biochrom NanoVue Plus User Manual

NanoVue Plus
Product User Manual
Biochrom US Telephone: 1-508-893-8999 84 October Hill Rd Toll Free: 1-800-272-2775 Holliston, MA Fax: 1-508-429-5732 01746-1388 support@hbiosci.com USA www.biochromspectros.com
2
5061-049 REV 1.4
1. INSTALLATION 3
1.1. Unpacking and positioning 3
1.2. Safety 3 4
2. INTRODUCTION 5
2.1. Your NanoVue Plus 5
2.2. File system 6
2.3. Data export 6
2.4. Sample treatment 7
2.5. Pathlength and Absorbance nomalization 7
2.6. Auto-Read 8
2.7. Quality Assurance 8
2.8. Keypad and display 9
2.9. Software style 10
3. OPERATION AND MAINTENANCE 11
3.1. Sample application guide 11
3.2. Sample Plate replacement 12
3.3. Cleaning and general maintenance 13
3.4. Return for repair 13
3.5. Lamp replacement 13
3.6. Replacing the printer paper 13
4. LIFE SCIENCE 14
4.1. File system 14
4.2. Nucleic acids 14
4.2.1. Theory 14
4.2.2. DNA measurement 16
4.2.3. RNA measurement 17
4.2.4. Oligonucleotide measurement 19
4.2.5. T
m
Calculation 20
4.2.6. CyDye™ measurement 23
4.3. Proteins 26
4.3.1. Theory 26
4.3.2. Protein UV 28
4.3.3. Protein A280 30
4.3.4. BCA 31
4.3.5. Bradford 34
4.3.6. Lowry 36
4.3.7. Biuret 39
5. APPLICATIONS 42
5.1. File system 42
5.2. Single Wavelength 42
5.3. Concentration 44
5.4. Wavescan 46
5.5. Kinetics 48
5.6. Standard Curve 50
5.7. Multiple Wavelength 53
5.8. Absorbance Ratio 54
6. FAVOURITES AND METHODS 56
7. UTILITIES 57
7.1. Date and time 57
7.2. Regional 57
7.3. Printer 58
7.4. Preferences 58
7.5. Contrast 58
7.6. About 58
7.7. Pathlength Check & Calibration 59
7.8. Games 62
8. ACCESSORIES 63
8.1. Printer installation 63
8.2. Bluetooth accessory installation 65
8.3. Fitting SD memory card accessory 67
8.4. After sales support 70
9. TROUBLESHOOTING 71
9.1. Pathlength calibration over-range 71
9.2. Error messages 72
9.3. Fault analysis 74
9.4. Frequently asked questions 74
10. SPECIFICATION AND WARRANTY 77
11. LEGAL
Page finder
3
5061-049 REV 1.4
1. INSTALLATION
1.1. Unpacking and positioning
• Removetheinstrumentfromitspackagingandinspectitforsignsofdamage.If
any are discovered, inform your supplier immediately.
• Theinstrumentmustbeplacedonastable,levelsurfacethatcantakeitsweight (~ 4.5 kg) and positioned such that air can circulate freely around the casing.
• Ensureyourproposedinstallationsiteconformstotheenvironmentalconditions
for safe operation.
• Theinstrumentisdesignedforindooruseonly,temperaturerange5°Cto35°C
and should be kept away from strong draughts.
• Ifyouusetheinstrumentinaroomsubjectedtoextremesoftemperature
change during the day, it may be necessary to recalibrate (by switching off and then on again) once thermal equilibrium has been established (2–3 hours).
• Atemperaturechangeofnomorethan4°C/hourandamaximumrelative humidityof80%at31°C,decreasinglinearlyto50%at40°Carerequired.
• Iftheinstrumenthasjustbeenunpackedorhasbeenstoredinacold
environment, it should be allowed to come to thermal equilibrium for 2–3 hours in the laboratory before switching on. This will prevent calibration failure as a result of internal condensation.
• Theinstrumentmustbeconnectedtothepowersupplywiththepoweradaptor
supplied. The adaptor can be used on 90–240V, 50–60Hz supplies. It will become warm once plugged into the power supply and should not be covered up.
• Switchontheinstrumentviathekeypad(
) after it has been plugged in. The
instrument will perform a series of self-diagnostic checks.
• Itisrecommendedthatusersreadthroughthismanualpriortouse.
• Contactyoursupplierifyouexperienceanydifficultieswiththisinstrument
• Werecommendthatyoucarryoutaquickcheckofthepathlengthbeforeusing
1.2. Safety
Spectrophotometer Health & Safety Document including General Operating Instructions are available as a booklet provided with each instrument. The booklet, translated into the European Union languages, is available on the delivered CD. The instructions provide the user with basic use, troubleshooting and how to use the instrument in a safe manner.
CAUTION
This instrument contains a UV source which generates a light beam that
is present at the sample platform. Do not attempt to divert the beam as prolonged exposure to the beam may cause permanent eye damage.
Note that the instrument prevents operation if the head is raised or the Xenon flash does no reach the detection system.
the NanoVue, please see instructions on how to do this in section 7.7.
1. INSTALLATION
1.1. Unpacking and positioning
• Removetheinstrumentfromitspackagingandinspectitforsignsofdamage.If
any are discovered, inform your supplier immediately.
• Theinstrumentmustbeplacedonastable,levelsurfacethatcantakeitsweight (~ 4.5 kg) and positioned such that air can circulate freely around the casing.
• Ensureyourproposedinstallationsiteconformstotheenvironmentalconditions
for safe operation.
• Theinstrumentisdesignedforindooruseonly,temperaturerange5°Cto35°C
and should be kept away from strong draughts.
• Ifyouusetheinstrumentinaroomsubjectedtoextremesoftemperature
change during the day, it may be necessary to recalibrate (by switching off and then on again) once thermal equilibrium has been established (2–3 hours).
• Atemperaturechangeofnomorethan4°C/hourandamaximumrelative humidityof80%at31°C,decreasinglinearlyto50%at40°Carerequired.
• Iftheinstrumenthasjustbeenunpackedorhasbeenstoredinacold
environment, it should be allowed to come to thermal equilibrium for 2–3 hours in the laboratory before switching on. This will prevent calibration failure as a result of internal condensation.
• Theinstrumentmustbeconnectedtothepowersupplywiththepoweradaptor
supplied. The adaptor can be used on 90–240V, 50–60Hz supplies. It will become warm once plugged into the power supply and should not be covered up.
• Switchontheinstrumentviathekeypad(
) after it has been plugged in. The
instrument will perform a series of self-diagnostic checks.
• Itisrecommendedthatusersreadthroughthismanualpriortouse.
• Contactyoursupplierifyouexperienceanydifficultieswiththisinstrument
• Werecommendthatyoucarryoutaquickcheckofthepathlengthbeforeusing
WARNING
High voltages exist inside the NanoVue Plus instruments. Repair and maintenance should only be carried out by individuals trained specifically to work on these instruments.
4
5061-049 REV 1.4
If the instrument is used in a manner not specified or in environmental conditions not appropriate for safe operation, the protection provided may be impaired and instrument warranty withdrawn.
There are no user-serviceable parts inside this instrument.
5
5061-049 REV 1.4
2. INTRODUCTION
2.1. Your NanoVue Plus
NanoVue Plus is a simple-to-use UV/Visible instrument with twin CCD array detectors (1024 pixels) and few moving parts, which contributes to its inherent reliability.
General Principles
In conventional UV/Visible spectrophotometers, the sample is usually contained within a glass or silica cuvette which is placed in the sample beam. The Absorbance of this is measured and then compared to that of a standard. From this comparison, the concentration is calculated.
However, when the quantity of sample is limited or highly concentrated, dilution ­or the use of ultra low volume cuvettes is required, which is time consuming, can introduce errors and presents cleaning difficulties.
NanoVue Plus was developed to overcome these problems. Using NanoVue Plus, typical sample volumes of 2 µl are pipetted onto a hydrophobic surface and then a very short pathlength of either 0.2 mm or 0.5 mm is created by lowering the sampling head onto the top of the sample.
With the sample so constrained, all of the general software features of the instrument, including wavelength scanning, single or multi-wavelength Absorbance and concentration measurement, kinetics, standard curves and Absorbance ratio, can be utilized. In all of these modes, the instrument utilizes either the 0.2 mm or 0.5 mm pathlengths. The benefit of these small pathlengths is that the instrument can measure smaller volumes of very concentrated or highly absorbing samples.
NanoVue Plus also has a range of more specific life science applications, including DNA and RNA concentration and purity, Oligonucleotide concentration, Tm Calculation, Cy Dye concentration and a variety of protein measurements. In all of these applications, instead of the actual Absorbance measurement being used, the value is normalized to reflect a standard pathlength of 10 mm, so that generally accepted factors (50 for DNA, 40 for RNA and 33 for oligos, for example) can be used to calculate concentration. As an example, if the absolute Absorbance of the sample is 0.025 A using a pathlength of 0.5 mm, the normalized Absorbance shown will be 0.500 A.
NanoVue Plus is therefore ideally suited to the life scientist where sample is limited and speed and convenience of analysis is key.
Figure 1: NanoVue Plus Spectrophotometer
UV-Visible
6
5061-049 REV 1.4
2.2. File system
After switch on and automatic checking the instrument defaults to a home page entitled ‘NanoVue Plus‘. This page displays five folders which form the topmost layer of a simple file tree which is the basis of the user-interface. The folder screens are reached by pressing the appropriate number on the keypad, with return to the top level by means of the ‘Esc’ key. The folders group various facilities together as follows:-
1. Life Science Standard life science applications such as nucleic acid and protein assays.
2. Applications General spectroscopic applications.
3. A folder to store your more frequently used methods.
(Inactive when empty)
4. Methods Contains 9 folders that can store less frequently used methods. Up to 9 methods per folder are allowed, totalling 81 methods.
5. Utilities Instrument set up options and games.
Figure 2: NanoVue Plus with Printer
Figure 2: NanoVue Plus with Printer
Favorites
Data can be transferred to a PC, either via a Bluetooth accessory (supplied pre­installed or as an optional accessory) or via a USB cable. The software to perform this task “PVC” (Print Via Computer) is a small application running under Windows 2000™, Windows XP™ or Windows 7, 8, 8.1. PVC can operate via USB and Bluetooth simultaneously.
PVC can store data either in a common directory or can be configured to save to independent directories by both file format and connection. The data may be stored as an Excel spreadsheet, an EMF graphics file, a comma delimited (csv) data file, a tab delimited (txt) data file, rich text format (RTF) which is compatible with Word or in native PVC format.
Some users may find it convenient to “print through” the PC directly to a printer already attached to it. The software to enable this option will be found on the CD provided.
Datrys Computer control software is also available for NanoVue Plus as an optional extra.
2.3. Data export
An integral printer is available for the instrument; this may be either supplied pre-installed or as an optional accessory. The installation procedure is described in section 8.1.
(Games only available if selected via the Preferences Screen)
7
5061-049 REV 1.4
2.4. Sample treatment
NanoVue Plus uses a unique sampling head which enables users to accurately measure the optical characteristics of samples volumes as low as 0.5 µl. The sample is applied to a horizontal plate having a hydrophobic surface, the sampling head is lowered into position and the reading taken. The pathlength is
automaticallyadjustedunlesstheuserselectsmanualadjustment.Thiscanbe
either 0.2 mm or 0.5 mm.
Application of the sample is most conveniently carried out using any commercially available pipetting tool and standard tips, a 10 µl pipette is recommended.
2.5. Pathlength and Absorbance normalization
For methods in the Application folder, the Absorbance values have not been normalized and the pathlength needs to be taken into account when the values are used for concentration calculations. NanoVue Plus has the ability to measure samples using either a 0.2 mm or a 0.5 mm pathlength. The 0.5 mm pathlength should be used for samples of low concentration and the 0.2 mm pathlength for more concentrated samples. It is best to use the 0.5 mm pathlength whenever possible. In the Life Science applications, the Absorbance readings are presented as their normalized 10 mm pathlength values, to allow the use of literature based factors for concentration measurements.
Within the DNA, RNA, Oligonucleotide, Tm Calculation, Cy Dye, Protein UV and Protein A280 methods the system defaults to having the pathlength chosen automatically. When this is active, the instrument will measure at the 0.5 mm pathlength first and if the Absorbance is high (> 1.7 A) then the measurement will be carried out at the 0.2 mm pathlength. If the concentration range of the samples is known beforehand, the relevant pathlength can be pre-selected; if it is unknown then the automatic option can be used. Within the other applications, the actual measured Absorbance value is reported, so if the Absorbance is high at the
0.5 mm pathlength (> 1.7 A), the sample can be re-measured at the 0.2 mm pathlength. In the automatic mode, the reference scan is always carried out using both pathlengths, so that if a sample is measured at one pathlength and subsequently re-measured at the other pathlength, a new reference scan is not required.
The pathlength that is used to carry out the measurement is printed/exported with each result and is also displayed in the top left hand corner of the header line on the instrument’s display. Care must be taken that the correct pathlength is being used when comparing sample results or using a stored calibration curve.
Figure 3a: Sample head
Figure 3b: Sample application
3a
3b
8
5061-049 REV 1.4
2.6. Auto-Read
NanoVue Plus senses when the sampling head has been lowered, so Auto-Read can be used to automatically make a measurement when this happens, obviating the need to press any keys. The instrument assumes that the first reading after starting any application will be a reference scan and that subsequent readings will be sample scans.
Switching Auto-Read on or off:
Go to Utilities (option 5 from the NanoVue Plus menu). Select Preferences (option 4). Use the arrow keys to go to the Lid Switch option and select On to turn Auto-Read
on, or Off to turn Auto-Read off. Press the confirm button. When Auto-Read is off, use the 0A/100%T and
keys to take the reference and
measurement scans respectively. When Auto-Read is on, the keys on the keypad are still functional. This enables
a user to carry out a new reference scan at any time by pressing the 0A/100%T key prior to lowering the sampling head, or to perform a series of readings or reference scans without raising the sampling head.
2.7. Quality Assurance
Quality Assurance is provided to help minimize the impact of pipetting errors. If Quality Assurance is switched on from the Preferences Screen, the instrument will ask the user for a second reference to compare with the first to ensure that the reading is reasonable. If the second reading is not sufficiently close to the first reading, the user will be asked to replace the reference again. The process will be repeated until two successive good readings (within 0.02 A) are identified and the reference will be the average of these readings.
Note: When the user is prompted for the second reference, the user is required to clean the surface and reapply the sample. The 0A/100%T key is disabled until
theuserliftsthesamplingheadtopreventtheuserfromjustre-readingthesame
sample.
0.5 mm pathlength
0.2 mm pathlength
9
5061-049 REV 1.4
Key Action
On/Off key Turns the instrument on/off.
Arrow keys Use the four arrow keys to navigate around
the display and select the required setting from the active (highlighted) option.
View options

View options for that application mode. Some of these are common to all applications and are described below. Options unique to an application are described in the relevant section.
Alphanumeric keys Use these to enter parameters and to write
text descriptions. Use repeated key presses to cycle through lower case, number and upper case. Leave for 1 second before entering the next character. Use the C key to backspace and the 1 key to enter a space.
Escape/Cancel
Esc
Escape from a selection and return to the previous folder.
Set Reference
OA
100%T
Set reference to 0.000 A or 100%T on a reference solution at the current wavelength in the mode selected. When in scan mode, make a reference scan.
Confirm selection
Confirms a selection.
Take measurement
Makes a measurement.
2.8. Keypad and display
The back-lit liquid crystal display is very easy to navigate around using the alphanumeric entry and arrow keys on the hard-wearing, spill-proof membrane keypad.
OA
100%T
Esc
Esc
1
2
3
abc
def
5
6
4
mno
jkl
ghi
7
8
9
wxyz
tuv
pqrs
C
0
+ / -
A
C
G
T/U
0A
100% T
Alphanumeric
Confirm selection
Escape/Cancel
On/Off
Set reference
Take measurement
View options
Arrow keys
Display screen
10
5061-049 REV 1.4
Options Sub Menu (select using keypad numbers)
1. View Parameters for the experiment.
2. Print the results. 3, 4, 5, 6 Described in the relevant application.
7. Define the Sample Number you wish to start from.
8. Save the parameters as a method to a defined folder name with
a defined method name.
9. Toggle Auto-Print on/off. Default is off.
Exit options by pressing Esc
Esc
or wait.
Experienced operators can use the numeric keys as a shortcut to the option required without needing to enter the Options menu.
2.9. Software style
The user interface is built around folders of files which are displayed on the first page when the instrument is switched on. Different folders are numbered and opened by using the associated number key on the keypad.
1. Standard life science applications such as nucleic acid and protein assays.
2. General spectroscopic applications.
3. A folder to store your more frequently used
methods. (Inactive when empty)
4. Contains 9 folders that can store less frequently used methods. Up to 9 methods per folder are allowed, totaling 81 methods.
5. Instrument set up options and games. (Games only available if selected via the
Preferences Screen)
Summary Function Keypad Number Description
11
5061-049 REV 1.4
3. OPERATION AND MAINTENANCE
3.1. Sample application guide
Step 1
Lift the sampling head to the vertical position and using a low-volume (0–10 µl) pipette take up approximately 2 µl of sample. A reliable pipette and matching high quality tips are strongly recommended.
When removing a very small aliquot from a larger volume, ensure that the sample is truly representative. Filtration may also be advantageous.
Step 2
Carefully apply the sample so that it sits over the black spot between the four alignment spots, (see Figure: 4).
Take care not to introduce bubbles into the sample.
After the plunger is depressed, lift the pipette up carefully so as not to drag the droplet off-target.
Step 3
Observe the position and shape of the drop; it should resemble the one shown in Figure: 6a rather than 6b. A noticeably spread-out sample indicates the target area may be contaminated and requires cleaning (see below). Gently lower the sampling head.
Figure 4
Figure 5
Figure 6a
Figure 6b
Return light path
12
5061-049 REV 1.4
Step 4
If a program has been selected and the Auto-Read function is set to On (see section 2.6), the sample reading process will begin automatically – a reference scan will be taken first, followed by the measurement scan.
If the Auto-Read function is set to Off, use the 0A/100%T and
buttons to take the reference and measurement scans respectively. Note: It is not recommended to raise and lower the head repeatedly to take multiple measurements of one sample, this can cause the droplet to disperse. If repeat measurements are required use the button
Step 5
If you wish to keep the sample it can be recovered after the reading has been taken using a pipette.
Step 6
After taking the reading, both the top and bottom plates should be cleaned by wiping away the sample using a soft lint-free tissue. Wipe the bottom plate towards you and the top plate upwards to avoid contaminating the return light path to the rear (see previous figure 6a).
It may be necessary to remove sticky or dried-on residues. Water or a dilute (2%) detergent solution should suffice. After using detergent, the plates should be wiped a second time with either water or Isopropanol (2-propanol).
3.2. Sample Plate replacement
If after careful cleaning the droplet shape is still unsatisfactory, or if either of the plates becomes damaged, or there is deterioration of the hydrophobic coating, then it will be necessary to replace them as a pair. The glass plates are permanently fixed within plastic housings which clip to the sample handling module. Replacements are available from your supplier.
Exchanging the sample plates is simple (see Figures: 8a and 8b). They are available from your supplier under part number 28-9244-06
1. Remove the top sample plate. Put one thumb at the base of the sample plate
and with the other hand gently lever down the tab at the top of the plate.
2. Replace the top sample plate. Insert the bottom tab and hold this in place with
one thumb and then gently push the tab at the top until the plate clicks into place. Once in position, two silver pins will be visible either side of the top sensor.
3. Remove the bottom sample plate. Gently lever the plate upwards using the
thumb tab at the front of the plate and then pull it towards you to remove it completely.
4. Replace the bottom sample plate. Gently insert the tabs at the back of the plate
as far as they will go and then push down the front of the sample plate so the tabs underneath are fully pushed in.
5. Always calibrate the system after replacing the sample plates.
Figure: 7a
28955301.
Locate the screw above the bottom sample plate. Turn counter clockwise to loosen. Remove the bottom sample plate. Gently lever the plate upwards using the thumb tab at the front of the plate and then pull it towards you to remove it completely.
Replace the bottom sample plate. Gently insert the tabs at the back of the plate as far as they will go and then push down the front of the sample plate so the tabs underneath are fully pushed in. Insert screw back in, turn clockwise to tighten.
13
5061-049 REV 1.4
3.3. Cleaning and general maintenance
Before cleaning the case of the instrument, switch off the instrument and disconnect the power cord.
Clean all external surfaces using a soft damp cloth. A mild liquid detergent may be used to remove stubborn marks.
3.4. Return for repair
The responsibility for decontamination of the instrument lies with the customer. The case may be cleaned with mild detergent or an alcohol such as ethanol or Isopropanol.
Examination or repair of returned instruments cannot be undertaken unless they are accompanied by a decontamination certificate signed by a responsible person.
Figure: 8a
Figure: 8b
3.3. Cleaning and general maintenance
Before cleaning the case of the instrument, switch off the instrument and disconnect the power cord.
Clean all external surfaces using a soft damp cloth. A mild liquid detergent may be used to remove stubborn marks.
3.4. Return for repair
The responsibility for decontamination of the instrument lies with the customer. The case may be cleaned with mild detergent or an alcohol such as ethanol or Isopropanol.
Examination or repair of returned instruments cannot be undertaken unless they are accompanied by a decontamination certificate signed by a responsible person.
This could be of the following form:-
3.5. Lamp replacement
The xenon lamp should not need replacement until used for several years. In the unlikely event that it does fail, this should be undertaken by a service engineer from your supplier.
3.6. Replacing the printer paper
Spare paper for the printer (20 rolls) is available from your supplier under part
number 28918226
Step 1 Keep the power on. Lift off the paper cover. Step 2 Lock the platen (horizontal position) and feed the paper into the slot.
The drive will engage automatically and take up the paper.
Step 3 It may help to release the platen lock (turn flat green catch clockwise) and
turn the green knob manually.
Step 4 Replace the cover.
Figure: 8a
Figure: 8b
Please contact technical support at support@hbiosci.com for assistance.
14
5061-049 REV 1.4
4. LIFE SCIENCE
4.1. File system
The Life Science Screen is organized into six sub-folders as shown below:
Folder Keypad Number Application Function
1 DNA Concentration and purity check for DNA samples 2 RNA Concentration and purity check for RNA samples 3 Oligo Concentration and purity check for Oligonucleotide samples 4 Tm Calculation DNA melting point calculator 5 Cy Dye Labeling efficiency measurement 6 Protein Folder containing methods for protein determination
Pressing 6 enters the Protein sub-folder which contains various standard methods for protein determination:-
4.2. Nucleic Acids
4.2.1. Theory
•Nucleicacidscanbequantifiedat260nmbecauseatthiswavelengththere
is a clearly defined peak maximum. A 50 µg/ml DNA solution, a 40 µg/ml RNA solution and 33 µg/ml solution of a typical synthetic Oligonucleotide all have an optical density of 1.0 A in a 10 mm pathlength cell. These factors (50, 40 and 33 respectively) can be inserted into the formula (1) below, although they do vary with base composition and this can be calculated more precisely if the base sequence is known.
(1) Concentration = Abs260 * Factor (where * means “multiplied by”)
15
5061-049 REV 1.4
•NanoVuePluswilldefaulttofactors50fordoublestrandedDNA,40forRNA
and 33 for single stranded DNA and Oligonucleotides. It also allows manual compensation for dilution - aided by a dilution calculator.
•Nucleicacidsextractedfromcellsareaccompaniedbyproteinandextensive
purification is required to remove the protein impurity. The 260/280 nm Absorbance ratio gives an indication of purity, however it is only an indication and not a definitive assessment. Pure DNA and RNA preparations have expected ratios of 1.7–1.9 and ≥ 2.0 respectively. Deviations from this indicate the presence of impurity in the sample, but care must be taken in the interpretation of results.
•The260nmreadingistakennearthetopofabroadpeakintheAbsorbance
spectrum for nucleic acids, whereas the 280 nm reading naturally occurs on a steep slope, where small changes in wavelength will result in large changes in Absorbance. Consequently, small variations in wavelength accuracy have a much larger effect at 280 nm than at 260 nm. It follows therefore that the 260/280 ratio is susceptible to this effect and users are warned that spectrophotometers of different designs may give slightly different ratios.
•Inpractice,concentrationalsoaffectsthe260/280ratioastheindividual
readings approach the instrument’s detection limit. If a solution is too dilute, the 280nm reading shows a greater proportional interference from background and as the divisor becomes smaller, has a disproportionate effect on the final result.
It is advisable to ensure that the Abs260 value is greater than 0.1 A for accurate measurements.
•AnelevatedAbsorbanceat230nmcanalsoindicatethepresenceofimpurities.
230 nm is near the Absorbance maximum of peptide bonds and may also indicate interference from common buffers such as Tris and EDTA. When measuring RNA samples, the 260/230 ratio should be > 2.0. A ratio lower than this is generally indicative of contamination with guanidinium thiocyanate, a reagent commonly used in RNA purification, which absorbs over the 230–260 nm range. A wavelength scan of the nucleic acid is particularly useful for RNA samples.
•NanoVuePlusdisplaystheindividualAbsorbancevaluesandthe260/280and
260/230 ratios on the left side of the screen and attention should be given to these as well as the final result on the right hand side.
Use of Background Correction
•BackgroundCorrectionatawavelengthwellapartfromthenucleicacidor
protein peaks is often used to compensate for the effects of background
Absorbance.Theprocedurecanadjustfortheeffectsofturbidity,stray
particulates and high-Absorbance buffer solutions.
•NanoVuePlususesBackgroundCorrectionat320nmbydefaultonallLife
Science Applications, particularly nucleic acid measurements. It is particularly recommended since very small samples are particularly susceptible to stray particulates. The Background function toggles On and Off with either left/right arrows from the relevant page.
•Ifitisused,therewillbedifferentresultsfromthosewhenunused,because
Abs320 is subtracted from Abs260 and Abs280 prior to use in equations: Concentration = (Abs260 – Abs320) * Factor Abs ratio 260/280 = (Abs260 – Abs320) / (Abs280 –Abs320) Abs ratio 260/230 = (Abs260– Abs320) / (Abs230 –Abs320)
16
5061-049 REV 1.4
Note:
•TheAbsorbancemaximumnear260nmandAbsorbanceminimumnear230nm.
•Theflatpeaknear260nmandsteepslopeat280nm.
•ThereisverylittleAbsorbanceat320nm.
4.2.2. DNA measurement
The procedure is as follows:
Step 1
Press 1 to select DNA mode.
Step 2
Select Pathlength using the left and right arrows. Options are 0.5 mm, 0.2 mm or Automatic. Press the down arrow.
Step 3 (Dilution Factor known) Enter the Dilution Factor using the keypad numbers (range 1.00 to 9999). Use the C button to backspace and clear the last digit entered.
OR
Step 3 (calculate Dilution Factor) Press
 to enter the Dilution Factor Screen (see second parameter screen to
the left).
Enter the Volume of the sample using the keypad numbers (range 0.01 to 9999). Press the down arrow.
Enter the volume of the Diluent using the keypad numbers (range 0.01 to 9999).
Press OK
to calculate the Dilution Factor and return to the Parameters Screen
OR Press Cancel to cancel the selections and return to the Parameters Screen.
Step 4
Background Correction at 320 nm will be on. It may be turned off with the left or right arrows. Press the down arrow.
Step 5
Select the Units of measurement using the left and right arrows. Options: µg/ml, ng/µl, µg/µl. Press the down arrow.
Step 6
Enter the Factor using the keypad numbers. The default value is 50, the range is
0.01 to 9999.
Step 7 Press OK
to enter the Results Screen and begin making measurements.
Typical spectral scan of a Nucleic Acid:-
17
5061-049 REV 1.4
OR
Esc
to return to the Life Science Screen.
Results Screen
Step 8
Pipette on the reference sample and lower the sampling head. If Auto-Read is off, press the 0A/100%T key. This will be used for all subsequent samples until changed. If QA is switched on the sample will need to be replaced and the 0A/100% T key pressed again.
Step 9
Clean the top and bottom plates, pipette on the sample and lower the sampling head. If Auto-Read is off, press
. This measures at the selected wavelengths and displays the results. The ratio of the Absorbance values at wavelengths 1 and 2 are calculated. The Concentration is based on the Absorbance at wavelength 1. Repeat step 9 for all samples.
Press
Esc
to return to the Life Science Screen.
Press
 to display available Options which are described below.
Options (select using keypad numbers)
1. Return to Parameters Screen (step 1 above).
2. Print result via selected method.
3. Toggle Graph on/off. The graph shows a Wavescan plot across the range
220 nm to 320 nm with cursors denoting 230, 260, 280 and (if Background Correction selected) 320 nm.
7. Sample Number – add a prefix to the sample number and reset the
incrementing number to the desired value.
8. Save Method – use the alpha-numeric keys to enter a name for the method and
press Save
.
9. Auto-Print – toggles Auto-Print on/off.
Exit Options by pressing
Esc
or wait.
4.2.3. RNA measurement
The procedure is as follows:
Step 1
Press 2 to select RNA mode.
Step 2
Select Pathlength using the left and right arrows. Options are 0.5 mm, 0.2 mm or Automatic. Press the down arrow.
Step 3 (Dilution Factor known)
Enter the Dilution Factor using the keypad numbers (range 1.00 to 9999). Use the C button to backspace and clear the last digit entered.
OR
Step 3 (calculate Dilution Factor)
Press
 to enter the Dilution Factor Screen (see second image to the left).
Enter the Volume of the sample using the keypad numbers (range 0.01 to 9999).
Press the down arrow.
Enter the volume of the Diluent using the keypad numbers (range 0.01 to 9999).
Press OK
to calculate the Dilution Factor and return to the Parameters Screen.
OR Press
Esc
to cancel the selections and return to the Parameters Screen.
18
5061-049 REV 1.4
Step 4
Select whether the Background Correction at 320 nm is to be used or not with the left and right arrows.
Press the down arrow.
Step 5
Select the Units of measurement using the left and right arrows. Options: µg/ml, ng/µl, µg/µl.
Press the down arrow.
Step 6
Enter the Factor using the keypad numbers. The default value is 40, the range is
0.01 to 9999.
Step 7
Press OK
to enter the Results Screen and start making measurements.
OR
Esc
to return to the Life Science Screen.
Results Screen
Step 8
Pipette on the reference sample and lower the sampling head. If Auto-Read is off, press the 0A/100%T key. This will be used for all subsequent samples until changed. If QA is switched on the sample will need to be replaced and the 0A/100% T key pressed again.
Step 9
Clean the top and bottom plates, pipette on the sample and lower the sampling head. If Auto-Read is off, press
. This measures at the selected wavelengths
and displays the results. The ratio of the Absorbance values at wavelengths 1 and 2 are calculated. The Concentration is based on the Absorbance at wavelength 1.
Repeat step 9 for all samples.
Press
Esc
to return to the Life Science Screen.
Press
 to display available Options which are described below.
Options (select using keypad numbers)
1. Return to Parameters Screen (step 1 above).
2. Print result via selected method.
3. Toggle Graph on/off. The graph shows a Wavescan plot across the range
220 nm to 320 nm with cursors denoting 230, 260, 280 and (if Background Correction selected) 320 nm.
7. Sample Number – add a prefix to the sample number and reset the incrementing number to the desired value.
8. Save Method – use the alpha-numeric keys to enter a name for the method and press Save
.
9. Auto-Print – toggles Auto-Print on/off.
Exit Options by pressing
Esc
, or wait.
19
5061-049 REV 1.4
4.2.4. Oligonucleotide measurement
The procedure is as follows:
Step 1
Press 3 to select Oligo mode.
Step 2
Select Pathlength using the left and right arrows. Options are 0.5 mm, 0.2 mm or Automatic. Press the down arrow.
Step 3 (Dilution Factor known) Enter the Dilution Factor using the keypad numbers (range 1.00 to 9999). Use the C button to backspace and clear the last digit entered.
OR
Step 3 (calculate Dilution Factor) Press
 to enter the Dilution Factor Screen.
Enter the Volume of the sample using the keypad numbers (range 0.01 to 9999).
Press the down arrow.
Enter the volume of the Diluent using the keypad numbers (range 0.01 to 9999).
Press OK
to calculate the Dilution Factor and return to the Parameters Screen.
OR Press
Esc
to cancel the selections and return to the Parameters Screen.
Step 4
Select whether the Background Correction is to be used or not with the left and right arrows.
Press the down arrow.
Step 5
Select the Units of measurement using the left and right arrows. Options: µg/ml, ng/µl, µg/µl and pmol/µl. If pmol/µl is selected the factor changes to a selection table denoting the ratios of the 4 bases in the structure.
Press the down arrow.
Step 6 (Units not pmol/µl) Enter the factor using the keypad numbers. The default value is 33, the range is
0.01 to 9999.
OR
Step 6 (Units pmol/µl) Enter the proportions of bases present using the keypad numbers and up and down arrows to move between boxes. The default is 10 for each, the range is 0 to
9999.
Step 7
Press OK
to enter the Results Screen and start making measurements
OR
Esc
to return to the Life Science Screen.
Results Screen
Step 8
Pipette on the reference sample and lower the sampling head. If Auto-Read is off, press the 0A/100%T key. This will be used for all subsequent samples until changed. If QA is switched on the sample will need to be replaced and the 0A/100% T key pressed again.
Step 9
Clean the top and bottom plates, pipette on the sample and lower the sampling head. If Auto-Read is off, press
. This measures at the selected wavelengths
20
5061-049 REV 1.4
and displays the results. The ratio of the Absorbance values at wavelengths 1 and 2 are calculated. The Concentration is based on the Absorbance at wavelength 1.
Repeat step 9 for all samples.
Press
Esc
to return to the Life Science Screen.
Press

to display available Options which are described below.
Options (select using keypad numbers)
1. Return to Parameters Screen (step 1 above).
2. Print result via selected method.
3. Toggle Graph on/off. The graph shows a Wavescan plot across the range
220 nm to 320 nm with cursors denoting 230, 260, 280 and (if Background Correction is selected) 320 nm.
7. Sample Number – add a prefix to the sample number and reset the incrementing number to the desired value.
8. Save Method – use the alpha-numeric keys to enter a name for the method and press Save
.
9. Auto-Print – toggles Auto-Print on/off.
Exit options by pressing
Esc
, or wait.
4.2.5. Tm Calculation
This utility calculates the theoretical melting point from the base sequence of a primer. It is done using nearest neighbor thermodynamic data for each base in the nucleotide chain in relation to its neighbor (Breslauer et al, Proc.Natl. Acad. Sci. USA Vol 83, p3746 1986). The data obtained are useful both in the characterization of oligonucleotides and in calculating Tm for primers used in PCR experiments.
The ACGT/U sequence entered into the utility is used to calculate the theoretical T
m
, the theoretical Absorbance (Absorption units/mmol) and the conversion factor (mg/ml). This is because the stability of a bent and twisted sequence of bases such as an oligonucleotide is dependent on the actual base sequence. The calculated
thermodynamicinteractionsbetweenadjacentbasepairshavebeenshownto
correlate well with experimental observations. This utility uses matrices of known, published thermodynamic values and extinction
coefficients to calculate T
m
and the theoretical Absorbance/factor respectively of an
entered base sequence.
Tm
This is calculated using the equation:-
Tm = ∆H × 100
- 273.15 +16.6 log [salt]
∆S + (1.987 × loge(c/4 +53.0822)
where ∆H and ∆S are the enthalpy and entropy values summed from
respective 2 × 4 × 4 nearest neighbor matrices
c is the Primer concentration of oligonucleotide (pmoles/ml) in the
calculated Tm or the measured concentration in measured Tm. In the latter case concentration is obtained from the equation:
c = Abs(260 nm) × Calculated factor × pathlength multiplier × 10 000 MW
21
5061-049 REV 1.4
Calculated factor and MW are defined below
[salt] is the buffer molarity plus total molarity of salts in the hybridization
solution (moles/l)
Weights for ∆Sareindexedbyadjacentpairedbases.
A similar equation applies to weights for ∆H,againindexedbyadjacentbases.
Note that bivalent salts may need normalizing using a multiplying factor of 100 because of their greater binding power.
Theoretical Absorbance
The Theoretical Absorbance is based on a calculation as follows:
Foreachadjacentpairofbases(nearestneighbors)anextinctioncoefficientweight
is accumulated using a 4 × 4 table (one for either DNA or RNA). This total weight is doubled and then for each internal base a counterweight is subtracted using another 1 × 4 table. The end bases are excluded from the latter summation.
That is:
Total Extinction Coefficient E = ∑ (2 × aTable[base_type][base(n)][base(n+1)])
- ∑ (tTable[base_type][base(n]])
Conversion factor
The Conversion Factor is given by = Molecular weight
ABCDE
∑ E
ABCDE
where E
ABCDE
= [ 2 × (EAB +EBC +ECD +EDE) –EB –EC –ED ]
•Themolecularweight(MW)ofaDNAoligonucleotideiscalculatedfrom:
MW (g/mole) = [(dA × 312.2) + (dC × 288.2) + (dG × 328.2) + (dT × 303.2.)] + [(MW counter-ion) × (length of oligo in bases)]
(for RNA oligonucleotide, (dT × 303.2) is replaced by (dU × 298.2)
 TheMWcalculatedusingthisequationmustbeadjustedforthecontributionof
the atoms at the 5’ and 3’ ends of the oligo.
For phosphorylated oligos: add: [17 + (2 × MW of the counter-ion)] For non-phosphorylated oligos: subtract: [61 + (MW of the counter-ion)]
The MW (g/mole) of the most common oligo counter-ions are; Na (sodium) 23.0 K (potassium) 39.1 TEA (triethylammonium) 102.2 Other Defaults to 1.0 (H) Variable 0.1–999.9 in next option box.
Calculated molecular weight: a weight is added for each base looked up from a table. The weight of the counter ion is added for every base from a small table for the known ions. If phosphorylated, then the system adds 17.0 plus two counter ions otherwise it subtracts 61.0 and one ion.
TheoreticalAbsorbance:foreachadjacentpairofbases(nearestneighbors)a
weight is accumulated using a table. For each internal base a weight is subtracted using another table. Separate tables are used for DNA and RNA.
22
5061-049 REV 1.4
Calculated factor:thisisjustthecalculatedmolecularweightdividedbythe
theoretical Absorbance.
The procedure is as follows:
Step 1
Press 4 to enter the Tm Calculation Parameters Screen.
Step 2 Select the Base Type: DNA or RNA.
Press the down arrow.
Step 3 Select whether the sample is Phosphorylated or not (Yes or No).
Press the down arrow.
Step 4 Enter the Primer Conc. using the keypad numbers (range 0.000 to 99.9) in pmole/ml.
Press the down arrow.
Step 5 Enter the Buffer Molarity (buffer molarity + total molarity of salt in moles/l) using the keypad numbers (range 0.000 to 10).
Press the down arrow.
Step 6 Select the Counter Ion: Na, K, TEA or Other.
Step 7 (if Other selected, see the previous section)
Enter the molecular weight (Other MW) of the Counter Ion used.
Step 8 Press Next
to select these parameters and go on to the Base Sequence
Screen
OR
Press Cancel
Esc
to return to the Life Science Screen.
Step 9 Select the Pathlength of the sample cells. Options are 0.5 mm, 0.2 mm or Automatic.
Step 10 Enter the known Base Sequence triplets using the number keys: 1 for A, 3 for C, 4 for G and 6 for T or U. Note that a comma is added after each triplet to improve readability
Step 11 Press OK
to select these parameters and start to measure Tm
OR
Press Cancel
Esc
to return to the Parameters Screen.
23
5061-049 REV 1.4
Results Screen
Step 12
Pipette on the reference sample and lower the sampling head. If Auto-Read is off, press the 0A/100%T key. This will be used for all subsequent samples until changed. If QA is switched on the sample will need to be replaced and the 0A/100% T key pressed again.
Step 13 Clean the top and bottom plates, pipette on the sample and lower the sampling head. If Auto-Read is off, press
. The unit measures the Absorbance and uses
this information to calculate the Measured Tm.
Repeat step 13 for all samples.
Press
Esc
to return to the Life Science Screen.
Press

to display available Options which are described below.
Options (select using keypad numbers)
1. Return to Parameters Screen (step 1 above).
2. Print result via selected method.
7. Sample Number – add a prefix to the sample number and reset the
incrementing number to the desired value.
8. Save Method – use the alpha-numeric keys to enter a name for the method and press Save
.
9. Auto-Print – toggles Auto-Print on/off.
Exit options by pressing
Esc
, or wait.
4.2.6. CyDye measurement
Measurement of the labeling efficiency of fluorescently labeled DNA probes before 2-color microarray hybridization ensures that there is sufficient amount of each probe to give satisfactory hybridization signals. These data also provide an
opportunitytobalancetherelativeintensitiesofeachfluorescentdyebyadjusting
the concentration of each probe before hybridization. The DNA yield is measured at 260nm while the incorporation of fluorescein, Cy3, Cy5 and other dyes are measured at their absorption peaks. This method may also be useful for measuring the yields and brightness of fluorescently labeled in-situ hybridization probes.
Equations used in the software are detailed below:
DNA Concentration = (A
260
– A
Background
) * NucleicAcid Factor
DNA Quantity = DNAConcentration * Volume
Dye Concentration = ((A
Dye
- A
Background
) / ExtinctionCoefficient
Dye
) * PathlengthFactor * Dilution_Factor
* 10
6
Dye Quantity = DyeConcentration * Volume
24
5061-049 REV 1.4
Dye Incorporation = DyeQuantity / DNAQuantity
Dye FOI = DyeQuantity * (324.5 / DNAQuantity)
Step 1 Press 5 to enter the Cy Dye Parameter Screen from the Life Sciences screen.
Step 2 Either select the Dye Name of the dye being used by using the left and right arrows, or select “Custom” if you wish to create a new entry. If a standard dye is selected, press Next to move to the next screen (Step 7). If “Custom” has been selected press the down arrow.
Step 3 (Custom only) Enter the Wavelength of the dye peak Absorbance.
Press the down arrow.
Step 4 (Custom only) Enter the dye Extinction Coefficient.
Press the down arrow.
Step 5 (Custom only) Enter the dye 260 nm Correction factor (if known), otherwise set at zero.
Step 6 (Custom only) Enter a name for the dye to identify it correctly.
To complete the entry press Next
to enter the second dye detail entry screen.
OR
Press Cancel
Esc
to return to the Life Science Screen.
Step 7 If there is no second dye select None, otherwise proceed as for Dye 1.
Press Next
to complete data entry and enter the Parameter screen
OR
Press Cancel
Esc
to return to the Dye 1 data entry screen.
Step 8
If the concentration is known select the appropriate Pathlength, otherwise leave it as Automatic
Step 9 Unless there are complicating absorbencies at the Background wavelength, leave this as On.
Step 10 If using standard dyes, the Background Wavelength is set automatically, but if there are interferences at this default wavelength, the user can select a wavelength where the background is at or close to the reference sample. If no such wavelength is available set the previous box Background Correction to Off. This will give rise to noisier results though
Loading...
+ 53 hidden pages