B&C Electronics OD 265.2 INSTRUCTION MANUAL

CERT. N. 9115 BCEL
INSTRUCTION MANUAL
UNI EN ISO 9001:2000
OD 265.2
CONTROLLER
POLAROGRAPHIC CELL
Rev. B - 03/07
B&C Electronics Srl - Via per Villanova 3 - 20040 Carnate (Mi) - Italy - P.IVA 00729030965 Tel +39 039 63 1721 - Fax +39 039 607 6099 – bc@bc-electronics.it - www.bc-electronics.it
B&C Electronics OD 265.2
Index
1 GENERAL INFORMATION....................................................................................................... 3
2 PRINCIPLES ................................................................................................................................ 3
3 FUNCTIONAL DESCRIPTION .................................................................................................. 4
4 PHYSICAL DESCRIPTION ........................................................................................................ 5
5 SPECIFICATIONS....................................................................................................................... 6
6 ELECTRICAL INSTALLATION ................................................................................................ 7
6.1 Connecting the power.......................................................................................................... 7
6.2 Warnings.............................................................................................................................. 7
6.3 Connecting the cell .............................................................................................................. 7
6.4 Connecting alarms, pumps, valves ...................................................................................... 7
6.5 Connecting a recorder.......................................................................................................... 8
6.6 Connecting the RTD............................................................................................................ 8
6.7 Checking.............................................................................................................................. 8
7 OPERATING THE SYSTEM ...................................................................................................... 9
7.1 Pre-operation check ............................................................................................................. 9
8 NORMAL OPERATION............................................................................................................ 10
9 CALIBRATING THE CONTROLLER ..................................................................................... 11
9.1 Electric zero....................................................................................................................... 11
9.2 Zero cell calibration........................................................................................................... 11
9.3 Slope calibration ................................................................................................................ 11
9.4 Polarization voltage of sensor............................................................................................ 11
10 TEMPERATURE COMPENSATION ....................................................................................... 12
10.1 Manual ............................................................................................................................... 12
10.2 Automatic .......................................................................................................................... 12
11 PREVENTIVE MAINTENANCE.............................................................................................. 13
11.1 Controller........................................................................................................................... 13
11.2 Sensor ................................................................................................................................ 13
12 TROUBLESHOOTING GUIDE ................................................................................................ 14
Instruction Manual - Rev. B - 03/07 - 2 -
B&C Electronics OD 265.2
1 GENERAL INFORMATION
This manual applies to the OD 265.2 digital bench controller. It explains the purpose of the equipment, describes the components of the series and the procedures for installing the equipment, operating it and calibrating it.
Some suggestions are also given for its maintenance.
2 PRINCIPLES
In a polarographic cell, if a constant Voltage is applied to the electrodes, the reduction of the Oxygen ions on the cathode generates a current proportional to the oxygen's partial Pressure.
The intensity of the current produced at a constant Temperature and at a given Oxygen Pressure depends on the quantity and composition of the electrolyte, the cathode's area and the sample's state. Response time depends on the electrolyte's and diaphragm’s characteristics. The value of the applied Voltage to the anode is chosen in a way that only Oxygen can react.
In absence of Oxygen, the current produced is not exactly zero. In fact, impurities and other factors generate a "dark current" which can be offset by the amplifier. The electrode consumes the Oxygen diffused through the permeable membrane and the electrolyte's film, therefore a concentration gradient is established between cathode and sample; the electrodes with a big cathode and with a high output current, consume a lot of Oxygen and are more sensitive to the sample's flow and therefore request a large flow and an adequate stirring of the sample.
The Temperature dependence of the output current is compensated by the automatic or manual Temperature compensation of the meter.
Instruction Manual - Rev. B - 03/07 - 3 -
B&C Electronics OD 265.2
3 FUNCTIONAL DESCRIPTION
This series of bench controllers is specifically designed for automatic activation of metering pumps and valves in fermentation processes. The controller provides a digital readout of the dissolved Oxygen in aqueous solutions. In order to satisfy the many requests of such measurements, four different ranges have been designed and may be selected on the meter:
- % Oxygen saturation
- % air saturation
- mm Hg
- mg/l (ppm)
RTD automatic Temperature compensation and manual compensation are employed to provide highly accurate measurements over the entire ranges of dissolved Oxygen and Temperature values. The Temperature value and the Temperature coefficient value are both visualized on the display by pushing the corresponding lever on the front panel.
A basic O2 monitoring system consists of two elements: a monitoring unit, and a probe or O2 cell. The system can be expanded by adding accessories for field applications; recorders, secondary regulators, proportional regulators, special probes and Temperature probes to suit particular applications.
The controller contains electronic circuits to control the operations of the entire system. It provides a digital readout of Oxygen on a 3 1/2 digit display. There is included an internal circuitry for controlling two external alarms, valves, pumps etc. Set-points are independent and programmed by a front-panel control to trigger at any level within the range of the meter. Red Leds on the panel indicate if the measured value is below the set point A level, or if it is above the set point B level. Red Leds also indicate the switching of corresponding relays.
The set points have the feature of being inverted and the two regulators have a delay action within 5 seconds by means of a rear trimmer
The set point value is visualized by pushing the button 6 on the front panel and the switching function of the relay is inverted by switches on the rear side of the meter.
When the relay is switched on, an AC Voltage is available on the rear outlet for driving pumps and valves.
The controller provides an output of 0/20 mA or 4/20 mA (selected by a switch on the rear panel) proportional to the meter reading, for driving a recorder, remote readout or regulators having a not grounded input. Zero and slope are calibrated by trimmers on the front panel.
Automatic Temperature compensation is carried out by means of a RTD Pt 100.
The meter's input is by a BNC connector for the electrodes and by a phone Jack for the RTD.
The digital Temperature readout, both in the manual and the automatic Temperature operation, is obtained by pushing a lever 10 on front panel, as measuring and controlling functions of the instrument are not altered. The manual Temperature compensation is adjusted by the trimmer 11. The digital readout of Temperature coefficient ( % / °C ), both in the manual and the automatic operation is obtained by pushing a lever 12, its value being adjusted by the trimmer 13.
The scale is chosen by the selector 14 and the corresponding Led 15 shows which scale is operating.
Instruction Manual - Rev. B - 03/07 - 4 -
B&C Electronics OD 265.2
4 PHYSICAL DESCRIPTION
The controller's case consists of an anodized aluminium front panel and an epoxidic-coated metallic case, to ensure the maximum anticorrosion characteristics.
Connections to the power supply, loads, recorders, RTD's, electrodes or probes are carried out by means of sockets or connectors on the rear side.
Instruction Manual - Rev. B - 03/07 - 5 -
B&C Electronics OD 265.2
5 SPECIFICATIONS
Voltage: 110 or 220 Volts (as requested) +/-10 % 50/60 Hz
Input: from polarographic O2 cell
Output: 0/20 mA dc, non isolated, 300 ohm max. 4/20 mA or others on request
Scales: 0.0/199.9 % O2
0.0/199.9 % air saturation
0.0/199.9 mm Hg
0.0/19.99 ppm
Zero: adjustment +/- 15% Slope: fine adjustment +/- 20% coarse adjust. 5 to 50 nAmp air saturation at 20 °C
Cell current in air: approx. 0.030 microAmp. Cell Voltage 675 mV Cell connector: BNC Factory calibration: half scale with 0.030 microAmp input.
Temperature: 0/50 °C Temperature compensation: manual and automatic 0/50° Celsius Temperature coefficient: 0 to 6 %/°C Temperature sensor: RTD Pt100 Temperature readout: -20/+120 °C
Regulators: +/- 0.25 % hysteresis (others as requested) Switching time: < 0.5 Seconds Regulator delay: 0 to 5 Seconds Regulator output: 220 V or 110 V (same as the power supply)
5 Amp. resistive load
Outlet fuse: 2 Amp. T
Circuit fuse: 80 mA T (110 V) internal
32 mA T (220 V) internal
Power: 3 VA + power of the loads Power fuse: 5 A Weight: 2,140 g Dimensions: 215 x 95 x 200 mm