Bender ISOMETER IRDH375, ISOMETER IRDH375B User Manual

ISOMETER® IRDH375

Insulation monitoring device for IT AC systems with galvanically connected rectifiers and converters and for IT DC systems IRDH375: Software-Version D0183 V1.8 IRDH375B: Software-Version D0184 V1.8

IRDH375B

IRDH375_D00124_05_M_XXEN/01.2020
Manual EN
Table of Contents
1. Important information .................................................................................... 7
1.1 How to use this manual ................................................................................. 7
1.2 Technical support: service and support ................................................... 8
1.2.1 First level support ............................................................................................. 8
1.2.2 Repair service ..................................................................................................... 8
1.2.3 Field service ........................................................................................................ 9
1.3 Training courses ............................................................................................. 10
1.4 Delivery conditions ....................................................................................... 10
1.5 Inspection, transport and storage ........................................................... 10
1.6 Warranty and liability ................................................................................... 11
1.7 Disposal ............................................................................................................ 12
2. Safety instructions ......................................................................................... 13
2.1 General safety instructions ........................................................................ 13
2.2 Work activities on electrical installations ............................................. 13
2.3 Device-specific safety information ......................................................... 14
2.4 Intended use ................................................................................................... 16
2.5 Directions for installation ........................................................................... 17
3. Function ........................................................................................................... 19
3.1 Common characteristics (IRDH375 and IRDH375B) .......................... 19
3.2 Additional characteristics IRDH375B ...................................................... 20
3.3 Product description ...................................................................................... 20
3.4 Function ............................................................................................................ 21
3.4.1 Self test .............................................................................................................. 22
3.4.2 Relay K3: device fault alarm and EDS common message ............... 23
3.5 Additional functions IRDH375B ............................................................... 25
IRDH375_D00124_05_M_XXEN/01.2020
3
Table of Contents
4. Commissioning flow chart (threepart) ...................................................... 30
5. Connection ..................................................................................................... 33
5.1 Wiring ................................................................................................................ 33
5.2 Wiring diagrams with coupling devices ................................................ 37
5.2.1 Connection with AGH150W-4 .................................................................. 37
5.2.2 Connection with AGH520S ........................................................................ 39
5.2.3 Connection with AGH204S-4 .................................................................... 40
6. Operation and setting .................................................................................. 42
6.1 Operating features and displays IRDH375(B) ...................................... 42
6.1.1 Display in the standard mode ................................................................... 43
6.1.2 Display in the menu mode ......................................................................... 44
6.1.3 Function keys .................................................................................................. 44
6.2 Menu structure and menu mode ............................................................ 47
6.2.1 Diagram menu structure ............................................................................ 49
6.3 Menu HISTORY INFO (IRDH375B) ............................................................ 50
6.3.1 Diagram HISTORY INFO (IRDH375B) ....................................................... 51
6.4 Menu ISO SETUP: Setting of the basic ISOMETER® functions ........ 52
6.4.1 Response values Alarm 1 and Alarm 2 ................................................... 52
6.4.2 Operating principle of the alarm relays ................................................ 52
6.4.3 Memory setting (on/off) .............................................................................. 55
6.4.4 Current output for external measuring instruments
(IRDH375B) ....................................................................................................... 55
6.5 Menu ISO ADVANCED: Setting of the extended functions ........... 56
6.5.1 External coupling devices (AGH: no) ...................................................... 56
6.5.2 Adaptation to the system leakage capacitance
(Cemax : 150 μF) ............................................................................................. 57
6.5.3 Changing the measuring principle from AMP to DC
(Measure: AMP) .............................................................................................. 57
6.5.4 Setting the repetition time for automatic self tests
(Autotest: 24h) ................................................................................................ 57
6.5.5 Setting the real-time clock (Clock) (IRDH375B) .................................. 58
4
IRDH375_D00124_05_M_XXEN/01.2020
Table of Contents
6.5.6 Setting the date (Date) (IRDH375B) ........................................................ 58
6.5.7 Specifying the starting time of the automatic self test (Test)
(IRDH375B) ....................................................................................................... 58
6.5.8 Diagram ISO ADVANCED ............................................................................ 59
6.6 Menu COM SETUP: Setting the BMS interface .................................... 60
6.6.1 Bus address „Addr:“ (IRDH375B) ............................................................. 60
6.6.2 ISOnet function (IRDH375B) ...................................................................... 60
6.6.3 ISO monitor (IRDH375B) ............................................................................. 61
6.6.4 Diagram COM SETUP (IRDH375B) ........................................................... 62
6.7 Menu PASSWORD ......................................................................................... 63
6.7.1 Activating and setting the password ..................................................... 63
6.7.2 Diagram PASSWORD .................................................................................... 64
6.8 Menu LANGUAGE .......................................................................................... 65
6.8.1 Setting the national language .................................................................. 65
6.8.2 Diagram Language ....................................................................................... 66
6.9 Menu SERVICE ................................................................................................ 66
6.10 Parameterization via Internet ................................................................... 67
7. Serial interfaces .............................................................................................. 68
7.1 RS-485 interface with IsoData protocol (IRDH375) ........................... 68
7.2 RS-485 interface with BMS protocol (IRDH375B) ............................... 70
7.3 Topology RS-485 network (IRDH375B) .................................................. 71
7.3.1 Correct arrangement ................................................................................... 71
7.3.2 Wrong arrangement .................................................................................... 71
7.3.3 Wiring ................................................................................................................ 71
7.4 BMS protocol (IRDH375B) .......................................................................... 72
7.4.1 BMS Master ...................................................................................................... 73
7.4.2 BMS Slave ......................................................................................................... 74
7.4.3 Commissioning of an RS-485 network with BMS protocol ............ 75
IRDH375_D00124_05_M_XXEN/01.2020
5
Table of Contents
8. Factory Settings ............................................................................................. 77
9. Technical data IRDH375(B) .......................................................................... 79
9.1 Data in tabular form ..................................................................................... 79
9.2 Standards, approvals and certifications ................................................ 83
9.3 Characteristic curves .................................................................................... 84
9.4 Ordering details ............................................................................................. 91
9.4.1 ISOMETER® ....................................................................................................... 91
9.4.2 Dust protection .............................................................................................. 92
9.4.3 Coupling devices ........................................................................................... 93
9.4.4 Measuring instruments ............................................................................... 93
INDEX .................................................................................................................... 95
6
IRDH375_D00124_05_M_XXEN/01.2020

1. Important information

1.1 How to use this manual

This manual is intended for qualified personnel working in electrical engineering and electronics!
Always keep this manual within easy reach for future reference.
To make it easier for you to understand and revisit certain sections in this man­ual, we have used symbols to identify important instructions and information. The meaning of these symbols is explained below:
This signal word indicates that there is a high risk of danger that will result in electrocution or serious injury if not
DANGER
WARNING
CAUTION
avoided.
This signal word indicates a medium risk of danger that can lead to death or serious injury if not avoided.
This signal word indicates a low-level risk that can result in minor or moderate injury or damage to property if not avoided.
IRDH375_D00124_05_M_XXEN/01.2020
7
Important information
This symbol denotes information intended to assist the user in making optimum use of the product.

1.2 Technical support: service and support

For commissioning and troubleshooting Bender offers you:

1.2.1 First level support

Technical support by phone or e-mail for all Bender products
Questions concerning specific customer applications
Commissioning
Troubleshooting
Telephone: +49 6401 807-760*
Fax: +49 6401 807-259
In Germany only: 0700BenderHelp (Tel. and Fax)
E-mail: support@bender-service.de

1.2.2 Repair service

Repair, calibration, update and replacement service for Bender products
Repairing, calibrating, testing and analysing Bender products
Hardware and software update for Bender devices
Delivery of replacement devices in the event of faulty or incorrectly
delivered Bender devices
Extended guarantee for Bender devices, which includes an in-house repair service or replacement devices at no extra cost
8
IRDH375_D00124_05_M_XXEN/01.2020
Important information
Telephone: +49 6401 807-780** (technical issues)
+49 6401 807-784**, -785** (sales)
Fax: +49 6401 807-789
E-mail: repair@bender-service.de
Please send the devices for repair to the following address:
Bender GmbH, Repair-Service, Londorfer Str. 65, 35305 Grünberg

1.2.3 Field service

On-site service for all Bender products
Commissioning, configuring, maintenance, troubleshooting of Bender products
Analysis of the electrical installation in the building (power quality test, EMC test, thermography)
Training courses for customers
Telephone: +49 6401 807-752**, -762 **(technical issues)
+49 6401 807-753** (sales)
Fax: +49 6401 807-759
E-mail: fieldservice@bender-service.de
Internet: www.bender-de.com
*Available from 7.00 a.m. to 8.00 p.m. 365 days a year (CET/UTC+1)
**Mo-Thu 7.00 a.m. - 8.00 p.m., Fr 7.00 a.m. - 13.00 p.m.
IRDH375_D00124_05_M_XXEN/01.2020
9
Important information

1.3 Training courses

Bender is happy to provide training regarding the use of test equipment.
The dates of training courses and workshops can be found on the Internet at www.bender-de.com -> Know-how -> Seminars.

1.4 Delivery conditions

Bender sale and delivery conditions apply.
For software products the "Softwareklausel zur Überlassung von Standard­Software als Teil von Lieferungen, Ergänzung und Änderung der Allgemeinen Lieferbedingungen für Erzeugnisse und Leistungen der Elektroindustrie" (software clause in respect of the licensing of standard software as part of de­liveries, modifications and changes to general delivery conditions for prod­ucts and services in the electrical industry) set out by the ZVEI (Zentralverband Elektrotechnik- und Elektronikindustrie e. V.) (German Electrical and Electron­ic Manufacturer's Association) also applies.
Sale and delivery conditions can be obtained from Bender in printed or elec­tronic format.

1.5 Inspection, transport and storage

Inspect the dispatch and equipment packaging for damage, and compare the contents of the package with the delivery documents. In the event of damage in transit, please contact Bender immediately.
The devices must only be stored in areas where they are protected from dust, damp, and spray and dripping water, and in which the specified storage tem­peratures can be ensured.
10
IRDH375_D00124_05_M_XXEN/01.2020
Important information

1.6 Warranty and liability

Warranty and liability claims in the event of injury to persons or damage to property are excluded if they can be attributed to one or more of the follow­ing causes:
Improper use of the device.
Incorrect mounting, commissioning, operation and maintenance of the
device.
Failure to observe the instructions in this operating manual regarding transport, commissioning, operation and maintenance of the device.
Unauthorised changes to the device made by parties other than the manufacturer.
Non-observance of technical data.
Repairs carried out incorrectly and the use of replacement parts or
accessories not approved by the manufacturer.
Catastrophes caused by external influences and force majeure.
Mounting and installation with device combinations not recom-
mended by the manufacturer.
This operating manual, especially the safety instructions, must be observed by all personnel working on the device. Furthermore, the rules and regulations that apply for accident prevention at the place of use must be observed.
IRDH375_D00124_05_M_XXEN/01.2020
11
Important information

1.7 Disposal

Abide by the national regulations and laws governing the disposal of this de­vice. Ask your supplier if you are not sure how to dispose of the old equip­ment.
The directive on waste electrical and electronic equipment (WEEE directive) and the directive on the restriction of certain hazardous substances in electri­cal and electronic equipment (RoHS directive) apply in the European Commu­nity. In Germany, these policies are implemented through the "Electrical and Electronic Equipment Act" (ElektroG). According to this, the following applies:
Electrical and electronic equipment are not part of household waste.
Batteries and accumulators are not part of household waste and must
be disposed of in accordance with the regulations.
Old electrical and electronic equipment from users other than private households which was introduced to the market after 13 August 2005 must be taken back by the manufacturer and disposed of properly.
For more information on the disposal of Bender devices, refer to our homepage at www.bender-de.com -> Service & support.
12
IRDH375_D00124_05_M_XXEN/01.2020

2. Safety instructions

2.1 General safety instructions

Part of the device documentation in addition to this manual is the enclosed "Safety instructions for Bender products".

2.2 Work activities on electrical installations

Only qualified personnel are permitted to carry out the work necessary to install, commission and run a device or system.
Risk of electrocution due to electric shock!
Touching live parts of the system carries the risk of:
An electric shock
DANGER
If the device is used outside the Federal Republic of Germany, the applicable local standards and regulations must be complied with. The European stand­ard EN 50110 can be used as a guide.
Damage to the electrical installation
Destruction of the device
Before installing and connecting the device, make sure
that the installation has been de-energised. Observe the rules for working on electrical installations.
IRDH375_D00124_05_M_XXEN/01.2020
13

2.3 Device-specific safety information

Children and unauthorised persons must not have access to or contact with the ISOMETER®.
WARNING
Make sure that the operating voltage is correct!
Prior to insulation and voltage tests, the ISOMETER® must be disconnected from the IT system for the duration of the test.
CAUTION
CAUTION
In order to check the correct connection of the device, a functional test has to be carried out before starting the system.
Make sure that the basic settings meet the requirements of the IT system.
In the event of an alarm message of the ISOMETER®, the insulation fault should be eliminated as quickly as possible.
Safety instructions
14
If the ISOMETER® is installed inside a control cabinet, the insulation fault message must be audible and/or visible to attract attention.
IRDH375_D00124_05_M_XXEN/01.2020
Safety instructions
When using ISOMETER®s in IT systems, make sure that only one active ISOMETER® is connected in each interconnected system. If IT systems are interconnected via coupling switches, make sure that ISOMETER®s not currently used are disconnected from the IT system and deactivated. IT systems coupled via diodes or capacitances may also influence the insulation monitoring process so that a central control of the different ISOMETER®s is required.
Prevent measurement errors!
When a monitored IT system contains galvanically coupled DC circuits, an insulation fault can only be detected correctly if the rectifier valves (e.g. rectifier diode, thyristors, IGBTs, frequency inverters, …) carry a minimum current of > 10 mA.
Unspecified frequency range
When connecting to an IT system with frequency components below the specified frequency range, the response times and response values may differ from the indicated technical data. However, depending on the application and the selected measurement method, continuous insulation monitoring is also possible in this frequency range. There is no influence on the insulation monitoring for IT systems with frequency components above the specified frequency range, e.g. within the range of typical switching frequencies of frequency inverters (2…20 kHz).
IRDH375_D00124_05_M_XXEN/01.2020
15
Safety instructions

2.4 Intended use

The ISOMETER® is intended for:
 monitoring the insulation resistance of IT systems
Use for the intended purpose also includes
 compliance with all information in the operating instructions
and
 compliance with test intervals.
In order to meet the requirements of the applicable standards, customised pa­rameter settings must be made on the equipment in order to adapt it to local
equipment and operating conditions. Please heed the limits of the range of application indicated in the technical data.
Any use other than that described in this manual is regarded as improper.
16
IRDH375_D00124_05_M_XXEN/01.2020
Safety instructions

2.5 Directions for installation

Risk of property damage due to unprofessional installation!
If more than one insulation monitoring device is connected
CAUTION
CAUTION
The terminals and KE shall be connected by a separate wire to the pro­tective conductor (PE). If the terminals L1, L2 of the device are connected to a system under operation, the terminals and KE must not be disconnected from the protective conductor (PE).
to a conductively connected system, the system can be damaged. If several devices are connected, the device does not function and does not signal insulation faults. Make sure that only one insulation monitoring device is connected in each conductively connected system.
Ensure disconnection from the IT system!
When insulation or voltage tests are to be carried out, the device shall be isolated from the system for the test period. Otherwise the device may be damaged.
Check proper connection!
Prior to commissioning of the installation, check that the device has been properly connected and check the device functions. Perform a functional test using an earth fault via a resistance that is suitable for the mains voltage.
IRDH375_D00124_05_M_XXEN/01.2020
17
Safety instructions
Prevent measurement errors!
When an AC system includes galvanically connected DC circuits, the following shall be considered: Insulation faults in DC circuits can only be monitored correctly when the rectifiers carry a continuous load of 5…10 mA.
18
IRDH375_D00124_05_M_XXEN/01.2020

3. Function

3.1 Common characteristics (IRDH375 and IRDH375B)

 ISOMETER® for IT AC systems with galvanically connected rectifiers
and for IT DC systems (IT = unearthed systems)
 The operating range of the nominal voltage U
coupling devices.
 Automatic adaptation to the existing system leakage capacitance  measuring principle (European Patent: EP 0 654 673 B1)  Two separately adjustable ranges of the response value1 kΩ …10 MΩ
(Alarm 1, Alarm 2)
 Two-line LC display  Connection monitoring (monitoring of the measuring leads)  Automatic device self test  Option "W":
This option provides: improved shock and vibration resistance for use in ships, on rolling stock and in seismic environment.
can be extended via
n
IRDH375_D00124_05_M_XXEN/01.2020
19
Function

3.2 Additional characteristics IRDH375B

 Memory with real-time clock to store all alarm messages with date and
time stamp.
 BMS interface (BMS protocol) for data exchange with other Bender
devices (RS-485 electrically isolated).
 Internal disconnection of the ISOMETER from the IT system to be moni-
tored (using a control signal; terminals F1/F2) , e.g. if several ISOMETER®s are interconnected.
 Current output 0(4)…20 mA (galvanically separated) in relation to the
measured insulation value.
 Remote setting of certain parameters via the Internet (option; COM465
additionally required)

3.3 Product description

The ISOMETER® type IRDH375 monitors the insulation resistance of IT systems. It is suitable for universal use in 3NAC, AC/DC and DC systems. AC systems may include extensive DC supplied loads, such as converters or thyristor-con­trolled DC drives. The device automatically adapts itself to the existing system leakage capacitance.
Suitable coupling devices are available to extend the nominal voltage range
.
U
n
The IRDH375B can be used in combination with a control and indicating de­vice, e.g. PRC1470 version 2 or higher, on the BMS (BMS = Bender Measuring Device Interface) bus.
20
IRDH375_D00124_05_M_XXEN/01.2020
Function

3.4 Function

The ISOMETER® IRDH375 is connected between the unearthed system (IT sys­tem) and the protective conductor (PE).
The response values and other function parameters are set via the function keys. The parameters are indicated on the LC display and are stored in a non­volatile memory (EEPROM) after the setting is completed.
A microprocessor-controlled pulsating AC measuring voltage is superim­posed on the IT system to be monitored ( measuring principle*). The measuring cycle consists of positive and negative pulses of the same am­plitude. The period of these pulses depends on the respective system leakage capacitances and the insulation resistances of the IT system to be monitored.
An insulation fault between the IT system and earth closes the measuring cir­cuit. From the measured current value, the microprocessor calculates the in­sulation resistance which is indicated on the LC display or the external kΩ measuring instrument.
The measuring time is determined by the system leakage capacitances, the in­sulation resistance, and the system-related interference disturbances. System leakage capacitances do not influence the measuring accuracy.
If the reading is below the selected response values Alarm 1/Alarm 2, the as­sociated alarm relays respond and the alarm LEDs "Alarm 1/2" light up and the measuring value is indicated on the LC display (in the event of DC insulation faults, the faulty supply line is indicated). If the terminals R1/R2 are bridged (external RESET button [NC contact] or wire bridge), the fault indication will be stored. Pressing the RESET button, resets the insulation fault message, provid­ed that the currently displayed insulation resistance is at least 25% above the actual response value when the reset is carried out. The fault memory behav­iour can also be set in the "ISO SETUP" menu, by selecting the sub menu "Memory: on/off".
The connections for external kΩ display supplied by the current output 0…400 μA or 0(4)…20 mA (IRDH375B) at M+/M- are galvanically isolated.
IRDH375_D00124_05_M_XXEN/01.2020
21
Function
*) measuring principle "adaptive measuring pulse", a measuring principle developed by Bender (European Patent: EP 0 654 673 B1).

3.4.1 Self test

A self test can be carried out manually using the TEST button or automatically. In order to guarantee high functional reliability, the ISOMETER® IRDH375 pro­vides comprehensive self test functions. After switching the supply voltage on, all internal measuring functions, the components of the process control such as data and parameter memory as well as system and earth connections are checked using the self test functions. The progress of the self test is indi­cated on the display by a bar graph. Depending on the conditions in the IT sys­tem to be monitored, the self test is running for 15…20 s, then the message "Test ok!“ appears on the LC display for approximately 2 s. Then the device re­turns to normal measuring mode and the current measuring value is dis­played after the expiry of the measuring time.
When a device or connection fault is found, the message "!Error!" appears on the display, the device fault LED lights up, the relay K3 (31-32-34) switches and the respective fault message (see table) is indicated. If such a device fault oc­curs, a self test is started again every minute. If no more malfunction is detect­ed, the fault message is deleted automatically and the device fault LED extinguishes.
During operation, the self test function can be started by pressing the TEST button (internal or external). The self test can also be started automatically every hour or every 24 h by selecting "ISO ADVANCED: Autotest" menu.
22
IRDH375_D00124_05_M_XXEN/01.2020
Function
The alarm relays Alarm1/2 only switch after starting the self test function by pressing the TEST button, that means if an automatic self test has been select­ed, the alarm relays do not switch.

3.4.2 Relay K3: device fault alarm and EDS common message

Rel ay K3 is in tended t o sig nal d evice an d con nect ion e rrors of th e ISO METER®.
K3 is permanently set to N/C operation, with the contacts 31-34 normally closed, that means when a fault occurs, the relay deenergizes (K3: 31-32 con­nected).
Further details are described in "Chapter 3.4.1 Self test".
The settings for K3 are preset and cannot be adjusted.
Behaviour of the analogue output
Setting Manual test Automatic test
0-20 mA 20 mA
while test procedure
4-20 mA 20 mA
while test procedure
0 mA The current value depends on the insulation value
4 mA The current value depends on the insulation value
IRDH375_D00124_05_M_XXEN/01.2020
23
Error message Meaning Steps to be taken
Function
System connection?
Connection PE? No low-resis-
Device error x Internal device
No low-resis­tance connec­tion of terminals L1, L2 to the IT system
tance connec­tion of the
terminals and KE to earth (PE)
error
If the on/off switching of the supply voltage is not possible for technical reasons, a RESET of the process control can be carried out by pressing the "INFO“, "RESET“ and "MENU“ key.
1. Check the wiring of terminal L1, L2 to the IT system.
2. Press the test button.
3. Switch the supply voltage off and on.
4. Check the fuses.
1. Check wiring of terminal
and KE to earth (PE).
2. Press TEST button.
3. Switch the supply voltage off and on.
1. Press TEST button.
2. Switch the supply voltage off and on.
3. Contact Bender.
24
IRDH375_D00124_05_M_XXEN/01.2020
Function

3.5 Additional functions IRDH375B

Current output for external measuring instrument
The current output of IRDH375B provides 0(4)…20 mA. The current output is galvanically isolated from the device electronics and the RS-485 interface. The ISO SETUP menu, on page 43, allows to switch over between 0…20 mA and 4…20 mA.
Real-time clock
The real-time clock serves as a time base for the memory and self test func­tions. At first, the correct time and date must be set in the menu "ISO ADVANCED". If time and date are not set, a "C" (clock) is flashing in the standard display. In the event of a supply voltage failure, time and date will be stored for at least thirty days.
If the 24 h selftest is activated in the "ISO ADVANCED" menu, a special time of day can be selected for the execution of the self test in the menu "TEST: 12:00". Then a self test will be started automatically once a day exactly at the preset time. If the 1 h auto test has been selected, the self test is automatically carried out every full hour.
Interconnected IT systems
When using ISOMETER®s in IT systems, make sure that only one active ISOMETER® is connected in each interconnected system. If IT systems are in­terconnected via coupling switches, make sure that ISOMETER®s not currently used are disconnected and deactivated via a control system. IT systems cou­pled via diodes or capacitances may also influence the insulation monitoring process. Hence, also in this case a central control of the different ISOMETER®s is required.
IRDH375_D00124_05_M_XXEN/01.2020
25
Function
Function input F1/F2 for connection or disconnection of IT systems being monitored
The ISOMETER® can be disconnected from the IT system and set to STANDBY mode with the function input F1/F2. If the input F1/F2 is bridged, the connec­tions L1/L2 are switched off via internal coupling relays, the measuring func­tion is stopped and the message "STANDBY" appears on the display. Software version 1.4 or higher does not indicate the measured insulation resistance during the disconnection, but indicates the value > 10 MΩ. Furthermore, the alarm relays and alarm LEDs no longer provide alarm messages. After opening the function input F1/F2, the connection to the IT system will be restored and a completely new measuring cycle for insulation monitoring is started.
With this function, selective disconnection of an IRDH375B in interconnected IT systems can be carried out via auxiliary contacts of the respective coupling switch. One coupling switch each in a line-type or ring-type arrangement can deactivate a subsequent IRDH375B. This arrangement guarantees that only one ISOMETER® is active in each galvanically connected system. In a ring-type arrangement with all coupling switches closed, it can be assumed that all ISOMETER®s are deactivated. In order to prevent this, a BMS Master (IRDH375B BMS address 1) monitors the condition of the function input F1/F2 of all Slave ISOMETER®s.
When all Slave ISOMETER®s are in the STANDBY mode, the insulation monitor­ing function of the Master ISOMETER® and hence the function input F1/F2 of the Master are without function in this mode.
Details are shown in the graphic below.
26
IRDH375_D00124_05_M_XXEN/01.2020
BMS bus (A/B, RS485)
F1/F2
PE PE
PE PE
IT system 1 IT system 2
IT system 3IT system 4
G
G
G
G
Addr. 1
Addr. 2
Addr. 4
Addr. 3
IRDH375B
IRDH375B
IRDH375B
IRDH375B
F1/F2
F1/F2
F1/F2
Function
IRDH375_D00124_05_M_XXEN/01.2020
27
Function
ISOnet Function (COM SETUP)
Select "ISOnet=ON" from the COM SETUP menu to activate this function. This function is a type of scanning function. The BMS Master activated via the ISOnet function controls the ISOnet Slave devices via the BMS bus. Once an ISOMETER® has finished its measurement cycle, the authorization for measur­ing the insulation resistance is passed on from the ISOnet Master to the next Slave. While an ISOMETER® is carrying out a measurement all other ISOMETER®s are in the STANDBY mode. In this way it can be prevented that the ISOMETER®s disturb each other in interconnected systems. In comparison to the solution coupling switches and function input F1/F2, the response time is prolonged, since no continuous measurement takes place. The advantage is that no auxiliary contacts of a coupling switch are required. Furthermore, this solution is ideally suited for capacitive IT systems or IT systems connected via diodes.
An ISOnet Slave checks whether there is a Master available in the network. If there is no Master available, the fault message "ISOnet Master?" appears on the display after approximately 1 h. Additionally, the LED for device errors lights and the Relay K3 switches. When the ISOnet function is activated, the function input F1/F2 will be disconnected.
28
IRDH375_D00124_05_M_XXEN/01.2020
BMS bus (A/B, RS485)
PE PE
PE PE
IT system 1 IT system 2
IT system 3IT system 4
G
G
G
G
Addr. 1
Addr. 2
Addr. 4
Addr. 3
IRDH375B
IRDH375B
IRDH375B
IRDH375B
Function
IRDH375_D00124_05_M_XXEN/01.2020
29

4. Commissioning flow chart (threepart)

Is the system to be monitored an unearthed system (IT system)?
The IRDH375 is not suitable for this application( contact BENDER).
Un is too high for direct connection. A coupling device providing the respective voltage range must be connected.
Device connection
Optional device
connection
no
no
yes
The terminals L1 and L2 must be connected to Un of the system to be monitored according to the wiring diagram.
An external kW measuring instrument at M+/M- with a display range of 10 kW...1000 kW, Scale centre point: 120 kW
Output current IRDH375: 0...400 mA Output current IRDH375B: 0/4...20 mA
Connect the supply voltage Us to the terminals A1/+ and A2/-. Consider the details indicated on the nameplate.
External TEST button (NO contact) to the terminals T1 and T2
External RESET button (NC contact) to the terminals R1 and R2
The output contacts of the alarm relays System fault (31-32-34) Alarm 1 (11-12-14) Alarm 2 (21-22-24)
When using the RS485 interface, take care that a 120 W resistor is connected at the beginning and the end of the network. Terminate IRDH...: S1 = ON
13
12
11
7
8
10
Deenergize the installation
before connecting the device!
Recommended wire cross section of
connecting cable
single wire 0.2...4 mm
2
flexible 0.2... 2.5 mm
2
Is the maximum nominal voltage Un AC 793 V or Un
DC 650 V
yes
The two PE connections
and KE must be connected separately to the equipotential bonding.
6
2 3 4 5
1
The encircled figures in the flow chart correspond to the figures in the legend to the wiring diagram.
Commissioning of the ISOMETER® (1)
30
IRDH375_D00124_05_M_XXEN/01.2020
Commissioning flow chart (threepart)
Commissioning of the ISOMETER® (2)
The ISOMETER® carries out a
self test. The display indicates
the insulation value after finishing
the measurement.
IRDH375_D00124_05_M_XXEN/01.2020
31
Loading...
+ 70 hidden pages