Windows XP and Windows Vista are trademarks of Microsoft Corporation.
LICENSE NOTICETO PURCHASER
Purchase of this instrument conveys a limited non-transferable immunity from suit for the
purchaser’s own internal research and development and for use in applied fields under one or
more of U.S. Patents Nos. 5,656,493, 5,333,675, 5,475,610 (claims 1, 44, 158, 160-163 and
167 only), and 6,703,236 (claims 1-7 only), or corresponding claims in their non-U.S.
counterparts, owned by Applera Corporation. No right is conveyed expressly, by implication or
by estoppel under any other patent claim, such as claims to apparatus, reagents, kits, or
methods such as 5’ nuclease methods. Further information on purchasing licenses may be
obtained by contacting the Director of Licensing, Applied Biosystems, 850 Lincoln Centre
Drive, Foster City, California 94404, USA.
Bio-Rad’s thermal cyclers and real-time thermal cyclers are covered by one or more of the
following U.S. patents or their foreign counterparts owned by Eppendorf AG: U.S. Patent Nos.
6,767,512 and 7,074,367.
Hard-Shell PCR plates are covered by one or more of the following U.S. patents or their
foreign counterparts owned by Eppendorf AG: U.S. Patent Nos. 7,347,977, 6,340,589,
6,528,302.
i
ii
Bio-Rad Laboratories Resources
Table 1 lists Bio-Rad resources and how to locate what you need.
Table 1. Bio-Rad resources
ResourceHow to Contact
Local Bio-Rad Laboratories
representatives
Technical notes and literature Go to the Bio-Rad Laboratories web site (www.bio-
Find local information and contacts on the Bio-Rad
Laboratories web site by selecting your country on the
home page (www.bio-rad.com). Find the nearest
international office listed on the back of this manual
rad.com). Type a term in the Search box and select
Literature to find links to technical notes, manuals, and
other literature
We staff our Technical Support department with
experienced scientists to provide our customers with
practical and expert solutions
To find local technical support on the phone, contact your
nearest Bio-Rad Laboratories office. For technical support
in the United States and Canada, call 1-800-424-6723
(toll-free phone), and select the technical support option
S1000 Thermal Cycler Manual
Warranty
The S1000 thermal cycler and associated accessories are covered by a standard Bio-Rad
warranty. Contact your local Bio-Rad Laboratories office for the details of the warranty.
Writing Conventions Used In This Manual
This manual provides instructions on how to safely set up and operate the S1000 thermal
cycler and uses the writing conventions shown in Table 2 to quickly provide relevant
information.
Table 2. Manual conventions
ConventionMeaning
TIP: Provides helpful instructions, including information explained in
further detail elsewhere in this manual
NOTE:Provides important information, including information explained in
further detail elsewhere in this manual
WARNING! Explains crucial information about a topic that may lead to injury
to the user, instrument damage, or data loss
Screen messageIndicates the one or more words on the screen the user should
select
iii
Safety and Regulatory Compliance
Table 2. Manual conventions (continued)
ConventionMeaning
NAME of control panel
key
Select XSelect X using the arrow keys. For example, select NEW means
Select X > YFrom menu X, select Y. For example, select MAIN
Press XPress the X key on the control panel. For example, press ENTER
Indicates a key on the thermal cycler control panel. For example,
these keys have the following names:
•The ENTER key is
•The right arrow key is
use the arrow keys to select the NEW option on the screen
means select the RUN option in the MAIN menu
means press the ENTER key on the control panel
Safety and Regulatory Compliance
>
RUN
The S1000 thermal cycler heats and cools very quickly during operation. We strongly
recommend that you follow the safety specifications listed in this section and throughout this
manual.
Safety Warning Labels
Warning labels posted on the instrument and in this manual warn you about sources of injury
or harm. Refer to Table 3 to review the meaning of each safety warning label.
Table 3. Instrument safety warning labels
IconMeaning
CAUTION: Risk of danger! This symbol identifies components that pose a risk of
personal injury or damage to the instrument if improperly handled. Wherever this
symbol appears, consult the manual for further information before proceeding
CAUTION: Risk of electrical shock! This symbol identifies components that pose a
risk of electrical shock if improperly handled
CAUTION: Hot surface! This symbol identifies components that pose a risk of
personal injury due to excessive heat if improperly handled
iv
S1000 Thermal Cycler Manual
Instrument Safety Warnings
The following warning labels display on the instrument, and refer directly to the safe use of this
S1000 thermal cycler (Table 4).
Table 4. Instrument safety warning labels
IconMeaning
Warning about risk of harm to body or equipment.
Operating the S1000 thermal cycler before reading this manual can constitute a
personal injury hazard. Only qualified laboratory personnel should operate this
instrument
Warning about risk of harm to body or equipment from electrical shock.
Do not attempt to repair or remove the outer case of this thermal cycler base, power
supply, heat pump, or other accessories. If you open these instruments, you put
yourself at risk for electrical shock and void your warranty. All repairs must be done
by an authorized repair service
Never remove the outer case of a thermal cycler base. This may cause you to
receive an electrical shock.
This thermal cycler uses neutral fusing, which means that live power could still be
exposed inside the instrument even when the fuse is blown or removed
Warning about risk of burning.
A thermal cycler generates enough heat to cause serious burns. Wear safety
goggles or other eye protection at all times during operation. Always allow the
sample block to return to idle temperature before opening the lid and removing
samples. Always allow maximum clearance to avoid accidental skin burns
Warning about risk of explosion.
The sample blocks can become hot enough during the course of normal operation
to cause liquids to boil and explode
Safety and Regulatory Compliance
This instrument has been tested and found to be in compliance with all applicable
requirements of the following safety and electromagnetic standards (Table 5).
Table 5. Safe use specifications
Safe User RequirementsSpecifications
Input powerRated100–240 Vac, 50–60 Hz
Fuses250 V, 10 A
TemperatureIndoor useAmbient temperature of 15–31°C. Relative
humidity maximum of 80%
(noncondensing)
AltitudeUp to 2,000 meters above sea level
Overvoltage
Categories
Pollution degree2
II
v
Safety and Regulatory Compliance
SAFETY COMPLIANCE
This instrument has been tested and found to be in compliance with all applicable
requirements of the following safety and electromagnetic standards:
• UL Std No. 61010A-1 Electrical Equipment for Measurement, Control, and
Laboratory Use, Part 1: General Requirements
• UL Std No. 61010A-2-010 Electrical Equipment for Measurement, Control, and
Laboratory Use, Part 1: General Requirements
• CAN/CSA C22.2 No. 1010.1-92 - Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use, Part 1: General Requirements
(includes Amendment 1)
- Safety Requirements for Electrical Equipment for Measurement, Control, and
Laboratory Use, Part 1: General Requirements
• CAN/CSA C22.2 No. 1010.2.010A-97 - Safety Requirements for Electrical
Equipment for Measurement, Control, and Laboratory Use, Part 2-010: Particular
Requirements for Laboratory Equipment for the Heating of Materials, Amendment
No. 1
• IEC 61010-1 Safety Requirements for Electrical Equipment for Measurement,
Control, and Laboratory Use, Part 1: General Requirements
• IEC 61010-1 Safety Requirements for Electrical Equipment for Measurement,
Control, and Laboratory use, Part 2: Particular Requirements for Laboratory
Equipment for the Heating of Materials
ELECTROMAGNETIC COMPATIBILITY (EMC)
• FCC Title 47 Part 15B as a Class A digital device
• EN61326 Class A Electrical Equipment for measurement, control, and laboratory use
- EMC Requirements
FCC WARNINGSAND NOTES
•Warning. Changes or modifications to this unit, not expressly approved by the party
responsible for compliance, could void the user’s authority to operate the equipment
•Note. This equipment has been tested and found to comply with the limits for a Class A
digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference when the equipment is operated in a
commercial environment. This equipment generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with the instruction
manual, may cause harmful interference to radio communications. Operation of this
equipment in a residential area is likely to cause harmful interference in which case the
user will be required to correct the interference, at his own expense
• Note regarding FCC compliance. Although this design of instrument has been tested
and found to comply with Part 15, Subpart B of the FCC Rules for a Class A digital
device, please note that this compliance is voluntary, for the instrument qualifies as an
“exempted device” under 47 CFR 15.103(c), in regard to the cited FCC regulations in
effect at the time of manufacture
• Note regarding Canadian EMC compliance: Le present appareil numerique n’emet
pas de bruits radioelectrique depassant les limites applicables aux appareils numeriques
de class A prescrites dans le reglement sur le brouillage radioelectrique edicte par le
Ministere des Communications du Canada
•Cables. Shielded cables must be used with this unit to ensure compliance with the Class
A FCC limits
• Use only Bio-Rad USB cable (catalog #184-8000) when using any 1000-series cycler
Read this chapter for information on setting up the S1000 thermal cycler.
• System overview (below)
• Reaction modules (page 3)
• Setting up the S1000 thermal cycler (page 5)
• Operating the reaction module lid (page 7)
System Overview
The S1000 thermal cycler base (Figure 1) includes:
• Reaction module bay – holds the inserted reaction module
• Reaction module locking bar – locks the inserted module in place
• Control panel – provides access to all the functions needed to create and run PCR
protocols
• Air vents – allow the thermal cycler to heat and cool quickly
Module locking bar
The back panel of the S1000 thermal cycler includes data ports (Figure 2).
• USB B port — connects the S1000 thermal cycler to a C1000™ thermal cycler
• USB A ports — currently inactive
Reaction module bay
Control panel
Air vents
Figure 1. Frontal view of the S1000 thermal cycler.
1
System Overview
• Test port — for service testing only
The control panel on the S1000 thermal cycler provides access to all the functions needed to
run the thermal cycler and includes the following components:
• Liquid crystal display (LCD) — displays the main menu and other screens
• Command, numeric, and navigation keys — use these keys to enter commands,
The main screen is displayed after booting is complete. Figure 3 shows the components of the
control panel.
USB B port
Power
switch
Power
input
Fuses
Cooling vents
Figure 2. Back panel of the S1000 thermal cycler.
numbers, or letters, and navigate various screens
Numeric and
LCD
command keys
Test port
USB A ports
Main menu
options
Navigation
keys
Screen
key
Figure 3. Components of the control panel on the S1000 thermal cycler.
Function of the Control Panel Keys
The control panel of the S1000 thermal cycler contains five sets of keys with the functions
listed in Table 6.
Table 6. Function of keys on control panel
KeyFunctionAdditional Notes
Numeric and Command Keys
1 through 9Enter numbers
0, INCUBATEInserts a zero or infinity, or starts
instant incubation
2
S1000 Thermal Cycler Manual
Table 6. Function of keys on control panel (continued)
KeyFunctionAdditional Notes
CANCEL (–)Enters a minus sign or cancels a
function
PAUSE (
SCREEN key
Navigation Keys
Right arrowMoves the cursor to the right
Left arrowMoves the cursor to the left
Up arrowMoves the cursor upPress the up arrow key to scroll
Down arrowMoves the cursor downPress the down arrow key to scroll
ENTERConfirms a selection
.
)Enters a decimal point or pauses
a protocol
Toggles between screens for
alternative views
Press this key to delete an entry,
cancel a function, stop a protocol,
end an incubation, or delete text
on the screen
Press this key to view the status of
a run
from A to Z on the screen. For
example, press the up key three
times to select the letter C
from Z to A on the screen. For
example, press the down key eight
times to select the letter S
Reaction Modules
The S1000 thermal cycler is compatible with any 1000-series reaction module. The reaction
modules come in three block sizes: the 96-, dual 48-, or 384-well block. Each block in the
reaction module includes a fully adjustable heated lid that is capable of running reliably with a
broad range of reaction vessels.
Recommended Sample Volume
When using the S1000 thermal cycler, the maximum sample volume is determined by the type
of reaction module used. -well reaction module is used. Table 7 lists the recommended
sample volume, as well as the maximum sample volume to be used with different reaction
modules.
Table 7. Size and volume limit for the 1000-series reaction modules
Number of Wells
Dual 48/48210–50 µl (50 µl limit)
96110–50 µl (50 µl limit)
38413–30 µl (30 µl limit)
Number of
Blocks
Recommended Sample
Volume (Upper Limit)
3
Reaction Modules
Specifications of Reaction Modules
Specifications for each 1000-series reaction module are listed in Table 8.
Table 8. Reaction module specifications
Feature96-Well FastDual 48/48 Fast384-Well
Sample capacity96 x 0.2 ml tubes2 x 48 x 0.2 ml tubes 1 x 384-well PCR
Gradient directionBack (upper
Gradient temperature
range
Gradient temperature
differential
Gradient accuracy±0.2°C of
Gradient (end row)
uniformity
Gradient calculator
accuracy
Heated lid temperature0–110°C0–110°C0–110°C
Average ramp rate3.3°C/sec3.0°C/sec2.0°C/sec
Maximum ramp rate5.0°C4.0°C2.5°C
Temperature range0–100°C0–100°C0–100°C
Temperature accuracy±0.2°C of
Temperature uniformity±0.4°C well-to-well
microplate
Back (upper
temperature) to front
(lower temperature)
of block
30–100°C30–100°C30–100°C
1–24°C1–24°C1–24°C
programmed
temperature at end
rows
±0.4°C well-to-well
(within row) within 10
sec of arrival at
target temperature
±0.4°C of the actual
well temperature
programmed target
at 90°C
within 10 sec of
arrival at 90°C
temperature) to front
(lower temperature)
of block
±0.2°C of
programmed
temperature at end
rows
±0.4°C well-to-well
(within row) within 10
sec of arrival at
target temperature
±0.4°C of the actual
well temperature
±0.2°C of
programmed target
at 90°C
±0.4°C well-to-well
within 10 sec of
arrival at 90°C
Back (upper
temperature) to front
(lower temperature)
of block
±0.2°C of
programmed
temperature at end
rows
±0.4°C well-to-well
(within row) within 10
sec of arrival at
target temperature
±0.4°C of the actual
well temperature
±0.2°C of
programmed target
at 90°C
±0.4°C well-to-well
within 10 sec of
arrival at 90°C
Each reaction module contains cooling fins for fast heating and cooling and a fully adjustable,
heated lid. Figure 4 shows the lid and cooling fins for the 96-well reaction module.
• Heated inner lid — adjusts the lid temperature to prevent condensation and
evaporation
• Sample/reaction block — holds reaction vessels, including tubes and microplates
4
S1000 Thermal Cycler Manual
Lid
Cooling fins
Figure 4. The lid and cooling fins of a 96-well reaction module.
The top of a reaction module lid includes a lid lever, lid force knob, and status LED (Figure 5).
• Lid lever — opens and closes the lid
• Lid force knob — sets lid force and seals the reaction
• Status LED — turns on to indicate that the block is selected or running
Lid force knob
Lid lever
Figure 5. A top view of a reaction module.
Setting Up the S1000 Thermal Cycler
The S1000 thermal cycler package includes:
• S1000 thermal cycler base
• Power cord
• Consumables selection guide
• Instruction manual
• Quick guide for system installation
Reaction modules for use with the S1000 thermal cycler are shipped in separate packaging.
Remove all packaging materials and store them for future use. If any item is missing or
damaged, contact your local Bio-Rad office.
Place the S1000 thermal cycler base on a flat, dry surface with sufficient cool airflow to run
properly. The instrument can run in two modes: stand-alone or software-controlled. When
running the system under software-controlled mode, make sure there is sufficient space for a
computer during setup.
Status
LED
5
Setting Up the S1000 Thermal Cycler
To insert either a 96-, dual 48-, or 384-well reaction module into the reaction module bay
of the thermal cycler base, follow these instructions:
1. With the locking bar in the down position and the lid lever of the reaction module
pointing to the front, lift the reaction module into the reaction module bay (Figure 6).
Leave about 1–2 cm of space in front of the module.
Lid lever (front)
Leave space here
Figure 6. Inserting the reaction module into the bay.
2. Pull the locking bar up to lock the reaction module in place (Figure 7). There is no space
at the front of the module when it is locked into the S1000 thermal cycler base.
TIP: Store the reaction module in the base when it is not in use.
Locking bar
(down position
when unlocked)
No space here
(in the locked position)
Figure 7. Locking the reaction module in place.
3. Plug the supplied power cord into the appropriate electrical outlet.
4. Turn on the thermal cycler using the power switch on the back panel of the thermal
cycler base.
NOTE: Before operating the thermal cycler, be sure to read the safety
specifications (“Safety and Regulatory Compliance” on page iv) and operating
requirements.
5. When the S1000 thermal cycler starts up, it goes through two screens: the black booting
and the self-test screens. Once the self-test is run to verify proper functions, the main
menu is displayed. Use the main menu to begin operating the thermal cycler.
To remove the reaction module from the thermal cycler base, follow these instructions:
1. Turn off the thermal cycler.
2. Unlock and release the reaction module by pushing the locking bar down.
Locking bar
(up position
when locked)
3. Carefully lift the reaction module out of the bay (Figure 8).
WARNING! Cooling fins may be hot immediately after running a protocol or
incubation. Before lifting the reaction module, make sure that the cooling fins are
not hot.
6
4. After removing the reaction module from the S1000 thermal cycler, store it on a clean,
flat surface where it cannot get bumped, scraped, or dropped.
Scraping the cooling fins of the reaction module or dropping the module on the fins
could compromise the ability of the module to heat and cool correctly.
Figure 8. Lifting the reaction module out of the bay.
Operating the Reaction Module Lid
S1000 Thermal Cycler Manual
Cooling fins
The inner lid of the reaction module applies heat and force to the reaction vessel lids (caps or
tape) to produce consistent and successful reactions. Heating the inner lid prevents
condensation, while applying force seals the reaction to prevent evaporation.
WARNING! After a run, the heated inner lid can remain hot. Use caution when
opening and closing the lid.
To open the lid, use the following steps:
1. Turn the lid force knob counterclockwise to release the inner lid (Figure 9).
Lid force knob
(Turn counterclockwise
to release the lid)
Figure 9. Turn the lid force knob counterclockwise to release the inner lid.
7
Operating the Reaction Module Lid
2. To open the lid, push the lid lever back and then lift it up (Figure 10).
3. Lift the lid lever completely until the reaction module stays open without assistance.
To close the lid, use the following steps:
1. Push the lid lever down (Figure 11), making sure that the front of the lid is secured
beneath the housing, and then lock it in place.
Lid lever
Figure 10. Lift the lid lever up to open the lid.
Lid lever
Figure 11. Push the lid lever down.
2. Adjust the lid force by turning the lid force knob (Figure 12).
• Turn the knob 1/4 clockwise (to the right) to increase the lid force
• Turn the knob 1/4 counterclockwise (to the left) to decrease the lid force
Adjust the lid force to a similar setting each time by turning it to the same position.
NOTE: The position marks on the lid indicate 1/4 turns.
Lid force knob
(Turn clockwise to
secure the lid)
Figure 12. Adjust the lid force by turning the lid lever.
8
S1000 Thermal Cycler Manual
Loading Sample Vessels into the Reaction Block
To ensure uniform heating and cooling of samples, vessels must be in complete contact with
the reaction block. Adequate contact is achieved by:
• Confirming that the block is clean before loading samples
• Firmly pressing the individual tubes or the microplate into the block wells
TIP: When using one or a few tubes, make sure to place them in the center of the
block to ensure uniform thermal cycling of all samples. Use the tube frame (catalog
#184-9000), or load at least one empty tube in each corner of the block to ensure
that the lid exerts even pressure on individual tubes.
Main Menu
The main menu (Figure 13) provides access to all thermal cycler operations and displays the
status of the reaction module and the name of the thermal cycler.
Name of
thermal cycler
Options
Status message
Figure 13. The main window of the S1000 thermal cycler.
Select the options in the main menu to start these instrument functions:
• RUN
• NEW
• EDIT — to modify stored protocol files
• FILES
• VIEW
• TOOLS
— to run an existing protocol file
— to create a new protocol file
— to copy, move, rename, delete, or secure protocol files and/or folders
— to review an existing protocol file
— to change thermal cycler settings or to view the last protocol that was run
The File Tree
All protocol files are stored in the file tree. The file tree displays when you need to select a
protocol to run, edit, or view.
The file tree (Figure 14) includes these folders:
• MAIN folder — stores the preinstalled protocols and cannot be deleted or renamed.
The preinstalled files can be run or copied by any user. Do not store user-created
protocols in the MAIN folder
9
Operating the Reaction Module Lid
• User folders — contain user-created protocol files. User folders and associated files
can be secured with a password. The files cannot be edited or deleted without using
the password
Selected folder
User folders
Selected file
Figure 14. The file tree of the S1000 thermal cycler.
NOTE: If a folder contains more than six protocols, use the arrow keys to scroll
down to see all the protocols. All folder names are displayed with angle brackets (<
and >) surrounding the name.
User-created
protocol file
Preinstalled
protocols
10
S1000 Thermal Cycler Manual
2Creating and Editing Protocols
Read this chapter for information on creating and editing protocols.
• Protocol steps (below)
• Creating a new protocol (page 12)
• Parameters for temperature or gradient steps (page 19)
• Editing an existing protocol (page 22)
• Sample volume and lid temperature (page 27)
Protocol Steps
Table 9 includes a list of steps in a protocol. The table also includes the limits and range of the
parameters.
Table 9. Protocol steps and parameters of the S1000 thermal cycler
Step NameParameters and RangesDescription
TEMP
(Temperature)
Temperature in °C: The target
temperature between 0.0 and
100.0°C in tenths of a degree
Hold time: The hold time
between 1 sec and 18 hr in the
format of hr:min:sec. To enter an
infinite hold, press the ∝ (infinite,
0) key
Instructs the thermal cycler to ramp to
the target temperature, and hold that
temperature for the specified amount
of time
11
Creating a New Protocol
Table 9. Protocol steps and parameters of the S1000 thermal cycler (continued)
Step NameParameters and RangesDescription
GRAD
(Gradient
range)
GOTOGOTO step: The step number of
END(No parameters)A protocol step that instructs the
Lower: The lower temperature
in the gradient. Enter a number
between 30.0 and 99.0°C in
tenths of a degree
Upper: The upper temperature
in the gradient. The maximum
temperature is 100°C. Enter a
temperature within 24.0°C of the
lower temperature
Time:The hold time between 1
sec and 18 hr in the format of
hr:min:sec. To enter an infinite
hold, press the ∝ key (infinite, 0)
key
the first step in the repeat.
ADDTNL REPEATS: The
number of additional times that
the steps repeat.
Instructs the thermal cycler to ramp to
the target temperature gradient across
the block, and hold that temperature
gradient for the specified amount of
time
A protocol step that instructs the
thermal cycler to repeat a set of steps
for the specified number of times.
NOTE: The total number of cycles in
the protocol is the number of GOTO
repeats, plus the first cycle.
thermal cycler to finish the protocol
Creating a New Protocol
NOTE: The internal memory of the S1000 thermal cycler can hold up to 400, 2-step
protocols.
To create a protocol:
1. Select NEW from the main menu (Figure 15). Press ENTER to confirm the selection.
Figure 15. Select NEW from the main menu.
2. Use the numeric keys to enter the name of the new protocol file. Enter a letter by
pressing the up or down arrow key and a number by pressing the numbered key. For
example, to select the letter C, press the up key 3 times. To select the letter S, press the
12
S1000 Thermal Cycler Manual
down key eight times. Press ENTER to continue to the next space. Press ENTER to
continue to the next screen.
NOTE: A protocol file name can contain 1–8 characters and must be unique to the
folder. To delete or change a letter, press CANCEL and select a new letter. To
delete the entire name, press CANCEL multiple times.
In Figure 16, the characters STD3 are entered, and the cursor is highlighting the next
space.
Figure 16. STD3 is entered as the protocol name.
3. (Optional) Enter a new lid temperature, and press ENTER to continue to the next screen.
NOTE: The lid temperature can range from 0 to 110°C. When the block is running
an infinite hold at a temperature below the Turn off below parameter, the lid heater
maintains 31.0°C. To change the default Turn off below parameter, select TOOLS
> DEFAULTS.
4. (Optional) Enter the sample volume in microliters (µl), and press ENTER to continue to
the next screen.
NOTE: Entering a sample volume between 1 and 50 selects Calculated Temperature control mode, which is the standard mode. Entering zero (0) in the
volume field selects Block mode. Calculated mode is the recommended mode
because it most accurately represents the actual sample temperature. For more
information about Temperature control modes, see page 27.
5. Using the arrow keys, select TEMP to enter a temperature step or GRAD to enter a
gradient temperature step in the protocol file. Press ENTER to continue to the next
screen.
In Figure 17, TEMP is selected as the temperature step.
Figure 17. TEMP is selected as the temperature step in this protocol file.
NOTE: The first step in a protocol must be either a TEMP or GRAD step.
6. Enter the target temperature between 0 (zero) and 100.0°C for the temperature step.
Press ENTER to continue to enter the next item in the protocol.
13
Creating a New Protocol
In Figure 18, the target temperature is 95°C.
7. Enter the hold time (TIME) in minutes and seconds using the numeric keys. The hold
time (TIME) ranges between 0:01 (one second) and 18:00:00 (18 hours). Entering 0 (zero)
adds an infinite hold and holds this step FOREVER. Press ENTER to continue to the
next field.
For example, to enter 4 minutes (4:00), type 400. To enter 30 seconds, type 30. In
Figure 19, the hold time is four minutes.
Figure 18. Target temperature of 95°C is used in this protocol.
Figure 19. The hold time is 4 minutes in this protocol.
8. Select YES, No, or Option by pressing the right and left arrow keys. Press ENTER to
continue to the next screen.
• YES — to confirm the current parameters for this protocol step
• No — to change a parameter in this protocol step
• Option — to add more parameters to this protocol step. For more information
about entering options, see “Adding an Increment to a Temperature Step”
(page 19)
9. (Optional) Enter a gradient temperature step by pressing the right arrow key to select
GRAD (Figure 20). Press ENTER to continue to the next screen.
NOTE: A temperature gradient is limited to a 24°C spread. The lowest possible
“lower” temperature in the gradient is 30°C and the highest “upper” temperature is
14
S1000 Thermal Cycler Manual
100°C. Therefore the lowest gradient is 30–54°C, and the highest possible gradient
is 76–100°C.
Figure 20. GRAD is selected in the gradient temperature step.
TIP: Check the temperature in each row of the block in a gradient by selecting the
gradient calculator tool (TOOLS > GRADCALC on page 58).
10.Enter the lower temperature in the gradient. The lower temperature is at the front (row H)
of the block.
In Figure 21, the lower temperature is 50°C.
Lower
temperature
(back row)
Figure 21. The lower temperature is 50°C in this protocol.
11.Enter the upper temperature in the gradient. The upper temperature is at the back (row
A) of the block.
NOTE: The range of temperature is limited by the widest available range for
gradient, which is 24°C. The highest value that can be entered for the upper
temperature is 100°C.
12.Enter a hold time between 0:01 (one second) and 18:00:00 (18 hours).
13.Select YES, No, or Option by pressing the right and left arrow keys, then press ENTER
to continue to the next screen:
• YES — to confirm the current parameters for this protocol step
• No — to change a parameter in this protocol step
• Option — to preview the temperature gradient. If Option is selected, select VIEW
on the next screen to view the gradient or EXT to add a hold time extension. Press
ENTER again to return to the previous screen
Range of
temperatures to
be chosen
15
Creating a New Protocol
In Figure 22, the gradient is formed on a 96-well block with a range from 55–75°C. This
screen displays the approximate temperature of each row of the block, and labels the
front and back rows.
14.Repeat the instructions in steps 6–9 to continue entering additional temperature steps. In
Figure 23, four steps are entered.
Entered steps
Figure 22. The approximate temperature of each row of the block.
Choose new
step type
Figure 23. Enter all the temperature steps.
NOTE: A protocol can contain up to 99 protocol steps. The first step must be a
temperature (TEMP) step, while the last step must be an END step.
15.(Optional) To enter a GOTO step immediately after the set of steps to be repeated in a
cycle, use the arrow keys and select GOTO. Press ENTER to continue to the next
screen.
For more information about how the GOTO step creates a cycle, see “Protocol Steps”
(page 11).
In Figure 24, step 5 is a GOTO step.
Figure 24. A GOTO step is selected.
NOTE: The GOTO step cannot be the first or the last step in the protocol.
16
16.Enter the step number for the first step in the GOTO repeats using the numeric keys.
Press ENTER to continue to the next screen.
S1000 Thermal Cycler Manual
In Figure 25, the first step is 2. The GOTO step instructs the thermal cycler to return to
step 2 and repeat all the steps between steps 2 and 5.
Steps that
repeat during
GOTO step
First step in
GOTO repeats
Figure 25. Enter the first step in the GOTO repeats.
17.Enter the number between 1 and 9999 for the additional repeats (ADDTNL REPEATS) in
the GOTO step. Then press ENTER to continue to the next screen.
NOTE: The GOTO step adds additional cycles to the PCR protocol. The first cycle
is not included in the GOTO step. For example, to run a PCR protocol with 31
cycles, enter 30 repeats in the GOTO step.
In Figure 26, the number of repeats is 30, and the total number of cycles is 31.
Steps that
repeat in the
GOTO step
Figure 26. Enter the number of repeats in a GOTO step.
18.Select YES to accept the GOTO step parameters (Figure 27), or select No to return to
the beginning of this step and change the GOTO step parameters. Then press ENTER to
continue to the next screen.
Figure 27. Confirm a GOTO step by selecting YES.
19.Enter the remaining steps by choosing the step type and adding parameters. Then press
ENTER to continue to the next screen.
TIP: To instruct the thermal cycler to emit a sound at the end of the protocol,
include a BEEP option in the final temperature step (page 21).
Number of
additional repeats
in a GOTO step
17
Creating a New Protocol
20.Select END using the arrow keys to instruct the thermal cycler to finish the protocol file.
Press ENTER to continue to the next screen. In Figure 28, the END step is selected.
NOTE: The END step must be the last step of a protocol; a protocol can only
contain one END step.
21.Select YES (Figure 29) to accept the protocol step parameters or No to return to the
beginning and select a different protocol step.
Figure 28. The END step in a protocol.
Figure 29. Select YES to accept the protocol parameters.
22.Use the arrow keys to select the folder where you want the new protocol file to be saved,
and press ENTER to save the protocol file.
NOTE: The file tree folder screen does not appear if there are no user-created
folders (i.e. the MAIN folder is the only folder on the list). Saving the protocol in the
MAIN folder is not recommended. If the protocol is saved in the MAIN folder, we
recommend moving it to a user-created folder. (See “Moving a Protocol File” on
page 41 for more information.)
In Figure 30, the STD3GRAD file is saved in the GRANT
Figure 30. The STD3GRAD file is to be saved in the GRANT folder.
To run a protocol, follow the instructions in “Preparing to Run a Protocol” on page 29.
folder.
18
S1000 Thermal Cycler Manual
Parameters for Temperature or Gradient Steps
Table 10 includes a list of options for temperature and gradient steps for the S1000 thermal
cycler. The table also includes the limits and range of the parameters.
Table 10. Options and parameters for protocols on the S1000 thermal cycler
Step NameParameters and RangesDescription
INC
(Increment)
EXT
(Extend)
RATE
(Ramp rate)
Beep(No parameters)Applies only to a temperature step (see
A temperature from –10.0 to
10.0°C per cycle in tenths of a
degree
A time from –60 to 60 sec per
cycle
A number from 0.1 to 5°C per
sec
Applies only to a temperature step (see
“Adding an Increment to a
Temperature Step” on page 19).
Instructs the thermal cycler to
increment (change) the target
temperature of a step with each cycle,
where a positive number increases the
temperature and a negative number
decreases the temperature
Applies to both temperature and
gradient steps (see “Extending the
Hold Time in a Temperature Step” on
page 20). Instructs the thermal cycler
to extend the hold time with each
cycle. A positive number increases the
hold time and a negative number
decreases the hold time
Applies only to a temperature step (see
“Changing the Ramp Rate in a
Temperature Step” on page 21).
Instructs the thermal cycler to ramp to
the target temperature at the specified
ramp rate in that step
“Adding a Beep to a Temperature
Step” on page 21). Instructs the
thermal cycler to beep to signal that
the thermal cycler has reached the
target temperature for that step
Adding an Increment to a Temperature Step
The increment (INC) parameter changes the target temperature of a protocol step. The
increment can increase or decrease the target temperature with each cycle in the protocol.
To add an increment:
1. Select OPTION using the arrow keys. Press ENTER to continue to the next screen.
19
Parameters for Temperature or Gradient Steps
In Figure 31, OPTION is selected to add an increment to a temperature step.
Figure 31. Select OPTION to add an increment to a temperature step.
NOTE: The INC parameter must be added to a step within the GOTO repeats in
order to increment with each cycle of the reaction.
2. Select INC using the arrow keys (Figure 32) to add an increment to the protocol step in
each cycle. Press ENTER to continue to the next screen.
Figure 32. Select INC to add an increment.
3. Enter the increment temperature using the numeric keys. To decrease the temperature
each cycle, enter a negative number by pressing the CANCEL (–) key. Press the PAUSE
(
.
) key to enter a decimal point. Then press ENTER to continue to the next screen.
NOTE: Enter an increment from –10.0 to 10.0°C per cycle in tenths of a degree,
and within the limits of a temperature step (0–100°C).
In Figure 33, the increment is +0.5°C. The target temperature for step 3 will increase by
0.5°C each cycle.
Figure 33. Enter the increment temperature.
4. To confirm the parameters of the protocol step, select YES and then press ENTER. To
change the parameters, select No and then press ENTER.
Extending the Hold Time in a Temperature Step
20
The EXT parameter changes the hold time for a temperature or gradient temperature step. The
extension increases or decreases the hold time with every cycle.
Loading...
+ 67 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.