BD FACSCanto Reference Manual

5 (1)
For In Vitro Diagnostic Use
bdbiosciences.com
Part No. 642241 Rev. A June 2007
BD FACSCanto
Flow Cytometer
Reference Manual
BD Biosciences
2350 Qume Drive San Jose, CA 95131-1807 USA Tel (877) 232-8995 Fax (800) 325-9637 facservice@bd.com
Asia Pacific
Tel (65) 6-861-0633 Fax (65) 6-860-1590
Europe Tel (32) 2 400 98 95 Fax (32) 2 401 70 94 help.biosciences@europe.bd.com Tel 81-24-593-5405
Brazil Tel (55) 11-5185-9995 Fax (55) 11-5185-9895
Japan
Nippon Becton Dickinson Company, Ltd. Toll Free 0120-8555-90
Fax 81-24-593-5761
Canada
Toll Free (888) 259-0187 Tel (905) 542-8028 Fax (888) 229-9918 canada@bd.com
Mexico
Toll Free 01-800-236-2543 Tel (52) 55 5999 8296 Fax (52) 55 5999 8288
© 2007, Becton, Dickinson and Company. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, stored in retrieval systems, or translated into any language or computer language, in any form or by any means: electronic, mechanical, magnetic, optical, chemical, manual, or otherwise, without prior written permission from BD Biosciences.
The information in this guide is subject to change without notice. BD Biosciences reserves the right to change its products and services at any time to incorporate the latest technological developments. Although this guide has been prepared with every precaution to ensure accuracy, BD Biosciences assumes no liability for any errors or omissions, nor for any damages resulting from the application or use of this information. BD Biosciences welcomes customer input on corrections and suggestions for improvement.
BD FACSDiva software © Becton, Dickinson and Company. This software is the property of Becton, Dickinson and Company. Each sale of a stored unit of this software grants the purchaser a nontransferable, nonexclusive, personal license. This software may not be duplicated, reproduced, or copied in any form or by any means whatsoever, except as otherwise permitted by law.
BD FACSCanto clinical software © Becton, Dickinson and Company. This software is the property of Becton, Dickinson and Company. Each sale of a stored unit of this software grants the purchaser a nontransferable, nonexclusive, personal license. This software may not be duplicated, reproduced, or copied in any form or by any means whatsoever, except as otherwise permitted by law.
BD, BD logo, and all other trademarks are property of Becton, Dickinson and Company.
Cy™ is a trademark of Amersham Biosciences Corp. Cy™ dyes are subject to proprietary rights of Amersham Biosciences Corp. and Carnegie Mellon University and are made and sold under license from Amersham Biosciences Corp. only for research and in vitro diagnostic use. Any other use requires a commercial sublicense from Amersham Biosciences Corp., 800 Centennial Avenue, Piscataway, NJ 08855-1327, USA.
JDS Uniphase is a trademark of JDS Uniphase, Inc.
Microsoft and Windows are registered trademarks of Microsoft Corporation.
Sapphire is a trademark and Coherent is a registered trademark of COHERENT, INC.
Teflon is a registered trademark of E.I. du Pont de Nemours and Company.
All other company and product names might be trademarks of the respective companies with which they are associated.
Patents
PerCP: US 4,876,190 APC-Cy7: US 5,714,386
FCC Information
WARNING: Changes or modifications to this unit not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
NOTICE: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, can cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his or her own expense.
Shielded cables must be used with this unit to ensure compliance with the Class A FCC limits.
This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.
Cet appareil numérique de la classe A respecte toutes les exigences du Réglement sur le matériel brouilleur du Canada.
Notice
BD Biosciences delivers software and workstations that are intended for running the cytometers supplied by BD Biosciences. It is the responsibility of the buyer/user to ensure that all added electronic files including software and transport media are virus free. If the workstation is used for Internet access or purposes other than those specified by BD Biosciences, it is the buyer/user’s responsibility to install and maintain up-to-date virus protection software. BD Biosciences does not make any warranty with respect to the workstation remaining virus free after installation. BD Biosciences is not liable for any claims related to or resulting from buyer/user's failure to install and maintain virus protection.
History
Revision Date Change Made
336928 Rev. A 1/04 Initial release
337969 Rev. A 4/04 Updated for CE IVD release
338619 Rev. A 9/04 Updated for BD FACSDiva software v4.1
343370 Rev. A 9/05 Updated for BD FACSCanto clinical software v2.0 and the Opticon LG2 Imager barcode reader
640801 Rev. A 5/06 Updated for BD FACSDiva software v5.0 and BD FACSCanto clinical software v2.1
642241 Rev. A 6/07 Updated for BD FACSDiva software v6.0

Contents

About This Manual ix
Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . x
Technical Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xi

Chapter 1: Introduction 13

Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
System Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Flow Cytometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Fluidics Cart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Computer Workstation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
BD FACS Loader (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Barcode Reader (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
System Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

Chapter 2: BD FACS Loader Option 33

Loader Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Using the Loader . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Changing Operational Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Maintaining the Loader . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Cleaning External Surfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
v

Chapter 3: Barcode Reader Option 45

Installing and Using the Barcode Reader . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Cleaning the Barcode Reader . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Barcode Symbologies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
1D Barcode Symbologies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
2D Barcode Symbologies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50

Chapter 4: Maintenance 51

Scheduled Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Emptying the Waste Container . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Purging the Fluidics Filters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Decontaminating the Fluidics System (Long Clean) . . . . . . . . . . . . . . . . 59
Replacing the Air Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Replacing Fluidics Filters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Unscheduled Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Changing a Cubitainer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Removing an Air Lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Cleaning External Surfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Removing Bubbles from the Flow Cell . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Cleaning the Flow Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Purging the Bubble Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Decontaminating the Fluidics System for Storage . . . . . . . . . . . . . . . . . . 80
Replacing the Bal Seal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Resetting the Cytometer Circuit Breaker . . . . . . . . . . . . . . . . . . . . . . . . . 84
Reconnecting the Ethernet and RS232 Cables . . . . . . . . . . . . . . . . . . . . . 85
Reconnecting the Fluidics Cart Tubing . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Replacing the Fluidics Level Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
Replacing the Fluidics Cart Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
vi BD FACSCanto Flow Cytometer Reference Manual

Chapter 5: Troubleshooting 95

Cytometer Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Fluidics Cart Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Loader Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102

Appendix A: Technical Overview 105

Flow Cytometry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Fluidics System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Optics System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Light Scatter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
Fluorescence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
Optical Filters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
Detectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Detector Arrays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Spillover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118
Electronics System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
Pulses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
Pulse Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122

Appendix B: Supplies and Replacement Parts 127

Cytometer Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128
Installation Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128
Other Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129
Bar Code Reader Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129
Consumables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
Cytometer Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
Reagents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
Labware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131
Contents vii

Appendix C: Technical Specifications 133

Cytometer Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
Optics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
Fluidics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
Signal Processing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
Fluidics Cart Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138
Capacity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138
Loader Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
Index 141
viii BD FACSCanto Flow Cytometer Reference Manual

About This Manual

This manual contains reference information about the BD FACSCanto™ flow cytometer and BD FACS™ Loader. Operating instructions are contained in the BD FACSCanto Instructions for Use.
Most cytometer functions are controlled by BD FACSCanto™ clinical software and BD FACSDiva™ software. BD FACSCanto clinical software contains modules for dedicated clinical applications with automatic gating algorithms, while BD FACSDiva software is non–application specific. You can use either software to perform cytometer quality control. Each software package has its own reference manual, provided on the documentation CD.
BD Biosciences recommends that first-time users of this cytometer take advantage of operator training offered with the sale of every new cytometer.
The BD FACSCanto Flow Cytometer Reference Manual assumes you have a working knowledge of basic Microsoft® Windows® operation. If you are not familiar with the Windows operating system, refer to the documentation provided with your computer.
ix

Conventions

The following tables list conventions used in this manual. Table 1 lists symbols that are used in this manual or on safety labels to alert you to a potential hazard. Text and keyboard conventions are shown in Table 2.
Tabl e 1 Hazard symbols
Symbol Meaning
CAUTION: hazard or unsafe practice that could result in material damage, data loss, minor or severe injury, or death
Electrical danger
Laser radiation
Biological risk
a. Although these symbols appear in color on the cytometer, they are in black and white throughout this
reference manual; their meaning remains unchanged.
Tabl e 2 Text and keyboard conventions
Convention Use
; Tip
NOTICE
Italics Book titles and new or unfamiliar terms
a
Features or hints that can save time or prevent difficulties
Provides information that supplements the topic material
> Menu selection
For example, “select File > Print” means select Print from the File menu.
Ctrl-X Press the indicated keys simultaneously
For example, Ctrl-P means hold down the Control key while pressing the letter p.
x BD FACSCanto Flow Cytometer Reference Manual

Technical Assistance

For technical questions or assistance in solving a problem:
Read the section of the manual specific to the operation you are performing.
See Chapter 5, Troubleshooting.
If additional assistance is required, contact your local BD Biosciences technical support representative or supplier.
When contacting BD Biosciences, have the following information available:
product name, part number, and serial number
software version number
•error messages
details of system performance
For cytometer support within the US, call (877) 232-8995.
For support within Canada, call (888) 259-0187.
Customers outside the US and Canada, contact your local BD representative or distributor.
About This Manual xi
xii BD FACSCanto Flow Cytometer Reference Manual
1

Introduction

The BD FACSCanto system uses a fixed-optics design and advanced digital electronics to support multicolor analysis of up to six fluorescent markers and two scatter parameters per assay.
The cytometer does not require special facilities—it plugs into a standard wall outlet, uses air-cooled lasers, and provides its own air pressure and vacuum for the fluidics and waste. You can prepare samples on the BD FACS™ Sample Prep Assistant II and import the sample prep worklist.
For further automation, use BD FACSCanto clinical software and the BD FACS Loader for sample acquisition. Alternatively, use BD FACSDiva software for more flexibility in acquisition and analysis.
See the following for more information:
Intended Use on page 14
System Components on page 14
System Requirements on page 31
13

Intended Use

The BD FACSCanto system is intended for use as an In Vitro Diagnostic device for identification and enumeration of lymphocyte subsets in human cells in suspension for flow cytometry.

System Components

The BD FACSCanto system consists of a benchtop flow cytometer, a self­contained fluidics cart, and the BD FACSCanto workstation. System options include an automated sample loader and a barcode reader.
For a description of system components, see:
Flow Cytometer on page 15
Fluidics Cart on page 25
14 BD FACSCanto Flow Cytometer Reference Manual
Computer Workstation on page 28
BD FACS Loader (Optional) on page 28
Barcode Reader (Optional) on page 30

Flow Cytometer

Cytometer power is controlled by the power button. All other cytometer and fluidics cart functions are controlled by BD FACSCanto clinical software and BD FACSDiva software (provided with the cytometer).
flow cell
access door
side door
data ports
optics
access door
sample injection tube
power button
fluidics cart
connections
The BD FACSCanto flow cytometer consists of a fluidics subsystem, an optics subsystem, and an electronics subsystem. For a more in-depth discussion of
Chapter 1: Introduction 15
fluidics, optics, electronics, and flow cytometry, see the Technical Overview on page 105.
Fluidics
The fluidics system consists of the sample injection tube (SIT), the aspirator arm, the flow cell, a pressurized interior reservoir, and a network of tubing that provides sheath and cleaning fluids to and removes waste from the flow cell.
flow cell
tubing
adapter lever
sample injection tube (SIT)
tube guide (Loader option only)
aspirator arm
16 BD FACSCanto Flow Cytometer Reference Manual
aspirator arm bar
The following table briefly describes these components.
flow cell where laser beam intercepts particles
tubing brings sheath and cleaning fluids to and aspirates waste from flow cell
SIT hollow metal tube that brings sample to flow cell
adapter lever lever used to change SIT from manual to automatic loading
aspirator arm movable waste aspiration port
aspirator arm bar
tube guide helps guide tube onto the SIT (Loader operation only)
metal bar used to push aspirator arm away from SIT during manual loading
When you install tubes onto the SIT, a pump within the fluidics cart pressurizes the interior reservoir (plenum), which then provides sheath fluid to the flow cell. At the same time, the sample tube is pressurized and sample is pushed up the SIT and into the flow cell.
When you remove tubes from the SIT, the cytometer cleans the SIT by flushing sheath solution down the inside and outside of the tube. The flushed sheath solution is aspirated by the aspirator arm.
To activate SIT cleaning, push the aspirator arm all the way to the left when you manually unload a tube. When you are using the Loader, SIT cleaning occurs automatically.
Do not leave a tube of distilled water on the SIT between sample tubes, or during or after daily shutdown.
Chapter 1: Introduction 17
Flow Cell
Once the sample moves into the flow cell, particles move in single file past the laser beams. The scattered and emitted light from these particles provides information about their size, shape, granularity, and fluorescence properties.
The flow cell is beneath the flow cell access door. For more information, see Fluidics System on page 106.
Figure 1-1 Flow cell
where lasers intercept
sample stream
obscuration bar
FSC diode
flow cell
Optics
The optics system in the BD FACSCanto cytometer is composed of both excitation optics and collection optics. Excitation optics bring light to the flow cell. Collection optics gather the light signals emitted or scattered by the particles.
Excitation Optics
The excitation optics consist of lasers, fiber optic cables, beam-shaping prisms, and an achromatic focusing lens, as shown in Figure 1-2 on page 20.
18 BD FACSCanto Flow Cytometer Reference Manual
The BD FACSCanto cytometer uses low-powered air-cooled and solid state lasers that do not have special power and cooling requirements.
Laser
Coherent® Sapphire™ Solid State
JDS Uniphase™ HeNe Air Cooled
Wavelength
(nm)
488 (blue) FITC, PE, PerCP,
633 (red) APC, APC-Cy7
Commonly Used Fluorochromes
PerCP-Cy™5.5, PE-Cy™7
Chapter 1: Introduction 19
Fiber optic cables direct the laser light onto beam-shaping prisms, which in turn transmit the laser light to a focusing lens. The lens directs the laser light onto the sample stream within the flow cell (Figure 1-2). See also Figure A-3 on page 110.
Figure 1-2 Optical pathway
1
3
2
5
6
interlock flow cell
1
prisms red He-Ne laser
2 5
focusing lens blue solid-state laser
3 6
4
When the flow cell access door opens, an interlock shutters the laser light and blocks its pathway for safety reasons.
4
20 BD FACSCanto Flow Cytometer Reference Manual
Collection Optics
From the flow cell, laser light is routed to the collection optics, which efficiently gather the signals emitted and scattered from each particle. The BD FACSCanto collection optics include two detector arrays, which consist of photomultiplier tubes (PMTs) arranged in one octagon and one trigon (Figure 1-3).
The octagon contains five PMTs and detects light from the 488-nm (blue) laser. One PMT in the octagon collects side scatter (SSC) signals.
The trigon contains two PMTs and detects light from the 633-nm (red) laser.
Figure 1-3 Octagon and trigon detector arrays
red-laser
A
585/42
488/10
E
signal
C
B
660/20
trigon
longpass
A
mirror
bandpass filter
735LP
780/60
C
octagon
PMT
H
F
780/60
735LP
D
502LP
530/30
655LP
670LP
B
556LP
G
blue-laser
signal
Chapter 1: Introduction 21
When light arrives at an array, a longpass mirror (filter) transmits the highest wavelengths to the first PMT in the series and reflects lower wavelengths to the next PMT. Similarly, the next PMT’s longpass mirror transmits the next highest wavelengths and reflect lower wavelengths, and so on around the array. A bandpass filter (or additional longpass mirror) in front of each PMT further screens unwanted light.
A
D
B
Fiber
E
In addition to the PMT detectors, a photodiode collects the stronger forward scatter signals. The obscuration bar prevents excess laser light from entering this diode (Figure 1-1 on page 18).
For additional descriptions of detector arrays, mirrors, and filters, see Optics System on page 109.
22 BD FACSCanto Flow Cytometer Reference Manual
At installation, the octagon and trigon arrays have the filter and mirror combinations shown in Table 1-1.
Table 1-1 Octagon and trigon optical filters
Detector Array
(Laser)
Octagon (488-nm blue laser)
Trigon (633-nm red laser)
PMT
Position
A 735 780/60 PE-Cy7
B 655 670 PerCP-Cy5.5 or
C 556 585/42 PE
D 502 530/30 FITC
Eblank
Fblank
Gblank
H— blank
A 735 780/60 APC-Cy7
B blank 660/20 APC
LP Mirror
optical holder
optical holder
optical holder
BP Filter or
LP Mirror
488/10 SSC
blank
optical holder
blank
optical holder
optical holder
Intended Dye
PerCP
Blank optical holders do not contain optical filters. They are used in the octagon and trigon to prevent unwanted light from interfering with fluorescence signal.
Chapter 1: Introduction 23
Electronics
The electronics system converts optical signals to electronic signals and digitizes them for computer analysis. The photodiode and PMTs generate signals proportional to the amount of light they detect. The cytometer’s onboard electronics amplifies and then converts the signals from continuous voltage values (analog) into discrete values (digital). Upon amplification and digital conversion, fluorescent light signals from consistently prepared and stained particles characteristically fall into certain channels, thus allowing analysis.
On the BD FACSCanto, electronic system components consist of power controls and connectors along with processing boards in the card cage. This section describes only the user-accessible power panel. For more information, see Electronics System on page 120.
Power Panel
Power to the cytometer, lasers, and fluidics cart is supplied by a power cord from the cytometer plugged directly into a standard electrical outlet. BD recommends using an uninterrupted power supply (UPS) unit to maintain cytometer power during a power outage. The system power button turns on the cytometer and fluidics cart, and powers the lasers (Figure 1-4).
Figure 1-4 Flow cytometer power panel
system AC power
cord plugs in here
system circuit breaker
The system circuit breaker is located next to the AC power cord. The breaker will need to be reset if there is a power surge in the laboratory.
24 BD FACSCanto Flow Cytometer Reference Manual
system power button

Fluidics Cart

The fluidics cart provides filtered sheath and cleaning fluids to the cytometer, and collects system waste products (Figure 1-5). The cart supplies the required air pressure and vacuum, which eliminates the need for an external source (although the cart can be hooked up to an in-house air source).
Figure 1-5 Fluidics cart
cubitainer
10-L waste container
pressure
gauge
door access
knob
filters
Chapter 1: Introduction 25
Containers and Ports
The fluidics cart holds a 10-L waste container, a 20-L BD FACSFlow™ cubitainer, a 5-L BD FACS shutdown solution cubitainer, and a 5-L BD FACSClean solution cubitainer.
Use the waste container provided with the system; do not substitute other containers.
Figure 1-6 Fluidics cart containers and ports
BD FACS™ shutdown solution port and cubitainer
waste port
BD™ FACSClean port and cubitainer
BD FACSFlow™ port
waste tank
BD FACSFlow cubitainer
Each solution has its own non-interchangeable fluid port and level-sensor connection. Fluid level alarms occur within BD FACSCanto clinical software and BD FACSDiva software.
26 BD FACSCanto Flow Cytometer Reference Manual
Controls
The fluidics cart connects to the flow cytometer unit by way of cables and tubing. When you turn on the power to the cytometer, the fluidics cart powers on also. Under ordinary circumstances, you do not need to adjust any of the switches on the cart’s power panel. Leave the auxiliary air supply switch off unless the cart has been attached to an in-house air supply by BD Biosciences service personnel (see Table 1-2 for details). Leave the cart circuit breaker on at all times.
Figure 1-7 Fluidics cart panel
auxiliary air supply switch, off
cart circuit breaker, on
Table 1-2 Auxiliary air supply switch positions
Position Air Source What This Means
Off Cart air The cart is providing its own air
pressure and vacuum.
On House air The cart is hooked up to the
building’s air pressure and vacuum.
Chapter 1: Introduction 27
Powering Off
To turn off the fluidics cart (and the cytometer, as well), press the system power button. Normally during cart shutdown, you hear a hiss, and a small amount of condensed water will discharge.

Computer Workstation

The workstation consists of a computer compatible with Microsoft Windows XP Professional operating system, a flat-screen monitor, and an optional printer.

BD FACS Loader (Optional)

The BD FACS Loader automatically introduces prepared samples to the cytometer. The Loader is controlled by BD FACSCanto clinical software or BD FACSDiva software. It can be added to your system at any time. The Loader consists of:
•a drawer
a cover
three optical sensors
an electronics module
•a tube lifter
a 40-tube carousel
For more information, see BD FACS Loader Option on page 33.
28 BD FACSCanto Flow Cytometer Reference Manual
Figure 1-8 BD FACS Loader
Chapter 1: Introduction 29

Barcode Reader (Optional)

The Opticon LG2 Imager is a hand-held barcode reader that plugs into the USB port on the BD FACSCanto computer workstation. A green light on the top of the barcode reader indicates that it is connected to the USB port and ready to use.
The barcode reader reads most barcode standards, including Codabar, Code 128, Code 39 with checksum, and PDF417. It reads information from the BD FACS™ 7-color setup beads label into BD FACSCanto clinical software, and also reads coded patient information into a worklist.
Figure 1-9 Typical barcode reader (example)
ready light
trigger
view from top
For information on installing and using the barcode reader with the BD FACSCanto, see Installing and Using the Barcode Reader on page 46. For detailed information including how to program the barcode reader for other barcode standards, refer to the information supplied by the manufacturer.
30 BD FACSCanto Flow Cytometer Reference Manual
Loading...
+ 114 hidden pages