This instrument carries out the following functions:
1) dissolved oxygen measuring when connected to the sensor
2) manual/automatic d.oxygen regulation, if suitable devices are connected to the output relays
3) temperature measuring, if an RTD Pt1000 is connected
4) manual or automatic compensation of temperature, pressure, salinity and relative humidity effects
6) D.O. and temperature acquisition, when connected to a recorder or a data logger
7) sending data via a serial interface, if the 091.701 option "RS232" is installed
8) external device activation for the sensor cleaning
Sensor
The B&C Electronics OD 8382 optical DO sensor or special galvanic sensor are suitable for this
instrument.
The sensor installation should allow a continuous contact with the sample, in a position with sufficient
stirring and exchange of sample.
In some applications the use of sensor cleaning devices is suggested.
Instruction manual - Rev. A - 07/09
- 3 -
B&C Electronics OD 7685.110
1.2 OPTICAL D.O. MEASURING PRINCIPLE
A light beam of a specific wavelength is sent to a special fluorescent layer in contact with the sample.
The absorbed light energy is partially released as a light pulse with an higher wavelength.
This phenomena is called fluorescence.
If oxygen molecules are in contact with the sensing layer, the fluorescing is reduced (quenching).
By measuring the amount of quenching it is possible to determine the oxygen concentration.
The advantages of this measuring method are the absence of electrolyte and membrane, the possibility
to measure the oxygen concentration in water or in air, a not sensitivity to the flow, and a good
sensitivity in a low oxygen concentration.
1.3 SENSORS AND ACCESSORIES
To be ordered separately:
OD 8382 Auto clean optical DO probe, 10 m cable
OD 8112 Pressure air cleaner+controller
BC 931.2 Enclosure for one unit. IP65
BC 931.3 Enclosure for two units. IP65
Instruction manual - Rev. A - 07/09
- 4 -
B&C Electronics OD 7685.110
2 SPECIFICATIONS
2.1 FUNCTIONAL SPECIFICATIONS
Input
The instrument accepts input from the OD 8382 optical DO probe or the AT amplified galvanic cell.
A second input is provided for 2 or 3 wires Pt1000 RTD temperature sensors.
Temperature compensation
The unit is supplied with manual or automatic temperature compensation and temperature
information may be displayed on the LCD.
The instrument detects of the absence or malfunctioning of the temperature sensor and automatically
switches to manual compensation.
Secondary parameters
The unit is supplied with pressure, salinity and relative humidity manual compensation.
Measuring ranges
The input range may be selected in PPM, mg/l, % air saturation and mmHg partial pressure.
The temperature range is selectable in °C or in °F.
Dual Analog output
Either a 0/20 mA or 4/20 mA isolated output may be selected, for use as an interface with computers
or data loggers.
A special routine allows selection of the analog output range.
The output current may be set anywhere from 0 to full scale.
Control relays
The D.O. monitor is equipped with two SPDT control relays.
Each control relay may be programmed for set-point, high/low, hysteresis or delay time actuation.
When the unit is configured as main display DO +SET, the full display indicates the current settings
and current status of each relay.
Cleaning function
The instrument features two relays (C-D) for autocleaning cycle.
The cleaning function may be activated automatically or manually.
The user may select:
- cleaning time
- holding time
- cycle repetition time
Instruction manual - Rev. A - 07/09
- 5 -
B&C Electronics OD 7685.110
During the CLEAN and HOLD time:
- flashing messages will appear
- analog outputs are maintained in hold
- relays A and B are deactivated
Calibration mode
The instrument may be automatically calibrated in air.
Manual calibration may also be performed.
Software filter
The unit is provided with a dual programmable software filter, to be inserted when the readout is not
stable.
Configuration
The electronics for the D.O. monitor system are designed to be as flexible as possible.
A number of programming functions are provided in the configuration menu and are protected by an
access number, which must be entered to allow changes in this setting.
The main display can be configured in order to show the D.O. and temperature values or the D.O.
value and set point status.
Keyboard lock
The keys on the front panel of the monitor can be used for both changing the display and for
calibrations and set-point adjustments.
When the monitor is shipped, all functions are accessible.
However, the adjustment and calibration functions may be locked in order to prevent unauthorized
adjustments to the instrument.
Options
091.404 24 Vac power supply
091.4143 9/36 Vdc power supply
091.701 RS232 isolated output
The output sends the data (D.O. value, °C) to the serial port
Instruction manual - Rev. A - 07/09
- 6 -
B&C Electronics OD 7685.110
2.2 TECHNICAL SPECIFICATIONS
The Default values are correspondent to the factory calibration values.
Parameters marked by "*" can be modified in the Configuration procedures.
1) DO SENSOR
* Main display: DO+SET/DO+TEMP
Input: from optical DO sensor
or from amplified ATI galvanic cell
Input signal at 20 °C (68 °F): 83.2/1040 mV
Sensor sensitivity: 20%/250 %
Temperature compensation: internal table
Nr. of bit: 8 bit
Nr. of stop bit: 1 bit
Parity: None
Isolated from measuring circuits
Example of data: ±20.00PPM ±50.0°C Z:-0.500mV S:100.0%
Data sent every: 0.4 seconds
9) 24 VAC POWER SUPPLY (option 091.404)
Default
Voltage: 24 Vac +/-10 % 50/60 Hz Power: 5 VA max
10) VDC POWER SUPPLY (option 091.4143)
Default
Voltage: 9 to 36 Vdc Power: 4 W max. Absorption max.: 320 mA at 12 Vdc 160 mA at 24 Vdc Isolation I/O: 1500 Vdc
Instruction manual - Rev. A - 07/09
- 9 -
B&C Electronics OD 7685.110
11) PARAMETERS ON CONFIG BLOCK (See for *)
Free calibration (access code not required):
Keyboard Locked/Unlocked.
LCD contrast (0/7).
Main display: DO+SET/DO+TEMP
Temperature unit: °C/°F
Under access code number (0):
Measuring unit (mmHg,%air,ppm,mg/l).
Range of measure (40.0/20.00)
Autorange on/off
Digital filter response time (large): 0.5/50.0"
Digital filter response time (small): 0.5/50.0"
Input connected to out 1 (O2/°C/°F)
Range of analog output 1 (0/20 4/20mA)
Point 1 for output 1
Point 2 for output 1
Input connected to out 2 (O2/°C/°F)
Range of analog output 2 (0/20 4/20mA)
Point 1 for output 2
Point 2 for output 2
Function of relay A (HI/LO).
Function of relay B (HI/LO).
Cleaning function (Auto/Manual/Disabled)
Number of cycle for cleaning function: 1/10
Charging time: 0.5/60.0"
Discharging time: 0.0/10.0"
Holding time before restarting of normal operation: 0/60.0'
Access Number: 0/999
12) GENERAL SPECIFICATIONS
Acquisition time: 0.4 seconds approx.
Alphanumeric display: 1 line x 16 characters
Operating temperature: 0/50 °C
Humidity: 95% without condensate
Power supply: 110/220 Vac +/- 10% 50/60 Hz
Isolation: 4000 V between primary and secondary (IEC 348)
Power: 5 VA max.
Terminal block: extractable
Weight: 850 g
Default
Unlocked
4
DO+SET
°C
ppm
20.00
off
2.0"
10.0"
O2
4/20
0.00
20.00
°C (°F)
4/20
0.0
50.0
LO
LO
Disabled
4
20.0"
3.0"
3.0'
0
Default
Instruction manual - Rev. A - 07/09
- 10 -
B&C Electronics OD 7685.110
2.3 PHYSICAL DESCRIPTION
The controller enclosure is designed for surface or panel mounting.
It consists of an anodized aluminum case built according to the standard DIN 43700, with an
aluminum panel coated with scratch-proof and non-corrosive polycarbonate membrane.
A transparent waterproof front door SZ7601 can be added to the housing, in
order to protect the unit from excessive moisture or corrosive fumes.
Signal and power cable connections are made by using two extractable terminal blocks placed in the
back of the instrument.
This makes wiring, installation and general maintenance of the probes and other devices easier.
The package is supplied complete with fixing clamps for panel-mounting.
Instruction manual - Rev. A - 07/09
- 11 -
B&C Electronics OD 7685.110
3 OPERATING THE SYSTEM
1. LCD display
electronics
2. Mode-Display key
3. Calibration key
4. Increase key
5. Decrease key
6. Enter key
Fig. 1
Instruction manual - Rev. A - 07/09
- 12 -
Loading...
+ 28 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.