Berthold MicroPolar LB 566 User Manual

Concentration Meters MicroPolar (++)
LB 566
User’s Guide Hardware Manual 41986BA2
Rev. Nr.: 06-E, 09/2017
MicroPolar (++) LB 566
The units supplied should not be repaired by anyone other than Berthold Technologies Service engineers or technicians by Berthold Technologies.
In case of operation trouble, please address to our central service department (address see below).
The complete user’s guide consists of the hardware manual and the software manual.
The hardware manual comprises:
mechanical components
installation
electrical installation
technical data
electrical and mechanical drawings
The software manual comprises:
operation of the evaluation unit
parameter description
basic setup
calibration
error messages
The present manual is the hardware description.
Subject to change without prior notice.
BERTHOLD TECHNOLOGIES GmbH & Co. KG Calmbacher Str. 22 D-75323 Bad Wildbad Switchboard: Service:
Phone +49 7081 177 0 Phone +49 7081 177 111 Fax +49 7081 177 100 Fax +49 7081 177 339 industry@Berthold.com Service@Berthold.com www.Berthold.com
Table of Contents
MicroPolar (++) LB 566 5
Table of Contents
Page
Safety Summary 7Chapter 1.
1.1 Symbols and Warnings 7
1.2 General Information 8
1.3 General Safety Instructions 9
General Information 11Chapter 2.
2.1 Use and Function 11
2.2 Frequency License 12
2.3 Intended Use 14
2.4 Explanation of Terms 15
System Description 17Chapter 3.
3.1 Principle of Measurement 17
3.2 Calculation of Measured Values 18
3.3 Temperature Compensation 19
3.4 Throughput Calculation and Output 20
3.5 Mechanical Components 21
3.5.1 Evaluation Units 23
3.5.2 FlowCell 25
3.5.4 Container Probe 27
3.5.5 High-frequency Cable 30
3.6 Assembly on a Pipeline 32
3.7 Assembly on a Container 33
Getting Started 35Chapter 4.
4.1 Transport 35
4.2 Commissioning the FlowCell 36
4.2.1 Installing the FlowCell 36
4.2.2 Installing the Evaluation Unit 38
4.2.3 Connecting the HF Cable 39
4.3 Commissioning the Container Probe 41
4.3.1 Installing the Container Probe 41
4.3.2 Setting Up the Evaluation Unit 43
4.3.3 Connecting the HF Cable 43
4.5 Connecting the Evaluation Unit 45
4.5.1 Pin Configuration of the Connector Strip 46
4.5.2 Digital Outputs, Relay 48
Service instructions 51Chapter 5.
5.1 General Information 51
5.2 Wear Parts 52
5.3 Instrument Cleaning 53
5.4 Battery 53
5.5 Fuse Replacement 54
Table of Contents
6 MicroPolar (++) LB 566
Technical Data 55Chapter 6.
6.1 Technical Data Evaluation Unit 55
6.2 Technical Data Sensors 58
6.3 Technical Data HF Cable 61
6.4 Format of Serial Data Output RS232 and RS485 62
Certificates 63Chapter 7.
7.1 EC Declaration of Conformity 63
7.2 Frequency License 65
Technical Drawings 73Chapter 8.
8.1 Dimensions Drawings Evaluation Unit Wall Housing 73
8.2 Electrical Wiring Diagram 75
8.3 Dimensional Drawings FlowCell 76
8.3.1 Type LB 5660-102-00X FlowCell DN 50 VFL. FOA 76
8.3.2 Type LB 5660-112-00X FlowCell DN 50 G-BS/M 77
8.4 Dimensional Drawings Container Probes 78
8.4.1 Type LB 5650-01 78
8.4.2 Type LB 5650-02 79
8.4.3 Type LB 5650-03 80
8.4.4 Type LB 5650-04 81
8.4.5 Type LB 5650-05 82
8.4.6 Type LB 5650-09 83
8.4.7 Installation Situation in Pipelines 84
8.5 Dimensional Drawings Container Flush Probes 85
8.5.1 Type LB 5651-01 85
8.5.2 Type LB 5651-02 86
8.5.3 Type LB 5651-03 87
8.5.6 Installation Situation in Pipelines 88
Chapter 1. Safety Summary
MicroPolar (++) LB 566 7
Safety SummaryChapter 1.
1.1 Symbols and Warnings
In this user manual, the term Berthold Technologies stands for the company Berthold Technologies GmbH & Co.KG.
To rule out bodily injury and property damage, please keep in mind the warning and safety instructions provided in this user manual. They are identified by the following sings: DANGER, WARNING, CAUTION or NOTICE.
Indicates imminent danger. If it cannot be avoided, death or most severe personal injuries may be the consequence.
Indicates a possibly dangerous situation. The consequences may be death or most severe personal injuries.
Indicates a possibly harmful situation The consequences may be minor or medium personal injuries.
Indicates a situation that may cause material damage if the instructions are not followed.
IMPORTANT
Paragraphs with this symbol provide important information on the product and how to handle it.
TIP
Contains user tips and other useful information.
Chapter 1. Safety Summary
8 MicroPolar (++) LB 566
Other symbols used in this documentation:
Warning: No intervention, do not alter anything
Requirement: Disconnect power
Requirement: Wear safety boots
1.2 General Information
The most important safety measures a summarized in this user manual. They supplement the corresponding regulations which must be studied by the personnel in charge.
Please pay attention to:
the national safety and accident prevention regulations
the national assembly and installation directions
the generally recognized engineering rules
the information on transport, assembly, operation, service,
maintenance
the safety instructions and information in these operating in-
structions
the enclosed technical drawings and wiring diagrams
the characteristic values, limit values and the information on
operating and ambient conditions on the type labels and in the data sheets
the signs on the devices
the country-specific licensing schemes
Chapter 1. Safety Summary
MicroPolar (++) LB 566 9
1.3 General Safety Instructions
IMPORTANT
The equipment housings have IP 65 protection and are suitable for outdoor applications. The units are factory tested and are delivered in a condition that permits safe and reliable operation.
For outdoor applications, the measuring systems must be protected from direct sunlight and rain, for example by a suitable shelter.
IMPORTANT
Never change the installation and the parameter settings without a full knowledge of these operating instructions, as well as a full knowledge of the behavior of the connected controller and the possible impact on the operating process to be controlled.
The systems may only be used in perfect technical condition and only for the intended use!
Only let persons work with the systems who are mandated to do this and are suitably qualified and adequately trained! Installations and modifications on the systems which may affect the operational safety are not permitted!
IMPORTANT
All system components require non-corrosive ambient conditions during transport, storage and operation.
IMPORTANT
If liquid gets inside the instrument, cut off the power supply. The equipment must be inspected and cleaned by an author­ized service center.
Ambient conditions
Chapter 1. Safety Summary
10 MicroPolar (++) LB 566
Electrical hazards
Disconnect power to ensure that contact with live part is avoided during installation and when servicing.
Disconnect the power supply before opening the instrument. Work on open and live instruments is prohibited.
Caution! Potential hazards, material damage! Device type: LB 566-12 MicroPolar ++ (ID no. 51833-02)
When connecting the 24 V DC power supply, the + and – poles must be connected correctly. There is no reverse polarity protection!
Spare fuses must match the rating specified by the device manufacturer. Short-circuiting or manipulation is not permit­ted.
IMPORTANT
The LB 566 and all additional equipment must be connected to mains via grounded connection.
IMPORTANT
The concentration meter LB 566 may be installed, serviced and repaired only by qualified specialists.
Qualified persons Qualified specialists are persons who through professional
training have acquired sufficient skills in the respective field and who are familiar with the relevant national industrial safety reg­ulations, accident prevention directions, guidelines and accepted engineering standards. They must be able to safely assess the results of their work and they must be familiar with the contents of these user manual.
Chapter 2. General Information
MicroPolar (++) LB 566 11
General InformationChapter 2.
2.1 Use and Function
The MicroPolar LB 566 has been designed as a concentration measuring system and may be used only for this purpose. If the devices are used in a manner that are not described in this user manual, the protection of the devices is compromised and the warranty will be lost.
Berthold Technologies is liable and guarantees only that the de­vices comply with its published specifications. The LB 566 may only be installed in an undamaged, dry and clean condition. Al­terations and modifications to the system components are not permitted.
The LB 566 is not qualified as a "safety-related measurement".
The standards and guidelines the LB 566 complies with are itemized in these device instructions in chapter 2.2 Frequency License and chapter 7.1 EC Declaration of Conformity.
The protection type of the LB 566 to IEC 60529 is max. IP 65.
The following use is inappropriate and has to be prevented:
The use under conditions other than the terms and condi-
tions stated by the manufacturer in their technical documen­tation, data sheets, operating and installation manuals and other specifications.
The use after repair by persons not authorized by Berthold
Technologies.
The use in a damaged or corroded state.
Operation with open or inadequately closed cover.
Operating with insufficiently tightened adapters and cable
glands.
Operation without the manufacturer’s recommended safety
precautions.
Manipulating or bypassing existing safety facilities.
Authorized persons are those who, by law, are permitted to per­form the respective activity, or who have been approved by Berthold Technologies for certain activities.
Conformity to standards
Protection type
Warning against misuse
Authorized persons
Chapter 2. General Information
12 MicroPolar (++) LB 566
2.2 Frequency License
This device complies with Part 15 of the FCC Rules and with In­dustry Canada licence-exempt RSS standard(s).
Operation is subject to the following two conditions:
(1) this device may not cause harmful interference, and
(2) this device must accept any interference received,
including interference that may cause undesired operation.
Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux appareils radio exempts de licence.
L'exploitation est autorisée aux deux conditions suivantes:
(1) l'appareil ne doit pas produire de brouillage, et
(2) l'appareil doit accepter tout brouillage radioélectrique
subi, même si le brouillage est susceptible d'en compromettre le fonctionnement.
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protec­tion against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the in­structions, may cause harmful interference to radio communica­tions. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and re-
ceiver.
License labels
Chapter 2. General Information
MicroPolar (++) LB 566 13
- Connect the equipment into an outlet on a circuit differ-
ent from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician
for help.
Changes or modifications made to this equipment not expressly approved by BERTHOLD TECHNOLOGIES may void the FCC au­thorization to operate this equipment.
The MicroPolar and MicroPolar ++ comply with the R&TTE Di­rective 1999/5/EC and thus meet all the requirements for this type of high-frequency devices. As a mark of conformity in ac­cordance with the CE mark, the devices bear the no. 0682 of the certification body. The certificate can be found in chapter
7.3 Frequency License.
IMPORTANT
The frequency approvals and conformities are only applicable in combination with the container probes and the FlowCell.
To comply with ETSI (European Telecommunications Stand­ards Institute), the LB 566 ++ in combination with horn and spiral antennae may only be operated with a 10dB attenuator pad in the sending path (MTX).
IMPORTANT
The LB 566 has been manufactured in compliance with the safety requirements for microwave devices. It will be the user’s responsibility to adhere to any special legal provisions regarding the use of microwaves.
IMPORTANT
Any change in the frequency or otherwise tampering with the microwave device will lead to a loss of the frequency license and may result in criminal consequences. The microwave modules do not include any replaceable components and must not be opened.
Chapter 2. General Information
14 MicroPolar (++) LB 566
2.3 Intended Use
The LB 566 can be used to determine the water or moisture content and the concentration of virtually any material. The fol­lowing sensors and evaluation versions are available:
1. The container probes have been designed for installation in­to containers and pipelines with a nominal width of 200 mm. The probe is mounted so that both measuring rods (transmitter and receiver) are immersed into the product being measured.
2. The FlowCell is a tubular probe, with the microwave trans­mitter and receiver being firmly welded onto the outside of the tube. The tube is lined with Teflon to ensure smooth walls. The FlowCell is either installed directly into the exist­ing pipeline system or into a bypass.
The evaluation unit is available in two versions: The standard model MicroPolar and the high dynamic version MicroPolar ++. The Micro Polar ++ requires a microwave signal attenuation of at least 40 dB. The MicroPolar must be used for lower micro­wave attenuation.
During operation, the concentration meters MicroPolar and Mi­croPolar ++ emit electromagnetic radiation in the frequency range between 2.4 GHz and 2.5 GHz (range limitations depend­ing on local regulations in your country). Microwaves are not dangerous to human beings and the environment (power radia­tion < 10 mW). The microwaves are emitted directed from the microwave window; the product is not altered by the micro­waves.
To ensure proper function of the measuring system, please pay attention to the following:
TIP
The material to be measured may be electrically conduc-
tive only to a limited degree.
The product must not contain any gas bubbles or gas
bubbles have to be compressed with adequate pressure when carrying out measurements in pipelines.
The ion concentration, for example, salt content must be
nearly constant.
Chapter 2. General Information
MicroPolar (++) LB 566 15
2.4 Explanation of Terms
Attenuation Weakening of microwave signals, microwaves measuring effect.
EVU Evaluation Unit
Factory setting In the factory setting all parameters have been set to default
values. In most cases this considerably facilitates the calibra­tion of the instrument. Despite factory setting, calibration always has to be performed.
FlowCell Tubular probe for simple integration into the existing pipeline
system.
Flush probe Container probe with flushing device.
HF cable High frequency cable
Microwaves Designation for electromagnetic waves in a specific frequency
range.
Phase Phase or phase shift, microwave measuring effect.
Quad cable Combination of four HF cables of equal length in a corrugated
tube.
Softkeys Software associated keys.
TC Temperature compensation
Chapter 3. System Description
MicroPolar (++) LB 566 17
System DescriptionChapter 3.
3.1 Principle of Measurement
As the microwaves pass through the product, their propagation velocity is slowed down (= phase shift) and their intensity is at­tenuated (= attenuation). Figure 3-1 illustrates the principle of measurement: Compared to a reference signal, the propagation velocity of microwaves passing through the product is slowed down (phase shift) and their intensity (attenuation) is reduced.
Phase shift
Messgut
Phase comparison
--> Phase Amplitude comp.
--> Attenuation
Referencepath
Referencesignal
Measurementsignal
Measured value: Concentration %DS
HF sources
Receiver
Transmitter
The prerequisite is that the product being measured shows die­lectric properties. Generally, water is a very distinct dielectric fluid. The water or dry mass concentration can therefore be de­termined by measuring the phase shift and/or attenuation.
The concentration to be detected in the product is therefore in good approximation linearly dependent on the phase shift and the attenuation. For this reason, we can measure the concen­tration or the dry matter content of the product using a linear calibration (see chapter 3.2 Calculation of Measured Values).
Figure 3-1:
Schematic diagram:
Change in
microwave by
product
Chapter 3. System Description
18 MicroPolar (++) LB 566
3.2 Calculation of Measured Values
The microwave parameters phase and attenuation are calibrat­ed according to an automatic plausibility analysis.
During calibration, the phase and/or attenuation of a concentra­tion value (or density value) are assigned through sampling. The calibration is done automatically and the sampling process is supported by the evaluation unit.
Which of the parameters (phase, attenuation or both) will be used for the calibration depends on the size and interference of the measuring effect. For example, the attenuation is signifi­cantly more sensitive to electrolytic conductivity (salt content).
In many cases, the pure phase measurement is recommended and the measured value is calculated as follows:
Measured value = A ∙ Phase + C Eq. 3-1
where:
Measured value Concentration / Moisture / Dry matter A, C Coefficients of the respective calibration
function
The LB 566 allows you to calibrate, display and output two con­centrations: Con1 and Con2. You have to enter the calibration coefficients separately for concentration 1 and 2. For more in­formation please refer to the Software Manual.
Limitations
Weakly bound water can be detected depending on the
strength of the binding. Thus, the measuring effect may be dependent on the grain size distribution and the chemical properties of the product being measured, provided this changes the binding of water to the solid matter.
Walls made of plastic, rubber or insulation materials with
fairly low dielectricity hardly affect the measurement and are calibrated at a constant level.
Ice and crystal water cannot be measured because the water
molecules cannot rotate freely (ice and crystal water are dry).
Chapter 3. System Description
MicroPolar (++) LB 566 19
Conductive materials such as graphite or coke cannot be trans­mitted by microwaves. Metal walls can also not be transmitted by microwaves.
Compensation In addition to the water content, the product temperature,
product density and a varying material load (varying microwave irradiation path) may have an influence on the phase and at­tenuation. This influence has to be compensated for during cali­bration.
3.3 Temperature Compensation
Temperature compensation (TC) is required for fluctuating product temperature. It is generally advisable to provide a tem­perature compensation, i.e. a temperature signal (0/4...20 mA or PT100) to be connected to the evaluation unit and, optional­ly, to enable the compensation in the evaluation unit. The eval­uation unit is designed so that the required TC’s can be calcu­lated automatically. The variation in temperature where TC be­comes absolutely essential is dependent on the product and on the water content. In the first approximation, 2° C should be set as fluctuation limit.
For example, if the product temperature is measured via the PT100 input, then Eq. 3-1 is expanded as follows:
Measured value = A · Phase + D · T
meas
+ C Eq. 3-2
where:
Measured value Concentration / Moisture / Dry matter A, D, C Coefficients of the calibration function T
meas
Product temperature
How to work with the temperature compensation is described in detail in the Software Manual.
Chapter 3. System Description
20 MicroPolar (++) LB 566
3.4 Throughput Calculation and Output
For pipeline applications, the LB 566 allows you to calculate the throughput (mass flow) and to output the result via a current output.
The calculation is based on the microwave measured value; if this value correlates with the product density, one can calculate the throughput, if some additional information is available. The additional information needed is: internal pipe diameter/cross section and the product speed. The product speed has to be fed via current input.
For details, please see the Software Manual, chapter 4.2.12
Mass Flow.
Chapter 3. System Description
MicroPolar (++) LB 566 21
3.5 Mechanical Components
The measuring system comprises an evaluation unit, a probe/antenna pair and one set of special high-frequency cables (short HF-cable). The evaluation unit is available in two ver­sions: the standard model MicroPolar LB 566 and the high dy­namics version MicroPolar ++ LB 566 (see Figure 3-4 and Fig­ure 3-5).
Figure 3-4:
Evaluation unit
MicroPolar
LB 566
Figure 3-5:
Evaluation unit
MicroPolar ++
LB 566
Chapter 3. System Description
22 MicroPolar (++) LB 566
The probes/antenna pairs are available in different versions, as pipeline and container probe with or without flushing device (see Figure 3-6 to Figure 3-7).
The FlowCell is available in the nominal pipe sizes 50 -150 mm (50, 65, 80, 100, 125 and 150) The following connections types are available:
Hygiene milk pipe screw connection DIN 11853
V flange EN 1092-1
Figure 3-6:
From left:
Container probe
LB 5650 and
LB 5651 with
flushing device
Figure 3-7:
FlowCell LB 5660-102-00x
nominal width 50 mm
with V flange
Chapter 3. System Description
MicroPolar (++) LB 566 23
3.5.1 Evaluation Units
The evaluation units comprise the evaluation computer and the microwave unit. The microwaves are generated, received and analyzed in the microwave unit. Signal processing and commu­nication take place in the evaluation computer. For simple oper­ation, the measuring system includes a display, 4 softkeys and an alphanumeric keypad. Different functions are assigned to the softkeys on the display.
Differences between MicroPolar ++ und MicroPolar
The MicroPolar ++ evaluation unit has an additional HF amplifier module in comparison to the standard model; therefore, it also has a larger wall housing (dimensions see chapter 6.1 Technical Data Evaluation Unit). Otherwise, the evaluation units differ on­ly in their application.
MicroPolar ++ The high dynamics version MicroPolar ++ permits higher
product attenuations. Larger measuring paths can be irradiated, i.e. flow cells with larger nominal diameters can be used. Which type of evaluation unit is used depends on the product attenua­tion. MicroPolar is used up to an attenuation of 50 dB; MicroPo­lar ++ is used for higher attenuations. The MicroPolar ++ gen­erally requires an attenuation of 40 dB. If this attenuation is not reached, the software displays an error message.
An RS232 interface is included on the bottom side of the in­strument.
and cable feed-through
Lock
Numerical
keypad
Softkey buttons
LED’s
LCD display
RS232 connection
HF connections for
signal cable reference cable
Micro-Polar
Figure 3-8:
Evaluation unit –
front view
Chapter 3. System Description
24 MicroPolar (++) LB 566
Interpretation of LEDs
Five LEDs on the instrument front panel indicate the current de-
vice status.
LED Function Run On: Device in measurement mode
Flashes + ERROR LED off: Device in warning state, on hold, paused or low load state. A display message with error code indicates the cause (see Software Manual, chapter 11. Error
Lists and Device States).
Error On: Device in error state.
A display message with error code indicates the cause (see Software Manual, chapter 11. Error
Lists and Device States).
Canceled after reset or if error has been eliminat­ed.
Signal 1 Display depending on the selected function of
relay 1, possible functions: Error, no product, alarm min., alarm max., measurement stopped, low load
Signal 2 Display depending on the selected function of
relay 2, possible functions: Error, no product, alarm min., alarm max., measurement stopped, low load
Comm Communication active, e.g. via RS232 and RS485
Figure 3-9:
Evaluation unit –
bottom view
Figure 3-10:
LEDs on the
front panel of
the evaluation unit
Run
Error
Signal 1
Signal 2
Comm
Cable feed-through
M 20 and M 16
High-frequency connections
Rs232
9 pole SubD-connector
M-TxM-Rx R-Tx
R-Rx
Chapter 3. System Description
MicroPolar (++) LB 566 25
Terminal block
The electrical connections of the LB 566 are located on a con­nector strip in the wall cabinet. The terminal block can be ac­cessed from the front by opening the cover of the housing. There, you also find the power cut-off switch and the fuses. The high-frequency terminals are located on the outside of the housing. All other elements, especially the voltage-carrying el­ements (on the motherboard) are provided with a protective cover.
3.5.2 FlowCell
The FlowCell is available in the nominal sizes of 50 to 150 mm (see fig. 3-11). As connection, the versions V flange EN 1092-1 or Hygiene milk pipe screw connection DIN 11853 are available. For technical data please see chapter 6.2 Technical Data Sen- sors.
A: with V flange EN 1092-1 B: with Hygiene milk pipe screw connection DIN 11853
Figure 3-11:
FlowCell versions
Chapter 3. System Description
26 MicroPolar (++) LB 566
The flow cell consists of a sturdy stainless steel body. The mi­crowave transmitter and receiver are firmly welded to the out­side of the pipe. The entire product pipe is PTFE-coated and thus meets the special requirements for use in food for use in food.
There are no objects that extend into the pipe (e.g. a measuring sensor). Depending on the version, the FlowCell can be mount­ed with the V flange or milk pipe screw connection to the piping. For the versions with V flange, ASA adapter flanges are availa­ble as accessory.
The FlowCell has two HF ports to feed in and output the micro­wave signals. The input and output can be allocated as needed (M-Tx, M-Rx). The microwave signals transmit the product over the entire pipeline cross-section.
For all versions, the following accessories are available:
1. Pipe-mounted PT100 or Inline PT100
2. Conductivity measuring device
3. Sampling valve
Chapter 3. System Description
MicroPolar (++) LB 566 27
3.5.4 Container Probe
The container probe is available in a version with and without flushing device (see Figure 3-12). The technical data are listed in chapter 6.2 Technical Data Sensors.
A: High-frequency connections B: Process connection, flanges of different sizes
Container probe type LB 5650 and type LB 5651
The container probe has been specially designed for concentra­tion measurements in containers. Both measuring rods are im­mersed into the product. Microwaves are emitted from one end of the rod and received by the other end of the rod; they are emitted only towards the opposite end of the rod. This direc­tional characteristic of the probe minimizes the interfering influ­ence of metal parts in the vicinity of the probe and allows instal­lation if only limited space is available. For example, the con­centration of sugar strike can be measured continuously to find the suitable inoculation time.
The plastic caps of the measuring rods meet the special re­quirements for application in foodstuffs.
Two different probe types are available:
The standard type is the container probe without flushing
device.
The flushing probe is used in processes where incrustations
are likely to occur, for example, due to increased deposi­tions. The flushing device prevents deposition on the micro­wave exit windows.
Figure 3-12:
Container probe
versions
Chapter 3. System Description
28 MicroPolar (++) LB 566
For the case depicted in Figure 3-13, the flow direction of the product being measured must be perpendicular to the drawing plane. This ensures a representative product between the measuring rods, provided the product is mixed thoroughly.
The container probe LB 5650 is the only one equipped with a PT100 connection and can be connected to the evaluation unit via a 4-core cable. The PT100 wiring diagram is described in chapter 4.6.1 Pin Configuration of the Connector Strip. To min­imize the danger of incrustation in the immediate vicinity of the measuring rods, the container flush probe is not provided with a PT100.
Warning: Possible material damage! Do not open the cover screws on the front of the container probes, see Figure 3-14.
Figure 3-13:
Container probe LB 5650
Figure 3-14:
Front view
container probe
PT100
Chapter 3. System Description
MicroPolar (++) LB 566 29
Container probe type LB 5651 with flushing device
The flushing probe LB 5661 has been designed for processes in which depositions, for example, due to incrustations are likely to occur on the probe.
The flushing probe has two flushing channels which keep the measuring rods free from incrustations; this ensures that the microwaves come into direct contact with the product being measured. All probe parts coming into contact with the product meet the special requirements for application in foodstuffs. Fig-
ure 3-15 shows the probe design.
The flushing slit width is the same for both probe rods and is depicted in Figure 3-16.
Figure 3-15:
Flushing probe
LB 5651
Figure 3-16:
Rod head with
flushing pipe
Plastic rods
Flushing pipe
Flushing
s
l
i
t
~
1.5
m
m
Chapter 3. System Description
30 MicroPolar (++) LB 566
3.5.5 High-frequency Cable
High-frequency cables (HF cable) are used to transmit micro­wave signals.
HF cables change their conductivity (for microwaves) with tem­perature and would therefore produce measurement errors with varying ambient temperature. This error is compensated for by enabling the automatic cable compensation. The influences of the ambient temperature on the signal cable are compensated for by means of the reference cable. To this end, the sum of the reference cables has to match the length of the sum of the sig­nal cables.
Two different HF cable types of different lengths are available:
Version 1: The so-called HF-cable quad: It consists of four sin­gle HF-cables of equal length, whose ends are terminated by one HF-connector (N-type). Available cable lengths: 2, 4, 6 and 10 m (see Figure 3-17).
The HF-cable quad is used exclusively for the container probe and the FlowCell. For the FlowCell, the reference line on the probe side is short-circuited using an N-connector (see Figure 3-
18).
Figure 3-17:
HF cable quad
Lengths 2, 4, 6 and 10 m
Corrugated tube
4 x N-connector
ø18.5 mm
0.35 m
0.35 m
4 x N-
4 x N-connector
Chapter 3. System Description
MicroPolar (++) LB 566 31
Figure 3-18:
HF cable quad
probe side:
The ends of the
reference line
R-Rx and R-Tx are
short-circuited
using an
N-connector.
N-connector (ID-no. 20608)
Sensor side
R
-
R
x
R
-
T
x
M
-
R
x
M
-
T
x
Version 2: It consists of a single HF-cable whose ends are ter­minated by an HF-connector (N-type). Available lengths: 2, 2.5, 3, 3.5 and 4 m (see Figure 3-19).
Figure 3-19:
HF cable
Lengths 0.5 to 4 m
ø10 mm
19 mm wrench
For further technical data see chapter 6.5 Technical Data HF- Cable.
During assembly, ensure that the end of the corrugated tube is bent down on the probe side. By doing so, the ingress of fluids into the corrugated tube is prevented.
Chapter 3. System Description
32 MicroPolar (++) LB 566
3.6 Assembly on a Pipeline
The evaluation unit is installed close to the FlowCell to keep the HF cable between evaluation unit and probe as short as possi­ble. The shorter the cable connection, the better the stability of the measurement. The standard length is 2 m and the maxi­mum length of the HF cable is 10 m. The HF cable must be at least 2 m long.
The FlowCell is integrated into the existing pipeline system or in a bypass. The orientation of the FlowCell can either be vertical or horizontal. To avoid possible sedimentary deposits, vertical installation in a riser is preferred (see Figure 3-20).
The installation should preferably be close to a sample sampling point to ensure representative sampling for calibration.
A representative temperature signal (current signal or PT100) should be connected to the evaluation unit for product tempera­ture compensation.
Figure 3-20:
Typical measurement
arrangement on
a pipeline
Chapter 3. System Description
MicroPolar (++) LB 566 33
3.7 Assembly on a Container
The evaluation unit is installed in the immediate vicinity of the container probe to keep the HF-cable between evaluation unit and probe fairly short. The shorter the cable connection, the better the stability of the measurement. The standard length is 2 m and the maximum length of the HF cable is 10 m. The HF cable must be at least 2 m long.
The installation should preferably be close to a sampling point to ensure representative sampling for calibration. A representa­tive temperature signal (current signal or PT100) should be connected to the evaluation unit for a possibly required product temperature compensation.
Our example in Figure 3-21 shows the measurement configura­tion on a process container. The probe is fixed to the container wall using a flange coupling so that both measuring rods are immersed into the product.
Figure 3-21:
Example:
Pipeline measurement
configuration on a
process container
Chapter 4. Getting Started
MicroPolar (++) LB 566 35
Getting StartedChapter 4.
4.1 Transport
Warning: Possible material damage! System parts may get damaged during transportation!
Transport all components in their original packaging. Protect parts against shocks. In particular, the measuring rods of the container probes must be protected against mechanical shock!
After unpacking, make sure all parts listed on the packing list have been delivered and show no sign of damage; if necessary, clean these parts.
If you detect any damage, please notify the forwarder and the manufacturer immediately.
The weight of the system components may exceed 25 kg, depending on the version. We recommend, therefore, that you wear safety boots.
Chapter 4. Getting Started
36 MicroPolar (++) LB 566
4.2 Commissioning the FlowCell
4.2.1 Installing the FlowCell
Note the following points when installing the FlowCell:
The FlowCell is integrated at a suitable location in the pipe-
line system. Keep in mind that material sampling directly behind the FlowCell should be possible for calibration.
The FlowCell should be installed in a vertical riser, if possi-
ble. It must be ensured that material deposits cannot form on the pipe walls and no bubble formation occurs in the product. In case of horizontal installation, please observe the correct orientation of the HF connections, see Figure 4-
2.
There should be a straight pipe section of at least 300 mm
and equal nominal width before and after the FlowCell to en­sure a fairly homogeneous flow profile and to rule out possi­bly occurring microwave reflections in the pipeline. Also, not fittings must be installed in these pipe sections.
No gas inclusions should be present in the product. If gas
bubbles cannot be ruled out, a pressure of at least 4 bar is required in the pipeline to minimize the influence of gas bubbles. Please observe the max. permissible operating pressure, see chapter 6.2 Technical Data Sensors.
The high-frequency cables should preferably be connected to
the FlowCell from below to prevent flowing water from get­ting to the connection sockets.
The signal and reference cable should as far as possible fol-
low the same path to make sure both cables are exposed to the same temperature and should not come into contact with hot pipelines. We recommend installing the HF cable through a single protection tube. If you are working with the HF cable quad, this function is taken over in good approxi­mation by the corrugated tube.
Chapter 4. Getting Started
MicroPolar (++) LB 566 37
Abb. 4-2:
Horizontal installation:
Orientation of
HF connectors
Abb. 4-1:
Installation in a
vertical riser
Chapter 4. Getting Started
38 MicroPolar (++) LB 566
4.2.2 Installing the Evaluation Unit
Note the following points when installing the evaluation unit:
Position the evaluation unit depending on the length of the
HF cable in the vicinity of the microwave probe.
The evaluation unit has to be protected against vibrations.
In some cases, it is advisable to set up the evaluation unit on a stand separated from the pipeline system.
For instrument installation you should foresee a cutoff de-
vice to allow easy and quick disconnection of the device from the power supply.
Provide an automatic separating device (line circuit breaker)
that disconnects the unit from power within 0.03 seconds in case of failure. The separating device must be matched to the cable cross-section of the supply line, but at least it must be designed for 1 A continuous current.
When installing the evaluation unit on a crystallizer, use a
distance rail to minimize the thermal radiation and conduc­tion. See Figure 4-3:
For outdoor applications, the evaluation unit must be pro-
tected from direct sunlight and rain, for example by a suita­ble shelter.
Figure 4-3: View from above: Installation of the
evaluation unit on a
hot container wall
Insulation container wall
Spacer
Evaluation unit
Chapter 4. Getting Started
MicroPolar (++) LB 566 39
4.2.3 Connecting the HF Cable
The FlowCell is connected with the evaluation unit via the HF cables. Two different options are available:
Version 1: 1 x HF cable quad and 1 x N-connector
Version 2: 2 x HF cables (as signal cable)
1 x HF cable (as reference cable)
The decision for a certain cable version is taken by Berthold Technologies in the planning stage. It is subject to the applica­tion and the desired distance between evaluation unit and FlowCell.
Prerequisite for a correct measurement is the correct installation of cables! Please keep in mind:
TIP
Make sure the cable does not get into contact with hot pipes over the entire length (corrugated tube and single cable section after splitting), e.g. direct contact with the device wall (not insulated). This will ensure that all individual cables are exposed to the same ambient conditions and that the com­pensation of the cable drift works properly.
Never bend the HF cable! The bending radius should not be less than 100 mm. Fix the HF cable with cable binders or other suitable means, so that the cable cannot slip anymore!
During assembly, ensure that the end of the corrugated tube is bent down on the probe side. By doing so, the ingress of fluids into the corrugated tube is prevented.
Version 1 The HF cable quad and the HF connectors on the evaluation unit
are labeled. Connect the FlowCell to the evaluation unit as shown in Figure 4-4; only cables with identical labeling can be combined. The two connections on the FlowCell are not labeled, the assignment of both cable connectors M-Tx and M-Rx is arbi­trary. Connect the cable connectors R-Tx and R-Mx with the N­connector (short-circuited).
Version 1 Connect the FlowCell to the evaluation unit as shown in Figure
4-5; make sure that the reference cable (ring line) is connected to R-Tx and R-Mx.
The reference cable must be as long as the sum of both signal cables.
Chapter 4. Getting Started
40 MicroPolar (++) LB 566
Hand tighten all screwed connections of the HF cable (2 Nm =
0.2 kg/m)! Before tightening, carefully screw on the cable by hand. Caution! Threaded joint jams easily.
Check occasionally if the screwed connection is still properly tightened. If the installation is exposed to vibrations, the screwed connection may come loose and this may result in in­accurate measurements or corrosion of the connections.
Abb. 4-4:
FlowCell connection,
version 1
Abb. 4-5:
FlowCell connection,
version 2
Chapter 4. Getting Started
MicroPolar (++) LB 566 41
As long as the cables are not connected, the coaxial sockets have to be covered with plastic caps and the cable connectors have to be protected against moisture and dirt.
4.3 Commissioning the Container Probe
4.3.1 Installing the Container Probe
For installation of the probes, please keep in mind:
Select the installation site such that good mixing and a ho-
mogeneous product are ensured and no bubbles are present in the probe. A sample tap should be available in the direct vicinity to allow representative sampling.
The probe must be flanged to the container so that the ma-
terial to be measured flows between the two measuring rods. That means the fork (both measuring rods) must be installed perpendicular to the material flow.
The distance between the measuring rod tips and any metal-
ized walls (heating element, stirrer, container wall) should be at least 60 mm.
The following mounting hole dimensions in the fitting
flange are required for installation of the probe:
Flange Minimum mounting hole size (mm) DN 65 / PN 6 102 0.3 others 104 0.5
For further installation dimensions, such as the required in-
stallation depth, please refer to chapter 8.6 and 8.7 Assem- bly Sheets.
Use the appropriate flat gasket (standard accessory) to
compensate for minor surface tolerances in the fitting flange.
Installation on process containers
Figure 3-26 shows the position of the container probe on the container. This position is also valid for the container flush probe. Follow the instructions in chapter 3.9 Assembly on a
Container.
The assembly sheets in chapters 8.6 and 8.7 include all the in­formation required for installation.
Depending on the process, the probe flange may stick to the process flange. In such case, during the dismounting process,
Dismounting in the event of a stuck flange
Chapter 4. Getting Started
42 MicroPolar (++) LB 566
remove the two eyebolts first. Then insert two longer screws (M8) into their bore holes up to the process flange. If you con­tinue to rotate the screws, you will push the probe away from the process flange.
Installation in pipelines
Container probes can be installed in pipelines with a nominal width 200 mm using an adapter flange. Note the position and orientation of the container probe, see the technical drawings in
chapter 8.4.7 and 8.5.6 Installation Situation in Pipelines.
Connecting the flushing pipes
The container probe with flushing device has two flushing con­nections with G3/8'' female thread (DIN ISO 228-1). The flush­ing connections should be sealed only at this thread. Sealing the probe cover, for example with silicone, is not permitted.
Flushing parameters (only for flushing probe)
The degree of deposition or incrustation is essential for the flush settings, i.e. flush interval and flush duration. The flushing pa­rameters must be matched to the product and the process.
The following product and process independent flushing param­eters must be strictly adhered to:
Flush solution Water, condensate Temperature of flush solution
Maximum 120°C
Pressure
3 bar, max. 8 bar Fittings 2 x G3/8'' female thread (DIN ISO 228-1) Supply pipe
1/2 inch
In general: The flushing connections can be flushed simultane­ously or sequentially. The flushing parameters are valid for each flushing connection.
TIP
The required flush duration has to take into account possible inertias of the system, e.g. valve openings. The flush supply pipes must be well insulated to prevent that the flush solution is initially colder.
Amounts of water With a 5 bar flushing pressure, the amount of water per flushing
connector is approximately 0.8 liters per second.
Chapter 4. Getting Started
MicroPolar (++) LB 566 43
4.3.2 Setting Up the Evaluation Unit
Set up the evaluation unit as described in chapter 4.2.2.
4.3.3 Connecting the HF Cable
Connect the container probe to the evaluation unit via the HF cable quad.
Prerequisite for a correct measurement is the correct installation of cables! Please keep in mind:
TIP
Make sure the cable does not get into contact with hot pipes over the entire length (corrugated tube and single cable section after splitting), e.g. direct contact with the device wall (not insulated). This will ensure that all individual cables are exposed to the same ambient conditions and that the com­pensation of the cable drift works properly.
Never bend the HF cable! The bending radius should not be less than 100 mm. Fix the HF cable with cable binders or other suitable means, so that the cable cannot slip anymore!
The HF cables and the HF connections on the evaluation unit and on the probe are labeled. Connect the probe to the evalua­tion unit as shown in Figure 4-6; only cables with identical la­beling can be combined.
Chapter 4. Getting Started
44 MicroPolar (++) LB 566
Hand tighten all screwed connections of the HF cable (2 Nm =
0.2 kg/m)! Before tightening, carefully screw on the cable by hand. Caution! Threaded joint jams easily.
Check occasionally if the screwed connection is still properly tightened. If the installation is exposed to vibrations, the screwed connection may come loose and this may result in in­accurate measurements or corrosion of the connections.
As long as the cables are not connected, the coaxial sockets have to be covered immediately with plastic caps and the cable connectors have to be protected by suitable provisions against moisture and dirt.
Figure 4-6:
Connecting the
container probe
to the
evaluation unit
Chapter 4. Getting Started
MicroPolar (++) LB 566 45
4.5 Connecting the Evaluation Unit
Electrical hazards:
Disconnect power to rule out any contact with live parts during installation and servicing. Turn off power before opening the instrument. NEVER work on open and live instruments.
Caution! Potential hazards, material damage! Device type: LB 566-12 MicroPolar ++ (ID no. 51833-02)
When connecting the 24 V DC power supply, the + and – poles must be connected correctly. There is no reverse polarity protection!
The line cross-section the for power supply must be at least 1.0 mm2.
Connect all desired input and output signals to the terminal
strip as shown on the following pages. Use the M feed­through to keep the degree of protection.
Check if the voltage indicated on the type plate matches
your local supply voltage.
Connect the deenergized power cable to the terminals 3(L1),
2(N) and 1(PE).
Verify that the test switch (power interruption) is in position
"ON" (see Figure 5-1).
Close the instrument housing and turn on the power supply.
Chapter 4. Getting Started
46 MicroPolar (++) LB 566
4.5.1 Pin Configuration of the Connector Strip
The connector strip of the evaluation unit includes the following terminals:
Power supply: Terminals 3 (L1, +), 2 (N, -) and 1 (PE, )
For MicroPolar, depending on device type, see name plate on the housing outer wall.
1.) 100…240 V AC, 45…65 Hz
2.) 24 V DC: 18…36 V
24 V AC: -20%, +5%, 40…440 Hz
For MicroPolar ++, depending on device type, see name plate on the housing outer wall.
1.) 100…240 V AC, 45…65 Hz
2.) 24 V DC: 18…36 V, no reverse polarity protection
Current input no. 1 (terminals 20+ and 8-), insulated Current input no. 2 (terminals 22+ and 10-), not insulat­ed
Input as 0/4 - 20 mA signal. For example, for temperature compensation or reference signal recording.
Current output no. 1 (terminals 27+ and 15-), insulated
Output as 4 - 20 mA signal. Output options: Concentrations (1 /
2), current input signals (1 / 2), PT100 signal, mass flow
Current output no. 2 (terminals 19+ and 7-), insulated Output as 0/4 - 20 mA signal. Output options same as for cur-
rent input no. 1.
Figure 4-7:
LB 566 wiring diagram
Chapter 4. Getting Started
MicroPolar (++) LB 566 47
PT100 (terminals 23+ and 11-)
Connection for temperature measurement.
When commissioning the container probe, connect the 4-core cable of the PT100 to the connector strip of the evaluation (see fig. 4-8). Connect the other end of the cable with the corre­sponding plug on the container probe (see fig. 4-6).
In the delivered state, the PT100 cable has a length of 10 meters. We recommend shortening the cable to the required length. Combine the two core pairs in accordance with figure 4-8.
After connection of the PT100 cable, the included ferrite core must be wound around the cable. The applicable position is within the housing, as close as possible to the cable bushing.
Digital input 1: DI1 (terminals 24+ and 12-)
Configuration options:
No function
Measurement: Start (closed) and stop (open)
Digital input 2: DI2 (terminals 25+ and 13-)
Configuration options:
No function
Average value: hold (closed) and continue averaging (open)
Product selection: product 1 (open) and product 2 (closed)
Figure 4-8:
PT100 connection
container probe
Figure 4-9:
Ferrite core
Chapter 4. Getting Started
48 MicroPolar (++) LB 566
Digital input 3: DI3 (terminals 26+ and 14-)
Configuration options:
No function
Start sampling, open: no action, closed: unique measure-
ment starts
Product selection
Relay 1: (Terminals 4, 5 and 6) and Relay 2: (Terminals 16, 17 and 18)
Changeover contacts (SPDT), insulated, configuration option:
No function
Error message
Stop measurement
Limit value min. and max.
No product
Below load limit
RS485 interface (terminals 21 (RS1) and 9 (RS2)) and RS232 interface (on instrument underside)
Serial data interface for output of the live data (all readings for every sweep (measuring cycle), the protocol and data logs. Data format: Data transfer rate 38400 baud, 8 data bits, 1 stop bit, no parity, no handshake
4.5.2 Digital Outputs, Relay
The status of the measurement is output via two relays:
Error
Alarm (alarm min. and max.)
No product
On the Plausibility menu, you can enter a min. attenuation for pause detection (e.g. for process pause, no product present); if this value is not reached, „No product“ is signaled via a relay and the current output drops to 0 or 4 mA.
Measurement stopped
Below load
The respective switching state is also signaled via LEDs on the front panel (LEDs signal 1 and 2).
Chapter 4. Getting Started
MicroPolar (++) LB 566 49
Relay no. Error, alarm, no product,
measurement stopped, Below load, currentless condition
Normal
1
2
The relays with changeover contacts can either be operated as make contact, terminals 4 & 5 (open at error, alarm ...) or as break contact, terminals 5 & 6 (closed at error, alarm ...).
com
456
com
161718
com465com1618
17
Chapter 5. Service instructions
MicroPolar (++) LB 566 51
Service instructionsChapter 5.
5.1 General Information
A malfunction of the measuring system is not always due to a defect in the instrument. Often the error is caused by incorrect operation, improper installation or irregularities in the product being measured. If a malfunction occurs, anyway, the measur­ing system helps you to identify and eliminate errors by display­ing error messages on the display, indicating operator errors and defects of the electronics.
Defective modules of the evaluation unit cannot usually be re­paired but must be replaced. The microwave module is firmly bolted to a screening hood and must not be opened.
For device disposal, please contact the Berthold Service and ap­ply for a recycling passport.
Chapter 5. Service instructions
52 MicroPolar (++) LB 566
5.2 Wear Parts
The evaluation unit does not include any parts that are subject to wear or components that require special maintenance.
The plastic caps of the measuring rods of the container probes and the PEEK Microwave window of the flow cell may be subject to abrasion over the course of time. A low to medium abrasive influences the measurement only very little and can be compen­sated for by a calibration. Therefore, check the wear parts in in­tervals of about 2 years. In case of heavy wear, the plastic caps of the measuring rods of the container probe and the PEEK Mi­crowave window of the flow cells can be replaced.
The plastic caps of the measuring rods can be replaced on site. For this purpose, please proceed as follows:
1
Unscrew the two plastic caps from the measuring rods and remove the four sealing rings (see figure 5-1, yellow arrow).
2
Clean the four grooves of the sealing rings with a lint-free (if necessary wet) cloth.
3
For each measuring rod, insert two new sealing rings into the grooves.
4
Screw on the two new plastic caps.
Caps set for container probe
Id.-Nr. 66049-S 2 pieces of PEEK plastic caps with 4 sealing rings
Figure 4-7:
PT100 connection
container probe
Chapter 5. Service instructions
MicroPolar (++) LB 566 53
5.3 Instrument Cleaning
Clean all system components exclusively with a damp cloth with no chemical cleaning agent. Parts coming into contact with the product (during regular operation) can be cleaned with warm water, taking into account the temperature limits, see chapter
6.2 Technical Data Sensors.
5.4 Battery
If the measuring system LB 566 is a long time without power supply (power failure or disconnected from the mains supply), the system clock is powered by the lithium battery on the motherboard. If the battery voltage is no longer sufficient, the error message CODE 14 "Battery voltage" appears after a restart of the eval­uation unit. After acknowledging the error message, the unit continues to work properly; however, the date and time should be checked and corrected, if necessary. Measurement data that are output via a serial interface can be fatally damaged by in­correct date and time information. We recommend changing the batteries immediately. The service life of the battery, even under continuous stress, is approximately 8 years. Replacement of batteries must be car­ried out in a device disconnected from mains. Battery type: 3 Volt lithium cell (button cell), type CR2032
Chapter 5. Service instructions
54 MicroPolar (++) LB 566
5.5 Fuse Replacement
The mains fuses of the LB 566 are located in the wall housing. Replace the fuses only if the instrument is disconnected from mains. Use only fuses with the correct rating, see chapter 6.1
Spare fuses must match the rating specified by the device manufacturer. Short-circuiting or manipulation is not permit­ted.
Figure 5-2:
View with open
housing wall
MicroPolar
0
I
Netzteil
Fuses
Line connector
Terminalstrip
(can be pulled off)
Battery
EEprom (can be pulled off)
Feed-through for
lineconnector
Testswitch Motherboard
Protectivecover
Chapter 6. Technical Data
MicroPolar (++) LB 566 55
Technical DataChapter 6.
General specifications
Method Microwave transmission measurement
Operating frequen-cy2.4 - 2.5 GHz (ISM band), depending on
local regulations
Transmission power
MicroPolar: < 0.1 mW (< - 10 dB) MicroPolar ++: < 10 mW (< 10 dBm) All coaxial line power
Application Concentration / moisture measurement in
containers and pipelines
6.1 Technical Data Evaluation Unit
Evaluation unit
Housing Wall housing made of stainless steel, see
dimensional drawing in chapter 8 MicroPolar:
HxWxD: 300 x 323 x 140 mm MicroPolar++:
HxWxD: 400 x 338 x 170 mm
Protection type IP 65
Weight MicroPolar: approx. 6.5 kg
MicroPolar ++: approx. 8.0 kg
Ambient conditions during operation
-20 ... +60 °C ( 253 K ...333 K ), no condensation
Relative humidity: max. 80 % Altitude: max. 2000 m
Ambient conditions during storage
-20 ... +70 °C ( 253 K ...343 K ), no condensation
Relative humidity: max. 80 % Achievable accuracy
0.2 weight % (standard deviation)
depending on product and sensor
Display Dot matrix LC display, 114 mm x 64 mm,
240 x 128 pixels, with back-lighting,
automatic contrast setting
Keyboard Freely accessible foil keypad, light-stable
and weatherproof: alphanumeric key-
board and 4 softkeys (software-assigned
buttons)
Chapter 6. Technical Data
56 MicroPolar (++) LB 566
Power supply For Micro Polar, depending on device type:
1.) 100 ... 240 V AC, 45 ... 65 Hz
2.) 24 V DC: 18...36 V; 24 V AC: -20%, +5%, 40 ... 440 Hz
For MicroPolar++, depending on device type:
1.) 100 ...240 V AC, 45...65 Hz
2.) 24 V DC: 18...36
V, no reverse polarity
protection
Power consumption For Micro Polar:
max. 30 VA (AC/DC), depending on configuration
For MicroPolar ++: max. (48/60) VA (AC/DC), depending on configuration
Fuses For Micro Polar:
2 x 2.0 A / slow-blow For MicroPolar ++: 2 x 2.0 A / slow-blow at 100 … 240 V AC or
2 x 6.3 A / slow-blow at 24 V DC
Battery type 3 V Lithium button cell, type CR2032
Measured value e.g. concentration, dry matter content
Inputs and outputs
Cable cross-section min. 1.0 mm² (mains supply)
Cable feed-through 2 x M20x1.5 for cable 5...14 mm (depending
on application) 4 x M16x1.5 for cable 5 ...8 mm (depending on application)
Sensor connection Inputs and outputs for signal and reference
channel, 50 N-socket
HF cable HF cable
Cable lengths: 2, 4, 6 and 10 m; 50 ; both sides with 4 N connectors
Current input 2 x current input 0/4 ...20 mA, ohmic
resistance 50 , 1x insulated, 1x instrument ground e.g. for temperature compensation
Current output Current output 1: 4...20 mA, ohmic re-
sistance max. 800 , insulated Current output 2: 0/4...20 mA, ohmic resistance max. 800 , insulated
e.g. for measured value or temperature output
PT100 connection Measuring range: -50 ... +200 °C (223 ...
473 K); measurement tolerance: < 0.4 °C
Chapter 6. Technical Data
MicroPolar (++) LB 566 57
Digital input 3 x digital inputs (DI1..3), for floating
connectors Configuration options:
DI1: none, measurement start/stop
DI2: none, measurement hold, product
selection DI3: none, sampling, product selection
Function description:
1. Measurement (Start/Stop), open: Meas­urement stopped, closed:
Measurement
started and/or measurement running
2. Hold measurement, open: measurement running, closed: measurement stopped, i.e. average values and current output are held
3. Product selection via a DI:
open: Product 1 (P1), closed: P2
Product selection via two DI’s: DI2 & DI3 open: P1 DI2 closed & DI3 open: P2 DI2 open & DI3 closed: P3 DI2 & DI3 closed: P4
4. Start sampling: open: no actions, closed: single measurement starts
Relay outputs 2 x relays (SPDT), insulated
Configuration options:
- Collective failure message
- Stop measurement
- Limit value (min. and max.)
- No product
- Low load
Load capacity: AC: max. 400VA
DC: max. 90W
AC/DC: max. 250V, max. 2A, non-inductive 150V: voltage must be grounded
The cable used at the relay output must
correspond to a mains cable.
Restrictions at 24 V AC/DC (DC: 18 ...36 V; AC: 24 V +5 %, -20 %) mains supply, if the ground conductor is not connected to terminal 1 (PE):
AC: max. 50 V DC: max. 70 V
Serial interfaces RS232 on the bottom side
RS485 via terminal strip
Data format: 38400 Bd, no handshake, 8 data
bits, 1 stop bit, no parity
Chapter 6. Technical Data
58 MicroPolar (++) LB 566
6.2 Technical Data Sensors
Flow cells
Application Microwave FlowCell with various nominal diame-
ters and flanges for measurement on pipelines
Material Stainless steel, PEEK microwave window
Process coupling
Two versions:
1. Hygiene milk pipe screw connection DIN 11853
2. Flange according to EN 1092-1 (V flange)
Optional adapter for the V flange version with clamp nozzles and ASA flange
Process pressure up to 16 bar (relative), depending on nominal
diameter and flange type, see table below
Temperature range Product temperature: 10...130 °C (283...403 K)
Ambient temperature: -20...60 °C (253...333 K) Storage temperature: 10...80 °C (283...353 K)
Connections 2 x HF connections: N female, 50
for HF cable with max. 10 m length
Versions Nominal pipe widths from 50 ... 150 mm
Dimensions See dimensional drawings in chapter 8.
Overview FlowCells with V flange
Designation ID no. Nominal
width [mm]
Flange Pressure
[bar]
LB 5660-102-00x 66744-001 50 DN 50 / PN 16 16
Further versions on request
Overview FlowCells with Hygiene milk pipe screw connection
Designation ID no. Nominal width [mm] Pressure
[bar]
LB 5660-112-00x 66744-002 50 16
Further versions on request
Chapter 6. Technical Data
MicroPolar (++) LB 566 59
Overview ASA flange adapter
Designation
ID no.
ASA flange adapter set for Flow Cell 50 62324
ASA flange adapter set for Flow Cell 65 62319
ASA flange adapter set for Flow Cell 80 62328
ASA flange adapter set for Flow Cell 100 62331
ASA flange adapter set for Flow Cell 150 62335
The kit consists of two adapters, screws and two seals.
Chapter 6. Technical Data
60 MicroPolar (++) LB 566
Container Probes
Application Container probes with and without flushing device
for concentration measurement in process containers and pipelines with nominal width 200 mm.
Material Plastic caps, stainless steel
PT100 connection cable: Silicon / Teflon
Process coupling Flange according to DIN EN 1092 type 05
DN65 / PN6, DN 80, 100, 150 / PN16; ASA flange 2.5’’ / 150 PSI (More on request)
Process pressure Up to 16 bar, depending on model
Temperature range Product temperature: 10...120 °C (283...393 K)
Ambient temperature: -20...60 °C (253...333 K) Storage temperature: 10...80 °C (283...353 K)
Connections 4 x HF connections: N female, 50
for HF cable with max. 10 m length
Dimensions See dimensional drawings in chapter 8.
Accessory sealing washer
Material Klingersil C-4400
Thickness 3 mm
Caps set for container probe
Id.-Nr. 66049-S 2 pieces of PEEK plastic caps with 4 sealing rings
Overview container probes and sealing washers
Designation ID no. Flange
Pres­sure [bar]
ID no. Seals
LB 5650-01 65464-01 DN 65 / PN6 6 32175
LB 5650-02 65464-02 DN 80 / PN16 16 33717
LB 5650-03 65464-03 DN 100 / PN16 16 46661
LB 5650-04 65464-04 DN 150 / PN16 16 46664
LB 5650-05 65464-05 ASA 2.5’’ / 150 PSI 16 46665
LB 5650-09 65464-09 ASA 3’’ / 150 PSI 16 50659
LB 5651-01 65937-01 DN 65 / PN6 6 32175
LB 5651-02 65937-02 DN 80 / PN16 16 33717
LB 5651-03 65937-03 DN 100 / PN16 16 46661
LB 5651-04 65937-04 DN 150 / PN16 16 46664
LB 5651-05 65937-05 ASA 2.5’’ / 150 PSI 16 46665
Chapter 6. Technical Data
MicroPolar (++) LB 566 61
6.3 Technical Data HF Cable
HF cable Quad
Material Corrugated tube: Polyamide (PA6)
Cable sheath: Polyethylene (PE)
Protection type IP 66
Temperature
In operation: -30 ... +70 °C When installing: -20 ... +70 °C
Cable length [m] ID no.
2 43431
4 43432
6 43433
8 43434
10 43435
HF cable Quad (solid cable)
Material Cable sheath: Polyethylene (PE)
Protection type IP 68 when unscrewed
Temperature
In operation: -40 ... +85 °C When installing: -40 ... +85 °C
Attenuation coefficient
about 0.3 dB/m
Cable length [m] ID no.
2.0 11476
2.5 11477
3.0 11478
3.5 11479
4.0 11480
Chapter 6. Technical Data
62 MicroPolar (++) LB 566
6.4 Format of Serial Data Output RS232 and RS485
Headline
Date·TimeStateStatusSynchronizerProductAttPhiR2TintIN1IN2PT100
CCmC2C2mMF1MF2
Following lines
01.01.2005·00:00:0000000010.435.300.070.00.00.00.0
1 2 3 4 5 6 7 8 9 10 11 12
75.3675.000.000.000.0000.000¶
13 14 15 16 17 18
Column no. Description Format 1 Date and time DD.MM.YY·HH:MM:SS 2 State 4 digits, HEX 3 Status: Information about the quality of the last
measurement
0 : Measurement OK < 0 : Error
4 Product synchronization 5: not active
1: still asynchronous 0: all values synchronous
-1: Error
-2: Time too short for syn.
-3: Speed outside range 5 Product number X (1 to 4) 6 Attenuation [dB] X.XX 7 Phase [°/GHz] X.XX 8 Dispersion of the phase regression X.XX 9 Device temperature [temperature unit] X.X 10 Current input 1 [unit of current input] X.X 11 Current input 2 [unit of current input] X.X 12 PT100 temperature [temperature unit] X.X
[…] with selection of the unit g/cm
3
13 Concentration 1 live X.XX [X.XXXX] 14 Concentration 1 averaged X.XX [X.XXXX] 15 Concentration 2 live X.XX [X.XXXX] 16 Concentration 2 averaged X.XX [X.XXXX] 17 Mass flow for concentration 1 X.XXX 18 Mass flow for concentration 2 X.XXX
Special characters
” Tabulation“¶” Carriage return + Line feed “·” Blank character
Chapter 7. Certificates
MicroPolar (++) LB 566 63
CertificatesChapter 7.
7.1 EC Declaration of Conformity
Chapter 7. Certificates
64 MicroPolar (++) LB 566
Chapter 7. Certificates
MicroPolar (++) LB 566 65
7.2 Frequency License
Chapter 7. Certificates
66 MicroPolar (++) LB 566
Chapter 7. Certificates
MicroPolar (++) LB 566 67
Chapter 7. Certificates
68 MicroPolar (++) LB 566
Chapter 7. Certificates
MicroPolar (++) LB 566 69
Chapter 7. Certificates
70 MicroPolar (++) LB 566
Chapter 7. Certificates
MicroPolar (++) LB 566 71
Chapter 7. Certificates
72 MicroPolar (++) LB 566
Chapter 8. Technical Drawings
MicroPolar (++) LB 566 73
Technical DrawingsChapter 8.
8.1 Dimensions Drawings Evaluation Unit Wall Housing
Chapter 8. Technical Drawings
74 MicroPolar (++) LB 566
Chapter 8. Technical Drawings
MicroPolar (++) LB 566 75
8.2 Electrical Wiring Diagram
Line in for MicroPolar:
1. / 2. depending on configuration
1. AC 100-240V, 45-65 Hz
2. DC 24 V (18-36 V) or AC 24 V -20/+5%, 40-440 Hz
-------------------------------------------
Line in for MicroPolar ++:
1. / 2. depending on configuration
Chapter 8. Technical Drawings
76 MicroPolar (++) LB 566
8.3 Dimensional Drawings FlowCell
8.3.1 Type LB 5660-102-00X FlowCell DN 50 VFL. FOA
Chapter 8. Technical Drawings
MicroPolar (++) LB 566 77
8.3.2 Type LB 5660-112-00X FlowCell DN 50 G-BS/M
Chapter 8. Technical Drawings
78 MicroPolar (++) LB 566
8.4 Dimensional Drawings Container Probes
8.4.1 Type LB 5650-01
Chapter 8. Technical Drawings
MicroPolar (++) LB 566 79
8.4.2 Type LB 5650-02
Chapter 8. Technical Drawings
80 MicroPolar (++) LB 566
8.4.3 Type LB 5650-03
Chapter 8. Technical Drawings
MicroPolar (++) LB 566 81
8.4.4 Type LB 5650-04
Chapter 8. Technical Drawings
82 MicroPolar (++) LB 566
8.4.5 Type LB 5650-05
Chapter 8. Technical Drawings
MicroPolar (++) LB 566 83
8.4.6 Type LB 5650-09
Chapter 8. Technical Drawings
84 MicroPolar (++) LB 566
8.4.7 Installation Situation in Pipelines
Chapter 8. Technical Drawings
MicroPolar (++) LB 566 85
8.5 Dimensional Drawings Container Flush Probes
8.5.1 Type LB 5651-01
Chapter 8. Technical Drawings
86 MicroPolar (++) LB 566
8.5.2 Type LB 5651-02
Chapter 8. Technical Drawings
MicroPolar (++) LB 566 87
8.5.3 Type LB 5651-03
Chapter 8. Technical Drawings
88 MicroPolar (++) LB 566
8.5.4 Installation Situation in Pipelines
Index
MicroPolar (++) LB 566 89
Index
A
Accuracy · 55 Adapter flange · 42 ASA flange adapter · 59 Assembly on a container · 33 Assembly on a pipeline · 32 Assembly sheets · 91
B
Battery · 53
C
Cable installation · 39, 43 Calculation of measured values · 18 CE mark · 13 Compatible with foodstuffs · 27 Compensation · 19 Components · 21 Conductive materials · 19 Connector strip · 46 Container probe · 27 Container probe installation · 41
D
Data format RS232 · 62 Data transfer rate · 48 Device disposal · 51 Digital outputs · 48 Distance rail · 38 Distance to metal · 41
E
Evaluation unit · 23 Evaluation unit installation, flow cell · 38 EVU · 15
F
Factory setting · 15 Fitting flange · 41 Flat gasket · 41 Flow cell · 25 Flow cell installation · 36 Flush probe · 15, 29 Flushing parameters · 42 Format RS232 · 62 Frequency license · 12, 65 Fuses · 54
G
Gas inclusions · 14, 36
H
HF cable · 30 HF-cable quad · 30
I
Incrustation · 29 Installation depth · 41 Installation situation in pipelines · 84, 90 Instrument cleaning · 53
L
LED’s · 24, 48 Limitations · 18
M
Mains fuses · 54
O
Overview container probes · 60 Overview flow cells · 58
Index
90 MicroPolar (++) LB 566
Overview sealing washers · 60
P
Pipeline pressure · 36 Power radiation · 14 Power supply · 45 Principle of measurement · 17
R
Recycling passport · 51 Relay · 48 RS232 interface · 48
S
Safety summary · 7 Salt content · 14 Service instructions · 51 Setting up the evaluation unit, container probe · 43 Symbols · 7
Synchronization · 20
T
technical data · 55 Technical data HF cable · 61 Technical data sensors · 58 Technical drawings · 73 Temperature compensation · 19 Throughput calculation · 20 Transmission power · 55 Transport · 35
V
Vertical riser · 36
W
Warning · 7 Wear parts · 52 Wiring diagram · 75
Notes
MicroPolar (++) LB 566 91
Notes
Notes
92 MicroPolar (++) LB 566
Notes
Concentration Meters MicroPolar (++)
LB 566
User’s Guide Software Manual 41986BA2
Rev. Nr.: 06-E, 09/2017
MicroPolar (++) LB 566 3
The units supplied should not be repaired by anyone other than Berthold Service en­gineers or technicians by Berthold.
In case of operation trouble, please address to our central service department (ad­dress see below).
The complete user’s guide consists of the hardware manual and the software manual.
The hardware manual comprises the
component description
assembly instructions
electrical installation description
technical data
certificates
dimensional drawings
The software manual comprises the description of the
operation
software functions
calibration
error messages
The present manual is the software description.
Subject to changes without prior notice.
BERTHOLD TECHNOLOGIES GmbH & Co. KG Calmbacher Str. 22 D-75323 Bad Wildbad Switchboard: Service:
Phone +49 7081 177 0 Phone +49 7081 177 111 Fax +49 7081 177 100 Fax +79 7081 177 339
industry@Berthold.com Service@Berthold.com www.Berthold.com
Table of Contents
4 MicroPolar (++) LB 566
Table of Contents
Page
Chapter 1. Safety Summary 7
Chapter 2. Communication with MicroPolar 9
Chapter 3. Getting Started Guide 11
Chapter 4. Software Functions 13
4.1 Information on the Menu Structure 13
4.2 Menu Structure 14
4.2.1 Start Menu 16
4.2.2 Diagnostic 16
4.2.3 Setup 18
4.2.4 Access Level 19
4.2.5 Language 19
4.2.6 Configuration 20
4.2.7 General Data 21
4.2.8 Measurement 21
4.2.9 Plausibility 22
4.2.10 Phase Measurement 23
4.2.11 Offline mode 23
4.2.12 Massflow 24
4.2.13 Calibration 25
4.2.14 System Adjust 25
4.2.15 Calibrate Concentration 26
4.2.16 Sampling 27
4.2.17 Sample Data (expanded) 28
4.2.18 Advanced 28
4.2.19 Calibration 29
4.2.20 Inputs / Outputs 30
4.2.21 Current Output 31
4.2.22 Current Out 1 31
4.2.23 Current Out 2 32
4.2.24 Current input 32
4.2.25 Current Input 1 33
4.2.26 Current Input 2 33
4.2.27 PT100 34
4.2.28 Digital Output 35
4.2.29 Digital Input 35
4.2.30 Service 37
4.3 Trend Display 39
Chapter 5. Configuration 41
5.1 Configuration Setup 41
5.1.1 General Data 41
5.1.2 Measurement 42
5.1.3 Plausibility 42
5.1.4 Microwave 43
5.1.5 Units 43
5.1.6 Marker 44
5.2 Start Calibration Coefficients 45
5.3 Configure Plausibility 46
Table of Contents
MicroPolar (++) LB 566 5
Chapter 6. Calibration 47
6.1 System Adjust 47
6.1.1 Verifying the Reference Values 49
6.2 Sampling 49
6.2.1 Entering the Lab Values 51
6.3 Calibration 52
Chapter 7. Start-up of MicroPolar ++ 53
Chapter 8. Calibration and Advanced 55
8.1 Plausibility | Configuring Phi/Att Ratio 55
8.1.1 Phi/att Ratio 55
8.1.2 Process Recording 56
8.2 Adjusting the Calibration 58
8.3 Output of the Start-up Protocol 60
8.4 Calibration 61
8.4.1 Calibration with Two Concentrations 64
8.4.2 Calibration with Split Value 68
8.5 Typical Calibration Coefficients/Start Values 70
Chapter 9. Password 71
9.1 Password Forgotten 71
Chapter 10. Error Lists and Device States 73
10.1 Hardware and Error Prompts 73
10.2 Input Error 73
10.3 Measurement Error and Error Prompts 74
10.4 Device States 75
Chapter 11. Start-up Protocol 77
11.1 Example of Start-up Protocol 80
Chapter 1. Safety Summary
MicroPolar (++) LB 566 7
Chapter 1. Safety Summary
Please observe all safety instructions in the Hardware Manual, es­pecially those in chapter 1 Safety Summary.
Parameter settings
Never change the installation and the parameter settings without a full knowledge of these operating instructions, as well as a full knowledge of the behavior of the connected controller and the possible influence on the operating process to be controlled.
Loading...