Berthold LB 444 K-40 Operating Manual

Page 1
Potassium Content
Measurement
LB 444 K-40
ID No. 55884BA2
Rev. Nr.: 00
Page 2
Page 3
01/10 347 LB 444 K-40
Operating Manual Potassium Meter
LB 444 K-40
Contents
1 GENERAL INFORMATION ..................................................... 1
2 OVERVIEW........................................................................... 2
3 SYSTEM DESCRIPTION......................................................... 4
4 INSTALLATION OF THE DETECTORS..................................... 5
4.1 Selection of the Measuring Site .............................................................. 5
4.2 Installation of the Detector LB 5430 on a Bunker................................... 6
4.3 Installation of the Detector LB 5402 in a Bunker.................................... 7
4.4 Measurement in a Pipe ........................................................................... 7
4.5 Measurement on a Conveyor Belt ........................................................... 9
4.6 Installation of the Evaluation Electronics ............................................. 10
4.7 Installation of the Detector .................................................................. 10
4.7.1 Installation in a Container .............................................................. 10
5 CONNECTIONS................................................................... 11
5.1.1 Evaluation Unit LB 444 K-40........................................................... 11
5.2 Evaluation Unit LB 444 K-40 ................................................................. 13
5.2.1 General Description....................................................................... 13
5.2.2 Display........................................................................................ 13
5.2.3 Keypad Function........................................................................... 13
5.2.4 Softkeys...................................................................................... 14
5.2.5 Menu Structure............................................................................. 14
5.2.6 Menu Potassium Content Measurement ............................................ 15
6 SOFTWARE FUNCTIONS AND SYSTEM CONFIGURATION ... 19
6.1 General Data......................................................................................... 19
6.2 Operating Mode .................................................................................... 20
6.3 Parameter............................................................................................. 21
6.4 Product Data......................................................................................... 23
6.5 Calibrate...............................................................................................24
6.6 Live Display .......................................................................................... 26
6.7 Service Menu ........................................................................................ 26
7 GETTING STARTED............................................................. 27
7.1 Quick Installation Overview.................................................................. 27
7.2 Getting Started ..................................................................................... 28
7.2.1 Basic Settings............................................................................... 28
7.2.2 Calibration ................................................................................... 29
7.2.3 Checking the Calibration ................................................................ 31
7.3 Live Display .......................................................................................... 31
7.4 Correcting the Results: Addition and Multiplication .............................. 32
7.4.1 Additive Constant.......................................................................... 32
7.4.2 Multiplication Factor ...................................................................... 32
7.5 Automatic Measuring Time Switchover................................................. 33
7.6 Batch Measurement..............................................................................34
7.7 Radiating Interference Detection.......................................................... 35
7.8 Error Messages ..................................................................................... 36
7.8.1 Acknowledging Error Messages ....................................................... 36
Page 4
01/10 347 LB 444 K-40
7.8.2 Error Messages during Operation .................................................... 36
7.8.3 Error Messages during Calibration ................................................... 36
7.8.4 Error Messages during Measurement ............................................... 37
7.9 System Start/Stop................................................................................ 39
8 TECHNICAL DATA ...............................................................40
8.1 Evaluation Unit LB 444 K-40................................................................. 40
8.2 Detectors.............................................................................................. 41
9 SERVICE INSTRUCTIONS....................................................42
9.1 General Safety Precautions................................................................... 42
9.2 Evaluation Unit LB 444 K-40................................................................. 42
10 SERVICE.............................................................................44
10.1 Service Menu Overview ........................................................................ 44
10.2 Service Menu ........................................................................................ 46
10.2.1 Plateau Check .............................................................................. 47
10.3 Detector ............................................................................................... 50
10.3.1 Checking the Crystal-Multiplier Assembly ......................................... 50
10.3.2 Testing and replacement of the photomultipliers in the detector LB 543051
10.4 Replacing the Evaluation Unit LB 444 K-40........................................... 52
10.5 Setup Protocol...................................................................................... 53
11 DIMENSIONAL DRAWINGS.................................................55
11.1 LB 5402 ................................................................................................ 55
11.2 LB 5430 ................................................................................................ 55
11.3 LB 444 K-40.......................................................................................... 56
II
Page 5
01/10 347 LB 444 K-40
Operating Manual
Potassium Meter
LB 444 K-40
Issue
No. Date Comments 1 01.01.10 First edition
III
Page 6
Page 7
01/10 347 LB 444 K-40
1 General Information
Safety Summary
Electrical Shock Hazard
Disconnect power to ensure that contact with energized part is avoided during installation and servicing.
Specific Warnings
Never change the installation or the parameter settings without a full knowledge of the relevant part of this manual, the connected con troller and the process, if it is controlled by this measuring device.
1
Page 8
01/10 347 LB 444 K-40
2 Overview
The Potassium Meter LB 444 K-40 is designed for measurements of the concentration of potassium in bulk materials, suspensions and solutions.
The concentration of potassium in bulk materials or liquid solutions is measured by detecting gamma radiation of the natural isotope K-40. This isotope is contained in natural potassium in a constant percentage (0.0119%). As the isotope K-40 decays, it emits gamma radiation with an energy of 1.46 MeV.
The detector LB 5430 detects radiation using an organic scintillator (PVT), the detector LB 5402 using a NaI crystal. The radiation triggers tiny flashes of light in the scintillator which are converted into electrical pulses by the photomultiplier. The count rate of this radiation is a direct measure for the potassium concentration. In addition , the scintillation detector responds to radiation emitted from the environment, for example, the walls and the floor which are covered with a layer of potassium.
Further radiation is generated by incident cosmic rays at the location of the detector. Background radiation remains more or less constant for each installation within a plant,
provided the location is not changed. To keep the statistical error of the measured result as small as possible, the scintillation detector, fixed at the outside of a bunker or below a belt conveyor, is shielded against background radiation by an additional lead shielding.
If the scintillation detector is inserted int o bulk material, solutions or suspensions inside a protection tube, the material itself is acting as a shielding against the background radiation.
Potassium can be found predominantly in the form of KCl. However, to this day the unit K2O is still used in many companies. The evaluation unit LB 444 K40 allows the calibration and measurement in both units. Important: the calibration and measurement have must be performed in the same unit. The following details for “K20” also apply for “KCI”.
There is a linear relationship between the count rate measured by the scintillation detector and the potassium concentration.
This relationship can be expressed as follows: K = a (N - N where:
= count rate measured by the scintillation detector if the bulk material or the
N
0
liquid contains 0% K “N” = count rate detected by the scintillation detector at a concentration “ K” of K “K” = concentration of K “a” = reciprocal value of the count rate per % K * This value can also be understood as the slope of the linear function. The term a x N
is the background expressed as a percentage of K
) (1)
0
O.
2
O.
2
O.
2
O.*
2
0
O.
2
2
Page 9
01/10 347 LB 444 K-40
The hardware and software of the LB 444 K-40 system makes it easy to adapt the system to rather different measuring geometries and measuring tasks. Therefore, the settings and parameters of the measuring instrument have to be defined with care for the respective measuring task when taking the system into operation. Important parameters may not be changed later, in order not to compromise the reliable operation of the system. The system should be taken into operation and settings changed only by persons who know how to work with the instrument. Therefore, all users should read these operating instructions carefully. We recommend documen t ing all set ting s in a set up protocol.
Before starting any work, please read this operating manual carefully!
3
Page 10
01/10 347 LB 444 K-40
3 System Description
The system is comprised of the following components:
Figure 1: Measuring system with LB 5430 detector
a) Detector LB 5430 for installation on the outside of the container or b) LB 5402 for installation in a protective pipe inside the container c) Connection cable d) Evaluation unit LB 444 K-40
A 19” rack can be supplied for installation of up to 4 devices or a protective hous ing for two devices for installation of the evaluation unit.
The device LB 444 K-40 with a software that has been expanded for potassium content measurements is used as evaluation unit. The functions of the density measurement are still available. This operating manual describes only those functions that are relevant to the potassium content measurement.
4
Page 11
01/10 347 LB 444 K-40
4 Installation of the Detectors
4.1 Selection of the Measuring Site
Assuming a homogeneous mixture of potassium in the material to be measured, the number of K-40 isotopes increases with increasing amount of material. This means that the count rate of the detector rises when the amount of material in front of the detector rises, although the material contains a constant percentage of the K-40 isotope. The radiation is partially absorbed between its point of origin and the detector. Consequently, material that is farther away from the detector does not contribute much to the result. If the material thickness in front of the detector is increased, or the detector is surrounded by material, the count rate will rea ch a saturation value and does not rise anymore even if the thickness of the material layer st ill increases. This material thickness is called saturation thickness which is expressed as follows:
d = 100/ρ cm, where ρ is the density of the bulk material in g/cm
Example: For a potassium sample with a bulk weight of 1.2 g/cm
thickness necessary for the measurement would be 100 cm /1.2 = 83 cm.
This means, for a surface probe, the inner diameter of a bunker should be
about 83 cm; a dip probe should be inserted in the center of a bunker having
an inner diameter of at least 166 cm. These conditions ensure that the measured value is independent of bulk density
variations. If this conditions cannot be fulfilled, one has to calculate if the errors that occur as a result of a smaller volume and a possible change of the bulk weight or the density are still acceptable.
The goal of a continuous measurement of the K content is to cover the detection device with the total amount of material to be measured, i.e. the position of the scintillation detector has to be selected such that the surrounding material thickness is not much thicker than the saturation thickness.
If this is not the case the material being further away from the detector than the saturation thickness would not contribute to the measurement and therefore its potassium content would not be taken into account (see Figure 2). Provided the product to be measured shows a homogeneous structure, no measurement errors will occur.
3
.
3
the saturation
2
1
S
0
2
S
S
S = saturation layer thickness 1 = measured material layer
Figure 2: Saturation layer
5
Page 12
01/10 347 LB 444 K-40
2 = material not covered by the measurement
4.2 Installation of the Detector LB 5430 on a Bunker
The LB 5430 detector is installed on the outside wall of a bunker. Typically, it is fixed to the flange of the detector.
Figure 3: Detector LB 5430
Depending on the version, the detector weights approx. 60 – 70 kg. Please take this weight into account in your planning. All screws and mounting parts have to be secured such that they cannot come loose during operation, so that the detector cannot fall down.
The detector must not be subject to heavy mechanical vibrations. If extreme vibrations are likely to occur at the installation site, the detector has to be mounted on a damping and shock absorbing arrangement. Care must be taken that no potassium containing material may be deposited between the surface probe and the outside wall. This would result in measurement errors.
The ambient temperature at the installation site of the detector must not exceed 50°C. If the bunker wall is hot (>50°C), you may insert a suitable insulating plate having a thickness of approx. 20 mm between the container wall and the detector or use a detector with water cooling.
6
Page 13
01/10 347 LB 444 K-40
4.3 Installation of the Detector LB 5402 in a Bunker
The detector is installed into a protective pipe, front side closed, and inserted into a bunker. The distance of the material surface in front of the detector head sensitive to radiation (position of the scintillation crystal) should on all sides correspond to the calculated saturation thickness. If hot liquids are used a cooling water jacket for the probe is necessary.
The detector temperature must never exceed 50°C.
No crystallization must occur on the protective pipe. If necessary, the protective pipe has to be provided with a thermal insulation and possibly with a heating jacket. Crystallizations would significantly falsify the measured results, since the potassium concentration in the crystallized layer is typically higher than in the liquid solution.
S = saturation layer thickness 1 = measured material layer 2 = material not covered by the measurement
Figure 4: Measurement in a container
4.4 Measurement in a Pipe
When measuring in a pipe, the saturation layer thickness is usually n ot achieved or at least not achieved in all directions. In the case of large pipes, the saturation layer thickness is usually achieved in the longitudinal direction of the pipeline, However, transverse to the pipeline axis, the saturation layer is not always achieved. This can have an effect on the measurement result, which is larger:
a) the smaller the pipe is b) the stronger the density varies in the pipeline, (see Figure 6).
The measurement of the saturation thickness is also affected by the diameter of the pipe, the pipe material and the wall thickness of the pipe. A correction of the influence is possible in some cases by an additional density measurement. This will require an extended period of time for the recording of
a) Density values b) Displayed K2O values and c) the appropriate, K2O values determined in the laboratory.
7
Page 14
01/10 347 LB 444 K-40
A correction function can then be calculated from these values.
Figure 5: Measurement in a pipeline
Potassium measurement. Saturation thickness for different densites resp.
bulk densities
100
90
80
70
60
50
cps in %
40
30
20
10
0
0 100 200 300 400 500 600 700 800 900 1000
Density =1,4
Density =1,0
Density =1,2
Layer thickness in mm
Figure 6: Influence of thickness and density
Installation:
8
Page 15
01/10 347 LB 444 K-40
The detector should be mounted in a horizontal pipeline at the top or side. This will prevent that the measurement is influenced by any deposits on the ground of the pipeline.
4.5 Measurement on a Conveyor Belt
Normally, the required saturation thickness cannot be reached on a conveyor belt, so the measured result will be influenced by changes of t he bulk weight and/or changes of the material layer thickness.
However, if it is ensured that
• variations of the bulk weight in the material cannot occur (uniform grain size)
• the height of the material layer is kept constant by leveling device.
a measurement can also be carried out on a conveyor belt. However, the measurement error is larger than with measurements in a bunker.
9
Page 16
01/10 347 LB 444 K-40
4.6 Installation of the Evaluation Electronics
The evaluation unit (protection type IP 20) can be installed in a 19” rack. This rack then has to be installed into a suitable measurement cabinet. The wall housing can be fixed directly to a wall using screws. For outdoor installation, you may have to install a sunshield to protect the evaluation unit against solar heating.
4.7 Installation of the Detector
Use a water cooling jacket to protect the detector against temperatures > 50°C. Cables have to be protected against temperatures > 70°C.
Measuring system installed outdoors have to be protected from rain and direct sunshine by a canopy. The cable entry must not be exposed to temperatures exceeding 70°C. The cable gland and the cable entry must be aligned such that no water can flow along the cable into the connection box. The detector installation site must not be subject to heavy vibrations as this could damage the detector.
Samples have to be taken for calibration of the measurement. A sampling point foreseen in the direct vicinity of the measuring site.
has to be
4.7.1 Installation in a Container
If the detector is installed in a protective pipe inside the container, the protective pipe has to be sealed to prevent the penetration of dust and humidity. The detector includes two suspension eyes. Wire ropes can be attached to these suspension eyes. It is not permitted to suspend the detector on the cable. The cable should form a loop to allow possibly penetrating water to drip off.
Suspension eyes
Figure 7: Installation of de­tector in a protective pipe
Loop
10
Page 17
01/10 347 LB 444 K-40
5 Connections
The power supply of the detector and the transfer of the measured values and important information to the evaluation unit LB 444 K-40 takes place via a two-wire screened cable. The possible cable length is dependent on the cross-section of the connection cable used.
The cable entry at the detector permits a cable diameter of 6-10 mm.
5.1.1 Evaluation Unit LB 444 K-40
Connect cable on the rear panel of the evaluation unit as shown in the wiring diagram in the appendix to this manual.
LB 444
cross-section in mm² Max. cable length in m
1 1000
1.5 1500
2.5 2500
Probe
2
1
Fuses
Voltage selection
2(-) 1 (+)
c a
2
4
4
6
Distanzpart
6
8
8 10 12 14 16 18 20 22 24 26 28 30 32
Relay 2 Relay 3 Relay 1 Dig. IN 1
Dig. IN 2 Dig. IN 3
RS 485 0/4 - 20 mA
Mains
Figure 8: Terminal connection evaluation unit (rear panel)
Connect the device only to the appropriate line voltage. All safety previsions regarding the power distributor have to be observed. The evaluation units do not include a separate mains switch. A separate
safeguarding and easily accessible shut-down have to be foreseen. The device must not be installed in explosive hazardous areas.
Refer to the wiring diagram in the appendix to this manual regarding the connections.
11
Page 18
01/10 347 LB 444 K-40
Detector (2a/2c) Relay 2 (12a/12c)
The relay can be set for the following functions: Min. – Alarm, Max. – Alarm, Detector temperature,
Relay 3 (4a/14c)
Functions: Min. – Alarm, Max. – Alarm, Detector temperature, Relay 1 (16a/16c) The relay is used to indicate errors. The contact opens in case of error.
External product selection
Digital input 1 (18a/18c) Digital input 2 (20a/20c) Calibration data of up to 4 different products can be defined in the evaluation unit LB 444 K-40. These data sets can be selected via digital inputs.
External Start/Stop Signal (22a/22c)
Option to interrupt measurement for the following special applications:
- Start/Stop in batch mode
- Stop of continuous measurement, e.g. to suppress unexpected malfunctions
0/4-20mA (26a+/26c-)
Isolated current output for measured value, max. load 500 Ω.
Current input (28a-/28c+)
Product temperature for temperature compensation (usually not needed for potassium content measurements).
Power supply (30a/30c)
Power supply 115V/230V AC or 24 V AC/DC depending on power supply unit (see label on instrument rear pane!). Fuses To replace fuses, you have to open the fuse holder with a screw-driver. Observe fuse type and rating!
Before turning on the power supply, carefully check all conn ections once more to rule out any damage to the instruments.
RS 232
Port for data transfer from evaluation unit to printer or PC (front panel of evaluation unit).
12
Page 19
01/10 347 LB 444 K-40
5.2 Evaluation Unit LB 444 K-40
5.2.1 General Description
The evaluation electronics is designed as a 19” module in the format 3 HE, 21 TE. It includes the microprocessor­controlled evaluation electro­nics and the power supply for the required operating voltage. A 32 bit microproc­essor is used for signal processing.
LCD displa y
Function keys
Softkeys
BERTHOLD LB 444 - 1 V 1.0 Density -Meter
enter
enter
clear
run
enter enter
clear run
more
Figure 9: Operating elements of the evaluation unit
The instrument is operated via six foil keys: Three keys work as softkeys which allow user-guided definition of all instrument settings and input of the required parameters. Three more keys serve as function keys. The front panel includes a RS 232 interface port. Data can be transferred to a PC using a terminal program. The terminal strip on the instrument rear panel includes all terminals for power supply, for the detector, and for the analog and digital output signals. The current output is isolated. The built-in relays for max.-min.-indication an d for error messages include an isolated contact. System malfunctions are signaled by error messages. Calibration data is stored in a FLASH memory and saved, so they will n ot be lost in case of power failure.
5.2.2 Display
The instrument’s illuminated display compri ses four lines. The first three lines show the menu titles, the currently selected parameters or the current measurement value. The bottom line shows the current function of the respective softkey button. If a measure­ment is running, the “run” status is displayed.
5.2.3 Keypad Function
The Potassium Meter is operated via the softkeys and function keys described below. With these keys you can select the operating level you want within a menu structure in order to select a function or enter parameters.
13
Page 20
01/10 347 LB 444 K-40
T
5.2.4 Softkeys
Softkeys are used to select different menu grou ps and operating levels within the menu structure. Depending on the current position in the menu structure, functions are assigned to these keys, as shown on the display above the respective key.
5.2.5 Menu Structure
The menu structure is illustrated on the following pages. Push <more> to select the various menu groups. From there you get to the respective menu by pushing <sk1> or <sk2>. Within the menu, push <more> to go to the individual windows and at the end of the menu push <done> to return to the menu group.
sk1: General Data sk2: Operating Mode
sk1 sk2 more
Relay 2 Minimum 50% Hysteresis 5%
^^^
<-<-<-
Time constant Value 20s
^^^ <-<-<- more
HV (500): 20/s HV (560): 123/s HV (560): 620/s + - exit
more
sk1 and sk2 more Go to next display or menu group done Shows the end of the menu and
^^^ Numerical values: increments the

+ and -
Go to the indicated menu
takes you back to the menu group
ext: scrolls through the various
selection options
number marked by the cursor by 1
Moves the cursor to the left and at the end of the input field again to the start position
Scrolls forward / back in a result list or in the live display
Figure 10: Softkey functions
14
Page 21
01/10 347 LB 444 K-40
Enter Accepts the entry and moves the
cursor to the next input field or
sk1 : General Data sk2 : Operating Mode
sk1 sk2 more
enter
clear run
Figure 11: Function keys
Run Starts or stops a measurement or
Clear
toggles between two input fields.
leads directly back to the display and at the start of a measurement automatically changes to the measurement value display.
Clears the numerical value.
5.2.6 Menu Potassium Content Measurement
Select the menu item Operation mode more  K20 ENTER to automatically select the parameters that are relevant for the potassium content
measurement:
• linear characteristic
• time constant 300 s
• measuring path “1”
• no decay compensation
In addition, the detector code has to be checked and changed, if necessary.
Detector LB 5430 with organic scintillator (PVT) Detector LB 5402 with NaI scintillator Detector code 0
Detector code 23
15
Page 22
01/10 347 LB 444 K-40
Password: xxxx
Date & Time Date : 01.02.00 Time : 12.21
Berthold LB 444 K-40 V. X.XX Density - Meter
Berthold Detector Software Version 1.02
Language
Deutsch
English Francais Espanol
Print Parameter?
(press Enter)
sk1: General Data sk2: Operating Mode sk1 sk2 more
Config instrument
Density without TK
Density with cur. Inp Mass Flow without TC Mass Flow with Pt 100
Radiation measure
Backscatter measure
K20/KCl
Error mode
in error case
stop measurement/
continue measurement
Relay in error case
Hold state
Current following
Current outp. error current: HOLD / VALUE Value: 12 mA
sk1: Parameter sk2: Product Data sk1 sk2 more
Measuring product No.: 1/2/3/4 Product 1
Detector und Isotope Code: 0 Isotope: Cs 137
Measuring Path
Value : 10,0 cm
Select: Measure Mode
Continuous Measure Batchmode
Time constant
Value: 10.0s
Rapid switchover
OFF / ON
Sigma: 3,0000
Radiating inter ference OFF/ ON Sigma: 5.0000
Product select Nr.: 1/2/3/4 Product 1
Current Output Range : 4 - 20 mA/
0 - 20 mA
Curr. Output Limits
0/4 mA 1,0000 20 mA 1,2000
Relais Nr. 2 setup
Minimum Maximum Det. Temp
Hysteresis : 5%
Relais Nr. 3 setup
Minimum
Maximum Det. Temp
Hysterese : 5%
Temperature coeff. TC 1 Value : 1.2345e-03
Product select Nr.: 1/2/3/4 Product 1
Suspensions measure
yes / no
Solid density Value: 2.6500 g/cm3
Liquid density Value: 1.0000 g/cm3
Water TC ? yes / No
Unit select [g/cm2] [Bx] [Be] [g/] Unit : g/cm3
sk1:Data input sk2:Calculate sk1 sk2 more
sk1:Calibrate sk2:Live Display sk1 sk2 more
Calibrate mode
one / lin / squ / cub / auto
Mode: linear
Zero countrate Io Value: 54321 cps
Coefficient a1 Value: - 6.64000e - 02
Coefficient a2 Value: - 1.23400e - 04
Coefficient a3 Value: - 1.23456e - 06
Square Error Value: 0.000000
16
Factory setting
Current input: Range: 0 - 20 mA
4 - 20 mA
Temp. Input 0 / 4 mA: 0 °C 20 mA: 10 0 °C
Radiating inter ference:-: Delay time: : 20 s
Maximum Rate: Value: : 100000 cps
Minimum Rate: Value: : 0 cps
Product selection
extern / intern
Temperature coeff. TC 2 Value : 1.2345e-06
Reference Temperature Value : 30 °C
Calibr. Data transfer
yes / no
Product 1->2 ?
1. Rate = 12345 cps
1. 1.0000 g/cm3
1. Temp. = 44,2 °C
10. Rate = 12345 cps
10. 1.0000 g/cm3
10. Temp. = 44,2 °C
Factor Value: 1.0000
Offset Value: 0.0000
Calibration value
1. 1.0000 g/cm3
2. 1.1200 g/cm3 + -
Page 23
01/10 347 LB 444 K-40
sk1: Service Menu sk2: Mass Flow
sk1 sk2
Page 16
sk1: Test calculate sk2: I / O - Test sk1 sk2 more
Testrate 12345 cps Output Curr : 10,0 mA 1,2345 g/cm3
Unit Mass Flow [kg/h, t/h] t/h
Range input {at 20 mA] Value: 100 m3/h
Flow Value 87.00 m3/h
Ext. Mass Counter Tons per cycle
Erase mass ­integrator no/ yes
Set outputcur Current : 9.0mA
View input cur. Current: 12.0mA
Relay 1 : ON Relay 2 : OFF Relay 1 : ON
Input 1: OFF Input 2: OFF Input 3: ON
sk1: HV Adjustment sk2: Status request
sk1 sk2 more
High Voltage Value : 800 V
Save Default
Value : 850 V
Read Default
Value : 850 V
Temperature =
45.2°C Det. Temp. = 28 °C
For Detector - Reset please press (sk 1 & clear)
sk1: Reset Detector sk2: Pt 100 adjust
sk1 sk2 more
For Pt 100-Adjust please press
sk1: Measure Plateau sk2: Request Plateau
sk1 sk2
Plateau measurement
running......
IIIIIIIII...........................
HV( 500): 817/s HV( 560): 884 /s HV( 620): 923/s + - done
sk1: Plateau sk 2: Adj. Current out sk 1 sk2 more
Current 1.8 mA Offset: 4123
-
Current 18 mA Offset: 345
-
Curr. Input Adjust Set value 1.8 mA Real value 1,7 mA + - more
Curr. Input Adjust Set value 18 mA Real value 17.5 mA + - more
sk1: Adj. Cur. input sk 2: Relay Delaytime sk 1 sk2 more
Relay delay Value: 50 ms
17
Page 24
01/10 347 LB 444 K-40
sk1: Service Menu sk2: Mass Flow
sk1 sk2 more
sk1: Interface sk1 more
Interface RS 232: Printer/PC
RS 485 Printer/LB 447
sk1 more
sk1:Calibrate sk2:Live Display sk1 sk2 more
L B 4 4 4 DENSITY
1.2345 g/cm3 + -
Value: 1.2345 g/cm3 Temperature: 45.2 °C
1.2345 g/cm3 + -
RS232 PC only
PC Access Control
Read only data / change of data
I-mean = 12345 cps I-actual = 12435 cps HV -auto = 745 V + - more
Detector Temperature: 47 °C
+ - more
18
Page 25
01/10 347 LB 444 K-40
6 Software Functions and System
Configuration
Depending on the selected configuration, some input dialog boxes are dimmed. In chapter 5.2.5 you find an overview of the menu structure After commissioning, enter the final operating settings in the Configuration list in the Appendix. Individual options are selected by pushing the <^^^>. Push <enter> to accept the selected value.
After the last entry, the instrument has to be connected to mains for at least 5 minutes to ensure that all values entered have been stored in the FLASH memory and will not be lost if power failure occurs.
6.1 General Data
Password:
You can enter a 6-digit number. Upon confirmation with <enter>, the system is protected against unauthorized manipulation of the parameters. Display: Keyboard locked. All parameters can still be viewed, however. Password protection is revoked as soon as you enter the correct numerical value and confirm your entry with <enter>. To change the password, enter a new password while the keyboard is unlocked.
Date:
Enter the current date in the format DD.MM.YY.
Time:
Enter the current time in the format HH.MM.
System/Version:
Shows the instrument type and software program version. Please have this dat a handy if you need to consult with the manufacturer.
Detector Software:
Shows the software version installed in the detector. Please have this data handy if you need to consult with the manufacturer.
Language:
German, English, French and Spanish can be chosen as dialog language by pressing the <^^^> key.
Print Parameter:
A printer with serial interface can be connected on the front panel. Push <enter> to print all parameters for documentation . You may also connect a PC and transfer the data to a PC for processing using a terminal program. See also 6.2
Factory setting: Push <sk1> to reset all parameters to the factory setting.
The current calibration data is lost when the instrument is reset!
19
Page 26
01/10 347 LB 444 K-40
6.2 Operating Mode
Measurement mode K20/KCl.
Error mode in case of error Stop measurement
in case of error Continue measurement
Relay in error case Hold state
Current following Current output error
The parameters that are relevant for the potassium content measurement are set automatically.
• linear characteristic
• time constant 300 s
• measuring path “1”
• no decay compensation
The measurement process stops if an error occurs. The error is indicated on the display. Push <enter> to clear the error message on the display. If the error has been eliminated, push <run> to start the measurement again. If you select “continue measurement” the error is displayed, but the measurement continues.
The max./min. relays behave corresponding to the measured Concentration. The max./min. relays follow the fault current (see chapter 6.4)
Current: HOLD VALUE
RS232 Interface:
The RS232 Interface is located on the front panel of the evaluation device. Select baud rate by pressing <^^^> and confirm with <enter>. Select device to which data is to be transferred: select either PC, printer or modem. Terminal program settings: In LB 444 K-40: Printer In terminal program: 9600 baud, 8 data bit, 1 stop bit, no parity, protocol: X on / X off. RS 485: Presently not used.
The output current is held at the last value where the error has occurred. In case of error, a pre-defined output current is selected (see chapter 6.4).
20
Page 27
01/10 347 LB 444 K-40
6.3 Parameter
Measuring product No. 1/2/3/4:
Parameters (detector type, current output range, limit values, etc.) for up to 4 different products can be entered and stored. These parameters have to be entered separately for each product. The calibration data measured for a product may be copied to other data sets. Thus, different current ranges (0/4-20 mA) can be defined for the same product and can be selected externally by selecting the “new” product , or different limit values can be used. In the normal measurement mode the product for the current measurement is selected here.
Detector and Isotope Code:
Detector-specific code number taking into account the parameters to be defaulted for the detector type.
Detector LB 5402 Enter code “0”. Detectors LB 5430 / LB 4430: Enter code “23”.
Measuring Path (0.1 to 9999.9 cm):
“1” is used for the potassium content measurement.
Select: Measure Mode: Continuous:
Batchmode
Continuous measurement. This is the standard setting for most measurements. With this type of measurement, a sliding average is calculated over the selected time constant. Batch measurement with external start-stop signal (digital input 22a and 22c). or via the <run> button. We recommend using the “Batch measurement” mode, for example, when a pipeline or container is filled briefly and then emptied again. In this case, shorter measuring times can be used. During batch measurement an arithmetic mean is calculated over the measurement time. To start a measurement process, either push the <run> button or briefly connect terminals 22a /22c.
A measurement stops as soon as the entered measurement time is over (time constant). The shortest batch time that can be used is approx. 2 s. or if you push the <run> button again. See also 7.6
Time constant:
Enter the time constant of the instrument system. It determines the average calculation of the counts supplied by the detector. In order to reduce statistical variations, you should select the highest time constant that is still permissible. T he typical input value is 300 s.
A measurement time can be preset for batch measurement. Once this time is over, the measurement stops automatically. The shortest batch time that can be used is approx. 2 s.
21
Page 28
01/10 347 LB 444 K-40
Rapid switchover: ON / OFF:
This function (= 1/10 of defined time constant) is needed to adjust the output signal quickly to the new value if sudden Concentration changes occur. This function is not used for potassium content measurements.
Sigma
Sigma defines a range (window) which the display change has to exceed before the selected time constant automatically switches over to a value that is smaller by a factor of 10 (max. 0.8 sec). When using the rapid switchover function, you shou ld at least enter “4” or better “5” as Sigma value to make sure that a switchover of the time constant is not triggered too often by statistical variations or minor Concentration changes. We recommend determining the optimum entry during routine operation (see also chapter 7.5).
Radiating interference:
This function is needed only when unforeseeable influences due to interfering radiation are to be expected, e.g. due to weld seam testing. It is enabled and disabled via <ON> and <OFF>. Sigma defines a limit value which the display change has to exceed before the measurement is canceled. The value has to be set such that short-term Concentration variations occurring during operation do not trigger this function.
Enabling radiating interference detection disables the rapid switchover function.
Radiating interference delay time: (only with Radi ating interference “ON”)
Interference due to radiation is normally only of short duration. After the delay time you have entered here is over, the measurement starts again automatically following detection of radiating interference.
Maximum rate:
The measurement stops automatically as soon as the count rate supplied by the detector exceeds the entered value. The output current signal is held at the last value. The error message “Measurement halted” is displayed. The measurement continues automatically as soon as the count rate has dropped below the entered value. The entered values are permanently corrected during decay compensation. The functions “Maximum Rate” and “Minimum Rate” can be used to “freeze” the measured value in case of malfunctions (e.g. empty measuring path, empty container, increased radiation due to weld seam testing, etc.) and to avoid long response times of the control devices. Select the input values such th at that they will not be reach ed during regular operation. If this function is not important for your operation, disable it by en tering a pulse rate of “0”.
Minimum rate:
The measurement stops automatically as soon as the count rate supplied by the detector falls below the entered value. If this function is not important for your operation, disable it by en tering a pulse rate of “0”. For the potassium content measurement you should always enter the value “0”.
Product selection [extern/intern]
Define if the product is to be selected via the keyboard of the evaluation unit (internal) or via digital inputs (external, 18a/c, 20a/c).
Product DI1 18
DI2 20 a/c
a/c
1 0 0 2 0 1 3 1 0 4 1 1
1= terminal a/c connected
22
Page 29
01/10 347 LB 444 K-40
6.4 Product Data
Enter product data for parameter setup and calibration separately for each product.
Product selection
No. 1/2/3/4: Enter product number for calibration.
Current Output:
Select current output signal 0 or 4 mA.
Current Output Limits: 0/4 mA: 20 mA: Example:
Measuring range 10 ... 30% K20.
Input: at 4 mA: 10 at 20 mA: 30
Current Output Error: Current:
VALUE HOLD
In case of error, the current output jumps to the defined value ( 0...22 mA)
In case of error, the current output holds the last value. Caution: If you have selected “continue in case of error”, the current output operates using a wrong value corresponding to the false value.
Relay No. 2 setup: Minimum / Maximum / Det. Temp. / Mass pulse:
Different functions can be assigned to the relay. As soon as the status to be signaled Concentration < minimum or Concentration > minimum Detector temperature > X °C has been reached, the relay is de-energized. If the displayed value drops again, the relay picks up again, delayed by the value of the selected hysteresis.
Hysteresis
The switching hysteresis should be about 5%.
Relay No. 3 setup:
Same as relay no. 2.
Minute printout Value
The measured value can be printed out in defined time intervals using a printer connected to the RS 232 interface.
23
Page 30
01/10 347 LB 444 K-40
6.5 Calibrate
Background for rod detector N/A Product selection
No. 1/2/3/4: Select product to be calibrated and confirm with <enter>.
Suspension measure No <enter>.
Unit select
g/cm³, t/m³,Bx, weight %, °Be, g/l, mm, g/m Select the unit for concentration measurement and area weight measurements (mm, g/m²).
% K20 <enter>.
Note: The selection of the displayed unit does not have any influence on the calibration or the measurement.
Data input
Enter or read in the calibration data for the requested number of calibrat ion points.
Calibr. Data transfer
Select “Yes” to copy the calibration data of a product that has already been calibrated. In the copy you can change the data then as needed.
Sequence: Yes <enter> Product X <enter> Y <enter> etc. Then run through Rate 1 to Rate 10 and correct them if necessary. You may correct the values, if necessary.
2
, % K20
1. Rate to 10. Rate
You need at least 2 calibration points (see chapters 6.5 and 7.2.2). You have to run through all 10 points. Calibration points with zero values are not calculated. Data can be entered manually or read in by pressing the <run> button. Its also possible to use the background radiation as a calibration point. This has to be measured
a) with empty vessel or b) with product in the vessel which does not contain any potassium or c) dismounted detector arranged on a concrete floor or min 20 mm lead or 40 mm
steel plate in front of the radiation window.
The count rate is read in automatically. As soon as the measured value is stable, press the <run> button again to stop the reading-in process. Press <enter> to confirm the read-in or entered value.
If you push <more> to go to the next item without pushing <enter> first, the
24
Page 31
01/10 347 LB 444 K-40
values will not be saved.
Rate: Concentration:
Count rate in cps is read-in or entered. Enter the concentration value %K20 determined in the lab.
sk2: Calculate
Calibrate mode Select the calibration mode: on/lin/squ/cub/auto (see also 6.5 and 7.2.2).
The calibration mode “linear” is used for the K2O measurement.
One-point
N/A
One-point calibration. For one-point calibration we only need one value pair (count rate and concentration value) and the absorption coefficient. For common products it can be taken from the table in the Appendix and entered as coefficient a1.
Linear
Linear curve fit. This mode is used if at leas t two value pairs are available. It should also be used when several calibration points are available which are very close together, so that not the entire measuring range is covered by samples.
Square
N/A
The square curve fit can be selected if at least 3 calibration points are available which are distributed fairly ev enly over the measuring range or if the measurement is carried out in %-concentration, °Bx or in another unit which is not in linear correlation with th e concentration.
Cubic
N/A
The curve fit with cubic coefficient should be applied when the same conditions exist as under square, but at least 4 value pairs are available and it becomes apparent during operation that values deviate in some parts when using square.
Auto N/A The program automatically selects the best curve fit.
At least 4 value pairs must be available for this function.
Select the calibration mode and push <enter> to calculate the calibration factors.
Coefficient a1
Linear absorption coefficient (range -10 to +10). The coefficient is calculated by the program. At least two calibration points have to be available.
Zero count rate Io
The zero count rate is automatically calculated for calibration. It corresponds to the zero point of the unit of measure, i.e. the concentration value “0”.
Coefficient a1
The calculated or entered coefficient a1 is displayed. Do not change it manu ally; this also applies to coefficients a2 and a3.
Coefficient a2 N/A
Square absorption coefficient. This value is automatically calculated for two- or multi­point calibration.
Coefficient a3 N/A
Cubic absorption coefficient. This value is automatically calculated for multi-point calibration.
Square error.
25
Page 32
01/10 347 LB 444 K-40
This value is calculated automatically for tw o- or multi-point calibration. It indicates the quality of the calibration curve for the selected calibration mode. The smaller the numerical value, the better the curve fit.
Factor
Enter a multiplication factor (0 to 10) to correct the measured values. Each measured value is multiplied by this factor. It allows you to change the slope of the calibration curve. When performing a new calibration, the factor is automatically set to “1” (see also chapter 7.4.2).
Offset Additive correction of the measured values; allows parallel offset of the curve. The offset entered here is added to each measured value. When performing a new calibration, the value is automatically set to “0” (see also chapter 7.4.1).
Calibration value N/A
(Only for temperature compensation via Pt 100 or current input:) The temperature compensated lab values are displayed which were entered at rates 1. to 10 (g/cm values.
3
). Scroll with <+> and <->. The calibration curve is calculated using these
6.6 Live Display
Start measurement
A measurement can be started from any menu item by pressing the <run> button (exception: Service menu). The measured values are displayed continuously in the selected unit for the selected product. RUN appears on the display. With <+> and <-> you can cycle through the displays below.
Menu sk2: Live Display
Results are displayed in this menu if RUN appears on the display; otherwise the values of the last measurement. With <+> and <-> you can cycle through the displays below.
Display:
The actual concentration can be displayed in the defaulted range, e.g. 17.8% K20.
I-mean
Shows the averaged count rate. The entered basic time constant is used as time constant.
I-actual
Shows the actual count rate. The entered basic time constant is used as time constant. HV auto: Shows the adjusted high voltage (HV). Stop measurement Select product
by pressing <run> again.
on the menu sk1: Parameter.
6.7 Service Menu
See chapter Service Instructions.
26
Page 33
01/10 347 LB 444 K-40
7 Getting Started
7.1 Quick Installation Overview
Page Install detector at the measurement site. 10 Install water cooling if temperatures exceed >50°C. 5 Install evaluation unit. Connect detector to LB 444 K-40 via two-wire cable. Connect cable to terminal 2a and 2c of evaluation unit.
External product selection With the external product selection you can measure up to 4 different products. Both digital inputs have to be used so that 4 different products can be selected. Digital input 1(DI 1): 18a / 18c Digital input 2 (DI 2): 20a / 20c
Product DI 1 DI 2 1 0 0 2 0 1 3 1 0 4 1 1
23
Digital input 3 (DI 3): 22a /22c. External Start/Stop signal Digital input 3 can be used for the following signals: a) Start/Stop in batch mode b) Stop continuous measurement. (“freezing” measurement; “HALT” appears on display)
Digital outputs (relays): These outputs can be used for the following signals: 12a, 12c: Rel. 2: Min. , max. , detector temperature, 14a, 14c: Rel. 3: Min. , max. , detector temperature, 16a, 16c: Rel. 1: Error message
27
Page 34
01/10 347 LB 444 K-40
7.2 Getting Started
7.2.1 Basic Settings
These steps describe how to take the system into operation and how to change the basic system settings.
Note: To enter numbers, select the entry position with <←←←> and the number you want with <^^^>. Confirm your selection or new entry with <enter>. Press <run> to return to the display mode directly any time (except in Service menu). If no entry is made, the system automatically switches from the current menu item to the display mode after several minutes.
Important:
After data input the system has to remain turned on for at least 5 minutes to ensure that all entered data have been saved.
1. Push <more> to select menu General Data / Operating Mode
Push <sk1> to select General Data and <more> to call parameters sequentially. Enter password and unlock or lock system with <enter> Check date & time and correct it, if necessary Check instrument version (e.g. LB444 K-40, version: 3.xx)
Select language Print Parameter, only if needed Factory Setting, only if needed (see also Software Functions and System
Configuration) Push <done> to return to menu group
2. Push <sk2> to select Operating Mode menu and <more> to select parameters. Select Config Instrument N/A and continue with <more> Select K20 Define Error mode
etc.
3. Push <more> to select menu Parameter / Product Data.
3.1 Push <sk1> to select Parameter menu and <more> to call parameters sequentially. Select product: 1 Detector (code “0” or “23” and isotope) Select “none” Measuring path in product “1” Select Measure mode (i.e. continuous) Enter time constant (e.g.: 300 s) Disable rapid switchover “Off” Disable Interference Radiation “OFF”.
Define max. count rate (e.g. 100000 cps) Define min. count rate (e.g. 0 cps)
Product selection Select “external” or “internal” (only if different products are to be is measured).
Push <done> to return to menu group.
28
Page 35
01/10 347 LB 444 K-40
3.2 Push <sk2> to select Product Data and <more> to call parameters sequentially. Select product: “1”
Select current output 0 –20 or 4 – 20 mA Define current output limit values 0/4 and 20 mA: x% K20 = 4 mA y% K20 = 20 mA Define behavior of the current output in case of error message. Define relay 2: Function (Min), switch point (e.g. 10% ) Hysteresis (e.g. 5%) Define relay 3: Function (Max), switch point (e.g. 20% ) Hysteresis (e.g. 5%) Periodic printout: e.g. 10 min Push <done> to return to menu group.
7.2.2 Calibration
Prerequisites
Device and detector have been installed. The device must be turned on at least 1 h prior to the start of calibration. Container or pipeline have to be filled with the product to be measured. The calibration is performed as follows: At different K
memory. The concentration values determined in the laboratory (analysis values) are entered in a further storage sequence. The correlation of the memories is described below using one calibration data set as an example.
At least 2 calibration points must be available. The calibration points should be distributed fairly equally over the entire measurement range. However, some calibration points may lie outside the measure- ment range. A calibration only with points outside the measurement range is not advisable.
Its also possible to use the background radiation as a calibration point. This has to be measured
d) with empty vessel or
0 concentrations, the counts supplied by the detector are read into the
2
Count rates:
Concentration K20 (lab values):
1: 405 10.2 2 : 583 17.3 3 : 785 25.4
10 : 0 0
29
Page 36
01/10 347 LB 444 K-40
e) with product in the vessel which does not contain any potassium or f) dismounted detector arranged on a concrete floor or min 20 mm lead or 40 mm
steel plate in front of the radiation window.
A calibration with several very closely adjacent concentration v alues can easily result in an incorrect calibration curve.
Calibration steps:
Select menu Calibrate / Live Display with <more>. Push <sk1> to call the Calibrate submenu. Select product. Select “no” for suspension measurement  more Select % K20. more
Push <sk1> to call the Data input submenu. Background for rod detector N/A: continue with more Deselect Calibr. Data transfer with “no”
1. Read in 1. Rate: cursor appears in the row “1. Rate = cps”. Push <run>. The count rate is read in. Wait until the measured value has become stable
(approx. 100…300 s). While reading-in the count rate, the concentration of the products must not change. While reading in the count rate, take a sample of the product from the pipeline or the container and determine the concentration of the sample in the lab. If the product is very inhomogeneous, take several samples in quick succession and calculate an average value from their concentration values.
2. Stop measurement by pressing the <run> button again.
3. Accept result with <enter>. Cursor jumps to next row.
4. Enter the concentration value of the sample(s) determined in the lab in % K
0 and
2
confirm with <enter>. Cursor jumps to third row. Push <enter> to skip the current product temperature.
Further calibration points:
Repeat steps 1. – 4. for other different concentrations. The values must always be available in pairs. If the associated concentration value is
missing, the count rate is automatically assigned the concentration value “0”. All unused calibration points must contain “0” as count rate. If “0” is already present,
push <more>. Otherwise, delete the values with <clear> and confirm each with <enter>. Value pairs where only the count rate but n o concentration is entered, will result in an incorrect calibration curve. The entered data can be corrected as needed. The calibration process is carried out in the menu item Calculate.
Push <sk2> to select Calculate. Select calibrate mode
30
Page 37
01/10 347 LB 444 K-40
10 15 20 25 30
linear <enter>.
The calibration is carried out. The coefficient is calculated by the program. The calculated zero count rate Io and the coefficient a1 and the square error are displayed.
If necessary, enter a factor, by which the measured values are multiplied. If necessary, enter an offset which is added to the measured values.
Push <done> to return to submenu group. If the current output limits change (see page 26), calculate the calibration curve once more. (Pos. 16.). When entering the concentration values, you need not observe a falling or rising order. However, you must make sure that the correlation of count rate/concentration value is correct.
7.2.3 Checking the Calibration
To discover errors during calibration, the calibration curve can be plotted, e.g. with EXCEL.
Calibration curve
900 800 700 600
cps
500 400 300 200 100
0
% K2O
Figure 12: Calibration curve
Thus, errors in the calibration (marking) can be detected more easily. Remedy: Correct or clear values and calculate calibration curve once more.
7.3 Live Display
Start measurement: Push <run>. Stoop measurement: Push <run> again.
Result display % K20:
31
Page 38
01/10 347 LB 444 K-40
Push +….- to call further value
• current measured value in the selected unit
• averaged count rate (averaging according to entered time constant)
• actual count rate
• display of the adjusted high voltage
7.4 Correcting the Results: Addition and Multiplication
This option should not be regarded as a substitute for careful calibration . It should only be used for minor corrections. Otherwise we recommend doing a re-calibration.
An additive constant can be entered at Offset, a multiplication factor at Factor Calibrate menu, submenu Calculate).
7.4.1 Additive Constant
The value stored under Offset is added to the concentration value calculated on the basis of the count rate. This allows a parallel shift of the calibrati on curve.
Example: Measuring range 10-30% K20 It is found that the measuring values are too low by 2% K20.
Remedy: Enter “2.0” at Offset. All measured values will be increased by 2.0% K20.
Instead of 10% K20, 12% K20 is now indicated and instead of 30% 32% K20. If the measured values are too high by 2% K20, you have to enter “- 2.0”.
7.4.2 Multiplication Factor
Each measured value is multiplied by the value stored as Factor. This allows you to change the gradient of the calibration curve.
Example: Measuring range 10 - 30% K20 If you enter 1.1 at Factor, 11% K20 will be display ed instead of 10% K20. At 30% K20
now the value 33% K20 appears. Thus, even the low point of the measuring range is increased, provided it is not at 0.
Remedy: This can be corrected by entering the appropriate data under Offset and Factor:
Example: For 10% K20 the indicated value is correct, instead of 20% K20 th e reading should show
22 K20. Input value for Factor:
32
Page 39
01/10 347 LB 444 K-40
−
−
−=−=−
2%102%22
LH
minmin
=
F
−
Input value for Offset:
H = upper value of the measuring range in g/cm H = lower value of the measuring range in g/cm
The same is true if upper and lower point of the measuring range are to be changed by different values.
Example: L
actual
L
nominal
H
actual
H
nominal
K = 12 – 11*1.0666 = 0.2674
= FLLK
min
= 11% K20
= 12% K20 = 26% K20 = 28% K20
−
1228
=F
=
−
1126
alnoalno
LH
=
actualactual
actualalno
0666.1
−
OKOK
2%102%20
OKOK
22.1*1010*
3
3
2.1
=
7.5 Automatic Measuring Time Switchover
To quickly update the measured value if sudden concentration changes occur, you can enable automatic measuring time switchover (Rapid switchover on the menu Parameter, see chapter 6.36.4).
The smallest value that can be entered is 8 s. If a sudden concentration change occurs which exceeds the entered threshold (Sigma),
the entered time constant is reduced to 1/10 of the original value, but not less than 0.8 s. Thus, the output signal adjusts much quicker to the new concentration value. The changed time response of the output signal is illustrated in Figure 13.
33
Page 40
01/10 347 LB 444 K-40
Response
100
80
1 Density value
60
40
Display and output signal in %
20
0
-3 2 7 12 17 22 27 32 37 42 47 52 57 62 67 72 77 82 87 92 97
Figure 13: Response
2 Signal trend with 3 Signal trend without automatic switchover of time constant
T in s
When rapid switchover is in effect, the measurement first works with the defined basic time constant. If sudden concentration changes occur which exceed the Sigma band, the system switches to 1/10 of the time constant and reading and output signal can follow the concentration change much quicker. At the same time, statistical variations increase by a factor of about 3. When the concentration value becomes stable again, the system automatically switches back to the longer basic time constant after a time period that corresponds to the basic time constant.
7.6 Batch Measurement
If the measurement is carried out on or in a container which is filled and emptied cyclically, the measurement can be carried out in the batch mode. With the continuous measurement, the measured value at empty container is 0% K2O. After the container has been filled, it takes a very long time until the new measured value is reached again. In the batch mode the measurement can be started when the container has been filled. The earlier available measured value is deleted and the measurement is started new. Thus, the new measured value is available in a much shorter time.
The batch mode is selected via the menu item “Measure mode” “Batch mode”.
Variable measuring time:
Enter “0” as time constant. The measurement is started in the batch mode by pushing the <run> button and stopped by pushing the Sample <run> button again
or by connecting the terminals 26a and 26c.
Fixed measuring time:
Enter a value > 2 as time constant.
34
Page 41
01/10 347 LB 444 K-40
The measurement is started in the batch mode by pushing the <run> button and stops automatically when the entered time is over
or or briefly (>1s) connecting the terminals 26a and 26c.
Once the entered time is over, the measurement is stopped automatically. In both cases, the current signal remains on the last value until a new measuremen t is
started.
7.7 Radiating Interference Detection
Select the function Radiating interference (menu Parameter, see page 24) to suppress interfering radiation (e.g. weld seam testing).
Caution:
When this function is on (Radiating Interference <ON>), the automatic switchover of the time constant (function Rapid switchover) is always off. When turning this function on, you have to define a Sigma value, which defines a window for the measuring signal. If the detector signal suddenly exceeds this threshold
• the measurement stops
• the error message “Interference Radiation” appears on the display.
The measured value and the current outputs are held on the last value. Reset the error message and start the measurement via the keyboard by confirming the error message with <enter>; then start the measurement with <run> or externally the digital input.
Note: Enter n > 5 to rule out false alarms with sufficient statistical safety.
For calculation it holds: Sigma =
Is/
Example: Count rate I
= 300 cps, n = 6
m
= Im + n * Im
I
S
= 300 + 6 *
I
S
= 404 cps
300
Thus, an alarm is signaled as soon as I/s exceeds the value of 404 cps.
via
Note:
Due to the dynamic behavior of radiating interfering detection, a quick increase of the pulse rate due to operative factors (e.g. very fast emptying of the vessel or major short­term concentration changes caused by stirrers) may be interpreted as interfering radiation. For example, opening the active beam channel on the shielding can also cause a quick increase of the pulse rate. You have to reset the alarm that is then triggered, or better, do not enable radiating interfering detection at first. Enable radiating interfering detection only after calibration has been performed.
35
Page 42
01/10 347 LB 444 K-40
7.8 Error Messages
7.8.1 Acknowledging Error Messages
All error messages must be reset with <enter>. Several simultaneously or consecutively occurring errors are stored in an error register in
the order of their appearance; they have to be reset individually by pushing <enter> several times.
All errors are signaled via the “Error” output (relay 16a/16c).
7.8.2 Error Messages during Operation
Error no.
32 Wrong
Message Cause, Remarks Remedy
password
Wrong password entered to unlock the keyboard.
Push “enter” to reset error message. Enter correct password.
7.8.3 Error Messages during Calibration
Error no.
10 Not calibrated Calibration incomplete Perform calibration 36 Missing calib.
37 Curve not clear System could not calculate a
38 Error one-point
39 Measuring path
56 Range overflow Measuring path = 0 Enter measuring path “1”
Message Cause, Remarks Remedy
points
calibration
not defined
Not enough points entered for selected calibration mode
clearly rising or falling curve.
Absorption coefficient = 0 Absorption coefficient = 0
Measuring path = 0 Enter measuring path “1”
Enter further calibration points or select another calibration mode Check calibration points for input errors. Select calibration mode “linear”.
36
Page 43
01/10 347 LB 444 K-40
7.8.4 Error Messages during Measurement
Error no.
1 Rate Overflow Count rate > 520000 High voltage wrong.
2 No Detector Rate Detector faulty Replace detector 3 Wrong HV Detector faulty Replace detector 4 Pt 100 wrong
5 4 –20 mA faulty Input current below 4 mA,
6 Detector
7 Trouble with
8 Measurement
9 Detector
10 Not
11 Power fail > 1
12 No input allowed Input locked via password Enter password 47 Rate < Minimum Enter actual count rate < than
48 Rate > Maximum Enter actual count rate > than
61 Replace
Message Cause, Remarks Remedy
Replace detector
temperature
temperature > 65°C
detector communication
stopped
temperature > 80°C
calibrated
month
photomultiplier
Temperature measurement activated and no Pt100 connected. Cable faulty.
evaluation unit faulty. Detector temperature too high, detector faulty
Cable detector – evaluation unit interrupted or too long or cable resistance too high, detector faulty evaluation unit faulty. Measurement process stopped via digital input or measurement stopped because count rate threshold has been exceeded or not reached. Detector temperature too high, Detector faulty Measurement was started without calibration Check date after long period of power failure
entered min. count rate: shielding container closed? detector fault?
entered max. count rate: detector fault? pipeline empty and shielding container not closed? High voltage increased too much.
Check cable, replace detector or Pt 100. Disable temperature measurement. Check input current, replace, evaluation unit. Cool detector, if necessary, replace it
Replace cable or use bigger wire cross-section.
Replace detector Replace evaluation unit. Open digital input. Disable count rate thresholds to identify fault.
Replace detector
Perform calibration
When the If this error message appears already after a brief period of power failure: replace Li-battery.
Replace detector Open shielding container
Replace detector Close shielding container
Check plateau. Possibly replace photomultiplier.
37
Page 44
01/10 347 LB 444 K-40
Message Cause, Remarks Measurement halted <HALT>
Measurement process stopped via digital input or measurement stopped because count rate threshold has been exceeded or not reached.
Rate Overflow Detector faulty. High voltage outside of the
plateau.
Power Fail Complete failure or drop of power supply below
tolerance level.
No Detector Rate *)
Detector supplies no pulses for at least 60 s.
Wrong HV *)
HV reference voltage of detector too high or too low. Measurement is stopped when this error occurs and has to be restarted by pushing the <run> button. Measurement is not stopped. Relay “Failure” is de-energized.
Error temperature measurement Temperature input faulty (see chapter
“Temperature Compensation” and “Service
Instructions”). Power failure > 1 month *) Enter New Date
After a long period of power failure you have to
check and, if necessary, correct the date to
ensure correct decay compensation.
Even if the date is correct, you have to enter
one digit of the year and confirm the input with
<enter>.
*) When these error messages are displayed, the measurement stops automatically and has to be restarted by pushing the <run> button.
38
Page 45
01/10 347 LB 444 K-40
7.9 System Start/Stop
To stop operation of the measuring system, proceed as follows: Turn system off. The water cooling device has to be used if the detector temperature may rise above 50°C
even though the instrument is not in operation. In case of sub-zero temperatures, empty the water cooling.
To start operation of the measuring system: Put water cooling in operation again, if it was disabled. Turn instrument on. If instrument has been disabled for more than one month, the error message “Power
failure > 1 month” is displayed. In this case, check and, if necessary, correct the date and time again in the menu sk1:
General Data.
39
Page 46
01/10 347 LB 444 K-40
8 Technical Data
8.1 Evaluation Unit LB 444 K-40
Assembly: 19” module 3 HE, 21 TE; protection type IP 20 Power supply: 115 V ± 10%, 230 V ± 10%, 18-32 V AC/DC Power consumption: approx. 30 V or 30 W Temperature range: Operating temperature: 0 - 50 °C; no condensation
Storage temperature: - 40 to + 70 °C CPU: 32 bit computer, data storage in FLASH- E- PROM. Weight: approx. 2 kg Display: LCD display with 4 x 20 characters, illuminated .
Data input via keyboard. Softkey guided dialog; Languages: German, English Data editing possible only after entry of a freely selectable password .
Time constant: 0.5 – 9999 s with automatic reduction to 1/10 of the
values in case of sudden concentration changes (can be turned off).
Adjustable response threshold for time constant
switchover. Automatic decay compensation: for Inputs/Outputs: Detector connection Ex II (2) G [EEX ib] II B or Ex II (2) G [EEX ib] II C Digital inputs: 2 digit a l inputs for external product selection
1 digital input for external start/stop of measurement: Stop of continuous measurement or
Restart after interruption of measurement due to
Analog output: 0/4 - 20 mA isolated (max. 500 Ω) for measured value Limit value outputs: 2 relay outputs for measured value max. / min.,
1 relay output for collective failure message Loading capacity: AC: Max. 250V, max. 1A, max. 200VA
241
Start/Stop in batch mode,
interfering radiation.
detector temperature, mass pulse, correlation freely
selectable
Am,
137
Cs, 60Co,
244
Cm, 85Kr and 90Sr .
40
Page 47
01/10 347 LB 444 K-40
DC: Max. 300V, max. 1A, max. 60 W non-reactive load Current input: 0/4 - 20 mA isolated
for input of temperature signal or
volume current signals
Interfaces: RS 232. Port for data transfer from evaluation unit to printer
or PC
8.2 Detectors
Type
LB 5401-03 50/50 18 0 No LB 5430 150/150 52 23 No
Protection type: IP 65,
Housing: Stainless steel housing
Cable entry: M16 external cable diameters 7...10 mm for connection cable
Temperature range:
Operating temperature:
Storage temperature:
Crystal ∅ / l in mm
detector - evaluation unit and M 12 for external cable diameter
7...10 for connection of Pt 100
- 20 to + 50°C. A water cooling device is availab le for higher temperatures. Monitoring of detector temperature and alarm if max. permissible temperature is exceeded.
- -40 to + 70 °C
Weight kg
Detector code Ex protection
Connection cable evaluation unit - detector
Cross-section in mm² Max. cable length in m
1 1000
1.5 1500
2.5 2500
41
Page 48
01/10 347 LB 444 K-40
9 Service Instructions
9.1 General Safety Precautions
Any time you are working on electrical components, you have to observe the relevant safety regulations. Please refer to the Safety Summary at the beginning of this operating manual.
9.2 Evaluation Unit LB 444 K-40
The evaluation unit is provided with an error control which also monitors the detector functions. An error code appears on the display if an error is detected. For information about the error and possible causes please refer to the error code list. In case of hardware errors the evaluation unit has to be replaced. If no error code is displayed, the electronics is working properly and all measured values and parameters lie in the normal possible range. Possible function problems must then have another cause. “Total Reset” should be performed only in exceptional cases and after consultation with Berthold. To perform a “Total Reset”, turn the instrument off; then push the Clear button while you turn the instrument on.
All entered date will be deleted and the default values restored.
Error code list
No. Problem Cause
1 Rate Overflow Count rate > 520000cps 2 No Detector Rate Error in detector or during transmission 3 Wrong HV HV < 500V or > 1500V, outside control range 6 Detector temperature >
65°C
7 Trouble with detector
communication 8 Measurement stopped Measurement stopped by closed contact 22 a/c 9 Detector temperature >
80°C 10 Missing calibration Calibration has to be terminated with <enter> and
11 Power fail > 1 month Instrument has been turned off for more than one
12 No input allowed Device locked by password or wrong password entered 33 Wrong password Wrong password entered 36 Missing calib. points Not enough calibration points available 37 Curve not clear Calibration not steady 39 Measuring path not defined Measuring path not yet entered 46 Radiating interference New start with <run> or via input 3 (22a/22c)
Detector temperature has exceeded > 65°C
Data transfer between detector and evaluation unit faulty
Detector temperature has exceeded > 80°C
<more>
month. Enter new date
42
Page 49
01/10 347 LB 444 K-40
50 I.Max/I.Min value Current output entered incorrectly
51 Current input error Current input range entered incorrectly
52 Current output limit value Current output range entered incorrectly
53 Current input error Current input too high > 20mA
54 Device not calibrated Calibration has to be terminated with <enter>
56 Range overflow Input value outside permissible range
57 Wrong input value Input value outside permissible range
58 Calibration faulty Wrong calibration
60 Digital inputs for product
selection allowed
Window for product selection in menu Parameter set to external.
Problem Possible Cause Potential Solution
No display No power supply Check mains cables
Check fuses
Display unreadable Processor error Note error code
Perform total reset
Replace evaluation unit No pulse rate (Error code 2)
Power supply for detector interrupted
Check cable to detector
Detector is faulty Replace detector Result reading shows drifts
Detector stabilization faulty
Replace detector
Photomultiplier faulty Replace photomultiplier Decay compensation
activated
43
Page 50
01/10 347 LB 444 K-40
10 Service
10.1 Service Menu Overview
44
Page 51
01/10 347 LB 444 K-40
sk1: Service Menu sk2: Mass Flow
sk1 sk2 more
sk1: Test calculate sk2: I / O - Test sk1 sk2 more
Testrate 12345 cps Output Curr : 10.0 mA
1.2345 g/cm3
Unit Mass Flow [kg/h, t/h] t/h
Range input {at 20 mA] Value: 100 m3/h
Flow Value 87.00 m3/h
Ext. Mass Counter Tons per cycle
OFF/0,01/0,1/1/10
Erase mass ­integrator no/ yes
Set output cur Current : 9.0mA
View input cur. Current: 12.0mA
Relay 1: ON Relay 2: OFF Relay 1: ON
Input 1: OFF Input 2: OFF Input 3: ON
sk1: HV Adjustment sk2: Status request
sk1 sk2 more
High Voltage Value : 800 V
Save Default
Value: 850 V
Read Default
Value: 850 V
Temperature = 45.2°C Det. Temp. = 28 °C Counts = 3210/s
For Detector - Reset please press (sk 1 & clear) sk1
sk1: Reset Detector sk2: Pt 100 adjust
sk1 sk2 more
For Pt 100-Adjust please press (sk2 & clear)
sk1: Measure Plateau sk2: Request Plateau
sk1 sk2 more
Plateau measurement
running ......
IIIIIIIII...........................
HV( 500): 817/s HV( 560): 884 /s HV( 620): 923/s + - done
sk1: Plateau sk 2: Adj. Current out sk 1 sk2 more
Current 1.8 mA Offset: 4123
-
Current 18 mA Offset: 345
-
Curr. Input Adjust Set value 1.8 mA Real value 1,7 mA + - more
Curr. Input Adjust Set value 18 mA Real value 17.5 mA + - more
sk1: Adj. Cur. input sk 2: Relay Delay time sk 1 sk2 more
Relay delay Value: 50 ms
45
Page 52
01/10 347 LB 444 K-40
10.2 Service Menu
The Service menu functions support input of test values (output current, count rates) to simulate instrument functions
• check of outputs and inputs adjustment of analog outputs and inputs
adjustment of Pt 100 input (N/A)
• plateau recoding of detector
• manual setting of high voltage for test purposes.
The individual functions are described below.
Test calculate:
Any pulse rate within the calibration range can be entered. The count rate supplied by the detector will then be ignored. The respective output current is calculated from the count rate and the concentration value. Output current and the respective concentration value are displayed as a result of the defaulted pulse rate.
The product temperature supplied via Pt 100 or as current signal is taken into account when calculating the measured value. Therefore, the temperature compensation for the test has to be disabled. To do this, select “Density without TC”.
After completion of the test, reset the test count rate to “0”!
I/O Test Output Current:
To use this function, the measurement must not be in the RUN mode. To test the connected instruments, you can set any output current between 0 – 22 mA.
View input current:
Shows the current measured at the current input.
Relay 1/2/3:
The relays can be switched to test the signal circuits (ON = relay energized)
Input 1/2/3:
Shows if the inputs are open or closed. (OFF = input open)
HV Adjustment: Set HV detector
With the standard setting “0” the detector works with automatic high voltage control. If you enter a high voltage value for special tests, the high voltage control is disabled. Th e entered value is used as fixed high voltage. In this case, the info HV-Test is displayed in the measurement menu or the live display (instead of HV auto).
Set this value to “0” again when you exit the Service menu. Otherwise, the measurement device will not work correctly.
Status request
This menu item includes the detector temperature current counts high voltage.
46
Page 53
01/10 347 LB 444 K-40
Reset Detector:
To reset the detector you have to push the buttons <sk1> and <clear> at the same time. This will interrupt the plateau recording and reset the detector electronics. A reset also has to be performed when the microprocessor of the detector has been replaced.
Pt 100 Adjust (N/A)
The Pt 100 input is adjusted by the manufacturer. New adjustment is necessary only when detector boards or the microprocessor of the detector have been replaced. Apply resistance with 100 Ω (with lowest tolerance) at Pt 100 input of the detector (terminals 3 and 4,). Select menu item “Pt 100 Adjust” Push <sk2> and <clear> at the same time. The message “Pt 100 adjustment running...” is displayed. The adjustment is finished after approx. 5 s. Then connect resistance thermometer to detector, terminals 3 and 4. Pt 100 adjustment should be performed by trained personnel only!
10.2.1 Plateau Check
Measure Plateau
A detector plateau can be recorded automatically in steps of 60 V, e.g. to check the proper function of the NaI crystal multiplier combination. Working principle of scintillation counters The detector (scintillation counter) is comprised of: a) an organic scintillator (PVT) or b) a sodium iodide crystal doped with Thallium c) photomultiplier d) electronics unit
As the gamma radiation hits the detector, it triggers minute flashes of light in the crystal which are invisible to the naked eye. These flashes of light release electrons from the light-sensitive cathode (photo cathode) of the photomultiplier. This electron current is amplified via a dynode system resulting in an electrical impulse at the photomultiplier output for each light flash. The high voltage required for operation of the photomultiplier is generated in the electronic unit. The scintillation counter is equipped with a drift compen sation w hich compensates f or the effects of aging and temperature-related gain variations, thus ensuring high long-term stability. Malfunctions of the scintillation counter are not alw ays indicated by a missing pulse rate. It is also possible that the specific Gamma sensitivity appears to have changed or obvious instabilities are apparent. This error can be detected only by means of a plateau check. The evaluation unit LB 444 K-40 includes a function for automatic plateau recording. Please keep in mind: The scintillation counter may remain at the measuring point if it is certain that the concentration in the measuring path will remain absolutely constant during the time of the check (approx. 10 min.). Otherwise, the check has to be carried out in the workshop. An appropriate amount of potassium salt (10 kg) has to be provided directly in front of the detector. Select Plateau / Measure Plateau in the Service menu. Upon completion of the plateau measurement, you may invoke the function Request Plateau to display the value pairs HV-steps/count rate. The detector works properly if a clear plateau is visible; the position of the plateau is
47
Page 54
01/10 347 LB 444 K-40
irrelevant.
16000 14000 12000 10000
8000
cps
6000 4000 2000
0
500 600 700 800 900 1000 1100 1200 1300
HV in V
Figure 14: Plateau curve with Na I crystal (LB 5402)
Plateau LB 5430
50000 45000 40000 35000 30000 25000
cps
20000 15000 10000
5000
0
560 660 760 860 960 1060
Setpoint of the HV control
HV in V
Figure 15: Plateau curve with organic scintillator (LB 5430)
Plot the count rate vs. the high voltage, as shown in the diagram above. If the count rate in the flattest part of the curve changes
a) by more than 5% per 100 V high voltage for the LB 5402, b) by more than 15% per 100 V high voltage for the LB 5430,
the scintillation counter will operate unstable. The detectors or at least the crystal
48
Page 55
01/10 347 LB 444 K-40
multiplier combination have to be replaced.
Request Plateau
The value pairs of the plateau measurement may be displayed here and may be plotted to assess the curve.
Adjust current output
Current 1.8 or 18 mA: Adjustment of output current to the values 1.8 mA and 18 mA by incrementing or decrementing the numerical value on the display.
Adjust current input
Target value/Actual value Adjustment of input current. Feed current of 1.8 mA. Confirm display with <enter>. RUN appears on the display, the adjustment is running. Push <more> to stop the adjustment and continue with the next item. Then feed 18 mA current and confirm with <enter> and push <more> to continue.
49
Page 56
01/10 347 LB 444 K-40
10.3 Detector
10.3.1 Checking the Crystal-Multiplier Assembly
A plateau that is too small or too steep indicates faults in the crystal-multiplier assembly. Please proceed as follows to perform a visual inspection of crystal and multiplier:
Switch off the scintillation counter before opening the instrument. Do not perform the check in bright daylight, as this may damage the photo
cathode of the photo multiplier.
Open the scintillation counter by removing the cover of the connection box first and then the screws of the base. The entire electronics (1) with the crystal/multiplier combination can then be detached from the housing. Remove photo multiplier combination from base (2) and unscrew ring nut (3) on front panel. The multiplier (4) including crystal (5) can now be detached from the Mu­metal shielding (6). A thin layer of silicone oil between crystal and mu ltiplier ensures optical coupling; silicone oil is rather viscous, particularly at low temperatures. Carefully detach the crystal from the multiplier window by gently sliding the crystal sideways. Do not wipe off the silicone oil layer if no new silicon oil is available! Check: The crystal must be perfectly clear inside and not show any dull areas. Its typical color is white. A yellowish to brownish coloring is a sign of thermal overload and indicates that the crystal has to be replaced. The surrounding white reflecting la yer must not be damaged. You can only check the photo multiplier for glass breakage or other mechanical damage. Other faults cannot be identified by visual checks alone. However, if the crystal does not show any faults, a bad plateau indicates that the photo multiplier is faulty.
1 = electronics part with
(a) CPU,(b) HV, (c) voltage divider 2 = base 3 = ring nut 4 = photomultiplier 5 = crystal 6 = Mu-metal shielding
(c) (b)
(a)
Figure 16: Assembly of the scintillation counter
3
5
4
6
2
1
50
Page 57
01/10 347 LB 444 K-40
The multiplier window is coated with a vapor-deposited layer acting as photocathode.
This layer gives the window a brownish tint similar to smoked glass. If this layer is no
longer present or if it is stained, then the photocathode has been destroyed (e.g. by
overheating, glass breakage, or incident light), and the multiplier must be replaced.
Faults caused by damage to the dynode systems (e.g. by excessive vibration) cannot be
identified by appearance. If in doubt, replace the multiplier .
The glass pane at the mating face to the photo multiplier must not show any cracks.
Before re-assembly, apply a drop of clean silicon oil between crystal and multiplier and
distribute it evenly by gentle rubbing to ensure sound optical connection between both
components. Attach the Mu-metal screen and fix it with screws, making sure that the
Mu-metal screen is only under light tension.
Re-assembly:
Set the crystal again onto the front face of the photo multiplier and twist both compo-
nents several times counter-clockwise. If you have to replace a component, apply 1 or 2
drops of silicone oil on the mating faces of the new and old component. It may also
suffice to press the new component against the mating face of the old one and to turn it
several times. Reassemble parts properly.
Just as after replacement of a complete detector, the calibration should be checked
immediately.
10.3.2 Testing and replacement of the photomultipliers in the
detector LB 5430
Fig­ure 17: As-
1
sembly of the scintillation counter
The organic scintillator is in the front housing part of the detector LB 5430 (1 in Fig. 17 ).
This housing should only be opened by the service staff from Berthold.
The detector electronics and the photomultiplier are located in the rear housing part (2 in
Fig. 17 )..
The same rules and criteria apply as for the detector LB 5430 regarding dismantling and
inspection.
The detector must stand upright on the flange for dismantling and installation of the
detector electronics. An installation of the electronics in horizontal position of the
detector is very difficult. The multiplier can get damaged.
1. Dismantle electronics with photomultiplier (see above)
2. Check photomultiplier and replace if necessary.
3. Put 1-2 drops of silicone oil on the frontal area of the ph otomultipliers before assembly
2
51
Page 58
01/10 347 LB 444 K-40
and rub it in gently with your finger.
4. Guide the electronics and the photomultiplier back into the housing until the photomultiplier stands up on the ground of the detector housing. Turn the electronics without exerting pressure several times at about 90 ° back and forth. This distributes the oil on the frontal area of the photomultipliers evenly.
5. Screw the electronics on at the base.
6. Screw on lid.
10.4 Replacing the Evaluation Unit LB 444 K-40
To replace the entire LB 444 K-40 evaluation unit, please proceed as follows: Connect new evaluation computer. In submenu Calculate (menu Calibrate), select the calibration mode (K2O) . Enter data using the set-up protocol and date of the original set-up. Start calibration process: select calibration more with <enter> and then push <more>. Start measurement again with <run>.
Example: Calibration point 1 405 10,2 01.02.96 2 583 16,9 01.02.96 3 785 22,5 01.04.96 4 558 22,1 01.04.96 5 703 19 01.10.96
Count rates % K2O Date
52
Page 59
01/10 347 LB 444 K-40
10.5 Setup Protocol
Measuring Point ............................................ Date: ...........................
Product: ........................................................
Parameter Range Value
Password
Instrument ID
Program version
Configure instrument
Radiation measure yes / no
Error mode Hold/continue
Product no. 1/2/3/4
Measuring path cm
Measure mode
Time constant s
Rapid switchover ON/OFF
Sigma for radiating
interference
Count rate limit (min) cps
Count rate limit (max.) cps
Current out for concentration 0-20mA or 4-20mA
0/4 mA
20 mA
Current output error Hold Value: 0-22mA
0-10
-1
-1
53
Page 60
01/10 347 LB 444 K-40
Parameter Range Value Relay 2 setup Min or Max: Hysteresis : 0-10% Relay 3 setup Min or Max: Hysteresis : 0-10%
Data input:
Count rate cps
Concentration (lab values) Date Temperature °C
1 2 3 4 5 6 7 8 9 10
Calculation:
Parameter Range Value Calibrate mode on/lin/squ/cub/auto Result a1 Background count rate Io Coefficient a1 Coefficient a2 Coefficient a3 Square error Factor 1 to 10 Offset
54
Page 61
01/10 347 LB 444 K-40
11 Dimensional Drawings
11.1 LB 5402
Weight approx. 6 kgs
Cable gland M16 for cable dia. 7-10 mm
approx. 40
dia. 105
346
208
dia.
75.5
11.2 LB 5430
LB 5430 Weight appr. 52 kg
55
Page 62
01/10 347 LB 444 K-40
11.3 LB 444 K-40
19” module
Installation in 19” rack
Installation in wall housing
56
Page 63
01/10 347 LB 444 K-40
2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a
32c2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a18a
2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a
30c 2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a32a
2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a
28c 2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a30a
2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a
26c 2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a28a
250/125 V ~
230/115 V ~
~~
0/4-20mA
potential
+ -
0/4-20mA
potential
+ -
freier
Eingang
freier
Ausgang
Isolated
Isolated
Netz / Line
24 V -
input
output
2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a
RS RS
Start /Stop
Ext.
Ext.
Batch
Produkt
Produkt
2 1
RS 485
Stop
Messung
Measur
ement
Product
2
0 V (+ 24 V)
Product
1
Fehler
Min.
12 3
Failure
Max.
Det./
Temp.
Digitale Eingänge
Digital Inputs
24c 2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a26a
2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a
22c 2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a24a
2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a
20c 2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a22a
2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a
18c
2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a
16c 2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a20a
2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a
14c 2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a16a
PA
2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a
12c 2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a14a
Rel.2 Rel.3 Rel.1
Min.
Max.
Det./
Temp.
Impuls
521
count
Pt 100
34
LB 44../LB 54..
Probe
Detektor
Für eigensichere Installationen
[EEx ib] Kabeladern mit 10 mm
Schrumpfschlauch überziehen
10 mm long heat shrink sleeving
must be used to comply with
the [EEx ib] safety certification.
NYMHY 2*1mm² Kabel/Cable:
+ 24 V
2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a10a 2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a10c
2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a 2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2c 2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a2a12a
+ -
Detektor
Probe
Anschlussplan / Connection diagramm LB 444 19 " Rack
57
Page 64
01/10 347 LB 444 K-40
NYMHY 2*1mm²,max. 1000 m
Detektor / Probe
65
64 63
Det. Temp. Min. / Max.
Kanal B
Channel B
Fehler / Failure
Start / Stop Batch /Reset
Product Ext.
potentialfrei /isolated Stromausgang / Current output
RS 485
0/4 - 20 mA Stromeingang / C urrent input
Masse / Ground
Aux. Voltage Hilfsspg. Ausg.
62
48
Rel.2 Rel.3
47
Rel.1
46
45
44
+ IN IN
43
42
RS 1 RS 2
39 40 41
37 38
GND
+ 24 V
36
GND
+ 24 V
35
60 61
59
58
57
51- 56 50
49
Relais
3
Für eigensichere Installationen
[EEx ib] Kabeladern mit 10 mm
Schrumpfschlauch überziehen
2
Digital inputs
1
Digitale Eingän ge
10 mm long heat shrink sleeving
Kabel/Cable:
Probe
Detektor
LB 44../LB 54..
54321
PA
must be used to comply with
the [EEx ib] safety certification.
Anschlussplan Wandgehäuse LB 4460 für Dichtemessung LB 444
Wiring Diagram Cabinet LB 4460 f or Density Measuring LB 444
NYMHY 2*1mm²,max. 1000 m Kabel/Cable:
Detektor / Probe
34
32 33
Pt 100 Pt 100
54321
must be used to comply with
the [EEx ib] safety certification.
+ -
Rel.3
Rel.1 Rel.2
IN IN IN
+
RS 1 RS 2
GND
+ 24 V
GND
+ 24 V
PE
31
16 17
29 30
15
28
13 14
26 27 12 11 10
9
678
20-25
1918 5
4
1 2 3
Relais
123
Digitale Eingänge
Für eigensichere Installationen
[EEx ib] Kabeladern mit 10 mm
Schrumpfschlauch überziehen
10 mm long heat shrink sleeving
Digital inputs
Det. Temp. Min. / Max.
Fehler / Failure
LB 4460 Wandgehäuse / wall mounted cabinet
Start / Stop Batch /Reset
Product Ext.
potentialfrei /isolated Stromausgang / Current output
RS 485
0/4 - 20 mA Stromeingan g / C urrent input
Masse / Ground
Aux. Voltage Hilfsspg. Ausg.
24 V AC/DC 230/ 115 V AC
Kanal A
Channel A
Netz / Power
Probe
Detektor
LB 44../LB 54..
PA
58