The Blood Glucose Monitoring System GM232C is a personal blood glucose monitoring system to be used
Rightest
for self-testing only. The glucose meter and lancing device are for single patient use. Do not use on multiple patients.
Do not share meter or lancing device with anyone including other family members.
A healthcare professional should be contacted when Customer Service is not available.
Please forward your warranty card to Bionime customer service.
Bionime customer service
1450 E. Spruce Street, Bldg. #B,
Ontario, CA 91761, USA
2
Caution
- Before using the Rightest Blood Glucose Monitoring System GM232C to test your blood glucose, please read all of
the instructions and conduct all of the tests including the quality control test (Refer to page 38).
- Please perform the quality control test regularly to make sure the test results are accurate.
- The Rightest Blood Glucose Meter can only be used with Rightest Blood Glucose Test Strips GM232C. Other brands’
test strips should not be used under any circumstances. The use of other brands’ strips may give inaccurate
results.
- The Rightest Blood Glucose Monitoring System GM232C is intended for in vitro diagnostic use only. The Rightest
Blood Glucose test results using fresh capillary whole blood samples from the fingertip, palm and forearm are
calibrated to be the equivalent to that of plasma testing.
- The Rightest Blood Glucose Monitoring System GM232C is intended for self-testing. It should not be used to screen
for or diagnose diabetes mellitus.
- If the Rightest Blood Glucose Meter GM232C and Test Strips are exposed to a substantial change in temperature,
please wait 45 minutes before measurement.
- Follow all environmental protection regulations when disposing of batteries.
- The Rightest Blood Glucose Meter GM232C will be set as mg/dL by default when sold in United States. If meter
display shows “mmol/L” during setting or test, please contact GE Diabetes Customer Support Center. Use of the
wrong unit of measurement may cause incorrect treatment.
3
Caution
Caution
- The minimum blood sample size to test using theBlood Glucose Monitoring System GM232C is 0.75µL: ( ) Rightest
Sample Size Example
Blood sample size above 3.0µL might contaminate the test strip port and the meter.
Blood sample size below 0.75µL may cause an inaccurate result or may prevent a meter reading. An Er4 reading will
be displayed if the sample size is too small. In this case, repeat the test with a new test strip.
4
0.75µL 1.0 1.5µL 2.0µLµL 3.0µL
Important Safety Notes:
- TheBlood Glucose Meter GM232C and Lancing Device are for single patient use. Do not share
Rightest
them with anyone including other family members.
- All parts of the kit are considered biohazardous and can potentially transmit infectious diseases, even
after following the cleaning and disinfecting procedures.
Disinfecting Procedures" on page 48.
- Users should wash their hands thoroughly with soap and water before and after handling the meter,
lancing device, or test strips.
- User should turn off meter or meter’s Bluetooth connection to keep it in flight mode “ ” while traveling
in flight. User can still perform blood glucose testing while meter is in flight mode “ ” .
- Action upon delivery:
a) Unpacking; b) checking delivery for completeness; c) checking for damage during transport.
Please refer to the section "Cleaning and
5
Table of Contents
About theBlood Glucose Monitoring System GM232C
The Blood Glucose Monitoring System GM232C...................................................
The Blood Glucose Meter GM232C...........................................................................................................
The Blood Glucose Test Strip GM232C...................................................................................................
Before Testing
Meter Activation and Battery Change.....................................................................................................................................
Setting Up Your Meter - Setting the Date, Time and Volume........................................................................................
Turning On/Off the Meter...............................................................................................................................................................
Turning On/Off the Bluetooth......................................................................................................................................................
Handling the GE006 Blood Glucose Test Strip........................................................................................................
Testing Procedure
Getting Ready for Testing..............................................................................................................................................................
Performing a Blood Glucose Test...........................................................................................................................................
Setting Markers...............
Alternative Site Testing...................................................................................................................................................................
Understanding Test Results and Messages..........................................................................................................................
Quality Control
About Quality Control Testing......................................................................................................................................................................................................
Performing a Quality Control Test........................................................................................................................................
Understanding Control Test Results.....................................................................................................................................
Test Memory
Recalling Test Results.................................................................................................................................................................
Recalling Average Test Results...............................................................................................................................................
Caring for Your Meter.................................................................................................................................................................
Cleaning and Disinfecting Procedures...............................................................................................................................
Error Messages and Troubleshooting................................................................................................................................
Real Time Synchronize....................................................................................................................................................................................................................
FEDERAL COMMUNICATIONS COMMISION (FCC) STATEMENT & NCC警語.............................................................................................................