monitor:

the current date and time

a white “Asteric” (“Freeze”-mode)

Note: If the displayed date and time is incorrect, press Setup to display the Setup menu.

Press F2 to display the Date /Time menu and use the track ball to go to the

field to adjust. Press to toggle the menu’s. Press F1 for Next Page

To exit the menu press Setup.

3-4

BI2102-D

System Falcon 2101/2101EXL & Hawk 2102/2102XDI/2102EXL/Surgical Hawk

Checking Procedure

3.2.5.Test Oscillators

1.Connect an electronic transducer to the 2101/2102.

2.Switch On the scanner.

3.Press Shift, Alt and to select the Test menu

4.Select the Scancon Test oscillator by pressing F3 and unfreeze

Toggle between Scancon test 1,2 and 3 by pressing F3, and compare to the images below. To bring up the test menu again, press Shift, Alt and Note that the actual image on the scanner depends on various settings e.g. size and transducer type.

Fig. 3.3-2. Scancon Test Osc. 1 and 2. Check that image is going from black to white

Fig. 3.3-3. Scancon Test Oscillator 3 - Greyscale horizontally.

Note ! this image might flicker dependable on transducer type and size of image

BI2102-D

3-5

Checking Procedure

System Falcon 2101/2101EXL & Hawk 2102/2102XDI/2102EXL/Surgical Hawk

5.Exit the Scancon Test by keeping F3 depressed for 1 second.

6.Select the Input Test Oscillator by pressing F1. Compare to the image below.

Fig. 3.3-4. Input Test Oscillator.

7.Exit the Input Test Oscillator by pressing F1.

3-6

BI2102-D

System Falcon 2101/2101EXL & Hawk 2102/2102XDI/2102EXL/Surgical Hawk

Checking Procedure

8.Switch off the scanner and disconnect all transducers from the 2101/2102.

9.Turn on the scanner and after the boot sequence press Shift, Alt and .

10.Select the Delay Test oscillator by pressing F2 and unfreeze The image should now look like the figure below i.e. concentric bands which gradually goes from black to white (light grey) and then back to black.

Note ! The actual number of concentric bands depends on the size selected.

Fig. 3.3-5. Delay Test Oscillator 1

BI2102-D

3-7

B-K Medical Falcon 2101 Service manual

Checking Procedure

System Falcon 2101/2101EXL & Hawk 2102/2102XDI/2102EXL/Surgical Hawk

11.Press Shift, Alt and to access the Test Oscillator again.

12.Press F2 to start Delay test oscillator 2. The image should now look like the figure below i.e. 6/8 concentric bands which consist of a number of white (light grey)blocks separated by 48/64 narrow black radial lines.

Note ! If bands are missing from the sides it could be because of the user setup. In that case make a backup of the 8802 set-up on a floppy disk and reset the 8802 set-up. Remember to load the customer setup again after the test.

Fig. 3.3-6. Delay Test Oscillator 2 on 2101

Fig. 3.3-7. Delay Test Oscillator 2 on 2102

3-8

BI2102-D

System Falcon 2101/2101EXL & Hawk 2102/2102XDI/2102EXL/Surgical Hawk

Checking Procedure

13.Turn the gain down to see if all bands can be adjusted to dark grey and turn the gain up again to see if the bands appears white again.

Fig. 3.3-8. Delay Test Oscillator 2 with gain turned down – test image should fade out

.

BI2102-D

3-9

Checking Procedure

System Falcon 2101/2101EXL & Hawk 2102/2102XDI/2102EXL/Surgical Hawk

14.Press F2 to start Delay test oscillator 3. The image should now look like the figure below i.e. 24/32 radial fields separated by narrow white lines. The fields should be white from top to bottom and then gradual decrease in length going from right to left.

Note ! On 2101 the left line should be a narrow line where as on the 2102 it should be a wide field.

Fig. 3.3-9. Delay Test Oscillator 3 (2101).

Fig. 3.3-10. Delay Test Oscillator 3 (2102).

15.Turn the gain down to see if all bands can be adjusted to dark grey and turn the gain up again to see if the bands appears white again.

16.Switch off the scanner

3-10

BI2102-D

 

System Falcon 2101/2101EXL & Hawk 2102/2102XDI/2102EXL/Surgical Hawk

Checking Procedure

2102 Only !

17.

Connect an electronic transducer to the 2102 and switch on the scanner.

(for 2101 continue

18.

Press CFM

 

with step 24)

 

19.

Select the test menu by pressing Press Shift, Alt and .

 

 

 

 

 

 

20.

Select DSP osc. by pressing F4 and then press Freeze and compare to the

 

 

 

figure below.

 

Fig. 3.3-11. Doppler test Oscillator 1.

BI2102-D

3-11

Checking Procedure

System Falcon 2101/2101EXL & Hawk 2102/2102XDI/2102EXL/Surgical Hawk

21.Press Shift, Alt and to activate the Test menu.

22.Press F4 to activate Test oscillator Disp. Osc.2

23.Press Freeze twice (freeze and then unfreeze) and compare to the Figure below.

Fig. 3.3-12. Doppler test Oscillator 2.

3-12

BI2102-D

System Falcon 2101/2101EXL & Hawk 2102/2102XDI/2102EXL/Surgical Hawk

Checking Procedure

24.If the 2101/2102 has a single input module then switch off the scanner and disconnect all transducers from the 2101/2102.

25.Turn on the scanner.

26.Start the Single Input Test Oscillator by pressing Shift, Alt, and Z, select e.g. 8539 (F2) and unfreeze. Compare to the image below.

Fig. 3.3-13. Single Input Test Oscillator.

27.Exit the Single Input oscillator by pressing Shift, Alt, and Z.

28.If the 2101/2102 has an ECG option then connect a transducer and unfreeze the image.

29.Press S (HR) to get the ECG Curve on the screen.

30.Select the ECG test by pressing Shift, Alt, S. Compare to the image below. Note ! The amplitude and position of test curve depends of the setting of the ECG potentiometers.)

Fig. 3.3-14. ECG Test Oscillator

BI2102-D

3-13

Checking Procedure

System Falcon 2101/2101EXL & Hawk 2102/2102XDI/2102EXL/Surgical Hawk

3.3.Monitor Checking

Press Shift, Alt, and G, to get a test picture on the monitor screen (The scanner must be in freeze mode).

1.One for brightness with two dark squares where one is 2/255 and the other is 4/255 of 100% white,

Brightness is adjusted so the 2/255 part disappear and the 4/255 is just become visible.

2.One for contrast with one square filled out white 100% white (255/255). Contrast has to be adjusted to 60 cd/m2 measured in the centre of the square

3.Test picture displaying a linear 8 step greyscale for final check of monitor adjustment

Use <Alt><Shift><G> to toggle through the test pictures. 1.

2.Verify that the external controls for Brightness and Contrast (on the monitor) provides sufficient adjustment range and set to give the best image.

3.Press the Shift, Alt and . Then press F5 to chose the Test picture. (Pressing F5 toggles between Geometric test picture, all pixel off, all pixels on)

2101-197 mm

2102 -10”:176mm, 12”:189mm, 14”:221mm 15:228mm

2101: 149 mm –10”:133mm, 12”:143mm,

14”:167mm 15”:171mm

2102

 

Fig. 3.3-15. The Geometric Test Picture..

4.Check the geometry of the test Picture.

5.If you have problems meeting the above requirements, consult Section 5, ADJUSTMENT PROCEDURE.

3.4.Transducer Inputs

1.

Move the transducer connector to the other input module(s), Press

and

 

confirm that the transducer can be identified.

 

2.Start the transducer to verify correct operation of the input module.

Note:

If available connect a Single-element transducer to the Single element socket

 

and verify correct operation of the transducer.

3-14

BI2102-D

Loading...
+ 140 hidden pages