Beha-Amprobe AM-500-EUR User manual

AM-500-EUR
Digital Multimeter
User Manual
EN FR DE
IT ES DU
PL SV NO
FI PT DA
Digital Multimeter
User Manual
10/2017, 4275548 A ©2017 Amprobe Test Tools. All rights reserved. Printed in China
English
Limited Warranty and Limitation of Liability
Your Beha-Amprobe product will be free from defects in material and workmanship for three years from the date of purchase unless local laws require otherwise. This warranty does not cover fuses, disposable batteries or damage from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling. Resellers are not authorized to extend any other warranty on the behalf of Beha-Amprobe. To obtain service during the warranty period, return the product with proof of purchase to an authorized Beha-Amprobe Service Center or to an Beha­Amprobe dealer or distributor. See Repair Section for details. THIS WARRANTY IS YOUR ONLY REMEDY. ALL OTHER WARRANTIES
- WHETHER EXPRESS, IMPLIED OR STATUTORY - INCLUDING IMPLIED WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE OR MERCHANTABILITY, ARE HEREBY DISCLAIMED. MANUFACTURER SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you.
Repair
All Beha-Amprobe tools returned for warranty or non-warranty repair or for calibration should be accompanied by the following: your name, company’s name, address, telephone number, and proof of purchase. Additionally, please include a brief description of the problem or the service requested and include the test leads with the meter. Non-warranty repair or replacement charges should be remitted in the form of a check, a money order, credit card with expiration date, or a purchase order made payable to Beha-Amprobe.
In-Warranty Repairs and Replacement – All Countries
Please read the warranty statement and check your battery before requesting repair. During the warranty period, any defective test tool can be returned to your Beha-Amprobe distributor for an exchange for the same or like product. Please check the “Where to Buy” section on beha-amprobe.com for a list of distributors near you.
Non-Warranty Repairs and Replacement – Europe
European non-warranty units can be replaced by your Beha-Amprobe distributor for a nominal charge. Please check the “Where to Buy” section on beha-amprobe.com for a list of distributors near you.
Beha-Amprobe
Division and reg. trademark of Fluke Corp. (USA)
* United Kingdom
Germany In den Engematten 14 52 Hurricane Way 79286 Glottertal Norwich, Norfolk Germany NR6 6JB United Kingdom
Phone: +49 (0) 7684 8009 - 0 Phone: +44 (0) 1603 25 6662 beha-amprobe.de beha-amprobe.com
The Netherlands - Headquarters Science Park Eindhoven 5110 5692 EC Son The Netherlands Phone: +31 (0) 40 267 51 00 beha-amprobe.com
*(Correspondence only – no repair or replacement available from this address. European customers please contact your distributor.) **single contact address in EEA Fluke Europe BV
**
AM-500-EUR Digital Multimeter
4
7
8
1
32
5
6
1
LCD Display
2
SELECT Button
5
Input Terminal for voltage, diode, resistance and
continuity measurement
6
Input Terminal for battery test and AC/DC mA or μA
measurement
7
Input Terminal for AC/DC A measurement to10A
8
COM (return) terminal for all measurements
3
DATA HOLD Button
4
Rotary Switch
Screen Display
1
2 3
4
5
1
The Meter selects the range with best resolution
2
Direct Current
3
Alternate Current
4
Negative reading
5
Data hold
6
Diode test
7
Continuity test
8
Measurement units for resistance test
9
Measurement units for voltage test
10
Measurement units for current test
11
Low battery indicator
12
Battery Test
6 7
8
11 12
10
9
CONTENTS
SYMBOLS ............................................................................ 2
SAFETY INFORMATION ......................................................2
UNPACKING AND INSPECTION ..........................................4
FEATURES ............................................................................4
MAKING MEASUREMENT .................................................. 5
Rotary Switch Positions .................................................. 5
SELECT Button ................................................................ 6
DATA HOLD Button ........................................................ 6
Auto Power OFF ............................................................. 6
Measuring AC and DC Voltage ...................................... 6
Measuring Resistance .................................................... 7
Measuring Continuity .................................................... 7
Measuring Diode ............................................................ 8
Battery Test .....................................................................8
Measuring AC and DC Current ...................................... 9
SPECIFICATION ....................................................................10
MAINTENANCE AND REPAIR .............................................14
BATTERY AND FUSE REPLACEMENT .................................. 15
1
SYMBOLS
Caution ! Risk of electric shock.
Caution! Refer to the explanation in this Manual
Alternating Current (AC)
Direct Current (DC)
The equipment is protected by double insulation
or reinforced insulation Earth (Ground)
Audible tone
Battery Complies with European Directives
Conforms to relevant Australian standards
Canadian Standards Association (NRTL/C)
Do not dispose of this product as unsorted
municipal waste. Contact a qualified recycler.
SAFETY INFORMATION
The Meter complies with: IEC/EN 61010-1 3rd Edition, UL61010-1 2nd Ed. and CAN/ CSA-C22.2 No. 61010-1-04 + CSA Update No.1: 2008 to Category III 600 Volts, Pollution degree 2 IEC/EN 61010-2-030 IEC/EN 61010-2-31 for test leads EMC IEC/EN 61326-1
“This product has been tested to the requirements of CAN/CSA-C22.2 No. 61010-1, second edition, including Amendment 1, or a later version of the same standard incorporating the same level of testing requirements”.
Measurement Category III (CAT III) is for measurements performed in the building installation. Examples are measurements on distribution boards, circuit- breakers, wiring, including cables, bus-bars, junction boxes,
2
switches, socket-outlets in the fixed installation, and equipment for industrial use and some other equipment, for example, stationary motors with permanent connection to the fixed installation.
CENELEC Directives
The instruments conform to CENELEC Low-voltage directive 2006/95/EC and Electromagnetic compatibility directive 2004/108/EC
Warning: Read Before Using
• To avoid possible electrical shock or personal injury, follow these instructions and use the Meter only as specified in this manual.
• Do not use the Meter or test leads if they appear damaged, or if the Meter is not operating properly. If in doubt, have the Meter serviced.
• Always use the proper function and range for measurements.
• Before rotating the function range selection switch, disconnect test probe from circuit under test.
• Verify the Meter’s operation by measuring on a known voltage source.
• Do not apply more than the rated voltage, as marked on the Meter, between the test probe or between any test probe and earth ground.
• Use the Meter with caution for voltages above 30 Vac rms, 42 Vac peak, or 60 Vdc. These voltages pose electrical shock hazards.
• Disconnect circuit power and discharge all high­voltage capacitors before testing resistance.
• Do not use the Meter around explosive gas or vapor.
• When using the test leads, keep your fingers behind the finger guards.
• Remove test leads from the Meter before opening the Meter case or battery door.
• if the meter is used in a manner not specified in the users manual, the protection provided by the equipment may be impaired
3
UNPACKING AND INSPECTION
Your shipping carton should include:
1 AM-500-EUR Multimeter 1 Pair of test leads 2 1.5V alkaline AAA battery (installed) 1 Users manual If any of the items are damaged or missing, return
the complete package to the place of purchase for an exchange.
FEATURES
Easy to use digital multimeter designed for the homeowner or DIY enthusiast. Install, troubleshoot or repair light fixtures, fans, appliances, or automotive electrical problems with easy and safety of a CAT III 600V rated product. Check for voltage in electrical sockets, extension cords, batteries and other electrical circuits. Let the power of a professional Amprobe multimeter keep you safe and help you solve all your electrical challenges.
• Measurements: Voltage up to 600V AC/DC, AC/DC
Current and Resistance
• Audible continuity
• Diode Test
• Data hold
• Auto power off
• Low battery warning
• Safety: CAT III 600V
4
MAKING MEASUREMENT

1. Use the proper function and range for measurements.
2. To avoid possible electrical shock, personal injury or damages to the Meter, disconnect circuit power and discharge all high-voltage capacitors before testing resistance and diode.
3. Connecting test leads:
• Connect the common (COM) test lead to the circuit before connecting the live lead.
• After measurement, remove live lead before removing the common (COM) test lead from the circuit.
4. Symbol “OL” is displayed on LCD when the measurement is out of range.
Rotary Switch Positions
Switch Position Measurement Function
AC or DC voltage measurement (use SELECT button for switching to AC or DC).
Resistance measurement Voltage measurement of diode PN
junction Continuity measurement For measurement of dry batteries of
9V
not exceeding 15Vdc For measurement of dry batteries of
1.5V not exceeding 2Vdc
AC or DC current measurement (use SELECT button for switching to AC or DC).
μA
V
mA A
5
Press to select altemate measurement
SELECT
Button
functions on the rotary switch.
DATA
Display freezes present reading
HOLD
SELECT Button
Press the yellow SELECT button to select alternate measurement functions on the rotary switch.
DATA HOLD Button
Press DATA HOLD button to freeze present reading on display. Press again to resume normal operation.
Auto Power OFF
Auto power off: approx. 15 minutes. When the Meter is in auto power off mode, press SELECT or DATA HOLD button to resume normal operation.
Measuring AC and DC Voltage
Press SELECT button to switch to DC voltage measurement function.
Do not apply on a voltage source higher than
AC/DC 600V.
6
Measuring Resistance
Disconnect circuit power and discharge all
high-voltage capacitors before testing resistance.
Note: On a higher resistance measurement (>1MΩ), the measurement may take a few seconds to get stable reading.
Measuring Continuity
Disconnect circuit power and discharge all high-
voltage capacitors before testing continuity.
7
Measuring Diode
Disconnect circuit power and discharge all high-
voltage capacitors before testing diode.
Battery Test
Applying a voltage source or incorrect battery type
under battery test may cause personal Injury or damage to the Meter. Battery 1.5V range is for dry battery not exceeding 2Vdc. The resistance load is around 30Ω. Battery 9V range is for dry battery not exceeding 15Vdc. The resistance load is around 1KΩ.The resistance load is around 1KΩ.
8
Measuring AC and DC Current
Press SELECT button for switching to DC current measurement function.
To avoid personal injury or damage to the Meter:
1. Do not attempt to make an in-circuit current measurement when the open-circuit potential to earth ground exceeding 600 V.
2. Switch to proper function and range for your measurement.
3. Do not place the test probe in parallel with a circuit when the test leads are connected to the current terminals.
4. Connect the test leads to the correct input A/mA μA current terminal and to the circuit before powering the circuit under test.
5. After measurement, switching OFF the circuit’s power before removing test leads from the circuit.
or
for
for
For current range from 8-10A, do not measure current for more than one (1) minute. Wait for 10 minutes before taking another measurement.
9
SPECIFICATION
Ambient temperature: 23°C ±5°C (73.4°F ±9°F) Relative temperature: 75% Accuracy: ±(% of reading + digits) Maximum voltage between input terminal and earth
ground: AC 600Vrms or DC 600V
Fuse for mA μA input:
0.5A H 700V fast-fuse, 6.3x32mm
Fuse for 10A input:
10A H 600V fast-fuse, 6x25mm
Maximum display: 1999, updates 2 to 3/sec Over-range indication: OL Range: Automatic Altitude: Operating 2000m Operating temperature: 0°C ~ +40°C (32°F ~ 104°F) Relative humidity: 0°C ~ +30°C (32°F ~ 86°F) 75%; +30°C
~ +40°C (86°F ~ 104°F) 50%
Storage temperature: -10°C ~ +50°C (14°F ~ 122°F) Electromagnetic compatibility: In an RF filed of 1V/m =
Specified accuracy ±5%
Battery: 2 x 1.5V AAA alkaline battery or equivalent Low battery indication: Dimensions (L x W x H): 150 mm x 83 mm x 40 mm (5.9 in
x 3.3 in x 1.6 in) Weight: Approx. 290g (0.64lb) with batteries installed
10
1. DC Voltage Measurement
Range Resolution Accuracy
200.0mV 0.1mV ±(0.8%+3dgt)
2.000V 1mV ±(0.8%+1dgt)20.00V 10mV
200.0V 100mV
600V 1V ±(1.0%+3dgt)
Input impedance: around 10M; (Input impedance > 3G for DC 200mV range)
Overload protection: 600VDC or AC rms
2. AC Voltage Measurement
Range Resolution Accuracy
2.000V 1mV ±(1.0%+3dgt)20.00V 10mV
200.0V 100mV
600V 1V ±(1.2%+3dgt)
Input impedance: around 10M Frequency response: 45Hz ~ 400Hz Overload protection: 600VDC or AC rms
3. Resistance Measurement
Range Resolution Accuracy
200.0Ω 0.1Ω
2.000kΩ 1Ω
200.0kΩ 100Ω
2.000MΩ 1kΩ ±(1.2%+2dgt)
20.00MΩ 10kΩ ±(1.5%+5dgt)
±(1.2%+5dgt) at 5Ω ±(1.2%+3dgt) at >5Ω
±(1.0%+2dgt)20.00kΩ 10Ω
11
200Ω range: Measured value = (Measured display value) – (Short-circuiting value of probe)
Open circuit voltage: around 0.5V Overload protection: 600V
4. : Continuity : Diode measurement
Range Resolution Accuracy
Open circuit voltage is around 0.5V. Resistance >150Ω, buzzer will not
0.1Ω
Overload protection: 600V
1mV
sound. Resistance 10Ω, buzzer will sound.
11 < Resistance < 150.........Not
Specified. Open-circuit voltage is around 1.5V.
Normal voltage is around 0.5V to
0.8V for silicon PN junction.
5. Battery Test
Range Resolution Accuracy
1.5V
9V
Overload protection:
10mV ±(10%+3dgt)
F1 fuse, 0.5A H 700V fast-fuse, 6.3x32mm For 1.5V range: Load resistance is around 30Ω. For 9V range: Load resistance is around 1kΩ.
12
6. DC Current Measurement
Range Resolution Accuracy
200.0μA 0.1μA
μA
2000μA 1μA
20.00mA 10μA
mA
200.0mA 0.1mA
2.000A 1mA
A
10.00A 10mA
Overload protection:
±(1.0%+2dgt)
±(1.2%+3dgt)
mA /μA input:
F1 fuse, 0.5A H 700V fast-fuse, 6.3x32mm
10 A input:
F2 fuse, 10A H 600V fast-fuse, 6x25mm
7. AC Current Measurement
Range Resolution Accuracy
200.0μA 0.1μA
μA
2000μA 1μA
20.00mA 10μA
mA
200.0mA 0.1mA
2.000A 1mA
A
10.00A 10mA
Frequency response: 45Hz ~ 400Hz Overload protection:
±(1.2%+2dgt)
±(1.5%+3dgt)
mA /μA input:
F1 fuse, 0.5A H 700V fast-fuse, 6.3x32mm
10 A input:
F2 fuse, 10A H 600V fast-fuse, 6x25mm
13
MAINTENANCE AND REPAIR
If the Meter fails to operate, check battery, test leads, etc., and replace as necessary. Double check the followings:
1. Replace the fuse or battery if the meter does not work.
2. Review the operating instructions for possible mistakes in operating procedure.
Except for the replacement of the battery, repair of the meter should be performed only by an Authorized Service Center or by other qualified instrument service personnel.
The front panel and case can be cleaned with a mild solution of detergent and water. Apply sparingly with a soft cloth and allow to dry completely before using. Do not use aromatic hydrocarbons, Gasoline or chlorinated solvents for cleaning.
14
BATTERY AND FUSE REPLACEMENT

WARNING
To avoid shock, injury, or damage to the Meter: Disconnect test leads before opening case. Use ONLY fuses with the amperage, interrupt, voltage, and speed ratings specified.
Replacing BATTERY follow below steps:
1. Disconnect the test lead probe from measuring circuit.
2. Turn the Meter to OFF position.
3. Remove the screws from the battery cover and open the battery cover
4. Remove the batteries and replace with two
1.5V Alkaline Batteries (AAA) or equivalent. Pay attention to the polarity signs.
5. Put the battery cover back and re-fasten the screw.
Battery: 1.5V Alkaline Batteries (AAA) or equivalent
15
Replacing FUSE follow below steps:
1. Disconnect the test lead probe from measuring
circuit.
2. Turn the Meter to OFF position and remove the
holster.
3. Remove the screws from the enclosure and open
the enclosure.
4. Remove the broken fuse and replace with new
specified fuse.
5. Put the enclosure back and re-fasten the screw.
Fuse: mA /μA input:
F1 fuse, 0.5A H 700V fast-fuse,
6.3x32mm
10 A input:
F2 fuse, 10A H 600V fast-fuse, 6x25mm
F1
F2
16
Multimètre numérique
Mode d’emploi
10/2017, 4275548 A ©2017 Amprobe Test Tools. Tous droits réservés. Imprimé en Chine.
Français
Limites de garantie et de responsabilité
Amprobe garantit l’absence de vices de matériaux et de fabrication de ce produit pendant une période d’un an prenant effet à la date d’achat, sauf disposition contraire prévue par la loi. Cette garantie ne s’applique pas aux fusibles, aux piles jetables ni à tout produit mal utilisé, modifié, contaminé, négligé ou endommagé par accident ou soumis à des conditions anormales d’utilisation et de manipulation. Les revendeurs n’ont pas l’autorisation de prolonger toute autre garantie au nom d’Amprobe. Pour bénéficier de la garantie, renvoyez le produit accompagné d’un justificatif d’achat auprès d’un centre de services agréé par Amprobe ou d’un distributeur ou d’un revendeur Amprobe. Voir la section Réparation pour tous les détails. LA PRÉSENTE GARANTIE EST LE SEUL ET EXCLUSIF RECOURS DE L’UTILISATEUR TOUTES AUTRES GARANTIES, EXPLICITES, IMPLICITES OU STATUTAIRES, NOTAMMENT LE CAS ÉCHÉANT LES GARANTIES DE QUALITÉ MARCHANDE OU D’ADAPTATION À UN OBJECTIF PARTICULIER SONT EXCLUES PAR LES PRÉSENTES. LE FABRICANT NE SERA EN AUCUN CAS TENU RESPONSABLE DE DOMMAGES PARTICULIERS, INDIRECTS, ACCIDENTELS OU CONSÉCUTIFS, NI D’AUCUNS DÉGATS OU PERTES DE DONNÉES, SUR UNE BASE CONTRACTUELLE, EXTRA­CONTRACTUELLE OU AUTRE. Étant donné que certaines juridictions n’admettent pas les limitations d’une condition de garantie implicite ou l’exclusion ou la limitation de dégâts accidentels ou consécutifs, il se peut que les limitations et les exclusions de cette garantie ne s’appliquent pas à votre cas.
Réparation
Tous les outils de test renvoyés pour être réparés au titre de la garantie ou pour étalonnage doivent être accompagnés des éléments suivants : nom, raison sociale, adresse, numéro de téléphone et justificatif d’achat. Ajoutez également une brève description du problème ou du service demandé et incluez les cordons de test avec l’appareil. Les frais de remplacement ou de réparation hors garantie doivent être acquittés par chèque, mandat, carte de crédit avec date d’expiration, ou par bon de commande payable à l’ordre de Amprobe®.
Remplacements et réparations sous garantie – Tous pays
Veuillez lire la déclaration de garantie et vérifiez la pile avant de demander une réparation. Pendant la période de garantie, tout outil de test défectueux peut être renvoyé auprès de votre distributeur Amprobe® pour être échangé contre un produit
identique ou similaire. Consultez la section « Where to Buy » sur le site www.amprobe.com pour obtenir la liste des distributeurs dans votre région. Les appareils sous garantie devant être remplacés ou réparés au Canada et aux États-Unis peuvent également être envoyés dans un centre de services Amprobe® (voir les adresses ci-dessous).
Remplacements et réparations hors garantie – Canada et États-Unis
Les appareils à réparer hors garantie au Canada et aux États­Unis doivent être envoyés dans un centre de services Amprobe®. Appelez Amprobe® ou renseignez-vous auprès de votre lieu d’achat pour connaître les tarifs en vigueur de remplacement ou de réparation.
Aux États-Unis et au Canada Amprobe Amprobe Everett, WA 98203 Mississauga, ON L4Z 1X9Canada Tél. : 877-AMPROBE (267-7623) Tél. : 905-890-7600
Remplacements et réparations hors garantie – Europe
Les appareils européens non couverts par la garantie peuvent être remplacés par votre distributeur Amprobe® pour une somme nominale. Consultez la section « Where to Buy » sur le site www. metermantesttools.com pour obtenir la liste des distributeurs dans votre région.
Adresse postale européenne* Amprobe® Test Tools Europe Beha-Amprobe GmbH In den Engematten 14 79286 Glottertal, Allemagne Tél. : +49 (0) 7684 8009 - 0 www.beha-amprobe.com *(Réservée à la correspondance – Aucun remplacement ou
réparation n’est possible à cette adresse. Nos clients européens doivent contacter leur distributeur.)
Multimètre numérique AM-500-EUR
4
7
8
1
32
5
6
1
Afficheur LCD
2
Bouton de sélection
5
Borne d’entrée pour les mesures de tension, de
résistance, le contrôle de diode et de continuité
6
Borne d’entrée pour le test des piles et les mesures mA
ou μA ac/dc
7
Borne d’entrée pour les mesures A ac/dc jusqu’à 10 A
8
Borne de retour COM pour toutes les mesures
3
Bouton de maintien d’affichage
4
Sélecteur rotatif
Affichage
1 6 7
2 3
4
5
1
Le multimètre numérique sélectionne la gamme
avec la meilleure résolution
2
Mesure continue
3
Mesure alternative
4
Lecture négative
5
Maintien des données affichées
6
Contrôle de diode
7
Contrôle de continuité
8
Unités de mesure du test de résistance
9
Unités de mesure du test de tension
10
Unités de mesure du test de courant
11
Témoin de piles faibles
12
Test des piles
8
11 12
10
9
TABLE DES MATIÈRES
SYMBOLES .......................................................................... 2
CONSIGNES DE SÉCURITÉ .................................................. 2
DÉBALLAGE ET INSPECTION .............................................. 4
FONCTIONNALITÉS ............................................................. 4
OPÉRATIONS DE MESURE ..................................................5
Positions du sélecteur rotatif ........................................ 5
Bouton SELECT ............................................................... 6
Bouton DATA HOLD ....................................................... 6
Mise en veille automatique ........................................... 6
Mesures de tension alternative et continue ................. 6
Mesures de résistance .................................................... 7
Mesures de continuité ................................................... 7
Contrôle de diode .......................................................... 8
Test des piles ...................................................................8
Mesures de courant alternatif et continu .................... 9
CARACTÉRISTIQUES ........................................................... 10
ENTRETIEN ET RÉPARATION ............................................... 14
REMPLACEMENT DES FUSIBLES ET DES PILES .................. 15
1
Loading...
+ 238 hidden pages