biochrom Libra S6 User Manual

Libra S6
User Manual
English
Deutsch
Français
Italiano
Declaration of Conformity
This is to certify that the Libra S6 and S6H Visible Spectrophotometers Part number 80-5000-10 and 80-5000-11
Serial number 88000 onwards
manufactured by Biochrom Ltd. conform to the requirements of the following Directives-: 73/23/EEC & 89/336/EEC
Standards to which conformity is declared
EN 61 010-1: 2001 Safety requirements for electrical equipment for measurement, control and
laboratory use.
EN 61326: 1998 Electrical equipment for measurement, control and laboratory use – EMC
requirements
Signed: Dated: 26
th
Oct 2004
David Parr Managin g Direc tor Biochrom Ltd
Postal address Telephone Telefax Biochrom Ltd +44 1223 423723 +44 1223 420164
22 Cambridge Science Park Milton Road e mail: enquiries@biochrom.co.uk website: http://www.biochrom.co.uk Cambridge CB4 0FJ England
Registered in England No: 974213 Registered Office: 22 Cambridge Scie nce Park, Milton Road, Cambridge CB4 4FJ, England
.
Biochrom Ltd
Certificate No. 890333
CONTENTS
Unpacking, Positioning and Installation 1 Essential Safety Notes 1
OPERATION 2
Introduction 2 Using the Instrument 3 Absorbance and % Transmission 4 Absorbance Ratio 5 Cell Density 5 Scan 6 Factor Concentration 7 Standard Curve 8 Kinetics 10 To recall a saved method 11
SET UP 12 ACCESSORIES 13 ERROR MESSAGES 13 OUTPUT OF RESULTS 13
Use with serial printer 13 Use with chart recorder 13
USE WITH PC AND THE GRAFICO PC UTILITY SOFTWARE 14
Installation 14 Introduction 14 Menu Descriptions 15 Practical Aspects 16
MAINTENANCE 17
After Sales Support 17 Lamp Replacement 17 Cleaning and gen eral care of th e instru men t 18 Changing cell holder or removal for cleaning 18 Changing the brightness of the display 18
SPECIFICATION AND WARRANTY 19
___________________________________________________________________ Issue 01 - 12/2004 Libra S6, English 1
Unpacking, Positioning and Installation
Inspect the instru ment for any signs of damage caused in tran si t. If any damage is
discovered, inform your supp lier immediately. Check the position of the metal lamp bracket insid e the lamp access area.
Ensure your proposed installation site conforms to the environmental conditions for safe
operation:
Indoor use only Temperature 5°C to 35 °C. No te that if you use the instr ument in a room sub ject to extremes of temperature change du r ing the day, it may be necessary to recalibrate (b y switching off and then on again) once thermal equilibrium has been established (2-3 hours). Maximum relative humidity of 80 % up to 31°C decreasing linearly to 50 % at 40°C
The instrument must be placed on a hard, flat bench or table that can take its weight (<2
kg) such that air is allowed to circulate freely around the inst rument.
This equipment must be connected to the power supply with the power cord supplied. It
can be used on 90 - 240V supplies.
Switch on the instrument via the display after it has been plugged in. The instrument
performs a series of self-diagnostic checks for lamp performanc e, wavelength calibration and diode array pixels; press F2 to proceed.
If the instru ment h as ju s t b een unpacked or h as be en stored in a co ld environmen t, i t should be allowed to come to thermal equilibrium for 2-3 hours in the laboratory before switching o n to prevent calibration failure as a result of internal condensation.
The cell holder supplied with the instrument accepts standard 10mm pathlength glass or
plastic cells (adap t er s are avai lable to convert it to accept 10, 12 and 16mm diameter test tubes). It can be removed for cleaning if spillages occur by undoing the screws that hold it or it can be flushed through with water in situ.
If the instrument has a heated cell holder option and it is on, allow 10 minutes for it to
come to thermal equilibrium. This cell holder cannot be removed.
If this equipment is used in a manner not specified or in environmental conditions not appropriate for safe operation, the protection provided by the equipment may be impaired and instrument w a rranty withdrawn.
Essential Safety Notes
There are a number of warning labels and symbols on your instrument. These are there to inform you where potential danger exists or particular caution is required. Before commencing installation, please take time to familiarise yourself with these symbols and their meaning.
Caution (refer to accompanying documents). Background colour yellow, symbol and outline black.
___________________________________________________________________ 2 Libra S6, English Issue 01 – 12/2004
OPERATION
Introduction
Your spectrophotometer is a simple-to-use, microprocessor controlled instrument. It is a diode array product (1024 pixels), has no moving parts and scans very quickly.
After switch on, calibrati on and press in g F2 to proceed the home page is shown offer ing the choice of
Repeat last op er ation
Make a measurement
Set up instrume nt
“Repeat last operation” returns the user to the last screen displayed when the instrument was switched off, and provides a short cut to the last test that was performed.
Within “Make a measurement” your spectrophotometer has facilities for:
measurement of absorba n ce, % transmission, r atio and conce n tr at ion values
cell culture optical density measurements at 60 0nm
entry of a multi point stan dard curve in memory
output of wavelength scan to display
output of kinetics assay to display
application of a factor to an absorbance change over a specified time interval for an
enzymatic determination (reaction rate)
storage of up to 99 user defined methods
Within “Set up instrument” your spectrophotometer can be set up to
select the display language option (English, French, German, Spanish, Italian)
link via a serial lead to either a serial printer for hardcopy ou tput or to a PC for
download of results to spreadsheet
link via a converte r lead to chart record e r
set the da te for print outs
The instrument is supplied with Grafico PC utility - on the accompanying CD - and a serial lead. These provide the user with th e means to cap t ure, print an d s tore data from the instrument to a PC. Specifically it
produces a printable graphical plot of the scan, in Abs
logs date, time and serial number with any output from the instrument
produces a results log in order to store, tabulate and subsequently print output from
the instrument
enables expor t of the outpu t from the instrument to Excel as a text file
A tutorial on UV/V isib l e spe ctrophotometry is included as part of the Grafico software.
___________________________________________________________________ Issue 01 - 12/2004 Libra S6, English 3
Using the Instrument
The back-lit liquid crystal display is very easy to navigate around using the function / select and arrow keys on the hard wearing, spill proof membrane keypad.
Keypad
F1, F2, F3, F4
The function select / entry soft keys on the keypad are situated next to the corresponding option on the display, and are used to select an appropriate mode
3456
When a parameter within a mode needs selecting or changing (as indicated by highlighted text on the display), the four arrow ke ys (3456) are used in conjunction with the function keys to make that selection or chan ge. Use F4 to implement change, followed by 34 to choose between options indicated, and 56 to enter alphanumeric characters (for example in th e selection of a wavelength o r en try of a method title). Then u se F 4 to acce p t th e ch an ge made.
to escape or stop making measurements
0A/100%T to set reference to 0.000AU or 100%T on a reference solution at the
current wavelength in the mode selected , or to do a referen c e scan if in scan mode
to start making measurements
Display
The following symbols will appear in bottom right hand corner and mean
the following: Use 3456 to select option R T Ready to set reference or r u n sample
Note that the light beam shines from LEFT to RIGHT through the cell chamber; ensure the cell is inserted in the corre ct ali gn ment. The optical height is 15mm, and the minimum volume that can be used is approx. 700µl in a semi-micro cell.
___________________________________________________________________ 4 Libra S6, English Issue 01 – 12/2004
Absorbance and % Transmission
This makes simple absorbance measurements on samples, measuring the amount of light that has passed through a sample relative to a blank (this can be air). The procedure is as follows:
Option on display or action Press Comment
Make a measurement F2 Single / Multi λ / Ratio
F1
Single λ
F1 Abs / % T F1 Alternates between the two Set λ F2, then 56
Select wavelength
Accept λ
F2 Insert reference
RT on display
0A/100%T
Used for subsequent samples until changed
Insert sample
Value is displayed
Repeat as necessa r y To exit
To make up to 4 absorbance measurements on the same sample:
Option on display or action Press Comment
Make a measurement F2 Single / Multi λ / Rati o
F1 Multi λ
F2 Set λ’s F1, then 56
Select first wavelength
Select λ F1, then 56
Select second wavelength Repeat as necessa r y All OK F4 Insert reference
RT on display
0A/100%T
Used for subsequent samples until
changed Insert sample
Absorbance values ar e d i s p la yed Repeat as necessa r y
To exit
___________________________________________________________________ Issue 01 - 12/2004 Libra S6, English 5
Absorbance Ratio
This makes simple absorbance ratio measurements on samples, measuring the amount of light that has passed through a sample relative to a blank (this can be air) at two wavelengths. The procedure is as follows:
Option on display or action Press Comment
Make a measurement F2 Single / Multi λ / Rati o
F1 Ratio F3 Remove this row Set λ1 F2, then 56
Select wavelength
Accept λ
F2 Set λ1 F2, then 56
Select wavelength
Accept λ
F2 Insert reference
RT on display
0A/100%T
Used for subsequent samples until changed
Insert sample
Ratio is displayed
Repeat as necessa r y To exit
Cell Density
This function should be used o make an OD600nm reading on a cell culture rather than a direct absorbance reading as it compensates for turbidity using an auto­correction; the absorbance at two wavelengths is measured simultaneously and an algorithm applied to compensate for scattered light. This enables better comparison with results obtained from other instruments. The procedure is as follows:
Option on display or action Press Comment
Make a measurement F2 Cell Densit y / Proteins F2 Insert reference RT on display
0A/100%T
Used for subsequent samples until changed
Insert sample
Value is di splayed; an aut o ­correction factor is app l i ed to th e
Absorbance valu e. Repeat as necessa r y To exit
___________________________________________________________________ 6 Libra S6, English Issue 01 – 12/2004
Scan
An absorption spectrum can be obtained from your instrument, enabling simple identification of peak height and position. The procedure is as follows:
Option on display or action Press Comment
Make a measurement F2 Scan F4 Abs / % T F1 Alternates between the two Insert reference RT on display
0A/100%T
Used for subsequent samples until changed
Insert sample
Scan is displayed
Repeat as necessa r y To identify pea k s: Move cross hairs
34 Abs and λ values appear at top
To zoom in on a region of interest:
Zoom F2, then
3456
Move box that appears on display to area
of interest Zoom in F1 Examine detail Zoom out F1 Return to original data To exit
___________________________________________________________________ Issue 01 - 12/2004 Libra S6, English 7
Factor Concentration
Factor concentration mode is used when a conversion factor is known, and is required to convert the absorbance measurement for a sample at a specific wavelength into a concentration, by a simple multiplication of absorbance x factor. The procedure to define a new method is as follows:
Option on display or action Press Comment
Make a measurement F2 Select a method F3
56
Select method number
New F1
Name is highlighted
Change Name
F4, then 56
Enter first character of name
4, then 56
Enter second character of name
4, then 56
Repeat as necessary
Accept F4
λ is highlighted
Change Wavelength
F4, then 56
Accept F4
Units is highlighted
Select Units
F4, then 56
Accept F4
Cal is highlighted
Select Calibration option
F4, then 34
Select Factor
Enter Factor
F4 then56
. . . .
F3 Moves decimal point Positive o r n egative? F2 Alternat es be tween the two Accept F4
Kinetics
Leave as no
All OK F1 Accept method protocol Run method F1 Insert reference RT on display
0A/100%T
Used for subsequent samples until changed
Insert sample
Concentration is displayed
Repeat as necessa r y To delete F3, then F1 To exit
Note: It is not necessary to enter the name, and this can be omitted for a quick measurement.
___________________________________________________________________ 8 Libra S6, English Issue 01 – 12/2004
Standard Curve
The construction of a multi-point calibration curve from standards of known concentration in order to quantify unknown samples is a fundamental use of a spectrophoto meter; this instrument has the advantage of being able to store this curve as a method, using up to 5 standards.
To include a zero concentration standard, include this in the number of standards to be entered and enter 0.00 f or concentration; use a blank when required to enter standard
The procedure to define a new method is as follows:
Option on display or action Press Comment
Make a measurement F2 Select a method F3
56
Select method number
New F1
Name is highlighted
Change Name
F4, then 56
Enter first character of name
4, then 56
Enter second character of name
4, then 56
Repeat as necessary
Accept F4
λ is highlighted
Change Wavelength
F4, then 56 Accept F4
Units is highlighted
Select Units
F4, then 56 Accept
F4, 6
Cal is highlighted
Select Calibration option
F4, then 34
Select Std
Accept F4
Goes to Calibration Curve page
Set Std
F1 1 is highlighted (maximum is 5) Change
F1, then 56
Enter concentration of standard 1
. . . .
F3 Moves decimal point Accept F4 2 is highlighted Change
F1, then 56
Enter concentration of standard 2
. . . .
F3 Moves decimal point Repeat as necessary Incorrect en try? F3, then F1,
56, F4
Clears entry prior to re-entry
Standards are all O K
F4 Accept Concentrations Insert reference RT on display
0A/100%T
Used for subsequent samples until changed
Insert Standard 1
Absorbance for Std 1 is measured Std 2 is highlighted
Insert Standard 2
Absorbance for Std 2 is measured
Std 3 is highlighted Repeat as necessary Incorrect entry? F3, then F1, Clears entry prior to re-entry
___________________________________________________________________ Issue 01 - 12/2004 Libra S6, English 9
56, F4
All OK F4 Accept Standar d s Change Curve Fit algorithm
F3, then 6, F4
Select linear least squares or
polynomial curve View Graph
F4
Accept graph F3 Can now run samples
To run samples All OK F1 Run F1 Insert reference RT on display
0A/100%T
Used for subsequent samples until
changed Insert sample
Concentration is displayed Repeat as necessa r y
To delete F3, then F1 To exit To delete F2, then F1 To exit
___________________________________________________________________ 10 Libra S6, English Issue 01 – 12/2004
Kinetics
Kinetics studies, where the change in absorbance needs to be followed as a function of time at a fixed wavelength, can be readily performed.
Reagent test kits are routinely used for the enzymatic determination of compounds in food, beverage and clinical laboratories by measuring NAD / NADH conversion at 340 nm. The change in absorbance over a specified time period can be used to provide useful information when an appropriate factor, defined in the reagent kit protocol, is applied. Reaction rate and enzyme activity can be calculated if the factor used takes account of the absorbance difference per unit time, as opposed to the absorbance difference per se.
For this reason, the change in absorbance per minute (A/min), concentration (A/min x factor) and correlation coefficient (cal culated from a best fit of the data points) are displayed. They may not be relevant for simple kinetics experiments.
The procedure to define a new method is as follows:
Option on display or action Press Comment
Make a measurement F2 Select a method F3
56
Select method number New F1
Name is highlighted Change Name
F4, then 56
Enter first character of name
4, then 56
Enter second character of name
4, then 56
Repeat as necessary Accept F4
Wavelength is highlighted Change Wavelength
F4, then 56
Accept F4
Units is highlighted Enter Units
F4, then 56
Accept F4
Cal is highlighted Select Calibration option
F4, then 34
Select Factor Enter Factor
F4 then 56
If required; this is used to co n vert
A/min to Concentration
. . . .
F3 Moves decimal point Positive o r n egative? F2 Alternat es be tween the two Accept F4
Kinetics
F4, then 34
Select Yes or Fixed time *
Accept F4 Start is highlighted Enter Start Time
F4, then
3456
Usually 00m 00s, unless there is a lag time
Accept F4 Interval is highlighted Enter tim e interval between each measurement
F4, then
3456
Minimum interval is 10 seconds
___________________________________________________________________ Issue 01 - 12/2004 Libra S6, English 11
Accept F4 End is highlighted Enter end time
F4, then
3456
Maximum time is 59m 59s after completion of start time Maxi mum number of reading s is 20, so maximum en d time is 20 x the
time interval Accept F4 Run method F1 Insert reference RT on display
0A/100%T
Used for subsequent samples until
changed Insert sample
Abs values displayed for each time
interval
At end of run, calculated ∆A/min,
correlation coefficient and
concentration are displayed To view data Use Page Up and Page Dn F2 or F3 To view graph F1
Return to values Repeat as necessa r y
To exit
* The Fixed time option is for a single time measurement after a specified time, and therefore no options for start time, time interval and graphics are available.
If the instrument is connected to a chart re co r d e r the o utp ut is linearly fitted
between data points as the software automatically interpolates these for the benefit of presentation.
If you have a factory fitted electrical heated cell holder version of the
instrument, go to Set-up to switch this facility on (37°C). Allow 10 minutes for the instrument to come to thermal equilibrium.
To recall a saved method
Option on display or action Press Comment
Make a measurement F2 Select a method F3
56
Select method number Accept F4 Selected method is recalled Run method
F1
Insert reference RT on display
0A/100%T
Used for subsequent samples until
changed Insert sample
___________________________________________________________________ 12 Libra S6, English Issue 01 – 12/2004
SET UP
Option on display or action Press Comment
Set up instrument F3
Set Language
F1
Select displ ay language
56
English, French, German,
Spanish, Italian Accept F1 To exit
Comms/Software Update
F2, F1
Select Communications Select serial printer or PC F1
Alternates between them, with
default settings for each option:
Printer 1 is S1000P
Printer 2 is Martell / Seiko DPU-414
PC is for download to spreadsheet
Set Baud of 9600 or 38400 F2 Alternates between the two Auto-print F3 Select on for automatic increment
of sample number and print out
after measurement Accept F4 To exit
Set Date / Time
F3 Select format F1 European or North American Enter values
3456
Enter as appropriate Accept F4 To exit
Heater control*
F4 May not be available*
Heated cell
34
Select on for thermostatting at
37°C Accept F4 To exit
*Heated cell holder factory fitted version only. This option cannot
be fitted
retrospectively to an instrument.
___________________________________________________________________ Issue 01 - 12/2004 Libra S6, English 13
ACCESSORIES
PC serial cable (spare) 80-3001-00 S1000P serial printer (includes serial printer cable) 80-3002-53 Chart recorder interface cable 80-3003-55 Test tube adapters (10, 12, 16mm) 80-2117-47
ERROR MESSAGES
After switch on, the instrument undergoes self-diagnostic tests for the tungsten lamp, wavelength calibration and diode array as part of its calibration procedure (9 for OK, X for fail). The results of this test are displayed and can be printed out or output to PC for filing and GLP (Good Laboratory Practice) purpose. The messages for tungst e n l amp and / waveleng t h c a l i bration are sel f e xp l anator y, involving checking that the cell compartment is clear or replacement of the tungsten lamp. In the unlikely event of a diode array fail message contact y our local supplier.
OUTPUT OF RESULTS
Use with serial printer
The instrument is designed to print to a 40 col umn serial printer at 96 00 Baud. Output is automatic when a measurement is made. Note that all results can be output to PC using the serial lead and Grafico software supplied on the user manuals CD.
Use with chart recorder
Kinetics results can be output to a chart recorder using the appropriate cable (80­3003-55). Voltage setting is 1V per 1 Absorbance unit (± 10 %) with an offset of 1V = 0.000 Abs on the chart recorder; corresponding %T values are 1V per 100%T (± 10 %) with 0V = 0%T. To make the chart cable yourself, you require a female 9 way D type at one end with two (1 red, 1 black) 4mm banana plugs at the other (depending on the chart recorder) and 2 metres of coaxial cable or screened twin core, with the shield connecting the black plug and pin 5 and the core connecting the red plug and pin 1.
___________________________________________________________________ 14 Libra S6, English Issue 01 – 12/2004
USE WITH PC AND THE GRAFICO PC UTILITY SOFTWARE
Your instrument is supplied with a serial lead and Grafico software (on the user manuals CD) that enables it to be connected to a PC so that results can be capt ured, stored, printed and transferred into other applications easily. In particular, a complete wavelength scan can be visualised on the PC and copied/pasted into a word document or powerpoint presentation. An informative tutorial on aspects of UV/Visible Spectrophotometry is available as part of the software.
Installation
The software takes up approximately 0.5Mb of hard disk space when installed. Proceed as foll ows to install the software:
1. Place CD into the CD drive of the PC
2. Use Windows explorer to locate the setup.exe file Grafico folder within the
appropriately named instrument folder on the user manuals CD
3. Double click on this so that the software installs, filling out the information as
requested.
4. The software can be started directly by Start > Programs > Grafico.
Introduction
When Grafico is selected, you are prompted to enter the file details (note that
the title entered here is used as the title of the wavelength scan graph). After pressing OK, the instrument (it should be already switched on and connected to the PC with the serial lead) is recognised by the software.
There are two parts to the Grafico software, data-logging and scan.
The default mode is data-logging; this receives instrument output from
absorbance, %T, concentration and rate measurements (including time and date stamp).
o Results can be copied from Grafico and pasted directly into Excel
for ease of data transfer. Alternativ ely results can be saved and opened up using Excel.
If scan mode is selected (View > Scan mode), the full 330-800nm wavelength
scan output from the instrument is shown (just press the run key as usual). Multiple peaks can be identified using a trace routine and labelled if required (by dragging the icon at the lef t side of the displayed graph and releasing at the appropriate point).
o Graphs can be co pi ed and pasted into Word, Excel or powerpoint o Graphs can be saved in a format that can be opened directly by
Excel
___________________________________________________________________ Issue 01 - 12/2004 Libra S6, English 15
Menu Descriptions
File
New Clears any existing data and starts a new report. Prompts for file
details (user name, organisation, title, descriptive text)
Save / Save As Saves the data file in the file format selected. The file details are
included with this data
Setup Displays a tabbed dialogue box so that automatic post process
options for saving of graph, printing of graph and the graph scaling parameters can be defined. The default data directory can be
defined and is used for a ll save ope ra tio ns. Print Prints the entire file, including a header if defined in File>New Print Setup Runs the Common Print Dialog function to set up the printer Exit Closes the application
Edit
Copy Copies the data to cl ip board for pasting into another ap p lic a tio n; in
data-logging and scan modes this is text and graphic, respectively Clear Clears the data from the data set Select All Selects data and header together
View
Scan mode Switches between scan and data logging modes. Successive scans
overwrite existing scans on the display and can be saved if the
autosave function is on File details Shows the file details entered at the start (or after File > New) and
allows modification of these details, if requir ed Autoscale Automatically sets the scale of the absorbance axis to optimise
presentation (2.5, 2.0, 1.5, 1.0, 0.5, 0.2 or 0.1A) Set scale Sets the scale to user preference (Full, Auto, Define ) Display grid Toggles on/off the grid on the graph (for presentation purposes) Toolbar View menu bar as icons Status bar View status bar at bottom of display
Help
Tutorial View tutorial on UV/Visible spectrophotometry Help topics View help topics About View version number etc
___________________________________________________________________ 16 Libra S6, English Issue 01 – 12/2004
Practical Aspects
Data logging mode
When exporting rate mode results you can add the time in 10 second intervals to
the spreadsheet manually (note that first data point is after 10 seconds, not zero seconds) and then graph the absorbance / time data (see scan mode – export to excel for more details).
Scan mode
Files can be saved as *.txt, *.csv (opens directly in Excel when double clicked)
or *.wmf (picture) formats
Label a peak by dragging and releasing the icon at the left side of the graph.
The absorbance/wavelength details are shown in the title bar. Dragging it again moves the label; moving it the left hand side takes the label away. Multiple peaks can be added.
Use display grid off for clearer presentation.
Data can be output in abs orbance only
Scan mode – export to Excel and graphing
If saving as a *.txt file, save the results to folder of choice.
o Use Excel to open this file; with files of type set to “all files” o Note that saving as a *.csv file and double clicking on it will open
Excel directly
Highlight the wavelength and absorbance values and click the graph icon
Select chart type “XY Scatter” and the curved lines (no data points) option
Label the axes etc as required
Double click on the x-axis, select Scale and minimum to 330 and maximum to
800
Set colour scheme to suit your prefere nces
___________________________________________________________________ Issue 01 - 12/2004 Libra S6, English 17
MAINTENANCE
After Sales Support
Support agreements that help you to fulfil the demands of regulatory guidelines concerning GLP/GMP are available.
Calibration, cert ific at io n using filters tracea b le to international standards
Certificated engineers and calibrated test equipment
Approved to ISO 9001 standard
Choice of agreement apart from break down coverage can include
Preventative maintenance
Certification
When using calibration standard filters, insert such that the flat surface is facing away from the spring end of the cell holder
Observe all necessary precautions if dealing with hazardous samples or solvents.
Lamp Replacement
A replacement lamp is available from your supplier using the following part numbers:
Tungsten Lamp, S1000L 80-2115-33 (use only this tungsten lamp as it is supplied with the connection wires; others will not operate correctly in this spectrophotometer)
The design of the lamp area is such that users are able to change their own
lamps. No lamp alignment is necessary as the lamp is pre-aligned.
The lamp becomes hot in use. Ensure it is cool before changing it.
Do not touch the optical surfaces of the lamp with your fingers (use tissue); if
touched, the area should be cleaned with iso-propanol.
Instructions for lamp change are provided with the lamp and overleaf.
___________________________________________________________________ 18 Libra S6, English Issue 01 – 12/2004
To change the lamp, proceed as f ollows:
1. Switch off the instrument, remove the sample from the cell holder and
disconnect the power supply cord
2. Remove the protective layers at the lamp access and plug in points on the
underneath of the instrument
3. Remove the l a mp wires from the gr oove by ge nt l y unc l i pping it
4. Remove the lamp by twisting the lamp assembly anti-clockwise
5. Remove the lamp connection end by gently pulling with your fingers
6. Replace with new lamp using the reverse of these actions
Cleaning and general care of the instrument
External cleaning
Switch off the instrument and disconnect the power cord. Use a soft damp cloth. Clean all external surfaces. A mild liquid detergent may be used to remove stubborn marks.
Changing cell holder or removal for cleaning
Undo the screws that are visible on the top of the cell holder using a small flat
headed screwdriver and lift the holder out by holding onto the projection; this may require pushing to the right as you do so in order to prevent fouling against the left side of the instrument cover.
Insert the test tube holder and secure in place using the same screws.
Note that as well removal for cleaning, spillages in the cell holder ca n be flushed
through using water from a squeeze bottle in order to prevent crystallisation / fermentation of residues.
Changing the brightness of the display
To change display brightness, proceed as follows:
1. Ensure the instrument is on and that there is no sample in the cell holder
2. Remove the protective layer at the lamp plug in point (underneath and at the rear
of instrument)
3. Place the instrument on its back, insert a small flat headed screwdriver into the
potentiometer slot and turn it right or left until a suitable level of brightness is obtained.
___________________________________________________________________ Issue 01 - 12/2004 Libra S6, English 19
SPECIFICATION AND WARRANTY
Wavelength range 330 - 800 nm Monochromator Flat grating Wavelength calibration Automatic upon switch on Spectral bandwidth 7 nm Wavelength a ccuracy
± 2nm
Wavelength rep ro ducibility
± 1nm
Light sources Pulsed Tungsten halogen Detector Diode array Photometric range - 0.300 to 2.500A, 0.3 to 199%T Photomet ric linearity
± 2.0 % or ± 0.010A to 1.000A at 546nm, whichever is the greater
Photometric reproducibility < 0.002 A at 0A and 500nm Stray Light < 1%T 340nm according to ANSI/ASTM E387-72 Stability
± 0.005A/h at 0A and 546nm after warm-up
Noise
± 0.002A near 0A and ± 0.020A near 2A at 600nm
Analogue output
1V per 1 Abs (±10%), 1V = 0A offset 1V per 100%T (±10%), 0V = 0%T offset
Digital output 9 pin serial Dimensions 180 x 270 x 390 mm Weight 1.75 kg Power input 90-265 V, 50/60 Hz, 15 VA
Safety standard EN61010-1 EMC emissions EN 61326-2.3 Generic emissions EMC immunity EN 61000-4-6 Generic immunity part 1 Mains harmonics EN 61000-3-2 Susceptibility standard IEC 801 Quality System Designed and manufactured in accordance with an ISO
9001 approved quality system
British Design Registration No. 2097049
Specifications are measured after the i n strument has warmed up at a constant ambi ent temperature and are typical of a production unit. As part of our policy of continuous development, we reserve the right to alter specifications without noti ce.
Warranty
Your supplier guarantees that the product supplied has been thoroughly tested to ensure that it meets its published specificati on. The warranty included in the conditions of supply is valid for 12 months only if the product has been used according to the instructions supplied. They can accept no l iability for loss or damage, h o wever caus ed , arising from the faulty or in co r rect use of this product. This product has been designed and manufactured by Biochrom Ltd, 22 Cambridge Science Park, Milton Road, Cambridge CB4 0FJ, UK.
Loading...