Bio-Rad Vacuum Manifold User Manual

Page 1
Aurum
Vacuum Manifold
Instruction Manual
For technical support, call your local Bio-Rad office, or in the US, call 1-800-4BIORAD (1-800-424-6723).
4110115 Rev B
Bulletin 0000 Rev A US/EG
Bio-Rad Laboratories, Inc.
Web site www.bio-rad.com USA 800 4BIORAD Australia 61 02 9914 2800 Austria 01 877 89 01 Belgium 09 385 55 11 Brazil 55 21 3237 9400 Canada 905 364 3435 China 86 21 6426 0808 Czech Republic 420 241 430 532 Denmark 44 52 10 00 Finland 09 804 22 00 Greece 30 210 777 4396 Hong Kong 852 2789 3300 Hungary 36 1 455 8800 India 91 124 4029300 Israel 03 963 6050 Italy 39 02 216091 Japan 03 6361 7000 Korea 82 2 3473 4460 Mexico 52 555 488 7670 The Netherlands 0318 540666 New Zealand 0508 805 500 Norway 23 38 41 30 Poland 48 22 331 99 99 Portugal 351 21 472 7700 Russia 7 495 721 14 04 Singapore 65 6415 3188 South Africa 27 861 246 723 Spain 34 91 590 5200 Sweden 08 555 12700 Switzerland 061 717 95 55 Taiwan 886 2 2578 7189 United Kingdom 020 8328 2000
France 01 47 95 69 65 Germany 089 318 84 0
00-0000 0000 Sig 0308
Page 2
Table of Contents
Section 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Section 2 Kit Components . . . . . . . . . . . . . . . . . . . . . . . . .1
Section 3 Care and Maintenance . . . . . . . . . . . . . . . . . . . .2
Section 4 Aurum™ Vacuum Manifold . . . . . . . . . . . . . . . . .3
Section 5 Vacuum Regulator . . . . . . . . . . . . . . . . . . . . . . .4
Section 6 Vacuum Manifold Setup . . . . . . . . . . . . . . . . . . .5
Section 7 Use for 96-Well Plates . . . . . . . . . . . . . . . . . . . . . . . .6
Section 8 Use for the Column Adaptor Plate (CAP) . . . . . .7
Section 9 Troubleshooting Guide . . . . . . . . . . . . . . . . . . . .8
Section 10 Ordering Information . . . . . . . . . . . . . . . . . . . .10
Page 3
Section 1 Introduction
The Aurum™ vacuum manifold and column adaptor plate (CAP)` comprise a system which allows a single vacuum manifold to be used with either 96-well plates or columns with luer ends. Separate manifolds are no longer required for different preparative formats; the Aurum vacuum manifold and CAP ensure fast, high-quality sample preparation in both formats, while maintaining the simplicity of handling associated with vacuum processing. The uniquely designed vacuum regulator allows for complete control of negative pressure within the Aurum vacuum manifold.
Section 2 Kit Components
The Aurum™ vacuum manifold contains the following components as shown in Figure 1:
Vacuum manifold consisting of:
- manifold top
- manifold base
Column adaptor plate
Vacuum regulator
Stage
Luer caps
Tubing (
5
/
16
ID)
1
Page 4
Section 3 Care and Maintenance
The Aurum™ vacuum manifold should be stored clean and dry at room temperature. After each use rinse with distilled water and wipe dry with paper towels or air dry. Never use alcohol or other organic reagents to clean the manifold. Never use abrasive materials to clean the manifold.
Gasket surfaces should be carefully rinsed to remove salts, which could inhibit contact between manifold parts and the column adaptor plate (CAP). Check the tubular and flat silicone gaskets for wear before and after use of the manifold. In the event that the bottom tubular gasket requires replacement, a length of
1
/
8
" OD silicone tubing may be used.
2
Section 4 Aurum™ Vacuum Manifold
The Aurum vacuum manifold is designed to accommodate the 96-well plates used in Bio-Rad's Aurum total RNA 96 kit. The manifold may also be compatible with other manufacturers' kits. For column processing, the column adaptor plate (CAP) is required.
The CAP allows the Aurum vacuum manifold to to be used for processing columns with a luer fitting. The CAP has 18 black luer fittings in a 6 x 3 array. Up to 18 Aurum miniprep columns can be accommodated on the CAP without the need for connectors or other manifold accessories. The CAP will also accommodate other columns with luer ends.
When vacuum is applied to the manifold, the CAP should self-seat, forming an airtight seal without the need to press down on the CAP. However, the application of gentle downward force may occasionally be required to facilitate seating.
The vacuum regulator is included and strongly recommended for measuring negative pressure, and for gradually increasing negative pressure during different protocols.
The Aurum manifold consists of two primary parts: the manifold base and top. The top has an attached tubular silicone gasket, allowing the manifold base and top to seal when a vacuum is introduced. Attached to the opening in the manifold top is a flat silicone gasket, enabling Aurum 96-well purification plates or the CAP to establish an airtight seal with the manifold top when a vacuum is applied. Collectively these two gaskets facilitate self-seating of all components, reducing the need to apply external pressure to the manifold to initiate vacuum. However, under certain conditions application of gentle downward force on the plates or CAP may be required to facilitate seating. In the event that the tubular silicone gasket becomes detached, it can easily be reinserted into the track on the underside of the manifold top.
Each Aurum vacuum manifold is supplied with a stage that centers and elevates a 96-well sample collection plate during elution from the purification plate. The stage should be placed with the printed "this side up" in the base with the groove side down to allow for airflow.
3
Fig. 1. Aurum vacuum manifold kit components.
Manifold top
Manifold base with
quick connect fitting
Vacuum regulator
A stage
Luer caps
Column adaptor plate (CAP)
Tubing (
5
/
16
ID)
Stage Waste collection tray
Page 5
Section 5 Vacuum Regulator
The vacuum regulator (see Figure 2) allows complete control of the negative pressures inside the Aurum™ vacuum manifold, allowing optimum levels of negative pressure for efficient sample preparation. This is particularly important when a gradual increase in negative pressure is required during 96-well plate purifications. Attached in-line between the vacuum source and filter flask, the regulator measures the difference between ambient pressure and pressure within the vacuum manifold in inches of mercury (inHg). The regulator is capable of measuring negative pressures up to –30 inHg. See Table 1 for pressure unit conversions.
At various points in the protocols, it is necessary to release the vacuum inside the manifold so that plates and columns can be removed. To release the negative pressure, rotate the regulator dial fully counter-clockwise until the gauge reads < 1 inHg. The vacuum is restored inside the manifold by rotating the control dial clockwise until the desired negative pressure is attained.
The regulator control dial is located off-center, toward the right side of the regulator. The vacuum source is attached to the right hose fitting and the vacuum manifold is attached to the left fitting (see Figure 3).
Table 1. Pressure unit conversions
To convert from inches of mercury (inHg) to:
Multiply by: millimeters of mercury or torr (mmHg, torr) 25.4 millibar (mbar) 33.85 atmospheres (atm) 0.03342 pounds per square inch (psi) 0.4912 kilopascals (kPa) 3.385
Fig. 2. Vacuum regulator and gauge.
4
Section 6 Vacuum Manifold Setup
Tubing provided is 4 ft long and must be cut in appropriate pieces before proceeding.
Prior to setup, you may ensure that the gauge pointer is adjusted to zero by removing the lens cover, followed by turning the adjustment pin located beneath the dial face.
Vacuum setup
1. Cut tubing into three pieces of appropriate length.
2. Use one piece of tubing to connect the vacuum source to the right side of the vacuum regulator.
3. Use another piece of tubing to connect the left side of the vacuum regulator to the sidearm of the filter flask.
4. Place a rubber stopper with hole into the mouth of the filter flask. Insert a serological pipette (or comparable) into the hole of the stopper.
5. Snap in the black sealed end of the quick connect fitting into the manifold base.
6. Finally, use the last piece of tubing to connect the filter flask to the quick connect fitting of the nozzle of the manifold.
Fig. 3. Vacuum setup conditions.
5
Page 6
6 7
Section 8 Use for the Column Adaptor Plate (CAP)
Manifold wash setup (see Figure 5)
1. Insert the CAP (luer ends up) into the depression in the vacuum manifold top. Ensure that the CAP rests evenly on the gasket.
2. Insert the luer ends of the desired columns into the available luer fittings, ensuring a tight fit.
3. Close the unused luer fittings with the caps provided. Close caps by rotating clockwise until light resistance is encountered. Excessive tightening of a cap may cause the luer fitting to dislodge when the cap is removed.
4. The manifold is now ready for column processing according to the vacuum protocol of the appropriate column purification kit.
5. When ready to elute, proceed with the appropriate spin elution step as recommended by the protocol.
6. After finishing the elution, rinse the CAP and Aurum vacuum manifold with water, and air dry or wipe with paper towels.
Fig. 5. Setup and use for column processing.
Section 7 Use for 96-Well Plates
Manifold wash setup (see Figure 4A)
1. Place the manifold top on the base, ensuring complete and uniform contact between the manifold top and base.
2. Place the desired 96-well purification plate on the manifold top and apply the recommended vacuum pressure for your application.
Manifold elution setup (see Figure 4B)
3. When eluting from the purification plate, place the stage on the manifold base, with the etched "this side up" facing up. Place a clean 96-well microtiter collection plate securely on top of the stage and replace the manifold top.
Note: If desired, waste can be collected into an optional pipet tip box lid instead of draining directly into the vacuum trap.
4. After use, rinse the Aurum™ vacuum manifold with water and air dry or wipe with paper towels. Failure to rinse the vacuum manifold at the end of each use will cause the manifold to become cloudy and pitted.
Fig. 4. Setup and use for 96-well plate processing.
A) 96-well wash; B) 96-well elution.
A.
B.
Page 7
Section 9 Troubleshooting Guide
Problem Possible Cause Possible Solution
Insufficient negative Open vacuum regulator Close vacuum regulator pressure in vacuum by turning completely manifold clockwise
Manifold top not seated Ensure uniform contact properly on base between manifold top
and base
Purification plate or CAP Press down gently on not fully seated the corners of the plate
or CAP to seal
Unused CAP luer fittings Close unused luer not closed completely fittings completely with
caps
Gasket residue preventing Rinse manifold with even contact between distilled water manifold parts
Lower gasket misaligned Press gasket back into
groove
Physical damage to Replace vacuum manifold or CAP (e.g., manifold or CAP chips and cracks)
Purge step in purification If closure of wells or protocol columns (e.g., with
sealing tape) restores negative pressure, no corrective action needed
Excessive noise Purification plate or CAP Press down gently on during vacuuming not fully seated plate or CAP to seat
Gasket residue preventing Rinse manifold with even contact between distilled water manifold parts
Worn manifold gasket Replace gasket
Problem Possible Cause Possible Solution
Incorrect negative Vacuum source Connect vacuum pressure from connected to left to right side of vacuum vacuum regulator side of vacuum
Clogging of Insufficient negative See “Insufficient purification plate pressure in vacuum negative pressure or CAP columns manifold in vacuum manifold”
problem above
98
Page 8
Section 10 Ordering Information
Catalog # Description
732-6470 Aurum Vacuum Manifold
732-6820 Aurum Total RNA Mini Kit
732-6830 Aurum Total RNA Fatty and Fibrous Tissue Kit
732-6800 Aurum Total RNA 96 Kit
732-6400 Aurum Plasmid Mini Kit